Electron Paramagnetic Resonance Biodosimetry in Teeth and Fingernails A. Romanyukha1,2, R.A. Reyes2, F. Trompier3, L.A. Benevides1, H.M. Swartz4 1Naval Dosimetry Center, 8901 Wisconsin Ave., Bethesda, MD, 20889, USA, Services University, 4301 Jones Bridge Rd., Bethesda, MD, 20814, USA, 3Institut de Radioprotection et de Sûreté Nucléaire, Fontenay-aux-roses, France, 4Dartmouth Medical School, Hanover, NH, 03755, USA 2Uniformed Outline • • • • • EPR dosimetry basics In vitro X and Q dosimetry in tooth enamel In vivo tooth L-band dosimetry EPR dosimetry in fingernails Conclusions What is Electron Paramagnetic Resonance (EPR) ? • Non-destructive magnetic resonance technique used to detect and quantify unpaired electrons. • Absorption of ionizing radiation generates unpaired electrons (i.e., paramagnetic centers). • The concentration of radiation-induced paramagnetic centers is proportional to the absorbed dose. EPR: Fundamentals and Principles • There is a net absorption of energy from the microwave field at resonance because of a greater population of electrons are in the lower energy state. • The process is nondestructive because the population difference reestablishes itself after the microwave field is turned off. • Thus, the history of radiation exposure is not destroyed by EPR measurements. Optical Imaging Electron Resonance Typical frequencies and wavelengths required for resonance of a free electron in EPR measurements Mw Frequency, Band GHz Magnetic Sample size field, T L 1.5 0.05 Small animals, whole human teeth, fingers in situ S 3.2 0.11 Whole teeth, fingers X 9.5 0.33 30 - 1000 mg (solid) K 20 0.70 10 – 30 mg (solid) Q 35 1.22 2 – 10 mg (solid) W 95 3.30 0.25 – 1 mg (solid) EPR dosimeters for partial body exposure Radiation-induced radicals are stable only in hard tissues: teeth, bone, fingernails and hairs. Depending on mw band EPR can be measured in vivo or in vitro using specially prepared samples from human hard tissues Finger- and toenails Finger- and toenails Characteristics of EPR dosimetry Non-invasive Based on a physical process Not affected by biological processes such as stress Not affected by simultaneous damage that is likely to occur with irradiation such as wounds & burns Applicable to individuals Measurements can be made at any interval after irradiation up to at least 2 weeks (fingernails) or indefinately (teeth) Can provide output immediately after the measurement Unaffected by dose rate Can operate in a variety of environments Systems can be developed so that they can be operated by minimally trained individuals In vitro measurements in tooth enamel samples (X and Q-bands) Extracted teeth can be available for in vitro EPR measurements Validation and Standardization Four successful International Dose Intercomparisons with totally more than 20 participating labs ICRU, 2002. Retrospective Assessment of Exposures to Ionizing Radiation. Report 68 (Bethesda, MD: ICRU). IAEA, 2002. Use of electron paramagnetic resonance dosimetry with tooth enamel for retrospective dose assessment. International Atomic Energy Agency, Vienna, IAEA-TECDOC-1331. EPR dosimetry with teeth is the only method which can reconstruct external gamma radiation doses (<100 mGy) individually. Steps of the method • • • • Tooth collections Tooth enamel sample preparation EPR measurements of radiation response Calibration of EPR radiation response EPR Biodosimetry (Teeth) EPR Biodosimetry (Teeth) • Hydroxyapatite constitutes: – ~95% by weight of tooth enamel – 70-75% of dentin – 60-70% of compact bones Romanyukha, et. al, Appl. Radiat. Isot. (2000) and IAEA-TECDOC-1331 EPR Biodosimetry (Dose Calibration) EPR Biodosimetry Applications (Epidemiological Investigations Using Tooth EPR) Description of group Survivors of abombing of Hiroshima, Japan Year of overexposure Number of reconstructed doses Values of reconstructed doses, Gy Reference 1945 100 0.3-4.0 Gy Nakamura et al., Int. J. Radiat. Biol. 73, 619-627 (1998) Mayak nuclear workers, Russia 1948-1961 ~100 0.2-6.0 Gy Wieser et al., Radiat. Env. Biophys. 2006 Romanyukha et al., Health Phys. 78, 1520 (2000) Techa riverside population 1948-1958 ~100 0.1-10 Gy Romanyukha et al., Health Phys., 81, 554-566 (2001) Romanyukha et al., Radiat. Environ. Biophys., 35, 305-310 (1996) Eye-witnesses of Totskoye nuclear test, Russia 1954 10 0.1-0.4 Gy Romanyukha et al., Radiat. Prot. Dosim., 86, 53-58 (1999). Chernobyl clean up workers, Ukraine 1986 660 0 - 2.0 Gy Chumak et al., Radiat. Prot. Dos. 77, 9195 (1998) Population of areas contaminated by Chernobyl fallout, Russia 1986 2500 ~ 0.1 Gy Stepanenko et al., Radiat. Prot. Dos. 77, 101-106 (1998) Semipalatinsk population 1950s 32 0.3-4.0 Gy Romanyukha et al., Appl. Mag. Res.., 22, 347-356 (2002) Conclusion • EPR X-band (9 GHz) dosimetry in tooth enamel works excellent (LLD<100 mGy, time after exposure when dose measurements are possible from 0.01 hr to 106 yr. • But it requires to have extracted or exfoliated teeth available for preparation of tooth enamel Alternatives to exfoliated/extracted teeth L-band (1.2 GHz) non-Invasive in vivo measurements Q-band (35 GHz) measurements in enamel “biopsy” samples (~2 mg) with followed up tooth restoration Q-band (35 GHz) measurements in enamel “biopsy” samples (~2 mg) with followed up tooth restoration Description of Q-band feasibility test Tooth enamel powder samples for test: 0; 0.1 Gy; 0.5 Gy; 1 Gy; 3 Gy; 5 Gy Each sample was recorded 3 times in X (100 mg) and Q bands (2, 4 mg) Recent publication Romanyukha A. et al. Q-band EPR biodosimetry in tooth enamel microprobes: Feasibility test and comparison with X band. Health Physics. 93, 631-635, (2007). X-band spectrum vs Q-band spectrum X-band (100 mg), 0.1 Gy Q-band, (4 mg) 0.1 Gy 0.08 0.01 0.00 0.04 EPR signal, a.u. EPR signal. a.u. 0.06 0.02 0.00 -0.02 -0.01 -0.02 -0.04 -0.03 -0.06 -0.08 3490 3500 3510 3520 Magnetic field, G 3530 3540 -0.04 12060 12080 12100 12120 12140 12160 12180 12200 Magnetic field, G 1. Q-band has significantly lesser amount of the sample required for dose measurements 2. Q-band has significantly better spectral resolution of dose response 12220 Dose dependence: X vs Q 0.12 X-band, 100 mg sample 1.2 Q-band, 4 mg sample 0.10 EPR radiation response, a.u. EPR radiation response, a.u. 1.0 0.8 0.6 0.4 0.2 0.0 0.08 0.06 0.04 0.02 0.00 0 1 2 3 Radiation dose, Gy 4 5 0 1 2 3 Radiation Dose, Gy 4 5 Dental Biopsy Technique With the enamel biopsy technique a small enamel chip is removed from a tooth crown with minimal damage to the structural integrity of the tooth. A high-speed compressed-air driven dental hand piece is used with appropriate dental burs for this purpose. Standard techniques for tooth restoration using light-cured composite resins rapidly restore the small enamel defect in the biopsied enamel surface of the crown. Whole Tooth Biopsy Preliminary study on discarded teeth have demonstrated the feasibility of removing 2 mg enamel chips, the desired size for sufficient sensitivity In collaboration with B. Pass, P. Misra, with Q-band EPR dosimetry. T. De (Howard University) Q-band biopsy experiment • Tooth enamel biopsy sample 2.2 mg was irradiated 4 times to the same dose - 4.3 Gy • After each irradiation angle dependence (12 positions) of biopsy sample was studied • Using average, maximum, minimum and median values of EPR radiation response at each dose (e.g. 4.3, 8.6, 12.9 and 17.1 Gy) and linear back extrapolation attempt to reconstruct dose of 4.3 Gy was made Angle dependence of radiation response 0.40 Dose = 8.6 Gy EPR peak-to-peak ampl., a.u. 0.35 0.30 0.25 0.20 0.15 0 50 100 150 200 Angle, degree 250 300 350 Possible approaches: 1. Use average value of radiation response at each dose; 2. Use maximum value of radiation response at each dose; 3. Use minimum value of radiation response at each dose; 4. Use median value of radiation response at each dose. Spectra in biopsy sample at different doses and dose dependences 0.4 4.3 Gy 8.6 Gy 12.9 Gy 17.1 Gy 0.2 0.1 0.0 -0.1 Average Maximum Minimum Median 0.50 0.45 Radiation response, a.u. EPR dose signal, a.u. 0.3 0.55 0.40 0.35 0.30 0.25 0.20 0.15 -0.2 0.10 12100 12150 12200 Magnetic field, G Appearance of tooth enamel spectrum (maximum) of the same biopsy sample 2.2 mg at different doses -2 0 2 4 6 8 10 12 14 Dose, Gy Dose dependences for average, maximum, minimum and median values of radiation response at each dose Results of attempt to reconstruct 4.3 Gy in biopsy sample (2.2 mg) using different approaches Approach Average values Result of linear back extrapolation 5.5 ± 0.8 Gy Maximum values 7.3 ± 3.6 Gy Minimum values 5.4 ± 0.7 Gy Median values 5.4 ± 1.4 Gy Preliminary conclusions • Tooth enamel biopsy spectra have slightly different shape from powder spectra, they are more narrow and have higher signal-tonoise ratio for the same dose than powder spectra. However existence of angle dependence for biopsy spectra makes difficult dose reconstruction. Possible solution is to use average, maximum, minimum or median values for each dose for dose reconstruction • Use of average and minimum EPR radiation response values gives the best results to reconstruct 4.3 Gy, e.g. 5.5 ± 0.8 Gy and 5.4 ± 0.7 Gy, respectively • A possible reason for some dose offset (~1 Gy) is a slope of a base line of the spectra for this sample • A possible solution is to apply base line correction to spectra before measurements of peak-to-peak amplitude of radiation response L-band in vivo Recent publications • • • Swartz H.M. et al. Measurements of clinically significant doses of ionizing radiation using non-invasive in vivo EPR spectroscopy of teeth in situ. Appl. Radiat. Isot. 62, 293-299 (2005) Swartz H.M. et al. In Vivo EPR Dosimetry to Quantify Exposures to Clinically Significant Doses of Ionizing Radiation. Radiat. Prot. Dosim. 120, 163-170 (2006). Swartz H.M. et al. In Vivo EPR for Dosimetry. Radiat. Meas. 42, 1075-1084, (2007). • L-band (1 GHz) of microwaves is better for realization of in vivo EPR than standard Xband (9 GHz) because it has • Greater tolerance for the presence of water • Relatively large sample volume sufficient for whole tooth. Components of in vivo EPR spectrometer • Resonators that will probe teeth in vivo • Magnet system that can comfortably and effectively encompass the human head • Software for EPR dose response determination • Dose calibration for in vivo L-band measurements Clinical EPR Spectrometers MAGNETIC FLUX LINES PATIENT MAGNET FIXTURES SPHERE OF HOMOGENEITY MAGNET COILS Retrospective Radiation Dosimetry In Vivo EPR Radiation Dosimetry Under practical conditions with an irradiated tooth in the mouth of a volunteer, the dose dependent signal amplitude is clearly observed. (Acq. time = 4.5 minutes/spectrum) EPR Dose Response 1.5 Slope=0.077, SEP=2.41Gy Slope=0.073, SEP=1.19Gy Slope=0.081, SEP=2.67Gy C #27 #11 P-value difference = 0.4 C #11 0.5 1.0 #22 C #21 #27 Patient V107 Patient V110 C #22 0.0 Average P2P RIS EPR signal in canine teeth 2.0 Dose-response relationship for two head-and-neck radiation patients 0 5 10 15 20 R adiation dos e giv en, Gy 25 30 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.0 Averaged over 3 days tooth-size adjusted P2P SE dose prediction = 46 cGy 0 2 5 10 15 30 Radiation given, Gy Dose-dependence for 6 in vivo teeth, with each tooth irradiated to a different dose and measured on 3 separate days. Linear regression analysis shows that the standard error of dose prediction is ± 46 cGy. EPR biodosimetry in tooth enamel for partial body dose assessment • X-band EPR is ready to use for forensic dose assessment. Could be carried out on compact and transportable (< 150 kg) EPR spectrometer. Dose level <100 mGy. • Q-band biopsy potentially is able to measure doses < 500 mGy in biopsy tooth enamel samples 2-4 mg. • L-band in vivo EPR potentially is able to measure doses as low as 3 Gy. Needs some additional development. Finger-and toenails facts The major component of fingernails is a a-keratin. This protein is built up from three, long a-helical peptide chains that are twisted together in a left-handed coil, strengthened by S – S bridges formed from adjacent cisteine groups. •Typical available amounts of nail parings are up to 120 mg for fingernails and up to 160 mg for toe nails • Nails grow all the time, but their rate of growth slows down with age and poor circulation • Fingernails grow at an average of onetenth of an inch (3 mm) a month. It takes 6 months for a nail to grow from the root to the free edge • Toenails grow about 1 mm per month and take 12-18 months to be completely replaced • The nails grow faster on your dominant hand, and they grow more in summer than in winter Recent development Romanyukha A. et al. EPR dosimetry in chemically treated fingernails. Radiat. Meas. 42, 1110-1113, (2007). Trompier F. et al. Protocol for emergency EPR dosimetry in fingernails. Radiat. Meas. 42, 10851088, (2007). Reyes R.A. et al. Electron paramagnetic resonance in human fingernails: the sponge model implication. To be published in Radiat. Env. Biophys. (2008) New insights in EPR fingernail dosimetry • Fingernails can be considered as a sponge-like tissue which behaves differently from in vivo fingernails when mechanically-stressed after clipping • Most of previously published results on EPR fingernail dosimetry were obtained on stressed samples and not applicable to life-scenario situation • Unstressed fingernails have more significantly stable and sensitive radiation response which can be measured with EPR Radiation-induced signal in unstressed fingernails 1 Gy 5 Gy 8 Gy 0.2 RIS, a.u. 0.1 0.0 -0.1 -0.2 3450 3500 Magnetic field, G RIS parameters: g=2.0088 DH=9 G 3550 3600 RIS spectra obtained by subtraction of BKS spectrum recorded prior irradiation Result of dose reconstruction in the sample irradiated to 4 Gy 5 days before reconstruction 0.50 0.3 Original signal after treatment +2 Gy +4 Gy +6 Gy +11 Gy 0.1 0.45 EPR dose response, a.u. EPR signal, a.u. 0.2 0.0 -0.1 Parameters of the data fit with Grun model Imax=0.5513 D0=7.3 Gy DE=3.66Gy Grun model 0.40 0.35 0.30 Grun model: A = Imax(1 - exp(-(D+DE)/D0)), where A= EPR dose response, Imax = max EPR dose response (saturation level), DE=the dose to be determined D0= characteristic saturation dose 0.25 -0.2 0.20 3420 3440 3460 3480 3500 3520 3540 Magnetic field, G 3560 3580 3600 0 2 4 6 Added dose, Gy Reconstructed dose 3.66 Gy, reduction 8 10 12 Variability of dose dependence in fingernails 0.70 Fresh 1 0.60 A Old Samples Fresh 2 Fresh 3 Fresh 4 Amplitude (a.u.) 0.50 Fresh 5 Fresh 6 Fresh 7 0.40 Fresh 8 Fresh 9 Fresh 10 0.30 Old 1 Old 2 0.20 Fresh Samples Old 3 Old 4 Old 5 0.10 Old 6 Old 7 Old 8 0.00 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Dose (Gy) Old 9 Old 10 Dosimetric properties of fingernails • Optimal sample mass is 15-20 mg (nailparings from 2-3 fingers) • Measurements time 5 minutes (10 scans) • Achievable lower dose threshold ~ 1 Gy • RIS fading half-time 300 hr (~2 weeks) Conclusions Part of EPR body band/freq Tooth X enamel Tooth Q enamel Tooth L LLD, Gy 0.1 Fingernails X in vivo/ Time amount stability 50 – 100 mg 106 yr 0.3-0.5 2-4 mg 106 yr 3-5 In vivo 106 yr 0.5-1 20-30 mg ~2 wks Acknowledgements G. Burke, E. Demidenko, C. Calas, I. Clairand, T. De, O. Grinberg, A. Iwasaki, M. Kmiec, L. Kornak, B. LeBlanc, P. Lesniewski, P. Misra, C. Mitchell, R.J. Nicolalde, B. Pass, A. Ruuge, D.A. Schauer, J. Smirniotopoulos, A. Sucheta, T. Walczak Disclaimer The views expressed in this presentation are those of the author and do not reflect the official policy or position of the Navy and Marine Corps Public Health Center, Navy Bureau of Medicine and Surgery, Department of the Navy, Department of Defense, or the U.S. Government. www.Biodose-2008.org