TENDER CIIT EPC 02 (2015-16) ES 01 DEPARTMENT OF ENVIRONMENT (BIOTECHNOLOGY) Prices are required on C& F Islamabad Basis S.No Item Name Lab Name Description of Items Qty 01 Autoclave GE 01 02 CO2 Incubator SL Interior Material: Stainless steel, Temperature controlled, Temperature setting method, temperature display method, timer, Chamber volume 23 lit minimum, Max operational pressure: 0.18-0.2 MPa Dimensions External dimensions (WxDxH)770 x 730 x 905 mmInternal dimensions (WxDxH)643 x 523 x 700 mmVolume230 LNet weight (approx)90 kgPerformance Temperature control range and fluctuationAT +5 ~ +50 0.1Temperature uniformity0.25CO2 control range and fluctuation0 ~ 20 0.15Humidity level and fluctuation95Control Temperature sensorThermistorCO2 sensorDual IRDisplayLCD touch screenConstruction Exterior materialPainted steel (rear cover not painted)Interior materialSS copper alloyedInsulation materialExtruded polystyreneDHA heating systemYes Outer door 1 Outer door lockoptionReversible doorYesInner door 1Shelves 4Max. load per shelf7 kgMax. total load20 kgMax. shelf capacity10 kgAccess port - positionReardiameter30Alarms Power failureRearOut of temperature settingV-B-RHigh temperatureV-B-ROut of CO2 settingVB-RDoor openV-BElectrical and noise level Power supply230 VFrequency50 HzNoise level25 dB Co2 Cylinder & Regulator 03 Shakers ML Capacity: 144 Flasks of 25 mL to 6 Flasks of 6000 mL Temperature Range: 0 to 80C Speed Range: 20 to 420 RPM (at 60 Hz) 02 01 Real-time PCR system for rapid cycling up to 96 samples. Applications include absolute and relative quantification, melting curve analysis, and endpoint genotyping. Precise Instrument thermal homogeneity and cycling speed, accurate and reproducible results in a very short time. optical detection system flexibly detects sequence-dependent probes, such as hydrolysis probes and sequence-independent dyes (i.e., SYBR Green I). Multiplex and multicolor capabilities to permit use of up to four different fluorescent dyes. Application and Instrument Software creates a simple analysis workflow with easy import and export functions, email notifications after each run, online monitoring flexibility, and server-based network interfacing. Features Plate-based real-time PCR system with 96-well block Temperature gradient (assay optimization) Standalone system (integrated PC) Touchscreen with separate instrument software (free to load on any PC) Dynamic range of 10 log 04 GE qPCR instrument: 01 TENDER CIIT EPC 02 (2015-16) ES 01 DEPARTMENT OF ENVIRONMENT (BIOTECHNOLOGY) Prices are required on C& F Islamabad Basis Sensitivity of one copy 4-plex multicolor Runtime of less than 1 hour Applications: Absolute Quantification, Relative Quantification, Endpoint Genotyping, Melting Curve Analysis Intuitive state-of-the-art software PC analysis software can be loaded free of charge on any PC Thermal cycling system: Peltier-based, 96-well block ,Max ramp rate heating: 4.4°C, Average ramp rate cooling: 2.2°C, Programmable temperature range: 37–98°C Temperature accuracy :±0.2°C of programmed target temperature TM uniformity Range (max–min) 0.4°C SD < 0.1°C Gradient function Yes Gradient operational range 37–98°C Gradient programmable span (X–Y °C) Max. 20°C gradient range Excitation High-power broad spectrum LED Measurement (integration time) Simultaneous data acquisition for all positions in 10–1000 ms (dynamic mode) Detection CCD Camera Optical system Fixed fiber optics with four excitation and four emission filters No moving scanning elements Cq uniformity Range (max–min) 0.8 SD < 0.2 (enabling resolution of 2-fold concentration differences) Electrical approvals CE, ICE, UL Reaction volumes 10–50 µl Sample format 96-well plates, 8-tube strips Runtime < 40 min for 3-step 40 cycles PCR Melting measurement Continuous with 1–25 data points/°C Operating systems Win 7 + Win XP etc, Data analysis Abs Quantification Rel Quantification TM Calling Endpoint Genotyping HRM* Qual Detection* Data export Result table: export as .txt file Charts: Data and image export Data import Sample information can be imported from Excel, .csv or .txt files Failure flagging Automated flagging for critical controls (e.g., positive control is negative or negative control is positive) User definable flagging for further criteria (e.g. standard curve with error greater than x)