IR change and dose rate linearity Mai-Britt K. Jørgensen, Lone Hoffmann, Lars H. Præstegaard, Maria F. Jensen Aarhus University Hospital Department of Medical Physics www.regionmidtjylland.dk Change of IR beam components Motivation The IR beams on Aarhus accelerators have never been changed. Therefore the voltage is much smaller than what Varian recommends. Varian recommends changing them ~once a year A mechanical way to change the DLG? Method Before and after change of IR components: 3 PF, 3 DRGS and 3 DRMLC tests with MLC initialisation in between. Change of IR beam components Results No change in deviations for the PF, DRGS and DRMLC tests Discussion Would you persue this isssue? Dose rate linearity Motivation Two types of dose rate control: Pulse drop servo and pulse leght servo 2 How well does the accelerator perform at different dose rates Method Arcs in service mode with different dose rates and same gantry speed Static gantry delivery in service mode All measurements performend with Farmer chamber in a square box of solid water Dose rate linearity: Pulse drop servo Dose rate linearity: Pulse length servo 2 0.60 0.40 0.20 Output 0.00 0 100 200 300 400 -0.20 -0.40 -0.60 -0.80 -1.00 DR Acc8 Acc5 500 600 700