place photo here Quantrainx50 7.2 EDS Basic 3-2010 Confidential EDS • Universal applications • Elemental analysis • 1% Accuracy • Beam Control / Imaging • Repeatable 2 References (“the Book” --Highly Recommended) • Scanning Electron Microscopy and X-Ray Microanalysis A Text for Biologists, Materials Scientists, and Geologists, Joseph I. Goldstein, Dale E. Newbury, Patrick Echlin, David C. Joy, A.D. Romig, Jr., Charles E. Lyman, Charles Fiori, Eric Lifshin, Plenum Press, New York, 1992. (ISBN -- 0-306-44175-6) 3 • *(Many thanks to Dr. Bob Anderhalt for graphics and Example of Quantax50/EDAX Integration OUT DATED IMAGE !!! 4 Your Basic Bohr Atom 5 N shell M shell L shell K shell Inelastic Collision N shell Primary Beam Electron M shell Secondary Electron L shell K shell Forward Scattered Electron 6 Inelastic Collision Emitted X-ray N shell M shell Secondary Electron L shell K shell Forward Scattered Electron 7 X-ray Transitions N shell g Transition M shell L shell K shell β Transition α Transition 9 K α Transition N shell M shell K α Transition 10 L shell K shell K β Transition K β Transition 11 N shell M shell L shell K shell L α Transition L α Transition 12 N shell M shell L shell K shell Inelastic Collision Summary • Primary beam electron interacts with atom • There is an energy transfer • An electron from the atom is liberated • Creates a secondary electron (SE) • SE has low energy • The atom is excited and wants to return to a relaxed state • Characteristic X-ray is emitted 13 Continuum X-rays -(Background Radiation) Incident Electron Beam Ejected Electron White radiation (Continuum) 14 Characteristic X-ray Emission Depth of X-rays of KV Vs.. Z (in Microns) SIMPLIFIED VALUES Z 4 5 11 12 13 14 19 20 22 24 24 26 27 28 29 30 32 38 40 42 46 47 79 15 SYMBOL Be C Na Mg Al Si K Ca Ti Cr Mn Fe Co Ni Cu Zn Ge Sr Zr Mo Pd Ag Au ELEMENT Beryllium Carbon Sodium Magnesium Aluminum Silicon Potassium Calcium Titanium Chromium Maganese Iron Cobalt Nickel Copper Zinc Germanium Strontium Zirconium Molybdenum Palladium Silver Gold 5KV .5 .4 .9 .5 .3 .3 .5 .2 .01 10KV 1.7 1.2 3.1 1.7 1.1 1.2 2.9 1.6 .5 .25 .2 .2 .1 .09 .06 .02 20KV 5.3 3.9 10.0 5.6 3.6 4.1 10.8 5.9 2.0 1.2 1.2 1.0 .9 .85 .8 1.0 1.2 1.2 .2 30KV 10.5 7.7 19.2 11.1 7.1 8.2 21.9 12.2 4.1 2.6 2.4 2.2 2.0 1.9 1.9 2.3 3.0 4.8 1.7 .9 .5 .4 .2 EDS Hardware SEM Column Monitor (MCA Display) Dewar FET Preamp Pole Piece Detector Analyzer SCSI Window Collimator 16 Sample Stage X-section of Window & Crystal -500 to 1000 volts +,charges Detector Vacuum Detector Window 8u Be or 0.3u Polymer 17 Detector SiLi to preamplifier (FET) X-section of Window & Crystal -500 to 1000 volts +,charges Detector Vacuum Detector Window 8u Be or 0.3u Polymer 18 Detector SiLi to preamplifier (FET) 3.8 eV for each charge pair X-section of Window & Crystal -500 to 1000 volts AlKa X-ray= 1.48KeV in +,charges Detector Vacuum Detector Window 8u Be or 0.3u Polymer 19 Detector SiLi to preamplifier (FET) 389 charge pairs out @ 3.8 eV each= 1.48KeV How a Spectrum Grows Kα Kb Lα Counts .938 20 Energy 8.044 Qualitative analysis - Peak ID • Identification of all possible peaks • Auto or Manual Peak ID • Do not trust system : it is not 100% reliable • Be aware of : - spectral artefacts - skirt effects (low vacuum systems) 21 Qualitative analysis - Peak ID • Identification of an element • Place cursor exact on the top of the peak (expand scale) • L and M line can show shift due to poor separation of /ß line • First check main lines such as L or M lines • Check the other lines of the same element 22 • Heavy elements will produce K Qualitative analysis - Peak ID Manual ID (Z- Z+) steps along elements Auto ID EPIC table Peak Identification chart) (Element HPD : halographic peak deconvolution Peak Fit / (to check overlapping elements) Display possibilities Marker options : ESC - SUM 23 Qualitative analysis Deconvolution • Deconvolution = peak strip method • Requires well calibrated system • Check deconvolution for overlapping elements and missing elements • Press HPD / Peak Fit button • Check other lines 24 What Affects Quantitative Analysis • K Ratio • Atomic Number (Z) Accelerating Voltage • Absorption (A) Take-Off Angle • Fluorescence (F) Atomic Matrix/Topography 25 What Affects Quantitative Analysis • K Ratio Unknown / Known ratio or Unknown Intensity / Pure Element Intensity 26 What Affects Quantitative Analysis • K Ratio • Atomic Number (Z) Accelerating Voltage 1.5 to 10 Times excitation energy Or 2 times the highest energy peak 27 Why should the overvoltage be less than 10 to 20 times the lowest energy peak? High overvoltage means a high absorption condition and a small peak and poor statsitics (again) X-Ray Generated Volume 28 Why should the overvoltage be at least 1.5 for the highest energy element? X-Ray Vol. Low overvoltage means a small, poorly excited peak and poor statistical quality in the spectrum Electron Volume 29 What Affects Quantitative Analysis • K Ratio • Atomic Number (Z) Accelerating Voltage • Absorption (A) Take-Off Angle 30 Take-off Angle • The take-off angle is the angle between the x-ray trajectory and the sample surface. • The angle is a combination of detector angle, its position, sample working distance, and sample tilt. • Typical angles will range from 25 - 40 degrees 31 Normal Take-Off Angle EDS Detector 35 º T-O Normal take -off Angle Allows Low Energy X-rays to Become Absorbed in Specimen 32 Greater Take-Off Angle EDS Detector Greater T-O Greater Take-Off Lets Low Energy X-rays Escape 33 Sample position is extremely important • Working distance is very important • Optimum sample position eucentric position = 10 mm (Sirion = 5 mm) • Field of view of the EDX detector : 9 -12 mm (Sirion : 5 - 6 mm) • Below 12 mm : inhomogeneous 34 Ideal (Set-up) Detector Geometry Scale =50, EA = 35, Azimuth=45, ID= 10 (5) Intersection Distance/ Working Distance Elevation Angle 35 Various Detector Geometries ID EA WD > ID TOA > EA 36 WD < ID TOA < EA Tilt > 0 TOA > EA • Intersection Distance • Elevation Angle • Working Distance • Take-Off Angle What Affects Quantitative Analysis • K Ratio • Atomic Number (Z) Accelerating Voltage • Absorption (A) Take-Off Angle • Fluorescence (F) Atomic Matrix/ Topography 37 Fluorescence EDS Detector 35 º T-O High Energy X-ray Excites a Lower Atomic Number Atom 38 Other Issues with Quantitative Analysis BSE-image useful to determine if sample homogeneous SE-image 39 BSE-image Quantitative Analysis- Background Subtraction • Background needs to be removed before quantification • Auto or Manual method (new method Conc. Method - v3.2) • Manual method may improve BKGD fit • Background shape will affect the quantitative results : normally a very small effect • Pay attention to absorption edges 40 • Glasses and minerals : Si absorption edge Quantitative analysis - Background Subtraction 41 Quantitative analysis - Deconvolution • Halographic peak fit procedure (HPD / Peak Fit) • Suitable to separate overlapping peaks • Sensitive to peak shift • Requires well calibrated system • Check deconvolution for missing elements • In case of poor results or 42 Quantitative analysis - Matrix correction • Remove Background • Calculation of net peak intensity • K-ratio calculation • K-ratio = int. unknown peak / int. standard peak • Matrix correction for Z - A - F • Conc. = K-ratio / [ ZAF] • Different correction models 43 Quantitative analysis - Matrix correction • Different correction models to calculated ZAF factors • Two models : - ZAF correction model - Phi-Rho-Z correction model • Phi-Rho-Z model very suitable for light element quantification 44 Quantitative analysis - Matrix correction • Standardless analysis (normalise to 100%) • Flexible, kV independent • System needs kV to make a matrix correction • Every identified peak will be quantified • EDAX standardless calculation : WT% = Intensity measured /(P.E.I.F). X (SEC) ______________________________ (Z.A.F) 45 • Quantitative analysis standard less 2 different methods methods • Standardless (normalisation to 100%) • Default SEC (standardless element co-efficient) • All SECs are set to 1.0 • Improved Standardless : updating the SECs - User table • Normally a few elements are updated to create more accuracy using standards • Only the updated elements are more accurate : rest still standardless 46 • Mainly used for light elements- can be dangerous!!! Quantitative analysis: SEC • SEC = standardless Element Coefficient • All default SECs are set to 1.0 • SEC’s have to be changed for the lighter elements only (B, C, N, O, F) 47 Poor Quantitative Analysis Conditions • Several situations where the calculation of the ZAF factor does not apply : - unsupported thin film - thin film on substrate - inclusion or particle - biological sample - particle on thin foil 48 Quantitative analysis - Sources of errors • Situation where the ZAF factors does not apply • X-ray interaction volume larger than phase size • Wrong coating (preferable Carbon-coating) • Poor statistics (acquisition time / countrate to low) • High count rate (high dead time) • Overlapping elements (trace elements) • Energy calibration errors • Improper background selection • Irregular specimen surfaces • Skirt effect (low vacuum SEM / ESEM) 49 Quantitative analysis - Summary • Background subtraction • Deconvolution : needs well calibrated system • Matrix correction models : ZAF,Phi-Rho-Z and Phi-ZAF • Several situations where the ZAF factors do not apply • Several errors possible 50 Effects of Specimen Surface on X-ray Emission EDS Detector Electron Beam Backscatter electrons Fluorescence X-rays Specimen Matrix Absorption of x-rays 51 Interaction volume Directionality Is a Major Effect Detector Direction A B C sample stage/mount Topography has a significant effect on spectrum count rate and on composition (take-off angle and absorption effects) 52 The Effect of Topography A= Lower low end peaks B= Normal C= Higher low end peaks A Take-off angle is highest at C and lowest at A. 53 B C 3 different spectra at 3 locations on the same particle with a uniform composition. Real Time • Real Time = Live Time + Dead Time (Real Time = Clock Time) • Live Time - time when detector is alive and able to receive an x-ray event • Dead Time - time when the detector or preamplifier is unable to accept a pulse because it is busy processing or rejecting an event(s). 54 X-ray Generation - Continuum radiation • Continuum radiation = Bremsstrahlung or background radiation due to inelastically scattering • Observed fall out at low energies due to X-ray absorption enroute to the detector • Background needs to be removed for quantification 55 Spectral artifacts - Sum peaks • Sum peaks due to pulse pile-up effects • Two X-rays are entering detector at the same time • Sum of the energies is seen as one energy • Sum peaks depending on (to high) count rates for the corresponding amplification time • Not only pure elements : combinations possible • Prevent sum peaks by keeping countrate in balance with the chosen Ampl.Time (Dead time 25 - 35%) 56 Spectral artifacts - Sum peaks Sum peak Cr Ka = 2 x 5.411 = 10.822 KeV 2 times line energy 57 Spectral artifacts - Escape peaks • Escape peak : result of losing Si K energy in the Si-dead layer of the crystal • Si K line = 1.74 KeV • Remaining energy is original energy minus Si K energy • Difficult to identify • Intensity of the escape peak belongs to the main peak • S/w can correct for escape peaks • Example : Fe escape peak = same position 58 Artifacts- Escape Peaks Si @ 1.74 SiLi crystal 59 Spectral artifacts - overview Artificial spectrum of Fe, showing background shape, escape and sum peaks and the absorption edge 60 The Effect of Detector Time Constant At faster time constants, the throughput is increased but the resolution broadens. Fast time constants are commonly used for mapping but not for the collection of spectra with subtle overlaps. 61 Spectral artefacts - Dead time • Dead time = system is busy with pulse shaping • Dead time = relationship of input and output count rate • Dead time depending on amplification time (TC) • Reasonable dead time 25 - 35 % • High dead time : system is slow • Result : sum peaks and peak broadening • Use live-seconds (corrected for dead time) • Keep dead time in balance with chosen TC 62 Calibration of EDS 63 X-ray peaks must be located accurately on the energy axis Therefore calibration needed Automatic s/w procedure: zero and gain adjustments Use two elements: Al and Cu Calibrate using a countrate as under normal operations Optimum countrate: dead time 25 35 % Calibration (cont.) • Calibration of amplification times • Slow ampl.time is used for quantification: good resolution (130 - 138 eV) • Fast ampl. time used for mapping: poor resolution (145 - 180 eV) • Calibration every 2 - 4 weeks (if temperature is constant) 64 Calibration Control page X-ray lines of Al (1) and Cu(2) Maximum full scale counts Number of attempts detector resolution (Mn Ka line) 65 Spectral artefacts - Warming of the detector Two type of dewars : 2.5 litre and 10 liter 66 • Spectral artefacts - Warming of the Large dewar (10 L):detector always LN2 • Not designed to run dry • Once a year let it run dry (EDAX tip 22) • Small dewar (2.5 L) : fill when needed • Designed to run dry* • When warming up : low end noise peak • Bias light still green • Bias light will turn red when no LN2 67 present anymore Warming Detector • As the detector warms the noise peak widens and may appear in the spectrum as a low-end noise peak. • All peaks will broaden and may shift in energy • Also note large incomplete charge collection area to the left of the Cu peak 68 EDS Summary • Quick lnformation • Consistency is a must • Accuracy is poor but repeatability is near perfect • Ignore multiple decimal points, round out to nearest whole number for consistency 69 Spectral artefacts - Peak overlap • EDS poor resolution : result peak overlap • Difference at least 60 eV to separate lines • Classical example: Pb M S K 2.345 KeV - 2.307 KeV difference = 38 eV 70 Spectral artefacts - Poor counting •statistics Low count rate results in ‘noisy’ spectrum = poor statistics • ‘Noisy’ spectrum will introduce PEAK ID errors • System will identify noise (Auto ID) Solution • Long acquisition time - disadvantage : limited number of specimens • High countrate in combination with appropriate ampl. time disadvantage : poor peak separation • Find the ideal combination for your own samples • qual/quan work : 2000 cps (DT30%) at Amp. Time 50 • acquisition time 100 Lsec 71 - How to perform an analysis? 72 Use calibrated system Choose suitable place on the sample Make an image : - SE image shows topography - BSE image shows atomic number contrast Focus (calculation of the TOA) Acquire a spectrum (label spectrum during acquisition) Unknown sample : 30 - 15 - 5 kV to find all elements Identification of all elements (using HPD / Peak Fit) Store spectrum to HD Built library of pure elements (reference spectra) If needed : perform quantification Microanalysis under low vacuum conditions Poor Vacuum Microanalysis Considerations “Skirt” electrons have almost full beam energy: 1. X-rays from the probe spot (actual information) 2. X-rays due to the gas 3. Information of surroundings (=skirt) 74 2 1 3 Steps to reduce skirt effect • shorten BGPL (use the cone) • lower pressure • use high acceleration voltage (25kV) • use beam stop method • correct via software Gas Compensation Module (EDAX 3.1) (last choice) 75 EDS Geometry With and Without Cone Working Distance (WD) vs. Beam Gas Path Length (BGPL) Electron beam Electron beam EDX EDX Detector Detector 10mm WD 10mm WD Sample Cone 76 No Cone EDS at Low Gas Pressures and Short BGPL (ESEM Configuration) ESEM Mode: 15kV, 2mm BGPL, 10mm WD, 1 torr Water Vapor Examples of data from Electron Flight Simulator 77 Microanalysis under low vacuum conditions • Low vacuum SEM : charge is eliminated by a gas (water, air or N2) • High kV possible, no limitation of excitation energies • Two major problems : - beam damage - beam spread (skirt effect) • Beam damage : because of high kV heating of sample • Beam spread : - Electron are scattered due to gas collision - X-ray generation outside the probe spot - X-ray information upto 500 micron from central spot (=skirt) 78 Microanalysis under low vacuum conditions • To reduce the skirt effect: • use short gas path (EDX cone in case of ESEM) • use high acceleration voltage (25kV) • use low pressure (0.1 - 0.3 mbar) • correct via s/w module (quant.) (Gas Compensation Module = GCM) • GCM available for ESEM in s/w version 3.1 (minimum particle size 20 um) 79 Mapping and Linescans Mapping and Line scans • Mapping and Linescan: via optional EDX s/w module EDX Multi element mapping program Linescan software covered by EDAX applications class 81 EDX Multi Element Mapping : Image collection and display Imaging: • • • 82 built-in scan generator built-in pixel averaging • High resolution images (8200 x 6400 pixels) • High resolution X-ray maps (2048 x 1600 pixels) • Up to 15 elements with simultaneous image collection • Overlay of maps, colour or grey levels • Full control of beam and stage Region of Interest Control Page Element with region of interest (keV window) Activated ROIs + is enabled (= activated) 83 Mapping – Windows or ROI 84 EDX Multi Element Mapping 85 Options for EDX Multi Element Mapping EDX Fast Mapping • • • Fast X-ray mapping: Continuous update during collection EDX Quantitative Mapping Quantitative mapping • True element distribution EDX Line Scan (by EDAX advanced class) • • • • 86 • Digital X-ray line scan Results must be transferred to MS Excel (EDAX advanced class) EDX Particle/Phase Analysis Automated area distribution and X-ray EDX Fast Mapping Fast X-ray mapping: Continuous update during collection 87 EDX Multi Element Mapping Quantitative Mapping Quantitative X-ray mapping: true element distribution in samples with overlapping peaks PbM = 2.35 KeV S K = 2.31 KeV 88 EDX Multi Element Mapping - Line Scan (EDAX advanced class ) Line scan in combination with Quant map also collection of quantitative line scans 89 EDX Particle/Phase Analysis (EDAX advanced class) Rapid, automated detection and characterisation of particles Chemical and morphological data Automated multifield run Classification in user defined classes Automated area distribution and X-ray classification 90 Microanalysis with Sirion Collimator with magnets W and ESEM-FEG instruments only SUTW or UTW Window with magnets to deflect BSE 92 If BSE reach the detector they will produce background anomalies --a hump in the background at high energies. EDS Operation of the Sirion • Semi-inlens system, using strong external magnetic field • Electron-trapless EDX detector collimator to give optimised X-ray collection (no magnets) • Backscatter electrons “trapped” by semi-immersed lens field • HR mode: BSEs will enter the detector EDX detector is “blinded” No X-ray microanalysis possible X-rays only below 5 kV (hardly any BSEs) 93 EDS Operation of the Sirion(cont.) • UHR mode : • Strong magnetic field to trap BSEs • Skirt effect caused by reflecting BSEs • X-ray microanalysis possible (skirt effect) Minimum magnification 1100x At 5 kV • EDX mode : Dedicated EDX-mode to trap BSEs in the field • Minimum magn. 130x (easy navigation) • Full range of kV at 5 mm WD Best EDX results 94 No skirt effect EDS Operation of the Sirion - Summary • Three modes of operation • HR mode • UHR mode • EDX mode • EDX only possible under UHR and EDX mode • Best results with EDX mode • SS-BSD can limit the “field of view” for EDX • optimum EDX WD around 5.5 mm 95 End of Quantrain 7.1 Options EDS 96 Title slide sub-title Confidential