AcrySof ® ReSTOR® Aspheric IOL AcrySof ® ReSTOR® Aspheric IOL AcrySof ® ReSTOR® Aspheric IOL SN6AD3 SN6AD1 Add Power: +4 D Spectacle Plane: 3.2 D Range: +10 D to +34 D A-Constant: 118.9 Add Power: +3 D Spectacle Plane: 2.4 D Range: +10 D to +34 D A-Constant: 118.9 2 AcrySof ® ReSTOR® Aspheric IOL Discussion Topics • Aspheric (SN6AD3) vs. Spheric (SN60D3) • Introduction of SN6AD1 • True performance at near, intermediate and distance • Comparison between SN6AD1 and SN6AD3 • High resolution image quality • Minimal visual disturbances 3 AcrySof ® ReSTOR® Aspheric IOL Anatomy of the Apodized Diffractive Technology Central 3.6 mm apodized diffractive structure Step heights decrease peripherally from 1.3 – 0.2 microns Anterior aspheric optic A +4 D at lens plane equaling +3.2 at spectacle plane Outer refractive zone 4 AcrySof ® ReSTOR® Aspheric IOL Anatomy of the Apodized Diffractive Technology 13.0 mm Anterior Apodized Diffractive Optic 6.0 mm Symmetric Biconvex 6.0 mm Anterior Aspheric Optic 5 AcrySof ® ReSTOR® Aspheric IOL Apodization • Gradual reduction or blending of the diffractive step heights. • Optimally manages light energy delivered to the retina as it distributes the appropriate amount of light to near and distant focal points, regardless of the lighting situation. • Designed to improve image quality while minimizing visual disturbances. 1.3 micron step 6 AcrySof ® ReSTOR® Aspheric IOL Technology of the AcrySof ® ReSTOR® IOL in “Human” Terms • Thickness of a Human Hair = 60 microns • Thickness of a Red Blood Cell = 7 microns • Step Height at periphery of the diffractive portion of the AcrySof ® ReSTOR® Aspheric IOL = 0.2 microns 7 AcrySof ® ReSTOR® Aspheric IOL Design Objective • Design considerations for the AcrySof ® ReSTOR® Aspheric IOL: – Induce negative Spherical Aberrations with the lens to compensate for positive corneal Spherical Aberrations 8 AcrySof ® ReSTOR® Aspheric IOL The Problem – Spherical Optics Marginal Rays Spherical IOL Light Rays Paraxial Rays Spherical Aberration • Spherical aberration occurs when light rays are over-refracted at the periphery of a lens system, resulting in a region of defocused light which can decrease image quality. 9 *Smith, G., Atchinson D.A., (1997) The Eye and Visual Optical Instruments. Cambridge University Press, Cambridge, United Kingdom, pp. 667. AcrySof ® ReSTOR® Aspheric IOL The Solution – Aspheric Optics Aspheric IOL Light Rays Aspheric Surface • Aspheric optics align the light rays to compensate for positive corneal spherical aberration, resulting in enhanced image quality. 10 *Smith, G., Atchinson D.A., (1997) The Eye and Visual Optical Instruments. Cambridge University Press, Cambridge, United Kingdom, pp. 667. AcrySof ® ReSTOR® Aspheric IOL AcrySof ® ReSTOR® Aspheric IOL (SN6AD3) Specifications OPTICS 13.0 mm 6.0 mm Multifocal Apodized Diffractive Optic Compensation for Positive Corneal Spherical Aberration Aspheric Optic Optic Type Proprietary Symmetric Biconvex Optic Diameter 6.0 mm Overall Length 13.0 mm MATERIAL Optic/Haptic Material AcrySof® Hydrophobic Acrylic Light Filtration UV and High-Energy Blue DESIGN IOL Design Single-Piece Haptic Design STABLEFORCE® Modified-L SPECIFICATIONS Diopter Range +10 D – +30 D in 0.5 D increments +31 D – +34 D in 1.0 D increments 11 Introducing AcrySof ® ReSTOR® Aspheric IOL +3 D AcrySof ® ReSTOR® Aspheric IOL Lens Objectives • Offer an additional choice of add power • Maintain existing optical design characteristics and manufacturing processes • Without affecting visual disturbances or image contrast 13 AcrySof ® ReSTOR® Aspheric IOL SN6AD1 Design Characteristics • Utilizes existing ReSTOR® Aspheric +4 D platform with identical asphericity, energy distribution profile, and shape factor • Modified add power from +4 D to +3 D – 9 diffractive rings vs. 12 diffractive rings – Slightly wider ring spacing to modify the add power 14 AcrySof ® ReSTOR® Aspheric IOL Physical Comparison +4 D, 12 zones +3 D, 9 zones • Both +4 D and +3 D have 3.6 mm Apodized Diffractive region • +4 D central zone diameter = 0.742 mm • +3 D central zone diameter = 0.856 mm 15 AcrySof ® ReSTOR® Aspheric IOL FRACTION OF ENERGY Physical Comparison ReSTOR ® Energy Balance 4.0 D add vs. 3.0 D add 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 ReSTOR® Aspheric +3 D Distance ReSTOR ® +4 D Add Distance ReSTOR® Aspheric +4 D Near ReSTOR® Aspheric +3 D Near 1.0 2.0 3.0 4.0 5.0 6.0 LENS DIAMETER (mm) • Diffractive step height profile and energy distribution are identical for +4 D and +3 D ReSTOR® Aspheric IOLs Data on File. Alcon, Inc. 16 AcrySof ® ReSTOR® Aspheric IOL Study Objective To confirm that the increase in the distance of near vision for the ReSTOR® Aspheric +3 D IOL does not negatively impact: • visual acuity at near, distance, or intermediate • contrast sensitivity • halo visual disturbance scores as compared to the ReSTOR® Aspheric +4 D IOL 17 AcrySof ® ReSTOR® Aspheric IOL Study Design • • • • • Prospective Randomized Subject masked 6 month duration with FDA submission at 3 months 300 subjects bilaterally implanted with either – AcrySof ® ReSTOR® Aspheric +3 D IOL Model SN6AD1 – AcrySof ® ReSTOR® Aspheric +4 D IOL Model SN6AD3 • Approximately 20 subjects at each site 18 True VA Performance • Near • Intermediate • Distance AcrySof ® ReSTOR® Aspheric IOL Mean Distance VA Uncorrected VA Best Corrected VA -0.1 LogMAR 20/20 0 20/25 0.1 20/32 0.2 20/40 0.3 ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 • Both VAs are less than one letter different 20 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Mean Near VA at Best Distance Uncorrected VA Best Corrected VA -0.1 LogMAR 20/20 0 20/25 0.1 20/32 0.2 20/40 0.3 ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 • Both VAs are less than two letters different and both 20/25 21 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Mean Distance of Best Near VA ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 Distance in cm 40 38 36 34 32 30 28 Uncorrected in cm Distance corrected in cm • ReSTOR® Aspheric +3 D has approximately 6.5 cm (2.6 in) increase in distance compared to ReSTOR® Aspheric +4 D Data on File. Alcon, Inc. 22 AcrySof ® ReSTOR® Aspheric IOL Mean Uncorrected Intermediate VA LogMAR 50 cm 60 cm 70 cm 0 20/20 0.1 20/25 0.2 20/32 0.3 20/40 0.4 ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 • VAs are approximately 1-1.5 line(s) better for ReSTOR® Aspheric +3 D Data on File. Alcon, Inc. 23 AcrySof ® ReSTOR® Aspheric IOL Mean Distance Corrected Intermediate VA LogMAR 50 cm 60 cm 70 cm 0 20/20 0.1 20/25 0.2 20/32 0.3 20/40 0.4 ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 • VAs are approximately 1.5 lines better for ReSTOR® Aspheric +3 D Data on File. Alcon, Inc. 24 AcrySof ® ReSTOR® Aspheric IOL Percent of Subjects Combined VA’s: Distance, Intermediate, Near 100 80 60 40 20 0 20/20 or 20/25 or 20/32 or 20/40 or better better better better ReSTOR® Aspheric +3 D N=138 worse than 20/40 ReSTOR® Aspheric +4 D N=131 25 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Bilateral Defocus Curve Refraction (D) 2 1.5 1 0.5 0 0.5 -1 -1.5 -2 -2.5 -3 -3.5 -4 -4.5 -5 -0.1 0 20/20 logMAR 0.1 0.2 0.3 20/40 0.4 0.5 0.6 0.7 ReSTOR® Aspheric +3 D Data on File. Alcon, Inc. ReSTOR® Aspheric +4 D 26 AcrySof ® ReSTOR® Aspheric IOL Clinical Defocus Curve 27 AcrySof ® ReSTOR® Aspheric IOL Distance Contrast Sensitivity Mean Photopic without Glare Log Units 2.5 2 1.5 1 0.5 0 3 cpd 6 cpd 12 cpd 18 cpd Spatial Frequency ReSTOR® Aspheric +3 D N=117 ReSTOR® Aspheric +4 D N=114 28 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Distance Contrast Sensitivity Mean Photopic with Glare Log Units 2.5 2 1.5 1 0.5 0 3 cpd 6 cpd 12 cpd 18 cpd Spatial Frequency ReSTOR® Aspheric +3 D N=117 ReSTOR® Aspheric +4 D N=114 29 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Distance Contrast Sensitivity Mean Mesopic without Glare Log Units 2.5 2 1.5 1 0.5 0 1.5 cpd 3 cpd 6 cpd 12 cpd Spatial Frequency ReSTOR® Aspheric +3 D N=117 ReSTOR® Aspheric +4 D N=114 30 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Distance Contrast Sensitivity Mean Mesopic with Glare Log Units 2.5 2 1.5 1 0.5 0 1.5 cpd 3 cpd 6 cpd 12 cpd Spatial Frequency ReSTOR® Aspheric +3 D N=117 ReSTOR® Aspheric +4 D N=114 31 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Distance Image Quality in Mesopic Conditions 40 30 21 22 20 18 11 10 10 0 5 mm Aperture ReSTOR® Aspheric +3 D ReSTOR® Aspheric +4 D TECNIS* MF IOL REZOOM* IOL ACRI.LISA* 366D MTF measured with model eye at 100 lp/mm. All IOL powers range from 20.0D to 22.0D Data on File. Alcon, Inc. *Trademarks are property of their respective owners. 32 Minimal Visual Disturbances AcrySof ® ReSTOR® Aspheric IOL Mean Visual Disturbances None Moderate Mild Severe Problems with Night Vision Problems with Color Perception Halos Glare/Flare Double Vision Distorted Near Vision Distorted Far Vision Blurred Near Vision Blurred Far Vision 0 1 2 3 ReSTOR® Aspheric +3 D N=138 Data on File. Alcon, Inc. 4 5 6 ReSTOR® Aspheric +4 D N=131 7 34 AcrySof ® ReSTOR® Aspheric IOL Mean Visual Disturbances ACRI.LISA* REZOOM* ReSTOR® Aspheric +4 D Images provided by Jim Schwiegerling at the University of Arizona *Trademarks are property of their respective owners. TECNIS* ReSTOR® Aspheric +3 D 35 AcrySof ® ReSTOR® Aspheric IOL Frequency of Spectacle Wear (Distance Vision) 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 Never Sometimes Always How often do you wear glasses? 36 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Frequency of Spectacle Wear (Overall) 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% ReSTOR® Aspheric +3 D N=138 ReSTOR® Aspheric +4 D N=131 Never Sometimes Always How often do you wear glasses? 37 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL Percent of Subjects 96% of ReSTOR +3 Patients Would Have the Same Implant Again 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% ReSTOR® Aspheric +3 D N=149 ReSTOR® Aspheric +4 D N=139 No Yes Would you have the same implant again? 38 Data on File. Alcon, Inc. AcrySof ® ReSTOR® Aspheric IOL AcrySof ® ReSTOR® Aspheric IOL Conclusions • True performance at near, intermediate, and distance • ReSTOR® Aspheric +3 D IOL provides a range of near vision – Best distance approximately 7cm out for ReSTOR® Aspheric +3 D IOL • • • • 1 – 1.5 line(s) improvement in intermediate VA Aspheric optic provides improved image quality No significant differences in visual disturbances High rate of patient satisfaction and spectacle independence 39