NOx & Ozone Gas Phase Titration (GP)Calibration (August 2008, rk) The Gas phase titration method to calibrate NO, NOx, NO2 and O3 is based on the one to one reaction of NO +O3 to form NO2. NO + O3 NO2 Things to turn on: 1. Check that NOx and O3 meters are on; usually they are on all the time. If thye are not, tune on the computer remote switch “S outlet mixing fan” 2. Check that the GPC air tank is on in the utility area. If it is empty you will have to use the back up pump on the wall (Figure 1) 3. Remote switch “S inlet mixing fan” shold be on. This turns on the solenoid on-off valve on the GPT air tank in the utility area, and the Tylan flow digital meters. It also switches the NOx and O3 meters to the GPT cal flask from the Mahogany sampling board. 4. When finished turn off the remote switch on data system, set to South inlet mixing fan to off). Doing this will return the NOx and O3 sampling lines to the chamber gas sampling manifold behind the particle instruments. Make sure the NO cal tank is off. The general Idea: The NO instrument is 1st calibrated by flowing and diluting known amounts of NO from a calibrated tank (usually at ~50 ppm + 2 %) with NOx free, clean air, into a flask. This produces a known concentration of NO at the outlet of the flask. A NOx instrument and an O3 instrument ”look” at these known concentrations. NOx free air usually flows into the flask, that will be called a GPT flask, at a flow of 2.5 to 4 L/min. (See Figures 1 and 2 below). C8urrently we are using a Tylan flow of 3.5 l/min. O3 generator Tylan flow meter purifil GPT flask 50 ppm NO Cal tank dryrite rotameter 3-way valve on/off valve 10 L/min ss metabelows pump GPT tank air “south inlet mixing fan relay” UV O3 calibrator or nebulizer particle size calibrator Figure 1 Schematic of GPT cal system Figure 2. Picture of GPT Cal system In a complete calibration, four different NO concentrations are added to the flask, starting with ~0.1 ppm and going to ~0.6 to 0.8 ppm. To do this, you will start with an NOx flow of 0.01 l/min and an air flow of ~4.00 l/min (~0.12 ppm NO). After a stable reading, you will then increase the NO flow to 0.02 l/min then 0.04, and finally to 0.06 (~0.74 ppm). We will then add O3 to the flask by turning on the inline O3 generator to titrate some NO and create NO2 so that there is a mixture of NO and NO2 in the GPT flask. Typically, NO is in excess over O3, so that not all of the NO reacts. This change in NO gives the amount of NO2 that forms and the amount of O3 that had to react to give the change in NO. Not all of the O3 will react in the flask, and the amount of ozone that makes in through the flask is recorded. The NO flow to the flask is then removed and the total O3, (un-reacted and reacted) is observed. The un-reacted O3 is subtracted from the total O3 and this equals the calibrated change in NO. 1. To begin: The NOx and O3 instruments in the new aerosol lab should be on for at least 2 hours. Make sure the NOx meter pump is on. The old NOx meter vacuum should be ~ 26 inches of Hg. There is a filter in back of the NOx meter and replacement of the filter is necessary after every 3-4 experiments. If you are using the new Monitor Labs NOx meter (currently it is broken), you need to check the dryer on the top of the NOx instrument. If it does not have 1 inch of blue zone, it needs to be replaced (pink indicates that the dryer is spent). We are currently using an old Bendix Nox meter (Figure 3). The vacuum pump should read at least 25”Hg. (Figure 4). Note, there is a back up Monitor Lab Nox meter in the utility room. There are two components to this NOx meter. They are in separate blue instrument boxes that are on the other side of the wall from the GPT cal area. The one on top is the pneumatics and pump box and the other is the electronics box. The directions on the meter say to power up the pneumatic box 5 minutes after the main power is turned on. I don’t know why , but this message is left over from the 1980s at the site. The power cord usually is plugged in. If you have to use this NOx meter, there is a sampling line that run into the For the old Bendix Nox meter that we are currently using, make sure the put the NO and NOx settings on the range scale of 1 ppm. Do not move the span or zero settings. If using the back up Monitor labs Nox meter, on the pneumatic box on top, see that the O 3 rotameter is 30-40% of full scale and the other two are ~80% of full scale. (If you add more than 1ppm of NOx, change the NOx & NO2 to Figure 3. Old Bendix 2ppm. Do this for both instruments) Chemilumenescent Nox meter 2. Bring up the NOx computer (usually it is on, but the monitor may be off; it is next to the black data logger). The Tylan flow controller boxes (see Figure 2) should be on, because they are turned on by remote switch S inlet mixing fan. If you do not see the digital numbers on the Tylan flow meters, turn on remote switch “South Inlet Mixing Fan. This also switches the sampling of the NOx and O3 instruments to the GPT cal flask form the Mahogany board chamber sampling manifold. 3. The general procedure to calibrate the NOx meter will be to first take zero NOx and O3 readings with just NOx free air flowing through the flask; then you will create dilute mixtures of NO in nitrogen in the GPT dilution flask. NOx free air to the GPT flask can come from an air tank in the utility area called GPT air, or if the tank is empty, Figure 4. Pressure Gauge on Old from a pump on the wall. Note the air goes Bendix Chemilumenescent NOx through a dryer column (should has at least 5 cm meter of blue indicator) and a purifil column to remove NOx. Adjust the air flow to 3.5 liters/min on the Tylan air flow controller read out, by adjusting the air flow rotameter in Figure 2. The Tylans read in liters/min at zero oC. Beginning the calibration: 5. It is very important to purge the NO regulator before conducting a cal: This procedure is performed so that any air in the regulator that results during no operation, does not back diffuse in the NO tank. If it does, O2 will slowly oxidize NO and ruin the tank. Typically, you may find the NO flow controller needle valve connected to the regulator and this supplies NO to the Tylan flow meter. You will have to disconnect this apparatus (see red male quick disconnect and the NO needle valve), since this is the normal position you will typically find the system in (see picture above, with labeled NO flow needle valve.). After removal replace this with another male quick disconnect that is attached to an 1/8” tube that vents directly into the vertical waste pipe behind the NOx instrument. To purge the regulator, close the low pressure regulator outlet valve near the quick disconnect. Then quickly open the main valve of the tank and close it quickly. You should see the pressure on the main tank gauge of the regulator to ~1000 psi. Now vent the regulator by opening the regulator outlet valve. When all of the pressure on the main regulator gauge goes to zero, close the outlet valve, and quickly open the main tank valve. Repeat this 5-7 times. Again, this is very important because it purges oxygen from the regulator which can back-diffuse into the tank and slowly oxidize NO to NO2. Now adjust the tank outlet pressure to 10 to 15 psi. 7. The Tylan flow controller Flows: It would be great if we could read the Tylans directly, apply a calibration factor and then correct them for the temperature in the GPT flask, to get the real flow. For now we are checking our previous Tylan air flow calibration with a “spot check flow” from a calibrated flow meter. If this is ok, there is no need for a more rigorous bubble meter flow calibration. The Tylan air flow when corrected to room air temperatures reads about 2-5% low. A check on the Tylan flow meter is performed by adjusting the air flow with the air flow rotameter valve on wall to ~2.5 to ~4 l/min. When the flow is stable on the Tylan read out, check the flow the large 496 ml soap bubble flow meter or the “dry cal test meter” Get 3 stable stopwatch times (say 12.36 sec+ 0.08 sec); record these and the air flow from the Tylan flow read-out in the spread sheet cells under the graph (Tylan air flow cal) on the left side of the spread sheet. Also enter the temperature from the temperature sensor near the GPT flask. The spread sheet will now estimate, based on the last cal, the predicted flow to the flask. This should agree with in 5% of the soap bubble meter. Now remove the air flow from the bubble meter and re-attach it to the GPT flask, adjust the air flow to ~ 3.5/min so we can get GPT zero readings with the NOx and O3 meters. The NO inlet to the GPT flask should be capped and the NO flow should not be attached to the GPT flask, because we will spot check its flow. Check the NO Tylan flow the same way you did the air flow, but use the 100 ml burette bubble flow meter. This, however, must be done for each NO flow that is used the Tylan flow meter is not stable from day to day. A note on the bubble meters; make sure you get good clean bubbles that are circular. Your should be level with the bubble as it crosses the line on the bubble meter at the bottom and at the top. Ideally, only one bubble should be in the bubble tower at a time. Readings should agree to 0.08 sec. 8. Adding NO to the GPT flask: Bring up the NOx-cal template excel spread sheet and give it today’s date. Chart readings NOx cal worksheet (change grey shaded cells only and save under a different name, blue cell values are calculated) OPERATORS: DiHu, SL DATE: 6/3/2007 Tank ID Tank NO conc= NO2 = Our new tank National specialty gases #cc51730, 2-20-98 50.5 ppm 0.82 ppm With NOx free clean air (often called zero air) flowing to the GPT flask and the NOx and O3 meters sampling form the rear of the flask record the the NOx and O3 GPT zero values. Note there are two places to record zero values. There is a bank set of cells the other says “GPT air concentrations only.” These should be the same, but if say were to humidify the air and look at how the NOx zeros change with different humidities we could record this in the top boxes. Typically we just record the GPT air concentrations only. NO 0.000 NOx O3 0.000 0 GPT air concentrations only NO 0.005 NOx -0.002 O3 0 The calibration can now proceed with either a one point or a, multi-point calibration. One point calibrations are done before each experiment. To do a 4 point calibration, 4 different NO flows have to be introduced to the to the flask (say 0.01, 0.02,0.03, and 0.04,l/min) waiting 7- 10 minutes between flows. The air flow can be left constant. Note in the spread sheet below the is a column to enter the air flow. Over the years we have found the Tylan air flow to be within 5% of the soap bubble meter so typically we just take the Tylan air flow reading and correct it from STP (0 oC) to the temperate at the GPT flask. The current version of the spread sheet reflects this. To increase the NO flows open the needle valve slightly till you get the desired flow. Record the raw Tylan air and NO flows and enter these into the spread sheet. 1st Calibration air flow NO flow l/min seconds 6.36 6.4 6.46 6.43 6.36 37.83 38.05 37.96 37.8 Soap bubble meter volume(cc) avg(sec.) 496 6.402 10 37.91 flow(L/min) 4.648547 0.01582696 3rd Calibration NO flow seconds 2nd Calibration airflow NO flow seconds seconds 15-Jan00 15.76 15.63 15.69 15.7 496.00 seconds L/Min 4.648547329 10 15.7 0.038217 seconds L/Min air flow seconds 496 #DIV/0! 10 #DIV/0! #DIV/0! #DIV/0! Again, enter the temperature at the GPT flask during each flow change. The spread sheet will compute the correct flows for air and Tank NO based on the temperature of the flask In the cells below, the spread sheet calculates the NO and NOx concentrations in the GPT flask. Record the the observed measured values in the appropriate cells. Based on these flows the following NO and NOx conc result from Cal Number #1 [NO ppm]tank x NO flow/total flow = [NOppm]flask predicted concentration NO NOx O3 NO meter readings NOx O3 0.732 0.718 0.720 0.744 0.004 Calibration factors NO measuredbkg cal factor #2 measuredbkg #3 Calfactor 3 #4 Calfactor 4 The way the NOx spread sheet computes the NO concentration in the GPT flask, is to start with a mass balance on the flask. That is: the mass/unit time of NO flowing into the flask = the mass/time fo NO flowing out of the flask or: [NOtank] x NO tank flow = [NOGPT] x total flowflask Say that the Tylan corrected air flow at 27oC = 4.00 liters/min to the GPT flask and the NO corrected flow = 0.02 L/min at 27oC, and the NO tank concentration is 50.08 ppm [NOGPT] = [NOtank] x NO tank flow / total flowflask= 50.08x 0.02/(4.00+0.02)= 0.249 ppm After you are finished with the cal, enter the calibration slopes into a file called NOx cal history. This way we can keep track of how the NOx meter is performing Make sure your NO predicted values do not exceed 0.9 ppm. The NOx meter overloads beyond 1 ppm. The multi-point cal spread sheet computes a regression plot in the spread sheet and will give you NO and NOx calibration lines in y= mx+ b format. 9. Calibrating O3 to the GPT flask: Starting with a stable NO and NOx concentration in the range of 0.90 to 0.6 ppm, record the O3 reading at the back of the flask. On a one point cal, the NO and NOx before the titration is keyed to the observed NO and NOx Ozone Calibration: Titration meter NOx 0.720 NO 0.718 before O3-back of flask NO after 0.323 adding O3 Corrected NO NO xcal factor O3 -0.0022 0.707 Corrected NO 0.407 0.075 NOx The reading for O3 should be equivalent to an O3 zero. Then turn on the O3 generator. This is the switch on the black box above the GPT flask on the wall (see picture). NO should decrease and NOx should more or less stay the same. Wait until NO and NOx are stable. This may take 10 minutes (because the UV lamp needs to warm up). Remember NOx should stay stable, even though NO and NO2 are changing. If NOx changes by more than 5% there may be a problem with the NOx catalytic converter. If we got a good NO and NOx cal first, even if NOx changes slightly due to flow changes, we can get back to the numbers. The assumption here is that the catalyst is working correctly. It is important to be taking O3 data at this point, because we need to assess the amount of O3 which does not react in the flask and makes it to the outlet of the flask before reaction (O3 flask outlet) . When NO and NO2 look reasonably stable after adding O3 to the GPT flask, record these values. A delta NO is calculated next, by subtracting uncorrected NO both before and after O3 is added. The correct NO is estimated by multiplying by the NO cal factor (slope). 10. After NO2 and NO are stable remove the NO flow from the flask and shut off the NO tank, this is very important. Also quickly cap the GPT flask and take O3 data. You will need to take at least 6-8 readings. Record the raw O3 value. remove NO flow to flask predicted O3= corrected delta NO corrected delta NO 0.407 raw O3 meter 0.409 Corrected meter O3 O3-O3fsk 0.334 NO flow factor 1.0166 O3 cal factor correct NO/O3=O3flask *NO flask factor 1.238744 Recall that the amount of O3 which made it through the flask is entered in or near cell D59. This will be subtracted from the total O3 measured when the NO tank flow is removed from the flask, and this difference is equal to the corrected NO. The corrected NO conc. is repeated in O3 calibration calculation above. Lastly, when we removed NO from the flask, the total flow to the flask flow decreased by a small amount equal to the NO tank flow. This must is taken into account, since the NO tank flow is no longer diluting the flask. It is computed by taking the original total flow and dividing it just by the air flow. This number is entered in cell for “NO flow factor” and then factored in computing the corrected meter O3. Finally, this corrected O3 is compared with the meter O3 and a calibration correction factor is estimated to be 1.24 in the above calculations. Record the O3 cal factor in the NOx cal history file. Turn off the O3 generator. Check that the NO tank is off 10-16-96 rk rev: 3-31-97 kl rev 3-23-04 rk rev aug 14, 2008, rk