SUR ­ BVFamily & Endura, Libra, Pulsera ­ Stand reboots during flouro: Mitra power supply issue
Applies To
Surgery, BV Endura, BV Family R2, BV Libra, BV Pulsera, BV Family R1.2
Problem Description Details
Stand reboots during fluoro
Problem Cause
The system demands a lot of power during fluoro. Sometimes the voltage drops too much, this causes the Midas board to restart.
Problem Solution
Check the Stand power supply.
The Stand has a built­in power supply. This power supply delivers power to the Midas board. The Midas board will restart when the voltage gets too low.
This power supply delivers:
Output 1 => 24 Volt
Output 2 => 5 Volt
Output 3 => 15 Volt
Output 4 => 15 Volt
Note: there is a resistor on Output. This resistor acts as a current limiter, and is required.
Check the following:
1. Check the input voltage of the power supply, are the wires connected correctly?
2. Check that the air flow is sufficient and not blocked. Ensure that the air inlet and outlet are dust free.
3. Check that the sense wire is connected correctly (between pins 1 and 2) See drawing below.
A. Measure the voltage on the sense wire. This voltage should not drop more than 0.5 Volt of Output 1 (+24 Volt).
B. If the sense voltage is too low check the connections.
You can also adjust the voltages on the output 2,3 and 4.
Source of Information
Tier 3 Helpdesk
Attachments
PE3238­40.pdf
Terms and Conditions for use of the Eureka Fault/Fix Solution Knowledgebase:
You may not use or access this knowledgebase without prior acceptance of these terms and conditions. Use of this knowledgebase constitutes your acceptance of these terms and conditions. You may only access this
knowledgebase as an authorized user. You must treat all materials displayed in this knowledgebase as company confidential. All materials displayed in this knowledgebase are the property of Philips Medical Systems B.V.
and are protected by copyright, trademarks, or other intellectual property rights afforded by law. You agree not to copy, reproduce, re­transmit, distribute, disseminate, sell, publish, broadcast, circulate, or otherwise publish
any such information in this knowledgebase to any third party. Transferring and/or copying in whole or in part of the knowledgebase to other databases is only allowed after prior written consent of Philips Medical Systems
B.V. As an authorized user of the knowledgebase you are allowed to download the content and other related information for use within Philips Medical Systems provided that you treat them as Company Confidential.
Knowledgebase information may be changed or updated by PMS B.V. without notice. Please note that information content may not apply to all system configurations due to variation in product or product upgrades. Use this
information only in conjunction with your professional knowledge and judgement. Copies of any kind of this information are uncontrolled. After being registered as an authorized user, you are requested to enter only once per
browser session, your user name and password to get access tothe knowledgebase information.
data:text/html;charset=utf­8,%3Ctable%20border%3D%220%22%20cellpadding%3D%220%22%20cellspacing%3D%220%22%20width%3D%22100%25…
1/1
Download