New Gilian 5000
5 LPM Air Sampling Pump
Gilian 5000 Features
•
•
•
•
•
•
•
Our most powerful 5 LPM pump
Digital flow display with 30-day calibration
Meets EN 1232 requirements
Intrinsically safe
Fast charge, 4 hours or less
Memory free NiMH battery pack
Intuitive 4-key operation
Wide Flow Rate Range
• 1000 to 5000 cc/min,
constant flow
• 20 to 750 cc/min with
LFCP adapter
• Digital Flow Display
Intuitive Keypad Operation
• Easy operation
• No tools, 4-key format
• Similar to Gilian 3500
Highest Back Pressure Specs
•
•
•
•
•
5000 cc/min…..…….to 20 inches H2O
4000 cc/min………...to 30 inches H2O
3000 cc/min…………to 50 inches H2O
2000 cc/min…………to 60 inches H2O
1000 c/min…………..to 70 inches H2O
Compare at 2 LPM
•
•
•
•
•
Gilian 5000
Gilian 3500
GilAir-3
GilAir-5
HFS-513
•
•
•
•
•
60 inches H2O
35 inches H2O
20 inches H2O*
37 inches H2O
40 inches H2O
•
* GilAir-3 spec is 20 inches at 2.5 LPM
Compare Also at 2 LPM
•
•
•
•
Gilian 5000
Casella APEX Pro
SKC 224-44XR
SKC AirChek XR5000
• MSA Escort Elf
• TSI SP-350
• Buck VSS-5
•
•
•
•
•
•
•
60 inches H2O
31.5 inches H2O
40 inches H2O
50 inches H2O
30 inches H2O
29 inches H2O
60 inches H2O
Compare at 5 LPM
•
•
•
•
•
Gilian 5000
GilAir-5
HFS-513
SKC 224-44XR
SKC AirChek XR5000
• Casella APEX Pro
• Buck VSS-5
•
•
•
•
•
•
•
20 inches H2O
10 inches H2O
15 inches H2O
N/A
10 inches H2O
Free flow
2 inches H2O
Operation of the Gilian 5000
Power Up/Ready Mode
Taking a Sample
Key Pad Lock-out
Setting the Flow Rate
Displayed Flow Calibration
Calibration Set Up
Low Flow Sampling
• Set Flow Rate at 1500
cc/min
• Attach adapter with
arrow pointing toward
the pump
• Can be attached in-line
or fixed mount
• Use holder with needle
valve for constant
pressure/multi-flow
sampling
Calibration Using Low Flow
Gilian 5000 Programmability
• Start-Stop Programmable
through Set/Cal key
• Up to four events of varying
duration
• Can also program a duty
cycle for extended time
sampling
• Can also activate and
deactivate certain functions
like auto re-start and instant
start.
Programming
• The Program Mode is accessed by pressing the set/Cal
key three times.
• A program is activated by pressing the up arrow to
display a “1”
• To create a new program press Enter, and enter values
for the following parameters
• 1) A delayed start
• 2) Up to four run times of various duration
• 3) Up to four delay-run times to determine holds between
runs
• 4) Number of cycles
Options
• Options are accessed by pressing the Set/Cal key four
times to display OPT
• Three Options may be activated by pressing the UP
arrow to display a “1”
• 1) Disable auto re-start
• 2) Enable auto lock
• 3) Enable auto start (by-passes the Ready mode)
Gilian 5000
Our most powerful
personal sampling pump
from the world leader.