LHC Project Note xxx 2009-09-02 bernhard.auchmann @cern.ch Magnetic Model of the Superconducting Separation Dipole MBRB (D4) B. Auchmann, A. Jain, and J. Miles, for the FiDeL team CERN, Technology Department Keywords: Superconducting Magnets, Magnetic Field Model, Harmonics, LHC. 1. Introduction Function in the machine: The separation dipoles are used in several insertions to change the nominal beam separation (194 mm) of the LHC arcs. In the RF insertion (IR4), the D3 (MBRS) and D4 (MBRB) superconducting magnets are used to create a beam separation of 414 mm to make space for the RF system [1]. The MBRB is a double-aperture magnet. The field polarity is the same in both apertures. The yoke width is evaluated in order to accommodate the return flux. The magnets were designed, manufactured, and tested by BNL [2-4]. The magnet cross section is shown in Fig. 1. It is very similar to the MBRC (D2), except for the beam separation (194 mm in D4, 188 mm in D2). Fig. 1: Cross-section of the MBRB (D4) cold mass, from Ref. [4] Numbers and variants: Two magnets plus one spare have been manufactured (see Table I). Table I: D4 number and measurements Name Installed Spare Rejected T otal MBRB (D4) 2 1 0 3 Measured at Measured at room 1.9 K temperature 3 3 This is an internal CERN publication and does not necessarily reflect the views of the LHC project management. Naming convention: Magnet IDs read as follows: HCMBRBA001-BL00000x (x = 1…3). The electrical circuits are given in Table. Expected operational cycles, range of current: Table II summarizes the excitation of individual magnets at three different energy levels. The required field is 0.25 T at injection, and 3.88 T at 7 TeV. Table II: Slot allocation and powering of the MBRB Magnet State E (GeV/c) Circuit B (T ) I (A) FiDeL HCMBRBA001-BL000002 injection 450 RD4.L4 0.25 394.69 HCMBRBA001-BL000001 injection 450 RD4.R4 0.25 394.72 HCMBRBA001-BL000002 collision 3500 RD4.L4 1.942 3068.09 HCMBRBA001-BL000001 collision 3500 RD4.R4 1.942 3068.09 HCMBRBA001-BL000002 collision 5000 RD4.L4 2.774 4385.12 HCMBRBA001-BL000001 collision 5000 RD4.R4 2.774 4385.12 HCMBRBA001-BL000002 collision 7000 RD4.L4 3.883 6175.73 HCMBRBA001-BL000001 collision 7000 RD4.R4 3.883 6177.62 Summary of manufacturing parameters, manufacturers, and operational temperature: The MBRB magnets were designed, manufactured, and tested at Brookhaven National Laboratory. The coils are straight, RHIC-type dipole coils. The field polarity in the two apertures is identical. Main parameters are summarized in Table. Table III: Main parameters of the superconducting separation dipole MBRB Magnetic length (m) 9.45 Operation temp. (K) 4.5 Coil inner diameter (mm) 80 Beam separation 194 Nominal field (T ) at nominal current 3.80 Measured field (T ) at nominal current 3.83 Nominal current (A) 6050 Min operational current (A) 395 Max operational current (A) 6178 2. Layout Slots and positions: The slot numbers of individual magnets are given in Table II. Fig. 2 shows the schematic layout drawing of IR 4. Fig. 2: Layout [1] of the right-hand side of the matching section in IR 4 (RF) -2- 3. Measurements Device: A 1-m-long rotating coil of 25 mm radius was used to measure all field harmonics in ten different longitudinal positions. Integral harmonics were computed from the local measurements. During room temperature measurements, the integrated dipole field was also measured using a 10-m-long non-rotating coil with two orthogonal dipole windings [3]. All magnets were measured horizontally in their cryostats at 4.5 K [8]. 3.1 ROOM TEMPERATURE MAGNETIC MEASUREMENTS Powering: Room temperature measurements were performed at approximately ±15 A. Available and missing measurements: Room temperature measurements are available for all magnets. Local measurements are available, as well as integral values. Rejected or faulty measurements: No rejected or faulty measurements have been observed. Use of the measurements in FiDeL: The room temperature measurements were not directly used for the FiDeL model, as 4.5 K measurements are available for the entire magnet family, but they can be used for cross-check. 3.2 MAGNETIC MEASUREMENTS AT 4.5 K Powering: Field quality at 4.5 K is available for the ramp up in the range of 200 – 6400 A, covering the entire operational range. For going from one current to the next, a ramp rate of 10 A/s was used. Measurements were taken 40 s after reaching the desired current level. Available and missing measurements: Measurements at 4.5 K are available for all MBRB magnets. Local measurements are available, as well as integral values. Pre-cycle: The field measurements were done after a cycle from 25 A to 6400 A at a constant ramp rate during up and down of 10 A/s. The cycle was followed by an 8 min waiting period before the measurements were started. Rejected or faulty measurements: No rejected or faulty measurements have been observed. Use of the measurements in FiDeL: The harmonics were rescaled from 25 mm to 17 mm reference radius. Only the integral measurements have been used. 4. Transfer function The geometric values for transfer functions are taken at 2500 A, see Fig. 3. In Table IV we give the room temperature and the geometric values at 4.5 K. In both cases the spread is about 5 units. No systematic difference between the apertures is observed. The 4.5 K measurements are about 40 units larger than room temperature measurements. Saturation starts to be significant at about 3600 A (see Fig. 3). At nominal current it has an impact on the transfer function of about 0.85% (85 units). Persistent-current effects are small and amount to 6 units at 400 A. For the DC-magnetization fits of the transfer functions the parameter DCMAG-q was fixed at 2 (this approach is called the “constrained fit” in the documents of J. Miles). Measurements and the FiDeL model [9] are shown in Fig. 2. Table V summarizes the fit parameters used for the static model of the MBRB transfer function, as well as fit errors, which are below one unit. -3- TF (Tm/A) 10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100 0 1000 2000 3000 4000 5000 Current (A) 6000 7000 Fig. 2: Measured transfer function of both apertures of the MBRB family (blue markers), average of measurements (black line), and the FiDeL fit (red line) Table I: MBRB transfer function at room temperature and geometric at 4.5 K Ap. Room temperature Geometric at 4.5 K HCMBRBA001-BL000001 (T m/A) 1 5.96E-03 5.98E-03 HCMBRBA001-BL000002 (T m/A) 1 5.96E-03 5.98E-03 HCMBRBA001-BL000003 (T m/A) 1 5.96E-03 5.98E-03 Average (T m/A) 1 5.96E-03 5.98E-03 (units) 1 4.8 4.6 HCMBRBA001-BL000001 (T m/A) 2 5.96E-03 5.98E-03 HCMBRBA001-BL000002 (T m/A) 2 5.96E-03 5.98E-03 HCMBRBA001-BL000003 (T m/A) 2 5.96E-03 5.98E-03 Average (T m/A) 2 5.96E-03 5.98E-03 (units) 2 3.3 4.6 Spread Spread Table V: Fit parameters for the transfer function of the MBRB magnets DCMAG-m DCMAG-p DCMAG-q DCMAG-h DCMAG-Tc0 DCMAG-Tmeas DCMAG-Ic DCMAG-Iinjref SAT1-s SAT1-I0 SAT1-s SAT1-Inomref Min error Max error RMS error (T m/A) (adim) (adim) (adim) (K) (K) (A) (A) (T m/A) (A) (adim) (A) (units) (units) (units) -4.11E-06 1.27 2 2 9.5 4.5 7000 400 3.36E-04 7850 3.12 6200 -0.9 1.1 0.6 5. Field errors In a dipole coil, allowed harmonics are b2n+1. For b3 and b5 multipoles we create a static FiDeL model using geometric, DC-magnetization, saturation, and residual magnetization components. The same models are used for both apertures. Moreover, DC-magnetization models of the quadrupole component were created for each aperture. -4- Room temperature measurements are summarized in Table VI. Average b3 is about -1 units, and b5 is about 0.2 units. There is a large systematic b2 of 3.5 units, with opposite signs in each aperture. As in all others MBR, there is a systematic a3 of -0.3 units due to the connection cables. The systematic b2, disappears in the geometric at 4.5 K, as in MBRC. In 4.5 K measurements at low currents, the b2 has the opposite sign than in room-temperature measurements. The rest of the data confirm the room temperature measurements. The quadrupole component in room-temperature- and 4.5-K-measurements has been analyzed extensively by BNL. The effect is due to cross-talk at very low as well as very high excitation. The reason for cross-talk at both, low and high, currents lies in the fact that the relative magnetic permeability of the yoke steel is not monotonous. It takes values around 200 at low excitation, then rises steeply to above 2000, and decreases again with saturation. The sign difference between room temperature and 4.5 K measurements is explained by the fact that the 4.5-K-data contains remnant magnetization, whereas it is subtracted from roomtemperature data. Geometric: Geometric values were taken at 3000 A. The values are summarized in Table VII Table VI: Room temperature data of the MBRB field harmonics (in units 10-4 at 17 mm reference radius) ap b2 a2 b3 a3 b4 a4 b5 a5 b6 a6 HCMBRBA001-BL000001 1 -3.92 0.78 -1.16 -0.25 0.00 -0.09 0.09 0.06 -0.01 -0.02 HCMBRBA001-BL000002 1 -2.94 0.35 -1.48 -0.16 0.04 0.23 0.16 0.08 0.01 0.03 HCMBRBA001-BL000003 1 -3.50 -0.86 -2.01 -0.31 -0.05 0.20 0.17 0.06 0.02 0.03 average 1 -3.45 0.09 -1.55 -0.24 0.00 0.11 0.14 0.07 0.01 0.01 spread 1 0.98 1.65 0.85 0.16 0.09 0.32 0.08 0.02 0.03 0.05 HCMBRBA001-BL000001 2 3.43 2.58 -0.74 -0.39 0.14 -0.03 0.18 0.02 -0.01 -0.03 HCMBRBA001-BL000002 2 3.37 -1.48 -1.30 -0.27 0.06 -0.13 0.20 0.04 0.00 -0.02 HCMBRBA001-BL000003 2 2.60 0.54 -0.77 -0.26 0.06 0.27 0.18 0.08 0.01 0.02 average 2 3.13 0.55 -0.94 -0.31 0.08 0.04 0.19 0.05 0.00 -0.01 spread 2 0.83 4.07 0.56 0.13 0.08 0.40 0.02 0.06 0.01 0.04 Table VII: Geometric values of the MBRB field harmonics (in units 10-4 at 17 mm reference radius) ap b2 a2 b3 a3 b4 a4 b5 a5 b6 a6 HCMBRBA001-BL000001 1 -0.29 0.73 -0.80 -0.24 0.05 -0.08 0.08 0.07 0.00 -0.01 HCMBRBA001-BL000002 1 0.11 0.27 -1.22 -0.18 0.04 0.17 0.12 0.07 0.00 0.03 HCMBRBA001-BL000003 1 -0.46 -0.82 -1.66 -0.35 -0.04 0.20 0.12 0.05 0.02 0.04 average 1 -0.21 0.06 -1.23 -0.25 0.02 0.10 0.11 0.07 0.00 0.02 spread 1 0.56 1.56 0.86 0.17 0.09 0.27 0.04 0.02 0.02 0.05 HCMBRBA001-BL000001 2 0.06 2.51 -0.45 -0.32 0.06 -0.02 0.16 0.04 0.00 -0.02 HCMBRBA001-BL000002 2 0.19 -1.31 -1.01 -0.21 0.02 -0.14 0.17 0.07 0.00 -0.02 HCMBRBA001-BL000003 2 -0.08 0.48 -0.47 -0.30 0.06 0.29 0.16 0.07 0.00 0.01 average 2 0.06 0.56 -0.64 -0.28 0.05 0.04 0.16 0.06 0.00 -0.01 spread 2 0.27 3.81 0.56 0.11 0.04 0.43 0.01 0.03 0.01 0.03 DC Magnetization and Saturation: The average of all 4.5 K measurements, corrected by the geometric values, is used to model DC-magnetization and saturation for b3, b5 and b2. Note that, in order to obtain high-quality fits in the low-field range, the constraint on the DCMAGq parameter in the DC-magnetization fit, which was used in the transfer function model, was dropped. Using Gnuplot for optimal fitting yields DCMAG-q values of about 12 for b3 and b5, and 6 for b2. A proper fit of the b5 data at injection current requires taking into account the penetration phase of the DC-magnetization. For this purpose, the residual magnetization component in the FiDeL model is used. The b2 has a large deviation at low fields, which is modeled through DC-magnetization. It has opposite sign in the two apertures. Two DCmagnetization models were therefore created, one per aperture. The fit agrees with -5- measurements within 0.2 units for b2 and b3, and within 0.01 units for b5. Fit parameters as well as fit errors are summarized in Table VIII. The comparison of fit and measurements is shown in Figs. 4 – Fig. 6. Table VIII: Fit parameters for the MBRB magnets b3 -geometric (units) Aperture DCMAG-m DCMAG-p DCMAG-q DCMAG-h DCMAG-Tc0 DCMAG-Tmeas DCMAG-Ic DCMAG-Iinjref SAT1-s SAT1-I0 SAT1-s SAT1-Inomref SAT2-s SAT2-I0 SAT2-s SAT2-Inomref RESMAG-r RESMAG-I_inj_ref Min error Max error RMS error (units) (adim) (adim) (adim) (K) (K) (A) (A) (units) (A) (adim) (A) (units) (A) (adim) (A) (adim) (A) (units) (units) (units) b3 1&2 -4.43 1.77 13.26 2 9.5 4.5 7000 400 -0.66 4966 4.43 6200 -0.146 0.180 0.066 b5 1&2 -0.25 2.28 12.09 2 9.5 4.5 7000 400 0.17 6177 4.62 6200 0.19 2.04 -0.006 0.006 0.003 b2 1 2.71 0.96 6.22 2 9.5 4.5 7000 400 -0.143 0.018 0.042 b2 2 -2.81 0.98 6.31 2 9.5 4.5 7000 400 -0.041 0.156 0.046 2 1 0 -1 -2 -3 -4 -5 -6 -7 -8 -9 0 1000 2000 3000 4000 Current (A) 5000 6000 7000 Fig. 3: Measured sextupole of both apertures of the MBRB family (markers), average of the measured magnets and apertures (black line), and the FiDeL fit (red line) -6- 0.6 0.5 b5 - geometric (units) 0.4 0.3 0.2 0.1 0.0 -0.1 -0.2 0 1000 2000 3000 4000 Current (A) 5000 6000 7000 Fig. 4: Measured decapole of the MBRB family (markers), average of the measured magnets and apertures (black line), and the FiDeL fit (red line) 8 6 b2- geometric (units) Aperture 1 4 2 0 -2 -4 Aperture 2 -6 -8 0 1000 2000 3000 4000 5000 Current (A) 6000 7000 Fig. 5: Measured quadrupole of the MBRB family (markers), average of the measured magnets per aperture (black lines), and the FiDeL fits per aperture (red lines) 6. Summary and critical points All magnets have been measured at 4.5 K and at room temperature. The transfer function spread among the three magnets (six apertures) is about 5 units. The transfer function has a saturation of 70 units at 7 TeV operational current. Field harmonics are optimized in the range 3.5-7 TeV operational current, where normal sextupole is around -1 units, and normal quadrupole is less than 0.3 units. The sextupole component of the field at injection current is around -4 units. At low currents, a quadrupole component is present in the field, which is of opposite sign in the two apertures, and reaches 3-4 units at injection. -7- Ackowledgements We wish to thank L. Bottura for providing additional data, and L. Deniau for his advice in generating the FiDeL models. References [1] O. Bruning, et al., CERN Report 2004-003 (2004). [2] E. Willen. Functional specification: Superconducting beam separation dipoles. LHC Pro ject Document LHC-MBR-ES-0001 rev. 2.0, CERN, June 2000. [3] J. Muratore and et al. Test results for LHC insertion region dipole magnets. Proceedings of the 2005 Particle Accelerator Conference, Knoxville, Tennessee, pages 3106–3108, 2005. [4] E. Willen and et al. Superconducting dipole magnets for the LHC insertion regions. Proceedings of EPAC 2000, Vienna, Austria, pages 2187–2189, 2000. [5] R. Ostojic. Engineering change order - Class I: LHC layout version 6.4: Layout of IR4, Alice compensator, D2 and D4 cold masses. LHC Project Document LHC-LS64-EC-0001, CERN, September 2001. [6] R. Ostojic. Engineering change order - Class I: Modification of the cold bore separation in D3 and D4 dipoles. LHC Project Document LHC-LBR-EC-0002, CERN, February 2003. [7] P. Hagen, private communication, 2009. [8] J. Muratore and et al. Test results for initial production of LHC insertion region dipole magnets. Proceedings of EPAC 2002, La Villette, Paris, pages 2415–2417, 2002. [9] Model specifications (EDMS 908232) -8-