Instrument QC and Qualification Overview Why QC is Important LSRII Optical Configuration LSRII QC Validation Optimization Calibration Standardization Practice Analysis Characterizing the Cytometer 3 Challenges… • Instrument - optical configuration, optimization, standardization, and calibration • Reagent - optimization and standardization • Sample processing • Staining protocols • Data Analysis - compensation & gating • Operators • Volume of data (death-by-excel!) Duke University Medical Center Two Methods of Instrument Charaterization BD CS&T: Cytometer Setup and Tracking Instrument Performance • Parameters CS&T Perfetto • • • • • Laser (amps vs mW) CV (resolution) Signal-to-noise ratio Linearity Specific fluorescence yes yes (target) NO Yes (range) NO yes yes (range) yes (range) yes (range) yes (target) • Automated YES NO Instrument QC Frequency Purpose Initial Characterization & After Optical Service Validation: optimal voltage range for each detector Once Per Assay Optimization: assay specific optimal voltage for each detector & assay specific target channels Each Experiment & Troubleshooting Calibration: set detectors to assay specific target channel values for specimen acquisition Standardization: Monitor trendlines using assay specific target channels over time Key Performance Factors in High-Quality Flow Cytometry Data • Resolution of subpopulations, including dim subpopulations • Sensitivity • Relative measured values of fluorescence • Linearity and accuracy • Reproducibility of results and cytometer performance • Tracking • Comparison of results across time and amongst laboratories • Standardization LSRII qualification Validation Linearity Resolution (CV) Signal-to-Noise Ratio (S:N) - LLOD & LLOQ Optimization Select Optimal Voltages for Inst Performance with AssaySpecific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F) Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trendlines Linearity Definition: Proportionality of output to input Method: Access linearity using the ratio of two pulses over voltage range of the PMT Significance: Linearity is important for compensation (Median Pk6 - Median Pk5) = Constant Median Pk5 Linearity • Defined as proportionality of output (MFI) to input (Fluorescence/ # of photons) 2X Abs 2000 MFI = X Abs 1000 MFI Ab = X Ab = 2X 0 1000 2000 3000 • Important for fluorescence compensation • Compensation of data in the last decade involves subtraction of large numbers • Small errors (non-linearity) in one or both large numbers can cause a large absolute error in the result Linearity: Effect on Compensation • Compensation of data in the last decade involves subtraction of large numbers • Errors (non-linearity) in one or both large numbers can cause a large absolute error in the result A Detector FITC PE B C D Median Fluorescence Intensity (MFI) 68 80 1796 75 5921 79 BD CompBeads stained with varying levels of FITC-Ab. Compensation was set using samples A and C. This cytometer had a 2% deviation from linearity above 50,000 units. 73,000 365 LSRII Qualification Validation Optimization Linearity Resolution (CV) Signal-to-Noise Ratio (S:N) - LLOD & LLOQ Select Optimal Voltages for Inst Performance with AssaySpecific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F) Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines cvs 15 Instrument Sensitivity: Two Definitions • Defining sensitivity 1. Threshold: Degree to which a flow cytometer can distinguish particles dimly stained from a particle-free background. Usually used to distinguish populations on the basis of Molecules of Soluble Equivalent Fluorochrome (MESF). 2. Resolution: Degree to which a flow cytometer can distinguish unstained from dimly stained populations in a mixture. • How to measure instrument-dependent sensitivity? • Resolution sensitivity is a function of three independent instrument factors: • • • Br Qr Electronic noise (SDen) Background contributions 17 Why is Br important? • High Br widens negative and dim populations. • High Br value = lower resolution • Low Br value = higher resolution Low Br High Br Br: Optical Background from Propidium Iodide • Example: It is common to use propidium iodide (PI) to distinguish live from dead cells. Propidium iodide was added in increasing amounts to the buffer containing beads, and Qr and Br were estimated. PerCP Br 9000 8000 7000 6000 5000 4000 3000 2000 1000 0 Br Qr 0.05 0.04 0.03 0.02 Qr 0.01 0 0 1 PI-free 2 3dye 4(µg) 5 6 • Residual PI in your sample tube will increase Br, which will reduce sensitivity. 20 Relative Q: Qr • Qr is photoelectrons per fluorescence unit and indicates how bright a reagent will appear on the sample when measured in a specific detector. Qr = # photoelectrons # fluorescence molecules PMT 1 Qr = 2 photoelectrons = 0.25 8 fluorescence molecules PMT 2 Qr = 1 photoelectron 8 fluorescence molecules = 0.125 Why is Qr important? A system with a higher Qr has a better resolution than a system with a lower Qr. Low Qr value = lower resolution High Qr value = higher resolution High Qr Low Qr What Factors Affect Qr? •Laser power •Optical efficiency •PMT sensitivity (red spectrum) •Poor PMT performance •Dirty flow cell •Dirty or degraded filter Spillover Decreases Resolution Sensitivity Spread from APC Cy-7 background Population resolution for a given fluorescence parameter is decreased by increased spread due to spillover from other fluorochromes. Summary: Instrument Performance and Sensitivity • Qr and Br are independent variables, but both affect sensitivity. • Increases in Br or decreases in Qr can reduce sensitivity and the ability to resolve dim populations. • Sensitivityrelative Qr Br • On digital instruments, BD FACSDiva™ software v6 and CS&T provides the capability to track performance data for all of these metrics, also allowing users to compare performance between instruments. Instrument performance can have a significant impact on the performance of an assay. • Resolution vs. Background Negative Population Positive Population Negative population has low background Populations well resolved Negative population has high background Populations not resolved Negative population has low background high CV (spread) Populations not resolved The ability to resolve populations is a function of both background and spread of the negative population. Measuring Sensitivity: The Stain Index • The Stain Index is a measure of reagent performance on a specific cytometer, a normalized signal over background metric. median positive median negative Brightness Stain Index Width of Negative 2 rSD negative Brightness Width of negative • • The brightness is a function of the assay (antigen density, fluorochrome used). The width of the negative is a function of: – Instrument performance (Qr, Br, and SDen) [single-color] – The assay (Fluorescence spillover / compensation) [multicolor] The cell population Stain Index: Normalized Signal/Background Stain Index (SI) D/W = Relative Brightness 70 60 Goal: Normalize the signal to the spread of background where background may be autofluorescence, unstained cells, or compensated cells from another dye dimension. D 50 40 Wunstained 30 20 Wcompensated dye 10 0 5.5 6.5 7.5 Stain Index (SI) D 1Stain 8.5 W 9.5 10.5 11.5 D = difference between positive and negative peak medians W = the spread of the background peak (= 2X rSDnegative) Index: metric used by Dave Parks, Stanford – Presented at ISAC 2004 Electronic Noise (SDEN): Determining Baseline PMT Voltages The BD FACSDiva™ 6.0 CS&T module analyzes dim particles, which are similar to dim cells’ brightness, allowing relevant detector baselines to be visualized by plotting fluorescence intensity vs (PMT gain, CV, and SD) PE: Detailed Performance Plot Dim Bead • For this detector, the SDEN = 18 • Fluorescence intensity of dim bead = 10 x SDEN = 180 CV Standard Deviation 10000 1000 CV or SD • Determine PMT voltage required to set the dim beads at 180 = 500 volts = baseline voltage • As PMT voltage is lowered, CV increases resolution decreases 500 V 1000 100 18 • As PMT voltage is increased CV unchanged resolution unchanged 10 1 10 100 10000 Median Fluorescence Intensity 180 29 1000 100 100000 PMT Voltage PMT Voltage PMT Voltages: Optimal Gains Can Reduce Classification Errors % Negative in CD4+ Monocyte Gate 12.0% CD4 dim monocytes 650 V CD4 negative CD4+ lymphocytes % Negative in CD4 Gate 550 V 10.0% 8.0% 6.0% 4.0% 2.0% 750 V 0.0% -100 0 100 PMT Voltage Offset 30 31 Methods Used for Validation Which Beads to Use What Values to Plot Selection Criteria Example LSRII Optical Configuration Blue Laser Duke LSRII (488nm) Red Laser Duke LSRII (635nm) FITC Alexa 680 515/20 710/50 SSC APC PE ミ Blue APC Blue Cy7 Green Laser Duke LSRII Violet Laser (532nm) Duke LSR II (407nm) QDot PE 655 Cy5.5 QDot PE TR 710/50 585 660/40 Am Cyan 535LP QDot PE- QDot Green 565 545 Cas Blue PE 33 Cy5 PE QDot Cy7 605 QDot 705 Duke University Medical Center Beads Used for LSRII Validation • 8 Peak Rainbow: 300-900v in 50v increments (all PMTs) • Non-fluorescent to Very Bright; Broad Spectrum Ex & Em • 8pks • Run: Once, then after optical service - “High” flow rate & LOG scale • Uses: Validation • Check Linearity: (MedianPk6-MedianPk5)/MedianPk5 • Check CV: CV Pk5 • Check S:N (LLOD & LLOQ): MedianPk5/MedianBlank • Unstained Comp Beads: 300-900v in 50v increments (all PMTs) • Non-fluorescent • 1pk • Run: Once, then after optical service- “High” flow rate & LOG scale • Uses: Validation • Check S:N (LLOD & LLOQ): MedianPk5/MedianBlank Example of Optimal PMT Performance 3000 3000 3000 3000 3000 2000 2000 1000 2000 1000 0 10 2 3 10 Green B-A 10 4 10 2000 1000 0 0 5 300 Volts 10 2 3 10 Green B-A 10 4 10 2000 1000 0 0 S S C-A 4000 S S C-A 4000 S S C-A 4000 S S C-A 4000 S S C-A 4000 5 0 0 350 Volts 1000 10 2 3 10 Green B-A 10 4 10 5 0 0 400 Volts 10 2 3 10 Green B-A 10 4 10 5 0 450 Volts 3000 3000 3000 3000 1000 2000 1000 0 0 0 10 2 3 10 Green B-A 10 4 550 Volts 10 5 2000 1000 10 2 3 10 Green B-A 10 4 600 Volts 10 5 10 2 3 10 Green B-A 10 4 650 Volts 10 5 4 10 5 1000 0 0 10 2000 1000 0 0 3 10 Green B-A S S C-A 3000 S S C-A 4000 S S C-A 4000 S S C-A 4000 S S C-A 4000 2000 2 500 Volts 4000 2000 10 0 0 10 2 3 10 Green B-A 10 4 700 Volts 10 5 0 10 2 3 10 Green B-A 10 4 750 Volts 10 5 Criteria for the Selection of Voltage Ranges Lowest CV possible Lowest Voltage possible Highest MFI possible Lowest background possible Example of Optimal PMT Performance & Voltage Selection Criteria (Green B) GreenB Voltage vs. CV, LIN and Ratio 70 500 450 60 400 50 350 300 40 250 30 200 150 20 100 10 50 0 0 300 350 400 450 500 550 600 650 700 Voltage Ratio = M1/B Linearity = (M2-M1)/M1 CV Lin Ratio Optimal Voltages: 450-650 750 R a tio C V o r L IN Selection Criteria: Linear Lowest CV Highest S:N Lowest Voltage Red A (APC-Cy7): Before (30Mar06) & After (10Apr06) Replacing PMT 30 March 2006 10 April 2006 5000 200 5000 180 4500 180 4500 160 4000 160 4000 140 3500 140 3500 120 3000 120 3000 100 2500 100 2500 80 2000 80 2000 60 1500 60 1500 40 1000 40 1000 20 500 20 500 0 0 300 350 400 450 500 550 600 650 700 750 C V o r L IN 200 0 0 300 350 400 450 500 Voltage CV Lin 550 600 650 700 750 Voltage Ratio Optimal Voltages: Indeterminate High CV’s; poor S:N CV Lin Ratio Optimal Voltages: 625-700 R a tio RedA Voltage vs. CV, LIN and Ratio R a tio C V o r L IN RedA Voltage vs. CV, LIN and Ratio Faulty PMT on install… Before After 4000 3000 3000 S S C-A S S C-A 4000 2000 2000 1000 1000 0 0 0 10 2 3 10 65010RedVolts A-A 4 10 5 0 10 2 3 10 Red C-A 10 650 Volts 4 10 5 LSRII Qualification Validation Optimization Linearity Resolution Signal-to-Noise Ratio (S:N) - LLOD & LLOQ Select Optimal Voltages for Inst Performance with AssaySpecific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F) Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines Beads Used for LSRII Optimization • STAINED Comp Beads: validated range in 50v increments (each PMT) • • • • Assay Specific fluorescence 1pk Run: Once, then after optical service- “High” flow rate & LOG scale Uses: Optimization • Optimize PRIMARY Fluorescence (Primary>Secondary) • Unstained Comp Beads: optimized voltages (all PMTs) • • • • Non-fluorescent 1pk Run: Once, then after optical service- “High” flow rate & LOG scale Uses: Validation • Check S:N (LLOD & LLOQ) • 1x or Midrange Rainbow: optimized voltages (all PMTs) • Moderate fluorescence, near cellular expression; Broad Spectrum Ex & Em • 1pk • Run: Once (after optimization) - “High” flow rate & LOG scale • Uses: Validation • Establish Target Channels Linearity: medians = assay specific target channels Criteria for the Selection Optimal PMT Voltages The primary fluorescence should be the highest in the respective detector relative to all secondary detectors. (Blue B-A): GeoMean = 19 0 10 2 10 3 10 4 (Green A-A): GeoMean = 5462 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 129 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 10 5 0 2 10 10 2 R ed B- A 3 10 4 10 5 0 10 2 10 10 4 3 10 4 10 5 0 10 2 10 4 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A (Violet D-A): GeoMean = 21.4 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 90.2 10 10 Violet B- A G reenD -A 0 3 (Violet C-A): GeoMean = 12.3 R ed C- A 3 10 (Violet B-A): GeoMean = 47 (Red C-A): GeoMean = 15 G reenC -A 10 2 Violet A- A (Green D-A): GeoMean = 99.2 0 10 (Red B-A): GeoMean = 5.69 (Green C-A): GeoMean = 26.3 0 3 R ed A- A (Green B-A): GeoMean = 44.7 0 10 (Violet A-A): GeoMean = 8.05 3 10 4 (Violet E-A): Geo Mea n = 7.82 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 32.2 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 9.43 Green A: TNF PE-Cy7 460v Baseline & Final 0 10 2 10 3 10 4 Violet G- A TNFa PE-Cy7 DukeJY_460v baseline.fcs (Violet H-A): GeoMean = 18.8 43 0 10 2 10 Violet H- A 3 10 4 (Blue B-A): GeoMean = 14.1 0 10 2 10 3 10 4 (Green A-A): GeoMean = 617 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 295 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 3 10 4 10 5 0 2 10 10 5 0 10 2 10 2 10 4 3 10 4 10 5 0 10 2 10 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 0.98 10 4 (Violet D-A): GeoMean = 8.57 G reenD -A 0 10 (Violet C-A): GeoMean = 6.55 R ed C- A 3 3 Violet B- A (Red C-A): GeoMean = 97.4 10 10 (Violet B-A): GeoMean = 206 R ed B- A G reenC -A 10 2 Violet A- A (Green D-A): GeoMean = 3.39 0 10 (Red B-A): GeoMean = 870 (Green C-A): GeoMean = 165 0 3 R ed A- A (Green B-A): GeoMean = 986 0 10 (Violet A-A): GeoMean = 1890 3 10 4 (Violet E-A): Geo Mea n = 3.85 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 32.2 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 9.64 Green B: CD8 PerCP-Cy5.5 450v Baseline 0 10 2 10 3 10 4 Violet G- A CD8 PerCPCy55 DukeJY_450vbasline.fcs (Violet H-A): GeoMean = 18.7 44 0 10 2 10 Violet H- A 3 10 4 (Blue B-A): GeoMean = 15.1 0 10 2 10 3 10 4 (Green A-A): GeoMean = 609 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 549 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 10 5 0 2 10 10 2 R ed B- A 3 10 4 10 5 0 10 2 10 10 4 3 10 4 10 5 0 10 2 10 4 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A (Violet D-A): GeoMean = 26.3 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 10.7 10 10 Violet B- A G reenD -A 0 3 (Violet C-A): GeoMean = 37.2 R ed C- A 3 10 Violet A- A (Red C-A): GeoMean = 114 G reenC -A 10 2 (Violet B-A): GeoMean = 206 (Green D-A): GeoMean = 18 0 10 (Red B-A): GeoMean = 1181 (Green C-A): GeoMean = 442 0 3 R ed A- A (Green B-A): GeoMean = 4343 0 10 (Violet A-A): GeoMean = 3418 3 10 4 (Violet E-A): Geo Mea n = 28.2 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 274 0 Green B: CD8 PerCPCy5.5 550v Final 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 192 0 10 2 10 3 10 4 Violet G- A CD8 PerCPCy55 DukeJY_550v final.fcs (Violet H-A): GeoMean = 91.4 45 0 10 2 10 Violet H- A 3 10 4 (Blue B-A): GeoMean = 23.7 0 10 2 10 3 10 4 (Green A-A): GeoMean = 1497 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 368 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 3 10 4 10 5 0 2 10 10 5 0 10 2 10 2 10 4 3 10 4 10 5 0 10 2 10 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 284 10 4 (Violet D-A): GeoMean = 53.2 G reenD -A 0 10 (Violet C-A): GeoMean = 25.6 R ed C- A 3 3 Violet B- A (Red C-A): GeoMean = 1663 10 10 (Violet B-A): GeoMean = 267 R ed B- A G reenC -A 10 2 Violet A- A (Green D-A): GeoMean = 305 0 10 (Red B-A): GeoMean = 1324 (Green C-A): GeoMean = 2747 0 3 R ed A- A (Green B-A): GeoMean = 6724 0 10 (Violet A-A): GeoMean = 321 3 10 4 (Violet E-A): Geo Mea n = 16.8 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 39.1 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 11.1 Green C: CD27 PE-Cy5 350v Baseline 0 10 2 10 3 10 4 Violet G- A CD27 PE-Cy5 VRC_350v baseline.fcs (Violet H-A): GeoMean = 19.4 46 0 10 2 10 Violet H- A 3 10 4 (Blue B-A): GeoMean = 23.4 0 10 2 10 3 10 4 (Green A-A): GeoMean = 1507 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 709 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 10 5 0 2 10 10 2 R ed B- A 3 10 4 10 5 0 10 2 10 10 4 3 10 4 10 5 0 10 2 10 4 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A (Violet D-A): GeoMean = 129 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 1689 10 10 Violet B- A G reenD -A 0 3 (Violet C-A): GeoMean = 114 R ed C- A 3 10 Violet A- A (Red C-A): GeoMean = 1904 G reenC -A 10 2 (Violet B-A): GeoMean = 252 (Green D-A): GeoMean = 1379 0 10 (Red B-A): GeoMean = 1805 (Green C-A): GeoMean = 17138 0 3 R ed A- A (Green B-A): GeoMean = 13446 0 10 (Violet A-A): GeoMean = 563 3 10 4 (Violet E-A): Geo Mea n = 82.4 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 306 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 196 Green C: CD27 PE-Cy5 450v Final 0 10 2 10 3 10 4 Violet G- A CD27 PE-Cy5 VRC_450v final.fcs (Violet H-A): GeoMean = 91.5 47 0 10 2 10 Violet H- A 3 10 4 (Blue B-A): GeoMean = 33.7 0 10 2 10 3 10 4 (Green A-A): GeoMean = 614 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 6.68 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 10 5 0 2 10 10 2 R ed B- A 3 10 4 10 5 0 10 2 10 10 4 3 10 4 10 5 0 10 2 10 4 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A (Violet D-A): GeoMean = 50.7 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 252 10 10 Violet B- A G reenD -A 0 3 (Violet C-A): GeoMean = 251 R ed C- A 3 10 (Violet B-A): GeoMean = 229 (Red C-A): GeoMean = 42.5 G reenC -A 10 2 Violet A- A (Green D-A): GeoMean = 22260 0 10 (Red B-A): GeoMean = 24.6 (Green C-A): GeoMean = 4539 0 3 R ed A- A (Green B-A): GeoMean = 2966 0 10 (Violet A-A): GeoMean = 139 3 10 4 (Violet E-A): Geo Mea n = 10.9 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 34.7 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 9.53 Green D: CD45RO PE-TR 430v Baseline & Final 0 10 2 10 3 10 4 Violet G- A CD45RO PE -T R VRC_430v baeline.fcs (Violet H-A): GeoMean = 19.1 48 0 10 2 10 Violet H- A 3 10 4 (Blue B-A): GeoMean = 44.3 0 10 2 10 3 10 4 (Green A-A): GeoMean = 269 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 3.41 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 10 5 0 2 10 10 2 R ed B- A 3 10 4 10 5 0 10 2 10 10 4 3 10 4 10 5 0 10 2 10 4 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A (Violet D-A): GeoMean = 925 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 28006 10 10 Violet B- A G reenD -A 0 3 (Violet C-A): GeoMean = 395 R ed C- A 3 10 (Violet B-A): GeoMean = 222 (Red C-A): GeoMean = 24.4 G reenC -A 10 2 Violet A- A (Green D-A): GeoMean = 22352 0 10 (Red B-A): GeoMean = 8.53 (Green C-A): GeoMean = 3200 0 3 R ed A- A (Green B-A): GeoMean = 1604 0 10 (Violet A-A): GeoMean = 95.4 3 10 4 (Violet E-A): Geo Mea n = 244 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 90.3 Green E: MIP1b PE 350v Baseline & Final 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 10 0 10 2 10 3 10 4 Violet G- A M IP1b PE DukeJY _450v baseline.fcs (Violet H-A): GeoMean = 18.6 49 0 10 2 10 Violet H- A 3 10 4 Blue Duke LSRII Optimization of Specific Fluorescence Primary Fluorescence Secondary Fluorescence Violet Red Green (Blue B-A): GeoMean = 15.3 0 10 2 10 3 10 4 (Green A-A): GeoMean = 9.54 10 5 0 10 2 10 3 10 4 (Red A-A): GeoMean = 28.4 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 2 10 4 10 5 0 10 2 10 3 10 4 10 5 0 10 2 10 2 10 3 10 4 10 5 0 10 2 10 4 3 10 4 10 5 0 10 2 10 4 10 5 10 3 10 4 10 5 2 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 Violet C- A (Violet D-A): GeoMean = 30.1 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 214 10 10 (Violet C-A): GeoMean = 38.8 G reenD -A 0 3 Violet B- A R ed C- A 3 10 Violet A- A (Red C-A): GeoMean = 41.2 G reenC -A 10 2 (Violet B-A): GeoMean = 85.3 (Green D-A): GeoMean = 250 0 10 R ed B- A (Green C-A): GeoMean = 72.1 10 10 (Red B-A): GeoMean = 42.1 G reenB -A 0 3 R ed A- A (Green B-A): GeoMean = 51.2 0 10 (Violet A-A): GeoMean = 10.6 3 10 4 (Violet E-A): Geo Mea n = 29.5 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): Geo Mea n = 235 0 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 178 Neg Comp Beads 0 10 2 10 3 10 4 Violet G- A Controls_Neg Comp.fcs (Violet H-A): GeoMean = 90.9 51 0 10 2 10 Violet H- A 3 10 4 LSRII Qualification Validation Optimization Linearity Resolution Signal-to-Noise Ratio (S:N) - LLOD & LLOQ Select Optimal Voltages for Inst Performance with AssaySpecific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F) Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines (Blue B-A): GeoMean = 4587 0 10 2 10 3 10 4 (Green A-A): GeoMean = 1505 10 5 0 10 2 10 3 10 4 (Red A-A): Ge oMe an = 2172 10 5 0 10 2 G reenA -A B lue B-A 10 2 10 3 10 4 10 5 0 10 2 G reenB -A 10 2 10 10 4 10 5 0 10 3 10 4 3 10 4 10 5 0 2 10 10 5 0 10 2 10 2 10 4 3 10 4 10 5 0 10 2 2 10 10 5 10 3 10 4 10 5 10 3 10 4 10 5 10 5 10 5 10 5 10 5 10 5 (Violet D-A): GeoMean = 2539 10 5 0 10 2 10 3 10 4 Violet D- A (Green E-A): GeoMean = 30710 10 4 Violet C- A G reenD -A 0 10 (Violet C-A): GeoMean = 3283 R ed C- A 3 3 Violet B- A (Red C-A): Ge oMe an = 3934 10 10 (Violet B-A): GeoMean = 970 R ed B- A G reenC -A 10 2 Violet A- A (Green D-A): GeoMean = 65203 0 10 (Red B-A): Ge oMe an = 3927 (Green C-A): GeoMean = 22057 0 3 R ed A- A (Green B-A): GeoMean = 11590 0 10 (Violet A-A): GeoMean = 712 3 10 4 (Violet E-A): GeoMea n = 3052 10 5 0 10 2 G reenE-A 10 3 10 4 Violet E-A (Violet F-A): GeoMea n = 12196 0 1x (Target Channel Values) 10 2 10 3 10 4 Violet F-A (Violet G-A): GeoMean = 27119 0 10 2 10 3 10 4 Violet G- A Controls_1x.fcs (Violet H-A): GeoMean = 4157 53 0 10 2 10 Violet H- A 3 10 4 Target Channels & Ranges LSRII Qualification Validation Optimization Linearity Resolution Signal-to-Noise Ratio (S:N) - LLOD & LLOQ Select Optimal Voltages for Inst Performance with AssaySpecific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F) Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines Duke University Medical Center Daily Calibration • Mid-Range or “1x” Rainbow Beads: TC settings • • • • Moderate fluorescence, similar to cells; Broad Spectrum Ex & Em 1pk Run: Morning & Before each Exp - “High” flow rate & LOG scale Uses: Calibration & Standardization • • • • Set Assay-Specific Target Channels (TCs) Determine Daily Voltage Settings Daily QC: P/F (CVs, voltages, trends) LSRII Performance Verification for Exp Run: P/F (CVs & TCs) • 8 Peak Rainbow & Unstained Comp Beads: TC Settings • • • • Non-fluorescent to Very Bright; Broad Spectrum Ex & Em 8pk and 1pk Run: Once Daily - “High” flow rate & LOG scale Uses: Standardization • • • • Check Linearity Check S:N (LLOD & LLOQ) Check Specificity Check Resolution • Ultra Rainbow Beads: Mean ch ~150,000 • • • • VERY Bright; Broad Spectrum Ex & Em 1pk Run: Once Daily - “Low” flow rate & linear scale Uses: Manufacturer Recommended • • • • Daily QC: P/F (CV & voltages) Check/Adjust Area Scaling Check/Adjust Time Delay Check/Adjust Window Extension Duke University Medical Center Daily Standardization • • • Mid-Range or “1x” Rainbow Beads: • Record Daily For each PMT & Scatter • CV • HV • Median • Plot Monthly For each PMT & Scatter • CV • HV vs Median 8 Peak Rainbow & Unstained Comp Beads: • Record Daily For each PMT & Scatter • Median • Neg Comp Beads • Peak 5 • Peak 6 • Frequency • Peak 5 • Peak 6 • CV: peak 5 • HV • Plot Monthly For each PMT & Scatter • Linearity: Medianpk6-Medianpk5/Medianpk5 • Signal:Noise: Medianpk5/MedianNeg • Q & B??? Ultra Rainbow Beads: • Record Daily For each PMT & Scatter • CV • HV • Median • Plot Monthly For each PMT & Scatter • CV • HV vs Median Duke University Medical Center 1x Blue B Trend line 1 X B lu e B (+/- 5 % o f M e a n ) 475 10,000 9,000 465 8,000 6,000 V o ltag e 455 5,000 445 M ed ian 7,000 4,000 3,000 435 2,000 1X V oltages 1X M edians 1,000 11/13/2006 11/6/2006 10/30/2006 10/23/2006 10/16/2006 10/9/2006 10/2/2006 9/25/2006 9/18/2006 9/11/2006 9/4/2006 8/28/2006 8/14/2006 8/21/2006 0 8/7/2006 425 58 Duke University Medical Center Ratio and Voltage Variation 700 500 300 B515 100 G 610 R660 V585 0 1 2 3 4 5 6 7 8 9 10 11 3000 3 .5 2500 3 .0 2 .5 2000 2 .0 1500 1 .5 1000 1 .0 R at io 500 CV 0 .5 0 0 .0 1 2 3 4 5 6 7 8 9 1 11 D ays Duke University Medical Center LASERs have a limited lifespan… Laser Current Laser Power Laser Performance 5.84 16.00 5.82 5.80 15.60 5.78 5.76 15.40 5.74 15.20 5.72 15.00 5.70 5.68 14.80 Date L a s e r P o w e r ( m W a tt s ) L a ser C u r r en t (A m p s) 15.80 Overview Why QC is Important LSRII Optical Configuration LSRII QC LSRII Validation LSRII Optimization LSRII Calibration LSRII Standardization Practice Analysis?? Duke University Medical Center Example: Stain Index, Qr, and Br Across Laboratories Example : FITC Channel Lab SI Qr Br 8 peak beads 12 00 90 0 1 77.4 0.015 233 60 0 30 0 0 0 2 10 3 10 4 10 5 10 10 00 80 0 2 134.5 0.028 216 High SI = Good Resolution; correlates with high Qr and low Br 60 0 40 0 20 0 0 0 2 10 3 10 4 10 5 10 10 00 80 0 3 55.6 0.015 976 60 0 40 0 20 0 0 0 2 10 10 3 10 4 10 5 12 00 90 0 4 80.5 0.01 298 60 0 30 0 0 0 2 10 3 10 10 4 5 10 80 0 60 0 5 66 0.01 613 40 0 20 0 0 0 10 2 10 3 10 4 10 5 Low SI = Bad Resolution; correlates with low Qr and high Br 6 13.7 0.007 2322 # C el l s 60 0 40 0 20 0 0 0 10 2 10 3 10 4 10 5 10 00 80 0 7 16.2 0.018 2768 60 0 40 0 20 0 0 0 10 2 10 3 10 4 10 5 Impact of Instrument Performance in Quality Assurance of Multicolor Flow Cytometry Assays. Jaimes MC,1 Stall A,1 Inokuma M,1 Hanley MB,1 Maino S,1 D’Souza MP,2 and Yan M.1 ISAC 2010 1BD Biosciences, San Jose, CA 95131; 2Division of AIDS, National Institute of Allergy and Low SI = Bad Resolution; correlates with high Br Correlation of SI, Qr, and Br with Assay Performance Lab 2 Lab 4 Lab 6 Plots gated on lymphocytes Plots gated on CD3+ cells Plots gated on CD3+CD8+ cells Impact of Instrument Performance in Quality Assurance of Multicolor Flow Cytometry Assays. Jaimes MC, Stall A, Inokuma M, et al. ISAC 2010 Lab 7 1 Assay – 4 Platforms BD FACSCanto II BD FACSAria III 64 CD4 V450 • CD8 PE • CD3 FITC BD LSR II • CD20 APC • BD LSRFortessa Thank You!! 65 Defining Instrument Performance and Sensitivity: Qr, Br, and SDen Qr: Anti-CD10 PE Example Dim Population The laser and detectors were attenuated by ND filters over a 30-fold range to illustrate the effects of decreasing detector sensitivity on population resolution. Qr=0.227 Transmission SI Corrected 100% 83 Qr=0.087 35.5% 55 Qr=0.038 17.8% 35 Qr=0.014 7.1% 20 Qr=0.007 3.5% 14 Relative Background: Br • Br is a measure of true optical background in the fluorescence detector, which helps indicate how easily (dim) signals may be resolved from unstained cells in that detector. Unbound antibody or fluorochrome Scatter from the flow cell and ambient light. Raman scatter Spectral overlap on a cell Cell autofluorescence • Factors affecting Br: dirty flow cell, damaged optical component Why is Br important? • High Br widens negative and dim populations. • High Qr value = lower resolution • Low Qr value = higher resolution Low Br High Br Br: Optical Background from Propidium Iodide • Example: It is common to use propidium iodide (PI) to distinguish live from dead cells. Propidium iodide was added in increasing amounts to the buffer containing beads, and Qr and Br were estimated. PerCP Br 9000 8000 7000 6000 5000 4000 3000 2000 1000 0 Br Qr 0.05 0.04 0.03 0.02 Qr 0.01 0 0 1 PI-free 2 3dye 4(µg) 5 6 • Residual PI in your sample tube will increase Br, which will reduce sensitivity. Electronic Noise (SDen) • SDen is the background signal due to electronics: • Contributed by: • PMT connections / PMT noise • Cables too near power sources • Digital error • Broadens the distribution of unstained or dim particles • Increases in electronic noise results in decreased resolution sensitivity Most important for channels with low cellular autofluorescence • APC-Cy™7, PE-Cy™7, PerCP-Cy™5.5 Cy™ is a trademark of Amersham Biosciences Corp. Cy™ dyes are subject to proprietary rights of Amersham Biosciences Corp and Carnegie Mellon University and are made and sold under license from Amersham Biosciences Corp only for research and in vitro diagnostic use. Any other use requires a commercial sublicense from Amersham Biosciences Corp, 800 Centennial Avenue, Piscataway, NJ 08855-1327, USA. Gel (from coupling) found all over the flow cell???? Green E Green D Green C Green B Green A 10000 1000 CV QuickTime™ and a TIFF (LZW) decompressor are needed to see this picture. 100 10 1 0 200 400 600 voltage 800 1000 77 78 79