Introduction of Microwave Test Bench Components 1 Microwave Test Bench It is a setup shown below which consists of some microwave devices. This whole setup with few alterations is able to measure guided wavelength, free space wavelength, cut-off wavelength, impedance, VSWR, power etc. 2 Klystron Power Supply The klystron power supply is shown in below figure. Klystron power supply is compact and regulated power source for reflex klystron tubes. It is generally used to generate microwave power in laboratories. Three power sources are required for reflex klystron operation: (1) positive voltage (often referred to as beam voltage), (2) negative voltage (often referred to as repeller voltage), and (3) filament power. It supplies regulated DC voltage to beam electrode, negative regulated voltage to reflector and 6.3 V AC to filament. Beam and repeller voltages can be continuously varied according to the requirement and can be monitored along with beam current regularly by a meter mounted on the front panel. The reflector supply can be internally modulated with a saw tooth wave for FM or with a square wave for obtaining AM. Square and saw tooth wave generators used for modulating the klystron output are integral part of the klystron power supply. The frequency of 1 KHz is provided. Typical Specifications: Beam Supply: Voltage: 250 to 300 volts Current : 50 mA (maximum) Reflector Supply: Voltage: 15 to 300 volts negative w.r.t cathode. Current: 10 μA (maximum) Filament Supply: Voltage: 6.3 volts DC ± 0.5 volts w.r.t cathode. Current: 10 A (maximum) 3 Reflex Klystron • The reflex klystron tube is a single cavity microwave tube that makes use of the principle of velocity modulation to generate microwave power. It is also known as electron beam modulated tube or cavity modulated tube or linear beam tube. • It is used mainly as a microwave signal source for general laboratory use and as a local oscillator in microwave receivers. • Efficiency of the reflex klystron is 20 to 30% and output power is 10 to 500 mW in frequency range of 1 to 25 GHz. • The frequency of operation can be changed by mechanical rotation of tuning screw – On doing so the dimension of resonant cavity changes and hence the frequency changes. 4 Klystron Mount Klystron mount is a waveguide of suitable length having octal base on the broad wall of waveguide and used to transmit microwave power from reflex klystron tube. It is designed with movable plunger at one end of the waveguide to direct the microwave power generated by reflex klystron tube. A small hole on the broad wall of the waveguide is provided through which coupling pin of reflex klystron tube enters into the waveguide. Maximum power transfer from reflex klystron tube can be achieved by tuning of movable plunger. 5 Waveguide Isolator • Isolator is a two port non-reciprocal microwave component used to isolate oscillators (reflex klystron in this case) from load i.e. to reduce load mismatch. • It means that it passes microwave signal in one direction and blocks signal going in the other direction. • A simple waveguide isolator contains a magnetized ferrite bar to which a resistive plate is attached. Microwave energy travelling in the forward direction is very little attenuated. In the reverse direction the energy is absorbed in the resistive plate. This effect is due to the non-reciprocal distortion of the RF field caused by the magnetized ferrite. 6 Variable Attenuator • An attenuator is a two port device that reduces the level of transmitted microwave power allowed to pass though it by either reflecting or absorbing the power. Waveguide attenuators consists of an attenuation plate (resistive surfaces known as resistive cards or thin metallic film or glass substrate) mounted within the waveguide by means of metal rods or by moving the plate in and out of the guide at a fixed location. • Attenuators are required to adjust the power flowing in a waveguide. Variable attenuators provide a convenient means of adjusting power level very accurately. 7 Frequency Meter • It consists of a cylindrical tunable cavity mounted with its axis perpendicular to the main waveguide and is coupled to the waveguide through small hole as depicted in Fig. (a). • The resonant frequency of the cavity can be varied by varying the position of plunger in the cavity. • A micrometer is attached to the plunger to indicate the frequency directly. • Output power versus frequency characteristics of the cavity frequency meter is shown in Fig. (b). The output of the cavity frequency meter decreases when the cavity is tuned through resonance. At frequency far from resonance the cavity presents an effective short circuit to the waveguide wall. • The frequency measurement would, therefore, consists of tuning the cavity of the frequency meter until maximum ‘dip’ occurs on the indicating VSWR meter and then reading the micrometer position and the corresponding frequency from the calibration chart provided with the frequency meter. (a) (b) 8 10.47 GHz 9.195 GHz 9 Waveguide Slotted Section • This system consists of a waveguide, a travelling probe carriage and facility for attaching instruments. The precession built probe carriage having centimeters scale with a vernier reading of 0.1 mm least count is used to measure the position of the probe. • The longitudinal slot is cut along the center of the waveguide broad wall. The probe (crystal detector) is made to move along the slotted wall which measures the electric field strength inside the waveguide. • In slotted line a small part of the electric field is fed to a crystal detector via a probe inserted in the waveguide. SWR or VSWR can be read directly on VSWR meter connected to the crystal detector output. This method is accurate only if: (a) the probe depth is small enough not to disturb the field in the waveguide. (b)the crystal detector works in the square law region i.e. the output voltage is proportional to the input power. 10 • The amount of coupling between waveguide and co-axial cable depends on diameter and depth of penetration of the probe. • The slotted section with tunable probe is used for monitoring standing wave patterns (determination of location of voltage standing wave maxima and minima along the line) into waveguide systems, measuring the VSWR and impedance of the unknown load. 14.49 cm 9.5 cm 11 Standing Waves and VSWR The electromagnetic field at any point of a waveguide may be considered as the sum of two travelling waves: the incident wave which propagates from the generator and the reflected wave which propagates towards the generator. The reflected wave is due to discontinuity in the waveguide or from a load impedance. The amplitude and phase of the reflected wave depends upon the amplitude and phase of the reflecting impedances and losses in the waveguide. The presence of two travelling waves gives rise to the standing waves along the waveguide. The electric and magnetic field varies periodically with distance. The maximum field strength is found where the two waves add in phase and the minimum is observed where the two waves add in opposite phase. Figure shows the standing wave pattern for different load impedances. The distance between two successive minima (or maxima) is half the guided wavelength. VSWR = Emax/Emin 12 VSWR Meter VSWR meter is essentially a low noise, high gain, tuned amplifier operating at fixed frequency generally 1 kHz, having an output meter calibrated to indicate directly in SWR (dB) when used with square law device. This device is primarily used for measuring SWR in conjunction with slotted line with a suitable detector. Specifications: Frequency: 1 kHz ± 2% Scale Selector: -5 dB, Normal and Expanded Meter Scale: SWR 1-4 SWR 3-10 Expanded SWR 1-1.3 Normal dB 0-10 Expanded dB 0-2 It reads more accurately on the upper half of the scale than on the lower half of the scale. Therefore, to bring all readings on the upper half of the scale a -5 dB attenuator is provided. 13 Matched Termination • Matched terminations are designed to absorb incident energy without appreciable reflections i.e. having VSWR of the order of 1.05. It is also known as matched load. It consists of a section of waveguide in which a tapered resistive card (absorbing material) is attached. • The reflections arising from the end are minimized by tapering the card. The card is placed parallel to the dominant TE10 mode at a place where the electric field is maximum to have maximum attenuation. • As the card has finite thickness, the reflections arising from it cannot be ruled out. Moreover, it is located at Emax which will give maximum reflected power. To avoid this the resistive card is kept closer to the side wall and its length is increased. 14 Procedure of Tuning Microwave Bench for Optimum Output Power • Make the connections as required according to the setup given. • Set the variable attenuator at maximum position. • Keep the control knobs of klystron power supply as below: Meter switch - ‘Off’ Mod-switch - AM Beam voltage knob - Fully anticlockwise Reflector voltage - Fully clockwise AM-Amplitude knob - Around fully clockwise AM-Frequency knob - Around mid-Position • If you are using VSWR meter, keep the control knobs of VSWR meter below: Range dB - 50 dB position Input switch - Crystal low impedance Meter switch - Normal position Gain (coarse & fine) - Mid Position • Switch ‘ON’ the klystron power supply, VSWR meter/CRO and cooling fan. • Rotate the meter switch of power supply to beam voltage position and set beam voltage at 250 V (you should not make beam voltage higher than 300V) with help of beam voltage knob (you should not touch this knob till the end of the experiment). • Adjust the reflector voltage to get some deflection in VSWR meter. • Maximize the deflection with AM amplitude and frequency control knob of power supply. • Tune the reflector voltage knob for maximum deflection. 15 • Tune the probe for maximum deflection in VSWR meter. Procedure of Measurements • Tune the frequency meter plunger to get a ‘dip’ on the VSWR scale and note down the frequency directly from the frequency meter. • Move probe along with the slotted line, the deflection in VSWR meter will vary. Move the probe to a minimum deflection position, to get accurate reading; it is necessary to increase the VSWR meter range dB switch to higher position. Note and record the probe position. • Move the probe to next minimum position and record the probe position again. • Calculate the guided wavelength as twice the distance between two successive minimum positions obtained as above. • Finally, calculate the cutoff wave length and broader dimension of the waveguide. Observations: • Resonant frequency from frequency meter: f0 = ______ GHz • For short circuit termination, note down the positions of four consecutive minima with the help of VSWR meter. Position of first minima, d1 = ______ cm Position of second minima, d2 = ______ cm Position of third minima, d3 = ______ cm Position of fourth minima, d4 =_____ cm 16 Calculations: • • • • • Take average, (let)A = [(d2-d1) + (d4-d3)] /2 Now λg can be calculated as 2*A Calculate λo from f0. From λg and λo, calculate λc with the help of equation (2). At last, from λc, calculate the broader dimension of waveguide, a = λc/2. 17