1 Major & Minor Losses Lucas White Group 4 Abstract The Modules used in this lab were the Technovate fluid circuit and the Edibon Energy Loss in Bends Module. From the gathered data using the Technovate fluid circuit, the friction coefficients for the two pipes of different diameters were calculated to confirm the benefit of using a pipe with a greater diameter to reduce head loss over the length of the pipe. These values were compared to the theoretical values calculated using the Colebrook Equation. The experimental error was greater for pipe 3 (25%) than pipe 4 (11.4%). This indicates the slower velocity makes for more consistent data. The data gathered using the Edibon Energy Loss in Bends Module was used to calculate and plot the minor loss in each fitting versus the squared velocity. The slopes of each trendline was obtained from the plots to calculate the experimental loss coefficient, K, and was compared to the theoretical K obtained from the textbook. When compared, the K values for the elbows fall within 10% experimental error, and the K values for the expanding and contracting fittings exceed 50%. This high error is somewhat due to human and systematic error, but is ultimately due to the method used in the calculation of the experimental K. Introduction Fluids flowing through a pipe carry a certain amount of energy. Some of that energy is lost due to friction. The roughness of pipes varies with material, even new pipes have an interior roughness. A higher roughness can cause Figure 1: Major Loss 𝑃1 𝑉1 + + 𝑧1 + ℎ𝑝𝑢𝑚𝑝 𝜌𝑔 2𝑔 𝑃2 𝑉2 = + + 𝑧2 + ℎ𝑇 + ℎ𝐿 𝜌𝑔 2𝑔 In the above equation, hL is the measure of energy lost in the form of pressure drops between two points in the flow. The total energy lost can be categorized as major loss or minor loss. The major loss (head loss) is the energy lost due to friction in the straight section of the pipe (Cengel & Cimbala, 2014). The friction acting on the fluid is created from contact with the surface of the pipe (Spellman, 2008). 𝐿 𝑉2 ℎ𝐿 = 𝑓 𝐷 2𝑔 (Image Source: Spiraxsarco, 2010) more turbulence in a flow, which leads to more head loss across the length of the pipe due to swirling and friction. The roughness is subject to change as a result of age and corrosion (Spellman, 2008). The length and diameter of the pipe are factors in the frictional contact with the flowing fluid. A pipe with shorter length has a loss due to friction. However, the larger the diameter of the pipe, the less head loss. A large diameter means less contact with the fluid per unit volume passing through the pipe. Also, a pipe with a smaller diameter 2 has a higher velocity than a larger diamerter pipe with the same volumetric flow rate. A higher velocity leads to a higher turbulence and head loss (Cengel & Cimbala, 2014). In addition to the losses due to the friction of the interior surface of the pipe, there are losses due to the flow through valves and fittings. There are several types of fittings: elbows of varying angles, expansions and contractions of diameter, Methods & Materials Materials The Techovate fluid circuit system consists of several flowmeters and copper pipes. Four manometers were used to take readings from two copper pipes of differing diameters and an orifice flowmeter. There Figure 3: Technovate Circuit System Figure 2: Minor Loss (Image Source: L. White) (Image Source: Slidesharecdn, 2013) and valves. The losses in these fittings and valves are rather minor when compared to the head loss, therefore, it’s referred to as minor loss (Cengel & Cimbala, 2014). 𝑉2 ℎ𝑓 = 𝐾 2𝑔 Each fitting and valve has a varying loss coefficient, K. For sudden enlargements in a pipe from a smaller diameter to a larger diameter, the K value is found to be: 𝐷1 2 𝐾 = [1 − ( ) ] 𝐷2 2 (Pope, 1997). For most fittings and valves, the manufacturers provide this value. were several connection points at specific locations on the pipes and flowmeters. Two pairs of clear tubes were used to get readings at theses connection points, which was taken from the manometers. The Edibon Energy Loss in Bends Module consisted of six fittings: a long elbow, expanding section, contracting section, medium elbow, short elbow, and a right angle fitting. Figure 4: Edibon Module All fittings were connected by pipes with an inner diameter of 20mm (excluding the 40 mm expansion and contraction pipe). Manometers were (Image Source: L. White) connected to 3 specific points along the module to measure the pressure drop at each fitting. additional trials. The clear tubes connected to both ends of pipe 3 were disconnected and connected to both ends of pipe 4. The process for pipe 3 was repeated for pipe 4. Methods The Techovate fluid circuit system was used to find the major loss on water flowing through two pipes of differing diameters and an orifice manometer. The clear tubes were connected to the specified pipe (Pipe #3) and the orifice flowmeter. The flow rate was adjusted until the maximum Pdrop across the orifice was obtained. The Pdrop across the orifice and pipe were recorded from the manometers. The flow rate was adjusted to be lower and the above process was repeated for 5 The Edibon Energy Losses in Bends Module was used to measure the minor losses in several types of fittings. The flow rate was adjusted until a Pdrop near 60mm could be observed in the right angle fitting manometers. The Pdrop of each fitting was recorded and the flow rate was determined by closing the drain of the dump tank and observing the rate of volume change in the tank. The flow rate was lowered and the process was repeated for 5 additional trials. Calculations The energy equation represents the constant amount of energy in a system. Whether the energy is transformed or transferred, it is neither created nor destroyed. 𝑃1 𝑉 𝑃 𝑉 + 2𝑔1 + 𝑧1 + ℎ𝑝𝑢𝑚𝑝 = 𝜌𝑔2 + 2𝑔2 + 𝑧2 + ℎ𝑇 + ℎ𝐿 (m) 𝜌𝑔 The head loss, hL, is obtained from the sum of head loss from the pipe and minor loss from any valves or fittings attached to the system, or from the simplified energy equation. 𝑃 𝑃 ℎ𝐿 = ℎ𝑚𝑎𝑗𝑜𝑟 + ℎ𝑚𝑖𝑛𝑜𝑟 = 𝜌𝑔2 − 𝜌𝑔1 (EQ. 1) The major loss was calculated from the pressure drop data taken from the manometers and the below equation was used to solve for the experimental friction coefficient, f. 𝐿 𝑉2 ℎ𝐿,𝑚𝑎𝑗𝑜𝑟 = 𝑓 𝐷 2𝑔 (EQ. 2) f = friction coefficient (no unit) L = length of section of pipe (m) D = inner diameter of pipe (m) V = velocity of fluid flow (m/s) The theoretical friction coefficient was calculated using the Colebrook Equation and the Goal Seek function in Microsoft Excel. 1 √𝑓 𝜀 𝐷 = −2.0 log( 3.7 + 2.51 𝑅𝑒√𝑓 ) (EQ. 3) 4 The minor losses were calculated from the pressure drop data taken from the manometers and the below equation was used to solve for the experimental loss coefficient, K. K= (slope of trendline) x 2g (EQ. 4) The theoretical loss coefficient for the expanding section (EQ. 4) and contracting section (EQ. 5) of the pipe was calculated using the following equation. 𝐷 2 2 𝐾 = [1 − (𝐷1) ] (EQ. 5) 2 Results & Discussion After the calculations of the velocity, head loss, and friction coefficient, the calculated data was plotted in the graph below. Figure 3 is a graph of the calculated friction coefficient values versus the squared velocity for two pipes of different diameters. According to the graph, the experimental friction coefficient values are slightly larger than the theoretical values. The average difference in theoretical and experimental friction coefficient values for pipe 3 is 0.00565 with an average experimental error of 25%. The error is due to human error in the gathering of data. Despite the large experimental error, the experimental and theoretical values were consistently within a 0.01 range. The average difference in theoretical and experimental data for pipe 4 was 0.00237 with an average experimental error was 11.4%. The data gathered from pipe 4 was more accurate and consistent with the theoretical values. The larger diameter of pipe 4 causes a lower velocity than pipe 3 when similar volumetric flow rates are applied. The lower friction coefficient values for pipe 4 confirms that a pipe with a larger diameter will have less head loss due to friction. The length of the pipe is a factor in the loss due to friction, however, both pipes are the same length. A longer pipe would mean more exposure to the interior surface which leads to more head loss. 0.035 0.030 f values 0.025 0.020 0.015 0.010 0.005 0.000 0 0.5 1 1.5 2 2.5 V squared Theoretical f (Pipe 3) Experimental f (Pipe 3) Experimental f (Pipe 4) Theoretical f (Pipe 4) Figure 4: Friction Coefficient values versus the squared velocity 3 5 Figure 4 is a plot of the minor loss of each fitting versus the squared velocity. From the plotted trendlines with a set zero intercept, the slope was used in Equation 4 to calculate the experimental loss coefficients of each fitting. According to the graph, the right angle fitting has the highest head loss at all velocities. This is due to the sharp 90 degree turn without any flanges or threads to cushion the fluid as it changes direction abruptly, which lead to more loss. The sudden expansion fitting has the least minor loss because of its sudden increase in diameter. The flowing fluid expands to fill the pipe and experiences less loss because of its limited contact with the interior of the pipe. The long elbow has more loss than the medium elbow but less loss than the short elbow. This is due to the long elbow slowing the direction change of the fluid. The medium elbow changes the fluids direction quicker with less head loss. The short elbow transitions quicker than the long and medium elbow and that causes more head loss than the other two elbows. The only notable difference between the right angle fitting and the short elbow is the flanges of the short elbow that lowers the loss. 0.06 y = 0.052x R² = 0.9751 0.05 y = 0.0385x R² = 0.9859 hL (m) 0.04 y = 0.0349x R² = 0.9545 0.03 y = 0.0166x R² = 0.9656 0.02 y = 0.0142x R² = 0.9464 0.01 y = 0.0084x R² = 0.9239 0 0.0 0.2 0.4 0.6 0.8 1.0 V squared Long Elbow Expansion Compression Medium Elbow Short Elbow Figure 6: Calculated minor loss versus squared velocity Right Angle 1.2 6 The theoretical and experimental loss coefficient values and their experimental errors are displayed in Table 1. The theoretical loss coefficients were calculated or found in the tables of the textbook. The experimental errors for the elbow fittings were within 10% error. However, the errors of the expansion and compression fittings exceeds 50%. This indicates the method used to calculate the experimental loss coefficients is not as valid for these two fittings as it is for the elbow fittings. Long Elbow Expansion Contraction Medium Short Right Angle K (Exp) 0.326 0.165 0.755 0.279 0.685 1.02 K (Theo) 0.300 0.563 0.500 0.300 0.700 1.10 % Exp Error 8.56 70.7 51.1 7.13 2.18 7.25 Table 1: Experimental K values, theoretical K values, and percent experimental error Conclusion After gathering data and calculating the friction coefficients for the two copper pipes of different diameters on the Technovate fluid circuit and the theoretical friction coefficients using the Colebrook equation, it’s noticeable that a pipe of larger diameter has less loss due to friction. A larger diameter would have less contact with the interior surface of the pipe per unit volume flowing through the pipe. The larger diameter would also have a lower velocity which made for more accurate data, a difference in experimental error of 13.6%. The lengths of both pipes were equal and did not change so the group was not able to observe the effect of length on the head loss. According to Cengel & Cimbala, a greater length would cause more head loss because of a longer contact to the interior of the pipe (Cengel & Cimbala, 2014). The data gathered from the Edibon energy losses in bends module was used to calculate and plot the head loss and squared velocity. This plot showed the right angle fitting to have the highest loss of all fittings used. This relatively high loss is due to the abrupt change in the direction of the fluid flow. From the trendlines of each plot, the slope was used to calculate the experimental K. The theoretical K was obtained from the textbook depending on the fitting. After comparison of the two K values for each fitting, it was found that the method used for calculating the experimental K was accurate within 10% experimental error of the theoretical K for the elbow fittings. However, it was less accurate for the sudden expansion and contraction fittings with an experimental error greater than 50%. Some of the error is due to systematic and human error but ultimately to the method used in calculation. 7 References Text Cengel, Y. A. & Cimbala, J. M., 2014, Fluid Mechanics: Fundamentals and Applications. New York City, New York: McGraw-Hill. Spellman, F. R., 2008, The Science of Water. Boca Raton, Florida: Taylor & Francis Group. Pope, J. E., 1997, Rules of Thumb For Mechanical Engineers. Houston, Texas: Gulf Publishing Company. Images Spiraxsarco, http://www2.spiraxsarco.com/images/resources/steam-engineeringtutorials/10/2/Fig_10_2_3.gif Slidesharecdn, http://image.slidesharecdn.com/8-fm9flowinpipesmajorlosesco3-copy130328071733-phpapp02/95/8-fm-9-flow-in-pipes-major-loses-co-3-copy-35638.jpg?cb=1364455131