NMR on the 750MHz Bruker Magnet Magic Angle Spinning: Spin controller: * options to operate in LOCAL or REMOTE mode use LOCAL * options to operate in MANUAL or AUTOMATIC mode use AUTOMATIC * Display: Dp = Drive Pressure Vd = spinning speed set to this value Bp = Bearing Pressure Va = actual spinning speed (Typical value for 10kHz spinning: Dp = 1080, Bp = 2660) Spin up new sample: 1. 2. 3. Mark the closed end of the rotor with black marker Press INSERT on the spin controller Type cfmas on computer to configure MAS for ‘insert air on time’ type 30s (this is how much time you have to put the sample into the magnet) ENTER also type 30s for eject time ENTER 4. Make sure that airflow is zero (at least under 1scfm) on VT controller 5. Press INSERT on spin controller 6. Place sample into magnet (you have 30s) * drive tip must be upward 7. Press INSERT on spin controller ( may need to press STOP) 8. Adjust SPINNING RATE by up and down arrows 9. Press GO * spin controller will automatically adjust the bearing and drive pressure 10. As soon as sample starts spinning increase airflow to 2scfm SLOWLY 11. If sample does not spin up press STOP EJECT STOP INSERT GO * if sample still does not spin up take it physically out and try spinning up again. * Possible problems with spinning: - drive tip may be damaged - airflow from VT must be zero Stop spinning: 1. Press STOP on spin controller or type mash on computer and then LOCAL on spin controller 2. Reduce airflow to zero on VT controller – Note: only Boris is allowed to do this 3. Press EJECT on spin controller * The cap covering the sample ejector must be about a 1cm off before ejecting the sample Amplifiers: Carbon X300 (300W output) X1000 (1kWatt output- use this) Proton H150 (150W output) H1000 (1kWatt output- use this) Pre-amplifiers: Low-Power Proton Preamplifier Low-Power Broad Power Preamlifier Deuterium Preamplifier High-Power Proton Preamplifier High-Power Broad Power Preamlifier Proton: Tuning: Proton Amplifier H150 output cable ‘#1 150W BLAH150’ (Bruker Linear Amplifier H channel) Low power preamp 750MHz filter 50ohm external reference High-tuning & taking spectra: Proton Amplifier H1000 output cable ‘#1 1000W BLAH1000’ (Bruker Linear Amplifier H channel) Bi-directional coupler 750MHz filter Probe (Note: proton high-power preamp is not working properly so we bypass it) Carbon: Carbon Amplifier X1000 output cable ‘#1 1000W BLAX1000’ (Bruker Linear Amplifier X channel) Bi-directional coupler High power preamp with 120-205MHz filter Probe Nitrogen: (Once Amplifier Gets Changed) Nitrogen Amplifier X1000 output cable ‘#1 1000W BLAX1000’ (Bruker Linear Amplifier X channel) Bi-directional coupler High power preamp with 70-125MHz filter Probe Probe: If in triple resonance mode and only two channels are used (for instance H and X), attach 50ohm dummy load to third channel (channel Y) How to Take Spectra Login: Make sure that you select KDE session To start NMR program: Type Xwinnmr in unix window To open data files: File search edit directory list Tuning LOW POWER TUNING: Proton 1. Set up cables the following way: Proton Amplifier H150 output cable ‘#1 150W BLAH150’ Low power preamp 750MHz filter 50ohm external reference 2. Open file tuneH – this is file with safe parameters for proton tuning 3. type eda to check/edit parameters 4. type edasp to edit wiring diagram which corresponds to physical wiring of channels/amplifiers/preamplifiers to the probe – double click on buttons to rewire For proton low power tuning have to following wiring scheme: Frequency BF1: 750.22000 SFO1: (calc.) OSFH1 Logical Channel F1 H Hardware Channel FCU2/SGU2 Amplifier H High/Low Preamplifier Power Stage H/F 150W 1H LNA Note: use current OFSH1 value (correction due to magnet drift) – on 06/04/03: -2500Hz 5. type wobb ext50 wobb = tune ext50 = external 50ohm reference resistance 6. type a for acquisition window 7. when spectrometer is done referencing a window will pop up and ask for attaching cable to the probe 8. take off 50ohm reference and attach cable to the probe channel H 9. tune and match 10. type stop Carbon: 1. Set up cables the following way: Carbon Amplifier X1000 output cable ‘#1 1000W BLAX1000’ Bi-directional coupler High power preamp with 120-205MHz filter Probe 2. Open file CNGly_CCP_highpowertune – this is file with safe parameters for carbon tuning 3. type eda to check/edit parameters 4. type edasp to edit wiring diagram which corresponds to physical wiring of channels/amplifiers/preamplifiers to the probe – double click on buttons to rewire (type edasp setpreamp if detected nucleus was N in previous experiment) For carbon low/high power tuning have to following wiring scheme: Frequency BF1: SFO1: (calc.) OSFH1: Logical Channel Hardware Channel F1 FCU1/SGU1 Amplifier X High/Low Power Stage X1000W Preamplifier HPHP XBB31P 13C Note: use current OFSH1 value (correction due to magnet drift) 5. type wobb 6. type a for acquisition window 7. tune and match 8. type stop Nitrogen: Note: with N always use FCU3/SGU3!!! The carbon and the nitrogen amplifiers do not generate the same power. 1. Set up cables the following way: Nitrogen Amplifier X1000 output cable ‘#3 1000W BLAX1000’ Bidirectional coupler High power preamp with 75MHz filter Probe 2. Open file 3. type eda to check/edit parameters 4. type edasp to edit wiring diagram which corresponds to physical wiring of channels/amplifiers/preamplifiers to the probe – double click on buttons to rewire (type edasp setpreamp if detected nucleus was C in previous experiment) For nitrogen low/high power tuning have to following wiring scheme: Frequency BF1: SFO1: (calc.) OSFH1: Logical Channel Hardware Channel F1 FCU3/SGU3 Amplifier X High/Low Power Stage X1000W 15N Note: use current OFSH1 value (correction due to magnet drift) 5. type wobb 6. type a for acquisition window 7. tune and match 8. type stop Preamplifier HPHP XBB31P HIGH POWER TUNING: 1. Rearrange cables so you bypass proton preamplifier the following way: Proton Amplifier H1000 output cable ‘#1 1000W BLAH1000’ Bi-directional coupler 750MHz filter Probe 2. open ‘highpowertune’ file 3. type edasp to edit wiring diagram which corresponds to physical wiring of channels/amplifiers/preamplifiers to the probe – double click on buttons to rewire For high power tuning have to following wiring scheme for HCN mode: (Note: Carbon setup is the same as for low power tuning, but Proton setup is different) Frequency BF1: SFO1: (calc) OSFH1: BF1: 750.22000 SFO1: (calc) OSFH1: BF1: SFO1: (calc) OSFH1: 4. 5. Logical Channel Hardware Channel Amplifier High/Low Power Stage F1 FCU1/SGU1 X X 1000W FCU2/SGU2 H H/F 1000W FCU3/SGU3 Y X 1000W Preamplifier HPHP XBB31P 13C F2 H F3 15N start pulsing by typing zg measure forward and reflected power Forward Power Reflected Power Bi-directional Coupler 6. tune and match until reflected power is 2-5% ex. 06/04/03 Nucleus Power Level Forward Power Reflected Power % Reflected C 5dB 1.6V 46mV 2.80% H 6dB 840mV (top of ramp) 43.6mV 5.20% 7. type stop General Notes Just good to know… * If you start pulsing the computer will automatically overwrite the previous dataset – if you want to pulse without saving new dataset type gs * To start a new acquisition which you want to save do the following: 1. open a file which was taken with the appropriate pulse program (ex. Test 1) 2. type wra which will copy the parameters into a new file which you distinguish from the previous one by giving it a different number (ex. Test 2) 3. type re 2 1 which will open (REad) ‘Test 2’ experiment where you can change the parameters you want to change Bug: * check BSMS (the controller where you can shim also)– sometimes it goes into “sweep” mode by itself – if it does so turn sweep mode off Safe Power Levels: * what may be confusing to a hardcore Varian user at first that the power levels of the pulses (aH, aX etc in Varian) are given in terms of attenuation. The highest attenuation is 120dB = 0 power to pulse (no pulse) The lowest attenuation is –6dB = highest power pulse * the highest safe power levels are listed for each probe next to the console by Boris Probes: * The sensitivity of HCN probe is 30% lower than that of HFX. Channels: * There are 4 channels on the instrument * Only 3 channels are functional : you can pulse on all 4 of them but you cannot detect on the 4th channel * Pulse sequences are written so you detect on Ch1 Pulses/pulse Sequences: * directory : /opt/xwinnmr/exp/stan/nmr/lists * abbreviations: cpd – decoupling programs gp – gradient programs pp – pulse programs wave – shaped pulses * if you edit an existing pulse sequence copy it into a new file name, and after you edit it make the extension your initials ex. If your name is Joseph Larmor yourfavoritepulseprogram.jl Pulses: * In system pulse programs: by default delays are in seconds, pulses are in microseconds (us) Warning: in user written pulse programs the pulses are in seconds, you must type us in order to pulse only for microseconds – if you forget to do so you may pulse for several seconds which may fry both your sample and the electronics * shaped pulses: ex. Rampup80100.100 means: that the power level at the beginning of the pulse is 80% of the maximum, the end is 100% power level, and there are 100 steps in the ramp Decoupling: * Recommended decoupling sequence for proteins: XIX * Advantages of XIX: - the power level (pl13) does not need to be optimized (recommended: 80kHz), only the pulse length (p31) - independent of spinning speed Data Processing: * type edp for processing parameters (edit process) S1 equivalent to zero fill FT Fourier transform – use fqc for quadrature detection dp1 & dp2 to check spectral window display Macro to Optimize Acqusition Parameters by Array: * Type popt a window will pop up where you can define the parameter you want to optimize for instance: PARAMETER pl1 OPTIMUS POS MAX STARVAL 31 ENDVAL 8 VARMOD LIN INC 0.5 means that the pl1 (power level 1, most likely on aXcp) will be arrayed from 31dB to 8dB in linear mode, with 0.5dB increment, and the best value (at POSition MAXimum) will be selected by the computer. You can set up several optimizations in queue, and the program will pick the optimum value automatically of each optimization and use that value from that point on. SAVE UPDATE START * Note: there is a bug in the MACRO – the first time you type parameters the program won’t save them type them again To Kill an Acquisition: * pull down menu: DISPLAYACTIVE COMMANDS ALLOW FOR KILLING …. Acqusition Parameters: * Type eda to edit * Type ased to display parameters used in this pulse program (note: some parameters may not be displayed) PARMODE Pulse Sequence TD – – – NS D – – SW FIDRES DIGTYPE DE PL – – – – – P – CNST NUCLEI DS FW RG – – – – – AQ – PRGAIN CPDPRG2 – – WBST WBSW – – IN O1 (?) – – dimensionality – has to be consistent with pulse program ex: zg – zero go (1 pulse) time domain - number of points = al If 2D pulse sequence: there will be 2 windows next to each other – 1st window = al (direct dim.), 2nd window = al2 (indirect dim.) number of scans delays – these are specific to pulse sequence defined from D0 to D31 ex: D0 = initial t1 D1 = pulse delay spectral width digital resolution always use HADC+ for solids preacqusition delay power level attenuation of pulses – Note: look at safe power levels above! defined from PL0 to PL31 lowest: 120dB highest: -6dB (do not use it!) pulse length defined from p0 to p31 constants – frequency wiring diagram dummy scans – recommended: 16 with this probe filter width – make it about 1.25X larger than SW receiver gain Typical values: Proton: 32 Proteins: 256 For very weak signal: 8K acqusition mode Usually: qsim (quadrature acquisition) preamp. gain decoupling Ex. XIX number of points (?) wobble spectrum width - it can act similar to spectrum analyzer increments – ex. 2D experiment increment delays frequency offset (??) PULSE SEQUENCES Note: you need to double-check the pulse sequence you are using for the appropriate pulse designations! (Pulse programs may had been modified or the same pulse program may had been written by several people.)