www.ssi.shimadzu.com EDX-720 Training E nergy Dispersive X-ray Energy Dispersive X-ray Fluorescence Spectrometer Fluorescence Spectrometer www.ssi.shimadzu.com An Atom Electrons Nucleus: K L M Orbital or shell Protons Neutrons www.ssi.shimadzu.com Atoms & Energy Atoms can absorb and release energy using electrons Atoms absorb energy by promoting an electron to a higher energy orbital K L M Atoms release energy by demoting electrons to a lower energy orbital Energy www.ssi.shimadzu.com What are X-Rays?.......Energy www.ssi.shimadzu.com Two types of X-Rays Continuous X-rays High energy Broad spectrum Energy in vs. Characteristic X-rays Low energy Discrete spectrum Energy out www.ssi.shimadzu.com Two types of X-Rays DETECTER Energy Released from the System (Characteristic X-ray) K L M Energy Put into the System (Continuous X-ray) X-ray Tube (Rh) www.ssi.shimadzu.com Characteristic X-Rays Types of transitions www.ssi.shimadzu.com X-Ray Tube and Spectra How X-Rays are generated www.ssi.shimadzu.com SiLi Detector www.ssi.shimadzu.com Filters Rh creates it own spectra MoNi Ag Mo Ti EDX-720 includes: Ag, Mo, MoNi, Ti Sensitivity is low in Introducing a filter reduces the Rh background these areas Increases Sensitivity! ! ! www.ssi.shimadzu.com Filters (Cont.) www.ssi.shimadzu.com Filters (Cont.) Without primary filter With primary filter • X-ray tube consist of characteristic X-ray (Rh) and continuous X-rays. • Continuous X-rays are background • Filters decrease background www.ssi.shimadzu.com Intensity vs. Matrix Intensity and Density are inversely related Density affects the analyte’s ability to Fluoresce through the matrix www.ssi.shimadzu.com Measurement Atmosphere Air, Helium, or Vacuum: Elements in the range of Na – Sc Measurement under Helium or Vacuum • Allows analysis of light elements whose x-ray emissions may be absorbed by air, and • Increasing sensitivity for light elements Solid - Air, Vacuum, or Helium Liquid - Air, Helium Powder - Air, Helium www.ssi.shimadzu.com Types of Analyses Two Types of Analyses: Qual/Quant Quantitative Analysis Conditions: FP (Fundamental Parameters Calibration Curve Sample Form: Bulk Films Filter Paper www.ssi.shimadzu.com Hardware Accessories Collimator Exchanger 10 mm 5 mm CCD Camera X,Y Positioner X Y 3 mm 1 mm www.ssi.shimadzu.com Sample Turret • Up to 16 samples • Analysis in air • Analysis under He/Vacuum for lighter elements • Continuous measurement using a sample turret www.ssi.shimadzu.com Sample Films and Containers Film (Covers)