Cardiovascular System II Einthoven’s triangle and the cardiac vector Electrocardiogram connections are termed “leads”. The simplest connections, Leads I, II and III, define a triangle. I like to think of this triangle as a box containing the heart. Each lead is like a peephole that lets us have a view of the heart in the box. If the bulk of the flow of current around the heart during ventricular depolarization is represented by a vector that sums all the currents, this vector would ordinarily point downward and to the left, pretty much along the physical axis of the heart. Lead I gives us a view of this vector pointing almost directly away from us. Lead III shows the vector pointing almost directly toward us, and Lead II gives a side-on view. Any 2 of leads I, II and III can define the vector in 2 dimensions Here, the relative heights of the R wave in leads I, II and III are plotted on the Einthoven triangle to yield the vector The 12-lead EKG • In modern clinical settings, EKG data is collected from 12 different leads • The 3 augmented leads simply add a second triangular box – if the standard Einthoven triangle and the augmented lead triangle were drawn together, they would form a 6-pointed star. • The 6 precordial leads utilize an exploring electrode that is placed at 6 points forming a ring around the left chest. Each of these points acts as a keyhole through which we can gain a different view of the cardiac vector, allowing it to be visualized in 3 dimensions. Axis deviations • As a rule, the vector shifts away from damaged, wasted or non-excited tissue that does not generate current - a scar from an old heart attack, for example. • The vector shifts toward parts of the ventricle that generate excessive current, such as a hypertrophied ventricle or a patch of freshly damaged muscle that is electrically leaky. Ventricular hypertrophy as a source of axis deviation • Systemic hypertension that overloads the left ventricle over time will cause it to hypertrophy, resulting in a left-axis deviation. • Long-term overload of the right ventricle (pulmonary hypertension or an interventricular shunt) will result in a rightaxis deviation. What kind of axis shift is happening here? This trace is from a 63 year old woman with long-term hypertension. She also has a 1st degree AV block, as shown by the prolonged P-R interval (>0.20 sec) Bundle branch block as a source of axis deviation • Ordinarily, the left ventricle is served by two branches of the Bundle of His • If one of these is damaged, a conduction delay or conduction block results, causing excitation to spread to that part of the ventricle more slowly, causing both an axis shift toward the blockage (i.e. a right axis shift for a right bundle block and vice versa) and a broadening or splitting of the QRS. Arrhythmias • Arrhythmias have 3 main causes – Failure within the SA node or conducting system – Presence of an ectopic pacemaker (ectopic means “wrong place) – Abnormal pattern of spread of excitation within the myocardium, or between the atria and the ventricles If the SA node fails, the AV node can take over Sinoatrial nodal block, with AV nodal rhythm during the block perod (Lead III). The SA node is driving the 1st and 2nd beats – see the P waves. When the SA node drops out, the AV node becomes the pacemaker –see no P waves - at a slower rate. 1st Degree AV block In 1st degree AV block, conduction into the AV node is even slower than usual, so the P-R interval is prolonged 2nd Degree AV Block In 2nd degree AV block, conduction from the SA node to the AV node fails entirely for some beats, so the heart skips a beat. In case of complete AV conduction failure, a ventricular pacemaker usually appears Notice that there is no relationship between the SA rhythm and ventricular contractions, and the shape of the QRS seems wrong What is happening here? PVC (= VPC) This trace is from a 62-year old woman who had a heart attack some years previously. Note that the QRS and T waves from the abnormal beat are different from the normal beats (why?) and there is a compensatory pause before the next normal beat (why?). PVC = premature ventricular contraction. This patient is at risk to fall into ventricular fibrillation if the PVC falls during the time interval in which some, but not all of the ventricle has repolarized. In fibrillation, the myocardium stops acting as a unit • In the fibrillating myocardium, excitation does not pass over the myocardium as a wave front, but continuously circles back to reenter areas that had already recovered from their refractory period. This is usually due to the presence of some damaged tissue that conducts excitation very slowly. Once it starts, the heart does not recover a normal rhythm spontaneously. • Defibrillation synchronously excites the whole myocardium, making it all refractory at the same time. This may reestablish a normal conduction pattern. Mechanical Aspects of the Heart Cycle Review of cardiac structures Beware of some confusing multiple names. I will refer to the tricuspid and mitral valves simply as the AV valves. Sometimes the mitral valve is referred to as the bicuspid valve. The pulmonary and aortic valves are sometimes referred to as the semilunar valves. Events of the heart cycle • Late diastole: AV valves are open – blood is flowing into the ventricles under venous pressure (diastasis or passive filling). Pulmonary and aortic valves are closed. • Atrial contraction (atrial systole): adds an additional amount of blood to the ventricles – the contribution is trivial except during exercise • AV delay: allows atrial contraction to be complete • Ventricular systole: immediately causes closure of AV valves, causing 2nd heart sound. Vent. Systole can be divided into – an initial period of isovolumetric contraction during which all heart valves are closed, – followed by a period of blood ejection, during which pulmonary and aortic valves are open, – then pulmonary and aortic valves close (2nd heart sound), leading to a period of isovolumetric ventricular relaxation. When ventricular pressure drops below venous pressure, the AV valves reopen and passive filling begins again. Pressure and volume changes in the heart cycle A must-know, mustunderstand slide. The pressure values are for the left ventricle and aorta– values for the right ventricle and pulmonary arteries would be about 1/3 as great The mechanical task of the heart • The heart fills under low pressure of venous blood (preload) -1 – +3 mmHg – it’s not a good suction pump, so venous values cannot and should not go far in the negative direction. • The heart ejects blood against the high pressure of arterial blood (afterload) average value about 100 mmHg. • As averaged over times longer than a few beats, the cardiac output of each ventricle must equal the venous return to that ventricle, and the cardiac output of each ventricle must match that of the other, or a physiological catastrophe would occur. Intrinsic regulatory properties of the healthy heart muscle prevent this catastrophe. Determinants of Cardiac Performance • Heart rate: determined by balance of chronotropic effects of adrenergic and cholinergic inputs • Intrinsic heart contractility or inotropic state: determined by the state of health of the myocardium, the inotropic effects of autonomic inputs, and anything else that affects the peak Ca++ level reached in the cytoplasm during activation. • End-diastolic ventricular volume – which determines the preload of the ventricule • Aortic or pulmonary arterial pressure, which determines the afterload of the ventricle. Performance characteristics of the isolated heart • The first step towards understanding the performance of the CV system is to study the performance of the heart in isolation. • These experiments were carried out by Frank for the frog heart and Starling for the dog heart. The effect of the preload: the Frank-Starling Law says the heart pumps out what comes in These cardiac function curves are basically about the relationship between the length of cardiac muscle fibers and their force development: over the normal operating range of ventricular end-diastolic volumes, increasing volume increases force, which increases stroke volume, which increases cardiac output. The curves reflect the performance of the isolated heart, in the absence of the vascular system. Cardiac function curves reflect the limits of potential performance – they don’t determine actual performance The inotropic state of the heart merely determines the range of possible cardiac outputs – for example, the maximum cardiac output for the heart in the absence of any positive inotropic effect is 10 L/min; with maximal sympathetic input it rises to almost 25 L/min. As we will see, the actual cardiac output in any real-life situation is determined by the interaction of the heart and the vascular system, and not by the heart alone. Cardiac performance versus the afterload + sympathetic input No input Parasympathetic input or disease This curve shows that the ventricle is rather insensitive to afterload, maintaining a constant output until arterial pressure reaches a value at which the heart is simply not strong enough to open the aortic valve.