Project DUNAMET no.

D36

Tomáš Valenta

Czech Metrology Institute

Project DUNAMET no. D36

• the inter-comparison among

Czech Republic ( CMI )

Austria ( BEV )

Hungary ( OMH ) and Slovak Republic ( SMU )

• took place from May 2004 to August 2004

• the inter-comparison of low pressure test facilities for gas meters in the range from 100 m 3 /h to m 3 /h

1000

Project DUNAMET no. D36

The axial turbine gas meter G650 was used for the comparison.

Manufacturer:

PREMAGAS

(ELSTER –AMCO

Group)

Type: DAE

Size: G650

DN150

Project DUNAMET no. D36

Calibration procedure:

The turbine gas meter G650 was tested in each laboratory by air in horizontal position .

It was necessary to use the upstream straightening pipe that was long at least 5x DN and the downstream straightening pipe that was long at least 3x DN .

The reference pressure from the turbine gas meter was measured from the output “ p r

”.

Project DUNAMET no. D36

The reference temperature from the turbine gas meter was measured in the distance (2

3)x DN downstream of the turbine gas meter.

The HF pulse emitter A1S (the pulse number 4579,88 pulses/m 3 ) was used for the tests.

The calibration had to be performed in the laboratory where the temperature was from 19,5°C to 23,5°C . No oil lubrication was used.

Project DUNAMET no. D36

The turbine gas meter was tested in 8 flow rates :

1000 m 3 /h, 800 m 3 /h, 650 m 3 /h, 500 m 3 /h, 400 m 3 /h, 300 m 3 /h,

200 m 3 /h, 100 m 3 /h.

Error of the meter was determined. The Error is value which shows the relationship in percentage terms of the difference between the volume indicated by the meter and the volume which has actually flowed through the meter , to the later value.

E

V i

V c

V c .

100 (%) where E is the error of the meter

V i is the indicated volume by the meter (m 3 )

V c is the real volume which has actually flowed through the meter (m 3 )

Project DUNAMET no. D36

Determination of the reference value

To determine the reference value tree different approaches were used.

Mean

:

Mean is the arithmetic average of a set of numerical data.

Median: Median is the middle value of an ordered set of numerical data.

Weighted mean error (WME):

If in a series of observations a statistical weight (1/u 2 ) is assigned to each value, a weighted mean y can be calculated by the formula: where y

 x

1

1 u u

2

2

 

1

 

1

....

....

 x

1

N u

2 u

2 x

 

N

 

N x i are values of errors of meter in different laboratories in one flow rate u (x i

) are uncertainties U(k=2) of errors of meter in different laboratories in one flow rate

Project DUNAMET no. D36

Results:

Flow rate in the meter m³/h

1000

800

650

500

400

300

200

100

Mean

%

-0,17

-0,03

0,01

0,02

-0,03

-0,03

0,15

0,93

Median

%

-0,14

0,01

0,05

0,03

-0,04

-0,05

0,12

0,91

Weighted mean error (WME)

%

-0,18

-0,03

0,02

0,03

-0,03

-0,03

0,14

0,93

1,20

1,00

0,80

0,60

0,40

0,20

0,00

-0,20

0

-0,40

-0,60

-0,80

-1,00

Project DUNAMET no. D36

Austria (BEV)

DUNAMET No. D36, turbine gas meter G650

100 200 300 400 500 600 700 800 900 1000 1100

BEV mean

WME median flow rate (m

3

/h)

Project DUNAMET no. D36

Hungary (OMH)

DUNAMET No. D36, turbine gas meter G650

1,20

1,00

0,80

0,60

0,40

0,20

0,00

-0,20

0

-0,40

-0,60

-0,80

-1,00

100 200 300 400 500 600 700 800 900 1000 1100 flow rate (m

3

/h)

OMH mean

WME median

1,20

1,00

0,80

0,60

0,40

0,20

0,00

-0,20

0

-0,40

-0,60

-0,80

-1,00

Project DUNAMET no. D36

Slovak Republic (SMU)

DUNAMET No. D36, turbine gas meter G650

100 200 300 400 500 600 700 800 900 1000 1100

SMU mean

WME median flow rate (m

3

/h)

Project DUNAMET no. D36

Czech Republic (CMI)

DUNAMET No. D36, turbine gas meter G650

1,20

1,00

0,80

0,60

0,40

0,20

0,00

-0,20

0

-0,40

-0,60

-0,80

-1,00

100 200 300 400 500 600 700 800 900 1000 1100 flow rate (m

3

/h)

CMI mean

WME median

Project DUNAMET no. D36

Conclusion:

All participants were successful. All determined errors of meter with their uncertainties U(k=2) in every flow rate cover the reference values

Mean, Median and Weighted Mean Error.

New EUROMET project

THIS PROJECT CAN BE REPEATED AMONG

EUROMET COUNTRIES IN 2006.

Note: I would like to coordinate quick and simple intercomparison. That is why the supposed time to make calibration and to deliver the meter to next laboratory will be only 1 week .

The form „

PROPOSED EUROMET PROJECT

“ will be sent to everybody by e-mail.

Please don´t applause because Mr. Lojek is sleeping.

Thank you for attention.