12 and 15 Lead Acquisition STEMI Recognition Class 12 Lead Rapid Acquisition • Module 1 Electrode Location • Module 2 Electrode Placement • Module 3 How to do a 15 Lead • Module 4 Demonstrations • Module 5 Reducing Artifact • Module 6 Tips and Techniques Module 1 Electrode Location Electrodes must be placed in the proper position to obtain an accurate 12 Lead ECG Module 1 Electrode Location 12 leads obtained from 10 electrodes • 4 on the limbs • 6 on the chest Module 1 Electrode Location Limb lead positioning is simple. The electrodes are placed off the torso, on the limbs. The most correct positioning is near the wrist and ankles. However, EMS generally places the electrodes on/near the torso to limit artifact while transporting. Module 1 Electrode Location Chest leads have specific anatomic locations V1, V2, V4 the rest are placed in relationship to these leads Module 1 Electrode Location Module 1 Electrode Location Module 1 Electrode Location Module 1 Electrode Location Module 1 Electrode Location Module 1 Electrode Location Module 1 Electrode Location Module 1 Electrode Location Module 2 Electrode Placement •The key to correctly placing the chest electrodes is finding the 4th intercostal space Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Here’s another approach to locating the 4th intercostal space Module 2 Electrode Placement Best view laterally Module 2 Electrode Placement • Locate the supersternal notch (1) at the top of the manubrium • Palpate down appox 2” until you find the sternal angle, slide your finger laterally to the right, you’re finger is now on the 2nd rib •Palpate down into the 2nd 3rd and 4th intercostal space Module 2 Electrode Placement After placing V2, palpated down to the 5th intercostal space midclavicular line and place V4 Module 2 Electrode Placement 15 Leads are simple Remove V4 and move it 5th intercostal space, midclivicular on the Right side of Pt’s chest V1 V2 V3 V4R V4 Module 2 Electrode Placement 15 Leads are simple Remove V4 and move it 5th intercostal space, midclavicular on the Right side of Pt’s chest Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 2 Electrode Placement Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations If you were the patient, where would you prefer to have you’re 12 Lead done, in the house or in the truck? Module 3 Demonstrations There’s a lot of people out there and they can see into the ambulance. You should obtain the 12 lead in the house Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Module 3 Demonstrations Practice Time Module 4 Reducing Artifact Unless you have a clear ECG to analyze, all you interruptive skills are of little use Module 4 Reducing Artifact Stress labs obtain clear ECG’s while the Patient is running on a treadmill We should be able to obtain a 12 lead while the Patient is laying still Module 4 Reducing Artifact As the heart depolarizes, an electrode on the Pt’s skin picks up the electrical activity Module 4 Reducing Artifact It can also pick up other electrical signals Module 4 Reducing Artifact To reduce artifact, we have to increase the heart’s signal and reduce the other electrical activity Module 4 Reducing Artifact Artifact Reduction Strategy: Helping the electrode gel to better penetrate the skin will increase the signal strength from the heart and reduce the signal strength from other sources. Module 4 Reducing Artifact Module 4 Reducing Artifact Remove hair with electric clippers Module 4 Reducing Artifact Module 4 Reducing Artifact Now the skin is prepared, we can attach our electrodes Module 4 Reducing Artifact Module 4 Reducing Artifact Module 5 Tips and Techniques Supine is the proper position, if the Pt will tollerate Module 5 Tips and Techniques When the Pt changes position, the heart moves within the chest. This can cause ECG changes similar to a misplaced electrode. Module 5 Tips and Techniques Do your best to maintain the modesty of a female Pt Module 5 Tips and Techniques You could try wide medical tape Module 5 Tips and Techniques A folded blanket or towel may help hold the electrodes in place Module 5 Tips and Techniques • Strand each lead out individually • When ECG cables are looped around IV lines, O2 tubing, BP cuff tubing, or dangling between squad bench and stretcher you will have more artifact • Make sure the Pt isn’t twiddling the ECG cables • If unable to lay supine for ECG, place them semifowlers and breathing normally • Do not allow Pt to prop themselves up by the arms or you will have muscle tremor artifact Module 5 Tips and Techniques • If the Pt is cold/shivering, cover with blanket or sheet prior to capturing the 12 Lead Module 5 Tips and Techniques Same Pt as before, covered with a towel Module 5 Tips and Techniques For some Pt’s obtaining a clear ECG will be difficult (e.g. respiratory distress Pt, sitting up) However, in most cases it is possible to a 12 Lead ECG with excellent, or at least acceptable data quality It just takes effort. A desire to obtain a clean 12 Lead and the knowledge to trouble shoot problems Module 5 Tips and Techniques Myocardial Infarctions are not like broken bones, and therefore, ECG’s are not like X-rays. If you’re treating a Pt with a broken hip. That x-ray could be taken now, 10 min’s from now, an hour from now and what would you see? A broken hip. With MI the events in the coronary artery can be changing moment by moment. The ECG can be very dynamic as well. There is a value to obtaining repeat ECG’s when you suspect MI. Making a habit of doing early and repeat ECG’s will help you identify a STEMI that could easily be missed. Module 5 Tips and Techniques Module 5 Tips and Techniques 12 lead Validation Does Lead I show Global Negativity? 12 lead Validation Limb Reversal! 12 lead Validation Look for R Wave Progression in the pre-cordial leads. The QRS should go from negative to more positive. Validate this 12 lead Module 5 Tips and Techniques This short course provides you with what you need to know in order to rapidly obtain a 12 Lead ECG that’s both clear and accurate Just as the case with ECG interpretation, acquisition also requires practice. After you’ve done this 20 or 30 times, you’ll become comfortable and confident FIN