Vacuum measurement

advertisement
3. Vacuum measurements
Measurement procedures in vacuum technique:
1) Pressure measurement.
2) Pumping speed, desorption speed, etc..
3) Leek testing.
4) Mass spectroscopy.
N.B. 2) is performed by means of pressure measurements (1), at two ends of a tube with a known
conductivity.
I=G  p1 -p2  =S1p1 -S2p2
3.1. Pressure measurements
There is no universal gauge that can measure from atmosphere to UHV pressure a (dynamic range of
1015). The instrument chosen depends on the pressure range and the residual gases in the vacuum.
Mechanical gauges in which solid or liquid diaphragm is moved by the gas molecules hitting it, give
absolute pressure measurement unaffected by gas properties Unfortunately they are ineffective below
10-5 Torr.
Gauges which measures some bulk gas property, such as heat conduction or viscosity, are dependent on
gas composition and are effective over limited pressure range below approximately 100 Tr and above
10-4 Tr. Gauges for high vacuum and ultra high vacuum are generally based on charge collection, that
is the residual gas molecules are ionised, and the resulting current measured. Although such gauges
will ionise vapours as well as permanent gases, their response depends on parameters other than
ionisation potential, making accurate total pressure measurement difficult in a gas mixture.
3.1.1 Barometers
1) Toricelli vacuum meter
2) Closed barometer -for the lower pressures.
3) “Shortened” barometer.
4) Mechanical membrane gauge.
5) Compressing absolute barometer (McLeod) Fig: 7a
insert figure:
Operation principles:
Bring the Hg level in the compression capillary 1 to y-y. The Hg level in the reference capillary 2,
identical to 1, will be higher, at the line z-z. Equilibrium condition: p2=H+p1 in [Tr] units, where H is
the difference between the Hg levels in 1 and 2. The compression ratio, k=V 1/V2, is set by the ration of
the volume of the container 3 (V1 , known and constant for a given gauge) to V2=(/4)d2h.. In the
equilibrium conditions p1V1=p2V2, i.e.,
V2
πd 2 h
p1 = (H+p1 )=
(H+p1 )=ch(H+p1 )
V1
4V1
p1 =
where
c=
chH
1-ch
πd 2
is the gauge constant. Typically ch
4V1
1 , and the measured pressure p1  chH
In the linear method (case a, see figure) we keep the level difference constant h=h0. Then:
p1 =
ch 0
H  c1H
1-ch 0
In the quadratic method (case b) we compress to have H=h, and
p1 =
cH
H  c2 H 2
1-cH
We want the compression ration k to be as high as possible, therefore we want d to be as small as
possible, but there is an obvious limit to reducing d. The best realistic k  104-105, allowing for
measuring down to p  10-6 Torr.
N.B. This is an absolute pressure gauge - important for calibration of relative vacuum meters.
The gauge can operate either with mercury or with vacuum oil. There is a problem of contamination by
the measuring agent and false effect due to the pumping by liquid. N 2 traps located at the system input.
Reduced rotary McLeod – see Edwards catalogue
3.1.2. Thermomolecular
Principle of operation Fig 7b
Insert figure
S - static plates are heated to T’, R - rotary plates at T
In an ideal vacuum the plates would remain parallel. At the presence of the gas in-between the plates,
the difference in velocity of molecules scattered from the hot plates and from cold gauge walls
generates a torque momentum.
From the side of the static plates the force /unite surface of a R plate is
Fs =2m0 v`f1`=2c3n1m0 v`v
From the opposite side, the R plates feel the force:
Fb =2m0 vf1 `=2c3n1m0 v 2
The net force:
 v`   T` 
ΔF=2c3n1m 0 v 2  -1  =p 
-1 
v   T 
p=cΔF=c`α
The constant depends on the type of the gas through the accommodation coefficient (depending also on
the state of the surface). The coefficient  accounts for thermal energy transfer - non-perfect scattering.
A particle scattered from the surface at temperature T does not fully accommodates the speed
corresponding to that temperature.
 For heavy molecules   1.
 For light molecules   0.3 -- 0.4 (dirty surface)
 For light molecules   0.1 (for H2) -- 0.03 (for He) (clean surface)
Measuring range is limited by :
On the low pressure the limiting factor is electromagnetic radiation. p lim10-6 Torr
On the high pressure the limiting factor is the viscosity of gas. p lim 10-1 Torr.
3.1.3. Friction gauge
Fig. 7c.
Gas molecules execute torque momentum on the measuring disk. The torque transfer efficiency
depends on the gas pressure. The pressure is measured by examining the torque applied on the upper
disk. In some gauges the discs are magnetically suspended.
Pressure range: 10-2 - 10-6 Torr
3.1.4. Heat conductivity meters (Pirani)
Heat conductivity of a gas depends on its pressure (through the number of molecules) The basic idea is
shown in Fig 7d:
Current carrying Pt or W wire (D) changes its resistance with changing temperature:
R T  R 0 1  T
1
1
β Pt =3×10-3   , β W =4.5×10-3  
K
K
If the current through the wire is kept constant then the voltage drop VI  I 0 R T will be a measure of
the pressure. That is a constant current gauge. The opposite case: the voltage is maintained constant by
adjusting the bias current (which is a measure of the pressure). That is a constant resistor gauge.
Power delivered to the wire P  I 0 R T  I 0 VT is dissipated by:
1) gas
2) radiation
3) through the wire holders
For low-pressure 2) and 3) start to dominate and set the measurements limit.
2
The relation V=f(p) is obtained from the calibration curve.
In the laboratory practice the calibration is obtained by setting: V to V min (beginning of the csale) at
p=760 Tr and V to Vmax ,(full scale) at p=0. p=0 in practice means p<10 -5 Tr – e.g., an operating
diffusion pump.
Zoology of heat conductivity meters:
1) constant current
2) constant resistance
3) thermistor
4) thermocouple
5) bimetal (mechanical read-out)
3.1.4.1. Thermocouple meter
insert figure
The pressure range between 10 and 10-3 Torr is indicated by measuring the voltage of thermocouple
spot welded to a filament exposed to system gas. The filament is fed from a constant current supply,
though its temperature depends on thermal losses to the gas. At higher pressure, more molecules hits
the filament and remove more heat energy’ causing the thermocouple voltage to change. These gauges
are used extensively in fore-line monitoring and to provide the signal to automatically switch the main
chamber from backing and high-vacuum pumps at the crossover pressure.
3.1.4.1. Pirani
In a Pirani gauge two filaments, often platinum, are used as resistance in to arms of a Wheatstone
bridge. The reference filament is immersed in a fixed gas pressure, while the measurement filament is
exposed to the system gas. The current through the bridge heats both filaments. As in T/C gauge, gas
molecules conduct heat away from the immersed filament and unbalance the bridge. Pirani gauges have
roughly the same pressure measurement range as T/C gauge and are used in identical application, but
generally provide faster response.
3.2. Ion vacuum-meters
For pressures lower then 10-3 Tr all measurement are based on ionising gas molecules and measuring
the ion current. Ionisation is achieved by
 radiation ( particle)
 accelerated electrons
 cold glow discharge
N.B. Sensitivity depends on the type of gas through deferent ionisation efficiency  No. of ions created
by 1 electron along the path of 1 cm at the pressure of 1 Tr.
If the gauge sensitivity for air is 1 then:
He
H2
N
Air
5
3
2.5
1
N2
0.9
O2
0.9
CO
0.9
A
0.8
CO2
0.7
Hg
0.3
The gauge measures the ion current collected by the collector circuit. Measurements with ion gauges
bear an error due to a spurious effect of pumping by ion adsorption in the gauge body.
3.2.1. Ionization gauge
Bayard – Alpert head
Insert figure
Covers the range between 10-4 to 10-9 Torr, with gauge sensitivity from 5 to 20 Toor -1. B-A gauges are
available with one or two filaments (the second acting as a spare) and with two filaments
material: thoria-coated iridium, used in oxygen reach application and for ‘burn-out’
resistance if the gauge is dumped to atmospheric, and tungsten, used for lower cost and in
residual gases containing halogen.
B-A heads are available in naked and dressed version, see Fig 7f,g
VA  100 - 150 Volts
Vc  -25 - -75 Volts
Ie  10 A - 10 mA
Gage constant k (Tr-1)
k=
I ion
Iep
,
p=
1
I ion =sI ion
kI e
typically k 10 - 30 Tr-1 , therefore for for Ie  1 - 5 mA s  10 Tr/A.
N.B. The ion current in a B-A head is about an order of magnitude smaller then the pressure in Tr.
The B-A gauge’s lower pressure limit (10-9 Torr range) is caused by emission of the soft X-ray
radiation generated by primary electrons hitting the anode. X-ray photons reaching the ion collector
electrode release photoelectrons. The electron current due to photoemission is indistinguishable from
the ion current of the positive ions collected by the collector electrode. Below 10 -9 Torr the
photoemission current becomes a large enough fraction of the ion current to distort the pressure reading
(for vacuum 10-12 Tr. we have to measure current of 10-13A!). Below 10-9 Tr the collector current is:

σ 
I c =kI e p+σIe =kIe p  1+ 
 kp 
where I e is the X-ray contribution. In order to have the measurement error < 10%:
σ
 0.1
kp
Therefore, the low pressure limit for the B-A head:
p min  10
σ
, ~ 10-9Tr range.
k
Readhead gauge
B-A head with extended measuring range: Redhead gauge equipped with an additional thin wire
modulator electrode located within the anode, see Fig 7j.
The procedure consists of taking take two measurements with different potential V M of the modulator:
1) Set VM =VA and measure Ic’
2) Set VM =Vc and measure Ic’’
Assuming that the X-ray contribution to both measurement is the same
I`c =k`Ie p+σIe and I``c =k``Ie p+σIe
Obviously, k’>k’’ and
p=
I`c -I``c
 k`-k`` Ic
Redhead low pressure limit falls into 10-11 - 10-12 Tr range.
Other low pressure extended range gauges
Groszkowski - screened collector protected by a tube against X-rays see Fig 7h.
Helmer-Haynord - applies magnetic field and bends the ion beam towards the hidden collector
Lafferty - magnetic field assisted ionisation + bending.
Cold cathode glow discharge head (Penning)
Low pressure limit -- problems with ignition and glow maintenance. Same solutions use spark plug to
start. At the high pressure side the limiting effect is the arc discharge. Range - 5*10-3 - 5*10-7 Tr.
3.3. Mass spectrometry
Partial pressure measurement - spectroscopy of the gas composition
Separation of masses can be performed by :
1) magnetic field bending of a particle trajectory.
2) Cyclotron resonance .
3) Time of flight.
3.3.1. Magnetic field
Emitter
Collector
R
k
H
M 0U
Solutions:
H = constant
R = constant
M0U = constant
M0U = constant
3.3.1.1. 180 mass spectrometer
Insert graph:
Typical spectra for air at 10-9 Tr: Spectra of DC 704 vacuum oil vapour.
Insert graph
H2
H2O
CH4
N2+CO
O2
Ar
2
18
16
28
32
44
R
M0
R0

M 0 s1  s2
Resolution:
S1 and S2 are the apertures of the emitter and collector boxes. Observe that we cannot reduce s 1 and s2
below certain limits because the overall sensitivity will decrease.
Typically:
M0
 102
M 0
3.3.2. Omegatron
The principle is based on a cyclotron resonance of charged gas molecules
Insert picture
Electric field between electrodes P1-P2 =Ecos(t), electrons- flow from cathode K to anode A along
(x-x’ path.), Gas ions are created along x-x, and accelerated by the electric field, in the presence of
magnetic field. The cyclotron resonance frequency:
  c 
eH
m0
At the resonance condition ions move along opening spiral tracks and each the collector. The resolution
m0
M0
eH 2 R 0


m0 M 0 2 Em0
The Omegatron is a simple and a small device, e.g., 20x20x20 mm. Typical parameters:
Ie ~ A, H ~ 3000 Oe, E ~ 10 V/cm, w for 2<M 0 < 100 2.3 MHz to 43 kHz,
M0
~ 10-20
M 0
3.3.3.Time of flight mass spectrometer (without magnetic field)
kA - ionisation by dc electron current .
ab - acceleration of the ioos by dc voltage
electrodes b-d and d-e biased with ac electric field in opposite phases
C
e
d
b
a
This type of mass spectrometer has low resolution.
A
k
3.3.4. Quadruple mass spectrometer.
Build around four rods through which the ionised gas particles are flowing. Between each pairs of rods
there is a voltage difference V0+Vcos(t).
Analysis is performed at  = const and at changing V/V0 ratio.
Collector is reacted only by ions with a mass: M0~V/2
.3.5. Direct flight pulse spectrometer
Square shape accelerating pulse; all particles acquire the same energy but due to difference in mass
they have different time of arrival to the collector.
Measurement of collector signal in time. Big complicated but good resolution:
M0
>100
M 0
3.4. Leak detection.
Leak detector = mass spectrometer tuned permanently to He + a pumping unit. (diffusion pump or TM
based). The pumping unit has precise tuning of the pumping speed (throttle valve) in order to provide
optimum measuring condition for various leaks.
In-side vs. outside operation (sneezing) .
Download