TAMPEREEN TEKNILLINEN YLIOPISTO TAMPERE UNIVERSITY OF TECHNOLOGY TLT-5500, TLT-5506 83950 Telecommunication Laboratory Course Network analyzer work E-mail: ari.asp@tut.fi Room: TG 212 Lab:TG205 INTRODUCTION You can find the information you need for the pre-lab part in the following book Hewlett & Packard: "Back to Basics" (later referred as B to B) Also Lehto/Räisänen: "Mikroaaltomittaustekniikka" is quite useful. NOTE! The sources mentioned above can be found in the shelf of TLT-LAB (TG206). 1. PRE-LAB ASSIGNMENT 1.1 Draw a block diagram of a network analyzer and explain operating principle. What can you measure with a network analyzer? 1.2 What are the different sources of measurement errors and calibration (=measurement calibration or work calibration) methods? (B to B s.64). Explain the calibration principle, i.e. how calibration corrects measurement errors (B to B s.71). 1.3 What are meant by the following terms? Frequency response, 3dB bandwidth, quality factor, insertion loss/gain, phase response, passing time delay, rejection, the 1-dB compression point of an amplifier and amplification. 2. MEASUREMENTS Measurements are carried out by HP network analyzer 8714C. Measurement range is from 300 kHz to 3 GHz. You should familiarize yourself with the theory of network analysis, network analyzer and other equipment used in the measurements before starting the lab. For calibration we use the kit HP 85032B. In this lab we measure characteristics of different circuit elements, manufactured by MINI CIRCUITS or HP. Amplifier Mini Circuit ZFL-1000 Attenuator Mini Circuit NAT-20 Terminator Mini Circuit NTERM-50 Filter K & L 71 MHz/0 Network analyzer basic use After turning on, you can immediately set the equipment to its basic operation mode by pressing PRESET –button. Now we have only one of the two channels in use. (Transmission measurements, 300 kHz -3 GHz, source power 0 dBm, etc). Hence we can use the equipment to measure transmittance (amplification/attenuation). It has to be said, however, that the measurements are inaccurate, if we don’t perform the measurement calibration. You can easily find the default settings for different measurements by pressing the BEGIN –button. You can find the settings for amplifiers, filters, broadband passive components and mixers from the menu. By pressing AMPLIFIER ->TRANSMISSN the range of scaling and also the bandwidth of the measurement system change. As we have two channels in our equipment, we can measure two parameters at the same time. We can define the measurements by pressing MEAS 1 and MEAS 2 buttons. So we measure transmission, reflection, power, etc. We can also find different options for different types of demodulators: broad- and narrowband. From the frequency menu FREQ, we can change the frequency band of measurement by entering the START and STOP frequencies. We can also set the CENTER (center frequency) and SPAN. By pressing CW button, we get into non sweeping measurement mode, in other words SPAN = 0 Hz and we can change the measurement frequency by keying in or by using the rotating disc of the equipment. By pressing the SWEEP button you find the options to chenge sweep parameters. By pressing the POWER button you find the options by which you can adjust the level of the signal source LEVEL, turn on/off the power source RF ON/off and also set the value for the source power when pressing the PRESET button Pwr Level at Preset. By pressing the SCALE button, you find some common setting possibilities, which you should already know from the spectrum analyzer, and a couple of new ones, which may be needed under some special circumstances. You shouldn’t have any difficulty to figure out what MARKER button does. By pressing the DISPLAY button, you get a menu with different options. In this work, however, we don’t use any of them, except for NORMALIZE and More Display > Split Disp FULL split, by which you can split the display into two “sub-displays” (for each channel). After pressing FORMAT you can choose the format in which you would like to see the measurement results. For example, after pressing FORMAT in REFLECTIONmeasurements you can see the reflection coefficient in logarithmic scale by choosing Log Mag option. This is naturally equivalent to return loss, while choosing Lin Mag, would give you SWR. By pressing Smith Chart you have RF-man’s dream on the display. Polar display, though used less frequently, is available too. CAL menu is a very important one, especially at the beginning of the measurements, and also when the reference level of calibration (reference line) changes, (mainly as a result of change in electrical length). We will shortly get back to this. By pressing the AWG button, you will find the mean of the measurements, so the equipment takes the average of the measurement results. This means that the equipment takes sweeps by the value determined by the mean coefficient, and then calculates the average. This would reduce the noise on the display. Included in AWG menu is also System Bandwidth setting, by which you can make the dynamics of the analyzer better, because the noise level reduces to a narrower bandwidth. Sweeping time however increases. Calibration At the beginning of measurements you must always perform measurement calibration, which is usually done using a calibration kit. There are different kits for different connector types. Look at (for instance) Räisänen/Lehto: Mikroaaltomittaustekniikka page.107. By performing the standard measurements (measurement of the known loads): “open”, “short”, “load” and “through” the analyzer can find the error parameters in the error model and compensate for different error sources. By pressing CAL button depending on what we are measuring - the equipment suggests different calibration methods. E.G: In transmission measurements, Transmission Cal gives different options for calibration: Restore Defaults, Response, Response & Isolation and Enhanced response. In each of these, there is a different error model, and so there are different numbers of standards to be measured. In reflection measurement, the equipment proposes: Restore Defaults or One Port. By pressing One Port the equipment asks you to connect different standards to be measured, and then calculates correction coefficients at the end of the measurements. Perform the measurements always at the calibration reference level; in other words don’t connect any adapter between the calibration reference level and the equipment to be measured. If you doubt the measurement results, perform CAL CHECK. AT THIS STAGE, YOU SHOULD ALREADY KNOW THAT THERE ARE TWO CHANNELS IN THE INSTRUMENT, AND THE REFLECTION MEASUREMENTS CAN ONLY BE CARRIED OUT FROM THE REFLECTION PORT. ALSO TRANSMISSION CAN ONLY BE MEASURED FROM THE REFLECTION PORT TO THE TRANSMISSION PORT. S-parameters can not be measured directly. Especially we can’t measure the four sparameters simultaneously, though it is possible with more expensive equipment. Transmission- and reflection measurements Set MEAS 2 = off and MEAS 1 = reflection. Press CAL and One Port. Connect a piece of cable (almost 30 cm long) to the reflection port. Perform the calibration. 2.1 Let’s examine the effect of the cable length on the accuracy of the calibration/measurements. Connect Mini-Circuit 50 terminator to the end of the cable and check the return loss: => FORMAT = Log Mag. You can also try the Smith Chart format. Is the value of the terminator 50 in the 300kHz - 3GHz range? Take the cable off and connect the terminator directly to the reflection port. How did the measurement result change? 2.2 Measure the reflection and transmission of Mini-Circuit 20 dB attenuator. Press the BEGIN button and try to see if there is any difference between Amplifier, Filter and Broadband Passive options. Choose Broadband Passive. Measure Reflection and Transmission, what can you say about the reliability of the results; especially about Transmission measurement? 2.3 Perform the Transmission Cal Enhanced Response and measure once more. Compare the results. 2.4 Measure the 3dB bandwidth, amplification and 1 dB compression point of the Mini Circuit ZFL-1000 amplifier. 2.5 Measure the frequency response, 3 dB bandwidth, quality factor, insertion loss, group delay, phase response, rejection, etc of Murata GSM or K & L 71MHz band pass filter. Note that you can measure the parameters above in different manners, for example using a -marker or directly. 3. POST-LAB ASSIGNMENTS 3.1 Write down the answers to the pre-lab assignments neatly and return it along with the post-lab assignments. (Hopefully now, after performing the measurements, you have a deeper understanding of the concepts). 3.2 Explain, what is the significance of calibration in network analyzer measurements? (Consider your observations during the measurements) 3.3 Answer to the questions presented in the second part of this lab assignment and add all the necessary figures. Compare the results obtained with those declared by the manufacturer. REFERENCES 1. 2. 3. 4. 5. 6. A. Lehto, A. Räisänen: MIKROAALTOMITTAUSTEKNIIKKA Pekka Eskelinen: NYKYAIKAISEN RADIOTEKNIIKAN MITTAUKSET KÄYTÄNNÖSSÄ A. Räisänen, A. Lehto: Radiotekniikka Hewlett & Packard: BACK TO BASICS Fluke: The ABC´s of OSCILLOSCOPES Mini-Circuits: RF / IF DESIGNER´S HANDBOOK