Chromatographs eluent tank detector pump PC injector column CHROMATOGRAM • Qualitative & • Quantitative information GasChromatography (GC) 1952: A.T. James & A.J.P. Martin •1956: van Deemter: kinetic theory •M. Golay: capillary columns High performancy Qualitative & Quantitative information Complicated samples Separation MOBILE PHASE: GAS STATIONARY PHASE: solid or liquid on solid support (GSC, GLC) COLUMN ELUTION TECHNIQUE Base of separation: 1. Boiling point (vaporization) 2. Structure GASCHROMATOGRAPHY: analysis in vapor phase ~12 billion organic compounds ~ 50 000: evaporative without destruction Evaporization depends on: •Molar mass •polarity Thermal stability GASCHROMATOGRAPHY (GC) Sample introduction to the mobile phase: gas/vapor Sample can be: 1. gas 2. liquid: vaporization 3. solid: dissolution in liquid Pressure and flow regulators Gas tank Gas cleaner Gaschromatograph (GC) GASCHROMATOGRAPH (GC) injector detector cleaner PC column thermostate Pressure controller Flow controller Gas tank Eluent gas Name pure Very pure Ultra pure sign % ppm 2.5 99,5 5000 3.0 99.9 1000 3.5 99,95 500 4.0 99,99 100 4.5 99,995 50 5.0 99,999 10 6.0 99,9999 1 7.0 99,99999 0,1 Depending on the type of detector: •H2 •Ar •N2 •He Reductor valve: Type depends on the quality and pressure of the gas Inside apparatus: Pressure and flow controllers Flow-rates Sample introduction 1. Injection in a very short time 2. Vapor/gas phase 3. Mixible with eluent gas Syringe For gas & liquid sample volume 0,1 l-1 ml Liquid vaporization: 100-10000 X volume increase „six-port” valve FLASH INJECTOR Septum (rubber) 1. Samle introduction 2. Vaporization 3. Inlet to column Eluent gas inlet Heating block (25 – 300 oC) liner (glass) column Packed columns: greater diameter: greater sample volume Capillary columns: small sample volume Flash injector Injection 1. Stick needle into the septum 2. Push the syringe piston 3. Remove syringe vaporization 1. Sample vaporization 2. Liquids: 100 – 1000 X volume increase 3. Mixing with eluent Quick injection solvent Slow injection Eluent gas moves the sample to the column. Type of injectors Split-injector •SPLIT •SPLITLESS •ON-COLUMN •PTV Carrier gas Septum wash split-gas Split/splitless ratio: determines amount of sample moving to the column 5:1 200:1 Splitless injector Purge Off Purge On On-column PTV (Programmed Temperature Vaporizer) Injection directly to the column Columns Capillary polyimid, 350 oC quarz d Stationary phase microbore: d < 150 m standard capillary: 150 m < d < 500 m widebore: d > 500 m Adsorption mechanism: PLOT (Porous Layer Open Tubular) Distribution mechanism: WCOT: Wall Coated OT SCOT: Support Coated OT Interaction: between stationary phase and sample Active side: silanol groups SiOH SiOH •„tailing” •Non-symmetric peaks SiOH SiOH SiOH SiOH Quartz surface desactivation: sylil reagents Si-O-Si(CH3)3 Si-O-Si(CH3)3 SiOH Si-O-Si(CH3)3 Stationary phases I. Thermal stability No „bleeding” Known chemical structure Chemical inertnees Low price Adsorbents (GSC) porous, with large special suface Organic adsorbents: •active carbon •polymers inorganic adsorbents: •silicagel •aluminium-oxide •zeolits (molekulasziták) modified adszorbents: Based on carbon or silicagel Analytes: Hydrocarbons with small molar mass, He, Ne, Ar, Kr, Xe (PLOT) Stationary phases II. (GLC) (absorption: dissolution of gas and liquids in liquids) Polymers: WCOT: polymers on the surface of capillary) Relative small number: 12-15 substituted polysiloxans (silicons): long lifetime R Si R R: substituents on polysiloxans Thermal stability: up 250-300 C R O Si O R Substituents:: Methyl phenyl Cianopropyl Trifluoropropyl n Methyl: -CH3 Phenyl: Cianopropyl: -CH2CH2CH2CN Trifluoropropyl: -CH2CH2CF3 O CH 3 Si CH 3 O Si O CH 3 Si CH 3 O CH 3 Si CH 3 •methyl-phenyl •cianopropyl-phenyl •etc. substitution: how much % of Si atoms 100 % metil 5 % fenil & 95 % metil Polyethyleneglycols (PEG) HO CH2 O CH2 O H n Carbowax Special separation Disadvantage: •Lower thermal stability •„oxygen-sensitivity” Polarity of stationary phase: •Structure of stationary phase •Quality of functional groups •Number of functional groups Apolar stationary phases: • 100 % methyl • 5 % phenyl Midium polar phases: •35 % phenyl •50 % phenyl Polar phases: •cyanopropyl •PEG Selectivity depends on: the interaction between stationary phase and analyte Interactions depend on: •Quality of analytes •Structure of stationary phase Thermostate Type of working: •Izotherm •Programmable heating column T (oC) thermostate •Large temperature range -50 – 400 C •Programmable heating: 0- 40 oC/min •„cooling” Decrease of analysis time Good peak shape t (min) Detectors Quantitative analysis: signal of detector is proportional with concentration of analytes in detector universal: signal for every compounds selective: signal for a groups of compounds specific: signal for special compounds destruktiv non destruktiv Dinamic range: change of concentration results a change in signal linearity: T= mc (deviation < 5 %) sensitivity: m (ratio of signal/concentration) Limit of detection (LOD): signal to noise ratio: 3 Limit of quantitation (LOQ): signal to noise ratio: 10 Detectors Thermal Conductivity Detector (katharometer) Change of impedance Wheatstone-bridge W-filaments: 100-200 mA heating current non destructív universal dinamic range: 105 LOD: 5-50 ng Carrier gas: H2, He N2 Flame-ionization detector (FID) hydrogen/air microburner with a pair of electrodes Carrier gas: non ionizable gas: N2, Ar, He, H2 Organic compounds leaving the column are burning in burner jet, ions are forming Ions result a small current Carbon-detector: it is good for organics, except formic acid destructív Dinamic range: 105-106 LOD: 0,05-0,5 ng High Performance Liquid Chromatography (HPLC ) Mobile phase: liquid Stationary phase: adsorbent (LSC) or liquid on a support(LLC) Column Elution technique Sample: liquid eluent tank Gas removal column thermostate pump detector injector PC HPLC Gas removal pump automated injector detector (thermostate) Eluent Should (have) be: Low viscosity inert: no reaction with analytes Chemical stability No corrosion No toxycity Higher boiling point Low price Good quality and purity Compatible with detector UV-absortion: low purity: HPLC grade Water and buffers too !!! Eluent Analytes distributed between stationary and mobile phase: interaction of analytes with both phases Polarity of molecule & mobile phase & stationary phase hexane chloroform tetrahidrofuran acetonitrile isopropanol ethanol methanol water P O L A R I T Y Change of polarity: •Change of quality of mobile phase •Mixing of solvents Mixed solvents: should be mixcible Eluent strength: determined on silicagel on the bease of heat of adsorption of solvents izoeluotrope mixture: eluent strength is the same: k’, Rs: may change !!! Izocratic elution: fixed mobile phase composition Gradient elution: eluent strength is increasing in time Use of buffers: adjusting of pH in the case of analysis of ionisable components Pumps To carry of eluent Should be: •pressure (400 bar) •Stable flow-rate •Compatible with different solvents:no corrosion •Small hold-up volume •No pulsation Flow-rates in classical analytical HPLC: 0,1-1,5 ml/min (0-5 ml/min) Syringe-type pump Reciprocating pump Pulsation: double pistons (phase-deviation) Volume: 10-100 l Change of flow rate: easy V time Gas removals Liquids: contain dissolved gases Effect of gas bubbles: In pump: •Pressure pulsation •Different flow-rates •Mechanical instability In detector: •Increased noise (retention time changes) Remove of gas from the solvent: Ultrasound: •cheap •Non effective Vacuum: •Higher price •effective He-purge: •Higher price •effective Sample loading 1. Quick 2. Sample should be mixable with eluent Sample volume: 10-50 l Micro syringe: „Six-port” valve Columns Function: separation Liquid chromatography: NP LC: Normal Phase RP LC: Reversed Phase NPLC: polarity of stationary phase > polarity of mobile phase RPLC: polarity of stationary phase < polarity of mobile phase Material of column: •Stainless steel •Glass •PEEK (poly(ether-ether-ketone) Size of column: •diameter: 2-5 mm •length: 5-25 cm Packing: regular spherical Modified silica gel OH OH SiO2 OH OH OH Modifying groups: C18: octadecyl: C18H37 C8: octyl: C8H17 C4: buthyl: C4H9 Amino: CH2CH2CH2NH2 Ciano: CH2CH2CH2CN Phenyl: C6H5 RPLC: C18 stationary phase & methanol/water mobile phase NPLC: silicagel stationary phase& hexane/alcohol mobile phase Guard column: avoid contamination of analytical column Detectors Quantitative analysis: signal of detector is proportional with concentration of analytes in detector universal: signal for every compounds selective: signal for a groups of compounds specific: signal for special compounds destruktiv non destruktiv Dinamic range: change of concentration results a change in signal linearity: T= mc (deviation < 5 %) sensitivity: m (ratio of signal/concentration) Limit of detection (LOD): signal to noise ratio: 3 Limit of quantitation (LOQ): signal to noise ratio: 10 UV-Vis spectrophotometer Application: UV-Vis range Lambeert-Beer: A = ε c l Measuring side splitter cuvette rés fényforrás I0 I I0 I0 Reference side monocromator Light source: UV: deuterium lamp Vis: wolfram lamp Detector: fotodiode Most usable HPLC detector 190 nm < < 800 nm D E T E C T O R Cuvette: quartz l=5-10 mm A = lg I0/I Dioda Array Detector (DAD) polychromator Light source lence cuvette Diode array Advantage: Spectra and chromatogram at the same time Paper and thin-layer chromatography Planar arrangement Stationary phase: paper silica gel or aluminium-oxide on a glass plate Evaluation of chromatogram: Dropping liquid sample on the one edge of the plate with capillary Evaporation (drying) the solvent Place the plate to the closed container saturated with vapors of developing solvent Running of analytes: based on capillary activity After development of chromatogram, remove plate from container and dry it Locating analytes on the plate: spraying with chemical reagents, like iodine, sulfuric acid or UV-light Selection of mobile phase:like in Normal Phase HPLC Qualitative data: retardation factor (Rf) Quantitative data: intensity of spots Advantages: •simple •cheap