Intercomparison of TLD Services in Slovenia Matjaž ŠTUHEC and Gorazd LAKOVIČ Jožef Stefan Institute , Jamova 39, 1001 Ljubljana, Slovenia Abstract Five personal-dosimetry services from Slovenia and neighbouring countries were examined, in order to intercompare the quality of their TLD systems. Irradiations were carried out with different irradiation conditions, where dose, energy and angle of incidence were varied. Comparison of the results with the ICRU performance test requirements showed a rather good agreement within the prescribed trumpet curve, with a few outliers on the upper edge or a few percent above the edge. An attempt was made to analyse the results according to the IEC type test. Although the statistics doesn't satisfy for decisive conclusions, the results indicate, that the investigated types of dosimeters, with the right energy compensation, tend to comply with the tested requirements. Introduction The quality of a supplier of personal-dosimetry service depends on the technical characteristics of the approved dosimetry system and the training and experience of the staff, together with the calibration procedures and quality assurance programmes. In Slovenia procedures for approving the dosimetry systems at the regulatory authority level are not yet established. As a preliminary investigation of performance quality of the dosimetry systems, the intercomparison was organised by the Slovenian National Laboratory for Dosimetry Standards. All three dosimetry services from Slovenia, one service from Ruđer Bošković Institute, Croatia and one from Bosnia and Herzegovina took part in the intercomparison exercise. In the first part a general performance test was made according to the ISO 14146 [1] recommendations. In the second part, the measured characteristics of four different dosimeter types were compared with the most important IEC1066 [2] type-test requirements: linearity, energy dependence and angular dependence. Irradiations The dosimeters were irradiated as sets of three on the ICRU slab water phantom with Hp(10) doses. All irradiation conditions complied at the ISO 4037 [3] standard. Overall uncertainty of the irradiation doses was 5%. Each group was irradiated with specified conditions for dose, energy and angle of incidence, no fields with mixed energies were used. 1.) Linearity was tested with three doses at Cs-137 in the range from 0.1 mSv to 5.5 mSv. The lower limit was chosen to be near the monthly recording level and the upper limit in the range of annual limit for monthly monitoring, recommended by the ICRP [4]. 2.) Energy dependence was tested with 660 keV Cs-137, 1.2 MeV Co-60 and with Xrays at 40, 60 and 100 kV tube voltages with the narrow spectra qualities N40, N60 and N100. 3.) For isotropy test, the dosimeters were irradiated with the N60 X-rays at 300 and 600 angels of rotation around vertical axis. Irradiations were split into two half time shots for plus and minus angles, to get symmetrical conditions for the dosimeters lying out of rotation axe. For the last two tests, the doses were around 0.5 mSv. The dosimeters were then sent for evaluation to participating laboratories. Unirradiated dosimeters were included to measure travelling background dose, which was accounted for in the evaluation of the results. Responses of the participating TLD systems under the defined conditions are reported and compared to the directives of international standards. Results and Discussion After receiving the measured doses from the participants, and subtracting the travelling dose, the response was calculated for individual dosimeter, as a ratio of measured with true dose, R=Dm/Dt. In the first test, responses of all dosimeters were compared with the so called ICRP trumpet curve [3]. Trumpet 2.0 R 1.5 1.0 0.5 0.0 100 1000 10000 dose Fig.1: Responses R in comparison with the ICRP trumpet. Figure 1 shows the responses for all the individual dosemeters as measured by the participants in comparison with the ICRP trumpet. No additional corrections were made to the evaluated doses at this stage of evaluation. Except for a few outliers, the responses comply with the requirements of so called performance test for personaldosimetry services. In the second stage of evaluation, an attempt was made to compare the measurement capabilities of four participant systems, having different technical characteristics, to the requirements of the IEC1066 type-test. Four different types of personal dosimetry systems were compared: Harshaw and Rados systems with the dosimeters based on LiF:Mg, Panasonic system with the combination of LiB and CaSO4 tablets in dosimeter badges, and IJS CaF2:Mg energy compensated dosimeters. Although the statistical sample of the measurements, where only three dosimeters for individual irradiation condition were used, doesn't allow conclusive judgements, we tried to get at least an indication of how good the performance can be in comparison with the standard. The results are in the table 1 for energy dependence and table 2 for isotropy and linearity. Table 1: Energy dependence - errors of measured doses as R-1 for systems I-IV. E(keV) I II III IV 33 0.154 0.318 0.169 0.190 48 -0.108 0.066 -0.103 -0.010 83 -0.270 -0.218 0.038 -0.170 660 0.032 -0.065 0.015 0.050 1200 0.052 -0.040 0.015 0.033 Table 2: Isotropy - errors as R/R0-1, and linearity - errors as R-1. I II III IV 300 0.024 0.023 0.253 0.018 600 0.169 0.085 0.350 0.132 Sv I II III IV 105 0.007 0.013 0.005 0.043 530 0.032 -0.065 0.015 0.050 5554 0.032 -0.017 0.039 0.059 As the result of energy dependence and linearity, for each participant (I-IV) and each irradiation condition, the relative error is calculated as R-1. R is the response of the average of three measured doses for given irradiation, compared to the true dose, <Dm>3/Dt. For isotropy, the relative response R/R0 is evaluated, that is the ratio of response for angular incidence to the normal incidence. The results for energy dependence are displayed in the figures 2,3 and the results for linearity and isotropy in figures 4,5. Error bars in the figures indicate the measurement uncertainty for the groups of three dosimeters, calculated as weighted standard deviation of the mean, for the confidence interval of 95% probability. 0.300 0.225 0.150 R-1 0.075 0.000 33 -0.075 48 83 660 1200 -0.150 -0.225 -0.300 E(keV) Fig.2: Energy dependence of the TLD systems I(red) and IV(blue). 0.270 R-1 0.180 0.090 0.000 -0.090 33 48 83 660 1200 -0.180 -0.270 E(keV) Fig.3: Energy dependence of the TLD systems II(pink) and III(green). 0.210 0.140 R-1 0.070 0.000 30 60 ` 105 530 5554 -0.070 -0.140 -0.210 X: 30/60deg Cs: 0.1 - 5.5 mSv Fig.4: Isotropy and linearity of the TLD systems I(red) and IV(blue). 0.350 0.250 R-1 0.150 0.050 -0.050 30 60 105 530 5554 -0.150 X: 30/60deg Cs: 0.1 - 5.5 mSv Fig.5: Isotropy and linearity of the TLD systems II(pink) and III(green). Conclusions The following conclusions can be drawn from the results for the individual types of TLD systems: I: The dosimeters have no energy compensation, measurement of energy dependence results an average energy response of 1.58 for the X-ray energies. With this correction factor taken into account as an average X-rays energy correction, we get the energy dependence within the 30% IEC requirements. Isotropy complies at 300 incidence angle and is just near above the edge of 15% requirements at 600. Linearity complies also at the lowest dose, together with the broader, 10% confidence interval. A good batch homogeneity is observed, with the overall spread of the individual readings in the groups of three dosimeters to be 3% on the average, when leaving out a higher value of 8% only at the lowest dose. II: The dosimeters are energy compensated, with the relative error for energy dependence on the edge of 30% IEC requirements for the lowest energy X-rays, N40. The measurements indicate, that isotropy complies with the 15% requirements, although with a rather broad 10% confidence interval. Linearity seems to be within the 10% requirements, except that the lowest dose shows a broader uncertainty because of an outlier measurement. Average relative uncertainty of the groups of three dosimeters is 6% at high energies (leaving out the lowest dose) and about 10% for low energy measurements. III: The measurements tend to comply with the energy dependence requirements, although with rather large 30% confidence interval for the N60, where the results is at the limit for energy dependence. A good result is achieved for linearity, including the lowest dose. On the other hand, a discrepancy is observed for isotropy at N60 quality Xrays, showing about 30% error, as well as measurement uncertainty. A further investigation is needed with a better statistics to enlighten the findings. IV: The measurements indicate an overall good result, showing tendency to comply with all requirements, if better measurement statistics would be achieved. Poor homogeneity of some of the tested groups seems to be the cause of the three outliers. One of them, where one of the three dosimeter is obviously wrong at N60, is corrected in the final result. In the rest of two, the doses of all three dosimeters are spread in about 20% confidence interval. Disregarding these outliers, a 4% overall average measurement uncertainty is achieved. Acknowledgments Mr.B. Vekić is kindly acknowledged for his help at the stage of irradiations. References [1] International Organization for Standardization. Radiation Protection-Criteria and Performance Limits for the Periodic Evaluation of Processors of Personal Dosimeters for X and Gamma Radiation. ISO 14146, Geneva (2000). [2] International Electrotechnical Commission. Thermoluminiscence dosimetry systems for personal and environmental monitoring. IEC 1066 (1991). [3] International Organization for Standardization. X and gamma reference radiation for calibrating dosimeters and dose rate meters and for determining their response as a function of photon energy. ISO/FDIS 4037 1-3 (1999). [4] International Commission on Radiation Protection. General Principles for the Radiation Protection of Workers Monitoring. ICRP 75 (1997).