Direct detection of C2H2 in
air and human breath by cw-CRDS
Florian Schmidt, Olavi Vaittinen, Markus Metsälä and Lauri Halonen
Laboratory of Physical Chemistry, University of Helsinki, Finland
Cavity ring-down spectroscopy (CRDS)
Laser
Detector
11

1
Scan
 int  Sn    
c  0 

Absorption enhancement due to cavity: 103 to 105

CRDS independent of laser power fluctuations

No calibration needed
June 21, 2010
0
2
Experimental CRDS setup - schematic
Tuning range: 1520-1575 nm
(6350-6575 cm-1)
Cavity Finesse: 390 000
Sensitivity: 7×10-11 cm-1
Acquisition time: 1 s - 20 min
Detected at ppt level:
C2H2, NH3, HCN
Other detectable species:
CH4, N2O, CO, CO2, H2O
ECDL = external cavity diode laser
OI = optical isolator
AOM = acousto-optic modulator
MFC = mass flow controller
June 21, 2010
3
Experimental CRDS setup
June 21, 2010
4
Motivation to measure acetylene









One of the most common hydrocarbons in the troposphere
Mostly anthropogenic sources (fossil- and biofuels, biomass burning)
Correlates with other combustions gases and VOCs
Trace pollution and exposure to combustion gases (exhausts, smoking)
Baseline of breath C2H2 in the healthy population? Is C2H2 a biomarker
for a medical condition (e.g. Proteus mirabilis bacterial infection)?
Not previously detected in ambient air with high time resolution and
without pre-concentration.
Not previously quantified in breath
GC-MS slow and expensive
Not measurable with PTR-MS and SIFT-MS with standard pre-cursors
June 21, 2010
5
Spectra from breath and ambient air
Air acetylene concentration 0.93 ppbv; Scan time 20 min; Detection limit 120 ppt
June 21, 2010
6
System response to C2H2 absorption
June 21, 2010
7
Outdoor air – continuous flow
F. M. Schmidt, O. Vaittinen, M. Metsälä, P. Kraus, L. Halonen, Appl. Phys. B (2010)
June 21, 2010
8
Day – Night comparison
F. M. Schmidt, O. Vaittinen, M. Metsälä, P. Kraus, L. Halonen, Appl. Phys. B (2010)
June 21, 2010
9
Indoor – outdoor air comparison
F. M. Schmidt, O. Vaittinen, M. Metsälä, P. Kraus, L. Halonen, Appl. Phys. B (2010)
June 21, 2010
10
Breath collection and sample handling

No adsorption/desorption phenomena in the bags, the gas system
tubing and the quartz-coated ring-down cavity

Acetylene can be collected and stored in aluminum breath bags
(statistical error for 10 bags: 0.4 ppb)

Acetylene concentration stays within 0.4 ppb for 5 days in the bag

No quantitative difference between alveolar/full breath samples and
nose/mouth breathing (non-smokers)

Thus: static measurements were possible

Spectrometer temperature: 22 degrees (room temperature)

Sample pressure in the cavity: 0.1 bar
June 21, 2010
11
Study - Background level of C2H2 in breath

40 healthy volunteers
22 male, 18 female, 20 to 63 years old, 31 nonsmokers, 9 smokers

Subjects requested to have stayed at sampling
location for 30 min prior to breath collection

One full breath sample collected to a breath bag

At the same time: one ambient air sample
collected to a second bag

Samples were analyzed on the same day or at
least within 48 hours.
June 21, 2010
12
Results – Breath acetylene minus air acetylene
Breath C2H2 baseline = Ambient air C2H2 level
Smokers could be identified with a sensitivity and specifity of 100 %
June 21, 2010
13
Timer after smoking
June 21, 2010
14
Exposure studies

Subject smokes a cigarette or takes a puff

Continuous breathing into CRDS system for up
to 30 min

C2H2 decay recorded standing on peak of
absorption profile (3 points per second)

For long-term decay: bag samples or shorter
intervals of continuous breathing

Fit exponential functions to decay

Simple 3-compartment model (respiratory tract,
blood, tissue)
June 21, 2010
15
Breath cycles
June 21, 2010
16
Acetylene residence times
June 21, 2010
17
Summary

CRDS suitable for fast and direct detection of C2H2 in air and breath

Mean C2H2 mixing ratio in Helsinki outdoor air of ~1 ppbv

C2H2 correlates with CO, human activity and temperature

Strong fluctuations (up to 60 ppbv on a minute time scale) in urban
outdoor air during daytime. Single bag measurements unreliable.

Breath C2H2 baseline is equal to ambient air C2H2 level

Smokers were identified with 100 % sensitivity and specifity

Breath acetylene returns to baseline level around 3 hours after
exposure. Urban air pollution could impair smoking status assessment

High-time resolution decay measurements could improve knowledge
about residence times of gases and related physiological processes