Z-2000

advertisement
Introduction for
HITACHI Z-2000 SERIES
Polarized Zeeman Atomic Absorption
Spectrophotometer
About HITACHI Polarized Zeeman AAS
Year
1973
1976
1980
1983
1987
1987
1988
1991
1996
2004
Model
501 (Flame)
170-70 (Furnace)
180-60 (Flame)
180-70 (Furnace)
180-80 (Tandem)
Z-6000 (Flame)
Z-7000 (Furnace)
Z-8000 (Tandem)
Z-9000 (Furnace)
Z-6100 (Flame)
Z-8100 (Tandem)
Z-8200 (Tandem)
Z-5000 series
Z-2000 series
Units
46
291
103
211
488
356
134
531
346
1,365
566
1,054
1,610
>450
Tandem alignment
Multi-elements analysis
PC control
3-D optics 三次元光学系
Dual detector system 二検知器
Hitachi introduced world’s first polarized Zeeman AAS Model 501 in 1973, and
have delivered a variety of units to about 7,000 customers in 25 countries
around the world.
New Model / Z-2000 Series
Product configuration
(1) Tandem Type
(2) Flame Type
(3) Graphite Furnace Type
Z-2300
Z-2000
Model Z-2000
Model Z-2300
Model Z-2700
Z-2700
High Quality
1. Direct-current Zeeman background correction
(both Flame method and Graphite Furnace
method)
直流ゼーマンバックグラウンド補正、フレームとファーネスの両方
2. Dual detector
二検知器
Photomultiplier for Sample
サンプル用光電子増倍管
Photomultiplier for reference
レファレンス用光電子増倍管
Z-2000 series Optics
Dual detector system enables simultaneous signal capture
of both Sample and Reference.
二検知器システムにより、サンプル光とレファレンス光の同時取り込みが可能になりました。
Timing Chart of Signal Processing
20 ms
20 ms
Exact simultaneous measurement of sample signal and reference signal with the
dual detector system realizes low noise and highly accurate analysis.
New optics with dual detection system
achieves better sensitivity.
Comparison data among Z-8200/Z-5000/Z-2000
Model Z-8200
Model
Model Z-5000
Z-5000
Detection
Limit
0.4 ug/L
Z-2000
0.15 ug/L
Model Z-2000
Element : As (193.7nm)
Graphite furnace method, As Standard Solution (1ug/L)
40uL injection
Stable baseline


Flame DC Zeeman correction and dual detector system
achieved very stable baseline for long time.
After the warm-up time of about 5 min., baseline can be
stabilized over one hour.
Baseline change (Flame method) : Cu lamp
Zeeman background correction
in Hydride Formation method
Lamp
Flame
Optics
Detector
Magnet
Heating Absorption Cell
HFS-3
Heating Absorption Cell
Burner Head
Z-5000
Z-2000
Magnet
descend
mechanism
Magnet is exposed
to Flame
Background correction
impossible
Shield
Plate
Background correction
possible
Zeeman background correction becomes possible, since
heating absorption cell can be set between magnet.
Zeeman background correction
in Hydride Formation method
As Standard Solution
0
5
10
15 ug/L
As Standard Solution
0
5
10
15 ug/L
River
Water
Z-5000
Z-2000
Measurement of As
(HFS-3, Heating Absorption cell)
Know-how for a better AAS expert
更なる原子吸光スペシャリストへの測定のコツ
Hitachi High-Technologies
FLAME
The Determination of
Cadmium, Copper and Lead
in Nickel Plating Bath
Solution.
Analysis of high-calorie food
高カロリー食品の分析
Requirements for analysis


分析に対する要求項目
Rapid analysis (use of autosampler) 迅速分析
Concentration difference in elements(different dilution factors)
元素により濃度差

Continuous measurement (multi-elements sequential analysis)
連続測定
Problem
問題点
1. Different dilutions, different viscosities cause different delay times
希釈倍率による粘度差で試料の到達時間に差が出る
2. Large number of samples多数検体
3. Long-term stability required 長時間安定性が必要
4. Clogging of burner バーナーの詰り
5. Corrosion by highly acidic solution 高酸性溶液による腐食
Analysis of high-calorie food
Solution
1.Different dilutions, different viscosities
cause different delay times 希釈倍率による粘度差で試料の到達時間に差が出る
→ Different delay time can be set for each element
元素毎に測定開始までの時間設定が可能
Delay time
Analysis of high-calorie food
2.Large number of samples 多量検体
→ Multiple autosampler racks can be used.
オートサンプラのラックが複数使えます。
Analysis of high-calorie food
3. Long-term Stability
Element
Conc.(mg/100mL)
RSD(%)
長時間安定性
Na
82
K
173
Ca
51
Mg
20
Fe
520
Zn
521
Mn
126
Cu
116
0.7
0.7
1.8
0.6
1.6
1.4
1.8
2.3
Analysis of high-calorie food
4. Clogging of burnerバーナーの詰り
→ Clean by setting the delay time (pre-spray)
予備噴霧時間の設定により洗浄
1~30min
Cleaning of nebulizer
ネブライザーの洗浄
OK
NG
insert from this side
Inlet
吸入口
Nebulizer tip
先端
Cleaning
wire
A cleaning wire has to be inserted from inlet.
クリーニングワイヤーは吸入口から入れる。
Cleaning of burner head
The burner head is
cleaned by thick paper.
バーナースロットは厚紙でクリーニング。
Overhaul cleaning is
required once a week.
分解清掃は週1回程度は必ず行う。
Analysis of high-calorie food
5. Corrosion by highly acidic solution 高酸性溶液による腐食
Engineering plastic
(chemical resistant)
エンジニアプラスチック
(耐薬品性)
Chemical Resistance of Flame Burner Chamber
Name of Chemical
Concentration
Temperature
Hydrochloric acid
Concentrated
Normal temp.
Hydrofluoric acid
Concentrated
Normal temp.
Nitric acid
10% or less
Normal temp.
Sulfuric acid
10% or less
Normal temp.
Aqua regia
3-fold dilution or lower
Normal temp.
NaCl solution
Saturated
80 C or less
Tetrahydrofuran

Normal temp.
Tri-n-octylamine

Normal temp.
Xylene

Normal temp.
1-methyl-2-pyrolidone

Normal temp.
Heptane

Normal temp.
Acetonitrile

Normal temp.
Acetone

Normal temp.
Propanol

Normal temp.
N, N-dimethylacetoamide

Normal temp.
Ethylene glycol

Normal temp.
Cyclohexane

Normal temp.
2-ethoxyethyl acetate

Normal temp.
Butyl acetate

Normal temp.
4-methyl-2-pentanone (MIBK)

Normal temp.
Concentration range can be changed by
analytical line selection.
分析線を変えることにより測定濃度範囲を変えられます
Sensitivity for 1mg/L Na
Use of oxydizing flame and reduction flame
酸化炎と還元炎の使い分け
Flame
condition
Oxydizing flame
(fuel lean)
Reduction flame
(fuel rich)
フレームの状態
酸化炎
還元炎
Applicable
elements
Cd, Pb, Mn, Fe...
Sn, Cr, Mo...
Difference of linearity in working curves
フレーム状態による検量線の違い
Oxydizing flame 酸化炎
Example
of Fe 0.2
Reduction flame
還元炎
Absorbance
Absorbance
0.2
0
Conc.
0
Conc.
Oxydizing flame may be changed to reduction flame by introduction of real sample.
Also, the flame condition may be changed acetone come from C2H2 cylinder when
residual pressure < 0.5MPa
実試料の導入によって酸化炎→還元炎に変わる場合がある。アセチレンガスの残圧が0.5MPa以下の場合、ボンベ
からのアセトンの混入も、炎の状態を変化させる。
Contamination
Laboratory environment: use of indoor shoes
室内環境 :上履き使用
Contamination
Laboratory environment: use of local ventilation
室内環境 :局所排気
Desktop
local hood
Contamination from environment
Zn signal
Aspirate DW
Flame on
tobacco
smoke
Zn signal in flame AAS
DW left for 3days
Contamination from environment
ABS
0.100
0.095
0.090
0.085
0.080
0.075
0.070
0.065
0.060
0.055
0.050
0.045
0.040
0.035
0.030
0.025
0.020
0.015
0.010
0.005
0.000
-0.005
-0.010
-0.015
-0.020
REF
0.20
0.15
0.10
0.05
0.00
Na signal
Aspirate DW
DW left for 3days
Tobacco
smoke
Flame on
0
100
200
時間(s)
Na signal in flame AAS
300
Open the drain valve of the compressor
after measurement
測定後は必ずドレインを開けてください
Drain valve
Otherwise, oil and water vapor may be introduced into the AAS,
resulting in data imprecision and malfunction of the instrument.
ドレイン抜きをしないとオイルを含んだ水分が装置に入り、データ不良、故障の原因になります。
Graphite Furnace
Determination of Pb in tap water(bad example)
水道水中の鉛の定量(悪い例)
Determination of Pb in tap water in Hitachi Lab.
装置条件
温度プログラム
Component of a typical borosilicate glass
代表的なホウケイ酸ガラスの組成
Component
(%)
SiO2
80.90
B 2 O3
12.70
Al2O3
2.30
Na2O
4.00
K 2O
0.04
Fe2O3
0.03
other
0.03
Si, B, Al, Na, K, Fe
are contained in %
level.
Others are contained
in 0.03%.
Contamination form containers
容器からの汚染
(1) Grass wares should be avoided for inorganic
analysis of ppb (ug/L) or lower level.
μg/L以下の無機分析で.ガラス容器の使用は避ける。
(2) Plastic wares such as polypropylene (PP),
polyethylene(PE) are suitable. But blank level
should be checked before use because metal
oxides may be added as a plasticizer.
PP, PE等の樹脂製の容器が適する。 ただし、可塑剤に金属酸
化物が使われている場合があるので、ブランク値の確認をする。
(3) An ultrapure grade or EL-grade acid is
preferable to analytical grade or AAS grade.
特級、原子吸光用よりもUltrapureやELグレードの酸が望ましい。
Use of matrix modifier
0.15
Absorbance
As 10μg/L
+ Pd 100mg/L
0.10
0.05
As 10μg/L
0
500
1000
1500
Ashing temp.灰化温度
Effect of matrix modifier on arsenic absorbance
Asの吸光度に対するマトリックス修飾剤の効果
Use of matrix modifier
Modifier:Mg(NO3)2
without
modifier
with modifier
Atomization signal of Al in soil extract.
土壌溶出液中Alの原子吸光信号
Use of matrix modifier
元素
As
灰化
差
温度
Pb
灰化
差
温度
400
600
none
Pd・Mg 1,400 1,000 1,400
1,400 1,000 1,200
Ni
1,300 900 1,200
Mg
Se
灰化
差
温度
Sb
灰化
差
温度
400
600
800 600 200 1,000
600 1,400 1,000 1,400
600 1,600 1,200 1,200
Cd
灰化
差
温度
400
400 1,000
800 1,000
600 400
600
600
0
Modifier concentration
Pd・Mg:500 ppm
Ni: 1,000 ppm
Mg: 1,000 ppm
The matrix modifiers for AAS are
provided from Kanto Chemicals
and Merck.
Use of 1000ppm AAS Standard
Solution as a matrix modifier is not
recommended because they are
contaminated with some trace
elements.
Determination of Pb in tapwater
Sample preparation 試料の調製
1)Use acid-washed plastic 25mL volumetric flasks.
酸洗浄した樹脂製のフラスコ25mLを用意する。
2)Dispense 0.2mL of ultra-pure nitric acid into each flask.
高純度硝酸0.2mLを各フラスコに入れる。
3)Dispense 0, 25, 75, 125, 250uL of 1.0mg/L Pb into each flask,
and fill with DW. Use these solution as working standards.
Pbの標準液1.0mg/Lをマイクロピペッターにて0,25,75 ,125, 250μLを入れ、純水で全量を25mLとし、こ
れを検量線用試料とする。
4)Use a mixture of Pd and Mg(NO3)2 as matrix modifier (1000mg/L
each as Pd and Mg).
マトリックス修飾剤は Pd+Mg(NO3)2混合液(Pd、Mgそれぞれ1000mg/L)を用意する。
Analysis of data imprecision
データ不良の解析
Injection hole
試料注入口
Graphite tube
試料注入口
Light
光源
Electrode電極
Construction of electrothermal atomizer
電気加熱原子化部の構造
Factors for imprecision
再現性不良の原因
Sampling error
サンプリング不良
Contamination from
environment
周囲環境からの汚染
Contamination of rinse port
洗浄ポートの汚染
Wrong temperature program
(sample boiling)
温度条件不適(突沸)
Contamination of electrode ring
電極リングの汚染
Factors for imprecision
再現性不良の原因
Sampling error
サンプリング不良
Factors for imprecision
再現性不良の原因
Adjusting nozzle position
ノズル位置の調整
Caution on nozzle height
ノズル高さに注意
OK
NG
Set the nozzle height very
close to injection hole
Nozzle tip
上下位置はキュベット穴に対し
ぎりぎりにセットする
0~0.5mm
Better for
low volume
low viscosity
Nozzle height can be
set without Graphite
Tube Camera
グラファイト炉カメラなしでノズル高さが合
わせられます。
The same distance
Caution on sample volume
注入試料に注意
up to100μL
Simple Aqueous solution
単純な水溶液
up to 30μL
Organic solvent
有機溶媒
Factors for imprecision
再現性不良の原因
Wrong temperature program
(sample boiling)
温度条件不適(突沸)
DC Zeeman method can monitor all stages
in heat program
直流ゼーマン法は全加熱ステージをモニターできます。
Analysis of Juice sample
Hitachi DC Zeeman method enables monitoring for all stages (Drying, Ashing, Atomization),
and this helps operator to set up optimal analysis condition.
AC Zeeman method can monitor only Atomization stage.
Optimum temperature can be set without
monitoring inside of the graphite tube by camera
グラファイトチューブ内をカメラでモニターしなくても、最適な温度が設定できます。
Dry
Nozzle
injection
Ash Atomize
Sharp peaks cased by
sample sputtering
Atomization peak
Effects of Drying Temperature (too high) on Performance.
Sharp peaks caused by sputtering can be recognized in the beginning of the drying stage.
Factors for imprecision
再現性不良の原因
Contamination from
environment
周囲環境からの汚染
Contamination from environment can be reduced
by the standard equipped clean cover
標準装備のクリーンカバーにより環境からの汚染を低減できます
Exhaust duct
排気ダクト
Graphite furnace
電気加熱炉
Autosampler
オートサンプラ
Flame atomizer
フレーム部
Clean cover
クリ-ンカバー
Contamination from environment can be reduced
by the standard equipped clean cover
標準装備のクリーンカバーにより環境からの汚染を低減できます
Determination of aluminum
アルミニウムの定量
Contamination
汚染の混入
Clean cover open
Clean cover closed
クリーンカバー開放
クリーンカバー閉じる
Factors for imprecision
再現性不良の原因
Contamination of rinse port
洗浄ポートの汚染
time interval
洗浄ポート
Rinse port
Factors for imprecision
再現性不良の原因
Contamination of rinse port
洗浄ポートの汚染
Contamination of rinse port
洗浄ポートの汚染
1.Clean with cotton swab
綿棒で清掃
2. Add 100uL of conc.HCl and
leave for 2-3 min.
塩酸100μLを入れ2~3分待つ
3. Rinse with DW
純水で洗浄
Factors for imprecision
再現性不良の原因
Contamination of electrode ring
電極リングの汚染
Absorption signal does not appear
at the beginning of the atomization,
but increases at the end of
atomization and cleaning.
Electrode ring
電極リング
最大加熱をかけた場合、初期にはピークが出ず後半に
出現する
Exchange of electrode ring
電極リングの交換
Ar gas
Water in/out
O-ring
Be aware the o-rings not to be sandwiched between electrode and base.
The o-rings should be attached to the base side.
O-リングを電極とベースの間に挟まないように注意。O-リングはベース側につけておく。
Another factors for imprecision
再現性不良の他の原因
Quartz window
石英窓
突沸や過剰冷却により窓が曇りノイズが増え再現性が悪くなります。
Condensation by overcooling of the quartz window
過冷却による石英窓の結露
Cooling circulator
循環冷却機
Normal
Overcooled
正常
過冷却
Set around room temperature
室温±5℃程度に設定
Sample volume in the Sample Cup
サンプルカップの液量
Half cup of sample is enough. Too mach sample may cause carryover.
カップ半分で十分。多すぎるとキャリーオーバーの原因になります。
Contamination with matrix modifier
マトリックス修飾剤による汚れ
High conc. of matrix modifier
(eg. Pd/Mg 1000ppm) piles up
at the ends of the cuvette.
高濃度のマトリックス修飾剤を使用すると、
キュベットの末端に修飾剤が付着してきま
す。
Clean the end of cuvette
with a cotton swab
each 100 measurements.
約100測定毎に、綿棒などで
清掃します。
Contamination with matrix modifier
マトリックス修飾剤による汚れ
Matrix modifier piles up also at
the electrode rings.
Clean the ring by passing
a paper like Kimwipe.
電極リングにもマトリックス修飾剤が付着し
てきます。
キムワイプなどをリングに通して
清掃します。
Another factors for imprecision
再現性不良の他の原因
Bubbling in syringe pumps
気泡の発生
Loosening of screw
接続部の緩み
Deterioration of packing
パッキンの劣化
Autosampler syringe pump
オートサンプラーシリンジポンプ
Analysis of semiconductor reagent (organic Si)
半導体試薬(有機Si)の分析
1-Ca 422.7 nm
ABS
0.25
REF
0.10
BKG
0.05
0.20
0.00
Sample+Ca5ng/mL
0.15
Sample+Ca2.5ng/mL
Sample×5Dilution
0.10
Xylene
0.05
0.00
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
時間(s)
Avoid contamination from environment and containers.
The sample is diluted with Xylene.
Add hydrofluoric acid (HF) to remove Si.
Add tetrahydrofuran (THF) to improve HF and organic Si.
→Matrix modifier : HF/THF (1:1)
Only GFAAS can analyze the sample.
Z-2000 series New Platform Cuvette
EPA method compatible


Cumbersome insert operation is eliminated by built-in platform. 面倒な挿入作業が不要
Long life. 長寿命
Sensitivity is enhanced for some elements (1.5 times for Cu and Cr).
いくつかの元素では感度向上
0.2
O m ega type
Insert
Abs orba nce

surface deteriorated
C onventional
0.1
Conventional type
baseline unstable
0
0
200
400
600
800
1000 1200
1400 1600
1800 2000
Firings
Repeated measurement for Cr standard solution.
Note: The lifetime of the cuvette will be
variable according to samples and heat conditions.
The data shown above is only rough reference for
simple aqueous solution and is not certified value.
New Omega type
キュベットの寿命は単純な標準液での目安であり、保証値ではありません。
Know-how for a better AAS expert
更なる原子吸光スペシャリストへの測定のコツ
Thank you !
Download