Troubleshooting Platform C HAMILTON-C1 HAMILTON-C2 HAMILTON-C3 HAMILTON-T1 HAMILTON-MR1 Troubleshooting Platform C Doc.no: 624485/00 May 2012 Page 1 of 42 Doc.no: 624485/00 May 2012 Troubleshooting Platform C Table of Contents 1 Copyright ................................................................................................................................................................................... 4 2 Conventions .............................................................................................................................................................................. 5 2.1 2.2 2.3 Typographic Conventions ........................................................................................................................................................................ 5 Expressions ............................................................................................................................................................................................. 5 Foreword.................................................................................................................................................................................................. 5 3 Technical faults ......................................................................................................................................................................... 6 4 Troubleshooting Service Software ........................................................................................................................................ 37 4.1 4.2 4.3 Adjustments and calibrations................................................................................................................................................................. 37 Component Tests .................................................................................................................................................................................. 38 System Tests ......................................................................................................................................................................................... 42 Device’s response alarm or message Alarm or message ............................. page Alarm or message ............................. page Alarm or message ............................. page Battery 1 Calibration required ......................... 21 Battery 1 Defective ......................................... 25 Battery 1 Temperature high ............................ 22 Battery 1 Wrong battery .................................. 24 Battery 2 Calibration required ......................... 22 Battery 2 Defective ......................................... 25 Battery 2 Temperature high ............................ 22 Battery 2 Wrong battery .................................. 24 Battery low ................................................ 23, 24 Battery power loss .......................................... 24 Blower service required .................................. 10 Check Flow Sensor tubing .............................. 12 CO2 calibration needed .................................. 15 CO2 sensor faulty ........................................... 15 Device temperature high ................................ 14 Disconnection on Ventilator side ...................... 7 Fan failure ....................................................... 25 Loss of external power .................................... 20 Oxygen supply failed ......................................... 7 Pressure not released ..................................... 11 Preventive maintenance required ................... 25 Real time clock failure ..................................... 19 Replace HEPA filter ........................................ 17 Safety ventilation 331001 ................................ 30 Safety ventilation 332001 ................................ 30 Safety ventilation 341001 ................................ 30 Safety ventilation 341002 ................................ 30 Safety ventilation 341003 ................................ 30 Safety ventilation 341004 ................................ 30 Safety ventilation 341005 ................................ 31 Safety ventilation 343001 ................................ 31 Safety ventilation 346002… ............................ 31 Safety ventilation 383001 ................................ 31 Safety ventilation 383002 ................................ 31 Safety ventilation 383003 ................................ 31 Safety ventilation 383004 ................................ 31 Safety ventilation 383005 ................................ 31 Safety ventilation 383006 ................................ 32 Safety ventilation 383007 ................................ 32 Safety ventilation 385001 ................................ 32 Safety ventilation 385002 ................................ 32 Safety ventilation 385003 ................................ 32 Safety ventilation 386001 ................................ 32 Technical Event 231002 .................................... 7 Technical Event 231003 .................................... 7 Technical Event 231004 .................................... 7 Technical Event 231005 .................................... 7 Technical Event 231007 .................................... 8 Technical Event 231008 .................................... 8 Technical Event 231009 .................................... 8 Technical Event 231010 .................................... 8 Page 2 of 42 Doc.no: 624485/00 May 2012 Troubleshooting Platform C Technical Event 231012 ................................. 10 Technical Event 231013 ................................. 10 Technical Event 231014 ................................. 10 Technical Event 231018 ................................. 11 Technical Event 231019 ................................. 11 Technical Event 231020 ................................. 11 Technical Event 231022 ................................. 11 Technical Event 232002 ................................. 11 Technical Event 232003 ................................. 11 Technical Event 232005 ................................. 11 Technical Event 232006 ................................. 12 Technical Event 232008 ................................. 13 Technical Event 232029 ................................. 14 Technical Event 232030 ................................. 14 Technical Event 232034 ................................. 15 Technical Event 232035 ................................. 15 Technical Event 233001 ................................. 15 Technical Event 233002 ................................. 15 Technical Event 233003 ................................. 16 Technical Event 233004 ................................. 16 Technical Event 233005 ................................. 16 Technical Event 233006 ................................. 17 Technical Event 243001 ................................. 17 Technical Event 243002 ................................. 17 Technical Event 243003 ................................. 18 Technical Event 243004 ................................. 19 Technical Event 243005 ................................. 19 Technical Event 243006 ................................. 19 Technical Event 244011 ................................. 24 Technical Event 244014 ................................. 25 Technical Event 246002 ................................. 25 Technical Event 246004 ................................. 25 Technical Event 246005 ................................. 25 Technical Event 246006 ................................. 26 Technical Event 246007 ................................. 26 Page 3 of 42 Technical Event 246008 ................................. 26 Technical Event 246009 ................................. 26 Technical Event 246010 ................................. 26 Technical Event 249001 ................................. 26 Technical Event 249002 ................................. 26 Technical Event 249003 ................................. 26 Technical Event 249004 ................................. 26 Technical Event 249010 ................................. 26 Technical Event 249011 ................................. 27 Technical Event 249012 ................................. 27 Technical Event 249020 ................................. 27 Technical Event 281001 ................................. 27 Technical Event 281002 ................................. 27 Technical Event 281003 ................................. 27 Technical Event 281004 ................................. 27 Technical Event 281005 ................................. 27 Technical Event 283001 ................................. 29 Technical Event 283003 ................................. 29 Technical Event 283004 ................................. 29 Technical Event 283005 ................................. 29 Technical Event 283006 ................................. 29 Technical Event 283007 ................................. 29 Technical Event 284002 ................................. 29 Technical Event 284003 ................................. 29 Technical Event 284004 ................................. 29 Technical Event 285001 ................................. 29 Technical Event 285002 ................................. 30 Technical Event 285003 ................................. 30 Technical Fault 431001 ................................... 33 Technical Fault 431002 ................................... 33 Technical Fault 431004 ................................... 33 Technical Fault 431005 ................................... 34 Technical Fault 431006 ................................... 34 Technical Fault 431007 ................................... 34 Technical Fault 431008 ................................... 34 Doc.no: 624485/00 Technical Fault 431009 ................................... 34 Technical Fault 431010 ................................... 34 Technical Fault 431011 ................................... 34 Technical Fault 431012 ................................... 34 Technical Fault 431013 ................................... 34 Technical Fault 431014 ................................... 35 Technical Fault 431015 ................................... 35 Technical Fault 431016 ................................... 35 Technical Fault 432001 ................................... 35 Technical Fault 432002 ................................... 35 Technical Fault 433001 ................................... 35 Technical Fault 444001 ................................... 35 Technical Fault 444002 ................................... 35 Technical Fault 444003 ................................... 35 Technical Fault 444004 ................................... 35 Technical Fault 444005 ................................... 36 Technical Fault 446001 ................................... 36 Technical Fault 446002 ................................... 36 Technical Fault 446003…446021 ................... 36 Technical Fault 446022 ................................... 36 Technical Fault 446023 ................................... 36 Technical Fault 446024 ................................... 36 Technical Fault 446025 ................................... 37 Technical Fault 446026 ................................... 37 Technical Fault 446027 ................................... 37 Technical Fault 446028 ................................... 37 Technical Fault 481001 ................................... 37 Technical Fault 481002 ................................... 37 Technical Fault 483001..483008 ..................... 38 Technical Fault 485001 ................................... 38 Technical Fault 485002 ................................... 38 Technical state failed ....................................... 38 Touch not functional ........................................ 28 Unknown part number ..................................... 37 Vent outlet temperature high ............................. 9 May 2012 Troubleshooting Platform C 1 Copyright © 2011 HAMILTON MEDICAL AG. All rights reserved. Printed in Switzerland. No part of this publication may be reproduced or stored in a database or retrieval system, or transmitted in any form or by any means, electronic, mechanical, or by photocopying, recording, or otherwise, without the prior written permission of HAMILTON MEDICAL AG. This document may be revised or replaced by HAMILTON MEDICAL AG at any time and without notice. You should ensure that you have the most current applicable version of this document; if in doubt, contact the technical support department of HAMILTON MEDICAL AG, Switzerland. While the information set forth is believed to be accurate, it is not a substitute for the exercise of professional judgement. Nothing in this document shall limit or restrict in any way HAMILTON MEDICAL AG’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, HAMILTON MEDICAL AG has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including software) described herein. The equipment must be operated, serviced or upgraded only by trained professionals. HAMILTON MEDICAL AG’s sole responsibility with respect to the equipment and its use is as stated in the limited warranty provided in the operator`s manual. Product and company names mentioned herein may be the trademarks of their respective owners. HAMILTON MEDICAL AG will make available on request, component parts lists, descriptions, calibration instructions, or other information that will assist the user’s appropriately trained personnel to repair those parts of the equipment designated by HAMILTON MEDICAL AG to be repairable. Manufacturer Representative in USA HAMILTON MEDICAL AG Via Crusch 8 CH-7402 Bonaduz Switzerland Phone: (+41) 81 660 60 10 Fax: (+41) 81 660 60 20 info@hamilton-medical.com www.hamilton-medical.com HAMILTON MEDICAL Inc 4990 Energy Way Reno, NV 89502 PO Box 30008 Reno, NV 89520 Toll-free: (800) HAM-MED 1 Or: (800) 426-6331 Fax: (775) 856-5621 Page 4 of 42 Doc.no: 624485/00 May 2012 Troubleshooting Platform C 2 Conventions 2.1 Typographic Conventions Effect Example Function [xx] [C2] HAMILTON-C2 only [C1] HAMILTON-C1/T1 only [T1] HAMILTON-T1 only Only valid for the product for the product within the brackets 2.2 Expressions Expression Example Explanation Trigger condition PventControl < (Set pressure - 2mbar) for more than 5s Conditions which lead to the triggering of the technical fault Cancel condition PventControl ≥ (Set pressure - 2mbar) Conditions which cancel the technical fault Priority H H High priority alarm L Low priority alarm 2.3 Foreword Follow suggested troubleshooting steps in order. Follow these steps when troubleshooting: 1. Confirm that test configuration is correct. 2. If replacing components, repeat the applicable test following each replacement. If the test passes successfully, repeat all tests in order (see Section Service Software). 3. If a problem cannot be resolved, enter a customer event report (CER) on the HAMILTON MEDICAL AG online help desk (https://hamilton.ky2help.com). Export event log and Device report for failure analysis. Check regularly the knowledge base of the ky2help for latest information and troubleshooting. Page 5 of 42 Doc.no: 624485/00 May 2012 Troubleshooting Platform C 231001 Pressure Controller Pressure Low 231002 Pressure Controller Pressure High Disconnection on Ventilator side H PventControl < (Set pressure 2mbar) for more than 5s PventControl ≥ (Set pressure - 2mbar) Technical Event 231002 PventControl ≤ (PSafetyLimit10mbar) 231003 Flow Controller Flow Low Technical Event 231003 231004 Flow Controller Flow High Technical Event 231004 231005 Inspiration Valve Leak 231006 O2 Controller Flow Low Technical Event 231005 Oxygen supply failed H [C1] PventControl or Paw > PSafetyLimit AND PventControl and Paw after 1s not ≤ (PSafetyLimit – 10mbar) [C2] PventControl > PSafetyLimit AND PventControl after 1s not ≤ (PSafetyLimit – 10mbar) H Qvent < (Set flow - 200ml/s) for more than 5s, where by Set flow = Qvent Set flow + Set flow from the Baseflow regulator H Qvent > (Set flow + 200ml/s) for more than 5s, where by Set flow = Qvent Set flow + Set flow from the Baseflow regulator H [C2] Qvent > 2l/min when Set flow = 0 for more than 250ms H QO2 < (Set flow - 20%) for more than 250ms AND Set flow ≥ 1ml/s AND Set flow < 100l/min Description Page 6 of 42 Trigger condition Cancel condition Service software page No. Device's response alarm or message Priority id number (internal use) 3 Technical faults Action required - Check for leaks - Check ambient valve [C2] - Check Pvent_Control - Check Pvent_Control - Check Paw [C1] - Perform pressure calibration - Perform flow sensor calibration [C1] - Check inspiratory valve - Perform inspiratory valve calibration 2108 2204 2201 2201 2341 2341 2347 2107 2342 Qvent ≥ (Set flow 200ml/s) - Inspiration outlet obstructed - Check Qvent sensor - Check Inspiratory valve [C2] 2202 2107 Qvent ≤ (Set flow + 200ml/s) - Check Qvent sensor - Check Inspiratory valve [C2] 2202 2107 Qvent = 0 when Set - Check Inspiratory valve [C2] flow = 0 - Check Qvent sensor QO2 ≥ (Set flow - Check HPO Supply if in specs (2.8 to 6 20%) for more than bar , Medical Oxygen) 10ms AND Set flow ≥ - If HPO is available check O2 Mixer 10ml/s Doc.no: 624485/00 2107 2202 2112 May 2012 231007 O2 Controller Flow High Technical Event 231007 H QO2 > (Set flow + 200ml/s) for more than 5s 231008 O2 Valve Leak Technical Event 231008 Description 231009 Blower Controller Speed Low 231010 Blower Controller Speed High 231011 Vent Output Temperature High Technical Event 231009 Technical Event 231010 Vent outlet temperature high Trigger condition Cancel condition QO2 ≤ (Set flow + 200ml/s) or „High Pressure O2“ inactive H O2 valve off AND „High QO2 < 10m/s when Pressure O2“ active AND QO2 > Set flow < 1ml/s or 10ml/s for more than 5s „High Pressure O2“ inactive H The blower speed is lower than the target speed-5% for more than 5s. H The blower speed is lower than the target speed+5% for more than 5s. H The internal temperature of the ventilator is higher than expected. [C2] Tvent > 66°C (Temperature sensor of the Qvent Flow sensor) for more than 1s Action required - Check O2 mixer 2112 Blower speed=> target speed -5% - Check the LPO/HPO inlet for leaks as this could lead to 231008 even though the O2 valve is working properly. If HPO is used then make sure no LPO connector is connected to the LPO inlet. - Install latest Software version - Check O2 Mixer - Replace the O2 mixer assembly - Install latest SW version - Check blower Blower speed=> target speed+5% - Install latest SW version - Check blower Tvent <=64°C - Remove the ventilator from the sun or other heat source. - Ensure the ambient temperature does not exceed 40°C as specified - Check the cooling fan and its filter. - If needed, clean or replace air intake dust filter and fan filter every month (or more often, if required). The blower pressure will be reduced by 8 mbar preventively. [C1] Tvent > 66°C (Temperature sensor of the Qvent Flow Page 7 of 42 Service software page No. Device's response alarm or message Priority id number (internal use) Troubleshooting Platform C Doc.no: 624485/00 If problem remains: - Check Tvent (Sensor Data 2) by comparing with external measured outlet temperature. 5101 2112 2203 5101 2104 2105 5101 2104 2105 2402 May 2012 Device's response alarm or message Trigger condition Cancel condition sensor) for more than 10s 231012 Qvent Flow Sensor Defect Technical Event 231012 H The read values from Qvent sensor are out of range Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required (Use e.g. TSI Flow analyzer) - Check the Qvent Flow sensor connector for loose connectors - Consider to replace the cable between mainboard and Qvent flow sensor. - Replace Qvent if Tvent displays unreal values. - Replace blower module in case of a real temperature rise The read values from - Check the Qvent Flow sensor connector for Qvent sensor are loose connectors - Check voltages of Qvent and Tvent sensor within range (Test pins and voltages described in service manual) - Consider to replace the cable between Qvent flow sensor and control board [C1] resp., mainboard [C2] - Check Qvent and replace if necessary Page 8 of 42 Doc.no: 624485/00 2202 2202 May 2012 231013 QO2 Flow Sensor Defect Device's response alarm or message Technical Event 231013 H The read values from flow sensor QO2 are out of range 231014 Ambient Valve Error Technical Event 231014 231017 Blower Service Required Blower service required 231018 self test with Error Technical Event 231018 231019 Insp.Valve ISensor Defect Technical Event 231019 231020 Regulator Data Log On 231022 Pexpvalve Sensor Defect Technical Event 231020 Technical Event 231022 Page 9 of 42 Trigger condition Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required The read values from - Check the QO2 connector for loose flow sensor QO2 connectors - Check voltages of Qvent and Tvent sensor sensor are out of (Test pins and voltages described in range service manual) - Consider to replace the cable between mainboard/Control board and Qvent flow sensor. - Check QO2 and replace O2 mixer assembly if necessary H [C2] The end stage of the Current within range - Check ambient valve ambient valve has detected an - Check ambient valve coil if within range over current (12..30 Ohm) L The blower has reached the end Replacement of the - Replace blower module of lifetime (Blower timer=100%) blower resets the - Update the technical state in service blower timer software The self test has encountered an The self test ends - Check event log for other technical faults / Error successfully events - Check all cable connections [C2] Inspiratory valve current No overflow or - Check Inspiratory valve sensor overflow or underflow for underflow - Replace mainboard if problem reappears more than 100ms Debug Log active - Install latest SW version - Contact Technical Support H [C1] Overflow or underflow for No overflow or - Check Pexpvalve 2s and Pressure regulator active underflow Doc.no: 624485/00 2203 2108 2204 1302 2111 5101 2111 2401 May 2012 Device's response alarm or message Trigger condition Cancel condition 232002 Pvent Monitor Sensor Defect Technical Event 232002 H Overflow or underflow for 1s No overflow or underflow for 1s 232003 Paw Sensor Defect Technical Event 232003 H Overflow or underflow for 2s No overflow or underflow for 2s 232004 Pressure not released Pressure not released H PventM ≥ (Pmax + 15mbar) or Paw ≥ (Pmax + 15mbar) for more than 5s 232005 Blower hot Technical Event 232005 H [C1] BlowerTemp > 92°C [C2] BlowerTemp > 80°C For more than 100ms PventM < (Pmax + 15mbar) and Paw < (Pmax + 15mbar) [C1] BlowerTemp ≤ 82°C [C2] BlowerTemp ≤ 70°C 232006 Blower Temperature Sensor Defect Technical Event 232006 Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Check Pvent_monitor - Check voltage of Pvent_monitor at test pins if the valid range (according Service Manual) - Check Paw - Check voltage of Paw at test pins if the valid range (according Service Manual) - Exhalation obstructed - Check Pvent_monitor pressure sensor - Check Paw pressure sensor 2341 2341 2341 - Check the cooling fan and its filter. - If needed, clean or replace air intake dust filter and fan filter every month (or more often, if required). - Check Tblower temperature sensor - [C2] Check heat pipe connections H The temperature of the blower The temperature can [C2] cannot be read or is out of range be read and is within - Check FFC cable on mainboard connector range J21 (Temp. sensor) for proper connection. - Check voltage of Tblower (Sensor Test pins P42, Mainboard) - Check the temperature in Service Software-> Sensor Data->Sensor 2-> Temperature sensor Tblower - Replace blower module if necessary 2402 2402 [C1] - Check connector P4 on Driver Board Page 10 of 42 Doc.no: 624485/00 May 2012 232007 Qaw Flow Sensor Defect 232008 Pambient Sensor Defect Device's response alarm or message Check Flow Sensor tubing Technical Event 232008 Trigger condition H The values from the external flow sensor are out of tolerance (compared to Qvent) H Overflow or underflow Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Check voltage of Tblower (Sensor Test pins P3, Control Board) - Check the temperature in Service Software-> Sensor Data->Sensor 2-> Temperature sensor Tblower - Replace blower module if necessary The values from the Install latest SW Version external flow sensor - Perform proximal flow sensor calibration are within tolerance - Check flow sensor tubing (compared to Qvent) - Replace proximal Flow sensor - Check Qvent and Qaw - Check voltage of Paw at test pins if the valid range (according Service Manual) No overflow or - Check Pambient pressure sensor at Sensor underflow Data-> Sensor 2 and compare with a external measurement (E.g. TSI Flow analyzer) - If value shows “---“ check FFC cable of Pfilter if connected properly (QSPI Bus blocked, also in combination with other TF’s) 5101 2202 2106 2109 2402 - Ensure latest SW version is installed (The SW version < 2.x is not compatible with Mainboard Rev. >13) [C2] - If problem remains (very rare) replace [C2] mainboard, [C1] Control Board Page 11 of 42 Doc.no: 624485/00 May 2012 232027 Device Temperature High Device's response alarm or message Trigger condition Cancel condition Tinst ≤ 55°C Device temperature H Tinst > 60°C high Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Remove the ventilator from the sun or other heat source. - Ensure the ambient temperature does not exceed 40°C as specified - Check the cooling fan and its filter. - If needed, clean or replace air intake dust filter and fan filter every month (or more often, if required). - Check Tinst temperature sensor at Sensor Data-> Sensor 2 If value shows “---“ check FFC cable of Pfilter if connected properly (QSPI Bus blocked, also in combination with other TF’s) 232029 Tinst Sensor Defect 232030 Clock Error Page 12 of 42 Technical Event 232029 Technical Event 232030 H Overflow or underflow H Clock error greater than 10% in 1min - Software error - ESM (FPGA) defect No overflow or underflow - If problem remains (very rare) replace [C2] mainboard, [C1] Control Board - Check Tinst temperature sensor at Sensor Data-> Sensor 2 If value shows “---“ check FFC cable of Pfilter if connected properly (QSPI Bus blocked, also in combination with other Technical Events/Faults) 2402 2402 - If problem remains (very rare) replace [C2] mainboard, [C1] Control Board Clock error less than - Install latest SW version 10% in 1min - Replace ESM if failure remains and contact Technical Support Doc.no: 624485/00 May 2012 232034 O2 Sensor Error Technical Event 232034 H O2 cell EEPROM communication error 232035 Pfilter Sensor Defect 232038 CO2 Calibration Needed 232039 CO2 Sensor Invalid 233001 Autozero Pvent_monitor Fail Technical Event 232035 CO2 calibration needed H Pfilter communication error L “Calibration needed” set in the sensor status CO2 sensor faulty L OEM ID incorrect Technical Event 233001 L Pvent_monitor autozero valve defect (short circuit or disconnected) 233002 Autozero Pvent_Control Fail Technical Event 233002 L Pvent_control sensor out of range during autozero Description Page 13 of 42 Trigger condition Cancel condition Service software page No. Device's response alarm or message Priority id number (internal use) Troubleshooting Platform C Action required No communication error - Ensure O2 cell is connected properly - Replace O2 cell - Replace O2 cell cable if problem remains Pfilter - Check Pfilter FFC cable if connected communication OK properly “Calibration needed” - Calibrate the CO2 sensor reset in the sensor status OEM ID correct - Replace the CO2 sensor Pvent_monitor autozero valve OK - Check autozero valve and listen to the switching - Check FFC cable to pressure sensor board proper connection - Check connector of the valve to pressure sensor board if properly connected - Replace Pressure sensor assembly if reappears Pvent_control sensor - Check autozero valve and listen to the OK switching - Check FFC cable to pressure sensor board proper connection - Check connector of the valve to pressure sensor board if properly connected - Check voltage if within range at test pins according to the service manual - Replace Pressure sensor assembly if problem reappears Doc.no: 624485/00 2203 2402 Preop Check Preop Check 2106 2109 May 2012 Device's response alarm or message 233003 Autozero Paw Fail Technical Event 233003 L 233004 Autozero Qaw Fail Technical Event 233004 L 233005 Pressure Sensor Tolerance Technical Event 233005 L Page 14 of 42 Trigger condition Prox. autozero valve defect (short circuit or disconnected) Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required Prox. autozero valve - Check autozero valve and listen to the OK switching - Check FFC cable to pressure sensor board proper connection - Check connector of the valve to pressure sensor board if properly connected - Replace Pressure sensor assembly if problem reappears Dist Autozero valve defect (short Dist Autozero valve - Check autozero valve and listen to the circuit or disconnected) OK switching - Check FFC cable to pressure sensor board proper connection - Check connector of the valve to pressure sensor board if properly connected - Replace Pressure sensor assembly if reappears Pvent_control ≠ (Pvent_monitor Pvent_control = - Check pressure sensors ± 2mbar) (Pvent_monitor ± - Check autozero valves 2mbar) for more than - Check tubing of pressure sensor assembly - Replace pressure sensor assembly if 100ms problem reappears Doc.no: 624485/00 2106 2106 2401 2106 May 2012 Device's response alarm or message Trigger condition 233006 Nebulizer Valve Technical Event Error 233006 L Nebulizer valve defect (short circuit or disconnected) 241001 Replace HEPA 232028 filter L Pfilter > (5mbar at 80l/min, 10mbar at 180l/min, linear calculation) for more than 2Min Replace HEPA filter 243001 Alarm Silence Technical Event Error 243001 243002 Alarm Unknown Technical Event 243002 243003 Loudspeaker Defect Technical Event 243003 L Alarm silence not removed after 130s H Alarming receives an alarm ID for which there is no alarm data available (Software error) H No sound detected when alarming Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required Nebulizer valve OK - Check Nebulizer valve and listen to the switching - Check Nebulizer valve if connected properly - Check the coil resistance of the nebulizer valve at the connector P1 “Nebulizer” on control board [C1], resp. connector P13 on mainboard [C2] Coil resistance= 200..400 Ohm (while device switched off) - Replace O2 mixer assembly if necessary Pfilter ≤ (4mbar at - Check air intake dust filter and replace if 80l/min, 9mbar at necessary - Check HEPA filter and replace if necessary 180l.min, linear calculation) for more - Check Pfilter than 30s Alarm silence OK - Check Alarm Monitor Sound detected 2116 2402 2117 2113 - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Check loudspeaker: - If sounds, the problem is due to microphone working improperly. Do as follows: 5101 2102 [C2] - Check FFC cable between Front panel board (J2) and mainboard (J18) if connected properly Page 15 of 42 Doc.no: 624485/00 May 2012 Device's response alarm or message Trigger condition Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - with Front panel board revision 4 and higher: - Check Jumper P12 on Front panel board for correct position: Pos 2 for SW >= 2.1.0 Pos 3 for SW < 2.1.0 - Ensure the technical state corresponds to the front panel board revision installed [C1] - Replace driver board If the loudspeaker doesn’t sound: [C1] - Check loudspeaker if connected properly to driver board and replace loudspeaker if necessary. [C2] -Replace front panel board Hint: Check microphone at voltage test pins between GND and SndLvl: During alarm the voltage > 1.21V [C2] Page 16 of 42 Doc.no: 624485/00 May 2012 243004 Buzzer Defect At Startup Device's response alarm or message Technical Event 243004 H No Buzzer sound detected during self test Trigger condition 243005 Loudspeaker Sounds continuously 243006 Develop. Sound 243007 RTC Reset Technical Event 243005 H Continuous sound detected for more than 10min Technical Event 243006 Real time clock failure L 244001 External Power Loss Loss of external power H External Power out of range -Mains only (Input voltage < 90VAC -Power Supply Output voltage < 21.6VDC or >26.4VDC L Cancel condition Action required - Install latest SW version - Check buzzer: if doesn’t sound replace mainboard [C2],resp. control board [C1] No sound detected Software compiled with the sound disabled Real Time Clock communication Communication OK error [C2] [T1] -when DC Only connected: DC input voltage < 11VDC and > 27VDC Page 17 of 42 Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Doc.no: 624485/00 2113 Hint: Check microphone at voltage test pins between GND and SndLvl: During buzzer sound the voltage > 1.21V - Install latest SW version - False positive alarm eliminated with SW version 2.1.0 [C2] - Alarm silence resets alarm - Install latest SW version - Set RTC and reboot - do not enter any option keys after the device has been rebooted after a RTC set [C1] - Check if external power is available and within specifications - Ensure the device is switched off and connected to external power, check plug if properly inserted (both sides) 2102 5101 1401 or User Mode [C2] - Check circuit breaker if switched OFF - Check wiring from AC power inlet to power supply and to mainboard - Check if 24V (voltage range see Service manual) is present at mainboard (measured between pin GND_Power and May 2012 Device's response alarm or message Trigger condition Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required pin +24V_PS) and replace power supply if +24V_PS is out of range. If +24V_PS is within range then replace mainboard [C1] - Ensure the device is switched off and connected to mains AC power -Check FFC60 cable (Driver Board<>Control Board) if connected properly. -Check voltage on Control Board (between Test pins +24V_PS_tp<-> GND) -Replace power supply if +24V_PS_tp voltage is out of range (13.57 - 13.85Vdc) -Replace driver board if voltage is within range (13.57 - 13.85Vdc) -Replace control board if problem remains (defective voltage monitoring). To prevent unintentional disconnection of the power cord, make sure it is well seated into the ventilator socket and secured with the power cord retaining clip. Always check the reliability of the AC outlet. The AC power symbol in the bottom righthand corner of the screen is displayed with a frame around it. Page 18 of 42 Doc.no: 624485/00 May 2012 244002 Battery 1 Calibration Required 244003 Battery 2 Calibration Required Device's response alarm or message Battery 1 Calibration required Battery 2 Calibration required L L Trigger condition “Calibration required” status set in the battery “Calibration required” status set in the battery [C2] [T1] Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required Hint: The battery charge indicator lights indicate the availability of external power “Calibration required” - Calibrate the battery with the external status reset in the charger PN369104 battery [C2] With the battery 369102 a recalibration is typically needed once per year, at the latest after 30 full-charge/-discharge cycles. With the new battery 369106 (since July 2010) [C2], 369108 [C1] a recalibration is rare and typically needed once per two years, at the latest after 100 full-charge/discharge cycles. Under special circumstances the battery can perform a recalibration by itself. “Calibration required” -Calibrate the battery with the external status reset in the charger PN369104 battery [C2] With the battery 369102 a recalibration is typically needed once per year, at the latest after 30 full-charge/-discharge cycles. With the new battery 369106 (since July Page 19 of 42 Doc.no: 624485/00 May 2012 244004 Battery 1 Temperature High 244005 Battery 2 Temperature High 244006 Battery Power Low see also 244010 Device's response alarm or message Trigger condition Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required 2010) [C2], 369108 [C1] a recalibration is rare and typically needed once per two years, at the latest after 100 full-charge/discharge cycles. Under special circumstances the battery can perform a recalibration by itself. - Check cooling air inlet and fan Battery 1 Temperature high H Battery temperature >= 70°C Battery temperature <= 65°C Battery 2 Temperature high H Battery temperature >= 70°C [C2] [T1] Battery temperature <= 65°C - Check cooling air inlet and fan Battery low H Battery capacity < 15% M Battery capacity < 20% Battery capacity > 15% Battery capacity >=20% HAMILTON MEDICAL recommends that the ventilator’s battery is fully charged prior to ventilate a patient. If the battery is not fully charged and AC power fails, always pay close attention to the level of battery charge. [C2] [T1] - Connect to AC or DC power source - Replace battery 2 with a fully charged battery [C1] - Connect to AC power source The use of one battery is mandatory. The battery is used as internal backup battery. Do not operate without the backup battery. Page 20 of 42 Doc.no: 624485/00 May 2012 Device's response alarm or message Trigger condition Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required Hint: Look at the battery charge indicator lights: [C2] - green solid indicates: External power available and the battery is being charged [C1] - Green solid indicates: External power is available and the battery is fully charged -green blinking indicates: External power is available and the battery is being charged 244007 Battery Power Loss 244008 Wrong Battery 1 244009 Wrong Battery 2 244010 Battery Power Low On Power see also 244006 Page 21 of 42 Battery power loss H Batteries missing Battery present Battery 1 Wrong battery Battery 2 Wrong battery Battery low L Not a HAMILTON battery A HAMILTON battery - Replace battery with a HAMILTON battery L Not a HAMILTON battery [C2] [T1] Battery capacity < 20% during battery operation OR Battery capacity < 15% during AC/DC operation A HAMILTON battery - Replace battery with a HAMILTON battery L - Insert a battery Battery capacity > - Prepare to change the battery during 20% during battery battery operation operation OR Battery - The alarm resets when the battery is charged during AC/DC operation capacity > 15% during AC/DC operation Doc.no: 624485/00 May 2012 244011 Battery System Manager Bus Error Device's response alarm or message Technical Event 244011 Trigger condition H Communication error with the batteries or battery manager SMBus2 Error Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Cancel condition Action required Communication OK [C1] Check wiring: Qvent connector on Control board, FFC cable Control board<>driver board, FFC cable pressure sensor assembly<-> control board -Replace Control board if problem reappears 244012 Battery 1 Permanent Failure 244013 Battery 2 Permanent Failure Battery 1 Defective Battery 2 Defective 244014 Battery Task Technical Event 244014 Life Sign timeout 246001 Service needed Preventive maintenance required Page 22 of 42 H Communication error with the battery -Battery in a permanent failure H Communication error with the battery -Battery in a permanent failure [C2] [T1] H - Battery Task Life Sign timeout -Task reacted too late or does not exists L Service time reached Communication OK [C2] - Check Event log for other technical faults, replace Qvent cable if TF431008 or 231012 appears after start-up - Replace mainboard if Technical Event: 244011 appears without the combination of TF431008,231012 - Also do check the FFC from mainboard to the battery pack compartment for properly connection - Replace the battery Communication OK - Replace the battery Battery Task Life Sign OK - Restart the device - Contact Technical Support Service timer reset - Perform preventive maintenance - Reset service timer Doc.no: 624485/00 1200 May 2012 Device's response alarm or message Trigger condition Cancel condition Action required Service software page No. Description Priority id number (internal use) Troubleshooting Platform C 246002 CPU Temperature High 246003 Fan Error Technical Event 246002 H Processor temperature > 75°C Processor temperature < 65°C - Check cooling air inlet and fan Fan failure M Fan is not rotating Fan is rotating 2113 246004 Processor Overload Technical Event 246004 H Processor load >95% for >5s - Software error Processor load <95% 246005 Alarm Monitor Defect 246006 EEPROM Defaults 246007 EEPROM Write Failed 246008 CPU Temperature Defect 246009 Watchdog Disabled 246010 Hardware Parameter Error 249001 Option File not Read Technical Event 246005 Technical Event 246006 Technical Event 246007 Technical Event 246008 H Alarm Monitor communication error H EEPROM error during self test (default values loaded) H EEPROM write error Communication OK - Check fan if is rotating - Check if mechanically blocked - Check if fan is connected properly - Replace fan if necessary - [C2] Replace mainboard if failure remains - [C1] Replace driver board if failure remains - Restart the device - Install latest SW version - Contact Technical Support if problem reappears - Check the Alarm Monitor Calibration values saved EEPROM write OK - Run all the calibration/ tests in the service software - Perform the calibrations 2321 2347 H CPU temperature out of range for more than 1min CPU temperature in range - Replace ESM board 2113 Technical Event 246009 Technical Event 246010 L - Alarm silence resets alarm - Install latest SW version - Contact Technical Support - Restart the device - Reload the software 5101 Technical Event 249001 H Option file read error Page 23 of 42 L Software compiled with the watchdog disabled Hardware parameter file read error Doc.no: 624485/00 - Set RTC and restart the device - Enter the options - Install latest SW version 5101 2113 5101 5101 May 2012 249002 Option File Not Written 249003 Option File Wrong Version 249004 Option File Default Generated 249010 Device Config File Error 249011 Setup Config File Error 249012 Last Setting Config File Error 249020 Invalid Option Board 281001 SDRAM Error Device's response alarm or message Technical Event 249002 Technical Event 249003 Technical Event 249004 Technical Event 249010 Technical Event 249011 Technical Event 249012 Technical Event 249020 Technical Event 281001 281002 Ethernet Error Technical Event 281002 281003 EEPROM Error Technical Event 281003 281004 RTC Error Technical Event 281004 Page 24 of 42 Trigger condition H Option file write error Cancel condition Reset with Alarm silence H Incorrect option file version H Option file missing , default Reset with Alarm generated silence or start-up [C2] At the first startup after the upgrade from SW 1.x to SW 2.x Technical event 249004 (OhOptionFileDefaultGenerated) occurs. L Device configuration file error Reset with Alarm silence L Setup file error Reset with Alarm silence L Last settings file error Reset with Alarm silence L Invalid option board during start- Valid option board up (Wrong Part Nr) during start-up Action required - Reenter the options - Restart device - Reload the software [C2] With SW version lower than 2.x no SW options are available. So the software needs to generate a default option file on the first start up 5101 - Reenter the options if failure persists after a second start-up - Save the device configuration, restart the device - Save the device setup, restart the device - Start ventilation to save new last settings file H RAM error during self test - Switch off the device - Install the correct option board - Start-up the device - Replace ESM H Ethernet error during self test - Replace ESM H ESM EEPROM error during self test H Real Time Clock error during self test - Replace ESM Doc.no: 624485/00 Service software page No. Description Priority id number (internal use) Troubleshooting Platform C - Replace ESM May 2012 281005 IDE NAND Error 281006 Touch Error Device's response alarm or message Technical Event 281005 Touch not functional Trigger condition Cancel condition H IDE NAND error during self test H Touch error during self test Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Replace ESM Start-up with a Even though the touch screen does not successful self test of work, all functions of the ventilation the touch screen software can be controlled redundantly over the press and turn knob. Additionally the alarm system is still working. Therefore the patient safety is not restricted. 2321 [C2] with SW Version < 2.0.3: If the self test fails the device switches to ambient mode. The following Technical Faults are displayed: TF 485001, TF 485002 and TF 281006. With SW Version 2.0.3 and higher: The device passes the self test and alarms with "Touch not functional" - Ensure touch was permanently pressed during start-up - Check touch screen if connected properly to front panel board - Check function of the touchscreen by measuring the resistance on the black connector from the middle pin to the other 4 pins. Page 25 of 42 Doc.no: 624485/00 May 2012 Device's response alarm or message Trigger condition Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required While pressing on touchscreen the values of the resistance should be < 5kOhm. While not pressed the values of the resistance should be over 1MOhm. If the values are out of range: -Replace Touch screen (Display Front Complete & Display Gasket [C2]) -Replace Mainboard [C2], resp. Control Board [C1] if the values are within the given range and the problem persists. Note: Once the alarm has been triggered the device needs to be restarted in order to reset the alarm message 283001 VGUI Startup Failed 283003 VGUI language Not Loaded 283004 VGUI device Config File Error 283005 VGUI setup Config File Error Page 26 of 42 Technical Event 283001 Technical Event 283003 Technical Event 283004 H Startup software failure - (Re-)Install latest software version H Language file missing L Device file error - (Re-)Install latest software version - Replace ESM board if problem remains - (Re-)Install latest software version - Replace ESM board if problem remains Technical Event 283005 L Configuration file error Doc.no: 624485/00 - Install latest software version - Set configuration settings and restart 5101 User Mode May 2012 283006 VGUI last Setting Config File Error Device's response alarm or message Technical Event 283006 L Trigger condition Cancel condition Last settings file error Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Set Real Time Clock (RTC) and restart the device 1404 or User Mode -Check if the Adult option is active and enter if necessary 283007 VGUI last Setting Error Technical Event 283007 284002 SGUI alarm Service High 284003 SGUI alarm Service Medium 284004 SGUI alarm Service Low 285001 Alarm Lamps Error Defect 285002 Warning Alarm Lamps Defect 285003 Backlight Defect 331001 Pvent Pressure Sensor Defect Technical Event 284002 Technical Event 284003 Page 27 of 42 Technical Event 284004 Technical Event 285001 Technical Event 285002 Technical Event 285003 Safety ventilation 331001 The adult option key can be obtained from technical support - Restart device H Last settings do not fit to the options during a restart after power fail Dummy alarm ID used in the service software Dummy alarm ID used in the service software - No action required, part of the test 2205 - No action required, part of the test 2205 - No action required, part of the test 2205 Dummy alarm ID used in the service software M Alarm lamp defect Alarm lamp OK - Check red lamp 2102 H Warning lamp defect Warning lamp OK - Check yellow lamp 2102 H [C1] Backlight defect Backlight OK - Check backlight 2102 - Check Pvent_control - Check cable connector pressure sensor board to mainboard [C2], resp. Control board [C1] for proper connection - Replace Pressure sensor assembly if problem remains 2401 S Overflow or underflow for >200ms Doc.no: 624485/00 May 2012 332001 Qaw Flow Sensor Error Safety ventilation 332001 S Qaw Overflow or underflow > 200ms during self test 341001 Breath Settings Not Accepted 341002 Returned Breath Settings Incorrect Safety ventilation 341001 Safety ventilation 341002 S - Software error 341003 Adaptive Settings Invalid Safety ventilation 341003 Description 341004 Breath Settings Safety ventilation Timeout 341004 341005 Controller Settings Invalid Safety ventilation 341005 343001 Communication Safety ventilation Timeout 343001 346002 Watchdog … Failed 346052 383001 Control Settings not accepted Page 28 of 42 Safety ventilation 346002… 346052 Safety ventilation 383001 Trigger condition S The returned Breath Settings are not the same as the send Settings - Software error S The calculated adaptive settings are invalid - Software error S The Breath Settings were not returned within 2 s - Software error S The calculated settings are invalid - Software error S Alarming failed to receive the dummy alarms - Software error S Watchdog - Software error S Control Settings not accepted - Software error Doc.no: 624485/00 Cancel condition Service software page No. Device's response alarm or message Priority id number (internal use) Troubleshooting Platform C Action required - Check Qaw - Check cable connector pressure sensor board to mainboard [C2], resp. Control board [C1] for proper connection - Replace Pressure sensor assembly if problem remains - Install latest SW version - Contact Technical Support - Install latest SW version - Contact Technical Support 2202 - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Contact Technical Support 5101 5101 5101 May 2012 383002 Returned Control Settings incorrect Device's response alarm or message Cancel condition Action required S Returned Control Settings incorrect - Software error - Install latest SW version - Contact Technical Support 5101 383003 Control Safety ventilation Settings 383003 validation failed S Control Settings validation failed - Software error - Install latest SW version - Contact Technical Support 5101 383004 Monitoring Safety ventilation channel timeout 383004 S Monitoring channel timeout - Software error - Install latest SW version - Contact Technical Support 5101 383005 Vent Mode Control response timeout 383006 Incorrect Device Settings returned 383007 Trend channel timeout 385001 Alarm channel timeout 385002 Safety Failure Detected Safety ventilation 383005 S Vent Mode Control response timeout - Software error - Install latest SW version - Contact Technical Support 5101 Safety ventilation 383006 S Incorrect Device Settings returned - Software error S Trend channel timeout - Software error S Alarm channel timeout - Software error S Safety Mode activated, Device has switched to safety mode - Install latest SW version - Contact Technical Support 5101 - Install latest SW version - Contact Technical Support - Install latest SW version - Contact Technical Support Check event log for other technical faults causing the safety mode 5101 Page 29 of 42 Safety ventilation 383002 Trigger condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Safety ventilation 383007 Safety ventilation 385001 Safety ventilation 385002 Doc.no: 624485/00 5101 3302 May 2012 Service software page No. Device's response alarm or message Priority id number (internal use) Troubleshooting Platform C 385003 GUI-> Power Management Lifesign error Safety ventilation 385003 S GUI-> Power Management Lifesign Error It is an aftereffect of the ambient mode and occurs always in combination with other TF numbers. Check the eventlog or even the display for other TF numbers to identify the actual root cause of the problem. 386001 GUIL bitmap Not Loaded 431001 Blower Fault Safety ventilation 386001 Technical Fault 431001 S Bitmaps could not be loaded during startup A Blower end stage has detected one or more of the following faults: -Invalid hall sensor signal -Short circuit on motor windings -Supply voltage too low (Vbb<18V) - Install latest SW version - Contact Technical Support - Check blower - Check blower if connected properly - Check mainboard [C2], resp. Driver board [C1] for burned components (endstage) and replace the affected board if necessary - Replace blower - [C2] Replace mainboard if failure remains - [C1] Replace driver board if failure remains Technical fault can be ignored if appears in the Alarm monitor test (Service Software) only - Check blower - Check blower if connected properly - Check voltages - Replace blower if necessary - [C2] Replace mainboard if failure remains - [C1] Replace driver board if failure remains Description Trigger condition Cancel condition No fault from endstage -Endstage too hot (>165°C) 431002 Blower Disconnected Page 30 of 42 Technical Fault 431002 A Blower test failed during self test Doc.no: 624485/00 Action required 5101 2104 2104 May 2012 431004 Inspiration Valve Over Current Technical Fault 431004 A [C2] Inspiratory valve over current 431005 Expiration Valve Disconnected Technical Fault 431005 A [C2] No current when voltage ≠ 0 for more than 200ms 431006 Expiration Valve Over Current Technical Fault 431006 A Expiration valve over current 431007 Vent Output Over Temperature Technical Fault 431007 431008 Qvent Flow Sensor Error 431009 QO2 Flow Sensor Error 431010 Control REGSPI Timeout Technical Fault 431008 Technical Fault 431009 Technical Fault 431010 A [C1] Tvent > 76°C for more than 10s [C2] Tvent > 76°C for more than 1s A Communications error during self test A Communications error during self test A Regulator timeout Description Page 31 of 42 Trigger condition Doc.no: 624485/00 Cancel condition Action required Service software page No. Device's response alarm or message Priority id number (internal use) Troubleshooting Platform C - Check inspiratory valve - Check inspiratory valve if connected properly - Check coil resistance (10…14 Ohm, typically 11Ohm) - Replace inspiratory valve if problem reappears - Replace mainboard if problem persists - Check expiration valve - Check expiratory valve if connected properly 2107 - Check expiration valve - Check expiratory valve if connected properly - Replace expiratory valve if problem reappears - Replace mainboard if problem persists - Check cooling air inlet and fan 2111 - Check Qvent - Check cable/ connector - Check QO2 - Check cable/ connector - Install latest SW version - Replace ESM if failure persists - Contact Technical Support 2401 2111 2401 May 2012 431011 Monitor REGSPI Timeout 431012 Flow Sensor Measurement Imprecise 431013 Calibration Read Failed 431014 iExp. Valve Sensor Defect 431015 Emergency Off Failed Device's response alarm or message Trigger condition Technical Fault 431011 A Monitoring timeout Technical Fault 431012 A Qvent coefficients are not zero - Software error - ESM (FPGA) defect A Calibration data read error Technical Fault 431013 Technical Fault 431014 Technical Fault 431015 Technical Fault 431016 Technical Fault 432001 A Overflow or underflow 432002 Device Over Temperature 433001 Breath Monitoring Tick Timeout Technical Fault 432002 Technical Fault 433001 A Tinst > 80°C A [C1] BlowerTemp > 112°C [C2] BlowerTemp > 100°C For more than 100ms A Breath Monitoring timeout - Software error Action required - Install latest SW version - Replace ESM if failure persists - Contact Technical Support - Check Qvent and replace if necessary - Set factory settings and then recalibrate the device - Check expiration valve - Check Expiration valve current - Check Alarm monitor A Expiration valve current sensor overflow or underflow A Emergency off during self test failed 431016 Pvent Pressure Sensor Defect 432001 Blower Over Temperature Page 32 of 42 Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C [C1] BlowerTemp ≤ 102°C [C2] BlowerTemp ≤ 90°C Tinst ≤ 75°C Timing OK Doc.no: 624485/00 2202 2111 2403 2113 2114 [C2] - Replace mainboard if problem reappears [C1] - Replace control board if problem reappears - Check Pvent_control 2401 - Check Tblower 2402 - Check cooling air inlet and fan - Check Tinst - Install latest SW version - Contact Technical Support 2402 May 2012 Device's response alarm or message Trigger condition Cancel condition A When the sum of the remaining charge of both batteries is <6%, ventilation is not possible anymore. A Battery temperature 10°C higher than “Battery hot” alarm A Battery temperature 10°C higher than “Battery hot” alarm A Hardware signals “voltage out of tolerance” A Software is still running 2s after the shutdown command is given - Mainboard / Control board defect Action required 444001 Batteries Total discharged Technical Fault 444001 444002 Battery 1 Over Temp 444003 Battery 2 Over Temp 444004 Voltage Out Of Tolerance 444005 Shutdown Failed Technical Fault 444002 Technical Fault 444003 Technical Fault 444004 Technical Fault 444005 446001 CPU Temperature Critical 446002 Safety Failed Technical Fault 446001 A Tcpu > 85°C Technical Fault 446002 Technical Fault 446003…446021 A Device did not switch to Safety after request A Watchdog - Software error - Install latest SW version - Contact Technical Support Technical Fault 446022 A 3.3V or ADCref or REGSPI timeout [C2] - Check voltages - Replace mainboard Technical Fault 446023 A Watchdog -Software error - Install latest SW version - Contact Technical Support 446003 Watchdog … Failed 446021 446022 Voltage Error 446023 Watchdog Failed Workload Page 33 of 42 Service software page No. Description Priority id number (internal use) Troubleshooting Platform C - Insert one or two charged batteries and restart the ventilator. - Check for defective power supply, see “Loss of external Power” - Check cooling air inlet and fan - Check Battery with external charger - Check cooling air inlet and fan - Check Battery with external charger - Check voltages - Install latest SW version - Contact Technical Support Tcpu < 80°C Doc.no: 624485/00 5101 - Check cooling air inlet and fan - Check event log for other error messages 5101 May 2012 446024 Voltage Error 3V3 Device's response alarm or message Technical Fault 446024 Trigger condition A 3V3 voltage error 446025 Adc Error Technical Fault 446025 A REGSPI timeout (FPGA), Software error 446026 ADCref voltage Error Technical Fault 446026 A ADCref voltage error 446027 Watchdog Failed Technical Fault 446027 A Watchdog - Software error 446028 Clock Error Technical Fault 446028 Technical Fault 481001 A System Tick > ±10% compared to the Alarm Monitor clock A Error during reprogramming the FPGA Technical Fault 481002 A Program checksum error 481001 First Run Migration Error 481002 CRC Error Page 34 of 42 Doc.no: 624485/00 Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Check voltages [C1] - Replace control board if voltage 3V3 is out of tolerance - Replace driver board if failure remains [C2] - Replace mainboard if voltage 3V3 is out of tolerance - Install latest SW version - Replace ESM if problem persists - Contact Technical Support - Check the following voltages and replace [C1] control board, [C2] mainboard if values are out of tolerance +2V5_Ref & +3V3 - Install latest SW version - Contact Technical Support - Install latest SW version - Contact Technical Support - Install latest available SW version - Replace ESM Board if failure persists - Contact Technical Support - Install latest SW version - Replace ESM Board if failure persists - Contact Technical Support 5101 5101 5101 5101 5101 May 2012 481003 Unknown Part Number Device's response alarm or message Unknown part number A Unknown Device part number in the EEPROM 481004 Technical State Technical state Error failed 483001.. Startup Failure 483008 Technical Fault 483001..483008 484001..484001 484001.. 484008 485001 Ambient Failure Technical Fault Detected 485001 Page 35 of 42 Trigger condition A Required Technical State data is not complete A Error during software startup - Software error A Ambient Mode activated Doc.no: 624485/00 Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - [C1] Check wiring (Pressure sensor board, Front panel board, Driver board, Qvent sensor, QO2 sensor) - Check error log file for the failing component, search for entries like “.._Startup.out:Error read EEprom 'Frontpanel Board' ..” - Check Technical State if contains wrong part numbers - Check Device Part Nr an enter if necessary - Contact Technical Support if failure remains - Install latest SW version - Check technical state for wrong entered part number, refer to service manual - Ensure O2 cell is connected properly [C1] - Check wiring (Pressure sensor board, Front panel board, Driver board, Qvent sensor, QO2 sensor) - Install latest SW version - Contact Technical Support if problem remains The device switched to ambient mode. Check event log for other technical faults 1102 1302 5101 5101 3302 May 2012 485002 Self test Failed Page 36 of 42 Device's response alarm or message Technical Fault 485002 Trigger condition A Self test failed Cancel condition Service software page No. Description Priority id number (internal use) Troubleshooting Platform C Action required - Enter Service Software and check event log for other technical faults - Check all cables and connections Doc.no: 624485/00 3000 May 2012 Troubleshooting Platform C 4 Troubleshooting Service Software 4.1 Adjustments and calibrations Adjustments and calibrations Touchscreen (Page No 2321) Insp. valve (Page No 2342) [C2] Symptom Required action Calibration cannot be performed - Check touch screen if connected properly to front panel board - Replace touch screen (Front panel complete) if failure persist - Check technical state if the revision number of the inspiratory valve corresponds to the installed valve. Restart the device If revision number was changed (Ventil. Info->Technical State > Modify , Page No 1102) - Check blower (Tests/Calib > Comp test > Pneumatics 1> Blower flow, Page No 2104) - Check expiratory valve membrane - Check for leaks - Ensure the external measurement (Pressure gauge) is using mbar and was zeroised prior the use - Replace pressure sensor assembly - Check the expiratory membrane and replace it if necessary - Check the cable of the expiratory valve for proper connection - Replace expiratory valve - Replace expiratory valve cover if problem remains - Check Qvent (Tests/Calib > System test > Flow ,Page No 2202) - Check the expiratory membrane and replace it if necessary - Check internal tubing of expiratory valve [C1] - Replace expiratory valve cover - Replace expiratory valve if problem remains - Replace O2 cell - Check O2 cell cable for proper connection and replace if necessary Calibration cannot be performed. Qvent doesn’t reach target value Pressure (Page No 2341) 50mbar (+/- 5mbar) cannot be reached Gain values are > 1000 +/- 50 Exp. valve (Page No 2343) [C2] Calibration takes more than 3 minutes or fails. Several or a certain pressure levels cannot be reached Exp. valve (Page No 2343) [C1] Calibration cannot be performed / Qvent doesn’t reach 4l/min O 2 cell (Page No 2346) Page 37 of 42 Gain calibration fails Doc.no: 624485/00 May 2012 Troubleshooting Platform C Adjustments and calibrations Flow sensor (Page No 2347) Symptom Required action Calibration fails - Check proximal flow sensor and replace if necessary - Ensure adult/pediatric flow sensor is used - Check for leaks - Check ambient valve [C2] (Tests/Calib > Comp test > Pneumatics 2> Ambi valve, Page No 2112) - Check the expiratory membrane and replace it if necessary - Perform component tests in order to locate the problem (Rinse flow, Autozero valve, Pressure sensors) and replace pressure sensor assembly if the corresponding tests fail - Check Qvent (Tests/Calib > System test > Flow ,Page No 2202) 4.2 Component Tests Component Tests Electronics > Alarm system (Page No 2102) Electronics > Alr. Mon. 1 (Page No 2113) Electronics > Alr. Mon. 2 (Page No 2114) Page 38 of 42 Symptom - Loudspeaker doesn’t sound - Loudspeaker sounds distorted or too quiet - Alarm lamp remains dark One or more tests fail One or more tests fail Required action - Check loudspeaker for proper connection - Replace loudspeaker if problem remains [C1] C2 with Front panel board revision 4 or higher] - Install latest SW version - Ensure the technical state corresponds to front panel board revision installed - Check jumper on front panel board for correct position - Ensure the foil of loudspeaker was removed (Rear Front panel Cover) - Replace loudspeaker if problem remains - Replace front panel board if problem remains [C2 with Front panel board revision < 4] - Replace front panel board if problem remains - Replace front panel board - Install latest SW version - Contact Technical Support - Replace control board - Install latest SW version - Contact Technical Support - Replace control board Doc.no: 624485/00 May 2012 Troubleshooting Platform C Component Tests Electronics > User I/F > (Page No 2115) Pneumatics 1> Binary valve (Page No 2106) Pneumatics 1> Autozero (Page No 2109) Pneumatics 1> Blower flow (Page No 2104) Pneumatics 1> Blower pressure (Page No 2105) Pneumatics 1> Exp. valve (Page No 2111) Page 39 of 42 Symptom Some or all of the knob range do not function Required action - Replace P&T encoder Some or all of the keys and LED do not function One or more tests fail with Result “NOT - Replace front panel board OK” and status shows “INTERRUPTION” One or more tests fail with Result “NOT OK” and status shows “SHORT CIRCUIT” or “OK” One or both tests fail with “NOT OK” One or more tests end with Status “NOT OK” One or more tests end with Status “NOT OK” Test ends with Status “NOT OK” - Check if autozero valves are connected to pressure sensor board - Check FFC cable if connected properly to pressure sensor board and to [C2] mainboard resp. to [C1] control board - Replace pressure sensor assembly - Replace pressure sensor assembly - Check for internal leaks - Replace blower module - Replace mainboard [C2], driver board [C1] if problem remains - Check for internal leaks - Check Ambient valve [C2] (Tests/Calib > Comp test > Pneumatics 2> Ambi valve , Page No 2106) - Check expiratory valve membrane - Check Pvent_monitor (Adj/Calib > Calibration> Pressure, Page No 2341) - Replace blower module - Replace mainboard [C2], driver board [C1] if problem remains Pexpvalve test: [C1] - Check Pvent_control (Adj/Calib > Calibration> Pressure, Page No 2341) - Check expiratory valve membrane - Check Pexpvalve tubing for proper connection - Calibrate expiratory valve - Check expiratory valve membrane and replace if necessary - Replace expiratory valve cover - Check for internal leaks - Replace expiratory valve Doc.no: 624485/00 May 2012 Troubleshooting Platform C Component Tests Symptom Test ends with Status “NOT OK” Test ends with Status “NOT OK” Pneumatics 1> Insp. Valve (Page No 2107) [C2] Test ends with NOT OK External measured flow does not correspond to Qvent Test ends with NOT OK Page 40 of 42 Required action Pressure test: - Check Ambient valve [C2] (Tests/Calib > Comp test > Pneumatics 2> Ambi valve , Page No 2106) - Check Pvent_control and calibrate if necessary (Adj/Calib > Calibration> Pressure, Page No 2341) - Calibrate expiratory valve - Check expiratory valve membrane and replace if necessary - Replace expiratory valve cover - Check for internal leaks - Replace expiratory valve Leakage test: [C1] - Check for internal leaks - Check Ambient valve [C2] (Tests/Calib > Comp test > Pneumatics 2> Ambi valve , Page No 2106) - Check expiratory valve membrane and replace if necessary - Check Qvent (Tests/Calib > System test > Flow , Page No 2202) Leakage: - Check for internal leaks - Check Qvent (Tests/Calib > System test > Flow , Page No 2202) Flow Control: - Check for internal leaks - Check ambient valve (Tests/Calib > Comp test > Pneumatics 2> Ambi valve, Page No 2112) - Calibrate inspiratory valve (Adj/Calib > Calibration> Insp. Valve, Page No 2342) - Check Qvent (Tests/Calib > System test > Flow, Page No 2202) - Check cable of inspiratory valve if connected properly Pressure: - Check for internal leaks - Check ambient valve (Tests/Calib > Comp test > Pneumatics 2> Ambi valve, Page No 2112) - Perform the calibration of the inspiratory valve (Adj/Calib > Calibration> Insp. Valve, Page No 2342) - Check Qvent (Tests/Calib > System test > Flow, Page No 2202) - Check Pvent_monitor (Adj/Calib > Calibration> Pressure ,Page No 2341) Doc.no: 624485/00 May 2012 Troubleshooting Platform C Component Tests Pneumatics 2> O2 input (Page No 2112) Symptom Test ends with NOT OK Pneumatics 2> Neb. valve (Page No 2116) Pneumatics 2> Ambi valve (Page No 2108) [C2] Pneumatics 2> Prox. Test (Page No 2110) Test ends with NOT OK Pneumatics 2> Check valve (Page No 2118) [C1] Pneumatics 2> Air entry (Page No 2117) Page 41 of 42 One or more tests end with Status “NOT OK” Required action - Ensure LPO connector is not inserted - Ensure HPO is connected (medical oxygen, 2.8…6bar ) - Check the cable of the O2 proportional valve for proper connection - Replace O2 mixer assembly - Ensure HPO is connected (medical oxygen, 2.8…6bar ) -Calibrate flow sensor - Check Qaw (Tests/Calib > System test > Flow , Page No 2202) - Replace O2 mixer assembly if problem remains - Check ambient valve if seated properly in bottom foam - Replace ambient valve if problem persists Flows through both flow sensor connectors are Rinse flow test: unequal - Replace pressure sensor assembly - Replace rinse flow assembly if failure remains [C2] Test ends with NOT OK Proximal pressure: - Check autozero valve (Tests/Calib > Comp test > Pneumatics 1> Autozero , Page No 2109) - Replace pressure sensor assembly Test ends with NOT OK Proximal flow: - Check Qvent (Tests/Calib > System test > Flow , (Page No 2202) - Replace pressure sensor assembly Test ends with NOT OK Rinse tank: [C1] - Check Paw (Calibration Pressure) - Replace pressure sensor assembly Test ends with NOT OK - Check for internal leaks - Check Qvent (Tests/Calib > System test > Flow, Page No 2202) - Replace inspiratory path (patient connection) With HEPA filter installed the pressure is - Check HEPA Filter and replace if necessary >2mbar - Check Pfilter if connected properly (Sensor board side) - Check Pambient if measuring correctly (compare to an external measurement e.g. TSI flow analyzer.(Tests/Calib > Sensor data > Sensor 2 , Page No 2402) - Replace Pfilter pressure sensor board if Pambient is measuring correctly and Control board [C1] resp. Mainboard [C2] if Pambient is measuring incorrectly. Doc.no: 624485/00 May 2012 Troubleshooting Platform C 4.3 System Tests System Tests Pressure (Page No 2201) Symptom Set values are out of tolerance Flow (Page No 2202) Set values are out of tolerance O2 mixer (Page No 2203) Set values are out of tolerance Displayed value of oxygen (O2 cell) only is out of range Tightness (Page No 2204) Alarming (Page No 2205) Page 42 of 42 Displayed value of QO2 is out of tolerance: Test fails Alarm priorities are incorrectly Required action - Check for leaks - Check expiratory valve - Check pressure sensors - Calibrate flow sensor - Replace flow sensor if necessary - Check blower (Tests/Calib > Comp test > Pneumatics 1> Blower flow , Page No 2104) - Replace Qvent if problem remains - Ensure LPO connector is not inserted - Ensure HPO is connected (medical oxygen, 2.8…6bar ) - Check Qvent (Tests/Calib > System test > Flow , Page No 2202) - Perform O2 cell offset and gain calibration - Replace O2 cell if problem persists - Check O2 mixer assembly if properly connected to blower module (tubing) - Replace O2 mixer assembly - Check for leaks - Check expiratory valve membrane and replace if necessary - Perform component tests Pneumatics 1 and 2 to locate the leak / faulty component - Install latest SW version - Contact Technical Support Doc.no: 624485/00 May 2012