TOC and Conductivity for USP and EP Water Systems Content 1. Conductivity 2. TOC analysis 3. Validating a TOC analyzer 4. Introduction to the PAT700 combined TOC and Conductivity Analyzer Dissociation and Conductivity • Conductivity measures ionic activity • Ionic compounds dissociate into ions, then re-associate back into the compound • E.g. Sodium Chloride (NaCl) made from Na+ and Cl• Add energy and the Na+ and Cl- split, or dissociate • The ions then contribute to the conductivity of the water • Some substances dissociate more readily and contribute more strongly to conductivity • Different temperature compensation curves for different waters Conductivity Conductivity Meter Cl- Na+ + + Measurement Cell Conductivity = TOC Conductivity varies with temperature Temperature USP<645> Conductivity & Temperature • Conductivity meters will report different results depending on: • The compounds present in the water • The temperature compensation curve selected • Hence pharmacopoeias call for un-compensated, e.g. USP<645>: Content 1. Conductivity 2. TOC analysis 3. Validating a TOC analyzer 4. Introduction to the PAT700 combined TOC and Conductivity Analyzer On-line TOC analysis – principle of operation Conductivity Meter - + + Measurement Cell UV lamp On-line TOC analysis – principle of operation Conductivity Meter Organic material turned to CO2 + ++ -- + UV UV On-line TOC analysis - Principle of Operation Conductivity Organics oxidised to CO2 d conductivity = TOC UV on Time Letter of the Law USP (‘till recently) “… share the objective of completely oxidizing the organic molecules in an aliquot of sample water to carbon dioxide…” EP states “… have in common the objective of completely oxidizing the organic molecules in the sample water to produce carbon dioxide…” Only Hach’s, End Point Detection method, strictly meets this requirement for online TOC analyzers Dynamic End Point Detection USP <643> Requirement: Instrument Must Completely Oxidize the Sample Content 1. Conductivity 2. TOC analysis 3. Validating a TOC analyzer 4. Introduction to the PAT700 combined TOC and Conductivity Analyzer Validating the TOC analyser 1. Validate that the instrument matches the URS 2. Validate that installation is as per manufacturer 3. Verify Calibration TOC Conductivity Temperature sensor 4. System suitability 5. Verify connections to remote systems 6. Performance qualification Calibration is stable Analyser works correctly Tests to validate the analyser 1. Validate that the instrument matches the URS 2. Validate that installation is as per manufacturer 3. Verify Calibration TOC Conductivity Temperature sensor 4. System suitability 5. Verify connections to remote systems 6. Performance qualification Calibration is stable Analyser works correctly Validation documents & procedures • Demonstrate URS requirements are met through, Commissioning, IQ and OQ • PQ demonstrates that the analyzer is stable and can support required calibration and system suitability intervals 6 monthly testing Week 1 IQ, Commissioning First 4 weeks Calibration stability, System suitability Tests to validate the analyser 1. Verify that the instrument matches the URS 2. Verify that installation is as per manufacturer 3. Verify Calibration TOC Conductivity Temperature sensor 4. System suitability 5. Verify connections to remote systems 6. Performance qualification Calibration is stable Analyser works correctly 1000 TOC Calibration – factory & on site 0 Reported TOC Factory Calibration 0 Injected TOC 1000 TOC Calibration on site TOC 750ppb 500ppb 250ppb Water system TOC value – modern system = 30 to 70ppb Certified traceable standards Calibration – As Found, As Left System Suitability Test Must Be Performed System Suitability is not calibration! System suitability is “the process of validating whether your system (i.e. TOC analyzer) is acceptable for providing useful analytical data without any bias.” This is typically done by: • Analyzing a material that is easy-to-oxidize (sucrose) • Analyzing a material that is difficult-to-oxidize (1,4 benzoquinone) • Calculating the ratio of the responses System suitability – USP<643> & EP2.2.44 System suitability calculation Sucrose analysis result = 475ppb TOC Benzoquinone analysis result = 485ppb TOC Calculation = Sucrose Benzoquinone = 475 = >85%<115% 485 Verifying conductivity – USP<645> 1. Verify meter accuracy (+/- 0.1mS) 2. Verify cell constant (+/-2%) 3. Verify temperature sensor (compensation) Temperature sensor verification Verifying conductivity – meter accuracy Accuracy of meter Verifying conductivity – meter accuracy Disconnect measurement cell Verifying conductivity – meter accuracy Meter = Resistor +/-0.1mS Replace cell with NIST traceable resistor Verifying conductivity – cell constant Function area and distance Area Distance Verifying conductivity – cell constant • Introduce certified conductivity standard • +/-2% of certified standard • Cell constant known within +/-2% Content 1. Conductivity 2. TOC analysis 3. Validating a TOC analyzer 4. Introduction to the PAT700 combined TOC and Conductivity Analyzer PAT700 TOC Analyzer with OASIS Onboard, Automated Standards Introduction System TM Introduction to the ANATEL PAT700 • On-line TOC analysis with complete sample oxidation • OASIS TM Onboard, Automated Standards Introduction System • Calibration and system-suitability test SOP built-in via color touch-screen • Individual 4-20mA outputs for TOC, temperature and conductivity • Main and standby auto-switching UV lamps with UV Detect™ technology • IP 56 stainless-steel enclosure improves protection from water and particulates PAT700 – Supports all test requirements • USP <643> Total Organic Carbon • USP <645> Conductivity • USP <788> Water for Injection • USP <789> Opthalmic Water • EP 2.2.38 Conductivity • EP 2.2.44 Total Organic Carbon • JP <60> • 21 CFR Part 11 TM OASIS Technology Saves time and delivers improved confidence in quality • All bottles for a test installed at once and tests automated • Information about each standard stored in RFID tag on the bottle – contents, concentration, certified value, lot number, expiration date – (Manual data entry possible for customer prepared standards) • Analyzer voids bottle following test • OASIS can be used for grab sample analysis Non-Hach Standards Warning Message If the user selects “No”, then the test is cancelled and the user is placed back at the Run Standards selection screen. If the user selects “Yes”, then the user is prompted to enter all of the bottle information manually and the test is allowed to continue as if the bottle did not have an RFID tag at all. PAT700 TOC Analyzer with OASIS Grab-sample analysis TM Grab Sample Analysis • Utilizes OASIS sample bay • Up to 4 grab sample bottles at a time • Grab Sample Bottle ID nine character alphanumeric – (ex. PW POU 01 ) PAT700 TOC Analyzer with OASIS OOS sample capture and root-cause analysis support TM OOS sample capture capability 1. Utilizes OASIS sample bay 2. OOS triggers automatic collection of two samples 3. Re-tests 1 bottle to confirm OOS 4. Saves one sample for analysis in laboratory to support root-cause analysis PAT700 TOC Analyzer with OASIS 21 CFR data export TM 21 CFR Part11 Data Export TOC Dataview PC Software Allows the viewing of encrypted data Allows decrypted data to be saved Encrypted file exported as filename.hef Read-only .pdf for record retention .txt files to load into LIMS database PAT700 TOC Analyzer with OASIS Clean-In-Place verification TM On-line Monitoring of CIP Final Rinse Water Verification Conductivity can be used for detecting acid wash residues TOC can be used as broad-spectrum test to detect presence of drug material residues Unit programmed to meet user’s specific criteria • Either conductivity threshold • Or time based After TOC analysis, reports results and return to idle mode On-line Monitoring of CIP Final Rinse Water Verification • Automated Mode (plc controlled): • Unit idle until called for • TOC analysis triggered via: • Digital input, or • Serial command, or • Modbus over Ethernet • Reports results and return to idle mode PAT700 TOC Analyzer with OASIS Laboratory TOC analysis TM PAT700 with auto-sampler • Autosampler for laboratory use can load up to 36 vials • e.g. point-of-use and rinse water samples.(0-2000ppb) • Provides a laboratory TOC analyzer without reagents or carrier gas PAT700 TOC Analyzer with OASIS Water loop monitoring and root-cause analysis support TM Dual-stream option PAT700 Two inputs • One on the PW/WFI loop start, one on the return – Brackets loop quality – Helps root-cause investigation to find contamination source, e.g. feed-water or point-of-use Bracketing POU for On-line Release Contamination Source ? Points-Of-Use (POUs) Contamination Source ? Pure Steam Generation Distillation - WFI PAT700 TOC Analyzer with OASIS 12-month service and calibration intervals TM UV DetectTM Diagnostics • Direct measurement of UV source performance – Diagnosis of UV status – Fast, real-time lamp feedback On-line: Verified after every TOC analysis Off-line: Can be conducted in diagnostic mode after lamp replacement Detects rouge contamination • Provides confidence for on-line water release – Standby UV lamp automatically switched on if main lamp fails – avoids gaps in analysis data • More reliable than simple “hours-of-operation” counters • Reduces cost-of-ownership – analyzer can support 12-monthly calibration and service intervals Questions? PAT700 TOC and Conductivity Analyzer