Gas/Vapor/Particulate Sampling, Calibration, Detector Tubes, CGM, PID, and Breathalyzer Lab Exercise Elements: Prepare a series of standards Calibrate air sampling and measurement devices Measure and record the concentrations using: o Detector Tubes o Photo Ionization Detector (PID) o Combustible Gas Meter (CGM) o Breathalyzer Complete the tables Plot the data Answer the questions Equipment: 10-Liter Tedlar gas bags Microliter syringes One rotameter Air sample pump calibrated to a known flow rate Solvent stock solution Colorimetric indicator tubes (Gastec Sensidyne) Detector Tube samplepump (GastecSensidyne, piston-type) Photo Ionization Detector Combustible Gas Meter Breathalyzer Calibration of Colorimetric Tube Hand Pumps using a Bubble Soap Meter: Materials needed are a 100 ml burette, Tygon tubing, a soap reservoir with water, a mechanism to secure the burette and a detector tube pump (Sensidyne). Leak Check for Detector Tube Pump: Perform a leak check. 1. Insert an unopened tube into the bung. 2. Pull the piston pump and watch for expansion of volume over several minutes. If a leak is detected, disassemble pump front plate and remove exhaust valve disc. Wash disc in warm, soapy water, rinse, dry and re-assemble. Repeat leak check. If leak still occurs, replace valve disc with a new one. Calibration of Stroke Volume for Detector Tube Pump: 1. A soap bubble is started by inverting the burette into the soap container. (This may be facilitated by rinsing the burette with soapy water first to lubricate the inside column walls.) 2. Attach a hand pump to the colorimetric tube at the end of the Tygon tubing connected to the burette. 3. Draw the bubble up the column to the first reference mark. 4. Draw the air sample, one pump stroke. 5. Allow piston to expand fully to the end of the stroke. Bubble will move up column and stop. This is the volume of air that the pump draws per stroke (called the stroke volume). 6. Repeat steps 3 through 5, three times and determine an average stroke volume. It should be 100 ± 5ml. 7. Once a stroke volume of 100 ml has been verified, airborne measurements can begin. Gas/Vapor/Particulate Sampling, Calibration, Detector Tubes, CGM, PID, and Breathalyzer Lab Exercise: 1. 2. 3. 4. 5. Remove the air out of a clean 10 Liter gas bag using the lab bench top vacuum connection. Fill the gas bag with 10 Liters of air using lab air passing through a rotameter. For example, if the lab air passing though the rotameter has been determined to be 2 LPM, use the stop watch to record the time required to fill the bag and stop flow to the bag after 5 minutes (i.e. total 10 Liter volume). For a single solvent, prepare two standards in the concentration range of interest (0 ppm, 50 ppm, 500 ppm, 1000 ppm and 5000 ppm are suggested) by injecting the calculated amount of liquid solvent into the filled 10 Liter gas bag through the bag septum. Record the actual calculated concentration (column 1 of Table 2) given the amount of solvent actually injected (Column 2 of Table 2). Following the directions for use of the detector tubes, CGM, PID and Breathalyzer, take 2-3 replicate sample readings, record the readings and the average values in Table 2, in the appropriate data columns. Plot the readings versus the concentration for the detector tubes, CGM and PID for the chemicals tested. For the chemicals shown in Table 1, record the molecular weight (MW), liquid density (Density), ionization potential (IP), vapor pressure (VP), Lower Explosive Limit (LEL), Permissible Exposure Limit (PEL), Threshold Limit Value (TLV) and Immediately Dangerous to Life and Health (IDLH) values. Refer to the latest editions of the NIOSH Pocket Guide and the TLV Booklet or literature. Table 1. Physical and Chemical Characteristics of Common Solvents. Solvent MW Density IP (eV) VP(mm Hg) LEL (%) PEL(ppm) TLV(ppm) IDLH(ppm) Acetone Benzene Ethanol Toluene Table 2. Gas Standard Preparation & Direct-Reading Measurements for Ethanol. Gas Standard (ppm) 0 ppm 250 ppm 1000 ppm (L) Detector Tube Readings 3 Replicates and Mean CGM Readings PID Readings 3 Replicates and Mean 3 Replicates and Mean Breathalyzer Readings 3 Replicates and Mean Gas/Vapor/Particulate Sampling, Calibration, Detector Tubes, CGM, PID, and Breathalyzer BW Gas Alert Micro 5 Gas Detector with PID Background: The Gas Alert Micro5 can simultaneously detect up to five atmospheric hazards including oxygen, percent LEL combustible gas,Hydrogen Sulfide, Carbon Monoxide and photoionization detector (PID) with correction factors for a number of Volatile Organic Compounds (VOCs). The PID uses ultraviolet light as a source of energy to remove an electron from neutrally charged molecules creating ions and an electrical current. The ionization energy, measured in eV, is the amount of energy needed to remove an electron from a specific molecule. The PID can only detect substances with ionization energies lower than the lamp’s output of 10.6 eV. 1. 2. 3. 4. 5. 6. 7. Press to turn on the unit. Watch as the unit cycles through the start-up routine: a. Battery Check b. Date/Time c. Datalogging d. Self-Test of Sensor and Power e. LEL and PID correction factors When the detector reaches the “Pump Test”, watch for the “Block inlet” notification and place your finger over the inlet and wait for “Pump Test Successful” message. Important Note: The O2 sensor test requires an ambient atmosphere of 20.9% O2at each start-up in order accurately calibrate before use. The final test shows the days before next calibration. The pump will now display ambient readings in ppm for H 2S, CO, and the PID (shown here as SO2), as well as % O2, and % LEL Pressing the and simulateously accesses the options menu from which alarm, span gas parameters, PID and %LEL correction factors, display, datalogging, time/date, and other selections can be made. Calibration can be done by pressing and simultaneously in a clean atmosphere. a. The unit will first auto zero the toxic gas sensors and calibrate the O 2 sensor automatically. b. Next it will enter “auto span mode.” The toxic gases can then be calibrated simultanously or one at a time depending on the type of span gas available. The correct span gas values must be set in the Options menu before beginning calibration. c. Attach span gas to the pump inlet and turn on regulator to begin calibration. d. Wait for completion of calibration (2 minutes), the meter will then prompt for the next span gas. e. If you choose not to span the other sensors, press to continue with the calibration process. f. Press to choose to set calibration dates on the sensors that were calibrated during this session or press to skip this step. g. TWA and STEL alarms can be chosen by pressing and following screen prompts or may be pressed to skip this step. h. Finally press to save calibration information. To turn off the unit press and hold and wait for the countdown to complete. Gas/Vapor/Particulate Sampling, Calibration, Detector Tubes, CGM, PID, and Breathalyzer TSI SidePak Personal Aerosol Monitor Model AM510 Background:An aerosol particle counter works on the principal of either light scattering or light blocking. An aerosol stream is drawn through a chamber with a light source (either Laser Based Light or White Light). When a particle is illuminated by this light beam, it is redirected or absorbed. Light scattered by a single particle in a specific direction in relation to the original direction has a unique signature which relates to the size of the particle. This allows for sizing and counting of individual particles. The SidePak Personal Aerosol Monitor is a miniature battery operated laser photometer that measures airborne particle mass concentration in units of milligrams per cubic meter (mg/m3). 1. 2. 3. 4. 5. 6. Locate prepared dust chamber and sample train tubing. Set up sample train from dust chamber to SidePak. Shake dust chamber to circulate Arizona Test Dust for sampling and wait 5 minutes. Press the PAGE key to turn on instrument. Wait for the concentration to stabilize and record the instrument’s output. Remove sample train and take a reading from the lab room ambient environment. BACtrack S80 Professional Breathalyzer Background:The fuel cell has two platinum electrodes with a porous acid-electrolyte material sandwiched between them. As the exhaled air from the suspect flows past one side of the fuel cell, the platinum oxidizes any alcohol in the air to produce acetic acid, protons and electrons.The electrons flow through a wire from the platinum electrode. The wire is connected to an electrical-current meter and to the platinum electrode on the other side. The protons move through the lower portion of the fuel cell and combine with oxygen and the electrons on the other side to form water. The more alcohol that becomes oxidized, the greater the electrical current. A microprocessor measures the electrical current and calculates the BAC. %BAC = g/deciliter (dl) of blood, USA assumes a blood:breath ratio of 2100:1. Example: A standard prepared at 50 ppm of ethanol is measured with a breathalyzer. breathalyzer output in %BAC? 50ppm -> x mg/m3 𝑥 𝑚𝑔 𝑚3 = (50𝑝𝑝𝑚)(46.1 𝑙 24.45 𝑚𝑜𝑙 𝑔 ) 𝑚𝑜𝑙 What should the 98.4 mg/m3 -> x g/dl (breath) = 98.3 𝑚𝑔 𝑥 𝑚3 𝑔 𝑑𝑙 = (98.3 𝑚𝑔 𝑚3 )( 1𝑔 1000𝑚𝑔 )( 1𝑚3 10000𝑑𝑙 ) = 0.00000983 𝑔 𝑑𝑙 0.00000984 g/dl (breath) -> x g/dl (blood) -> %BAC 𝑥 %𝐵𝐴𝐶 = (0.00000983 1. 2. 3. 𝑔 𝑑𝑙 𝑏𝑟) ( 2100 1 ) = 0.0206 ~ 0.02 %𝐵𝐴𝐶 Prepare BACtrack S80 sample train by inserting the mouth piece and fit tubing between mouthpiece and prepared standard bag. Press START button on BACtrack and wait for countdown to complete and press down on to the prepared tedlar bag for 5 seconds or until a double beep sound. Wait and record results. Gas/Vapor/Particulate Sampling, Calibration, Detector Tubes, CGM, PID, and Breathalyzer Questions for Gas and Vapor Sampling 1. How many microliters are required to make a 100 ppm toluene standard in a 10 liter gas bag? Describe how the standard is prepared. 2. What analyte do the detector tubes used actually measure? (Chemical reactions) 3. How many milliliters did the detector tube piston pump actually draw? 4. How long did it take to draw the detector tube sample? 5. What do the CGM and the PID used actually measure? 6. Explain how the breathalyzer determines Blood Alcohol Content? 7. How precise were you able to prepare the gas standards? To what do you attribute the variation or errors? 8. How did the replicate detector tube, CGM, PID and breathalyzer meter readings compare? What factors would explain the similarities and/or differences? 9. If a PID reads 10 ppm when measuring an acetone standard, what is the concentration (in ppm) of the acetone standard? (Assuming a calibrant gas of isobutylene and a response factor for acetone of 5.2 relative to the calibrant gas) 10. Convert a 4000 ppm PID reading for toluene to %LEL. 11. What is the % concentration at saturation for ethanol (vapor pressure of 44 mm Hg) in a room which has a barometric pressure of 756 mm Hg? 12. Which method would you choose to monitor the vapors (tested in this lab) for compliance with the TLV or IDLH? Why? 13. What is the appropriate course of action if the data needed to conduct the evaluation is not found in the NIOSH Pocket Guide or TLV Booklet or TLV Documentation? 14. What concentrations of particulates were detected from the dust chamber and the ambient lab room atmosphere? 15. How does the laser photometer detect airborne particulate concentrations?