Operating Manual BIOSTAT ® B BB-8821051 | BB-8821050 Fermenter | Bioreactor 85037-545-02 85037-545-02 The enclosed CD contains the operating instructions as a PDF file in various international languages. If the CD is missing, you can obtain a copy form us by specifying the order number: Order number: 85037-545-13 Sartorius Stedim Biotech GmbH Technical Editorial Department August-Spindler-Strasse 11 37079 Goettingen, Germany tech.pubs@sartorius-stedim.com www.sartorius-stedim.com System Requirements: – Windows, MacOS X – Browser with JavaScript enabled – PDF-Reader Use start.html to run the application Contents Contents 1. About this Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.1 Validity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.2 Accompanying Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.3 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 2. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2.1 General Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2.2 Informal Safety Measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 2.3 Symbols Used on the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 2.4 Intended Use and Foreseeable Misuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 2.5 Residual Risks from Use of the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.6 Danger Due to Electrical Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.7 Danger Due to Components Under Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 2.8 Hazards Arising from Bursting Culture Vessel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 2.9 Danger Due to Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 2.9.1 Danger Due to Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 2.9.2 Danger Due to Nitrogen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.9.3 Danger Due to Carbon Dioxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.10 Danger Due to Escaping Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.11 Danger Due to Hot Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 2.12 Danger Due to Rotating Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 2.13 Danger Due to Use of Incorrect Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 2.14 Safety and Protective Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 2.14.1 Main Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 2.14.2 Safety Valves and Pressure Reducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 2.14.3 Overheating Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 2.15 Personal Protective Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 2.16 Emergency Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 2.17 Operator Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 2.18 Personnel Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 2.18.1 Personnel Qualification Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 2.18.2 Personnel Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 2.18.3 Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 2.18.4 Unauthorized Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 2.18.5 Training and Instruction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 3. Device Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 3.1 Control/Supply Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 3.1.1 BIOSTAT ® B-MO Single / Twin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 3.1.2 BIOSTAT ® B-CC Single / Twin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 3.1.3 Control Elements and Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 3.1.4 Aeration Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 3.1.4.1 Modules “Additive Flow 2-Gas” (BIOSTAT ® B-MO Single / Twin) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 3.1.4.2 Modules “Additive Flow 4-Gas” (BIOSTAT ® B-CC Single / Twin) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 3.1.5 Peristaltic Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 3.1.6 External Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 3.2 Culture Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 3.2.1 UniVessel® Glass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 3.2.2 UniVessel® SU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 3.2.3 BIOSTAT ® RM 20 | 50 Rocker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 3.2.4 BIOSTAT ® RM 200 Rocker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 3.3 Stirrer Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Operating Manual BIOSTAT ® B 3 Contents 4. Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 4.1 User Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 4.2 System Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 4.3 Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 4.3.1 Device-specific User Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 4.3.2 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 4.3.2.1 Header Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 4.3.2.2 Work Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 4.3.2.3 Footer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 4.3.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 4.3.4 Overview of the Main Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 4.3.5 Overview of Selection Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 4.3.6 Direct Function Keys for Selection of Submenus . . . . . . . . . . . . . . . . . . . . 46 4.3.7 Selection Lists and Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 4.4 Password Protection of Individual Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 4.5 User Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 4.6 Bug Handling and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 4.7 Locking Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 5. Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 5.1 Inspection Upon Acceptance by the Recipient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 5.1.1 Report and Document Transport Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 5.1.2 Check Completeness of the Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 5.2 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 5.3 Instructions for Transport Within the Company . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 6. Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 6.1 Acclimatization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 6.2 Ambient Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 6.3 Work Surfaces and Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 6.4 Laboratory Energy Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 6.4.1 Electricity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 6.4.2 Manufacturer’s ID Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 6.4.3 Tempering Medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 6.4.4 Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 7. Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 7.1 Installation Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 7.2 Connecting the Device to the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 7.3 Connecting Laboratory Water Supply to the Device . . . . . . . . . . . . . . . . . . . . . . . . . 64 7.4 Connecting Laboratory Gas Supply to the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 7.5 Connecting the Stirrer Driver (only UniVessel® Glass / UniVessel® SU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 7.6 Connecting the UniVessel® SU Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 7.7 Connecting the Barcode Scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 7.8 Connect BIOSTAT ® RM Rocker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 7.9 Connecting the Sensor Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 7.10 Connecting the Tubes for Aeration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 7.11 Connecting the Temperature Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 7.11.1 Jacketed Culture Vessels / Single-walled Culture Vessels with Heating | Cooling Jacket (only UniVessel® Glass / UniVessel® SU) . . . . 69 7.11.2 Tempering the culture bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 7.11.3 Heating Jacket (only UniVessel® Glass / UniVessel® SU) . . . . . . . . . . . . . . 72 7.12 Connecting the Exhaust Cooling Hoses (only UniVessel® Glass) . . . . . . . . . . . . . . 73 7.13 Connecting external pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 7.14 Turning the Device On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 4 Operating Manual BIOSTAT ® B Contents 8. Preparing and Running the Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 8.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 8.2 Preparing the Glass Culture Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 8.3 Connecting Transfer Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 8.4 Filling the Culture Vessel with Culture Medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 8.4.1 UniVessel® Glass / UniVessel® SU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 8.4.2 Culture bags . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 8.5 Sterilizing Glass Culture Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 8.6 Preparing the Cultivation Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 8.6.1 Mounting the Agitator Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 8.6.2 Heating | Cooling Jacket Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 8.6.3 Installing the Heating Blanket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 8.6.4 Connecting the Aeration Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 8.6.4.1 Conducting Preliminary Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 8.6.4.2 UniVessel® SU Safety Valve Station . . . . . . . . . . . . . . . . . . . . . . . . 85 8.6.4.3 Connecting the “Additive Flow 2-gas” Aeration System . . . 86 8.6.4.4 Connecting the “Additive Flow 4-gas” Aeration System . . . 87 8.6.5 Preparing the Corrective Solution Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 8.7 Performing a Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 8.7.1 Setting Up the Measurement and Control System . . . . . . . . . . . . . . . . . . . 91 8.7.2 Guaranteeing Sterility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 8.7.3 Carrying out the Cultivation Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 8.8 “Main” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 8.8.1 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 8.8.2 Process Displays in the “Main” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 8.8.3 Direct Access to Submenus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 8.9 “Trend” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 8.9.1 “Trend” Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 8.9.2 Configuring the “Trend” Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 8.9.2.1 Setting the Trend Display for Parameters . . . . . . . . . . . . . . . . . 97 8.9.2.2 Setting the Parameter Display Range . . . . . . . . . . . . . . . . . . . . . 97 8.9.2.3 Resetting the Display Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 8.9.2.4 Setting the Trend Display Color . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 8.9.2.5 Defining a New Time Range as “Time Range” . . . . . . . . . . . . . 98 8.10 “Calibration” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 8.10.1 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 8.10.2 pH Calibration (conventional sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 8.10.2.1 General Information on the pH Sensor . . . . . . . . . . . . . . . . . . . . 99 8.10.2.2 “Calibration pH-#” Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 8.10.2.3 Perform Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 8.10.2.4 Direct Input of the Zero Offset and Slope . . . . . . . . . . . . . . . . 102 8.10.2.5 Perform Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 8.10.3 DO Calibration (conventional sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 8.10.3.1 General Information on the pO2 Sensors . . . . . . . . . . . . . . . . . 103 8.10.3.2 Submenu “Calibration pO2-#” . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 8.10.3.3 Perform Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104 8.10.3.4 Direct Input of the Zero Offset and Slope . . . . . . . . . . . . . . . . 105 8.10.4 Optical pH and pO2 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 8.10.4.1 Signal Quality of the Optical Probes . . . . . . . . . . . . . . . . . . . . . 106 8.10.4.2 Notes on Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 8.10.5 pH Calibration (optical sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 8.10.5.1 “Calibration pH-#” Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 8.10.5.2 Enter Initial Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 8.10.5.3 Performing Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 8.10.5.4 Configuring the Measurement Cycle for pH Measurement . . 111 Operating Manual BIOSTAT ® B 5 Contents 8.10.6 DO Calibration (optical sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111 8.10.6.1 Submenu “Calibration pO2-#” . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 8.10.6.2 Perform Initial Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 8.10.6.3 Performing Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 8.10.6.4 Configuring Measurement Cycles for DO Measurement . 114 8.10.7 Totalizer for Pumps and Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 8.10.8 Scale Taring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 8.11 “Controller” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 8.11.1 Functional Principle and Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 8.11.2 Controller Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 8.11.3 General Controller Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 8.11.4 Setpoint Profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 8.11.5 General Controller Parameterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 8.11.5.1 Output Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 8.11.5.2 Dead Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 8.11.5.3 Controller Parameterization Menu Screen . . . . . . . . . . . . . . . 124 8.11.5.4 PID Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 8.11.5.5 PID Controller Optimization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 8.11.6 Guide and Slave Controller (TEMP, JTEMP) Temperature Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 8.11.7 Temperature Measurement Without Slave Controller (TEMP) . . . . . . 127 8.11.8 Speed Regulation (STIRR) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 8.11.9 Antifoam Controller “FOAM” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129 8.11.10 Level Control with Level Sensor (LEVEL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 8.11.11 Adjusting the “VWEIGHT” Gravimetric Level Controller . . . . . . . . . . . . . 131 8.11.12 Gravimetric Filling Pump Controller “FLOW” . . . . . . . . . . . . . . . . . . . . . . . 132 8.11.13 Filling Pump Controller (SUBS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133 8.11.14 Gas Controller (Gas Filling Controller / Gas Flow Controller) . . . . . . . 134 8.11.15 pH Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135 8.11.15.1 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135 8.11.15.2 Controller Operator Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135 8.11.15.3 Parameter Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136 8.11.15.4 Change the process value source . . . . . . . . . . . . . . . . . . . . . . . . . 136 8.11.15.5 Acid, Base, and CO2 Supply-driven pH Control . . . . . . . . . . 136 8.11.16 DO Control Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137 8.11.16.1 Adjusting the DO (pO2) Sensor Process Value Source . . . . 138 8.11.16.2 pO2 Controller CASCADE (cascade controller) . . . . . . . . . . . . 139 8.11.17 DO (pO2) Controller ADVANCED (Polygon Controller) . . . . . . . . . . . . . . 142 8.11.18 Glucose Controller (GLUCO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150 8.11.19 Controller Functions on BIOSTAT ® RM Rocker . . . . . . . . . . . . . . . . . . . . . . 151 8.11.19.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151 8.11.19.2 Angle Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 8.11.19.3 Position Settings “POSITIONING” . . . . . . . . . . . . . . . . . . . . . . . . . 153 8.11.19.4 Aeration Rate (BIOSTAT ® RM 20 | 50 Rocker) . . . . . . . . . . . . . 155 8.11.20 Additional information — only for BIOSTAT ® RM 20 | 50 Rocker . . . . 156 8.11.21 Additional Information — For BIOSTAT ® RM 200 Rocker Only . . . . . . 157 8.11.21.1 Activating Single-Bag Function . . . . . . . . . . . . . . . . . . . . . . . . . . 158 8.11.21.2 Activating the Twin-Bag Function . . . . . . . . . . . . . . . . . . . . . . . 160 8.12 “Settings” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163 8.12.1 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163 8.12.2 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165 8.12.3 Measuring Range Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165 8.12.4 Manual Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167 8.12.4.1 Manual Operation for Digital Inputs . . . . . . . . . . . . . . . . . . . . . 168 8.12.4.2 Manual Operation for Analog Inputs . . . . . . . . . . . . . . . . . . . . . 172 8.12.4.3 Manual Operation of Analog Outputs . . . . . . . . . . . . . . . . . . . . 173 8.12.4.4 Manual Operation for Controllers (“Control Loops”) . . . . . 175 8.12.5 Manual Operation of Sequence Control (“Phases”) . . . . . . . . . . . . . . . . . 176 8.12.6 Externally Connected Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177 8.12.7 Service and Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178 6 Operating Manual BIOSTAT ® B Contents 9. Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179 9.1 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179 9.2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179 9.3 Hardware-related Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180 9.3.1 Fault Table “Contamination” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180 9.3.2 Troubleshooting Table “Counter Cooling System” . . . . . . . . . . . . . . . . . . 181 9.3.3 Troubleshooting Table “Aeration and Ventilation” . . . . . . . . . . . . . . . . . . 181 9.4 Process-related Faults / Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182 9.4.1 Alarm Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182 9.4.2 “Alarm Overview” Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183 9.4.3 Process Value Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183 9.4.4 Alarms for Digital Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186 9.4.5 Alarms, Meaning and Corrective Measures . . . . . . . . . . . . . . . . . . . . . . . . . 187 9.4.5.1 Process Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187 9.4.5.2 System Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188 10. Cleaning and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189 10.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190 10.1.1 Cleaning the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190 10.1.2 Cleaning the Culture Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191 10.1.3 Cleaning and Maintaining Heating Jackets . . . . . . . . . . . . . . . . . . . . . . . . . 191 10.2 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193 10.2.1 Carry out Maintenance Work on Function Elements . . . . . . . . . . . . . . . 193 10.2.2 Safety Component Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193 10.2.3 Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194 11. Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196 12. Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197 12.1 General Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197 12.2 Hazardous Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197 12.3 Decontamination Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198 12.4 Disposing of the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198 13. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199 13.1 BIOSTAT ® B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199 13.1.1 Dimensions and Weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199 13.1.2 Safety Valves and Pressure Reducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199 13.1.3 Culture Vessels and Culture Bags . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 13.2 Energy Connections Inside the Laboratory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201 13.2.1 Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201 13.2.2 Process Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201 13.2.3 Cooling Water Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201 13.3 Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201 13.4 Stirrer Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202 13.5 External Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202 13.6 Ambient Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202 13.7 Water Hardness Conversion Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202 14. Conformity & Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203 14.2 GNU Licensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203 15. Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205 15.1 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205 15.2 Decontamination Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205 15.3 Setup Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207 Operating Manual BIOSTAT ® B 7 About this Document 1. About this Document All information and instructions in this operating manual have been compiled in consideration of applicable standards and guidelines, the state of technology and our many years of experience and knowledge. These operating instructions provide you with all the information necessary to install and operate the BIOSTAT ® B-MO and BIOSTAT ® B-CC (referred to as “device” in the following). The operating instructions inform you about how to −− operate the device according to safety requirements, −− maintain the device according to guidelines, −− clean the device according to guidelines, −− take appropriate measures should a fault occur. The operating instructions must be read, understood and used by all personnel entrusted with the operation, maintenance, cleaning and troubleshooting of the device. This applies particularly to the safety instructions listed. ttPlease read these operating instructions carefully and completely before using this device. ttThese instructions are part of the product. Keep them in a safe and easily accessible place near the device's site of installation. ttIf the instructions should be lost or misplaced, please contact Sartorius for a replacement or download the latest version from our website: www.sartorius.com This description is based on known bioreactor models. The equipment supplied does not necessarily cover every available type of equipment; some may differ from the description while others may contain equipment not described here. The names, characteristics and information can differ from the ones in the technical documentation because these are each adapted to the equipment supplied. Documentation on customer-specific equipment can be delivered with the documentation provided to the customer and | or separately supplied and are available on request. The device may only be used under the operating conditions described in the Technical Data Sheet [ Chapter “3. Device Overview“]. The user must be qualified to handle the device, media and cultures and be aware of the hazards potentially associated with the process. The process may make it necessary to equip the device or the work area with additional safety equipment or to take other measures for the protection of personnel and the work environment. The documentation does not include more detail about such conditions or legal or otherwise obligatory guidelines. Safety and danger instructions in the documentation only apply to this unit and supplement the guidelines of the operator in the work area for the process in question. In addition to the operating manual, all generally valid, legal and otherwise binding regulations for accident protection and environmental protection of the country of use must be observed. 8 Operating Manual BIOSTAT ® B About this Document The operating manual must always be kept at the place of use of the device. −− Any guarantees and warranties shall apply to manufacturing-related defects and malfunctions. −− The unit is designed for usual laboratory conditions and techniques. 1.1 Validity These operating instructions apply to the BIOSTAT ® B-MO (microbial), BIOSTAT ® B-CC (cell culture) in single and twin variants in combination with the following culture vessels (operating volume): −− UniVessel® Glass, single-walled (SW) | double-walled (DW): −− 1 L −− 2 L −− 5 L −− 10 L −− UniVessel® SU Single Use Bioreactor, single-walled (SW): −− 2 L −− BIOSTAT ® RM 20 | 50 Rocker with culture bags: −− Flexsafe® RM 1L basic | optical | perfusion −− Flexsafe® RM 2L basic | optical | perfusion −− Flexsafe® RM 10L basic | optical | perfusion −− Flexsafe® RM 20L basic | optical | perfusion −− Flexsafe® RM 50L basic | optical | perfusion −− BIOSTAT ® RM 200 Rocker with culture bags: −− Flexsafe® RM 100L basic | optical | perfusion ATF −− Flexsafe® RM 100L basic | optical | perfusion ATF A detailed overview of the available models can be found in [Æ Chapter “3. Device Overview“]. 1.2 Accompanying Documents The operating manual describes how to operate the device along with the standard equipment intended for this purpose. In addition to this operating manual, you can find all necessary technical documentation on the device, e.g. P&I diagrams, a spare parts list, installation plans, technical drawings etc., in the folder “Technical Documentation”. If customer-specific modifications were implemented, the appropriate documents can either be integrated into the “Technical Documentation“ folder or they can be delivered together with the device as separate documents. If the documents supplied do not match the equipment delivered to you or if any documents are missing, please contact the person responsible for your area or get in touch with Sartorius Stedim Biotech directly. In addition to this operating manual, please observe the information in the documentation for the culture vessels. Operating Manual BIOSTAT ® B 9 About this Document 1.3 Symbols Used As a means of instruction and direct warning against hazards, all text statements to be particularly noted in these operating instructions will be marked as follows: This instruction denotes a possible danger with medium risk that death or severe injury may result if it is not avoided. This symbol denotes a possible danger with risk that moderate or minor injury may result if it is not avoided. This symbol denotes a danger with low risk that could result in property damage if the risk is not avoided. This symbol −− is an indication of a function or setting on the device. −− that caution should be exercised while working. −− identifies useful information. The following presentations will also be used: −− Texts that follow this mark are lists. tt Texts that follow this mark describe activities that must be carried out in the specified order. yy Texts that follow this mark describe the result of an action. “ ” Texts in quotes are references to other chapters or sections. [Æ] Texts following this symbol make reference to other chapters, sections or documents. 10 Operating Manual BIOSTAT ® B Safety Instructions 2. Safety Instructions Failure to observe the following safety precautions can have severe consequences: −− Danger to personnel due to electrical, mechanical or chemical influences −− Failure of important functions of the unit Read the safety precautions and dangers listed in this section thoroughly before starting the unit. In addition to the instructions in this operating manual, also observe all generally valid safety and accident protection guidelines. In addition to the instructions in this operating manual, the operator/user must observe all existing national workplace, operating and safety guidelines. Existing internal plant guidelines must also be observed. 2.1 General Safety Instructions −− The unit may only be set up, started or serviced after gaining familiarity with this operating manual. −− Use the device only for its intended purpose [Æ Chapter “2.4 Intended Use and Foreseeable Misuse“]. −− The device is not ATEX (ATmosphère EXplosive)-certified. The unit may not be operated in potentially explosive atmospheres. −− During operation of the unit, do not permit any work method that hinders the safety of the unit. −− Always keep the working area of the unit clean and orderly, in order to avoid danger from dirt or scattered parts. −− Always squat to work on low-lying components; do not bend over. Carry out work on high components in an upright, straight body position. −− Do not exceed the technical performance data (see data sheet for the unit). −− Keep all safety precautions and danger descriptions at the unit in legible condition and replace them as needed. −− Operation as well as work on the device may only be carried out by trained personnel. −− Do not start the unit if other people are in the danger zone. −− In case of malfunction, immediately stop the unit. Have the fault corrected by appropriately trained personnel or by your Sartorius Service office. Operating Manual BIOSTAT ® B 11 Safety Instructions 2.2 Informal Safety Measures −− Always keep the operating manual in the location where the unit is in use. −− In addition to the operating manual, follow all general and local regulations for accident prevention and environmental protection. 2.3 Symbols Used on the Unit The following symbols are used on the device: Especially hazardous area or hazardous use of the peristaltic pump! Follow the instructions in the accompanying documents. Danger of crushing your hand or fingers in the peristaltic pump – do not attempt to put your hand inside rotating parts, e.g. the pump head! To thread the tubing or adjust the rollers, be sure to switch off the pump first. −− These pumps must be switched off in the measurement and control system (in the main menu or associated controller operator screen). Danger of burns! Motor and culture vessel equipment becomes hot during operation. −− Avoid accidental, unintentional contact. −− Use protective gloves when operating the equipment. −− Prior to removing the motor from the stirrer drive, allow the motor housing to cool down. −− Let the culture vessel and equipment cool down before carrying out assembly work. Some of the markings and labels on the equipment were attached by the equipment’s manufacturers. These may not always correspond to the safety labels commonly used by Sartorius Stedim Biotech. Be sure to follow the instructions given in this manual. Keep all safety precautions and danger descriptions at the unit in legible condition and replace them as needed. 12 Operating Manual BIOSTAT ® B 2.4 Safety Instructions Intended Use and Foreseeable Misuse The operational safety of the unit is only ensured when it is used for its intended purpose and operated by trained personnel. The device is used for the cultivation of prokaryotic and eukaryotic cells solely in aqueous solution. The device may only be used indoors. The intended use also includes: −− That all instructions in the operating manual are followed. −− Observation of inspection and maintenance intervals. −− The use of oils and greases which are suitable for use with oxygen. −− The use of operating and auxiliary materials according to applicable safety guidelines. −− Observation of operating and repair conditions. All other applications are not considered the intended use. They may include risks that cannot be estimated and they are the sole responsibility of the operator. Claims of any kind resulting from damages due to use other than the intended use are excluded. Sartorius Stedim Biotech bears no liability for use other than the intended use. Danger due to use other than the intended use! Any use beyond and/or other than the intended use of the unit can lead to dangerous situations. The following uses are considered not to be the intended use and are strictly forbidden: −− Processes using biological materials in Safety Classes 3 and 4 −− Cultivation in non-aqueous solutions −− Operation of the device outside the performance limits −− Operation outdoors Operating Manual BIOSTAT ® B 13 Safety Instructions 2.5 Residual Risks from Use of the Unit This unit is a state-of-the-art machine and was designed and built in accordance with recognized safety standards. Nevertheless, use of the device may pose bodily or physical harm to users or third parties as well as potentially cause impairments to the testing system itself or other material damage. Anyone assigned to the installation, commissioning, operation, maintenance or repair of the device must have read and understood the operating manual. The device may only be used: −− For its intended use, −− with its safety systems in perfect working order, −− by qualified and authorized personnel. Furthermore, the following rules must be followed: −− All moving parts must be lubricated as required. −− All screw connections must be checked at regular intervals and tightened if necessary. 2.6 Danger Due to Electrical Power Danger to life caused by electrical voltage! Electrical switching elements are installed in the device. Contact with parts under voltage represents a direct danger of death. Damage to the insulation or individual components can be fatally dangerous. −− Never open the device. The device may be opened only by authorized personnel of the Sartorius Stedim Biotech Company. −− Work on the electrical equipment of the device may only be carried out by Sartorius Service or authorized technicians. −− Check the electrical equipment of the device regularly for defects such as loose connections or damage to the insulation. −− In case of defects, turn the power supply off immediately and have the defects corrected by Sartorius Service or authorized technicians. −− If work is required on parts under voltage, ask a second person to be ready to turn off the unit’s main switch if needed. −− During all work on the electrical equipment, disconnect it and check that voltage is no longer applied. −− During maintenance, cleaning and repair work, turn the power supply off and secure it against reactivation. −− Keep moisture away from parts under voltage, as it can lead to short circuits. −− Have the following tested by a qualified electrician according to the national regulations: electrical components, connection cables with plugs and extension and device connector cables with their sockets, given they are used. 14 Operating Manual BIOSTAT ® B 2.7 Safety Instructions Danger Due to Components Under Pressure Danger of injury from leaked substances! If individual components are damaged, gaseous and liquid materials may escape under high pressure and e.g. cause injury to eyes. Therefore: −− Do not start the culture vessel without a safety valve or comparable overpressure safety (e.g. a burst disk). −− Turn the unit off and secure it against reactivation when working on pressurized components. −− Release the pressure from system sections and pressure lines to be opened before starting any repair work. −− Regularly check all lines, hoses and connections under pressure for leaks and externally detectable damage. 2.8 Hazards Arising from Bursting Culture Vessel Danger of injury from shattered glass! Damaged and bursting glass can cause cuts and eye injuries. Therefore: −− Provide operating personnel with training regarding glass breakage due to external influences. Ensure the culture vessel remains stable at all times. −− Wear personnel protective equipment. −− Make sure that the culture vessel is properly connected to the supply and control units. −− Ensure that the culture vessel remains within the maximum permissible pressure limits. −− Make sure that the cooling water flows back without pressure. −− Regularly check all lines, hoses and connections under pressure for leaks and externally detectable damage. 2.9 Danger Due to Gases 2.9.1 Danger Due to Oxygen Danger of explosion and fire! −− Keep pure oxygen away from flammable materials. −− Avoid sparks in the vicinity of pure oxygen. −− Keep pure oxygen away from ignition sources. −− Keep the whole aeration segment free from oil and grease. Make sure that the cooling water flows back without pressure. Operating Manual BIOSTAT ® B 15 Safety Instructions Reactions with other substances! −− Ensure that pure oxygen does not come into contact with oil or grease. −− Use only materials and substances suitable for use with pure oxygen. 2.9.2 Danger Due to Nitrogen Danger of suffocation due to escaping nitrogen! Escaping gas at high concentrations may force the air out of closed rooms and can cause loss of consciousness, leading to suffocation. −− Check the gas lines and culture vessels for leaks. −− Ensure good ventilation of the installation site. −− Keep a breathing device independent of ambient air ready for emergencies. −− If personnel appear affected by symptoms of suffocation, immediately provide them with a breathing device independent of ambient air. Bring them into fresh air, make them comfortable and keep them warm. Call a doctor. −− If a person stops breathing, initiate first aid measures with artificial respiration. −− Monitor the limit values at the system and in the building (sensors recommended). −− Check process gas lines and filters on leakages regularly. 2.9.3 Danger Due to Carbon Dioxide Danger of poisoning by escaping carbon dioxide! −− Check the gas lines and culture vessels for leaks. −− Ensure good ventilation of the installation site. 2.10 Danger Due to Escaping Materials Danger of scalding due to defective components! −− Inspect the unit before starting the process. −− Check the connections of containers and the connections to the supply unit. −− Regularly check hoses for leaky places and replace any leaking hoses. 16 Operating Manual BIOSTAT ® B Safety Instructions Danger of injury from escaping feed and culture media! −− Use the specified hoses only. −− Use hose fastenings on connecting pieces. −− Empty the feed hoses before loosening the hose connection. −− Wear personnel protective clothing. −− Wear safety glasses. Danger of contamination from escaping feed and culture media! −− Empty the feed hoses before loosening the hose connection. −− Wear personnel protective clothing. −− Wear safety glasses. 2.11 Danger Due to Hot Surfaces Danger of burns upon contact with hot surfaces! −− Avoid contact with hot surfaces like temperature-controlled culture vessels and motor housings. −− Block the danger zone off. −− Wear protective gloves when working with hot culture media. 2.12 Danger Due to Rotating Parts Danger of limbs being pulled into the rotation pump and crushed! −− Do not remove the safety mechanisms. −− Allow only qualified and authorized personnel to work on the device. −− Disconnect the device from power when performing maintenance and cleaning tasks. −− Block the danger zone off. −− Wear personnel protective equipment. 2.13 Danger Due to Use of Incorrect Consumables Danger of injury due to incorrect consumables! −− Incorrect or erroneous consumables can lead to damage, malfunction or the total failure of the unit, as well as affecting safety. −− Use only original consumables. Purchase consumables through Sartorius Stedim Biotech. You can find all necessary specifications for consumable materials in the folder “Technical Documentation“. Operating Manual BIOSTAT ® B 17 Safety Instructions 2.14 Safety and Protective Systems 2.14.1 Main Switch The main switch is on the operator side of the control unit. The main switch is simultaneously the emergency-off switch used to turn the unit on and off. 2.14.2 Safety Valves and Pressure Reducer Danger of injury from bursting culture vessels and lines! −− Do not start the unit without safety valves and pressure reducer or comparable overpressure safety equipment. −− Have your safety valves and the pressure reducer serviced regularly by Sartorius Service. −− Observe the information in the folder “Technical Documentation”. Overpressure Valve Aeration Segment Always use only the integrated aeration module. Overpressure valves are built into the device’s aeration segment for aeration of the sparger and overlay. The overpressure valves limit the aeration pressure. Different overpressure valves are installed depending on the type of culture vessel: −− UniVessel®: Limited to 1.0 bar −− BIOSTAT ® RM Rocker: Limited to 0.1 bar When using the UniVessel® SU, connect the UniVessel® SU safety valve station with the aeration segments of the BIOSTAT® B supply unit. The pressure is limited to 0.5 barg (7.25 psig) with the help of the UniVessel® SU safety valve station. Pressure Reducer Cooling System The pressure reducer is installed in the unit. The cooling water within the temperature control and exhaust systems is limited to 1.2 bar using the pressure reducer. 18 Operating Manual BIOSTAT ® B Safety Instructions 2.14.3 Overheating Protection Danger of burns from overheated components! If individual components are damaged, gaseous and liquid materials may escape under high pressure and e.g. cause injury to eyes. −− Do not start the unit without overheating protection. −− Have your overheating protection serviced regularly by the Sartorius Service. −− Observe the information in the folder “Technical Documentation”. The overheating protection within the unit limits the maximum permissible temperature for the temperature control system. The following temperature control systems can be used: −− Temperature control system water circulation system −− Temperature control system heating jacket 2.15 Personal Protective Equipment When operating the unit, personal protective equipment must be worn in order to minimize dangers to health. −− During work, always wear the protective equipment needed for that work. −− Follow any instructions posted in the work area pertaining to personal protective equipment. Wear the following personal protective gear during all work: Protective work clothes Protective work clothing is tight-fitting work clothing with low tear resistance, tight sleeves and without any projecting parts. It is primarily intended to protect you from being caught by moving machine parts. Do not wear rings, chains or other jewelry. Head covering To protect your hair from being caught and pulled into moving parts of the unit, wear a head covering. Safety gloves Wear safety gloves to protect your hands from process materials. Safety glasses Wear safety glasses to protect yourself from media escaping under high pressure. Safety boots Wear non-slipping safety boots to protect against slipping on smooth floors. Operating Manual BIOSTAT ® B 19 Safety Instructions 2.16 Emergency Instructions Preventive Measures −− Always be prepared for accidents or fire. −− Keep first aid equipment (bandages, blankets, etc.) and fire extinguishers handy. −− Familiarize personnel with accident reporting, first aid, fire extinguishing and rescue facilities. −− Keep entry and rescue routes free for rescue vehicles and rescue personnel. Measures to be Taken After Accidents −− Trigger an emergency stop at the main switch. −− Save personnel from the danger zone. −− In case of stopped heart or breathing, initiate first aid measures immediately. −− Report personal injuries to the first aid officer and an emergency doctor and/or the rescue service. −− Keep entry and rescue routes free for rescue vehicles and rescue personnel. −− Extinguish fires in the electrical equipment with a CO2 extinguisher. 2.17 Operator Responsibilities The device is used in the commercial sphere. The operator of the device is therefore subject to the legal obligations for workplace safety. In addition to the safety instructions in this operating manual, the safety, accident prevention and environmental protection regulations valid for the location of use of the device must also be observed. This applies particularly: −− The operators must inform themselves of the applicable workplace safety regulations and carry out a hazard evaluation to determine additional dangers that may result due to the special working conditions at the location of use of the unit. These must be expressed in the form of operational instructions for the operation of the unit (hazard prevention plan). −− During the entire period of use of the unit, the operator must check whether the operating instructions correspond to the current state of regulations and adapt them as needed. −− The operator must clearly regulate and determine responsibilities for operation, maintenance and cleaning. −− The operator may only allow trained and authorized individuals to work with the device. Trainees such as apprentices or auxiliary staff members are only allowed to work with the unit under supervision of qualified technicians [ Chapter “2.18 Personnel Requirements“]. −− The operator must ensure that all employees involved with the unit are suitable in terms of physical capacity, person and character to operate the unit responsibly. 20 Operating Manual BIOSTAT ® B Safety Instructions −− The operator must also ensure that all employees are familiar with the basic regulations concerning workplace safety and accident prevention, trained in handling the unit and have read and understood the operating manual. −− Furthermore, the operator must evaluate the safety-conscious work of personnel at regular intervals and demonstrate training of personnel and notification of danger. −− The operator must avoid stress situations during operation of the unit by means of technological and organizational work preparation. −− The operator must provide adequate workplace lighting in the place of operation of the unit in accordance with locally applicable workplace guidelines. −− The operator must provide personnel with personal protective equipment. −− The operator must ensure that no-one works on the unit whose reaction time is impaired, for example by drugs, alcohol, medications or the like. Furthermore, the operator is responsible for ensuring that the unit is always in technically perfect condition. The following therefore applies: −− The operator must ensure that the maintenance intervals described in this operating manual are observed. −− The operator must regularly have the safety systems tested for functionality. 2.18 Personnel Requirements Danger of injury if personnel qualifications are insufficient! Improper use can lead to significant personal injury and/or property damage. Have all activities performed only by qualified personnel. Only those individuals are permitted of whom it can be expected that they will carry out their work reliably. No-one may work on the unit whose reaction time is impaired, for example by drugs, alcohol, medications or the like. 2.18.1 Personnel Qualification Requirements In the operating manual, the following qualifications are cited for different areas of activity: Trainee A trainee such as an apprentice or an auxiliary staff member does not know all the dangers that can occur during operation of the unit. They may only perform work on the unit under the supervision of technicians. Trained Person A trained person has been informed in a training session by the operator about the tasks assigned to them and the possible dangers of improper behavior. Operating Manual BIOSTAT ® B 21 Safety Instructions Technician Due to his or her technical education, knowledge and experience, as well as knowledge of applicable regulations, a technician is capable of carrying out tasks assigned to him or her and independently detecting and avoiding possible dangers. Electrician Due to his or her technical education, knowledge and experience, as well as knowledge of applicable standards and regulations, an electrician is capable of carrying out work on electrical equipment and independently detecting and avoiding possible dangers. 2.18.2 Personnel Responsibilities Before undertaking any work with the unit, all personnel are obliged to: −− Pay attention to the basic occupational safety and accident prevention regulations. −− Read the safety instructions and warnings in this operating manual and confirm by signature that they have understood. −− Follow all safety and operation instructions in this operating manual. 2.18.3 Responsibilities The areas of responsibility of the personnel concerning the operation, maintenance and cleaning of the unit must be clearly defined. 2.18.4 Unauthorized Personnel Danger to unauthorized personnel! Unauthorized personnel who do not meet the qualification requirements for personnel do not know the dangers in the work area. Therefore: −− Keep unauthorized personnel away from the work area. −− In case of doubt, speak to personnel and instruct them to leave the work area. −− Stop work as long as unauthorized personnel remain in the work area. 22 Operating Manual BIOSTAT ® B Safety Instructions 2.18.5 Training and Instruction Personnel must regularly receive instruction from the operator. Log the performance of this training for better tracking. Date Name Type of training Training given by Signature Operating Manual BIOSTAT ® B 23 Device Overview 3. Device Overview The devices BIOSTAT ® B-MO, BIOSTAT ® B-CC are designed for the cultivation of microorganisms and cells in discontinuous and continuous processes. They were designed for cultivating microorganisms and cells at various reactor volumes. The devices can be used to conduct studies on developing and optimizing fermentation procedures and to perform limited-volume production fermentation processes in a reproducible way. The measurement and control systems permit measurement, control and evaluation of process parameters (such as temperatures, pH and pO2 values) online, independent monitoring of the progress of fermentation or cultivation in each culture vessel (Twin variant), as well as reproducible process control in combination with the SCADA software BioPAT®MFCS/win by performing the user-defined parameter sets specified in the recipes. The device is made up of the following components (the actual equipment depends on configuration): Control Unit −− Control unit in single and twin models −− DCU measurement and control system −− Aeration modules “MO” (BIOSTAT ® B-MO) for aeration of the culture vessel with air and oxygen, e.g. for microbial cultures −− Aeration modules “CC” (BIOSTAT ® B-CC) for aeration of the culture vessel with air, O2, N2 and CO2, e.g. tissue cell cultures containing animal cells in the suspension culture −− Temperature control modules with associated fittings (e.g. heating jacket and cooling finger) −− Cooling water circuit for the exhaust cooler −− Exhaust filter heating −− Peristaltic pump modules (up to 4 modules for single version | up to 8 modules for twin version) Culture Vessels −− UniVessel® Glass single-walled, jacketed, UniVessel® SU, BIOSTAT ® RM Rocker −− Culture vessel volume −− UniVessel® Glass: 1 L, 2 L, 5 L, 10 L −− UniVessel® SU: 2 L −− RM culture bags: 1L, 2L, 10L, 20 L, 50 L, 100 L, 200 L −− Equipment components for microbial cultures and cell cultures Stirrer Driver −− Top drive with direct stirrer drive motor −− Drive with magnetic coupling between motor and stirrer −− 6-blade disk impeller or 3-blade disk impeller The illustrations in the following sections show general system configurations. The actual equipment depends on your configuration and may deviate from the devices shown here. 24 Operating Manual BIOSTAT ® B 3.1 Control/Supply Units 3.1.1 BIOSTAT ® B-MO Single / Twin Device Overview Fig. 3-1: Example of the BIOSTAT ® B-MO Twin with UniVessel® Glass 3.1.2 BIOSTAT ® B-CC Single / Twin Fig. 3-2: Example of the BIOSTAT ® B-CC Twin with UniVessel® Glass Operating Manual BIOSTAT ® B 25 Device Overview Variants with different culture vessels 26 Operating Manual BIOSTAT ® B UniVessel® Glass – UniVessel® Glass UniVessel® SU – UniVessel® SU BIOSTAT ® RM 20 | 50 Rocker – BIOSTAT ® RM 20 | 50 Rocker UniVessel® Glass – BIOSTAT ® RM 20 | 50 Rocker UniVessel® Glass – UniVessel® SU UniVessel® SU – BIOSTAT ® RM 20 | 50 Rocker Single-Bag-configuration Twin-Bag-configuration BIOSTAT ® RM 200 BIOSTAT ® RM 200 3.1.3 Device Overview Control Elements and Connections Fig. 3-3: Front view / detail view BIOSTAT ® B-CC Twin Pos. Description BIOSTAT ® B-CC (MO) Twin with UniVessel® Glass / UniVessel® SU BIOSTAT ® B-CC Twin with BIOSTAT ® RM Rocker 1 Operator display (touch panel) Operator display (touch panel) 2 Main switch / power switch Main switch / power switch 3 Flow meter (rotameter) Flow meter (rotameter) 3a AIR “Overlay” (BIOSTAT ® B-CC)* - 3b AIR “Sparger” (BIOSTAT ® B-CC, MO) AIR “Overlay” 3c O2 “Sparger” (BIOSTAT ® B-CC, MO) O2 “Overlay” 3d N2 “Sparger” (BIOSTAT ® B-CC)* N2 “Overlay” 3e CO2 CO2 “Overlay” 4 USB data interface USB data interface 5 Peristaltic pump Peristaltic pump “Sparger” (BIOSTAT ® B-CC)* * BIOSTAT ® B-MO screen Operating Manual BIOSTAT ® B 27 Device Overview Fig. 3-4: Back view / detail view BIOSTAT ® B-CC Twin Pos. Description 1 Power connection / Potential equalization 1a Potential equalization (if available in lab) 1b Power connection 2a Network port 2b Common alarm connection 3a Tempering medium inlet d 10 mm, laboratory side connection 3b Tempering medium return d 10 mm, laboratory side connection 4 Aeration (laboratory side connection) 4a Air (BIOSTAT ® B-CC, MO) Serto gland d 6 mm 4b O2 (BIOSTAT ® B-CC, MO) Serto gland d 6 mm 4c N2 (BIOSTAT ® B-CC)* Serto gland d 6 mm 4d CO2 (BIOSTAT ® B-CC)* Serto gland d 6 mm * BIOSTAT ® B-MO screen 28 Operating Manual BIOSTAT ® B Device Overview 1b 2b 2d 1 2f 2h 2 2j 2l 3 2a 2c 2e 2g 2i 2k 2m Fig. 3-5: Side view / detail view BIOSTAT ® B-CC, UniVessel® Glass, UniVessel® SU 3a 3c 3b 3d 3e 4 4 Pos. Description Comments 1 1a Aeration Overlay (BIOSTAT ® B-CC)* Serto gland d 6 mm 1b Sparger Serto-screw d 6 mm 2 2a 2b 2c Sensors Temp Serial-A pH/Redox-A 2d Serial-B 2e pO2-A 2f Serial-C 2g 2i 2j 2k 2l 2m 3 3a 3b 3c Foam Serial-D1/ Turbidity sensor, RS-232, Lemo plug Turb-1 Level Level sensor, M12 plug connection Ext.Sig. A/B External signal input, M12 female connector Pump-B External pump, M12 plug connection Ext.Sig. C/D External signal input, M12 female connector Pump-B External pump, M12 plug connection Temperature control | cooling Exhaust Exhaust cooling return, Serto gland d 10 mm Heating Blanket Heating jacket, Amphenol plug Exhaust Exhaust cooling inlet, Serto gland d 10 mm 3d Thermostat Temperature control return, Serto gland d 10 mm 3e 4 Thermostat Temperature control inlet, Serto gland d 10 mm Stirrer motor connection 2h 1a Temperature sensor, M12 plug connection Scales connection (FWEIGHT), RS-232, M12 plug connection pH sensor, Redox sensor, VP8 plug Scales connection (FWEIGHT / VWEIGHT), RS-232, M12 plug connection. pO2 sensor, VP8 plug UniVessel® SU Holder, BioPAT ® ViaMass, RS-485, M12 plug connection Foam sensor, M12 plug connection * BIOSTAT ® B-MO screen Operating Manual BIOSTAT ® B 29 Device Overview 1a 2a 2b 2c 1 2e 2d 2f 2 2g 2h 3 2i 3a 3b Fig. 3-6: Side view / detail view BIOSTAT ® B Pos. Description Comments 1 Aeration 1a Overlay 2 Sensors 2a pH-A Opt. Optical pH-sensor, VP8 plug 2b Serial-A Scales connection (FWEIGHT), RS-232, M12 plug connection 2c Serial-B Scales connection (FWEIGHT), RS-232, M12 plug connection 2d pO2-A Opt. Optical pO2 sensor, VP8 plug 2e Serial-C BIOSTAT ® RM Rocker, RS-232, M12 plug connection 2f Serial-D Serial interface, RS-232 | RS-485*, M12 plug connection 2g Ext.Sig. A/B External signal input, M12 female connector 2h Pump-B External pump, M12 plug connection 2i Pump-C External pump, M12 plug connection 3 Temperature control | cooling 3a Thermostat Temperature control return, Serto gland d 10 mm 3b Thermostat Temperature control inlet, Serto gland d 10 mm (BIOSTAT ® B-CC) Serto gland d 6 mm * Depending on the configuration of the device. 30 Operating Manual BIOSTAT ® B 3.1.4 Device Overview Aeration Modules The supply units of the devices can be equipped with a range of different aeration modules. Every device can only be equipped with one of the aeration modules described below. The laboratory gas supplies for each of the gases must be preset to 1.5 barg (21.76 psig). The pressure in the vessel supply lines is limited to max. 1 barg (14.5 psig) by the safety valves in the aeration modules. Supplemental Information The built-in variable area flow meters are calibrated to the following standard conditions. Calibration Parameters Gas type: Air Temperature: 20 °C = 293 K Pressure: 1.21 bar (absolute) When gases with deviating pressures pass through, higher or lower values can be displayed. These must be recalculated to evaluate the flow rates. The manufacturer of the flow rate meters make tables with conversion factors available. Using the conversion tables, flow rates for the different processes can be recalculated. Specific data for gas Density [kg/m3] Carbon dioxide (CO2) 1.977 Air (AIR) 1.293 Oxygen (O2) 1.429 Nitrogen (N2) 1.251 3.1.4.1 Modules “Additive Flow 2-Gas” (BIOSTAT ® B-MO Single / Twin) Aeration modules “MO” are used for air supply and oxygen enrichment, e.g. for microbial cultures. −− Supplies air and O2 to each culture vessel through 2/2-way solenoid valves. Flow regulated by DCU system pO2 controller: −− Operating modes available: “man”, “auto”, “off” in the operating menu. Fig. 3-7: BIOSTAT ® B-MO rotameter −− In the “man” operating mode, you can set the gas flow rate on the variable area flow meter. −− Sparger outlet for supplying gas to the culture medium. −− Up to two mass flow controllers for air (AIR) and O2. Operating Manual BIOSTAT ® B 31 Device Overview Connections supply unit: BIOSTAT ® B-MO Single: “Sparger-1” BIOSTAT ® B-MO Twin: “Sparger-1, -2” Fig. 3-8: BIOSTAT ® B-MO connections 3.1.4.2 Modules “Additive Flow 4-Gas” (BIOSTAT ® B-CC Single / Twin) Module “Additive Flow 4 Gas”, when a BIOSTAT ® RM Rocker is connected. Module “Additive Flow 5 Gas”, when a UniVessel® Glass and | or UniVessel® SU is connected. Aeration modules “CC” are used to supply up to 4 gases. These are by default: −− Supply of air −− N2 for decreasing the O2 content, or O2 for increasing the oxygen content; Fig. 3-9: BIOSTAT ® B-CC rotameter −− CO2 for adjusting the pH or use as a carbon (C) source Air and CO2 can be fed both into the medium in the culture vessel (sparger) and into the headspace (overlay), while the other gases are fed into the supply line for the culture medium (sparger) by default. These modules are designed for tissue cell cultures, e.g. suspension cultures that contain animal cells. They are also suitable for use with cultures with special aeration requirements (if CO2 needs to be used as the carbon source, e.g. in anaerobic bacteria or algae cultures). −− Regulates N2 and O2 flow by means of 2/2-way solenoid valves, controlled by the DCU system DO controller (pO2). −− Regulates CO2 flow by means of a solenoid valve that is controlled by the DCU system pH controller (acidity controller). −− Operating modes available for selection in the controller operating menu: man, auto, off −− Gas volume can be adjusted at the variable area flow meter or by means of optional mass flow controllers. Fig. 3-10: BIOSTAT ® B-CC connections −− Sparger outlet for supplying gas to the media and overlay for supplying gas to the headspace in the culture vessel. −− Up to four optional mass flow controllers. Connections supply unit: BIOSTAT ® B-CC Single: “Sparger-1” / “Overlay-1” BIOSTAT ® B-CC Twin: “Sparger-1, -2” / “Overlay-1, -2” 32 Operating Manual BIOSTAT ® B 3.1.5 Device Overview Peristaltic Pumps The peristaltic pump units WM 114 are located on the supply unit; they convey the correction media and nutrient media through hoses into the vessel. Up to 4 peristaltic pump modules can be installed on the BIOSTAT ® B Single. Up to 8 peristaltic pump modules can be installed on the BIOSTAT ® B Twin. The peristaltic pump modules can be installed on the supply unit in different specifications (see table below). Type Fig. 3-11: Peristaltic pump module WM 114 WM 114 Speed control 5 - 150 rpm 24 V | DC WM 114 Speed control 0.15 - 5 rpm 24 V | DC Inner Flow rate (ml/min) diameter Min Max of the hose Min Max 0.50 6.0 180 0.10 3.0 Flow rate (ml/h) 0.80 0.20 6.0 12.0 360 1.60* 0.70* 21.0* 42.0* 1,260* 2.40 1.45 43.5 87.0 2,610 3.20* 2.35* 70.5* 141.0* 4,230* 4.80 4.25 127.5 255.0 7,650 0.50 0.003 0.10 0.18 6 0.80 0.006 0.20 0.36 12 1.60* 0.021* 0.70* 1.26* 42* 2.40 0.044 1.45 2.61 87 3.20* 0.071* 2.35* 4.23* 141* 4.80 0.128 4.25 7.65 255 * = tubing sizes supplied as standard 3.1.6 External Pumps External pumps can be connected to the supply unit. The connections for the external pumps and the signal transmission are located at the sensor field of the supply unit [ Chapter “3.1.3 Control Elements and Connections“]. Maximum Speed of External Pumps The BIOSTAT ® B control unit can control external pumps in a range of 1:100. The controllable rpm range of the external pump is determined by the maximum speed configured for the pump. Example: The maximum speed of the external pump is set to 50 rpm. The BIOSTAT ® B control unit can control the external pump in a range of 0.5-50 rpm. Set the maximum speed of the external pump so that the minimum required speed is within the control range of the control unit. Notes on setting the speed can be found in the operating instructions for the external pump. Operating Manual BIOSTAT ® B 33 Device Overview 3.2 Culture Vessels In the following figures, the functional elements will be displayed using examples of the UniVessel® Glass 1 L and UniVessel® SU 2 L (made of pre-sterilized polycarbonate). More information about the culture vessels (single-walled, double-walled, volumes) can be found in the [UniVessel® Glass, UniVessel® SU culture vessel, BIOSTAT ® RM 20 | 50 and BIOSTAT ® RM 200 Rocker operating manuals]. 3.2.1 UniVessel® Glass Fig. 3-12: Functional elements, UniVessel® Glass 1 L Pos. Description 34 Operating Manual BIOSTAT ® B 1 Exhaust cooler 2 Stirrer driver 3 Cover plate with ports | retainers for sensors, supply media, sampling, aeration 4 Culture vessel support column 5 Glass vessel: −− temperature control by double wall (jacketed) −− heating jacket and cooling finger (single-walled) 6 Supply bottle with holding ring 3.2.2 Device Overview UniVessel® SU 1 2 3 4 Fig. 3-13: Functional elements, UniVessel® SU 2 L with holder Pos. Description 1 Stirrer with connecting pieces for motor adapter of different control units 2 Cover plate with ports/retainers for sensors, supply media, sampling, aeration, exhaust air 3 Plastic vessel (temperature control by heating jacket or heating/cooling jacket) 4 Holder Operating Manual BIOSTAT ® B 35 Device Overview 3.2.3 BIOSTAT ® RM 20 | 50 Rocker 1 2 3 4 5 6 Fig. 3-14: Functional elements RM 20 | 50 Rocker Pos. 1 2 3 4 5 6 Description Hood Culture bags Culture bag holder Supply and control unit (RM 20 | 50 Rocker) Touch panel Connections, left side 3.2.4 BIOSTAT ® RM 200 Rocker 1 2 3 4 4 5 Fig. 3-15: Functional elements RM 200 Rocker Pos. 1 2 3 4 5 36 Operating Manual BIOSTAT ® B Description Hood Culture bags Rocker platform Connections Control cabinet (service purposes only) 3.3 Device Overview Stirrer Driver Fig. 3-16: Stirrer driver Pos. Description 1 Stirrer driver for culture vessel coupling 2 Power supply 3 Sleeve The top drive includes a direct drive for the stirrer shaft and a magnetic coupling. Available drive motors: −− Motor 200 W, rpm range 20 ... 2000 1/min Rpm Ranges The standard stirrer shaft is sealed with a rotating mechanical seal. The optional magnetic coupling is sealed with a rotating mechanical seal, too; the motor coupling on the outer side, however, is sealed in an enclosure and attached to the driver motor by means of a magnetic coupling [ UniVessel® Glass operating manual]. Glass vessels UniVessel® SU 1 L/2 L 5L 10 L 0.53 gal 20 – 2000 1/min 20 – 1500 1/min 20 – 800 1/min 20 - 400 1/min Operating the stirrer at impermissible high speeds can affect the culture vessel’s stability and cause damage to its fittings. Depending on the size and equipment of the culture vessels, the permissible speed may be limited, e.g. to max. 300 min-1 in vessels with aeration rigs for bubble-free aeration. Operating Manual BIOSTAT ® B 37 Software 4. Software 4.1 User Information This operating manual shows the standard functions of the DCU software. DCU systems can be customized according to the customer’s specifications. Therefore, this documentation may describe functions that a delivered configuration does not contain or a system may contain functions that are not described here. Information about the actual scope of functionality can be found in the configuration documents. Additional functions can be described in the technical data sheet in the “Technical Documentation”. Illustrations, parameters and settings in this documentation are only examples. They do not show the configuration or operation of a DCU system in terms of a particular unit, unless they explicitly refer to that unit. Information about the exact settings are provided in the configuration documents or have to be obtained empirically. Usage Instructions, Structure and Functions The DCU system can be integrated into higher-level automation systems. For example, the industrially tested MFCS/Win system can take on host PC functions like process visualization, data storage, process logging etc. This operating manual shows operating values and settings that are default values and examples. Only if explicitly specified as such will they show settings for the operation of a particular bioreactor. The configuration documents contain the specifications on the settings permissible for a bioreactor and the specifications for a customer’s system. Only system administrators or authorized, trained and experienced users may change the system configuration. 4.2 System Start The control unit is turned on together with the whole system using the main switch. After power on and program start (or restoration of voltage after a power outage), the system starts in a defined basic state: −− The system configuration is loaded. −− Any user-defined parameters from a previous process are stored in a memory (CF card) and can be used for the next process: −− Setpoints −− Calibration parameters −− Profiles (given there are any) −− All regulators are turned off and actuators (pumps, valves) are in the rest position. 38 Operating Manual BIOSTAT ® B Software For interruptions in operation, the startup behavior of outputs and system functions that have a direct effect on the associated end unit (regulators, timers, etc.) depend on the type and duration of the interruption. There are several different types of interruptions: −− Turning the unit off | on at the main switch of the control unit. −− Failure of power supplied from the connection in the laboratory (power outage). In the “System Parameters” submenu of the main “Settings” menu, a maximum duration for power interruptions can be set as “Fail Time”: Fig. 4-1: “System Parameters” submenu, [Æ see Chapter “8.12 “Settings” Menu“]. In case of a power outage shorter than “Fail Time”, the system continues as follows: −− A “Power failure” error message shows the outage time. −− Controllers continue to work with the set target value and mode. −− Timer and setpoint profiles continue to be processed. If the power outage lasts longer than the configured “Failtime”, the DCU system acts as though the user had turned the unit off normally, that is, it starts in the defined basic state. After the next restart, an alarm message appears [ alarm messages in Chapter “9. Faults“], specifying the date and time at which the power failure occurred. Operating Manual BIOSTAT ® B 39 Software 4.3 Principles of Operation 4.3.1 Device-specific User Interfaces The user interfaces of the DCU are variable, depending on the device version and culture vessel type. The following versions are possible: −− BIOSTAT ® B-MO Single UniVessel® Glass −− BIOSTAT ® B-MO Twin UniVessel® Glass −− BIOSTAT ® B-CC Single UniVessel® Glass −− BIOSTAT ® B-CC Single UniVessel® SU −− BIOSTAT ® B-CC Single BIOSTAT ® RM 20 | 50 Rocker −− BIOSTAT ® B-CC Single BIOSTAT ® RM 200 Rocker (Single-Bag Configuration) −− BIOSTAT ® B-CC Twin BIOSTAT ® RM 200 Rocker (Twin-Bag Configuration) −− BIOSTAT ® B-CC Twin UniVessel® Glass | UniVessel® SU −− BIOSTAT ® B-CC Twin UniVessel® Glass | BIOSTAT ® RM 20 | 50 Rocker −− BIOSTAT ® B-CC Twin UniVessel® SU | BIOSTAT ® RM 20 | 50 Rocker 4.3.2 User Interface The user interface offers a graphical overview of the controlled unit, with symbols for the reactor, components of the gas supply (e.g. valves, mass flow controllers, probes, pumps, dispensing counters, and if applicable, additional peripherals with their typical locations relative to the reactor. The user interface is divided into 3 sections: −− Header −− Work area −− Footer 4.3.2.1 Header Line Display of the system status, time, date: −− Time in format [hh:mm:ss] −− Date in format [yyyy-mm-dd] Alarm display (area marked in red / bell symbol): −− Time of the triggered alarm. −− Type of malfunction. −− Alarm triggered, information on the triggered alarm in the alarm message [Æ see alarm messages section “Alarm Overview Menu” in Chapter “9. Faults“]. −− All alarm messages are shown in the “Alarm” menu. 40 Operating Manual BIOSTAT ® B 4.3.2.2 Software Work Area Fig. 4-2: Example BIOSTAT ® B-CC Twin: Menu “Main” for the Unit “1” (above illustration) and for Unit “1” and Unit “2” (middle/lower illustrations) Operating Manual BIOSTAT ® B 41 Software The measuring range shows the function elements and submenus of the active main function: −− Preselected process values with current measured or set value −− Pumps or dispenser counters with process values, e.g. flow rates or dispensing volumes for correction materials and gases −− Controllers, e.g. for temperature, speed, mass flow controller (MFC) etc., with current setpoints −− Probes, e.g. for pH, pO2, antifoam etc., with measured values −− Peripheral devices, e.g. weighing system, with measured values or current setpoints *Function elements, tags, parameters and subsystems actually available depend on configuration. 4.3.2.3 Footer Fig. 4-3: Control unit, BIOSTAT ® B Single Fig. 4-4: Control unit, BIOSTAT ® B Twin The footer shows the menu buttons [Æ Chapter “4.3.4 Overview of the Main Function Keys“] for: −− Accessing the main menus of the associated menu buttons: −− “Main” −− “Calibration” −− “Controller” −− “Trend” −− “Settings” −− Toggle between the overviews for the two units (“All”) and for individual units (“1” and “2”) - BIOSTAT ® B Twin only −− Activation of additional functions: −− “Shutdown” −− “Remote” (operation via host interface) −− “Alarm” (with an overview of alarms) Example “Main” and “1”: −− Most important and most frequently used parameters for Unit “1”. −− Display of all parameters for Unit “1”. 42 Operating Manual BIOSTAT ® B Software Display: −− Selected main function: Button light gray, activated −− Function not selected: Button dark gray, deactivated Depending on the configuration, the BIOSTAT ® B can be equipped with one or two culture vessels. Operation is specific for any culture vessel: −− The DCU system is operated directly on the display by selecting a main function and the associated submenus. The function elements in the work area and the menu buttons in the footer contain touch buttons. By pressing them, you can activate the assigned submenus; this is necessary e.g. for inputting data and setpoints or selecting modes. −− Available functions, tag names, parameters and submenus depend on the culture vessel used and the configuration (for example, heating and cooling system, aeration type etc.). Configuration of the process parameters and monitoring of the process values: Process 1 (left culture vessel) 4.3.3 Process 2 (right culture vessel) Process 1 and 2 (left and right culture vessel) Display The display of function elements is shown in the following table: Symbol Display Meaning, use Function element Button with gray underline [PV tag]: Field for short name (“tag”) for the function element, e.g. TEMP, STIRR, pH, pO2, ACID, SUBS, BALANCE MV [Unit]: Field for measured or set value in its physical unit −− Submenu or function can be selected by pressing Function element Button with green underline Measured value collection or output of function element is active, with measured value or set value as shown (The button is not highlighted in green for pure data collection; the button turns green if the controller is active). Function element Button with light green underline Output of function element is active, controller in cascade mode Function element Button with yellow underline Display of function if in “Manual” mode; (switched on or off); automatic control not possible [Tag PV] MV [Unit] No underline No submenu assigned (function cannot be selected) “r”, “s”, “v”, “w” Arrow button Move forward or back in specified menu or function Operating Manual BIOSTAT ® B 43 Software Symbol Display Meaning, use Pump off Auto on Line gray green Direct access to submenu to select the mode Pump off Manual on yellow underline, pump gray green Submenu to select mode [Æ Example in Chapter “8.8 “Main” Menu“] Valve off auto on Line gray green Direct access to submenu to select mode, example for 2/2-way valve Valve off Manual on yellow underline, flow direction green Valve symbol also shows flow direction (possibly changed) Submenu to select the operating mode [Æ Example in Chapter “8.8 “Main” Menu“] Additional Functional Elements for BIOSTAT ® RM Rocker Symbol Display Meaning, Use Gas supply pressure Access to the menu where alarm limits can be configured Shaker drive operation [r/min] Direct access to the submenus to: −− Enter the setpoint for the shaker −− Select operating mode for ROCKS controller −− Switch to ROCKS controller menu Wobble angle in [°] Access to the menu where alarm limits can be configured Additional Functional Elements — BIOSTAT ® RM 200 Rocker with BIOSTAT ® B in Twin-Bag Configuration Only Symbol Display Meaning, Use For control unit BIOSTAT ® B in Twin-Bag configuration only Direct access to the submenu for culture bags: Depending on the number and type of mounted culture bags, the Single or Twin-Bag function must be set. 44 Line green Single-Bag function switched on −− When using a 200 L culture bag. Line gray Twin-Bag function switched on −− When using one or two 100 L culture bags. Operating Manual BIOSTAT ® B Software Examples for function elements, short descriptions, measured values, working values and submenus that can be called by selecting touch buttons [Æ see Chapter “8.8 “Main” Menu“ and Chapter “4.3.6 Direct Function Keys for Selection of Submenus“]. 4.3.4 Overview of the Main Function Keys Button, symbol Meaning, use Main function “Main” Start screen with graphical overview of the culture vessel: −− Display of components of the current configuration −− Overview of measured values and process parameters −− Direct access to important menus for operating input Main function “Calibration” Menus for calibration functions, for example: −− Measurement sensors for pH, pO2 −− Totalizer for all pumps (BASE, etc.) −− Totalizer for aeration rates for valves −− Scales Main function “Controller” Operating and configuration menus for controllers, for example: −− Temperature regulation TEMP −− Speed regulation STIRR −− pH adjustment and DO (pO2) controller −− Control of correction medium pumps (e.g. pH, SUBS) −− Aeration rate regulation (valves or mass flow controllers). Main function “Trend” Display of process sequences, selection of up to 8 parameters from: −− Process values −− Setpoints from control loops −− Outputs of controllers Main function “Settings” Basic system settings, e.g.: −− Measurement ranges of process values −− Manual operation, e.g. for inputs and outputs, controllers, etc. −− External communication (e.g. with printers, external computers) −− Selection, modification of configurations (password-protected, only for authorized service technicians) Main function “1”, “All”, “2” Area selection: −− Area 1 −− Both areas −− Area 2 Main function “Shutdown” Shutdown function: Pressing the “Shutdown” function switches all outputs in the defined safety position. This does not affect any other functional sequences for controllers, timers, profiles, formulations or sterilization cycles. Main function “Remote” Operation with external computer systems (central computer): Pressing the menu button switches to remote operation; instructions on configuration [Æ Chapter ““Settings” Menu“] Main function “Alarm” Overview table of alarms that have occurred: −− If alarms occur, the symbol changes color and an acoustic signal sounds. −− Display red: Table still contains unconfirmed alarms. −− Pressing the menu button opens an overview menu of all alarm messages. Main functions can be selected at any time during a running process. The title of the main function shown in the work area is also displayed in the header. Operating Manual BIOSTAT ® B 45 Software 4.3.5 Button Overview of Selection Keys Meaning, use Cancel Changes will not be saved Confirmation of input Further controller parameters Cancel Changes will not be saved Deleting characters Selection of sign when entering a value Selection list of process values 4.3.6 Direct Function Keys for Selection of Submenus The function elements in the work area of the “Main” main menu can contain function keys that can be used to directly active submenus for important functions: −− for the numerical input of setpoints, conveying and flow rates etc. −− for the configuration of alarm limits −− for the selection of controller modes Which functions can be reached from the main menu depends on the configuration. Press the function keys to view the available functions in the supplied configuration. 46 Operating Manual BIOSTAT ® B Software In this section, an example of screens and submenus accessible via direct function keys is shown. Detailed instructions for the associated functions and possible inputs can be found in the Chapters “8.10 “Calibration” Menu“ and “8.11 “Controller” Menu“. Example: Input of Temperature Setpoint Fig. 4-5: Setpoint input and selection of the “TEMP” controller mode from the “Main” menu ttIn the work area of the “Main” menu, press the TEMP function element or in the work area of the “Controller” menu select the TEMP controller (TEMP function element). yyWhen the “Main” menu is accessed, a submenu opens with a keypad on the left side for data entry and a selection field for possible operating modes. Operating Manual BIOSTAT ® B 47 Software Fig. 4-6: Set value input and selection of the “TEMP” controller mode from the “Controller” menu yyWhen the “Controller” menu is accessed, the “Setpoint” touch key can be used to enter a set value (after pressing the touch key, a screen keyboard also opens). The touch button “off” can be used to select the mode. ttEnter the new setpoint using the screen keyboard (note the permissible value range under the input field). If you want to correct the value entered, press the BS key. If you don’t want to save the new value, leave the submenu by pressing the C key. ttConfirm by pressing the “OK” key. The submenu window closes. The setpoint is active and is displayed. Example: Selecting the Controller Mode “Mode”: ttIn the work area of the main menu, press the “TEMP” function element or select the “Controller” function and there the TEMP controller. ttPress the function key for the desired operating mode “Mode” on the right side. ttConfirm by pressing the “OK” key. The function (of the controller) is activated and displayed. You can access the full operator screen of the controller through . This corresponds to activating the “Controller” menu button and selecting the TEMP controller there in the overview screen [Æ Chapter “8.11 “Controller” Menu“]. 48 Operating Manual BIOSTAT ® B 4.3.7 Software Selection Lists and Tables If submenus contain lists of elements, short names or parameters that cannot be displayed in a window, a scrollbar with a position marker is displayed: Fig. 4-7: Access to values accessible from the submenu after assignment of a channel in the trend display. To page through lists that contain more entries than can be displayed in the window, the following options are available: ttPress the arrow keys “s” (down) or “r” (up). ttPress the position mark (light gray field in the scrollbar) and push it. ttPress directly in the scrollbar at the relative height where the channel tag could be located. Operating Manual BIOSTAT ® B 49 Software 4.4 Password Protection of Individual Functions Only disclose this information to authorized users or service staff. If required, remove this page from the manual and keep it in a special place. Certain system functions and settings that should only be accessible for authorized personnel are protected by a password. These include, in the regulator menu, the settings for the regulator parameters (e.g. PID), in the “Settings” menu”: −− Process value setting “PV” −− At the manual operation level (“Manual Operation”), the interface parameter setting for digital and analog process inputs and outputs or for simulation controllers. The “Service” submenu of the “Settings” menu is only accessible via a special service password. This is only provided to authorized service technicians. When selecting password-protected functions, a key field is displayed automatically with a prompt to enter the password. The following passwords can be defined: −− Standard password (factory-set: 19) −− Customer-specific standard password1 −− Service password 4.5 User Management The user management function regulates the access of users to the DCU system. The function makes it possible to grant or restrict access permissions, for example, to prevent incorrect operation of the DCU system. ttObserve the user management operating instructions for DCU4 systems. 4.6 Bug Handling and Troubleshooting If the DCU system should encounter technical problems, contact Sartorius Service. 4.7 Locking Functions Locking functions are permanently configured; the user cannot change them. In the “Settings” menu [Æ Chapter “8.12 “Settings” Menu“] locked inputs and outputs are highlighted with a colored marking. The extent of the locking functionalities is system-specific and is predefined during configuration. This is documented in the configuration lists enclosed with every system. 1 You will receive this information by post or together with the Technical Documentation 50 Operating Manual BIOSTAT ® B Transport 5. Transport The device will be delivered by Sartorius Service or by a transport company engaged by Sartorius. 5.1 Inspection Upon Acceptance by the Recipient 5.1.1 Report and Document Transport Damage Upon acceptance of the unit by the customer, the unit must be inspected for visible transport damage. ttReport transport damage immediately to the delivering office. 5.1.2 Check Completeness of the Delivery The delivery includes all required valves, connector elements, lines, hoses and cables. ttCheck that the delivery is complete by comparing it against your order. Connection lines to supply facilities are not in the scope of delivery. Components that do not correspond to the specifications of Sartorius Stedim Biotech GmbH may not be used. 5.2 Packaging The packaging used for transport and protection of the device consists primarily of the following materials suitable for recycling: −− Corrugated cardboard −− Styrofoam −− Polyethylene film −− Pressed particle board −− Wood Do not dispose of the packaging in the garbage. Dispose of all packaging material in accordance with local regulations. Operating Manual BIOSTAT ® B 51 Transport 5.3 Instructions for Transport Within the Company When moving the device, it is particularly important to do so in such a way as to prevent damages by force or careless loading and unloading. Danger of severe personal injury and property damage due to improper transport! −− The device may only be moved by technical personnel. −− The load capacity of the lifting system (forklift) must be at least the weight of the unit (you can find weight specifications in the data sheets in the “Technical Documentation” folder). −− Wear protective work clothing, safety boots, safety gloves and a hard hat during these tasks. −− The unit may only be transported with the transport locks in place. To install the transport locks, contact Sartorius Service if necessary. −− Transport locks may only be removed at the place of installation. −− Lift the unit only at suitable points with lifting accessories. −− Always lift the device slowly and carefully, to ensure stability and safety. −− Secure the device from falling during transport. −− During transport of the unit, ensure that no personnel are in the path. −− These devices require two or more people to transport and set up. −− Wear protective work clothing and safety boots during these tasks. −− Lift the device only at suitable points. −− Always lift the device slowly and carefully, to ensure stability and safety. −− Secure the device from falling during transport. Protect the device during transport against −− moisture, −− impact, −− falling, −− damage. Loading | Unloading When loading and unloading, pay attention to the following aspects: −− Do not unload the unit outdoors during rain or snow. −− If necessary, cover the unit with plastic sheeting. −− Do not leave the unit outdoors. −− Use only suitable, clean, undamaged load lifting accessories. 52 Operating Manual BIOSTAT ® B Setup 6. Setup The guide for setup of the device is the setup drawing. The setup of the device is to take place according to contractual conditions, −− by Sartorius Service, −− by Sartorius authorized specialist personnel, −− by the customer’s authorized specialist personnel. Setting up the bioreactor involves the following main steps: −− Ensuring that the ambient conditions have been fulfilled for the installation location [ Chapter “13. Specifications“]. −− Ensuring that there is sufficient and suitable work space [ Chapter “6.3 Work Surfaces and Loads“]. −− Ensuring that the laboratory energy sources correspond to the specifications [ Chapter “6.4 Laboratory Energy Sources“]. −− Setting up the supply unit BIOSTAT ® B, the culture vessels used and other equipment and facilities needed for the process. Danger of severe personal injury or property damage due to improper setup of the unit! The proper setup of the device is essential for the safe operation of the unit. −− Observe the guidelines for building and laboratory equipment. −− Observe the laboratory and process-related safety rules and guidelines on setting up your workplace and securing it from access by non-authorized persons. −− This device requires two or more people to transport and set up or the use of appropriately rated lifting equipment. −− Ensure that only authorized persons have access to the device. −− Follow the instructions in the following sections. 6.1 Acclimatization Condensation from humidity can form on the surfaces of a cold device when it is brought into a substantially warmer area. You should therefore let a device that has been disconnected from its power source acclimatize for approximately 2 hours before reconnecting it to the power. 6.2 Ambient Conditions You will find information on ambient conditions in the [ Chapter “13.1 BIOSTAT ® B“]. Operating Manual BIOSTAT ® B 53 Setup 6.3 Work Surfaces and Loads The device is designed as a bench-top system and should be set up on a stable laboratory table. The work surface must be big enough for the equipment required for the fermentation process. It must be easy to clean and, where relevant, to disinfect. Danger of injury if access to emergency shutdown equipment and shut-off devices is blocked! Equipment for emergency shutdown and shut-off devices, e.g., for the power supply, water or gas feed, as well as the particular equipment connections, must be kept clear and easily accessible. −− Make sure that the device is set up with sufficient distance to the wall to provide enough space for ventilation and easy access to the rear of the device. The recommended wall distance is about 300 mm. Danger of injury due to insufficient stability of the culture vessels! −− Follow any other manufacturer operating manuals for individual system parts and additional components. −− Follow the construction guidelines required to ensure that the device stands stably. −− Ensure that the laboratory table is dimensioned sufficiently for the weight of the device, the culture vessels and the process media in use [ Chapter “13. Specifications“]. −− Ensure that the laboratory table is level. −− Ensure that the workplace is dimensioned in such a way that the device is easily accessible for in-process operation, maintenance and service work. The space requirements also depend on the peripheral devices to be connected. 54 Operating Manual BIOSTAT ® B Setup Example setup for UniVessel® Glass, UniVessel® SU Fig. 6-1: Example setup for BIOSTAT ® B-CC Twin / Single Pos. Description 1 Control unit, BIOSTAT ® B-CC Twin 2 UniVessel® Glass (2 L, jacketed) 3 UniVessel® SU (2 L, Single Use) 4 Control unit, BIOSTAT ® B-CC Single Example Setup, UniVessel® Glass, BIOSTAT ® RM 20 | 50 Rocker Fig. 6-2: Example setup, BIOSTAT ® B-CC Twin with UniVessel® Glass, 2 L and RM 20 | 50 Rocker Pos. Description 1 Control unit, BIOSTAT ® B-CC Twin 2 UniVessel® Glass (2 L, jacketed) 3 RM 20 | 50 Rocker Operating Manual BIOSTAT ® B 55 Setup Setup examples for BIOSTAT ® RM 200 Rocker 1 4 2 3 2 Fig. 6-3: Setup example for BIOSTAT ® B-CC Single with RM 200 Rocker | SingleBag configuration (using one culture bag) No. 1 Fig. 6-4: Setup example for BIOSTAT ® B-CC Twin with RM 200 Rocker | TwinBag configuration (using one or two culture bags) Description Control unit BIOSTAT ® B-CC Single Single-Bag configuration (using one culture bag) 2 Laboratory cart for control unit 3 RM 200 Rocker 4 3 Control unit, BIOSTAT ® B-CC Twin Twin-Bag configuration (using one or two culture bags) Set Up Dimensions Equipment for emergency shutdown and shut-off devices, e.g., for the power supply, water or gas feed, as well as the particular equipment connections, must be kept clear and easily accessible. The required lab table dimensions and distances of the culture vessels to the device are shown in the installation plans [ Chapter “15.3 Setup Drawings“]. The required installation area of the UniVessel® SU 2L holder corresponds to about the installation area of the culture vessel UniVessel® Glass, 10 L DW. The equipment (e.g. stirrer motor) can be stored on the storage dish (1). 1 Fig. 6-5: Storage dish for equipment 56 Operating Manual BIOSTAT ® B 6.4 Setup Laboratory Energy Sources The connections for energy and supply systems must be prepared before installation of the device in the work area. They must be easily accessible, correctly installed, set in accordance with the device’s specifications and ready to operate. Risk of death from energy that is unexpectedly released, such as electric shocks! Energy supply lines may be incorrectly dimensioned and not protected against impermissible fluctuations and faults. Safety equipment must be available and fully functional: −− Ground fault circuit interrupters (residual current protection) for mains connections −− Fittings for shutting off water, compressed air and gas supplies. Observe the energy specifications on the type plates [ “6.4.2 Manufacturer’s ID Label“], [ Chapter “13. Specifications“]. The connections for supply media are located on the rear of the device. The following supply media are connected to the device: −− Power supply, potential equalization The BIOSTAT ® RM Rocker requires a separate power supply. −− Tempering medium −− Gases: −− Compressed air (Air) −− Oxygen (O2) −− Nitrogen (N2) −− Carbon dioxide (CO2) Make sure that the inlets for electricity, water, compressed air and gases are complying with the specifications of the device. Ensure that the intakes are equipped with suitable valves for blocking and emergency shutoff. 6.4.1 Electricity Risk of death from electrical shock! The power supply (mains) in your laboratory must meet the equipment specifications. −− Check whether the devices match your voltage supply rating [ “6.4.2 Manufacturer’s ID Label“]. −− Do not switch on the devices if the laboratory’s mains voltage is incorrect. −− The laboratory’s mains connections must be grounded, free from interference and splash-protected. −− Emergency shut-off equipment (ground fault circuit interrupters, power switch) must always be in perfect working order. −− The laboratory’s power supply [Mains wall outlet] must be equipped with a protective grounding conductor. −− Do not use multiple socket outlets for connecting the bioreactor modules to an AC power outlet (mains outlet). Operating Manual BIOSTAT ® B 57 Setup −− The power cables must have the correct plugs that match your laboratory AC outlet. Do not use any damaged mains cables, e.g. with broken insulation, and in particular if the wires are exposed. −− Do not repair defective mains cables or replace incorrect connectors. Please contact Sartorius Service for this purpose. Risk of damage to the device from voltage dips and spikes! The laboratory supply voltage may not fluctuate by more than 10% from its nominal rating. Keep access to the emergency shut-off devices inside the laboratory and the device’s power connection clear. If you need to switch off the device in an emergency, first press the emergency off switch in the lab then disconnect the power cables from the device. To clean and perform user maintenance, be sure to shut off the power to the device and unplug the power cable from the outlet. 6.4.2 Manufacturer’s ID Label All information concerning the correct power supply can be found on the manufacturer’s ID label. The manufacturer’s ID label can be found at the rear of the device. Type BB-8821051 Type BB-8821050 No./ Year 01000 / 14 No./ Year 01000 / 14 V 230 V 120 A 10 A 12 Hz 50 Hz 60 Fig. 6-6: Manufacturer’s ID labels, version 230 V / 120 V 58 Operating Manual BIOSTAT ® B 6.4.3 Setup Tempering Medium Water is used as tempering medium for the device and applies to the following functions: −− Temperature control of a jacketed culture vessel, −− of a UniVessel® SU with heating | cooling jacket, −− of a culture bag in connection with the optional tempering coil (BIOSTAT ® RM 20 | 50 Rocker) heating | cooling plate (BIOSTAT ® RM 200 Rocker) −− Cooling liquid of the exhaust cooler and the cooling finger (for single-walled glass vessels) Risk of damage to the heat circulation pump, fittings and thermostat system! Inadequate water quality can affect the functioning of the heat circulation pump and fittings in the thermostat system. The following impairments are possible: −− Formation of scale if water is hard −− Corrosion from distilled or demineralized water −− Faults resulting from contaminants or corrosion residues. Malfunctions and damages arising from unsuitable water quality are excluded from the warranty granted by Sartorius Stedim Biotech. Green microbes inside the double wall are a sign of algae formation caused by organic contaminants in the water. Do not use water contaminated in this way. Water supply connection values −− Water pressure: 2 to 8 bar −− Flow volume: min. 10 L/min −− Drain: pressureless ttCheck whether the water is clean before connecting it to the device. ttFlush the laboratory supply pipes. ttIf necessary, fit a suitable prefilter in the laboratory or the supply pipe leading to the device. ttUse tap water with max. 12 dH, no distilled or demineralized water. Using water with a maximum hardness of 12 dH minimizes scale buildup in the thermostat loop and in the double wall of the culture vessels. Operating Manual BIOSTAT ® B 59 Setup 6.4.4 Gas Supply Gas supply comprises the following gases (depending on the integrated aeration module): Aeration Modules BIOSTAT ® B-MO “Additive Flow 2-Gas” BIOSTAT ® B-CC / BIOSTAT ® RM Rocker “Additive Flow 5-Gas” “Additive Flow 4-Gas” AIR (air) AIR (air) Oxygen (O2) Oxygen (O2) Nitrogen (N2) Carbon dioxide (CO2) Danger of explosions and fire due to escaping oxygen! There is danger of explosions and fire when high amounts of oxygen are released in an uncontrolled manner. Pure oxygen can give rise to chemical reactions that could cause substances to self-combust. Escaping gases that contain C can cause chemical reactions and cause a fire. −− Keep pure oxygen away from flammable materials. −− Avoid sparks in the vicinity of pure oxygen. −− Keep pure oxygen away from ignition sources. −− Keep the whole aeration segment free from oil and grease. −− Check the tightness of the connections. Danger of suffocation due to escaping gases! There is danger of suffocation due to CO2. −− Ensure good ventilation of the installation site. −− Keep a breathing device independent of ambient air ready for emergencies. −− If personnel appear affected by symptoms of suffocation, immediately provide them with a breathing device independent of ambient air, bring them into fresh air, make them comfortable and keep them warm. Call a doctor. −− If a person stops breathing, initiate first aid measures with artificial respiration. −− Do not eat, drink or smoke during work. −− Monitor the limit values at the system and in the building (sensors recommended). −− Check process gas lines and filters regularly. −− Check the tightness of the connections. 60 Operating Manual BIOSTAT ® B Setup Risk of malfunctions and damages to gas-carrying components! Soilings like oil and dust can impair the function of gas-carrying components and lines. −− When corrosion-causing gases needed for certain processes are used in the gas supply, −− The gas-carrying components have to be corrosion-resistant (e.g. ammonia can cause corrosion to gas-carrying components made of brass). −− Make sure that the supply gases are dry and free of dirt, oil and ammonia. −− Install suitable filters, if necessary. −− Malfunctions and damages arising from contaminated gas media are excluded from the warranty granted by Sartorius Stedim Biotech. Operating Manual BIOSTAT ® B 61 Startup 7. Startup Starting up the bioreactor involves the following main steps: −− Connecting the device to the power supply [ “7.2 Connecting the Device to the Power Supply“] −− Connecting the laboratory’s water supply [ “7.3 Connecting Laboratory Water Supply to the Device“] −− Connecting the laboratory’s gas supply [ “7.4 Connecting Laboratory Gas Supply to the Device“] −− Connecting the stirrer motor to the supply unit [ “7.1 Installation Material“] −− Connecting the holder (UniVessel® SU) [ “7.6 Connecting the UniVessel® SU Holder“] −− Connecting the barcode scanner [ “7.7 Connecting the Barcode Scanner“] −− Connecting the BIOSTAT ® RM Rocker with culture bags [ operating manual BIOSTAT ® RM 20 | 50, BIOSTAT ® RM 200 Rocker] −− Connecting the sensor cable [”7.9 Connecting the Sensor Cable“] −− Connecting the tubes for aeration [”7.10 Connecting the Tubes for Aeration“] −− Connecting the temperature control system [ “7.11 Connecting the Temperature Control“] −− Connecting the exhaust cooling [ “7.12 Connecting the Exhaust Cooling Hoses (only UniVessel® Glass)“] −− Connecting external pumps ( “7.13 Connecting external pumps“] −− Switching on the device [ “7.14 Turning the Device On and Off“] 7.1 Installation Material The equipment supplied with the bioreactor includes a connection kit. −− Only use lines and fittings approved for use with the bioreactor or whose suitability has been confirmed in writing by Sartorius Stedim Biotech. −− Only replace damaged components and wear parts with Sartorius Stedim Biotech approved parts. Danger of operating faults and breakdowns! Sartorius Stedim Biotech does not accept any liability for operating faults and breakdowns related to the use of equipment that has not been approved for use with the bioreactor, as well as any secondary damage arising from this. 62 Operating Manual BIOSTAT ® B 7.2 Startup Connecting the Device to the Power Supply The connections for the power supply (2) and the potential equalization (1) can be found on the rear of the device. Mains Connection −− The device can be supplied in the following voltage versions: −− 230 V (± 10%), 50 Hz at a power consumption of 10 A or −− 120 V (± 10%), 60 Hz at a power consumption of 12 A ttUse the power cable supplied with the device in accordance with the power specifications of your country. ttConnect the reserved mains connection cable to the device and connect the device to the laboratory power supply. Fig. 7-1: Mains and potential equalization connection ttConnect the reserved potential equalization to the device and connect the device to the laboratory potential equalization if available. Power Supply Malfunctions ttCheck the position of the main switch. Contact Sartorius Service if the power supply continues to malfunction. Fig. 7-2: Main Switch Information on operating the DCU system can be found starting in [ Chapter “4.3 Principles of Operation“]. Operating Manual BIOSTAT ® B 63 Startup 7.3 Connecting Laboratory Water Supply to the Device Risk of injury due to bursting culture vessel! If the pressure in the temperature control circuit is too high, there is a risk that the double-walled versions of the culture vessels will burst. Therefore: −− Make sure that the cooling water supply and cooling water return line (“Cooling Water” connecting area) have been properly connected. −− Avoid creating kinks in the lines. The water must flow freely into the outlet. −− When connection to a closed (laboratory) cooling circuit system, the water must not jam back and build up pressure on the outlet connection. The water input pressure is limited by a pressure reducer. A flap valve does not allow water to access the system if the water supply was inadvertently connected to the water outlet. The connections for tempering media are located on the rear of the supply unit [ Fig. 7-3]. −− For connecting the water supply, use the hose clips and hoses supplied (or components with equivalent specifications). −− Carefully fasten the connections and protect them from accidentally coming loose. −− Make sure that the preliminary pressure of the laboratory is adjusted correctly prior to opening the inlets to the supply unit. −− Make sure that there are no kinks in the hose and lay it in such a way that there is no risk that water pockets form. Regularly check that any excess water can freely drain off. Water outlet Water intake Fig. 7-3: Water supply on the supply unit’s connection panel Connecting External Cooling Equipment You can connect a laboratory cooling water circuit or cooling device to the “Cooling water” inlet and outlet. For the external cooling devices, following specifications apply: −− Water pressure: 2 to 8 bar −− Flow volume: min. 10 L/min −− Drain: pressureless −− Temperature min. = 4 °C −− Connection: Nozzle | outer diameter = 10 mm Make sure that the inlet and outlet are connected in the right order: −− Connect the outlet of the external loop or the cooling device to the inlet of the supply unit. −− Connect it from the outlet of the supply unit to the laboratory return pipe or the inlet of the cooling device. Operate the cooling device or the external cooling loop at ambient pressure. Prevent the cooling medium from flowing back into the device’s outlet. 64 Operating Manual BIOSTAT ® B 7.4 Startup Connecting Laboratory Gas Supply to the Device The gases | compressed air supplied inside the laboratory must correspond to the supply unit’s specifications. The connections for the gas supply are located at the back of the supply media unit [ Fig. 7-4]. Please observe the following information and the P&I diagram: Specifications of the Supply Units −− Compressed air, preset to 1.5 positive pressure (29 psig) −− Gas flow rate 0.02 – 2 vvm (depending on the size of the culture vessel) −− Gas required, e.g. O2, N2 or CO2, preset to 1.5 barg (21.76 psig) −− Gases must be dry and clean, i.e., free of condensate and contaminants from the tubing | piping. −− The unused inlets of the aeration modules “O2 Enrichment” and “Additive Flow 2-Gas” are sealed with dummy plugs. Dimensions of Variable Area Flow Meters The variable area flow meters must be designed for the planned gases. Their measuring jets have been calibrated for standard conditions. You will find all specifications labeled on the glass tube or holder, for example: −− Gas type: Air −− Temperature: 20° C = 293 K −− Pressure: max. 1.21 barg (21.76 psig) If used with different gases with a different pressure or temperature, the variable area flow meters’ flow rates may be too high or low. To determine the actual gas flow volumes, you will need to convert the flow rates measured. ttVariable area flow meter manufacturers provide tables and monograms for calculating the correction factors for flow rates under defined operating conditions [ Manufacturer’s Information], e.g., the “Technical Documentation” folder. ttPrepare the laboratory supply points, where necessary, by fitting suitable filters to ensure that the supplied gas is free from oil and grease. ttConnect the laboratory supply points to the device using the appropriate adapters [ Fig. 7-4]. 1 2 3 4 1 Air 2 Oxygen (O2) 3 Nitrogen (N2) 4 Carbon dioxide (CO2) Fig. 7-4: Gas inlets on the rear panel of the device Operating Manual BIOSTAT ® B 65 Startup 7.5 Connecting the Stirrer Driver (only UniVessel® Glass / UniVessel® SU) Risk of injury when motor is running! The motor can be started up for testing purposes before being fitted by switching on the DCU system. Reaching into the running drive can cause injuries to the fingers. −− Do not reach into the protective sleeve with your fingers. −− Leave the motor controller switched off (except if the power is switched off and you are connecting the drive to perform a function test) until you have fastened the motor to the stirrer shaft. −− Make sure that the motor control is turned off when you plug the motor plug into the connection coupling on the motor. Danger of damage to the stirrer drive! The main switch must be turned off prior to fitting or detaching the motor cable; otherwise there is risk of short circuits and the motor can be damaged. Make sure that the motor is not yet fitted to the stirrer shaft. The following figures show connecting cable plugs and their corresponding female connectors on the stirrer motor. The following illustrations show possible sleeve and stirrer shaft coupling models. The actual model supplied may differ from the illustration. 1 1 2 2 Fig. 7-5: Stirrer driver power supply 66 ttConnect the motor plug to the motor couplings as displayed in item (1) und firmly tighten the connections (2). Operating Manual BIOSTAT ® B 7.6 Startup Connecting the UniVessel® SU Holder The UniVessel® SU holder is for mounting the UniVessel® SU culture vessel and | or to compile and analyze measurement signals from the optical pH and DO sensors in the UniVessel® SU culture vessel. The holder transmits the measurement signals via the digital interface. The holder and adapter ring ensure the stability and proper operation of the culture vessel. Furthermore, the process data are exchanged with the device via the interface. Sensor plate Serial-C Conn.Cable UniVessel® SU RS485, M12-8 | M12-8, 2m UniVessel® SU Holder Side View Front Serial Fig. 7-6: Connecting the UniVessel® SU Holder to the device ttConnect the data cable to the holder at the connection “Serial” and on the side of the device at the connection “Serial-C” [ Chapter “3.1.3 Control Elements and Connections“]. 7.7 Connecting the Barcode Scanner The calibration data is determined during production and delivered with the culture vessel for optical single-use pH and DO sensors. The pH and DO calibration data are located on the calibration sticker on the UniVessel® SU cardboard box. Fig. 7-7: Barcode scanner (optional equipment) The calibration data can either be entered manually or with the barcode scanner and then transmitted to the device via the USB port. ttConnect the barcode scanner to the USB port on the front side of the BIOSTAT ® B. ttBefore the process, carry out the pH and DO calibration [ÆChapter “8.10 “Calibration” Menu“]. Operating Manual BIOSTAT ® B 67 Startup Further information on the UniVessel® SU, UniVessel® SU Holder, adapter ring and barcode scanner can be found in the following operating instructions: −− “UniVessel® SU culture vessel installation instructions” −− “UniVessel® SU Holder operating instructions” −− “Adapter ring installation instructions” 7.8 Connect BIOSTAT ® RM Rocker ttConnect the “D-LINK 2” connector of the BIOSTAT ® RM Rocker with the ’Serial-C’ connector of the BIOSTAT ® B. −− See operating instructions for RM Rocker (“BIOSTAT ® RM 20 | 50” or “BIOSTAT ® RM 200 Rocker”). 7.9 Connecting the Sensor Cable ttConnect the sensor cable to the side of the device [ Chapter “3.1 Control/Supply Units“]. The temperature sensor Pt-100 is firmly attached with the connection cable. 7.10 Connecting the Tubes for Aeration Risk to health from gases! The gases used or formed during the fermentation process can be hazardous to health. −− Make sure that there is adequate ventilation at the work site. −− If you are using large amounts of CO2, e.g. for regulating pH, or if CO2 is created as a result of the cell’s metabolism, connect the exhaust air connection of the culture vessel to a laboratory exhaust air treatment system. −− Establish the potential amounts of hazardous gases that might occur and could escape. −− If necessary, fit suitable equipment for monitoring the air in the room. The device is fitted with aeration systems that have independently controllable spargers, depending on the specifications: −− The models “Additive Flow 2-Gas” only have one adjustable “Sparger” outlet for medium aeration. −− The versions “Additive Flow 4-Gas” have an adjustable “Sparger” outlet for medium aeration and an “Overlay” outlet for headspace aeration. ttConnect the hoses for aeration to the front of the device. 68 Operating Manual BIOSTAT ® B Startup 7.11 Connecting the Temperature Control 7.11.1 Jacketed Culture Vessels / Single-walled Culture Vessels with Heating | Cooling Jacket (only UniVessel® Glass / UniVessel® SU) Danger of injury from shattered glass! Excess pressure can cause the glass culture vessels to break. Bursting glass culture vessels can cause cuts and damages to the eyes. −− Make sure that the hose on the return connection leading to the supply unit is not kinked or disconnected. Running the system on dry can damage the circulation pump in the temperature control system! Always fill the temperature control system before activating the temperature controller. To ensure optimum heat transfer, the double wall must be completely filled. Check the filling level every time before the equipment is sterilized and before starting a process. Hose Kits Hose kits to connect jacketed culture vessels or the UniVessel® SU heating | cooling jacket are included with the equipment supplied. The exhaust coolers are supplied with the hose kits needed for connecting to the associated outlet of the device. Operating Manual BIOSTAT ® B 69 Startup 1a 2 9 3 4 5 8 7 6 Fig. 7-8: Hose kit | temperature control for jacketed culture vessels 1b 2 3 4 5 9 8 7 6 Fig. 7-9: Hose kit | temperature control for single-walled culture vessels with heating | cooling jacket Pos. 1a 1b 2 3 4 5 6 7 8 9 70 Operating Manual BIOSTAT ® B Description Culture vessel, jacketed Heating | cooling jacket Hose with plug-in sleeve Sealing coupling Hose with sealing coupling for return (length 600 mm) Supply unit connection (return) Supply unit connection (inlet) Hose with sealing clip for inlet (length 600 mm) Plug-in sleeve Hose with sealing coupling Startup Fill Tempering Medium The device is switched on [ Chapter “7.14 Turning the Device On and Off“]. ttConnect the inlet hose (7) to supply unit connection (6) and then to connection (9). ttConnect the return hose to supply unit connection (5) and then to connection (2). ttSwitch on the device. ttActivate the temperature control function via touchscreen of the controller. 5 ttThe filling process can be stopped as soon as the water is exiting from the laboratory’s outlet. 6 Fig. 7-10: Temperature control connections During the fermentation process, the cooling water is only fed into the temperature control circuit if the vessel needs to be cooled. The cooling water supply to the exhaust cooler is configured in such a way that, once the laboratory supply point has been opened, there will be a constant flow of fresh water. External Cooling Equipment The minimum culture vessel temperature is around 8 °C above the ambient temperature. To operate the bioreactor at lower temperatures, it needs to be connected to an external cooling system. If you are connecting the equipment to external cooling circuit in the laboratory or a cooling thermostat, the temperature control circuit must be operated at zero pressure (at ambient pressure). 7.11.2 Tempering the culture bag BIOSTAT ® RM 20 | 50 Rocker Tempering of the culture medium can be carried out using electrical heating mats or tempering coils flushed with water. −− The heating mats are part of the basic equipment included and are connected directly to the RM Rocker 20 | 50 at startup of the RM Rocker 20 | 50 [ “BIOSTAT ® RM 20 | 50 operating manual”]. −− The tempering coils are optionally available and are connected when the RM Rocker 20 | 50 is installed on the BIOSTAT ® B [ “BIOSTAT ® RM 20 | 50 operating manual”]. BIOSTAT ® RM 200 Rocker Depending on the version of the device, tempering of the culture medium can be carried out with electrical heating plates or heating | cooling plates flushed with water. The heating plates are connected directly to the RM 200 Rocker at startup, the heating | cooling plates are connected to the control unit BIOSTAT ® B [ “operating manual BIOSTAT ® RM 200 Rocker”]. Operating Manual BIOSTAT ® B 71 Startup 7.11.3 Heating Jacket (only UniVessel® Glass / UniVessel® SU) The heating jackets are designed for heating single-walled culture vessels. Danger to life caused by electric shock if heating blanket is defective! The heating blankets have to be in perfect condition. −− Observe the accompanying safety instructions. The power consumption of the heating blanket used may not exceed 780 watts. −− Only use the parts specified by Sartorius Stedim Biotech. Special versions and especially models from other suppliers require the prior written agreement of Sartorius Stedim Biotech. Supplying the heating blanket with the wrong kind of voltage will damage the heating blanket. The heating jackets must only ever be connected to the female connector on the supply unit – never to a power supply in the laboratory. The only connection that provides the correct voltage is the “Heating blanket” connection, which is controlled by the supply unit’s temperature controller. Construction of Heating Jacket 3 1a 5 4 1 1b 2 3 5 1 4 1b 1a Fig. 7-11: Heating jacket Pos. Description 1 Power cable 1a Cable connection with overheating protection 1 1b 2 6-pin Amphenol power cord Protective film of heating coil Pos. Description 2 Protective film of heating coil (vessel side) 3 4 5 Heating coil Silicone foam sleeve Velcro fastener Fitting the Heating Jacket to the Device ttMake sure that the supply unit has been switched off at the main switch [Æ Chapter “7.14 Turning the Device On and Off“]. ttInsert the plug of the heating jacket connecting cable into the connector “Heating Blanket” (1) on the device. Fig. 7-12: “Heating Blanket” connection 72 Operating Manual BIOSTAT ® B Startup 7.12 Connecting the Exhaust Cooling Hoses (only UniVessel® Glass) 1 2 3 5 4 Fig. 7-13: Hose kit, exhaust cooling for culture vessels Pos. 1 2 3 4 5 Description Exhaust cooler Hose with sealing coupling for return Supply unit connection (return) Supply unit connection (supply line) Hose with sealing clip for inlet Connecting Hoses to the Supply Unit ttConnect the supply hose (5) to the supply unit connection (4). ttConnect the return hose (2) to supply unit connection (3). 3 4 Fig. 7-14: Exhaust cooling connections During the fermentation process, the cooling water is only fed into the temperature control circuit if the vessel needs to be cooled. The cooling water supply to the exhaust cooler is configured in such a way that, once the laboratory supply point has been opened, there will be a constant flow of fresh water. Operating Manual BIOSTAT ® B 73 Startup 7.13 Connecting external pumps ttConnect the external pumps to the supply unit. −− The connections for the external pumps and the signal transmission are located at the sensor field of the supply unit [ Chapter “3.1.3 Control Elements and Connections“]. ttSet the maximum speed of the external pump so that the minimum required speed is within the control range of the control unit. −− Observe the notes in Chapter [ “3.1.6 External Pumps“ on page 33]. −− To configure the speed, observe the operating instructions of the external pump. 7.14 Turning the Device On and Off Prerequisites The system must be properly installed and connected in accordance with the specifications. You must also have gained familiarity with the safety instructions [Æ Chapter “2. Safety Instructions“]. ttEnsure that all required supply energies are connected. Switching On You can carry out two independent processes on the BIOSTAT ® B-MO and BIOSTAT ® B-CC twin variant. ttTurn the device on at the main switch [Æ Chapter”3. Device Overview“]. ttChoose the culture vessel you wish to use for the process on the DCU operator display [Æ starting in Chapter “8.8 “Main” Menu“]. If a culture bag is used as a culture vessel, the BIOSTAT ® RM Rocker must also be turned on [ operating manual “BIOSTAT ® RM 20 | 50”, “BIOSTAT ® RM 200 Rocker”]. Fig. 7-15: Main Switch Switching Off the System −− If no other process is running (twin variant), turn the device(s) off at the main switch upon completion of the process. 74 Operating Manual BIOSTAT ® B Preparing and Running the Process 8. Preparing and Running the Process Read the operating manual carefully before carrying out processes on the unit. This is especially important for the safety instructions [ Chapter “2. Safety Instructions“]. 8.1 Overview Process preparation of the bioreactor during the relevant process involves the following main steps, depending on the culture vessels used: −− Equipping and changing the equipment of the culture vessels [ UniVessel® Glass Operating Manual]. −− Connecting and installing the UniVessel® SU components [ UniVessel® SU Holder Operating Manual and installation instructions for the delivered UniVessel® SU components]. −− Connecting BIOSTAT ® RM Rocker with culture bags [ BIOSTAT ® RM 20 | 50, BIOSTAT ® RM 200 Rocker operating manual]. −− Connecting the culture vessels and set up the bioreactor at the site designated for the fermentation process. −− Autoclaving the UniVessel® Glass culture vessels and the accessories that are to be connected aseptically [ UniVessel® Glass Operating Manual]. −− Performing a process. 8.2 Preparing the Glass Culture Vessels Risk of injury when handling heavy culture vessels! Fully equipped and filled culture vessels are heavy, e.g. an UniVessel® Glass with an operating volume of 5 liters weighs > 18 kg. −− The culture vessels must be handled with care. −− Use suitable transport equipment and lifting equipment. −− Only lift the culture vessels using the handles provided for this purpose. Equip the culture vessels only with those components that are needed for the process [ UniVessel® Glass operating manual]. General Measures Make sure that the vessel equipment is in perfect condition and clean before installing it in the culture vessel. −− Remove all residues, contaminations or microbes from the previous fermentation process from the culture vessel and its fittings. −− Carefully check all equipment, and glass culture vessels, seals and silicone hoses in particular, for damage. Replace all damaged and worn out parts. Operating Manual BIOSTAT ® B 75 Preparing and Running the Process Measures Required Before Installing and Connecting Certain Parts −− pH sensor (see operating instructions of the manufacturer): −− Calibrate the pH sensor before autoclaving the culture vessel. −− Calibrate the zero point and slope of the sensors using the buffers in accordance with the scheduled measuring range. −− pO2 sensor (see operating instructions of the manufacturer): −− Test the sensor as recommended by the manufacturer and service it if required. For example, replace the membrane and the electrolyte for measurement. −− The DO sensor (pO2) must be calibrated after the culture vessels have been sterilized in readiness for the fermentation process. −− Redox sensor (optional, where included): −− Test the sensor as recommended by the manufacturer using reference buffers. 8.3 Connecting Transfer Lines The transfer lines are connected between the correction medium bottle(s) and the culture vessel. Transfer line for UniVessel® Glass , UniVessel® SU Standard designation Material Inner diameter of the hose Wall thickness 0.8 x 1.6; VMQ 7621; 55° Shore Silicone hose transparent 0.8 mm 1.6 mm 1.6 x 1.6; VMQ 7621; 55° Shore Silicone hose transparent 1.6 mm 1.6 mm 3.2 x 1.6; VMQ 7621; 55° Shore Silicone hose transparent 3.2 mm 1.6 mm Transfer Line for BIOSTAT ® RM Rocker Standard designation Material Inner diameter of the hose Wall thickness 0.8 x 1.6; VMQ 7621; 55° Shore Silicone hose transparent 0.8 mm 1.6 mm 1.6 x 1.6; VMQ 7621; 55° Shore Silicone hose transparent 1.6 mm 1.6 mm Correction Media Bottles ttPrepare the bottles for acid, base, antifoaming agent or nutrient solutions and connect the transfer lines. Information about setup, equipment and installation of the corrective solution bottles can be found in the [ UniVessel® Glass operating manual]. 76 Operating Manual BIOSTAT ® B Preparing and Running the Process Fitting the Transfer Lines Risk of chemical burns from acids and bases! If the hoses are not securely fastened, there is risk that they will slide off, in which case the correction media will leak out. −− Wear personal protective equipment: −− Protective clothing, safety gloves, protective goggles −− Use the hoses included with the supplied equipment. −− Make sure that the hoses are securely fastened. In the BIOSTAT ® B, each of the culture vessels can be supplied with correction media from up to 3 different bottles. ttFit a piece of silicone hose onto the tube connection on the correction media bottle to which the riser pipe is fitted. ttConnect the free end of the hose to the inlet on the culture vessel. The hoses must be long enough to ensure that they can be easily fitted into the associated hose pumps after having been set up at the supply unit. ttSecure all of the hose connections with hose clamps. ttBefore autoclaving, clamp off the tubing connected to the immersion pipes using tube clamps. The reason is that when positive pressure builds up in the bottles, media can be forced out of the bottles, and this must be prevented. ttPut the corrective solution bottles in the bottle holder. If the bottles need to be connected to the culture vessels at a later time, they can be autoclaved separately. Fit the transfer lines with STT quick release couplings to create a sterile connection to the culture vessel: −− Fit the STT quick release coupling plugs to the transfer line. −− Fit the coupling to the supply line of the culture vessel. For detailed information on connecting the STT quick release couplings, please refer to the [ UniVessel® Glass Operating Manual]. Operating Manual BIOSTAT ® B 77 Preparing and Running the Process 8.4 Filling the Culture Vessel with Culture Medium 8.4.1 UniVessel® Glass / UniVessel® SU Heat-Resistant Culture Medium ttBefore autoclaving, fill the culture medium into the culture vessel via the lid port. Non-Heat-Resistant Culture Medium ttFill the culture vessel with a small amount of water (approx. 200 - 300 ml) and autoclave the culture vessel. ttFill the culture vessel with the culture medium after autoclaving. ttBe sure that the culture medium is sterile when fed in. UniVessel® SU: The UniVessel SU is delivered sterile. The culture vessel must not be autoclaved. The culture vessel must be filled with sterile medium. 8.4.2 Culture bags To fill the culture bag with culture medium, follow the instructions in the [ BIOSTAT ® RM 20 | 50, BIOSTAT ® RM 200 operating instructions]. 8.5 Sterilizing Glass Culture Vessels Risk of breaking the culture vessels! Excessive pressure during heating and any vacuum created during cooling can destroy the glass vessels. The sterile filter of the exhaust air segment makes sure that the pressure between the inside of the vessel and the surrounding atmosphere is equalized in a sterile manner. −− These exhaust air paths may not be kinked. In double wall vessels, pressure is equalized through the outlet (connection piece at the top, piece of hose with the male end of the coupling). −− This piece of hose must not be kinked, disconnected or sealed. Do not use vacuum autoclaves. At the end of the sterilization process, vacuum can cause heavy foaming in the medium. Foam penetrating into the inlet or exhaust filters can cause them to block and render them inoperable. The double wall of the culture vessel must be filled for optimal heat transfer in the autoclave and during the process. ttAt the culture vessel, pinch off transfer lines connected to dip tubes and the hose between air inlet filter and sparger pipe with hose clamps. ttAutoclave the culture vessels at 121°C. The autoclave dwell time required to ensure sterilization must be empirically determined [ Documentation on the autoclave]. 78 Operating Manual BIOSTAT ® B Preparing and Running the Process To ensure reliable sterilization (e.g., to kill off thermophilic spores), the temperature in the culture vessels must be maintained at sterilization temperature for at least 30 minutes. 8.6 Preparing the Cultivation Process Danger of burns due to hot surfaces! The premature removal of culture vessels from the autoclave can cause burns. −− Leave the culture vessels to cool down inside the autoclave. −− Wear protective gloves to transfer them afterwards. Risk of injury when handling heavy culture vessels! Fully equipped and filled culture vessels are heavy, e.g. an UniVessel® Glass with an operating volume of 5 liters weighs > 18 kg. −− Use suitable transport equipment and lifting equipment. −− Only lift the culture vessels using the handles provided for this purpose. ttCarefully transport the culture vessels to the work site and place the culture vessels in front of their supply unit in such a way that all of the lines and peripheral devices can be easily connected. ttFit the stirrer driver to the stirrer shaft [ Chapter “8.6.1 Mounting the Agitator Drive“] Temperature control system – UniVessel® Glass jacketed: ttConnect the supply and drain tubing of the temperature control system with the ports on the culture vessel. Temperature control system – UniVessel® Glass single-walled / UniVessel® SU (Single Use): ttConnect the supply and drain lines of the temperature control system with the heating | cooling jacket and fit to the culture vessel. [ Chapter “8.6.2 Heating | Cooling Jacket Installation“] or (depending on the equipment) ttFit the heating jacket to the culture vessel. [ Chapter “8.6.3 Installing the Heating Blanket“] Operating Manual BIOSTAT ® B 79 Preparing and Running the Process Exhaust Cooling ttConnect the supply and return tubing of the exhaust cooling to the ports of the exhaust cooler at the culture vessel. Exhaust Heater – UniVessel® SU (Single Use): ttFit the exhaust filter heater to one of the exhaust filters and connect the plug to the mains supply [ installation instructions “Heater for Exhaust Filter”]. Sensors ttConnect the sensors to the associated cables. Aeration Modules ttConnect the aeration to the culture vessel. [ Chapter “8.6.4 Connecting the Aeration Modules“] Corrective Solution Supply ttPlace the transfer hoses in the peristaltic pumps on the device. [ Chapter “8.6.5 Preparing the Corrective Solution Supply“]. Exhaust Filter Pressure Regulation ttStick a silicone hose on the exhaust filter and connect it with the connection “Press in” on the front of the supply unit. 8.6.1 Mounting the Agitator Drive Risk of injury when motor is running! The motor can be started up for testing purposes before being fitted by switching on the DCU system. Reaching into the running drive can cause injuries to the fingers. −− Do not reach into the protective sleeve with your fingers. −− Leave the motor controller switched off (except if the power is switched off and you are connecting the drive to perform a function test) until you have fastened the motor to the stirrer shaft. The illustrations show possible sleeve and stirrer shaft coupling models. The actual model supplied may differ from the illustration. 1 2 The coupling (1) of the motor is equipped with a rubber compensation element (2). The compensation element establishes a positive connection to the coupling of the stirrer shaft, ensuring silent force transmission of the drive. The stirrer drive motor can be fitted to the following stirrer shafts: −− UniVessel® Glass (single-walled / jacketed) −− UniVessel® SU (with corresponding adapter) Fig. 8-1: Motor Coupling 80 Operating Manual BIOSTAT ® B Preparing and Running the Process Assembly of UniVessel® Glass Culture Vessels 1 2 3 4 Fig. 8-2: UniVessel® Glass stirrer coupling ttPrior to placing the device, take the motor (1) and connect the coupling with the sleeve (2) to the stirrer shaft. ttGently twist the motor housing to the left or right until the motor’s coupling and the coupling (3) on the stirrer shaft engage. ttIn order to fasten the motor securely to the stirrer shaft, tightly screw the fastening screw (4) of the sleeve. Assembly of UniVessel® SU Culture Vessels It’s not possible to fit the motor for the stirrer shaft directly to the coupling when using UniVessel® SU culture vessels. An adapter is required in order to fit the motor. The adapter is not part of the device’s standard equipment. You can order the adapter with enclosed installation instructions from Sartorius Stedim Biotech. 3 1 4 5 2 Fig. 8-3: UniVessel® SU stirrer coupling Operating Manual BIOSTAT ® B 81 Preparing and Running the Process ttFit the adapter (1) to the coupling of the stirrer shaft (2) [ “Motor adapter installation instructions”]. ttPrior to placing the device, take the motor (3) and connect the coupling with the sleeve (4) to the adapter. ttGently twist the motor housing to the left or right until the motor’s coupling and the coupling on the adapter engage. ttIn order to fasten the motor securely to the stirrer shaft, tightly screw the fastening screw (5) of the sleeve. 8.6.2 Heating | Cooling Jacket Installation Danger of burns upon contact with hot surfaces! The heating | cooling jacket can heat up to 55° Celsius. −− Avoid contact with hot surfaces. −− Wear safety gloves when working with heaters and hot culture media. 1 2 3 5 4 Fig. 8-4: Heating | cooling jacket on the UniVessel® SU Pos. Description 82 Operating Manual BIOSTAT ® B 1 UniVessel® SU culture vessel or single-walled culture vessel 2 Heating | cooling jacket 3 Connection for return temperature control (Rectus quick connect coupling half) 4 Connection for supply temperature control (Rectus quick connect coupling half) 5 Velcro fastener of the jacket Preparing and Running the Process Fitting the Heating | Cooling Jacket to the Culture Vessel The heating | cooling jacket is filled with tempering medium and connected to the temperature control system hoses [ Chapter “7.11.1 Jacketed Culture Vessels / Single-walled Culture Vessels with Heating | Cooling Jacket (only UniVessel® Glass / UniVessel® SU)“]. ttPlace the jacket (2) around the culture vessel (1). ttFasten the Velcro fasteners (5) such that the jacket lies closely around the culture vessel. Observe further notes in the installation instructions “Heating | cooling jacket for culture vessel”. 8.6.3 Installing the Heating Blanket Objects with points or sharp edges may damage the heating coils! Sharp-edged or heavy objects can damage the heating coil and cause a short circuit. Never put any objects on the heating jacket. ttCarefully lift and hold the jacket at the edge opposite to the cable connection. Tensile stress can damage the cable attachment! −− Do not lift the heating and cooling jacket using the power cord. This can damage the cable attachment. −− Do not roll the blanket together more tightly than the shape of the culture vessel allows. −− Do not bend or fold up the heating jacket. −− When using the heating jacket on the UniVessel® SU, the heating jacket should be installed in the lower part of the culture vessel for optimal heat transfer. Mount the heating jacket such that the power cord can be fed upwards. This prevents the connection cable from kinking. ttWrap the heating blanket around the culture vessel with foil-shielded side touching the vessel. yyThe side insulated with silicone foam should face outwards. The isolated layer prevents burns when touching the device. ttSecure the Velcro fasteners so that the blanket is lying flat on the culture vessel without folds, warping or dents. Fig. 8-5: Heating jacket at culture vessel Operating Manual BIOSTAT ® B 83 Preparing and Running the Process Operating the Heating Jacket Danger of burns at the heating jacket! Depending on the operating temperature intended for the culture vessel, the heating jacket can heat up to approx. 80°C. −− When in operation (above 40 °C), never touch the heating jacket with your bare hands. −− When you have to handle the culture vessel, always use safety gloves. ttSwitch on the device. ttConfigure the temperature measurement on the user interface and activate it if needed for the process. Whenever the culture vessel needs to be heated up or cooled down, the measurement and control system activates either the power supply of the heating jacket or the cooling water supply for the cooling finger Assembly of cooling fingers: [ UniVessel® Glass Operating Manual). ttDuring the process, check the heating jacket regularly. Black discolorations occurring on the power cord connection or on the silicone foam along the heating coil indicate that the heating coil and/or cable are defective. Immediately interrupt the operation and replace the heating jacket. ttWhenever there is contact with splashing water or media, interrupt heating operations, remove the heating jacket from the culture vessel, clean and dry it thoroughly. 84 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.6.4 Connecting the Aeration Modules 8.6.4.1 Conducting Preliminary Steps The culture vessels must be fitted with the equipment needed for medium aeration [ UniVessel® Glass operating manual]: −− Aeration pipe with ring sparger or microsparger or an aeration basket with a silicon tube membrane, −− Supply air filter, −− Exhaust cooler with exhaust filter, −− Supply air filter for headspace aeration when using the “Additive Flow” aeration module. Aeration tubing Overlay/Sparger for UniVessel® Glass, UniVessel® SU Standard designation Material Inner diameter of the hose Wall thickness 3.2 x 1.6; VMQ 7621; 55° Shore 3.2 mm Silicone hose transparent 1.6 mm 6 x 9; Silicone Transparent; 55° Shore rubber tubing Aeration tubing Sparger for BIOSTAT ® RM Rocker Standard designation Material 3.2 x 1.6; Silicone hose VMQ 7621; 55° Shore transparent Inner diameter of the hose Wall thickness 3.2 mm 1.6 mm 6 x 9; Silicone rubber Transparent; 55° Shore tubing Exhaust tube for BIOSTAT ® RM Rocker Standard designation Material Inner diameter of the hose Wall thickness 6 x 9; Silicone rubber Transparent; 55° Shore tubing The culture vessels must be autoclaved together with the supply air and exhaust air filters and then set up next to the associated supply unit. Configure the calibration parameters for the DO sensor and select the aeration mode using the DCU system [ Chapter “8.10 “Calibration” Menu“]. After autoclaving and before aeration with air and oxygen, you can carry out the zero point calibration of the pO2 sensor with nitrogen. Observe the instructions for zero point calibration of the pO2 sensor with nitrogen in vessels with “O2 Enrichment” and “Gasflow Ratio” aeration modules [ Chapter “8.6.4.3 Connecting the “Additive Flow 2-gas” Aeration System“]. 8.6.4.2 UniVessel® SU Safety Valve Station The safety valve station ensures that the specified maximum operating pressure of the culture vessel UniVessel® SU is not exceeded. Operating Manual BIOSTAT ® B 85 Preparing and Running the Process 4 2 5 3 The safety valve station is interconnected between bioreactor control unit and culture vessel. This prevents inadmissible overpressure in the culture vessel. ttSet up the safety valve station on a stable foundation in proximity to the bioreactor control unit. ttSet up the safety valve station such that the front (1) is facing you. 1 ttConnect the hoses on the outputs Overlay and Sparger with the inputs of the safety valve station (2) and (3) [ safety valve station installation instructions]. ttConnect the outlets of the safety valve station (4) and (5) to the corresponding inlets of the culture vessel UniVessel® SU [ UniVessel® SU operating instructions]. When laying and connecting the hoses, make sure not to bend or stretch them. 8.6.4.3 1 2 Connecting the “Additive Flow 2-gas” Aeration System Zero Point Calibration To calibrate the DO sensor’s zero point by feeding nitrogen into the vessel through the aeration system “O2 Enrichment” and “Gasflow Ratio”, proceed as follows: ttFor the zero point calibration, connect the laboratory's nitrogen supply to the “AIR” (3) connection at the rear of the supply unit. ttConnect the tubing from the outlet “Sparger” (1) to the supply air filter of the culture vessel. ttOpen the laboratory nitrogen supply and the variable area flow meter at the outlet “Sparger” (2). ttAerate the culture medium with nitrogen and calibrate the zero point [ Chapter “8.10 “Calibration” Menu“]. ttAfter zero point calibration, remove the laboratory’s nitrogen supply tubing from the “AIR” (3) connection. Fig. 8-6: Connection and gas flow control 3 ttConnect the laboratory's air supply to the “AIR” (3) inlet on the supply unit. ttAerate the culture medium with air and calibrate the slope [ Chapter “8.10 “Calibration” Menu“]. ttSet the gas flow rate that you want to use at the start of the process using the “Sparger” variable area flow meter or the DCU System's gas flow controller. If the supply unit is fitted with mass flow controllers for the gas supply, set the variable area flow meter for the “Sparger” outlet to the maximum gas flow rate. Fig. 8-7: Nitrogen supply connection to “AIR” 86 Operating Manual BIOSTAT ® B 8.6.4.4 1 2 3 Preparing and Running the Process Connecting the “Additive Flow 4-gas” Aeration System ttConnect the tubing from the outlet “Sparger” (1) to the supply air filter of the culture vessel. ttAerate the culture medium with nitrogen and calibrate the zero point [ Chapter “8.10 “Calibration” Menu“]. ttAerate the culture medium with air and calibrate the slope. ttConnect the tubing from the outlet “Overlay” (2) to the supply air filter of the culture vessel. ttSet the gas flow for your process at the variable area flow meters (1). If the supply unit is fitted with mass flow controllers for the gas supply, set the variable area flow meter for the “Sparger” and “Overlay” outlets to the maximum gas flow rate. Fig. 8-8: Connection and gas flow control Operating Manual BIOSTAT ® B 87 Preparing and Running the Process 8.6.5 Preparing the Corrective Solution Supply The supply unit is fitted with up to 8 integrated peristaltic pumps WM 114 for supplying correction media (acid, base, anti-foam agents or nutrient solutions/ substrates). Preliminary Steps: The culture vessels must be fitted with the following equipment needed for supplying correction media or media removal [UniVessel® Glass Operating Manual]: −− pH sensor, feed pipe for acid and base −− Anti-foam sensor, feed pipe for anti-foam agent −− Harvest pipe for removing the medium The bottles must be prepared [ Chapter “8.3 Connecting Transfer Lines“]. Danger of limbs being pulled into the rotation pump and crushed! −− Allow only qualified personnel to work on the device. −− Prior to putting the hoses in, turn the peristaltic pumps to “off”. Configuring the Tube Holder for the Peristaltic Pump It's possible to put hoses with different hose diameters into the peristaltic pumps. The tube holder must be set up according to the used hose diameter. ttIn order to conduct adjustments, lift up the cover of the peristaltic pump. The position of the tube holder can be determined with the markings on the tube holder (1) and on the housing (2, 3). 1 2 The following table can be used to determine the position of the tube holder depending on the inner diameter of the hose. 3 Inner diameter of the hose 0.5 mm 0.8 mm 1.6 mm 2.4 mm 3.2 mm 4.0 mm 4.8 mm Position of the tube holder 3 3 2 2 2 2 2 Cross-section 1 2 3 Fig. 8-9: Position of the tube holder Using bigger hoses (4.0 - 4.8 mm inner diameter) with the tube holder in position (3; for small hoses) will result in decreasing feeding volumes and lifespan. Using smaller hoses (0.5 - 0.8 mm inner diameter) with the tube holder in position (2; for big hoses) will increase the risk that the hose gets into the pump head and thus bursts. 88 Operating Manual BIOSTAT ® B Preparing and Running the Process Changing Position of the Tube Holder Changing to smaller hose diameter: Prior to changing the tube holder position, turn off the pump. In order to re-position the lower tube holders on both sides of the pump head, use a pointed object (e.g. a ballpoint pen). ttInsert the pointed object into the recess (1) and push the instrument (e.g. a ballpoint pen) downwards. ttPush the tube holder’s flange towards position (2), until the flange snaps into the new position. 1 3 2 yyBy now, the marking of the tube holder should be covering the marking for the small hose diameter (2). ttReduce the pressure laid on the instrument. yyThe flange should rise up and be aligned correctly. If the flange does not rise up, repeat the procedure and maintain downward pressure until release. The tube holder on the other side of the pump head is adjusted accordingly. Changing to bigger hose diameter: Fig. 8-10: Position of the tube holder Conduct the steps as described in the section above. Push the tube head towards the opposite direction, allowing the flange to snap in position (3). Soiling on the Peristaltic Pump Mechanisms When no hose is inserted after setting up the tube holder position, the cover of the peristaltic pump must be sealed. Soilings in the mechanisms can result in malfunctions and decrease the lifespan of the peristaltic pump. Inserting and Removing Hoses Check if the tube holders on both sides of the pump head are set up according to the used hose size [Æ “Changing Position of the Tube Holder“]. ttLift up the cover completely. ttMake sure that there is sufficient hose available for the curvature inside the pump's track. The hose must be positioned between the rotor-roller and the track; press and hold it towards the inner wall of the pump head. The hose must not be seated to the rolls in a twisted or bended manner. ttTip down the cover until it snaps into the closed position. yyThe track closes automatically and the hose will be properly bent. ttIn order to remove the hose element, conduct the steps in reverse order. Fig. 8-11: Inserting the hose Operating Manual BIOSTAT ® B 89 Preparing and Running the Process 8.7 Performing a Process Danger of injury from shattered glass! The culture vessel can burst due to unallowed overpressurization, resulting in cuts and injury to the eyes inflicted by glass splinters. −− Only operate the temperature control circuit of double-walled culture vessels at ambient pressure. Do not exceed a max. pressure of 0.8 barg (18.85 psig) when aerating the culture vessels [ UniVessel® Glass operating manual). −− Make sure that the culture vessel is positioned stably. −− Wear personnel protective equipment. −− Make sure that the culture vessel is properly connected to the supply unit. −− Make sure that the culture vessel does not overflow. −− Observe the filling of the culture vessel and ensure that none of the connected collection container exceed the volume available in the culture vessel. −− Make sure that the cooling water flows back without pressure. −− Regularly check all lines, hoses and connections under pressure for leaks and externally detectable damage. Danger of contamination from escaping feed and culture media! Accidentally released hazardous substances, infectious cultures and corrosive media can result in health hazards. −− Follow the company’s safety instructions (e.g. for processes which pose special requirements to your workplace, the use of components or on handling of media and contaminated components). −− Empty the feed hoses before loosening the hose connection. −− Wear personnel protective clothing. −− Wear safety glasses. Danger of contamination from media and cultures used and products generated in the process! The media and cultures used for the fermentation process and the products generated during it can be hazardous to health. −− Where necessary, always disinfect or sterilize contaminated equipment. This can be done by filling the UniVessel® Glass and the accessories that have come into contact with the culture with water before dismantling and cleaning it, and autoclaving it again. −− It may be sufficient to heat the UniVessel® Glass to >65 °C for approx. 1 hour. This will be sufficient for killing most live cells, but will not kill spores or thermopile microorganisms. −− When using non-hazardous cultures and media, the UniVessel® Glass only needs to be carefully rinsed with water. 90 Operating Manual BIOSTAT ® B Preparing and Running the Process Danger of chemical burns from acids and bases! Excess acids and alkalines in correction media bottles can cause chemical burns in the event of uncontrolled leakage! −− In order to neutralize the acids and alkalines, empty the lines into appropriate containers. −− All other equipment that has come into contact with acids, bases or (potentially) hazardous media must be treated with appropriate cleaning solutions or safely disposed of. Danger of burns upon contact with hot surfaces of culture vessels! In double wall vessels, the outlets on the temperature control module, the temperature control system hoses and the culture vessel, can become very hot and cause burns. In single wall culture vessels, the heating jackets become hot. ttWear protective gloves when working with hot culture media. Danger of burns upon contact with hot surfaces of the stirrer drive motor! The stirrer motors can become hot if operated over long periods of time, at high speed and when stirring high viscosity media. −− Take note of the motor’s safety label. This label will become discolored at high temperatures. −− Avoid accidental contact with the motor and only touch stirrer drive motors during the fermentation process wearing gloves. Operating the stirrer at impermissible high speeds can affect the culture vessel’s stability and cause damage to its fittings. Depending on the size and equipment of the culture vessels, the permissible speed may be limited, e.g. to max. 300 min-1 in vessels with aeration rigs for bubble-free aeration. 8.7.1 Setting Up the Measurement and Control System Carry out the following steps: ttSwitch on all peripheral devices (e.g. exhaust filter heater). ttCheck for any malfunctioning or failure. Error messages from the DCU system are shown on the operator display [ Chapter “9.4 Process-related Faults / Alarms“]. ttSelect the measurement and control system and enter the parameters required for the process: Operating Manual BIOSTAT ® B 91 Preparing and Running the Process 8.7.2 Guaranteeing Sterility Sterility Test Before starting the fermentation process, it is possible to perform a sterility test. The sterility test can be used to establish whether the culture vessels and the connected equipment have been safely sterilized or are contaminated. ttEnter all of the process parameters as specified (temperature, speed, aeration, pH regulation, etc.). ttLeave the bioreactor running for 24 hours and monitor it for signs of error, e.g.: −− change in the pH value −− unexpectedly high oxygen consumption −− cloudiness of the medium −− unusual odors in the exhaust air These signs could indicate that the equipment has not been properly sterilized or that germs from the environment have made their way inside through defective or insufficiently tightened connections and seals. Corrective measures: ttSterilize with new medium for longer sterilization times. Do not raise the sterilization temperature! ttWhen performing operations with a Single Use vessel (UniVessel® SU, culture bag): Dispose of this one and install a new one. ttDismantle all of the vessel equipment and connections and check all of the seals and lines for damage. 8.7.3 Carrying out the Cultivation Process ttTransfer the inoculation culture into the culture vessel. ttPerform the scheduled process steps. ttTake samples, given this is necessary to monitor the process flow. ttOn completion of the process, harvest the culture and transfer it to the next point of use (scale-up, processing, etc.). Switching Off the System ttIf no other processes are being performed using the supply unit, turn them off at the main switch. 92 Operating Manual BIOSTAT ® B 8.8 “Main” Menu 8.8.1 General Information Preparing and Running the Process The “Main” menu opens after the control unit has been switched on. This is the central starting point for in-process operation. Example with glass culture vessels: Fig. 8-12: Start screen Twin variant “Main All” menu. Fig. 8-13: Start screen single variant in “Main” menu. The graphical display of the system structure simplifies the overview of the system components and uses function elements implemented as touch buttons to provide access to the submenus for the most important or most frequently used settings. If practical, the function elements also show the currently entered or configured data and settings. Which function elements are actually shown depends on the configuration of the DCU system, the unit controlled, e.g. the type of bioreactor, and customer specifications. Operating Manual BIOSTAT ® B 93 Preparing and Running the Process 8.8.2 Process Displays in the “Main” Menu The function elements can display associated process values: −− Values measured by connected probes such as pH, pO2, foam, etc. −− Calculated variables like pump filling amounts, calculated values of arithmetic functions, etc. −− Process duration displays −− Measured data and key figures from the responses of external components such as speed regulation, mass flow controllers, scales, etc. 8.8.3 Direct Access to Submenus The following menu screens show examples of submenus accessible from the “Main” menu and configuration options for the measurement and control system. Which submenus are available and which parameters can be configured depends on the configuration: −− Setpoint specifications and mode selection for headspace aeration (overlay) for air and CO2 −− Setpoint specifications and mode selection for supplying the media with gas (sparger); for all gases, example menu “AIROV-#” −− Settings for alarm limits and activation of alarm monitoring for totalizer; example “ACIDT-#” −− Mode selection for correction medium pumps, example “SUBS-A#” 94 Operating Manual BIOSTAT ® B Preparing and Running the Process −− Mode selection for agitator speed “STIRR-#” −− Mode selection for level control “LEVEL-#” −− Analogous for foam monitoring “FOAM-#” −− “LEVEL #” pump control mode selection (automatic and manual pump control) Fig. 8-14: Menu screens for functions accessible directly from the “Main” menu Operating Manual BIOSTAT ® B 95 Preparing and Running the Process 8.9 “Trend” Menu 8.9.1 “Trend” Display Using the “Trend” display, the user can graphically display process values for a period of up to 72 hours. This overview of the process flow gives for example a quick impression as to whether the process is running as expected or whether irregularities or disruptions are present. The trend display works retroactively from the current point in time and offers: −− Up to 8 (selectable) channels −− Time basis freely selectable from 1 to 72 hours Trend readouts are not saved. Use a host system (e.g. MFCS) to log these data and permanently document the process value time profiles. Operator Screen Fig. 8-15: Start screen “Trend” menu BIOSTAT ® B (no recording active) Field Value Function, entry required Key line 1…8 Display and set channels Diagram 1…8 Line chart of the selected channels (y) over time (x) Upper Upper limits of the selected display ranges for each channel Middle Colored line chart Lower Lower limits of the selected display ranges for each channel HH:MM Time scale Subtitle 96 Operating Manual BIOSTAT ® B 8.9.2 8.9.2.1 Preparing and Running the Process Configuring the “Trend” Display Setting the Trend Display for Parameters ttSelect the “Trend” menu button. ttPress the key of the channel that you want to set: yyThe “Channel # Settings” window opens. Fig. 8-16: Menu for parameter selection and setting ttTo change the parameter for the channel, press “PV”. yyThe “Select Buffered Channel” menu shows the default values: ttPress “Cfg” to display all parameters in the configuration. If the desired parameter is not visible, you can scroll through the table. ttPress the key for the parameter to select it. yyThe parameter will be activated immediately. ttTo deselect a parameter without having to re-assign the channel, press “.....”. Fig. 8-17: Overview table of default parameters 8.9.2.2 Setting the Parameter Display Range ttSelect the “Channel # Settings” window and press “Min” and | or “Max”. ttEnter the upper and | or lower limit. The display limit values for the parameter are shown underneath the data window. ttConfirm the input with “OK”. Fig. 8-18: Example for setting the upper temperature limit Operating Manual BIOSTAT ® B 97 Preparing and Running the Process 8.9.2.3 Resetting the Display Range ttPress “Reset Range” on the “Channel # Settings” screen to reset a modified display range back to the factory setting for “Max” and “Min”. Fig. 8-19: Resetting a running trend recording 8.9.2.4 Setting the Trend Display Color The color for every parameter can be selected from a table. ttSelect the “Channel # Settings” screen and press the key with the name of the preselected color. ttPress the key with the name of the new color to be used. yyThe selection is instantly assigned and activated. Fig. 8-20: Assigning a color to the selected parameter 8.9.2.5 Defining a New Time Range as “Time Range” ttPress the [“h”] key in the header. ttSelect the desired time range. yyThe time scale below in the working environment changes automatically. yyThe parameter trend will be displayed over the new time range. Fig. 8-21: Selecting the display range 98 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.10 “Calibration” Menu 8.10.1 General Information In the “Calibration” menu, all calibration actions required for routine operation can be activated: −− Calibration routines for sensors: e.g. pH, pO2 −− Calibration of pump filling counter: e.g. acid, base, substrate −− Calibration of gas filling counter: e.g. N2, O2, CO2 Fig. 8-22: “Calibration” menu (configuration example) After pressing the “Calibration” menu button, the calibration menu opens. Selectable touch buttons show the status of the associated calibration functions and open the associated submenu to carry out calibration routines. Operating instructions on the individual steps and required entries on the display lead the user through the menus. The calibration parameters remain stored when the DCU system is switched off. After power is restored, the DCU system uses the saved figures until a new calibration is carried out. 8.10.2 pH Calibration (conventional sensor) 8.10.2.1 General Information on the pH Sensor Conventional pH sensors are calibrated using a two-point calibration with buffer solutions. During pH measurement, the system calculates the pH value based on the sensor potential according to the Nernst equation while taking zero point deviation, slope and temperature into consideration. Sensors are calibrated before installation at the point of measurement, e.g. in the culture vessel. Sterilization can alter the sensor zero point. To recalibrate the pH sensors, you can measure the pH value externally in a sample taken from the process and enter the value into the calibration menu. The calibration function compares the pH value measured online with the one determined externally, calculates the resulting zero offset and displays the corrected process value. Operating Manual BIOSTAT ® B 99 Preparing and Running the Process The effects of heat during sterilization and reactions of the diaphragm and/or electrolytes with components of the medium can influence the measurement properties of the pH sensors. Test and calibrate the pH sensors before each use. −− Whenever possible, use buffer solutions manufactured by the sensor manufacturer as contained in the equipment supplied with the pH sensor. Information on reordering is available on request. −− If the “zero offset” and “slope” values are known and the process allows, you can also enter these values directly into the relevant fields. −− The sensor’s service life is limited and depends on the in-process working and operating conditions. Whenever a function check or calibration points to a malfunction, the pH sensor should be serviced and replaced as needed. −− The pH sensors must be serviced or replaced when the following values are outside the specified range*: −− Zero point offset (“zero”) outside – 30 to + 30 mV −− Depending on the type and design of the sensors supplied, the menus, sequence and operation of the calibration function can differ from the information provided herein. Please refer to the configuration documents or to the function specifications of the bioreactor, if available. *These values apply for pH sensors from the manufacturers Hamilton and Mettler Toledo. If you use other manufacturers” products, please refer to the manufacturer’s documentation. 8.10.2.2 “Calibration pH-#” Submenu The submenu “Calibration pH-#” shows both the pH value and the measurement chain voltage of the sensors, as well as the zero offset (“zero”) and slope sensor parameters. That allows you to easily check the functionality of the pH sensors. ttPress the touch button of the sensor to be calibrated (“pH-#”) in the “Calibration” menu. yyThe submenu “Calibration pH-#” opens: Field Value Function, entry required Mode Measure −− Open the submenu “Calibration pH-# Mode“ −− Automatic switch to pH measurement after calibration routine is completed Calibrate −− Perform complete calibration Recalibrate −− Perform recalibration Calibrate Zero −− Perform zero point calibration as a single step Calibrate Slope −− Perform slope calibration as a single step pH pH Display of pH measured value Electrode mV Combination electrode voltage (raw signal) TEMP °C Temperature value for temperature compensation Zero mV Display / entry of the zero point offset Slope mV | pH Display / entry of the slope Manual Auto 100 Operating Manual BIOSTAT ® B Temperature compensation with entry of a temperature measured manually outside the culture vessel Temperature compensation with the temperature measured in the culture vessel (DCU system) Preparing and Running the Process ttPress the touch button “Measure” in the submenu “Calibration pH-#”. yyThe submenu “Calibration pH-# Mode” opens: 8.10.2.3 Perform Calibration Depending on your choice, only the zero point (Calibrate Zero) or the slope (Calibrate Slope) is calibrated, or a full calibration (Calibrate) is carried out. Select/enter Temperature Compensation ttPress the touch button “Calibrate” in the submenu “Calibration pH-# Mode”. ttSelect the type of temperature compensation. ttIf “Manual” is selected: Enter the value for temperature compensation. yyThe input window “pH-#: Zero Buffer” for zero point calibration is displayed. Zero Point Calibration ttHold the pH sensor in a buffer solution (generally 7.00 pH). ttIn the input window “pH-#: Zero Buffer”, input the pH value of the buffer solution. ttObserve the measured value display in the window “pH-#: Zero Value”. Once the display is stable, confirm the measurement with “OK”: Operating Manual BIOSTAT ® B 101 Preparing and Running the Process Slope Calibration ttHold the pH sensor in the second buffer solution. ttIn the input window “pH-#: Slope Buffer”, input the pH value of the second buffer solution. ttObserve the measured value display in the window “pH-#: Slope Value”. Once the display is stable, confirm the measurement with “OK”: yyThe pH sensor is calibrated. 8.10.2.4 Direct Input of the Zero Offset and Slope Zero Offset ttPress the touch button “Zero” in the submenu “Calibration pH-#”. ttIn the input window “pH-#: Zero Buffer”, input the pH value. ttIn the input window “pH-#: Zero Value”, input the measured value for the zero offset. Slope ttPress the touch button “Slope” in the submenu “Calibration pH-#”. ttIn the input window “pH-#: Slope Buffer”, input the pH value. ttIn the input window “pH-#: Slope Value”, input the measured value for the slope. 8.10.2.5 Perform Recalibration You can only recalibrate a single pH sensor. By following the operating steps described below, you can adapt the calibration of a pH sensor to changed measuring conditions after a sterilization cycle in the autoclave or during the process as needed: ttTake a sample from the process. ttMeasure the pH value of the sample with a calibrated pH measurement device. ttPress the touch button “Re-Calibrate” in the submenu “Calibration pH-# Mode”. ttEntered the pH value measured with the measurement device. yyThe DCU system calculates the zero offset and displays the corrected pH value. 102 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.10.3 DO Calibration (conventional sensor) 8.10.3.1 General Information on the pO2 Sensors Calibration of the pO2 sensor is based on a two-point calibration. Measurement is performed in [% oxygen saturation]. Calibration determines the sensor parameters zero current (“zero”) and slope (“slope“). The reference parameter for “zero” is the oxygen-free medium in the culture vessel. Air-saturated medium can be defined as 100 % saturated and be the basis for determination of the slope. Since you will be calibrating the sensor after sterilization, take changes in the measuring properties that can result from heat exposure or effects of the medium during sterilization into consideration. Special Notes Prior to first use or whenever the pO2 sensor has been disconnected from the power supply (measurement amplifier) for longer than 5 to 10 min., it has to be polarized. Polarization can take up to 6 hours (less time when the sensor was only disconnected from the measurement amplifier for a few minutes). This does not apply to optical pO2 sensors (e.g. VISIFERM, Hamilton). Follow the sensor manufacturer’s instructions. 8.10.3.2 Submenu “Calibration pO2-#” In addition to pO2 saturation, the submenu “Calibration pO2-#” also shows the current sensor current as well as the zero current and slope with calibration conditions. This allows easy regulation of the sensor’s functions. ttPress the touch button of the sensor to be calibrated (“pO2-#”) in the “Calibration” menu. yyThe submenu “Calibration pO2-#” opens: Field Value Function, entry required Mode Measure −− Open submenu Mode −− Automatic switch to DO measurement after calibration routine is completed Calibrate −− Perform complete calibration Calibrate Zero −− Perform zero point calibration as a single step Calibrate Slope −− Perform slope calibration as a single step DO (pO2) pH Display DO saturation Electrode mV Combination electrode voltage (raw signal) TEMP °C Temperature value for temperature compensation Zero mV Display / entry of the zero point offset Slope mV | pH Display / entry of the slope Manual Temperature compensation with entry of a temperature measured manually outside the culture vessel Temperature compensation with the temperature measured in the culture vessel (DCU system) Auto Operating Manual BIOSTAT ® B 103 Preparing and Running the Process ttPress the touch button “Measure” in the submenu “Calibration pO2-#”. yyThe submenu “Calibration pO2-# Mode” opens: 8.10.3.3 Perform Calibration Depending on your choice, only the zero point (Calibrate Zero) or the slope (Calibrate Slope) is calibrated, or a full calibration (Calibrate) is carried out. The pO2 sensor must be serviced if: −− the zero point is not within the range of 0 to … +10 nA (window “pO2-#: Zero Value”), −− the sensor current at maximum aeration with air is lower than 30 nA (window “pO2-#: Slope Value”). Select/enter Temperature Compensation ttPress the touch button “Calibrate” in the submenu “Calibration pO2-# Mode”. ttSelect the type of temperature compensation. ttIf “Manual” is selected: Enter the value for the temperature compensation and confirm the entry with “OK”. yyThe input window “pO2-#: Zero Buffer” for zero point calibration is displayed. Zero Point Calibration Zero point calibration of the DO sensor can be performed as follows: −− on the laboratory table in a gel sample (0% oxygen saturation), −− in medium gassed with nitrogen (only BIOSTAT ® B-CC) (as described in the following): ttInstall the DO sensor on the culture vessel. ttSet the aeration “N2” to 100%, all other aeration to 0%. ttSet the stirrer speed (STIRR) to approx. 80% to 100%. ttIn the input window “pO2-#: Zero Buffer”, input the pO2 value (typically 0%). ttWait until the oxygen dissolved in the medium has been displaced. yyWhen the raw electrode signal stabilizes near the 0 nA value, the oxygen saturation is approaching the minimum. ttObserve the measured value display in the window “pO2-#: Zero Value”. Once the display is stable, confirm the measurement with “OK”: 104 Operating Manual BIOSTAT ® B Preparing and Running the Process Slope Calibration Slope calibration of the DO sensor can be performed as follows: −− on the laboratory table in the ambient air, −− in medium gassed with air (only BIOSTAT ® B-CC) (as described in the following): ttSet the aeration “AIR” to 100%, all other aeration to 0%. ttSet the stirrer speed (STIRR) to approx. 80% to 100%. ttIn the input window “pO2-#: Slope Buffer”, input the pO2 value (typically 100 %). yyWhen the raw electrode signal stabilizes near the 60 nA value, the oxygen saturation is approaching the maximum (this value is only applicable to Hamilton sensors). ttObserve the measured value display in the window “pO2-#: Slope Value”. Once the display is stable, confirm the measurement with “OK”: yypO2 sensor is calibrated. 8.10.3.4 Direct Input of the Zero Offset and Slope Zero Offset ttPress the touch button “Zero” in the submenu “Calibration pO2-#”. ttIn the input window “pO2-#: Zero Buffer”, input the DO value. ttIn the input window “pO2-#: Zero Value”, input the measured value for the zero offset. Slope ttPress the touch button “Slope” in the submenu “Calibration pO2-#”. ttIn the input window “pO2-#: Slope Buffer”, input the DO value. ttIn the input window “pO2-#: Slope Value”, input the measured value for the slope. 8.10.4 Optical pH and pO2 Sensors Sartorius Stedim Biotech’s optical sensor technology makes it possible to measure the pH and dissolved oxygen (DO) values non-invasively. The sensors are integrated into various systems. On the UniVessel® SU, the sensor patches are located on the bottom of the single-use vessels where they are read off directly via free-space optoelectronics. In culture bags, the sensor patches are pre-installed on the flexible bag wall or inserted through a sensor port. In this case, the readout takes place via a flexible fiber cable. All systems are evaluated for the cultivation of cell culture and microbial fermentations with scalability ranging from the process development to the production scale. Sunlight or long exposure to daylight damages the pH sensors. The optical pH sensors are useless if they are exposed to approximately 8 days of daylight or 2 hours of direct sunlight. −− Calibrate the optical pH sensor just before inoculation and after setting up the culture bag or the UniVessel® SU. Operating Manual BIOSTAT ® B 105 Preparing and Running the Process 8.10.4.1 Signal Quality of the Optical Probes Culture bags in the version “Optical” are equipped with optical single-use pH and DO (pO2) sensors. An optical fiber is used for the sensor connection. The sensor is located at the end of a hose, in the interior of the culture bag. The fiber-optic cable transmits light at a specific wavelength from the measurement amplifier to the sensor and the sensor’s luminescent response back to the measurement amplifier. Good contact between the optical fiber and the sensor is a prerequisite for accurate measurement. The amplitude of the signal is indicative of the signal quality. When the optical fiber is not properly inserted into the end of the tube, the signal quality is adversely affected. In the menu “Calibration” [Æ Fig. 8-23], the values for the pH and DO amplitudes are displayed (without dimensions). You can accept values between 0 and >50000. The value is displayed in steps of 1000. Example: The display value “31” represents an amplitude value of 31,000. −− The PV “pO2_Ampl” shows the signal strength (amplitude) of the DO sensor. −− The PV “pH_Ampl” shows the signal strength (amplitude) of the pH sensor. For precise measurement, the values “pO2_Ampl” and “pH_Ampl” should be greater than 10,000 units, after the sensors were wettened for at least 2 hours. If the value is less than 10,000 units, this indicates a fiber optic cable was not correctly installed. ttMake sure that the fiber optic cable is properly installed and secured with the securing clips. Fig. 8-23: Main menu “Calibration 2” (culture bags) with display of the optical probe signal quality 106 Operating Manual BIOSTAT ® B 8.10.4.2 Preparing and Running the Process Notes on Calibration The indicator strip for the measured value sensors decays when exposed to light. The measured value drifts by approx. 0.13 pH based on 10,000 measurements. To compensate for this drift, enhanced DCU configurations provide a “recalibration” function. Measurements should therefore be taken as seldom as possible during the process. Additionally, the acceptable measurement accuracy for the process and the possible number of resulting measurements (measurement cycle duration) can be determined empirically. −− pH sensor: During typical pH calibration, reference values for zero point and slope are measured in calibration buffers. This is not possible with disposable optical pH sensors, as they are pre-sterilized before incorporation into the culture bags or UniVessel® SUs. Instead, a method was developed that determines the typical measured values pH0 | pH and phase-angle f (min) | f (max) for a batch of sensors. This information is provided with the culture bags or UniVessel® SUs (label sticker with calibration parameters) and entered in the pH calibration menu [ Chapter “8.10.5 pH Calibration (optical sensor)“]. −− DO sensor (pO2): The typical measured value at 0% and 100% DO (pO2) is determined for a batch of sensors. This information is provided with the culture bags or UniVessel® SUs (label sticker with calibration parameters) and entered in the DO calibration menu [ Chapter “8.10.6 DO Calibration (optical sensor)“]. 8.10.5 pH Calibration (optical sensor) General notes on optical sensors can be found in [Æ Chapter 8.10.4 on page 105]. To calibrate optical pH sensors, as follows: ttEnter the initial calibration data [Æ Chapter 8.10.5.2 on page 109]. ttWait until the medium has reached the process temperature. Let the probes soak in the medium for at least 2 h. ttTake an offline sample and perform a re-calibration [Æ Chapter 8.10.5.3 on page 110]. It is recommended to recalibrate the pH sensor daily. A recalibration is also necessary if the ionic strength of the medium is changed by adding feed etc. Operating Manual BIOSTAT ® B 107 Preparing and Running the Process 8.10.5.1 “Calibration pH-#” Submenu Field Mode Value Function, entry required Display of the active operating mode: Measuring, Calibrating, Recalibrating −− Inactive [Inactive] Appears after commissioning, before the first calibration −− Calibrate [Calibrate] Appears when going through the calibration steps −− Measure [Measure] Indicates that the measurement is active in process −− Hold [Hold] Indicates that the measurement in process has been paused −− Re-Calibration [Re-Calibration] Appears during re-calibration in process pH pH Current pH measured value TEMP °C Type of temperature compensation; switch between: −− Automatic compensation for pH measurement in process −− Manual compensation to calibrate the pH sensor (do not use during normal operation) Samp. Rate s Lot No. Measurement cycle (waiting time between individual measurements) −− Setting range: 5 to 3600 s; recommended (default value) 30 s −− Choose a measurement cycle that produces a maximum number of measurements at acceptable accuracy [ Chapter “8.10.5.4 Configuring the Measurement Cycle for pH Measurement“]. Manufacturer reference for released batch of culture vessels Temp Comp °C Reference temperature for calibration f (max) ° Phase reference, reference pH (reference measurement deviating from the zero point) f (min) ° Phase reference, zero point pH (during reference measurement for “zero point”) dpH pH Reference pH for sensor batch (typical deviation) pHO pH Typical zero point pH sensor batch Meas. Cnts. Number of measurements performed Act. Sample Re-calibration reference value Parameter Display of the calibration parameters ttPress the touch button of the sensor to be calibrated (“pH-#”). yyThe operator screen “Calibration pH-#” opens: 108 Operating Manual BIOSTAT ® B 8.10.5.2 Preparing and Running the Process Enter Initial Calibration Data The calibration data to be entered are printed on the culture bag or UniVessel® SU used. These data must be entered, as no (valid) pH measurement is possible beforehand. (When using the UniVessel® SU, the calibration data can be scanned with the barcode scanner). ttPress the button “Inactive” to bring up the window “Mode Calibration pH #”. Entering the Parameters ttPress the touch button “Enter init. Parameters” to enter the parameters. Parameters can be entered in two ways: −− Scanning the parameters with the barcode scanner (only in conjunction with UniVessel® SU) −− Manual entry of the parameters ttScanning the parameters from the culture vessel label. (only in conjunction with UniVessel® SU) yyWait until [ok] is active. tt[Manual]: Check the scanned parameters or enter the following parameters from the culture vessel label successively into the respective input window and confirm the entry with [Enter] or [ok]. −− “Lot-No.” −− Temperature compensation −− “pH f (max)” −− “pH f (min)” −− “pH dpH” −− “pH pHO” tt[ok]: Confirm the parameters. ttCheck the parameters displayed. ttBy pressing the respective key, the corresponding parameter can be modified as necessary. ttConfirm the manually-entered | scanned parameters with [ok]. Operating Manual BIOSTAT ® B 109 Preparing and Running the Process Transferring the Parameters yyThe data are transferred to the DCU system. ttWait until the parameters have been transferred. yyThe initial calibration of the pH sensor is now complete. ttTake a sample and recalibrate the pH sensor in case of deviations. 8.10.5.3 Performing Recalibration ttPress the button “Inactive” to bring up the window “Mode Calibration pH #”. ttPress the touch button “Re-Calibrate“ to begin recalibration. ttPress the touch button “Act. Sample”. ttTake a sample from the process. ttMeasure the pH value of the sample taken with a calibrated pH measurement device. ttEntere the pH value measured with the measurement device. yyConfirm the input with [ok]. The DCU system calculates the zero offset and displays the corrected pH value. Depending on the operating mode the device switches to the operating mode [Measure] automatically or must be manually switched to the operating mode [Measure]. −− After successful initialization | calibration, the operating mode [Measure] is switched to automatically. −− After the operating mode [Hold] the operating mode [Measure] must be switched to manually. 110 Operating Manual BIOSTAT ® B 8.10.5.4 Preparing and Running the Process Configuring the Measurement Cycle for pH Measurement Optical pH sensors show decay of the indicator dyes, e.g. photo-bleaching. This degradation depends on the amount of light and increases with the pH value (for alkaline media). The pH sensors used in culture bags are designed for 20,000 measurement points. Calculation of the Measurement Cycle The measurement cycle can be configured so that 20,000 measurements are possible over the total process time. Example of calculation specifications: −− Overall process time = 666 hours (approx. 28 days) −− Maximum number of measurements = 20,000 Calculation: 20,000 measurement cycles / 666 hours = 30 measurement cycles / hour = one 120-second measurement cycle per measurement Modification of the Measurement Cycle ttIn the operator screen “Calibration pH-#”, press the touch button “Samp. Rate” to modify the measurement cycle. ttEnter the standard password “19” and confirm with [ok]. ttChange the value for the pH measurement cycle according to the above calculation. ttConfirm the input with [ok]. 8.10.6 DO Calibration (optical sensor) General notes on optical sensors can be found in [Æ Chapter 8.10.4 on page 105]. To calibrate optical DO sensors, proceed as follows: ttEnter the initial calibration data [Æ Chapter 8.10.6.2 on page 113]. ttWait until the medium has reached the process temperature. Let the probes soak in the medium for at least 2 h. ttTake an offline sample and perform a re-calibration [Æ Chapter 8.10.6.3 on page 114]. Operating Manual BIOSTAT ® B 111 Preparing and Running the Process 8.10.6.1 Submenu “Calibration pO2-#” Field Value Mode Function, entry required Display of the active operating mode: Measuring, Calibrating, Recalibrating −− Inactive [Inactive] Appears after commissioning, before the first calibration −− Calibrate [Calibrate] Appears when going through the calibration steps −− Measure [Measure] Indicates that the measurement is active in process −− Hold [Hold] Indicates that the measurement in process has been paused −− Re-Calibration [Re-Calibrate] Appears during re-calibration in process DO (pO2) % Current DO measured value TEMP °C Type of temperature compensation; switch between: −− Automatic compensation for DO measurement in process −− Manual compensation to calibrate the DO electrode (do not use during normal operation) Samp. Rate s Measurement cycle (waiting time between individual measurements) −− Setting range: 5 to 3600 s; recommended (default value) 5 s −− Choose a measurement cycle that produces a maximum number of measurements at acceptable accuracy [Æ Chapter “8.10.6.4 Configuring Measurement Cycles for DO Measurement“]. Lot No. Manufacturer reference for released batch of culture vessels Temp Comp °C Reference temperature for calibration 0% sat % Typical reference zero point (zero DO) of the batch 100 % sat % Typical reference slope (slope DO) of the batch Meas. Cnts. Number of measurements performed Act. Sample Re-calibration reference value Parameter Display of the calibration parameters ttPress the touch button of the sensor to be calibrated (“pO2-#”). yyThe submenu “Calibration pO2-#” opens. 112 Operating Manual BIOSTAT ® B 8.10.6.2 Preparing and Running the Process Perform Initial Calibration The calibration data to be entered are printed on the culture bag or UniVessel® SU used. These data must be entered, as no (valid) DO measurement is possible beforehand. (When using the UniVessel® SU, the calibration data can be scanned with the barcode scanner). ttPress the touch button of the sensor to be calibrated (“pO2-B#”). ttPress the key “Inactive” to start initial calibration. Entering the Parameters ttPress the touch button “Enter init. Parameters” to enter the parameters. Parameters can be entered in two ways: −− Scanning the parameters with the barcode scanner (only in conjunction with UniVessel® SU) −− Manual entry of the parameters ttScanning the parameters from the culture vessel label. (only in conjunction with UniVessel® SU) yyWait until [ok] is active. tt[Manual]: Check the scanned parameters or enter the following parameters from the culture vessel label successively into the respective input window and confirm the entry with [Enter] or [ok]. −− “Lot-No.” −− Temperature compensation −− “pO2 0 %” −− “pO2 100 %” tt[ok]: Confirm the parameters. ttCheck the parameters displayed. ttBy pressing the respective key, the corresponding parameter can be modified as necessary. ttConfirm the manually-entered | scanned parameters with [ok]. Operating Manual BIOSTAT ® B 113 Preparing and Running the Process Transferring Parameters yyThe data are transferred to the DCU system. ttWait until the parameters have been transferred. yyThe initial calibration of the DO sensor is now complete. 8.10.6.3 Performing Recalibration ttPress the button “Inactive” to bring up the window “Mode Calibration pH #”. ttPress the touch button “Re-Calibrate“ to begin recalibration. ttPress the touch button “% sat”. ttTake a sample from the process. ttMeasure the DO value of the sample taken with calibrated DO measuring equipment. ttEnter the DO value measured with the measurement device. ttConfirm the input with [ok]. yyThe DCU system calculates the zero offset and displays the corrected DO value. Depending on the operating mode the device switches to the operating mode [Measure] automatically or must be manually switched to the operating mode [Measure]. −− After successful initialization | calibration, the operating mode [Measure] is switched to automatically. −− After the operating mode [Hold] the operating mode [Measure] must be switched to manually. 8.10.6.4 Configuring Measurement Cycles for DO Measurement Optical DO sensors show decay of the indicator dyes, e.g. photo bleaching. Calculation of the Measurement Cycle The measurement cycle can be configured so that 200,000 measurements are possible over the total process time. Example of calculation specifications: −− Overall process time = 1666 hours (approx. 69 days) −− Maximum number of measurements = 200,000 Calculation: 200,000 measurement cycles / 1666 hours = 120 measurement cycles / hour = two measurement cycles / minute 114 Operating Manual BIOSTAT ® B Preparing and Running the Process Modification of the Measurement Cycle ttIn the submenu “Calibration pO2-B#”, press the touch button “Samp. Rate” to modify the measurement cycle. ttEnter the standard password “19” and confirm with [ok]. ttChange the value for the pO2 measurement cycle according to the above calculation. ttConfirm the input with “OK”. 8.10.7 Totalizer for Pumps and Valves Function To document the correction medium consumption, the DCU system adds up the running times of the pumps or proportioning valves. It calculates the feed volumes from the running times, taking the specific flow rates into account. You can determine unknown pump feed rates using the calibration menus of the pumps or proportioning valves. Depending on the hoses and pumps used, known specific production rates can be entered directly into the calibration menu. The calibration and filling counter functions are the same for all pumps and proportioning valves. Calibration is described using the example “LEVELT-#”. Submenu Field Value Function, entry required Mode Calibrate −− Start calibration Totalize −− After completion of “Calibrate”, the system automatically switches to “Totalize” Reset −− Reset the filling counter to zero Display the volume of liquid being pumped for: LEVELT-# ml Level pump (typically a digital pump) FOAMT-# ml Antifoam pump (typically a digital pump) ACIDT-# ml Acid pump (typically a digital pump) BASET-# ml Base pump (typically a digital pump) SUBST-A1...C1 ml Substrate pump A to C (typically an analog pump) Flow Directly enter the specific feed rate of the pump or the flow rate of the proportioning valve, if known ml/min Operating Manual BIOSTAT ® B 115 Preparing and Running the Process Preparing for Pump Calibration Always use the same type of tubing with the same dimensions for calibrating and pumping the media. To calibrate, it is recommended that a suitable scale be used, as this is more accurate. Prior to calibration, the hose must first be filled. To do this, proceed as follows: ttInsert the hose into the pump. ttHang the hose end from the pump inlet into a water-filled beaker. ttHang the hose from the pump outlet into a measuring cup that you can use to measure the feed rate. The analog pumps (SUBST-A1, B1, C1) must be controlled by the Subs controller. ttActivate the pump with “on”. yyLeave the pump activated until the hose is completely filled. ttDeactivate the pump. Perform calibration ttPress the touch button of the pump to be calibrated (“LEVELT-#”). Before the first calibration, the mode “Off” is displayed. ttPress the touch button “Calibrate” in the submenu “LEVELT-# Mode”. yyThe “START calibration with OK” menu is displayed. ttStart the pump calibration with “OK”. yyThe “STOP calibration with OK” menu is displayed. The pump pumps the medium. ttWait until a sufficient volume has been pumped. ttStop the calibration by pressing “ok”. ttRead the feed volume off the measuring beaker. ttIn the input window “LEVELT-#: Volume”, enter the feed volume and confirm with “OK”. yyThe filling counter is reset and the display mode changes to “Totalize”. The DCU system calculates the pumping rate automatically from the internally registered pump run time and the feed volume calculated. The pumping rate is displayed in the “Calibration LEVELT-#” submenu in the “Flow” field. Direct Entry of the Feed Rate If the flow rate as a function of the hoses and pumps used is known, the feed rate can also be entered directly. 116 Operating Manual BIOSTAT ® B Preparing and Running the Process ttPress the touch button “Flow” in the submenu “Calibration LEVELT-#”. ttIn the input window “LEVELT-#: Flow”, enter the feed rate and confirm with “OK”. Resetting the Filling Counter ttPress the touch button “Reset” in the submenu “LEVELT-# Mode”. yyThe filling counter is reset. Activating the Filling Counter The filling counter is reset after calibration. The filling counter is automatically activated after switching on the pump and | or the associated controller. 8.10.8 Scale Taring Function The weight of culture vessels, feed bottles or media or harvest containers can be measured on weighing platforms or pressure gauges. Any tare corrections required, e.g. after re-equipping the culture vessel or refilling a holding bottle, can be made during running operations. To do so, determine the net weight and adapt the tare weight to the change in weight caused by the changed equipment. Submenu “Calibration VWEIGHT-#“ Field Value Mode Function, entry required Display of the active operating mode −− Measure [Measure] Indicates that the measurement is active in process −− Tare [Tare] Zero taring −− Hold [Hold] Indicates that the measurement in process has been paused Display net weight (WEIGHT = gross tare): FWEIGHT-A#/B# g/kg Substrate or harvest container weight (UniVessel® Glass / UniVessel® SU) VWEIGHT-B# g/kg Culture vessel weight (UniVessel® Glass / UniVessel® SU) RWEIGHT-# g/kg BIOSTAT ® RM Rocker weight Tare g/kg Tare weight display or entry (DCU system) Gross g/kg Gross weight display (measured value of the scale) Operating Manual BIOSTAT ® B 117 Preparing and Running the Process Zero-taring on the example scale/culture vessel ttPress the touch button “tare” for zero-taring in the submenu “VWEIGHT-# Mode”. yyThe display “Tare” (measured value in the DCU system) is set to zero. The gross weight “Gross” (measured value of the scale) remains unchanged. Tare Correction during Running Operations ttPress the touch button “Hold” in the submenu “VWEIGHT-# Mode”. yyThe display “Tare” is frozen. ttMake changes to the equipment: for example, re-equipping the culture vessel or refilling a holding bottle. ttStop the tare correction by pressing “ok”. yyDespite a change in the gross weight, the display “Tare” remains. Changing the Tare Weight via Direct Entry ttPress the touch button “Tare” in the submenu “VWEIGHT-# Mode”. ttIn the input window “VWEIGHT-#: Tare”, input the new weight via the screen keyboard. ttConfirm the weight change with “OK”. yyThe display “Tare” (measured value in the DCU system) is set to the entered value. The gross weight “Gross” (measured value of the scale) remains unchanged. 118 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11 “Controller” Menu 8.11.1 Functional Principle and Equipment The control loops in the DCU system work as PID controllers, setpoint generators or two-point controllers and are adapted to their control circuits. PID controllers can be parameterized to match the control task. The controller outputs control their control elements either continuously or using pulse-width modulation. Single-sided and split range control are implemented. Which controllers are implemented in a DCU system depends e.g. on the terminal unit (e.g. bioreactor). Controllers can be customized. Available controllers in the DCU software include: Controllers Function “TEMP” temperature controller PID cascade controller with pulse-width modulated split range outputs for the control of the heater and | or valve on the cooling water intake with the measured value of the culture vessel temperature as controlling value “JTEMP” double wall temperature controller Slave controller for temperature control: −− with TEMP controller “off”, possible as setpoint generator for heating | cooling Speed regulation (STIRR) Setpoint generator for external motor controller controlling the agitator motor Regulator “ROCKS” (BIOSTAT ® RM Rocker) Setpoint controller for shaker – Controller for the shaker speed ROCKS Regulator “ANGLE” (BIOSTAT ® RM Rocker) Setpoint controller for shaker - Controller for rocker angle ANGLE “pH” controller PID controller with pulse-width modulated split-range outputs: −− Controls the acid pump and | or the CO2 supply and the alkaline pump DO controller “pO2” PID cascade controller for controlling up to 4 slave controllers: −− Gas filler controller Air, O2 or N2 −− Gas flow controller −− Speed regulator −− Controller for substrate supply Gas filler controller: Module “Additive Flow 4-Gas” BIOSTAT ® RM Rocker −− AIROV-# Slave controller or setpoint generator for gas proportioning valves, pulsed feed: −− Air for the headspace (overlay) −− O2OV-# −− O2 for headspace aeration (overlay) −− N2OV-# −− N2 for headspace aeration (overlay) −− CO2OV-# −− CO2 for headspace aeration (overlay) Module “Additive Flow 4-Gas” UniVessel® Glass | UniVessel® SU −− AIROV-#, AIRSP-# Slave controller or setpoint generator for gas proportioning valves, pulsed feed: −− Air for headspace aeration (overlay) and medium aeration (sparger) −− O2OV-# −− O2 for headspace aeration (overlay) −− N2OV-# −− N2 for headspace aeration (overlay) −− CO2OV-# −− CO2 for headspace aeration (overlay) Operating Manual BIOSTAT ® B 119 Preparing and Running the Process Controllers Function Module “Additive Flow 5-Gas” UniVessel® Glass / UniVessel® SU −− AIROV-#, AIRSP-# Slave controller or setpoint generator for gas proportioning valves, pulsed feed: −− Air for headspace aeration (Overlay) and medium aeration (Sparger) −− O2OV-#, O2SP-# −− O2 for headspace aeration (Overlay) and medium aeration (Sparger) −− N2OV-# −− N2 for headspace aeration (Overlay) −− CO2OV-# −− CO2 for headspace aeration (Overlay) Gas flow controller Slave controller or setpoint generator for mass flow controller −− Each of the gases listed in each segment Antifoam controller “FOAM” Pulse pause controller for introduction of antifoam agent “AFOAM” Level controller “Level-#” Pulse pause controller for level controller “LEVEL” Substrate controller “SUBS-A#, -B#, -C#” Target value generator for filling pumps Gravimetric level controller “VWEIGHT”, “RWEIGHT” PID controller with pulse-width modulated outlet for pump (harvest and filling operation); works with the weight of the culture vessel “VWEIGHT“, “RWEIGHT” as master variable Gravimetric filling control “FLOW” Setpoint generator for internal or external filling pump; works with the weight of the substrate vessels “BWEIGHT”, “FWEIGHT” as master variable: −− Only controlled units with associated weight measurement Controller Mode You can switch between controller operating modes: off Controller switched off with defined output Auto Controller activated Manual Manual access to control element profile Selection of previously defined profile. If no profile is defined, the controller automatically switches to “auto” mode 8.11.2 Controller Selection You can access the operator screens on the controllers of a configuration in various ways: −− For the controllers most frequently used, from the “Main” menu or from the “Controller” menu, both in the “All” view. −− For other frequently used controllers, from the “Main” menu in the detail views for “unit 1”, etc. −− For all controllers, from the “Controller” menu in the detail views for “unit 1”, etc. 120 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.3 General Controller Operation For the most part, operation of the controller is uniform. It comprises setting the setpoints and alarm limits and the selection of the control operating mode. If a controller can control more than one output, the controller output is assigned by means of the parameterization functions accessible with a password. This also applies to controller settings not required during routine operation. Controller Operator Screen Fig. 8-24: Example, selection of the temperature controller TEMP-1. Field Display Function, entry required Controller mode Selection Input of the controller mode Mode off Controller and slave controller switched off Auto Controller switched on, slave controller in “Cascade” operating mode Manual Manual access to control output profile Selection of previously defined profile. If no profile is defined, the controller automatically switches to “auto” mode Actual value TEMP-1 Nominal value of the process value in its physical unit, e.g. degC for temperature, rpm for revolutions per minute, pH for pH value, etc. Target Setpoint Setpoint of process value in the physical unit, e.g. °C for temperature Controller output Out Display of controller output in % Alarm parameters Alarm parameters Enter alarm limits (high limit, low limit) and alarm status (enabled, disabled) Profile Parameter Profile Param. Possibility to input a time-dependent setpoint profile (max. 20 spikes) Function key Function key Access to controller parameters (with password) for cascade controllers: Selection of the slave controller (see Chapter “8.11.5 General Controller Parameterization“, Page 123) ok Confirm entries with “Ok” Operating Manual BIOSTAT ® B 121 Preparing and Running the Process 8.11.4 Setpoint Profile Controller Profile The “Profile Parameter” function can be used to navigate to the setpoints of the individual controllers. Time-based setpoint profiles can be set up. Up to 20 steps can be configured. Any pre-installed DCU system can be additionally retrofitted with control functions by changing the configuration. Moreover, control blocks available in the software can also be used to configure special controllers. These configuration changes may only be carried out by Sartorius Service. off Controller switched off with defined output auto Controller activated Manual Manual access to control element profile Selection of previously defined profile. If no profile is defined, the controller automatically switches to “auto” mode Most control loops can be operated with time-dependent setpoint profiles (Control Loop Profiles). The profile is input into a table using the operator terminal. Jumps and ramps are possible in the profile; however, a profile can have a maximum of 20 spikes. You can start and stop profiles at any time. The elapsed time is displayed for started profiles. Access Operator Screens ttCall up the appropriate controller. ttCall up the following operator screen using the button “Profile Param.” Fig. 8-25: Operator screen using the example of the AIRSP Profile Field Value Add Mode Setpoint 122 Operating Manual BIOSTAT ® B Function, entry required Adding a profile spike off Setpoint profile not active profile Setpoint profile has been started and is being processed [PV] Display of the current controller setpoint in the physical units of the process value, e.g. °C for temperature Preparing and Running the Process Field Value Function, entry required Elapsed Time h:m:s Display of the elapsed time since the profile start in [hours:minutes:seconds] Graphical display of the elapsed time on the profile screen No. 1-20 Number of the profile spike Time h:m:s Input of the time for the profile spike Setpoint [PV] Input of the setpoint for the profile spike in the physical units of the process value, e.g. °C for temperature Del Deleting a profile spike Operation We recommend that you create a sketch with spikes and associated setpoints for your profile. The time and setpoints to be programmed can be read directly from the spikes entered into the sketch. A profile must contain at least one profile spike with a non-zero time in order to be started. Special notes −− When starting the setpoint profile, the controller mode will automatically be switched to “profile” in the “Controller” menu. −− If you do not input the time “00:00 h:m” for the first spike, the system will use the current setpoint as the starting time. −− In the case of a setpoint jump, the same time is programmable for both spikes. −− When starting a “pO2” profile, whichever profile for “STIRR”, “AIR”, or “PRESS” has been started will be automatically stopped and switched to “cascade” mode. 8.11.5 General Controller Parameterization For optimum adaptation of the controller to each control segment, the controller parameters can be changed using the parameterization screens: Controller parameterization using the example of a TEMP controller Field Display Function, entry required MIN, MAX Value in % Minimum and maximum output limit for the controller output DEADB Value example: °C Dead zone setting (PID controller only) XP, TI, TD Value in %, sec PID parameters (PID controller only) Parameterization screens are accessible after selecting in the controller operator screen and password entry. DCU systems are configured at the factory with parameters that ensure the stable operation of the control circuits in the bioreactor. Factory-configured parameters can be found in the customer-specific configuration documentation. After entering the password, you have access to the parameterization screen to set PID parameters, output limits, and if necessary a dead zone. In “Remote” operation, the host PC defines the setpoints and operating modes. Operating Manual BIOSTAT ® B 123 Preparing and Running the Process As a general rule, it is not required to change the control parameters. The exceptions are controlled loop paths, the behavior of which is strongly influenced by the process, e.g. pH and DO (pO2) control loops. 8.11.5.1 Output Limits You can limit the controller output for the target value generator and PID controller downwards (MIN) and upwards (MAX). In this way you can avoid unintentional, extreme control element controls or limit the target value range for the slave controller during cascade control. −− The limits are entered in the MIN (minimum limit) and MAX (maximum limit) fields. The setting is made relative to the overall controller range in %. −− The following limits apply to the full modulation of the controller output: −− One-sided controller output: MIN = 0%, MAX = 100% −− Split-range controller output: MIN = -100 %, MAX = 100% 8.11.5.2 Dead Zone A dead zone can be set up for PID controllers. If the control tolerance remains within this dead zone, the controller output maintains a constant value and | or is set to zero (pH controller). If the nominal values fluctuate stochastically, the dead zone enables more stable control operations with minimized control element movements. For controls with split-range outputs, this prevents oscillation of the controller output (e.g. constantly changing acid/alkaline proportioning on the pH controller). −− The dead zone is displayed in the DEADB field or configured in the associated submenu. Example for pH controller: Set dead zone ± 0.1 pH Set target value 6.0 pH −− In that case, the control loop is inactivated at nominal values between 5.9 pH and 6.1 pH. Fig. 8-26: Submenu for controller parameterization using the example of the pH controller 124 Operating Manual BIOSTAT ® B 8.11.5.3 Controller Parameterization Menu Screen Field Value Function, entry required MIN % Minimum output limitation, limit value for switch to upstream slave controller MAX % Maximum output limitation, limit value for switch to downstream slave controller DEADB pH Dead zone in the unit of the process value XP % P share (proportional range); signal amplification of the control response proportional to the input signal TI sec Integral portion; time function. With a higher I portion control will react more slowly (and vice versa) TD sec Differential portion: Damping, greater D portion, damps the controller response (and vice versa) OUT Controller output 1 (only in configurations in which the switching of the output is provided) OUT2 Controller output 2 (only in configurations in which the switching of the output is provided) 8.11.5.4 Preparing and Running the Process PID Parameters The PID controllers can be optimized using the PID parameters XP, TI and TD. The implemented digital controllers run according to the position control algorithm. They allow structural toggles (P, PI, PD, PID) and changing the parameters during ongoing operations. −− The controller structure can be configured by setting individual PID parameters to zero: P controller: Æ TI = 0, TD = 0 PI controller: Æ TD = 0 PD controller: Æ TI = 0 PID controller: All PID parameters defined 8.11.5.5 PID Controller Optimization Knowledge about control technology are prerequisite in order to be able to optimally tune a PID controller to a controlled loop path; otherwise empirically tested tuning methods (e.g. Ziegler Nichols) can be found in the pertinent literature. As a general guideline: −− Only switch the D portion (TD) if the nominal values are relatively stable. For stochastically variable actual values, the D portion makes fast, large changes to the output. This causes unstable control. −− As a rule, the TI : TD ratio should be around 4 : 1. −− Periodic oscillations in the control circuit can be counteracted by increasing XP and/or TI / TD. −− If the adjustments are too slow after setpoint jumps and/or in the case of nominal value drift, you can lower XP and/or TI / TD. 8.11.6 Guide and Slave Controller (TEMP, JTEMP) Temperature Measurement Temperature measurement with guide and slave controller is only possible in conjunction with double wall vessels. The temperature control works like a cascaded regulation. The TEMP controller uses the temperature measured in the culture vessel as a master value and acts on the mode of the JTEMP slave controller. JTEMP’s output controls the assigned control elements through pulse-duration modulated or constant outputs in the split-range operation. Associated control elements can include: −− Cooling water supply valves (double wall culture vessel, heating | cooling jacket, tempering coil) When the value approaches the setpoint, the guide controller switches the controller structure from “PD” (starting condition) to “PID”, preventing overshoot. In the temperature control circuits, like on bioreactors, a digital output also switches off the circulating pump as well as the heating protection when the temperature controller is switched off. Operating Manual BIOSTAT ® B 125 Preparing and Running the Process Master Controller TEMP Operator Screen Fig. 8-27: Master Controller TEMP-1 Operator Screen Refer to Chapter “8.11.3 General Controller Operation“for notes on the fields, entered values and entries. Operation Observe the maximum permissible temperatures of the component groups and fixtures your bioreactor is equipped with. Culture vessel Maximum permissible temperatures for “TEMP” master controller UniVessel® Glass jacketed (thermostat) 80 °C UniVessel® Glass single-walled (heating jacket) 60 °C UniVessel® SU heating jacket 50 °C UniVessel® SU (heating | cooling jacket) 50 °C Culture bag (BIOSTAT ® RM 20 | 50 Rocker) Heating mat Tempering coil Culture bag (BIOSTAT ® RM 200 Rocker) Heating plate Heating | Cooling plate 40 °C 40 °C 40 °C 40 °C The temperature cascade controller is operated from the master controller. You can only change setpoints and operating modes on the master controller “TEMP-#”. All operations of the slave controller “JTEMP-#” are triggered automatically. −− For routine operation, you only need to configure the master controller “TEMP-#” (setpoint, mode and alarm limits). −− Direct configuration for heating and cooling can be done on the slave controller “JTEMP-#” when the master controller “TEMP-#” is turned off (“manual” mode). 126 Operating Manual BIOSTAT ® B Preparing and Running the Process Special notes −− In “auto” mode of the “TEMP-#” master controller, the “JTEMP-#” slave controller automatically switches to “cascade” mode. In the “off” setting of the master controller, the slave controller is also automatically “off”. −− On certain systems, a setpoint limit must be parameterized for the slave controller using the “MAX” output limit of the master controller. −− Example, UniVessel® Glass double wall: max. out = 62 % for max. temperature = 80 °C −− The output limits required for safe operation are preset in the system configuration. User-defined output limits that deviate from this must be reset after a system reset. 8.11.7 Temperature Measurement Without Slave Controller (TEMP) Fig. 8-28: Operator screen when called from “Controller – #” Operating Manual BIOSTAT ® B 127 Preparing and Running the Process 8.11.8 Speed Regulation (STIRR) Function The speed controller works like a target value generator for an external motor controller, which controls the speed of the stirrer motor. In addition to its function as a single controller, the speed controller can also be used as a slave controller in pO2 control. Controller operator screen Operator entries, output of the analog setpoint signal for the motor controller and the display of the speed signal from the controller are all done on the controller operator screen. Field Display Function, entry required STIRR-1 rpm Display of the current stirrer speed SetPoint rpm Configuration of the target speed in the mode “auto” Out % Display of the speed limitation (MIN / MAX) and setting configuration of the target speed in the operating mode “manual” Alarm Param. Entry of the alarm limits (Highlimit, Lowlimit) and switching the alarm on/off Profile Param. Input of a time-dependent setpoint profile (max. 20 spikes) Function key Entry of the speed limitation (MIN/MAX) Operation High speeds can damage vessel attachments. Sometimes, only a certain maximum stirrer speed is allowed depending on the vessel type, volume and equipment. Higher stirrer speeds can damage vessel attachments, e.g. a tubing aeration system. Vessels can become unstable and move across the surface of the floor. Note the maximum permissible stirrer speed for your bioreactor: Culture vessel Maximum stirrer speeds BIOSTAT ® B UniVessel® Glass, 1 L, 2 L 2000 rpm UniVessel® Glass, 5 L 1500 rpm UniVessel® Glass, 10 L 800 rpm UniVessel® SU, 2 L 400 rpm More information can be found in the folder [Æ “Technical Documentation”]. When inputting MIN | MAX output limits or making direct entries into the OUT field, the permissible speed controller range must be considered. ttSet the desired stirrer speed via “Setpoint”. Speed limitation When selecting the speed control MIN / MAX 0 - 100% for speed range 0 - 2000 rpm and 1200 rpm as a permissible max. stirrer speed, a value of “OUT”: MAX 60% must be configured. If the MIN | MAX setting is changed after a system reset, you must reset them new limits to the range permissible for the bioreactor! 128 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.9 Antifoam Controller “FOAM” Function The autoclavable foam sensor is installed in the culture vessel. The sensor is adjustable in height, so that the sensor tip can be adjusted to the maximum level of the medium. A threshold signal generated by the foam sensor and amplified by a measurement amplifier serves as the input signal of the foam regulator “Controller FOAM-#”. This is activated, as long as the sensor is in foam. The trigger sensitivity “Sensitivity” of the amplifier can be adjusted. The output of the foam controller modulates a correction medium pump and switches it on and off (Cycle / Pulse) periodically when a sensor signal is emitted. Controller operator screen Field Display Function, entry required Mode off Controller switched off auto Controller switched on Manual Manual activation of the controller output; pump runs permanently as a function of Cycle/Pulse Cycle hh:mm:ss Total cycle time in [hours: minutes: seconds] Pulse hh:mm:ss Pump runtime (dispensign time) in [hours: minutes: seconds] Sensitivity −− Low −− Medium Low −− Medium High −− High Trigger sensitivity for the foam sensor Alarms Param. Switching the alarm on/off Operation ttSet the cycle time “Cycle” and the pump runtime “Pulse” according to process requirements. ttConfigure the trigger sensitivity “Sensitivity” of the sensor. To prevent proportioning errors resulting from leakage currents and sensor growth, you should set the response sensitivity as low as possible. ttSwitch the mode to “auto”. In the operating mode “manual”, the pump periodically runs in continuous operation as a function of the settings “Cycle” and “Pulse”. Special notes −− The measurement amplifier is equipped with a response lag time mechanism (approx. 5 sec), that prevents activation after splashing liquid. −− Switching to the “auto” or “manual” mode automatically also activates the filling counter “FOAMT-#”. Operating Manual BIOSTAT ® B 129 Preparing and Running the Process 8.11.10 Level Control with Level Sensor (LEVEL) Function The autoclavable level sensor is installed in the culture vessel. The sensor is adjustable in height, so that the sensor tip can be adjusted to the maximum level of the medium. A threshold signal generated by the level sensor and amplified by a measurement amplifier serves as the input signal of the level regulator “Controller LEVEL-#”. It is active when the level of the medium rises to where it is in contact with the level sensor. The trigger sensitivity “Sensitivity” of the amplifier can be adjusted. The level controller is normally operated during the harvesting operation. By changing the direction of hoses and switching from “Pump” to “Feed”, the level controller can also be used during addition. The harvesting operation is described in the following. The output of the level controller controls a harvesting pump. The pump speed is constant. If the medium has not been in contact with the level sensor, the pump stops pumping after a specified period. If the level is controlled via the level sensor, an additional harvesting pipe should be installed. Controller operator screen Field Display Function, entry required Mode off Controller switched off auto Controller switched on Manual Manual activation of the controller output; pump runs permanently Pump Harvest Feed Pump during the harvesting operation Pump during addition Pulse hh:mm:ss Pump runtime (harvest time) in [hours: minutes: seconds] Sensitivity −− Low −− Medium Low −− Medium High −− High Trigger sensitivity for the sensor Alarms Param. Switching the alarm on/off Operation ttSelect the pump operation “Harvest”. ttSet the harvest time “Pulse” according to the process requirements. ttConfigure the trigger sensitivity “Sensitivity” of the sensor. To prevent proportioning errors resulting from leakage currents and sensor growth, you should set the response sensitivity as low as possible. ttSwitch the mode to “auto”. In the operating mode “manual”, the pump runs continuously. 130 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.11 Adjusting the “VWEIGHT” Gravimetric Level Controller With gravimetric leveling, it is possible to maintain a particular medium volume in the culture vessel. The pump speed is controlled automatically depending on the weight change in the culture vessel. When the medium volume is regulated, there is a difference between: −− Feed operation −− Harvest operation The controller is set to feed operation by default. Feed Operation When the weight of the culture vessel drops below the setpoint: It activates a variable speed pump. The substrate is added to the culture vessel until the setpoint is reached again. Harvest Operation When the weight of the culture vessel exceeds the setpoint: It activates a variable speed pump. The medium is harvested until the setpoint is reached again. 1 Procedure ttConfigure the controller according to requirements, for example: No. Parameter Value Description 1 Shows the operating mode selection window Operating mode Off Switch off gravimetric level control Auto Adjust the media volume to a weight using the “Setpoint” key Manual Adjust the feed/harvest pump to a constant pump speed using the “Out” key (pump runs permanently) Profile 6 5 7 4 3 Cascade Gravimetric level control is controlled by a master controller 2 8 9 Have the medium volume controlled by a set weight profile 2 Out % In “manual” mode only: Adjust the controller output (feed/harvest pump) to a constant pump speed 3 SetPoint kg Enter weight 4 Profile Param. 5 VWEIGHT 6 Alarm Param. Adjust alarm parameters 7 PUMP Shows the “pump” selection window for choosing the pump operating mode 8 Harvest Activate harvest operation 9 Feed Activate feed operation Create weight profile kg Displays the current weight: medium with culture vessel Operating Manual BIOSTAT ® B 131 Preparing and Running the Process 8.11.12 Gravimetric Filling Pump Controller “FLOW” Function The controller “FLOW-#” is a precise gravimetric filling pump controller. It is used a weighing system and an analog filling pump. Because the control algorithm in the DCU system works directly with the weight measured on the scale/balance, the gravimetric filling controller allows accurate proportioning over days and weeks. Controller operator screens Refer to Chapter “8.11.3 General Controller Operation“ for notes on the fields, entered values and entries. Operation Operation with supply container and filling controller: ttTare the scale to zero and place the vessel on the balance. [Æ Chapter “8.10.8 Scale Taring“, Page 117] ttSet the desired quantity to add via “Setpoint”. ttSwitch the mode of the filling pump controller to “auto”. A negative weight readout on the scale | balance or on the DCU indicates the feed volume. Special notes −− The feed volume of the filling pump has an important influence on the controlled loop path. The pump output must therefore be adjusted to the required flow [Work Min]; [Work Max] in the parameter menu. −− For accurate proportioning, the working range of the controller output (“Out”) must lie in the range from 5 to 90%. For that purpose, you can adapt the feed range of the pump to the working range of the controller. You can use hoses with a different diameter that offer the desired conveying range. 132 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.13 Filling Pump Controller (SUBS) Function To introduce nutrient solution, the filling pump controller can control an internal or external pump. The controller function works as a setpoint generator, handles control, and emits an analog setpoint signal for the pump. Controller operator screen Fig. 8-29: Fig. 17-32: Parameterization screen Fig. 8-30: Operator screen for the controller Refer to Section “8.11.3 General Controller Operation“ for notes on the fields, entered values and entries. Operation ttSet the desired quantity to add via “Setpoint”. ttSwitch the mode of the filling pump controller to “auto”. Special notes −− Matching connecting cables are available for certain pumps, like WM 120, WM 323. Ordering information is available on request. −− Pumps from other manufacturers can be connected if they have an external setpoint input from 0 to 10 V. Operating Manual BIOSTAT ® B 133 Preparing and Running the Process 8.11.14 Gas Controller (Gas Filling Controller / Gas Flow Controller) Gas controllers control gas supply for the corresponding gas segment, e.g. “AirOV-#”, “AirSp-#”, “O2SP-#”, “N2SP-#”, “CO2OV-#”, or “CO2Sp-#” and introduce gases into the aeration segments “Overlay” or “Sparger”. The following types of gas controllers can be used: −− Gas filling controllers (solenoid valves) −− Gas flow controllers (massflow controller) The massflow controller makes it possible to aerate the culture vessel with a continually changing gas flow. The controllers normally work as slave controllers for the DO or pH control loop. With the DO controller turned off, they can be used as setpoint generators. Controller operator screen Field Display Function, entry required Mode off Controller switched off, output on stand-by Manual Manual access to control output auto Automatic operation, control with a predefined target value AIRSP-1 rpm Display of current total gas flow SetPoint rpm Configuration of a set point for the gas flow controller % Configuration of a set point for the gas filling controller Out % Alarm parameter Entry of the alarm limits (Highlimit, Lowlimit) and switching the alarm on/off Profiles Param. Input of a time-dependent setpoint profile (max. 20 spikes) Function key Entry of the lower (MIN) and upper (MAX) output limit, setting range 0 to 100% of the control range and other control parameters To operate the gas controller as a setpoint generator, the master controller must be switched off. Check its operating mode in the “Main” or “Controller” menu, and switch the mode of the master controller to “off” if it is active. −− Select the “Main” or “Controller” view in the detail view “1”... in which you want to set up the gas filling controller. −− Select the function key with the current display of the setpoint “0.01 lpm”. Enter the setpoint in the window with the numerical keypad. −− Set the alarm limits, if needed, and activate alarm monitoring. −− Select the function key for the operating mode and select the “auto” operating mode. −− Press “OK” to activate the controller. Special notes −− Select the setpoint of 100% to configure the flow rate on the variable area flow meter (rotameter) and calibrate the filling counter (if the calibration function is included in the configuration). Oxygen then flows continuously in the air supply. −− For manual gas supply, select the desired setpoint in the range 0..100 %. −− When activating the “auto” mode on the master controller, the gas filling controller automatically switches to “cascade” mode. In that event, settings in the gas filling controller are not possible or will be ignored. 134 Operating Manual BIOSTAT ® B Preparing and Running the Process Follow the instructions regarding the “Parameter Settings in the System” in the “Configuration Documentation”. −− MIN/MAX output limits are entered in % of the control range of the gas feed. When entering values directly in the OUT field, take the measurement range for the aeration rate into consideration. −− If the gas flow controller is a slave controlled in the DO control cascade, enter the MIN | MAX values in the “DO controller” parameterization menu. The settings will then act as a switching criterion for cascade control. −− Switching off the GASFL flow controller (select “off” and after an emergency shut off due to overpressure) closes the control valve in the mass flow controller. Culture bags are only pressure-resistant to a certain extent and may burst if subjected to excess pressure. The pressure is monitored in the gas supply. When the pressure limit is exceeded (435.11 mpsig for standard culture bags), e.g. by air blockage, the flow controller is shut off. The gas supply remains closed as long as the pressure is too high (> 435.11 mpsig). Note the specifications for the measurement|control range of the aeration rates of the bioreactor. With a bioreactor operated with overpressure, the counter pressure might cause the maximum aeration rate not to be reached. 8.11.15 pH Controller 8.11.15.1 Function The pH control normally works with PID control characteristics. It controls correction medium pumps for acids and bases and/or proportioning valves or mass flow controllers for CO2 in split-range mode using pulse-width modulated outputs. This enables bilateral control. −− pH control with base is configured by default. −− pH control with acid and CO2 depends on the configuration. −− The negative controller output acts on the acid pump (or the CO2 feed) and the positive output on the base pump. −− The pH controller does not activate the control signals until the control deviation is located outside of a configurable dead zone. This prevents any unnecessary acid/ alkaline proportioning. 8.11.15.2 Controller Operator Screen Refer to Chapter “8.11.3 General Controller Operation“ for notes on the fields, entered values and entries. Operating Manual BIOSTAT ® B 135 Preparing and Running the Process 8.11.15.3 Parameter Settings A DEADB dead zone can be entered in the pH controller parameterization screen. The controller remains inactive as long as the measured value remains within the dead zone around the setpoint. Example: Set dead zone: ± 0.05 pH Set target value: 6.0 pH yyIn that case, the control loop is inactivated at nominal values between 5.95 pH and 6.05 pH. 8.11.15.4 Change the process value source Multiple, different pH sensors (conventional, optical) can be installed on the device. One of these sensors can be selected as the process value source for pH regulation. Procedure ttPress the “pH-A1” key (1) in the advanced configuration window. yyThe “Input pH-1” selection window appears. ttPress the “pH-##” key (2) to select the desired process value source: Description/process pH sensor type value source pH-A1 conventional pH sensor pH-B1 UniVessel® SU optical pH sensor ttConfirm the input in the advanced configuration window with the “OK” key. 1 2 8.11.15.5 Acid, Base, and CO2 Supply-driven pH Control Acid | Base Supply-driven Control The pH controller output “+Out” normally controls the acid pump with a positive output signal (0 to +100 %). Correspondingly, the controller output “-Out” controls the base pump with a negative output signal (0 … -100%) and adds base. To disable the addition of acid or base, the regulator value of 100% (+ | -) must be set to 0%. 136 Operating Manual BIOSTAT ® B Preparing and Running the Process For special configurations, the acid or alkaline pump can be assigned to substrate controllers if they are not needed for pH regulation. To do this, “+Out” must be set to “None” (instead of “Acid” or “CO2”) and “-Out” must also be set to “None”. CO2 Supply-driven Control For bioreactors for cell culture, a CO2 valve or a CO2 massflow controller can work as control element of the pH control in place of the acid pump. Culture bags are only pressure-resistant to a certain extent and may burst if subjected to excess pressure. The pressure is monitored in the gas supply. When the pressure limit is exceeded (435.11 mpsig for standard culture bags), e.g. by air blockage of the exhaust filter, the flow controller is shut off. The gas supply remains closed as long as the pressure is too high (> 435.11 mpsig). In configurations for cell culture, the output “+Out” can be connected to the CO2 feed. After switching to “CO2”, the output controls the CO2 valve (or the massflow controller of the CO2 segment) to introduce CO2 into the culture vessel. Special Notes −− When the “auto” or “manual” operating modes are activated, the “ACIDT-#” | “CO2T-#” and “BASET-#” filling counters automatically switch to the “Totalize” operating mode. 8.11.16 DO Control Methods The DCU system features various methods of DO control. Which of them is possible, required or sensible for the controlled terminal unit depends on the configuration or process. −− When aerating with air, either the oxygen portion can be reduced by adding nitrogen or the air can be enriched with oxygen. −− The mixture can be influenced e.g. by controlling the agitator speed. −− Cell growth can be influenced by adding or reducing substrate. The DO control (pO2) works like a cascaded regulation. The output of the DO controller (master controller) modulates the target value input of the slave controller, which then acts on the control element (e.g. the valves or mass flow controllers for N2 or O2 or the stirrer). The following control strategies are possible: −− 1-stage control cascade, i.e. the pO2 control only affects one of the available setting variables. −− Up to 4-stage simultaneous control cascade, during which the pO2 control is affected up to 4 setting variables according to their priority. A range (MIN | MAX) can be defined in the DO controller, in which the DO controller defines the target value for each slave controller. In multi-stage cascaded regulation, the output of the DO controller modulates the slave controller after sequential switch-on as follows: −− The pO2 controller acts on the slave controller with the priority 1 (Cascade 1) and defines its setpoint. The slave controller 2 receives the target value defined in the DO controller as “MIN”. Operating Manual BIOSTAT ® B 137 Preparing and Running the Process −− When the preset target value of the 1st slave controller (Cascade 1) reaches its maximum, the output of the pO2 controller switches to the target value input of the 2nd slave controller (Cascade 2) after an adjustable delay time “Hyst.” and defines the following target values: −− Slave controller (cascade) 1: with defined maximum −− Slave controller (Cascade) 2: controlled output of the DO controller −− This sequence continues for the other control elements according to the predefined priority “Cascade #”. −− If the need for oxygen drops, the controllers are reset to in the reverse order. In this way, the control can regulate the in-process DO value, even if there are considerable fluctuations in the need for oxygen in the culture. In order to still be able to additionally optimally adapt the control to the behavior of the controlled loop path, the PID parameters of the slave controller are parameterizable independently of one another. 8.11.16.1 Adjusting the DO (pO2) Sensor Process Value Source The DO (pO2) controller must be set to the process value to be used for control with reference to the DO (pO2) sensor used. Types of DO (pO2) Sensors Description Type pO2-A “Unit #” conventional DO (pO2) sensor (amperometric or optical, e.g., Visiferm, manufactured by Hamilton) pO2-B “Unit #” optical DO (pO2) sensor UniVessel® SU ttOn the “Controller pO2-##” controller screen, press the “pO2-##” key. yyThe “Input pO2-##” selection window appears. ttSelect the process value source by pressing the corresponding “pO2-##” key. ttConfirm the input on the controller screen by pressing “ok.” 138 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.16.2 pO2 Controller CASCADE (cascade controller) Operator Screen Fig. 8-31: pO2 cascade controller menu on the “Controller – All” operator screen Refer to Chapter “8.11.3 General Controller Operation“ for notes on the fields, entered values and entries. The operator screen also include the following input fields: Field Value Function, display, entry required Setpoint % sat Preset Target Value in the Master Controller Setpoint Cascaded Controller Preset setpoint for the slave controller in the cascaded regulation, in the sequence of the priority predefined in the parameterization screen: Mode off Selected slave controllers will automatically be switched to “off” auto Selected slave controllers will automatically be switched to “cascade” mode profile With the profile, selected slave controllers will automatically be switched to “cascade” mode Alarms param. −− Input of the limit values “High” “Low” −− Input of the time delay −− Enable/Disable Alarm Profile Param. Inputting the profile parameter Sub-menu – parameterization screens Operating Manual BIOSTAT ® B 139 Preparing and Running the Process DO Cascade Controller Parameterization Screen Fig. 8-32: Example: Configuration of the operator screen 140 Operating Manual BIOSTAT ® B Field Value Function, display, entry required DEADB % Entry of the deadband Cascade # [Controllers] Slave controller with the relevant parameters MIN % Minimum output limit, corresponding to the minimum target value for the slave controller. MAX % Maximum output limit, corresponding to the maximum target value for the slave controller. XP % P share (proportional range); signal amplification of the control response proportional to the input signal TI sec Integral portion; time function. With a higher I portion control will react more slowly (and vice versa) TD sec Differential portion; damping, greater D portion, damps the controller response (and vice versa) End Mode off, auto Mode for slave controller when the master controller is “off” or “disabled”. Hyst. m:s Lag time for switching between the slave controllers. Mode off Selected slave controllers will automatically be switched to “off” auto Selected slave controllers will automatically be switched to “cascade” mode profile With the profile, selected slave controllers will automatically be switched to “cascade” mode Preparing and Running the Process Operation of the Multi-stage Cascade Controller ttSelect the slave controller according to the desired priority in the “Cascade Parameter pO2-#” submenu. ttSet each minimum and maximum controller setpoint limit for the selected slave controller using the output limits MIN or MAX in the parameterization image of the pO2 controller. ttWhen switching on the DO (pO2) controller, the slave controller modulated by the DO (pO2) controller is displayed as “active”. Special notes −− In modes “auto” and “profile” of the DO controller, the selected slave controllers are automatically switched to “cascade” mode. −− In the “off” mode of the pO2 controller, the selected slave controllers remain in the cascade last reached and may need to be switched off individually. −− Switching from slave controller 1 to the downstream controller and vice versa is not done until the respective output limit for the time span defined in the “Hyst.” field of the parameterization screen has been over or undershot. After this time has elapsed, check the switch conditions once again and only switched back if they have been met. −− An inverted control direction for slave controllers, such as the substrate controller, can be achieved by inverting the setpoint limit (MIN > MAX). −− The DO master controller always uses the MIN | MAX limits of the respective slave controller as the working range. −− The difference between MIN anx MAX must always be more than 2% of the specific measurement range. Operating Manual BIOSTAT ® B 141 Preparing and Running the Process 8.11.16.3 DO (pO2) Controller ADVANCED (Polygon Controller) The advanced DO (pO2) controller monitors and regulates the DO (pO2) in the bioreactor or the controlled end device for which the DCU system was designed. The controller acts as the master controller during DO (pO2) regulation. It acts on a configurable selection of slave controllers for the intake of media or to control actuators that influence the DO (pO2) in the process. Examples of such media include gases like N2, air, O2, or nutrient solutions. The DO (pO2) value measured in the process depends on the media introduced, the oxygen consumed during cell growth and cell metabolism, and material distribution from mixing. The master controller works as a PID controller with configurable control behavior. It uses the DO (pO2) measured at a measurement point (up to two measurement points can be selected) as the actual value. In case of deviation from the setpoint, the master controller sends an output signal to the slave controllers. Due to the variety of possible slave controllers, the output signal is relative to the control range 0 to 100%. One configuration can include up to six slave controllers, of which five can be selected simultaneously for the polygon controller. They control their actuators using analog or digital output signals. Each slave controller can be assigned up to five setpoints in the physical units of the set value, dependent on the output “Out” of the master controller. The controller screen shows this graphically as a polygonal curve above the output “Out.” In comparison with conventional DO (pO2) cascade controllers, the advanced DO (pO2) polygon controller supports the parallel operation of the slave controllers, that is, all actuators are controlled simultaneously. In combination with the determination of multiple setpoints dependent on the “Out” of the master controller, this results in an easy-to-understand and convenient-to-operate DO (pO2) regulation. Screen Fig. 8-33: DO (pO2) controller menu on the “Controller – All” screen 142 Operating Manual BIOSTAT ® B Preparing and Running the Process Settings for the Advanced DO (pO2) Controller Operating Display for the Master Controller Field Value Function, display, input required Mode Off Controller switched off, output on stand-by Auto Controller active, controls the actuator if necessary Manual DO (pO2) Manual access to controller output Display of the DO (pO2) Setpoint % Setpoint; in % relative to the control range 0 to 100% Out % Current controller output; in % relative to the control range 0 to 100% Access to the parameterization menu with standard password “Cascade Param.” Alarm Param. Access to the selection menu for the slave controllers, via standard password Alarm Parameter Input of the alarm limits (high limit, low limit) and Switch the alarm on/off Operating Menus for Configuring Slave Controllers Field Value Function, display, input required N2-SP1 Tag Slave controller assigned to this channel N2, O2, AIR, etc. Tag Media feed (gas, substrate) or function (e.g., speed controller) SP etc. Tag Feed to the culture vessel, e.g., sparger or overlay 1, 2, etc. # The unit assigned to the controller output, e.g., culture vessel 1, 2 End mode “Off” “Auto” Mode for slave controllers when the master controller is “off” or “disabled”; mode restored after emergency shut off or power-on Mode “Disable” “Enable” Activate or deactivate the slave controller for DO (pO2) regulation (only available when the master controller is in operating state “off” or “disabled”) Example: Input (modification) of the DO (pO2) setpoint Since the selection of slave controller can be changed according to process requirements, the setpoint of the DO (pO2) controller output is set in % relative to the control range. The slave controllers control their actuators with setpoints in their physical units. ttPress “pO2” in the “Controller” menu. ttPress “Setpoint.” ttUse the numeric keypad to enter the set value. Confirm with “OK.” ttPress the function key of the slave controller to be configured, e.g., “N2-SP1.” Enter up to five setpoints that depend on the “Out” output of the master controller. The settings are graphically displayed with a polygonal curve. ttActivate the DO (pO2) controller by switching to the “auto” mode and confirming with “OK.” Operating Manual BIOSTAT ® B 143 Preparing and Running the Process Parameterization of the DO (pO2) Master Controller Parameters Unit Value Function, display, input required MIN % 0 … 100 Minimum output, within 0 ... 100% of the control range MAX % 0 … 100 Maximum output, within 0 ... 100% of the control range DEADB % 0.5 Dead zone; controller output remains unchanged as long as DO (pO2) deviates from the setpoint by less than DEADB XP % 0.1 ... 1000 P share (proportional range); signal amplification of the control response proportional to the input signal; in % of the measurement range TI s 1 … 10,000 Integral portion; time function of the control response. With a higher I portion, the controller will react more slowly (and vice versa). TD s 0 … 1000 Differential portion; damping of the controller. With a greater D portion, the controller response is reduced (and vice versa) Normally you only change the parameters “MIN,” “MAX,” and “DEADB.” ttIn the “Controller” menu, select “pO2” of the corresponding assembly to be configured and open the controller screen. ttPress the key. ttSelect the parameter to be set (“MIN,” “MAX,” or “DEADB”), enter the value and confirm with “OK.” Setting the “P,” “I,” or “D” Controller Parameters: The adaptation of PID controllers requires knowledge of control theory. The setting options listed here are rough guidelines. Only qualified personnel should carry out controller optimization. Depending on the process (e.g., stability of gas intake or actuator), it may be necessary to change the parameters “P,” “I,” or “D” to adapt control behavior. You can test the following changes: −− If the measured DO (pO2) value (process value) fluctuates around the setpoint and does not stabilize, you can reduce the “P” portion. −− If the actual value only approaches the setpoint very slowly or does not reach it, you can increase the “P” portion. −− With a low “I” portion, the controller will react more quickly; as the “D” portion falls, it will react more strongly to setpoint deviations. However, this can create a tendency for the controller to overshoot. −− By increasing the “I” portion, we make the controller react more slowly, and by increasing the “D” it will react more weakly to deviations in actual value. This will make the controller response (the control behavior) more sluggish. 144 Operating Manual BIOSTAT ® B Preparing and Running the Process Selecting and Configuring Cascade Parameters Fig. 8-34: Selecting cascade parameters Field Value Cascade # Function, display, input required Slave controller to be assigned to the position “Cascade #”: −− Up to six slave controllers are possible −− Up to five slave controllers can form a polygon controller N2, O2, AIR etc. Tag Feed of media (gases, substrate) or actuators (e.g., drives) SP, OV Tag Introduction to control segment (e.g., sparger “SP”, headspace aeration “OV” in the culture vessel or container, mass flow controller “FL”) 1, 2 # The unit assigned to the controller output, e.g., number 1, 2 Out % Output signal “Out” from master controller in control range 0 to 100% to which the setpoints of the slave controllers should be assigned Setpoint PV Input of the setpoint for the master controller Mode Disable Enable Manually switchable mode of the slave controllers (only available when the master controller is in operating state “off” or “disabled”) Operating Manual BIOSTAT ® B 145 Preparing and Running the Process Selecting Slave Controllers ttActivate “Cascade Param.” to open the submenu for selection of slave controllers and change the previous selection. ttEnter the password. ttPress the key for position “Cascade #” for which another slave controller should be selected or the existing one deselected. Changes to a controller “Cascade #” deletes the subsequent selection. You must reassign all downstream controllers. Since the slave controllers control their actuators simultaneously, the order of the controllers has no effect on the control. 1 Adjusting the Polygonal Curve of a Slave Controller (Firmware Version 7.8 and Higher) The following description uses the “AIRSP” slave controller as an example. 2 ttActivate the function key of the slave controller you want to configure, e.g., “AIRSP-#.” ttAdjust the polygonal curve of the slave controller to the requirements, for example: 3 4 7 5 No. Parameter Description 1 No. Pressing shows the “Set Profile Set Point” window for configuring the values of the polygon point 2 Del Pressing deletes the polygon point 3 ADD Pressing shows the “Set Polygon Set Point” window for entering another polygon point 4 Delete All Pressing deletes all polygon points 5 Mode Pressing shows a window for activating/deactivating the slave controller 6 End Mode Selection of the slave controller mode when the master controller is “off” or “disabled” 7 Output Pressing shows a window for entering the output value; output is 0–100% related to the DO (pO2) controller output 8 Setpoint Pressing shows a window for entering the setpoint value 6 8 yyAfter closing the submenu with “OK,” the setpoints for the slave controller are graphically displayed as a polygonal curve above the “Out” of the master controller. ttSet the polygonal curves of the other slave controllers. 146 Operating Manual BIOSTAT ® B Preparing and Running the Process Special Notes The slave controllers work as long as the master controller is active, that is, in “auto” or “manual” mode. After the master controller is turned off (in state “off”), the slave controllers can be operated manually, either individually or together in the selected combination. The behavior of the master controller is based on sampled settings for the delay time and switching hysteresis. These settings are determined internally and not accessible for user modification. If necessary, they must be changed in the configuration. The following settings are saved for the master controller and slaves: −− The setpoint −− The settings for alarm monitoring −− The PID parameters of the master controller −− The settings of the slave controllers in relation to the output of the master controller profile parameters As a result, these settings then become available again after a power outage or after the DCU system or the controlled end device is turned off. They will be restored for the next process after power returns or the controller is switched back on. A reset of the DCU system (“Settings” menu) restores the factory settings. You must therefore store process or user-specific settings before the reset if you want to use them again later. After loading a new system configuration, the DCU system initially starts up with the factory settings. Here, too, you must reenter any process or user-specific settings. Application Instructions and Examples of Applied Control Strategies Other control strategies, such as Exclusive Flow, can be implemented by selecting and configuring the polygon controller: Example for “Advanced Additive Flow” Aeration Strategy ttGive “N2SP-#” a setpoint in the range “Out” = 0 to 20%, with the maximum at 0%. ttGive “AIRSP-#” a setpoint in the range “Out” = 0 to 20%, with the maximum at 20%. Leave “Out” constant for 20 to 100%. ttSet “O2SP-#” between “Out” = 20 to 40%, with the maximum at 40%. Leave “Out” constant for 40 to 100%. ttSet “STIRR-#” between “Out” = 0 to 40% and increase to a maximum at 60%. Leave “Out” constant for 60 to 100%. ttLeave “SUBS-A” constant in the range “Out” = 0 to 60% and increase to a maximum at 80%. yyThis activates the slave controller in the sequence shown, based on the deviation between the actual and setpoints and the output signal of the master controller. If the actual value approaches the setpoint, the slave controllers switch back in the reverse order. Operating Manual BIOSTAT ® B 147 Preparing and Running the Process Example for “O2 Enrichment” Aeration Strategy (Air, O2) In the “O2 Enrichment” aeration strategy, air is first used for the enrichment of the medium. If this is not sufficient, the air is then continuously enriched with pure oxygen in order to ensure a sufficiently high content of oxygen in the medium. ttSelect “AIRSP-1” and “O2SP-1” as slave controllers. ttFor “AIRSP-1,” set a minimum setpoint for “Out” = 0% and a maximum setpoint in the control range “Out” = 20 to 100%. ttFor “O2SP-1,” set −− a minimum setpoint for “Out” = 0 to 20% and −− a setpoint climbing to 100% in the control range “Out” = 20 to 100%. Fig. 8-35: Configuration of the aeration strategy “O2 Enrichment” yyThis cascade control initially leads to oxygen enrichment in the control range “Out” = 0 to 20%. Then, the oxygenation capacity in the control range “Out” = 20 to 100% is continuously increased via the addition of O2. 148 Operating Manual BIOSTAT ® B Preparing and Running the Process Example for “Gasflow Ratio” Aeration Strategy (Air, O2) In the “Gasflow Ratio” aeration strategy, a constant amount of gases is supplied to the culture vessel. ttSelect “AIRSP-1” and “O2SP” as slave controllers. ttFor “AIRSP-1,” set −− a maximum setpoint for “Out” = 0% −− a minimum sinking setpoint in the control range “Out” = 100%. ttFor “O2SP-1,” set −− a minimum setpoint for “Out” = 0% −− a maximum climbing setpoint in the control range “Out” = 100%. Fig. 8-36: Configuration of the aeration strategy “Gasflow Ratio” yyIn this cascade control, only air is supplied within the control range “Out” = 0%. The supply of air is continuously reduced. To the same extent, the supply of O2 is increased until solely oxygen is supplied within the control range “Out” = 100%. Operating Manual BIOSTAT ® B 149 Preparing and Running the Process 8.11.17 Glucose Controller (GLUCO) Function The glucose controller controls the addition of glucose in the culture vessel, so that the glucose concentration in the culture is maintained at a constant level. The measured value calculated by the BioPAT® Trace is used as the input signal of the “GLUCO-#” glucose controller. As soon as the glucose concentration falls below the setpoint in the culture, the speed of the addition pump is increased by the DCU system. More substrate containing glucose is added to the culture vessel until the setpoint for the glucose concentration is reached again. If the glucose concentration exceeds the setpoint in the culture, the speed of the addition pump is decreased by the DCU system. Less substrate containing glucose is added to the culture vessel until the glucose in the culture vessel breaks down biologically and the setpoint is reached again. Controller Screen Field Display Function, input required Mode Off Controller switched off Auto Controller switched on Manual Manual activation of the controller output; pump runs permanently g/l When the set value is not reached or is exceeded, the pump output changes accordingly SetPoint Alarm Param. Input of the alarm limits (high limit, low limit) and switching the alarm on/off Profile Param. Input of a time-dependent setpoint profile (max. 20 spikes) Function key Input of the weight limit (MIN/MAX) and other control parameters Operation ttSet the desired concentration via “Setpoint.” 150 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.18 Controller Functions on BIOSTAT ® RM Rocker In this section, the special controller functions angle control, aeration rate, sensor signal quality and additional features of the BIOSTAT ® RM 20 | 50 Rocker in the version “Optical” are described. Fig. 8-37: Main screen “Controller” of a configuration with RM 20 | 50 Rocker with culture bag Additional Functional Elements on BIOSTAT ® RM Rocker Symbol Display Meaning, Use Gas supply pressure Access to the menu where alarm limits can be configured Shaker drive operation [r/min] Direct access to the submenus to: −− Enter the setpoint for the shaker −− Select operating mode for ROCKS controller −− Switch to ROCKS controller menu Wobble angle in [°] Access to the menu where alarm limits can be configured 8.11.18.1 Introduction Angle Electronic configuration of the angle. Manual Positioning The “Phases” function can be used to bring the bag holder into the back or front position. The angle can be configured manually. This function can be used to harvest the culture when cultivation has ended. It can also be used to carry out sampling. Fig. 8-38: RM 20 | 50 Rocker optical | perfusion Aeration Rates Depending on culture bag size and the maximum desired pressure, different aeration rates are recommended. Operating Manual BIOSTAT ® B 151 Preparing and Running the Process Signal Quality of the Optical Sensors Display of the raw sensor data in the menu “Calibration” for evaluation of the signal quality of the optical sensors. 8.11.18.2 Angle Control This bioreactor has electronic angle control (“ANGLE”) Angle control Fig. 8-39: Main menu of the BIOSTAT ® B Configuration of the process value “ANGLE” ttIn the work area of the “Main” menu, press the “ANGLE” function key or select the “Controller” main function and then the “ANGLE” controller. yyWhen the “Main” menu is accessed, a submenu (Fig. 8-40) opens with a keypad on the left side for data entry and a selection field for possible operating modes. ttEnter the new setpoint (observe the permissible values between “Min” and “Max”). If you don’t want to save the new value, leave the submenu and press the C key. To turn on the controller, press the button “auto”. Fig. 8-40: Direct access to input and selection of the “Angle” controller mode 152 Operating Manual BIOSTAT ® B ttPress the parameter key Preparing and Running the Process to view the graphic controller output. ttPressing the parameter button again opens a password entry screen. Fig. 8-41: Representation of the “Angle” controller outputs ttSet the controller parameterization and confirm the entry with “OK”. yyThe submenu window closes. The setpoint is active and is displayed. 8.11.18.3 Position Settings “POSITIONING” The “POSITIONING” function is used to send position information to the BIOSTAT ® RM Rocker and receive status information. −− The RM Rocker platform can be moved into a front or rear position. −− The angle can be configured manually. −− The sample function can be activated. If the “Sample” control button is pressed, then the platform of the RM Rocker moves into a 10° forward-inclined position. In order to avoid local overheating, heating is switched off while the Sample Function is active. After stopping the Sample Phase, the rocker moves into its rear position again and the heater turns on. A safety feature ensures that the process automatically starts after a period of time set by the user in the RM Rocker SPS. This prevents the user from accidentally forgetting to manually start the RM Rocker after sampling. −− The HEAT_PID function is used to transfer the PID parameters of the heating system to the RM Rocker, where they are stored locally. Operating Manual BIOSTAT ® B 153 Preparing and Running the Process Fig. 8-42: Function “POSITIONING” Configuration of the position: ttIn the work area of the “Controller“ menu, press [Æ Fig. 8-42] the function key “ANGLE” yyThe menu “POSITIONING” is displayed in the upper right of the screen. ttPress the touch button “FRONT-#” (or “BACK-#”, “HEAT_PID-#”, “SAMPLING-#”). Example yyThe screen shows the phase “FRONT-#”. ttPress the touch button . ttEnter the standard password and confirm with [ok]. yyThe “Phase Parameter FRONT-#” window opens. ttPress the input field “MANPOS-#”. yyA keypad appears. ttEnter the desired angle using the keypad and confirm with [ok]. ttClose the window “Phase parameters FRONT-#”. ttPress the touch button “State”. yyThe window “Phase Mode” opens. ttTo start the phase, press the touch button “start”. 154 Operating Manual BIOSTAT ® B Preparing and Running the Process yyThe phase window “Phase FRONT-#” opens. ttConfirm the start of the phase by pressing the touch button “YES”. yyThe platform of the RM Rocker now moves to the forward position, the status changes to “Running”. The operation of the phases “BACK-#”, “HEAT_PID-#”, and “SAMPLING-#” is analogous to the described phase “FRONT-#”. 8.11.18.4 Aeration Rate (BIOSTAT ® RM 20 | 50 Rocker) The flow range of the mass flow controller in the BIOSTAT ® B can be chosen when ordering. For operation with a Culture Bag Holder 20, a mass flow controller is supplied by default for total flow with a flow range of up to 1 slpm. For a Culture Bag Holder 50, the flow range of the BIOSTAT ® B is restricted to a maximum of 3 slpm by default. The dynamic pressure in the culture bag changes depending on the selected aeration rate (see the following figure). We recommend selecting an aeration rate that keeps the dynamic pressure well below 30 mbar and, where appropriate, to remove the reducing valve from the exhaust filter. Fig. 8-43: Dynamic pressure in the culture bag depending on the aeration rate Operating Manual BIOSTAT ® B 155 Preparing and Running the Process 8.11.19 Additional information — only for BIOSTAT ® RM 20 | 50 Rocker Functions via RM 20 | 50 Rocker Touch Panel Please observe that the following operations can be made solely on the RM 20 | 50 Rocker Touch Panel: −− All calibration work on the RM 20 | 50 Rocker −− The duration of the sampling position must be configured | changed in the Settings menu −− The bag configuration −− Query of the service interval for the RM 20 | 50 Rocker −− RM 20 | 50 Rocker functions in the menu “technician level” You can find details about the above functions in the operating instructions for the RM 20 | 50 Rocker. Temperature Control Measuring and Control Range The temperature control range of the RM 20 | 50 Rocker system is between 15 °C and 40 °C. In the DCU Control system, an input range from 0 to 40 °C is implemented. Please observe that, when using a heating mat for temperature control, the temperature range 0 °C to 40 °C is used only for display. Temperature control in these ranges is not possible. 156 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.11.20 Additional Information — For BIOSTAT ® RM 200 Rocker Only Additional functional elements — RM 200 Rocker with BIOSTAT ® B in Twin-Bag configuration only Symbol Display Meaning, Use Line green Single-Bag function switched on −− When using a 200 L culture bag. Line gray Twin-Bag function switched on −− When using one or two 100 L culture bags. For BIOSTAT ® B Control Unit in Twin-Bag Configuration Only: You must set the Single or Twin-Bag function depending on the number and type of mounted culture bags. This setting affects the parameter settings for the culture bag(s): Single-Bag Function When using a 200 L culture bag: Only set the process parameters for Unit-1. Do not set the process parameters for Unit-2. Twin-Bag Function When using one or two 100 L culture bags. −− Two culture bags: Set the process parameters for both Units. −− One culture bag: Set the process parameters for one Unit, e.g. Unit-1. Set all the regulators for the other Unit to “Off”. After restarting the system, the Twin-Bag function is activated by default! Operating Manual BIOSTAT ® B 157 Preparing and Running the Process 8.11.20.1 Activating Single-Bag Function Fig. 8-44: “Main” menu of the configuration BIOSTAT ® B with RM 200 Rocker (Twin-Bag configuration), Twin-Bag function activated yyA 200 L culture bag is installed. yyThe temperature controller “TEMP” is deactivated. The temperature controllers “TEMP-1” and “TEMP-2” must be deactivated before you can configure the Single-Bag function. If one of the temperature controllers is active, an error message appears (see Page 159). ttPress the “Calibration” function key in the footer. ttIn the work area of the “Main” or “Controller“ menu, press the function key “SINGLEBAG”. yyThe “SINGLEBAG” window opens. ttTo activate the Single-Bag function, press the touch button “On”. yyThe “New MESSAGE” window opens. 158 Operating Manual BIOSTAT ® B Preparing and Running the Process Observe the note “You must not operate UNIT 2!”: If the Single-Bag function is activated, the process parameters for process 1 (Unit-1) must be set. Process parameters must not be set for “Unit-2”! The Single-Bag function automatically applies the temperature value of Unit-1 to Unit-2. Control of other parameters (aeration, pH control, etc.) is carried out by Unit-1. −− Only set the process parameters for Unit-1. −− Do not set the process parameters for Unit-2. Parameter settings for Unit-2 can lead to faults in the operating sequence. ttTo confirm the Single-Bag function (Single-Bag operation), press the touch button “Acknowledge”. yyThe Single-Bag function is activated. ttOnly set the process parameters for Unit-1. Error message when temperature controller is activated The temperature controllers “TEMP-1” and “TEMP-2” must be deactivated before you can configure the Single-Bag function. If one of the temperature controllers is active, an error message appears. yyThe temperature controller(s) TEMP-1 and | or TEMP-2 is | are activated. yyActivate the Single-Bag function (see Page 158). yyThe “New MESSAGE” window opens. Observe the note “Switch off TEMP control loops before change to Single-Bag Operation!” The Single-Bag function can only be activated if you disable the temperature controller. ttConfirm the error message by pressing the touch button “Acknowledge”. yyThe “New MESSAGE” window opens. ttConfirm the activated Twin-Bag function by pressing the touch button “Acknowledge”. Operating Manual BIOSTAT ® B 159 Preparing and Running the Process ttTurn off the temperature controller(s) TEMP-1 and | or TEMP-2. ttActivate the Single-Bag function (see Page 158). 8.11.20.2 Activating the Twin-Bag Function Fig. 8-45: “Main” menu of the configuration BIOSTAT ® B with RM 200 Rocker (Twin-Bag configuration), Twin-Bag function activated yyOne or two 100 L culture bags are installed. yyThe temperature controller “TEMP” is deactivated. The temperature controllers “TEMP-1” and “TEMP-2” must be deactivated before you can configure the Single-Bag function. If one of the temperature controllers is active, an error message appears (see Page 161). ttPress the “Calibration” function key in the footer. ttIn the work area of the “Main” or “Controller“ menu, press the function key “SINGLEBAG”. 160 Operating Manual BIOSTAT ® B Preparing and Running the Process yyThe “SINGLEBAG” window opens. ttTo activate the Twin-Bag function, press the touch button “On”. yyThe “New MESSAGE” window opens. ttTo confirm the Twin-Bag function (Twin-Bag operation), press the touch button “Acknowledge”. yyThe Twin-Bag function is activated. When using only one 100 L culture bag, you must configure the process values for one unit (e.g. Unit-1). The controller for the other Unit must be set to “off”. One culture bag (100L): ttSet the process parameters for Unit-1. Two culture bags (100L): ttSet the process parameters for Unit-1 and Unit-2. Error Message When Temperature Controller Activated The temperature controllers “TEMP-1” and “TEMP-2” must be deactivated before you can configure the Twin-Bag function. If one of the temperature controllers is active, an error message appears. yyThe temperature controller(s) TEMP-1 and | or TEMP-2 is | are activated. yyActivate the Twin-Bag function (see Page 160). yyThe “New MESSAGE” window opens. Operating Manual BIOSTAT ® B 161 Preparing and Running the Process Observe the note “Switch off TEMP control loops before change to Twin-Bag Operation!” The Twin-Bag function can only be activated if you disable the temperature controller. ttConfirm the error message by pressing the touch button “Acknowledge”. yyThe “New MESSAGE” window opens. ttConfirm the activated Single-Bag function by pressing the touch button “Acknowledge”. ttTurn off the temperature controller(s) TEMP-1 and | or TEMP-2. ttActivate the Twin-Bag function (see Page 160). 162 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.12 “Settings” Menu The “Settings” menu permits changes to the system configuration. Malfunctions that have unforeseeable impacts on safe operation can result from settings that are not permissible or are unsuited for a certain terminal unit. Settings that impact safe operation are password-protected. Only trained and experienced persons may change these settings. The standard password [ see Chapter “4.4 Password Protection of Individual Functions“] may only be disclosed to authorized users and the service password [ separate notice] only to authorized service technicians. 8.12.1 General Information In the “Settings” menu, the DCU system provides various functions for system maintenance and troubleshooting: −− General settings like date, time, fail time, password-protected screen saver, parameter settings for communicating with external devices (“Internet Configuration”). −− Defining Process Values (PV) and their ranges and limits. −− Manual operation of digital and analog inputs and outputs or simulation controllers, for example. −− Service function, e.g., for resetting the system (Reset) or to select the system configuration on multiple configurations. “Settings” operator screen Fig. 8-46: “Settings” menu (system settings) Operating Manual BIOSTAT ® B 163 Preparing and Running the Process Functions Available for Selection Touch button Function System parameters Changing general system settings [ see Chapter „8.12.2 Settings“] PV Ranges Configuring measurement ranges of process values [ see Chapter „8.12.3 Measuring Range Settings“] Manual Operation Switch process inputs and outputs to manual operation [ see Chapter „8.12.4 Manual Operation“] External View status of externally connected devices, e.g. scales [ see Chapter „8.12.6 Externally Connected Devices“] Service Service and diagnostic interventions [ see Chapter „8.12.7 Service and Diagnosis“] System information displayed Field Hardware Value PCM 9363 Function, entry required Version of the DCU hardware Firmware X.YY Version of the system’s firmware Configuration XX_YY_ZZZZ Version of the configuration For inquiries about the system or for contacting the service department in the event of a malfunction, please always state the firmware indicated here and the configuration of your system. 164 Operating Manual BIOSTAT ® B Preparing and Running the Process 8.12.2 Settings Using the “System Parameters” touch button, you can change general system settings, for example setting the real-time clock on the DCU system. To open the “System parameters” submenu, you will need to enter the standard password [ see Chapter “4.4 Password Protection of Individual Functions“] . Field Value Function, entry required Time hh:mm:ss Enter the current time, format: hh:mm:ss Time Synchronize Synchronize: enabled/ Enable and disable time synchronization disabled IP Address Entry of the IP-address Time Zone Select the time zone Date dd.mm.yyyy Enter the current date, format: dd:mm:yy Beeper enabled/ disabled Turns acoustic signals on | off, e.g. alarm tones Fail time hh:mm:ss Enter power outage time to tell system how to behave when switched back on, format: hh:mm:ss Power outage time < FAILTIME: The system continues to run on the previous settings Fig. 8-47: Submenu “System Parameters” Power outage time > FAILTIME: System switches to default mode Screensaver hh:mm Enter the time of inactivity after which the screen saver will be turned on, Format: hh:mm:ss (00:00:00 = switched off) Internet Config 12-digit The DCU system’s address in the IP network binary number Changes to the date and time will only take effect in the first 5 minutes after the DCU system is turned on. 8.12.3 Measuring Range Settings The beginning and end of the measuring range (“PV Ranges”) for all process values can be changed in the “Settings” menu. Measuring ranges configured specifically to devices or customer specifications are factory-set in the bioreactor [Æ Configuration Documentation]. Only personnel authorized to do so may change the menu settings. To make settings in the menu, the standard password needs to be entered [Æ Chapter “4.4 on page 50“]. Operating Manual BIOSTAT ® B 165 Preparing and Running the Process Operator screens −− After pressing the “PV ranges” touch button and entry of the standard password, the “Process Value Ranges” submenu opens: Fig. 8-48: Table of process values (or ranges) configured −− By pressing the “Ch.” (Channel) touch button, the process values (ranges) can be configured: Fig. 8-49: Manual configuration of process values using example “TEMP-1” (channel 1) Field Value Ch. Min Max Decimal Point Alarm Low °C Alarm High °C Alarm disabled enabled Delay s 166 Operating Manual BIOSTAT ® B Function, entry required Channel Minimum value Maximum value Decimal point display Lower alarm limit in the physical unit Upper alarm limit in the physical unit Alarm monitoring deactivated Alarm monitoring alarms active Alarm lag time Preparing and Running the Process 8.12.4 Manual Operation When starting up operations and troubleshooting, all analog and digital process inputs and outputs as well as DCU internal parameters can be switched to manual operation (“Manual Operation” touch button). −− To open the “Manual Operation” submenu, you will need to enter the standard password [ see Chapter “4.4 Password Protection of Individual Functions“] . −− You can disconnect inputs from the external signal generators and preset input values to simulate measuring signals. −− You can separate outputs from internal DCU functions and directly influence them in the operator screen, for example, to test the effect of certain settings. Settings during manual operation have the highest priority; their effects on the inputs and outputs of the DCU system supersede those of other functions. Color Displays of Inputs | Outputs −− If an input or output is in “Manual” operation, the display in the “Value” column has a green background. −− If a controller is in cascade control mode, the display in the “Setpt” column has a light green background (controllers only). −− If a phase is acting on an output, the display in the “Value” column has a turquoise background. −− If an input or output is in “Manual” operation, the display in the “Value” column has a yellow background. −− If an input/output is locked, the display in the “Value” column has a violet background. −− If an emergency off is triggered during process, the displays of all outputs in the “Value” column have a red background. −− If no function is accessing an input | output, the display in the “Value” column has a gray background. −− If the process control system is accessing an output, the display in the “Value” column has a white background. Operating Manual BIOSTAT ® B 167 Preparing and Running the Process 8.12.4.1 Manual Operation for Digital Inputs −− For manual operation, disconnect the digital input from the external sensor, e.g. limit value sensors and simulate the input signal from the “ON” or “OFF” input. Operator Screen Fig. 8-50: Manual configuration of digital inputs, example “HEATC-1” (simulation for signal of the power-on status of the heating) Field Value Function, entry required Day Description Display of digital input Port Description Hardware address Value PV Display of signal level of switching state 0 V = off 24 V = on, Input for “AUTO” or “MANUAL ON | OFF” operating modes: “AUTO“: Normal operation, external input acts on DCU “MANUAL”: Manual operation, manual specification of digital input A Display active status I: on = turned on (signal level 24 V) N: on = turned on (signal level 0 V) off : turned off AL Alarm state A = activated – = not activated PV Switch status of the digital input off = turned off on = turned on Special notes −− The following signal levels apply to the switch status (status): off On 168 Operating Manual BIOSTAT ® B 0V… 24 V for process inputs (DIP) Preparing and Running the Process After working on the manual level, you have to switch all inputs back to the “AUTO” operating mode. Otherwise, the function of the DCU system will be limited. −− During manual operation, disconnect the digital output from the internal DCU function and manipulate it directly. For static digital outputs, e.g. controlling valves, switch the output on or off. For pulse-duration modulated outputs, set the switch-on ratio in [%] by hand. −− Multiple functions may act internally on a digital output. After the field is selected, the currently active function will be displayed in the VALUE column in the corresponding submenu. If several functions are activated (e.g. on controller outputs that interact with sterilization), the following priority applies: Highest priority Lowest priority Shutdown Manual Operation (Manual Level) Locking Pump calibration Controllers, timers, sensors, scales/balances Operating state (OPS) Operating Manual BIOSTAT ® B 169 Preparing and Running the Process Operator Screen Fig. 8-51: Manual configuration of digital outputs, example “HEAT-1” (simulation for signal controlling the heating) Field Value Function, entry required Day Description Display of digital input Port Description Hardware address Val off on nn % Switch status of the digital output off = turned off on = turned on % = power-on ratio (0 … 100 %) for pulse-duration modulated digital outputs Input for “AUTO” or “MANUAL ON | OFF” operating modes: “AUTO”: Normal operation, external input affects DCU “MANUAL”: Manual operation, manual specification of digital output A Display active status I = turned on (signal level 24 V) N = turned on (signal level 0 V) off = turned off Ty Upstream function cl = controller expr = logical function – = without SRC 170 Operating Manual BIOSTAT ® B nn % | off Upstream controller output Display of output value: off –100 % to +100 % Preparing and Running the Process Special notes −− The following signal levels apply to the switch status (status): off on 0V… 24 V for process outputs (DO) −− On pulse-width modulated digital outputs, the relative power-on time is displayed and | or preset. The cycle time is defined in the specific configuration. Example: Cycle time 10 sec, PWM*1 output 40%: −− Digital output 4 sec on and 6 sec off. After working on the manual level, you have to switch all outputs back to the “AUTO” operating mode. Otherwise, the function of the DCU system will be limited. 1 PWM: Pulse-width modulation Operating Manual BIOSTAT ® B 171 Preparing and Running the Process 8.12.4.2 Manual Operation for Analog Inputs You can disconnect all analog inputs from the external circuitry during manual operation, e.g., a measurement amplifier and simulate them by entering a relative signal level (0...100%). Operator Screen Fig. 8-52: Manual configuration of analog inputs, example “JTEMP-1” (simulation for input signal for temperature measurement in heating circuit) Field Value Function, entry required Day Description Display of analog input Port Description Hardware address Value PV Input signal 0 to 10 V or 0/4 to 20 mA Input for “AUTO” operation or “MANUAL ON | OFF” PV Process value Unit Physical variable Special notes −− For analog inputs (A), the signal level can be configured between −− 0 to 10 V (0 to 100 %) −− 0 to 20 mA (0 to 100 %) −− 4 to 20 mA (0 to 100 %) −− During manual operation, only the relative signal level (0 to 100 %) of the analog inputs is displayed or entered. The allocation to the physical value is a product of the measuring range of the affected process value. After working on the manual level, you have to switch all inputs back to the “AUTO” operating mode. Otherwise, the function of the DCU system will be limited. 172 Operating Manual BIOSTAT ® B 8.12.4.3 Preparing and Running the Process Manual Operation of Analog Outputs You can disconnect analog outputs from the internal DCU functions and influence them directly using signals with a relative level (0 to 100 %). Output signals have the following priorities: Highest priority Lowest priority Shutdown Manual Operation (Manual Level) Locking Controllers, etc. Operator screen Fig. 8-53: Manual configuration of analog outputs, example “STIRR-1” (simulation of control signal for the speed regulation of the motor drive) Field Value Function, entry required Day Description Display of the analog output, e.g. STIRR-1 Port Description Hardware address, e.g. 1AO05 Value PV Output signal 0 to 10 V or 0|4 to 20 mA Input for “AUTO” or “MANUAL ON | OFF” operating modes: “AUTO”: Normal operation, external input affects DCU “MANUAL”: Manual operation, manual specification of analog output Ty SRC Upstream function cl = controller expr = logical function – = without nn % | off Upstream controller output Display of output value: off –100 % to +100 % Operating Manual BIOSTAT ® B 173 Preparing and Running the Process Special notes −− The physical signal level of the analog outputs (AO) can be configured between: −− 0 to 10 V (0 to 100%) −− 0 to 20 mA (0 to 100%) −− 4 to 20 mA (0 to 100%) After working on the manual level, you have to switch all outputs back to the “AUTO” operating mode. Otherwise, the function of the DCU system will be limited. 174 Operating Manual BIOSTAT ® B 8.12.4.4 Preparing and Running the Process Manual Operation for Controllers (“Control Loops”) You can simulate controllers in manual operation by entering a setpoint. Operator Screen Fig. 8-54: Manual configuration of controller, example “TEMP-1” (simulation of control signal of temperature controller) Field Value Function, entry required Day Description Display of the controller, e.g. TEMP-1 PV Process value Setpt Display of setpoint Input for “OFF” or “AUTO” operating modes: “OFF”: Controller is turned off “AUTO”: Normal operation, setpoint for controller can be configured Unit Physical variable C Display of active cascade 0 = no cascade 1 to n = specific cascade for cascade controller Out Calculated output value Special notes After working on the manual level, you have to switch all outputs back to the “AUTO” operating mode. Otherwise, the function of the DCU system will be limited. Operating Manual BIOSTAT ® B 175 Preparing and Running the Process 8.12.5 Manual Operation of Sequence Control (“Phases”) You can simulate sequences in manual operation (e.g. during startup or in case of problems in the sequence execution during sterilization) by starting a sequence. Operator Screen Fig. 8-55: Manual start of a sequence, example “FILL1” (simulation of control signal for double wall filling) Field Value Function, entry required Day Description Display of sequence, e.g. FILL-1 State Display of sequence status | step Start | stop of a sequence (“START” | “STOP”) Continuation to next sequence step (“STEP”) Step 176 Operating Manual BIOSTAT ® B Display of current sequence step Preparing and Running the Process Special Notes Type and number of sequence steps of individual sequences depends on the configuration of your system. After working on the manual level, you must stop all sequences. Otherwise, the function of the DCU system will be limited. 8.12.6 Externally Connected Devices The “External” menu button can be used to view and set the status of externally connected units (e.g. scales). Only personnel authorized to do so may change the menu settings. To make settings in the menu, the standard password needs to be entered [ Chapter “19 Appendix”]. Operator Screen After pressing the “External” touch buttons and entry of the standard password, the “External System” submenu opens: Fig. 8-56: Display of externally connected devices in the “External System” submenu (configuration example) Operating Manual BIOSTAT ® B 177 Preparing and Running the Process Field Value Function, entry required Day Description Display of the connection, e.g. SERIAL-A1 Interface Description Display of interface Alarm Display and configuration of alarm status: enabled = activate alarm disabled = deactivate alarm Status Display of status of connected device (offline | online) 8.12.7 Service and Diagnosis This operating level is only accessible for interventions by authorized service technicians or associates of Sartorius Stedim Biotech. 178 Operating Manual BIOSTAT ® B Faults 9. Faults 9.1 Safety Instructions Danger to life caused by electrical voltage! Contact with parts under voltage represents a direct danger of death. −− Work on the electrical equipment of the device may only be carried out by a competent electrician. −− Before any work, turn the unit off and disconnect it from power. −− During all work on the electrical equipment, disconnect it and check that voltage is no longer applied. Danger of limbs being pulled into the rotation pump and crushed! −− Do not remove the safety mechanisms. −− Allow only qualified and authorized personnel to work on the device. −− Disconnect the device from power when performing maintenance and cleaning tasks. −− Block the danger zone off. −− Wear personnel protective equipment. Danger of burns upon contact with hot surfaces! −− Avoid contact with hot surfaces like vessels, motor housings and pipelines carrying steam. −− Let the vessels/culture vessels cool before troubleshooting. −− Block the danger zone off. 9.2 Troubleshooting Always proceed according to the following steps when faults occur on the device. 1. Switch off the device and unplug it from the power supply (pull power plug) if the fault (e.g. smoke or odors, abnormally high surface temperatures) represents a direct danger to personnel or property. 2. Inform management on site about the fault. 3. Determine the cause of the fault and remedy it before switching the device back on [ see Chapter “7.14 Turning the Device On and Off“]. If the fault cannot be remedied, please consult your Sartorius Service [ Chapter “15.1 Service“]. Operating Manual BIOSTAT ® B 179 Faults 9.3 Hardware-related Faults Danger of injury if personnel qualifications are insufficient! Improper use can lead to significant personal injury and/or property damage. It is therefore important that all troubleshooting activities be carried out by technical personnel. 9.3.1 Fault Table “Contamination” We recommend that you perform a sterility test before each process. Duration 24 - 48 h. Conditions for a Sterility Test: −− The culture vessels must be filled with the prescribed culture medium or a suitable starter medium and autoclaved in accordance with the specifications. −− All of the components, peripheral devices, correction medium feed lines and sampling systems to be tested must be connected to the culture vessels. −− The system must be set to the prescribed operating conditions (e.g., temperature, stirrer speed, aeration). Contamination Possible causes Corrective measures Generalized and widespread, even without having inoculated the culture (during the sterility test phase) Insufficiently autoclaved culture vessel. Check the autoclave settings. Increase the autoclaving time. Perform sterility tests using test spores. Air inlet line or air inlet filter defective. Replace the tubing. Check the filter and replace if necessary. Generalized and gradual (even without inoculating the culture) Seals on the culture vessel or the integrated components are damaged (e.g. hairline cracks) Carefully check the integrated parts. If suspected to be damaged (rough, porous surfaces or dents), replace seals. After inoculation, wide-spread Contaminated inoculum culture Non-sterile inoculation equipment Take control samples of the inoculation culture and test inoculated culture medium from the vessels (e.g. on test nutrient solutions). Incorrect inoculation Check the inoculation procedure. Carefully practice the inoculation process. Supply air filter or connection has become non-sterile or defective Check the filter and replace if necessary. Replace the connection line. Supply air filter or connection has become non-sterile or defective Check the filter and replace if necessary. Replace the connection line. Accidental or unauthorized manipulation of equipment Take organizational measures at the work site to prevent the equipment from being manipulated without authorization. During the process, rapid 180 Operating Manual BIOSTAT ® B Faults Contamination Possible causes Corrective measures During the process, gradually Seals on the culture vessel or the integrated components are defective (e.g. hairline cracks or porosity) If possible, continue process to the end. Once finished, dismantle the vessel and carefully check the integrated parts. If suspected to be damaged (rough, porous surfaces or dents), replace seals. Exhaust air filter(s) or connection has become unsterile or is defective (contaminated from the exhaust air line). Check the filter (if possible, perform a validity test) and replace if necessary. Replace the connection line. 9.3.2 Troubleshooting Table “Counter Cooling System” The counter-cooling system does not work or does not provide sufficient cooling power. Fault Possible causes Corrective measures The cooling water is not fed into the system The laboratory’s supply line is blocked or the valves in the cooling water supply are defective. If all other potential causes can be excluded (see below), contact Sartorius Service. The cooling water supply valve does not work or the non-return valve has become stuck because of contaminated cooling water or scale deposits. Check the water hardness (no more than 12 dH). Flow rate too low The minimum operating temperature is around 8° C above the cooling water temperature. Insufficient cooling power Cooling water temperature too high Check the non-return valve. Feed clean cooling water into the system (if necessary, install a pre-filter). If necessary, install an upstream cooling device. 9.3.3 Troubleshooting Table “Aeration and Ventilation” Aeration or ventilation system does not work or does not provide sufficient gas/ ventilation. Fault Possible causes Corrective measures Air inlet line blocked Air inlet filter blocked Check the air feed (dry and free of oil and dust). If necessary, install a pre-filter. The gas or air supply is blocked or decreases suddenly Check the hose and filter and, if necessary, fit new sterile filters. Hose kinked or disconnected Exhaust air filter blocked (e.g. as a result of moist air and the formation of condensate or foam) Operating Manual BIOSTAT ® B 181 Faults 9.4 Process-related Faults / Alarms Faults in the operating sequence are displayed as alarms on the operator terminal. To correct these process-related faults, read the following sections. The DCU system makes a distinction between alarms and messages. Alarms have higher priority and are displayed first ahead of the messages. 9.4.1 Alarm Triggering When alarms are triggered, they automatically are displayed in a window that superimposes all other windows. The color of the soft button alarm bell turns red. The color of the alarm bell stays red as long as at least one unconfirmed alarm remains in memory. “New ALERT” operator screen Fig. 9-1: Alarm message: “New ALERT” pop-up screen (new alarm) −− Closing the window: , the alarm is stored as an “unacknowledged alarm” (UNACK) −− After pressing in the alarms list and the alarm symbol remains activated. −− The alarm window closes after acknowledgement of the alarm with “Acknowledge”. The alarm message disappears in the header. 182 Operating Manual BIOSTAT ® B 9.4.2 Faults “Alarm Overview” Menu The alarm overview can be selected as follows: ttPress the “Alarm” function key. Operator screen “Alarm” Fig. 9-2: Alarm table, accessible through the “Alarm” function key Field Function, entry required ACK ALL Acknowledges all activated alarms ACK Acknowledges the selected alarm RST Resets and deletes the selected alarm 9.4.3 Process Value Alarms The DCU system has limit value monitoring routines that monitor all process variables (measured data and calculated process values) to ensure that they are within the alarm limits (High | Low). The alarm limits must be within the measurement range limits. After entering the alarm limits, you can release or lock the limit value monitoring individually for every process parameter. The DCU system can lock certain process outputs after process value alarms. Operating Manual BIOSTAT ® B 183 Faults “Process value alarms” operator screen Fig. 9-3: Submenu for configuring alarm monitoring, example “TEMP-1”, called from the “Controller” menu, overview “All” Field Value Function, entry required High limit °C Upper alarm limit in the physical unit of the PV Low limit °C Upper alarm limit in the physical unit of the PV Alarm 184 Operating Manual BIOSTAT ® B Status for alarm monitoring disabled Alarm monitoring, high | low alarms locked enabled Alarm monitoring, high | low alarms activated Faults Operating Instructions Alarms are displayed on the operator screen and must be acknowledged: −− If the value falls outside the alarm limits, an alarm window opens above the active screen. An acoustic signal sounds. The alarm display is displayed in the header line of the operator screen. The process value display also shows a small alarm symbol: Operator screen example: exceeding the alarm limit Fig. 9-4: Alarm message, exceeding the alarm limit for pH-1. −− The alarm window closes after acknowledgement of the alarm with “Acknowledge” or after pressing . −− After the alarm is confirmed with “Acknowledge”, the alarm symbol disappears. ”, the alarm is stored in the alarm list as an unacknowledged −− After pressing “ alarm and the alarm symbol remains active (the alarm bell stays red). −− If several alarms have been triggered, the next, still unconfirmed alarm will be displayed after the active alarm window is closed. Special notes The DCU system continues to display limit value alarms as long as the process value remains outside the alarm limits. Operating Manual BIOSTAT ® B 185 Faults 9.4.4 Alarms for Digital Inputs Digital inputs can be prompted in response to alarm conditions as well. These can be used to monitor components like limit contacts (antifoam | level sensors), motor protection switches or circuit breakers. When an alarm is triggered, an alarm message with the time of the alarm event and an acoustic confirmation signal is emitted. The DCU system can lock certain process outputs after process value alarms. “Alarm monitoring” operator screen Fig. 9-5: Activating and deactivating alarm monitoring Fig. 9-6: Fig. 19-6: Alarm deactivated, alarm activated Field Value Alarms Param. 186 Operating Manual BIOSTAT ® B Function, entry required Alarm monitoring operating mode disabled Alarm monitoring locked for the input enabled Alarm monitoring activated for the input Faults Operating Instructions A new alarm is indicated in two ways: −− When an alarm is triggered for the first time, a message appears in the display and an acoustic signal is emitted. −− The alarm symbol is displayed in the header line of the operator screen. ttEliminate the cause of the alarm. Check the function of the component that is producing the input signal, the corresponding connections, and if necessary the regulator settings. ttConfirm the alarm with “Acknowledge” or press “X”. yyThe alarm window closes. −− After the alarm is confirmed with “Acknowledge”, the alarm symbol disappears (the alarm bell turns white). The alarm is recorded in the alarm list as a confirmed alarm (“ACK”). −− After pressing “X”, the alarm is stored in the alarm list as an unacknowledged alarm and the alarm symbol remains active (the alarm bell stays red). Special notes For an overview of alarms that have occurred, you can open the alarm table with the “Alarm” menu button. 9.4.5 9.4.5.1 Alarms, Meaning and Corrective Measures Process Alarms The user can switch on and off the individual alarms listed in the following table: Text in the alarm line Meaning Remedy [Name] State Alarm Digital input alarm Confirm alarm with “ACK” [Name] Low Alarm The corresponding process value has exceeded its lower alarm limit Confirm alarm with “ACK” [Name] High Alarm The corresponding process value has exceeded its upper alarm limit Confirm alarm with “ACK” Jacket Heater Failure Overheating protection in the temperature circulation of the double wall has triggered The tempering system must be refilled Motor Failure Overheating protection of the motor responded Allow the motor to cool down OVP Overvoltage protection Operating Manual BIOSTAT ® B 187 Faults 9.4.5.2 System Alarms The alarms in the following table are system-generated messages that the user cannot switch off: Text in the alarm line Meaning Remedy Source: Factory Reset Confirmation message for a system reset started from the “Settings” menu Confirm alarm with “ACK' [Name] Watchdog Timeout Confirmation message for a Watchdog Timeout, triggered by malfunctions in the DCU including reference to the source of failure Note down the alarm and report it to the Service Department. Confirm alarm with “ACK' Power Failure Power lost at [yyyy-mm-dd hh:mm:ss] Power failure with the date and time Confirm alarm with “ACK' Power Failure, Process Stopped System in Standby Power lost at [yyyy-mm-dd hh:mm:ss] Power failure with the date and time; maximum power outage time exceeded Confirm alarm with “ACK”. Shut down Unit # “Shut down” was pressed on the bioreactor Switch the bioreactor back on with “Shut down” 188 Operating Manual BIOSTAT ® B Cleaning and Maintenance 10. Cleaning and Maintenance Incorrect cleaning and maintenance can lead to erroneous process results, causing high production costs. Regular cleaning and maintenance is thus essential. Among other factors, the operational safety and effective performance of fermentation also depend on proper cleaning and maintenance. The cleaning and maintenance intervals largely depend on the stress placed on the culture vessel and equipment by aggressive components contained in the media (e.g. acids and bases used to regulate pH) and the level of contamination from culture and metabolic product residues attached to this equipment. Danger to life caused by electrical voltage! Electrical switching elements are installed in the device. Contact with parts under voltage represents a direct danger of death. −− Never open the device. The device may be opened only by authorized personnel of the Sartorius Stedim Biotech Company. −− Work on the electrical equipment of the device may only be carried out by Sartorius Service or authorized technicians. −− During repair work and cleaning, turn the power supply off and secure it against reactivation. −− Keep moisture away from parts under voltage, as it can lead to short circuits. −− Check the electrical equipment of the device regularly for defects such as loose connections or damage to the insulation. −− In case of defects, turn the power supply off immediately and have the defects corrected by Sartorius Service or authorized technicians. −− Have the electrical components and stationary electrical equipment checked by an electrician at least every 4 years. Danger of limbs being pulled into the rotation pump and crushed! −− Do not remove the safety mechanisms. −− Allow only qualified and authorized personnel to work on the device. −− Disconnect the device from power when performing maintenance and cleaning tasks. −− Block the danger zone off. −− Wear personnel protective equipment. Danger of burns upon contact with hot surfaces! −− Avoid contact with hot surfaces like vessels, motor housings and pipelines carrying steam. −− Block the danger zone off. −− Wear protective gloves when working with hot culture media. Operating Manual BIOSTAT ® B 189 Cleaning and Maintenance Danger from projecting components! −− Ensure that danger points such as corners, edges and projecting components are covered. Preliminary Steps Always take the following preliminary steps when performing cleaning and maintenance: ttTurn the device off at the main switch. ttRemove the power supply from the laboratory connection. ttTurn off all supply media in the lab (water and gas supply). ttEnsure that the connections and hoses have been depressurized. ttIf necessary, remove the supply media lines from the device. 10.1 Cleaning Danger of corrosion and damage to the device and the culture vessel by unsuitable cleaning agents. −− Avoid strongly caustic or alkaline and/or chloride-containing detergents. −− Avoid solvent-based cleaning agents. −− Ensure that the cleaning agents used are compliant materials. Observe the safety instructions for the cleaning agents. The use and disposal of cleaning agents, and water containing the same, may be subject to legal or environmental protection regulations. 10.1.1 Cleaning the Device ttClean the device housing with a slightly damp cleaning cloth; for more severe contamination, use a mild soap. ttClean the operator display with a slightly damp, lint-free cleaning cloth; for more severe contamination, use a mild soap. Be sure not to scratch the device or operating display. Future contamination is otherwise more difficult to remove. 190 Operating Manual BIOSTAT ® B Cleaning and Maintenance 10.1.2 Cleaning the Culture Vessels In some cases, it may be sufficient to carefully rinse the culture vessels (UniVessel® Glass) with water. If the bioreactor is not used for a while, always fill the culture vessels with water, as this will protect the integrated sensors from drying out. Thorough cleaning is required if components of the culture or media adhere to the inside surfaces of the culture vessels and components installed. −− Culture vessels and glass containers can be cleaned in a dishwasher. Before cleaning culture vessels inside a dishwasher, always remove the support frame, the cover plate and the vessel fittings. −− Glass surfaces that are contaminated with organic substances can be cleaned with conventional laboratory glass cleaners. To remove stubborn organic contaminants, you may use mechanical means. −− Inorganic deposits must be removed using diluted hydrochloric acid. Once these contaminants have been removed, carefully rinse the culture vessel with water. −− You can clean metallic parts (cover plate, etc.) mechanically, if needed with the help of mild detergents or alcohol. −− The seals and O-rings must be cleaned mechanically. If deposits on the seals and O-rings prove to be persistent, replace the seals and O-rings. Detailed information on cleaning culture vessels, vessel equipment and sensors can be found in the [ UniVessel® Glass Operating Manual]. 10.1.3 Cleaning and Maintaining Heating Jackets Danger of damage from unsuitable cleaning agents and cleaning procedures. Do not use any cleaning agents or solvents that can corrode the power supply, silicone foil or silicone foam and make them porous. Do not use any hard and | or sharp objects to remove stubborn soiling. The heating jackets are insensitive to water and the media used in conventional culture procedures. Test the resistance to the laboratory acids, alkaline agents and solvents used. ttClean soiled heating jacket carefully using only a wet cloth and warm water or mild soapy water. ttBefore each use, make sure to check that the following parts are in perfect condition: −− the power cord, especially where it connects to the heating jacket −− the silicone foil on the heating side, −− the silicone foam isolation −− the Velcro fasteners Operating Manual BIOSTAT ® B 191 Cleaning and Maintenance Possible Damage Danger to life caused by electric shock if heating jackets are defective! None of the parts should be porous, folded, kinked or chipped. The silicone foil should not be discolored. This is a sign of short circuiting due to broken heating coils or a defective power cord. −− In this case, switch out the heating and cooling jackets and discontinue its use. 4 2 2 1a 1b 3 Fig. 10-1: Picture of damage 1a Cracks, porosity of the cable connection 3 Short circuiting of the heating coils, evidenced by discoloration on the silicone foil 1b Cracks, porosity of the power cord 4 Cracks, porosity of the Velcro straps 2 Cracks, porosity of the silicone foil above the heating coils After using the heating blanket, only store it in a clean, dry location. Do not expose it to sunlight for extended periods. When in perfect condition, the heating jackets allow for safe heating of the culture vessels. Malfunction and dangerous operating states can occur if damage was overlooked during the pre-use check. Spare and Wearing Parts Heating jackets do not contain spare or wearing parts. If worn or defective, they must be replaced. 192 Operating Manual BIOSTAT ® B Cleaning and Maintenance 10.2 Maintenance 10.2.1 Carry out Maintenance Work on Function Elements Maintenance performed by the user is restricted to the following tasks: −− Maintaining the pH, pO2 or Redox sensor as per the manufacturer's | supplier's specifications. −− Checking and replacing parts subject to wear as well as disposables, e.g., glass vessels, filters, tubing and seals with identical components with the same specifications [ Spare parts list]. −− Replacing O-rings, seals, filters, hoses and disposable items, e.g. inoculation membranes. Detailed instructions on the maintenance of culture vessels, vessel equipment and sensors can be found in the [ UniVessel® Glass Operating Manual]. The internal modules of the device, and the safety equipment, pump modules, drive motors and stirrer shaft couplings, must only be serviced by qualified and correspondingly authorized service personnel. Any servicing instructions for internal equipment, electrical modules and safety equipment contained in this manual and the technical documentation must be forwarded to the servicing personnel. Please return the device to Sartorius Service if it is defective. Observe the Decontamination Declaration. 10.2.2 Safety Component Maintenance Non-return Valve The wastewater outlet in the temperature control module consists of a check valve [ P&I Diagram]. This valve protects the system from excessive pressure buildup if the water feed is accidentally connected to the outlet of the thermostat system, if water dams up, or if water flows back into the supply unit from the outlet. Non-return valves that are defective must be replaced. Fig. 10-2: Non-return valve If the pressure in the temperature control circuit is too high, the culture vessels can burst. In jacketed glass vessels, this can cause the wall to burst. Non-return valves are only designed for assuring the direction of flow. They must not be used as safety valves. If a sealed external cooling circuit has been connected to the system, it must be ensured that this circuit operates at zero-pressure. The non-return valve’s function must be checked before the bioreactor is commissioned and thereafter checked once a year. The function test and replacement of the non-return valve, where necessary, is carried out by Sartorius Service. Operating Manual BIOSTAT ® B 193 Cleaning and Maintenance 10.2.3 Maintenance Intervals The cyclical maintenance of the device depends on its service life. The following table lists the maintenance intervals as they are assigned to the components: Before every process After 10-20 autoclave cycles Component If unsterile Activity 1 + yearly Glass culture vessel Pressure hold test Leaktightness Test x Gas leak test Visual inspection x Leak test Temperature control system Visual inspection x Replace x Visual inspection, replace if needed x Replace Leak test x Control Unit Connections to the culture vessel, air and water Tapping septa O-rings x x Air inlet and exhaust filters Filter elements Integrity test Replace x x x x Holding bottles Sampling bottles Visual inspection, replace if needed Gaskets, ventilation filters Replace x x x Rotating mechanical seal Check for contamination and damage Visual inspection x Visual inspection, replace if needed x pH probe Calibration, visual inspection for damage x pO2 probe Calibration, visual inspection for damage x Peristaltic pumps Pump hoses Probes Membrane body, electrolyte Visual inspection, (Clark probes) replace if needed 194 Operating Manual BIOSTAT ® B x Cleaning and Maintenance Before every process After 10-20 autoclave cycles Component If unsterile Activity 1 + yearly Sensor cap (optical O2 probe) x Foam probe Calibration, visual inspection for damage x Level probe Calibration, visual inspection for damage x Temperature sensors Calibration, visual inspection for damage x Plugs, contacts, lines x Visual inspection x Maintenance according to maintenance schedule Only to be carried out by Maintenance and Sartorius experts. functional test according to Please contact Sartorius maintenance report Service. x Operating Manual BIOSTAT ® B 195 Storage 11. Storage If the device is not set up immediately after delivery or temporarily not used, it must be stored under the conditions listed in the Chapter “13.6 Ambient Conditions“. Only store the device in dry buildings and do not leave the device outdoors. In case of improper storage, no liability will be assumed for resulting damage. 196 Operating Manual BIOSTAT ® B Disposal 12. Disposal 12.1 General Notes Packaging The packaging is made of environmentally friendly materials that can be used as secondary raw materials. If the packaging is no longer needed, it can be disposed of by local waste disposal authorities. Device The equipment, including accessories and empty non-rechargeable and rechargeable batteries, does not belong in your regular household waste; this equipment is manufactured from high-grade materials which can be recycled and reused. European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) requires that electrical and electronic equipment be collected and disposed of separately from other unsorted municipal waste, with the aim of recycling it. The crossed-out waste bin symbol indicates that separate collection is required. In Germany and several other countries, Sartorius itself assumes responsibility for the return and legally compliant disposal of its electronic and electrical products. These products may not be placed with household waste or brought to collection centers run by local public disposal operations – not even by small commercial operators. Please contact the Sartorius Service Center. In countries that are not members of the European Economic Area (EEA) or where no Sartorius subsidiaries or dealerships are located, please contact your local authorities or a commercial disposal operator. Prior to disposal and | or scrapping of the equipment, any batteries should be removed and disposed of at local collection points. Sartorius will not take back equipment contaminated with hazardous materials (ABC contamination) – either for repair or disposal. Addresses for Disposal Detailed information with service addresses for disposal of your device can be found on our website (www.sartorius.com). 12.2 Hazardous Materials The device does contain any hazardous materials that would necessitate special disposal measures. The cultures and media (e.g. acids, bases) used during the fermentation processes are potentially hazardous materials that could cause biological or chemical hazards. Note in accordance with European directive on hazardous substances! According to the EU directives, the owners of devices that come into contact with hazardous substances are responsible for properly disposing of these devices and to declare such devices when transporting them. Operating Manual BIOSTAT ® B 197 Disposal Corrosion When using corrosive gases, the fittings must be chosen accordingly, (e.g. made of stainless steel instead of brass). Please contact Sartorius Service for retrofitting. We do not accept any responsibility for operating faults and defects resulting from the use of unsuitable gases. 12.3 Decontamination Declaration Danger of injury from improperly performed work! The disassembly and disposal of the unit may only be carried out by technical personnel. Warning of dangerous electrical voltage! Work on the electrical equipment may only be carried out by a competent electrician. Carry out the following work steps for the disassembly of the unit: ttEmpty the culture vessel, pipelines and hoses of all culture media and additives. ttClean the entire unit. ttSterilize the entire unit. ttTurn the unit off via the unit’s main switch and secure the unit against being turned back on. ttDisconnect the unit from power and the supply lines. 12.4 Disposing of the Unit Danger of severe injury due to ejected or falling parts! When disassembling the unit, pay particular attention to those components that contain parts under mechanical tension that could spring out during scrapping, leading to injury. There is also danger due to moving parts and falling objects. −− The device may only be disassembled by technical personnel. −− Disassemble the unit carefully and in a safety-conscious manner. −− Wear the following personal protective equipment during work [ see also Chapter “2.15 Personal Protective Equipment“]: −− Safety gloves −− Protective work clothes −− Safety boots −− Safety glasses. ttDisassemble the unit until all unit parts have been assigned to a material group and can be appropriately disposed of. ttDispose of the device in an environmentally friendly manner. Follow the regulations applicable in the local region. 198 Operating Manual BIOSTAT ® B Specifications 13. Specifications 13.1 BIOSTAT ® B Specification Version BIOSTAT ® B-MO (microbial) BIOSTAT ® B-CC (cell culture) Housing Display Touch screen, 12”, glass, capacitive Resolution SCADA communication Potential-free alarm Value Stainless steel AISI 304 125 dpi Industrial Internet 13.1.1 Dimensions and Weights Specification Dimensions Width × height × depth Supply unit weight BIOSTAT ® B Single | Twin (Weight depends on version) Weight of the culture vessels UniVessel® Glass 1 L DW/SW UniVessel® Glass 2 L DW/SW UniVessel® Glass 5 L DW/SW UniVessel® Glass 10 L DW/SW UniVessel® 2 L SU without vessel holder UniVessel® 2 L SU with vessel holder Weight of the RM Rocker BIOSTAT ® RM 20 | 50 Rocker (including culture bag holder) BIOSTAT ® RM 200 Rocker Unit Value mm 350 × 822 × 430 kg approx. 40 - 55 kg kg kg kg kg kg approx. 10 approx. 14 approx. 20 approx. 34 approx. 1.5 approx. 15 kg approx. 30-40 kg approx. 197 Unit Value 13.1.2 Safety Valves and Pressure Reducer Specification Gas pressure safety valve for UniVessel® Glass | SU for BIOSTAT ® RM Rocker Water input pressure reduction barg (psig) 1.0 (14.5) barg (psig) 0.1 (1.45) barg (psig) 1.5 (21.76) Operating Manual BIOSTAT ® B 199 Specifications 13.1.3 Culture Vessels and Culture Bags Specification Operating volume | max. total volume UniVessel® Glass UniVessel® Glass 1 L UniVessel® Glass 2 L UniVessel® Glass 5 L UniVessel® Glass 10 L Material: Steel | glass UniVessel® SU UniVessel® SU 2 L BIOSTAT ® RM 20 | 50 Rocker Flexsafe® RM 1L Flexsafe® RM 2L Flexsafe® RM 10L Flexsafe® RM 20L Flexsafe® RM 50L BIOSTAT ® RM 200 Rocker Flexsafe® RM 100L Flexsafe® RM 200L 200 Operating Manual BIOSTAT ® B Unit Value L L L L 1 / 1.5 2/3 5 / 6.6 10 / 13 L 2 / 2.7 L L L L L 0.5 / 1 1/2 5 / 10 10 / 20 25 / 50 L L 50 / 100 100 / 200 Specifications 13.2 Energy Connections Inside the Laboratory 13.2.1 Electrical Specifications Specification Power connection for a 230 V supply unit Voltage Frequency Power consumption Power connection for a 120 V supply unit Voltage Frequency Power consumption Protection class Unit Value V Hz A 230 (± 10%) 50 10 V Hz A IP 120 (± 10%) 60 12 21 Unit Value 13.2.2 Process Gas Supply Specification Supply pressure* Compressed air [AIR], preset to O2, preset to N2, preset to CO2, preset to Supply rate* Compressed air [AIR], max O2, max N2, max CO2, max Connection: Nozzle | outer diameter barg (psig) 1.5 (21.76) barg (psig) 1.5 (21.76) barg (psig) 1.5 (21.76) barg (psig) 1.5 (21.76) L/min L/min L/min L/min mm 20 20 20 20 6 * All gases dry and free of particles 13.2.3 Cooling Water Supply Specification Supply rate, min. Supply pressure Water hardness, max. Temperature, min. Connection: Nozzle | outer diameter Depressurized outlet Water quality: clean water, free of particles Unit Value L/min 10 barg (psig) 2-8 (29-116) °dH 12 °C +4 mm 10 13.3 Temperatures Specification Operating temperatures, max. Operating temperatures, min. (cooling water) Unit °C °C Value +80 +8 Operating Manual BIOSTAT ® B 201 Specifications 13.4 Stirrer Driver Specification Motor 1 L, 2 L culture vessel 2 L culture vessel UniVessel® SU 5 L bioreactor 10 L bioreactor Unit W 1/min 1/min Value 200 20-2000 20-400 1/min 1/min 20-1500 20-800 Unit Value 1:100 13.5 External Pumps Specification Control range when using the BIOSTAT ® B control unit 13.6 Ambient Conditions Ambient Conditions Installation location: Conventional laboratory rooms, max. 2,000 m above sea level Ambient temperatures of between [°C]: 5 – 40 Relative humidity [%]: < 80% for temperatures up to 31 °C (87.8 °F), decreasing linearly < 50% at 40 °C (104 °F) Impurities: Pollution degree 2 (Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must be expected) Acoustic emission [dB (A)]: Max. sound pressure level < 70 13.7 Water Hardness Conversion Table If customer-specific modifications were implemented, the appropriate documents can either be integrated into the “Technical Documentation“ folder or they can be delivered together with the bioreactor as separate documents. 202 Operating Manual BIOSTAT ® B Alkaline Alkaline German CaCO3 earth ions earth ions hardness English French hardness hardness [mmol/l] [mval/l] [°dH] [ppm] [°eH] [°fH] 1 mmol/l 1.00 alkaline earth ions 2.00 5.50 100.00 7.02 10.00 1 mval/l 0.50 alkaline earth ions 1.00 2.80 50.00 3.51 5 1° German hardness 0.18 [°dH] 0.357 1.00 17.80 1.25 1.78 1 ppm CaCO3 0.01 0.020 0.056 1.00 0.0702 0.10 1° English hardness 0.14 [°e] 0.285 0.798 14.30 1.00 1.43 1° French hardness 0.10 [°fH] 0.200 0.560 10.00 0.702 1.00 Conformity & Licenses 14. Conformity & Licenses 14.1 EC Declaration of Conformity The attached declaration of conformity (see Page 204) hereby confirms compliance of the devices BIOSTAT ® B-MO und BIOSTAT ®B-CC with the directives cited. 14.2 GNU Licensing DCU systems contain software subject to the license terms of the “GNU General Public License (GPL)” or the “GNU LESSER General Public License (LGPL)”. If applicable, the terms of the GPL and LGPL as well as information about the options for access to GPL code and LGPL code used in this product are available upon request. The GPL code and LGPL code contained in this product are published without any guarantee and subject to the copyright of one or more authors. You can find detailed information in the documentation about the enclosed LGPL code and in the GPL and LGPL terms and conditions. Operating Manual BIOSTAT ® B 203 Conformity & Licenses X 204 Operating Manual BIOSTAT ® B Appendix 15. Appendix 15.1 Service Repairs may be performed by authorized service personnel or by the responsible service representative. Please contact Sartorius Service for all service needs, and in the event of warranty claims. Returning Devices You can send defective devices or parts to Sartorius. Returned devices must be clean and in hygienically flawless condition and packed carefully. Transport damage as well as measures forsubsequent cleaning and disinfection of the parts by Sartorius shall be charged to the sender. Service Addresses Detailed information with service addresses for returning your device for repair can be found on our website (www.sartorius.com). 15.2 Decontamination Declaration When returning equipment, copy the following form as required, carefully complete it and enclose it with the delivery documents. The recipient must be able to inspect the completed declaration before removing the device from the packaging. Operating Manual BIOSTAT ® B 205 Appendix Decontamination Declaration Declaration Concerning the Decontamination and Cleaning of Equipment and Components To protect our personnel, we require that all equipment or components which come into contact with our personnel at customers' facilities be free of biological, chemical or radioactive contaminants. Therefore, we can only take an order if: −− the devices and components have been adequately CLEANED and DECONTAMINATED. −− this declaration has been filled out, signed and returned to us. We ask for your understanding of our measures to ensure a safe and non-hazardous work environment for our employees. Description of the Devices and Components Description | cat. no.: Serial no.: No. of invoice | delivery note: Delivery date: Contamination | Cleaning Attention: Please specify exactly the biological, che- Attention: Please describe the cleaning and mical or radioactive contaminant decontamination procedure | method The equipment was contaminated with It has been cleaned and decontaminated by: Legally Binding Declaration I | we hereby certify that the information provided on this form is true and complete. The equipment and components have been adequately decontaminated and cleaned according to the legal requirements. No chemical or biological or radioactive risks remain that could endanger exposed persons' safety or health. Company | institute: Address | country: Phone: Name of the authorized person: Position: Date | signature: Fax: Please pack the equipment properly and send it ex recipient to your Sartorius Service. 206 Operating Manual BIOSTAT ® B Appendix 15.3 Setup Drawings On the following pages, you will find setup drawings for the following configurations: −− BIOSTAT ® B, Single with UniVessel® Glass, DW (Page 208) −− BIOSTAT ® B, Single with BIOSTAT ® RM 20 | 50 Rocker (Page 209) −− BIOSTAT ® B, Single with UniVessel® SU, SW (Page 210) −− BIOSTAT ® B, Twin with UniVessel® Glass, DW (Page 211) −− BIOSTAT ® B, Twin with BIOSTAT ® RM 20 | 50 Rocker (Page 212) −− BIOSTAT ® B, Twin with UniVessel® SU, SW (Page 213) −− BIOSTAT ® B, Twin with UniVessel® SU, SW | UniVessel® Glass, DW (Page 214) −− BIOSTAT ® B, Twin with BIOSTAT ® RM 200 Rocker (Page 215) Operating Manual BIOSTAT ® B 207 Operating Manual BIOSTAT ® B A B 16 15 14 1L UniVessel DW 13 Supply Unit Twin 12 Wasseranschluß Rücklauf Ø10mm Water return Connection Ø10mm Wasseranschluß Zulauf Ø10mm Water Supply Connection Ø10mm Netzanschluß Main Connection PE-Anschluß PE-Connection 12 11 11 507 9 10 9 Thermostat Rücklauf / Serto Verschraubung Ø10mm Thermostate Return / Serto Connection Ø10mm Thermostat Zulauf / Serto Verschraubung Ø10mm Thermostate Supply / Serto Connection Ø10mm Abluft Rücklauf / Serto Verschraubung Ø10mm Exhaust Return / Serto Connection Ø10mm Abluft Zulauf / Serto Verschraubung Ø10mm Abluft Supply / Serto Connection Ø10mm Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection° Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection Trübung / Lemo-Stecker Turbidity / Lemo-Connection LEVEL / M12 Steckanschluß LEVEL / M12 plug-in Connection FOAM / M12 Steckanschluß FOAM / M12 plug-in Connection Sparger Zulauf / Serto Verschraubung Ø6mm Sparger Supply / Serto Connection Ø6mm 10 507 C Gas Anschluß CO2 / Serto-Verschraubung Ø6mm Gas Connection CO2 / Serto Connection Ø6mm Gas Anschluß N2 / Serto-Verschraubung Ø6mm Gas Connection N2 / Serto Connection Ø6mm Gas Anschluß O2 / Serto-Verschraubung Ø6mm Gas Connection O2 / Serto Connection Ø6mm A(1:3) 13 8 Sensorfeld Sensor panel 8 800 7 Heizmanschette / Amphenol Stecker Heating Blanket / Amphenol Plug-In Connection Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection pO2 / VP8 Stecker pO2 / VP8 Connection pH / VP8 Stecker pH / VP8 Connection Temp / M12 Steckanschluß Temp / M12 plug-in Connection Overlay Zulauf / Serto Verschraubung Ø6mm Overlay Supply / Serto Connection Ø6mm 7 800 D Ethernet Host Ethernet Host 14 Common Alarm Anschluß Common Alarm Connection Gas Anschluß AIR / Serto-Verschraubung Ø6mm Gas Connection AIR / Serto Connection Ø6mm 15 1496 E F G H I J K 16 6 1496 6 1496 208 5 Ansicht A 5 1200 L 4 Ablagebox Utilitx Box 4 800 800 Index 3 Änderungen/Revision Datum/Date Name Allowable tolerances unless otherwise specified according to EN ISO 13920-A/-AE Zul. Abweichungen für Maße ohne Toleranzangaben nach EN ISO 13920-A/-AE 3 1 Datum/Date Ursprung/Origin 28.03.2012 2 TKösters Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1028702 Dokument-Nr./Document-no. 1L Univessel MU 1 1 /1 Blatt/Sheet Maßstab/Scale Arrangemanet Plan BIOSTAT B, single, dw Aufstellungsplan BIOSTAT B, Single, DW Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 2 A B C D E F G H I J K L Appendix 1200 1622 1882 1622 1882 Operating Manual BIOSTAT ® B A B 16 16 15 Thermostate return / Serto connection 10mm Thermostat Rücklauf / Serto-Verschraubung 10mm Thermostate supply / Serto connection 10mm Thermostat Zulauf / Serto-Verschraubung 10mm Ext. Signal / M12 Steckanschluss Ext. signal / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Overlay supply / Serto connection 6mm Overlay Zulauf / Serto-Verschraubung 6mm 15 14 Sensorfeld für RM Sensor panel for RM 14 13 Pumpe / M12 Steckanschluss Pump / M12 plug-in connection Pumpe / M12 Steckanschluss Pump / M12 plug-in connection Optischer Stecker PreSens Visual connector PreSens Optischer Stecker PreSens Visual connector PreSens 13 12 12 11 11 RM 20/50 10 10 1344 1344 1600 1600 9 9 8 Supply Unit Single 8 7 810 7 810 C D E F G H I J K L 6 6 Common Alarm Anschluss Common alarm connection Ablagebox Utility box 5 4 Netzwerkanschluss Ethernet hist 4 Gas connection CO2 / Serto Connection 6mm Gas Anschluss CO2 / Serto-Verschraubung 6mm Gas connection N2 / Serto Connection 6mm Gas Anschluss N2 / Serto-Verschraubung 6mm Gas connection O2 / Serto Connection 6mm Gas Anschluss O2 / Serto-Verschraubung 6mm Gas connection AIR / Serto Connection 6mm Gas Anschluss AIR / Serto-Verschraubung 6mm A 5 Index Water supply connection 10mm Wasseranschluss Zulauf 10mm 3 Datum/Date Name Ursprung/Origin 13.08.2013 Datum/Date Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 1 2 ABernhard Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1033224 Dokument-Nr./Document-no. RM 20/50 1 1 /1 Blatt/Sheet Maßstab/Scale Arrangement plan BIOSTAT B, Single Aufstellungsplan BIOSTAT B, Single Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH Water return connection 10mm Allowable tolerances unless otherwise specified according to ISO 2768-mH Änderungen/Revision Netzanschluss Main connection 2 Wasseranschluss Rücklauf 10mm Zul. Abweichungen für Maße ohne Toleranzangaben nach ISO 2768-mH PE-Anschluss PE-Connection A(1:3) 3 A B C D E F G H I J K L Appendix 1632 1504 800 800 1882 209 1632 1504 1882 210 Operating Manual BIOSTAT ® B A 16 15 14 13 12 11 10 10 604 9 Thermostat Rücklauf / Serto Verschraubung Ø10mm Thermostate Return / Serto Connection Ø10mm Thermostat Zulauf / Serto Verschraubung Ø10mm Thermostate Supply / Serto Connection Ø10mm Abluft Rücklauf / Serto Verschraubung Ø10mm Exhaust Return / Serto Connection Ø10mm Abluft Zulauf / Serto Verschraubung Ø10mm Abluft Supply / Serto Connection Ø10mm Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection° Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection Trübung / Lemo-Stecker Turbidity / Lemo-Connection LEVEL / M12 Steckanschluß LEVEL / M12 plug-in Connection FOAM / M12 Steckanschluß FOAM / M12 plug-in Connection Sparger Zulauf / Serto Verschraubung Ø6mm Sparger Supply / Serto Connection Ø6mm 9 8 Sensorfeld Sensor panel 8 1492 800 6 7 6 Heizmanschette / Amphenol Stecker Heating Blanket / Amphenol Plug-In Connection Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection pO2 / VP8 Stecker pO2 / VP8 Connection pH / VP8 Stecker pH / VP8 Connection Temp / M12 Steckanschluß Temp / M12 plug-in Connection Overlay Zulauf / Serto Verschraubung Ø6mm Overlay Supply / Serto Connection Ø6mm 7 1622 1622 B 11 604 C Wasseranschluß Zulauf Ø10mm Water Supply Connection Ø10mm Wasseranschluß Zulauf Ø10mm Water Supply Connection Ø10mm 12 Supply Unit Single Netzanschluß Main Connection PE-Anschluß PE-Connection 13 2L UniVessel single Use A(1:3) 14 800 D Gas Anschluß CO2 / Serto-Verschraubung Ø6mm Gas Connection CO2 / Serto Connection Ø6mm Gas Anschluß N2 / Serto-Verschraubung Ø6mm Gas Connection N2 / Serto Connection Ø6mm Gas Anschluß O2 / Serto-Verschraubung Ø6mm Gas Connection O2 / Serto Connection Ø6mm Gas Anschluß AIR / Serto-Verschraubung Ø6mm Gas Connection AIR / Serto Connection Ø6mm Common Alarm Anschluß Common Alarm Connection 15 1492 E F G H I J K 16 5 5 Ansicht A 1200 L 4 Ablagebox Utilitx Box 4 800 800 Index 3 Änderungen/Revision Datum/Date Name Allowable tolerances unless otherwise specified according to EN ISO 13920-A/-AE Zul. Abweichungen für Maße ohne Toleranzangaben nach EN ISO 13920-A/-AE 3 1 Datum/Date Ursprung/Origin 28.03.2012 2 TKösters Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1028689 Dokument-Nr./Document-no. 2L UniVessel SU 1 1 /1 Blatt/Sheet Maßstab/Scale Arrangemanet Plan BIOSTAT B, single, sw Aufstellungsplan BIOSTAT B, Single, SW Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 2 A B C D E F G H I J K L Appendix 1200 1882 1882 Operating Manual BIOSTAT ® B A B C 15 Gas Anschluß CO2 / Serto-Verschraubung Ø6mm Gas Connection CO2 / Serto Connection Ø6 Gas Anschluß N2 / Serto-Verschraubung Ø6mm Gas Connection N2 / Serto Connection Ø6 Gas Anschluß O2 / Serto-Verschraubung Ø6mm Gas Connection O2 / Serto Connection Ø6 16 Ethernet Host Ethernet Host 14 14 2L UniVessel DW Common Alarm Anschluß Common Alarm Connection Gas Anschluß AIR / Serto-Verschraubung Ø6mm Gas Connection AIR / Serto Connection Ø6 15 13 Supply Unit Twin A(1:3) 13 12 2L UniVessel DW Wasseranschluß Rücklauf Ø10mm Water return Connection Ø10mm Wasseranschluß Zulauf Ø10mm Water Supply Connection Ø10mm Netzanschluß Main Connection PE-Anschluß PE-Connection 12 11 11 9 10 9 Thermostat Rücklauf / Serto Verschraubung Ø10mm Thermostate Return / Serto Connection Ø10mm Thermostat Zulauf / Serto Verschraubung Ø10mm Thermostate Supply / Serto Connection Ø10mm Abluft Rücklauf / Serto Verschraubung Ø10mm Exhaust Return / Serto Connection Ø10mm Abluft Zulauf / Serto Verschraubung Ø10mm Abluft Supply / Serto Connection Ø10mm Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection° Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection Trübung / Lemo-Stecker Turbidity / Lemo-Connection LEVEL / M12 Steckanschluß LEVEL / M12 plug-in Connection FOAM / M12 Steckanschluß FOAM / M12 plug-in Connection Sparger Zulauf / Serto Verschraubung Ø6mm Sparger Supply / Serto Connection Ø6mm 10 8 Sensorfeld Sensor panel 8 7 Heizmanschette / Amphenol Stecker Heating Blanket / Amphenol Plug-In Connection Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection pO2 / VP8 Stecker pO2 / VP8 Connection pH / VP8 Stecker pH / VP8 Connection Temp / M12 Steckanschluß Temp / M12 plug-in Connection Overlay Zulauf / Serto Verschraubung Ø6mm Overlay Supply / Serto Connection Ø6mm 7 507 800 D E F G H I J K 16 507 800 L 6 1496 1496 6 5 1200 1200 5 Ablagebox Utilitx Box 4 4 800 800 Index 3 Änderungen/Revision Datum/Date Name Allowable tolerances unless otherwise specified according to EN ISO 13920-A/-AE Zul. Abweichungen für Maße ohne Toleranzangaben nach EN ISO 13920-A/-AE 3 Ursprung/Origin 26.03.2012 Datum/Date Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 1 2 TKösters Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1028603 Dokument-Nr./Document-no. 1L Univessel MU 1 1 /1 Blatt/Sheet Maßstab/Scale Arrangemanet Plan BIOSTAT B, twin Aufstellungsplan BIOSTAT B, Twin Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH 2 A B C D E F G H I J K L Appendix 1622 1882 211 1622 1882 212 Operating Manual BIOSTAT ® B A B C D E 16 15 Thermostate return / Serto connection 10mm Thermostat Rücklauf / Serto-Verschraubung 10mm Thermostate supply / Serto connection 10mm Thermostat Zulauf / Serto-Verschraubung 10mm Ext. Signal / M12 Steckanschluss Ext. signal / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Overlay supply / Serto connection 6mm Overlay Zulauf / Serto-Verschraubung 6mm 15 14 Sensorfeld für RM Sensor panel for RM 14 1882 Pumpe / M12 Steckanschluss Pump / M12 plug-in connection Pumpe / M12 Steckanschluss Pump / M12 plug-in connection Optischer Stecker PreSens Visual connector PreSens Optischer Stecker PreSens Visual connector PreSens 1882 F G H I J K 16 13 13 12 12 11 RM 20/50 11 10 10 9 2200 2335 2335 2200 Supply Unit Twin 9 8 8 7 7 RM 20/50 6 6 1506 1506 800 L 5 5 A 4 Index Gas connection CO2 / Serto Connection 6mm Gas Anschluss CO2 / Serto-Verschraubung 6mm Gas connection N2 / Serto Connection 6mm 3 A ( 1:3 ) 3 Änderungen/Revision Datum/Date Ursprung/Origin 14.08.2013 Datum/Date Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 2 Water return connection 10mm Wasseranschluss Rücklauf 10mm Water supply connection 10mm Wasseranschluss Zulauf 10mm Netzanschluss Main connection 1 2 ABernhard Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1033253 Dokument-Nr./Document-no. 1 1 /1 Blatt/Sheet Maßstab/Scale RM 20/50 - RM 20/50 Arrangement plan BIOSTAT B, Twin Aufstellungsplan BIOSTAT B, Twin Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH PE-Anschluss PE-Connection Name Allowable tolerances unless otherwise specified according to ISO 2768-mH Zul. Abweichungen für Maße ohne Toleranzangaben nach ISO 2768-mH Netzwerkanschluss Ethernet hist Gas Anschluss N2 / Serto-Verschraubung 6mm Gas connection O2 / Serto Connection 6mm Gas Anschluss O2 / Serto-Verschraubung 6mm Gas connection AIR / Serto Connection 6mm Gas Anschluss AIR / Serto-Verschraubung 6mm Common Alarm Anschluss Common alarm connection Ablagebox Utility box 4 A B C D E F G H I J K L Appendix 1632 810 800 1632 810 Operating Manual BIOSTAT ® B A B C D 16 15 Gas Anschluß CO2 / Serto-Verschraubung Ø6mm Gas Connection CO2 / Serto Connection Ø6 Gas Anschluß N2 / Serto-Verschraubung Ø6mm Gas Connection N2 / Serto Connection Ø6 Gas Anschluß O2 / Serto-Verschraubung Ø6mm Gas Connection O2 / Serto Connection Ø6 Gas Anschluß AIR / Serto-Verschraubung Ø6mm Gas Connection AIR / Serto Connection Ø6 A(1:3) 14 2L UniVessel single Use 14 13 12 12 2L UniVessel single Use Wasseranschluß Rücklauf Ø10mm Water return Connection Ø10mm Wasseranschluß Zulauf Ø10mm Water Supply Connection Ø10mm Netzanschluß Main Connection Supply Unit Twin PE-Anschluß PE-Connection 13 11 11 10 10 9 Thermostat Rücklauf / Serto Verschraubung Ø10mm Thermostate Return / Serto Connection Ø10mm Thermostat Zulauf / Serto Verschraubung Ø10mm Thermostate Supply / Serto Connection Ø10mm Abluft Rücklauf / Serto Verschraubung Ø10mm Exhaust Return / Serto Connection Ø10mm Abluft Zulauf / Serto Verschraubung Ø10mm Abluft Supply / Serto Connection Ø10mm Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection° Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection Trübung / Lemo-Stecker Turbidity / Lemo-Connection LEVEL / M12 Steckanschluß LEVEL / M12 plug-in Connection FOAM / M12 Steckanschluß FOAM / M12 plug-in Connection Sparger Zulauf / Serto Verschraubung Ø6mm Sparger Supply / Serto Connection Ø6mm 9 8 Sensorfeld Sensor panel 8 1622 6 7 6 Heizmanschette / Amphenol Stecker Heating Blanket / Amphenol Plug-In Connection Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection pO2 / VP8 Stecker pO2 / VP8 Connection pH / VP8 Stecker pH / VP8 Connection Temp / M12 Steckanschluß Temp / M12 plug-in Connection Overlay Zulauf / Serto Verschraubung Ø6mm Overlay Supply / Serto Connection Ø6mm 7 604 800 E F G Ethernet Host Ethernet Host Common Alarm Anschluß Common Alarm Connection 15 1496 1622 1496 H I J K 16 604 800 L 5 5 Ablagebox Utilitix Box 4 4 800 800 Index 3 Änderungen/Revision Datum/Date Name Allowable tolerances unless otherwise specified according to EN ISO 13920-A/-AE Zul. Abweichungen für Maße ohne Toleranzangaben nach EN ISO 13920-A/-AE 3 Ursprung/Origin 26.03.2012 Datum/Date Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 1 2 TKösters Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1028579 Dokument-Nr./Document-no. 2L UniVessel SU 1 1 /1 Blatt/Sheet Maßstab/Scale Arrangemanet Plan BIOSTAT B, twin Aufstellungsplan BIOSTAT B, Twin Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH 2 A B C D E F G H I J K L Appendix 1200 1882 213 1200 1882 214 Operating Manual BIOSTAT ® B A 16 15 14 13 12 1881 11 2L UniVessel single Use 11 9 Abluft Rücklauf / Serto Verschraubung Ø10mm Exhaust Return / Serto Connection Ø10mm Abluft Zulauf / Serto Verschraubung Ø10mm Abluft Supply / Serto Connection Ø10mm Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection° Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection Trübung / Lemo-Stecker Turbidity / Lemo-Connection LEVEL / M12 Steckanschluß LEVEL / M12 plug-in Connection FOAM / M12 Steckanschluß FOAM / M12 plug-in Connection Sparger Zulauf / Serto Verschraubung Ø6mm Sparger Supply / Serto Connection Ø6mm 10 507 10 9 Thermostat Rücklauf / Serto Verschraubung Ø10mm Thermostate Return / Serto Connection Ø10mm Thermostat Zulauf / Serto Verschraubung Ø10mm Thermostate Supply / Serto Connection Ø10mm 507 B C Supply Unit Twin Wasseranschluß Rücklauf Ø10mm Water return Connection Ø10mm Wasseranschluß Zulauf Ø10mm Water Supply Connection Ø10mm Netzanschluß Main Connection 12 1882 D 13 PE-Anschluß PE-Connection 2L UniVessel DW A(1:3) 14 8 Sensorfeld Sensor panel 8 1496 1622 7 Heizmanschette / Amphenol Stecker Heating Blanket / Amphenol Plug-In Connection Serial / M12 Steckanschluß Serial / M12 plug-in Connection Pumpe / M12 Steckanschluß Pump / M12 plug-in Connection Ext. Signal / M12 Steckanschluß Ext. Signals / M12 plug-in Connection pO2 / VP8 Stecker pO2 / VP8 Connection pH / VP8 Stecker pH / VP8 Connection Temp / M12 Steckanschluß Temp / M12 plug-in Connection Overlay Zulauf / Serto Verschraubung Ø6mm Overlay Supply / Serto Connection Ø6mm 7 604 800 E Gas Anschluß CO2 / Serto-Verschraubung Ø6mm Gas Connection CO2 / Serto Connection Ø6 Gas Anschluß N2 / Serto-Verschraubung Ø6mm Gas Connection N2 / Serto Connection Ø6 Gas Anschluß O2 / Serto-Verschraubung Ø6mm Gas Connection O2 / Serto Connection Ø6 Gas Anschluß AIR / Serto-Verschraubung Ø6mm Gas Connection AIR / Serto Connection Ø6 Common Alarm Anschluß Common Alarm Connection Ethernet Host Ethernet Host 15 1622 1496 F G H I J K 16 604 800 L 6 6 A 5 Ablagebox Utilitx Box 5 800 800 4 4 Index 3 Änderungen/Revision Datum/Date Name Allowable tolerances unless otherwise specified according to EN ISO 13920-A/-AE Zul. Abweichungen für Maße ohne Toleranzangaben nach EN ISO 13920-A/-AE 3 1 Datum/Date Ursprung/Origin 15.03.2012 2 AHalt Name Ers. f./Repl. for Artikel-Nr./Article-no. Ers. d./Repl. by DE-1028439 Dokument-Nr./Document-no. 1 1 /1 Blatt/Sheet Maßstab/Scale 2L UniVessel SU, 1L Univessel MU Arrangemanet Plan BIOSTAT B, twin Aufstellungsplan BIOSTAT B, Twin Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH Oberfläche/Finish: Bearbeiter Drawn Geprüft Checked 2 A B C D E F G H I J K L Appendix 1200 1200 Operating Manual BIOSTAT ® B A 1882 16 15 Thermostate return / Serto connection 10mm Thermostat Rücklauf / Serto-Verschraubung 10mm Thermostate supply / Serto connection 10mm Thermostat Zulauf / Serto-Verschraubung 10mm Ext. Signal / M12 Steckanschluss Ext. signal / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Serial / M12 Steckanschluss Serial / M12 plug-in connection Overlay supply / Serto connection 6mm Overlay Zulauf / Serto-Verschraubung 6mm 14 Sensorfeld für RM Sensor panel for RM Pumpe / M12 Steckanschluss Pump / M12 plug-in connection Pumpe / M12 Steckanschluss Pump / M12 plug-in connection Optischer Stecker PreSens Visual connector PreSens Optischer Stecker PreSens Visual connector PreSens 13 BIOSTAT B 13 900 1596 12 Schlauchhalter Hose support 12 800 11 D11 Flaschenhalter Bottle support 1882 B C D E F G 14 10 10 D(1:5) 9 RM 200 9 2883 8 8 1753 7 7 1622 1496 H I J 15 604 1154 K 16 6 6 A 5 5 1200 1080 L 4 Index 3 Datum/Date Name Allowable tolerances unless otherwise specified according to ISO 2768-mH Datum/Date 02.12.2014 Ursprung/Origin Bearbeiter Drawn Geprüft Checked 2 PE-Anschluss PE-Connection Water return connection 10mm Wasseranschluss Rücklauf 10mm Water supply connection 10mm Wasseranschluss Zulauf 10mm Netzanschluss Main connection 1 2 AHalt Name Ers. f./Repl. for Artikel-Nr./Article-no. RM 200 Ers. d./Repl. by DE-1038178 Dokument-Nr./Document-no. 1 1 /1 Blatt/Sheet Maßstab/Scale Arrangement plan BIOSTAT B/RM200 Aufstellungsplan BIOSTAT B/RM200 Für diese Zeichnung behalten wir uns alle Rechte vor / This drawing is the property of Sartorius Stedim Systems GmbH Oberfläche/Finish: A ( 0,40 : 1 ) Zul. Abweichungen für Maße ohne Toleranzangaben nach ISO 2768-mH 3 Änderungen/Revision Gas connection CO2 / Serto Connection 6mm Gas Anschluss CO2 / Serto-Verschraubung 6mm Gas connection N2 / Serto Connection 6mm Gas Anschluss N2 / Serto-Verschraubung 6mm Gas connection O2 / Serto Connection 6mm Gas Anschluss O2 / Serto-Verschraubung 6mm Gas connection AIR / Serto Connection 6mm Gas Anschluss AIR / Serto-Verschraubung 6mm Netzwerkanschluss Ethernet hist 800 Common Alarm Anschluss Common alarm connection 4 A B C D E F G H I J K L Appendix 800 215 1722 Sartorius Stedim Biotech GmbH August-Spindler-Strasse 11 37079 Goettingen, Germany Phone: +49.551.308.0 Fax: +49.551.308.3289 www.sartorius.com The information and figures contained in these instructions correspond to the version date specified below. Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice. Masculine or feminine forms are used to facilitate legibility in these instructions and always simultaneously denote the other gender as well. Copyright notice: This instruction manual, including all of its components, is protected by copyright. Any use beyond the limits of the copyright law is not permitted without our approval. This applies in particular to reprinting, translation and editing irrespective of the type of media used. © Sartorius Germany Last updated: 07 | 2017 Printed in the EU on paper bleached without chlorine. Publication No.: SBT6034-e170704