Product Information
The product is the GE Healthcare Pre Installation manual for the
INNOVA IGS 620-630 FINAL STUDY system. It is a set of drawings and
specifications that provide guidance for site design and preparation
before installing the GE Healthcare product. The manual includes
various sections such as general notes, equipment layout, structural
details, HVAC, electrical layout, power requirements, and
interconnections. It also emphasizes the importance of referencing the
Pre Installation manual to ensure complete documentation for site
design and preparation. Additionally, the manual mentions that GE
Healthcare does not take responsibility for damages resulting from
changes made by others to the drawings. It also advises against scaling
from printed pdf files and states that GE accepts no responsibility or
liability for defective work due to scaling from these drawings. The
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
3 of 77
https://device.report/manual/8787491
manual is in A3 format with a scale of 1/4=1'-0. The file name is ENVAS-TYP-IGS-6X0-WEB.DWG. There is also a reference to the PIM
Manual 5694389-1-1EN, which provides further information and
specifications for the product.
Product Usage Instructions
1. Before starting the installation of the GE Healthcare INNOVA IGS
620-630 system, it is essential to refer to the Pre-Installation manual
provided.
2. Ensure that all necessary site design and preparation documentation
is complete by following the guidelines in the manual.
3. Familiarize yourself with the different sections of the manual,
including general notes, equipment layout, structural details, HVAC,
electrical layout, power requirements, and interconnections.
4. Take note of the warnings and disclaimers mentioned in the manual,
such as GE Healthcare not taking responsibility for changes made by
others to the drawings and the potential issues with scaling from
printed pdf files.
5. Follow the specified format and scale of the drawings, which are in A3
size with a scale of 1/4=1'-0.
6. Make sure to access the PIM Manual 5694389-1-1EN for additional
information and specifications related to the product.
7. Adhere to the customer responsibilities mentioned in the manual,
such as ensuring room dimensions, ceiling support structure,
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
4 of 77
https://device.report/manual/8787491
cleanliness, cableways, and adequate illumination meet GE
specifications.
8. Install the system according to the guidelines provided in the manual,
including the installation of countertops, vascular baseplates, and
table baseplates if applicable.
9. Follow any specific requirements mentioned in the manual for floor
finish and protection for the IGS 730/740 models.
10. If there are any uncertainties or questions regarding the installation
process, contact a GE Healthcare representative for assistance.
DISCLAIMER
GENERAL SPECIFICATIONS
GE is not responsible for the installation of developers and associated
equipment, lighting, cassette trays and protective screens or
derivatives not mentioned in the order.
The final study contains recommendations for the location of GE
equipment and associated devices, electrical wiring and room
arrangements. When preparing the study, every effort has been made
to consider every aspect of the actual equipment expected to be
installed.
The layout of the equipment offered by GE, the dimensions given for
the premises, the details provided for the pre-installation work and
electrical power supply are given according to the information noted
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
5 of 77
https://device.report/manual/8787491
during on-site study and the wishes expressed by the customer.
The room dimensions used to create the equipment layout may
originate from a previous layout and may not be accurate as they
may not have been verified on site. GE cannot take any responsibility
for errors due to lack of information.
Dimensions apply to finished surfaces of the room.
Actual configuration may differ from options presented in some
typical views or tables.
If this set of final drawings has been approved by the customer, any
subsequent modification of the site must be subject to further
investigation by GE about the feasibility of installing the equipment.
Any reservations must be noted.
The equipment layout indicates the placement and interconnection of
the indicated equipment components.
There may be local requirements that could impact the placement of
these components. It remains the customer's responsibility to ensure
that the site and final equipment placement complies with all
applicable local requirements.
All work required to install GE equipment must be carried out in
compliance with the building regulations and the safety standards of
legal force in the country concerned.
These drawings are not to be used for actual construction purposes.
The company cannot take responsibility for any damage resulting
therefrom.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
6 of 77
https://device.report/manual/8787491
CUSTOMER RESPONSIBILITIES
It is the responsibility of the customer to prepare the site in
accordance with the specifications stated in the final study. A detailed
site readiness checklist is provided by GE. It is the responsibility of the
customer to ensure all requirements are fulfilled and that the site
conforms to all specifications defined in the checklist and final study.
The GE Project Manager of Installation (PMI) will work in cooperation
with the customer to follow up and ensure that actions in the
checklist are complete, and if necessary, will aid in the rescheduling of
the delivery and installation date.
Prior to installation, a structrual engineer of record must ensure that
the floor and ceiling is designed in such a way that the loads of the
installed system can be securely borne and transferred. The layout of
additional structural elements, dimensioning and the selection of
appropriate installation methods are the sole responsibility of the
structural engineer. Execution of load bearing structures supporting
equipment on the ceiling, floor or walls are the customer's
responsibility.
RADIO-PROTECTION
Suitable radiological protection must be determined by a qualified
radiological physicist in conformation with local regulations. GE does
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
7 of 77
https://device.report/manual/8787491
not take responsibility for the specification or provision of radioprotection.
THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND
APPROVED THE PLANS IN THIS DOCUMENT.
DATE
NAME
SIGNATURE
GLOBAL SITE READINESS CHECKLIST (DI)
DOC1809666 Rev. 7
Site Ready Checks at Installation
EHS Site Requirements
Overall access route to the scan room free from obstruction / high ha
zards.
Enough space to store tools, equipment, parts, install waste and the g
eneral area free from obstruction and trip hazards.
Enough necessary facilities for the GE employees available.
No 3rd parties working in the area that may affect the safety of the in
stallation activity.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
8 of 77
https://device.report/manual/8787491
Area free from any chemical, gas, dust, welding fume exposure and h
as painting been completed and dry.
All emergency routes identified, signed and clear from obstruction.
Accessible single source lockable panel that LOTO can be applied to f
or GE equipment installation (MDP and/or PDU).
There are no other conditions or hazards that you have observed or h
ave been made aware of by the customer or contractors on site.
Required for Mechanical Install start
Room dimensions, including ceiling height, for all Exam, Equipment/T
echnical & Control rooms meets GE specifications.
Ceiling support structure, if indicated on the GE drawing, is in the corr
ect location and at the correct height according to the Original Equip
ment Manufacturer specifications.
Levelness and spacing has been measured, and is ready for the install
ation of any GE supplied components.
Overhead support Structure (unistrut) has been confirmed with custo
mer/contractor to meet required GE provided criteria.
Finished ceiling is installed. If applicable ceiling tiles installed per PMI
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
9 of 77
https://device.report/manual/8787491
discretion.
Floor levelness/flatness is measured and within tolerance, and there a
re no visible defects per GEHC specifications.
Entry door threshold meets PIM requirement
Floor Strength and thickness have been discussed with customer/cont
ractor and they have confirmed GE requirements are met.
Rooms that will contain equipment, including staging areas if applica
ble, are construction debris free. Precautions must be taken to preven
t debris from entering rooms containing equipment.
Cable ways (floor/wall/ceiling/Access Flooring) are available for install
ation of GE cables are of correct length and diameter.
Cable ways routes per GE Final drawings and cable access openings a
reas installed at a time determined by GEHC PM. Surface floor duct ca
n be installed at time of system installation.
Adequate room illumination installed and working.
Customer supplied countertops where GE equipment will be installed
are in place.
Vascular baseplates preparation complete per GE requirements.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
10 of 77
https://device.report/manual/8787491
For IGS 730/740: Floor finish is according to the GE Specifications and
protection is installed . Specifications for concrete substrate & Mono
pur 7 mm flooring have been met. Table baseplate installed and flush
to the finished floor.
For IGS 730/740: Room Interventional Reference Point (RIRP) value ha
s been defined with the customer. Either 1120mm, 1278 mm or 1508
mm.
Ensure that all third party suppliers are identified and have been infor
med about the project dates and how they need to proceed in accord
ance with their needs for interfacing to our equipment.
Required for Calibration start
HVAC systems Installed, and the site meets minimum environmental
operational system requirements.
System power & grounding (PDB/MDP) is available as per GE specific
ations.
System power & grounding (PDB/MDP) is installed at point of final co
nnection and ready to use. Lock Out Tag Out is available.
PMI to confirm all feeder wires and breaker are size appropriately. EP
O installed if needed.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
11 of 77
https://device.report/manual/8787491
PMI to confirm with electrician all power and signal cables are well ter
minated ensuring there are no loose connections.
Network outlets installed.
Computer network available and working.
Lead doors and windows complete or scheduled to be installed. If ap
plicable, radiation protection (shielding) finished & radioprotection re
gulatory approval for installation obtained.
Note: The details shown here are only an extract from DOC1809666.
For the complete document please contact your PMI.
INTERFERENCE SPECIFICATIONS
MAGNETIC INTERFERENCE SPECIFICATIONS
Image intensifiers must be located in ambient static magnetic fields
of less than 1 gauss to guarantee specified imaging performance.
X-ray tubes must be located in ambient static magnetic fields of less
than 10 gauss to guarantee specified performance.
System electronics must be located in ambient static magnetic fields
of less than 10 gauss to guarantee data integrity.
Operators console equipment must be located in ambient static
magnetic fields of less than 10 gauss to obtain specified geometric
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
12 of 77
https://device.report/manual/8787491
linearity.
CUSTOMER SITE READINESS REQUIREMENTS
Any deviation from these drawings must be communicated in writing
to and reviewed by your local GE healthcare installation project
manager prior to making changes.
Make arrangements for any rigging, special handling, or facility
modifications that must be made to deliver the equipment to the
installation site. If desired, your local GE healthcare installation project
manager can supply a reference list of rigging contractors.
New construction requires the following;
1. Secure area for equipment,
2. Power for drills and other test equipment,
3. Capability for image analysis,
4. Restrooms.
Provide for refuse removal and disposal (e.g. crates, cartons, packing)
For CT, MR, PET/CT, and SPECT systems it is required to minimize
vibrations within the scan room. It is the customer's responsibility to
contract a vibration consultant/engineer to implement site design
modifications to meet the GE vibration specification. Refer to the
system preinstallation manual for vibration specifications.
ELECTROMAGNETIC INTERFERENCE
The system is suitable for use in the specified electromagnetic
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
13 of 77
https://device.report/manual/8787491
environment. The purchaser or user of the system should assure that it
is used in an electromagnetic environment as described below:
EMISSION
TEST COMP
S
LIANCE
Radio–F
requency E
Group1 Cla
ss A limits
ELECTROMAGNETIC ENVIRONMENT
The system is suitable for use in all establ
ishments other than domestic and those
missions C
directly connected to the low voltage po
ISPR11
wer supply network that supplies buildin
gs used for domestic purposes.
Group1 Cla
ss A limits
The system uses RF energy only for its int
ernal function. Therefore, the RF emissio
n is very low and not likely to cause any i
nterference in nearby electronic equipme
nt.
Harmonic
Not applica
The system is suitable for use only in esta
emissions I
ble
blishments not directly connected to a p
EC 61000–
ublic low voltage power supply network.
3–2
Voltage flu
ctuations/
flicker emi
Not applic
able
The system is suitable for use only in est
ablishments not directly connected to a
public low voltage power supply networ
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
14 of 77
ssions IEC
https://device.report/manual/8787491
k.
61000–3–3
A
GE Supplied
D
Available from GE
B
GE Supplied/contractor installe
E
Equipment existing in roo
d
C
Customer/contractor supplied
m
*
Item to be reinstalled from
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
15 of 77
https://device.report/manual/8787491
and installed
another site
MAX
B
ITEM
Y
MAX
DESCRIPTIO
HEAT
WE
HEAT O
WE
N
OUTPU
IG
UTPUT
IG
T
HT
HT
(lb
(k
s)
g)
(btu)
D
1
Advantage wo
(W)
3412
22
1000
10
1405
19.
410
9
rkstation
A
2
Operator cons
ole
D
3
Reference mo
6
1706
18
500
8
5435
80
1590
36
nitor
A
4
C1 Frontal Ca
binet
A
5
C1 Lateral Cab
inet
9
3969
69
2
7
1160
31
4
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
16 of 77
A
6
C2 Cabinet
https://device.report/manual/8787491
6171
57
1810
8
A
7
Detector cond
709
32
2
210
itioner
A
8
COOLIX 4100
9
Chiller Autotra
14.
5
21496
water chiller
A
26
26
6300
5
12
0
709
66
210
30
3412
25
1000
11
nsformer
A
10
Large Display
Cabinet
A
11
3kVA LDM UP
4
5
1257
79
370
36
2216
85
650
38
S
B/
12
D
A
Power Distrib
ution Box
13
3kVA PDB UP
9
9.5
1257
79
370
36
4720
10
1380
54
S
D
14
20kVA Fluoro
UPS UL
65
1
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
17 of 77
https://device.report/manual/8787491
A
15
Xray buzzer
-
1
-
0.5
A
16
LC gantry
5517
15
1620
70
56
A
17
LP Gantry
5517
17
5
1620
57
A
18
Longitudinal s
79
7
-
68
-
31
-
-
-
-
614
16
180
75
tationary rail f
or LP
Gantry
B
19
Cable drape r
ail for LP Gant
ry
A
20
OMEGA V pati
ent table
A
21
Monitor suspe
64
4.5
-
-
-
-
-
68
-
31
nsion short bri
dge
A
22
Longitudinal s
tationary rail
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
18 of 77
D
23
Large Display
https://device.report/manual/8787491
-
Monitor with t
88
-
0
39
9
wo backup m
onitors
D
24
Mavig rad shi
-
eld and LED la
14
-
65
-
3
3
mp with 2.5m
ceiling track
D
25
External Transf
-
6
ormer for LED
Surgical lamp
Circuit breaker or equivalent with LOTO must be
installed in the mains line to the PDB
C
26
this device must be compatible with the power i
nput specifications of the system. the
customer is responsible for procurement, delive
ry, and installation.
C
27
Cable drape rail.
C
28
Control wall to ceiling with lead glass viewing w
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
19 of 77
https://device.report/manual/8787491
indow.
C
29
Scrub sink
C
30
Countertop for equipment- provide grommeted
openings as required to route cables
C
31
Minimum door opening for equipment delivery
is 44 in. w x 83 in. h [1118mm x 2108mm], conti
ngent on a 96 in. [2438mm] Corridor width
C
32
Storage Cabinet
EXAM ROOM HEIGHT
FINISHED FLOOR-TO-FALSE CEILING
Require
Require
d
d
2845mm
9'-4" +/0.2"
Door openings: Exam room min 1160 [50 in], Technical room min 800
[36 in], rec 1000 [40 in]
For Accessory Sales: (866) 281-7545 Options 1, 2, 1, 2 or mail to:
gehcaccessorysales@ge.com CONTROL ROOM VIEW
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
20 of 77
https://device.report/manual/8787491
EXAM ROOM VIEW
TECHNICAL ROOM VIEW
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
21 of 77
https://device.report/manual/8787491
TECHNICAL ROOM VIEW
GANTRY
Center of Gravity
SCALE 1:50
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
22 of 77
https://device.report/manual/8787491
LATERAL POSITIONER
OMEGA V PATIENT TABLE
Center of Gravity
Scale 1:20
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
23 of 77
https://device.report/manual/8787491
INNOVA CABINETS
CENTER OF GRAVITY
CABIN
C1 (AP)
C1 (LAT)
C2
ET
DIMEN
mm
in
mm
in
mm
in
A
452
17.7
506
19.9
400
15.7
B
283
11.1
338
13.3
325
12.8
C
904
35.5
839
33.0
948
37.3
SION
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
24 of 77
https://device.report/manual/8787491
LARGE DISPLAY CABINET
DIGITAL ENERGY SG SERIES 10-20kVA UPS
3kVA UPS
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
25 of 77
https://device.report/manual/8787491
X-RAY TUBE CHILLER
COOLIX AUTOTRANSFORMER
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
26 of 77
https://device.report/manual/8787491
DETECTOR CONDITIONER
DELIVERY
The customer should
Provide an area adjacent to the vascular suite for delivery and
unloading of the GE equipment.
Ensure that the dimensions of all doors, corridors, ceiling heights are
sufficient to accommodate the movement of GE equipment from the
delivery area into the definitive installation room.
Ensure that access routes for equipment will accommodate the
weights of the equipment and any transportation, lifting and rigging
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
27 of 77
https://device.report/manual/8787491
equipment.
Ensure that all necessary arrangements for stopping and unloading
on public or private property not belonging to the customer have
been made.
DIMENSIONS OF DELIVERY WITH DOLLY TRANSPORT EQUIPMEN
T
EQUIPM
DIMENSIONS
WEIGHT
ENT
LC GAN
TRY
LP GAN
TRY
LP RAIL
S (BOX)
LENGTH
2820 mm
111 in
WIDTH
1230 mm
48.4 in
HEIGHT
2000 mm
79 in
LENGTH
2790 mm
109.8 in
WIDTH
1135 mm
44.7 in
HEIGHT
2152 mm
84.7 in
LENGTH
6070 mm
239 in
WIDTH
260 mm
10 in
HEIGHT
300 mm
11.8 in
1060 kg
2340 lbs
1225 kg
2700 lbs
95 kg
209 lbs
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
28 of 77
https://device.report/manual/8787491
SHIPPING DOLLY FOR LP GANTRY
The minimum door width needed (to accommodate the IGS Lateral
shipping dolly) is 1135 mm [44.7 in] with the dolly stabilizers
retracted.
SHIPPING DOLLY FOR LC GANTRY
Both ends of the dolly can be removed which will shorten the LC
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
29 of 77
https://device.report/manual/8787491
gantry dolly to 2300 mm [90 in], also the right and left top handles
can be removed which will make the width 1160 mm [46 in].
STRUCTURAL NOTES
All steel work and parts necessary to support ceiling-mounted tube
hanger or other equipment are to be supplied by the customer or his
contractors. The unistrut or equivalent structure should run
continuous with no fittings extending below face of unistrut channel,
run wall to wall, be parallel, square and in the same horizontal plane
flush with finished ceiling. The system is to be cross braced vertically,
horizontally and diagonally to allow no movement and a maximum of
1,58mm(1/16") deflection. (10) 12,7mm (1/2") dia. X 38,1mm (1 1/2")
long bolts with unistrut 12,7mm (1/2") nuts with springs are to be
provided by customer or his contractors for each stationary and
auxillary support rail. Closure strips shall be provided for areas of
unistrut exposed and without mounting units.
Methods of support for the steelwork that will permit attachment to
structural steel or through bolts in concrete construction should be
favored. Do not use concrete or masonry anchors in direct tension.
All units that are wall mounted or wall supported are to be provided
with supports where necessary. Wall supports are to be supplied and
installed by the customer or his contractors. See plan and detail
sheets for suggested locations and mounting hole locations.
All ceiling mounted fixtures, air vents, sprinklers, etc. To be flush
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
30 of 77
https://device.report/manual/8787491
mounted, or shall not extend more than 6,35mm (1/4") below the
finished ceiling.
Control walls with tube hanger passage above shall be constructed to
2130mm (7'-0") high.
Floor slabs on which equipment is to be installed must be level to
3,17mm (1/8") in 3050mm (10'-0")
Minimum floor thickness of 203mm (8").
Dimensions are to finished surfaces of room.
The customers contractor must provide all penetrations in posttension floors.
The customers contractor must provide and install any non-standard
anchoring. Documents for standard anchoring methods are included
with GE equipment drawings for geographic areas that require such
documentation.
Customers' contractor must provide and install hardware for "through
the floor" anchoring and/or any bracing under access floors. This
contractor must also provide floor drilling that cannot be completed
because of an obstruction encountered while drilling by the GE
installer such as rebar etc.
It is the customer's responsibility to perform any floor or wall
penetrations that may be required. The customer is also responsible
for ensuring that no subsurface utilities (e.g., electrical or any other
form of wiring, conduits, piping, duct work or structural supports (i.e.
post tension cables or rebar)) will interfere or come in contact with
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
31 of 77
https://device.report/manual/8787491
subsurface penetration operations (e.g. drilling and installation of
anchors/screws) performed during the installation process. To ensure
worker safety, GE installers will perform surface penetration
operations only after the customer's validation and completion of the
"GE surface penetration permit"
IF ACCESS IS NOT READILY AVAILABLE IT IS RECOMMENDED TO
PROVIDE A TRAPDOOR IN THE CEILING TO ALLOW SERVICE ACCESS
FOR CABLE MANAGEMENT.
STRUCTURAL LAYOUT ITEM LIST LC GANTRY AND TABLE
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
32 of 77
https://device.report/manual/8787491
ANCHORING WITH NO BASEPLATE
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
33 of 77
https://device.report/manual/8787491
1. 1: LC baseplate mounting location: 12 bolts Pullout strength on
each bolt 736 daN
M20 Through-Bolts recommended (supplied by GE) Alternates:
M16 Mechanical anchors (supplied by GE)
Chemical anchors (not supplied by GE): HILTIHVU adhesive
capsule + HAS Anchor rod
2. 2: Table mounting location: 4 bolts required Pullout strength on
each bolt 4432 daN
M20 Through-Bolts recommended (supplied by GE) Alternates:
M16 Mechanical anchors (supplied by GE)
Chemical anchors (not supplied by GE): HILTIHVU adhesive
capsule + HAS Anchor rod
3. 2b: Alternate bolt holes for seismic zones 1 and 2
FLOOR REQUIREMENTS AND CABLE MANAGEMENT
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
34 of 77
https://device.report/manual/8787491
FLOOR REQUIREMENTS
The maximum pullout force per GE supplied anchor was calculated
assuming :
A concrete compression strength of 17.24 MPa at 28 days (which
is the minimum required compression stength).
Anchors installed to the required hole depth of 165.1 mm [6.5 in]
minimum.
Center of anchor hole to concrete edge distance 79.4 mm [3.1
in].
Make sure to obtain data on compression strength of the
concrete before using floor anchors.
The floor slab on which the equipment is to be installed must be flat
and level (1 mm [0.04 in]/1 m [40 in] where equipment is installed and
5 mm [0.2 in]/2 m [79 in] general levelness).
Anchoring to the floor is intended to the structural elements and not
to common screed.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
35 of 77
https://device.report/manual/8787491
Do not glue the floor covering in the gantry zone.
JUNCTION BOX BELOW FLOOR
NOTE: PIPE, JUNCTION BOX AND DUCT OR CONDUIT ARE TO BE
SUPPLIED AND IS TO BE INSTALLED BY CUSTOMER OR CUSTOMER'S
CONTRACTOR.
XT RADIOGRAPHIC SUSPENSION, INBOARD MOUNTING
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
36 of 77
https://device.report/manual/8787491
Structural Support system is not supplied or installed by GE
Healthcare DETAIL 1
DETAIL 2
MAVIG SUSPENSION MOUNTING METHOD
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
37 of 77
https://device.report/manual/8787491
Weight up to: 94 kg [207 lb] (75 kg [165 lb] system + 19 kg [42 lb]
track)
The required factor of safety is "4" for attaching to Unistrut or
equivalent rails and "6" for attaching to the concrete ceiling.
CONSULT MAVIG INSTALLATION MANUAL REV: POR03001 TO
DESIGN AND MOUNT THE CEILING SUPPORT. SUSPENSION COLUMN
LENGTHS AND INSTALLATION DETAILS
MONITOR SUSPENSION RAIL MOUNTING SPECIFICATIONS
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
38 of 77
https://device.report/manual/8787491
CEILING SUSPENSION DISCLAIMER
Safety and precautionary comments Only qualified, licensed
technicians can perform electrical connections, installation, removal and
repair. It is strongly recommended that at least two persons perform the
installation. Installing the system: Prior to installation, a structural
engineer must confirm that the mounting structure is strong enough to
provide proper support for the entire system and any attached end
devices. Installation must be completed according to local building
codes. Determination of required installation hardware and torque
values for installation of the ceiling column and ceiling track is the sole
responsibility of the structural engineer. Ceiling mounted systems must
be installed properly. Failure to follow the instructions provided may
lead to a potentially dangerous and unstable condition of the system.
GE and/or MAVIG is not responsible for unauthorized modifications
made to the system or use of the system for unintended purposes. GE
and/or MAVIG cannot be held liable for improper operation and
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
39 of 77
https://device.report/manual/8787491
modifications. Since improper modifications may impair proper
operation, safety or reliability of the system, product modifications
require written authorization from MAVIG. Under GE responsibility or
under Customer responsibility, for all pre-installations, whatever is the
supporting structure (bridge, chair, Unistrut channel, other channels,
direct anchorage in concrete, transversal beam, etc. …) a certificate must
be obtained from a structural engineer. This certificate shall include the
definition of fasteners and of their tightening torque, especially for the
non-standard cases described in MAVIG PIM and for which the standard
anchoring/screws delivered with product shall not be used but shall be
defined (and implemented in most cases) by the structural company.
WARNING: It is prohibited to alter the length of the ceiling column or
remove any securing screws. LOCATION OF LATERAL POSITIONER
AND MONITOR SUSPENSION RAILS- SECTION
MONITORS CENTERED
1
Monitor suspension axis
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
40 of 77
https://device.report/manual/8787491
2
Monitor cable suspension axis
3
Lateral positioner rail axis
4
Lateral positioner cable suspensi
on axis
Note: no mounting hardware can protrude below the finished ceiling
height (top surface of lateral positioner stationary rails), such as Unist
rut mounting bolts, support brackets, sprinklers, air vents etc. in betw
een LP stationary rails
LATERAL POSITIONER RAIL MOUNTING SPECIFICATIONS
WARNING: STRUCTURE SHOULD NOT ALLOW VIBRATIONS
TRANSMISSION EQUAL OR LOWER THAN 10Hz
CAUTION
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
41 of 77
https://device.report/manual/8787491
The maximum load per bolt will not exceed 1912 N [430 lb].
Each bolt must not "pull out" otherwise fail under a vertically
downward "dead" load of 7633 N [1717 lb].
OTS SUSPENSION RAILS MOUNTING SPECIFICATIONS
Distance between holes axis 660.4 mm [26 in], Maximum load per screw
is 155.7 daN [350 lb], however each mounting screw must not "PULL
OUT" or otherwise fail under a vertically downward dead load of 622.8
daN [1400 lb]. Bolts for mounting stationary rails on structural supports
supplied by GE (1/2" - 13 headed bolts)
TEMPERATURE AND HUMIDITY SPECIFICATIONS
IN-USE CONDITIONS
EXAM ROOM
CONTROL ROO
TECHNICAL RO
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
42 of 77
https://device.report/manual/8787491
M
Temperatur
Min
e
Rec
Ma
om
x
Min
OM
Rec
Ma
om
x
Min
Rec
Ma
om
x
me
me
me
nde
nde
nde
d
d
d
15
22
32
15
22
35
10 °
18 °
32 °
°C
°C
°C
°C
°C
°C
C [5
C [6
C [9
[59
[72
[90
[59
[72
[95
0°
4°
0°
°F]
°F]
°F]
°F]
°F]
°F]
F]
F]
F]
Temperature
≤ 10 °C/h [50 °F
≤ 10°C/h [50 °F/
≤ 10 °C/h [50 °F
gradient
/h]
h]
/h]
Relative hum
30% to 70%
30% to 75%
30% to 75%
≤ 10%/h
≤ 10%/h
≤ 10%/h
idity (1)
Humidity gra
dient
STORAGE CONDITIONS
Temperature
+10 °C [50 °F] to +40 °C [104 °F]
Relative humidity (1)
10% to 80%
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
43 of 77
https://device.report/manual/8787491
Material should not be stored for more than 90 days.
(1) Non-condensing
AIR RENEWAL
According to local standards.
NOTE: In case of using air conditioning systems that have a risk of
water leakage it is recommended not to install it above electric
equipment or to take measures to protect the equipment from
dropping water. HEAT DISSIPATION
RO
DESCRIPTION
OM
Ex
amin
ation
room
LC Gantry / LP Gant
HEAT DISSIPATIO
HEAT DISSIPATIO
N (kW)
N (BTU/hr)
STA
AVE
MA
STA
AVE
MA
ND
RAG
X
ND
RAG
X
BY
E*
BY
E*
0.61
1.21
207
412
551
6
8
7
170
170
170
6
6
6
1.62
ry
Patient table
Large Display Moni
tor (LDM)
0.50
0.50
0.50
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
44 of 77
https://device.report/manual/8787491
Typical injector
0.09
0.09
0.09
320
320
320
TOTAL
1.20
1.80
2.21
410
615
754
2
4
3
DL user area with 1
0.16
0.16
0.16
546
546
546
-
-
1.00
-
-
341
TFT monitor
Advantage Worksta
tion (AW)
Additional LCD mo
2
0.12
0.12
0.12
409
409
409
0.28
0.28
1.28
79'-
79'-
363
7"
7"
'-1
nitor(s)
TOTAL
1"
C1 Frontal cabinet
0.71
1.29
1.59
T
echni
cal ro
C1 Lateral cabinet
0.31
0.86
1.16
242
441
543
1
2
5
105
294
396
7
6
9
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
45 of 77
om
C2 Cabinet
0.29
Coolix 4100 chiller
Frontal & Lateral @
https://device.report/manual/8787491
1.34
4.6
6.8
1.81
989
12.
457
617
1
1
15
23
42
4
8
60
832
474
991
0.12
0.14
0.18
410
478
614
0.42
0.42
0.42
141
141
141
8
8
8
221
221
221
6
6
6
125
125
125
7
7
7
125
125
125
7
7
7
341
341
341
2
2
2
60Hz (x2)
Chiller autotransfor
mer @60Hz(x2)
Detector condition
er (x2)
Power Distribution
0.6
0.65
0.65
Box (PDB)
UPS 3 KVA
UPS 3 KVA for LDM
LD cabinet
0.37
0.37
1.00
0.37
0.37
1.00
0.37
0.37
1.00
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
46 of 77
Fluoro UPS UL
TOTAL
1.38
https://device.report/manual/8787491
1.38
1.38
472
472
472
0
0
0
10.2
14.7
21.5
291
418
612
1
0
3
5'-
0'-
1'-
9"
1"
8"
WARNING
The list contains only the principal components of the system and
doesn't contain any non-GE supplied equipment.
This average corrisponds to 11 cases in 10 hours.
ELECTRICAL NOTES All wires specified shall be copper stranded,
flexible, thermo-plastic, color coded, cut 10 foot long at outlet boxes,
duct termination points or stubbed conduit ends. All conductors, power,
signal and ground, must be run in a conduit or duct system. Electrical
contractor shall ring out and tag all wires at both ends. Wire runs must
be continuous copper stranded and free from splices.
1. Aluminum or solid wires are not allowed.
2. Wire sizes given are for use of equipment. Larger sizes may be
required by local codes.
3. It is recommended that all wires be color coded, as required in
accordance with national and local electrical codes.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
47 of 77
https://device.report/manual/8787491
4. Conduit sizes shall be verified by the architect, electrical engineer or
contractor, in accordance with local or national codes.
5. Convenience outlets are not illustrated. Their number and location
are to be specified by others. Locate at least one convenience outlet
close to the system control, the power distritbution unit and one on
each wall of the procedure room. Use hospital approved outlet or
equivalent.
6. General room illumination is not illustrated. Caution should be taken
to avoid excessive heat from overhead spotlights. Damage can occur
to ceiling mounting components and wiring if high wattage bulbs are
used. Recommend low wattage bulbs no higher than 75 watts and
use dimmer controls (except MR). Do not mount lights directly above
areas where ceiling mounted accessories will be parked.
7. Routing of cable ductwork, conduits, etc., must run direct as possible
otherwise may result in the need for greater than standard cable
lengths (refer to the interconnection diagram for maximum usable
lengths point to point).
8. Conduit turns to have large, sweeping bends with minimum radius in
accordance with national and local electrical codes.
9. A special grounding system is required in all procedure rooms by
some national and local codes. It is recommended in areas where
patients might be examined or treated under present, future, or
emergency conditions. Consult the governing electrical code and
confer with appropriate customer administrative personnel to
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
48 of 77
https://device.report/manual/8787491
determine the areas requiring this type of grounding system.
10. The maximum point to point distances illustrated on this drawing
must not be exceeded.
11. Physical connection of primary power to GE equipment is to be made
by customers electrical contractor with the supervision of a GE
representative. The GE representative would be required to identify
the physical connection location, and insure proper handling of GE
equipment.
12. GEHC conducts power audits to verify quality of power being
delivered to the system. The customer's electrical contractor is
required to be available to support this activity.
All junction boxes, conduit, duct, duct dividers, switches, circuit
breakers, cable tray, etc., are to be supplied and installed by
customers electrical contractor.
Conduit and duct runs shall have sweep radius bends
Conduits and duct above ceiling or below finished floor must be
installed as near to ceiling or floor as possible to reduce run length.
Ceiling mounted junction boxes illustrated on this plan must be
installed flush with finished ceiling.
All ductwork must meet the following requirements:
1. Ductwork shall be metal with dividers and have removable,
accessible covers.
2. Ductwork shall be certified/rated for electrical power purposes.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
49 of 77
https://device.report/manual/8787491
3. Ductwork shall be electrically and mechanically bonded together
in an approved manner.
4. PVC as a substitute must be used in accordance with all local and
national codes.
All openings in raceway and access flooring are to be cut out and
finished off with grommet material by the customers contractor.
General contractor to insert pull cords for all cable run conduits
between the equipment room and the operators control room.
10 foot pigtails at all junction points.
Grounding is critical to equipment function and patient safety. Site
must conform to wiring specifications shown on this plan.
CONNECTIVITY REQUIREMENTS Your new GE Healthcare imaging
modality will require local and remote connectivity to enable our
full range of digital support:
Local connectivity: This allows your system to connect to local
devices such as PACS and modality worklist. We will require network
information to configure the system(s), and a live ethernet port(s)
prior to the delivery of the system(s).
Remote connectivity: Your GE Healthcare service warranty includes
InSite™ (applicable to InSite capable products), a powerful
broadband-based service which enables digital tools that can help
guard your hospital against equipment downtime and revenue loss
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
50 of 77
https://device.report/manual/8787491
by quickly connecting you to a GE Healthcare expert.
Depending on product family and software version, imaging
systems can be connected in one of the following methods:
1. TLS over TCP Port 443 (Preferred method for new products) via:
DNS resolution
Customer-provided Proxy or
GE Proxy (Available in some regions)
2. Site-to-Site IPsec VPN tunnel
Please provide the GE project manager with the contact information for
the resource that can provide information required to set up these
connections. GEHC will send out communication to these contacts,
which will include the project's Connectivity requirements, and a
Connectivity form. This form will need to be completed and returned to
GEHC prior to delivery of the system to ensure the system is tested and
connectivity is enabled prior to the completion of the installation.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
51 of 77
https://device.report/manual/8787491
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
52 of 77
IT
https://device.report/manual/8787491
ELECTRICAL LAYOUT ITEM LIST
E
M
1
24" x 24" x 12" [600 x 600 x 300] Box below floor, (1) 6" [150] di
a. suitable length threaded
pipe, (2) 6" [150] dia. locknuts and (4) 1" [25] dia. locknuts. (LC g
antry)
2
(1) GE supplied fitting. (1) 12" x 12" x 6" [300 x 300 x 150] Box,
(1) 6" [150] dia. bushing, (4) 1"
[25] dia. bushing for water lines (LC Gantry)
3
4"x4"x4" [100 x 100 x 100] flush wall junction box 12" [300] belo
w finished ceiling (x-ray
buzzer)
4
Empty 3" [75] conduit below floor for waterlines (LC chiller)
5
10" x 3 1/2" [250 x 89] surface wall duct with minimum 2 divider
s
6
18" x 3 1/2" [450 x 89] surface wall duct with minimum 2 divider
s
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
53 of 77
https://device.report/manual/8787491
7
8" x 8" x 6" [200 x 200 x 150] box under table (PDM/TRAM)
8
8" x 8" x 6" [200 x 200 x 150] box below floor (patient monitorin
g equipment)
9
Local service disconnect (LOTO)
10
Power Distribution Box
11
10" x 3 1/2" [250 x 89] below floor duct with minimum 2 divider
s
12
18" x 3 1/2" [450 x 89] below floor duct with minimum 2 divider
s
13
18" x 3 1/2" [450 x 89] above ceiling duct with minimum 2 divid
ers
14
10" x 3 1/2" [250 x 89] above ceiling duct with minimum 2 divid
ers
15
6" x 4" [150 x 100] riser duct from duct below to 1" [13] above fi
nished floor with minimum 1 2 divider
16
18" x 3 1/2" [250 x 89] vertical duct from duct above ceiling to fi
nished ceiling with minimum 2
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
54 of 77
https://device.report/manual/8787491
dividers
17
6"x6"x6" [150 x 150 x 150] flush ceiling box for waterlines (LP ga
ntry)
18
6"x6"x6" [150 x 150 x 150] flush ceiling box for waterlines (LP chi
ller)
Electrical Outlet Legend Additional Conduit Runs
Additional Conduit Runs
(Contractor Supplied and Installed)
From (B
To (Bubble # / Item)
Qty
ubble # /
Usable le
Size (in)
ngth
Item)
18
LAT
17
LAT Gant
1
59 ft.
3
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
55 of 77
Water Lin
https://device.report/manual/8787491
ry
e
4
AP
2
1
59 ft.
3
Warning light
1
-
12
6
1
-
12
120-V 1 phase power
1
-
As Req'd
Spooler
1
-
As Req'd
120-V 1 phase power
1
-
As Req'd
Spooler
1
-
Cables
Water Lin
AP Gantr
y
e
Light Sig
naling Co
ntrol Box
Light Sig
naling Co
C2 Cabin
et
ntrol Box
Light Sig
naling Co
ntrol Box
LED Trans
former
LED Trans
former
LED Lam
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
56 of 77
https://device.report/manual/8787491
p
come wit
h
spooler
3
X-R
6
1
90 ft.
11 2
1
90 ft.
11 2
2
-
3
2
68 ft.
As Req'd
Emergency off
1
-
12
Emergency off
1
-
12
ay Buzzer
3
X-R
net
5
ay Buzzer
6
Larg
C1F Cabi
Control R
oom
7
e Displa
TRAM/P
DM
y Cabinet
10
Po
wer Distri
6
20 kVA U
PS
bution B
ox
10
Po
wer Distri
bution B
ox
10
Po
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
57 of 77
https://device.report/manual/8787491
wer Distri
bution B
ox
10
Po
6
wer Distri
C1F Cabi
2
68 ft.
21 2
2
68 ft.
11 2
1
68 ft.
11 2
1
68 ft.
As Req'd
2
68 ft.
21 2
net
bution B
ox
10
Po
6
wer Distri
C1F Cabi
net
bution B
ox
10
Po
6
wer Distri
C2 Cabin
et
bution B
ox
10
Po
6
wer Distri
C2 Cabin
et
bution B
ox
10
Po
wer Distri
6
C1L Cabi
net
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
58 of 77
https://device.report/manual/8787491
bution B
ox
10
Po
15
Table (Ta
wer Distri
ble Powe
bution B
r)
1
-
Req'd
ox
9
LOT
2; As
480-V 3 phase power
1
-
As Req'd
10
1
-
As Req'd
1
-
3
2
-
3
O Discon
nect
9
LOT
Power Di
O Discon
stribution
nect
Box
5
Pati
16
Monitor
ent Moni
Bridge /
toring Co
Boom
nsole
8
Pati
ent Moni
7
TRAM/P
DM
toring Co
nsole
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
59 of 77
https://device.report/manual/8787491
POWER REQUIREMENTS
POWER SUPPLY
3 PHASES+N+G 380/400/415/480 V ±1
0%
FREQUENCIES for 380/400
50/60 Hz ± 3 Hz
/415 V
FREQUENCY for 480 V
60 Hz ± 3Hz
MAXIMUM INPUT POWER
150 kVA
(0.1 sec max)
NOMINAL LOAD (at 125 k
60 kVA
V, 100 ms, 640 mA):
MAXIMUM LINE RESISTAN
380 V : 0.09 Ω / 400 V : 0.096 Ω / 415 V :
CE PER 2 PHASES WIRES
0.102 Ω / 480 V : 0.12 Ω
(Ohm)
An EMI filter, provided by GE, is inside the PDB when Fluoro UPS CE is
used.
Neutral is mandatory for UPS control.
TNS neutral point connection must be used.
In case of IT or delta configuration without neutral, an isolation
transformer is needed (supplied by customer)
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
60 of 77
https://device.report/manual/8787491
Three-phase, 5 conductors (3 phase conductors, 1 neutral and 1
protective earth conductor)
Power supply should come into a power distribution box (PDB)
containing the protective units and controls.
The section of the supply cable should be calculated in accordance
with its length and the maximum line resistance per 2 phases.
SUPPLY CHARACTERISTICS
Power input must be separated from any others which may generate
transients (elevators, air conditioning, radiology rooms equipped with
high speed film changers ...)
All equipment installed with IGS system components must be
powered separately (e.g. lighting, power outlets)
Transients must be less than 2,000 V peak in common mode and
1,000 V in differential mode, with a duration limited to a few
microseconds.
GROUND SYSTEM
At least 35mm² copper from main ground point to the PDB.
The equipotential link will be by means of an equipotential bar. This
equipotential bar should be connected to the protective earth
conductors in the ducts of the non IGS cableways and to additional
equipotential connections linking up all the conducting units in the
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
61 of 77
https://device.report/manual/8787491
rooms where IGS units are located.
CABLES
Power and cable installation must comply with the distribution
diagram.
All cables must be isolated and flexible, cable color codes must
comply with standards for electrical installation.
The cables from signalling and remote control (Y, SEO, L...) will go to
PDB with a pigtail lenght of 1.5 m, and will be connected during
installation.
Each conductor will be identified and isolated (screw connector).
CABLEWAYS The general rules for laying cableways should meet the
conditions laid down in current standards and regulations, with regard
to :
Protecting cables against water (Cableways should be waterproof),
Protecting cables against abnormal temperatures (Proximity to
heating pipes or ducts),
Protecting cables against temperature shocks,
Replacing cables (Cableways should be large enough for cables to be
replaced) ,
Only GE cables are running inside cableways,
Metal cableways should be grounded.
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
62 of 77
https://device.report/manual/8787491
MANDATORY LOTO REQUIREMENTS
Core system: A wall circuit breaker or equivalent device with LOTO
capability must be installed on the mains line to the PBD. This device
must be compatible with the power input specifications of the
system. The customer is responsible for the procurement, delivery
and installation of this breaker.
LDM option: A wall circuit breaker or equivalent device with LOTO
capability must be installed on the mains line to the LDM UPS. The
rating of this device shall be 30A for UL and 16A for CE
configurations. The customer is responsible for the procurement,
delivery and installation of this breaker.
TYPICAL EQUIPOTENTIAL CONNECTIONS
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
63 of 77
https://device.report/manual/8787491
POWER DISTRIBUTION
L X-Ray room warning light control panel
L1 XRay ON light - 24V
SEO1-2 Emergency OFF button with two normally closed (NC)
contacts (1)
RL Room light circuit: room lights off during X-Ray generation
BR Injector wall outlet 10/16A+G
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
64 of 77
https://device.report/manual/8787491
(1) Location and/or quantity: refer to layout
NOTES: Max size of terminal block for PDB input cables: 4x3/0 AWG
FEEDER TABLE
MIN. FEEDER WIRE
SIZE, AWG OR MC
M (sq. mm)/VAC
480 VAC
MINIMUM FEEDER WIRE LENGTH - ft (m)
50
100
150
200
250
300
350
400
(15)
(30)
(46)
(61)
(76)
(91)
(10
(12
7)
2)
*1/
*1/
*1/
1/0
3/0
4/0
4/0
300
0 (5
0 (5
0 (5
(55)
(85)
(10
(10
M
5)
5)
5)
7)
7)
(15
0)
GENERAL NOTES
IN ALL CASES QUALIFIED PERSONNEL MUST VERIFY THAT THE FEEDE
R (AT THE POINT OF TAKE-OFF) AND THE RUN TO THE GE SYSTEM M
EET ALL THE REQUIREMENTS STATED IN THE PIM
FOR A SINGLE UNIT INSTALLATION, THE MINIMUM TRANSFORMER S
IZE IS 225KVA, WITH 2.4% RATED REGULATION AT UNITY POWER FAC
TOR. RESULTANT MAXIMUM ALLOWABLE FEEDER REGULATION IS 3.
6%
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
65 of 77
https://device.report/manual/8787491
THE GROUND WIRE TO EARTH SHALL BE A MINIMUM OF AWG #2 (U
L) OR 35 MM² (CE) OR THE SAME SIZE (100%) AS FEEDER WIRES, WHI
CH EVER IS LARGER.
WHEN A FLUORO UPS IS OR WILL BE INSTALLED, A NEUTRAL LINE IS
MANDATORY. IF IT SCHEME AS EARTHING SYSTEM IS USED, AN ISOL
ATION TRANSFORMER IS REQUIRED WITH DELTA-WYE OR DELTA-STA
R CONNECTION.
* MINIMUM WIRE SIZE FOR CIRCUIT BREAKER, BASED ON RECOMME
NDED OVERCURRENT PROTECTION.
INTERCONNECTIONS
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
66 of 77
https://device.report/manual/8787491
A
mandatory component of this drawing set is the GE Healthcare PreInstallation manual. Failure to reference the Pre Installation manual will
result in incomplete documentation required for site design and
preparation. Pre Installation documents for GE Healthcare products can
be accessed on the web at: www.gehealthcare.com/siteplannin GE does
not take responsibility for any damages resulting from changes on
drawings made by others. Errors may occur by not referring to the
complete set of final issue drawing. GE cannot accept responsibility for
any damage due to the partial use of GE final issue drawings, however
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
67 of 77
https://device.report/manual/8787491
caused. All dimensions are in millimeters unless otherwise specified. Do
not scale from printed pdf files. GE accepts no responsibility or liability
for defective work due to scaling from these drawings.
Documents / Resources
GE Healthcare INNOVA IGS 620-630 Interventional Neuro
radiology [pdf] Instruction Manual
INNOVA IGS 620-630 Interventional Neuroradiology, INN
OVA IGS 620-630, Interventional Neuroradiology, Neuror
adiology
References
Site Planning - Support | GE HealthCare (United States)
Documents - ge-healthcare,ge healthcare – INNOVA IGS 620-630
[pdf]
D Documents and Settings All Users Application Data Aut
odesk RAC 2011 Imperial Templates Annotations 5 114R S
heet A 1 INNOVA IGS 620 630 RENDERING 212058015 GE
HC SP www3 gehealthcare ~ media us support site planni
ng interventional gehc sp igs 114r |||
GE Healthcare has made reasonable efforts to
ensure this BIM file conforms to requirements
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
68 of 77
https://device.report/manual/8787491
of the applicable pre-installation manual. This
BIM file was produced to assist with site planni
ng and is not intended for construction or final
building modeling. GE Healthcare is not respo
nsible for any dama...
lang:en score:44 filesize: 225.78 K page_count:
3 document date: 2012-08-21
[pdf]
C Users 502673615 Documents 4 72 GEHC SP IGS 630 72R
gehealthcare jssmedia documents us global support site p
lanning interventional imaging innova igs 6x0 gehc sp 72r
rev 1 hash A5776D558BCD6D7F6B1CCEE42B44C1EA |||
GE Healthcare Project Management GE Health
care has made reasonable efforts to ensure thi
s BIM file conforms to requirements of the app
licable pre-installation manual. This BIM file wa
s produced to assist with site planning and is n
ot intended for construction or final building
modeling. GE Healthca...
lang:en score:33 filesize: 1.73 M page_count: 4
document date: 2017-01-24
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
69 of 77
[pdf]
User Manual
https://device.report/manual/8787491
Documentation
E6 Interconnections GEHC SP IGS 630 4 72F gehealthcare s
g jssmedia de4507ae2f594b12b563d972f68b24a3 la en |||
TYPICAL ---- ---- REV DATE MODIFICATIONS 0
1 - C1 - Cover Sheet 02 - C2 - Disclaimer - Site
Rea ... ucts can be accessed on the web at: ww
w.gehealthcare.com/siteplanning GE Healthcar
e Drawn by PMM INNOVA IGS 620-630 FINA
L STUDY Verified by Concession S.O. GON TST
- - GE does not take resp...
lang:en score:32 filesize: 2.27 M page_count: 2
2 document date: 2022-07-13
[pdf]
User Manual
Documentation
E6 Interconnections GEHC SP IGS 620 5 114F gehealthcare
ca jssmedia documents us global support site planning int
erventional imaging innova igs 6x0 gehc sp 114f rev 1 has
h 39287F231651918D4856137BE51A6B66 |||
TYPICAL ---- ---- REV DATE MODIFICATIONS 0
1 - C1 - Cover Sheet 02 - C2 - Disclaimer - Site
Rea ... ucts can be accessed on the web at: ww
w.gehealthcare.com/siteplanning GE Healthcar
e Drawn by PMM INNOVA IGS 620-630 FINA
L STUDY Verified by Concession S.O. GON TST
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
70 of 77
https://device.report/manual/8787491
- - GE does not take resp...
lang:en score:32 filesize: 2.2 M page_count: 22
document date: 2023-05-17
[pdf]
User Manual
Documentation
E6 Interconnections GEHC SP IGS 620 5 114F gehealthcare
ca jssmedia documents us global support site planning int
erventional imaging innova igs 6x0 gehc sp 114f rev 1 has
h 39287F231651918D4856137BE51A6B66 |||
TYPICAL ---- A 06/Dec/2018 Final study based
on DC- ---- REV DATE MODIFICATIONS 01 - C
1 - Co ... ucts can be accessed on the web at: w
ww.gehealthcare.com/siteplanning GE Healthc
are Drawn by PMM INNOVA IGS 620-630 FIN
AL STUDY Verified by Concession S.O. GON TS
T - - GE does not take resp...
lang:en score:31 filesize: 2.15 M page_count: 2
2 document date: 2022-03-17
[pdf]
User Manual
Documentation
E6 Interconnections GEHC SP IGS 620 5 114F gehealthcare
jssmedia documents us global support site planning interv
entional imaging innova igs 6x0 gehc sp 114f rev 1 hash 3
9287F231651918D4856137BE51A6B66 |||
TYPICAL ---- A 06/Dec/2018 Final study based
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
71 of 77
https://device.report/manual/8787491
on DC- ---- REV DATE MODIFICATIONS 01 - C1
- Co ... ucts can be accessed on the web at: ww
w.gehealthcare.com/siteplanning GE Healthcar
e Drawn by PMM INNOVA IGS 620-630 FINA
L STUDY Verified by Concession S.O. GON TST
- - GE does not take resp...
lang:en score:30 filesize: 2.24 M page_count: 2
2 document date: 2021-09-24
[pdf]
User Manual
Documentation
E6 Interconnections GEHC SP IGS 6X0 5 122F gehealthcare
jssmedia documents us global support site planning interv
entional imaging innova igs 6x0 gehc sp 122f rev 1 hash E
0BCB14B99DB0C13CA453FE3CAE213C9 |||
TYPICAL ---- A 23/Sep/2019 Initial release per
PIM revision 3 ---- REV DATE MODIFICATIONS
01 ... ucts can be accessed on the web at: www.
gehealthcare.com/siteplanning GE Healthcare
Drawn by RET INNOVA IGS 620-630 FINAL ST
UDY Verified by Concession S.O. GON TST - GE does not take resp...
lang:en score:30 filesize: 2.18 M page_count: 2
2 document date: 2021-12-10
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
72 of 77
[pdf]
User Manual
https://device.report/manual/8787491
Documentation
Accessories
Catalog
US Catalog GE Healthcare U S Accessories Optional record
er prints ECG strips on demand for manual review Touchsc
reen Operation trophon Sonex HL 6 bottles box 80 ml bott
le system reliability Operations included to supplement UP
S documentation UL HL™ E8350MC The cartridge contain
s a proprietary solution of ts us global products accesories
supplies new docs accessories catalog916cma gehealthcar
e media d9410ff1f36544ff940922f80cccc1bf Ceiling mount
configurations require catalog E8007PJ OCS III mounting p
late 100 kVA Bypass Panel Use with E4502FB latam jssmedi
a la es pa widen gehealthcarecom migrated rev f84fc8b29
42840a3ad1bad694377efe0 hash 7ACCCB43C0F63963E8A
F71773B8FC0A3 1 |||
GE Healthcare Accessories Supplies U. S. Catal
og CT, PET/CT MR Nuclear Medicine Ultrasoun
d X-ray Mammography Power Quality Adventu
re Series This catalog is searchable using contr
ol F on a PC or command F on a Mac. Rev. 01/
21 JB00708US Save time. Order accessories onl
ine with Service Sh...
lang:en score:28 filesize: 23.84 M page_count:
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
73 of 77
https://device.report/manual/8787491
308 document date: 2021-01-14
[pdf]
Section 1 Introduction GEMS ography acquisition systems
gehc ihe integration innova igs 620 630 5458749 8en rev2
latam gehealthcare jssmedia widen gehealthcarecom migr
ated rev b1890f5a5c7d44bb96ca4c9a6bd3f493 hash 7FA1
55CA3905C8C94AB7FE3E23E51123 |||
GE Healthcare InnovaTM IGS 620, 630 Integrati
ng the Healthcare Enterprise IHE Integration St
atement 5458749-1-8EN Rev 2 2012, General E
lectric Company, All Rights Reserved IHE Integr
ation Statement 5458749-1-8EN / rev 2 Revisio
n History Date Part/Rev January 10, 2012 5458
749-1-8EN rev 1 C...
lang:en score:27 filesize: 124.87 K page_count:
5 document date: 2012-03-22
[pdf]
Integrating the Healthcare Enterprise IHE Integration State
ment ConnectPro HIS RIS DICOM conformance statement
William Balloni GE Yokogawa Medical Systems ography ac
quisition systems gehc ihe integration innova igs 620 630
5714994 1 8en rev2 gehealthcare jssmedia widen gehealth
carecom migrated rev hash C5209FB8921DFEF5BA337912
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
74 of 77
https://device.report/manual/8787491
2C677C0D |||
Technical Publications GE Healthcare 57149941-8EN Revision 2 InnovaTM IGS 620, InnovaTM
IGS 630 Integrating the Healthcare Enterprise I
HE Integration Statement Copyright 2016 by G
eneral Electric Company Revision History Date
Part/Rev November 23, 2015 5714994-1-8EN r
ev 1 Change Descrip...
lang:en score:26 filesize: 56.46 K page_count: 6
document date: 2016-02-26
[pdf]
Accessories
Catalog
U S Accessories Supplies XrayInterventionalImaging US Ca
talog ChapterJuly 2021JB00644US gehealthcare jssmedia
gehc us images products imaging accessories xrayinterven
tionalimaging supplies catalog chapterjuly 2021jb00644us
rev 1 |||
GE Healthcare XR X-ray U.S. Accessories Suppli
es Rev. 07/21 JB00644US XR How to Order Sho
p online: gehealthcare.com/accessories Availab
le 24/7, GE Healthcare Service Shop makes ord
ering supplies online fast and easy. Search real
-time inventory of compatible, OEM products T
ake advantage of Auto R...
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
75 of 77
https://device.report/manual/8787491
lang:en score:26 filesize: 4.11 M page_count: 6
6 document date: 2021-07-30
[pdf]
Accessories
Catalog
Xray Accessories Supplies US Catalog Chapter 2016916JB3
9292USg1 latam gehealthcare jssmedia files x xray accesso
ries supplies us catalog chapter 2016916jb39292usg1 rev
bff463cf70b54945a9d08794ddaec654 hash CF5FCA6611B8
C000E3F8BBD68C0BA796 |||
GE Healthcare XR X-ray U.S. Accessories Suppli
es Rev. 01/21 JB00644US XR How to Order Sho
p online: gehealthcare.com/accessories Availab
le 24/7, GE Healthcare Service Shop makes ord
ering supplies online fast and easy. Search real
-time inventory of compatible, OEM products T
ake advantage of Auto R...
lang:en score:26 filesize: 4.29 M page_count: 6
6 document date: 2021-01-14
12/10/2023, 11:08 PM
GE Healthcare INNOVA IGS 620-630 Interventional Neuroradiology In...
76 of 77
[pdf]
Accessories
https://device.report/manual/8787491
Catalog
Imaging Accessories Supplies US CatalogJuly 2021JB00708
US gehealthcare media GEHC Files Products rev a0c8610c
b6cc4e7880fbbbd58f68bc5a |||
GE Healthcare Accessories Supplies U. S. Catal
og CT, PET/CT MR Nuclear Medicine Ultrasoun
d X-ray Mammography Power Quality Adventu
re Series This catalog is searchable using contr
ol F on a PC or command F on a Mac. Rev. 07/
21 JB00708US Save time. Order accessories onl
ine with Service Sh...
lang:en score:19 filesize: 26.68 M page_count:
308 document date: 2021-08-17
[pdf]
Documentation
interoperability dicom xray gehc conformance innovaigs62
0 630 5435421 1 8en rev2 gehealthcare jssmedia widen ge
healthcarecom migrated rev hash AC19A30FE290720B316
2EC6E21CE8B95 |||
GE Healthcare InnovaTM IGS 620, 630 Confor
mance Statement of DICOM OPERATING DOC
UMENTATION 5435421-1-8EN Revision 2 GE H
ealthcare InnovaTM IGS 620, 630 Conformance
Statement of DICOM Direction 5435421-1-8E
12/10/2023, 11:08 PM
N, Revision 2 2 GE Healthcare InnovaTM IGS 6
20, 630 Conformance Statement of DICOM Dir
ec...