Table S1: Sample preparation, instrumental analysis, validation parameters and application of sent standard mixture for laboratory coded 1. COMPOUND IF SAMPLE PREP INTRUM ANALYSIS VALIDATION PARAM Sample preparation (sample pre-treatment: pH adjustment, filter pore size, material and trade name, extraction method and conditions, additional clean-up, derivatisation, internal standard …) Instrumental analysis (operational parameters: separation and detection condition,…) Validation parameters (sensitivity, accuracy, recovery, reproducibility, repeatability, …) 1º) pH adjustment to 2 with HCl 2º) Addition of isotopically labelled surrogate standards (SS) 3º) Dilution of samples B, C and D 1:1 with HPLCwater 4º) Filtration through 0.45 µm cellulose acetate membranes 5º) Concentration by on-line SPE-LC-MS/MS: sample volume, 5 mL; cartridge, PLRP-s (10×2 mm, Spark Holland, Emmem, The Netherlands); washing, 0.5 mL HPLC-water - Analysis by LC-MS/MS - Equipment: SymbiosisTM Pico extraction system (Spark Holland, Emmem, The Netherlands) coupled on-line with a 4000QTRAP hybrid quadrupole-linear ion trap mass spectrometer equipped with a Turbo Ion Spray source from Applied Biosystems-Sciex (Foster City, California, USA). - LC separation on a reversed-phase column Purospher STAR RP-18e (125 x 2 mm, 5 µm particle size) from Merck (Darmstadt, Germany), maintained at 25 ºC. Ultrapure water (A) and methanol (B), both containing 0.1% of formic acid, were employed as mobile phase (flow-rate 0.2 mL min-1). LC gradient: 0–1 min, 5% B; 2 min, 20% B; 12 min, 80% B; 25–30 min, 100% B; 35–40 min, 5% B. - MS/MS conditions: positive ESI; SRM - Quantification by the isotope dilution method with calibration curves prepared in HPLC-water - Linearity: determination coefficient r2 = 0.9995 - Sensitivity: method limit of quantification (s/n 10) = 2.0 ng L-1 in surface water and in wastewater. - Accuracy: relative recovery (calculated with respect to the SS, at 20, 500 and 5,000 ng L-1 spiking levels, in surface and wastewater, n=5) = 89-107%. - Repeatability: relative standard deviation (RSD, in surface and wastewater, spiking levels 20, 500 and 5,000 ng L-1, n=5) = 1 - Application of sent mixture a. Standard mixture was used for calibration b. Standard mixture was used for check-up c.Standard mixture was not used b quantification transition, 261.1 > 92.0; SRM confirmation transition, 261.1 > 154.0; collision energy, 35 eV and 31 eV, respectively; declustering potential, 81 V. CP 1º) pH adjustment to 2 with HCl 2º) Addition of isotopically labelled surrogate standards (SS) 3º) Dilution of samples B, C and D 1:1 with HPLCwater 4º) Filtration through 0.45 µm cellulose acetate membranes 5º) Concentration by on-line SPE-LC-MS/MS: sample volume, 5 mL; cartridge, PLRP-s (10×2 mm, Spark Holland, Emmem, The Netherlands); washing, 0.5 mL HPLC-water. ETO 1º) pH adjustment to 2 with HCl 2º) Addition of isotopically labelled surrogate standards (SS) 13 %. - Analysis by LC-MS/MS - Equipment: SymbiosisTM Pico extraction system (Spark Holland, Emmem, The Netherlands) coupled on-line with a 4000QTRAP hybrid quadrupole-linear ion trap mass spectrometer equipped with a Turbo Ion Spray source from Applied Biosystems-Sciex (Foster City, California, USA). - LC separation on a reversed-phase column Purospher STAR RP-18e (125 x 2 mm, 5 µm particle size) from Merck (Darmstadt, Germany), maintained at 25 ºC. Ultrapure water (A) and methanol (B), both containing 0.1% of formic acid, were employed as mobile phase (flow-rate 0.2 mL min-1). LC gradient: 0–1 min, 5% B; 2 min, 20% B; 12 min, 80% B; 25–30 min, 100% B; 35–40 min, 5% B. - MS/MS conditions: positive ESI; SRM quantification transition, 261.1 > 140.0; SRM confirmation transition, 261.1 > 106.1; collision energy, 33 eV and 25 eV, respectively; declustering potential, 86 V. - Quantification by the isotope dilution method with calibration curves prepared in HPLC-water - Linearity: determination coefficient r2 = 0.9999 - Sensitivity: method limit of quantification (s/n 10) = 2.0 ng L-1 in surface water and 3.0 ng L-1 in wastewater. - Accuracy: relative recovery (calculated with respect to the SS, at 20, 500 and 5,000 ng L-1 spiking levels, in surface and wastewater, n=5) = 83-115%. - Repeatability: relative standard deviation (RSD, in surface and wastewater, spiking levels 20, 500 and 5,000 ng L-1, n=5) = 1 13 %. b - Analysis by LC-MS/MS - Equipment: SymbiosisTM Pico extraction system (Spark Holland, Emmem, The Netherlands) coupled on-line with a 4000QTRAP hybrid quadrupole-linear ion trap mass spectrometer equipped - Quantification by the isotope dilution method with calibration curves prepared in HPLC-water - Linearity: determination b 3º) Dilution of samples B, C and D 1:1 with HPLCwater 4º) Filtration through 0.45 µm cellulose acetate membranes 5º) Concentration by on-line SPE-LC-MS/MS: sample volume, 5 mL; cartridge, PLRP-s (10×2 mm, Spark Holland, Emmem, The Netherlands); washing, 0.5 mL HPLC-water. GEM 1º) pH adjustment to 2 with HCl 2º) Addition of isotopically labelled surrogate standards (SS) 3º) Dilution of samples B, C and D 1:1 with HPLCwater 4º) Filtration through 0.45 µm cellulose acetate membranes 5º) Concentration by on-line SPE-LC-MS/MS: sample volume, 5 mL; cartridge, PLRP-s (10×2 mm, Spark Holland, Emmem, The Netherlands); washing, 0.5 mL HPLC-water with a Turbo Ion Spray source from Applied Biosystems-Sciex (Foster City, California, USA). - LC separation on a reversed-phase column Purospher STAR RP-18e (125 x 2 mm, 5 µm particle size) from Merck (Darmstadt, Germany), maintained at 25 ºC. Ultrapure water (A) and methanol (B), both containing 0.1% of formic acid, were employed as mobile phase (flow-rate 0.2 mL min-1). LC gradient: 0–1 min, 5% B; 2 min, 20% B; 12 min, 80% B; 25–30 min, 100% B; 35–40 min, 5% B. - MS/MS conditions: positive ESI; SRM quantification transition, 589.0 > 229.0; SRM confirmation transition, 589.0 > 185.0; collision energy, 15 eV and 10 eV, respectively; declustering potential, 71 V. coefficient r2 = 0.9997 - Sensitivity: method limit of quantification (s/n 10) = 43 ng L-1 in surface water and 65 ng L-1 in wastewater. - Accuracy: relative recovery (calculated with respect to the SS, at 500 and 5,000 ng L-1 spiking levels, in surface and wastewater, n=5) = 76106%. - Repeatability: relative standard deviation (RSD, in surface and wastewater, spiking levels 500 and 5,000 ng L-1, n=5) = 2-12 %. - Analysis by LC-MS/MS - Equipment: SymbiosisTM Pico extraction system (Spark Holland, Emmem, The Netherlands) coupled on-line with a 4000QTRAP hybrid quadrupole-linear ion trap mass spectrometer equipped with a Turbo Ion Spray source from Applied Biosystems-Sciex (Foster City, California, USA). - LC separation on a reversed-phase column Purospher STAR RP-18e (125 x 2 mm, 5 µm particle size) from Merck (Darmstadt, Germany), maintained at 25 ºC. Ultrapure water (A) and methanol (B), both containing 0.1% of formic acid, were employed as mobile phase (flow-rate 0.2 mL min-1). LC gradient: 0–1 min, 5% B; 2 min, 20% B; 12 min, - Quantification by the isotope dilution method with calibration curves prepared in HPLC-water - Linearity: determination coefficient r2 = 0.9993 - Sensitivity: method limit of quantification (s/n 10) = 9.3 ng L-1 in surface water and in wastewater. - Accuracy: relative recovery (calculated with respect to the SS, at 20, 500 and 5000 ng L-1 spiking levels, in surface and wastewater, n=5) = 93-115%. - Repeatability: relative standard deviation (RSD, in b MTX 1º) pH adjustment to 2 with HCl 2º) Addition of isotopically labelled internal standards (IS) 3º) Dilution of samples B, C and D 1:1 with HPLCwater 4º) Filtration through 0.45 µm cellulose acetate membranes 5º) Concentration by on-line SPE-LC-MS/MS: sample volume, 5 mL; cartridge, PLRP-s (10×2 mm, Spark Holland, Emmem, The Netherlands); washing, 0.5 mL HPLC-water. 80% B; 25–30 min, 100% B; 35–40 min, 5% B. - MS/MS conditions: positive ESI; SRM quantification transition, 264.2 > 112.0; SRM confirmation transition, 264.2 > 95.0; collision energy, 25 eV and 63 eV, respectively; declustering potential, 71 V. surface and wastewater, spiking levels 20, 500 and 5,000 ng L-1, n=5) = 1 15 %. - Analysis by LC-MS/MS - Equipment: SymbiosisTM Pico extraction system (Spark Holland, Emmem, The Netherlands) coupled on-line with a 4000QTRAP hybrid quadrupole-linear ion trap mass spectrometer equipped with a Turbo Ion Spray source from Applied Biosystems-Sciex (Foster City, California, USA). - LC separation on a reversed-phase column Purospher STAR RP-18e (125 x 2 mm, 5 µm particle size) from Merck (Darmstadt, Germany), maintained at 25 ºC. Ultrapure water (A) and methanol (B), both containing 0.1% of formic acid, were employed as mobile phase (flow-rate 0.2 mL min-1). LC gradient: 0–1 min, 5% B; 2 min, 20% B; 12 min, 80% B; 25–30 min, 100% B; 35–40 min, 5% B. - MS/MS conditions: positive ESI; SRM quantification transition, 455.2 > 308.2; SRM confirmation transition, 455.2 > 175.1; collision energy, 33 eV and 59 eV, respectively; declustering potential, 91 V. - Quantification by the isotope dilution method with calibration curves prepared in HPLC-water - Linearity: determination coefficient r2 = 0.9995 - Sensitivity: method limit of quantification (s/n 10) = 2.0 ng L-1 in surface water and in wastewater. - Accuracy: relative recovery (calculated with respect to the SS, at 20, 500 and 5000 ng L-1 spiking levels, in surface and wastewater, n=5) = 84-104%. - Repeatability: relative standard deviation (RSD, in surface and wastewater, spiking levels 20, 500 and 5,000 ng L-1, n=5) = 115 %. b