Evaluation of AcurosXB deterministic algorithm for heterogeneous dose calculation in lung cancer with RPC thorax phantom Tao Han1, Firas Mourtada1,2, Roman Repchak1, Jacqueline Tonigan1, Justin Mikell 1, Rebecca Howell1, Mohammad Salehpour1, Andrea Molineu1, and David Followill1 1Department of Radiation physics, UT MD Anderson Cancer Center; 2 Department of clinical physics, Helen F. Graham Cancer Center Introduction AAA AAA AXB_mm Axial Modern radiation therapies such as intensity-modulated radiation therapy (IMRT) and volume modulated arc therapy (VMAT) demand from dose calculation algorithms higher accuracy and computation speed Sagittal Although the MC method can be considered as the gold standard in accuracy given sufficient particle histories, calculation times may not be short enough for clinical use with these advanced techniques. Recently, AcurosXB (AXB), a novel deterministic method based on the grid-based Boltzmann transport equation solver (GBBS), was introduced for external radiotherapy dose calculation and has shown poentials to improve the dose predictions over currently widely used convolution methods in heterogeneous media AXB_mm Coronal Fig 3.1: IMRT Fig 3.2: RapidArc Fig 3: Distribution of gamma index with a 5%/3mm criteria between film measurements to TPS calculations for IMRT (Fig 3.1) and RapidArc (Fig 3.2). Film position Fig 1: Screen capture from Eclipse TPS depicting the RPC thorax phantom, structure contours (heart, lung, cord, PTV, film inserts), and one CT slice depicting the locations of TLD. The goal of this study was to verify the dosimetric performance of AXB in IMRT and VMAT plans of lung cancer, in which the lung tissue heterogeneity may plays important role in dose calculation. We compare the AXB dose prediction with measured data from both TLD and film. We also compared with the Anisotropic Analytical Algorithm (AAA). Dose grid size: 0.1x0.1x0.1 cm3 AAA AXB_mw AXB_mm a) IMRT plan Axial Sagittal Coronal b) RapidArc Axial Sagittal Coronal Dose grid size: 0.3x0.3x0.3 cm3 AAA AXB_mw AXB_mm 89.2 90.4 91.5 95.1 91.1 96.7 95.7 91.4 96.1 89.9 90.5 91.6 95.2 92.1 96.5 95.6 92.8 96.2 87.2 88.3 89.5 91.6 92.2 94.2 92.1 92.6 94.5 87.6 88.9 89.7 91.8 92.4 94.1 92.4 92.9 94.7 Table 2: Percent of points passing gamma analysis with acceptability criteria of 5%3 mm. IMRT Methods & Materials • RPC thorax phantom • 4 TLD tublets • 3 EBT2 films (axial,sagittal, coronal) • Varian Eclipse TPS 11.0 • AAA 10.0.24 Fig 2: IMRT (left), RapidArc (right) plans and dose distributions • AXB 11.0.03 in axial and sagittal views. • Dose-to-water in medium (AXB_mw) IMRT • Dose-to-medium in Measure Dose grid size 0.1x0.1x0.1 cm Dose grid size 0.3x0.3x0.3 cm TLD medium (AXB_mm) ment position AAA AXB_wm AXB_mm AAA AXB_wm AXB_mm • Clinically equivalent IMRT and PTV_Inf (cGy) 642.8 -4.3 -1.6 -2.6 -4.0 -1.1 -2.4 PTV_Sup 642.0 -4.5 -1.6 -2.7 -4.2 -1.4 -2.4 VMAT (RapidArc) plans were Heart 175.0 3.8 3.3 3.1 4.3 4.0 4.2 generated on Eclipse Cord 139.5 -2.8 -0.5 -0.5 -2.5 -0.4 -0.4 RapidArc • 9 fields IMRT PTV_Inf 598.0 -3.2 -0.1 -1.7 -4.0 -0.2 -2.2 • 2 arcs RapidArc PTV_Sup 604.0 -2.7 0.5 -0.9 -2.5 1.0 -0.8 Heart 87.7 -5.3 -1.1 -2.5 -5.6 -0.8 -2.7 • Each plan delivered 3 times 98.7 -5.8 -3.8 -3.8 -6.4 -4.4 -4.4 • Dose grid sizes: 0.1x0.1x0.1 & Cord Note: percentage difference = (TLD-calculation)/TLD*100 0.3x0.3x0.3 cm3 • In-house gamma analysis Table 1: Percentage differences of AAA, AXB_mm, and AXB_wm software for TLD dose measurements in (a) IMRT and (b) RapidArc plan. 3 size do not improve the agreement to TLD data. Fig. 3 shows the distribution of gamma index with 5%/3mm criteria for first delivery of IMRT and RapidArc plans. The averaged gamma analysis for all three deliveries were summarized in Table 2. The AXB_mm gives the best agreement to film (all over 90%), while some of AAA predictions did not pass the 5%/3mm criteria. Fig.4 shows the comparison of DVH. Their differences are within 1% for normal tissue and 2% for PTVs. Table 3 shows the computation times. The AAA and AXB computation times were comparable for IMRT but AXB was 4-6 times faster than AAA for RapidArc plan. Conclusions The AXB was determined to be accurate using the RPC thorax phantom measurements and in equal or better agreement to both TLD and films than AAA. AXB dose-to-water in medium and AXB dose-to-medium in medium showed similar agreements to TLD and film measurements. AXB shorts the computation time 4 times over AAA for RapidArc plan. AXB shows promise for future dose calculations.oth in both accuracy and computation speed for lung cancer. Acknowledgements RapidArc Fig 3: Comparison of DVHs calculated by AAA, AXB_mm and AXB_mw for IMRT and RapidArc plans. 3 IMRT plan RapidArc Dose grid size: 0.1x0.1x0.1 cm3 AAA AXB_mw AXB_mm 8.7 15.8 15.8 75 21.8 21.7 Dose grid size: 0.3x0.3x0.3 cm3 AAA AXB_mw AXB_mm 2.3 3.1 3.1 16 4.2 4.3 Note: unit is in minutes Table 3: The computation times of AAA and AXB for IMRT and RapidArc Plans. National Institutes of Health grant 2R44CA105806-02, CA010953 and MDACC Support Grant CA016672 References 1.Wareing, T., J. Morel, and J. McGhee, Coupled electron-photon transport methods on 3-D unstructured grids. Trans Am Nucl Soc, 2000. 83: p. 240-242. 2.Gifford, K.A., et al., Optimization of deterministic transport parameters for the calculation of the dose distribution around a high dose-rate 192Ir brachytherapy source. Med Phys, 2008. 35(6): p. 2279-85. Results 3. Vassiliev, N.O., et al., Validation of a new grid- based Boltzmann equation solver for dose calculation in radiotherapy with photon beams. Phys. Med. Biol. 2010. 55, 581-598 Table 1 shows the comparison between TLD measurements with the calculated dose from AAA, AXB_mw, and AXB_mm. All of AAA and AXB_mm are within 5% except for the RapidArc cord position; dose calculation with smaller grid 4. Han, T., et al., Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media. Med Phys, 2011. 38. 2651-2663