GASS™-2040 Gas Analysis Sampling Systems PERMA PURE 2040-001 GASS-2040 • New product introduction at AWMA • Combines capabilities of Probe / Mini-GASS™ and GASS™-II systems • Can be mounted anywhere, including directly on stack flange with optional integral sample probe • Most complete sample conditioning system on the market • GASS-II discontinued, all other GASS systems still available PERMA PURE 2040-002 The General Problem Gas analysis for Emissions or Process Monitoring involves three steps: 1. Sampling 2. Conditioning Samples are typically: Hot Analyzers require samples to be: Cool 3. Analysis Wet Dirty Dry Clean Sample conditioning receives little attention, but is often the most difficult step of the three. PERMA PURE 2040-003 Sampling Methods 1. In-situ analysis 2. Extraction a. Wet (filtered only) b. Dilution c. Drying 1) Absorption 2) Condensation 3) Selective permeation PERMA PURE 2040-004 Typical Problems in Gas Samples Gases • Rarely a problem Liquids • Water • Acid mists (sulfuric acid) Solids • Ammonia salts • Excessive particulates PERMA PURE 2040-005 Water Problems • Clogging or damage to analyzers or components • Requirement for heated lines or analyzers • Long calibration times due to adsorption/desorption • Interference in infrared (or other) measurements • Loss of water-soluble analytes • Participation in unwanted chemical reactions PERMA PURE 2040-006 Sulfur Dioxide Solubility in Water SO2 Solubility (ml SO2/100 ml H2O) 100 90 80 70 60 50 40 30 20 10 0 0 10 20 30 40 50 60 70 80 90 100 Temperature (Deg. C) PERMA PURE 2040-007 Sulfur Dioxide Solubility in Water Temperature Degree C SO2 Solubility in H2O ml SO2 / 100 ml H2O 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 79.8 77.2 74.7 72.2 69.8 67.5 65.2 63.0 60.8 58.7 56.6 54.7 52.7 50.8 49.0 47.3 45.6 43.9 Degree C SO2 Solubility in H2O ml SO2 / 100 ml H2O 18 19 20 21 22 23 24 25 26 27 28 29 30 35 40 90 100 42.4 40.8 39.4 38.0 36.8 35.3 34.0 32.8 31.6 30.4 29.3 28.2 27.2 22.5 18.8 2.1 .0 Temperature PERMA PURE 2040-008 Unwanted Reactions 1. Formation of Sulfuric Acid a. 2SO2 + O2 2SO3 SO3 + H2O H2SO4 b. NO2 + SO2 + H2O H2SO4 + NO 2. Formation of Ammonia Salts a. NH3 + SO2 + H2O NH4HSO3 b. NH3 + SO3 + H2O NH4HSO4 PERMA PURE 2040-009 Acid Mist Problems • Mists are very corrosive, viscous, clogging • Difficult to remove; slow to coalesce into sizeable droplets when cooled • Special coalescing filters helpful PERMA PURE 2040-010 Ammonia Salts Problems Present special problems • Ammonia is gas, passes through probe filter • Forms clogs if allowed to cool, react with SO2 or SO3 and H2O to form ammonium salts • Requires special absorbents to remove ammonia before cooling to prevent clogs PERMA PURE 2040-011 Excessive Particulates Easiest to solve • Probe filter with heated blowback • Secondary filter with finer porosity • Inertial bypass filters if necessary PERMA PURE 2040-012 The Solution GASS™-2040 Gas Analysis Sampling Systems • Retain water-soluble analytes (HCl, SO2, NOX) • Achieve low final sample dew points (down to -25oC) • Accept high sample flow rates (up to 25 L/minute) • Accommodate nearly any sample (customized) • Eliminate heated sample lines (stack mounting) PERMA PURE 2040-013 ™ GASS -2040 Gas Analysis Sampling System PERMA PURE 2040-014 ™ GASS -2040 Overview • Stainless steel NEMA-4X enclosure • Integral probe with heated filter and blowback • Hastelloy heat exchanger • Coalescing/particulate filter with automatic drain • Ammonia scrubber(s) • Wide range of dryers, double dryers • PLC with Ethernet for remote control • Safety interlocks and alarms • Sample pump, other options to customize system PERMA PURE 2040-015 Integral Probe, Filter and Blowback PERMA PURE 2040-016 Integral Probe, Filter and Blowback • Range of probe lengths and materials Directly connected, no heated line • Probe filter stainless steel or glasscoated Temperature controlled by PLC Avoids losses of analytes in filter • Blowback heated, closely coupled Blowback timing controlled by PLC Avoids condensation during blowback PERMA PURE 2040-017 Coalescing / Particulate Filter and Drain PERMA PURE 2040-018 Coalescing / Particulate Filter PERMA PURE 2040-019 Coalescing / Particulate Filter PERMA PURE 2040-020 Coalescing / Particulate Filter • Two-layer, coalescing and particle element Temperature controlled by PLC Removes particles >0.1 micron and mists • Hastelloy heat exchanger Improves acid mist removal • Available in stainless steel or PVDF Kynar Corrosion resistant ® PERMA PURE 2040-021 Automatic Drain PERMA PURE 2040-022 Automatic Drain • Coalesced liquid removed by eductor Drain timing controlled by PLC Reliable, no peristaltic pump • Eductor entirely PTFE Teflon® Full corrosion resistance • Teflon check valve Positive seal when not operating PERMA PURE 2040-023 Ammonia Scrubber PERMA PURE 2040-024 Ammonia Scrubber PERMA PURE 2040-025 Ammonia Scrubber PERMA PURE 2040-026 Ammonia Scrubber • Proprietary scrubbing media Long-lasting, selective for ammonia • No tools needed for media replacement Easy maintenance • Stainless steel with stainless steel shell or polysulfone shell for visibility Corrosion resistant PERMA PURE 2040-027 ® Nafion Gas Dryer PERMA PURE 2040-028 NAFION Chemical Structure ® [ (CF2 CF2 )m CF CF2 ] n O CF2 CF WHEN: m = 5 to 13.5 n = ca 1,000 z = 1,2,3... x = 1 to 13 CF3 Z O CF2 CF2 SO3 H ● [H2O]X PERMA PURE 2040-029 NAFION Gas Dryer Principle PERMA PURE 2040-030 NAFION Dryer Schematic PERMA PURE 2040-031 PD-Series Polytube Dryers PERMA PURE 2040-032 NAFION Selectivity • Nafion permeation selectivity is based on chemical reactivity, not size • Only compounds that chemically associate with sulfonic acid permeate through Nafion • Bases associate with sulfonic acid and permeate through Nafion • Very few bases are gases at typical operating temperatures, so very few compounds permeate PERMA PURE 2040-033 Effect on Analytes • Removes water directly as water vapor. No condensate forms to dissolve analytes. • Removes compounds with hydroxyl groups (-OH) H2O, some organics • Retains important compounds O2, O3, CO, CO2, NOX, SO2, SO3, H2S, Hg, HCl PERMA PURE 2040-034 PD-Series Performance Chart 12” 24” 48” DEWPOINT (oC) 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40 -45 -50 PD-50T ...0 PD-100T ...0 PD-200T ...0 MODEL 72” 1 2 4 2 4 8 3 4 5 6 7 8 9 10 6 8 10 12 14 16 18 20 12 16 20 24 28 32 36 40 FLOW (liters/min.) PERMA PURE 2040-035 Low Final Dew Points • Dew point as low as -45oC can be achieved. • Dew point of -10oC to -25oC is typical. • Performance is a function of dryer size and sample flow rate. • Analytical results and reliability improved by reducing water interferences. • Acid mist problems overcome. PERMA PURE 2040-036 PLC PERMA PURE 2040-037 PLC • Digital selection of operating conditions Precise control of operating parameters • Standard unit, easy programming Easy maintenance and updating • Optional Ethernet connection Remote control and data display PERMA PURE 2040-038 Optional Heatless Dryer PERMA PURE 2040-039 Optional Heatless Dryer PERMA PURE 2040-040 Optional Heatless Dryer PERMA PURE 2040-041 Optional Heatless Dryer • Steady supply of dry air for system use Continuous 60 L/min. to -45ºC dewpoint • No line pressure surges Inactive canister repressurized before reuse • Very reliable with no maintenance Electronic switching, permanent springloaded molecular sieve PERMA PURE 2040-042 Optional Sample Pump PERMA PURE 2040-043 Optional Sample Pump • Pump head mounted in heated zone Prevents sample condensation • Pump motor mounted in unheated zone Improves pump reliability • Pump head stainless steel or Teflon Corrosion-resistant, long life • Sample transported under pressure Avoids contamination from minor leaks PERMA PURE 2040-044 GASS-2040 Options • Integral Probe • Dryer Sizes • Probe Filter • Purge Heatless Dryer • Blowback • Ethernet Connection • Bypass Filter • X- or Z-Purge • Coalescing Filter • Kynar® Fittings • Automatic Drain • Sample Pump • Ammonia Scrubbers • Purge Eductor PERMA PURE 2040-045 Summation ™ • GASS -2040 combines all necessary sampling and conditioning in one package •Proven technology, years of experience • Opportunity to eliminate heated lines • Many options; buy only what is needed • Smaller systems available for less demanding applications PERMA PURE 2040-046