SERVICE MANUAL FUJI DRI-CHEM CLINICAL CHEMISTRY ANALYZER FUJI DRI-CHEM 4000i FUJI DRI-CHEM 4000ie Safety and Precautions 1 Installation 2 Compositions and Name of Parts 3 Functions 4 Control Units 5 Software 6 Disassembly and Reassembly 7 Checking and Adjustment 8 Troubleshooting 9 Slide Information 10 Second Edition Z70N0167A January 2009 Introduction The FUJI DRI-CHEM 4000 series (henceforth, the FDC4000) is an in vitro diagnostic medical device to analyze blood or urine by colorimetric end-point, rate and potentiometric tests, using the FUJI DRICHEM SLIDEs. The FUJI DRI-CHEM 4000 series has 2 models, according to the destinations as follows: FUJI DRI-CHEM 4000i (henceforth, FDC4000i): For North America FUJI DRI-CHEM 4000ie (henceforth, FDC4000ie): For Europe or Asia NOTE:The functions of both models are same. NOTE:The FDC4000 indicates the FDC4000i or the FDC4000ie in this manual. CAUTIONS 1. No part or all of this manual may be reproduced in any form without prior permission. 2. The information contained in this manual may be subject to change without prior notice. 3. FUJIFILM shall not be liable for malfunctions and damages resulting from installation, relocation, remodeling, maintenance, and repair performance by other than dealers specified by FUJIFILM. 4. FUJIFILM shall not be liable for malfunctions and damages of FUJIFILM products due to products of other manufacturers not supplied by FUJIFILM. 5. FUJIFILM shall not be liable for malfunctions and damages resulting from remodeling, maintenance, and repair using repair parts other than those specified by FUJIFILM. 6. FUJIFILM shall not be liable for malfunctions and damages resulting from negligence of precautions and operating methods contained in this manual. 7. FUJIFILM shall not be liable for malfunctions and damages resulting from use under environment conditions outside the range of using conditions for this product such as power supply, installation environment, etc. contained in this manual. 8. FUJIFILM shall not be liable for malfunctions and damages resulting from natural disasters such as fires, earthquakes, floods, lightning, etc. Copyright reserved. Copyright 2007-2009 by FUJIFILM Corporation. “DRI-CHEM" is a registered trademark of FUJIFILM Corporation. FDC4000 Service Manual (i) Revision History NO. Issue Date Reason 0 Sep. 24, 2007 Training edition 1 Oct. 19, 2007 1 st edition Added: Sec.7.1.32 Replacing the Fuse Revised: Sec. 5.1 Block Dia. FAN901 => Not in use Revised: Sec. 6.13 Mode Description E036 description Added: Sec.9.2.1 Cover Lock Release 2 Jan. 1, 2009 Pages Affected P7-43 P5-1 P5-4 P6-33 to 39 P8-30, P9-24 P9-96 2 nd edition Added: Sec. 5.2 Connection diagram P5-2 Comment: Fan (incubator cooler) is not equipped. P5-5 Data file: “dc4_jklm.dat” => “dc4_di.dat” P6-3 OD calibration equation changed. (factor “i” added) P6-5 Number of tests: 0 to 20 slides => 0 to 30 slides P6-16 Mode type: Administrator => User (M49, M52, M103) Added: Mode 76 P6-31, 32 Data file: “dc4_jklm.dat” => “dc4_di.dat” P6-36, 38, 39 Revised for model year 2009 or later Sec. 7 (all) Statement added (in frame): For adjustment software startup P8-3 Added: “Do not perform this function other than when replacing the interfer- P8-29 ence filter.” NOTE: Before performing this function, clean the optical parts (the black and white plate, photometer head, and the lens of the lamp unit). Added: “Normally, this function is used only in the factory.” P8-30 Possible causes added for “E”: (3) A slide in the incubator moved during test processing. (4) Step-out of the syringe operation P9-7 Reviced detection level for drift error: (2.5) => (2.0) for CL (45 to 60 sec.) P9-10 Added in [W022]: “The counts of the consumables (slides, tips)” is also reset. P9-17 [W085] deleted. [W081] added. P9-29 Clogging detection level revised. P9-33, 35, 39, 41, 42 Added in [E0120] Check Points (8): “To check: ("Sampler 1" => "Sample")” P9-38 Added in [E0123] Check Points (6): “To check: ("Sampler 1" => "Diluent”)” P9-40 Added in [E0129] Check Points (6): “To check: ("Sampler 2" => "Reference P9-43 fluid")” Added in [E0143] Check Points (4): “To check: ("Sampler 1" => "Diluent")” P9-47 Added in [E0144] Check Points (4): “To check: ("Sampler 2" => "Reference P9-48 fluid")” (ii) Added in [E0145] Check Points (6): “To check: ("Sampler 1" => "Sample")” P9-49 Added in [E0146] Check Points (1): “(No foam on the bottle’s openning)” P9-50 FDC4000 Service Manual NO. Issue Date Reason Revised in [W191] Error description: “the designated slide lot” => “the first slide’s lot” Check Points (1): => “Load the same lot of slide as the first.(W191)” Pages Affected P9-55 Possible cause added in [W500], [W501]: (2) Cable disconnection P9-60 Added in Check Points (2): “For model year 2009 or later, the FAN901 is not equipped.” Possible cause added in [E0500]: “(2) The backup switch (S3) on the CPU40 P9-61 board is OFF.” Possible cause added in [E30xx]: “(2) The backup switch (S3) on the CPU40 P9-87 board is OFF. (Master crash error)” Check points (1) and (2) revised. Possible cause (1) deleted in [E50xx] Check points (1) deleted. Comment added: “(Use designated (recommended) CF cards.)” P9-91 Added: [E5510] description P9-94 Test codes changed: HDL-C (22 => 26), AMYL (36 => 43) P10-1 FDC4000 Service Manual (iii) (iv) FDC4000 Service Manual Introduction Revision History 1. Safety and Precautions................................................................................ 1-1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 1.10 1.11 1.12 Definition of Specific Safety Precautions ...................................................... 1-1 Precautions before Operating This Equipment............................................. 1-1 Biohazards and Disposal .............................................................................. 1-2 Disinfection ................................................................................................... 1-3 Cleaning Solutions........................................................................................ 1-3 Explosive Hazards ........................................................................................ 1-3 Electrical Hazards......................................................................................... 1-4 Safety Devices.............................................................................................. 1-4 Moving Parts................................................................................................. 1-4 Heavy Objects .............................................................................................. 1-5 General Precautions ..................................................................................... 1-5 Warning Labels............................................................................................. 1-6 2. Installation.................................................................................................... 2-1 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 Installation Space ......................................................................................... 2-1 Installation Site Requirements ...................................................................... 2-1 Environmental Requirements ....................................................................... 2-1 Electrical Requirements................................................................................ 2-1 Unpacking..................................................................................................... 2-2 Package Contents ........................................................................................ 2-2 Removing Tape and Fixtures........................................................................ 2-3 Preparations Before Turning the Power on .................................................. 2-3 Setting up the Optional Accessories............................................................. 2-3 2.9.1 2.9.2 Sample Barcode Reader .................................................................................. 2-3 Communication with Host Computer or PC...................................................... 2-3 2.10 Turning the Power on ................................................................................... 2-3 2.11 Acceptance ................................................................................................... 2-4 3. Compositions and Name of Parts ................................................................ 3-1 3.1 Component Names....................................................................................... 3-1 4. Functions ..................................................................................................... 4-1 4.1 Measurement Principle of Colorimetric Tests ............................................... 4-1 4.1.1 4.1.2 4.1.3 Preparation for Measurement........................................................................... 4-1 Process of Measurement.................................................................................. 4-1 Photometer and Calculation of Analyte Concentration..................................... 4-2 4.2 Measurement Principle of ISE Tests ............................................................ 4-2 4.2.1 4.2.2 Preparation for Measurement........................................................................... 4-2 Process of Measurement.................................................................................. 4-2 4.3 Sampling....................................................................................................... 4-3 4.3.1 4.3.2 FDC4000 Mounting Tip and Sealing................................................................................. 4-3 Sample Surface Detection................................................................................ 4-3 (v) 4.3.3 4.3.4 4.3.5 4.3.6 Sample Aspiration ............................................................................................ 4-4 Dilution.............................................................................................................. 4-4 Spotting Sample ............................................................................................... 4-5 Tip Disposal...................................................................................................... 4-5 4.4 Slide Transfer ............................................................................................... 4-6 4.4.1 4.4.2 4.4.3 4.4.4 Slide Dot Code Reading ................................................................................... 4-6 Slide Transfer from Slide Cartridge .................................................................. 4-6 Insertion into the Incubator ............................................................................... 4-6 Slide Disposal................................................................................................... 4-6 4.5 Incubation ..................................................................................................... 4-7 4.6 Photometer ................................................................................................... 4-7 4.6.1 4.6.2 4.6.3 4.6.4 4.6.5 Light Source Lamp ........................................................................................... 4-7 Cutting Infrared Rays off................................................................................... 4-8 Interference Filter ............................................................................................. 4-8 Photometric Head............................................................................................. 4-8 Built-in Reference Black Plate (RBP) and Reference White Plate (RWP) ....... 4-8 4.7 Standard Curve............................................................................................. 4-9 4.7.1 4.7.2 Calculation of ODr (Optical Reflection Density)................................................ 4-9 Calculation of Analyte Concentration Density .................................................. 4-9 4.8 Analyzer Calibration...................................................................................... 4-11 4.9 QC System ................................................................................................... 4-11 4.10 Calibration..................................................................................................... 4-11 4.10.1 Calibration for CRP Slide.................................................................................. 4-11 5. Control Units ................................................................................................ 5-1 5.1 Block Diagram .............................................................................................. 5-1 5.2 Connection Diagram ..................................................................................... 5-2 5.3 Circuit Board Functions ................................................................................ 5-3 5.3.1 5.3.2 5.3.3 5.3.4 5.3.5 5.3.6 5.3.7 5.3.8 5.3.9 CPU40 Circuit Board ........................................................................................ 5-3 DRV40 Circuit Board ........................................................................................ 5-3 CCD40 Circuit Board ........................................................................................ 5-3 LHA-AMP Circuit Board.................................................................................... 5-3 PRES1 Circuit Board ........................................................................................ 5-3 KEY40 Circuit Board......................................................................................... 5-3 EPM Circuit Board ............................................................................................ 5-3 RAC40 Circuit Board ........................................................................................ 5-3 USB40 Circuit Board ........................................................................................ 5-3 5.4 Sensor List.................................................................................................... 5-4 5.5 Motor List ...................................................................................................... 5-5 5.6 Others ........................................................................................................... 5-5 (vi) FDC4000 6. Software....................................................................................................... 6-1 6.1 Software Version .......................................................................................... 6-1 6.2 DIP Switch Functions.................................................................................... 6-1 6.2.1 6.3 6.4 6.5 6.6 6.7 CPU40 Board ................................................................................................... 6-1 Specific Functions of Power-on .................................................................... 6-2 Data Files...................................................................................................... 6-3 Data Collection ............................................................................................. 6-4 Version Upgrade........................................................................................... 6-4 Calculation Process ...................................................................................... 6-5 6.7.1 6.7.2 Colorimetric Tests............................................................................................. 6-5 ISE Tests .......................................................................................................... 6-6 6.8 Communication Interface.............................................................................. 6-7 6.9 2-way communication (requesting test information) [Type 1] ....................... 6-8 6.9.1 6.9.2 6.9.3 6.9.4 6.9.5 6.9.6 6.9.7 6.9.8 About This Specifications ................................................................................ 6-8 Applicability....................................................................................................... 6-8 Purpose ............................................................................................................ 6-8 System Configuration ....................................................................................... 6-8 Interface Specifications .................................................................................... 6-9 Communication Data Format............................................................................ 6-10 Communication Data Details ............................................................................ 6-12 Data Transmission Flow ................................................................................... 6-24 6.10 One-way communication [Type 3 (FDC3500 protocol)]................................ 6-26 6.10.1 Communication Parameters ............................................................................. 6-26 6.10.2 Data Transmission Specification ...................................................................... 6-26 6.11 Communication Cable Connection ............................................................... 6-30 6.11.1 D-SUB 9 pins (COM1A).................................................................................... 6-30 6.12 Sample Barcode Reader .............................................................................. 6-30 6.12.1 Manufacturer, Model number ........................................................................... 6-30 6.12.2 Data format....................................................................................................... 6-30 6.13 Mode Description.......................................................................................... 6-31 6.13.1 6.13.2 6.13.3 6.13.4 6.13.5 6.13.6 6.13.7 6.13.8 FDC4000 Mode List .......................................................................................................... 6-31 Mode 0 Changing Mode Type .......................................................................... 6-32 Mode 107 Deleting Data Files .......................................................................... 6-33 Mode 111 Inputting Parameters of RBP Level Check ...................................... 6-34 Mode 121 Displaying Cumulative Test Count .................................................. 6-35 Mode 122 Importing a Selected Data File from CF Card ................................. 6-36 Mode 123 Exporting All Data Files to CF Card................................................. 6-38 Mode 129 Inputting Detection Threshold for Un-spotted Slide......................... 6-40 (vii) 7. Disassembly and Reassembly..................................................................... 7-1 7.1 Disassembling and Reassembling................................................................ 7-3 7.1.1 7.1.2 7.1.3 7.1.4 7.1.5 7.1.6 7.1.7 7.1.8 7.1.9 7.1.10 7.1.11 7.1.12 7.1.13 7.1.14 7.1.15 7.1.16 7.1.17 7.1.18 7.1.19 7.1.20 7.1.21 7.1.22 7.1.23 7.1.24 7.1.25 7.1.26 7.1.27 7.1.28 7.1.29 7.1.30 7.1.31 7.1.32 7.1.33 7.1.34 7.1.35 7.1.36 CPU40 Board ................................................................................................... 7-3 USB40 Board <For model year up to 2008> .................................................... 7-5 USB40 Board <For model year 2009 or later> ................................................. 7-6 Operation Side Covers ..................................................................................... 7-7 Left and Right Covers....................................................................................... 7-9 Top Support Cover ........................................................................................... 7-10 Sampler Nozzle ................................................................................................ 7-11 Sampler Tube ................................................................................................... 7-12 Syringe ............................................................................................................. 7-14 Syringe Cleaning .............................................................................................. 7-15 PRES1 Board and Sampler Solenoid Valve <For model year up to 2008> ..... 7-18 PRES1 Board and Sampler Solenoid Valve <For model year 2009 or later> .. 7-20 Sampler Cover Lock Solenoid .......................................................................... 7-21 Sampler Cover Switch ...................................................................................... 7-22 Probes .............................................................................................................. 7-23 Probe Unit......................................................................................................... 7-24 Photometer Head <For model year up to 2008> .............................................. 7-26 Photometer Head <For model year 2009 or later>........................................... 7-29 Incubator Bearing ............................................................................................. 7-31 Exhaust Fan <For model year up to 2008> ...................................................... 7-32 Exhaust Fan <For model year 2009 or later>................................................... 7-33 DRV40 Board ................................................................................................... 7-34 RAC40 Board ................................................................................................... 7-34 CCD40 Board ................................................................................................... 7-35 Cooling Fan <For model year up to 2008>....................................................... 7-37 Interference Filter and Lens.............................................................................. 7-38 Printer ............................................................................................................... 7-39 KEY40 Board.................................................................................................... 7-40 Display.............................................................................................................. 7-41 QC Card Reader............................................................................................... 7-42 Top Cover ASSY .............................................................................................. 7-43 Power Supply ................................................................................................... 7-45 EPM Board ....................................................................................................... 7-46 Incubator Cable ................................................................................................ 7-47 Incubator Pressure Plate .................................................................................. 7-48 Replacing the Fuse........................................................................................... 7-49 8. Checking and Adjustment ............................................................................ 8-1 8.1 Mechanical Adjustment................................................................................. 8-1 8.1.1 8.1.2 8.1.3 8.1.4 8.1.5 8.1.6 8.1.7 (viii) Mechanical Adjustment Menu .......................................................................... 8-1 Starting up the Adjustment Software (Inspection Tool) .................................... 8-2 Incubator - Photometer position ....................................................................... 8-5 Incubator - Insert position ................................................................................. 8-6 Incubator - ISE position .................................................................................... 8-7 Sampler 1 - Tip pick 1....................................................................................... 8-8 Sampler 1 - Tip pick 2....................................................................................... 8-9 FDC4000 8.1.8 8.1.9 8.1.10 8.1.11 8.1.12 8.1.13 8.1.14 8.1.15 8.1.16 8.1.17 8.1.18 8.1.19 8.1.20 8.1.21 8.1.22 8.1.23 8.1.24 8.1.25 8.1.26 8.1.27 8.1.28 Sampler 1 - Sample.......................................................................................... 8-10 Sampler 1 - Diluted sample 1 ........................................................................... 8-11 Sampler 1 - Diluted sample 2 ........................................................................... 8-12 Sampler 1 - Diluent........................................................................................... 8-13 Sampler 1 - Spot 1............................................................................................ 8-14 Sampler 1 - Spot 2............................................................................................ 8-15 Sampler 2 - Tip pick 3....................................................................................... 8-16 Sampler 2 - Tip pick 4....................................................................................... 8-17 Sampler 2 - Reference fluid.............................................................................. 8-18 Sampler 2 - Spot 3............................................................................................ 8-19 Sampler 2 - Spot 4............................................................................................ 8-20 Sampler 2 - Tip eject ........................................................................................ 8-21 Sampler 2 - Spot standard position .................................................................. 8-22 Slide - CM spot position.................................................................................... 8-23 Slide - ISE sample spot position....................................................................... 8-24 Slide - ISE reference fluid spot position............................................................ 8-25 Slide - Slide insert position ............................................................................... 8-26 Slide - Insert standard position ......................................................................... 8-27 Filter - HP adjust............................................................................................... 8-28 Filter - First data sampling ................................................................................ 8-29 Filter - Reference black plate initial setting....................................................... 8-30 8.2 Mechanical Test Mode.................................................................................. 8-31 8.2.1 8.2.2 8.2.3 8.2.4 8.2.5 8.2.6 8.2.7 8.2.8 8.2.9 Mechanical Test Mode Menu ........................................................................... 8-31 Starting up the Mechanical Test Mode ............................................................. 8-31 Sensor .............................................................................................................. 8-32 Filter.................................................................................................................. 8-33 Slide.................................................................................................................. 8-34 Sampler ............................................................................................................ 8-35 Motor ................................................................................................................ 8-36 Serial Com Test................................................................................................ 8-36 Sample BCR Read Test ................................................................................... 8-36 8.3 Analyzer Calibration...................................................................................... 8-37 8.4 Voltage Inspection ........................................................................................ 8-38 8.4.1 8.4.2 8.4.3 Power Supply ................................................................................................... 8-38 CPU40 Board ................................................................................................... 8-40 DRV40 Board ................................................................................................... 8-42 9. Troubleshooting ........................................................................................... 9-1 9.1 Troubleshooting Guide ................................................................................. 9-1 9.1.1 9.1.2 9.1.3 9.1.4 9.1.5 9.1.6 9.1.7 9.1.8 9.1.9 FDC4000 H, L, >, <, @ (Warning for test results) High or low value or qualitative range 9-1 $ (Warning for test results) Thermistor disconnection ..................................... 9-3 + (Warning for test results) Temperature error ................................................ 9-4 - (Warning for test results) Temperature error ................................................. 9-4 * (Warning for test results) Fluctuation of RWP readout ................................ 9-5 ? (Warning for test results) RWP readout is outside of the range ................... 9-6 E (Warning for test results) Photometer malfunction ....................................... 9-7 & (Warning for test results) Abnormally high testing value .............................. 9-8 # (Warning for test results) Expiration of slide or calibration ........................... 9-8 (ix) 9.1.10 9.1.11 9.1.12 9.1.13 9.1.14 9.1.15 9.1.16 9.1.17 9.1.18 9.1.19 9.1.20 9.1.21 9.1.22 9.1.23 9.1.24 9.1.25 9.1.26 9.1.27 9.1.28 9.1.29 9.1.30 9.1.31 9.1.32 9.1.33 9.1.34 9.1.35 9.1.36 9.1.37 9.1.38 9.1.39 9.1.40 9.1.41 9.1.42 9.1.43 9.1.44 9.1.45 9.1.46 9.1.47 9.1.48 9.1.49 9.1.50 9.1.51 9.1.52 9.1.53 9.1.54 9.1.55 9.1.56 9.1.57 (x) ¥ (Warning for test results) Un-spotted slide ................................................... 9-9 ERR= *** (Errors for ISE test results) ........................................................ 9-10 C (Control mode) ............................................................................................. 9-13 W010 Keyboard open ...................................................................................... 9-14 E010 Keyboard open ....................................................................................... 9-15 W020 Disposal box open ................................................................................. 9-16 E021 Disposal box open .................................................................................. 9-16 W022 Disposal box is full ................................................................................ 9-17 W025 Sampler cover open .............................................................................. 9-18 E025 Sampler cover open ............................................................................... 9-19 W026 Cover lock failure .................................................................................. 9-20 W027 Sampler cover unlock error ................................................................... 9-21 W030 Light source lamp replacing warning ..................................................... 9-22 E035 Light intensity over range ....................................................................... 9-23 E036 Dirty RBP ............................................................................................... 9-24 W040 Wrong sample type warning .................................................................. 9-25 E050 Lamp replacement error ......................................................................... 9-26 W060, W061 Insufficient sample / Insufficient calibrator ................................. 9-27 W070 Degraded interference filter ................................................................... 9-28 W081 No sample rack for calibration ............................................................... 9-29 W090, W097, W099 Communication error ...................................................... 9-30 E0100 Faulty tip sensor ................................................................................... 9-31 W110 No tip warning ....................................................................................... 9-32 E0110, E0111 Clogging error during aspirating sample .................................. 9-33 W112 Wrong test programming ....................................................................... 9-34 E0112, E0113 Clogging error during spotting sample ..................................... 9-35 W115, W116 Tip ejection warning ................................................................... 9-36 W120 Tip error ................................................................................................. 9-37 E0120 No sample found .................................................................................. 9-38 E0121, E0122 Clogging error (diluent) ............................................................ 9-39 E0123 No diluent found ................................................................................... 9-40 E0124, E0125 Clogging error (diluted sample) ............................................... 9-41 E0126, E0127, E0128 Clogging error (reference fluid) ................................... 9-42 E0129 No reference fluid found ....................................................................... 9-43 W140 Incorrect direction loading of ISE slide .................................................. 9-44 E0140, E0141, E0142 Excessive liquid volume (beyond the upper limit) ....... 9-45 W141 Impossible to start ISE tests .................................................................. 9-46 E0143 Insufficient diluent ................................................................................ 9-47 E0144 Insufficient reference fluid .................................................................... 9-48 E0145 No calibrator found ............................................................................... 9-49 E0146 Excessive calibrator volume (beyond the upper limit) .......................... 9-50 W150 Dilution tests over .................................................................................. 9-51 W160 Dilution factor error ................................................................................ 9-51 W170 No QC information ................................................................................ 9-52 W173 Slide reading error ................................................................................. 9-53 W175 No slides loaded .................................................................................... 9-54 W180 No calibration performed ....................................................................... 9-55 W190, W191 Wrong slide loading in calibration process ................................ 9-55 FDC4000 9.1.58 9.1.59 9.1.60 9.1.61 9.1.62 9.1.63 9.1.64 9.1.65 9.1.66 9.1.67 9.1.68 9.1.69 9.1.70 9.1.71 9.1.72 9.1.73 9.1.74 9.1.75 9.1.76 9.1.77 9.1.78 9.1.79 9.1.80 9.1.81 9.1.82 9.1.83 9.1.84 9.1.85 9.1.86 9.1.87 9.1.88 9.1.89 9.1.90 9.1.91 9.1.92 9.1.93 9.1.94 9.1.95 W200 Sample barcode reader error (No response) ........................................ 9-56 E0200 - E0210 Feed motor error ..................................................................... 9-57 E0300, E0301, E0302 Gain error .................................................................... 9-59 W500, W501 Fan rotation stop ........................................................................ 9-60 E0500 RTC backup error ................................................................................. 9-61 E0501 RTC control error ................................................................................. 9-61 E0509 ADC calibration error ............................................................................ 9-62 E0510 A/D conversion error ............................................................................ 9-62 E510 RWP reading error ................................................................................. 9-63 E0530 ISE self-test error (gain) ....................................................................... 9-64 E0531 ISE self-test error (offset) ..................................................................... 9-65 E0532 A/D conversion error for ISE ................................................................ 9-66 E0600, E0601, E0602, E0603 Incubator motor error ...................................... 9-67 E0700, E0701, E0702 Probe motor error ........................................................ 9-69 E0900 - E0903 Filter motor error ..................................................................... 9-70 E1000 - E1002 Sampler vertical drive error .................................................... 9-71 E1100 - E1103 Sampler rotational drive error ................................................. 9-72 E1200 - E1204 Syringe motor error ................................................................. 9-73 E1500 CM temperature not ready ................................................................... 9-75 E1501 CM temperature control error ............................................................... 9-76 E1502 CM thermistor disconnection ................................................................ 9-77 E1503 CM heater disconnection ..................................................................... 9-78 E1504 ISE temperature not ready ................................................................... 9-79 E1510 ISE temperature control error ............................................................... 9-80 E1511 ISE temperature sensor disconnection ................................................ 9-81 E1512, E1513 Abnormal high temperature ..................................................... 9-82 E1514 Temperature warning ........................................................................... 9-83 E1521 High environmental temperature .......................................................... 9-83 E30xx Memory error ........................................................................................ 9-84 E4000 - E4010 QC card error .......................................................................... 9-88 W5000, E5100 Printer error ............................................................................. 9-89 E5000 - E5034 CF card error .......................................................................... 9-90 E5050 No CF card has been set. .................................................................... 9-92 E5201 - E5204 Malfunction of the slide reader ............................................... 9-93 E5500 Spotting control error ............................................................................ 9-94 E5510 Slide position error ............................................................................... 9-94 E9999 Impossible to continue testing .............................................................. 9-96 EFF01 - EFF0F Calculation error .................................................................... 9-97 9.2 Others ........................................................................................................... 9-98 9.2.1 Cover Lock Release ........................................................................................ 9-98 10.Slide Information......................................................................................... 10-1 FDC4000 (xi) (xii) FDC4000 1 Safety and Precautions 1 Safety and Precautions 1.1 Definition of Specific Safety Precautions Specific safety precautions are noted by the terms WARNING, CAUTION, IMPORTANT, and additional information by NOTE. The respective meanings are as follows: 1 WARNING Indicates hazardous situations that may lead to serious injury, even death or transmission of infectious agent if the precaution is not followed. CAUTION Indicates hazardous situations that may lead to minor or moderate injury or physical damages if the caution is not followed. IMPORTANT Indicates improper handling that could have an adverse effect on the accuracy of the measurement values. NOTE: Indicates procedures requiring special attention, instructions that must be followed, supplementary explanations, etc. 1.2 Precautions before Operating This Equipment CAUTION Before servicing this equipment, please read this Manual carefully so that you can operate it correctly. CAUTION Whenever you operate this equipment, be sure to observe the precautions in the manual. Failure to do so may cause you to subject to injuries, property damage, or incorrect test results to occur. CAUTION This equipment is an in vitro diagnostic medical device. This equipment is classified as an in vitro diagnostic medical device under DIRECTIVE 98/79/EC. Intended use of this equipment is to quantitate the concentration or the activity of the components in blood by using the FUJI DRI-CHEM SLIDEs. Do not use the equipment for other purposes. Please read the Instructions for Use included with the slides carefully to follow the instructions. FDC4000 Service Manual 1-1 1 Safety and Precautions CAUTION This equipment is only to be operated by personnel appropriately trained for the intended use and the operations. Patients do not come into direct contact with the equipment. 1 CAUTION Do not remodel the FDC4000. Otherwise, the safety will not be guaranteed. IMPORTANT Maintenance or repair may affect the accuracy of test results. Perform quality control (running control fluids) to ensure the reliability of the test results. Refer to Section 6 of the Instruction Manual for details. 1.3 Biohazards and Disposal WARNING As used (contaminated) consumables (e.g., FUJI DRI-CHEM SLIDEs, FUJI DRI-CHEM AUTO TIP, FUJI DRI-CHEM MIXING CUPS S and sample tubes) and contaminated swabs or cloths used for cleaning the equipment are infectious waste, process the waste correctly in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization or disinfection. WARNING When discarding the FDC4000 body that may be contaminated with blood samples, be sure to process it correctly in compliance with any applicable regulations in your country because it cannot be disposed of as a general waste. WARNING When handling samples (blood or urine) and performing maintenance (cleaning the analyzer), always follow biohazard procedures (e,g., wearing gloves, lab coat, and safety goggles). If any part of the body comes in contact with samples, immediately rinse the contaminated body part thoroughly under running water and then use ethyl alcohol as a disinfectant. Seek medical assistance if necessary. WARNING Do not touch used slides and tips with bare hands as this may cause contamination. If any part of the body comes in contact with contaminated consumables, immediately rinse the contaminated body part thoroughly under running water and then use ethyl alcohol as a disinfectant. Seek medical assistance if necessary. WARNING When samples come in contact with the analyzer components, immediately clean and disinfect the components. 1-2 FDC4000 Service Manual 1 Safety and Precautions 1.4 Disinfection WARNING To avoid biohazards during maintenance and repair work, Safety Control authorized by your site is required. The following are general cautionary measures: - You must always take care to prevent infection. - Always follow biohazard procedures (e,g., wearing gloves, lab coat, and safety goggles) during the work. - Before starting the work, disinfect all possible places that have come in contact with blood samples. - For disinfection, use ethyl alcohol for disinfection. (Only for disposal box cleaning, 0.5% of sodium hypochlorite acid disinfectant can be used.) - When blood adheres, remove the blood stain with a cotton swab moistened with water. For example: Sample rack, Sampler nozzle, Spotting part, Transfer unit, Incubator, Disposal box, etc. 1.5 Cleaning Solutions CAUTION Refer to each section of the instruction manual for information about usable solvents for cleaning. Do not use cleaning solution containing ammonia on or near the analyzer. CAUTION Do not use alcohol for cleaning the sampler cover (translucence), otherwise the surface will be damaged. 1.6 Explosive Hazards WARNING As this equipment is not explosion-proof type, be sure not to use flammable and explosive gas around the equipment. FDC4000 Service Manual 1-3 1 1 Safety and Precautions 1.7 Electrical Hazards WARNING The power supply voltage applied to the equipment is AC100-240V. To avoid electrical shock, observe the following precautions: - Avoid installation sites where water may splash, etc., on the equipment. - Make sure that the equipment is properly grounded to a protective earth lead for indoor wiring. - Make sure that all cables have been properly connected. 1 WARNING Plug the power cable of the equipment into an outlet with grounding receptacle. If the equipment is not grounded to a protective earth, this may cause electrical shock. WARNING When plugging the power cable in or out of an outlet, be sure to hand the plug body, not the cable to prevent damage to the cable. WARNING Before starting the maintenance, be sure to shut off the power and unplug the power cable to avoid electric shock, burn, and equipment damage. Use due care during maintenance since equipment contains parts that are not fully discharged and those which are burning hot (such as a lamp house) even after power off. Exercise due care to avoid electric shock and other hazards if the power cannot be turned off. CAUTION The equipment includes parts and assemblies sensitive to electrostatic discharge damage. Use caution to prevent damage during all service procedures. 1.8 Safety Devices WARNING Note that the safety devices such as fuses, covers, and interlock switch meet their specified fault-proof requirements. Never attempt to make modification or alternation to those devices. 1.9 Moving Parts WARNING To avoid danger, be sure to shut off the power before inspection and adjustment. 1-4 FDC4000 Service Manual 1 Safety and Precautions WARNING Do not place your hands near moving parts (sampler, slide transfer lever, incubator) whenever operating the equipment including maintenance. Also use care no to get your hands, hair, clothing, or accessories caught in moving parts. 1 CAUTION Do not wear a necktie, necklace, or other accessories that may get caught in moving parts. 1.10 Heavy Objects CAUTION Get an assistant to help you during installation and removal of heavy objects. Otherwise, use appropriate machinery to handle them. 1.11 General Precautions Use the genuine parts listed on the Parts List. Use the standard tools for adjustment and inspection. FDC4000 Service Manual 1-5 1 Safety and Precautions 1.12 Warning Labels 1 Warning labels and safety labels on the FDC4000 are: - High temperature caution label - Lamp caution label - Biohazard label - Chemicals caution label - Power switch - Name-plate Name-plate 405N100565 High temperature caution label Lamp caution label Refer to Sec.5.5 (P5-13) in the Instruction manual. Power switch | : Depress to power ON O : Depress to power OFF Biohazard label Chemicals caution label Refer to Sec.1.13 (P1-9) in the Instruction manual. 1-6 FDC4000 Service Manual 2 Installation 2 Installation NOTE: Install the FDC4000, following the "Quick Guide for Installing the FUJI DRI-CHEM 4000" and this chapter. 2.1 Installation Space OO OO OO 2 (&% NOTE: Empty space on the back is necessary to unplug the power cable connector from the appliance inlet of the analyzer. 2.2 Installation Site Requirements CAUTION Avoid the following installation sites: Places where spills or water leakage may occur. Places where the equipment is exposed to direct sunlight. Places near hot sources such as heaters. Places where temperature drastically changes. Places on an unstable table or where the equipment is subject to vibration. 2.3 Environmental Requirements Location: Illumination: Altitude: Indoor use Below 6,000 cd/m2 (lux) (Below 3,000 cd/m2 (lux) when using the sample barcode reader) Up to 2,000m Transient overvoltage category: II Pollution degree: Temperature Relative humidity 2.4 2 Operation +15 to +32 °C (59 to 89 °F) Transit or storage -10 to +50 °C (14 to 122 °F) 30 - 80 %RH 10 - 90 %RH Remarks No condensation during operation and transportation / storage Electrical Requirements Plug the power cable of the equipment into an outlet with grounding receptacle. If the equipment is not grounded to a protective earth, this may cause electrical shock. Voltage limit: 100 - 240V~ Frequency: 50 - 60 Hz Supply voltage fluctuations: ± 10% Rated wattage: FDC4000 Service Manual 200VA 2-1 2 Installation Phase: Type of protection against electrical shock: Single CLASS 1 EQUIPMENT NOTE: Plug the FDC4000 into an independent AC outlet separate from other devices. NOTE: AC power cable is not packed as an accessory. Please prepare an AC power cable conforming to the specifications below: 2 <The requirements of the cable specifications for the FDC4000> Power voltage: 100V-120V Requirements for the plug/connector: AC125V 10A Requirements for the cable: SVT 3/18AWG 60°C Power voltage: 200V-240V Requirements for the plug/connector: AC250V 10A Requirements for the cable: GTCE-3 1.0mm2 70°C NOTE: The AC power cable is required to be compliance with any applicable regulations in your country. 2.5 Unpacking CAUTION To lift or carry the analyzer, be sure to grip the handle portions with both hands as shown the figure. In winter, be sure to leave the equipment in its packing condition at room temperature for a sufficient time to avoid moisture condensation on the equipment. Handle portion 2.6 Package Contents After unpacking, perform the installation, following the "Quick Guide for Installing the FUJI DRI-CHEM 4000" and this chapter. In accordance with the Packing List including with the package, check that the contents comply with the List. 2-2 FDC4000 Service Manual 2 Installation 2.7 Removing Tape and Fixtures Remove tape and fixtures (placed on the outer covering) for transportation. After opening the sampler cover, remove tape and fixtures inside of the cover. 2.8 Preparations Before Turning the Power on In accordance with the "Quick Guide for Installing the FUJI DRI-CHEM 4000", perform the following: (1) Take the incubator (B) from the accessories box and press the red switch of the thermostat. (2) Assemble the incubator (B) into the analyzer. (3) Assemble sample racks. (4) Set recording paper. (5) Connect the power cable. 2.9 2.9.1 Setting up the Optional Accessories Sample Barcode Reader CAUTION Do not connect a barcode reader other than specified for the FDC4000. When using the sample barcode reader, refer to the following: - Section 4.5 "Sample Barcode Reader" (Instruction manual) - Section 8.2.20 "Mode 46 Selecting Communication Destinations" (Instruction manual) - Section 6.12 "Sample Barcode Reader" (Service manual) 2.9.2 Communication with Host Computer or PC CAUTION Do not connect the FDC4000 to the host computer or PC which has not been approved by IEC/UL60950-1. When connecting the analyzer to host computer of PC, refer to the following: - Section 9.2 "Data Communication" (Instruction manual) - Section 8.2.20 "Mode 46 Selecting Communication Destinations" (Instruction manual) - Section 6.8 "Communication Interface" (Service manual) - Section 6.11 "Communication Cable Connection" (Service manual) 2.10 Turning the Power on In accordance with the "Quick Guide for Installing the FUJI DRI-CHEM 4000", perform the following: (1) Turn the power switch on. (2) Adjust date and time (using Mode 20). (3) Read QC cards. NOTE: The clock IC on the CPU40 board is backed up by a lithium battery (rechargeable). Fully charging the battery needs 40 hours, so the board is not fully charged when shipping and the backup will expire about 1 month after shipping. Be sure to check the data & time when installing the analyzer. FDC4000 Service Manual 2-3 2 2 Installation 2.11 Acceptance Check the system accuracy using the FUJI DRICHEM CONTROL QP-L and/or QP-H and FUJI DRICHEM SLIDEs by following Sec. 6 "Quality Control" in the instruction manual. Check the system accuracy whether the test results are within the expected ranges of the controls. If required, perform the correlation test between the user's conventional measuring system and the FDC analyzer and input correlation coefficients (a, b) by following Sec. 8 "Mode 36" in the instruction manual to adjust the FDC analyzer to the user's system. 2 2-4 FDC4000 Service Manual 3 Compositions and Name of Parts 3 Compositions and Name of Parts 3.1 Component Names CF card slot Printer Display Sampler cover 3 Keyboard Cover lock LED (G) Sampler cover (A) Keyboard (H) Sampler (I) Slide cartridge (B) Incubator cable (C) Incubator A 3 (K) Sample rack (D) Spotting part (L) Card reader (E) Air filter (inside) (F) Disposal box FDC4000 Service Manual (M) Slide reader (inside) 3-1 3 Compositions and Name of Parts Card reader Power switch 3 Communication select switch USB connector Fuse holder Power plug RS-232C connectors Disposal box Light source lamp Air filter (insaide) 3-2 FDC4000 Service Manual 3 Compositions and Name of Parts FUJI DRI-CHEM MIXING CUPS S Tip setting position “b” (for diluent) Diluent hole “D” Reference fluid setting position Blood collection tube Tip setting position “c” (for electrolyte test sample) Tip setting position “a” (for sample) Tip setting position “d” (for reference fluid) 3 D d c b MIX CUP a RE FDC4000 Service Manual 3-3 3 Compositions and Name of Parts 3 3-4 FDC4000 Service Manual 4 Functions 4 Functions 4.1 Measurement Principle of Colorimetric Tests Sampler nozzle Slide cartridge Slide Te mperature sensor Incubator Heater Tip Bar Potentiometer Slide disposal box Slide reader Photometric head Droplet Spotting position Measurement process 4.1.1 Preparation for Measurement Before starting measurements, QC card included with the slide package should be read to compensate for the variation between the slide lots. When performing measurements, load slides to be tested in the slide cartridge, and load a sample and a tip in the sample rack followed by the START key. For measurements that require dilution, additionally load a mixing cup S with diluent. 4.1.2 Process of Measurement When the START key is pressed, the CCD device reads dot information printed on the back of the slide in the slide cartridge. The dot information contains the slide's test name and the slide production lot number. According to the information, the analyzer aspirates and spots the predetermined spotting volume for the slide. First, the sampler nozzle picks a tip up from the sample rack. Next, a tip detecting sensor checks the presence of the tip fitting to the nozzle. The sampler moves and stops above the sample fluid, then lowers. The analyzer monitors the change in air pressure as the tip reaches the sample by blowing air, and determines the sample surface. The point of the tip is controlled so that it does not enter deeply into the sample. The analyzer aspirates the sample fluid and then dispenses it before the first spotting. This function is to keep the accuracy of the spotting volume for the first spotting and the following spottings. The analyzer monitors the air pressure in the nozzle and judges the stickiness of the sample. If a sample is abnormal, the alarm will be informed. After the sample aspiration, the nozzle moves to the spotting position, and lowers the tip to the fixed position above the slide. After that, the analyzer dispenses a fixed volume of sample fluid to make a round droplet at the orifice of the tip. The analyzer lowers the nozzle more to make the droplet touch the slide surface, so that the sample fluid spreads on the surface of the slide. The spotted slide is inserted into the incubator to be incubated in the closed space for preventing evaporation. The incubator contains a heater and a temperature sensor to keep the temperature at 37±0.2°C. The incubator also has apertures at the bottom of the cells. Through the apertures, the photometer measures the changed color of the slide. The incubator reciprocally rotates and the analyzer measures the changed color of the slide periodically. Also, the photometer reads the reflectance of the reference black plate (RBP) & the reference white plate (RWP) in each rotation. After the measurement completed, the slide is pushed out to the center of the incubator to be discarded into the disposal box. FDC4000 Service Manual 4-1 4 4 Functions 4.1.3 Photometer and Calculation of Analyte Concentration Filter Wheel Slide Photo-sensor Glass Plate Interference Filter Head Amplifier Condenser Lens Focus Lens Lamp Focus Lens Optical Fiber IR Filter Optics 4 The analyzer uses the separated light of a halogen lamp through the interference filter. The light of the halogen lamp is gathered by the condenser lens and collimated. The IR filter blocks the infrared rays. The interference filter on the filter wheel separates a narrow band of the spectrum. After the collimated spectrum is focused on the end of the optical fiber by the focus lens, it is guided to the photometric head. In the photometric head, the light from the optical fiber is focused by another focus lens to project equally over the reading area of the slide. The reflected light on the reading area is detected by the photo sensor and converted into an electric signal. The signal from the photo sensor is amplified with the head amplifier and the pre-amplifier. The signal range varies with each of the photometric wavelength so the gain of the pre-amplifier is adjusted to make the output signal level the same. The output signal is input into the A/D converter circuit. The output value after A/D conversion is compared with the output value of the RBP and RWP, and the reflected OD value is calculated. The reflected OD value is converted into an analyte concentration value using the pre-determined standard curve. In this calculation, the information from the QC card compensates for the result. The analyzer has 4 kinds of compensation methods: - The compensation of correlation between FDC and other methods (a, b) - The slide production lot calibration using the QC card (c, d, e) - The compensation (Analyzer Calibration) of OD value (j, k) - The compensation of analyte density value (l, n) 4.2 4.2.1 Measurement Principle of ISE Tests Preparation for Measurement For ISE tests, the preparation is fundamentally same as CM tests. Additionally, load reference fluid for ISE tests. 4.2.2 Process of Measurement After the analyzer identifies an ISE slide as same way as CM tests, the analyzer picks tips for both the sample nozzle and the reference fluid nozzle to aspirate sample and reference fluid. The analyzer dispenses sample and reference fluid into the holes on the ISE slide. After spotting, the slide is inserted into the incubator. The slide is inserted into the specified cell in the incubator to be incubated at 30 +/1°C (or environmental temperature + 2°C). Six testing probes (3 pairs) rise up and touch the ion selective electrodes on the slide for testing genetic potential. Due to the mechanical structure, CM tests and ISE test cannot be performed at the same time. After the test completed, the slide is disposed of into the disposal box. 4-2 FDC4000 Service Manual 4 Functions 4.3 Sampling 4.3.1 Mounting Tip and Sealing For the sampling of sample fluid, a disposal tip (FUJI DRI-CHEM AUTO TIPS) is used. The outside of the top part is waterproofed. The nozzle is inserted into the tip using a fixed force and the joint surface is sealed with an O-ring. Insufficient inserting force or O-ring defection due to damage or aging causes leakage between the tip and the nozzle and may result in a suction error or a spotting error. Also, if there is soil something like oil on the tip sealing point or the nozzle surface, it may cause the tip dropout. Before picking up After picking up Nozzle Tip Sealing point O-ring 4.3.2 Sample Surface Detection The sample surface is detected by monitoring the pressure change caused by blowing air from the orifice of the tip onto the surface of the sample fluid. When the sampler nozzle lowers, the syringe piston moves forward and air is blown from the orifice at the tip. A pressure sensor is connected to the suction tube. When the tip is touches the sample surface, the sensor detects the pressure change. The sensor output is computed in differential calculus for increasing sensitivity. When its value exceeds a threshold value, the comparator turns on and determines the sample surface. When detecting the sample surface, the sampler stops lowering. The distance from the starting point to the detecting point is memorized as the sample surface position. After the first aspiration, the decreased sample surface is calculated with the previous aspiration volume to determine the next sample position. The orifice of the tip is usually inserted about 2mm into the sample fluid. It is essential that sample has no bubbles on the surface. Otherwise, the bubble membrane will be falsely detected and the orifice of the tip will not reach the correct position, which may cause empty aspiration and spotting. Comparator Tube Sampler Sensor output Threshold Pressure sensor Tip Syringe Differentiated signal Sensor output Time Sample fluid FDC4000 Service Manual 4-3 4 4 Functions 4.3.3 Sample Aspiration During the sample aspiration, the analyzer pre-aspirates and pre-dispenses in order to moisten the inside-wall of the tip before the first aspiration. This pre-aspiration improves the precision of the spotting volume. Subsequently the sample fluid for the first spotting is aspirated. The aspiration volume is determined by QC information of the slide. Aspiration errors caused by sample stickiness or fibrin are detected from the pressure change during the pre-aspiration, pre-dispense and the final aspiration. After the aspiration, the sampler pulls up the tip from the sample surface at a fixed slow speed. This prevents the sample fluid from adhering to the outside wall of the tip. If the sample is adhered to the outside wall of the tip, it may cause a spotting error. 4.3.4 Dilution The tests that require dilution can be performed by selecting dilution factors. The sampler dispenses sample and diluent into the right hole of mixing cup, and after mixing, the diluted sample is spotted onto the slide. <Dispensing volume into mixing cup> 4 4-4 Dilution factor Sample (µL) Diluent (µL) 2 50 50 3 30 60 4 25 75 5 20 30 10 10 90 21 5 100 FDC4000 Service Manual 4 Functions 4.3.5 Spotting Sample The sampler nozzle turns to the spotting position and lowers to the spotting preparation position, then it dispenses a proper volume of sample according to the slide information. The dispensed sample fluid makes a round droplet at the orifice. The tip lowers to a fixed position. Then the round droplet spreads on the slide. If the position is too close to the slide, the dispensed volume is likely to increase. If the position is too far, the dispensed volume is likely to decrease. The spotting position of the slide should be the center of the slide. In addition, the position must match the center of the photometric reading area when the slide is inserted in the incubator. Making droplet Spotting Tip Droplet 4 Slide 4.3.6 Tip Disposal The used tip is ejected at the tip eject part to be disposed of into the disposal box. FDC4000 Service Manual 4-5 4 Functions 4.4 4.4.1 Slide Transfer Slide Dot Code Reading The dot information printed on the back of slide contains the slide's test name (including sample type), the slide production lot number and the slide loading direction. The CCD device located under the slide cartridge loading position reads the dot information. A LED illuminates the dot printing part of the slide. 4.4.2 Slide Transfer from Slide Cartridge The stacked slides in the slide cartridge are transferred one by one from the bottom to the spotting position. A weight must be put on top of the stacked slides. This makes the slides stable to prevent miss-transfer by the slide transfer bar. Weight Incubator Slide cartridge Spotting position Slides Cover 4 Slide transfer bar Disposal box 4.4.3 Insertion into the Incubator The spotted slide is inserted into the photometer reading position in the incubator. If there is misalignment for the transfer base height or the cell position, a transfer error will occur. The analyzer performs photometric measurements when stopping the incubator rotation during the rotation process. A leaf spring gives small force on top of the pressure plate to prevent out of slide position due to the centrifugal force and the vibration caused by the incubator's rotation. 4.4.4 Slide Disposal After the measurement, the slide is pushed out to the center of the incubator, and the slide is discarded into the disposal box through the center hole in the incubator. 4-6 FDC4000 Service Manual 4 Functions 4.5 Incubation Upper Incubator block Incubator cable Pressure plate Te mperature sensor Spring Heater Shaft Slide 4 Reflection reading aperture Lower incubator block The incubation works by pressing the spotted slide between the upper incubator block (kept at 37 °C) and the lower incubator rotor. The incubator keeps the slide at this temperature and prevents the evaporation of the sample fluid from the slide during incubation. The upper incubator is made of thick aluminum, and contains a heater and a temperature sensor. It has pressure plates to close the top of the slide tightly. Also, the roles of the pressure plates are conducting heat from the heater and preventing out of the slide position. The incubator has springs to press the pressure plates on the slides. Owing to atmospheric temperature, there is a little temperature deviation between the part where the temperature sensor is mounted and the slide. In order to compensate for the deviation, the analyzer has an external temperature sensor on the keyboard to measure atmospheric temperature. The surface of the slide is shut tightly by the pressure plate, which is covered with special plastic to prevent adhesion of pollutants such as gas which may cause contamination for other slides. The pressure plate must be cleaned regularly to prevent the adhesion of the pollutants. An ISE slide is controlled at 30 °C in another specific cell (Cell 7) of the incubator. 4.6 4.6.1 Photometer Light Source Lamp The analyzer uses a 10w halogen lamp. The surface of the halogen lamp during lighting is kept at a fixed high temperature so that the halogen cycle may keep in an appropriate way in the lamp. If something like a fingerprint stains the surface of lamp, the light intensity will decrease. The analyzer determines whether the light intensity is sufficient or not by measuring the reflection reading of the built-in RWP at 400nm wavelength. If the intensity is not sufficient, lamp warnings or errors will occur. Because the light intensity is not stable during warming-up, the analyzer waits until the warming-up process completed after the power is turned on. When the analyzer is not used, the lamp turns off automatically to extend the life of the lamp. Once the light is turned off, a wait of several minutes is required before restart. FDC4000 Service Manual 4-7 4 Functions 4.6.2 Cutting Infrared Rays off Infrared rays from the lamp are blocked by the infrared cut filter (IR filter). Double wavelength light from the interference filter is also blocked. 4.6.3 Interference Filter If the stop position of the filter rotation deviates from the center, the precision of optical reading will be degraded because the light passes through whole area of the interference filter. During the measurement, the analyzer monitors the RWP value to maintain the stability of the measurement precision. When the value is unstable, photometric errors will occur. The interference filter has a certain deviation allowance for the peak wavelength. The peak of the color spectrum of some slides does not match the peak wavelength of the filter, so that the difference may affect the accuracy of the test results. The analyzer calibration compensates for this mismatch by software calculation. 4.6.4 4 Photometric Head Through the focus lens, the light conducted through the optical fiber from the light source is projected onto the colored area of the slide. The reflected light from the slide is measured by 2 photo-sensors. The difference between the center of the projected spot and the center of the colored part of the slide causes a reading error. The OD for the slide reading is calculated by using the RBP/RWP readings. Therefore, the reading deviation is small enough for the reading range close to the RBP/RWP, but the deviation exists around the intermediate range. This performance is called "linearity". For some test names, this linearity affects to the measured value. Therefore the analyzer performs compensation automatically for the specific test names. A glass plate covers the top of the photometric head for protection. Dust on the glass plate gives same level of effect on the reading accuracy for both RBP/RWP readings and the test slide readings, so that the effect can be cancelled. But regular cleaning is necessary to reduce the effect on the linearity. If the distance between the photometric head and the slide is changed, the measurement value is affected. Because the signal from the photo-sensor is low-intensity, the head amplifier is located near the photometric head, and the signal line from the sensor to the head amplifier is highly insulated. 4.6.5 Built-in Reference Black Plate (RBP) and Reference White Plate (RWP) The specific OD values for the RBP and the RWP are predetermined. The analyzer calculates the analyte concentration of the slide by using these OD values as reference. The RBP and the RWP sometimes are soiled with dust, smoke, oil vapor, etc. from air, so that periodic cleaning is necessary. 4-8 FDC4000 Service Manual 4 Functions 4.7 4.7.1 Standard Curve Calculation of ODr (Optical Reflection Density) The Optical Reflection Density (ODr) is calculated using the following equation: ODr = log { ( W - B) / [10-ODW( S - B ) - 10-ODb( S - W ) ] } (4.7.1.1) Where 4.7.2 W B S ODw ODb :Measured RWP reflection value :Measured RBP reflection value :Measured slide reflection value :OD value of RWP given by the standard reflectometer :OD value of RBP given by the standard reflectometer Calculation of Analyte Concentration Density (1) Method of End-Point Measuring time of the end-point tests is between 2 and 6 minutes. When the measurement is completed, the reflectance ODr is determined by least squares of 4 readings. ODr 1.0 0 㧡minutes Time The analyte concentration is calculated using the following equation: Analyte concentration = a0 + a1 ( ODr ) + a2 ( ODr )2 + a3 ( ODr )3 + a4 ( ODr )4 (4.7.2.1) Where an is defined by the standard curve coefficient of each slide. FDC4000 Service Manual 4-9 4 4 Functions (2) Method of Rate Test Measuring time of the rate tests is between 2 and 5 minutes When the measurement is completed, ODr (OD reading in the Window of the test) is converted into ODt. The Window is between 1 and 3 minutes (depends on the test name). A cubic equation or William-Clapper equation is used for the conversion from ODr to ODt. The rate of reaction v (v= ⊿ ODt/ ⊿ t) is calculated using ODt. The enzyme activity is calculated using the following equation: Enzyme activity = a0 + a1 v + a2 v2 + a3 v3 + a4 v4 (4.8.2.2) Where an is defined by the standard curve coefficient of each slide. ODr Window 1.0 4 0 㧡minutes Time (3) Determination of the CRP concentration The measuring time of CRP is 6 minutes. The Window is set at 4 - 6 minutes. The change of ODr [v= ⊿ ODr (4-6min)] within the Window Time is converted to the CRP concentration using the following equation: CRP concentration = a0 + a1v + a2 v2 + a3 v3 (4.8.2.3) Where an is defined by the standard curve coefficient of each slide. (4) ISE tests The potential after 1 minute from starting a test is converted to the analyte concentration using the following equation: Analyte concentration = 10 (E+S)/K (4.8.2.4) Where E: Measured potential S: Intercept of the standard curve K: Slope of the standard curve 4-10 FDC4000 Service Manual 4 Functions 4.8 Analyzer Calibration There are 2 ways to compensate variation between analyzers. One is compensation for OD value; another is for analyte concentration. 4.9 QC System The QC system compensates the variations between FUJI DRI-CHEM SLIDE's production lots. The system enables to maintain measurement accuracy without liquid calibration as long as the slides are handled in the defined manner and the analyzer's quality control is maintained. The compensation is performed using the following equation: Where, X: Measurement value before compensation Compensated concentration = c + dX + eX2 (4.9.1) Coefficients (c, d, e) are recorded on QC card included with the slide package. The FDC4000 is able to memorize maximum 2 lots of QC card data for each test name. When more than 2 lots of QC information are read, the oldest QC information will be deleted. It is different from the FDC3500 that the FDC4000 cannot start measurements if the QC card corresponding to the stacked slide's production lot has not been read NOTE: The "e" value recorded on the QC card represents the coefficient e x 1000. 4.10 Calibration The Fuji Dri-Chem System requires calibration for special tests. At present, this function is used for the CRP slide. 4.10.1 Calibration for CRP Slide Calibration is based on the standard curve, the assigned values of the calibrator CP-1, 2, 3, and dODr, which are obtained by measuring the spotted CRP slides with CP-1, 2, 3. The standard curve has been stored by a prior reading of the QC card, and the assigned values of CP-1, 2, 3 are keyed in during the calibration mode. The values of dODr are obtained from the measurement values of the CP. STEP1: The concentration values of CRP (CRP') are calculated from dODr and the standard curve. This is measurement value before calibration. STEP2: The calibration coefficients are calculated by following a linear equation so that the measured CRP concentration values are compensated to the values which equal the assigned values of CP1, 2, 3. Calibrated CRP value = g x ( CRP’) + f (4.10.1.1) The coefficients (g, f) can be browsed using Mode 61. FDC4000 Service Manual 4-11 4 4 Functions 4 4-12 FDC4000 Service Manual FDC4000 Service Manual AC100-240V Amplifier Potentiometer Power supply Amplifier Photometer Magnetic card reader Printer AD converter 16bit AD converter 16bit Interface CPU Hitachi SH7051 Clock: 20MHz Slide reader CPU Hitachi SH7055 Clock: 40MHz Driver circuit FROM2 SRAM (Backup) FROM1 Temperature controller Interference filter Slide transfer Sampler External interface ISE unit 5.1 VFD Keyboard Sample surface sensor Photometer 5 Control Units 5 Control Units Block Diagram 5 5-1 5 Control Units SOL01 MSA01 RAC40 board Sampler cover SW FANA01 PH803 Sampler cover lock solenoid Sampler cover LED40C lock LED PHC02 Sampler cover lock sensor Disposal box sensor K Na Cl EPM board DRV2 CN1 CPU15 SV1 SV1 V301 V301J CN7 CN7A Syringe HP CN4 IFTH3J Sampler rotation motor IFTH3P IFTH2P IFTH2J FG Sampler up/down HP IFHT1 Sampler rotation HP Sampler spot sensor PH202 CN1 TB2 LHA-AMP TB1 board LAMP HTIF board CNJ6 CNJ6A PH201 CN6A CN6 PH101 M101 Tip detect PD1,2 LAMP PH501 M501 TH401 H402 TH402 PH401 M401 PH103 PH102 PH801 VFD KEY40 board CPU5 CPU16 CPU10 CPU8 CPU11 CN5 CN3 PC6 PC6 PHA01 CN-J3 CN-P3 CN6 PH203 FIF board Sampler up/down motor M202 CN2 CN2A CF card CPU26 CPU4 DRV8 PH301 Printer CN1 CN1A DRV1 CPU2 CPU19 FG TISE TISE CPU9 CPU6 DRV5 DRV10 CPU40 board Syringe motor Solenoid valve V301 CPU3 CPU14 CPU7 M301 Power supply H601 TB7 ISE probes TB3 TB4 TB5 TB6 Keyboard sensor CPU12 Probe HP TB2 DRV4 CN2 CN1 Exhaust fan CN3 USB40 board FANA01 FANA01 CCD40 board DRV11 DRV9 Magnetic card reader Sample barcode reader (option) RS232C_I/F CN1 CN2 S1 CCD3 DRV6 ISE heater ISE temp. sensor TH601 CPU18 CPU13 CPU17 CPU29 CPU30 DRV40 board CPU1 DRV12 5 USB_I/F LED40 CCD2 CCD1 DRV7 PH601 DRV3 Slide transfer motor Slide feed HP Slide feed insert sensor Slide feed spot sensor Incubator motor Incubator HP Thermostat Incubator heater CM incubator temp. sensor Filter motor Filter HP Light source Photometer FDC4000 Service Manual 5-2 Connection Diagram 5.2 D-sub25 M601 Probe motor PRES1 board Syringe pressure sensor M201 5 Control Units 5.3 5.3.1 Circuit Board Functions CPU40 Circuit Board This board has a CPU and ROM which control the analyzer. The CPU calculates an analyte concentration value or enzyme activity using reflected optical density from the photometer and voltage from the potentiometer. The clock IC and memory IC are backed up by a lithium battery. The backup term during the power-off is at least 2 months. 5.3.2 DRV40 Circuit Board This board has pulse motor drivers to drive motors and other parts by the CPU40 board's signals. 5.3.3 CCD40 Circuit Board This board has a CCD device to take an image of the slide lot code printed on the slide, and sends the data to the CPU40 board. 5.3.4 LHA-AMP Circuit Board This board is a preamplifier for the photo-diode of the photometer. The output current of the photodiode is converted to a voltage signal and transferred to the CPU40 board. 5.3.5 5 PRES1 Circuit Board This board has a pressure sensor to obtain the pressure data and generate the sample surface detecting signal for the sampler unit. Variable resisters are mounted to compensate the offset and the gain of the pressure sensor and have been adjusted before shipping. 5.3.6 KEY40 Circuit Board This board has key switches for the operation to send the signals to the CPU40 board. Also, the board has a role of the signal interface between the display and the CPU40 board. 5.3.7 EPM Circuit Board This board is connected to the probes of the potentiometer. The measured voltage is converted to digital data to be transferred to the CPU40 board. 5.3.9 5.3.8 RAC40 Circuit Board This board has 3 photo-interrupters to detect the sample rack type. 5.3.9 USB40 Circuit Board This board is a serial interface board to generate signal of RS-232C or USB2.0. FDC4000 Service Manual 5-3 5 Control Units 5.4 Sensor List <ON/OFF Sensors> No. 5 Name Code 1 Slide feed HP PH101 2 Slide feed insert PH102 3 Slide feed spot PH103 4 Sampler rotation HP PH201 5 Sampler spot PH202 6 Sampler up/down HP PH203 7 Syringe PH301 8 Incubator HP PH401 9 Filter HP PH501 10 Probe HP PH601 12 Tip detect PH801 13 Disposal box PH803 14 Rack PH804 15 Keyboard PHA01 16 Cover lock PHC02 17 Sampler cover SW MSA01 Remarks <Analog Sensors> No. 5-4 Name Code Remarks 1 CM incubator temperature TH401 2 ISE incubator temperature TH601 3 Environmental temperature On the KEY40 board 4 Sample detect On the PRES1 board FDC4000 Service Manual 5 Control Units 5.5 Motor List Name Code Distance per pulse (mm/pulse) 1 Slide Transfer M101 0.1118 - 2 Sampler rotation M201 0.1254 0.0625 3 Sampler up/down M202 0.015 - 4 Syringe M301 0.0317 - 5 Incubator M401 0.047 0.125 On the photometer head 6 Interference filter M501 0.408 0.625 On the center of the filter 7 Probe M601 - 0.125 No. 5.6 Remarks On the circular arc of sampler nozzle Others Name No. 1 Rotation (degree/ pulse) Fan (exhaust) Code Remarks FANA01 Upper left side * Not in use * This fan is not equipped for model year 2009 or later. 2 * Fan (incubator cooler) FAN901 3 Magnetic card reader GR801 4 CM incubator heater H402 5 ISE incubator heater H601 6 Thermostat TH402 CM incubator (55 °C OFF) 7 Solenoid valve V301 Sample/Diluent nozzle switch 8 Light source lamp LP501 9 Cover lock solenoid SOLC01 10 Cover lock LED LEDC01 FDC4000 Service Manual 5 5-5 5 Control Units 5 5-6 FDC4000 Service Manual 6 Software 6 Software 6.1 Software Version V1.0 P01 Minor change and correction Major change and function additioin 6.2 6.2.1 DIP Switch Functions CPU40 Board For the FDC4000, the DIP switch settings are defined as below. When replacing the CPU40 board, the S4 of the spare part must be set as below. S4 <Type i (ISE unit equipped)> 1 CPU40 OFF 8 S4 1 㪦㪝㪝 OFF 4 ON ON Only No.5 =>ON <For Vet use (Type i)> S2 㪦㪝㪝 㪦㪝㪝 S2 6 No.5, 6, 7 =>ON <S4 functions> DIP SW No. Function (when SW is ON) 1 2 3 (Fixed to OFF) (Fixed to OFF) Lamp & temperature ready skip mode PF unit equipped ISE unit equipped For Vet use Spotting height 0.5mm down (Fixed to OFF) 4 5 6 7 8 FDC4000 Service Manual Remarks Factory use Factory use Memory area for reference intervals Factory use 6-1 6 Software 6.3 Specific Functions of Power-on By turning the power switch on with specific keys continuously pressed on at the same time, the analyzer starts up a specific mode listed below. NOTE: Press the keys continuously on until [Data storing] is displayed after the power switch on. Specific key MODE + CAL Functions Starts up the adjustment software. 6 6-2 FDC4000 Service Manual 6 Software 6.4 Data Files Data file name Description Remarks dc4_prm.dat dc4_std.dat Specific parameters Analyzer serial number, Adjustment data, etc. dc4_set.dat Setting information Setting information for mode operation, etc. dc4_di.dat DI information dc4_ab.dat Correlation coefficients dc4_rang.dat Reference intervals dc4_col.dat QC information for CM tests dc4_imm.dat QC information for immunochemical tests dc4_ise.dat QC information for ISE tests dc4_pft.dat PF information dc4_fdb.dat Time course dc4_prn.dat Reprint data dc4_info.dat Real time information dc4_cal.dat Time course of calibration dc4_err.log Error logs dc4_evnt.log Event logs dc4_data.log Data logs anml.mes Reference interval names prn_err.mes Error messages (Printer) lcd_err.mes Error messages (Display) prn_mode.mes Messages for mode operation (Printer) lcd_mode.mes Messages for mode operation (Display) prn_data.mes Messages (Printer) lcd_data.mes Messages (Display) (a, b) PF control information, PF calibration coefficients Lamp's cumulative illumination time, etc. 6 NOTE: “dc4_jklm.dat” is not used for V1.2-P01 or later. FDC4000 Service Manual 6-3 6 Software 6.5 Data Collection To export data files to a CF card, use Mode 123. NOTE: Refer to Sec. 6.13.7 for Mode 123. 6.6 Version Upgrade To upgrade the analyzer’s software, perform the following procedure: (1) Insert a CF card for version upgrade into the slot while the power is turned off. NOTE: Pay attention to the insert direction of the CF card. (2) Turn the power switch on. The upgrade procedure will start automatically. NOTE: Do not turn the power off during the upgrade processing. (3) After the upgrade process is completed, pull the CF card out from the slot. (4) Turn the power switch off and on again to check that the analyzer starts up successfully. NOTE: Check the printed software version. 6 NOTE: Both upgrade and downgrade can be performed. * In case that the version stored in the analyzer is different from the version of the CF card, the analyzer performs version upgrade (or downgrade) process. 6-4 FDC4000 Service Manual 6 Software 6.7 6.7.1 Calculation Process Colorimetric Tests The analyzer performs calculation to obtain test results according to the following order. X Concentration value OD OD value (1) OD calculation Performs the OD calculation using the reading values of the photometer. Refer to Section 4.8.1 for the calculation of optical reflection density. (2) OD calibration Performs the OD calibration using (i, j, k). OD’ = i x OD2 + j x OD + k (3) Calculating analyte concentration using standard curve Refer to Section 4.8.2 for the calculation of analyte concentration. (4) Compensation for slide production lot Compensates using the lot compensation coefficients (c, d, e). X' = c + d x X + e x X2 (5) Dilution factor calculation 6 Multiplies the data (X) by the dilution factor. X' = X x Dilution factor (6) Concentration calibration Performs the concentration calibration using (l, n). X' = X x l + X2 x n (7) PF compensation Performs the compensation using the PF coefficients (pf_a, pf_b). X' = X x pfa + pfb (8) Correlation coefficients compensation Performs the compensation using the correlation coefficients (a, b). X' = (X - b) / a (9) Determination range check Data are checked by the determination range which has already be multiplied by the dilution factor and adjusted in proportion to the coefficients (a, b). FDC4000 Service Manual 6-5 6 Software 6.7.2 ISE Tests The analyzer performs calculation to obtain test results according to the following order. (1) Voltage calculation Refer to Section 4.7.2 for the calculation of analyte concentration. (2) Compensation for slide production lot Compensates using the lot compensation coefficients (c, d). X' = c + X x d (3) PF compensation Performs the compensation using the PF coefficients (pf_a, pf_b). X' = X x pfa + pfb (4) Correlation coefficients compensation Performs the compensation using the correlation coefficients (a, b). Reference intervals are adjusted in proportion to the coefficients (a, b) and the dilution factors. X' = (X - b) / a 6 6-6 FDC4000 Service Manual 6 Software 6.8 Communication Interface Communication interface has 3 types as following. The FDC4000 has 2 new types of protocols on top of the FDC3500's protocol (Type 3): one is 2-way communication requesting test information (Type 1), another is 1-way communication using the Type 1 protocol (Type 2). All protocols are able to use a RS-232C connector (COM1A, 9pins) or a USB connector (COM1B) alternatively. (1) 2-way communication (requesting test information) [Type 1] Type 1 is a newly added protocol and is able to request test information (patient names, test names, etc.) from the FDC4000 to PC. Type 1 sends test results, but the protocol is different from the FDC3500's one-way protocol (Type 3). When using this protocol, select [Type 1] in Mode 46. (2) One-way communication (without requesting test information) [Type 2] Type 2 is a one-way transmission of test results and its protocol is same as Type 1. When using this protocol, select [Type 2] in Mode 46. (3) One-way communication [Type 3] Type 3 is one-way transmission as same as the protocol of the FDC3500. When using this protocol, select [Type 3] in Mode 46. 6 FDC4000 Service Manual 6-7 6 Software 6.9 6.9.1 2-way communication (requesting test information) [Type 1] About This Specifications The specification describes the system configuration, the operation procedures, and the process flow between the FDC analyzer and PC. 6.9.2 Applicability This specification is applied to the communication procedures between PC and the FDC analyzer. 6.9.3 Purpose The current models in the market identify each patient and sample by using its coded information. When using the current model, the analyzer’s operators input an ID defined by PC in advance. So, the operators have to take care of sample treatments to prevent mistakes. The new system defined by this document requests PC for the test information to enable the operator to check the information (sample ID, patient name, test menu, etc.) on the display or the printout of the analyzer, so as to reduce the risk of the sample treatment mistakes, and the operational cost by cutting the work sheet procedures. 6.9.4 System Configuration The system configuration is shown as below. FDC Analyzer PC task Built-in printer 6 Worklist USB LIS PC Test results Error codes Worklist Error explanation/action Onboard help Printer Hematology Analyzer Analyzer 6-8 FDC4000 Service Manual 6 Software 6.9.5 Interface Specifications We will use an USB-UART bridge chip to establish USB interface with PC. The following IC will be used: Maker: Silicon Laboratories Model: CP2102 Home Page: http://www.silabs.com/tgwWebApp/public/index.htm The maker’s Web sites are providing the virtual com port drivers. By using this driver, the application software of PC can be developed as same method of Windows standard com port access. (1) Communication Interface USB 2.0 Full Speed Compatible (2) Control Specifications (a) Time-out monitoring for receiving reply Condition: Five seconds have passed without receiving the reply after sending the request command. Action: The analyzer clears the request and waits the next event (request). (b) Re-transmission No re-transmission supported 6 FDC4000 Service Manual 6-9 6 Software 6.9.6 Communication Data Format (1) Communication Data System (a) Text format [STX (02H)], Command type, Parameter1, parameter2, ・・・ [ETX (03H)] [BCC] The parameters are added as required. The " , ” (breakpoint) is inserted between a command type and a parameter and between parameters. To connect some blocks, [ETB (17H)] is added at last. Ex.) STX Command type , Parameter 1 , Parameter 2 ETX BCC Command type , Parameter 1-1 , Parameter 1-2 ETB Ex.) STX Parameter 2-1 , Parameter 2-2 ETX BCC (b) BCC The calculation range for BCC is from the next of STX (command type) to ETX. The calculation method is exclusive OR. Ex.) STX Command type Parameter 1 , 6 , Parameter 2 ETX BCC Calculation range (2) Communication Data (a) Command type list Request / Info (Analyzer=>PC) Command type W 6-10 Description Request for sample info I Request for Worklist index S Test start info R Test results info E Error info Reply (PC=>Analyzer) Command type W I Description Reply for sample info Reply for Worklist index FDC4000 Service Manual 6 Software (b) Definition Name Command type Sample No.*1 Patient ID Description Given by one alphabetical character Assigned sample No. given by PC Unique ID for a patient (e.g patient’s file No.) Patient name Patient name Sample type W or P or U Number of test Test name Worklist index Species Number of tests for a sample Requested test name A set of a sample No., a patient ID, a patient name and others Unique No. assigned for species Reception No. No. for a patient (unique No. within a working day) *1 Sample No.: When PC assigns a number, the format is “mmdd**” [“Month” + “Day” + “Number (starting 01)”]. (Ex. 122001) In the off-line mode, the format is “**” [Number (starting 01)] without month and day 6 FDC4000 Service Manual 6-11 6 Software 6.9.7 Communication Data Details (1) Request for Worklist index This is used when the analyzer requests index information of the Worklist from PC. Item Character Size Description Header STX 1 STX: 02h Command type “I” 1 Request command for Worklist index (“ I ”: 49h) Breakpoint ‘,’ 1 Sample No. Numerical 13 Breakpoint ‘,’ 1 Number of requests Numerical 2 Maximum number of indexes to be requested(1 to 99) Delimiter EXT 1 ETX: 03h Max. 13 characters BCC NOTE: - Sample No. is the first key to search the Worklist. Blank data is allowable. - The search direction is forward direction of the Worklist. But, the Worklist data which has already been informed to start the measurements will be searched last. - The search results will be sent with the reply for the request (Worklist index). <Ex. 1> When requesting index information of 3 samples starting from sample ID “061201”: (Transmit data) 1 6 1 2 3 4 5 6 7 8 9 10 11 12 STX I , 0 6 1 2 0 1 , 3 ETX 13 14 15 = Space <Ex. 2> When requesting index information of 3 samples without assigning a starting ID for search: (Transmit data) 1 1 2 3 4 5 6 STX I , , 3 ETX 7 8 9 10 11 12 = Space 6-12 FDC4000 Service Manual 13 14 15 6 Software (2) Reply for request (Worklist index) This is used when PC sends index information of the Worklist to the analyzer. Item Character Size Description Header STX 1 STX: 02h Command type ‘I’ 1 Request command for Worklist index (‘ I ‘: 49h) Breakpoint ‘,’ 1 Number of indexes Numerical 2 Breakpoint ‘,’ 1 Sample No. 1 Numerical 13 Breakpoint ‘,’ 1 Patient ID 1 Alphabetical & numerical 13 Breakpoint ‘,’ 1 Patient name 1 Alphabetical & numerical 13 Breakpoint ‘,’ 1 Species 1 Numerical 2 Breakpoint ‘,’ 1 Sex 1 Numerical 1 Breakpoint ‘,’ 1 Age 1 Numerical 3 Block breakpoint ETB 1 Sample No. n Numerical 13 Breakpoint 7,’ 1 Patient ID n Alphabetical & numerical 13 Breakpoint ‘,’ 1 Patient name n Alphabetical & numerical 13 Breakpoint ‘,’ 1 Species n Numerical 2 Breakpoint ‘,’ 1 Sex n Numerical 1 Breakpoint ‘,’ 1 Age n Numerical 3 [999: Defined in (6)] Delimiter ETX 1 ETX: 03h Number of following indexes Max. 13 characters Max. 13 characters Max. 13 characters 0 to 99 0: Male 1: Female [9: Defined in (6)] [999: Defined in (6)] 6 ... Max. 13 characters Max. 13 characters Max. 13 characters 0 to 99 0: Male 1: Female [9: Defined in (6)] BCC FDC4000 Service Manual 6-13 6 Software NOTE: - Number of indexes is essential. - When no data in the Worklist, the “Number of indexes” should be 0 (zero). However, the sample No. send with the “Request for Worklist index” should be inserted into the data of the “Sample No. 1”. - When the Worklist has test orders, the sample No. is essential, and either the patient ID or the patient name is essential in the Worklist data. - In case that the sample No. is blank, or both the patient ID and the patient name are blank (even if the sample No. has data), the field is invalid. - If a number of index fields is larger than the “Number of indexes”, overflowed fields (larger than the number of indexes) are ignored. - If a number of index fields is smaller than the “Number of indexes”, the process will continue without error. - A block breakpoint (ETB) must NOT be added at the end of the last index field. <Ex.> When replying with 2 samples data: 1st sample info: 2nd sample info: 6 Sample ID 2006061201 Patient ID ABCDEFGHIJKLM Patient name Taro Fuji Species 2: Feline Sex 1: Female Age 3 years Sample ID 2006061202 Patient ID 12345ABCD Patient name Lucy Heska Species 1: Canine Sex 0: Male Age 1 year (Transmit data) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 1 STX I , 2 , 2 0 0 6 0 6 1 2 0 1 2 , A B C D E F G H I J K L M , 3 T a r o F u j i , 2 , 1 , 3 4 ETB 2 0 0 6 0 6 1 2 0 2 , 1 2 3 5 4 5 A B C D , L u c y H e s 6 k a , 1 , 0 , 1 ETX BCC = Space 6-14 FDC4000 Service Manual 6 Software (3) Request for sample information This is used when the analyzer requests sample information of the Worklist from PC. Item Character Size Description Header STX 1 STX: 02h Command type 'W' 1 Request command for sample info ('W': 57h) Breakpoint ',' 1 Sample No. Numerical 13 Breakpoint ',' 1 Patient ID Alphabetical & numerical 13 Breakpoint ',' 1 Patient name Alphabetical & numerical 13 Max. 13 characters Delimiter ETX 1 ETX: 03h Max. 13 characters Max. 13 characters BCC NOTE: - Sample No., patient ID, and patient name are used for search keys. At least one of those information is required. - Priority for the search keys is as follows: sample No. (1st), patient ID (2nd), patient name (last). - The character data length for sample No., patient ID, and patient name is max. 13 characters. If the data length is under 13, it is NOT necessary to fill the data area with "space". <Ex.> When requesting sample information for the 2nd sample: 2nd sample info: Sample ID 2006061202 Patient ID 12345ABCD Patient name Lucy Heska (Transmit data) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 1 STX W , 2 0 0 6 0 6 1 2 0 2 , 1 2 2 3 4 5 A B C D , L u c y 3 e s k a H ETX BCC = Space FDC4000 Service Manual 6-15 6 6 Software (4) Reply for request (sample info) This is used when PC sends sample information of the Worklist to the analyzer. Item Character Size Description Header STX 1 STX: 02h Command type 'W' 1 Request command for sample info ('W': 57h) Breakpoint ',' 1 Sample No. Numerical 13 Breakpoint ',' 1 Patient ID Alphabetical & numerical 13 Breakpoint ',' 1 Patient name Alphabetical & numerical 13 Breakpoint ',' 1 Number of tests Numerical 2 Breakpoint 1 ',' 1 Test name 1 Alphabetical 8 Breakpoint n ',' 1 Test name n Alphabetical 8 Max. 8 characters Delimiter ETX 1 ETX: 03h Max. 13 characters Max. 13 characters Max. 13 characters 0 to 30 slides Max. 8 characters ... BCC 6 NOTE: - Sample No. and number of tests are essential. - If the "Number of tests" is "0", it means the Worklist does not have test orders. - Test name field has maximum of 20 fields. - If a number of test name fields is larger than the "Number of tests", overflowed fields (larger than the number of tests) are ignored. - If a number of test name fields is smaller than the "Number of tests", the process will continue without error. - <Processing of received data> The received data ("Number of tests" and/or "Test name") is used as printout data for the built-in printer, and then will be deleted after the printout completed. To re-print out the information, perform from the beginning of the process ("Request for Worklist index"). Also, the received data will not be used other purposes; e.g. for checking test results matching with the received data. 6-16 FDC4000 Service Manual 6 Software <Ex. 1> When replying for sample information of the 2nd sample: Sample info: Sample ID 2006061202 Patient ID 12345ABCD Patient name Lucy Heska Number of tests 4 tests Test names BUN, CRE, GLU, ALP (Transmit data) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 1 STX W , 2 0 0 6 0 6 1 2 0 2 , 1 2 2 3 4 5 A B C D , L u c y 3 e s k a , 4 , B U N , C R 4 G L U , A L P H E , ETX BCC = Space <Ex. 2> When replying for sample information of the 1st sample: Sample info: Sample ID 2006061201 Patient ID ABCDEFGHIJKLM Patient name Taro Fuji Number of tests 1 panel Test names Panel A 6 = Space FDC4000 Service Manual 6-17 6 Software (5) Test start information This is used when the analyzer sends test start information to PC. Item Character Size Description Header 02h 1 STX Command type 'S' 1 Test start information command ('S': 53h) Breakpoint ',' 1 Selecting control measurements or normal Test condition Alphabetical measurements 7 < CONTROL or NORMAL : Space (20h)> Breakpoint ',' 1 Test date Numerical 10 Breakpoint ',' 1 Test time Numerical 5 Breakpoint ',' 1 Sample No. Numerical 13 Breakpoint ',' 1 Patient ID Alphabetical & numerical 13 Breakpoint ',' 1 Patient name Alphabetical & numerical 13 Breakpoint ',' 1 Sample position Numerical 2 Sample position on the sample disk Delimiter 03h 1 ETX 6 Year, month, and day when the measurements were performed. (Ex. 2006-09-25) Time when the measurements were performed. (Ex. 10:50) Unique ID for each sample Unique ID for each patient Patient name BCC - Each data area is a fixed-length field, so that data should be embedded from the left and spaces (20h) should be embedded in the remaining spaces. - The sample position should be fixed as "01". <Ex.> When the test for the 1st sample starts on June 12, 2006 at 10:50: Sample info: Sample ID 2006061201 Patient ID ABCDEFGHIJKLM Patient name Taro Fuji (Transmit data) 1 2 3 4 5 6 7 8 9 1 STX S , N O R M A L 2 - 0 6 - 1 2 , 1 0 3 0 6 0 6 1 2 0 1 4 D E F G H I J K F u j i 5 L 10 : 11 12 13 14 15 , 2 0 0 6 5 0 , 2 0 , A B C a r o M , T , 0 1 ETX BCC = Space 6-18 FDC4000 Service Manual 6 Software (6) Test results information (For two-way & one-way communication) This is used when the analyzer sends test results to PC. Item Character Size Header 02h 1 STX Command type 'R' 1 Test results inform command ('R': 52h) Breakpoint ',' 1 Test condition Alphabetical 7 Breakpoint ',' 1 Test date Numerical 10 Breakpoint ',' 1 Test time Numerical 5 Breakpoint ',' 1 Sample No. Numerical 13 Breakpoint 1 Breakpoint ',' Alphabetical & numerical ',' Alphabetical & numerical ',' Patient ID Breakpoint Patient name 13 13 Numerical 2 ',' 1 Sex Numerical 1 Breakpoint ',' 1 Age Numerical 3 Breakpoint ',' 1 Sample position Numerical 2 Breakpoint ',' 1 Number of tests Numerical 2 Breakpoint ',' Alphabetical & numerical ',' 1 Equal sign Breakpoint 8 1 1 Breakpoint Dilution factor Numerical 2 Breakpoint Reference interval lower limit Breakpoint Reference interval upper limit Breakpoint ',' 1 Numerical 5 ',' 1 Numerical 5 Unit Warning Test name breakpoint Delimiter FDC4000 Service Manual Time when the measurements were performed.(Ex: 10:50) Unique ID for each sample Unique ID for each patient Patient name 00 - 99 0: Male 1: Female 9: Stand-alone mode 999: Stand-alone mode 6 Sample position on the sample disk 00 to 20 slides Test name (5 characters) + Sample type (2 characters) 1 ',' Alphabetical & numerical Alphabetical & numerical ',' Test result Year, month, and day when the measurements were performed. (Ex: 2006-09-25) 1 Species Breakpoint Selecting control measurements or normal measurements 1 Breakpoint Test name Description Equal sign or unequal signs 9 Test result (9 characters) 6 Unit (6 characters) 1 Used dilution factor (from 01) Test result field Setting of reference interval lower limit Setting of reference interval upper limit ',' 1 Alphabetical 11 As stated following ',' 1 Breakpoint 03h 1 ETX 6-19 6 Software NOTE: - The test result field has "Number of tests" of test results. - A test name breakpoint must NOT be added at the end of the last test result field. - Each data area is a fixed-length field, so that data should be embedded from the left and spaces (20h) should be embedded in the remaining spaces. - "Sex" and "Age" fields (See the Description of the "Sex" and "Age" (previous page).) - For 2-way communication, given data from PC [see (2)] will be embedded (copied). For stand-alone mode, "9" will be embedded in the Sex fields, "999" will be embedded in the Age field. Warning characters The below table indicates warning character allocation in the warning data area. When no error has occurred for each, a space (20h) will be embedded. Position (sending order) Warning character H Testing value exceeds the upper limit of the preset reference interval. L Testing value falls below the lower limit of the preset reference interval. 2nd @ Testing value is outside of the determination range. 3rd # The valid term of the slide has expired. $ Temperature control error (thermistor disconnection) + Temperature control error (Incubator temperature has exceeded the upper limit of the range.) - Temperature control error (Incubator temperature has fallen below the lower limit of the range.) * Photometer malfunction (white plate fluctuation) ? Photometer malfunction (out of the white plate range) 1st 4th 6 5th 6th 7th Always a space (20h) & 8th Abnormally high testing value Always a space (20h) 9th E Calculation error or malfunction of ISE test results 10th ¥ Un-spotted slide 11th 6-20 Description Always a space (20h) FDC4000 Service Manual 6 Software Transmission example <Printout example> 2006-06-12 10:50 ID:ABCDEFGHIJKLM Name: Taro Fuji GLU-PS @+*? ¥# E =75 mg/dl (10) (50.0 - 100.0) AMYL-PS H # > 1500 U/l (200 - 1500) <Transmit data> 1 2 3 4 5 6 7 8 9 1 STX R , N O R M A L 2 - 0 6 - 1 2 , 1 0 3 0 6 0 6 1 2 0 1 4 D E F G H I J K F u j i 5 6 3 , 0 1 , 7 S , = 7 5 8 g / d l , 1 9 1 0 0 . 0 , , A M Y 10 11 5 0 0 12 1 , 5 13 # 0 0 L 0 L 2 10 : 11 12 13 14 15 , 2 0 0 6 5 0 , 2 0 , A B C T a r o , 1 , U - P M , , 2 , G L m 0 , 5 0 @ # + * - P S u / l , 1 5 . , E , 0 0 0 > , 1 , 0 , H ETX BCC = Space FDC4000 Service Manual 6-21 6 6 Software (7) Error information This is used when the analyzer sends error information to PC. Item Character Size Description Header 02h 1 STX Command type 'E' 1 Error information command Breakpoint ',' 1 Error event date Numerical 10 Breakpoint ',' 1 Error event time Numerical 8 Breakpoint ',' 1 Error number Alphabetical & numerical 5 Breakpoint ',' 1 Number of added info Numerical 1 Breakpoint ',' 1 Added info 1 Alphabetical & numerical 6 Breakpoint ',' 1 Added info n Alphabetical & numerical 6 Measured voltage etc. Delimiter 03h 1 ETX Year, month, and day when the error occurred. (Ex: 2006-06-12) Time when the measurements were performed.(Ex: 10:30:50) Error number Number of added info (0 to 9) Added info field Measured voltage etc. ... 6 NOTE: - The added info field has "Number of added info" of data. - A breakpoint must NOT be added at the end of the last added info field. <Ex.> When suction clogging error (E0110) has occurred: [Printout example by the built-in printer] E0110 Clogging>sample>suc Clogging = 1.000 V Check sample (Transmit data) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 1 STX E , 2 0 0 6 - 0 6 - 1 2 , 1 2 0 : 3 0 : 5 0 , E 0 1 1 0 , 1 3 , 1 . 0 0 0 ETX BCC = Space 6-22 FDC4000 Service Manual 6 Software (8) Test results information (One-way communication) In the case of one-way communication, this function is used to send tests results from the analyzer to PC. The communication data format for the one-way communication is same as "(6) Test results information (For two-way & one-way communication)". 6 FDC4000 Service Manual 6-23 6 Software 6.9.8 Data Transmission Flow (1) Single sample test (for FDC4000) Operator Analyzer PC Set a sample Press the WORKLIST key Request for index infomation of the Worklist Worklist index Reply for the request (Worklist index) Browse Worklist Display Worklist (Top patient info) key Renew Worklist on the display Press the PRINT key Select the patient Request for sample information of the Worklist just displayed on Reply for the request (sample info) Test menu is printed out 6 Sample info Test menu information Load the slides Press the START key Select the sample just displayed on Test start information Start the tests Test processing info. displayed Test results printed Tests completed Transmit test results Store test results - Pressing the WORKLIST key requests Worklist index information. - Pressing the PRINT key selects the sample just displayed on the display to request sample information of Worklist. - When using a sample barcode reader, reading a barcode selects a patient ID to request sample information using the patient ID as a key. - Pressing the START key selects the patient's sample which is just displayed on the display to start the testing. 6-24 FDC4000 Service Manual 6 Software (2) Re-transmitting test results This function is used to re-transmit test results manually. (This is useful when electric power failure or cable disconnection caused the communication interruption.) Operator Analyzer PC Search re-transmit data Edit re-transmit data Transmit test results Store test results (3) Transmitting test results (One-way communication) Operator Analyzer PC Load slides Press the START key Tests completed Tests results printed Transmit test results Store test results 6 FDC4000 Service Manual 6-25 6 Software 6.10 One-way communication [Type 3 (FDC3500 protocol)] 6.10.1 Communication Parameters Data transfer rate 1200, 9600, 19200bps Selectable Communication type Serial interface Communication method Asynchronous communication method Character length ASCII 8 bits Stop bit 1 bit Parity No BCC No or Yes (EOR) [selectable] DSR check: No Hand shaking CTS/RTS control 6.10.2 Data Transmission Specification Transmission data is fixed-length and the unused data space is expressed as a space data (20H). NOTE: = Space data (20H) (1) Transmission timing All data of one sample is transmitted after the last test of-belonging to the sample has been completed. (2) Transmission interval 6 The transmission interval between the data and the previous data is at least 3 seconds. (3) Test conditions Test results of calibration are not transmitted. Test information whether in the normal or in the control mode is transmitted. (4) Transmission data format Transmission order: [STX] => [Test results] => [ETX] => [BCC] STX (02H) 1 character data length Test condition Test information whether in the normal or in the control mode 7 characters data length Test date Test date (Ex. 2014-07-25) 10 characters data length Test time Test time (Ex. 17:59) 5 characters data length Sequence number Specific No. for 1234567890123) (Ex. 13 characters data length Sample ID Specific ID for the sample (Ex. ABCDEFGHIJKLM) 13 characters data length Sample position on the sample disk data (01) for the FDC4000 2 characters data length Sample tion 6-26 posi- the sample Fixed FDC4000 Service Manual 6 Software Test name/ Sample type Total 7 characters data length (5 characters data length for a test name and 2 characters length for a sample type) Sample type: "-W" (Whole blood), "-P" (Plasma/serum), "-U" (Urine) (Ex.) 7 characters data length AMYL-P , CRP In case that characters are less than 7, spaces are embedded. In case of CRP, 4 spaces will be embedded at the end (no "-" mark). Equal sign or unequal sign Ex. = , <, or > 1 characters data length Test results [Test results (9 ch) + Unit (6 ch) + Dilution factor (2 ch) + Warning indications (11 ch) x number of test names Ex. 123456.78(+)mg/ 28 characters data length dl (+)02(+)L@#$*F&PE\T ETX (03H) 1 characters data length BCC Exclusive OR from STX to ETX 1 character data length (5) Warning characters on test results Warning character 1 L, H The testing value is out of the preset Reference interval. 2 @ Qualitative range 3 # The slide has expired. 4 $, +, - Temperature control error 5 *, ? Fluctuation of light source intensity 6 F PF pressure error 7 & Abnormal testing value 8 P Not in use (validity term of calibration has expired) 9 E Calculation error or malfunction of ISE test results 10 ¥ Un-spotted slide 11 FDC4000 Service Manual 6 Description Not in use 6-27 6 Software When ISE test results have errors (drift error, impedance error, outside of range), the warning "E" is added on the 9th warning position of the transmitted data (on K and CL). 0Q 0C25O'SN -25" O'SN %N25O'SN '44 6 6-28 FDC4000 Service Manual 6 Software (6) Transmission example 0Q ).725" (' OIFN #/;.7* OIFN The below transmission example is for the above test result. 1 2 3 4 5 6 7 STX N O R M A L 8 9 10 11 12 13 14 15 16 17 18 19 20 2 0 1 4 - 0 7 - 2 5 1 7 A B C D = 1 @ # + 0 0 : 5 9 1 2 3 4 5 6 7 8 9 0 1 2 3 E F G H I J K L M 0 1 G L U - P 2 3 4 5 m g / d l 1 5 * F A M Y L - U > 1 d l 0 2 H # E ¥ m g / 0 ETX BCC NOTE: 6 Space data (20H) FDC4000 Service Manual 6-29 6 Software 6.11 Communication Cable Connection 6.11.1 D-SUB 9 pins (COM1A) %& % %& % 4:& 4& 4:& 4& 6:& 5& 6:& 5& '4 &64 '4 &64 )0& 5) )0& 5) &54 &4 &54 &4 465 45 465 45 %65 %5 %65 %5 4+ 4+ (TCOG (TCOG (&%ޓ%1/㧔RKPU㧕 6 *QUV㧛2%㧔RKPU㧕 6.12 Sample Barcode Reader 6.12.1 Manufacturer, Model number Manufacturer: Opto Electronics Co., Ltd. Model number OPT-6125-RS (25P) 6.12.2 Data format 6-30 Data format CODE-39, NW-7, ITF 2 of 5, CODE-128, WPC Data character length 13 FDC4000 Service Manual 6 Software 6.13 Mode Description 6.13.1 Mode List Mode type definition U: User mode A: Administrator mode (password required) S: Service mode (password required) F: Factory use (password required) Described Mode number Mode type ○ 0 U Administrator mode 19 U Turns on the control mode (a, b canceled) 20 U Sets date and time 23 U Displays and resets the lamp's cumulative illumination time 24 A Unit conversion [Unit (A) / Unit (B) Switch] 25 U Data retransmission to host computer 26 U Reprints test results 27 A Sample No. and sample ID settings 28 A Switches display method for values outside of the determination range 29 U Prints out slide lot numbers 30 A Work list display item setting 35 A Edits sample No. and sample ID 36 A Correlation coefficients (a, b) settings and printout 37 A Lot compensation coefficients (c, d, e) settings and printout 39 A Reference interval settings and printout 40 A Spotting count setting 42 A Leak check 44 A Lamp off selection 45 A Dilution factor settings 46 A Selects communication destinations 49 U Prints out error logs 52 U Reference plate level check 55 A Selects language 61 A Inputs immuno-chemical calibration parameters 76 U Prints out DI card information 81 A Beep sound settings FDC4000 Service Manual Mode description 6 6-31 6 Software Mode number Mode type 82 A Display brightness and print density 83 A Test result print sheets setting 85 A Display order of reference intervals 86 A Edits and inputs reference Interval names 103 U Displays temperature and humidity ○ 107 S Deletes data files ○ 111 S Inputs parameters of RBP level check ○ 121 S Displays cumulative test count ○ 122 S Imports a selected data file from CF card ○ 123 S Exports all data files to CF card ○ 129 SS Described Mode description Inputs detection threshold for un-spotted slide 6.13.2 Mode 0 Changing Mode Type Refer to the Instruction manual for Mode 0 operation. Mode 6 6-32 Password Administrator mode 4000 Service mode 753674 FDC4000 Service Manual 6 Software 6.13.3 Mode 107 Deleting Data Files This mode is used to delete time course or reprint data files which are memorized in the analyzer. (1) Enter into Mode 107. MODE ENTER (2) Select a menu (data file) to be deleted. 5GNGEVKQPFKCNQI /&GNGVKPIHKNGU 6KOGEQWTUG㨯 /&GNGVKPIHKNGU 4GRTKPVFCVC /&GNGVKPIHKNGU 'TTQTNQIU㨯 5GNGEV Select a data file using the press the ENTER key. keys and (3) Select whether to delete the file or not. Selecting [NOT delete] returns to the selection dialog of (2). (4) When selecting [Delete]: The analyzer prints out the deleted file name and returns to the selection dialog of (2). ENTER 㧔'ZCORNG㧕9JGPUGNGEVKPI=6KOGEQWTUG? /6KOGEQWTUG &GNGVG016FGNGVG /6KOGEQWTUG &GNGVG016FGNGVG To quit the mode, press the STOP key. 5GNGEV 6 ENTER ENTER 6QUGNGEVKQPFKCNQI &GNGVKPIFCVCHKNGU 6KOGEQWTUG &GNGVGF 6QUGNGEVKQPFKCNQI FDC4000 Service Manual 6-33 6 Software 6.13.4 Mode 111 Inputting Parameters of RBP Level Check This mode is used to input a default value and a detection threshold of RBP level. (1) Enter into Mode 111. MODE ENTER The current values are displayed. %WTTGPVUGVVKPIU (2) Input a default level for the RBP. /4$2NGXGNRCTC +PKV6JTGU 㧔'ZCORNG㧕+PKVKCN &GHCWNV 6JGUJQNF ENTER (3) Input a threshold value for detecting a warning or an error of dirty RBP. /4$2NGXGNRCTC +PKV6JTGU ENTER (4) The new input data will be printed out. %JGEMVJGKPRWVXCNWGU +PKV6JTGU 6 /+PRWVFCVC1-! ;'501 /+PRWVFCVC1-! ;'501 5GNGEV ENTER ENTER (5) Make sure that the printed new values are correct. Select a menu (YES or NO) using the keys, and press the ENTER key. (a) When selecting [YES], the analyzer saves the new values and quits the mode operation. (b) Selecting [NO] returns to the selection dialog of (2). 4$2NGXGNRCTC %JCPIGF #HVGTUCXKPIFCVC 3WKVU/QFG 6-34 6QKPRWVFKCNQI FDC4000 Service Manual 6 Software 6.13.5 Mode 121 Displaying Cumulative Test Count This mode is used to display cumulative test count. (1) Enter into Mode 121. MODE ENTER (2) The cumulative test count is displayed. /6GUVEQWPVGT 6GUV%QWPV The cumulative test count will be displayed for 5 seconds. The analyzer quits the mode automatically. &KURNC[UHQTUGE ENTER 3WKVU/QFG 6 FDC4000 Service Manual 6-35 6 Software 6.13.6 Mode 122 Importing a Selected Data File from CF Card This mode is used to import a selected data file from a CF card. Selectable files to be imported are as follows: Data file name Description dc4_prm.dat dc4_std.dat Specific parameters dc4_set.dat Setting information dc4_di.dat DI information dc4_ab.dat Correlation coefficients dc4_rang.dat Reference intervals dc4_col.dat QC information for CM tests dc4_imm.dat QC information for immuno-chemical tests dc4_ise.dat QC information for ISE tests dc4_pft.dat PF information anml.mes Reference interval names (1) Enter into Mode 122. MODE 6 ENTER 5GNGEVKQPFKCNQI (2) Select a data file name to be imported. /+ORQTVKPIHKNG 5RGEKHKERCTC㨯 /+ORQTVKPIHKNG 5GVVKPIKPHQ /+ORQTVKPIHKNG &+KPHQ /+ORQTVKPIHKNG %QTTGNCVKQPEQGHH /+ORQTVKPIHKNG 4GHKPVGTXCNU /+ORQTVKPIHKNG %/3%KPHQ /+ORQTVKPIHKNG +OOWPQ3%KPHQ /+ORQTVKPIHKNG +5'3%KPHQ /+ORQTVKPIHKNG 2(KPHQ /+ORQTVKPIHKNG 4GHKPVGTXCNPCOGU Select a data file using the the ENTER key. 5GNGEV ENTER 6-36 FDC4000 Service Manual keys and press 6 Software (3) Import the data file. 㧔'ZCORNG㧕6QKORQTV ̌5GVVKPIKPHQ̍ Insert a CF card into the slot and press the START key. 5GVVKPIKPHQ 5GV%(ECTF 56#46 START 5GVVKPIKPHQ +ORQTVKPI +ORQTVKPIHKNG 5GVVKPIKPHQ The status of the import processing will be displayed. (4) After the import is completed, the analyzer prints out that the process has completed. (KPKUJ (5) The display returns to the selection dialog of (2). 6QUGNGEVKQPFKCNKI To quit the mode, press the STOP key. 6 FDC4000 Service Manual 6-37 6 Software 6.13.7 Mode 123 Exporting All Data Files to CF Card This mode is used to export data files to a CF card. Data files to be exported are as follows: Data file name 6 6-38 Description dc4_prm.dat dc4_std.dat Specific parameters dc4_set.dat Setting information dc4_ab.dat Correlation coefficients dc4_rang.dat Reference intervals dc4_di.dat DI information dc4_info.dat Real time information dc4_err.log Error logs dc4_col.dat QC information for CM tests dc4_imm.dat QC information for immuno-chemical tests dc4_ise.dat QC information for ISE tests dc4_pft.dat PF information anml.mes Reference interval names dc4_data.log Data logs dc4_evnt.log Event logs dc4_cal.dat Calibration data for immuno-chemical tests dc4_fdb.dat Time course dc4_prn.dat Reprint data FDC4000 Service Manual 6 Software (1) Enter into Mode 123. MODE ENTER (2) Select a menu to designate files to be exported. /'ZRQTVKPIHKNGU 6KOG2TO#NN /'ZRQTVKPIHKNGU 6KOG2TO#NN /'ZRQTVKPIHKNGU 6KOG2TO#NN 5GNGEV <Exported files for the menu> [Time] - dc4_fdb.dat [Prm] - dc4_di.dat - dc4_prm.dat - dc4_std.dat ENTER [All] => All files /'ZRQTVKPIHKNGU 5GV%(ECTF 56#46 (3) Insert a CF card into the slot and press the START key. START /'ZRQTVKPIHKNGU (KNGPCOG &KURNC[UHQTUGE /'ZRQTVKPIHKNGUޓޓ 2WNN%(ECTFQWV The status of the export processing will be displayed. 6 (4) After the left message appears, pull the CF card out from the slot. After [Pull CF card out] appears on the display for a few seconds, the analyzer quits the mode operation. 'ZRQTVKPIHKNGU #NN (KPKUJ 3WKVU/QFG FDC4000 Service Manual 6-39 6 Software 6.13.8 Mode 129 Inputting Detection Threshold for Un-spotted Slide This mode is used to input a threshold value for detecting un-spotted slide. (1) Enter into Mode 129. MODE ENTER (2) Input a threshold value for detecting un-spotted slide. /7PURQVVGFUNKFG 6JTGU After the new value is printed out, the analyzer quits the mode operation. 㧔'ZCORNG㧕+PRWV ENTER 7PURQVVGFUNKFG 6JTGU 3WKVU/QFG 6 6-40 FDC4000 Service Manual 7 Disassembly and Reassembly 7 Disassembly and Reassembly WARNING Before starting the work, be sure to shut off the power and unplug the power cable to avoid electric shock, burn, and equipment damage. Use due care during maintenance since equipment contains parts that are not fully discharged and those which are burning hot (such as a lamp house) even after power off. Exercise due care to avoid electric shock and other hazards if the power cannot be turned off. CAUTION The equipment includes parts and assemblies sensitive to electrostatic discharge damage. Use caution to prevent damage during all service procedures. WARNING To avoid biohazards during maintenance and repair work, Safety Control authorized by your site is required. The following are general cautionary measures: - You must always take care to prevent infection. - Always follow biohazard procedures (e.g., wearing gloves, lab coat, and safety goggles) during the work. - Before starting the work, disinfect all possible places that have come into contact with blood sample. - For disinfection, use ethyl alchol for disinfection. (Only for disposal box cleaning, 0.5% of sodium hypochlorite acid disinfectant can be used.) - When blood adheres, remove the blood stain with a cotton swab moistened with water. For example, Sample rack, Sampler nozzle, Spotting parts, Transfer unit, Incubator, Disposal box, etc. WARNING As used (contaminated) consumables (e.g., FUJI DRI-CHEM SLIDEs, FUJI DRI-CHEM AUTO TIP, FUJI DRI-CHEM MIXING CUPS S and sample tubes) and contaminated swabs or cloths used for cleaning the equipment are infectious waste, process the waste correctly in compliance with any applicable regulations in your country, such as by incineration, melting, sterilization or disinfection. WARNING Do not touch used consumables (slides and tips) and contaminated swabs or cloths used for cleaning the equipment with bare hands as this may cause contamination. If any part of the body comes in contact with contaminated consumables, immediately rinse the contaminated body part thoroughly under running water and then use ethyl alcohol as a disinfectant. Seek medical assistance if necessary. CAUTION Refer to each section of the manual for information about usable solvents for cleaning. Do not use cleaning solution containing ammonia on or near the analyzer. FDC4000 Service Manual 7-1 7 7 Disassembly and Reassembly CAUTION Do not use alcohol for cleaning the sampler cover (translucence), otherwise the surface will be damaged. IMPORTANT Maintenance or repair may affect the accuracy of test results. Perform quality control (running control fluids) to ensure the reliability of the test results. Refer to Section 6 of the Instruction Manual for details. NOTE: Due to the design change, the contents of “Disassembly and Reassembly” have been revised in December 2008. The design change has been applied to the machines which serial numbers are over 3000. In this manual, the machines before the design change are described as “model year up to 2008”; the machines after the design change are described as “model year 2009 or later”. - Model year up to 2008 = (Example) Serial No. 42110001 or later - Model year 2009 or later = (Example) Serial No. 42113001 or later 7 7-2 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1 7.1.1 Disassembling and Reassembling CPU40 Board (1) In case that the mode operations are available, export all data files memorized in the analyzer using Mode 123. (Refer to Sec. 6.13.7) (2) Loosen 6 screws to remove the rear panel. (3) Unplug all connectors from the CPU40 board. Remove 8 screws. (4) For model year 2009 or later, lower the CPU40 board once and then take it out toward you to remove. NOTE: For mode year 2009 or later, a bracket is equipped to prevent dropping a CF card, so that it is necessary to lower the CPU40 board once before removing. 7 <Reassemble in the reverse order> NOTE: For the following sections, reassemble in the reverse order. (5) For a new CPU40 board (spare part), perform the following: (as the old board was.) - Set SW4. => Refer to Sec. 6.2.1. - Check SW2. => Only No.1 is ON (Refer to Sec. 6.2.1.) - Set SW3 to ON. (battery backup). FDC4000 Service Manual 7-3 7 Disassembly and Reassembly (6) After reassembling, import the all data files using Mode 122. IMPORTANT If the import process is not performed, the specific parameters (including adjustment data and coefficients (a, b), etc.) are not input into the analyzer. This will cause analyzer malfunction or wrong test results. IMPORTANT After the import process is completed, check that the data are properly imported using Mode operations. For example, use Mode 36 for checking (a, b) values. (7) Run the control fluids to check the performance. 7 7-4 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.2 USB40 Board <For model year up to 2008> NOTE: For model year 2009 or later, see the next section. (1) Unplug the connector (CPU4) from the CPU40 board. (2) Remove 2 screws. (3) Remove the bracket with the USB40 board from the body. (4) Remove 2 hexagon head screws to release the connector (COM1A, 9 pins). 7 (5) Unplug the connector (USB40). (6) Loosen 4 screws to remove the USB40 board. FDC4000 Service Manual 7-5 7 Disassembly and Reassembly 7.1.3 USB40 Board <For model year 2009 or later> NOTE: For model year up to 2008, see the previous section. (1) Remove the CPU40 board beforehand. (2) Remove 2 hexagon head screws to release the connector (COM1A, 9 pins). (3) Unplug the connector (USB40). (4) Loosen 4 screws to remove the USB40 board. 7 7-6 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.4 Operation Side Covers (1) Pull the disposal box out and remove the screw. (2) Remove 2 screws. (3) Lift the upper cover a little up to release the 2 hooks, and remove the front-side cover. NOTE: When assembling the cover, hook the 3 points (under part) at first, and then hook the 2 points (upper part). 7 FDC4000 Service Manual 7-7 7 Disassembly and Reassembly (4) Loosen 3 screws to remove the cover. (5) Remove 4 screws. (6) Remove the cover with turning the sampler. 7 (7) Remove the upper cover. 7-8 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.5 Left and Right Covers (1) Remove the screw of the right-side cover. (2) Remove 2 screws. (3) Pull the cover to remove. (4) Remove the screw of the left-side cover. 7 (5) Remove 2 screws. (6) Pull the cover to remove. FDC4000 Service Manual 7-9 7 Disassembly and Reassembly 7.1.6 Top Support Cover (1) Remove 2 screws back of the sampler. (2) Lift the top support cover a little up to release 2 hooks, and pull it down through the inner space of the sampler cover. 7 7-10 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.7 Sampler Nozzle (1) Remove the sampler arm cover. (2) Pull 2 tubes out from the nozzles. (3) Loosen 2 screws to remove the bracket. NOTE: When reassembling, the tube (for sample) with an O-ring must be connected to the backside. 7 (4) Lift the nozzle up to remove. NOTE: After reassembling, perform the leak check using Mode 42. FDC4000 Service Manual 7-11 7 Disassembly and Reassembly 7.1.8 Sampler Tube (1) Remove the rear cover of the sampler in advance. (2) Take the solenoid body from the original position a little and pull 2 tubes out from the solenoid. NOTE: When reassembling, the tube (for sample) with an O-ring must be connected to the original joint. - For model year up to 2008, connect to the under side (right lower in the photo). (The solenoid body is cylindrical style.) - For mode year 2009 or later, connect to the backside. (The solenoid body is rectangular solid style.) (3) Remove the tube from the back of the sampler. 7 (4) Open the clamp to the right to remove the tube. (5) Remove the sampler arm cover, and pull the tubes out from the nozzles. NOTE: Pull out the rubber joints of the tube's top in advance. (6) Pull the tube on the sampler arm out to the right. NOTE: Before removing the tube, confirm the present tube routing for reassembling. NOTE: When reassembling, the tube (for sample) with an O-ring must be connected to the backside. 7-12 FDC4000 Service Manual 7 Disassembly and Reassembly (7) When reassembling a new tube, pass the tube through a hole from the bottom of the sampler arm, and plug it to the sampler nozzle. NOTE: When passing the tube through the hole, remove the top's joints of the tube in advance, and plug them in again before connecting to the nozzles. (8) For routing under the sampler arm, turn the tube one and half around the shaft, and then fix it at the clamp. (9) From the clamp to the solenoid, pass the tube behind the sampler in the same way it was. (10) For mode year 2009 or later, be sure to insert the solenoid cable into the slit of the bracket. NOTE: Otherwise, the cable may interfere the timing belt. NOTE: After reassembling, perform the leak check using Mode 42. Sampler (11) For mode year 2009 or later, be sure to fix the sampler tube as the left. NOTE: Otherwise, the tube may interfere the timing belt. NOTE: Pay attention not to break it at the loop. Approx. 15m m FDC4000 Service Manual 7-13 7 7 Disassembly and Reassembly 7.1.9 Syringe (1) Remove the rear cover of the sampler in advance. (2) Unplug the tube. NOTE: For model year up to 2008, the syringe body has 2 nozzles. When mounting the syringe, the nozzles must point to up. NOTE: For model year 2009 or later, the syringe body has only 1 nozzle. When mounting the syringe, the nozzle must point to the operator side. (3) Open the hook (fixing the plunger of the syringe) to the right to release the plunger. 7 (4) Pull the syringe out to remove. CAUTION Be careful not to bend the plunger of the syringe. CAUTION Be careful not to pull the plunger out from the syringe body. NOTE: After reassembling, perform the leak check using Mode 42. 7-14 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.10 Syringe Cleaning Remove the syringe from the analyzer. Unscrew the left side screw in the figure. A Remove the plunger. B (inside) Clean 4 rings, an O-ring, and 2 X-rings individually. Blow air from the 3 directions. (Easy cleaning) Rub away from both sides using the plunger and neutral detergent. (Careful cleaning) (1) After removing the syringe, perform cleaning as follows. (2) Unscrew the left side screw. (3) Unscrew the plunger part (B) from the syringe body (A). NOTE: Remove locking adhesive attached to A or B parts. If locking adhesive remains here, it is impossible to screw in properly when reassembling. (4) Remove the 4 rings from the plunger. Remove an O-ring and the 2 X-rings from the rings. Clean 4 rings, an O-ring, and 2 X-rings individually. NOTE: It is effective for cleaning to use neutral detergent. After using the detergent, be sure to wash the parts with water and dry them well. (5) Clean the body by blowing air over it from 3 directions. NOTE: It is effective for cleaning to rub away from both sides using the plunger and neutral detergent. After using the detergent, be sure to wash the parts with water and dry them well. (6) Wipe stains off the plunger. After cleaning, reassemble, referring to the guide on the following page. FDC4000 Service Manual 7-15 7 7 Disassembly and Reassembly 7 7-16 FDC4000 Service Manual 7 Disassembly and Reassembly 7 FDC4000 Service Manual 7-17 7 Disassembly and Reassembly 7.1.11 PRES1 Board and Sampler Solenoid Valve <For model year up to 2008> NOTE: For model year 2009 or later, see the next section. (1) Unplug the tube from the syringe. +(6*2 (2) Remove 3 screws and 2 connectors (IFTH2P, IFTH3P) to remove the bracket. +(6*2 (3) Unplug 2 connectors (V301, PRES1). 8 24'5 (4) Loosen the screw of the bracket to pull out the bracket with the PRES1 board. 7 7-18 FDC4000 Service Manual 7 Disassembly and Reassembly (5) Unplug 2 tubes from the solenoid and remove the bracket from the body. NOTE: When reassembling, the tube (for sample) with an O-ring must be connected to the under side (right lower in the photo). (6) Release 4 spacers to remove the PRES1 board. NOTE: After reassembling, perform the leak check using Mode 42. 7 FDC4000 Service Manual 7-19 7 Disassembly and Reassembly 7.1.12 PRES1 Board and Sampler Solenoid Valve <For model year 2009 or later> NOTE: For model year up to 2008, see the previous section. (1) Unplug the tube from the syringe. (Refer to Sec. 7.1.9 (P7-14).) (2) Remove the top support cover. (Refer to Sec. 7.1.6 (P7-10).) (3) Loosen 2 screws of the bracket to take out the bracket with the PRES1 board. NOTE: As the top support cover was removed in Step (2) beforehand, you can access a screw driver from above. (4) Unplug 2 tubes from the solenoid and remove the bracket from the body. NOTE: When reassembling, the tube (for sample) with an O-ring must be connected to the back side. 7 (5) Release 3 spacers to remove the PRES1 board. NOTE: After reassembling, perform the leak check using Mode 42. NOTE: When reassembling the sampler tube, follow the fixing instruction of the tube of Sec. 7.1.8 (11). 7-20 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.13 Sampler Cover Lock Solenoid (1) Remove 2 screws. (2) Unplug the connector (DRV6). &48 (3) Loosen 2 screws to remove the solenoid. NOTE: The left photo is for model year up to 2008. NOTE: For model year 2009 or later (the left photo), a bracket is made of black plastic material. FDC4000 Service Manual 7-21 7 7 Disassembly and Reassembly 7.1.14 Sampler Cover Switch (1) Remove 2 E-rings. /5# (2) Unplug 2 connectors (MSA01/COM, MSA01/ ON) to remove the switch. %1/ 10 /5# NOTE: When reassembling, assemble the white rings as they were. 7 7-22 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.15 Probes CAUTION Do not touch the needle-shaped probes. Otherwise, injuries may result. CAUTION The probes are easily deformed, pay attention not to bend the probes. (1) Raise the probes using the Mechanical adjustment (sec.8.1.5). (Remove the incubator block.) NOTE: Although it is possible to replace the probes at the lower position (home position), it needs skill (difficult). (2) Clamp a probe and pull it up using a pair of needle-nose pliers. (3) Insert a new probe into the socket. Plastic material (4) Using a plastic material, press the top of the probe until it meets the top of the socket. (5) Check the height of the probe within the range using the Mechanical adjustment (Sec. 8.1.5). 7 FDC4000 Service Manual 7-23 7 Disassembly and Reassembly 7.1.16 Probe Unit (1) Loosen 4 screws to remove the incubator rotor. (2) Loosen 2 screws to remove the cover (black, plastic). 7 (3) Unplug the connector (M601). Unplug the connector (H601). / 7-24 * FDC4000 Service Manual 7 Disassembly and Reassembly (4) Loosen 2 screws to remove the probe unit (with cable). CAUTION Pay attention not to bend the probes. (5) Remove 2 screws. (6) Remove the probe motor. 7 (7) Removing the probe motor enables pulling the probe block up to remove. NOTE: After reassembling, adjust the incubator’s insert position (Sec. 8.1.4) of the Mechanical adjustment. FDC4000 Service Manual 7-25 7 Disassembly and Reassembly 7.1.17 Photometer Head <For model year up to 2008> NOTE: For model year 2009 or later, see the next section. (1) To remove the inner box of the disposal box, remove the bottom cover and then remove 2 screws. (Lay the analyzer left side down.) NOTE: Refer to the section of the DRV40 board for removing the bottom cover. NOTE: When laying the analyzer side down, place the left side down. And, place the analyzer on cloth or sheet to avoid scratches. (2) Remove the screw. 7 (3) Remove the inner box of the disposal box. 7-26 FDC4000 Service Manual 7 Disassembly and Reassembly (4) Remove the CPU40 board. NOTE: Refer to the section of the CPU40 board. (5) Loosen 2 screws to remove the bracket of the communication connectors. (6) Loosen 5 screws to remove the rear plate of the CPU40 board. (7) Loosen the screw (locking the optical fiber) a little, and pull the fiber out to the left with holding the metal part of the fiber. (8) After releasing the lock, pull the optical fiber out to the operator side. FDC4000 Service Manual 7-27 7 7 Disassembly and Reassembly (9) Remove the incubator rotor and the black cover beforehand. NOTE: Refer to the section of the Probe unit. (10) Remove the connector and 2 screws to remove the LHA-AMP board. (11) Loosen the black ring nut (locking the photometer head). NOTE: In case of replacing only the optical fiber, loosen the metal nut's part of the fiber to separate it from the photometer head. (12) Remove the photometer head (including the LHA-AMP board) from the body. NOTE: After reassembling, adjust the incubator’s insert position (Sec. 8.1.4) of the Mechanical adjustment. 7 (13) After reassembling, perform the cleaning of the black & white reference plates, and then perform the reference black plate initial setting (Sec. 8.1.28) of the Mechanical adjustment. (14) Last, run the control fluids to check the performance. <= After removing the photometer head assembly from the body 7-28 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.18 Photometer Head <For model year 2009 or later> NOTE: For model year up to 2008, see the previous section. (1) Loosen the screw (locking the optical fiber) a little, and pull the fiber out to the left with holding the metal part of the fiber. And then, pull the optical fiber out to the operator side. (2) Remove the incubator rotor and the black cover. (3) Remove the connector and 2 screws to remove the LHA-AMP board. (4) Remove the left cover and then loosen 2 screws to remove the left window (metal cover). 7 (5) Loosen the black ring nut (locking the photometer head). NOTE: In case of replacing only the optical fiber, loosen the metal nut's part of the fiber to separate it from the photometer head. FDC4000 Service Manual 7-29 7 Disassembly and Reassembly (6) Remove the photometer head (including the LHA-AMP board) from the body. NOTE: After reassembling, adjust the incubator’s insert position (Sec. 8.1.4) of the Mechanical adjustment. (7) After reassembling, perform the cleaning of the black & white reference plates, and then perform the reference black plate initial setting (Sec. 8.1.28) of the Mechanical adjustment. (8) Last, run the control fluids to check the performance. <= After removing the photometer head assembly from the body 7 7-30 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.19 Incubator Bearing (1) Remove the incubator rotor, the photometer head, and the receptacle bracket of the incubator cable connector beforehand. (2) Loosen the screw of the timing belt tension. (3) Remove 3 screws. (4) Lift the incubator bearing up to remove. NOTE: When reassembling, secure the screw which was loosened in (2). NOTE: After reassembling, adjust the incubator’s insert position (Sec. 8.1.4) of the Mechanical adjustment. (5) After reassembling, perform the cleaning of the black & white reference plates, and then perform the reference black plate initial setting (Sec. 8.1.28) of the Mechanical adjustment. (6) Last, run the control fluids to check the performance. FDC4000 Service Manual 7-31 7 7 Disassembly and Reassembly 7.1.20 Exhaust Fan <For model year up to 2008> NOTE: For model year 2009 or later, see the next section. (1) Remove the CPU40 board and its rear plate beforehand. (2) Remove the connector (FANA01) and 2 screws to remove the fan. (#0# 7 7-32 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.21 Exhaust Fan <For model year 2009 or later> NOTE: For model year up to 2008, see the previous section. (1) Unplug the connector (FANA01) (2) Remove the right cover and then loosen 2 screws. (3) Take the bracket with the fan out from the rear side. (4) Loosen 2 screws to replace the fan. FDC4000 Service Manual 7-33 7 7 Disassembly and Reassembly 7.1.22 DRV40 Board (1) Lay the analyzer left side down. NOTE: When laying the analyzer side down, place the left side down. And, place the analyzer on cloth or sheet to avoid scratches. (2) Loosen 6 screws to remove the cover. (3) Unplug all connectors from the DRV40 board. ( ) is a name on the board. • DRV1 • DRV2 • DRV3 • DRV4 • M201(DRV5) • DRV6 • DRV7 • H601(DRV8) • M601(DRV9) • M101(DRV10) • DRV11 • DRV12 (4) Remove 6 screws. NOTE: For model year 2009 or later, remove 4 screws. (The two points ( △ ) were replaced with plastic fixtures.) 7 (5) Remove the DRV40 board. 7.1.23 RAC40 Board (1) Unplug the connector (RAC1). (2) Loosen 2 screws to remove the RAC40 board. 4#% NOTE: For model year 2009 or later, the screws were replaced with plastic spacers. NOTE: After reassembling, perform the sensor test (Sec. 8.2.3) of the Mechanical test mode. 7-34 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.24 CCD40 Board (1) Remove the DRV40 board beforehand. (2) Loosen 4 screws to remove the plate. NOTE: For model year up to 2008, use a short screw (L=4mm) for the triangular position (the others are L=6mm) when reassembling. For model year 2009 or later, use same screws (L=6mm). (3) Unplug 2 connectors (CCD1, CCD3). %%& (4) Loosen 4 screws to remove the CCD40 board. %%& (5) Before replacing the board, check the DIP switch setting (S2) as follows. 10 FDC4000 Service Manual 7-35 7 7 Disassembly and Reassembly (6) In case of replacing only the simple CCD board or the LED40 board separately, disassemble and reassemble them as follows. (7) In case of removing the LED40 board, remove the connector (LED40B) and the screw. .'&$ (8) In case of removing the lens, remove 2 screws. 7 7-36 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.25 Cooling Fan <For model year up to 2008> NOTE: For model year 2009 or later, the cooling fan is not equipped. (1) Remove the CCD40 board beforehand. (2) Remove the inner box of the disposal box beforehand. (Refer to the section of the photometer head.) (3) Loosen 2 screws to remove the plate. (4) Remove the connector (FAN901) and 2 screws to remove the fan. (#0 7 FDC4000 Service Manual 7-37 7 Disassembly and Reassembly 7.1.26 Interference Filter and Lens (1) Remove the lamp unit beforehand. (2) Remove 2 screws ( ○ ). NOTE: For model year 2009 or later, remove 3 screws( ○ and △ ). Pull the plate with the lens out to remove. (3) Remove 2 screws. (4) Remove the interference filter. (5) Clean the black & white reference plate, referring to Sec. 5.2 of the Instruction manual. 7 (6) Clean the lens, referring to Sec. 5.5 of the Instruction manual. (7) After replacing it with a new filter, perform the filter adjustments of the Mechanical adjustment. - HP adjust (Sec. 8.1.26) - First data sampling (Sec. 8.1.27) - Reference black plate initial setting (Sec. 8.1.28) (8) Perform the analyzer calibration (Sec. 8.3). (9) Last, run the control fluids to check the performance. (10) To remove the lens, loosen the nut. 7-38 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.27 Printer (1) Remove 4 screws. (2) Unplug the connector (CN-P). %02 NOTE: If it is difficult to unplug the connector, remove the exhaust fan. It is easy to access it from the under side. NOTE: After replacing, check the print density. Perform the density adjustment using Mode 82 if necessary. FDC4000 Service Manual 7-39 7 7 Disassembly and Reassembly 7.1.28 KEY40 Board (1) Loosen 4 screws to remove the cover. -'; (2) Remove the connectors (KEY1, VFD) and 2 screws to remove the keyboard. 8(& (3) Loosen 9 screws to remove the KEY40 board. 7 7-40 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.29 Display (1) Remove the keyboard beforehand. (2) Loosen 4 screws to remove the display. NOTE: After replacing, check the brightness. Perform the brightness adjustment using Mode 82 if necessary. 7 FDC4000 Service Manual 7-41 7 Disassembly and Reassembly 7.1.30 QC Card Reader (1) Remove the right side cover beforehand. (2) Remove 3 screws. (3) Loosen the screw to remove the bracket. (4) Unplug the connector and remove the bracket (plastic) of the QC card reader. 7 (5) Loosen 2 screws to remove the QC card reader. 7-42 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.31 Top Cover ASSY (1) Remove the right and left covers beforehand. (2) Remove 2 E-rings (at the left side) to remove the white rings (2 pcs.). (3) Loosen 4 screws to remove the side plate. NOTE: The other rings (2 pcs.) can be removed from the white ring's positions. Be careful not to lose them. (4) Remove 2 E-rings (at the right side) to remove the white rings (2 pcs.) in the same way. (5) Loosen 4 screws to remove the side plate. NOTE: The other rings (2 pcs.) can be removed from the white ring's positions. Be careful not to lose them. FDC4000 Service Manual 7-43 7 7 Disassembly and Reassembly (6) Remove 2 screws (near the printer). (7) Remove 2 screws of the right side. (8) Remove the top cover ASSY from the body. 7 7-44 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.32 Power Supply (1) Remove the CPU40 board and its rear plate beforehand. (2) Remove screws (fixing the power supply). For model year upto 2008, remove 5 screws. For model year 2009 or later, remove 3 screws (circled). (3) Remove the power supply from the body. NOTE: The volumes on the power supply (spare part) have already been adjusted at the factory. Basically, the adjustment is not necessary. 7 FDC4000 Service Manual 7-45 7 Disassembly and Reassembly 7.1.33 EPM Board (1) Remove the receptacle bracket of the incubator cable connector beforehand. NOTE: For model year up to 2008, refer to Sec. 7.1.11 (P7-18). NOTE: For model year 2009 or later, remove 2 screws and 2 connectors (IFTH2P, IFTH3P) to remove the bracket, referring to the left photo. (2) Loosen 3 screws ( ○ ) to remove the cover. NOTE: For model year 2009 or later, remove 2 screws ( △ ). (3) Remove 6 cables soldered to the board and unplug the GND cable. NOTE: Before removing the cables, check that the cable’s colors are described in (5). If not, make a memo, and solder the cables as they were, when reassembling. 7 (4) Release 4 spacers to remove the EPM board. (5) When reassembling, solder the cables (removed in (3)) and plug the GND cable in the board as they were. <Cable colors (from the top)> • Pink • White/black spiral • Blue • White • Light green • Gray IMPORTANT Make sure that the cable’s colors are right. Otherwise, wrong results may occur. 7-46 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.34 Incubator Cable (1) Remove 3 thumbscrews and 2 screws, and then remove the cover (black). (2) Loosen 4 screws to remove the cable clamp parts (2 pcs.). (3) Unplug 2 connectors and remove the incubator cable. FDC4000 Service Manual 7-47 7 7 Disassembly and Reassembly 7.1.35 Incubator Pressure Plate (1) Remove the black cover of the incubator beforehand. (2) Remove the E-ring to remove the incubator pressure plate. (3) Clean the outer surface of the pressure plate and the hole in which the plate was inserted. (for 6 pressure plates) NOTE: Use cloth moisten with ethyl alcohol for cleaning. NOTE: When removing the E-ring, the groove of the E-ring may be damaged (burr). Check that there are no burrs at the groove. If they are, remove them with sandpaper. (4) When reassembling, check that 7 leaf springs are mounted on the ceiling of the black cover. 7 (5) After the reassembly is completed, check that all pressure plates can smoothly slide up and down. 7-48 FDC4000 Service Manual 7 Disassembly and Reassembly 7.1.36 Replacing the Fuse WARNING Before replacing the fuse, always unplug the power cable from the analyzer. Replacement procedures (1) Unplug the power cable from the analyzer. (2) Replace the fuse. F2 AC250V F 10A F1 AC250V F 10A Pull out the fuse holder by turning it 1/4 turn counterclockwise using a screwdriver. If the fuse has burned out, replace it with a new one. Insert the fuse holder and lock it by turning it 1/4 turn clockwise. CAUTION DOUBLE POLE/NEUTRAL FUSING Fuse rating: 250V F 10A Fuse maker: LITTLE FUSE 7 Model No.: 314010 CAUTION Be sure to use the above model of fuse (packed as an accessory). Otherwise, the safety cannot be guaranteed. NOTE: If the fuse immediately burns out after the power switch is turned on, contact your technical support for assistance. FDC4000 Service Manual 7-49 7 Disassembly and Reassembly 7 7-50 FDC4000 Service Manual 8 Checking and Adjustment 8 Checking and Adjustment 8.1 8.1.1 Mechanical Adjustment Mechanical Adjustment Menu <Menu tree> Photometer position Incubator Insert position ISE position Tip pick 1 Tip pick 2 Sample Sampler 1 Diluted sample 1 Diluted sample 2 Diluent Spot 1 Spot 2 Tip pick 3 Tip pick 4 Reference fluid Sampler 2 Spot 3 Spot 4 Tip eject Spot standard position 8 CM spot position ISE sample spot position Slide ISE reference fluid spot position Slide insert position Insert standard position HP adjust Filter Initial data sampling Reference black plate initial sampling FDC4000 Service Manual 8-1 8 Checking and Adjustment 8.1.2 Starting up the Adjustment Software (Inspection Tool) 5GV59VQ%1/$ 4GCTRCPGN COM2 COM1A SW1 (1) When using an USB cable for communication, set SW1 (on the rear panel) to the COM1B. COM1B (2) Connect the analyzer (COM1B) and a PC in which the adjustment software has already been installed with an USB cable. %QPPGEVYKVJ 75$ECDNG NOTE: When using a RS232C cable, set SW1 (on the rear panel) to COM1A and plug the cable in the COM1A connector. For the PC’s configuration, select “COM4” for the communication port. #FLWUVUQHVYCTGJCUCNTGCF[DGGPKPUVCNNGF (3) Turn the power switch on with the MODE key and the CAL key continuously pressed at the same time. MODE 㧗 CAL 2QYGT10 2TGUUDQVJMG[U NOTE: Press the keys continuously on until the analyzer displays [Data storing... ] (about 10 seconds) 2TGUUMG[UEQPVKPWQWUN[ CDQWVUGE &CVCUVQTKPI &QPQVRQYGT59QHH (4) The analyzer displays [Adjust Mode]. #FLWUV/QFG 8 8-2 5VCTVWR=+PURGEVKQP6QQN?QPVJG2% (5) Start up the [Inspection Tool] on the PC. FDC4000 Service Manual 8 Checking and Adjustment * The following frame statement is added. The dialog of (9) is revised. (2 nd Edition) Normally, Step (6) & (7) are not necessary. NOTE: When performing the adjustment from the beginning by malfunction of the CPU40 board, [Input Machine No.] is necessary. (6) Click Input machine No. (7) Input a machine No. and click OK. 㧨#FLWUVVQROGPW㧪 %NKEM㨇+PRWV/CEJKPG0Q㨉 㧨/CEJKPG0QKPRWVFKCNQI㧪 +PRWVCOCEJKPG0Q CPFENKEM=1-? Cancel (8) Click Connect. (9) In the dialog, input the machine No. again and click OK. 㧨#FLWUVVQROGPW㧪 %NKEM=%QPPGEV? 8 %JGEMVJGOCEJKPG0Q CPFENKEM=;GU? Yes FDC4000 Service Manual 㧨/CEJKPG0QEJGEMFKCNQI㧪 No 8-3 8 Checking and Adjustment (10) At the top menu for the FDC4000 adjustment, click Mechanical adjust to start up. 㧨#LWUVVQROGPW㧪 5VCTVWR %NKEM =/GEJCPKECNCFLWUV? 6QSWKV=+PURGEVKQP6QQN? ENKEM=':+6? 㧨/GEJCPKECNCFLWUVVQROGPW㧪 8 6QSWKVENKEM=':+6? 8-4 FDC4000 Service Manual 8 Checking and Adjustment 8.1.3 Incubator - Photometer position (3) (4) 2JQVQRQU (6) (6) (5) (7) (1) Click Mechanical adjust at the top menu dialog. (2) Remove the incubator upper block. (3) Click Incubator tab. (4) Click Photo pos. 8 (5) Click GO. => The incubator (cell 1) rotates to the photometer position. (6) Place a slide on cell 1. Adjust the position so that the light projects on the center of the slide by clicking L or R. NOTE: Clicking “-5” or “5” moves the object by 5 pulses. (For other adjustment dialogs, the coarse adjustment buttons are also prepared. Later, the instructions are skipped.)) (7) Click Save. FDC4000 Service Manual 8-5 8 Checking and Adjustment 8.1.4 Incubator - Insert position (1) (4) +PUGTVRQU (4) (2) (5) (1) Click Insert pos. (2) Click GO. => The incubator rotates to the slide insert position. (3) Place slide A (maximum slide). (4) Adjust the position (rotation) as slide A can smoothly slide into the cell by clicking L or R. 8 (5) Click Save. 8-6 FDC4000 Service Manual 8 Checking and Adjustment 8.1.5 Incubator - ISE position (1) (3) +5'RQU (6) (3) (6) (2) (4) (5) (5) (7) (1) Click ISE pos. (2) Click GO. The incubator rotates to the ISE test position. (3) Adjust the rotation by clicking L or R so as to align the probes with the center of the ISE cell’s holes. (4) Click Raise to raise the probes. Check the rotation position. 8 (5) To readjust the rotation, click Lower and GO and return to (3) to adjust the rotation again. (6) Next, adjust the probe height by clicking Up or Down. (Range: 3.7 - 4.2 mm) (7) Click Save. FDC4000 Service Manual 8-7 8 Checking and Adjustment 8.1.6 Sampler 1 - Tip pick 1 (1) (2) 6KRRKEM (4) (8) (4) (8) (3) (5) (7) (8) (8) (10) (9) (1) Click Sampler 1 tab. (2) Click Tip pick 1. (3) Click GO. => The nozzle rotates to hole a. (4) Adjust the rotation by clicking L or R so as to align the nozzle with the center of the hole. 8 (5) Click Go Home pos. => The nozzle goes up to the home position. (6) Set Master tip into hole a. (7) Click Lower. => The nozzle goes down to the tip pick position. (8) Adjust the dimension (press-in) by clicking Up or Down. (Range: 3 +/- 0.3 mm) After the adjustment, click Go Home pos and Lower, and then check it again. (9) Click Go Home pos to raise the nozzle and remove the Master tip by hand. (10) Click Save. 8-8 FDC4000 Service Manual 8 Checking and Adjustment 8.1.7 Sampler 1 - Tip pick 2 (1) 6KRRKEM (7) (3) (3) (7) (2) (4) (6) (7) (7) (8) (8) (1) Click Tip pick 2. (2) Click GO. => The nozzle rotates to hole b. (3) Adjust the rotation by clicking L or R so as to align the nozzle with the center of the hole. (4) Click Go Home pos. => The nozzle goes up to the home position. 8 (5) Set Master tip into hole b. (6) Click Lower. => The nozzle goes down to the tip pick position. (7) Adjust the dimension (press-in) by clicking Up or Down. (Range: 3 +/- 0.3 mm) After the adjustment, click Go Home pos and Lower, and then check it again. (8) Click Go Home pos to raise the nozzle and remove the Master tip by hand. (9) Click Save. FDC4000 Service Manual 8-9 8 Checking and Adjustment 8.1.8 Sampler 1 - Sample (1) (4) 5CORNG (4) (3) (5) (6) (1) Click Sample. (2) Set Master tip into hole a and the sample position tool on the sample rack. (3) Click GO. => The nozzle picks up the tip and rotates to the sample position. (4) Adjust the rotation by clicking L or R so as to align the tip’s top with the center of the tool. 8 (5) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. (6) Click Save. 8-10 FDC4000 Service Manual 8 Checking and Adjustment 8.1.9 Sampler 1 - Diluted sample 1 (1) &KNUCORNG (8) (4) (4) (8) (3) (7) (5) (4) (10) (9) (1) Click Dil sample 1. (2) Set Master tip into hole a and the dilution position tool (with holes) on the sample rack. (3) Click GO. => The nozzle picks up the tip and rotates to the diluted sample position. (4) Click 1mmUP and adjust the rotation by clicking L or R so as to align the tip’s top with the center of the tool’s hole. (5) Click Go Home pos to raise the nozzle. (6) Remove the dilution position tool and set the dilution height tool (no holes). (7) Click GO. => The nozzle rotates to the diluted sample position. (8) Adjust the height by clicking Up or Down so as to align the tip’s top with the surface of the tool. NOTE: This adjustment must be accurate, By using a circuit tester, check the electrical contact between the tip and the tool. Move the nozzle up step by step and adopt the point that the electrical contact is just lost. (9) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. (10) Click Save. FDC4000 Service Manual 8-11 8 8 Checking and Adjustment 8.1.10 Sampler 1 - Diluted sample 2 (1) &KNUCORNG (4) (4) (3) (5) (6) (1) Click Dil sample 2. (2) Set Master tip into hole b and the dilution position tool (with holes) on the sample rack. (3) Click GO. => The nozzle picks up the tip and rotates to the diluted sample position. (4) Adjust the rotation by clicking L or R so as to align the tip’s top with the center of the too’s hole. 8 (5) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. (6) Click Save. 8-12 FDC4000 Service Manual 8 Checking and Adjustment 8.1.11 Sampler 1 - Diluent (1) &KNWGPV (3) (5) (6) (1) Click Diluent. (2) Set Master tip into hole b and the dilution position tool (with holes) on the sample rack. (3) Click GO. => The nozzle picks up the tip and rotates to the diluent position. (4) Check that the tip’s top enters into the tool’s hole. (5) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. 8 (6) Click Exit. FDC4000 Service Manual 8-13 8 Checking and Adjustment 8.1.12 Sampler 1 - Spot 1 (1) (5) 5RQV (8) (5) (8) (3) (6) (7) (5) (10) (9) (1) Click Spot 1. (2) Set Master tip into hole a. (3) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. 8 (4) Place the spot position tool on the transfer base. (5) When the tip’s top touches the tool, use 1mmUp to raise. the nozzle. Adjust the rotation by clicking L or R so as to align the tip’s top with the center groove of the tool. (6) Click Go Home pos. => The nozzle goes up to the home position. (7) Click Lower. => The nozzle goes down to the spotting position. (8) Adjust the height by clicking Up or Down so as to align the tip’s top with the second step (lower) of the tool. NOTE: This adjustment must be accurate, By using a circuit tester, check the electrical contact between the tip and the tool. Move the nozzle up step by step and adopt the point that the electrical contact is just lost. (9) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (10) Click Save. 8-14 FDC4000 Service Manual 8 Checking and Adjustment 8.1.13 Sampler 1 - Spot 2 (1) 5RQV (8) (5) (5) (8) (3) (6) (7) (5) (10) (9) (1) Click Spot 2. (2) Set Master tip into hole b. (3) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. 8 (4) Place the spot position tool on the transfer base. (5) When the tip’s top touches the tool, use 1mmUp to raise. the nozzle. Adjust the rotation by clicking L or R so as to align the tip’s top with the center groove of the tool. (6) Click Go Home pos. => The nozzle goes up to the home position. (7) Click Lower. => The nozzle goes down to the spotting position. (8) Adjust the height by clicking Up or Down so as to align the tip’s top with the second step (lower) of the tool. NOTE: This adjustment must be accurate, By using a circuit tester, check the electrical contact between the tip and the tool. Move the nozzle up step by step and adopt the point that the electrical contact is just lost. (9) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (10) Click Save. FDC4000 Service Manual 8-15 8 Checking and Adjustment 8.1.14 Sampler 2 - Tip pick 3 (1) (2) 6KRRKEM (4) (8) (4) (8) (3) (5) (7) (8) (8) (10) (9) (1) Click Sampler 2 tab. (2) Click Tip pick 3. (3) Click GO. => The nozzle rotates to hole c. (4) Adjust the rotation by clicking L or R so as to align the nozzle with the center of the hole. 8 (5) Click Go Home pos. => The nozzle goes up to the home position. (6) Set Master tip into hole c. (7) Click Lower. => The nozzle goes down to the tip pick position. (8) Adjust the dimension (press-in) by clicking Up or Down. (Range: 3 +/- 0.3 mm) After the adjustment, click Go Home pos and Lower, and then check it again. (9) Click Go Home pos to raise the nozzle and remove the Master tip by hand. (10) Click Save. 8-16 FDC4000 Service Manual 8 Checking and Adjustment 8.1.15 Sampler 2 - Tip pick 4 (1) 6KRRKEM (7) (3) (3) (7) (2) (4) (6) (7) (7) (9) (8) (1) Click Tip pick 4. (2) Click GO. => The nozzle rotates to hole d. (3) Adjust the rotation by clicking L or R so as to align the nozzle with the center of the hole. (4) Click Go Home pos. => The nozzle goes up to the home position. 8 (5) Set Master tip into hole d. (6) Click Lower. => The nozzle goes down to the tip pick position. (7) Adjust the dimension (press-in) by clicking Up or Down. (Range: 3 +/- 0.3 mm) After the adjustment, click Go Home pos and Lower, and then check it again. (8) Click Go Home pos to raise the nozzle and remove the Master tip by hand. (9) Click Save. FDC4000 Service Manual 8-17 8 Checking and Adjustment 8.1.16 Sampler 2 - Reference fluid (1) 4GHHNWKF (3) (5) (6) (1) Click Ref fluid. (2) Set Master tip into hole d. (3) Click GO. => The nozzle picks up the tip and rotates to the reference fluid position. (4) Check that the tip’s top is the reference fluid position. 8 (5) Click Go Home pos to raise the nozzle and remove the Master tip by hand. (6) Click Exit. 8-18 FDC4000 Service Manual 8 Checking and Adjustment 8.1.17 Sampler 2 - Spot 3 (1) (5) 5RQV (5) (3) (6) (5) (7) (1) Click Spot 3. (2) Set Master tip into hole c. (3) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. 8 (4) Place the spot position tool on the transfer base. (5) When the tip’s top touches the tool, use 1mmUp to raise. the nozzle. Adjust the rotation by clicking L or R so as to align the tip’s top with the front groove of the tool. (6) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (7) Click Save. FDC4000 Service Manual 8-19 8 Checking and Adjustment 8.1.18 Sampler 2 - Spot 4 (1) 5RQV (5) (5) (3) (6) (5) (7) (1) Click Spot 4. (2) Set Master tip into hole d. (3) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. 8 (4) Place the spot position tool on the transfer base. (5) When the tip’s top touches the tool, use 1mmUp to raise. the nozzle. Adjust the rotation by clicking L or R so as to align the tip’s top with the back groove of the tool. (6) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (7) Click Save. 8-20 FDC4000 Service Manual 8 Checking and Adjustment 8.1.19 Sampler 2 - Tip eject (1) 6KRGLGEV (3) (5) (7) (1) Click Tip eject. (2) Set an AUTO TIP into hole a. (3) Click GO. => The nozzle picks up the tip and rotates to the tip eject position. (4) Check that the tip eject position is proper. 8 (5) Click Go Home pos. => The nozzle goes up to the home position. (6) Remove the tip by hand. (7) Click Exit. FDC4000 Service Manual 8-21 8 Checking and Adjustment 8.1.20 Sampler 2 - Spot standard position (1) 5RQVUVFRQU (2) (3) (1) Click Spot std pos. (2) Clicking GO starts the automatic adjustment. (3) Click Save. 8 8-22 FDC4000 Service Manual 8 Checking and Adjustment 8.1.21 Slide - CM spot position (1) (2) %/URQVRQU (6) (4) (6) (7) (8) (1) Click Slide tab. (2) Click CM spot pos. (3) Set Master tip into hole a. (4) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. (5) Place the spot position tool on the transfer base. (6) Adjust the transfer by clicking L or R so as to align the tip’s top with the center groove’s end of the tool. (7) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (8) Click Save. FDC4000 Service Manual 8-23 8 8 Checking and Adjustment 8.1.22 Slide - ISE sample spot position (1) +5'UCORNGURQV (5) (5) (3) (6) (7) (1) Click ISE sample spot. (2) Set Master tip into hole c. (3) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. 8 (4) Place the spot position tool on the transfer base. (5) Adjust the transfer by clicking L or R so as to align the tip’s top with the front groove’s end of the tool. (6) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (7) Click Save. 8-24 FDC4000 Service Manual 8 Checking and Adjustment 8.1.23 Slide - ISE reference fluid spot position (1) (5) +5'TGHHNWKFURQV (5) (3) (6) (7) (1) Click ISE ref fluid spot. (2) Set Master tip into hole d. (3) Click GO and open the spotting part cover when picking up the tip. The nozzle rotates to the spotting position. NOTE: The sampler moves continuously. So, open the spotting part cover quickly. 8 (4) Place the spot position tool on the transfer base. (5) Adjust the transfer by clicking L or R so as to align the tip’s top with the back groove’s end of the tool. (6) Click Go Home pos to raise the nozzle and remove the Master tip and the tool by hand. Close the spotting part cover. (7) Click Save. FDC4000 Service Manual 8-25 8 Checking and Adjustment 8.1.24 Slide - Slide insert position (1) +PUGTVRQU (3) (6) (1) Click Insert pos. (2) (Open the spotting part cover.) (3) Click GO. => The incubator rotates to the insert position and the bar moves to the incubator. 8 (4) Place the insert position tool on the cell. Adjust the tool’s hole (φ10mm) and the incubator reading aperture (φ9mm) to be concentric circles. (5) Remove the tool by hand. (6) Click Save. 8-26 FDC4000 Service Manual 8 Checking and Adjustment 8.1.25 Slide - Insert standard position (1) +PUGTVUVFRQU (2) (3) (1) Click Insert std pos. (2) Clicking GO starts the automatic adjustment. (3) Click Save. 8 FDC4000 Service Manual 8-27 8 Checking and Adjustment 8.1.26 Filter - HP adjust (1) (2) *2CFLWUV (5) (4) (6) (1) Click Filter tab. (2) Click HP adjust. (3) Be sure to set the incubator and its cover, and close the disposal box. (4) Click GO. => The analyzer starts the automatic adjustment of the home position. 8 (5) Check that the judge on the PC is OK. (6) Click Save. 8-28 FDC4000 Service Manual 8 Checking and Adjustment 8.1.27 Filter - First data sampling This purpose is memorizing the initial data for detecting the deterioration of the interference filter. When replacing the interference filter, be sure to perform this function. Do not perform this function other than when replacing the interference filter. (1) +PKVKCNUCORNKPI (3) (2) (4) (1) Click Initial sampling. NOTE: Be sure to set the incubator and its cover, and close the disposal box. NOTE: Before performing this function, clean the optical parts (the black and white plate, photometer head, and the lens of the lamp unit). 8 (2) Click GO. => The analyzer performs reading and setting automatically. (3) Check that the reading values have been input into the White column and the values in the Initial light column have been renewed. (4) Click Save. FDC4000 Service Manual 8-29 8 Checking and Adjustment 8.1.28 Filter - Reference black plate initial setting Reference black plate initial sampling This purpose is memorizing the initial (fresh) reading of the reference black plate for detecting the dirty error. Normally, this function is used only in the factory. (1) &CTMRNCVGUCORNKPI (2) (3) (1) Click Black plate sampling. NOTE: Be sure to set the incubator and its cover, and close the disposal box. (2) Click GO. => The analyzer performs reading and setting automatically. 8 (3) Click Save. NOTE: The “Initial black” is obtained by the difference (“Black” - “Dark”). The “Initial black” is the “Init” in Mode 111. (Refer to Sec. 6.13.4 “Mode 111”.) NOTE: In the following condition, E036 will occur. “Current Black” - “Current Dark” - “Initial black” > “ERR detect” (Threshold) * Revised (Oct. 19, 2007) 8-30 FDC4000 Service Manual 8 Checking and Adjustment 8.2 8.2.1 Mechanical Test Mode Mechanical Test Mode Menu Menu 8.2.2 Description 1 Sensor Observes sensor status 2 Filter Deterioration check of the interference filter 3 Slide Slide transfer test 4 Sampler For factory use (leak check) 5 Motor Motor tests 6 Serial com test For factory use (serial communication test) 7 Sample BCR read test For factory use (sample barcode reader test) Starting up the Mechanical Test Mode Start up [Inspection Tool]. (Refer to the mechanical adjustment mode to start up the software.) At the top menu for the FDC4000 adjustment, click Mechanical Test to start up. Click EXIT to terminate the mechanical test. 8 FDC4000 Service Manual 8-31 8 Checking and Adjustment 8.2.3 Sensor (1) (2) (3) (1) Click Sensor tab. (2) Click GO. => The current sensor conditions are displayed. (3) Click STOP. 8 8-32 FDC4000 Service Manual 8 Checking and Adjustment 8.2.4 Filter (1) (2) (1) Click Filter tab. (2) Click GO. => Deterioration (variable rate) for each wavelength will be displayed. 8 FDC4000 Service Manual 8-33 8 Checking and Adjustment 8.2.5 Slide (1) (2) (3) (1) Click Slide tab. (2) Select a menu and load slides. (3) Click GO. => The analyzer starts the slide transfer test. 8 8-34 FDC4000 Service Manual 8 Checking and Adjustment 8.2.6 Sampler (1) (2) (3) This function is for factory use. The function is the same as Mode 42. Using Mode 42 is convenient in the field. 8 FDC4000 Service Manual 8-35 8 Checking and Adjustment 8.2.7 Motor (1) (2) (3) (4) (1) Click Motor tab. (2) Select a motor to perform the test. (3) Click GO. (4) Click STOP to stop the movement. 8 8.2.8 Serial Com Test This function is for factory use (serial communication test). 8.2.9 Sample BCR Read Test This function is for factory use (sample barcode reader test). 8-36 FDC4000 Service Manual 8 Checking and Adjustment 8.3 Analyzer Calibration (1) (3) (4) (1) Click Analyzer calib. at the top menu dialog. (2) Input the center wavelength data of the interference filter. (3) Click GO. (4) Click Exit to quit. FDC4000 Service Manual 8 8-37 8 Checking and Adjustment 8.4 Voltage Inspection 8.4.1 Power Supply +5V 1 ADJ. +5V 2 ADJ. DRIVE1 (CN2) +24V 2 NFB 6 1 +24V NFB 8 1 1 F600 125V T6. 3A +24V 1 NFB 3 9 DRIVE2 (CN5) 12 6 1 1 CPU 2 PRN (CN6) (CN1) 3 1 4 7 LAMP-CNT (CN3) LAMP (CN4) +6V ADJ. (1) Voltage check points Measure voltages at the pins of the connectors on the condition of the connectors (CN1, CN2, CN5, CN6) unplugged. 8 Connector Voltage CN1 +5V +15V -15V +5V +24V +24V +24V +24V +24V +5V +15V -15V +5V CN2 CN5 CN6 8-38 Check point Normal range (V) Pin1 - Pin3 Pin5 - Pin6 Pin7 - Pin6 Pin1 - Pin2 Pin3 - Pin5 Pin4 - Pin6 Pin1 - Pin7 Pin2 - Pin8 Pin3 - Pin9 Pin4 - Pin10 Pin5 - Pin11 Pin6 - Pin11 Pin1 - Pin2 +4.75 to +5.25 +14.25 to +15.75 -14.25 to -15.75 +4.75 to +5.25 +22 to +25 +22 to +25 +22 to +25 +22 to +25 +22 to +25 +4.75 to +5.25 +14.25 to +15.75 -14.25 to -15.75 +4.75 to +5.25 FDC4000 Service Manual 8 Checking and Adjustment (2) Voltage check for lamp power (6V) Turn on the power and wait until the lamp is turned on. Measure the voltage between pin1 and pin4 of the CN4 after unplugging the CN4 connector. [Normal range] 5.85 - 6.15 (V) <All signal information for connectors> Con- Pin nector No. CN1 1 2 3 4 5 6 7 8 CN2 1 2 3 4 5 6 CN4 1 2 3 4 Signal +5V (1) LGND +15V AGND -15V FGND +5V (2) DGND +24V (1) +24V (2) DGND (1) DGND (2) +6V Connector CN5 Pin No. 1 2 3 4 5 6 7 8 9 10 11 12 Signal +24V (1) +24V (2) +24V (3) +5V (1) +15V -15V DGND (1) DGND (2) DGND (3) LGND AGND LGND Connector CN6 Pin No. 1 2 3 4 5 6 7 8 9 10 Signal +5V (1) LGND +5V (1) LGND +5V (1) LGND +5V (1) LGND +5V (1) LGND LGND 8 FDC4000 Service Manual 8-39 8 Checking and Adjustment 8.4.2 CPU40 Board TP12 DGND CF card TP5 A-12V TP11 DGND D54 TP3 A+5V S4 TP9 DGND D11-18 D51 1 ON TP4 AGND D55 TP6 A-5V BAT_ON TP2 A+12V D52 4 TP10 DGND ON TP45 D+3.3V 1 S2 F2 F1 2-4:OFF 1 :ON TP42 PGND TP41 P+24V D56 TP8 D+5V D50 S1 OFF S3 D57 TP43 P+5V (1) Voltage check points Signal Check point LED D+5V +4.75 to +5.25 TP8 D50 D+3.3V +3.15 to +3.45 TP45 D51 DGND 8 Normal range (V) - TP9, TP10, TP11, TP12 - A+5V +4.75 to +5.25 TP3 D54 A-5V -4.75 to -5.25 TP6 D55 A+12V +11.4 to +12.6 TP2 D52 A-12V -11.4 to -12.6 TP5 D53 AGND - TP4 - P+5V +4.75 to +5.25 TP43 D57 P+24V +22 to +25 TP41 D56 PGND 8-40 - TP42 - FDC4000 Service Manual CPU40 D53 OFF 8 8 Checking and Adjustment (2) Others Name Function Remarks S1 Reset switch S2 Control selection (Fixed) S2-1 => ON, Others (2 - 4) => OFF (Fixed) S3 Backup switch Spare parts setting is OFF. When installing in an analyzer, set to ON. S4 Function selection See Sec. 6.2.1. D11 - D14 Temperature control D15 - D18 CPU processing Blinking means CPU is running. 8 FDC4000 Service Manual 8-41 8 Checking and Adjustment 8.4.3 DRV40 Board TP23 DGND TP8 VCC(+5V) +24V2 D2 TP1 TP2 P+5V P+5V +24V2 D4 D5 VCC(+5V) TP3 +24V1 TP28 PGND TP30 PGND DRV40 D3 +24V1 TP21 +24V2 TP20 +24V2 TP14 +24V1 TP12 +24V1 TP13 +24V1 +24V2 TP19 F2 TP31 PGND TP29 PGND F1 (1) Voltage check points Signal VCC(+5V) DGND 8 Normal range (V) +4.75 ~ +5.25 - Check point LED TP8 D5 TP22, TP23 - P+5V +4.75 ~ +5.25 TP1 D4 +24V1 +22 ~ +25 TP3, TP12, TP13, TP14, D3 +24V2 +22 ~ +25 TP2, TP19, TP20, TP21 D2 PGND - 8-42 TP22 DGND TP28, TP29, TP30, TP31 - FDC4000 Service Manual 9 Troubleshooting 9 Troubleshooting 9.1 Troubleshooting Guide 9.1.1 H, L, >, <, @ (Warning for test results) High or low value or qualitative range [Display] [Printer] 6JG ̌*̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV 6225* IFN [Error Description] Testing values are outside of the preset reference intervals that are set by Mode39. => (H, L) Testing values are out of the measurement ranges. => (<, >) Testing values are in qualitative ranges. => (@) Refer to the table as shown below. [Possible Causes] (1) (These mark are not malfunction of the analyzer.) === Check Points ======================================================= < Warning indications for CM tests > Out of determination range: @ Indication Outside of qualitative range Measuring range (Determination range) Indication <, > or = > > Qualitative range @ = Outside of reference interval H = Reference interval H = Outside of reference interval L = Qualitative range @ = Outside of qualitative range FDC4000 Service Manual <, > or = Out of determination range: <, > = = L = 9 < @ = 9-1 9 Troubleshooting < Warning indications for ISE tests > Out of determination range: @ Indication Outside of qualitative range Measuring range (Determination range) Outside of reference interval <, > or = Indication > H Reference interval = L = Qualitative range @ = H = = L = < @ = 9 9-2 <, > or = > = Outside of reference interval Outside of qualitative range Out of determination range: <, > FDC4000 Service Manual 9 Troubleshooting 9.1.2 $ (Warning for test results) Thermistor disconnection [Display] [Printer] 6JG ̌̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV 6225 IFN [Error Description] The thermistor in the incubator is disconnected. [Possible Causes] (1) Thermistor disconnection === Check Points ======================================================= (1) Measure the resistance of the thermistor: => Refer to Section E1502. (2) Check the thermistor and its cable. => Refer to Section E1502. 9 FDC4000 Service Manual 9-3 9 Troubleshooting 9.1.3 + (Warning for test results) Temperature error [Display] [Printer] 6JG ̌ ̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV 6225 IFN [Error Description] [Possible Causes] Incubator temperature has exceeded the upper limit of the range. (1) Incubator cable disconnection (2) Room temperature is high. === Check Points ======================================================= (1) Check the incubator cable. => Refer to Section E1502 and E1503. (2) Check the room temperature. Use Mode 103 to browse the room temperature that the analyzer is recognizing. 9.1.4 - (Warning for test results) Temperature error [Display] [Printer] 6JG ̌̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV 6225 IFN [Error Description] [Possible Causes] 9 Incubator temperature has fallen below the lower limit of the range. (1) Incubator cable disconnection (2) Room temperature is low (out of the specification). === Check Points ======================================================= (1) Check the incubator cable. => Refer to Section E1502 and E1503. (2) Check the room temperature. Use Mode 103 to browse the room temperature that the analyzer is recognizing. 9-4 FDC4000 Service Manual 9 Troubleshooting 9.1.5 * (Warning for test results) Fluctuation of RWP readout [Display] [Printer] 6JG ̌ ̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV $7025 6225 OIFN IFN [Error Description] Reference White Plate (RWP) readout has changed more than 200 from the previous readout. [Possible Causes] (1) Lamp life expired or lamp mounting failure (2) RWP mounting failure (3) Step-out of the incubator rotation (4) Malfunction of the power supply (5) Malfunction of the LHA-AMP board (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the cumulative illumination time of the lamp using Mode 23. (2) Check the lamp unit and the interference filter. (Proper mounting, no loose screws, no stains, no dust) (3) Check the RBP/RWP and the photometer head. (Proper mounting, no loose screws, no stains, no dust) (4) Check the lamp voltage is within the acceptable range (5.85 - 6.15[V]). => Refer to Section 8.2.3 "Voltage Check of the Power Supply". (5) Replace the photometer head (including LHA-AMP board). (6) Replace the CPU40 board 9 FDC4000 Service Manual 9-5 9 Troubleshooting 9.1.6 ? (Warning for test results) RWP readout is outside of the range [Display] [Printer] 6JG ̌!̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV $7025! 6225! OIFN IFN [Error Description] [Possible Causes] RWP readout (A/D) is outside of the range (25900 - 30200). (1) Old light source lamp (2) Malfunction of light source (lenses, filters) (3) Malfunction of the RWP (4) Malfunction of the power supply (5) Defective optical fiber cable (6) Malfunction of the LHA-AMP board (7) Malfunction of the CPU40 board (8) Step-out of the incubator rotation === Check Points ======================================================= (1) Check the cumulative illumination time of the lamp using Mode 23. (2) Check the lamp unit and the interference filter. (Proper mounting, no loose screws, no stains, no dust) (3) Check the RBP/RWP and the photometer head. (Proper mounting, no loose screws, no stains, no dust) (4) Check the lamp voltage is within the acceptable range (5.85 - 6.15[V]). (5) Check the degradation of the optical fiber cable. (6) Replace the photometer head (including LHA-AMP board). (7) Replace the CPU40 board. 9 9-6 FDC4000 Service Manual 9 Troubleshooting 9.1.7 E (Warning for test results) Photometer malfunction [Display] [Printer] 6JG ̌'̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV 6225' IFN [Error Description] (1) The analyzer interrupted the measurement by error such as slide transfer error. (2) Readout of the slide is darker than RBP or brighter than RWP. (3) RWP readout is darker than RBP readout or an impossible calculation (divided by zero) has occurred. [Possible Causes] (1) The disposal box or the keyboard is open during testing. (The photometer unit was exposed to a bright light.) (2) Filled disposal box caused slide jam. (3) A slide in the incubator moved during test processing. (4) Step-out of the syringe operation (5) Malfunction of the LHA-AMP board (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check other errors and perform the necessary troubleshooting. 9 FDC4000 Service Manual 9-7 9 Troubleshooting 9.1.8 & (Warning for test results) Abnormally high testing value [Display] [Printer] 6JG ̌̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV %42 OIFN %C25 OIFN [Error Description] [Possible Causes] Abnormal time course (1) The sample for the immuno-chemical tests may not be diluted during diluting process. The glucose concentration of the sample is too high in immuno-chemical tests. (2) The testing value is abnormally high in general tests. === Check Points ======================================================= (1) Check the sample and perform the test that requires dilution. 9.1.9 # (Warning for test results) Expiration of slide or calibration [Display] [Printer] 6JG ̌̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV ).725 OIFN [Error Description] [Possible Causes] 9 The valid term of the slide or the calibration has expired. (1) The valid term of the slide has expired. (2) The valid term of the calibration for immuno-chemical tests has expired. (3) The calibration for immuno-chemical tests has not been carried out yet. === Check Points ======================================================= (1) Check the valid term of the slide. (2) Check the calibration log (date, valid term) for the immuno-chemical tests. => Refer to Mode 61. 9-8 FDC4000 Service Manual 9 Troubleshooting 9.1.10 ¥ (Warning for test results) Un-spotted slide [Display] [Printer] 6JG ̌>̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV %42> OIFN [Error Description] Detected a time course that resulted in the sample not being spotted onto the slide. [Possible Causes] (1) Abnormal sample (2) Faulty tip (waterproof failure) (3) Leakage of the syringe or piping system === Check Points ======================================================= (1) Check the sample. (2) Check the tip. (3) Check the adjustment of the spotting height. (4) Perform the leak check using Mode 42. (5) Analyze the time course data. (6) Check the threshold for detecting spotting-failure using the Mode 129. 9 FDC4000 Service Manual 9-9 9 Troubleshooting 9.1.11 ERR= *** (Errors for ISE test results) [Display] [Printer] 0C2514O'SN -25ZZZZO'SN %N25 14O'SN '44 [Error Description] Errors for ISE test results (Refer to the error code table as shown below.) A self test is performed automatically during and after ISE measurements. If any errors are detected, [E= *** ] will be printed along with the test results. The 3-digit figures are error codes of Na, K, and CL respectively. '44㧖㧖㧖 'TTQTEQFGHQT%. 'TTQTEQFGHQT'TTQTEQFGHQT0C Error codes (*) will be displayed with numbers 0-8 and each description is listed below. Code 9 Details 0 No problem 1 Drift error 2 Impedance error 3 Outside of the measurement range (This is not an analyzer malfunction.) 4 Drift error and impedance error 5 Impedance error and outside of the measurement range 6 Drift error and outside of the measurement range 7 Drift error, impedance error, and outside of the measurement range 8 Impossible measurement (1) Drift error Summary: Measured voltage fluctuation exceeds the acceptable range. Description: The analyzer measures the voltage after 30, 45, 60 seconds from starting a ISE test. An analyte concentration is calculated using a measured voltage of 60 seconds. If the voltage difference between 30 and 45 seconds and the difference between 45 and 60 seconds exceed the acceptable range, the analyzer indicates "Drift error". The acceptable range of the drift error [The values in the parentheses are for urine samples.] 9-10 Na (mV) K (mV) CL (mV) Difference between 30 and 45 sec. 0.7(1.5) 1.5(1.5) 2.5(2.5) Difference between 45 and 60 sec. 0.7(1.5) 1.2(1.5) 2.0(2.0) FDC4000 Service Manual 9 Troubleshooting (2) Impedance error Summary: Abnormal output impedance of the slide is observed after completing the test Description: The circuit as shown below is formed when the analyzer is performing the impedance check. + 4U 8 + 8U 4T 8T 2/UNKFG Current values flowing into Rs and Rr are equal: ( V – Vs ) ⁄ Rs = ( Vr – V ) ⁄ Rr .... (1) Therefore: Rs = Rr × ( V – Vs ) ⁄ ( Vr – V ) .... (2) Where: Vr = 208mV Rr ( Na ) = 10MΩ ....(3) Rr ( K ) = 10MΩ ....(4) Rr ( CL ) = 100KΩ ....(5) Substituting these into the equation (2) gives: Rs ( Na ) = 10MΩ × ( V – Vs ) ⁄ ( 208mV – V ) Rs ( K ) = 10MΩ × ( V – Vs ) ⁄ ( 208mV – V ) Rs ( CL ) = 100KΩ × ( V – Vs ) ⁄ ( 208mV – V ) Using these formulas, the analyzer calculates the slide impedance. The analyzer determines as normal if the calculated impedances are between 0.1-fold and 10-fold of the standard impedance (Na: 10MΩ, K:10MΩ, CL:100kΩ). In short, acceptable ranges are listed below. Na, K: 1M - 100M Ω CL: 10K - 1MΩ If the calculated impedance is outside of the range, the impedance error is printed on the test result. (Example: ERR - 222) Check Points: High impedance results indicate that effective circuit is not formed on the slide. Sample and/or reference fluid may not be sufficiently dispensed into a slide. Check the following: Aspirated fluids (sample and reference fluid) in the tips are almost equal in volume. Check the spotting process that each fluid is dispensed into the spotting holes on the slide. Replace the O-rings if the analyzer cannot aspirate fluids normally. Check the spotting position onto the slide. Adjust it if necessary. FDC4000 Service Manual 9-11 9 9 Troubleshooting (3) Outside of the measurement range (Error code: “3”) Summary: Outside of the measurement range of the analyte concentration The measured value is within the voltage measurement range of the analyzer (+/- 100 mV), but the calculated analyte concentration is outside of the analyte measurement range. (4) Impossible measurement (Error code: “8”) Summary: Outside of the voltage measurement range The measured voltage is outside of the voltage measurement range (+/-100 mV). Check Points: Check if the spotting process was performed normally. Aspirated fluids (sample and reference fluid) in the tips are almost equal in volume. Check the spotting process that each fluid is dispensed into the spotting holes on the slide. The test result may be affected by the coexisting substances in the sample. Dilute the sample with purified water and rerun the test. 9 9-12 FDC4000 Service Manual 9 Troubleshooting 9.1.12 C (Control mode) [Display] [Printer] 6JG ̌%̍ OCTMKURTKPVGFQWVQPVGUVTGUWNV 0Q % =? ).725OIFN [Error Description] Testing value is a result in the control mode. (This is not an analyzer malfunction.) [Possible Causes] (1) The test results have been measured in the control mode (Mode 19). === Check Points ======================================================= 9 FDC4000 Service Manual 9-13 9 Troubleshooting 9.1.13 W010 Keyboard open [Display] 9 %NQUGMG[DQCTF [Printer] 9 %NQUGMG[DQCTF [Error Description] [Possible Causes] The keyboard is open when the START key is pressed. (1) The keyboard is open during testing (manual pipetting mode). (2) Faulty sensor (PHA01) (3) Cable disconnection (4) Malfunction of the KEY40 board (5) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the sensor function using Mechanical test mode by opening/closing the keyboard. (2) Check the cable connection. (3) Replace the sensor (PHA01). (4) Replace the KEY40 board. (5) Replace the CPU40 board. 9 9-14 FDC4000 Service Manual 9 Troubleshooting 9.1.14 E010 Keyboard open [Display] ' %NQUGMG[DQCTF [Printer] ' %NQUGMG[DQCTF 4GTWPVJGVGUVU [Error Description] [Possible Causes] Keyboard is open during testing (1) The keyboard is open during testing (manual pipetting mode). (2) Faulty sensor (PHA01) (3) Cable disconnection (4) Malfunction of the KEY40 board (5) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the sensor function using Mechanical test mode by opening/closing the keyboard. (2) Check the cable connection. (3) Replace the sensor (PHA01). (4) Replace the KEY40 board. (5) Replace the CPU40 board. 9 FDC4000 Service Manual 9-15 9 Troubleshooting 9.1.15 W020 Disposal box open [Display] 9 %NQUGFKURQUCNDQZ [Printer] 9 %NQUGFKURQUCNDQZ [Error Description] [Possible Causes] The disposal box is open when the START key is pressed. (1) Faulty sensor (PH803) (2) Cable disconnection (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the sensor function using Mechanical test mode by opening/closing the disposal box. (2) Check the cable connection. (3) Replace the sensor (PH803). (4) Replace the CPU40 board. 9.1.16 E021 Disposal box open [Display] ' %NQUGFKURQUCNDQZ [Printer] ' %NQUGFKURQUCNDQZ 4GTWPVJGVGUVU [Error Description] [Possible Causes] 9 The disposal box is open during testing. (1) Faulty sensor (PH803) (2) Cable disconnection (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the sensor function using Mechanical test mode by opening/closing the disposal box. (2) Check the cable connection. (3) Replace the sensor (PH803). (4) Replace the CPU40 board. 9-16 FDC4000 Service Manual 9 Troubleshooting 9.1.17 W022 Disposal box is full [Display] 9 'ORV[&KURQUCNDQZ No error message printed Number of slides or tips in the disposal box has exceeded its capacity. Slide capacity: 80 slides Tip capacity: 40 tips Either condition has been established. [Possible Causes] (1) Disposal box is full of slides or tips. === Check Points ======================================================= (1) Empty the disposal box. The counts of the consumables (slides, tips) and the error are reset by opening the disposal box for the specified time. [Printer] [Error Description] 9 FDC4000 Service Manual 9-17 9 Troubleshooting 9.1.18 W025 Sampler cover open [Display] 95CORNGTEQXGT %NQUGEQXGT [Printer] 9 %NQUGUCORNGTEQXGT [Error Description] [Possible Causes] The sampler cover is open when the START key is pressed. (1) Sampler cover is open (2) Faulty sensor for detecting sampler cover closed (MSA01) (3) Cable disconnection === Check Points ======================================================= Check the sensor function using Mechanical test mode by opening/closing the sampler cover. (1) Check the cable connection. (2) Replace the micro switch (MSA01). 9 9-18 FDC4000 Service Manual 9 Troubleshooting 9.1.19 E025 Sampler cover open [Display] '5CORNGTEQXGT 1RGPGF [Printer] ' 5CORNGTEQXGTQRGPGF 4GTWPVJGVGUVU [Error Description] [Possible Causes] The sampler cover is open during testing. (1) The analyzer failed to lock the sampler cover, so that the sampler cover is open. (2) Faulty sensor for detecting sampler cover closed (MSA01) (3) Cable disconnection === Check Points ======================================================= (1) Check the sensor function using Mechanical test mode by opening/closing the sampler cover. (2) Check the cable connection. (3) Replace the micro switch (MSA01). 9 FDC4000 Service Manual 9-19 9 Troubleshooting 9.1.20 W026 Cover lock failure [Display] 95CORNGTEQXGT %CPPQVNQEMEQXGT [Printer] 9 %CPPQVNQEMURNEQXGT &QPQVQRGPEQXGT [Error Description] Although the analyzer sent a command to lock the sampler cover lock solenoid (SOLC01), the signal from the cover lock sensor (PHC02) has not changed. [Possible Causes] (1) The sampler cover has not been completely closed. (2) Faulty cover lock solenoid (SOLC01) (3) Faulty cover lock sensor (PHC02) (4) Cable disconnection === Check Points ======================================================= (1) Check that the cover lock solenoid works well. (2) For instance, press the START key without tips and a sample (with the sampler cover closed). (3) Check the sensor function using Mechanical test mode. (4) Check the cable connection. (5) Replace the cover lock solenoid (SOLC01). (6) Replace the cover lock sensor (PHC02). 9 9-20 FDC4000 Service Manual 9 Troubleshooting 9.1.21 W027 Sampler cover unlock error [Display] 95CORNGTEQXGT %CPPQVWPNQEMEQXGT [Printer] 9 %CPPQVWPNQEMURNEQXGT 2QYGTQHHQP [Error Description] Although the analyzer sent a command to unlock the sampler cover lock solenoid (SOLC01), the signal from the cover lock sensor (PHC02) has not changed. [Possible Causes] (1) The sampler cover has not been completely closed. (2) Faulty cover lock solenoid (SOLC01) (3) Faulty cover lock sensor (PHC02) (4) Cable disconnection === Check Points ======================================================= (1) Check that the cover lock solenoid works well. (2) Check the sensor function using Mechanical test mode. (3) Check the cable connection. (4) Replace the cover lock solenoid (SOLC01). (5) Replace the cover lock sensor (PHC02). 9 FDC4000 Service Manual 9-21 9 Troubleshooting 9.1.22 W030 Light source lamp replacing warning [Display] 9.QYNKIJV %NGCPRJQVQOGVGT [Printer] 9 .QYNKIJVKPVGPUKV[ %NGCPRJQVQOGVGT PO=O8? [Error Description] [Possible Causes] RWP readout (400 nm, gain=1) is within 19.8 - 22.1 [mV]. (1) Old light source lamp (2) Degraded interference filter (3) Malfunction of the RWP (4) Malfunction of the power supply (5) Defective optical fiber cable (6) Malfunction of the LHA-AMP board (7) Malfunction of the CPU40 board (8) Step-out of the incubator rotation === Check Points ======================================================= (1) Check the cumulative illumination time of the lamp using Mode 23. (2) Clean the lamp unit and the interference filter. (3) Clean the RBP & the RWP and the photometer head. (4) Check the voltage of the lamp power source is within the acceptable range. Measure the voltage at the lamp connector using a circuit tester. Acceptable range: 5.85 - 6.15 [V] (5) Replace the lamp. (After the replacement, reset the cumulative illumination time of the lamp using Mode 23.) (6) Replace the optical fiber. (7) Replace the photometer head (including LHA-AMP board). (8) Replace the CPU40 board. 9 9-22 FDC4000 Service Manual 9 Troubleshooting 9.1.23 E035 Light intensity over range [Display] '.KIJV QXGT %QPVCEV[QWTFGCNGT [Printer] ' .KIJVVQQKPVGPUG %QPVCEV[QWTFGCNGT PO=O8? [Error Description] [Possible Causes] RWP readout (400 nm, gain=1) exceeds 198 [mV]. (1) Interference filter malfunction (2) Malfunction of the RWP (3) Malfunction of the LHA-AMP board (4) Malfunction of the CPU40 board (5) Malfunction of the power supply === Check Points ======================================================= (1) Check the interference filter by visual check. (2) Check that the RBP/RWP ASSY is mounted properly. (3) Replace the photometer head (including LHA-AMP board). (4) Replace the CPU40 board. (5) Check the voltage of the lamp power source is within the acceptable range. Measure the voltage at the lamp connector using a circuit tester. Acceptable range: 5.85 - 6.15 [V] 9 FDC4000 Service Manual 9-23 9 Troubleshooting 9.1.24 E036 Dirty RBP [Display] '$NCEMRNCVG'44 %NGCPTGH$.-RNCVG [Printer] ' %NGCPTGHDNCEMRNCVG ZZZZZ=FGI? [Error Description] [Possible Causes] The RBP readout has exceeded the threshold to detect dirtiness. (1) Dirty (stained) Reference Black Plate (RBP) (2) The detection threshold is too low (sensitive). === Check Points ======================================================= (1) Perform reference plate check using Mode 52 to check the dirty level. (2) Clean the RBP (as well as the RWP). (3) After cleaning, check that the dirty level has been improved by Mode 52. (4) Check the input threshold for detecting dirtiness using Mode 111. (5) The factory setting of the threshold is 200. (October, 2007) < Error printout data > This error occurs in the following condition: Printed black plate readout > Threshold NOTE: Printed black plate readout = Current black readout - Light interrupted readout - Initial black readout NOTE: The “Initial black readout” and the “Threshold” can be browsed and input by Mode 111. NOTE: The “Initial black readout” is obtained by Sec. 8.1.28 “Filter - Reference black plate initial setting” of the Mechanical adjustment. 9 9-24 FDC4000 Service Manual 9 Troubleshooting 9.1.25 W040 Wrong sample type warning [Display] 9 %JGEMUCORNGV[RG [Printer] 9 +PEQTTGEVUCORNGV[RG [Error Description] Read slide code information does not correspond to the specimen type setting of the analyzer. [Possible Causes] (1) User operation mistake === Check Points ======================================================= (1) Slide sample type must correspond to the specimen type setting of the analyzer. 9 FDC4000 Service Manual 9-25 9 Troubleshooting 9.1.26 E050 Lamp replacement error [Display] '.QYNKIJV 4GRNCEGNCOR [Printer] ' .QYNKIJVKPVGPUKV[ 4GRNCEGNCOR PO=O8? [Error Description] [Possible Causes] RWP readout (400 nm, gain=1) is less than 19.8 [mV]. (1) Burned out lamp (2) Degraded interference filter (3) Malfunction of the RWP (4) Malfunction of the power supply (5) Defective optical fiber cable (6) Malfunction of the LHA-AMP board (7) Malfunction of the CPU40 board === Check Points ======================================================= (1) Open the lamp unit cover and turn the power on. Check that the lamp is lit at the initializing process. (2) Check the interference filter by visual check. (3) Check the voltage of the lamp power source is within the acceptable range. Measure the voltage at the lamp connector using a circuit tester. Acceptable range: 5.85 - 6.15 [V] (4) Check the optical fiber cable. (5) Replace the photometer head (including LHA-AMP board). (6) Replace the CPU40 board. 9 9-26 FDC4000 Service Manual 9 Troubleshooting 9.1.27 W060, W061 Insufficient sample / Insufficient calibrator [Display] 90Q5CORNG %JGEMUCORNGXQNWOG 90Q%2HQWPF %JGEMECNKDTCVQT%2 [Printer] 9 0QUCORNGHQWPF %JGEMUCORNGXQNWOG 9 0QECNKDTCVQT%2HQWPF %JGEM%2XQNWOG [Error Description] Error code Error description W060 (Insufficient sample) Although the sample surface has been detected, the sampling volume is insufficient. W061 (Insufficient calibrator) Although the calibrator surface has been detected, the sampling volume is insufficient. [Possible Causes] (1) Liquid volume is less than the lower limit to aspirate. (2) Clogging of the piping system, or sample surface detecting failure caused by leakage of the piping system (3) Malfunction of the PRES1 board (4) Improper adjustment ("Sample position") (5) Cable disconnection (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the liquid volume. (2) Check if the sample rack selecting matches the sample tube type. (3) Reconfirm that the sample rack assembly is correct. (4) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (5) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (6) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment ("Sampler 1" => "Diluted sample 1"). (7) Check the cable connection. (8) Replace the CPU40 board. FDC4000 Service Manual 9-27 9 9 Troubleshooting 9.1.28 W070 Degraded interference filter [Display] 9(KNVGT9#40+0) %QPVCEV[QWTFGCNGT [Printer] 9 (KNVGTEJGEM9#40+0) %NGCPNGPU %QPVCEV[QWTFGCNGT POZZZ [Error Description] The analyzer has detected degradation of the Interference filter during the gain setting process. 400nm: Rate of the degradation exceeds 80%. Others: Rate of the degradation exceeds 50%. [Possible Causes] (1) Degraded interference filter (2) Malfunction of the RWP (3) Malfunction of the power supply (4) Malfunction of the LHA-AMP board (5) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check variable rate of light intensity for all wavelength using mechanical test ("Filter"). (2) Check the interference filter by visual check. (3) Check that the RBP/RWP ASSY is mounted properly. (4) Check the voltage of the lamp power source is within the acceptable range. Measure the voltage at the lamp connector using a circuit tester. Acceptable range: 5.85 - 6.15 [V] (5) Replace the photometer head (including LHA-AMP board). (6) Replace the CPU40 board. 9 9-28 FDC4000 Service Manual 9 Troubleshooting 9.1.29 W081 No sample rack for calibration [Display] 90QUCORNGTCEM [Printer] 9 5GVUCORNGTCEM [Error Description] [Possible Causes] No sample rack loaded when starting calibration. (1) User operation mistake (2) Malfunction of the RAC40 board (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that the sample rack for calibrator CP is set on the analyzer. (2) Check the sensor function using Mechanical test mode. (3) Check the cable connection. (4) Replace the CPU40 board. 9 FDC4000 Service Manual 9-29 9 Troubleshooting 9.1.30 W090, W097, W099 Communication error [Display] 9Z%QO'44 %JGEMEQOECDNG [Printer] 9Z %QOOWPKECVKQP'44 %JGEMEQOECDNG [Error Description] Error code Error description W090 The CTS signal is still set. W097 Mismatched BCC W099 Overflowed receive buffer [Possible Causes] (1) Wrong cable connection (2) Incompatible protocol (3) Malfunction of the host computer (4) Cable disconnection (5) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the cable specifications (pin connection). (2) Check that the host computer is ready to communicate. (3) Check the protocol compatibility between the host computer and the analyzer. (Refer to Mode 46.) (4) Check the cable connection. (5) Replace the CPU40 board. 9 9-30 FDC4000 Service Manual 9 Troubleshooting 9.1.31 E0100 Faulty tip sensor [Display] '6KRUGPUQT'44 %NGCPFKURQUCNDQZ [Printer] ' 6KRUGPUQT'44 [Error Description] The analyzer has detected a tip mounted on the nozzle before the tip detecting process. [Possible Causes] (1) The disposal box is full of tips. (2) Faulty sensor (PH801) (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that the disposal box is not full of tips and is not stuck with tips. (2) Check that the sensor's interrupter moves well. (3) Check the sensor function using Mechanical test mode. (4) Check the cable connection. (5) Replace the CPU40 board. 9 FDC4000 Service Manual 9-31 9 Troubleshooting 9.1.32 W110 No tip warning [Display] 90QUCORNGVKR %JGEMVKRU CVQF [Printer] 9 0QVKR %JGEMVKRU CVQF [Error Description] The analyzer has not detected a tip mounted on the nozzle after the tip mounting process. [Possible Causes] (1) No tip has been set on. (2) Faulty sensor (PH801) (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that the sensor's interrupter moves well. (2) Check the sensor function using Mechanical test mode. (3) Check the cable connection. (4) Replace the CPU40 board. 9 9-32 FDC4000 Service Manual 9 Troubleshooting 9.1.33 E0110, E0111 Clogging error during aspirating sample [Display] '5CORNGUWEM'44 %JGEMUCORNG '5CORNGUWEM'44 %JGEMUCORNG [Printer] ' %NQIIKPI UCORNG UWEM %JGEMUCORNG ' %NQIIKPI UCORNG UWEM %JGEMUCORNG [Error Description] Error code Error description E0110 (Clog in sample pre-aspirating process) Difference of the air pressures in the piping system between pre-aspiration and post-aspiration has exceeded the threshold level to detect clogging during the sample pre-aspirating process. * Difference of the air pressures to detect clogging: 250 (mV) for under 10μL pipetting 700 (mV) for over 10μL pipetting E0111 (Clog in sample aspirating process) Difference of the air pressures in the piping system between pre-aspiration and post-aspiration has exceeded the threshold level to detect clogging during the sample aspirating process. * Difference of the air pressures to detect clogging: 250 (mV) for under 10μL pipetting 700 (mV) for over 10μL pipetting [Possible Causes] (1) Fibrin deposit (2) Clogging of the piping system (3) Malfunction of the PRES1 board (4) Malfunction of the CPU40 board (5) Cable disconnection === Check Points ======================================================= (1) Check the sample fluid (fibrin deposit). (2) Check clogging of the piping system. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (3) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (4) Replace the CPU40 board. (5) Check the cable connection. FDC4000 Service Manual 9-33 9 9 Troubleshooting 9.1.34 W112 Wrong test programming [Display] 90QUCORNGVKR %JGEMVKRU CVQF [Printer] 9 0QVKR %JGEMVKRU CVQF [Error Description] Loaded tips have already used up until the previous measurement so that there are no tips for the next measurement. [Possible Causes] (1) The recommended slide stacking rules are not followed. === Check Points ======================================================= (1) Follow the recommended slide stacking rules described in the instruction manual. 9 9-34 FDC4000 Service Manual 9 Troubleshooting 9.1.35 E0112, E0113 Clogging error during spotting sample [Display] '5CORNGURQV'44 %JGEMUCORNG '5CORNGURQV'44 %JGEMUCORNG [Printer] ' %NQIIKPI UCORNG URQV %JGEMUCORNG ' %NQIIKPI UCORNG URQV %JGEMUCORNG [Error Description] Error code Error description E0112 (Clog in sample spotting process) Difference of the air pressures in the piping system between pre-spotting and post-spotting has exceeded the threshold level to detect clogging during the sample spotting process. * Difference of the air pressures to detect clogging: 250 (mV) for under 10μL pipetting 700 (mV) for over 10μL pipetting E0113 (Clog in sample pre-dispensing process) Difference of the air pressures in the piping system between pre-spotting and post-spotting has exceeded the threshold level to detect clogging during the sample predispensing process. * Difference of the air pressures to detect clogging: 250 (mV) for under 10μL pipetting 700 (mV) for over 10μL pipetting [Possible Causes] (1) Fibrin deposit (2) Clogging of the piping system (3) Malfunction of the PRES1 board (4) Malfunction of the CPU40 board (5) Cable disconnection === Check Points ======================================================= (1) Check the sample fluid (fibrin deposit). (2) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (3) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (4) Replace the CPU40 board. (5) Check the cable connection. FDC4000 Service Manual 9-35 9 9 Troubleshooting 9.1.36 W115, W116 Tip ejection warning [Display] 96KRGLGEV'44 96KRGLGEV'44 [Printer] 9 6KRGLGEV'44 4GOQXGVKRHTQOPQ\\NG 9 6KRGLGEV'44 4GOQXGVKRHTQOPQ\\NG [Error Description] Error code Error description W115 A tip is being detected on the sample nozzle after the tip ejection. W116 A tip is being detected on the dilution nozzle after the tip ejection. [Possible Causes] (1) The disposal box is full of tips. (2) Faulty sensor (PH801) (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that the disposal box is not full of tips and is not stuck with tips. (2) Check that the sensor's interrupter moves well. (3) Check the sensor function using Mechanical test mode. (4) Check the cable connection. (5) Replace the CPU40 board. 9 9-36 FDC4000 Service Manual 9 Troubleshooting 9.1.37 W120 Tip error [Display] 96KR'44 4GOQXGVKRU [Printer] 9 4GOQXGVKRUHTQOPQ\\NGU 'ORV[&KURQUCNDQZ [Error Description] [Possible Causes] Regardless of the tip remaining error, the START key has been pressed. (1) A tip remains on the nozzle. (2) Faulty sensor (PH801) (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Remove the tip from the sampler nozzle by hand and empty the disposal box. (2) Check that the disposal box is not full of tips and is not stuck with tips. (3) Check that the sensor's interrupter moves well. (4) Check the sensor function using Mechanical test mode. (5) Check the cable connection. (6) Replace the CPU40 board. 9 FDC4000 Service Manual 9-37 9 Troubleshooting 9.1.38 E0120 No sample found [Display] '0QUCORNG %JGEMUCORNGXQNWOG [Printer] ' 0QUCORNGHQWPF %JGEMUCORNGXQNWOG [Error Description] [Possible Causes] 9 The analyzer could not detect the sample surface. (1) A sample has not been set in the sample rack correctly. (2) Malfunction of the PRES1 board (3) Leakage or clogging of the piping system (4) Clogging of the nozzle (5) O-ring damage (6) Syringe malfunction (7) Cable disconnection (8) Malfunction of the CPU40 board (9) Malfunction of the RAC40 board === Check Points ======================================================= (1) Check the sample volume. (2) Check if the sample rack selecting matches the sample tube type. (3) Reconfirm that the sample rack assembly is correct. (4) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (5) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (6) Make sure that the O-ring does not have any damage. (7) Check the syringe. (8) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment: To adjust: ("Sampler 1" => "Diluted sample 1") To check: ("Sampler 1" => "Sample") (9) Check the cable connection. (10) Replace the CPU40 board. (11) Replace the KEY40 board. 9-38 FDC4000 Service Manual 9 Troubleshooting 9.1.39 E0121, E0122 Clogging error (diluent) [Display] '&KNUWEM'44 %JGEMFKNWGPV '&KNURQV'44 %JGEMFKNWGPV [Printer] ' %NQIIKPI FKN UWEM %JGEMFKNWGPV ' %NQIIKPI FKN URQV %JGEMFKNWGPV [Error Description] Error code Error description E0121 (Clog in diluent aspirating process) Difference of the air pressures in the piping system between pre-aspiration and post-aspiration has exceeded the threshold level to detect clogging during the diluent aspirating process. * Difference of the air pressures to detect clogging: 700 (mV) E0122 (Clog in diluent dispensing process) Difference of the air pressures in the piping system between pre-dispensing and post-dispensing has exceeded the threshold level to detect clogging during the diluent dispensing process. * Difference of the air pressures to detect clogging: 700 (mV) [Possible Causes] (1) Fibrin deposit (2) Clogging of the piping system (3) Malfunction of the PRES1 board (4) Malfunction of the CPU40 board (5) Cable disconnection === Check Points ======================================================= (1) Check the sample fluid (fibrin deposit). (2) Check clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (3) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (4) Replace the CPU40 board. (5) Check the cable connection. FDC4000 Service Manual 9-39 9 9 Troubleshooting 9.1.40 E0123 No diluent found [Display] '0QFKNWGPV %JGEMFKNWGPV [Printer] ' 0QFKNWGPVHQWPF %JGEMFKNWGPV [Error Description] [Possible Causes] 9 The analyzer could not detect the diluent surface. (1) Diluent has not been set in the sample rack correctly. (2) Malfunction of the PRES1 board (3) Leakage or clogging of the piping system (4) Clogging of the nozzle (5) O-ring damage (6) Syringe malfunction (7) Cable disconnection (8) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the diluent volume. (2) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (3) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (4) Make sure that the O-ring does not have any damage. (5) Check the syringe. (6) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment: To adjust: ("Sampler 1" => "Diluted sample 1") To check: ("Sampler 1" => "Diluent") (7) Check the cable connection. (8) Replace the CPU40 board. 9-40 FDC4000 Service Manual 9 Troubleshooting 9.1.41 E0124, E0125 Clogging error (diluted sample) [Display] '5CORNGUWEM'44 %JGEMFKNWVGFUCORNG '5CORNGURQV'44 %JGEMFKNWVGFUCORNG ' %NQIIKPI FKNURN UWEM ' %NQIIKPI FKNURN URQV [Printer] [Error Description] Error code Error description E0124 (Clog in diluted sample aspirating process) Difference of the air pressures in the piping system between pre-aspiration and post-aspiration has exceeded the threshold level to detect clogging during the diluted sample aspirating process. * Difference of the air pressures to detect clogging: 250 (mV) for under 10μL pipetting 700 (mV) for over 10μL pipetting E0125 (Clog in diluted sample spotting process) Difference of the air pressures in the piping system between pre-spotting and post-spotting has exceeded the threshold level to detect clogging during the diluted sample spotting process. * Difference of the air pressures to detect clogging: 250 (mV) for under 10μL pipetting 700 (mV) for over 10μL pipetting [Possible Causes] (1) Fibrin deposit (2) Clogging of the piping system (3) Malfunction of the PRES1 board (4) Malfunction of the CPU40 board (5) Cable disconnection === Check Points ======================================================= (1) Check the sample fluid (fibrin deposit). (2) Check clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (3) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (4) Replace the CPU40 board. (5) Check the cable connection. FDC4000 Service Manual 9-41 9 9 Troubleshooting 9.1.42 E0126, E0127, E0128 Clogging error (reference fluid) [Display] '4GHUWEM'44 %JGEMTGHHNWKF '4GHUWEM'44 %JGEMTGHHNWKF '4GHURQV'44 %JGEMTGHHNWKF ' %NQIIKPI TGH UWEM %JGEMTGHGTGPEGHNWKF ' %NQIIKPI TGH URQV %JGEMTGHGTGPEGHNWKF [Printer] ' %NQIIKPI TGH UWEM [Error Description] Error code Error description E0126 (Clog in reference fluid preaspirating process) Difference of the air pressures in the piping system between pre-aspiration and post-aspiration has exceeded the threshold level to detect clogging during the reference fluid pre-aspirating process. * Difference of the air pressures to detect clogging: 700 (mV) E0127 (Clog in reference fluid aspirating process) Difference of the air pressures in the piping system between pre-aspiration and post-aspiration has exceeded the threshold level to detect clogging during the reference fluid aspirating process. * Difference of the air pressures to detect clogging: 700 (mV) E0128 (Clog in reference fluid dispensing process) Difference of the air pressures in the piping system between pre-dispensing and post-dispensing has exceeded the threshold level to detect clogging during the reference fluid dispensing process. * Difference of the air pressures to detect clogging: 700 (mV) [Possible Causes] 9 (1) Clogging of the piping system (2) Malfunction of the PRES1 board (3) Malfunction of the CPU40 board (4) Cable disconnection === Check Points ======================================================= (1) Check clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (2) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (3) Replace the CPU40 board. (4) Check the cable connection. 9-42 FDC4000 Service Manual 9 Troubleshooting 9.1.43 E0129 No reference fluid found [Display] '0QTGHHNWKF %JGEMTGHHNWKF [Printer] ' 0QTGHHNWKFHQWPF %JGEMTGHGTGPEGHNWKF [Error Description] [Possible Causes] The analyzer could not detect the reference fluid surface. (1) Reference fluid has not been set in the setting position correctly. (2) Malfunction of the PRES1 board (3) Leakage or clogging of the piping system (4) Clogging of the nozzle (5) O-ring damage (6) Syringe malfunction (7) Cable disconnection (8) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the reference fluid volume. (2) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (3) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (4) Make sure that the O-ring does not have any damage. (5) Check the syringe. (6) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment: To adjust: ("Sampler 1" => "Diluted sample 1") To check: ("Sampler 2" => "Reference fluid") (7) Check the cable connection. (8) Replace the CPU40 board. 9 FDC4000 Service Manual 9-43 9 Troubleshooting 9.1.44 W140 Incorrect direction loading of ISE slide [Display] 99TQPI+5'FKT +5'UNKFGFKUECTFGF [Printer] 9 +5'UNKFGFKTGEVKQP'44 [Error Description] The CCD sensor has detected that a ISE slide is incorrectly loaded in the slide cartridge. [Possible Causes] (1) Incorrect slide loading into the slide cartridge (user operation mistake) === Check Points ======================================================= (1) Load the ISE slide in correct direction in the slide cartridge. 9 9-44 FDC4000 Service Manual 9 Troubleshooting 9.1.45 E0140, E0141, E0142 Excessive liquid volume (beyond the upper limit) [Display] '5CORNG'44 'ZEGUUKXGUCORNG '&KNWGPV'44 'ZEGUUKXGFKNWGPV '4GHHNWKF'44 'ZEGUUKXGTGHHNWKF [Printer] ' 5CORNG'44 'ZEGUUKXGUCORNG %JGEMUCORNG ' &KNWGPV'44 'ZEGUUKXGFKNWGPV %JGEMFKNWGPV ' 4GHHNWKF'44 'ZEGUUKXGTGHHNWKF %JGEMTGHGTGPEGHNWKF [Error Description] Error code Error description E0140 (Excessive sample) The sample surface is beyond the upper limit to aspirate. E0141 (Excessive diluent) The diluent surface is beyond the upper limit to aspirate. E0142 (Excessive reference fluid) The reference fluid surface is beyond the upper limit to aspirate. [Possible Causes] (1) Excessive liquid (2) Malfunction of the PRES1 board (3) Clogging of the piping system (4) Clogging of the nozzle === Check Points ======================================================= (1) Check the liquid volume. (2) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (3) Check clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. FDC4000 Service Manual 9-45 9 9 Troubleshooting 9.1.46 W141 Impossible to start ISE tests [Display] 9+5''44 +5'VGUVUECPPQVTWP [Printer] 9 +5'VGUVUECPPQVTWP [Error Description] After "E0530" to "E0532" occurred, ISE tests are required without shutdown. [Possible Causes] (1) Refer to the above. === Check Points ======================================================= (1) Perform troubleshooting for "E0530" - "E0532" before the "W141", and then turn the power off and on. 9 9-46 FDC4000 Service Manual 9 Troubleshooting 9.1.47 E0143 Insufficient diluent [Display] '0QFKNWGPV %JGEMFKNWGPV [Printer] ' 0QFKNWGPVHQWPF [Error Description] Although the diluent surface has been detected, the volume for aspirating is insufficient. [Possible Causes] (1) Liquid volume is less than the lower limit to aspirate. (2) Clogging of the piping system, or sample surface detecting failure caused by leakage of the piping system (3) Malfunction of the PRES1 board (4) Improper adjustment ("Sample position") (5) Cable disconnection (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the diluent volume. (2) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (3) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (4) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment: To adjust: ("Sampler 1" => "Diluted sample 1") To check: ("Sampler 1" => "Diluent") (5) Check the cable connection. (6) Replace the CPU40 board. 9 FDC4000 Service Manual 9-47 9 Troubleshooting 9.1.48 E0144 Insufficient reference fluid [Display] '0QTGHHNWKF %JGEMTGHHNWKF [Printer] ' 0QTGHGTGPEGHNWKF [Error Description] Although the reference fluid surface has been detected, the volume for aspirating is insufficient. [Possible Causes] (1) Liquid volume is less than the lower limit to aspirate. (2) Clogging of the piping system, or sample surface detecting failure caused by leakage of the piping system (3) Malfunction of the PRES1 board (4) Improper adjustment ("Sample position") (5) Cable disconnection (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the reference fluid volume. (2) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (3) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (4) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment: To adjust: ("Sampler 1" => "Diluted sample 1") To check: ("Sampler 2" => "Reference fluid") (5) Check the cable connection. (6) Replace the CPU40 board. 9 9-48 FDC4000 Service Manual 9 Troubleshooting 9.1.49 E0145 No calibrator found [Display] '0Q%2HQWPF %JGEMECNKDTCVQT%2 [Printer] ' 0QECNKDTCVQT%2HQWPF %JGEM%2XQNWOG [Error Description] [Possible Causes] The analyzer could not detect the calibrator (CP) surface. (1) A calibrator has not been set in the sample rack correctly. (2) Malfunction of the PRES1 board (3) Leakage or clogging of the piping system (4) Clogging of the nozzle (5) O-ring damage (6) Syringe malfunction (7) Cable disconnection (8) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the remaining volume of the calibrator. (2) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (3) Check leakage and clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. (4) Make sure that the O-ring does not have any damage. (5) Check the syringe. (6) Check the adjustment of "Diluted sample 1” position (height). Refer to the mechanical adjustment: To adjust: ("Sampler 1" => "Diluted sample 1") To check: ("Sampler 1" => "Sample") (7) Check the cable connection. (8) Replace the CPU40 board. FDC4000 Service Manual 9-49 9 9 Troubleshooting 9.1.50 E0146 Excessive calibrator volume (beyond the upper limit) [Display] '%2'44 'ZEGUUKXGECNKDTCVQT [Printer] ' %2'44 'ZEGUUKXGECNKDTCVQT%2 %JGEM%2XQNWOG [Error Description] [Possible Causes] The calibrator surface is beyond the upper limit to aspirate. (1) Excessive liquid (2) Malfunction of the PRES1 board (3) Clogging of the piping system (4) Clogging of the nozzle === Check Points ======================================================= (1) Check the calibrator volume. (No foam on the bottle’s openning) (2) Check the PRES1 board as follows. Check the LED on the board is lit when a tip is not mounted on the nozzle. Check the LED blinks while tapping the orifice of the nozzle with a finger. (3) Check clogging of the piping system as follows. Perform the leak check using Mode 42. Make sure there is no material which could clog the tubes or the pipes, particularly at the joints of the tubes and the pipes. If the analyzer aspirated the liquid into the nozzle: Remove the sampler nozzle cover. Disconnect the tube from the nozzle and blow air from the pipe of the nozzle to remove the liquid. 9 9-50 FDC4000 Service Manual 9 Troubleshooting 9.1.51 W150 Dilution tests over [Display] 9&KNVGUVUQXGT 5GVEQPUWOCDNGU [Printer] 9 &KNVGUVUQXGT %JGEMTGHHNWKF [Error Description] Sampling from a mixing cup was required more than 3 times. [Possible Causes] (1) User operation mistake === Check Points ======================================================= (1) Number of sampling from a mixing cup is allowed up to 3 times. 9.1.52 W160 Dilution factor error [Display] 9&KNHCEVQT'44 5GVEQPUWOCDNGU [Printer] 9 &KNHCEVQT'44 %JGEMTGHHNWKF [Error Description] [Possible Causes] FDC4000 Service Manual Tests that require different dilution factors were programmed for a sample. (1) User operation mistake 9-51 9 9 Troubleshooting 9.1.53 W170 No QC information [Display] 90Q3%KPHQ 4GCF3%ECTF [Printer] 9 0Q3%KPHQ 4GCF3%ECTF ZZZZ 5NKFGEQFG [Error Description] The QC card corresponding to the slide lot has not been read. [Possible Causes] (1) No QC card has been read. === Check Points ======================================================= (1) Read the QC card corresponding to the slide. (2) Check the slide lot No. which have already been read using Mode 29. 9 9-52 FDC4000 Service Manual 9 Troubleshooting 9.1.54 W173 Slide reading error [Display] 9 5NKFGKPHQTGCF'44 [Printer] 9 5NKFGKPHQTGCF'44 [Error Description] Slide reader could not read dot information printed on the reverse side of the slide. [Possible Causes] (1) Dirty slide reader unit (CCD) (2) The slide is out of the standard position in the slide cartridge, so that the CCD cannot see the printed dot information. (3) Unclear printing of the slide's dot information (4) Malfunction of the CCD40 board (5) Cable disconnection (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that no slides lunge out of the slide cartridge. (2) Check the printing quality of the slide's dot information. (3) Clean the slide reader (glass window). (4) Check that the illuminating LED on the CCD40 board is lit when reading a slide. (5) Check the cable connection. 9 FDC4000 Service Manual 9-53 9 Troubleshooting 9.1.55 W175 No slides loaded [Display] 0QUNKFGNQCFGF [Printer] 9 5NKFGUJCXGPQVDGGPUGV [Error Description] [Possible Causes] No slides are loaded in the slide cartridge. (1) Regardless of loading slides to be tested, the START key has been pressed. (2) Dirty slide reader unit (CCD) (3) The slide is out of the standard position in the slide cartridge, so that the CCD cannot see the printed dot information. (4) Unclear printing of the slide's dot information (5) Malfunction of the CCD40 board (6) Cable disconnection (7) Malfunction of the CPU40 board === Check Points ======================================================= (1) Load slides to be tested, and rerun the tests from the beginning. (2) Check that no slides lunge out of the slide cartridge. (3) Check the printing quality of the slide's dot information. (4) Clean the slide reader (glass window). (5) Check that the illuminating LED on the CCD40 board is lit when reading a slide. (6) Check the cable connection. 9 9-54 FDC4000 Service Manual 9 Troubleshooting 9.1.56 W180 No calibration performed [Display] 90Q%CNKDKPHQ 2GTHQTO%CNKDTCVKQP [Printer] 9 2GTHQTOECNKDTCVKQP [Error Description] The calibration for a detected immuno-chemical test slide has not been performed. [Possible Causes] (1) Refer to the above. === Check Points ======================================================= (1) Perform calibration for the immuno-chemical test slide. 9.1.57 W190, W191 Wrong slide loading in calibration process [Display] 9%CNKD'44 %JGEMUNKFGPCOG 9%CNKD'44 %JGEMUNKFGNQV [Printer] 9 9TQPIUNKFGHQTECN 9 9TQPIUNKFGNQVHQTECN [Error Description] Error code Error description W190 (Wrong test name) A slide that does not require calibration has been detected on the slide cartridge in calibration process. W191 (Wrong slide lot) A slide that is different from the first slide’s lot has been detected on the slide cartridge in calibration process. 9 [Possible Causes] (1) Refer to the above. === Check Points ======================================================= (1) Load the same lot of slide as the first.(W191) FDC4000 Service Manual 9-55 9 Troubleshooting 9.1.58 W200 Sample barcode reader error (No response) [Display] 9$CTEQFGTGCFGT 0QEQPPGEVKQP [Printer] 9 $CTEQFGTGCFGT'44 0QEQPPGEVKQP [Error Description] [Possible Causes] The sample barcode reader is not connected to the analyzer. (1) Disconnection of the sample barcode reader (2) Malfunction of the sample barcode reader === Check Points ======================================================= (1) Check that the sample barcode reader is specified for the FDC4000. (2) Check the sample barcode reader connected to the COM2 on the rear panel. (3) Check that the LED of the sample barcode reader is lit. (4) If the sample barcode reader cannot be operated, change the setting by Mode 46 to set the sample barcode reader for disuse. 9 9-56 FDC4000 Service Manual 9 Troubleshooting 9.1.59 E0200 - E0210 Feed motor error [Display] '(GGFOQVQT'44 5GPUQT QP HCKNWTG '(GGFOQVQT'44 5GPUQT QHH HCKNWTG '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' (GGFOQVQT'44 5GPUQT QP HCKNWTG ' (GGFOQVQT'44 5GPUQT QHH HCKNWTG ' (GGFOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT [Error Description] Error code Error description E0200 The slide feed HP sensor (PH101) could not detect the interrupter after sending the command during initialization. E0201 The slide feed HP sensor (PH101) is still detecting the interrupter after sending the command during initialization. E0202 When transferring a slide to the spotting position, the slide feed spot sensor (PH103) could not detect the interrupter. E0203 Before transferring a slide to the spotting position, the slide feed spot sensor (PH103) is previously detecting the interrupter. E0204 When transferring a slide to the insert position, the slide feed insert sensor (PH102) could not detect the interrupter. E0205 Before transferring a slide to the insert position, the slide feed insert sensor (PH102) is previously detecting the interrupter. E0206 The spotting position is out of the acceptable range. E0207 The insert position is out of the acceptable range. E0208 The slide feed insert sensor (PH102) has detected transfer error during slide eject (forward process). E0209 The slide feed insert sensor (PH102) has detected transfer error during slide eject (back process). E0210 The software could not control the motor. [Possible Causes] FDC4000 Service Manual (1) (2) (3) (4) (5) (6) (7) (8) 9 Presence of foreign matter Disposal box is full of slides or tips. Faulty appropriate sensor Cable disconnection Malfunction of the DRV40 board Faulty slide feed motor (M101) Malfunction of the CPU40 board Improper timing belt tension 9-57 9 Troubleshooting === Check Points ======================================================= (1) Get to know the error conditions (the place of the error occurrence, the position of the slide insert bar, and the sensor which has detected the error) to inspect the mechanical interference. (2) Empty the disposal box. (3) Check the following sensors: Slide feed HP sensor (PH101) Slide feed insert sensor (PH102) Slide feed spot sensor (PH103) Check the sensor function using Mechanical test mode. (4) Replace the CPU40 board. (5) Replace the slide feed motor (M101). (6) Replace the CPU40 board. (7) Check the timing belt tension. 9 9-58 FDC4000 Service Manual 9 Troubleshooting 9.1.60 E0300, E0301, E0302 Gain error [Display] '2JQVQOGVGT'44 %QPVCEV[QWTFGCNGT '2JQVQOGVGT'44 %QPVCEV[QWTFGCNGT '2JQVQOGVGT'44 %QPVCEV[QWTFGCNGT ' 2JQVQOGVGT'44 %QPVCEV[QWTFGCNGT POZ=FGI? ' 2JQVQOGVGT'44 %QPVCEV[QWTFGCNGT POZ=FGI? [Printer] ' 2JQVQOGVGT'44 %QPVCEV[QWTFGCNGT POZ=FGI? Z=FGI? [Error Description] Error code Error description E0300 The A/D reading value during the gain setting (gain = n) is outside of the range (25800 - 30200). E0301 The A/D reading value of RBP during the gain setting (gain = n) exceeds 25800. E0302 The A/D reading value at the light shielding position during the gain setting (gain=n) exceeds 25800. [Possible Causes] (1) Old light source lamp (2) Degraded interference filter (3) Malfunction of the RWP (4) Malfunction of the power supply (5) Defective optical fiber cable (6) Malfunction of the LHA-AMP board (7) Malfunction of the CPU40 board (8) Loose connection === Check Points ======================================================= (1) Check that there are no unplugged connectors or loose connection for the lamp, the photometer, and the LHA-AMP board. (2) Check the cumulative illumination time of the lamp using Mode 23. (3) Check the lamp unit and the interference filter. (4) Check the RBP/RWP and the photometer head. (5) Check the voltage of the lamp power source is within the acceptable range. (6) Check the degradation of the optical fiber cable. (7) Replace the photometer head (including LHA-AMP board). (8) Replace the CPU40 board. FDC4000 Service Manual 9-59 9 9 Troubleshooting 9.1.61 W500, W501 Fan rotation stop [Display] 9(CWNV[HCP 9(CWNV[KPEWHCP [Printer] 9 (CWNV[HCP %QPVCEV[QWTFGCNGT 9 (CWNV[KPEWHCP %QPVCEV[QWTFGCNGT [Error Description] Error code Error description W500 The rotation of the fan (FANA01) is stopping. W501 The rotation of the fan (FAN901) is stopping. [Possible Causes] (1) Faulty fan (2) Cable disconnection (3) Malfunction of the CPU40 board (4) Malfunction of the power supply === Check Points ======================================================= (1) Check the cable connection. (2) Check the fan's rotation stops and replace the fan. The FANA01 starts the rotation by turning the power on. For model year up to 2008, the FAN901 is equipped, but not in use. For model year 2009 or later, the FAN901 is not equipped. (3) Check the power voltage (+24V). (4) Replace the CPU40 board. 9 9-60 FDC4000 Service Manual 9 Troubleshooting 9.1.62 E0500 RTC backup error [Display] '&CVG'44 4WP/QFG UGVFCVG [Printer] ' &CVG'44 4WP/QFG 5GVFCVGVKOG [Error Description] [Possible Causes] The real time clock IC has not been backed up. (1) The power of the analyzer is being turned off for a long time. (2) The backup switch (S3) on the CPU40 board is OFF. (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Set the date and time using Mode 20. Turn the power off. After a while, turn on the power. If the error still occurs, replace the CPU40 board. NOTE: The clock IC on the CPU40 board is backed up by a lithium battery (rechargeable). Fully charging the battery needs 40 hours, so the board is not fully charged when shipping and the backup will expire about 1 month after shipping. When turning on the power at the first installation, this error will occur depending on the term in warehouse. 9.1.63 E0501 RTC control error [Display] '46%'44 %QPVCEV[QWTFGCNGT [Printer] ' 4WP/QFG 5GVFCVGVKOG %QPVCEV[QWTFGCNGT 9 [Error Description] Data transfer error for the real time clock IC on the CPU40 board [Possible Causes] (1) Malfunction of the CPU40 board === Check Points ======================================================= (1) Set the date and time using Mode 20. Turn the power off. After a while, turn on the power. If the error still occurs, replace the CPU40 board. FDC4000 Service Manual 9-61 9 Troubleshooting 9.1.64 E0509 ADC calibration error [Display] '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' %QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Error Description] Calibration of the A/D converter for CM test reading has not completed. [Possible Causes] (1) Malfunction of the CPU40 board === Check Points ======================================================= (1) Replace the CPU40 board. 9.1.65 E0510 A/D conversion error [Display] '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' %QPVTQNNGT'44 %QPVCEV[QWTFGCNGT 9 [Error Description] A/D conversion for CM test reading has not completed. [Possible Causes] (1) Malfunction of the CPU40 board === Check Points ======================================================= (1) Replace the CPU40 board. 9-62 FDC4000 Service Manual 9 Troubleshooting 9.1.66 E510 RWP reading error [Display] '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' %QPVTQNNGT'44 %QPVCEV[QWTFGCNGT =PO?=FGI? [Error Description] [Possible Causes] RWP readout (A/D) is outside of the range (25800 - 30200). (1) Old light source lamp (2) Malfunction of the CPU40 board (3) Malfunction of the LHA-AMP board === Check Points ======================================================= (1) Replace the lamp. (2) Replace the CPU40 board. (3) Replace the photometer head (including LHA-AMP board). 9 FDC4000 Service Manual 9-63 9 Troubleshooting 9.1.67 E0530 ISE self-test error (gain) [Display] '+5'UGNHVGUV %QPVCEV[QWTFGCNGT [Printer] ' +5'UGNHVGUVICKP'44 /WNHWPEVKQPQH+5'WPKV %QPVCEV[QWTFGCNGT [Error Description] [Possible Causes] A gain setting error has occurred during ISE self-test. (1) Malfunction of the EPM board (2) Malfunction of the CPU40 board (3) Malfunction of the power supply (4) Cable disconnection === Check Points ======================================================= (1) Check the time course. Acceptable voltage range: 196 - 220 (mV) (2) Check the EPM board as follows. Make sure that the voltage between the TP14 and the TP5 (AGND) is within the range (2.500 +/0.0001 [V]) (3) Replace the EPM board. 9 9-64 FDC4000 Service Manual 9 Troubleshooting 9.1.68 E0531 ISE self-test error (offset) [Display] '+5'UGNHVGUV %QPVCEV[QWTFGCNGT [Printer] ' +5'UGNHVGUVQHHUGV'44 /CNHWPEVKQPQH+5'WPKV %QPVCEV[QWTFGCNGT [Error Description] [Possible Causes] An offset error has occurred during ISE self-test. (1) Malfunction of the EPM board (2) Malfunction of the CPU40 board (3) Malfunction of the power supply (4) Cable disconnection === Check Points ======================================================= (1) Check the time course. Acceptable voltage range: -1.0 to 1.0 (mV) (2) Replace the EPM board. 9 FDC4000 Service Manual 9-65 9 Troubleshooting 9.1.69 E0532 A/D conversion error for ISE [Display] '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' %QPVTQNNGT'44 /CNHWPEVKQPQH+5'WPKV %QPVCEV[QWTFGCNGT [Error Description] [Possible Causes] A/D conversion for ISE test reading has not been completed. (1) Malfunction of the EPM board (2) Malfunction of the CPU40 board (3) Cable disconnection === Check Points ======================================================= (1) Replace the EPM board. (2) Replace the CPU40 board. (3) Check the cable connection. 9 9-66 FDC4000 Service Manual 9 Troubleshooting 9.1.70 E0600, E0601, E0602, E0603 Incubator motor error [Display] '+PEWOQVQT'44 5GPUQT QP HCKNWTG '+PEWOQVQT'44 5GPUQT QHH HCKNWTG '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT '+PEWOQVQT'44 %JGEMFKURQUCNDQZ [Printer] ' +PEWDCVQTOQVQT'44 5GPUQT QP HCKNWTG ' +PEWDCVQTOQVQT'44 5GPUQT QHH HCKNWTG ' +PEWDCVQTOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT ' +PEWDCVQTOQVQT'44 %JGEMFKURQUCNDQZ [Error Description] Error code Error description E0600 The incubator HP sensor (PH401) could not detect the interrupter after sending the command during initialization. E0601 The incubator HP sensor (PH401) is still detecting the interrupter after sending the command during initialization. E0602 The software could not control the motor. E0603 The incubator HP sensor (PH401) has detected step-out of the incubator rotation. [Possible Causes] (1) Presence of foreign matter (2) Disposal box is full of slides or tips. (3) Faulty sensor (PH401) (4) Cable disconnection (5) Malfunction of the DRV40 board (6) Faulty incubator motor (M401) (7) Malfunction of the CPU40 board (8) Improper timing belt tension === Check Points ======================================================= (1) Inspect the mechanical interference by turning the incubator by hand or removing the upper incubator. (2) Empty the disposal box. (3) Check the incubator HP sensor (PH401). Check if the PH401 works or not by turning the incubator by hand. Check the sensor function using Mechanical test mode. (4) Check the cable for PH401. (5) Replace the DRV40 board. (6) Replace the incubator motor (M401). (7) Replace the CPU40 board. FDC4000 Service Manual 9-67 9 9 Troubleshooting (8) Check the timing belt tension. 9 9-68 FDC4000 Service Manual 9 Troubleshooting 9.1.71 E0700, E0701, E0702 Probe motor error [Display] '2TQDGOQVQT'44 5GPUQT QP HCKNWTG '2TQDGOQVQT'44 5GPUQT QHH HCKNWTG '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' 2TQDGOQVQT'44 5GPUQT QP HCKNWTG ' 2TQDGOQVQT'44 5GPUQT QHH HCKNWTG ' 2TQDGOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT [Error Description] Error code Error description E0700 The probe HP sensor (PH601) could not detect the interrupter after sending the command during initialization. E0701 The probe HP sensor (PH601) is still detecting the interrupter after sending the command during initialization. E0702 The software could not control the motor. [Possible Causes] (1) Presence of foreign matter (2) Faulty sensor (PH601) (3) Cable disconnection (4) Malfunction of the DRV40 board (5) Faulty probe motor (M601) (6) Malfunction of the CPU40 board (7) Mechanical malfunction === Check Points ======================================================= (1) Check that no foreign matter is present to interfere with the probe vertical operation. (2) Check the probe HP sensor (PH601). Check the sensor function using Mechanical test mode. (3) Check the cable of the probe HP sensor (PH601). (4) Replace the DRV40 board. (5) Replace the probe motor (M601). (6) Replace the CPU40 board. 9 FDC4000 Service Manual 9-69 9 Troubleshooting 9.1.72 E0900 - E0903 Filter motor error [Display] '(KNVGTOQVQT'44 5GPUQT QP HCKNWTG '(KNVGTOQVQT'44 5GPUQT QHH HCKNWTG '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT '(KNVGTOQVQT'44 #DPQTOCNTQVCVKQP [Printer] ' (KNVGTOQVQT'44 5GPUQT QP HCKNWTG ' (KNVGTOQVQT'44 5GPUQT QHH HCKNWTG ' (KNVGTOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT ' (KNVGTOQVQT'44 #DPQTOCNTQVCVKQP [Error Description] Error code Error description E0900 The filter HP sensor (PH501) could not detect the interrupter after sending the command during initialization. E0901 The filter HP sensor (PH501) is still detecting the interrupter after sending the command during initialization. E0902 The software could not control the motor. E0903 The filter HP sensor (PH501) has detected step-out of the filter rotation. [Possible Causes] 9 (1) Mechanical interference for rotation (2) Faulty sensor (PH501) (3) Cable disconnection (4) Malfunction of the DRV40 board (5) Faulty filter motor (M501) (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Turn the filter wheel by hand and check that the filter wheel rotates smoothly without any mechanical contact with other parts. Particularly, check that the interrupter of the filter wheel does not touch the HP sensor (PH501). (2) Check the filter HP sensor (PH501). Check if the PH501 works or not by turning the filter wheel by hand. Check the sensor function using Mechanical test mode. (3) Check the cable of the filter HP sensor (PH501). (4) Replace the DRV40 board. (5) Replace the filter motor (M501). (6) Replace the CPU40 board. 9-70 FDC4000 Service Manual 9 Troubleshooting 9.1.73 E1000 - E1002 Sampler vertical drive error [Display] '8GTVOQVQT'44 5GPUQT QP HCKNWTG '8GTVOQVQT'44 5GPUQT QHH HCKNWTG 'EQPVTQNNGT'44 %QPVCEV[QWTFGCNGT ' 5CORNGTXGTVOQVQT'44 5GPUQT QHH HCKNWTG ' 5CORNGTXGTVOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT [Printer] ' 8GTVOQVQT'44 5GPUQT QP HCKNWTG [Error Description] Error code Error description E1000 The sampler up/down HP sensor (PH203) could not detect the interrupter after sending the command during initialization. E1001 The sampler up/down HP sensor (PH203) is still detecting the interrupter after sending the command during initialization. E1002 The software could not control the motor. [Possible Causes] (1) Presence of foreign matter (2) Faulty sensor (PH203) (3) Cable disconnection (4) Malfunction of the DRV40 board (5) Faulty sampler up/down motor (M202) (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that no foreign matter is present to interfere with the sampler vertical operation. (2) Check the sampler up/down HP sensor (PH203). Check if the PH203 works or not by lifting/lowering the sampler arm by hand. Check the sensor function using Mechanical test mode. (3) Check the cable of the sampler up/down HP sensor (PH203). (4) Replace the DRV40 board. (5) Replace the CPU40 board. (6) Check the timing belt tension. FDC4000 Service Manual 9 9-71 9 Troubleshooting 9.1.74 E1100 - E1103 Sampler rotational drive error [Display] '5RNTQVOV'44 5GPGQT QP HCKNWTG '5RNTQVOV'44 5GPUQT QHH HCKNWTG '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT ' 5CORNGTTQVOQVQT'44 5GPUQT QHH HCKNWTG ' 5CORNGTTQVOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT '5RNTQVOV'44 #DPQTOCNTQVCVKQP [Printer] ' 5CORNGTTQVOQVQT'44 5GPUQT QP HCKNWTG ' 5CORNGTTQVOQVQT'44 #DPQTOCNTQVCVKQP [Error Description] Error code Error description E1100 The sampler rotation HP sensor (PH201) could not detect the interrupter after sending the command during initialization. E1101 The sampler rotation HP sensor (PH201) is still detecting the interrupter after sending the command during initialization. E1102 The software could not control the motor. E1103 The sampler rotation HP sensor (PH201) has detected step-out of the sampler rotation. [Possible Causes] 9 (1) Presence of foreign matter (2) Faulty sensor (PH201) (3) Cable disconnection (4) Malfunction of the DRV40 board (5) Faulty sampler rotation motor (M201) (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that no foreign matter is present to interfere with the sampler rotational operation. (2) Check the sampler rotation HP sensor (PH201). Check if the PH201 works or not by turning the sampler arm by hand. (3) Check the cable of the sampler rotation HP sensor (PH201). (4) Replace the DRV40 board. (5) Replace the CPU40 board. (6) Check the timing belt tension. 9-72 FDC4000 Service Manual 9 Troubleshooting 9.1.75 E1200 - E1204 Syringe motor error [Display] '5[TKPIG'44 5GPGQT QP HCKNWTG '5[TKPIG'44 5GPUQT QHH HCKNWTG '5[TKPIG'44 #DPQTOCNOQXGOGPV '5[TKPIG'44 %QPVCEV[QWTFGCNGT '%QPVTQNNGT'44 %QPVCEV[QWTFGCNGT [Printer] ' 5[TKPIGOQVQT'44 5GPUQT QP HCKNWTG ' 5[TKPIGOQVQT'44 5GPUQT QHH HCKNWTG ' 5[TKPIGOQVQT'44 #DPQTOCNOQXGOGPV ' 5[TKPIGOQVQT'44 #DPQTOCNOQXGOGPV %QPVCEV[QWTFGCNGT ' 5[TKPIGOQVQT'44 (CWNV[EQPVTQNNGT %QPVCEV[QWTFGCNGT [Error Description] Error code Error description E1200 The syringe sensor (PH301) could not detect the interrupter after sending the command during initialization. E1201 The syringe sensor (PH301) is still detecting the interrupter after sending the command during initialization. E1202 The software could not control the motor. E1203 The syringe sensor (PH301) has detected step-out of the syringe operation. E1204 The syringe sensor (PH301) has detected the step-out of the syringe operation again during the recovery process. [Possible Causes] (1) Syringe malfunction (2) Faulty sensor (PH301) (3) Cable disconnection (4) Malfunction of the DRV40 board (5) Faulty syringe motor (M301) (6) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the syringe. Check that the syringe moves smoothly by hand. This inspection should be done for both the syringe unit and the syringe itself. (2) Check the syringe sensor (PH301). Check if the PH301 works or not by sliding the syringe's plunger by hand. Check the sensor function using Mechanical test mode. (3) Check the cable of the syringe sensor (PH301). (4) Replace the DRV40 board. (5) Replace the CPU40 board. FDC4000 Service Manual 9-73 9 9 Troubleshooting (6) Check the timing belt tension. 9 9-74 FDC4000 Service Manual 9 Troubleshooting 9.1.76 E1500 CM temperature not ready [Display] '6GORPQVTGCF[ %JGEMTQQOVGOR [Printer] ' %/VGORPQVTGCF[ %JGEMTQQOVGOR %NGCPCKTHKNVGT [Error Description] The analyzer is not ready to start CM tests after 30 minutes from turning on the power. [Possible Causes] (1) The thermistor in the incubator is disconnected. (2) The environmental temperature is outside of the range. (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the CM incubator temperature and the environmental temperature using Mode 103. (2) Measure the resistance of the thermistor using a circuit tester. Pull out connector CPU3 from the CPU40 board and measure the resistance between pins No.1 and No.2 of the cable side. Temperature: 37°C => Resistance: About 8KΩ Temperature: 25°C => Resistance: About 12.5 KΩ Temperature: 10°C => Resistance: About 23 KΩ (3) Check the CPU40 board. Measure the voltage between the TP28 and AGND using a circuit tester. Reference voltage: Temperature 35°C 37°C 39°C Voltage 0V 2.5V 5.0V 9 FDC4000 Service Manual 9-75 9 Troubleshooting 9.1.77 E1501 CM temperature control error [Display] '6GOREVTN'44 %JGEMTQQOVGOR [Printer] ' %/VGOREQPVTQN'44 %JGEMTQQOVGOR [Error Description] CM incubator temperature is continuously outside of the range (+0.5°C, -0.2°C) for 30 seconds. [Possible Causes] (1) The thermistor in the incubator is disconnected. (2) The environmental temperature is outside of the range. (3) Incubator cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the CM incubator temperature and the environmental temperature using Mode 103. (2) Measure the resistance of the thermistor using a circuit tester. Pull out connector CPU3 from the CPU40 board and measure the resistance between pins No.1 and No.2 of the cable side. Temperature: 37°C => Resistance: About 8KΩ Temperature: 25°C => Resistance: About 12.5 KΩ Temperature: 10°C => Resistance: About 23 KΩ (3) Replace the incubator cable. (4) Check the CPU40 board. Measure the voltage between the TP28 and AGND using a circuit tester. Reference voltage: Temperature 35°C 37°C 39°C Voltage 0V 2.5V 5.0V 9 9-76 FDC4000 Service Manual 9 Troubleshooting 9.1.78 E1502 CM thermistor disconnection [Display] '6GORUGPUQT'44 %JGEMKPEWECDNG [Printer] ' (CWNV[%/VGORUGPUQT %JGEMKPEWECDNG [Error Description] [Possible Causes] The CM thermistor is disconnected. (1) The thermistor in the incubator is disconnected. (2) Incubator cable disconnection (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the CM incubator temperature using Mode 103. (2) Measure the resistance of the thermistor using a circuit tester. Pull out connector CPU3 from the CPU40 board and measure the resistance between pins No.1 and No.2 of the cable side. Temperature: 37°C => Resistance: About 8KΩ Temperature: 25°C => Resistance: About 12.5 KΩ Temperature: 10°C => Resistance: About 23 KΩ (3) Replace the incubator cable. (4) Check the CPU40 board. Measure the voltage between the TP59 and AGND using a circuit tester. If the measured voltage is over 150mV, replace the CPU40 board. * When the voltage is below 150mV, the signal means “Disconnected”. 9 FDC4000 Service Manual 9-77 9 Troubleshooting 9.1.79 E1503 CM heater disconnection [Display] '%/JGCVGT'44 %JGEMKPEWECDNG [Printer] ' (CWNV[%/JGCVGT %JGEMKPEWECDNG [Error Description] [Possible Causes] The CM heater is disconnected. (1) Heater disconnection (2) The safety thermostat (overheat switch) has been activated (circuit disconnected). (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Measure the resistance of the heater using a circuit tester. Pull out connector CPU8 from the CPU40 board and measure the resistance between pins No.3 and No.4 of the cable side. Reference resistance: About 19 KΩ (2) Reset the safety thermostat. (3) Check the cable connection. (4) Replace the incubator cable. (5) Replace the CPU40 board. 9 9-78 FDC4000 Service Manual 9 Troubleshooting 9.1.80 E1504 ISE temperature not ready [Display] '+5'VGOR'44 %JGEMTQQOVGOR [Printer] ' +5'VGORPQVTGCF[ %JGEMTQQOVGOR %NGCPCKTHKNVGT [Error Description] The analyzer is not ready to start ISE tests after 30 minutes from turning on the power. [Possible Causes] (1) The temperature sensor in the ISE incubator is disconnected. (2) The environmental temperature is outside of the range. (3) Cable disconnection (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the ISE incubator temperature and the environmental temperature using Mode 103. (2) Check the ISE temperature sensor. Between the pins No.1 and No.3: 5V Between the pins No.2 and No.3: Temperature x 10mV (3) Check the CPU40 board. Measure the voltage between the U77-10 and AGND using a circuit tester. Reference voltage: Temperature 0°C 20°C 30°C Voltage 0V 200mV 300mV Measure the voltage between the TP37 and AGND using a circuit tester. Measured voltages should be 10 times of above voltages. 9 FDC4000 Service Manual 9-79 9 Troubleshooting 9.1.81 E1510 ISE temperature control error [Display] '+5'VGOR'44 %JGEMTQQOVGOR [Printer] ' +5'VGOREQPVTQN'44 %JGEMTQQOVGOR [Error Description] ISE incubator temperature is continuously outside of the range (+5°C, -2°C) for 30 seconds. [Possible Causes] (1) The temperature sensor in the ISE incubator is disconnected. (2) The environmental temperature is outside of the range. (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the ISE incubator temperature and the environmental temperature using Mode 103. (2) Check the ISE temperature sensor. Between the pins No.1 and No.3: 5V Between the pins No.2 and No.3: Temperature x 10mV (3) Check the cable connection. (4) Check the CPU40 board. Measure the voltage between the U77-10 and AGND using a circuit tester. Reference voltage: Temperature 0°C 20°C 30°C Voltage 0V 200mV 300mV Measure the voltage between the TP37 and AGND using a circuit tester. Measured voltages should be 10 times of above voltages. 9 9-80 FDC4000 Service Manual 9 Troubleshooting 9.1.82 E1511 ISE temperature sensor disconnection [Display] '6GORUGPUQT'44 %QPVCEV[QWTFGCNGT [Printer] ' +5'VGORUGPUQT'44 [Error Description] [Possible Causes] The ISE temperature sensor is disconnected. (1) The temperature sensor in the ISE incubator is disconnected. (2) Cable disconnection (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the ISE incubator temperature and the environmental temperature using Mode 103. (2) Check the ISE temperature sensor. Between the pins No.1 and No.3: 5V Between the pins No.2 and No.3: Temperature x 10mV (3) Check the cable connection. (4) Check the CPU40 board. Measure the voltage between the U77-10 and AGND using a circuit tester. Reference voltage: Temperature 0°C 20°C 30°C Voltage 0V 200mV 300mV Measure the voltage between the TP37 and AGND using a circuit tester. Measured voltages should be 10 times of above voltages. 9 FDC4000 Service Manual 9-81 9 Troubleshooting 9.1.83 E1512, E1513 Abnormal high temperature [Display] '6GOREVTN'44 %JGEMTQQOVGOR '+5'VGOR'44 %JGEMTQQOVGOR [Printer] ' %/VGOREQPVTQN'44 %JGEMTQQOVGOR 6WTPRQYGT59QHH ' +5'VGOREQPVTQN'44 %JGEMTQQOVGOR 6WTPRQYGT59QHH [Error Description] Error code Error description E1512 CM incubator temperature has exceeded 38.2°C. E1513 ISE incubator temperature has exceeded 40.9°C. [Possible Causes] (1) The environmental temperature is outside of the range. (2) The thermistor or the temperature sensor is disconnected. (3) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check the CM & ISE incubator temperatures and the environmental temperature using Mode 103. (2) Check the thermistor and its cable. => Refer to Section E1502. (3) Check the ISE temperature sensor and its cable. => Refer to Section E1511. (4) Replace the CPU40 board. 9 9-82 FDC4000 Service Manual 9 Troubleshooting 9.1.84 E1514 Temperature warning [Display] '6GORYCTPKPI 6GUVUECPPQVTWP [Printer] ' 6GORYCTPKPI 6GUVUECPPQVTWP 2NGCUGYCKVCOQOGPV [Error Description] When starting the measurements, the incubator temperature is not ready. [Possible Causes] (1) The incubator temperature (CM or ISE) is not ready yet. === Check Points ======================================================= (1) Wait a moment until the temperature becomes ready. 9.1.85 E1521 High environmental temperature [Display] '4QQOVGOR'44 %JGEMTQQOVGOR [Printer] ' 4QQOVGOR'44 %JGEMTQQOVGOR %NGCPCKTHKNVGT [Error Description] [Possible Causes] Environmental temperature has exceeded 43°C. (1) The environmental temperature is outside of the range. (2) Faulty environmental temperature sensor === Check Points ======================================================= - 9 Environmental temperature sensorOn the KEY40 board (1) Check the environmental temperature using Mode 103. (2) Replace the KEY40 board which has the environmental temperature sensor. FDC4000 Service Manual 9-83 9 Troubleshooting 9.1.86 E30xx Memory error [Display] 'ZZ/GOQT['44 %QPVCEV[QWTFGCNGT [Printer] 'TTQTV[RG &CVCPCOGYJGTGVJGOGOQT[GTTQTQEEWTGF ' $CEMWR'44=RCTCOGVGT? %QPVCEV[QWTFGCNGT [Error Description] Error code 9 9-84 Memory crash area Error type Example printout E3000 Specific parameters Unrecoverable error E3001 Setting info Unrecoverable error E3002 Correlation coefficients Unrecoverable error E3003 Reference intervals Unrecoverable error E3004 Reference interval names Unrecoverable error E3005 Calibration data Unrecoverable error E3006 Revision info Unrecoverable error E3007 QC info (CM) Unrecoverable error E3008 QC info (Immuno) Unrecoverable error E3009 QC info (ISE) Unrecoverable error E300A PF info Unrecoverable error E300B Reprint data Unrecoverable error E300C Time course Unrecoverable error E300D Error logs Unrecoverable error E300E Message data Unrecoverable error <Example> E3000 (Specific parameters) ' $CEMWR'44=RCTCOGVGT? %QPVCEV[QWTFGCNGT FDC4000 Service Manual 9 Troubleshooting Error code Memory crash area Error type E3020 Specific parameters Master crash error E3021 Setting info Master crash error E3022 Correlation coefficients Master crash error E3023 Reference intervals Master crash error E3024 Reference interval names Master crash error E3025 Calibration data Master crash error E3026 Revision info Master crash error E3027 QC info (CM) Master crash error E3028 QC info (Immuno) Master crash error E3029 QC info (ISE) Master crash error E302A PF info Master crash error E302B Reprint data Master crash error E302C Time course Master crash error E302D Error logs Master crash error E302E Message data Master crash error Example printout <Example> E3025 (Analyzer calibration data) ' 5CXG'44=EQGHH LM ? %QPVCEV[QWTFGCNGT 9 FDC4000 Service Manual 9-85 9 Troubleshooting Error code E3040 E3041 E3042 E3043 E3044 E3045 E3046 E3047 E3048 E3049 E304A E304B E304C E304D E304E E304F 9 9-86 Memory crash area Error type Specific parameters FROM clear error / writing error Setting info FROM clear error / writing error Correlation coefficients FROM clear error / writing error Reference intervals FROM clear error / writing error Reference interval names FROM clear error / writing error Calibration data FROM clear error / writing error Revision info FROM clear error / writing error QC info (CM) FROM clear error / writing error QC info (Immuno) FROM clear error / writing error QC info (ISE) FROM clear error / writing error PF info FROM clear error / writing error Reprint data FROM clear error / writing error Time course FROM clear error / writing error Error logs FROM clear error / writing error Message data FROM clear error / writing error Calibration data FROM clear error / writing error Example printout <Example> E3048 (QC info) ' 5CXG'44=3%KPHQ? %QPVCEV[QWTFGCNGT FDC4000 Service Manual 9 Troubleshooting Error code Memory crash area Error type E3060 Specific parameters SRAM writing error E3061 Setting info SRAM writing error E3062 Correlation coefficients SRAM writing error E3063 Reference intervals SRAM writing error E3064 Reference interval names SRAM writing error E3065 Calibration data SRAM writing error E3066 Revision info SRAM writing error E3067 QC info (CM) SRAM writing error E3068 QC info (Immuno) SRAM writing error E3069 QC info (ISE) SRAM writing error E306A PF info SRAM writing error E306B Reprint data SRAM writing error E306C Time course SRAM writing error E306D Error logs SRAM writing error E306E Message data SRAM writing error Example printout <Example> E306D (Error logs) '& 5CXG'44='44NQI? %QPVCEV[QWTFGCNGT [Possible Causes] (1) The power was shut down while storing the data. (Master crash error) (2) The backup switch (S3) on the CPU40 board is OFF. (Master crash error) (3) Faulty FROM or SRAM (4) Malfunction of the CPU40 board === Check Points ======================================================= (1) Check that the backup switch (S3) on the CPU40 board is ON. (2) Turn the power off and on, and then check that the designated error data is correct. (The analyzer boots up by using the backup data, so that revised data is not saved. So, checking the setting data is necessary.) (3) Replace the CPU40 board. 9 FDC4000 Service Manual 9-87 9 Troubleshooting 9.1.87 E4000 - E4010 QC card error [Display] '3%'44 4GCF3%ECTFCICKP '3%'44 4GCFCHVGTVGUVGPF 5COGOGUUCIGHQT'' '0Q3%V[RGKPHQ '3%'44 5CXG'44 [Printer] ' 0Q3%V[RGKPHQ 4GCF3%ECTF VQKPRWVV[RGKPHQ ' 0QCTGCHQTUCXKPI3% %QPVCEV[QWTFGCNGT [Error Description] Error code 9 Error description E4000 Overrun error Or, the QC card data is outside of the range. E4001 Three QC cards for a test name have been read during test processing. E4002 QC card reading has not been completed within 3 seconds. E4003 Failed to detect the starting point of the card data. E4004 QC card data is undefined (not between "0x1b" and "0x7f"). E4006 Horizontal parity error for the 1st track E4007 Horizontal parity error for the 2nd track E4008 Vertical parity error for the 1st track E4009 Vertical parity error for the 2nd track E4010 When reading a QC card for panel slide, no corresponding type information are memorized. E4011 There are no memory area for saving QC information of the panel slide. [Possible Causes] (1) Scanning speed is not proper (not constant, too high, or too low). (2) Dirty head of the card reader (3) Faulty card reader (4) Malfunction of the QC card === Check Points ======================================================= (1) Read the QC card again. (2) Replace the card reader. (3) Read a QC card which has the corresponding type information of the slide. (E4010) 9-88 FDC4000 Service Manual 9 Troubleshooting 9.1.88 W5000, E5100 Printer error [Display] 92TKPVGT'44 5GVRCRGT 92TKPVGT'44 .QYGTRTKPVGTJGCF '2TKPVGT'44 %QPVCEV[QWTFGCNGT [Error Description] Error code Error description W5110 No paper W5120 The printer head has not been lifted down (locked). E5100 The printer is not ready to print. (Cable disconnection or malfunction of the printer) [Possible Causes] (1) W5110: No paper (2) W5120: The printer head has not been lifted down (locked). (3) E5100: Paper jam in the printer Cable disconnection Faulty printer === Check Points ======================================================= For E5100: (1) Remove the jamming paper. (2) Check the cable connection. (3) Replace the printer. 9 FDC4000 Service Manual 9-89 9 Troubleshooting 9.1.89 E5000 - E5034 CF card error [Display] 'ZZ(KNG'44 %QPVCEV[QWTFGCNGT [Printer] 'TTQTV[RG &CVCPCOGYJGTGVJGHKNGGTTQTQEEWTGF ' 4GCF'44=RCTCOGVGT? %QPVCEV[QWTFGCNGT [Error Description] Error code 9 9-90 Memory crash area Error type E5000 Specific parameters Read error E5001 Setting info Read error E5002 Correlation coefficients Read error E5003 Reference intervals Read error E5004 Reference interval names Read error E5005 Calibration data Read error E5006 Revision info Read error E5007 QC info (CM) Read error E5008 QC info (Immuno) Read error E5009 QC info (ISE) Read error E5010 PF info Read error E5011 Reprint data Read error E5012 Time course Read error E5013 Error logs Read error E5014 Message data Read error Example printout <Example> E5000 (Specific parameters) ' 4GCF'44=RCTCOGVGT? %QPVCEV[QWTFGCNGT FDC4000 Service Manual 9 Troubleshooting Error code Memory crash area Error type E5020 Specific parameters Save error E5021 Setting info Save error E5022 Correlation coefficients Save error E5023 Reference intervals Save error E5024 Reference interval names Save error E5025 Calibration data Save error E5026 Revision info Save error E5027 QC info (CM) Save error E5028 QC info (Immuno) Save error E5029 QC info (ISE) Save error E5030 PF info Save error E5031 Reprint data Save error E5032 Time course Save error E5033 Error logs Save error E5034 Message data Save error E5035 Calibration data Save error Example printout <Example> E5025 (Analyzer calibration data) ' 5CXG'44=EQGHH LM ? %QPVCEV[QWTFGCNGT [Possible Causes] (1) Faulty CF card (2) Malfunction of the CPU40 board === Check Points ======================================================= (1) Replace the CF card. (Use designated (recommended) CF cards.) (2) Replace the CPU40 board. 9 FDC4000 Service Manual 9-91 9 Troubleshooting 9.1.90 E5050 No CF card has been set. [Display] '0Q%(ECTF [Printer] ' 0Q%(ECTF [Error Description] When reading / writing a CF card, the analyzer has detected no CF card inserted into the slot. [Possible Causes] (1) No CF card has been inserted into the slot. (2) Faulty CF card === Check Points ======================================================= (1) Insert a CF card into the slot. (2) Replace the CF card. 9 9-92 FDC4000 Service Manual 9 Troubleshooting 9.1.91 E5201 - E5204 Malfunction of the slide reader [Display] 'Z.QVEVTN'44 %QPVCEV[QWTFGCNGT [Printer] ' .QVEQOUGVWR'44 6WTPRQYGT59QHHQP VQTGUVCTV ' .QVEQPVTQN'44 VGUV %QPVCEV[QWTFGCNGT ' .QVEQPVTQN'44 TGCF %QPVCEV[QWTFGCNGT ' .QVEQPVTQN'44 VGUV %QPVCEV[QWTFGCNGT [Error Description] Error code E5201 E5202 E5203 E5204 Error name Error description Initialization timeout During the initialization of the slide reader, there is no reply from the unit. Self-test timeout During the self-test of the slide reader, there is no reply from the unit. Read timeout During the slide reading by the slide reader, there is no reply from the unit. Self-test error The slide reader unit could not adjust the light intensity in the self-test process. [Possible Causes] (1) Malfunction of the slide reader (CCD40 board) (2) Malfunction of the CPU40 board (3) Cable disconnection === Check Points ======================================================= (1) Replace the CCD40 board. (2) Replace the CPU40 board. (3) Check the cable connection. FDC4000 Service Manual 9 9-93 9 Troubleshooting 9.1.92 E5500 Spotting control error [Display] '%QPVTQN'44 %QPVCEV[QWTFGCNGT [Printer] ' %QPVTQN'44 %QPVCEV[QWTFGCNGT [Error Description] The time between spotting and inserting into the incubator is out of the range (2.3 sec. to 5.0 sec.). [Possible Causes] (1) Malfunction of the CPU40 board. === Check Points ======================================================= (1) Perform the troubleshooting for the error which occurred before this error. (2) Turn the power off and on to restart. (3) Replace the CPU40 board. 9.1.93 E5510 Slide position error [Display] '+PEWDCVQT'44 %QPVCEV[QWTFGCNGT [Printer] ' 2TGUUWTGRNCVG'44 %QPVCEV[QWTFGCNGT [Error Description] 9 The slide is off position in the incubator during measurement from the analysis of the time course. (The analyzer determines photometric measurement error for the slide and the test result will have a indication “E”. The analyzer continues the measurement process.) [Possible Causes] (1) Loosened thumb screws of the upper incubator block. (2) Sliding movement (up/down) of the pressure plates of the upper incubator is not smooth. (3) Malfunction of the leaf springs in the upper incubator cover === Check Points ======================================================= (1) Tighten the three thumb screws of the upper incubator. (2) Check that the sliding movement (up/down) of the pressure plates is smooth. (3) Check that the leaf springs are properly set in the upper incubator cover. (4) Replace the upper incubator block. 9-94 FDC4000 Service Manual 9 Troubleshooting [Description] The slide position error is detected by the trend analysis of the obtained time course after the measurement. The time course curve differs by the test name. So, the parameters [time course trend (up/ down), OD change, a number of OD changes, etc.] for each test are prepared to detect the error. <Example: Malfunction detected> 㪫㫀㫄㪼㩷㪺㫆㫌㫉㫊㪼㩷㩿㪞㪣㪬㪀㪑㩷㪤㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅㩷㪻㪼㫋㪼㪺㫋㪼㪻 㪈㪅㪋 㪈㪅㪉 㪈 㪤㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅 㪻㪼㫋㪼㪺㫋㪼㪻 㪇㪅㪏 㪥㫆㫉㫄㪸㫃 㪇㪅㪍 㪤㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅 㪻㪼㫋㪼㪺㫋㪼㪻 㪥㫆㫉㫄㪸㫃 㪇㪅㪋 㪇㪅㪉 㪇 㪇 㪌 㪈㪇 㪈㪌 㪉㪇 㪉㪌 㪊㪇 The analyzer performs the detection by counting how many wrong OD changes are existed, compared with the normal time course. So, in some cases, the analyzer cannot detect the error. <Example: Malfunction NOT detected> 㪫㫀㫄㪼㩷㪺㫆㫌㫉㫊㪼㩷㩿㪘㪣㪧㪀㪑㩷㪤㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅㩷㩷㫅㫆㫋㩷㪻㪼㫋㪼㪺㫋㪼㪻 㪇㪅㪏 㪇㪅㪎 㪤㫌㫃㪽㫌㫅㪺㫋㫀㫆㫅㩷 㫅㫆㫋㩷㪻㪼㫋㪼㪺㫋㪼㪻 㪇㪅㪍 㪥㫆㫉㫄㪸㫃 㪤㫌㫃㪽㫌㫅㪺㫋㫀㫆㫅 㪻㪼㫋㪼㪺㫋㪼㪻 㪇㪅㪌 9 㪇㪅㪋 㪇㪅㪊 㪇 FDC4000 Service Manual 㪌 㪈㪇 㪈㪌 㪉㪇 㪉㪌 㪊㪇 9-95 9 Troubleshooting 9.1.94 E9999 Impossible to continue testing [Display] ' 6WTPRQYGT59QHHQP [Printer] ' 6WTPRQYGT59QHHQP VQTGUVCTV [Error Description] A critical error has occurred (testing cannot continue). [Possible Causes] (1) Malfunction of the CPU40 board === Check Points ======================================================= (1) Perform the troubleshooting for the error which occurred before this error. (2) Turn the power off and on to restart. (3) Replace the CPU40 board. 9 9-96 FDC4000 Service Manual 9 Troubleshooting 9.1.95 EFF01 - EFF0F Calculation error [Display] '((Z%CNEWNCVKQP'44 %QPVCEV[QWTFGCNGT [Printer] '((Z %CNEWNCVKQP'44 %QPVCEV[QWTFGCNGT [Error Description] [Possible Causes] Calculation error has occurred when calculating the analyte concentration. (1) Slide transfer error (2) Un-spotted slide (3) Abnormal sample === Check Points ======================================================= (1) Perform the troubleshooting for the error which occurred before this error. 9 FDC4000 Service Manual 9-97 9 Troubleshooting 9.2 9.2.1 Others Cover Lock Release (1) When the sampler cover is locked and you cannot open the cover, the release hole is prepared. Remove the disposal box and insert a thin rod (a toothpick or a micro screwdriver) into the hole to give a slight press. 9 9-98 FDC4000 Service Manual 10 Slide Information 10 Slide Information Slide Code Slide Filter Wavelength [nm] Sample Volume [µl] Test Code Sample Code GLU - P 10 50 505 10 BUN - P 11 50 625 10 UA - P 13 50 650 10 TCHO - P 14 50 505 10 NH3 - P 15 50 600 10 TG - P 16 50 600 10 CRE - P 17 50 600 10 TP - P 18 50 540 10 ALB - P 20 50 625 10 TBIL - P 21 50 540 10 Ca - P 23 50 625 10 IP - P 24 50 650 10 DBIL - P 25 50 577 10 HDL-C - P 26 50 600 10 Mg - P 28 50 650 10 GGT - P 30 50 400 10 GOT/AST - P 31 50 650 10 GPT/ALT - P 32 50 650 10 CPK - P 33 50 540 10 LDH - P 34 50 540 10 ALP - P 35 50 400 10 LAP - P 37 50 400 10 CKMB - P 38 50 540 10 CHE - P 39 50 650 10 AMYL - P 43 50 400 10 Na 91 00 - 50 K 92 00 - 50 Cl 93 00 - 50 CRP - S 51 50 650 10 Hb - W 12 00 540 10 NH3 - W 15 00 600 10 10 NOTE: Slide codes are subject to change without notice. Slide codes are printed out on the slide’s outer package. FDC4000 Service Manual 10-1 10 Slide Information 10 10-2 FDC4000 Service Manual 26-30, Nishiazabu 2-chome, Minato-ku, Tokyo 106-8620, Japan PARTS LIST FUJI DRI-CHEM CLINICAL CHEMISTRY ANALYZER FUJI DRIDRI-CHEM CHEM 4000i FUJI DRIDRI-CHEM CHEM 4000ie Upper Cover 1 Under Frame 2 Base 3 Transfer/ISE Unit 4 Sampler 5 Photometer 6 Cylinder 7 Lamp Unit 8 Accessories 9 Optional parts 10 Part No. Index List 11 Third Edition Z73N0111B April 2009 CAUTION: The Parts List is intended for exclusive use by technical service personnel authorized by FUJIFILM Corporation. (including personnel to whom technical servicing is commissioned by FUJIFILM Corporation.). To avoid possible electric shock or other personal injury, anyone else but the authorized technical service personnel must in no case be allowed to attempt to gain access to the interior of the equipment by screwing the cover or other parts off, using the Parts List as a guide. Use of the Parts List by anyone else but the authorized technical service personnel should be forbidden. 1 The necessary replacement part should be sought from the exploded drawing of the pertinent section of the instrument. 2 The exploded drawing numbers correspond to the drawing reference numbers in the parts lists. From the parts lists, record the part number and name, but also the instrument name and serial number should be provided. NOTE: Improved part substitutions will in certain instances result in part number changes without prior notice. NOTE: The numbers indicated in the parts list "Qty" column indicate the quantity of the designated part used in each unit. FDC4000 Parts List 2009.4 1 1 S4x10 M4x10 2 N4x10 M4x10 3 T4x10 M4x10 4 STa4x10 M4x10 5 NTa4x10 M4x10 6 TTa4x10 M4x10 7 PK4x10 M4x10 8 PT4x10 M4x10 9 PH4x10 M4x10 10 FK4x5 M4x5 11 FT4x5 M4x5 12 FH4x5 M4x5 13 R4x10 M4x10 14 Q4x10 M4x10 15 Na4 M4 16 W4 M4 17 SW4 M4 18 E4 Nominal diam.4 19 CG4 Nominal diam.4 20 CA4 Nominal diam.4 21 CP4 Nominal diam.4 22 KP4 Nominal diam.4 23 A4x10 M4x10,with tlat and washer 24 B4x10 M4x10, with tlat and spring washer D4x8 M4x8 1. 2. 5. 4. 8. 9. 10. 11. 12. 13. 14. 17. 20. 26 Screw,TP,hex head 27 28 Setscrew,hex socket head W point 29 30 2 23. 24. 26. 27. Screw,TP, hex head Refer to attached table 29. 15. 18. 21. Refer to attached table 19. 22. 25. 28. Setscrew hex socket head (W point) 30. Refer to attached table PLX3x6 Screw,Tapping hex head WP 6. 7. 16. 25 3. M3x6 Screw,Tapping hex head (WP) Refer to attached table FDC4000 Parts List 2009.4 FDC4000 Parts List 2009.4 3 INDEX INDEX 1 INDEX Upper Cover INDEX 4 7 INDEX 10 Optional parts INDEX INDEX 8 INDEX 6 Photometer INDEX Lamp Unit 9 Accessories 11 FDC4000 PARTS NO. INDEX LIST NOTE: Remember that the exploded view Parts are shaded FDC4000 Parts List 2009.4 5 3 Base Sampler Cylinder INDEX INDEX Under Frame Transfer/ISE Unit INDEX 2 can only be supplied as unitized blocks. 5 1 FDC4000 Upper Cover Na4 Na4 4 W4 1 W4 3 5 10 6 7 11 B2x6 BR3x6 TP3x6 9 TP3x6 1 8 B2x4 3 BR3x6 E3 2 12 E3 21 2 E3 13 TP3x6 20 E3 4 TaWP3x8 TP3x6 14 BR3x8 TP3x6 TP3x6 TaWP3x8 5 TP3x6 BR3x8 15 TP3x6 18 17 BR3x8 16 19 6 FDC4000 Parts List 2009.4 1 FDC4000 Ref. No. Part No. Part Name Qty 1 113Y100186 PCB assembly 1 16 350Y100566B Cover 1 2 350N100591A Cover 1 17 405N100646A Label 1 ie 17 405N101123 Label 1 2009 expression & up 3 347N100181A Spacer 1 4 388S1089 Spring 2 Remarks KEY40 board Ref. No. Part No. Part Name Qty Remarks It is compatible 5 113Y100177 PCB assembly 5 6 347N100188 Spacer 4 PHA01 board 17 405N100558B Label 1 i 7 608Y100001B Display 1 17 405N101125 Label 1 2009 expression & up 8 316S1197 Tie, wire 12 It is compatible 9 145S0013 Printer 1 i ie 10 350Y100568A Cover 1 Printer cover 18 350N100593C Cover 1 10 350N100600C Cover 1 2009 expression & up 19 376N100020B Filter, mechanical 1 It is compatible 20 363N100429D Guide 1 Printer cover 21 350Y100567B Cover 1 1 11 350N100602A Cover 1 12 350Y100584A Cover 1 It is compatible ① 350Y100877 Cover 12 350N100685A Cover 1 2009 expression & up ② 350Y100878 Cover 1 It is compatible ③ 350Y100879A Cover 1 4 ④ 350Y100880 Cover 1 ⑤ 358Y100050A Box 1 13 322N100184 Plain bearing 14 363N100428C Guide 1 15 347N100198A Plate, adjustment 4 FDC4000 Parts List 2009.4 Air filter 7 2 FDC4000 Under Frame TP3x6 BR3x6 3 4 TP3x6 BR3x6 26 2 16 5 6 7 TP3x6 TP3x6 8 1 TP4x8 15 9 17 BR3x6 12 11 13 TP3x6 25 TP3x6 14 23 18 TP3x6 TP3x6 21 BR3x30 TP3x6 TP3x6 TP3x6 TP3x6 TP3x6 TP3x6 TP3x6 BR3x6 TP3x6 BR3x10 BR3x6 19 20 8 FDC4000 Parts List 2009.4 2 FDC4000 Ref. No. Part No. Part Name Qty Remarks Ref. No. Part No. Part Name Qty 1 316S1033 Clamp, cord 6 14 316S1140 Clamp, cord 29 2 316S0258 Strap 26 15 113Y100172B PCB assembly 1 16 316S0258 Strap 26 17 125H10007B Power supply 1 3 113Y100171G PCB assembly 1 4 316S1197 Tie, wire 12 CPU40 board 4 316S1198 Tie, wire 9 18 136Y101242D Cable 1 5 318S1195 Bushing 2 2009 expression & up 19 340N100041B Handle 2 6 316S1027 Clamp, cord 12 20 367N100021A Foot 4 7 316S1027 Clamp, cord 12 21 316S1140 Clamp, cord 29 8 318S1195 Bushing 2 22 364S0027 Guard 2 23 119S0042A Fan 1 24 316S1033 Clamp, cord 7 Remarks USB40 board Up to 2008 expression 9 316S0256 Strap 6 10 347S0506 Spacer 5 11 316S0256 Strap 6 25 364S0003 Guard 1 12 316S0258 Strap 26 26 316S1140 Strap 29 2009 expression & up 13 316S0174 Strap 3 FDC4000 Parts List 2009.4 Up to 2008 expression Up to 2008 expression 9 3 FDC4000 Base 1 BR3x10 2 1 TP3x6 4 TP3x6 TP3x6 44 BR3x10 BR3x6 6 BR3x6 3 10 5 11 9 8 7 W3 12 48 13 TP3x6 TP3x6 TP3x6 42 41 40 19 18 14 17 E2 50 15 BR3x6 TP3x6 35 TP3x6 20 BR3x6 16 TP3x6 37 47 TP3x6 BR3x6 BR3x6 BR3x6 BR3x6 BR3x6 E1.5 21 A3x6 32 31 22 33 TP3x6 BR3x10 BR3x12 30 46 4 45 5 3 24 29 BR3x6 25 TP3x6 TP3x6 28 TP3x6 23 BR3x6 49 10 FDC4000 Parts List 2009.4 3 FDC4000 Ref. No. Part No. Part Name Qty Remarks EPM board Ref. No. Part No. Part Name Qty 1 113Y100192B PCB assembly 1 30 363N100425G Guide 1 2 316S0256 Strap 15 Up to 2008 expression 31 107Y100008A Solenoid 1 Remarks 3 147N100001 Magnetic head 1 32 316S0258 Strap 26 4 356N101590C Bracket 1 33 113Y100178 PCB assembly 7 4 356N101284 Bracket 1 2009 expression & up 34 347S0506 Spacer 5 Up to 2008 expression It is incompatible 35 113Y100207 PCB assembly 1 LED40C board PH801 sensor 36 309S0157 Washer, plastic 2 Up to 2008 expression 1 PHC02 board 5 146Y100011 Photo interrupter 6 347S0890 Spacer 1 37 341N100113A Arm 1 7 309S0220003 Washer, spring 5 37 341N100223 Arm 1 8 316S1197 Tie, wire 12 8 316S1198 Tie, wire 9 2009 expression & up 38 318S1261 Bushing 2 Up to 2008 expression It is compatible 39 316S2027 Strap 4 Up to 2008 expression 9 316S1140 Clamp, cord 29 40 316S2027 Strap 3 10 316S0258 Strap 26 41 309S0155 Washer, plastic 5 11 316S1033 Clamp, cord 6 42 128S1050 Microswitch 1 12 316S1140 Clamp, cord 29 43 316S0256 Strap 15 Up to 2008 expression 13 316S1027 Clamp, cord 12 44 394F0112 Magnet 1 14 316S1027 Clamp, cord 12 44 394N100007 Magnet 1 2009 expression & up 15 316F7057 Strap 1 16 316S2027 Strap 3 45 113Y100174B PCB assembly 1 CCD40 board 17 316S0258 Strap 26 46 113Y100175 PCB assembly 1 LED40 board 18 605N100013 Plate, glass 1 47 309S1055 Washer, plastic 5 2009 expression & up 19 316S0174 Strap 3 20 113Y100179 PCB assembly 1 48 316S0258 Strap 26 2009 expression & up 21 316S1140 Clamp, cord 28 22 316S0256 Strap 15 49 347S0908 Spacer 2 23 113Y100173A PCB assembly 1 DRV40 board 50 388N100595 Spring, tension 1 24 388N100250E Spring, torsion 3 24 388N100401 Spring, torsion 3 2009 expression & up ① 136Y101706 Cable 1 It is incompatible ② 350Y100893 Cover 1 25 366N100018B Actuator 3 ③ 356Y100658 Bracket 1 26 362N100114C Support 1 Up to 2008 expression ③ 356Y100850 Bracket 1 27 356N101610C Bracket 1 Up to 2008 expression RAC40 board 28 113Y100176 PCB assembly 1 29 356N101608D Bracket 1 29 356N103189A Bracket 1 2009 expression & up It is incompatible MSA01 switch It is compatible It is compatible PH803 board It is compatible 2009 expression & up 2009 expression & up With washer and nut 2009 expression & up It is incompatible ④ 852Y100024A Reader unit 1 ⑤ 363Y100249 Guide 1 2009 expression & up It is compatible It is incompatible FDC4000 Parts List 2009.4 11 4 FDC4000 Transfer/ISE Unit 12 3 11 13 10 1 9 S2.6x5 S2x5 1 4 14 B3x15 A3x25 A3x6 7 8 5 A3x6 6 15 2 B3x15 6 A3x6 A3x6 TP3x6 4 3 2 BR3x8 TP3x6 16 E3 17 WP3x5 7 A3x6 TP3x6 N2.6x5 A3x6 18 E5 A3x6 19 A3x6 20 26 S3x6 BR3x16 24 TP3x6 BR3x18 E5 WP3x5 INDEX 04-1 21 23 22 12 27 FDC4000 Parts List 2009.4 4 FDC4000 Ref. No. Part No. Part Name Qty Remarks Ref. No. Part No. Part Name Qty 19 113Y100178 PCB assembly 7 PH103 board 20 113Y100177 PCB assembly 5 PH101 board 1 144S0001 Electrode 6 2 113Y100178 PCB assembly 7 3 316S0256 Strap 6 21 324F3198B Timing belt pulley 1 4 316S0258 Strap 26 22 118Y100051A Motor 1 22 118N100052A Motor 1 PH601 board Up to 2008 expression Remarks Slide transfer motor 5 319N100571A Shaft 1 6 322SY357 Plain bearing 10 It is compatible 7 319F3437 Shaft 1 Slide transfer motor 7 319N101245 Shaft 1 2009 expression & up 23 356Y100389B Bracket 1 It is compatible 23 356Y100648 Bracket 1 10 2009 expression & up 2009 expression & up 8 367F0248 Foot 2 9 350F5595 Cover 1 24 322SY357 Plain bearing Spring, 1 25 309S0220003 Washer, spring 5 Up to 2008 expression 26 898H10113A Electrode kit 1 50pcs 27 136Y101831 Cable 1 2009 expression & up 1 10 388S1454 It is incompatible compression 11 346F2806B Plate, auxiliary 1 12 350F5597 Cover 1 13 350F5596 Cover 1 ① 357Y100219 Mount 14 346F1522A Plate, auxiliary 1 ② 118Y100095 Motor 1 15 394F0031 Magnet 4 ③ 350Y100882 Cover 1 16 346F1514 Plate, auxiliary 2 16 346N100362 Plate, auxiliary 2 17 323F3128 Timing belt 1 18 113Y100178 PCB assembly 7 FDC4000 Parts List 2009.4 ④ 350Y100883 Cover 1 2009 expression & up ⑤ 357Y100220 Mount 1 It is compatible ⑥ 368Y100028 Mount 1 ⑦ 368Y100029 Mount 1 PH102 board 13 5 FDC4000 Sampler 1 2 3 B3x8 B3x8 B2.6x6 1 B2.6x6 B2.6x6 4 A3x6 B2.6x6 A3x6 5 6 7 8 B2.6x10 4 2 5 9 10 B3x6 A3x6 11 13 TP3x6 BR3x6 WP3x5 14 12 25 27 BR3x12 15 TP3x6 16 WP3x5 TP3x6 TP3x6 BR3x16 24 23 BR3x16 E5 17 E5 TP3x6 20 BR3x12 TP3x6 BR3x18 22 21 18 3 WP3x5 19 26 14 FDC4000 Parts List 2009.4 5 FDC4000 Ref. No. Part No. Part Name Qty Remarks Sampler arm cover Ref. No. Part No. Part Name Qty Remarks 1 350F5593 Cover 1 2 385H0007 Tube 1 It is compatible 3 350F3806C Cover 1 Sampler rotation 4 350F3807B Cover 1 motor 5 322FY093 Bearing 2 20 322SY357 Plain bearing 6 322SF018 Bearing 2 21 323F3056 Timing belt 1 7 319F3432 Shaft 2 22 356Y100394B Bracket 1 8 319H0019C Shaft 1 22 356Y100649 Bracket 1 9 322F2033 Plain bearing 2 10 322SF018 Bearing 2 23 322SP204 Plain bearing 2 11 360F0449B Housing 1 24 322SP212 Plain bearing 4 12 324F3099A Timing belt pulley 4 24 322SY357 Plain bearing 4 13 113Y100178 PCB assembly 7 14 136Y101280 Cable 1 15 118Y100053B Motor 1 15 118N100052A Motor 1 19 118N100052A Motor 1 2009 expression & up 10 2009 expression & up It is incompatible PH201,PH202 board 2009 expression & up It is incompatible 25 323F3057 Timing belt 1 Sampler vertical motor 26 136Y101829A Cable 1 2009 expression & up 2009 expression & up 27 136Y101828 Cable 1 2009 expression & up ① 371Y100020A Nozzle 1 Sampler nozzle It is compatible Sampler vertical motor 16 324F3099A Timing belt pulley 4 ② 322Y100008B Bearing 2 17 356Y100395B Bracket 1 ③ 375H0005 Seal 1 17 356Y100650 Bracket 1 2009 expression & up ④ 350Y100884 Cover 1 It is incompatible ⑤ 113Y100280 PCB assembly 1 18 324F3098A Timing belt pulley 1 19 118Y100052A Motor 1 O-ring 4pcs PH203 board Sampler rotation motor FDC4000 Parts List 2009.4 15 6 FDC4000 Photometer 1 4 11 N3x30 2 A3x6 1 12 3 2 3 B3x16 A2.6x10 4 TP3x6 BR3x12 A2.6x10 E5 5 TP3x6 BR3x16 A3x6 B3x6 BR3x16 B3x6 E8 13 14 8 10 A3x6 WP3x5 5 15 A3x6 16 A4x10 9 6 16 FDC4000 Parts List 2009.4 6 FDC4000 Ref. No. Part No. Part Name Qty Remarks Ref. No. Part No. Part Name Qty 1 323F3055 Timing belt 1 11 340F0395A Knob 3 2 322SP212 Plain bearing 4 12 388F5052A Spring, leaf 7 2 322SY357 Plain bearing 4 2009 expression & up 13 401F0384A Plate, pressure 6 It is incompatible 14 401F0776B Plate, pressure 1 3 322SP204 Plain bearing 2 15 606F10001 Optical fiber 1 4 356Y100396B Bracket 1 16 136Y101832A Cable 1 4 356Y100651 Bracket 1 2009 expression & up It is incompatible ① 350Y100885 Cover 1 5 113Y100177 PCB assembly 5 HP401 board ① 350N101066A Cover 1 6 347S0026 Spacer 2 Up to 2008 expression 7 309S0220003 Washer, spring 5 Up to 2008 expression ② 815Y100028A Mount 1 8 324N100053A Timing belt pulley 1 ③ 136H0990A Cable 1 9 118Y100056A Motor 1 Incubator motor ④ 322Y100008C Bearing unit 1 9 118N100053 Motor 1 2009 expression & up ⑤ 357Y100221A Mount 1 It is compatible ⑥ 815H0007B Photometric unit 1 Remarks 2009 expression & up 2009 expression & up It is compatible Incubator motor 10 340F0522 Knob FDC4000 Parts List 2009.4 2 17 7 FDC4000 Cylinder 2 3 26 2 BR3x16 WP3x5 TP3x6 1 TP3x6 BR3x12 1 E5 E5 A3x6 6 4 7 11 BR3x6 9 B3x6 TP3x6 13 12 BR3x6 25 23 22 21 16 19 20 18 15 24 17 18 FDC4000 Parts List 2009.4 7 FDC4000 Ref. No. Part No. Part Name Qty 1 356Y100390C Bracket 1 1 356Y100652 Bracket 1 2 324F3099A Timing belt pulley 4 Remarks Ref. No. Part No. Part Name Qty 15 385F0654 Coating 6 2009 expression & up 16 134Y100010B Solenoid valve 1 It is incompatible 16 134H10011 Solenoid valve 1 Motor 1 Syringe motor 17 385F0495 Tube 3 3 118N100052A Motor 1 2009 expression & up 18 385F0495 Tube 3 It is incompatible 19 385F0654 Tube 6 Syringe motor 20 372F1871 Fitting, tube 3 PH301 board 21 385N100015 Tube 1 PCB assembly 5 2009 expression & up It is incompatible 3 118Y100054A 4 113Y100177 Remarks 5 316S0256 Strap 15 Up to 2008 expression 22 372F10011 Fitting, tube 1 6 322SY357 Plain bearing 10 23 385N100014 Tube 1 7 323S3249 Timing belt 1 24 136Y101886 Cable 1 2009 expression & up 8 319N100574A Shaft 1 25 356N103167A Bracket 1 2009 expression & up 26 136Y101885 Cable 1 2009 expression & up ① 378Y100007 Cylinder 1 ① 378Y100024 Cylinder 1 Up to 2008 expression 9 322SY357 Plain bearing 10 10 317F51003 Stopper 1 11 385F0654 Tube 6 12 113H0356D PCB assembly 1 13 316S0256 Strap 11 14 316S0256 Strap 15 Up to 2008 expression FDC4000 Parts List 2009.4 Up to 2008 expression PRES1 board 2009 expression & up It is compatible ② 375Y100022 Seal 1 Syringe seal set 19 8 FDC4000 Lamp Unit 1 2 3 1 4 5 A3x6 6 A3x6 A2.6x6 2 7 B3x8 B3x10 8 TP3x6 B3x8 S3x6 9 WP3x5 B3x8 10 11 20 FDC4000 Parts List 2009.4 8 FDC4000 Ref. No. Part No. Part Name Qty Remarks Ref. No. Part No. Part Name Qty Remarks 1 308F0391 Screw, special 1 9 118Y100055A Motor 1 Filter motor 2 308F0210 Screw, special 1 9 118N100052A Motor 1 2009 expression & up 3 401F0420 Plate, keep 2 It is compatible 4 613H0006A Housing, optics 1 Filter motor 5 603Y100038 Filter 1 6 113Y100178 PCB assembly 7 7 316S0258 Strap 26 8 316S1197 Tie, wire 12 8 316S1198 Tie, wire 9 FDC4000 Parts List 2009.4 PH501 board 2009 expression & up 10 322FC263A Bearing 1 11 136Y101833 Cable 1 ① 360Y100167 Housing 1 ② 122F1052 Lamp 1 2009 expression & up Lamp ASSY 21 9 Accessories FDC4000 1 2 3 4 7 6 5 8 1 11 12 15 14 13 16 17 18 2 9 10 3 19 22 20 FDC4000 Parts List 2009.4 9 FDC4000 Ref. No. Part No. Part Name Qty Remarks Slide cartridge Ref. No. Part No. Part Name Qty Remarks 1 358N100045C Box 2 2 319F2278A Shaft 1 14 347F1947 Spacer 1 For 75mm tube 2 319N101017 Shaft 1 2009 expression & up 15 347F1948 Spacer 2 For 100mm tube It is compatible 16 347F1946 Spacer 1 For PF 3 137S1297A Fuse 2 AC250V F 10A 17 347N100228A Spacer 1 For mixing cup 4 346F2566 Plate,auxiliary 2 Slide weight 18 347N100227A Spacer 1 For CP 5 350N100836B Cover 1 19 889H10004 Paper (6pcs) 1 Optional Parts 6 122F1052 Lamp 1 Lamp ASSY 20 136F10001A Cable 1 i 7 350N100684C Cover 1 ISE cover 8 890Y100018 Special tool 2 Leak check tool ① 358Y100048 Box 1 Sample rack kit 9 358N100042E Sample rack ② 347Y100013A Spacer 1 ③ 317Y100005 Pin 1 Box 4 10 405N100642B Label 1 11 347N100236 Spacer 1 For φ13mm tube 12 347N100211B Spacer 1 For FUJI tube 1.5mL FDC4000 Parts List 2009.4 13 347N100212C Spacer 1 For FUJI tube 0.5mL INDEX9-9 (3pcs) Spacer kit Pins for sample rack (9pcs) 23 10 Optional parts FDC4000 1 1 4 3 2 6 8 7 11 10 5 9 13 12 14 PF 15 24 16 17 FDC4000 Parts List 2009.4 10 FDC4000 Ref. No. Part No. 1 852H10001 2 890H10040 2 890Y100018 Part Name Sample barcode Qty Remarks 1 Ref. No. 9 Part No. 890F10007 Part Name Slide C Qty 1 reader Sampler leak check Special tool 1 2009 expression & up FDC4000 tool Special tool FDC7000 10 890N100004 Sample position 1 adjustment tool 11 890H10010 PF leak check tool 890F10013 Slide insert 1 adjustment tool Special tool 1 4 890F10014 Spotting position 1 adjustment tool 12 890F10016 13 890F10031 890N100003 6 890F10015 7 890F10032 Dilution position 1 release tool Special tool FDC4000 15 114N100019 Memory device 1 FDC4000 16 114Y100160 Memory device 1 1 Special tool 1 Special tool FDC3500 FDC4000 Special tool FDC4000 FDC7000 1 Special tool adjustment tool Master tip Interlock Special tool FDC3500 FDC4000 PF sensor 1 adjustment tool Dilution position 1 adjustment tool 14 890N100005 FDC7000 5 PF position check tool Special tool FDC3500 FDC4000 Special tool FDC3500 FDC4000 FDC3500 FDC4000 FDC7000 Special tool FDC4000 FDC4000 3 Special tool FDC3500 FDC4000 1 tool Sampler leak check Remarks No program installed Driver software (for version upgrade) 17 114Y100161A Memory device 1 Application software (for version upgrade) FDC4000 FDC7000 8 890F10006 Slide A 1 Special tool ① 898Y100303 Maintenance kit 1 Special tool FDC3500 FDC4000 FDC7000 FDC4000 Parts List 2009.4 25 11 FDC4000 PARTS NO. INDEX LIST Part No. 26 Index No.-Ref.No. Part No. Index No.-Ref.No. 107Y100008A 3-31 11 136Y101833 8-11 21 113H0356D 7-12 19 136Y101885 7-26 19 113Y100171G 2-3 136Y101886 7-24 113Y100172B 2-15 9 137S1297A 9-3 23 113Y100173A 3-23 11 144S0001 4-1 13 113Y100174B 3-45 11 145S0013 1-9 7 113Y100175 3-46 11 146Y100011 3-5 11 113Y100176 3-28 11 147N100001 3-3 11 113Y100177 1-5 308F0210 8-2 21 113Y100177 4-20 308F0391 8-1 21 113Y100177 6-5 17 309S0155 3-41 113Y100177 7-4 19 309S0157 3-36 113Y100178 3-33 11 309S0220003 3-7 113Y100178 4-18 13 309S0220003 4-25 113Y100178 4-19 13 309S0220003 6-7 113Y100178 4-2 113Y100178 5-13 113Y100178 8-6 113Y100179 3-20 113Y100186 1-1 113Y100192B 3-1 9 7 13 19 11 11 11 13 17 309S1055 3-47 11 316F7057 3-15 11 316S0174 2-13 9 316S0174 3-19 11 7 316S0256 2-11 11 316S0256 2-9 9 11 13 15 21 11 9 113Y100207 3-35 11 316S0256 3-2 113Y100280 114N100019 5- ⑤ 10-15 15 316S0256 3-22 25 316S0256 3-43 114Y100160 10-16 25 316S0256 4-3 114Y100161A 10-17 25 316S0256 7-13 118N100052A 4-22 13 316S0256 7-14 118N100052A 5-15 15 316S0256 7-5 118N100052A 5-19 15 316S0258 2-12 118N100052A 7-3 19 316S0258 2-16 118N100052A 8-9 21 316S0258 2-2 118N100053 6-9 17 316S0258 3-10 11 118Y100051A 4-22 13 316S0258 3-17 11 118Y100052A 5-19 15 316S0258 3-32 11 118Y100053B 5-15 15 316S0258 3-48 118Y100054A 7-3 19 316S0258 4-4 13 118Y100055A 8-9 21 316S0258 8-7 21 118Y100056A 6-9 17 316S1027 2-6 9 118Y100095 119S0042A 4- ② 2-23 13 316S1027 2-7 9 9 316S1027 3-13 122F1052 122F1052 8- ② 21 9-6 23 316S1027 3-14 316S1033 2-1 125H10007B 2-17 9 316S1033 2-24 9 128S1050 3-42 11 316S1033 3-11 11 134H10011 7-16 19 316S1140 2-14 9 134Y100010B 7-16 19 316S1140 2-21 9 136F10001A 9-20 23 316S1140 2-26 9 136H0990A 136Y101242D 6- ③ 2-18 17 316S1140 3-12 11 9 316S1140 3-21 136Y101280 5-14 15 316S1140 3-9 11 136Y101706 136Y101828 3- ① 5-27 11 316S1197 1-8 7 15 316S1197 2-4 9 136Y101829A 5-26 15 316S1197 3-8 11 136Y101831 4-27 13 316S1197 8-8 21 136Y101832A 6-16 17 316S1198 2-4 9 11 11 13 19 19 19 9 9 9 11 11 11 9 11 FDC4000 Parts List 2009.4 FDC4000 PARTS NO. INDEX LIST Part No. Index No.-Ref.No. Part No. Index No.-Ref. No. 316S1198 3-8 11 346F2566 9-4 316S1198 8-8 21 346F2806B 4-11 13 316S2027 3-16 11 346N100362 4-16 13 316S2027 3-39 11 347F1946 9-16 23 316S2027 3-40 11 347F1947 9-14 23 317F51003 7-10 19 347F1948 9-15 317Y100005 318S1195 9- ③ 2-5 9 23 347N100181A 1-3 7 347N100188 1-6 7 318S1195 2-8 347N100198A 1-15 7 318S1261 3-38 347N100211B 9-12 23 319F2278A 9-2 23 347N100212C 9-13 23 319F3432 5-7 15 347N100227A 9-18 23 319F3437 4-7 13 347N100228A 9-17 23 319H0019C 5-8 15 347N100236 9-11 319N100571A 4-5 13 347S0026 6-6 319N100574A 7-8 19 347S0506 2-10 319N101017 9-2 23 347S0506 3-34 319N101245 4-7 13 347S0890 3-6 322F2033 5-9 15 347S0908 3-49 322FC263A 8-10 347Y100013A 350F3806C 9- ② 23 5-3 15 9 11 21 23 23 23 17 9 11 11 11 322FY093 5-5 322N100184 1-13 7 350F3807B 5-4 15 322SF018 5-10 15 350F5593 5-1 15 322SF018 5-6 350F5595 4-9 13 322SP204 5-23 350F5596 4-13 322SP204 6-3 350F5597 4-12 322SP212 5-24 350N100591A 1-2 322SP212 6-2 350N100593C 1-18 7 322SY357 4-24 350N100600C 1-10 7 322SY357 4-6 350N100602A 1-11 322SY357 5-20 15 350N100684C 9-7 322SY357 5-24 15 350N100685A 1-12 322SY357 6-2 17 350N100836B 9-5 322SY357 7-6 19 322SY357 7-9 19 350N101066A 350Y100566B 6- ① 1-16 7 322Y100008B 5- ② 350Y100567B 1-21 7 322Y100008C 323F3055 6- ④ 17 6-1 17 350Y100568A 1-10 7 350Y100584A 1-12 7 323F3056 5-21 15 350Y100877 1- ① 7 323F3057 5-25 15 350Y100878 1- ② 7 323F3128 4-17 13 350Y100879A 1- ③ 7 323S3249 7-7 350Y100880 1- ④ 7 324F3098A 5-18 15 350Y100882 4- ③ 13 324F3099A 5-12 15 350Y100883 4- ④ 13 324F3099A 5-16 15 350Y100884 5- ④ 15 324F3099A 7-2 350Y100885 6- ① 17 324F3198B 4-21 324N100053A 6-8 350Y100893 356N101284 3- ② 11 3-4 11 340F0395A 6-11 17 356N101590C 3-4 340F0522 6-10 17 356N101608D 3-29 11 340N100041B 2-19 9 356N101610C 3-27 11 341N100113A 3-37 11 356N103167A 7-25 19 341N100223 3-37 11 356N103189A 3-29 11 346F1514 4-16 13 356Y100389B 4-23 346F1522A 4-14 13 356Y100390C 7-1 FDC4000 Parts List 2009.4 15 15 15 17 15 17 13 13 15 19 19 13 17 11 13 13 7 7 23 7 23 17 11 13 19 27 11 FDC4000 PARTS NO. INDEX LIST Part No. 28 Index No.-Ref.No. Part No. Index No.-Ref.No. 356Y100394B 5-22 15 385N100015 7-21 19 356Y100395B 5-17 15 388F5052A 6-12 17 356Y100396B 6-4 388N100250E 3-24 11 356Y100648 4-23 13 388N100401 3-24 11 356Y100649 5-22 15 388N100595 3-50 356Y100650 5-17 15 388S1089 1-4 356Y100651 6-4 17 388S1454 4-10 13 356Y100652 7-1 19 394F0031 4-15 13 356Y100658 3- ③ 11 394F0112 3-44 11 356Y100850 3- ③ 11 394N100007 3-44 11 357Y100219 4- ① 13 401F0384A 6-13 357Y100220 4- ⑤ 13 401F0420 8-3 357Y100221A 358N100042E 6- ⑤ 17 9-9 23 401F0776B 6-14 17 405N100558B 1-17 7 358N100045C 9-1 405N100642B 9-10 23 358Y100048 9- ① 23 405N100646A 1-17 7 358Y100050A 360F0449B 1- ⑤ 5-11 7 405N101123 1-17 7 15 405N101125 1-17 360Y100167 362N100114C 8- ① 3-26 21 603Y100038 8-5 11 605N100013 3-18 363N100425G 3-30 11 606F10001 6-15 363N100428C 1-14 7 608Y100001B 1-7 7 363N100429D 1-20 7 613H0006A 8-4 21 363Y100249 364S0003 3- ⑤ 2-25 11 815H0007B 6- ⑥ 9 364S0027 2-22 9 815Y100028A 852H10001 366N100018B 3-25 367F0248 4-8 17 23 11 13 852Y100024A 889H10004 11 7 17 21 7 21 11 17 17 6- ② 17 10-1 25 3- ④ 9-19 11 23 367N100021A 2-20 9 890F10006 10-8 25 368Y100028 4- ⑥ 13 890F10007 10-9 25 368Y100029 4- ⑦ 13 890F10013 10-3 25 371Y100020A 372F10011 5- ① 7-22 15 890F10014 10-4 25 19 890F10015 10-6 25 372F1871 7-20 19 890F10016 10-12 375H0005 5- ③ 15 890F10031 10-13 375Y100022 376N100020B 7- ② 1-19 19 890F10032 10-7 7 890H10010 10-11 378Y100007 7- ① 19 890H10040 10-2 25 378Y100024 385F0495 7- ① 7-17 19 890N100003 10-5 25 19 890N100004 10-10 385F0495 7-18 19 890N100005 10-14 385F0654 7-11 19 890Y100018 10-2 25 385F0654 7-15 19 890Y100018 9-8 23 385F0654 7-19 19 898H10113A 4-26 13 385H0007 5-2 898Y100303 10- ① 25 385N100014 7-23 15 25 25 25 25 25 25 19 FDC4000 Parts List 2009.4