Biological Microscope
CX43/CX33
CX3 Series
Comfortable, High-Throughput Routine Microscopy
Comfortable for Long Periods of Routine Microscopy
The CX33 and CX43 microscopes enable users to remain comfortable during routine microscopy. The microscope frame
well fit the hands and the location of the control knobs maximize ergonomics to improve work efficiency. Users can quickly
set a specimen with one hand, while adjusting the focus and operating the stage with the other hand with minimal
movement. Both microscopes also feature a camera port for digital imaging.
CX43
1
Maintain Preferred Observation Conditions with Minimal Adjustments
Uniform Illumination with Consistent Color
Temperature
The color temperature of the CX LED illumination produces
daylight conditions, so specimens can be viewed with their
natural colors. The color temperature is consistent at any
brightness, so users don’t have to spend time making
adjustments when they change brightness. The LEDs have a
long 60,000-hour lifetime in the design value, helping reduce
cost, and the brightness level remains stable throughout the
LED’s life.
Select and Set Your Contrast Level
Users can preserve their favorite contrast by locking the
aperture diaphragm. It stays fixed at the optimally chosen
position if it is accidentally touched while changing slides.
Change Magnification without Adjusting the
Condenser
Users can change the magnification from 4X to 100X without
moving the top lens on the condenser. 2X magnification is also
available by simply setting the objective and the condenser
turret to 2X position.
Excellent Optical Performance for Flat Images
The microscope employs Plan Achromat objectives, which
provide clear images with high image flatness over a wide field
of view. This helps users view specimens clearly and evenly
during routine microscope observations.
Simple Fluorescence Observation
Fluorescence observation is simple and easy. Plug the compact
fluorescent illuminator into the microscope frame for
fluorescence observation. Its LED light source is pre-centered,
and the transmitted illumination is shuttered by simply setting
the condenser turret to the FL position. This reduces
background noise in the fluorescence image from incidental
light coming from the top lens of the condenser.
FN 20
2
Remain Comfortable during Extended Usage
Single-Handed Sample Placement
Use Up to Five Objectives
A specimen can be quickly slid in and out using one hand. The
specimen holder opens a little and firmly retains the specimen
during operation. The versatile holder accommodates a variety
of slide types, including a hemocytometer.
For added flexibility, up to five objectives can be supported by
the revolving nosepiece. In addition to general objectives, users
can select a 2X objective for wide area observation or objectives
for phase contrast. These objectives with long working
distances help keep specimens from getting damaged.
Ergonomically-Positioned Focus Knob
The low-positioned focusing knob enables users to make
observations while keeping their hands and forearms rested on
the desk, helping provide comfort. The focusing stopper
prevents a specimen from accidentally hitting an objective when
working under high magnification.
Smooth Magnification Change
The low-positioned revolving nosepiece enables users to quickly
change magnifications with minimal arm movement between
focusing, greatly improving work efficiency during prolonged
use.
3
Ergonomic Stage and Eyepiece Position
Specimen Holders that Match Your Observation Style
The low-positioned stage is designed to enhance comfort and
reduce fatigue. The stage surface can be widely seen from the
eye point position, which enables users to smoothly set and
check specimens on the stage. The stage knob can be
controlled with just a light touch and can be adjusted at the same
time as the focusing knob, since they are located close together.
Stage accessories improve efficiency when users need to
observe a large number of specimens. With the specimen
holder sheet, a specimen can be freely operated by a finger on
the sheet and can be precisely adjusted using the stage knob.
The double specimen holder can retain a large specimen or two
specimens.
Simplified Fluorescence Observation
Fluorescence observation can be easily set up on the standard
configuration while keeping the eye point the same as other
observation methods. Simply plug the compact fluorescent
illuminator into the back of the microscope frame.
4
Versatile Applications
The universal condenser offers a variety of observation methods and future upgradability. In combination with the five-position
revolving nosepiece, multiple applications can be covered using the single microscope frame.
Brightfield
Brightfield
Leukocyte (minimum iris aperture)
Phase Contrast
Urinary Cast (minimum iris aperture)
Fluorescence
HeLa cells
Renal Glomerulus
Accessories
CX33 Microscope
Simple polarizing intermediate
attachment/CX3-KPA
Offers polarized observation of urate
crystals and amyloid in combination with a
polarizer and analyzer.
For less demanding requirements using only brightfield and
darkfield, the CX33 microscope is a great option. The lowpositioned nosepiece and stage, focusing lock, specimen
holder, and inward quadruple revolving nosepiece make the
CX33 microscope is well-suited for everyday observations in
one easy configuration.
Eyepoint adjuster/ U-EPA2
Raise the eyepoint position by 30 mm for
added comfort.
Arrow pointer/ U-APT
Insert an LED arrow into your image; great
for digital imaging and presentations.
Dual observation attachment/U-DO3
Enables dual, simultaneous observation of a
single specimen from the same direction with
equal magnification and brightness for both
operators. A pointer can be used to indicate
specific sections of the specimen to simplify
the training process and enhance discussion.
5
CX33
CX43 System
CAMERA
CAMERA ADAPTOR
WHB10X
WHB10X-H
Eyepieces
WHB10X
WHB10X-H
Eyepieces
A
U-CBI30-2
Binocular tube
U-CTR30-2
Trinocular tube
B
U-CTBI
Economical tilting binocular tube
(10X eyepieces incorporated. FN 18)
A
U-TRU
Trinocular intermediate attachment
UIS2 Objectives
A
CX43-RFAB
B excitation
fluorenscence
illuminator
U-TRUS
B
Trinocular intermediate unit
U-DO3
Dual observation attachment
U-DAL10X
Drawing attachment 10X
CX3-SHP
Specimen
hold plate
U-DA
Drawing attachment
U-EPA2
Eyepoint adjuster
CX3-HLDT
Specimen holder
U-ECA
Magnification changer 2X
CX43RF
Biological Microscope
U-POT
Polarizer
U-APT
Arrow pointer
ø45 filter
U-GAN
Analyzer for urate crystals observation
CX3-KPA
Simple Polarizing Intermediate Attachment
U-CA
Magnification changer
Dimensions
(Unit: mm)
140
140
375
383
430
CX33
393
CX43
155
211
261
376
Weight: Approx. 7.3 kg
155
211
261
397
Weight: Approx. 7.0 kg
6
CX43 Specifications
Optical System
Illumination System
Focusing
Revolving Nosepiece
Stage
Observation Type (anti-fungal)
Tube
Eyepiece (anti-fungal)
Tube Inclination
Light Path Selector
Interpupillary Distance Adjusting Range
Condenser
Observation Methods
Objectives
Fluorescence Light Source
Rated Voltage/Electric Current
UIS2 (universal infinity-corrected) optical system
· Built-in transmited illumination system
· Köhler illumination (fixed field diaphragm)
· LED power consumption 2.4 W (nominal value), precentered
· Stage height movement (coarse movement stroke: 15 mm )
· Stroke per rotation for coarse adjustment knob: 36.8 mm, Focusing stopper
· Torque adjustment for coarse adjustment knob
· Fine focus knob (minimum adjutsment gradations: 2.5 μm)
Fixed quintuple nosepiece with inward tilt
· Wire movement mechanical fixed stage, (W × D): 211 mm × 154 mm
· Traveling range (X × Y): 76 mm × 52 mm
· Single specimen holder (optional: double specimen holder, sheet holder)
· Specimen position scale
· Stage XY movement stopper
Binocular
Trinocular
Tilting binocular
10X Field Number (FN): 20
10X Field Number (FN): 20
10X Field Number (FN): 18
30°
30°
30° 60°
None
None (eyepiece/camera port = 50/50 fixed) None
48 75 mm
· Abbe condenser NA 1.25 with oil immersion
· Universal condenser with 7 turret positions: BF (4 100X), 2X, DF, Ph1, Ph2, Ph3, FL
· Condenser turret lock pin (BF only)
· Built-in aperture iris diaphragm
· AS lock pin
Brightfield, simple polarization, fluorescence, phase contrast, darkfield
Plan achromat (UIS2), anti-fungal
2X
NA 0.06
W.D. 5.8 mm
4X
NA 0.1
W.D. 18.5 mm
10X
NA 0.25
W.D. 10.6 mm
10XPH
NA 0.25
W.D. 10.6 mm
20X
NA 0.4
W.D. 1.2 mm
20XPH
NA 0.4
W.D. 1.2 mm
40X
NA 0.65
W.D. 0.6 mm
40XPH
NA 0.65
W.D. 0.6 mm
60X
NA 0.8
W.D. 0.2 mm
100XO
NA 1.25
W.D. 0.13 mm
100XOPH
NA 1.25
W.D. 0.15 mm
100XOI
NA 1.25 0.6
W.D. 0.13 mm
Easily add an LED reflected fluorescence illuminator (peak excitation wavelength 470 nm: B excitation only),
precentered
AC 100 240 V 50/60 Hz 0.4 A
CX33 Specifications
Optical System
Illumination System
Infinity optical system
· Built-in transmited illumination system
· Köhler illumination (fixed field diaphragm)
· LED power consumption 2.4 W (nominal value), precentered
· Stage height movement (coarse movement stroke: 15 mm )
· Stroke per rotation for coarse adjustment knob: 36.8 mm, Focusing stopper
· Torque adjustment for coarse adjustment knob
· Fine focus knob (minimum adjutsment gradations: 2.5 μm)
Fixed quadruple nosepiece with inward tilt
· Wire movement mechanical fixed stage, (W × D): 211 mm × 154 mm
· Traveling range (X × Y): 76 mm × 52 mm
· Single specimen holder (optional: double specimen holder, sheet holder )
· Specimen position scale
· Stage XY movement stopper
· 30° inclined trinocular tube (anti-fungal)
· Light path selector: eyepiece/camera port = 100/0 or 0/100
· Interpupillary distance adjusting range: 48 75 mm
· Eyepoint adjustment: 375.0 427.9 mm
· 10X Field Number (FN): 20
· 15X Field Number (FN): 16 (optional)
· Abbe condenser NA 1.25 with oil immersion
· Built-in aperture iris diaphragm
Brightfield, darkfield
Plan achromat, anti-fungal
4X
NA 0.1
W.D. 27.8 mm
10X
NA 0.25
W.D. 8.0 mm
20X
NA 0.4
W.D. 2.5 mm (optional)
40X
NA 0.65
W.D. 0.6 mm
100X
NA 1.25
W.D. 0.13 mm (optional)
AC 100 240 V 50/60 Hz 0.4 A
Focusing
Revolving Nosepiece
Stage
Observation Tube
Eyepieces (anti-fungal)
Condenser
Observation Methods
Objectives
Rated Voltage/Electric Current
•
•
•
www.olympus-lifescience.com
is ISO14001 certified.
is ISO9001 certified.
is ISO13485 certified.
• All company and product names are registered trademarks and/or trademarks of their respective owners.
• Specifications and appearances are subject to change without any notice or obligation on the part of the manufacturer.
Shinjuku Monolith, 2-3-1 Nishi-Shinjuku, Shinjuku-ku, Tokyo 163-0914, Japan
Printed in Japan N8600645-032017
Download