Sterilization

advertisement
Sterilization
Lecture 7
Tahir
A fermentation product is produced by the culture of a certain organism, or
organisms in a nutrient medium. If the fermentation is invaded by a
foreign microorganism then the following consequences may occur:
Avoidance of contamination may be
achieved by:
1. Using a pure inoculums to start the
fermentation.
2. Sterilizing the me
dium to be employed
3. Sterilizing the fermnter vessel
4. Sterilizing all material to be added to the
fermentation during the process.
5. Maintaining aseptic condition during the
fermentation.
Medium sterilization
•
•
•
Media may be sterilized by filtration, radiation ultrasonic treatment, chemical treatment or heat.
However, for practical reasons, STEAM is used universally.
Major exception is the use of filtration for the sterilization of media for animal cell culture – such
media are completely soluble and contain heat labile components making filtration the method of
choice. ‘
The destruction of micro organisms by steam may be described as a first order chemical reaction
and thus may by represented by the following equation:
•The relationship in above Fig would
be observed only with the
sterilization of a pure culture in one
physiological form, under ideal
sterilization conditions
•The value of K is not only species
dependent, but depended on the
physiological form of the cell.
•Eg. Endopsore of the genus Bacillus
are far more heat resistant than the
vegetative cells.
• As with any first order reaction, the reaction rate increases
with increase in temperature due to an increase in the
reaction rate constant, which, in the case of the destruction
of micro-organisms, is the specific death rate (K).
• Thus K is the true constant only under constant
temperature conditions.
• The relationship b/w Temp & the reaction rate constant
was demonstrated by Arrhenius & may be represented by
the equation:
•It is clear that the same degree of sterilization may be obtained over a wide range
of time & temp.
•This kinetic description of bacterial death enables the design of procedure for the
sterilization of fermentation broths.
•By choosing a value for Nt procedure may be designed having a certain probability
of achieving sterility, based upon the degree of risk that is considered acceptable.
•Only microbial contaminants present are spores of Bacillus stearothermophilus
(most heat resistant)
•Bacillus macerans suspended in oil were ten times more resistant to sterilization if
they were dry than if they were wet.
• Time & temp may now be determined to achieve the
desired Del factor.
• Deleterious reactions may occur in the medium during the
sterilization process, resulting in loss of nutritive quality.
•Deleterious effect of increasing medium sterilization
time on the yield of product of subsequent of
fermentations.
•Initial rise in yield is due to some component of the
medium being made more available to the process
micro-organism by the “Cooking Effect” of a brief
sterilization period.
•Two types of reactions contribute to the los of
nutrient quality during sterilization:
1. Interactions b/w nutrient components of the
medium
2. Degradation of heat liable components.
• The thermal destruction of essential media components conforms
approximately with first order reaction kinetics, may be describe by
equation similar to derived for destruction of bacteria:
Decline in conc. Of the nutrient components, along with decline in the no of
contaminants. The effect of temp on the reaction rate constant may be expressed by
the Arrhenius equations:
Therefore, a plot of natural logarithm of the reaction rate against 1/T will give a straight
line, slope - (E/R). As R (gas constant) is fixed, the slope of the graph is determined by
the value of the activation energy (E). E.g the value of E for B. stearothermophilus
spores has 67.7Kcal/mole, whereas the thermal destruction of the nutrients is 10 -30
kcal/mole (Richard 1968)
•From this plot it may be seen that as temp increased, the reaction rate rises more rapidly
for the reaction with higher activation energy.
Thus, considering the difference b/w activation energies for spore destruction & nutrient
degradation, an increase in temp would accelerate spore destruction more than medium
de-naturation.
• Del factor could be achieved over a range of temp/time regimes. So high temp for a short
time to achieve desired probability of sterility, with minimum nutrient degradation.
Advantages of continuous sterilization over batch sterilization
•
•
•
•
•
•
Superior maintenance of medium quality
Ease of scale up
Easier automatic control
The reduction of surge capacity for steam.
Reduction of sterilization cycle time
Under certain circumstances, the reduction of fermenter corrosion.
Advantages of batch sterilization over continuous sterilization
• Lower capital cost
•Lower risk of contamination – continuous processes requires the aseptic transfer
of the sterile broth to the sterile vessel.
•Easier manual control
•Easier to use with media containing a high proportion of solid matter
Early continuous sterilizers (plate heat exchangers), unsuitable on two accounts
1. Failure of the gas kit b/w the plates resulting in mixing of sterile & unsterile
streams
2. Particulate components in the media would block the heat exchangers.
However, modern continuous sterilizers use double spiral heat exchangers in
which two streams are separated by a continuous steel division. Spiral
exchangers are less susceptible to blockage. Further precision is controlled by
sophisticated computerized monitoring.
DESIGN OF BATCH STERILIZATION PROCESS
• It is less successful in avoiding the destruction of nutrients than a
continuous one. But still used with minimum loss of nutritive quality.
• Highest temp 121ºC, but procedure should be designed for the exposure
of medium to this temp kept minimum.
• This is achieved by heating and cooling periods. Following information
must be available for the design of batch sterilization process:
•Knowing the original no of organisms present in the fermenter & the risk of
contamination considered acceptable, the required Del Factor may be calculated.
•A frequently adopted risk of contamination is 1 in 1000 which indicate that Nt
should equal to 10-3 of the viable cells.
e.g: if a specific case in which unsterile broth was shown to contain 1011 viable
organism then the Del factor may be calculated, thus:
•
Therefore the overall Dell Factor required is 32.2. however, the destruction of cells
occurs during the heating and cooling of the broth as well as during the period of
121ºC, thus the over all Del Factor may re presented as:
Method of batch sterilization: (advantages of separate sterilization vessels)
1. Cooker is used to sterilized the medium
2. Sterilized in a cooler in more concentrated form than used in the
fermentation, then diluted in the fermenter by sterile water
3. If medium is viscous then require more power in fermenter, thus if
requirement for agitation is removed it need less power, but sterilization
kettle would have with power motor.
4. Fermenter would be spared the corriosn at high temperature
Disadvantage of a separate sterilization vessels:
1. Cost of constructing a batch medium sterilizer is much the same as that of the
fermenter
2. If cooker server a large no of fermenter complex pipe work would be
necessary to transport the sterile medium with inherent danger of
contamination.
3. Mechanical failure in a cooker will stop all the process, it maintenance costly.
THE DESIGN OF CONTINUOUS STERILIZATION PROCESS
•
The design of continuous sterilization cycles may be approached in exactly the
same way as for batch sterilization system.
• The continuous system includes a time period during which the medium is heated
to the sterilization temp, a holding time at the desired temp & a cooling period to
restore the medium to the fermentation temp.
• The length of the holding period is dictated by the length of the coil & the flow
rate of the medium. First it maintain the temp for heating up the incoming
medium & second using cooling water.
Advantage:
• A much higher temp may be utilized, thus reducing the holding time and reducing
the degree of nutrient degradation.
• The required Del Factor may be achieved by the combination of temp & holding
time which gives an acceptably small degree of nutrient decay.
• Furthermore, b/c of continuous process involves treating small increments of
medium the heating up and cooling down periods are very small compared with
those in a batch system.
• There are two types of continuous sterilizer which is used for the treatment of
fermentation media:
1. indirect heat exchanger
2. direct heat exchanger (steam)
• The most suitable indirect heat exchanger are of double spiral-type
which consist of two sheets of high grade stainless steel which have
been curved around a central axis to form a double spiral.
• To achieve sterilization temp steam is passed
through one spiral & medium through the other
in countercurrent stream.
• Spiral heat is also used to cool the medium after
passing through the holding coil.
• Incoming unsterile medium is used as the cooling
agent in the first cooler so that the incoming
medium is partially heated before it reaches the
sterilizer & thus, heat is conserved.
Advantages of spiral heat:
1. avoid contamination.
2. Clearances – self cleaning reduce the risk of
sedimentation, fouling & burning on.
• Indirect plate heat exchanger consist of
alternating
plates
through
which
the
countercurrent streams are circulated.
• The plates are separated by gas kit & failure of
these gaskets can cause cross contamination b/w
two streams.
• Clearances b/w the plates i.e suspended solids in
the medium may block the exchanger & thus
system is only useful in sterilizing completely
soluble media
• However, the plate exchanger is more adaptable
than the spiral system in that extra plates may be
added to increase its capacity.
•
The continuous steam injector injects steam directly into the unsterile broth.
Advantage:
1.
Very short heating up time.
2.
May be used for suspended solids
3.
Low capital cost
4.
Easy cleaning & maintenance.
5.
High steam utilization efficiency.
Disadvantages:
1.
Foaming occur during heating
2.
Direct contact of the medium with steam are requires, condense dilution & requires clean steam, free from
anticorrosion additives.
In some cases a combination of direct & indirect heat exchangers may be used.
•
•
•
•
•
The most widely used continuous sterilizations system is that based on the spiral heat
exchangers.
Plant is sterilize by circulating a hot (sterilized) water through the plant in a closed circuit.
Heat conservation is achieved by cooling the sterile medium against cold, incoming unsterile
medium which will then be partially heated before it reaches the sterilizer.
Del factor to be achieved in a continuous sterilization process has to be increased with an
increase in scale eg. If the volume to be sterilized is increased from 1000dm3 to 10,000 dm3 &
the risk of failure is to remain at 1 in 1000 then Del factor must be increased from 34.5 to
36.8.
Advantage of the continuous process is that temp may be used as variable in scaling up
continuous process so that the increased Del factor may be achieved, maintaining the
nutrient quality constant.
Download