Site layout proposal
General
1
2
3
4
C-1
A-1
AL-1
AD-1
Cover sheet. Table of contents
Site layout
Site requirements / Equipment legend
Schematic cross section(s)
Site layout approval:
11-07-2023
Rev Date
VMI
KTH
Proposal
Drw
Chk
Proj. phase
Description
Cover sheet
Table of contents
0
Sheet Subject
Revision history
H
G
F
E
D
C
B
A
YE-23-00003
Typical configuration shown (not site specific)
Taiz, Yemen
CathLab Room
Subject
Azurion 3 F15
CardioVascular Center - Taiz
Type
Project
Sheet
Drawing
Important notes
Signature customer
The information in this package is provided as a customer convenience,
and is not to be construed as architectural drawings or
construction documents.
Philips assumes no liability nor offers any warranty for the fitness
or adequacy of the premises or the utilities available at the premises
in which the equipment is to be installed, used or stored.
Signature project manager
Philips contact
Mr. A. Mahir Sadi, +971 55 1003732, abdul.khalek.sadi@philips.com
Philips Healthcare
Customer ID
Technical ID
A3
A business of Royal Philips N.V.
Philips Medical Systems Nederland B.V. Business Center EMEA
Veenpluis 4-6, 5684 PC, Best, The Netherlands
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Quote No
Order No
© Koninklijke Philips Electronics N.V. 2020
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
Table of contents
mm
C-1
Sheet
1/4
0
1
2
3
5m
4
1m
Dimension origin
at finished wall,
near floor level.
(6550)
1.2
(3460)
(3460)
Y
3200
1.9
1.8 1.3 1.4 1.1
4.2
Examination area
CathLab Room
Control area
2300
2.4
2.5
3.3
3.2
3.4
3.6
4.3
2.6
2.2
1.6
4.2
1
Wall connection box transmiter
(WCBT)
Workstation Hemo Control room
To be decided by customer
Dimensioning legend
Existing doors/walls to be removed
Beams or other building construction elements
Columns (materials and strength unknown)
XYZ
System position dimension
(ISO-centre)
(XYZ)
Architectural dimension
to be checked on site
Preferred service area
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Dimension from origin
A3
Quote No
Order No
Technical ID
New walls
11-07-2023
VMI
Walls
YE-23-00003
0
Legend
Site layout
1.5
Technical area
UPS Room
Sheet Subject
1.7
3.5
Project
4.1
Taiz, Yemen
CathLab Room
3.7
X
(4500)
(4500)
3.1
Drawing
© Koninklijke Philips Electronics N.V. 2020
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
2.1
Azurion 3 F15
CardioVascular Center - Taiz
2.3
1 : 50
mm
A-1
Sheet
2/4
· The equipment proposed in this room layout produces X-ray radiation. It is the
customer's responsibility to comply with the local regulation. Normally X-ray shielding has
to be applied to avoid hazards to patients, visitors and staff. Consult the Philips contact for
details and support.
Remote service diagnostics
To establish this feature, a RJ45 type Ethernet connector must be
installed with access to the customer's network (10/100/1000 Mbps).
Environmental requirements
Operating temperature range is 15°C - 28°C.
at 20%-80% relative humidity (non-condensing).
In case of UPS/Battery in the technical room, operating temperature must be 15 - 25°C.
Azurion 3 F15
Resp No
Description
Mass
[kg]
Examination Area (Basic system excluding options)
Heat Noise
[W] [dB(A)]
1600
45-60
A
1.1
Patient table (AD7 swivel and tilt)
450
-
-
A
1.2
Clea floor (FD15)
1160
-
-
C
1.3
Monitor ceiling suspension 3 Fold
206
-
-
A
1.4
Auxiliary connection box
3
-
-
C
1.5
Fixed column 800mm
22
-
-
A
1.6
Examination Lamp mounted on fixed column
14
-
-
A
1.7
Radiation shield mounted on fixed column
20
-
-
A
1.8
Table mounted radiation shield
30
-
-
A
1.9
Injector Medrad Mark 7 Arterion
82
1200
-
570
<50
Control Area (Basic system excluding options)
A
2.1
Operating station Basic 2x 24" Monitor
52
-
-
B
2.2
Table top
-
-
-
A
2.3
Firewall Cisco ASA-5506
-
-
-
A
2.4
DVD Writer
-
-
-
A
2.5
Intercom
-
-
-
A
2.6
Workstation Hemo Control room
38
300
-
3700
68
Technical Area (Basic system excluding options)
A
3.1
Main cabinet with Cabinet Rear Cover Deep
375
-
-
A
3.2
Peripheral cabinet with Cabinet Rear cover
200
-
-
A
3.3
Generator cabinet with Cabinet Rear cover
145
-
-
A
3.4
Documentation cabinet (trolley)
80
-
-
B
3.5
Mains switch
-
-
-
B
3.6
Earth reference terminal
-
-
-
A
3.7
Isolation transformer Hemo
-
-
-
B
4.1
Lead glass window
-
-
-
B
4.2
X-ray ON warning light
-
-
-
A
4.3
Wall connection box transmitter (WCBT)
5
10
-
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Quote No
Order No
Technical ID
11-07-2023
VMI
YE-23-00003
Responsibilities
A Delivered and installed by Philips
B Delivered and installed by customer/contractor
C (Pre) Delivered by Philips, installed by customer
D (Pre) Delivered by customer and installed by Philips
E Existing
Drawing
© Koninklijke Philips Electronics N.V. 2020
0
Sheet Subject
Project
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
General
Taiz, Yemen
CathLab Room
Mains power
Main cabinet:
· Input Ratings for the M-cabinet (L1,L2,L3,PE)
3~380V ± 10 %, 80A...125A, 50/60Hz
3~400V ± 10 %, 76A...125A, 50/60Hz
3~415V ± 10 %, 73A...125A, 50/60Hz
3~440V ± 10 %, 69A...125A, 50/60Hz
3~480V ± 10 %, 63A...125A, 50/60Hz
· Hospital mains fuse: 63A-125A with gG characteristics (slow blow).
· Input wiring at least: 13.3mm², max. 67.4mm².
· Maximum required power: 100 kVA.
· Peak currents max. 8.33ms.
· All connected phase wires shall have an upstream disconnect switch
with a rating of <=125A.
· Mains supply impedance, requirements at mains switch:
380V , 50/60Hz, max.74mΩ
400V , 50/60Hz, max.140mΩ
415V , 50/60Hz, max.215mΩ
440V , 50/60Hz, max.325mΩ
480V , 50/60Hz, max.465mΩ
If the mains resistance can't be met, an additional transformer is required.
Azurion 3 F15
CardioVascular Center - Taiz
Site requirements (Rel. 2.1 >)
· The feasibility of this project is based on site information (like drawings) provided to
Philips. Philips cannot assume any liability for the accuracy of this information and the
consequences.
Site requirements
Equipment legend
Project notes
A3
mm
AL-1
Sheet
3/4
© Koninklijke Philips Electronics N.V. 2020
B
B
Situation (scale 1:100)
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
A3
Sheet
Quote No
Order No
Technical ID
A
Taiz, Yemen
CathLab Room
Azurion 3 F15
CardioVascular Center - Taiz
Project
Schematic cross section B-B
Schematic cross sections
Sheet Subject
Schematic cross section A-A
11-07-2023
VMI
0
4
YE-23-00003
3
(3100)
(3100)
2
2900
2900
1
Drawing
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
0
5m
1m
A
1 : 50
mm
AD-1
4/4