I. PURPOSE : To provide a procedure for the validation of analytical methods under use at II. SCOPE : The procedure is to guide the analytical method validation studies of all the analytical methods being developed and used for the analysis of different products produced at Hetero Labs Limited. The procedure also provides the steps involved in the qualification of the compendial methods. III . RESPONSIBILITY : A. Quality Control shall : i. Prepare, Review and approve the method validation protocol with an assigned protocol number. ii. Carry out analysis & provide supporting data with a conclusive report on method validation carried out. B. Quality assurance shall : i. Review and approve the method validation protocol. ii. Should authorized any documents being used during the validation study ii. Review and approve the method validation report IV. DEFINITIONS : Method : A set of all the written procedures and instructions involved in the collection, processing, storage and analysis of chemical matrix for an analyte. System Suitability : Specific tests to ascertain the suitability and effectiveness of the operating system when employing chromatographic methods such as pressurized liquid chromatography and gas chromatography. Validation: The obtaining and documenting of evidence to demonstrate that a method can be relied upon to produce the intended result under stated conditions within defined limits. It is the process of establishing that the performance characteristics of a method (expressed in terms of analytical parameters) to meet the requirements for the intended application of the method. The normal analytical parameters used in method validation are defined as follows . Protocol: A pre approved document ,which should define the scope, objective, responsibilities, method details , experimental design and acceptance criteria with consideration to the intended application of the study is being performed. Report: An approved document ,which should explain the details found during the study and also should be concluded the status of the study against to the predetermined acceptance criteria Specificity: The ability of a method to measure accurately and specifically the analyte (the constituent being tested or analysed) in the presence of components that may be expected to be present in the sample. It is a measure of the degree of interference in the analysis of complex sample mixtures such as blank, bulk drug substances containing degradation products, related substances etc. Precision: The degree of agreement among individual test results when a method is applied to multiple samplings of a homogeneous sample. It is a measure of either the degree of reproducibility (agreement under different conditions) or of repeatability (agreement under same conditions) of the method. Limit of detection (LOD) : The lowest concentration of analyte in a sample that can be detected, but not necessarily quantitated under the stated experimental conditions. It is a parameter of limit tests. Limit of Quantitation (LOQ) :: The lowest concentration of analyte in a sample that can be determined with acceptable precision and accuracy under the stated experimental conditions. It is a parameter of quantitative determinations of low levels of compounds in sample mixtures, such as impurities in bulk drugs . Linearity: The ability of a method to produce results that are directly or indirectly proportional to the concentration of the analyte in samples within a given range. Range : The interval between the upper and lower levels of analyte (including these levels) that have been shown to be determined with precision, accuracy, and linearity using the method as written.. Accuracy: The closeness of test results obtained by a method to the true value. It is a measure of the exactness of the method. Ruggedness ( Intermediate precision ) : The Intermediate precision of an analytical method is the degree of reproducibility of test results obtained by the analysis of the same samples under a variety of normal test conditions ( with the combination of different analyst, different instrument, different lots of reagents, different day at normal laboratory conditions etc.) Intermediate precision is normally expressed as the lack of influence on test results of operational and environmental variables of the analytical method. Robustness : The robustness of an analytical procedure is a measure of its capacity to retain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage. V. PROCEDURE : 01. Before an analytical method is validated, the method must be completely developed and a detailed procedure shall be documented. 02. A method must be validated for intended use, employing an approved protocol. 03. Based on the analyte, the technique, a specific procedure shall be identified to assure the optimum operation of the system employed (as system suitability). 04. Validation studies shall be unique for each method, and shall be consistent with the methods purpose, concentration range to be tested, the sample and the technique used for analyte determination. 05. Validation experimental plans (protocols) shall be supported by appropriate statistical methodologies to assist in the interpretation of the results obtained. 06. Any modification of an analytical method shall be subject to revalidation. 07. Documentation of the successful completion of laboratory studies is a basic requirement for determining whether a method is suitable for its intended applications. 08. The procedure used to evaluate each validation parameter as described in the approved protocol will be focused into the following steps : i. Identification of appropriate validation parameters. ii. Design of experimental plan for parameters evaluation (validation protocol). iii. Execution of the validation. iv. Documentation to support of the conclusion that the method has been validated. v. The documentation shall contain the following : [ a. Approved protocol b. Raw data – Analysis work books/test data sheet, Chromatograms, Charts, Thermogram, Histograms (if any) etc., c. Validation report with conclusion of the study. 09. The experimental design and acceptance criteria of validation parameters can be altered to other than the conditions specified in this SOP VI. EXPERIMENTAL DESIGN: A) 1.1 For Inhouse methods : Determine the category (see Table I ) of the test/s which are to be validated. Prepare a Method Validation Protocol in the specimen format given in the Annexure for each item. Indicate in the protocol the experimental design comprising the tests or methods to be validated, the relevant analytical performance parameters to be checked, on the basis of the category given in the Table, the experimental details for carrying out these checks and the acceptance criteria for each parameter. Guidelines for experimental designs for some commonly done tests and assays are given below Category of tests and selection of validation parameters for In-house methods Category of test Specificity Precision Linearity Accuracy Intermediate & range precision Robustness LOQ LOD Assay by HPLC Yes Yes Yes No Yes Yes** No No Chromatograph ic purity by HPLC OVI &Residual solvent test Chromatograph ic purity (By TLC) Assay Potentiometry Yes Yes Yes Yes Yes Yes** Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes No No Yes No No Yes Yes Yes Yes No Yes Yes No No Particle size by laser diffraction Rinse analysis in cleaning validation No Yes No No Yes Yes No No Yes Yes Yes Yes Yes No Yes Yes Yes Yes Yes Yes Yes No Yes Yes Swab analysis in cleaning validation ** Carry out stability study for both mobile phase and test solution. B) For Pharmacopoeial methods : 1.2 Determine the category (see Table I ) of the test/s which are to be validated. Prepare a Method Validation Protocol in the specimen format given in the Annexure for each item. Indicate in the protocol the experimental design comprising the tests or methods to be validated, the relevant analytical performance parameters to be checked, on the basis of the category given in the Table, the experimental details for carrying out these checks and the acceptance criteria for each parameter. 1.3 Guidelines for experimental designs for some commonly done tests and assays are given below Category of tests and selection of analytical parameters for Pharmacopoeial methods Category of test Assay by HPLC Specificity Precision Yes Yes Linearity Accuracy Intermediate & range precision No No No Robustness LOQ LOD No** No No Chromatographic purity by HPLC Yes Yes No Yes No No** Yes Yes Chromatographic purity (By TLC) Yes Yes No No No No No Yes Assay by Potentiometry Yes Yes Yes No Yes No No No ** Carry out stability study for both mobile phase and test solution. Experimental design for the method validation of Assay by HPLC Analytical parameters System suitability Experimental design . 1. Establish the system suitability through out the validation followed by 2 procedures. Acceptance criteria As applicable (As per method details) 1 e.g. : For chromatographic methods suitable parameters such as theoretical plates, tailing factor must be determined. 2.Procedure1: Establish the system suitability Whenever the chromatographic system changed or disturbed. 2. As applicable (As per method details) 3. Procedure2: When there is no disturbance in chromatographic system , establish the online system suitability between each parameter by injecting the standard solution(single time) and calculate the %RSD for the values obtained from initial values and each online standard. 3. The cumulative RSD for the values obtained from initial (system suitability)values and the individual online standard solution should be less than 2.0% Experimental design for an assay by HPLC method validation Analytical parameters Experimental design Acceptance criteria Specificity 1. Diluent, impurity interference 2. Separate injection of individual impurities for identification and assay in case of chromatographic method. 2.The impurities of impurities should not interfere at the retention time of main compound 3. Compare the Relative retention times of each component obtained from individual solutions and blend solution. 3. The relative retention times (with respect to drug substance) of each impurity obtained from the blend solution should be comparable(within + 0.06) to the relative retention times obtained from the individual solutions of each impurity. 1 System suitability as applicable 1. Establish the system suitability based on the condition Precision 1. There should be no interference of the diluent or impurities & the RT of drug substance. 2. Repeatability of 5 replicate determination of standard solution at assay level concentration. 3. Method precision of atleast 6 replicate assay dermination of a sample. Determine the %RSD of assays. 2 The % RSD for peak area response of replicate standards should not be more than 2.0 % 3 % RSD of assay should not be more than 2.0 % Experimental design for the method validation of Assay by HPLC Analytical parameters Experimental design Acceptance criteria Linearity & 1. Determination of minimum 7 levels of 1. standard in the range of at least 50 to range 150% (50%,80%,90%,100%,110% 120% and 150%) of the assay level concentration .Carry out linearity for 2. first & last level in five replicates & 3. other levels in triplicate and take mean response for calculation and also calculate the % RSD for peak area 4. response of three replicates of each level . 1. Linearity co-efficient of mean response of replicate determination plotted against respective concentration should not be less than 0.999. The %y-intercept as obtained from the linearity data should be within 2.0 % determined using the formula : 100 x y intercept = ------------------------------------Response of assay level 2. Determine linearity co-efficient and % y intercept. 2. The % RSD for peak area response of three replicates of each level should not be more than 2.0 % . 3. Determine the precision of lower (1st level) and higher level (last level) in five replicates. 3. The % RSD for peak area responses of five replicates of first & last level should not be more than 2.0 % Experimental design for the method validation of Assay by HPLC Analytical parameters Experimental design Acceptance criteria 1. Establish the system suitability Ruggedness (Intermediate precision) 2. Repeatability of 5 replicate determination of Standard solution at assay level concentration. Determine the % RSD of peak area response & retention time. 1.System suitability as applicable 2 . The % RSD for peak area response of replicate standards should not be more than 2.0 % 3. Intermediate precision of Six replicate determination of assay of a sample (Which is assayed under precision study) with the combination of different analyst / day / instrument / different column of same make. Determine the %RSD of assays. 3 % RSD of assay should not be more than 2.0 % 4. Determine the cumulative% RSD of assays of precision study and Intermediate precision study together. 4 Cumulative % RSD of assay of precision study &intermediate precision should not be more than 2.0 % Experimental design for the method validation of Assay by HPLC Analytical parameters Experimental design Acceptance criteria Robustness 1. Deliberately alter any 2 critical parameters(as per requirement) by minor variation. a Mobile phase composition (higher concentration by 2 % or lower concentration by 5-10 %) b c pH of mobile phase ( 0.2 ) Column oven temperature ( 2.0 °C) d Mobile phase flow rate 2. Establish the system suitability 3. Method precision of three replicate determination of assay of a sample (Which is analysed under precision study) under each individual condition of the robustness study . Determine the %RSD of assays. 4.Determine the cumulative % RSD of Assay of Precision study & Robustness study together. 1. Select as per the applicability of the method 2. System suitability as applicable. ( The % RSD for peak area response of replicate standards should not be more than 2.0 % 3 % RSD of assay should not be more than 2.0 % 4. The cumulative % RSD of assay of precision study along with the results of Robustness study should not be more than 2.0 % Experimental design for the method validation of Assay by HPLC Analytical parameters Experimental design Acceptance criteria Solution stability 1. Establish the system suitability 1 System suitability as applicable. 2. Prepare two samples of a batch and calculate the assay of each preparation and note down the result as initial assay and store one sample is on bench top and another one is in refrigerator simultaneously for about 24 hours. After the specified time interval evaluate the assay of each sample using the same chromatographic conditions(same mobile phase, column &system) with freshly prepared standard solution. Also monitor the pattern of chromatogram of sample solution 2. a. The % variation between the initial results and the results of solution stability(individually) should be less than 1.0% 3. When there is no solution stability for 24 hours in any of the above said conditions, determine the solution stability at predetermined time interval ( 2 hours, 4 hours, 8, hours,12 hours ) at laboratory temperature. If sample is sensitive carry out solution stability study by keeping the sample solution in refrigerator at 2-8 °C at same time interval of laboratory temperature. Note: After finding the time interval of solution at which is not stable no need to evaluate solution stability for the samples of remaining time intervals. The time intervals can be reduced and altered for sensitive drug substances . 3. Report the solution stability in hours (solution stability is the last time interval of the test solution is found stable) b .If any extraneous peak of impurities or degradants are observed in chromatogram of sample than that of initial pattern consider the solution is not stable. Experimental design for the method validation of Assay by HPLC Analytical parameters Experimental design Acceptance criteria Mobilephase stability 1. Prepare mobile phase and store one portion in refrigerator and the other portion is at laboratory temperature for at least two days 2. Establish the system suitability 3. 1. Establish the mobile phase stability for 2 to 7 days at each condition 2. System suitability as applicable Repeatability of 5 replicate determination 3 .The % RSD for peak area response of replicate standards should not be of Standard solution at assay level more than 2.0 % concentration. 4. Method precision of three replicate determination of assay of a sample (Which is assayed under precision study) using each storage condition of mobile phase. Determine the %RSD of assays. 5. Determine the cumulative% RSD of assays of each condition of mobile phase stability study and precision together. 4. % RSD of assay should not be more than 2.0 % 5. Cumulative % RSD of assay of precision study & each condition of the mobile phase stability study should not be more than 2.0 % Experimental design for the method validation of Assay Titremetry Analytical parameters Specificity Experimental design Check the blank interference of diluent. . Precision Method precision of at least 6 replicate assay determination of a sample. Determine the % RSD of assays. Acceptance criteria The Burette reading value should be negligible ( Around 0.1 ml ). * Blank reading should be taken for the calculation of assay . % RSD of assay should not be more than 2.0 % Linearity & range Determination of minimum 6 levels of standard in the range of at least 50 to 150% of the assay level concentration including assay level concentration. Carry out linearity for first & last level in six replicates & other levels in triplicate and take mean burette reading for calculation. Determine linearity co-efficient and % y intercept. Determine the precision of assay content of lower (1st level) and higher level (last level) in six replicates. Ruggedness Reproducibility of Six replicate (Intermediate determination of assay of a sample (Which is assayed under precision study) precision) by a different analyst / day at normal laboratory condition / instrument (if available)Determine the cumulative % RSD of assays of precision study and Intermediate precision study together. Linearity co-efficient of mean burette reading of replicate determination plotted against respective concentration should not be less than 0.999.The %y-intercept as obtained from the linearity data (without extrapolation through origin 0,0) should be within 2.0 % determined using the formula : 100 x y intercept = ------------------------------------B.R. of assay level The % RSD for % assay of replicate (1st level) and higher level (last level) should not be more than 2.0 %. 1 System suitability as per method 2 % RSD of assay should not be more than 2.0 % 3 Cumulative % RSD of assay of precision study & intermediate precision should not be more than 2.0 % Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical parameters System suitability Experimental design . 1. Establish the system suitability through out the validation followed by 2 procedures. 2.Procedure1:Establish the system suitability when ever the chromatographic system changed or disturbed. Acceptance criteria 1. As applicable (As per method details) e.g. : For chromatographic methods suitable parameters such as resolution, RRT , Peak to valley ratio must be determined. If not specified keep following criteria . 2. As applicable (As per method details) 3. Procedure2: When there is no 3. Should complies the system suitability disturbance in chromatographic system , establish the online system suitability between each parameter and observe the compliance of system suitability. Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical parameters Specificity Experimental design Acceptance criteria 1.Blank interference Inject the blank solution and conclude the interference due to blank at the retention times corresponding to each known impurity and the main compound 1.There should not be interference due to blank at the at the retention times corresponding to each known impurity and the main compound Inject the individual standard solutions of each impurity and the main compound at limit level concentration. And Establish the RRT for each component with respective to the drug substance retention time obtained from the blend solution. 2. The elution order and the RRT of each component obtained from individual solutions and the blend solution should be comparable (with in + 0.06) Prepare the blend solution containing all components (each impurity and the main compound) and inject. 1. Determine the signal-to-noise ratio at Limit of quantitation 0.01% level for each component and calculate the S/N ratio. If the S/N ratio (LOQ) is less than 10.0 for 0.01% solution, then establish the LOQ concentration based on the S/N ratio which should give the S/N ratio between 10 and 15 Method-1: Using soft ware Procedure: as appropriate by respective software 1.System suitability as applicable 2. The 0.01%(with respective to sample concentration) concentration is acceptable as LOQ level if the signal-to-noise ratio is more than 10 . or the solution which should give the S/N ratio is between 10 and 15 Method-2: Manual Experimental design for the method validation of chromatographic purity/ Signal to noise ratio can be calculated by substances using software or theRelated following formula. by HPLC Analytical 2 HExperimental design parameters S/N = -----------H Acceptance criteria where H = height of peak corresponding to the component concerned in the chromatogram obtained with the prescribed reference solution, measured from the maximum of ht peak to the extrapolated baseline of the signal observed over a distance equal to 20 times the width at half height. h = Range of the background noise in the chromatogram obtained after injection or applicable of a blank, observed over a 2. The % RSD for peak area response of replicate LOQ solution should not distance equal to 20 times the width at half be more than 5.0 % & 1.0 % for height of the peak in the chromatogram retention time. obtained with the prescribed reference Limit of quantitation (LOQ) Method-2: Manual Signal to noise ratio can be calculated by using software or the following formula. 2H S/N = -----------h where H = height of peak corresponding to the component concerned in the chromatogram obtained with the prescribed reference solution, measured from the maximum of ht peak to the extrapolated baseline of the signal observed over a distance equal to 20 times the width at half height. h = Range of the background noise in the chromatogram obtained after injection or applicable of a blank, observed over a distance equal to 20 times the width at half height of the peak in the chromatogram obtained with the prescribed reference solution & if possible, situated equally around the place where this peak would be found. 1. The 0.01%(with respective to sample concentration) concentration is acceptable as LOQ level if the signal-to-noise ratio is more than 10 . or the solution which should give the S/N ratio is between 10 and 15 Experimental design for the method validation of chromatographic purity/ 2. Determine the Related % RSD for peak area by HPLC substances Analytical parameters Limit of quantitation (LOQ) responses & retention time. Experimental design Note: Can be used either one of the methods mentioned above for the evaluation of S/N ratio. 2. Determine the % RSD for peak area responses . Note: Can be used either one of the methods mentioned above for the evaluation of S/N ratio. Acceptance criteria 2. The % RSD for peak area response of replicate LOQ solution should not be more than 5.0 % . Limit of detection (LOD) 1. Determine the signal-to-noise ratio of solutions prepared at appropriate concentrations based on the signalto-noise ratio obtained during the determination of LOQ. 1.The concentration is acceptable as LOD level if the signal-to-noise ratio is between 3 to 5 Calculate the Signal to noise ratio as choosen for LOQ study Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical Experimental design parameters Linearity & 1.Determination of minimum 6 levels of range all impurities standards for known impurities and drug solution for unknown impurities in the range of atleast 25 to 150% of the limit level concentration including limit of quantitation as first level and limit level concentration. 2.Carry out linearity for first & last level in six replicates & other levels in triplicate and take mean response for calculation. Note: Determine the precision at each level and calculate the % RSD ( which should not be more than 5.0%) before considering the mean value of each 3.Determine linearity co-efficient and % y intercept using Excel macros. (Continue linearity with LOQ chapter if possible). 4.Determine the precision of lower (1st level)and higher level (last level) for 6 replicates 6 replicates. Acceptance criteria 1.Linearity co-efficient of mean response of replicate determination plotted against respective concentration should not be less than 0.998. 2.The %y-intercept as obtained from the linearity data should not be more than 5.0 % determined using the formula : 100 x y intercept value = ------------------------------------Response of limit level Note: When the % Y intercept is not meeting the acceptance criteria, establish the linearity study separately at LOQ concentration considering the LOQ as level –1 and evaluate the linearity study separately and confirm the range. 3. The % RSD for peak area response of replicate first & last level should not be more than 5.0 % Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical parameters Precision Accuracy Experimental design Acceptance criteria A. System precision: 1 System suitability as applicabe 1. Repeatability of 6 replicate determination of Impurity solution for known impurities and drug for unknown impurities at the limit level concentration. 2 The % RSD for peak area response 2.Determine % RSD of the peak area of replicate standards should not be responses of individual known impurity and more than 5.0 % drug substance . B.Method precision: 1.Reproducibility of at least 6 replicate determination of a sample of a batch.. 2.Determine the % RSD of content of known impurities, single maximum unknown impurity and total impurities. 3 The % RSD of known, single maximum unknown impurity & Total impurities should not be more than 15.0 %. To a sample in which content of impurities is already known (Analysed under precision study), individual linearity levels of known impurities are added and the recovery is calculated. Carry out this exercise in minimum three replicates for atleast four levels each covering the LOQ,50,100 & 150 % of limit level concentration. Recovery should be between 80% and 120 % at LOQ level and for the other levels should be between 85% and 115%. Report % recovery in range at each level . Experimental design for the method validation of chromatographic purity Related substances by HPLC Analytical parameters Experimental design Acceptance criteria Ruggedness 1 Carry out system suitability (Intermediate 2. Repeatability of 6 replicate determination of precision) Impurity solution for known impurities and drug for unknown impurities at the limit level concentration. Determine % RSD of the peak area responses & RT of individual known impurity and the drug 1 System suitability as applicable 2 The % RSD for peak area response of replicate standards should not be more than 5.0 % 3 .Intermediate precision of six samples & inject once (which is analysed under precision study) by a different analyst / day / instrument / different lot column of same make. 4.Determine the % RSD of known, single maximum unknown impurity & Total impurities. 3. The % RSD of known, single maximum unknown impurity & Total impurities should not be more than 15.0 %. 5. Determine the cumulative % RSD of known, single maximum unknown impurity & Total impurities obtained from the results of precision study & ruggedness study. 4. The cumulative % RSD of the results obtained from precision study and Ruggedness study should not be more than 15.0 % Experimental design for the method validation of chromatographic purity/Related substances by HPLC Analytical Experimental design Acceptance criteria parameters 1. Deliberately alter any 2 critical parameters(as Robustness per requirement) by minor variation. Select as per the applicability of the method a Mobile phase composition (higher concentration by 2 % or lower concentration by 5-10 %) b pH of mobile phase ( 0.2 ) c Column oven temperature ( 5.0 °C) d Mobile phase flow rate ( 0.2 ml/min) 1. Establish the system suitability 2.Repeatability of 6 replicate determination of 2 Impurity solution for known impurities and drug for unknown impurities at the limit level concentration. Compare the RRT’S of individual known impurities &Determine % RSD of the peak area responses of individual known impurity and the drug. 1. System suitability as applicable 2.The RRT’Sof known impurities should be comparable with the RRT’S obtained from the specificity study 3.The % RSD for peak area response of replicate standards should not be more than 5.0 % 3. Method precision of three replicate determination of a4. The % RSD of known, single sample (Which is analysed under precision study) under maximum unknown impurity & Total impurities should not be more each individual condition of the robustness study . than 15.0 %. Determine the %RSD of known impurity, single max unknown impurity and total impurities. 5. The cumulative % RSD of the results 4. Determine the cumulative % RSD of known, obtained from precision study and single maximum unknown impurity & Total robustness study should not be more impurities obtained from the results of precision than 15.0 % study & robustness study. Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical parameters Experimental design Acceptance criteria Solution stability 1.Prepare 2 sets of samples of a batch in 1. System suitability as applicable. triplicate and store one set of samples is on bench top and another set is in refrigerator simultaneously for about 24 hours. 2.After completion of the specified time interval establish the system suitability and inject both the conditions of the solution 2.a. The cumulative % RSD of the results obtained from precision study and robustness study should not be more than 15.0 % stability samples individually and calculate the % RSD between the results obtained b.If any extraneous peak of from precision study and each condition of impurities or degradants are solution stability study individually . observed in chromatogram of Also monitor the pattern of chromatogram sample than that of initial pattern of sample solution. consider the solution is not stable. Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical parameters Experimental design Acceptance criteria Solution Stability 3.When there is no solution stability for 24 hours 3.Report the solution stability in in any of the above said conditions, determine the hours(solution stability is the last solution stability at predetermined time interval time interval of the test solution is ( 2 hours, 4 hours, 8, hours,12 hours ) at laboratory found stable) temperature. If sample is sensitive carry out solution stability study by keeping the sample solution in refrigerator at 2-8 °C at same time interval of laboratory temperature. Note: After finding the time interval of solution at which is not stable no need to evaluate solution stability for the samples of remaining time intervals. The time intervals can be reduced and altered based on the intended application.. Experimental design for the method validation of chromatographic purity/ Related substances by HPLC Analytical parameters Experimental design Acceptance criteria Mobilephase stability 1. Prepare mobile phase and store one portion in refrigerator and the other portion is at laboratory temperature for at least 2 to7 days 1. Establish the mobile phase stability for 2 to 7 days at each condition 2. Establish the system suitability 2. System suitability as applicable 3. Repeatability of 6 replicate determination of Impurity solution for known impurities and 3 . The % RSD for peak area response drug for unknown impurities at the limit level of replicate standards should not be concentration. Determine % RSD of the peak more than 5.0 % area responses of individual known impurity and the drug 4 Method precision of three replicate determination of a sample (Which is analysed under precision study) under each individual condition of the robustness study . Determine the %RSD of known impurity, single max unknown impurity and total impurities. 4. The % RSD of known, single maximum unknown impurity & Total impurities should not be more than 15.0 %. 5. Determine the cumulative % RSD of known, single maximum unknown impurity & Total impurities obtained from the results of precision study & each mobile phase stability study 5.The cumulative % RSD of the results obtained from precision study precision study and each mobile phase stability study should be less than 15.0% Experimental design for the method validation of chromatographic purity by TLC Analytical parameters Specificity Experimental design Acceptance criteria 1. System suitability determination. 1 The impurity should be well separated from drug substance. 2 .Check interference of the diluent / 2 There should be no interference of the blank diluent / blank at the Rf value of drug substance and impurities Limit of Detection (LOD) Prepare a different concentrations of solutions of impurity & drug The intensity of the spot corresponding to lowest concentration , which is clearly visible is detection level . substances and check the lowest detection level . Precision Method precision of at least 6 sample The intensities of all impurities should preparations Check the impurities and be comparable . compare the intensity of impurity spots visually . Ruggedness (Intermediate precision) Prepare the LOD solution and observe. 1.The intensity of the spot corresponding to LOD solution should be visible clearly (for the concentration of LOD derived ) Experimental design for the method validation of Residual solvents by GC Analytical parameters System suitability Experimental design Acceptance criteria 1. Establish the system suitability through out 1. As applicable (As per method details) the validation followed by 2 procedures. 2.Procedure1: Establish the system suitability 2. As applicable (As per method When ever the chromatographic system details) changed or disturbed. 3. Procedure2: When there is no disturbance 3. Should comply with the system suitability requirements in chromatographic system , establish the Specificity online system suitability between each parameter and observe the compliance of system suitability. 4. a) Resolution must be determined. 4. System suitability determination. (Resolution should be more than 1.0 (Repeatability of 6 replicate injections of between each solvent). solvent standard at the working level. Determine the % RSD of peak area response) b) The % RSD for peak area response should not be more than 15 % 1. 1. Diluent interference. 2. 1.There should be no interference of the diluent. If interference is observed, the response of interference should not be 2. Separate injection of individual solvents for more than 5.0 % to the response of Individual identification. standard concentration. Consider the interference observed for calculation. 22. Compared the retention time of individual solvents with respective retention time of solvents in standard solution. Experimental design for the method validation of Residual solvents by GC Analytical parameters Experimental design Acceptance criteria 1. Determine the signal-to-noise ratio at lower levels (equal to the first level of linearity & higher than LOD) Method-1: Using soft ware Procedure: as appropriate by respective software Method-2: Manual Signal to noise ratio can be calculated by using software or the following formula Signal to noise ratio can be calculated by using the following formula. 2H Limit of quantitation (LOQ) 1.System suitability as applicable 2. The concentration is acceptable as LOD level if the signal-to-noise ratio is between 10 to 15 S/N = -----------h where H = height of peak corresponding to the component concerned in the chromatogram obtained with the prescribed reference solution, measured from the maximum of ht peak to the extrapolated baseline of the signal observed over a distance equal to 20 times the width at half height. h = Range of the background noise in the chromatogram obtained after injection or applicable of a blank, observed over a distance equal to 20 times the width at half height of the peak in the chromatogram obtained with the prescribed reference solution & if possible, situated equally around the place where this peak would be found. Experimental design for the method validation of Residual solvents by GC Analytical parameters Limit of quantitation (LOQ) Experimental design Acceptance criteria 2. The % RSD for peak area Determine the % RSD for peak area responses response 2. The % RSD for peak area & retention time from 6 replicate injections. response of replicate LOQ solution of replicate LOQ solution should should not be more than 15.0 % not Note: Can be used either one of the methods be more than 5.0 % mentioned above for the evaluation of S/N ratio. Limit of detection (LOD) 1. Determine the signal-to-noise ratio of solutions 1.The concentration is acceptable prepared at appropriate concentrations based on as LOD level if the signal-to-noise the signal-to-noise ratio obtained during the ratio is between 3 to 5 determination of LOQ. 2. Calculate the Signal to noise ratio as detailed in LOQ study Experimental design for the method validation of Residual solvents by GC Analytical parameters Linearity & range Experimental design 1 Determination of minimum 6 levels of all solvents in the range of atleast 50 to 150% of the limit level concentration including limit of quantitation as first level and limit level concentration. 2.Carry out linearity for first & last level in six replicates & other levels in triplicate and take mean response for calculation. Note: Determine the precision at each level and calculate the % RSD ( which should not be more than 15.0%) before considering the mean value of each 3.Determine linearity co-efficient and % y intercept using Excel macros. (Continue linearity with LOQ chapter if possible). 4.Determine the precision of lower (1st level)and higher level (last level) for 6 replicates 6 replicates. Acceptance criteria 1. Linearity co-efficient of mean response of replicate determination plotted against respective concentration should not be less than 0.995. The %y-intercept as obtained from the linearity data (without extrapolation through origin 0) should not be more than 5.0 % determined using the formula : 100 x y intercept value = ------------------------------------Response of limit level Note: When the % Y intercept is not meeting the acceptance criteria, establish the linearity study separately at LOQ concentration considering the LOQ as level –1 and evaluate the linearity study separately and confirm the range. 2. The % RSD for peak area response of replicate first & last level should not be more than 15.0 % . Experimental design for the method of Residual solvents by GC Analytical parameters Precision Experimental design Acceptance criteria 1. Repeatability of 6 replicate injections of solvent standard at the standard level. Determine the % RSD of peak area response & retention time of individual solvent. 1 System suitability as applicable 2. Method precision of atleast 6 replicate determination of a sample. Determine % RSD of the content of each solvents. 3. The % RSD for content of solvents should not be more than 15.0%. 2.The RSD should not be more than 15.0% for peak area response 3. If solvents are not detectable in sample then 4. The % RSD of % recovery of spike working level concentration in sample. (Six solvents should not be more than sample preparation & inject once). To be 15.0%. established the method precision. (Consider this level as a level 3 of Accuracy). Accuracy To a sample in which content of solvents are already known (analysed in precision study), individual linearity levels are added and the recovery is calculated. Carry out this exercise in minimum three replicates for at least four levels each covering LOQ, 50,100 & 150 % of working level concentration. Recovery should be between 85 and 115 % at each level Report % recovery in range at each level . Experimental design for the method validation of Residual solvents by GC Analytical parameters Experimental design Acceptance criteria Ruggedness (Intermediate precision) 1. Carry out system suitability 1 System suitability as applicable 2. Repeatability of 6 replicate determination of each solvent at the limit level concentration. , determine % RSD of the peak area responses of individual solvent 2.The % RSD for peak area response of replicate standards should not be more than 15.0 % 3.Diluent interference 4.Intermediate precision of six replicate determination of trace solvents of a sample (which is analysed under precision study) by a different analyst / day / instrument / different column of same make, calculate the % RSD of solvents present in sample 5.Calculate the cumulative % RSD of content of solvents in precision study and Intermediate precision study together. 3.There should be no interference of the diluent. If interference is observed, the response of interference should not be more than 5.0 % to the response of standard concentration. Consider the interference observed for calculation. 4.% RSD for content of solvents should not be more than 15.0 % 1. 5.Cumulative % RSD for content of 2. solvents of precision study & intermediate precision together should not be more than 15.0 % Experimental design for the method validation of Residual solvents by GC Analytical parameters Experimental design Acceptance criteria Robustness 1. Deliberately alter any 2 critical parameters(as per requirement) by minor variation. a. Carrier gas flow rate b.Initial column oven temperature ( 2.0 C) 1. Select as per the applicability of the method 2.Establish the system suitability 2. System suitability as applicable 3. Repeatability of 6 replicate determination of 3. The % RSD for peak area response of replicate standards should not be each solvent at the limit level concentration more than 15.0 in each condition , determine % RSD of the peak area responses of individual solvent. 4. There should be no interference of 4. Diluent interference the diluent. If interference is observed, the response of interference should not be more than 5.0 % to the response of standard concentration. Consider the interference observed for calculation. 5. precision of three replicatedetermination 5. % RSD for content of solvents of trace solvents of a sample (which is should not be more than 15.0. analysed under precision study) in each condition Calculate the % RSD of solvents present in sample 3. 6.Cumulative % RSD for content of 6. Calculate the cumulative % RSD of solvents of precision study & content of solvents in precision study and robustness study together each condition of robustness study together. should not be more than 15.0 . Experimental design for the method validation of Residual solvents by GC Analytical parameters Experimental design Acceptance criteria 4. Cumulative % RSD for the content of solvents obtained from precision study & each condition of robustness study together should not be more than 15.0 %. Robustness 1. Deliberately alter any 2 critical parameters(as per requirement) by minor variation. a Carrier gas flow rate b Initial column oven temperature ( 2.0 C) 1. Select as per the applicability of the method 2.Establish the system suitability 2. System suitability as applicable 3 Repeatability of 6 replicate determination of 3. The % RSD for peak area response of replicate standards should not be each solvent at the limit level concentration more than 15.0 % in each condition , determine % RSD of the peak area responses of individual solvent. 4. Diluent interference 5. precision of three replicate determination of trace solvents of a sample (which is analysed under precision study) in each condition Calculate the % RSD of solvents present in sample 4.There should be no interference of the diluent. If interference is observed, the response of interference should not be more than 5.0 % to the response of standard concentration. Consider the interference observed for calculation. 5. % RSD for content of solvents should not be more than 15.0 %. 5. 6.Cumulative % RSD for content of solvents of precision study & 6. Calculate the cumulative % RSD of robustness study together content of solvents in precision study and should not be more than 15.0 %. each condition of robustness study together. Experimental design for the method validation of rinse sample analysis Analytical Experimental design Acceptance criteria parameters System suitability determination. System suitability as applicable Specificity 6. Cumulative % RSD for the of the 1. There should be no interference content of solvents obtained from diluent. precision study & each condition 2. For UV methods the interference due to of robustness blank/blank swab should not be more studyattogether be Diluent interference than 0.005 specifiedshould wave not length. more than 15.0 %. 1. Determine the signal-to-noise ratio Limit of quantitation for 1 ppm or higher levels .Carry out 6 replicate determinations at this (LOQ) concentration. 2. Determine the % RSD for peak area response/ absorbance Evaluation of S/N ratio A) For HPLC methods: Method-1: Using soft ware Procedure: as appropriate by respective software Method-2: Manual Signal to noise ratio can be calculated by using software or the following formula Signal to noise ratio can be calculated by using the following formula. 1. The1ppm solution is acceptable as LOQ level if the signal-to-noise ratio is more than 10 . or the solution which should give the S/N ratio is between 10 and 15 2. The % RSD for peak area response / Absorbance of replicate LOQ solution should not be more than 5.0 % 2H S/N = -----------h Experimental design for the method validation of rinse sample analysis Analytical Experimental design Acceptance criteria parameters Limitof quantitation (LOQ) where H = absorbance at specified wave length due to standard sample. ----- h = absorbance at specified wave length due to blank. consider always the h=0.005 (worst case) as default. C.For analytical techniques where signal-to-noise ratio cannot be measured, then calculate the quantation limit by thefollowing formula : 10 QL = ---S where = standard deviation of the responses as estimated by measuring magnitude by analysing 3 replicates at lower levels (lower or equal to the first linearity level and higher than LOD) and calculating standard deviation of these responses. S = slope of the calibration curve of the analyte as determined in the linearity and range study Analytical parameters Limit of detection (LOD) Experimental design for the method validation of rinse sample analysis Experimental design Acceptance criteria 1. Determine the signal-to-noise ratio of solutions prepared at appropriate concentrations based on the signal-tonoise ratio obtained during the determination of LOQ. 3. Follow the procedure as choosen for LOQ study 1. The concentration is acceptable as LOD if the signal-to-noise ratio is should be between 3-5). Linearity & range 1. Determination of minimum 6 levels of drug solution in the range of atleast 50 to 150% of the limit level concentrationi including limit of quantitation as first level and limit level concentration. Linearity co-efficient of mean response of replicate determination plotted against respective concentration should not be less than 0.995. 2. Carry out linearity for first & last level in six replicates & other levels in triplicate and take mean response / absorbance for calculation. and determine linearity co-efficient. 3. Determine the precision of lower (1stlevel) and higher level (last level) for 6 replicates. Analytical parameters Precision Accuracy The % RSD of first & last level for peak area response / Absorbance of replicate standards should not be more than 5.0 Experimental design for the method validation of rinse sample analysis Experimental design Acceptance criteria Reproducibility of % content of trace analyte obtained in the working level concentration of Accuracy level in six replicates. (100 % concentration) Prepare standard stock solutions. Transfer known quantity of stock solution on SS plate having curved edges to hold the solution. Clean the plate with diluent used for cleaning in production. Collect the combined washings in a dish. and reconstitute to the known volume and carry out determination. Prepare standard solutions of each level as above. Carry out recovery for at least three levels covering 50 to 150 % of limit level for at least 3 replicate each. and calculate % recovery . Report the least recovery (out of all the levels )result as the recovery factor of the method . % RSD of the content of trace analyte should not be more than 10.0 %. Recovery should be between 75% and 125%. Note: If the recovery factor is obtained between 85 to 115 %,there is no need to consider the recovery factor for the trace level determination. Ruggedness Repeat the precision study on a different (Intermediate instrument (if available) /different day / with different analyst and determine the precision) reproducibility of replicate determinations. Calculate together the %RSD of recoveries obtained from precision and intermediate precision study. % RSD of the content of trace analyte should not be more than 10.0 %. Cumulaqtive % RSD of content of trace analyte obtained from intermediate precision study and the precision study should not be more than 15%. Experimental design for the method validation of swab sample analysis Analytical Experimental design Acceptance criteria parameters System suitability determination. As applicable Specificity There should be no interference of the Diluent / Swab interference diluent. For UV methods the interference due to blank/blank swab should not be more than 0.005 at specified wave length Reproducibility of % content of trace analyte % RSD of the content of trace analyte Precision obtained in the working level concentration should not be more than 10.0 %. of Accuracy level in six replicates on each surface (100 % concentration) Accuracy (Recovery from surface) Prepare standard stock solutions. Transfer known quantity on SS plate, evaporate to dryness on water bath. Swab the surface of plate with swab. Transfer the swab in volumetric flask. Add known amount of diluent, sonicate for about 2 minutes and carry out determination. Prepare standard solutions of each level as above without evaporation. [In case of heat sensitive material, do not evaporate, add known volume of stock solution (about 1 ml) on SS plate, swab the plate and proceed as above]. Carry out recovery study for atleast 50 to 150 % of each level l in atleast three replicate and calculate % recovery. Report the least recovery (out of all the levels )result as the recovery factor of the method . Recovery should be between 75 and 125%. Note: If the recovery factor is obtained between 85 to 115 %,there is no need to consider the recovery factor for the trace level determination. Experimental design for the method validation of swab sample analysis Analytical Experimental design Acceptance criteria parameters Swab Releasing capacity (Recovery from swab) Transfer a swab stick into a test tube. Add Recovery should be between 85 % to 1 ml of the standard stock solution of 115 % highest Concentration study on swab surface. Allow the swab to soak the solution. Add known amount of diluent, swirl the flask and carry out determination. Prepare the standard by diluting quantity of the same standard stock solution to known volume as above. Calculate % recovery using the values Obtained from swab solution and standard solution(directly prepared) Carry out the above exercise using the standard stock solution of the 10ppm concentration and determine the % recovery. Ruggedness Repeat the precision study on a different (Intermediate instrument(if available) /different day/ precision) with different analyst and determine the reproducibility of replicate(6times) determinations. Calculate together the RSD of the results obtained for precision study and intermediate precision study. % RSD of the content of trace analyte should not be more than 10.0 %. Cumulative % RSD of content of trace analyte obtained from intermediate precision study and the precision study should not be more than 15%. Experimental design for the method validation of particle size analysis by lazer technique Analytical Experimental design Acceptance criteria parameters Precision Determine the stability of sample dispersion 1. RSD should not be more than 15 for 6 timesat different time intervals( with 5 or 25 % (decide based on the For Dry methods minutes interval) and calculate the % RSD of requirement) (Stability of d(10) , d(50) & d(90) values. dispersion) 2. RSD should not be more than 15 Method precision of atleast 6 sample Forliquid or 25 % (decide based on the preparation ,determine the % RSD of d(10) methods requirement) , d(50) & d(90) values. Ruggedness Intermediate precision of atleast 6 sample 1.RSD should not be more than or (Intermediate preparation, determine the % RSD of d(10) , 25 % (as decided in precision study) precision) d(50) & d(90) values. Also determine the cumulative %RSD of precision study and 2. Cumulative RSD of precision study and reproducibility study reproducibility study together . together should not be more than 25 % Robustness Deliberately alter any 2 critical parameters(as per requirement) by minor variation. Change in RPM speed of pump / magnetic stirrer/pressure / different obscuration/ change in sonication time for sample Method precision from 3 sample preparation in each condition, Determine the % RSD of of d(10) , d(50) & d(90) values. Also determine the cumulative %RSD of precision study and each condition of robustness study together . 1..RSD should not be more than or 25 % (as decided in precision study) 2. Cumulative RSD of precision study and robustness study together should not be more than 25 % Calculations : Standard deviation : Definition: . The standard deviation is a measure of how widely values are dispersed from the average value (the mean). Formula: Standard deviation (s) = ∑(Xi –X )2 n-1 Where n = no of observations in the set , X i = i th observation , X = average % Relative standard deviation Definition :The variability of data may often be better described as a relative variation rather than an absolute variation. This can be accomplished by calculating the coefficient of variation (CV) that is the ratio of the Standard deviation to the mean Formula RSD = standard deviation X 100 mean