Oxygen conserving devices Brian L Tiep, MD UpToDate performs a continuous review of over 375 journals and other resources. Updates are added as important new information is published. The literature review for version 15.3 is current through August 2007; this topic was last changed on September 8, 2005. The next version of UpToDate (16.1) will be released in March 2008. INTRODUCTION — The functional disability experienced by patients with severe pulmonary disease relates both to physiological impairment and to their ability to adapt and maintain a nearly normal lifestyle. As patients develop a medical need for supplemental oxygen, breathlessness and lack of energy are compounded by the increased demand of carrying oxygen equipment. Oxygen conserving devices have been introduced as a means of making oxygen therapy more efficient, more portable, and less intrusive [1,2]. In addition, hypoxemia can be prevented more readily in those patients with high oxygen flow requirements with oxygen conserving devices, such as reservoir cannula and transtracheal catheters. Escalating costs and reduced reimbursement have also fostered the availability of these systems. This review will compare traditional, continuous flow oxygen delivery by nasal cannula with a variety of oxygen conserving devices. The indications for long-term supplemental oxygen, the use of oxygen in hypercapnic patients, and issues regarding oxygen therapy during air travel are discussed separately. (See "Long-term supplemental oxygen therapy" and see "Use of oxygen in patients with hypercapnia" and see "Traveling with oxygen aboard commercial air carriers"). CONTINUOUS FLOW NASAL CANNULA — Continuous flow nasal cannula oxygen delivery via nasal prongs is the standard against which all developments or improvements should be compared [3]. The nasal cannula is universal and generally well accepted by patients and public alike, and its effectiveness in meeting the oxygen needs of the stable chronic lung disease patient is essentially unquestioned. However, it is inefficient, as only a small percentage of the oxygen delivered to the nose actually reaches the alveoli. Oxygen flowing through the standard nasal cannula is nearly 100 percent pure. However, because the oxygen becomes entrained in a much larger volume of inspired atmospheric air, containing 20.9 percent oxygen, the patient actually receives a blend of pure oxygen and room air [4]. The resultant inspiratory oxygen concentrations via usual flow settings are approximated in Figure 1 ( show figure 1). Assuming the most effective oxygen delivery occurs in the first 200 msec of inspiration, the fraction of inspired oxygen (FiO2) can be calculated and expressed as a percentage. For example, the typical flow setting of 2 L per min would raise the FiO2 to about 28 percent. This is a small oxygen enrichment, but it adequately corrects hypoxemia in most patients during rest [5]. A model of inspiration is shown in Figure 2 (show figure 2) [4]. Assuming a tidal volume of 450 mL, the first gas to reach the alveoli is the 150 mL of oxygen-poor gas already occupying the airways; this is followed by the first 300 mL of inspired gas. The final 150 mL of inspired gas is destined to fill the airways at the end of inspiration (becoming the dead space gas), never reaching the alveoli. Ideally, all oxygen delivery should occur prior to this last 150 mL. Figure 3 shows that airflow slows at the end of inspiration, so that dead space inspiration occupies a disproportionate 50 percent of inspiratory time (show figure 3). Consequently, the ideal oxygen delivery should occur within the first 0.5 seconds of inspiration, as oxygen delivered beyond this time is largely wasted. One study, for example, demonstrated that nasal oxygen delivered incrementally earlier in inhalation progressively increased oxygen saturation (show figure 4) [6]. TYPES OF OXYGEN CONSERVING DEVICES — There are three types of devices for enhancing the efficacy of oxygen therapy: Reservoir cannulas Transtracheal catheters Demand oxygen pulsing devices Each device meets the goal of improving oxygen delivery, but each has unique features that recommend for or against its use in the individual patient. Reservoir cannulas — Reservoir cannulas function by storing oxygen during exhalation, making that oxygen available as a bolus upon the onset of the next inhalation. Oxygen is conserved because the patient can be adequately saturated on a substantially reduced oxygen flow. These cannulas are available in two configurations: A moustache configuration (Oxymizer®), in which the reservoir is located directly beneath the nose [7]. A thin, compliant membrane in the reservoir is pushed forward during exhalation, creating a chamber between the membrane and the posterior wall (show figure 5). This enables oxygen to be stored during most of exhalation. When the patient is ready to inhale, he/she receives the stored oxygen along with the continuously flowing supply oxygen. A pendant configuration (Pendant®), in which oxygen is stored both in a reservoir located on the anterior chest [8]. The reservoir membrane is pushed forward during exhalation, creating a chamber (show figure 6). This enables oxygen to be stored during exhalation in the reservoir. When the patient is ready to inhale, he/she receives the stored oxygen along with the continuously flowing supply oxygen. Both reservoir cannulas are simple, reliable, inexpensive, and disposable. They operate in response to the patient's nasal airflow. While the efficacy of both cannulas is similar, there are differences in design that affect patient preference: The Oxymizer tends to be more comfortable than the Pendant and, in some cases, the standard nasal cannula. However, it is noticeable on the face, causing some patients to refuse to wear it. The Pendant is less noticeable. Previous pendant configurations utilized ear loops and wide tubing that some found to be uncomfortable. The present configuration has thinner tubing and no ear loops. Both devices are partial rebreather systems; as they return some of the patient's warmed moisture, they effectively increase the relative humidity of the oxygen. Reservoir cannulas can be used by patients with high flow requirements because they improve the efficacy of oxygen delivery [9]. In the inpatient setting, reservoir cannulas have been used as a transition device from the Venturi mask, since they are less intrusive and facilitate eating and communication [10]. In the home setting, oxygen cylinders with limited storage and oxygen concentrators with limited flow range present barriers for patients on long term oxygen, particularly as the flow requirements reach 6 L/min. For these patients, a conserving device may provide the only opportunity to live at home and maintain adequate oxygen saturation. Transtracheal catheters — Delivery of oxygen by transtracheal catheter is discussed elsewhere in detail. (See "Transtracheal oxygen therapy"). However, it is also mentioned here for comparison with the other conserving devices. Transtracheal catheters function by delivering oxygen directly into the trachea through a small opening in the neck [11]. Continuously flowing oxygen is stored in the upper airways toward the end of exhalation and is delivered during early inhalation along with supply oxygen, bypassing the dead space of the upper airways. Transtracheal catheters have been successfully utilized to oxygenate patients with refractory hypoxemia [12]. This mode of delivery has the additional advantages of high cosmetic acceptability and, when used with higher flows, the potential for reduced work of breathing by lowering minute ventilation (show table 1) [12,13]. Demand oxygen pulsing devices — Demand devices deliver a pulse of oxygen to the patient only during inhalation. Interposed between the pressurized oxygen source and the nasal cannula, these devices consist of a sensor, logic circuitry, and a solenoid valve (show figure 7). The patient's inspiratory flow is detected through the nasal cannula or transtracheal catheter; the solenoid then opens and delivers a short oxygen pulse. The aim is to discharge all oxygen during the earliest part of inhalation [14]. Pulsing devices are available as stand alone modules or integrated into a liquid oxygen system, compressed gas system, or portable oxygen concentrator. Because they are high technology devices, a variety of protocols and formats may be included in their design. Some devices are electronic with high precision settings while others are pneumatically controlled and driven. Settings for oxygen delivery are adjusted in some devices by varying the length of the oxygen pulse. Other devices deliver an early pulse each time they discharge, avoiding lengthening of the pulse into the dead space portion of inhalation. With this high efficiency of delivery, oxygen saturation can be maintained by delivery of a pulse on every other breath, leading to lower flow requirements [14]. Demand conservers vary in efficacy from 3:1 to 7:1 compared to continuous flow delivery. The batteries may last from three hours to three weeks, depending on the delivery protocol. Some devices have a feature in which they automatically revert to continuous flow if they fail to detect a breath in 20 to 30 seconds, but this feature uses more battery current. Pneumatically driven units do not require batteries. The demand devices can be very efficient but are mechanically complex, so that failure is possible. They can be coupled with transtracheal catheters to combine the advantages of better inspiratory signal and cosmetic acceptability with high efficacy. One potential problem with some pulsing devices is their inability to adequately oxygenate some patients during exertion. This limitation can be prevented by modifying the device to deliver a larger pulse during the earliest part of inhalation or use a higher setting [15,16]. It is always advisable to test each patient during rest and exertion to assure adequate saturation on the device being prescribed. Oxygen conserving technology has spawned the development of newer oxygen demand delivery devices that therapeutically adapt to the patient's physiological requirements [15,17]. One such device has an activity sensor that automatically adjusts to the exercise setting and remains at that setting for 50 seconds following the discontinuation of activity [17]. Such devices not only conserve oxygen, but also respond to meet the higher energy requirements during exertion. Consequently, these devices may be therapeutic beyond simple oxygen supplementation. EFFICACY OF VARIOUS DEVICES — The efficacy of these systems is expressed as the ratio of oxygen flow required by continuous flow delivery to the flow required by the conserver to achieve equivalent oxygen saturation. Efficacy = Continuous flow ÷ flow to conserving device (at equivalent saturation) The data presented in this section represent statistical means, and the actual equivalency may vary between patients. Thus, it is advisable to draw a blood gas or measure oximetry to determine the actual prescription. Reservoir systems — The efficacy of reservoir cannulas compared to continuous flow delivery is shown in Figure 8 (show figure 8) [7,8,18-21]. Oxygen supplied at 0.5 L/min via the reservoir cannulas achieves equivalent saturation to continuous flow at 2 L/min. Similarly, 1 and 2 L/min reservoir settings are equivalent to 3 and 4 L/min continuous flow, respectively. Thus, the efficacy of reservoir cannulas ranges between 2:1 and 4:1. For higher flows, reservoir cannulas add 2 L per min to continuous flow settings. For example, a reservoir cannula setting of 4 L/min is equivalent to 6 L/min by continuous flow. Studies performed during exercise and sleep generally yield results consistent with resting values, albeit with some variation [20-22]. Transtracheal systems — The efficacy of transtracheal oxygen delivery ranges between 2:1 and 3:1 over continuous flow delivery. While the efficacy is reduced during exercise [13,23], patients have greater exercise tolerance and reduced work of breathing via transtracheal delivery [23,24]. (See "Transtracheal oxygen therapy"). Pulsed oxygen devices — The efficacy of pulsed demand oxygen delivery varies between 3:1 and 7:1 (show figure 9) [14,25-28]. In general, this efficacy is maintained during exercise and sleep [15,26-28]. Pulsing oxygen through a transtracheal catheter maximizes the efficiency of oxygen delivery at 7:1 over continuous flow [6,29]. Small lightweight compressed gas cylinders — The total weight of an oxygen system consists of the cylinder, regulator and the oxygen conserving device. Composite cylinders are manufactured from aluminum liners and strengthened by carbon fiber in an epoxy resin matrix. These cylinders weight less than 50 percent of standard aluminum cylinders. They accept higher pressures up to 3000 psi and thus improve the portability of the portable oxygen system. CLINICAL USE — Not all patients require an oxygen conserving device under all conditions. In general, an oxygen conserving device should be considered for patients who would benefit from a lower flow of oxygen or a lighter weight system, or those who have refractory hypoxemia. This would include patients who are ambulatory and wish to maintain an active lifestyle. Also, patients who live a long distance from their HME supplier should have a wider variety of options, as they must endure the practical limitation of less frequent home deliveries. Figure 13 demonstrates the effect of oxygen conserving devices on extending cylinder life (show figure 10). Each conserving device has inherent advantages and drawbacks which should be weighed by the physician and patient. The choice between systems often requires compromise. Table 2 compares the strengths and weakness of each device ( show table 2). Reservoir cannulas are the simplest, least expensive, and easiest for patients to convert from standard cannulas, but they may be obtrusive for some patients. Transtracheal oxygen delivery excels in cosmetics and may reduce the work of breathing, but it entails a minor surgical procedure, the catheter may form mucous balls at its tip, and it requires significant patient training. The demand devices are more efficient for both nasal and transtracheal delivery. These devices have enabled the development of integrative systems that maximize oxygen delivery. However, they make audible pulses that may be distracting and mechanical failure, albeit rare, is possible. Some pulsing devices fail to oxygenate patients adequately during exertion [16]. Economic considerations — The cost of long term oxygen therapy is significant and includes oxygen, equipment, and service to the home. In the United States, the Center of Medicare and Medicaid Services (CMS) has reported paying $2 billion for long term oxygen therapy. In 1989, CMS instituted a system of fixed reimbursement based on liter flow prescription for home oxygen. Since then CMS has instituted several significant reductions in reimbursement for home oxygen. The profit margin for the home medical equipment (HME) suppliers can be maintained only by reducing their overhead. As a result, the HME supplier is economically driven to provide the least expensive system to purchase and maintain and one that requires less frequent home service calls. However, in localities that have several competitive HME suppliers, those who supply patients with a more portable and ambulatory system may have a competitive advantage. Liquid oxygen, which offers the highest efficiency in oxygen storage with safe and easy patient transfilling to his/her ambulatory unit, is unfortunately the most expensive system to purchase and maintain, and it requires more frequent home deliveries. Consequently, there is a financial disincentive to offer liquid oxygen to patients. Two integrated systems of oxygen concentrators that refill oxygen cylinders, Venture Home Fill® (Invacare Inc, Elyria, OH), and Total O2® (CHAD Therapeutics Inc, Chatsworth CA), have become available. They obviate the necessity for home deliveries, which is the most costly aspect in providing domiciliary oxygen. The most commonly dispensed system is an oxygen concentrator with a backup E cylinder designated to serve in the event of power failure or for infrequent portable use. Concentrators contain a synthetic aluminum silicate that traps and removes nitrogen molecules and thus yields a gas mixture of approximately 95 percent oxygen and five percent argon [3]. Given their economic disincentive to supply liquid oxygen, some HME suppliers are considering alternatives. Oxygen conserving devices have expanded the delivery life of the smallest oxygen cylinders from one hour of continuous flow to seven hours of pulsed flow. Some of these compressed oxygen systems weigh less than 1.8 kg, comparing favorably to standard 4.3 kg liquid systems supplying a similar delivery life. Newer liquid oxygen systems have also been developed that are lighter and more portable, owing to the integration of oxygen conserving devices. The Helios weighs 1.5 kg and lasts 10 hours between refills, thus providing the dual advantages of portability and transfillability. When considering the cost savings of oxygen conserving methodologies, the cost of the oxygen conserving device must be included. This actually happens in the marketplace, since the HME supplier must decide which equipment it will make available to patients. In addition, patients who require higher oxygen flows or have refractory hypoxemia would derive clinical benefit beyond convenience and portability, since the alternative may require the patient to tolerate hypoxemia. The Medicare system of reimbursement also has some financial disincentives to using oxygen conserving devices. If the patient, as a result of requiring less oxygen, falls into a lower reimbursement category based on a lower prescribed liter flow, the HME supplier would receive a smaller payment. With a device that is more expensive, this incentive seems backwards. Reservoir cannulas and transtracheal catheters are particularly affected by lower liter-flow settings, as opposed to the pulsing devices, whose settings (in absolute numbers) correspond to continuous flow settings. SUMMARY — Oxygen conserving devices offer greater versatility in portability, reduced costs, and feasibility of using oxygen systems previously deemed impractical. They all involve a series of tradeoffs. Each patient who is considered for an oxygen conserving device should have arterial blood gases or oximetry tested using that device during wakeful rest, usual exercise, and perhaps during sleep to ensure an adequate prescription. The development of oxygen conserving devices has spawned further developments such as a portable oxygen concentrator, oxygen concentrators that refill portable cylinders, and activity adaptive oxygen delivery devices. Use of UpToDate is subject to the Subscription and License Agreement. REFERENCES McCoy, R. Oxygen-conserving techniques and devices. Respir Care 2000; 45:95. 2. Keller, RR. Long-term oxygen therapy: advances and perspectives in technical devices. Monaldi Arch Chest Dis 1999; 54:75. 3. Rees, RJ, Dudley, F. ABC of oxygen: Provision of oxygen at home. BMJ 1998; 317:935. 4. Tiep, BL. Continuous flow oxygen therapy and basis for improving the efficiency of oxygen delivery. In: Portable Oxygen Therapy: Including Oxygen Conserving Methodology, Tiep, BL (Ed), Futura Publishing, Mt Kisco, New York, 1991, pp. 205231. 5. Mithoefer, JC, Keighley, JF, Karetzky, MS. Response of the arterial PO2 to oxygen administration in chronic pulmonary disease. Interpretation of findings in a study of 46 patients and 14 normal subjects. Ann Intern Med 1971; 74:328. 6. Tiep, BL, Christopher, KL, Spofford, BT, et al. Pulsed nasal and transtracheal oxygen delivery. Chest 1990; 97:364. 7. Soffer, M, Tashkin, DP, Shapiro, BJ, et al. Conservation of oxygen supply using a reservoir nasal cannula in hypoxemic patients at rest and during exercise. Chest 1985; 88:663. 8. Gonzales, S, Huntington, D, Remo, R, Light, R. Efficacy of the Oxymizer Pendant in reducing oxygen requirements of hypoxemic patients. Respir Care 1986; 31:681. 9. Collard, Ph, Wautelet, F, Delwiche, J, et al. Improvement of oxygen delivery in severe hypoxaemia by a reservoir cannula. Eur Respir J 1989; 2:778. 10. Sheehan, JC, O'Donohue, WJ. Use of a reservoir nasal cannula in hospitalized patients with refractory hypoxemia. Chest 1996; 110:s1. 11. Heimlich, HJ. Respiratory rehabilitation with transtracheal oxygen system. Ann Otol Rhinol Laryngol 1982; 91:643. 12. Christopher, KL, Spofford, BT, Brannin, PK, Petty, TL. Transtracheal oxygen therapy for refractory hypoxemia. JAMA 1986; 256:494. 13. Christopher, KL, Spofford, BT, Petrun, MD, et al. A program for transtracheal oxygen delivery: assessment of safety and efficacy. Ann Intern Med 1987; 107:802. 14. Tiep, BL, Nicotra, B, Carter, R, et al. Low-concentration oxygen therapy via a demand oxygen delivery system. Chest 1985; 87: 636. 15. Fuhrman, C, Chouaid, C, Herigault, R, et al. Comparison of four demand oxygen delivery systems at rest and during exercise for chronic obstructive pulmonary disease. Respir Med 2004; 98:938. 16. Tiep, BL, Barnett, J, Schiffman, G, et al. Maintaining oxygenation via demand oxygen delivery during rest and exercise. Respir Care 2002; 47:887. 17. Tiep, BL, Murray, R, Barnett, M, Carter, R. Auto-adjusting demand oxygen delivery system that minimizes SaO2 swings between rest and exertion. Chest 2004 (Abstract). 18. Tiep, BL, Nicotra, B, Carter, R, et al. Evaluation of a low-flow oxygen conserving nasal cannula. Am Rev Respir Dis 1984; 130:500. 19. Pesce, L, Lauro, S, Testi, R, et al. Effetti della somministrazione de ossigeno attraverso un nuovo dispositivo economizzatore in 20 pazienti affetti da insufficcienza respiratoria. Min Pneum 1987; 25:143. 20. Carter, R, Williams, J, Berry, J, et al. Evaluation of the Pendant oxygen conserving nasal cannula during exercise. Chest 1986; 89:806. 21. Hagarty, EM, Skorodin, MS, Stiers, WM, et al. Performance of a reservoir nasal cannula (Oxymizer) during sleep in hypoxemic patients with COPD. Chest 1993; 103:1129. 22. Cuvelier, A, Muir, JF, Czernichow, P, et al. Nocturnal efficiency and tolerance of a demand oxygen delivery system in COPD patients with nocturnal hypoxemia. Chest 1999; 116:22. 1. Hoffman, LA, Wesmiller, SW, Sciurba, FC, et al. Nasal cannula and transtracheal oxygen delivery: comparison of patient response after six months use of each technique. Am Rev Respir Dis 1992; 145:827. 24. Benditt, J, Pollock, M, Roa, J, Celli, B. Transtracheal delivery of gas decreases the oxygen cost of breathing. Am Rev Respir Dis 1993; 147:1207. 25. Garrod, R, Bestall, JC, Paul, E, Wedzicha, JA. Evaluation of pulsed dose oxygen delivery during exercise in patients with severe chronic obstructive pulmonary disease. Thorax 1999; 54:242. 26. Carter, R, Tashkin, D, Djahed, B, et al. Demand oxygen delivery for patients with restrictive lung disease. Chest 1989; 96:1307. 27. Franco, MA, Conner, SA, Gougenheim, C. Pulse dose oxygen delivery during exercise. Respir Care 1985; 30:888. 28. Bower, JS, Brook, CJ, Zimmer, K, Davis, D. Performance of a demand oxygen saver system during rest, exercise, and sleep in hypoxemic patients. Chest 1988; 94:77. 29. Yaeger, ES, Goodman, S, Hoddes, E, Christopher, KL. Oxygen therapy using pulse and continuous flow with a transtracheal catheter and a nasal cannula. Chest 1994; 106:854. 23.