MR Imaging System Model MRP - 7000 TECHNICAL GUIDE ; w-w Installation & Adjustment --- HITACHI HMSA MEDICAL CORPORATION Tokyo, Japan .a L2E-BM0434-2 CONFIDENTIALITY : This material contains proprietary information of Hitachi Medical Corporation and it shall not be disclosed to the third party or reproduced without the prior written permission of Hitachi Medical Corporation. Copyright@ Hitachi Medical Corporation. 1991.Nl rights reserved. Strict Prohibitions During Installation, Adjustment, Maintenance and Inspection The inside of the MRI room is a strong magnetic field. So it is requested to strictly observe the following and provide a cautions marking at the entrance of the MRI room. 1. The servicing engineer carrying a magnetic substance such as an artificial joint, surgical clips, etc. should never be permitted to enter the MRI room. 2. Especially, any person using a pacemaker must be strictly prohibited to enter the MRI room. 3. The cash dispensing ID card, mechanical wristwatch, floppy disk, magnetic tape, etc. are affected by the magnetic field. It must be prohibited to carry them into the MRI room. 4. It should be strictly prohibited to use magnetic jigs and tools inside the MRI room. The jigs and tools to be used in "The following the room must be of a nonmagnetic material. may be used in the MRI, but more than lm off the gantry. (1) Impedance meter (2) Synchroscope -.- (1) 1 L2EBM0434 Page Contents Section I ", l-1 ,..'cly Procedures of Installation and Adjustment . Section II Carry-in, Installation and Assembling 2.1 Carrying-in, installation 2-l and assembling the gantry . . . . . . . . . . . . . 2.1.1 Carrying-in and installing gantry . . . . . 2-l 2.1.2 Detaching gantry protection plate . . . . . 2-5 2.1.3 Leveling and fixing gantry . . . . . . . . . 2-5 2.1.4 Assembling gantry . . . . . . . . . . . . . . . 2-6 2.1.5 Attaching gantry covers . . . . . . . . . . . 2-18 2.2 Installing patient table 2.2.1 2-26 . . . . . . . . . . . . . Positioning, leveling and fixing patient table . . . . . . . . . . 2-26 2.2.2 Attaching table covers .. ; . . . . . . . . . 2-28 2.2.3 Attaching sponge pad ............. 2-29 2.3 Installing central control console ....... 2-30 Notes for installation . . . . e . . . . . . . 2-30 Installation procedures . . . . . . . . . . . . 2.4 MFU unit . . . . . . . . . . . . . . . . . . . . . . . 2-35 2.3.1 2.3.2 2-44 2.4.1 MRI unit . . . . . . . . . . . . . . . . . . . . 2-44 2.4.2 Filter box . . . . . . . . . . . . . . . . . . . 2-44 W 2 L!zE-mm434 '3 Section III , ‘\ W i r i n g 3.1 Laying the cable between system components . . . . 3-l . . . 3-4 table . . . . . . . . . . . . . . . . . . . . . . . 3-4 3.2.2 Wiring central control console . . . . . . . . 3-7 3.2.3 Wiring MRI unit . . . . . . . . . . . . . . . . . 3-9 3.2.4 Wiring filter box . . . . . . . . . . . . . . . 3-11 3.2.5 Wiring to ECG monitor . . . . . . . . . . . . 3-13 4.1 Setting magnetic circuit temperature . . . . . . * 4-l . . . 4-4 ‘5 3.2 Connecting cable to each unit 3.2.1 Section IV . . . . . . . . Connecting cables to gantry and patient Adjustment 4.2 Switching on power supply . . . . . . . . . . . 4.3 Cperation check of central control console . . . 4-5 4.4 Operation check of thermostatic control circuit 4-6 .’ 4.4.1 4.4.2 Connection check of thermostatic heater circuit . . . . . . . . . 4-6 Checking magnetic circuit temperature . . . 4-6 ....... 4-8 ............ 4-8 4.5 Adjusting static field homogeneity 4.5.1 Preliminary procedure 4.5.2 Adjustment of static field homogeneity . . 4-9 ...... 4-22 .................. 4-27 4.6 Adjustment of gradient field build-up 4.7 Tuning the coils 4.7.1 Transmitter coil ............... 4-27 4.7.2 Receiver coils ................ 4-32 of cable for adjustment . . . . . 4-32 ......... 4-33 ............... 4-34 (ii) 3 L2E-BM0434 4.7.2.1Adjustment _- 4.7.2.2Entry 4.7.2.3Tuning of receiver coil data adjustment 4.8' Adjusting transmission gain and bias Offset 4.8.1 Bias 4.8.2 Gain . . . . . . . . . . . . . . . . . . . l . . 4-42 l . ...................... 4.8.2.1Initial setting of gain dial ........ 4-44 4.8.2.2Adjusting gain ................ 4-45 4.8.3 ..................... 4-51 Offset 4.9 Adjusting light localizers 4.10Adjusting . . . . . . . . . . . 4-52 . gradient field strength (FOV) . . . . . 4-57 ........ 4-57 4.10.1 FOV adjustment for spin echo 4.10.2 FOV adjustment for gradient echo 4 - 6 2 ....... .............. 4-68 .......... 4-70 4.13 Checking intercom (inter-communication) .... 4-71 4.14 Write of hospital name into magnetic disk ... 4-72 ......... 4-73 .......... 4-74 4.11Signal phase adjustment 4.12 Setting of date display order 4.15 Check of image analysis software 4-16 Check of ECG-gate measurement Section V Check of Image Quality 5.1 Check of FOV accuracy ............... . 5-l 5.1.1 SE sequence .................. . 5-l 5.1.2 GE sequence ................... . 5-8 ................ . 5-9 5.2 Check of distortion 5.3 ............ S-10 Check of S/N ratio when HS/MR ......... head receiver coil is used S-10 Check of S/N ratio when HS/MR body coil is used . . . . . . . . . . . . . 5-15 Signal-to-noise (S/N) ratio 5.3.1 5.3.2 wi - ~ ui (iii) 4 L2E-BM0434 5.3.3 Check of S/N ratio when joint coil is used 5.3.4 Check of S/N ratio when large flexible body coil is used . . . 5.3.5 5.3.6 5.3.7 5-16 5-17 Check of S/N ratio when small flexible body coil is used . . . 5-19 Check of S/N ratio when large extremity coil is used . . . . . 5-21 Check of S/N ratio when small extremity coil is used . . . . . 5-22 Section VI Reference Information 6.1 Connection betwein multiformat camera and central control console . . . . . . . . . . . . . 6-l 6.2 Filming of test pattern by multiformat camera . . 6-2 Procedures . . . . . . . . . . .-. . . . . . . . . 6-2 6.3 Performance data . . . . . . . . . . . . . . . . . . . 6-3 6.2.1 6.3.1 FOV accuracy . . . . . . . . . . . . . . . . . . . . 6-3 6.3.2 Geometrical distortion . . . . . . . . . . . . . 6-3 6.3.3 Signal-to-noise (S/N) ratio . . . . . . . . . . 6-3 6.3.4 Background noise 6-3 . . . . . . . . . . . . . . . . 6.4 Recording sheets . . . . . . . . . . . . . . . . . . (iv) 6-14 5 L!2E-BM0434 Section I Procedures of Installation and Adjustment 2.1.1 Carrying in and installing the gantry 2.1.2 Detaching the 2.1.3 Leveling 2.1.4 Assembling and gantry fixing the protective the panels gantry gantry Start the thermostat control of shielded room I 4.1 Setting the magnetic circuit temperature 2.3, 2.4 Installing the central c o n t r o l console,MRI unit and others , 1 S e e rr’ Note 1 Section III Wiring 1 2.1.5 Attaching the gantry See Note 2 covers 1 4.3 4.4.1 I Operation check of central control console Connection check of thermostat heater 1 4.6 Adjusting the rise-up waveform of gradient field I I $ 4.12 Setting of date displaying order 4.14 Write of hospital name into magnetic disk 4.15 Check of image analysis software 1 - 1 6 L2E-BM0434 static Bas the thermostat control circuit been energized for more than 60 hours after the magnetic circuit temperature has been set? field + 4.4.2 Temperature check .. of magnetic I circuit I t I 4.5.5 Adjusting the static 1 field uniformity I J I Removing the static field uniformity measuring instrument 2.2 Installing table and fixing the patient table, and attaching I the cover 3.2.1 Wiring the patient table J I 4.7 Tuning and adjusting the coils I 4.8 Transmission gain, bias and I offset adjustment I1 I t I I I 4.9 Adjusting the light localixers I J t I I 4.10 Adjusting I 4.11 Signal the phase gradient I I magnetic field strength (FOV) I adjustment I t 5.2 to 5.4 I 4.13 Check of image quality Checking the intercom, and I attaching the-covers I 4.16 Check of ECG-gate measurement A END 1 - 2 7 LZE-BM0434 NOTE: 1 2 During the start-up heating of the magnetic circuit, the cable set of the start-up heater should be le'ox%$ connected. UJ attached until the Keep all the gantry covers completion 0; the static field uniformity adjustment, since the uniformity will be shifted by detaching the gantry covers. l - 3 Section II Carry-in, Installation and Assembling 2.1 Carrying in, installing and assembling the gantry 2.1.1 Carrying in and installing the gantry (1)Place veneer panels (9t) on the floor of the carry-in route and steel plates on them (more than 3.2t) to prevent the floor surface from being damaged. (2)Lift the gantry using the jack and set the chilled roller under the gantry legs. (3)For carrying in the gantry, anchor nuts (M24) are provided as buried in the carry-in route shielded room . floors. .Screw an eyebolt (M24) into this anchor nut. Using a nylon sling, wire rope and winch as shown in Fig. 2.1, carry in the gantry. (4)At a curve of the carry-in route, tool to rotate the rollers. 2-l use the dedicated 9 L2E-BMQ434 u-4 0 2-2 10 L2E-BM0434 1. Weight of components -7 2. : 16000kg (1) Gantry (2) Patient table : Space necessary for 350kg installation of gantry and patient table 3800(W)X5000(D)X2350(H) mm M8 anchor bolt Center of magnet (SUS304) at 4 places 3. Dimensions and weight of components when carried in (1) Gantry 222O(W)Xl460(D)X1803(8) mm, 16000kg provided with (2) (without covers, rollers) Patient table 710(W)X2273(D)X630(H) mm 350kg (with casters) 112 anchor bolt (SUS304) at 4 places 3705 Unit: mm F i a . 2 . 2 Lavout P l a n o f G a n t r v a n d T a b l e o f MRP-7000 2080 22;o Plywood * cover I I t 2360 Lifting hang=-' I I ul zC Qo m ID C Jack boss f m lln- Fig. 2.3 Dimensions of Gantry when Carried in (in mm) 2 - 4 ei 12 L2E-BM0434 2.1.2 Detaching protection plate of gantry Use non-magnetic tools for detaching the plate. (Refer to Fig. 2.4) (l)Remove 8 protective plates (made of plywood). (Required tool : Hexagon key wrench No. 5) (2)Remove 32 studs mounting the protection plates. (Required tool : Wrench No. 17) 2.1.3 Leveling and fixing gantry The leveling work is to be carried out by specialist of the transportation company. (1)Referring to Fig. 2.5, remove the rollers. (2)Place the rubber plates (to be supplied with the gantry) at four places of the gantry base and install the gantry. (3)Using the level gauge placed on the points shown in Fig. 2.6 in X and Y directions, insert the liners It or 2t (to be supplied with the gantry) under the rubber plates so that the gantry is placed horizontally. Referring to Fig. 2.4, remove the jack bosses and fix the gantry as shown below. (1)Referring to Fig. 2.7, drill holes at the gantry fixing places shown in Fig. 2.2. If any steel bar obstructs the hole, make holes at the places shifted referring to Fig. 2.7. In such case, arrange the fixtures to be placed inside the side covers openning. (2)Fix the gantry, referring to the Fig. 2.7. 2 - 5 13 L2E-EM0434 l 2.1.4 Assembling gantry Attach the heat insulation covers as shown in Fig. 10. (1)Attach 8 heat insulation covers for the corners referring to Fig. 2.10. (2)After attaching the heat insulation covers, push in blind caps to assure an efficient heat insulation as shown in Fig. 2.11. (3)Place the insulation plates for floor, referring to Fig. 2.12. (4)Mount the table supporter, referring to Fig. 2.13. Attach the light markers as shown in Fig. 2.8 and 2.9. (1)Attach the right and left side light markers referring to Fig. 2.8. (2)Attach the center light marker referring to Fig. 2.9. (3)Make a connection of 40, 41 .and 42CN at the light markers. Attach the connector plates as shown in Fig. 2.12. (1)Attach the connector plate for RF coil to the gantry right rear leg. (2)Attach the connector plate for start-up heater to the gantry right rear leg. (3)Attach the core mounting plate and connect the cables to the preamplifier and connector plate for RF coil. 2 - 6 14 L2E-BM0434 1 5 UE-BM0434 2 - 8 16 LZE-BM0434 f-i aJ 3 17 LZE-BM0434 Gantry base / Gantry fixture Ml2 nonmagnetic hex. socket head bolt I I Rubber plate, accesory to gantry \ I\ B.. \ Copper foil I .-Ii / ’ I A II i II I To be cut so as to touch the bolt Unit : mm Nonmagnetic anchor nut; prepared hole of 18 in diameter and counterbore of 30 in diameter Fig. 2.7 Fixation of gantry 2 - 10 base on floor 1 8 L2E-BM0434 - . Light marker (left) H4X12e nonmagnetic hex. socket head bolt. 2 places - M4Xl2e nonmagnet i c socket head b o l t , Light Fig. 2.8 marker (right) Attaching the light markers 2 - 11 1 9 L2E-BM0434 hex. socket N I I-J N F i g . 2.9 . Attaching the lsght marker _- 2 - 13 21 L2E-BM0434 s u 4 2 - 14 22 L2E-BM0434 Floor heat insulation material 5 It- - Magnet l e g Bottom view of gantry Fig. 2.12 Layout of floor heat insulation material 2 - 15 2 3 L2E-BM0434 $ se- -w- 0 m X N 5-4 r $ JJ 4 m i-l . N --- 2 - 16 24 LZE-BM0434 / 25 L2E-BM0434 2.1.5 Attaching the gantry covers + Attach the covers'in the following procedure and as shown in Fig. 2.15 (also see Fig. 2.16). - ~ '.""-I 3 iii (1)Attach the front cover 0. Before attaching the front cover, connect the L-lCN, ClCN, R-lCN, 75CN and grounding wire to the control panel on the cover. See Fig. 2.17. (2)Attach the rear cover 0. (3)Attach the ceiling cover 0. (4)Attach the right and left lateral covers @ and @ with their upper sides hung down. NOTE: Keep all the gantry covers attached till the static field uniformity completion of the adjustment, since the uniformity will be shifted by detaching the gantry cover. d'j 2 - 18 26 L2E-BM0434 C o v e r , Front @ Cover, Rear @ C e i l i n g Cover @ L a t e r a l Cover, Right @ L a t e r a l Cover, Left (Q Control panel Cover @ Blind cover for cable 8 RF cover Ql Blind cover for connectors gig. 2.15 Gantry Covers 27 2 - 19 UE-BMO434 head SEMS H4 X 161 nonmagnetic pan head SEMS screv at 4 places (accessory to gantry) N I 0 5 N 0 YAYlCP head SEW screv at 2 places (accessory to gantry) \I ws \ Fig. 4‘._ a -- - Dan head SEW screv at 4 places (accessory to gantry) 2.16 Gantry Covers t. . ( pan HuX~ZQ nonmagnetic head SEMS screw Y) (accessory to gantr Fig. 2.17 Wire connections to the front cover 29 L2E-BM0434 2 - 21 W6XlO nonmagnetic hex. Socket head bolt at 12 places H8X12 nonmagnetic hex. Socket head bolt at 4 places. : -“i Caster Fig. 2.18 Attaching and Detaching Casters 2 - 22 30 LZE-BM0434 - Ml0 nonmagnetic hex. nut at 2 places Fig, 2.19 Fixation of gantry to table 2 - 23 3 1 L2E-BM0434 I ( 1040 Unit : mm *I Deiail o f p o r t i o n A : 4 330 *I I I m RF cover Cover front Detail of portion B : Hex. socket screw t , ( a c c e s s o r y t o rry Linlers Liner Fig. 2.20 Positioning and Leveling of Patient Table key (nominal size.5) . nonmagnetic - hex. bolt ~8 n o n m a g n e t i c h e x . n u t M8 Table base nonmagnetic spring W8 n o n m a g n e t i c s p e c i a l plain washer \ Insulation bush Liner In ,’ \ I I . I /’ 2Plywood Lonleum I Copper foil \ / Plywood I i I Nonmagnetic anchor nut : drilled hole of 12 in diameter t Unit : mm Fig. 2.21 Fixation of Table to Floor 2 - 25 3 3 LzE,BM0434 2.2 Installing patient table 2.2.1 Positioning, leveling and fixing patient table Referring to Fig. 2.18, 2.19, 2.20 and 2.21, position and level the patient table in the following procedures. (1)Connect 4CN and grounding the cables lCN, 3CN, conductor to the table and raise the table to the top position. (After the table is raised up to the top, disconnect all cables.) (2)Attach the four up/down adjust caster and remove the two casters being attached. (3)Turning the screws of the casters, adjust the table height so that the table supporter and table base can be connected. (4)Fix the table supporter and table base. (At this time do not fasten the screws too tight SO -) that the table height can be adjusted.) (5)Adjust the table height by turning the screws of the casters so that the tabletop can be moved in smoothly without any obstruction. .* (6)Fix the table supporter and table base firmly. (7)To make the longitudinal movement direction of the tabletop perpendicular to the gantry, slide the tabletop to the backmost position behind the gantry and adjust the direction of the table so that the distances from the RF cover in right and left direction become s a m e (175mm) 3 (8)Mark the table base and the four anchoring positions on the floor. 2 - 26 34 L2E-BM0434 _- (9) Remove the nuts fixing the table supporter and table base, and move the table so that the anchor nut can be set. (10) Drill holes position. of 12mm diameter at the anchoring (11) Drive in the anchors. (12) Once again fix the table supporter and table base, and the table base and floor. (At this time do not fasten the screw too tight so that the table height can be adjusted.) (13) Attach the adjustment volts at the table height adjusting position. (Place the liners under the table. so that the floor may not be scratched.) (14) Detach the four up/down adjustment casters. (15) Adjust the table height by turning the adjustment bolts so that the gap between the roller and the RF cover becomes 4mm when the, roller at the head side of the tabletop reaches just above the RF cover and the connection part in front of the cover. (16) Adjust the parallelism of the tabletop with the level gauges placed on the cover under the tabletop in X and Y directions by sliding the tabletop toward the gantry and turning the adjustment bolts. (17) Insert the liners into the designated places. (18) Detach the adjustment bolts and liners being used for protecting the floor, and once again check procedures (7), (15) and (16). (19) Fix the table tightly. 2 - 27 35 L2E-BM0434 0 Foot Cover, Right 8 F o o t Cover, L e f t 0 Foot Cover, Front @ Foot Cover, Rear 0 Cable Cover Q Sponge . M4XlOC nonmagnetic screws at 14 places (accessary t o t a b l e ) Fig. 2.22 Covers of patient table 2.2.3 Attaching sponge pad Referring to Fig. 2.22, attach the sponge pad on the edge of the table using adhesives. (Measure the gap between the front cover of the gantry and table cover and select the sponge pad with the thickness suitable for the gap.) 2 - 29 3 7 L2E-BM0434 2.3 Installing central control console 2.3.1 Notes for installation wj Before installing the central control COnSOlet carry out works with care to the followings : (l)When lifting up the central control console, hook ropes to 4 bolts of the transport jigs and lift it up. Fig. 2.3.1 . Fig. 2.3.2 Watch c3 bump 1 I , E b Watch adjusters Fig. 2.3.3 (2)Remove the transport jigs to When the move the console. console is moved being attached jigs, a -i I with the transport special attention must be paid to the people around as the jigs are extended beyond the corners of the console. (Refer to the Appendix 1) (3)The casters and adjusters are provided at the legs of the control console. Make sure that the adjusters are at the highest the position before moving console. (If not, adjust them to the highest position.) Take care that the adjusters do not hit bump when the console is moved on the bumpy floor. 2 - 30 30 LZEBM0434 control the carry-in (4)When console into the operation room through a narrow door, lift up the control panel base as shown Refer to 2.3.4. in Fig. Appendix 2. Fig. 2.3.4 console control central (5)The should not be placed under the (Water drops air cooling unit. from the cooling unit may cause trouble.) Fig. 2.3.5 (6)The front panel can be opened up when the 5" optical disk unit is to be installed or the PCBs and EL display are to be checked. Refer to Appendix 3. Fig. 2.3.6 2 - 31 3 9 L!JE-BM0434 Appendix 1 Detaching the transport jigs 1) Remove 4 screws and detach the front transport jig. * Transport -1 jig Fig. 2.3.7 2) Remove 2 bolts and detach the rear transport jig. ansport jig hexagon wrench Fig. 2.3.8 2 - 32 40 L2E.BM0434 Appendix 2 Opening and closing control panel base _- _- 1) Remove the front cover. 2) Remove 4 volts using Ml0 hexagon wrench. 3) Lift up the control panel base. 4) Hook the stopper located at the center of the base to the stand side. r - \ Front cover Fig. 2.3.9 2 - 33 4 1 L2E-BM0434 Appendix 3 Opening and closing front panel . '-%_I -) 1) Remove 3 set screws fixing the top rear cover. 2) Lift up the bottom of the cover and detach it from the front panel. 3) Disconnect the monitor cable detaching the blind patch. from the cover after 4) Remove the top rear cover. Top rear cover Fig. 2.3.10 5) Raise the front cover holding both the side ends. (The cover is stopped at approx. 45O with the arm stopper.) 6) TO close the front panel, raise the panel 0 frontward and push the stopper @ backward to release it. Arm stopper Fig. 2.3.11 2 - 34 42 L2E-BM0434 2.3.2 Installation procedures (1)Detach the front cover by removing 6 screws. Front cover Fig. 2.3.12 (2)Detach the rear cover by removing 6 screws. (3)Detach the top rear cover. (Refer to Appendix 3 Opening and closing front panel) Top rear cover \ Rear cover Fig. 2.3.13 . 2 - 35 43 LZEBM0434 (4)Open the front cover. -. Install the optical disk unit at this place after removing the top panel. \Top p a n e l Fig. (5)'Detach by removing the 6 screws. the top,panel (6)Installation of 2.3.14 5" optical disk unit. Install the ODR unit and formatter. Filter guide N o t e : Remove the filter guide prior to installation Formatter IF PCB Formatter unit Fig. 2 . 3 . 1 5 NOTE: Remove the formatter and IF PCB from the formatter unit prior to installation. 2 - 36 4 4 L2E-BM0434 ‘. ,.- Guide plate Fig. 2.3.16 Fig. 2.3.17 (6)-l Remove the guide plate. ODR unit 45 L2E-BM0434 / Fig. 2.3.18 (6)-2 Install the ODR unit with 4 M4xlO screws. 2 - 37 Fig. 2.3.19 (6)-3 Referring to Fig. 2.3.17, fix the ODR unit so that the hole diameter of the ODR cover becomes approx. 7mm. 4 Fig. 2 . 3 . 2 0 (6)-4 Mount the filter guide with 2 M4xlO SEMS screws. Guide plate Fig. 2.3.21 (6)-S Attach the guide plate which has been detached in Fig. 2.3.15. 2 - 38 46 LZE-BM0434 PCB base Fig. 2.3.22 (6)-6 Mount the PC3 base with 4 M4xlO SEMS screw Support screw IF PCB Fig. 2.3.23 (6)-7 Mount the IF PCB with 4 support screws and connect the signal cable (b) and power cable (c). Fig. 2.3.24 (6)-E! Mount the formatter with 4 M3x8 SEMS screws. 2 - 39 47 L2E-BM0434 i 9 h d e Fig. 2.3.25 (6)-9 Connect the following cables. Fig. 2.3.26 (6)-10 Connect the cable (j). 2 - 40 40 L2E-BM0434 (7)Installation of magnetikdisk unit Fig. 2.3.27 (7)-l Install the magnetic disk unit on the MD disk base with 4 M4x8 screws. MD power cable (l-53) MD signal cable (l-J11 Fig. 2.3.28 (7)-2 Connect the MD signal cable (l-Jl) and MD power cable (l-53). Refer to Fig. 2.3.28. 2 - 41 49 L2E-BM0434 (8) Attach the front’ cover with 6 screws. Fig. 2.3.29 (9) Attach the rear cover with 6 screws. NOTE: Attach it after turning on the FFB. (10) Attach the top rear cover with 3 screws. Top rear cover Rear cover Fig. 2.3.30 NOTE: Attach the front and rear covers after connecting all the cables. 2 - 42 5 0 L2E-Bh10434 (11) Place the monitor. Fig. 2.3.31 (12) Connect the monitor cable. .w‘7 Shield Clamp Fig. 2.3.32 2 - 43 51 LZE-BM0434 2.4 MRI Unit 2.4.1MRI unit 4 (l)The MRI unit is provided with a cooling fan at the rear side. So place the unit at least 15 cm away from the wall. (2) Secure a space for maintenance work at the front and left sides of the unit. (3)After installation and wiring, fix the unit with the casters raised by the floor adjuster. Wall s Unit 15 cm min. / Fig. 2.23 Installation of MRI Unit 2.4.2 Filter box The filter box is to be installed during the work of shielding the MRI room. If it is necessary to install the unit during installation of the MRI system, do it in the following procedure. 2 - 44 52 L2E-BM0434 Fig. 2.24 shows the installation of the filter box. The dash line indicates the outside dimensions of the filter box. The filter box consists of two parts which are to be shielded installed outside and inside the room, respectively. For ensuring the RF shielding effect, fix the filter box to the shielding panel at intervals of 50 mm along the entire circumference. The filter box should be fixed with M6 through-bolts and nuts at the points indicated with small circles, and with M4 tapping screws at the points indicated with black spots to provide secure contact between the shielding panel and filter box. 1 After mounting the filter box to the shielded wall connect the accessory grounding wire between the grounding stud of the inside unit and ground terminal of the outside unit. 2 - 45 53 LZE-BM0434 Inside shielded room I -------_------ - - ’ I I. : I - - - - - - - - - - - - - - - - - - ’ Outside shielded r o o m v-e-a_---------, I Shielding panel !- 686 635 * 10 x 50 pitches = 500 1 I * IC Fixation with M6 bolt and nut \ I ; \ ;$I Fixation to -w> panel with M4 t a p p i n g 1: Opening for shielding panel 600X900 I screw I---ffmw----s.t--------*----- ‘-I 4-b’ 1 Cable cover L---,,-,-e-m---$ ! Cable i ! inlet/outlet 4 Fig. 2.24 Installation of Filter 2 - 46 - Unit : mm Box 54 LZE-BM0434 Section III Wiring 3.1 Laying the cables between system components Refer to the wiring diagram and cables lists. the cable, take the following in consideration: When laying -The cables generally include the RF transmission/reception cables (RF), gradient magnetic field power cable (GC), signal cables (SIG), power cables (POW) and grounding For preventing image quality deterioration cables (GND). cables and malfunction, separate cables by laying them in pits as shown below and shielding (see the CABLE TYPE columns in the cables lists). - First connect the grounding cables. (l)Inside the shielded room (2)Outside the shielded room Example wiring pits (in mm) 3 -1 55 L2E-BM0434 (007) (124)-----------.m----y ,-------------------- i (106) (334) FIL (107) : I I I I I (002) (311) (336) ::yi; (brown) I I I (307) (308) ( 3 0 9 ) (003) (335) (310) 013) (337) (312) (114) (115) (327) (328) (116) (329) (390) r-------------------------~ (11.7) (330) MRI unit (118) (331) -a- MRI green) - - J’ (yellow) ww (302) (102) 1304) (305) (301) DP (white) (306) clPatient table TB (blue) IMAGER / .. Note Cable are l l * to be prepared at site. The color name shown below the abbreviation of each unit name indicates the ID color of a tape on each cable connected to the Unit. 3 -‘2 5 6 LZE-BM0434 102 MRI DP 0.7 103 1 MRI f F I L 1 1. 104 MKI ! FIL 0.6 I 0.7 I 1.0 I 0.7 14.1 ULSTSB 3Xl4AWC I ss.3 luLsr-sa 1.0 3xIAWG 0 I . I I I I I I 0 I I I KA212973.A 1KAZ12973.F 26.3 ULsr-SB XX6AWG 0 I ma IULST-SB 9 X6AWG 0 6.6 UL2464SS 2xIOAWG 141 llJLsS8 3xl4AWG --e% ~212073.G Kh212D73-H KA212974.L Kh21297CN KA212974.N Kh212974-P 114 FlL C A N - 0 ‘KA213046-M KA213046.N 2 6 . 7 ULST 1 x6AWG 0 1 6 . 7 ULsr 2xbAWG 0 - 6 . 6 hJL2464.66 2 x2OAWG KA213646.T I 1 24+x66 2465x68 I I I 301 I URI I DP I 302 LUU DP 304 MRI DP I 327 MRI FIL 1 FIL I 1 0.3 1 0.7 0.3 0.7 0.3 1 0.6 1 0.3 1 I I I I 0.3 1.7 FIL CAN 1.2 3 2 8 FIL C A N I I I 3 2 9 FIL CAN 330 0.6 I 311 1DP 1 FIL 1 312 I 1 3 0 5 1 MRI 1 D P 1 I I 1.3 I I 11.0 0.6 ZOPx36AWGI i 0 1 1 I&6,7.0 2xZOAWG I x96AWG 6.6 6D2W 8.6 6DZW 1.3 1.6 9 . 0 CG-MA-VVSB lOPx16AWG IUu464-SB 3xlOAWG UIl464.SB 1 xlOAWG 13 1.6 1.4 - 6.6 333 ECG DP 0 1.0 6.0 334 ECG FIL I.3 3.0 336 FIL CAN 0.9 1.0 336 DP FIL 0.1 1.1 1.2 2.2 1 1 0 UL2464-SB 0 0 0 ax20AWG KA212976.G 33.6+x71 33.6@X71 KA212976.H 33.66X71 33.69X71 KA212976-K zoxaxn I I zoxaxn KA212973-C I IKA~I~WI-A KA213061.Q 1 10~77x72 20X77X72 20~77x72 20~77x72 KA21306l.I: I 20X62X72 1 20~62x72 I KA21306bE I 27+X63 I I 14+x34 270x63 KA324mB 36x20.6x62.4 KA324931.C 23+x36.6 23+x36.6 KA324933-f ( 3 6 X 2 0 . 6 X 6 2 . 4 1 14+x34 1 KA32493bC 36x20.6X52.4 36X20.6X62.4 KA32394Qf I 1 0 I I 6.0 aczVA 6.6 I I PI KMW 1 KA213061-P lm~ 16X35X30 [KAll2976J 146x30 IKA327176-h CO-MA-VV.SB lOPX26AWG 101 9.0 ICO-MA.VV.SB fOPxZdAWG( 1 0 1 0.0 CO-MA-VV-SB IOPXZIAWG 101 274x63 1 ,4+X66 lol 9.0 3 - 3 1 20X49X72 1 20X49x72 101 tCZVA 1 I 1 6D.2W 0.6 1 1.0 l,.O SP I jUL2464.SS 7.0 CAN TB 1 1 12.0 b4A.VV-SB FlL FIL 1 1 FlL 338 I 101 6.6 I I 1 0 1 Cc-MA.VV.SB 1.0 I 1.2 I 1.0 I 36PxZ6AWG 332 I CG.WA-VV.SB 0 I 96PXZlAWGI 331 33.sgx71 I I I I I I I~PX~~AWGI I ICC-MA-VV.SB 13.0 f.0 337 I FIL IGAN I I 1 13.0 0.9 CAN KA21297kF =I= 33.6+x71 1 0 1 1 36X16X40 KA2 I297U I 169x37 16+x37 KA321367-D ~36X2O.SX62.4~ I4bX34 1 KA32493l-C 20X49X72 20X49X72 KA213061-R I 20X49X72 I 20X49X72 1KA213001-I. 1 20X49X12 KA213061-L’ 20X49X72 6 7 L2E-BM0434 Remove the h e a t insul tive cover Keayide I/ p(/ W I a Cleat lo-13CN Fig. 3.2 Wiring at Right Lateral Side of Gantry -i-k/ Fig. 3 . 3 W i r i n g a t L e f t L a t e r a l Side of Gantry 3.2 Connecting the cable to each unit 3.2.1 Connecting the cables to gantry and patient table (l)Referring to Fig. 3.1, lay the cables to the gantry and patient table. (2)Wire the gantry referring to Figs. 3.2 and 3.3. (3)Wire the patient table referring to Fig. 3.4. Cable (one) led from inside to lower right of gantry Run# .de 7 Run! Cable pit Runs Fig. 3.1 Laying the Cables to the Gantry and Table 3 - 5 59 L2E-BM0434 Inn Gantry n - n Side = C l e a t Shield clamp Fig. 3.4 Wiring the Patient Table 3 - 6 CD LX-BM0434 3.2.2 Wiring central control console - (1)Connect the external signal cables. (2)Connect the power cable and GND cable and fix them with clamps. Detail of portion B : Runfl Run4 Detail of portion A : 02 002 P rlP Shield Shield’ Fi g. 3.5 3 - 7 6 1 L2EBM0434 (3)Attach the front cover. - " I 'YI?, (4)Attach the rear cover after turning on the FFB. (5)Place the central position. control console at 4; the planned (6)Fix the consoJe with the adjusters of the legs. Slightly floating Fig. 3.6 Turn the adjuster with driver and fix the console so that the front casters are slightly floating. 3 -8 62 L2E-BMO434 3.2.3 Wiring the MRI unit Referring to Fig. 3.5, wire the MRI unit in the following procedure: 4 (1)Detach the right, left and rear covers from the MRI unit. (2)Remove the cable retainer at the rear lower portion of the MRI unit. (3)Connect each cable to the MRI unit (grounding cable should be connected first), and ground the cable at the unsheathed portion using the shield clamp. (4)Fix the cable with the clamp. (5)Retain the cable with the retainer for the cable not to touch the cut portion of the rear cover. (6)Re-install the covers having been detached. A power of 100 V AC is delivered at the output terminals (between 2TN-U6 and -V6) to the warning lamp. The rated current is 1A. 3-9 63 L2G-EM0434 rI vYJ5-4 I 3 - 10 64 LZE-BM0434 3.2.4 Wiring filter box Wire the filter box from inside and outside the shielded Referring to Fig. 3.2.9, connect the cables as room. follows. (l)In case the filter box must be modified for installation to a shield wall of a nonstandard thickness, modify the cables from inside before. Cut the external sheath only of the cable and shift it. (2)Detach the cover from the filter box. (3)Detach the internal wiring duct cover, lay and connect the cables. (4)Remount the wiring duct cover. (5)Ground the cable at the unsheathed portion using the shield clamp. (6)Fix the cable with the clamp. (7)Remount the cover having been detached. In case the filter box is installed as directed laterally, the cables can also be led in and out in the direction of arrow. lines (run 8309 and #329) and reception signal lines (run 8308, #312, #328 and #335) are not provided with any shield clamp. NOTE : The transmission signal 3 - 11 65 L2E-BM0434 EEL 33631 c 310307 W I E! P h) 113 114 I 115 Shield clamp and cleat I I I Shield clamp and cleat I Outside of shielded room Fig. Inside of shielded room 3.8 Wiring to Filter Box 3.2.5 Wiring to ECG monitor (1)Connect the power cable of the ECG monitor to the image control the central monitor power outlet inside console. (2)Connect the signal cable. Power cable Fig. 3.9 NOTE: The run #333 cable is connected in and supplied with the central control console when the cable length is standard. . 3 - 13 67 L2E-BM0434 Section IV Adjustment 4.1 Setting the magnetic circuit temperature It is necessary to set at 32.5OC the magnetic circuit Set the temperature having changed during the transport. temperature in the following procedure: (1)Connect the start-up heater controller between the stepdown transformer and the gantry. Step down transformer hermostatic control for upper heater Thermostatic control for lower heater To Gantry 71 to 74cn At the right side rear leg portion > 200 V AC ground (2)Turn on the circuit breaker in the stepdown tarnSfOrmer --) and the circuit breaker in. the start-up heater controller. The pilot lamp of the controller for Startup heater. (3)Counterclockwise turn to 30.0°C the dial of the thermostatic control at the start-up heater controller. (4)In case the4 ON lamp at the upper right of the thermostatic control is on, it is necessary to raise the magnetic circuit temperature in the following procedure: (a)Gradually turn the dial of the thermostatic control until the ON lamp turns off. (b)The divisions of temperature over which the dial has been turned are the difference between t h e temperature at present and the set temperature. -1 4 -1 68 LZG-EM0434 - (c)The time required for temperature rise is referred to Fig. 4.1. (d)Set the thermostatic control dial to 27OC. Check that the ON lamp of the thermostatic control turns on and the relay at the left is turned on at that time. If the wiring has been completed and the gantry fans are powered, set the dial to 30°C. (c)If the magnetic circuit temperature has risen and the thermostatic control has automatically been turned off, the temperature setting is complete. However, if the wiring is not yet completed even at this stage, keep the start-up heater turned on. (f)Turn off the circuit breaker in the stepdown transformer the start-up heater and disconnect controller. (5)If the ON lamp at the upper right of the thermostatic control is not lit, it is necessary to lower the magnetic circuit temperature in the following procedure: (a)Gradually turn the dial of the thermostatic control until the ON lamp lights turns on. (b)The divisiins of temperature over which the dial has been difference between the turned are the temperature at present and set temperature. (c)If the temperature difference exceeds 4OC, detach the heat-insulative covers at the right and left lateral sides of the gantry to cool the gantry inside. At this time, lower the temperature in the MRI room .as much as possible. (d)Turn off the circuit breaker of the stepdown transformer and disconnect the cable set for start-up heater. 4 - 2 69 L2E-EM0434 (c)When a time required for raising the magnetic circuit temperature has passed or when the pointer of the heater power meter at the lower front left of the MRI unit has started deflecting after the power supply is turned on, re-install the heat-insulative covers. 1 I I l.O'c/h J / / 10 20 Temp. difference ("C) Fig. 4.1 Time Required for Temperature Rise 4 Ambient temp. 20°C -0.25OC/h Temp. difference ("Cl Fig. 4.2 Time Required for Cooling 4 -3 70 L2E-BM0434 ..‘,‘?\ , ul@: c 4.2 Switching on power supply Before switching on the power supply to the system, check as follows: (1) If you felt something strange at any point during wiring, check that point once again. (2) Check the insulation between each power line and the ground potential at the terminal chassis located at the rear lower left portion of the MRI unit. (3) Pushing in all the circuit boards, check the circuit boards and connectors to see that they are securely connected. Check as follows with the power on: The (l)Turn on the MAINS FFB (101 FFB) of the MRI unit. 4 gantry and thermostatic control circuit (heater unit in the MRI unit) which are to be continuously energized will be energized and the HEATER lamp on the central control console be illuminated. (2)Next, press the CPU ON/OFF switch on the central control console, and the latter will be energized. (3)Then, press the MR ON/OFF switch on the central control console, and the MRI unit and patient table will be energized. With the above operations, all the component units of the system are energized. 4 - 4 71 L2E-BM0434 4.3 Operation check of central control console Check that circuit boards are correctly and securely inserted, and also check the cables and DC sources (by checking whether power corresponding LEDs are lit or not) No Yes Protocol changeable? 9 No+ Yes Image r e a d / w r i t e from/into OD No !I Initialize and loading of all accessory programs NOTE: In some cases, the software having been registered in the HD at the time of shipment is different in version from the software in the OD. So, reload all the accessory software referring to the section II of Technical Guide-Check 6 Service Programs. by See NOTE. check programs. The CPU POWER switch must be kept turned on also in the subsequent procedure. I f i t i s n o t , MR POWER switch cannot be turned on. 4 - 5 72 L2E-BM0434 4.4 Operation check of thermostatic control circuit - -, 4.4.1 Connection check of thermostatic heater circuit After switching on the power supply as in item 4.2, check the thermostatic heater as in the following: (1)Tilt to ON/OFF CONT position the switches 2015 and 202s on the heater unit located at the front lower left portion of tQe MRI unit. (2)Make sure that the voltmeter pointer of the power unit for the heater unit deflects to 35 V dc. (3)Check that the ammeter of the power unit for the heater unit indicates a value with the following range: - Righthand power source (upper) : 7.0 to 9.0 A - Lefthand power source (lower) : 10.0 to 12.5 A (4)Return the switches 2015 and 202s to NORMAL position. (5)Record the measured values. 4.4.2 Checking the magnetic circuit temperature After setting the magnetic circuit temperature as in item 4.1, check as follows after a predetermined time for temperature stabilization has passed: (l)Turn off the CPU POWER switch on the central control console, and then turn off the MAINS 101 FFB at the front portion of the MRI unit. (2)Detach the cover from the filter box from the control room, and disconnect the connector 7CN. 4 - 6 73 L2E-BM0434 (3)Using a digital voltmeter, measure the resistance (resistance of the thermistor for detection of the temperature of the upper magnet) between the connectors 7CN-1 and -13, and that (resistance of the thermistor for detection of the temperature of the lower magnet) between the connectors ~CN-2 and -14. (4)Reconnect the connector 7CN, and turn on the MAINS FFB (101 FFB) to put the heater control circuit into operation. (5)The values follows: measured at the step (3) should be as Between 7CN-1 and -13 : 808 to 844n (+0.5OC) Between 7CN-2 and -14 : 808 to 844Q (~t0.5~C) Difference between the value measured 7CN-1 and -13 and that between 7CN-2 and -14 : 9L? at maximum (+O.lS°C) (6)Record the measured values. 4-7 74 LZE-BM0434 4.5 Adjusting the static field homogeneity 4.5.1 Preliminary prosedure When (l)Wait until the magnet temperature gets stable. the following two conditions are met, the magnet is stable in temperature. (a)60 hours has passed after the thermostatic control circuit is put into operation. (b)Referring to the separate Technical Guide "Static field homogeneity tester", measure the static field The fluctuation should homogeneity every 3 hours. For DEG, enter be less than 1 ppm at 21, X and Y. " 9 0" . (2)Referring to Fig. 4.10.10-1.2, do as follows: Attach the accessory dial gauge fixture @ (4 pieces) After attaching the spindle to the adjusting cap @. to the accessory dial gauge 8, insert the dial gauge into the hole in the dial gauge fixture @ and turn the Tighten fixing bolt 0 by hand to fix the dial gauge. the fixing bolt @ as fitted in the groove of the dial gauge. Do not try to tighten the fixing bolt using a If the fixing bolt is tightened using a tool, tool. the spindle cannot be moved. Do not loosen the fixing bolt @ (4 X 4 pieces) or the upper yoke @ may possibly fall. (3)Set the dial gauge to "0". (4)Put the accessory dedicated the dial gauge. 4 - 8 heat insulation cover to 75 L2E-BM0434 4.5.2 Adjustment of static field homogeneity (1)Set the probe plate of 400 x 350 mm of the magnetic W3 field tester. (2)Adjust according to the instructions given in the separate Technical Guide "Static field homogeneity tester". For DEG, enter "90". (3)Adjust concerning 21, ~2, ~3, 24, ZY, X, Y according to the field homogeneity adjusting flow chart. (4)Adjust each of the above items according to the magnetic field analysis results and Table 4.1. NOTE 1 : When the field homogeneity has been done concerning,Zl, 22, 23 and 24, measure the static field at the 3 points: center, top and bottom, and then compare the frequency at each point on the CRT. If there is found a different of more than 100 Hz between the frequencies, press the &. space key and select "E" to end the measurement. Retry the adjustment. (5)If the field homogeneity is within the permissible value (PK/PK = 60 ppm be taken a measure; only a small number of measuring points is available for a measurement with 90° and the PK/PK value will be smaller than in the measurement with 300), end the measurement with 90°. If the measured field homogeneity does not meet the permissible value, repeat the adjustment and measurement until it is within the permissible value. (6)When the measured value with 90° is found within the permissible value, effect a measurement with 30° to check the final data. 4 - 9 7 6 L2E-BM0434 (7)Once the static field homogeneity is found within the permissible value, end the adjustment. If not, repeat the adjustment and measurement until it is attained. (8)When the measured static field homogeneity is found within the permissible value, check its reproducibility after the fixation. The fixation should be done as follows: Adjusting yoke 0: Evenly tighten the adjusting bolt @ (at 3 Places) and fixing bolt @ (at 3 places) using a hex. wrench key, not any pipe wrench. Adjusting cap @: With the dial gauge set to "O", CW turn the adjusting Uniformly bolt @ until the gauge indicates 0.07 mm. tighten the fixing bolts 0 until the gauge indication returns to "0". NOTE2 : During the field homogeneity adjustment, be careful as follows for no difference to occur in temperature between the magnet top and bottom: insulation 1)For the adjustment, the heat materials for the upper adjusting yoke (at upper center) and adjusting caps (at upper four corners) should not be left detached for more than 30 min. 2)Also the heat insulation materials other than the above should not be left detached for more than 30 min. 4 - 10 77 L2E-BM0434 (9)When the measured magnetic field measured with the conditions of 400 x 350 mm and 30* is found within the %.> permissible value, select the magnetic field analysis -"! mode and analyze the magnetic field through calculation Make with parameters of DSV X, Y, Z = 200 and DEG. 30. sure that the Pk-Pk value is less than 14 ppm. 4 - 11 78 L2E-BM0434 @ Upper yoke 8 Upper adjusting yoke 0 Column @ Fixing bolt, Ml2 hex. socket (nominal size 10) 0 Adjusting bolt, 1412 hex. socket (nominal size 6) Q Adjusting cap 8 Fixing bolt, M20 hex. socket (nominal size 17) @I Adiustinu bolt, H36 hex. socket (nominal size 14) @ -Dial gauge fixture (spanner 22) 6 Dill gauie @ F i x i n g b o l t , 744 h e x . socket (nominal size 3) @ @ @ @ Lower adjusting yoke Adjusting device for lower adjusting yoke Accessory handle Fixing bolt, Ml6 hex. socket (painted yellow) Detail of portion A Detail of portion C LGl Upper adjusting yoke: The yoke is moved 1.75 rni by turning the bolt@one full turn. Lower adjusting yoke: The yoke is moved 0.05 mm by turning the accessory handle @ one full turn. Fig. 4.10.-l Homogeneity Adjusting Point (1) @ Upper yoke adjusting bolts 8 pcs , I420 (nominal size 10, T wrench) 0 Lower yoke adjusting bolts 2 PCS, M20 (nominal size 10, T wrench) Rear side Fig. 4.10.10-2 Homogeneity Adjusting Points (2) @zY Adjustment I The digital values in the -+ flov chart are not any allowable values but ones in the process of adjustment. Field homogeneity measurement I IYes I I @23adjustment I Field homogeneity measurement I NO No I 0~4 adjustment I Field homogeneity measurement I I t= @@Z2 adjustment 1 Field homogeneity measurement @@Z2 /Field homoseneity 1 I I NOTEl: u NOTEZ: adjustment The Nos. 0 to @ indicate for Allowable values nonhomogeneity Zl, 22, x, Y +lOppm max. 23 fl5ppm max. 24 k2Oppm max. Others ~ I ?:lOppm max. I the step Nos. in Table 4.1. 4 - 14 81 L2E-BM0434 Table No. I tern Q 21 4.1 Adjusting point Wwe r adjusting yoke 8 Adjustment Sign of Adjusting magnetic fields method Variation of other item Loosen the fixing bolt @ in the upper adjusting yoke 8 (see d e t a i l o f p o r t i o n B). N e x t adjust, referring to Pig. 4.10.11, so that the difference i n m,agnetic f i e l d i n t e n s i t y between two points Z+(O, 0, +lOO) mm and z-(0, 0, -1OO)mm is less t h a n O.OlG (43Hz). + Turn the adjusting bolt 8 CCW to lower the adjusting yoke C3. oZ2=0.24 x21 - Turn the adjusting bolt @ CW to raise the adjusting yoke 8. Lower adjusting yoke @ Adjust, referring to Fig. 4.10.11, so that the difference, in magnetic field intensity between two points Z+ (0, 0, +lOO) mm and Z- (0, 0, -100) mm *is less than O.OlG (43 Hz). + Turn the handle @ CW to lower the adjusting yoke 0. - Turn the handle @. CCW to raise the adjusting yoke @. c3 23 Upper yoke adjusting bolts @ + Extract all eight upper yoke adjusting bolts 16 to a same length at a s a m e time. - Insert all eight upper yoke adjusting bolts 16 to a s a m e length at a s a m e time. nz1=-0.95xnz3 Do not turn the adjusting bolts more than +/-31 turns from the position adjusted at factory before shipment. 6 24 Upper yoke Q + Raise the upper yoke a. - Lower the upper yoke @. At this time, the gap moving amount should be as follows: 0.0028x(10.8Z4-Z2)mm The adjusting method is referred to No. @@. 4 - 15 az2=-2.93 xaz4 82 L2E-BM0434 No. I tern Adjusting point @ x Upper yokea Sign Adjusting method Variation of other item (1) Evenly loosen the fixing bolts @ in the adjusting cap @‘(four each). At this time, the dial gauge indication deviates +O.OS to 0.1 mm, but do not move the scale. (2) For smooth move of the upper yoke, each move at two points should be less than 0.1 mm. ( 3 ) R e f e r r i n g t o F i g . 4.10.14, determine a necessary gap moving amount for correction of nonhomogeneity item of x. + - 1) I n c a s e o f 22X): Turn the adjusting bolts 8 at LGl and LG2 CCW to lower the left side of the upper yoke a. 2) In case of 22<0: Turn the adjusting bolts @ at LG3 and LG4 CW to raise the right side of the upper yoke a. 11 2) aI Y Upper yokea In case of 22>0: Turn the adjusting bolts 8 at LG3 and LG4 CCW to l o w e r t h e l e f t .side o f the upper yoke Q. I n c a s e o f 22eO: Turn the adjusting bolts 8 at LGl and LG2 CW to raise the left side of the upper yoke a. (1) Evenly loosen the fixing bolts @ in the adjusting cap 6) (four each). At this time, the dial gauge indication deviates +0.05 to 0.1 mm, but do not move the scale. (2) For smooth move of the upper yoke, each move at two points should be less than 0.1 mm. 4 - 16 83 L2EBM0434 1 No. I tern Adjusting point Q Y Upper yokea Adjusting Sign ( 3 ) R e f e r r i n g t o F i g . 4.10.14, determine a necessary gap moving amount for correct i o n of nonhomogeneity item o f Y. + 1) In case of 22>0: Turn the adjusting bolts 8 at LG2 and LG4 CCW t0 lower the front side of the upper yoke a. 2) In case of 22<0: Turn the adjusting bolts (&, at LGl and LG3 CW to raise the rear side of the upper yoke 0. 1) I n c a s e o f 22>0: Turn the adjusting bolts (&, at LGl and LG3 CCW to lower the rear side of the upper yoke @. 2) In case of 22x0: Turn the adjusting bolts 8 at LG2 and LG4 CW to raise the front side of the upper yoke a. b - 8 0 c3 0 22 Upper and lower adjusting yokes Q and @ Upper yoke @ Variation of other item method + Insert both the upper and lower adjusting yokes 8 and @. A24= O.OSXA22 - *Extract both the upper and lower adjusting yokes 8 and @. (1) Evenly loosen the fixing bolts 0 (detail of portion A) in the adjusting cap 8 At this time, (four each). the dial g a u g e indication deviates +0.05 to 0.1 mm, but do not move the scale. (2) For parallel move of the upper yoke, each move at four points LGl, LG2, LG3 and LG4 should be less than 0.1 mm. A26= 0.05XA22 ’ A24= -0.34XA22 A26= 0.07XA22 ( 3 ) R e f e r r i n g t o F i g . 4.10.12, determine a necessary gap moving amount for correction of nonhomogeneity item of 22. 4 - 17 54 L2E-BM0434 No. Item Adjusting point Sign - .” . . Upper c3 yoke 0 + 0 Adjusting method Variation of other item T u r n t h e a d j u s t i n g b o l t BCCW to lower the upper yoke aa necessary moving amount. - T u r n t h e a d j u s t i n g b o l t aCcW to raise the upper yoke aa necessary moving amount. @ ZY Upper yoke adjusting + b o l t s @’ (front side) and lower yoke adjusting bolts @ (rear side) a’ Insert 2 bolts @ at front side of upper yoke, and extract 2 bolts 0 at rear side of lower yoke. AZ21 -1.OSxAZY - Extract 2 bolts @ at front side of upper yoke, and insert 2 bolts 0 at rear side of lower yoke. Do not turn the adjusting bolts more than +/-31 turns from the position adjusted at factory befor shipment. 4 - 18 8 5 L2E-BM0434 c c w c 4c w Adjusting handle CzYfxrglIf 4 an N 0 N * a d j u s t i n g yoke(mm) t 0 . f N adjusting a P 0 - K W . N . ~ I-L 4I z a3 a c P - - / U . c ul - yoke(mm) . 22:43.1 ppm/mm (0.023 mm/ppm) / &I C 3 1 P Id 14.7 ppm/mm c" -4 si E (0.068 mm/ppml 0 -1 -30 -20 -10 0 lo 20 30 22, 24 (ppm) Fig. 4.10.13 Gap Moving Amounts for 22, 24 I 0 X: 24.6 ppm/O.l mm\ (0.0041 mm/ppm) 0 10 Y: 56.2 p&0.1 I 20 30 mm (0 .0018 mm/ppm) I * 40 X,i(ppm) Fig. 4.10.14 Gap Moving Amount for X, Y 4 - 20 87 L2E-BM0434 See NOTE 0.175 1 ppm/rot 0.125 ppm/rot 10 e U 20 30 -4 -2 Fig. 4.10.15 2 0 23 (ppm) Adjusting Bolt Rotation for 23 0.325 -10 '4 -5 0 5 See NOTE ppm/rot 10 2Y 1 ppm) Fig. 4.10.16 Adjusting Bolt Rotation for ZY NOTE : The adjusting bolts U?J and CD for 23 and ZY should not be rotated more than 31 tunrs from the position adjusted at factory before shipment. e' a 4 - 21 88 L2E-BM0434 4.6 Adjustment of gradient field build-up - The equipment has already been adjusted at factory before shipment so that the gradient field is built up appropriately, but it is recommended to check at site and For the fine make a fine adjustment if necessary. adjustment, refer to the instruction manual for the gradient field tester. (1)Install the search coil stand so that the search coil is at the left side on the X-axis and at feet side on the Yaxis. (2)Mount the search coil on the search coil stand so that its portion to which the cable is connected is upside. (3)Set up the gradient field tester by making necessary connections to the tester, and set the switches on the G CONT2 board of the GC unit inside the MRI unit as follows: 5sw PULSE 4sw X 3sw ZGND YGND XTEST 2sw 1sw (4)Observe the SIG OUT1 and SIG OUT2 waveforms of the gradient field tester using an oscilloscope with storage (so-called storage-type oscilloscope), and make offset adjustment referring to the instruction manual. (5)Make sure that the peak value of the output signal If not, adjust waveform at SIG OUT1 is 2.5 V +/- 0.5 V. the X-axial position by turning the adjusting screws on the search coil stand. 4 - 22 89 L2E-BMOd34 Also, set the switches on the G CONT2 board as shown in Table 4.2. Make sure that the peak value of the SIG OUT Output signal waveform is 2.5 V +/- 0.5 V when each of If a deviation is the X-axis to Z-axis is adjusted. found, make a positional adjustment in each of the Xaxial to Z-axial directions. After completion of the checking or adjustment, securely tighten the screws on the search coil stand. Table 4.2 Switches on G CONTZ X-axis Y-axis I Z-axis I 5sw PULSE PULSE I PULSE I 4sw X 3sw ZGND zsw YGND YTEST 1sw XIPEST XGND trimmer 18VR-20VR ZSVR-31VR board Adjusting YGND I I XGND I I (6)Make sure that the output signal waveform in each of the axes (SIG OUT1 and SIG OUT2) meets all the following conditions. If not, make a fine adjustment by turning the adjusting trimmer for each of the axial directions. 4 - 23 90 L2E.BM0434 -?.., w)' Condition 1: SIG OUT1 waveform J~~~~~-7~ ----------- --e-m --------4 - Av II 4 I1 t i II II SIG OUT2 waveform I II I ! ‘1. Sms’ Fig. 4.11 Observing conditions: : 1 V/div. : 20 mV/div. : 500 ps/div. - SIG OUT1 waveform - SIG OUT2 waveform - Time base The output signal waveform in each of the axial directions under the conditions shown in Fig. 4.11 should meet the following relation: AV/V x 100 I 0.7 Condition 2 : --------SIG OUT2 Fig. 4.12 4 - 24 91 L2E-BM0434 Observing conditions: : 1 V/div. : 20 mV/div. : SO ms/div. - SIG OUT1 waveform - SIG OUT2 waveform - Time base The output signal waveform in each of the axial directions under the conditions shown in Fig. 4.12 should meet the following relation: nv/v x 100 I 0.7 Condition 3 : SIG OUT1 waveform 4 Fig. 4.13 Observing conditions: : 0.5 V/div. : 100 ps/div. - SIG OUT1 waveform - Time base The output signal waveform in each of the axial directions under the conditions shown in Fig. 4.12 should meet the following relation: X-Y axis 340 +/- 80 ps TTO : 490 +/- 80 /IS 9'50 : 2 axis T50 : T70 : 4 - 25 220 zk 80 ps 300 5 80 ps 92 L!2E-BM0434 NOTE: - indicate lengths of time from the waveform starts rising until SO%, 70% of the peak value V are reached, respectively. WI and T70 (7)After completion of the checking and adjustment, set the switches on the G CONTZ board as follows: 5sw 4sw 3sw : : : 2sw : 1sw : DC X ZDEMAND YDEMAND XDEMAND Power cord I Gradient field tester Coaxial Oscil .loscope with storage I /- cables 4 / Search coil stand NOTE: Only thicker spacer should be inserted between the search coil and stand. Fig. 4.14 4 - 26 93 L2E-BMW34 4.7 Tuning the coils The coils have been already adjusted at factory at the time ?, of shipment. We believe that the adjustments have not been ep$ so much deteriorated during the transport, but it is recommended to effect fine adjustment and check of them. Record the values before and after the adjustments. - Adjusting conditions: The adjustment should be done in a condition in which a measurement can be done after other adjustments except for the coil and transmission gain adjustments have already been completed. 4.7.1 Transmitter coil (1)Using the frequency search (FREQUENCY SEARCH : 48-7) program among the service measure the programs, resonant frequency. If the tuned state of the coil is greatly different from that having been done at factory before shipment (it varies depending upon the cable length, etc.), an error will be caused and the d frequency search is impossible. In such case, tune the transmitter and receiver coils with reference to a resonant frequency obtained through the adjustment of field uniformity, thereby enabling the frequency search. The frequency search is referred to item 5.6 in the Technical Guide - Check & Service Programs. (2)Detach the two lids located on the rear cover of the gantry as shown in Fig. 4.15 to adjust the tuning board and to connect the impedance meter. .. 4 - 27 94 LZE-BM0434 a : Tuning adjustment part 0 : Impedance meter connection part Fig. 4.15 (3)Short-circuit the cross diode using.the shorting bar on the tuning board. (4)Disconnect, at the 9CN side, the BNC of the cable connected from the tuning board to the output connector 9CN, and connect the probe of the impedance meter as shown below. 9CN (BNC) --) To impedance Tuning board meter \ Type T-BNC conversion COnneCtOr, accessory to the system (5)Set the frequency of the impedance meter to the resonant frequency, and adjust the variable capacitors 1VC (Cr) and 2VC (Cm) on the tuning board for the impedance to be within a range of 50 210 ohms and phase to be within a range of 0 ~00~. 4 - 28 95 L2E-BM0434 (6)Disconnect the impedance meter probe and connect the '^. _ cable from the tuning board to the 9CN (return to the 4.; normal state). Also make the shorting bar on the tuning board in release status. (7)Using the SIGNAL GENERATION CHECK of the PSC check program, perform measurement sequence operation to search the f*requency which makes the reflection power minimum and readjustment by shifting the frequency at the time of tuning adjustment by the difference from the resonance frequency'. Repeat this adjustment procedures ((3) - (7)) until the reflection power is reduced less than 15%. Procedures : (a)Manually set the resonant frequency using a dial on SYNTH 2 module. (b)Select the standard protocols setting of BRAIN and ,& SCANOGRAM, but set only TR to 1OOms (the protocol BRAIN must be selected). Start the SIGNAL GENERATION check (3) program among the PSC check program (47-5) under the following conditions : FRQ = 00000(resonant Tune = 18 (tuned value) GAIN = 14 (gain) frequency) . (c)Check the reflection power by the meter and meter range select switch on the front panel of the power amplifier. 4 - 29 96 L2E-BM0434 Pilot lamp \ Reflection /detection lamp i Meter indication select switch connector monitor Front panel of power amplifier -With the meter indication select switch @ to FWD (output power) and adjust the. transmission gain control (GAIN) of the TRANS4 module for the pointer of the meter 0 to deflect to 80%. - Set the switch @to RFD (reflection power) and read the deflection of the pointer of the meter 0. The reading is the reflection power. (d)If the reflection power is less than 15%, the If more than 15%, adjustment can be terminated. make readjustment as per the following procedures. (Be sure to check the reflection power by setting the resonant frequency.) (c)Changing the frequency by the frequency dial of the SYNTH2 (with the switch set to Local side), search the frequency which makes the reflection power Obtain the difference between this minimum. frequency fl and the resonant frequency(fo), and adjust the frequency fo' to be set at readjustment to the value shifted reversely by the value of the difference. 4 - 30 97 L2E-BM0434 fo’ = fo .8 + (fo - fl) e k.\ Carry out the readjustment with this fo' as per di (3). (8)Change over the switch of the SYNTHZ module to REMOTE (usual state) and remount the lids on the rear cover of the gantry. .I 4 - 31 98 L2E-BM0434 . . 4.7.2 Receiver coils - Each of receiver coils should be tuned and checked as in The items variable with each the following procedures. The optional coils should coil are listed in Table 4.3. be checked and adjusted as instructed in the documents accessory to‘the coils. 4.7.2.1 Adjustment of cable for adjustment Adjust the A/2 cable for adjustment. (This cable is not provided with the system.) h/Z,cable (for adjustment) c Impedance meter probe Trimmer (1)Connect meter. the h/2 condenser cable to the probe of the impedance Leave the other end of the cable open. (2)Set the frequency of the impedance meter to the resonance frequency of the system. (3)Adjust the trimmer condenser to make the impedance maximum (approx.l,4kQ-1.7kQ). 4 - 32 99 L2E-BM0434 4.7.2.2 Entry of receiver coil data Select and run the service program COIL DATA ENTRY (4831), set a coil name and RF reception line and enter the For attenuation amount for adjustment of the coil gain. the program, refer to item 5.17 in the Technical Guide This entry is to be done for Check c Service Programs. all the coils used. Setting : - COIL NAME : Set a name for each coil with 16 characters and enter it. at maximum, -'RF Line : Set "3" (RF LINE B) for the existing coil, and "2" (RF LINE A AND B) for the HS/MR coil (QD coil). - RF ATT : Enter an attenuation amount for the HS/MR coil (QD coil). This setting is not necessary for other coils. This value is selected and set at the time of'adjustment. - TUNE STEP : Set "3" Set the voltage variation step at auto-tuning. for all the settings for the time being. (If other values are set, pre-scan may be prolonged or S/N may be lowered.) Ex amp le of settinq : . 4 - 33 100 L2E-BM0434 4.7.2.3 Tuning adjustment Carry out the tuning adjustment and checking of each coil as per the following procedures. (The flexible body coil is not needed for the turning adjustment except for the check.) (1)Detach the lid and cover of the tuning box of each Detaching steps are shown in the following. coil. 0 HS/MR HEAD COIL Cover /' ' Cover \ J @ HS/MR BODY COIL 4 - 34 Cover 1 0 1 LZE-BM0434 (2)Set the phantom for adjustment in receiver coil in the center of the magnetic field. Phantom No.3 / 2Omm thick mat for abdominal immobilization HS/MR HEAD COIL Fix the phantom with aDaomlna1 immobilization belt 15mm t h i c k m a t HS/MR BODY COIL phantom 70mm thick pillow mat - No.2 2Omm thick abdominal immobilization mat 15mm Ii3 n$ thickness I _ r F I 1 JOINT/LARGE EXTREMITY/SMALL EXTREMITY COIL 4 - 35 102 L2E-BM0434 SMALL phantom No.3 BODY (NO.l) phantom No.4 FLEXIBLE 50mm mat COIL 50mm thick mat 15mm thick mat (double) SMALL (No.2~No.4)/LARGE (No.1) FLEXIBLE BODY COIL 2Omm thick tion phantom No.4 50mm thick 50mm t mat mat 15mm thic mat (double) LARGE FLEXIBLE BODY (No.2, No.3) coil 70mm thick phantom No.4 50mm thick mat 50mm mat 15mm thick mat (double) LARGE FLEXIBLE BODY 4 - 36 (No.4) coil IO3 LZE-BM0434 (3)Py r u n n i n g the PSC check program among the check program, set a tuning voltage as in the following : “-1 d‘! (a)Selection of PSC check program (47-5) (b)Selection of SIGNAL GENERATION check program (5) (c)With the following set, start the program : FREQ = 000 ( resonant frequency) TUNE = 00 (enter a tuned value to be set as specified in Table 4.3) GAIN = 14 (gain) Other than the tuned value has no direct relation with Stop (by pressing the SPACE key) the this adjustment. A tuning run of the program in several seconds. voltage is thus completely set. (4)Connect the receiver coil to be adjusted to the receiving Connect the h/2 cable for connector. I adjustment t"o the BNC connector for adjustment, and -‘ connect the impedance meter to this connector. Connector for adjustment (LINE B) Solenoid side of QD coil and single coil /for adjustment A/2 cable Connector for ad'ustment (LINE A) Saddle side o z QD coil for adjustment . 4 - 37 Impedance -probe meter 104 LZE-BM0434 (5)Set the frequency of the impedance m e t e r t o the resonant frequency, and adjust the trimmer condenser for tuning so that the impedance and phase become equal to those shown in Table 4.3. Adjust the solenoid and saddle side coils of the HS/MR HEAD and BODY COILS respectively. Check only the maximum impedance and phase for FLEXIBLE SPINE COIL (trimmer condenser is not provided). Table 4.3 Receiver coil adjustment values Tune Coil # Coil name Phantom # value to be set 1 HS/MR Head 2 HS/MR Body 3 Impedance adjust.(n) Phase adjust. Tuned value range to be checked 3 16 200+100 ItlO" OC-4E 4 18 200+100 fl0” OC-4E Joint/Neck 2 18 60-150 *lo” OC-36 4 Large Extremity 2 18 60-150 *lo” OC-36 5 Small Extremity 2 18 60-150 fl0" OC-36 6 Large Flexible 4 -- zoo* 100 f15” OC-72 3 Be 550_+100 f15" OC-72 4 -- 200-+100 &15" OC-72 4 Body No.l-4 7 Small Flexible Body No.1 Small Flexible Body No.2 to 4 8-31 As specified in the document attached to the coils Location of trimmer condensers of each coil adjustment are as follows : Other trimmer condensers should not be adjusted. 4 - 38 for 105 LZE-BM0434 1 I Solenoid Saddle &ide side HS/MR Head coil JOINT/LARGE and SMALL EXTREMITY coil 4 - 39 106 LZE-BM0434 Saddle ajde Solenoid side HS/MR Body coil 4 - 40 107 LZE-BM0434 (6)For determination of attenuation value (ATT value) of ::, HS/MR coil (QD coil), measure SN ratio of solenoid and saddle side of HS/MR coil respectively, and obtain the ~1 ATT value from the SN ratio. Follow the same procedures for HS/MR head and body .I coils. Procedures : 0 Set the phantom (in the same way as in tuning adjustment). 0 Change RF LINE,by COIL DATA ENTRY program. Solenoid side : RF LINE = 3 Saddle side : RF LINE = 1 0 Scan an image for SN ratio measurement and measure SN ratio. Refer to Item 5.3 for scan procedures. Scan is to be performed in single slice. Be sure to scan only after the RF GAIN adjustment is done for each scan. (Refer to Item 4.8) CD Calculate ATT value as follows : ATT value = ZO*log(Solenoid 0 s i d e SN r a t i o / S a d d l e s i d e SN r a t i o ) Round off the ATT value obtained in @and set this At this time value by COIL DATA ENTRY program. reset the RF LINE to "2". 4 - 41 108 L!lEBMOd34 w . 4.8 Adjusting the transmission gain and bias and offset The amplitude and offset and bias of a signal (Sine function) supplied to the transmitter coil for generating of RF magnetic field so that the MR signal can be efficiently received. - Adjusting conditions: This adjustment should be done after other adjustments and checking have been completed. - Description of TRANS 4 module Offset I W-P adjust dial Gain adjust dial Bias adjust dial 4.8.1Bias (l)Short-circuit the end of loop cable (installation jig) of the probe (1 : 1) of an oscilloscope as shown below, from it into a loop shape and set it in the center of the transmitter coil so that the loop is nearly perpendicularly. At the time of the second and subsequent check and adjustment, the waveform may be observed on an oscilloscope with the accessory coaxial cable (50 ohms) connected to the output monitor connector of the RF power amplifier and with the input of the oscilloscope (Use the accessory SO-ohm terminated with 50 ohms. terminator and T type BNC connector.) 4 - 42 109 LZE.BM0434 0 To oscilloscope, (2)By running the SIGNAL GENERATION check program among the PSC check programs, effect a serial-scan operation. Protocol conditions : Sequence Flip angle (FA) TR NSA : GE : 300 : 100 :2 The other protocol particularly. not conditions are specified (3)Observing the transmission waveform on the oscilloscope, adjust the Bias dial to make the pulse amplitude not change at every transmission. Lock the dial after adjustment. Adjust the Bias dial to make the amplitude not change at every transmission. Magnified \ 4.8.2 Gain The transmission gain of each receiver coil is automatically adjusted using the service program. However, initially set the gain dial by manual coarse adjustment. 4 - 43 110 LZE-BM0434 , 4.8.2.1Initial setting of gain dial This adjustment is to be done only during installation of The initial gain setting should be done the equipment. for the craniocervical receiver coil alone. (1)Initialize the RF gain value in the manual mode of Set RF GAIN the RF GAIN ADJUSTMENT program (48-8). Details are referred to for the coil No. 1 "32767". the Technical Guide - Check & Service Programs. (2)Using the accessory coaxial cable, connect an oscilloscope to the coaxial connector AF OUT O" or 90° on the front panel of the ADC in the RF Observe the received signal transmitter/receiver. just before it is converted from analogue to digital. (3)Set the adjusting phantom at the center of the magnetic field inside the receiver coil. - (4)Set the LOCAL/REMOTE switch on the SYNTHZ board in the RF transmitter/receiver to LOCAL position, and set a resonant frequency by the FREQUENCY dial. Set measuring protocols as shown below (all in SE mode): Coil No. Region Section TR 1 BRAIN TRS (5)Set the coil No. operation. TE Thickness 500 38 1 by 10 Multislice Echo NSA 1 1 2 the PATIENT DATA setting (6)Set the switches on the GC CONT board in the GC unit within the MRI unit to prevent X- and Z-directional gradient fields from being produced. 3sw 1sw : 2 aGND : X GND 4 - 44 111 UE-BM0434 (7) Observe the received signal on the oscilloscope to check that the resonant signal is developed, and ::-.,; then raise the frequency 1 kHz by the FREQUENCY dial ‘3'i on the SYNTH2'board. (8) Observing the received signal on the oscilloscope, adjust the GAIN dial on the TRANS4 board in the RF for the amplitude of this transmitter/receiver signal to be maximized. At this time, the range of the oscilloscope should be minimized as much as possible. (9) Set the GAIN dial for a value of the value adjusted at the step (8) above plus 10%. (10) Readjust .and check the bias (as in item 4.8.1) (since the offset changes if the gain is largely changed). 4.8.2.2Adjusting the gain Adjust the transmission gain for each receiver coil. a First set the phantom and receiver coil at the imaging center as in a real imaging. Type in "48" and press (l)Press the SELECT MODE key. the ENTER key, and type in "8" and press the ENTER key. (2)The following message is displayed: ** RF GAIN ADJUSTMENT ** 1 . . . MANUAL 2 . . . AUTO SELECT= 4 - 45 1 1 2 L2E-BM0434 (3)After typing in "2" and pressing the ENTER key, the following message and RF gains list are displayed and the dial-indicated values currently set in the system Type in a value table are displayed as defaulted. indicated by the dial on the TRANS4 board and press the ENTER key. ** RF GAIN ADJUSTMENT ** DIAL NUMBER=(to be typed in at this step) COIL NUMBER=(to be typed in at step (4)) COIL # 1 RF GAIN xxxxxx COIL # 17 RF GAIN xxxxxx 16 xxxxxx 31 xxxxxx RF pulse gains list (4)After input of the dial-indicated value, type in a coil No. to be adjusted according to the following message displayed, and press the ENTER key. When the CANCEL key is pressed, the operation goes back to the input of the. dial-indicated value (step (3)). COIL NUMBER= (5)After the coil No. is typed in, the automatic gain adjustment is started with display of the following message (blinks during a data measurement) : 4 RF GAIN SEARCH IN PROGRESS When the automatic adjustment has been successfully complete, the following message is displayed: 4 - 46 113 LZE.BM0434 RF GAIN SEARCH COMPLETED _ '".b‘i & The results of the adjustment and the data measured ~1 while changing the gain are graphically displayed, and the RF pulse gains list display is changed. ** RF GAIN ADJUSTMENT ** DIAL NUMBER=X.XX COIL NUMBER=XX RF GAIN SEARCH COMPLETED =XxXx (result of adjustment) RF GAIN 0 Graphic display of measured data (to the right of the RF pulse gains list): - ---------------- 4,. Measured data (max.) I +I +++ -k+ I + + + I + + + ‘+ + + Measured data + (min.) Dial setting (max.) Dial setting (min. I The gain with which the measured data is the maximum is taken as the result of adjustment. Measuring points : 30 points After an error message is displayed, the operation goes back to .the step of input of dial-indicated value in The case the error message is SEARCH RANGE OVER. process is terminated in case of any other error message. 4 - 47 114 L2E-BM0434 When the space key is pressed in the course of the automatic adjustment, the process is suspended with displayed of the following message: CONTINUE (Y/N)= By typing in "Y" and pressing the ENTER key, the By typing in "N" and process is continuously done. pressing the ENTER key, the operation goes back to the step (2). (6)When the automatic adjustment has. been successfully complete, the system waits for an input by the operator with display of the following message: SET OK? (Y/N)= By typing in "Y" and pressing the ENTER key, the results of the adjustment are entered into the system By typing in table and the process is terminated. "N" and pressing the ENTER key, the operation goes back to the step (2). 4 - 48 115 L2E-BM0434 . (7)Error messages list *; No. Error message Meaning Operation 1 MRI UNIT ERROR An error has occurred This after program error ends. in the HRI unit. 2 RF GAIN SEARCH ERROR Time-over -do- Measured data are not -do- (PSC TIME OVER) 3 RF GAIN SEARCH ERROR within the predetermined range even after 5 tries (data overflow or 4 underflow). SEARCH RANGE OVER SET Search DIAL NUMBER (peak cannot be searthe step (input of ched w i t h i n t h e s e a r c h d i a l - i n d i c a t e d value). n x.xx range-over Operation goes back to range (RF pulse gain of 3 2 3 6 7 to 21167)). -Dial-indicated value (maximum) of the graph displayed. displayed at the end E i t h e r “>*’ o r gl<l@ . 4 - 49 116 L2EBM0434 4 s(CAUTION >c 1) If the error item 4 (SEARCH RANGE OVER) has occurred, do as follows: / Taking the Measured data are graphically displayed. graph as reference, change the setting of the dial on the TRANS4 board and the entered dial-indicated value in a direction specified by “>” or “<” in the message. Change them by about 0.5. When the dial setting is changed, the gain settings for other coils having already been adjusted are automatically adjusted so that the total gain is kept constant according to the following equation: (dial setting before change) x (gain setting before change) = (dial setting after change) x (gain setting after change) coils need no readjustment. Therefore, these However, in case the gain after changed exceeds Since 32767, the correction will result in 32767. this result is incorrect, it is necessary to readjust the other coils. 2) If a time of more than 30 minutes has passed from the If the last scan, do a scanogram measurement once. I central frequency shows a large deviation, no signal measurement can be done as the case may be. 4 - 50 1 1 7 L2E-BM0434 . 4.8.3 Offset (1)As similarly to the case of the Bias adjustment (4.8.1- J\ short-circuit the tip of the oscilloscope probe (211, to make it in loop and set it vertically in the center of the transmission coil. (Do not use the output monitoring connector of the RF power Amp.) (2)Perform the continuous measurements by using the signal generation check of the PSC check. Protocol conditions t : Sequence Flip angle(FA) TR NSA : GE : 30 : 100 :2 the waveform on (3)Observing the transmission oscilloscope, adjust the offset dial to make the waveform ratio equal to the undermentioned value. Lock the dial after adjustment. 0 A t 4 - 51 A 22 2 2 -= B 100 1 1 8 LZEBM0434 I k&l@,; 4.9 Adjusting the light localizers -- (1) As shown in Fig. 4.12, install the FOV phantom mount to the head rest. (2) Place the FOV phantom as shown in Fig. 4.13. (3) Adjust the 1ef.k light localizer for the light beam to coincide with the lines Y = 0 and 2 = 0 (as shown in Fig. 4.15). (4) Move the tabletop in the automatic mode by pressing the SET button and take a sagittal scanogram of the FOV phantom under the conditions of HF/S (patient setting), . FOV = 260, slice position = 0 mm and slice thickness = 10 mm). (5) Display an image, and measure in the image the deviation between the center of the image and that of the FOV phantom. In the example B in Fig. 4.15, Y = -2 mm and Z = -2 mm. Change these parameters Y and Z to Y = YO and Z = ZO, respectively. .(6) Return the tabletop 771mm. (7) Set the right and left light localizers to Y = YO and Z = zo. (8) Move the tabletop as indexed again and check that the image center is Y = YO and Z = ZO. (9) Place the FOV phantom as shown in Fig. 4.14. (10) Move the tabletop as indexed and take a coronal scanogram of the phantom under the same conditions as in (4) above. (11) Set the phantom so that the image of the FOV phantom is straight circumferentially. 4 - 52 119 L!2E-BM0434 (12) Adjust the median-line light localizer for the light beam to coincide with the line of X = 0. Adjust the position and alignment referring to Fig. 4.17. . ‘% e‘i (13) Move the tabletop as indexed and take a coronal scanogram under the same conditions as in (4) above. (14) Display an image and measure in the image the deviation between the image center and that of the FOV phantom. In the example B in Fig. 4.16, X = 2. (15) Return the tabletop 771 mm. (16) Set the median-line light localizer to X = X0; (17) Move the tabletop as indexed and make sure that the image center is X = X0. 4 - 53 1 2 0 1.2E-BM0434 . Fig. 4.13 Fig. 4.14 Fig. 4.15 4 - 54 1 2 1 LZE-BM0434 . k ., 4 - 55 122 L2E-BM0434 socket head bolt M4 hex. socket head bolt M4 hex. socket head bolt 4 I sn I /X=0 Fig. 4.17 Adjustment of Medeian-line Light Localizer v Circumferential direction I x=+2 4.10 Adjusting the gradient field strength (FOV) 4.10.1 FOV adjustment for spin echo (1)Set the FOV phantom in the center of the craniocervical receiver coil (coronal plane) using the phantom mount, and move it to the center of the magnetic field. (2)Take scanograms of planes. the transverse and sagittal (3)Make sure that the gradient of the phantom is within 2 deg. with respect to the coronal plane. Also check that the phantom center is not deviated more than 10 . mm from the‘center of the magnetic field. (4)Next, set scan -parameters, and make a scan of a coronal plane as positioned in the scanogram image. The angle for positioning should be zero. SEQUENCE 2D/3D SEQ. No REPHASE PHASE DIR REJECTION RECTANG ANTI WRAP PRE SAT FOV TR TE THICKNESS MULTI SLICE' ECHO NSA PROJECTION : SE :2D :3 : OFF : R-L : OFF : OFF : OFF : OFF : 260 : 200 : 38 : 10 :l : 1 : 1 : 256 4 - 57 124 L2GBM0434 . (5) Magnify 2 .times thus obtained. the central portion of the image Make sure that the gradient of the line connecting the pins is less than f2 deg. and measure the Xdirectional distance X. The measurement should be done using the check program No. 31 FOV C/K. Y X Coronal image P (6) Adjust 9VR on the G CONT2 board in the MRI unit for the above-mentioned distance to be 100.0 +/- 0.5 mm. after Record the values the (X p encoded) adjustment. (7) Change the PHASE DIR. from R-L to H-F. (8) Take a scan of the coronal plane again, and magnify two times the central portion of the image. (9) Measure the Y-directional distance Cy between the The measurement should be done using the pins. check program FOV C/K. r) (10) Adjust 8VR on the G CONTZ board for the distance ey to be 100.0 +/- 0.5 mm. Record the values (Y P encoded) after the adjustment. (11) Measure the X-directional distance Px between the pins. 4 - 58 125 L2E-BM0434 (12) Rewrite the X-directional parameters of F-ENCD into the value derived from the following equation under T+., the servicwlj the GC ADJUSTMENT program among programs: r----7 ‘X L-J 100 r----7 L-A \ Value z Default measured at step (11) value (13) Change the PHASE DIR from H-F to R-L. (14) Make a scan of a coronal plane again, and magnify two times the central portion of the image. (15) Measure pins. the Y-directional distance P, between the i (16) Rewrite the Y-directional parameters of F-ENCD into the value derived from the following equation under-) the service the GC ADJUSTMENT program among programs: r----i LA‘X 100 r----7 :-,: -4 2 \ Value Default measured at step (15) value (17) Set the FOV phantom in the sagittal plane. (18) Take scanograms of the transverse and coronal planes, and make sure that the gradient of the phantom with respect to the sagittal plane is less than +_2 deg. Also check that the phantom center is not deviated more than f10 mm from the center of the magnetic field. 4 - 59 126 L2E-BM0434 (19) Set the following scan parameters, and take an image plane as positioned in the in the sagittal The angle for positioning should, be scanogram. zero. : SE SEQUENCE : 2D 2D/3D :3 SEQ. No : OFF REPHASE : A-P PHASE DIR : OFF REJECTION : OFF RECTANG : OFF ANTI WRAP : OFF PRE SAT : 260 FOV : 200 TR : 38 TE : 10 THICKNESS :l MULTI SLICE :l ECHO :l NSA : 256 PROJECTION . (20) Magnify 2 times the central portion of the image thus obtained. Make sure that the gradient of the line connecting the pins is less than +2 deg. and measure the Zdirectional distance 4,. e, Y - -.- _.-.-.-.-.- P I : iI i i i i i +_2# M max .-.-.-.-.- f L i I i 4 - 60 127 L2E-BM0434 (21) Adjust 7VR on the G CONTZ board in the MRI unit for the above-mentioned distance to be 100.0 +/- 0.5 mm. ‘-.'i, 4 ‘; Record the values (z P encoded) after the adjustment. (22) Change the PHASE DIR. from A-P to H-F. (23) Take a scan of the coronal plane again, and magnify two times the central portion of the image. (24) Measure the Z-directional distance Pz between the pins. (25) Rewrite the Z-directional parameters of F-ENCD into the value derived from the following equation under the GC ADJUSTMENT program among the service programs: r----1 :-,, ‘X 4 \ 100 r----i :-,,: \Value measured at step (24) .t Default value 4 - 61 1 2 8 LZEBM0434 4.10.2 FOV adjustment for gradient echo This adjustment must be carried out after adjustment of gradient magnetic field strength (FOV adjustment of SE) by use of a spin echo (SE) image. (1)Align the FOV phantom at the center of the receiving coil for HS/MR head (on coronal plane) by using the phantom mounting table and introduce it to the center of magnetic field. (2)Take each scanogram image on the transverse and sagittal planes. (3)Make sure that the gradient of phantom is within +2O Also confirm that the center of the coronal plane. of phantom is not deviated beyond 210 mm from the center of magnetic field. (4)Arrange the conditions for next imaging and determine the position on a scanogram image. Take the image on For positioning, angle must be the coronal plane. set at 0. SEQUENCE 2D/3D SEQ. No REPHASE PHASE DIR REJECTION RECTANG ANTI WRAP PRE SAT FOV TR TE FA THICKNESS MULTI SLICE N S A PROJECTION : GE : 2D :2 : OFF : R-L : OFF : OFF : OFF : OFF : 260 : 200 : 23 : 90 : 10 : 1 :l : 256 4 - 62 129 L2E-BM0434 .. (5)Enlarge the. central enlargement ratio of 2x. area of the image at an "?. &1 Make certain that the line, which connects pins, has a slope of f2O max., and measure the distance {x between pins in the X direction and that Py between pins in the Y direction. Measurement must be performed via FOV C/K. 2 i I -.-.-.-.-.-.I-- i. 2 2O max. 1Px .-.-.-.-.-.-_ 1 ! i i 3 I I 4 I i X Coronal image e (6)Rewrite the X-direction parameter of P-ENCD in gradient echo to a value given by the formula below under the service program "GC ADJUSTMENT." r----i L-J ‘X 10‘0 r----i :--: \ Measured value ex in (5) \ Default value .. 4 - 63 _ 130 LZE-BM0434 (7) Rewrite the Y-direction parameter of F-ENCD in gradient echo to a value given by the formula below under the service program "GC ADJUSTMENT." r----l 100 :,-J ‘X r----7 :--: \ z Measured value ey in (51 value Default (8) Make a scan again. Measure CX and 4y, and make sure that the measured values satisfy 100 + 0.5 mm. In case the measured values are outside the range, effect steps (6), (7) and (8) again. After accuracy values. confirmation, record PX and Cy (9) Change the PHASE DIR from R-L to H-F. (10) From the image on the coronal plane again and enlarge its central area at an enlargement magnification of 2x. (11) Measure the distance PX between pins in the X direction and Py between pins in the Y direction. (12) Rewrite the X-direction parameter of F-ENCD in gradient echo to a value given by the formula below under the service program "GC ADJUSTMENT." r - - - - 1 L-J ‘X 100 r - - - - i I :--J \ z Default 4 Measured value Qx in (11) value - 6 4 131 NE-EM0434 (13) Rewrite the Y-direction parameter of P-ENCD in -gradient echo to a value given by the formula below +I'.]\, under the service program "GC ADJUSTMENT." r--w-, I L-,--J 4 'X 100 r----i L.-A Measured value ey in (11) Default value \ z (14) Present the image again, and measure Cx and PY to make sure that the measured values satisfy 100 20.5 mm. In case the measured values are -outside the range, effect'steps (12), (13) and (14) again. After accuracy values. confirmation, record lx and Py (15) Set the FOV phantom on the sagittal plane. I (16) Take each scanogram image on the transverse and coronal planes. Confirm that the slope of phantom is within +2O of the sagittal plane. Also ascertain that the center of phantom is not deviated beyond 210 mm from the center of magnetic field. 4 - 65 132 LZE-BM0434 -"y; Make (17) Arrange the conditions for next imaging. positioni?g on a scanogram image and take the image For positioning, angle must on the sagittal plane. be 0. SEQUENCE 2D/3D SEQ. No REPHASE PHASE DIR REJECTION RECTANG ANTI WRAP PRE SAT FOV TR TE FA THICKNESS MULTI SLICE NSA PROJECTION GE : 2D :2 : OFF : A-P : OFF : OFF : OFF : OFF : 260 : 200 : 23 : 90 : 10 :l :l : 256 : (18) Enlarge the central area of the image at an Make certain that the enlargement ratio of 2x. line, which connects pins, has a slope of +2O max., and measure the distance Pz between pins in the 2 direction. -max. 4 - 66 _. 133 L2G-BM0434 (19) Rewrite the Z-direction parameter of P-ENCD in gradient echo to a value given by the formula below --,,%i w'j under the service program "GC ADJUSTMENT." r----7 L-A 100 ‘X r----1 :-A \ Measured 2 value in (18) Default value (20) Make a scan again. Measure t?~, and make sure that the measured values satisfy 100 +0.5 mm. In case the measured value is outside the range, effect steps (19) and (20) again. After accuracy confirmation, record Qz value. (21) Change PHASE DIR. from A-P to H-F. (22) Make a scan on the sagittal plane again and enlarge the central area at an enlargement magnification of 2x. (23) Measure the direction. (24) Rewrite distance the I d-i pz between pins in the Z- Z-direction parameter of F-ENCD in gradient echo to a value given by the formula below under the service program "GC ADJUSTMENT." r----1 L--J ‘X 3 100 r---- 1 I I L-,--J \ z Heasur ,ed value i n (231 Default value (25) Make a scan again. Measure Qz and make sure that it is within a range of 100 f0.5 mm. If outside the range, effect steps (24) and (25) again. After accuracy confirmation, record the 4'~ value. 4 4 - 67 134 L2E-BM043P 4 . 4.11 Signal phase adjustment For imaging by the gradient echo method, it is necessary to adjust for minimizing the difference in phase between the O" and 90° signals in the RF receiver circuit since Note that the phase difference causes wavy artifacts. this adjustment is not necessary during installation When because it has been done at factory before shipped. the circuit board replacement or the similar maintenance has been done in the RF transmitter/receiver, this adjustment is required. -Adjustment (l)Set the Ni solution phantom middle (No.3) to the HS/MR Head coil as shown below. Phantom No.3 4 \ Set the phantom at the pos ition s h i f ted slightly downward from them a g n e t i c f i e l d centrer. I field ‘Phantom center (2)Set the imaging factors or conditions as follows: Sequence Scan plane FOV TR TE FA THICKNESS MULTISLICE ECHO NSA PROJECTION : GE : TRS : 260 : 300 ms : 35 ms : 9o” 10 mm :l :l :2 : 128 : A 4 - 68 135 LZE-BM0434 (3)Adjusting the phase adjustment trimmer condenser on the AF AMP module in the MRI unit, perform scans so that the differential area between the real image and imaginal image becomes smallest. . ,Imaginal i m a g e (shaded area) For phase adjustment 0 I Adjust one of them. (Phase 1 or phase 2) -Real AF AMP module front panel image Image example Adjust the angle of the cross position of the screw head of the trimmer condenser and scan an image, then compare the real and imaginal images. First turn the condenser by about 45O (l/8 turn) right and leftward and fix the turning direction to find approximate optimum position. Next, adjust the condenser by a fine angle (l/16 turn) each to reach the smallest area of the imaginal image. 4 4 - 69 136 L2E-BM0434 -\; 4: 4.12 Setting of date displaying order It is possible to change the displaying order of the date digits (year, month and day) displayed with LEDs, and on When it is the monitor, on the central control console. desired to change the date displaying order, set the DIP switches on the PCON-C board (CC4165) inside the control console according to any of the displaying orders shown in Fig. 4.18. All the dates displayed with the LEDs, and on the monitor, on the control console are changed in order Also by changing only the settings of these DIP switches. the date input order in adjustment of the clock (SET CLOCK operation), birth date, etc. follows the date displaying order after changed. DIP switches at address 10F on the PCON-C board (CC4165) in the central control console: : aYear, month, day 1 aMonth, d a y , year ODay, month, year 2 1 2 ON 3 4 5 6 7 8 3 4 5 6 7 8 4 5 6 7 8 Fig. 4.18 Setting of Date Displaying Order a 4 - 70 137 L2E-BM0434 4.13 Checking the intercom (inter-communication) (l)Talking from control room to MRI room 3 The sound picked up by the microphone in the control room is conveyed to the MRI room by pressing the Adjust the sound volume by PATIENT INTERCOM button. adjusting the control OC + GANTRY located at the right bottom of the keyboard. As the PATIENT INTERCOM button room is is pressed, the sound from the MRI automatically decreased. (2)Talk from MRI room to control room Make sure that the sound volume can be changed by adjusting the potentiometer on the control panel or the control GANTRY + OC located at the right bottom of the keyboard. If any of these controls is set at MIN Also the sound position, no sound will be heard. volume obtainable by adjusting the potentiometer 1 depends upon the position at which the control on the Set the control at a -It right bottom of the keyboard. position for an appropriate sound volume which can be obtained by adjusting the potentiometer. . 4 Rotary potentiometer (on the operation panel) Pot. position for best hearing Pot. position for best hearing Rotary pot. at bottom of Keyboard panel (3)Auto voice control Adjust volume of Auto voice by the volume of OC + GANTRY (AV) located at the right-bottom of the keyboard. 4 - 71 138 L2E-BMOd34 e 4.14 Write of hospital name into magnetic disk The hospital procedure: name can be entered in the following (l)Press the SELECT MODE key to select the process menu. Type in "48" and press the ENTER key to select the service mode. (2)Select and type in "2" from the service program menu and press the ENTER key to select HOSPITAL NAME ENTRY. (3)"HOSPITAL NAME=" is displayed. Then type in a hospital name with a maximum of 20 characters. - For correction of any incorrect input, press the' ERASE key to erase characters one by one or press the CANCEL key to erase characters in each line. -After entry of the hospital name, make a scan of a image on the phantom. Display a reconstructed monitor and check that the hospital name entered is displayed in the lower right portion of the monitor. If you se: it there, the hospital name has been surely entered. 4 - 72 139 L2E-BM0434 4.15 Check of image analysis software According to the instructions given in the Instruction Manual, effect the following operations for image analysis to check that they are successfully completed: .A? 4.) -Level detection -Background elimination -Positive-negative reversal -Nonlinear windowing - T2 calculation - Tl calculation -Window selection -Multi-framed display - Image magnification - Image rotation -Right-left reversal -Measurement of distance and angle between 2 points -Profiling - ROI setting I( -Scale display - Edge enhancement/smoothing -Comment display -Histogram -Pixel value display - Subtraction/addition -Nonlinear window display and entry - Simultaneous display and selection of positioning images . -Auto window setting -Synthesize image -Gray scale display and selection -Text page display - MPR - ECG display -Uniformity check - FOV check - S/N check 4 - 73 140 lL!E-BP40434 4.16 Check of ECG-gated measurement - Make a measurement according to the instructions given in the instruction manual for the ECG-gated measurement to check that the ECG waveform is stably displayed and the However, it should measurement is done as stably gated. be noted that some noises appear in the ECG waveform during a measurement. 4 - 74 141 L2E-BM0434 Section V Check of Image Quality 5.1 Check of FOV accuracy 5.1.1 SE sequence (l)Press the PATIENT DATA key to enter the patient name, and set the patient orientation H, patient's posture S and coil No. (2)Set the FOV phantom in the center of the HS/MH head receiver coil (in the transverse plane) using the phantom mount, and move it to the center of the magnetic field. (3)Take scanograms in a sagittal and coronal planes. (4)Make sure that the gradient of the phantom with respect Also to the transverse plane is less than +/-2 deg. check that the distance between the center of the 4.j phantom and that of the image is less than +/-lo mm. These checkings are to be done by pressing the DIST key to select the image analysis program for measurement of the distance and angle between two points. f2" max. (5)Take a scanogram in the transverse plane. (6)Check that the rotation of the phantom is within +/-2 deg. f2O max. r; - 1 142 LZEBM0434 (7)Set the following scan parameters and position a scan For this plane at the center of the phantom. positioning, set the angle 0 and the positional tolerance 0 +/- 10 mm. : : 2D/3D a: SEQ. No. : REPHASE : PHASE DIR. : REJECTION : RECTANG. : ANTI-WRAP : PRESAT. : FOV : TR : TE : THICKNESS : MULTI SLICE : ECHO : NSA : PROJECTION : SLICE PLANE SEQUENCE TRS SE 2D 3 OFF A-P OFF OFF OFF OFF 260 200 38 10 1 1 2 256 (8)Make a scan and magnify the central portion of the phantom image with a magnification factor of 2 under the image analysis program for image magnification selected by pressing the MAG button. (9)Measure the distance & and Qx (in mm) between the centers of the pins in the magnified phantom image. X (frequency 5 - 2 encoding) 143 L2E-BM0434 The measurement is to be done using the analysis y. program No. 42 "FOV C/K. ".:I (a)Press the ANALYSIS key, type in “42” and press the ENTER key to select the FOV C/K program. l (b)A square frame enclosing the phantom pin pattern is If the pin pattern is out of the frame, displayed. operate the trackball to change the frame size so that the pin pattern fully enters the frame. (c)By pressing the ENTER key, the distance between the centers of the right and left, top and bottom pin patterns and the angle of intersection between the lines drawn between the right and left pins, and between the top and bottom pins are displayed as follows: X(F-ENCD)=OOO.O mm) + Inter-pin distance in X direction (in Y (SLICE)=***.* Z(P-ENCD)=OOO.O + Inter-pin distance in Z direction (in mm) < =OOO.O + Angle of intersection 8 between lines drawn between right and left pins, and between top and bottom pins (in degrees) NOTE : F-ENCD : Frequency encoding direction P-ENCD : Phase encoding direction SLICE : Slice direction (to always displayed with ***.*) 5 - 3 be 1 4 4 JLSBM0434 (10) Using the following equation, determine FOV accuracy for each of the Z and X directions: FOV accuracy (%) = (measured distance in mm - lOO)/lOO x 100 (11) The FOV accuracy should be within the standard value in item 6.5. (12) Write the measured FOV accuracies into the recording sheet. (13) Set the phantom in a sagittal plane and move it to the center of the magnetic field. (14) Take scanograms in a coronal and transverse planes. (15) Check that the gradient of the phantom with respect Also to the sagittal plane is less than +/-2 deg. check that the distance between the center of the phantom and that of the image is less than +/-lo mm. (16) Take a scanogram in the sagittal plane. (17) Make sure that the phantom rotation is less than +/-2 deg. (18) Set the following scan parameters and position a For this slice plane at the center of the phantom. positioning, set the angle 0 and the positional tolerance 0 +/- 10 mm. 5 -4 145 L2E-BM0434 . 4 : SAG : SE : 2D : 3 : OFF : A-P : OFF : OFF : OFF : OFF : 260 : 200 : 38 : 10 :l :l : 2 : 256 SLICE PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI -WRAP PRESAT. FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION (19) Make a scan and magnify the central portion of the*' phantom image with a magnification factor of 2. (20) Measure the distance 4y and 4z (in mm) between the centers of the pins in the magnified phantom image. 1 (21) * Z (phase encoding) Determine FOV accuracy directions. for each of the Y and 2 (22) The FOV accuracy should be within the standard value in item 6.5. 5 - 5 1 4 6 L2E-BM0434 (23) Write the measured FOV accuracies into the recording sheet. (24) Set the phantom in a coronal plane and move it to the center of the magnetic field. (25) Take scanograms in a transverse and sagittal planes. (26) Check that the gradient of the phantom with respect to the sagittal plane is less than +/-2 deg. and also that the distance between the center of the phantom and that of the image is less than +/-lo mm. (27) Take a scanogram in the coronal plane. (28) Make sure that the phantom rotation is less than t/-2 deg. (29) Set the following scan parameters and position a For this slice plane at the center of the phantom. positioning, set the angle 0 and the positional tolerance 0 t/- 10 mm. SLICE PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI-WRAP PRESAT. FOV TR TE THICKNESS MULTI SLICE., ECHO NSA PROJECTION : COR : SE : 2D :3 : OFF : R-L : OFF : OFF : OFF : OFF : 260 : 200 : 38 : 10 :l :l :2 : 256 5 - 6 1 4 7 L!2E-BM0434 (30) Make a scan and magnify the central portion of the I:.,,, phantom image with a magnification factor of 2. w-f (31) Measure the distance 4x and Qy (in mm) between the centers of the pins in the magnified phantom image. -.-.dI i cx l I ; C Y -t -.-.d- X (phase encoding) (32) Determine FOV accuracy directions. for each of the X and Y (33) The FOV accuracy should be within the standard value in item 6.5. (34) Write the measured FOV accuracies into the recording -YI sheet. l - 5.1.2 GE sequence (l)With the phantom set in the same manner as for the sequence SE, make scans in the transverse, sagittal and coronal planes with the following scan parameters: SEQUENCE : GE : 2D 2D/3D SEQ. No. : 2 REPRASE : OFF : OFF REJECTION : OFF RECTANG. : OFF ANTI-WRAP : OFF PRESAT. : 260 FOV : 300 TR : 23 TE FA : 90 THICKNESS : 10 MULTI SLICE :l ECRO :l NSA : 2 PROJECTION : 256 PHASE DIR. for scan in transverse plane : A-P : A-P PHASE DIR. for scan in sagittal plane : R-L PHASE DIR. for scan in coronal plane (2)Similarly to the sequence SE, determine FOV accuracy for each of the planes. (3)The FOV accuraiy.should item 6.5. be within the standard value in (4)Write the measured FOV accuracies into the sheet. 5 - 8 recording 1 4 9 UEBM0434 5.2 Check of distortion (l)Magnify two times the central portion of the phantolGY image, in the transverse plane through the FOV phantom, which has been taken at step (8) of item 5.1 "Check of FOV accuracy" (FOV adjustment for spin echo). (i!)Using the analysis program No. 42 "FOV C/K (FOV check program, measure the intersecting angle between the lines drawn between the centers of the right and left pin patterns and top and bottom pin patterns in the phantom image. (3)The intersecting angle 8 should be within the standarc' 4 value in item 6.5. (4)Write the intersecting angle 8 into the recording sheet. (S)Similarly, measure the intersecting angles 8 in the first echo images, of the sagittal and coronal planes, which have been taken at steps (19) and (30) in item 5.1 "Check of FOV accuraby8;, and write them into the recording sheet. 5 - 9 150 L2EBM0434 5.3 Signal-to-noise (S/N) ratio 5.3.1 Check of S/N ratio when HS/MR head receiver coil is used (l)When the S/N ratio is checked using a phantom, it depends upon the temperature of the phantom. If the temperature of a room where the phantom is stored changes, it takes several hours until the changed room temperature coincides with the phantom temperature. Therefore, for check of the S/N ratio, an adjusting phantom (Ni-solution phantoms) should be left for several hours in the MRI room of which the temperature is set in a range of 20°C to 25OC before the check is to be done. the temperature of the MRI room should preferably be 25OC. Keep the set room temperature as constant as possible. The temperature variation should desirably be less than +/-1OC. (2)Press the PATIENT DATA key to enter the patient name, and set the patient orientation H, his posture S and coil No. (3)Place the adjusting phantom middle) on the head rest. (Ni-solution phantom, Ni solution phantom No.3 (165 dia.) 'Abdomen immobilizing (20 mm thick) matt Example phantom setting 5 - 10 151 IZE-BM0434 (4)Slide the receiver coil to a position where it abuts Have the light mark defined by the ';* the head rest. ', light localizer coincide with the slice mark on the -"y! receiver coil (the light mark cannot be made coincident Press the SET with the cross (+) mark on the phantom. button to set the phantom in the center of the magnetic field. (5)Take scanograms of a sagittal and transverse planes. (6)Make sure that the center of the phantom is less than 10 mm from the center of the image in the scanogram of the sagittal plane. Also check that in the scanogram of the transverse plane, the center of the phantom image is less than +/-lo mm from the center of the image. Sagittal image (7)Set the following scan parameters, and make three scans without any positioning: SLICE PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI - WRAP PRESAT. C.POSITION : TRS : SE : 2D :3 : OFF : A-P : OFF : OFF : OFF : OFF : 0 l 5 - 11 152 L2E-BM043j FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION : 260 : 500 : 40 : 5 : 1 : 1 :2 : 256 (8) Press the DISPLAY button to display the scanned image. (9) Using the analysis program No. 43 "SNR C/K (S/N ratio check program), measure the signal-to-noise ratio. (a) Press the ANALYSIS key, type in "43" and press the ENTER key to select the SNR C/K program. (b) ROIs (A to E) are set as shown below: An S/N ratio calculated by the following equation is displayed in the form of SNR = 00.0. S/N = (Mean value of E) / (Mean SD of A, B, C and D) (10) Determine the S/N ratios of the three scanned images and take the arithmetic mean of these three S/N ratios as the S/N ratio. 5 - 12 153 L?E-BM0434 (11) Measure the phantom temperature, and determine a reference range of the S/N ratio from the standard -Y=:; performance data as in item 6.5. The reference rang&j varies depending upon the temperature since the S/N ratio of the phantom image depends upon the phantom temperature. (12) The measured S/N ratio should be within the reference range determined from the above graph. (13) Write the results of S/N ratio measurement, phantom range of S/N ratio temperature and reference (visually determined from the graph) into the For measuring the phantom recording sheet. temperature, read the thermometer attached on the The green figure on the thermometer phantom. indicates the phantom temperature. (14) Set the following scan parameters, make two scans and determine the S/N ratios in the multi-slice and multiecho modes. 4,; SLICE PLANE : TRS SEQUENCE : SE : 2D 2D/3D SEQ. No. :3 REPHASE : OFF, PHASE DIR. : A-P REJECTION : OFF RECTANG. : OFF ANTI-WRAP : OFF PRESAT. : OFF C.POSITION : 0 : 260 FOV TR : 1500 TE : 40 THICKNESS :5 INTERVAL : 10 MULTI SLICE :8 : 2 ECHO :2 NSA PROJECTION : 256 5 - 13 154 KE-BM0434 (15) Determine the scanned images, and a mean S/N ratio of the first echo images and also a mean S/N ratio of Take these mean S/N ratios the second echo images. for the respective echoes. (16) The measured S/N ratio should be within the reference range determined from the above graph. (17) Write the results of S/N ratio measurement, phantom ratio temperature and reference range of S/N (visually determined from the graph) into the recording sheet. 5 - 14 1 5 5 L2E-EM0434 5.3.2 Check of S/N ratio when HS/M.R body coil is used ";,' Similary to the check of the head coil, determine the S/N+,) ratio. (1)Place the adjusting phantom (No.4, 212 dia.) and cover the phantom with the rece.iver coil. phantom No.4 / Fix the phantom with abdominal immobilization belt 15mm t h i c k m a t \ 50mm thick abdominal immobilizatio Example phantom setting (2)Set the following scan parameters and make three scans without any positioning. SLICE PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI WRAP PRESAT. C.POSITION FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION : TRS : SE : 2D :3 : OFF : A-P : ON : OFF : OFF : OFF : 0 : 320 : 500 : 40 : 10 :l :l :2 : 256 5 - 15 156 L2E-EM0434 5.3.3 Check of S/N ratio when joint coil is used Similarly to the 4check of the head coil, determine the S/N ratio. (1)Place the adjusting phantom (No.2, 121 dia.) and cover the phantom with the receiver coil. phantom Abdomen immobili P i l l .ou matt (70 J itatt Example phantom setting (2)Set the following scan parameters and make three scans without any positioning. SCAN PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI-WRAP : TRS : SE : 2D :3 : OFF : A-P : OFF : OFF : OFF PRESAT. : OFF C.POSITION FOV TR : 0 TE THICKNESS MULTI SLICE ECHO NSA PROJECTION : 260 : 500 : 40 :5 :l : 1 :2 : 256 5 - 16 157 L2E-BM0434 5.3.4 Check of S/N ratio when large flexible body coil is used Similarly to the check of the head coil, determine the S/IQ&:/ ratio. (1)Place the adjusting phantom (No.4) phantom with the receiver coil. and cover the (a)connector No. 1 5Omm thick 1 I ----se-- (b)Connector / I phantom No. 4 1J No. 2, 3 phantom No. 4 5omm mat 5O~mn thick mat 15mm thic mat / (double) (c)connector No. 4 70mm thick pillow mat phantom No. 4 5Omm thick mat Omm thick rat 15mm thick Jmat (double) Example phantom setting 5 - 17 h&)' 158 L2E-BM0434 (2)Set the following scan parameters without any positioning. SCAN PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI - WRAP PRESAT. C.POSITION FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION and make three scans : TRS : SE : 20 :3 : OFF : A-P : ON : OFF : OFF : OFF : 0 : 320 : 500 : 40 : 10 : 1 : 1 :2 : 256 5 - 18 159 L2E-BM0434 5.3.5 Check of S/N ratio when small flexible body coil is used ,( :-"+\ & Similary to the check of the head coil, determine the S/h ratio. (1)Place the adjusting phantom (No.4, No.31 and cover the phantom with the receiver coil. (a)connector No.1 / phantom No.3 70mm thick pillow mat (b)connector No.2,3,4 /phantom No.4 Example phantom setting 5 - 19 160 L2E-BM0434 (2)Set the following scan parameters and make three scans without any positioning. SCAN PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI WRAP PRESAT. C.POSITION FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION : TRS : SE : 2D :3 : OFF : A-P : ON : OFF : OFF : OFF : 0 : 320 (260mm in case connector No.1) : 500 : 40 : 10 (5mn in case connector No.1) : 1 : 1 :2 : 256 5 - 20 . 161 L2E-BM0434 5.3.6 Check of S/N ratio large extremity coil is used Similary to the check of the head coil, determine the S,-) ratio. (1) Place the adjusting phantom (No.2) phantom with the receiver coil. , 70n IUII thick pi‘llow m. a t phantom and cover the No.2 immobilization 15mm thick mat Same in 5Omm Example phantom setting (2)Set the following scan parameters and make three scans without any positioning. SCAN PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI WRAP * PRESAT. C.POSITION FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION : TRS : SE : 2D : 3 : OFF : A-P : OFF : OFF : OFF : OFF : 0 : 260 : 500 : 40 : 5 :l :l :2 : 256 5 - 21 1 6 2 LZE-BM0434 5.3.7 Check of S/N ratio when small extremity coil is used Similary to the check of the head coil, determine the S/N ratio. (1)Place the adjusting phantom (No.2) phantom with the receiver coil. 70mm thick and cover the 2Omm thick abdominal immobilization 15mm thick mat Example phantom setting (2)Set the following scan parameters and make three scans without any positioning. SCAN PLANE SEQUENCE 2D/3D SEQ. No. REPHASE PHASE DIR. REJECTION RECTANG. ANTI-WRAP PRESAT. C.POSITION FOV TR TE THICKNESS MULTI SLICE ECHO NSA PROJECTION l : TRS : SE : 2D :3 : OFF : A-P : OFF : OFF : OFF : OFF : 0 : 260 : 500 : 40 :5 :l : 1 :2 : 256 5 - 22 163 L!2E-BM0434 5.4 Check of background noise (1)Compare the images obtained by scans taken during the S/N A+ d..l ratio (signal-to-noise ratio) check and images of which the S/N ratio were measured at the time of shipment from factory and which are stored in the OD accessory to the system (these images were obtained by scans with the same parameters). (2)Adjust the window level and width so that the back-ground of the image is seen, and referring to the images taken at the shipment from factory, make sure that the image is free from zonal noise, streak noise and other artifacts. (3)Write into the recording sheet the state of the image background such as existence or nonexistence of zonal noise, streak noise or.other artifacts. .a .- I. I f Streak noise 5 - 23 Zonal n o i s e 164 L2E-BM0434 Section VI ReferenceMaterials --. 6.1 Connection between multi-format camera and central control console The connection between the multi-format camera and central control console can be connected so that the power switch of the multi-format camera can be turned ON and OFF in The connection can, interlock with the MRI system. however, be made only for AC 200V power supply voltage. Disconnect terminator, and connect coaxial cable instead. Uultiformat camera Coaxial cable, multiformat camera Provide terminator Filter GND cable, accessory to - box I T o BSRI u n i t 6 -1 165 L2E-EM0434 6.2 Filming of test pattern by multiformat camera 6.2.1Procedure (l)Read an image from the following image OD accessory to thesystem: MRP SMPTE TEST PATTERNS (2)Press the DISPLAY button to display the image. (3)Press the SELECT MODE key, type in "7" and press the ENTER key to start the AUTO WINDOW selecting process. In response to a message displayed, type in "1" and press the ENTER key to select the normal window. (4)Press the U/L, WINDOW key to take a preset window for display of the test pattern image. (5)Select the standard format (number of frames) of the Unless otherwise specified, multiformat camera used. :' the format is 9 frames (on 11x14" film), 4 or 6 frames wli (on 8x10" film). (6)Write the date of filming into the image by the comment process or write it in the circumferential margin of the film using a magic ink. NOTE: l.The filming is to be done when more than 30 min has passed after the power supply to the multiformat camera is turned on. 2.After an intended filming, take the similar procedure to that at step (3) to start the AUTO WINDOW selecting process, thereby restoring the auto window state. 6-2 1 6 6 UE-BM0434 6.3 Performance data 6.3.1 FOV accuracy (1)SE sequence Less than +/-1.5% for all the transverse, sagittal and coronal planes (2)GE sequence Less than +/-1.5% for all the transverse, sagittal and coronal planes 6.3.2 Geometrical distortion ‘Less than 90 +/- 2.0 for all the transverse, sagittal and coronal planes 6.3.3 Signal-to-noise ratio (S/N ratio) Should be as shown in the graphs given below. 6.3.4 Background noise Compare the images obtained during the S/N ratio check and the S/N ratio check images in the accessory OD (S/N ratio No zonal noise, check done at shipment from factory). linear noise or other artifact should be found in the images. 6 -3 1 6 7 LZE-BM0434 23( 22c m---m-- : : I I : I I : I : I I 41 -----,1,,,-,,,-,,,,1----, -,,,:-,-I I I I \ I I I 210 I I : t I I : t I ,,----I----,,,,,-,,~,,,,,,,,,,,,, I I I I I 200 I I I I - - - - - - 190 i I I i------------~------------- I I : I- - - - - - I : I 1 I : - - - - - - I f I t I r I I I I I I I I I I ------f------i--------- -------- 180 I I : : I I I I I I I I 170 160 -- 150 -.e 0 Lower l&it 0 ,I 20 25 30 Temp. ("C) Reference range of S/N ratio when HS/MR Head receiver coil is used (SE 500/40, 1 slice, 1 echo) 6 - 4 168 L2E-BM0434 , 230 220 210 I I I I , i I I I i I : i I ,,,,,-I,,-,,--,-,,- J ,,,,,,,,,,,,-L,,----, 200 I I I I I I I I I I I I ! I I I I I I I I I I I I I : -m---e ~------------~-------------‘------- 190 I I I I I --s-s- : ------------~------------- 180 i I i I I I I I I I I : I I I I : ------r-------------~-------------' 170 I I ! I I ------- t I : I I : r------I I I 160 150 rc Lower limit 0 0 ., 20 25 30 Temp. ('C) Reference range of S/N ratio when HS/MR Head receiver coil is used (SE 1500/40, 8 slices, first one of two echoes) 6 - 5 1 6 9 ME-BMO434 80 limit _------- 75 0 70 Z ‘1 m 65 60 55 i ------+---------w-I 50 i i ..#-------------c - - - - - - I I c c 0 0 \\ 20 25 Temp. 30 (“C) Reference range of S/N ratio when HS/MR Head receiver coil is used (SE 1500/40, 8 slices, second one of two echoes) f - c *7n 1 ?P.RMrldRd 280 - ; I I ; I 1 I , I I I i i i I 260 - i 0 i i 240 - i , --,,,-A,,,,,,-, i . . I -,~--,,-,-------L------- 220 - ------&--j - - - - - - - i 200 - i I iI I ,,,,,,l,,,,,-,,,,,,J,,,-,,,,,,-,, I I i- - - - - - , .i I I 180 - ------ 3------------~-------------~------- I 160 - ------ f p+& ------8 Lower limit I 140 - I I I I I I I I - 0L 0 I I I * 20 25 30 Temp. ("C) Reference range of S/N ratio when HS/MR body receiver coil is used (SE 500/40, 1 slice, 1 echo) 6 - 7 171 L!ZE-BM0434 280 I I I I I 260 I I I I I I I I I I I - - - - - - a.------------ J-----------I I I I : I f t Upper-limit 240 ---s-e I I I i :- - - -- - I I I I : L- - - - - - I I I : I :- - - - -- I 220 i I I : I I I i i - - - - - - : _-----_----- :---------I I I I I I t I I I 200 180 ----me 160 - - - - - - I I : I- - - - - - I I I : I I I f - - - - - - - - - - - - ~ - - - - - - - - - - - - - l - - -: - - - I 1% - - - - I I Lower limit I I I I I I ~------- I I I i -----a‘-“-‘T’--“--‘-“‘;--“----------t- 0 .. 25 20 30 Temp. ("Cl Reference range of S/N ratio when joint receiver coil is used (SE 500/40, 1 slice, 1 echo) . 6 -a 1 7 2 L2E-BM0434 . .- - 24t I I ; I : : I I ! , : 1 8 ------L------------~------------- 220 I I I I 0 .+ cu L 200 t tn 180 I t I I r 1 I I I I I : - - - - - -I I I , ------ ~pz& ------I I : : : I I I I ,-,-,,I,,-,,,,-,,,,~,-,-,,---,-,, :- - - -- - iI t I t0 I III I !I 1I II I -----+---------------------I------I I I :I I1 II , I, : : ; ----k ----------~------------- I- - - - - - II ,! I I I 160 140 +J m 120 I Lover limit I 1 ------ i -------------~------------- : ;- - - - - - - 100 _----- f----+-- -- - - - - --- - ; - - - - - -cc 0 0 \\ 20 25 30 Temp. ('C) . Reference range of S/N ratio when flexible large body receiver coil is used (SE 500/40, 1 slice, 1 echo) 6 - 9 173 LZE-BM0434 220 I I I I : -- -- -- ~------------~----------------a- 200 I I 0 .+ lu L I I Upper limit , I I / I I ,,,-,,J.----,-,-,a------------I i I I I : Ii I I I 180 --..,,,I---,--,----- j----------- 160 I i I I I I 8 I I ,,,,,-I,,,,,,-----1 I 140 I I I I 120 ------ I I I I L- - - -- - I I : I I L,,,,,-I I : I I I I ~---------- ; -- -- -- - I I I I I I : I I ~~-j I I t I : : ----me -------------~------------f 100 I : ------- I I I t ------- 80 .* 0 \\ cl .a 20 25 Temp. 30 ("C) Reference range of S/N ratio when flexible small body receiver coil is used (SE 500/40, 1 slice, 1 echo) (Connector No.1) 6 - 10 174 L2E-$M0434 240 I I I ; I I ; I I I I I I I I I I : --s---I : I I Upper limit I -------L---------- -4------------I J : : I : 220 I I I : I : I ,-,,-,I-,-,-,,,-,,,~-,,,,,,------------ \ 200 I I I I I I 1 i : r I’ I I I -,---,L-,--,-,--,-,~-,,-,,,--,,,-L 180 : - - - - - - - I I I i I I -----mm I I i I 1 I’ I I t I 160 - - - - - - 140 ~*--j . . I , ---em.. 120 i : : “‘-““““-,“““‘““” I I I I : I I I ----+------- 100 .. I ! - - - - - - I I I 1 . cc 0 I I ;-------------f------- l 0 - - - - - - - 20 25 30 Temp. ("C) Reference range of S/N ratio when flexible small body receiver coil is used (SE 500/40, 1 slice, 1 echo) , (ConnectorNo.2-4) 6 - 11 176 LZE-BM0434 , S t a n d a r d r a n g e o f S/N r a t i o ---------I----------, i --------- ~----.I ; II , i I I I I I t I I I I I I I I I I I ! I - - - t----------:---------:---------i---------~--------I I I I I I I I I I I f I I I I I I I I I f I I I I i i I I I I ““““‘~“^“““~“““---’ I I I I I I I I I . 320 I ------a.--I I I I 300 Upper limit I ----w ~-------------L------1 3' II I I : I I I : I ---,,,I,,,,-,-,-,,,~,,,----------L \ 280 I I I : I I I I I -----me : I 1 I 260 --m-e- .----,,-----,~------------- i I I 1 I I i I : - - - - - - f I I i i I : I --,,,,I,,-,,,,,,,.., I 240 L -0 G -0 220 -mm- t lu + * m ------i-------------i-------------;------- 200 i ------‘-----------,-------------t------T. 180 II 0 0 \. 20 25 30 Temp. ('C) range of S/N ratio when extremity small receiver coil is used (SE 500/40, 1 slice, 1 echo) Reference 6 - 13 1 7 7 L!2E-EM0434 6.4 Recording sheets Adjustment Certificate/Data Recording Sheet 1. 2. 3. Heater current (see item 4.4.1) - Upper system (7.0 to 9.0 A) - Lower system (10.0 to 12.5 A) Magnetic circuit temperature (see item 4.4.2) - Upper magnet thermistor (808 to 844Q) - Lower magnet thermistor (808 to 84452) - Difference between upper and lower (less than 9Q) Static field homogeneity (see item 4.5) 4 4. 5. (less than 60 ppm) Transmitter coil tuning (see item 4.7.1) - Impedance (SOilO~) - (00f100) Phase Receiver coil tuning (see item 4.7.2) solenoid saddle - HS/MR Head - HS/MR Body - Joint/NecK Impedance (200lt100R) Phase (O"flOo) Tuned value (OC to 4E) Impedance (200f100R) Phase (00f100) Tuned value (OC to 4E) Impedance (60~15Ofi) Phase (OOflOO) Tuned value (OC to 36) 6- 14 178 LZE-BM0434 - Large Extremity - Impedance - Phase - Tuned value - Small Extremity - Impedance - Phase - Tuned value - Large Flexible Body No.1 No.2 No.3 No.4 - Small Flexible Body l Connector No.1 - Impedance (550+1ofi) - Phase (215O) - Tuned value (OC-72) 6 - 15 179 I+-ml0434 6. Transmission gain and bias and offset (see item 4.8) - Bias - Gain - Offset Coil No. 1 7. Adjustment value Coil No. l Adjustment value 9 2 10 3 11 4 12 5 13 6 14 7 15 8 16 FOV (see item 4.10) - X axis (P encoded) (100.0f0.2 - Y axis (P encoded) (100.0f0.2 mm) - Z axis (P encoded) (100.0f0.2 mm) 6 - 16 mm) 180 L!%BM0434 GC gain adjustment value: Axis Slicing Spin echo Gradient echo X Y Phase encoding X Y Frequency encoding X .- HZ 10. Center frequency 6 - 17 1 8 1 LZEBM0434 Image quality check recording sheet 1. FOV accuracy check 1.1 Sequence SE - Slice plane Coordinates : FOV accuracy (less than 1.5% in reference value) - Transverse - Sagittal - Coronal X direction : % (frequency encoding) Z direction : % (phase encoding) Y direction : % (frequency encoding) Z direction : % (phase encoding) X direction : % (phase encoding) Y direction : ' % (frequency encoding) 1.2 Sequence GE - Slice plane Coordinates : FOV accuracy (less than 1.5% in reference value) *'j - Transverse - Sagittal - Coronal 2. X direction : % (frequency encoding) Z direction : % (phase encoding) Y direction : % (frequency encoding) Z direction : % (phase encoding) X direction : % (phase encoding) Y direction : % (frequency encoding) Distortion check - slice plane Intersection angle (900+/-2.0° in reference value) 0 - Transverse : - Sagittal : 0 - Coronal : 0 6 - 18 182 L2E-BM0434 3. S/N ratio J 3.1 When HS/MR Head receiver coil is used: 3.1.1 SE 500/40 1 slice, 1 echo - S/N ratio : - Phantom temperature : - Reference range of S/N ratio : - Tuned value : - Gain : - Center frequency : OC max. min. 3.1.2 SE 1500/40 8 slices, 2 echoes * First echo: - S/N ratio : - Phantom temperature : - Reference range of S/N ratio : - Tuned value : - Gain : - Center frequency : OC min. max. min. max. * Second echo: : - S/N ratio - Reference range of S/N ratio : 6 - 19 183 L2E-BM0434 3.2 When HS/MR body receiver coil is used (SE 500/40 echo): - S/N ratio : - Phantom temperature : : - Gain : - Center frequency : 3.3 When joint receiver echo): max. min. coil is used (SE 500/40 - S/N ratio : - Phantom temperature : : - Gain : - Center frequency : 1 slice, -1 OC max. min. - Ref. range of S/N ratio : - Tuned value ."" dj OC - Ref. range of S/N ratio : - Tuned value 1 slice, 1 3.4 When large flexible body receiver coil is used (SE slice , 1 echo): 500/40 Connector I - S/N ratio II I - Tuned value II I- Gain - Phantom temperature I I I I I I I I I I I OC : - Ref. range of S/N ratio : - Center frequency I I min. max. : 6 - 20 184 LZE-BM0434 1 3.5 When small flexible body receiver coil is used (SE 500/40 slice, 1 echo): 1 .- --. No.1 No.2 Connector - S/N ratio - Tuned value - Gain - Ref. range of S/N ratio - - Phantom temperature : - Center frequency : - OC 3.6 When large extremity receiver coil is used (SE 500/40 slice, 1 echo): - S/N ratio : - Phantom temperature : - Ref. range of S/N ratio : - Tuned value : - Gain : - Center frequency : 1 OC min. max. 3.7 When small extremity receiver coil is used (SE 500/40 1 slice, 1 echo): - S/N ratio : - Phantom temperature : - Ref. range of S/N ratio : - Tuned value : - Gain : - Center frequency : 6 - 21 OC min. max. 185 L2E-BM0434 4. Background noise check .-. 6 - 22 166 L!ZE-BM0434