Supporting Information An advanced gas/vapor permeation system for barrier materials; design and applications to poly(ethylene terephthalate) Jong Suk Leea, Preeti Chandraa,b, Steven K. Burgessa, Robert Kriegelc, William J. Korosa,* a School of Chemical and Biomolecular Engineering Georgia Institute of Technology, Atlanta, GA-30332, USA b c Praxair Inc, Tonawanda, NY-14150, USA The Coca-Cola Company, Atlanta, G-30313, USA Determination of the methanol activity for gas/vapor permeation The total number of moles for O2/CO2/CH3OH in the upstream can be estimated by converting the mass flow rate in the upstream bypass to the number of moles. Let the molar flow rate in the upstream bypass line be BTOT. Then, the total number of moles per minute in the feed stream is estimated as: 1 mol ccSTP nTOT BTOT min 22, 414 ccSTP (S.1) On the other hand, the number of moles for CH3OH vapor is estimated by using the volumetric flow rate of CH3OH from the syringe pump based on the assumption that all CH3OH molecules are evaporated prior to reaching the permeation cell. Let the volumetric flow rate of CH3OH be VCH3OH , then: cc 1 mol l g nCH3OH VCH3OH CH3OH 103 min cc l M CH3OH g (S.2) where, CH3OH is the density of CH3OH (i.e. 0.792 (g/cc)) at 35ºC and M CH3OH is the molecular weight of CH3OH (i.e. 32 (g/mol)), so the respective partial pressure of CH3OH is obtained as: 1 pCH3OH nCH3OH nTOT pTOT (S.3) where, pTOT is the total pressure of the feed stream. As long as the partial pressure of the CH3OH * vapor component is less than the corresponding vapor pressure at 35ºC (i.e. pCH 35o C = 202.3 3OH mmHg), condensation can be avoided. By adjusting the volumetric flow rate of CH3OH and the mass flow rate of retentate flow, the desirable CH3OH activity level can be easily achieved. Gas Chromatography for a multicomponent permeation Gas Chromatography (GC) has been used to analyze the composition of the permeate in the case of mixed gas/vapor feeds. The feed composition of custom prepared gas/vapor mixtures has also been analyzed similarly. The instrument is a 6890N from Agilent Technologies (Palo Alto, CA). The valve diagram of the GC is shown below: Figure S. 1. Valve configuration for gas sampling A 2 cc sample loop has been used. The ‘sample in’ port is connected to the permeate and the feed side, whereas the ‘sample out’ port is connected to a 100 Torr MKS BaratronTM transducer and vacuum pump in series. The following are the operating conditions at the time of a GC run. 2 Table S. 1. GC operation conditions Parameter Oven Temperature Total Run Time Back Inlet Split Ratio Column Column Flow Carrier Gas Detector Reference Gas Detector Make-up Gas Flow Heated Valve Box Sample with methanol 120 ºC 4.00 min 150 ºC 10:1 HP Plot Q, 30m long, 0.32 mm diameter, 0.20 coating width 3.4 mL/min at 20 psi Helium 20.0 mil/min 7.0 mL/min 130 ºC The GC was calibrated using pure gases and vapor. Pure CO2 and O2 was expanded into the sample loop from the downstream of the permeation box and allowed to equilibrate for 3 minutes. Different pressures in the range of 0-10 Torr for the gases were injected. Methanol vapor was expanded into the GC from a vapor source volume that was kept outside the permeation box. The vapor was prepared by taking liquid methanol in a vial. Dissolved and headspace air was removed by at least five freeze-pump-thaw cycles. The vapor in the headspace of the liquid vial was then allowed to expand into a 1000 cc colume that was connected to the vial. This volume had been evacuated previously. Vapor from this volume was expanded into the GC sample loop. Methanol vapor in the range of 0-0.5 Torr was used. These pressures were chosen because this was the expected range of partial pressures in the permeate. All the transfer lines were heated to least 70 ºC to ensure that methanol did not adsorb on the transfer lines which would results in smaller peaks. Upon expansion into the sample loop, the sample (either a gas or vapor) was allowed an equilibration time of 3 minutes if the pressure was higher than 5 Torr and 5 minutes of the pressure was less than 5 Torr. Care had been taken to ensure that the leak rate of atmospheric air into the transfer lines was minimal and that no leak peak for air was observed during blank runs. The calibration factors obtained are shown below: 3 Table S. 2. Calibration factors for CO2, O2, and CH3OH Parameter Carbon dioxide Oxygen Methanol Sample with methanol (Area/Torr) 67.748 46.136 46.465 During a permeation run, the mixed gas/vapor feed was analyzed by expanding the feed mixture into the sample loop. Nearly 20-30 Torr of the mixture was injected. This ensured that methanol partial pressures lay in the calibration range. The mole fraction of methanol in the feed was calculated using equation S.1: YMeOH = AMeOH MeOHpinjected (S.4) pinjected is the pressure read off by the transducer in the GC transfer line. AMeOH is the measured area, and βMeOH is the calibration factor. When sufficient permeate was accumulated, which was in the range of 3-6 Torr, it was expanded into the sample loop (the sample loop is made of 1/16” tubing). Care was also taken to ensure that methanol partial pressures did not exceed 0.5 Torr in the downstream in most cases. This is because control tests of methanol adsorption were carried out only up till 0.5 Torr. The permeate was allowed to expand into the sample loop for 5 mins after which the run was started. Areas of the species were obtained and their mole fraction was calculated using the following expression: YA AA A AA A AB B AC C (S.5) Ai is the area of the species and is the calibration factor. The permeability of the species was then calculated using the following equation: 4 dp dpA YA TOT dt dt (S.6) dpA dt 101325 V 106 22414 L 14696 PA 1010 Barrers 60 760 8.314 T A p 76 (S.7) PA is the permeability; dpA/dt is the rate of pressure rise for a given species in Torr/min. For oxygen, the actual dp/dt is evaluated by subtracting out the measured leak rate. This was because it is air that leaks in, and contributes only to the oxygen peak, not the other gases. V is the downstream volume in cc. R = 8.314 J/mol/K, L is the film thickness in cm, ∆p, in psia, is the pressure differential across the film which is the upstream pressure in this case. A is the film area in cm2. 5