Uploaded by wissam.nsiri

M503-E028K Op manual

advertisement
Manual No. : M503-E028
Revision
:K
MOBILE X-RAY SYSTEM
(With FPD Upgrade Option)
OPERATION MANUAL
Read the operation manual thoroughly before you use the product.
Keep this operation manual for future reference.
About the Symbols Appearing in this Operation
Manual
Throughout the text in this manual, warnings and other information essential when using this unit, such as
cautionary or prohibited items, appear classified as per the following:
Mark
Description
Indicates an imminently hazardous situation which, if not avoided, will result in
serious injury or death.
Indicates a potentially hazardous situation which, if not avoided, could result in
serious injury or possibly death.
Indicates a potentially hazardous situation which, if not avoided, may result in minor
to moderate injury or equipment damage.
Emphasizes additional information that is provided to ensure the proper use of this
product.
*
Provides useful information about operation of this system. Please read the
description when required.
^ Reference
Provides reference information.
Revision History
Revision
Date
First Edition
Aug. 2008
Changes
⎯
Added "CXDI-55G/55C and 60C", "5.12 Second FPD Kit" , "5.17 FPD
Bag", "8.10 Manufacturer Information".
E
Nov. 2009
F
Sep. 2010
Correct "3 Basic Operation", "4 Advanced Operation", "8 Specifications".
G
Feb. 2011
Added "CXDI-70C Wireless".
H
Jul. 2011
Added "5.15 External Monitor Interface", "5.16 Barcode Reader".
J
Mar. 2012
Added "CXDI-80C Wireless", and compliant with Standards.
K
May 2012
Compliant with Standards.
Correct "2 Part Names and Descriptions", "3 Basic Operation", "4 Advanced
Operation", "8 Specifications".
Preface
Thank you for purchasing Shimadzu's mobile X-ray system, the MobileDaRt Evolution
(herein referred to as "the system"). This Operation Manual is designed to explain the
correct operation of this system. Be sure that you have fully read and understood this
Operation Manual prior to using the system. Also, for handling of MobileArt Evolution (FPD
Upgrade option), refer to this operation manual. For information regarding the DR system,
please refer to the Canon CXDI manual. Be sure only to use this system after you have
fully read and understood all included documents, including this Operation Manual,
attached documentation for the Canon CXDI-50G, CXDI-50C, CXDI-55G, CXDI-55C,
CXDI-60G, CXDI-60C, CXDI-70C Wireless, and CXDI-80C Wireless ("CXDI-50G/50C/
55G/55C/60G/60C/70C Wireless/80C Wireless"), and all warnings, cautions, and notes
contained in the Canon CXDI-50G/50C/55G/55C/60G/60C/70C Wireless/80C Wireless
manual. This manual does not cover the X-ray tube unit. Read the following manual for the
X-ray tube unit and Heating & Cooling Curve:
X-ray Tube Unit Model Name
X-ray Tube Unit
Document Number
0.7/1.3U163C-36
M535-E318
This manual should be kept available for future reference. If the user or usage location
changes, ensure that this operation manual is always kept together with the system.
Periodically check to be sure that the operation manual and the warning labels are not
missing or damaged. If they are, contact your Shimadzu service representative for
replacement. Original version is approved in English.
Notice
•
All copyrights regarding this manual are property of Shimadzu Corporation.
Neither all nor part of these contents shall be reproduced or duplicated without the express
permission of Shimadzu Corporation.
•
Content of this manual may be changed for improvement without notice. Although every
possible effort has been made to avoid errors while creating this manual, immediate revision
may not be possible in the event that errors or missing information are detected.
•
Screen images and illustrations contained in this manual may differ from those in actual use,
and are intended for example purposes only. Illustrations may also use partial images.
•
Mobile Art and Mobile DaRt are registered trademarks of Shimadzu Corporation in the
United States and Europe (Office for Harmonization in the Internal Market).
•
CXDI is a registered trademark of Canon Inc.
•
Microsoft and Windows are registered trademarks of Microsoft Corporation in the United
States and/or other countries.
•
Other company names and product names mentioned in this manual are trademarks or
registered trademarks of their respective companies.
•
The TM and ® symbols are omitted in this manual.
©2008-2012 Shimadzu Corporation. All rights reserved.
Preface
Operating Precautions
"Operating Precautions (for Both Safety and the Prevention of
Danger) in the Use of Electric Medical Equipment"
1.
Nobody without the following experience and knowledge should use the
system.
(1) Medical (radiographic) training (if particular qualifications are required in the
country concerned, those qualifications must be held).
(2) The capacity to read and understand the operation manual.
2.
When installing the system, pay attention to the following items:
(1) Do NOT install it near water faucet or similar equipment.
(2) Install it away from potential sources of problems such as abnormal pressure,
temperature or humidity, drafts, direct sunlight, dust, chlorine or sulphur gas.
(3) During transportation and operation of the system, avoid tilting, vibration and
sharp impact against it.
(4) Keep the system away from areas where chemicals or gases are stored.
(5) Use only the correct electrical power source with matching frequency, voltage
and current (or wattage).
(6) Check the conditions of the battery power source (power and polarity) before
operating the system.
(7) Properly ground the system.
3.
Before operating the system, pay attention to the following items:
(1) Check the conditions of switch contacts, polarity, dial settings, and meters,
and make sure the system performs correctly.
(2) Confirm that the ground is connected properly.
(3) Check all wiring for proper and correct connections.
(4) Pay attention when using more than one unit at a time, because it may lead to
an incorrect diagnosis and cause danger.
(5) Check the condition of the external electric circuit, which will be directly
connected to a patient.
(6) Check the condition of the battery power source.
ii
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
4.
1
While operating the system, pay attention to the following items:
(1) Do NOT exceed time or the amount of equipment use needed for diagnosis or
therapy.
1
(2) Observe the system and patient continuously for early detection of problems.
(3) When a problem is detected with the system or patient, take proper action to
stop the system without harming the patient.
1
(4) Do NOT let the system touch the patient.
5.
After operating the system, pay attention to the following items:
(1) Turn off the switches and return the dial to their original positions before use in
the prescribed order. Then, turn off the main power switch.
1
(2) Do NOT pull the power cable forcibly from the outlet.
(3) When storing the system, pay attention to the following items:
(i)
Keep it away from water.
(ii)
Store it away from potential causes of problems such as abnormal
pressure, temperature and humidity, draft, direct sunlight, dust, chlorine or
sulphur gas.
1
1
(iii) During transportation and storage of the system, avoid tilting, vibration and
sharp impact against it.
(iv) Store the system away from areas where chemicals and gases are stored.
(5) Keep the system clean to avoid problems during the next use.
1
When the system is found to be out of order, do NOT try to repair it.
Display an appropriate sign to indicate that the system is out of order, and
contact your Shimadzu service representative for repair.
1
(4) Clean all attachments, cables and contacts, and store them in one place.
6.
7.
Do NOT modify any part of the system.
8.
Preventive maintenance
1
(1) The system and its parts should be periodically checked.
(2) If the system has not been in operation for an extended period of time, test it
prior to actual operation to make sure it works correctly and safely.
9.
1
Concerning other items, operate properly according to the operation manual.
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
iii
Preface
Precautions in Usage
When using the system, please observe the following precautions for safety of the operator and
patient:
The responsibility for management of use and maintenance of medical
equipment lies with the user.
This system is restricted to use by, or under supervision of, a diagnostic
radiology technician or a person with a certificate indicating equal proficiency.
Repair and inspection of the inside of the system is dangerous. Be sure to
contact your Shimadzu service representative for repair and inspection.
Never modify the system.
In general, modifications are strictly prohibited by the Regulatory
requirements of the law of the country where the device is installed. Please
contact your Shimadzu service representative if it is necessary to modify the
system.
Perform periodic inspection.
Preventive maintenance is required to maintain long-term safety and performance of the
system.
The "7 Maintenance and Inspection" chapter in this manual gives detailed descriptions of
daily and periodic maintenance and inspection that a user should perform.
As for the maintenance and inspection that only specially trained experts can perform,
utilize the maintenance agreement program offered by Shimadzu.
iv
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
1
1
Repair and maintenance of the inside of the system can only be
performed by engineers assigned by Shimadzu.
Maintenance must be assigned to specially trained experts. Contact your
Shimadzu service representative for repair and maintenance.
1
1
1
In order to ensure safety when using the system, read the operation
manual provided with each system component for details on usage and
relevant precautions.
1
1
If the operator has no experience in operating the system, be sure that
he or she receives instruction on how to operate it from Shimadzu
service personnel or someone who has adequate experience in using
the system.
1
In order to operate the system safely, an explanation of the operation needs
to be given. When installing the system, Shimadzu service personnel explain
the operating procedure using this operation manual. Follow their directions
and operate the system correctly.
1
^ Reference "1.6 Operator Profile" P.1-9
1
1
Secure the means for the operator and the patient to communicate with
each other.
If system usage is deemed to put the patient at risk due to his or her
condition, refrain from conducting the study or treatment.
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
v
Preface
Be Sure to Read the Following to Prevent Explosion, Electric Shock,
or Injury
Do NOT use any potentially flammable or explosive gas, such as
disinfectant sprays, near the system.
Use of such gas may cause an explosion.
Check the condition of the patient before conducting a study.
If system usage is deemed to put the patient at risk due to the his or her
condition, refrain from conducting the study or treatment.
Do NOT use the system at places where the liquid may enter. Do NOT
spill liquids onto the surface or the inside of the system.
Otherwise, electric shock may occur.
In case of liquid spillage to the system, immediately turn off the power, and
contact your Shimadzu service representative.
When there is any abnormality in operation, or unusual smell or smoke
emission during operation, stop operation immediately and contact
your Shimadzu service representative.
Continued use may damage the system and cause injury.
vi
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
1
1
Do NOT open the covers of the system.
Otherwise, electric shock may result. When opening the covers for
maintenance, contact your Shimadzu service representative.
1
1
Do NOT use in a location where metal fragments may enter the system.
1
This can result in electric shocks.
1
Always be very careful when moving the system to avoid contact with
the patient or operator and to ensure that the patient or operator does
not become caught between the system and any neighboring devices.
1
Otherwise, it may cause injury.
1
1
Charge the battery in a well-ventilated room. Do NOT cover the system
while charging the battery.
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
vii
Preface
Cautions on Use and Storage
Connect the power cable plug to a 3-pin grounded socket.
Do NOT use the system in an oxygen-rich environment.
The use in an oxygen-rich environment may cause fatal or serious injuries or
damage to the system due to easy ignition.
Be sure to use the system under the following environmental
conditions:
The installation of a dedicated air-conditioner in the examination room
is recommended if the building air-conditioner cannot meet the
necessary environmental conditions.
• Atmosphere:
No explosive or corrosive gases
• Using/Storage temperature: 10 to 35 °C
• Relative humidity:
30 to 75 % (no condensation)
• Atmospheric pressure:
800 to 1060 hPa
• Environment luminosity:
150 to 500 lx
Note also that there must be no sudden change in temperature or
humidity.
This causes condensation, which can lead to system failure.
viii
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
Cautions on Radiography
1
1
Restrict all persons other than the patient from accessing the
equipment in accordance with local regulations.
1
To avoid unnecessary exposure, acceptable distances (maximum access
values) to the equipment by any person other than the patient are defined for
each region.
1
1
The equipment can be operated only by qualified personnel, such as
radiology technicians or those with equivalent qualifications.
1
1
No person but the patient is allowed to stay in the examination room
during X-ray irradiation.
1
If the equipment is not used correctly, the operator, the patient, and other
persons may receive a greater dose of radiation than necessary. If for some
reason another person has to be in the examination room, that person must
control X-ray irradiation from a distance over 2 m from the focal spot and the
X-ray beam and also must take adequate measures to protect themselves
against radiation (protective apron, screen, etc.).
1
* During radiography, the radiography indicator illuminates and the buzzer
sounds an audible warning.
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
ix
Preface
Perform X-ray irradiation carefully and according to the doctor's
directions when using the system with expectant mothers, women who
suspect they are pregnant, lactating women, or children.
Particular ways of using the system may increase the scatter dose absorbed
into the patient, which may cause a radiation hazard.
Always check the X-ray exposure region using the collimator lamp.
Irradiating a patient with X-rays outside the required region risks exposure of
the patient to unnecessary radiation.
During X-ray irradiation, ensure that the X-rays irradiate the necessary
region only.
To avoid unnecessary exposure, narrow down the collimator and take
protection measures, such as wearing a protective apron.
Do NOT place any unnecessary object in the location within the X-ray
exposure region.
Doing so may result in unnecessary radiation exposure to the patient.
x
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
1
1
Be sure to carry out a warm-up (running-in operation of the X-ray tube
unit) before taking an X-ray radiograph.
Follow the warm-up procedure described in the X-ray high voltage generator
operation manual.
1
^ Reference "3.2 Warm-up the X-Ray Tube Unit" P.3-6
1
1
Perform the warm-up if an arc occurs.
Suddenly using the X-ray tube unit near the nominal X-ray tube voltage
(above 100 kV) after using the unit at a relatively low tube voltage (80 kV
max.) for a prolonged period may result in arc. This arc occurs due to loss of
the warm-up effect at high tube voltage after the X-ray tube unit is used at a
relatively low tube voltage for a prolonged period.
1
^ Reference "3.2 Warm-up the X-Ray Tube Unit" P.3-6
1
1
Do NOT perform unnecessary standby operations.
1
If standby status continues after the radiography preparation button is
pressed, wire disconnection or withstand voltage failure may result owing to
the evaporation of the X-ray tube filament.
1
1
In order to minimize the radiation dose on the patient, make the
distance between the focus and the patient's body surface as long as
possible (Minimum 45 cm).
1
The shorter the distance becomes, the greater the amount of scatter dose
absorbed into the patient, which may cause a radiation hazard.
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xi
Preface
Pay extra attention when irradiating X-rays for a long time or repeatedly.
It may cause a radiation hazard.
Avoid prolonged contact with the X-ray tube unit directly after
radiography.
Prolonged contact may cause low temperature burn.
System Movement Cautions
When moving the arm unit or system, the operator must pay close
attention to ensure that the limbs of both operator and patient are not
caught between the system and any peripheral units.
Be especially careful when near a wall, changing the direction of the system
or moving it backwards.
xii
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
1
1
Avoid collisions and overturning.
Collisions and overturning can not only damage the system, but can also
cause serious injury.
1
Be careful of the following points when moving the system:
• When on a slope, in an entrance, or encountering obstacles, move the
system as slowly as possible. Be especially careful when descending/
ascending a steep slope.
1
• Do NOT move the system on a slope steeper than 7 degrees or over steps
higher than 2 cm. When stopping the system, always park the system on a
level place, avoiding slopes, since there is a danger of the system rolling
down the slope and overturning.
1
• Moving the system along surfaces that cause strong vibrations, such as
roads outdoors, over large steps and down steep slopes is dangerous.
Avoid these if possible. Do NOT stop the system on a slope.
1
• When moving the system, look out for any obstacles in front of it.
1
Cautions on Cleaning and Disinfection
1
1
Be sure to turn the system power OFF before cleaning and disinfecting
the system.
Otherwise, a malfunction may occur in the system, or the system may
operate in an unintended way.
1
Also, thoroughly ventilate the room before turning ON the power after
disinfection work is complete.
1
1
Be sure to clean and disinfect the system.
Cleaning and disinfection is very important to ensure that the system can be
used hygienically and safely. Strictly follow the methods prescribed.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
1
xiii
Preface
Do NOT directly apply rubbing alcohol or water, or spray it onto the
system.
Wipe the surface of the system with a cloth soaked with rubbing alcohol.
If rubbing alcohol gets inside the system, it can cause failure or accidents.
To clean the dose area product meter, wipe the surface with cloth soaked in
pH-neutral cleaner and which is thoroughly wrung out.
Do NOT use an organic solvent.
Organic solvents may change the surface color. If an organic solvent adheres
to the surface, wipe it out immediately.
Do NOT use the following disinfectants:
If any of the following disinfectants are applied, the system performance and
safety cannot be guaranteed.
• Chlorine-based disinfectants
• Disinfectants that corrode metals, plastics, rubber, or paint
• Disinfectants unsuitable for metals, plastics, rubber, or paint
• Spray-gas type disinfectants
• Volatile disinfectants
• Disinfectants that may enter the system
xiv
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
1
Use disinfectants at a minimum.
1
Repeated disinfection over a long time may lead to discoloring and cracking
on the system surface, and deterioration of rubber and plastic. If any
abnormality is found on the system after disinfection, stop using the system
immediately. Contact your Shimadzu service representative for repair.
1
1
•
Never use paint thinners or other organic solvents, other than rubbing alcohol for
disinfection.
•
Film-developing or fixing solutions will cause discoloration.
Wipe off immediately with a damp cloth.
•
Make sure no moisture enters the dose area product meter during cleaning. After
cleaning, dry it thoroughly before use.
1
1
1
Cautions Relating to Cellular Telephones
1
1
Do NOT bring any cellular telephones or related devices into the
examination room with their power ON.
Such devices can exceed the EMC standard limitations, and under some
conditions this can impair the proper functioning of the system. In the worst
case, this can cause serious injuries or clinical errors.
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xv
Preface
Cautions on Electromagnetic Compatibility
This system needs special precautions regarding EMC.
Install and use the system according to the EMC information provided in this
operation manual.
^ Reference "8.1 Environmental Conditions of EMC (Electromagnetic Compatibility)" P.8-2
Make sure that electromagnetic compatibility is obtained.
All peripheral devices must satisfy EMC standards regarding emission of
electromagnetic energy and susceptibility to electromagnetic environment.
Devices that do not satisfy these standards may disturb the correct
functioning of the system. In the worst case, this can cause serious injuries or
clinical errors.
^ Reference "8.1 Environmental Conditions of EMC (Electromagnetic Compatibility)" P.8-2
Do NOT use this system adjacent to, or stacked with, other equipment.
If adjacent or stacked use is necessary, check to be sure that this system
works properly in the environment.
^ Reference "8.1 Environmental Conditions of EMC (Electromagnetic Compatibility)" P.8-2
xvi
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
Cautions on Wireless Communication Using the CXDI-70C Wireless/
80C Wireless
1
1
This system operates in the same frequency band as premises radio stations
for identifying mobile devices in factory production lines (license required)
and specified low-power radio stations (license not required) in addition to
industrial devices (ex. microwave ovens), scientific instruments, and medical
devices. Other equipment may interfere with the system even if such
equipment complies with CISPR EMISSION requirements.
The use of this system may result in RF interference with the above
mentioned equipment and radio stations. Make sure you understand and
heed the following cautions when operating the system.
1
1
1
• Before using this system, make sure that no RFID premises radio stations
and specified low-power radio stations or similar equipment are used in
the immediate area.
• If this system causes damaging RF interference to affect an RFID
premises radio station, stop using the system immediately and contact
your Shimadzu service representative.
1
1
• If this system causes damaging RF interference to affect an RFID
specified low-power radio station or amateur radio station, contact your
Shimadzu service representative.
• This system may suffer interference from other equipment that emits radio
waves (such as microwave ovens, Bluetooth devices, and digital cordless
telephones). Use the system after moving such devices as far away as
possible to prevent interference.
1
1
1
1
Do NOT use the selected frequency channel (2.4 GHz/5 GHz band) for
other wireless devices.
Mutual interference between the system and other devices may result in
decreased image data transfer speeds.
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xvii
Preface
Do NOT cover the wireless antenna of the imaging unit with your hands.
Covering the antenna may hinder the wireless communication properties
(throughput and range) of the system.
xviii
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
Cautions Regarding Connection with an External Device
1
1
When using the system, only the following electrical devices can be
connected to the system's LAN connector, USB connector, and DVI
(Digital Visual Interface) connector.
1
• Medical electrical devices which conform to IEC60601-1
1
• Non-medical electrical devices which conform to related IEC safety
standards
• Non-medical electrical devices which have safety equal to devices which
conform to IEC safety standards
1
In addition, please note the following:
1.
When connecting medical electrical devices which conform to
IEC60601-1
1
Connect the system and a connected device with a common protective
ground, or connect an additional protective ground of the connected
device when that device is in a room not used for treatment and
diagnosis (office and storage room, etc.).
2.
1
When connecting other electrical devices
(1) Connect an additional protective ground of the connected device, or
supply electrical power to the connected device from an isolated power
source when the connected device is in the room used for treatment and
diagnosis.
1
(2) Connect the system and a connected device with common protective
ground, or connect an additional protective ground of the connected
device when that device is in a room not used for treatment and diagnosis
(office and storage room, etc.).
1
1
Table.
Necessary measures when connecting electrical device to this system
Location of Connected Device
Connected Device
Diagnosis and Treatment Room
Medical electrical devices
which conform to
IEC60601-1
Other electrical devices
⎯⎯⎯⎯⎯⎯
Additional protective ground or
power supply from an isolated
power source
Other
1
Common protective ground or
additional protective ground or
separation device
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xix
Preface
Scope of Software License
Do NOT change any of the OS settings.
The system uses Windows XP or Windows 7 as its operating system (OS).
(Settings differ depending on the FPD.)
If the OS settings are changed, the system operation is not guaranteed.
1.
Definition of licensed software
The licensed software ("software") refers to all computer programs
used by the system, in addition to all related documentation.
2.
Ownership of license
The software used in this system is copyrighted by Shimadzu
Corporation ("Shimadzu"), who possesses all rights, including
sublicenses for those rights (for copyrights, etc.) held by third parties.
The system with the installed software is sold on the basis of Shimadzu
licensing the software ownership to the customer. Accordingly, when
the customer uses the software, they must observe the items outlined
below:
(1) The customer shall only use the software for use with a single system.
(2) The intellectual property rights for the software are not transferred to the
customer.
(3) The customer, or any third party, is prohibited from performing any of the
following actions:
•
•
•
•
xx
Duplicating the software.
Changing the software, in whole or part.
Transferring, loaning, or sublicensing the software.
Transferring the software outside Japan without the prior permission of the
Japanese or U.S. government.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
Data Security
1
1
•
•
If individual data stored in the system is either lost or destroyed, Shimadzu may
provide system maintenance to prevent future data loss. Shimadzu does not
guarantee the retrieval of any lost or damaged data.
1
Data stored in the system is always subject to the risk of loss or damage. Loss of
or damage to data may result from system failure or malfunction, and may also
arise as a result of the deterioration of part performance due to age. To avoid data
loss or damage, users are recommended to save all essential data to a hard
media, such as film, or store it on an external file system. Data stored on external
file systems are also subject to the risk of loss or damage. Regularly check the
contents of data stored on external file systems.
1
1
1
1
1
1
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xxi
Preface
Warranty
The guarantee period of this system is one (1) year from the date of purchase.
The system is warranted to be free from defects in material and workmanship for one year from the date of
delivery. If found to be defective, the system must be offered to Shimadzu for inspection and examination. Upon
examination, Shimadzu, at its sole option, will repair or replace at no charge, the system or any part found to be
defective. Components which wear are not warranted.
This warranty extends to original purchaser or the lessee of the new system only.
If the system is to be resold or delivered to a third party, such third party must be provided with a copy of this
manual, the installation manual and the technical manual supplied with the system.
This warranty does not apply to the following:
1.
Failure or damage due to any installation, relocation, or service not
provided by your SHIMADZU service representative or a SHIMADZU
designated contractor.
2.
Failure or damage caused by the product of other companies (except
those purchased from SHIMADZU).
3.
Failure or damage due to repairs using non-SHIMADZU certified service
parts.
4.
Failure or damage due to non-compliance with the notices and
procedures set forth in this manual or in the Operation Manual for Canon
parts.
5.
Failure or damage due to any operating environment deviating from the
requirements set forth in this manual or in the Operation Manual for
Canon parts.
6.
Failure or damage due to natural disasters such as power surge, rain,
fire, earthquake, flood, and thunder.
7.
Failure/damage caused by installing software or connecting peripheral
devices not specified by Shimadzu.
8.
Failure/damage caused by changing or deleting the basic software's
configuration files (autoexec.bat, config.sys, etc.) or DR system specialuse files.
Service after the expiration of the warranty is available at a reasonable cost and should be performed
by your Shimadzu service representative.
IN NO EVENT SHALL SHIMADZU AND ITS AFFILIATED ENTITIES BE LIABLE TO ANY
PERSON OR ENTITY FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL OR
CONSEQUENTIAL DAMAGES (INCLUDING, WITHOUT LIMITATION, ANY DAMAGES
RESULTING FROM LOSS OF USE, BUSINESS INTERRUPTION, LOSS OF PROFITS, LOSS
OF SAVINGS, THE COST OF PROCUREMENT OF SUBSTITUTED GOODS, SERVICES OR
TECHNOLOGIES OR FOR ANY MATTER ARISING OUT OF OR IN CONNECTION WITH
THE USE OR INABILITY TO USE THE SYSTEM.In some jurisdictions, some of the foregoing
warranty disclaimers or damage limitations may not apply.
Shimadzu will be indemnified for any claim, liability, or damage arising out of the misuse or noncompliance with this manual by the purchaser or lessee of the system.
xxii
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Preface
Software Version
1
This operation manual corresponds to version 1.4 of the system control software.
The system control software does not include the software for the CXDI-50G/50C/55G/55C/60G/
60C/70C Wireless/80C Wireless.
Service Life
1
1
The system lifetime is 7 years (based on Shimadzu's criteria) assuming the specified maintenance
checks are performed.
Disposal Precautions
1
1
1
When disposing of the system, contact your Shimadzu service
representative.
An improper disposal of this system may pollute the environment by
substances contained in parts.
1
Action for Environment (WEEE)
1
To all users of Shimadzu equipment
in the European Union:
1
Equipment marked with this symbol indicates that it was sold on or after 13th August 2005, which
means it should not be disposed of with general household waste. Note that our equipment is for
industrial/professional use only.
1
1
WEEE Mark
Contact your Shimadzu service representative when the equipment has reached the
end of its life. They will advise you regarding the equipment take-back.
With your co-operation we are aiming to reduce contamination from waste electronic and electrical
equipment and preserve natural resource through re-use and recycling.
Do not hesitate to ask Shimadzu service representative, if you require further information.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xxiii
1
1
Preface
Warning and Caution Labels
The following safety labels, which describe handling precautions, are attached to the system. With
adequate understanding of the contents on these labels and the warning/caution items in this manual,
operate the system safely.
Inspect the safety labels periodically (once a year).
If any label is peeled or unreadable by stain or scratch, replace it with a new one.
For new labels, contact your Shimadzu service representative.
1
23
89
6
0
4
5
xxiv
MobileDaRt Evolution OPERATION MANUAL
7
M503-E028
Preface
1
1
2
1
3
1
4
1
5
1
1
6
1
7
1
8
9
1
1
0
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
xxv
Preface
Abbreviations
The following are the abbreviations used in this manual.
Abbreviations
xxvi
Description
APR
Anatomical Program
DICOM
Digital Imaging and Communication in Medicine
EMC
Electromagnetic Compatibility
FPD
Flat Panel Detector
LED
Light Emitting Diode
MWM
Modality Worklist Management
PACS
Picture Archiving and Communication Systems
SID
Source Image Distance
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Contents
Contents
Preface
Precautions in Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Be Sure to Read the Following to Prevent Explosion, Electric Shock, or
Injury . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vi
Cautions on Use and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Cautions on Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
System Movement Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Cautions on Cleaning and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Cautions Relating to Cellular Telephones . . . . . . . . . . . . . . . . . . . . . . . xv
Cautions on Electromagnetic Compatibility . . . . . . . . . . . . . . . . . . . . . . xvi
Cautions on Wireless Communication Using the CXDI-70C Wireless/80C
Wireless . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvii
Cautions Regarding Connection with an External Device . . . . . . . . . . . xix
Scope of Software License . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xx
Data Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxi
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxii
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiii
Service Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiii
Disposal Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiii
Action for Environment (WEEE)
To all users of Shimadzu equipment
in the European Union: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiii
Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . .xxiv
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xxvi
Chapter 1 Overview
1.1 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2 Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.4 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.4.1 Operation Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.4.2 Transportation and Storage Environment . . . . . . . . . . . . . . . . . . . . . . 1-5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
Contents-1
Contents
1.4.3 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1.4.4 Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.4.5 Wireless Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.5 Classification of Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Protection Method Against Electric Shock . . . . . . . . . . . . . . . . . . . . . . 1-8
Classification of Applied Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Operation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Degree of Protection Against Liquid Ingress . . . . . . . . . . . . . . . . . . . . 1-8
For Use in an Oxygen-rich Environment . . . . . . . . . . . . . . . . . . . . . . . 1-8
For Use in Flammable Atmosphere . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Classification of Installation Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.6 Operator Profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
1.7 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Chapter 2 Part Names and Descriptions
2.1 External Appearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2 Grip Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Emergency Brake Release Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.3 Power Supply Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Status Indicator Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.4 X-Ray Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2.5 Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Status Indicator Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.6 Hand Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.7 Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2.8 Apron Hanger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
2.9 Disposable Wipes Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2.10 Imaging Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.10.1CXDI-50G/50C/55G/55C/60G/60C . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Usage of CXDI-55G/55C/60G/60C . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.10.2CXDI-70C Wireless/80C Wireless . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Usage of CXDI-70C Wireless/80C Wireless . . . . . . . . . . . . . . . . . . . 2-21
2.11 Storage Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2.11.1CXDI-50G/50C/55G/55C/60G/60C . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
One Imaging Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Two Imaging Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Contents-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Contents
1
2.11.2CXDI-70C Wireless/80C Wireless . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2.12 Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.13 Main Circuit Breaker and Additional
Protective Grounding Wire . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.13.1Main Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.13.2Additional Protective Grounding Wire . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.14 LAN Connector for External Connection . . . . . . . . . . . . . . . . 2-27
1
1
1
Chapter 3 Basic Operation
3.1 Starting and Shutting Down the System . . . . . . . . . . . . . . . . . . 3-2
3.1.1 Starting the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.1.2 Shutting Down the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Auto-Power-OFF Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2 Warm-up the X-Ray Tube Unit . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
1
1
3.2.1 Warm-up Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.3 Moving the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3.3.1 Imaging Unit Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.3.2 Travel Mode and Inch-Mover Buttons . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Normal Travel Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Low-Speed Travel Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Inch-Mover Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.3.3 Obstruction Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Obstruction Detector Cancel Function . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3.3.4 Emergency Stop Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Releasing the Emergency Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
1
1
1
3.4 Arm Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.4.1 Locking and Releasing the Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Releasing the Arm Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Setting the Arm Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.4.2 Positioning the X-Ray Tube Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3.4.3 Setting the Irradiation Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Adjusting the Irradiation Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Irradiation Field Reference Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
1
1
3.5 Setting the Radiography Parameters . . . . . . . . . . . . . . . . . . . 3-21
M503-E028
3.5.1 Setting the Radiography Parameters . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
3.5.2 Setting the Radiography Parameters
Using the Anatomical Programs (APR) . . . . . . . . . . . . . . . . . . . . . . . 3-21
Selecting an Anatomical Program for DR Radiography Mode . . . . . . 3-21
1
Selecting an Anatomical Program for Cassette Radiography Mode . 3-22
1
MobileDaRt Evolution OPERATION MANUAL
Contents-3
Contents
3.5.3 Manually Setting the Radiography Parameters . . . . . . . . . . . . . . . . . 3-23
Setting the X-ray Tube Voltage (kV) . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Setting the Current-Time Product (mAs) . . . . . . . . . . . . . . . . . . . . . . 3-23
Changing the Focal Spot (only for High Power Type (32 kW)) . . . . . 3-23
3.6 X-Ray Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
3.6.1 DR Radiography Preparation
(for the CXDI-70C Wireless/80C Wireless) . . . . . . . . . . . . . . . . . . . .
3.6.2 DR Radiography Procedure
(Changing to the DR Radiography Mode) . . . . . . . . . . . . . . . . . . . . .
3.6.3 Transferring Images to an External Printer or PACS System . . . . . .
3.6.4 Saving Images to External Media . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.6.5 Cassette Radiography Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.6.6 Low-Battery Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current-Time Product Limits for Low Battery Voltage . . . . . . . . . . . .
3-24
3-25
3-29
3-30
3-31
3-33
3-33
3.6.7 Additional Radiography Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
3.7 Charging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
3.7.1 Charging After Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
3.7.2 Charging Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
3.8 Imaging Unit Storage (CXDI-70C Wireless/80C Wireless) . . . 3-38
Chapter 4 Advanced Operation
4.1 User Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.1.1 Entering the User Settings Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Registering Changed Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Not Registering Changed Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.1.2 Personal ID Setting [Pid] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.3 Power-Assist Setting [PAS] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.4 Acceleration Setting [Acc] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.5 Steer Control Setting [stc] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.6 Backward Power-Assist Setting [bAC] . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.7 Auto-Power-Off Time Setting [APO] . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.8 Buzzer Setting [bon] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.9 Collimator Lamp Control Setting [LAP] . . . . . . . . . . . . . . . . . . . . . . . .
4.1.10APR Settings at Startup [aPr] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.11Operation Sound Setting [Snd] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pattern Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-4
4-4
4-5
4-5
4-6
4-6
4-7
4-7
4-8
4-9
4-9
Setting the Pattern Tone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Setting the Sound Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Setting the sound emitted during X-ray exposure. . . . . . . . . . . . . . . . 4-10
Setting the sound emitted when radiography preparation is finished. 4-11
Contents-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Contents
1
Setting the sound emitted during system travel . . . . . . . . . . . . . . . . . 4-11
Checking the Sound Pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.2 Changing the Operation Sound Volume . . . . . . . . . . . . . . . . . 4-12
4.2.1 Changing the Sound Volume Emitted During X-Ray Exposure
or When Radiography Preparation Is Finished . . . . . . . . . . . . . . . . . 4-12
4.2.2 Using the Collimator Lamp Button to Adjust Volume . . . . . . . . . . . . . 4-13
4.3 Changing Anatomical Program Parameters . . . . . . . . . . . . . . 4-14
4.3.1 Registering Anatomical Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Registration Procedure for DR Radiography . . . . . . . . . . . . . . . . . . . 4-14
Registration Procedure for Cassette Radiography . . . . . . . . . . . . . . 4-16
4.3.2 Deleting Unused Anatomical Programs . . . . . . . . . . . . . . . . . . . . . . . 4-17
Deleting Unused Anatomical Programs . . . . . . . . . . . . . . . . . . . . . . . 4-17
How to Use the Deleted Anatomical Program . . . . . . . . . . . . . . . . . . 4-18
Chapter 5 Optional Features
1
1
1
1
1
5.1 Remote Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.1.1 Using the Remote Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.1.2 Operable Range of the Remote Controller . . . . . . . . . . . . . . . . . . . . . 5-4
5.1.3 Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
1
5.2 Grid Unit and Grid Frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5.2.1 Grid Unit (CXDI-50G/50C/55G/55C/60G/60C) . . . . . . . . . . . . . . . . . . 5-6
5.2.2 Grid Frame (for the CXDI-70C Wireless/80C Wireless) . . . . . . . . . . . . 5-7
1
5.3 Protective Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.3.1 Using the Shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.4 Dose Area Product Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
5.4.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
VacuDAP 1570015 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
VacuDAP 1580015 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
5.4.2 Using the Dose Area Product Meter . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
5.4.3 Cleaning the Dose Area Product Meter . . . . . . . . . . . . . . . . . . . . . . . 5-12
1
1
5.5 Dose Calculation Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5.5.1 Using the Dose Calculation Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
5.6 Distance Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
1
5.7 Keyless Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5.7.1 Using the Keyless Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5.8 Changing the Grip Bar Height . . . . . . . . . . . . . . . . . . . . . . . . 5-19
1
5.9 Luminous Hand Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
5.9.1 Using the Luminous Hand Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
M503-E028
MobileDaRt Evolution OPERATION MANUAL
Contents-5
1
Contents
5.10 Additional Hand Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5.11 Wireless LAN Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
5.11.1Using the Wireless LAN Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
5.11.2Wireless LAN Adapter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
5.12 Second FPD Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
5.12.1For the CXDI-55G/55C/60G/60C . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Using the Sensor Unit Attach/Detach Function . . . . . . . . . . . . . . . . . 5-24
Storage case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.12.2For the CXDI-70C Wireless/80C Wireless . . . . . . . . . . . . . . . . . . . . . 5-27
Using the Sensor Unit Attach/Detach Function . . . . . . . . . . . . . . . . . 5-27
Storage case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
5.13 Wired Connection Kit
(for the CXDI-70C Wireless/80C Wireless) . . . . . . . . . . . . . . . 5-28
5.13.1Switching from a Wireless to Wired Connection . . . . . . . . . . . . . . . . 5-28
5.13.2Switching from a Wired to Wireless Connection . . . . . . . . . . . . . . . . 5-30
5.14 Handle Unit
(for the CXDI-70C Wireless/80C Wireless) . . . . . . . . . . . . . . . 5-31
5.15 External Monitor Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
5.15.1Using the External Monitor Interface . . . . . . . . . . . . . . . . . . . . . . . . . 5-33
5.15.2Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
5.16 Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-35
5.16.1Using the Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-35
5.17 FPD Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-37
Chapter 6 Troubleshooting
6.1 Warning Indicator Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.2 Malfunction Indicator Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6.3 Releasing the Brake in Emergency . . . . . . . . . . . . . . . . . . . . . . 6-5
6.3.1 Activating the Brake Release Function . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.3.2 Releasing the Brake in Emergencies . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.4 Notification Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.4.1 Confirming the Usage Period, Count and Distance Information
for Each Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
6.4.2 Temporarily Disabling the Replacement Time Display Function . . . . . 6-8
6.5 Storing the System for a Long Period . . . . . . . . . . . . . . . . . . . 6-10
6.5.1 Storage Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Contents-6
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Contents
1
Chapter 7 Maintenance and Inspection
1
7.1 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
7.1.1 Daily Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Inspection for Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . 7-6
7.1.2 Periodic Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Periodic Inspection (by Shimadzu Service Representative) . . . . . . . . 7-6
7.2 Cleaning the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
7.3 Periodic Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
1
1
Fuse Rating List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
1
Chapter 8 Specifications
8.1 Environmental Conditions of EMC
(Electromagnetic Compatibility) . . . . . . . . . . . . . . . . . . . . . . . . 8-2
1
Classification of EMI in Accordance with EN/IEC60601-1-2: 2007 . . . 8-2
Performance to Be EMC Immunity Tested (Essential Performance) . . 8-2
Cable List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
1
Option List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Guidance and Manufacturer's Declaration
- Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Guidance and Manufacturer's Declaration
1
- Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Guidance and Manufacturer's Declaration
- Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
1
Recommended Separation Distances between Portable and Mobile
RF Communications Equipment and the EQUIPMENT or SYSTEM . . 8-8
8.2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8.2.1 Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8.2.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8.2.3 High-Voltage Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
8.2.4 X-ray Tube Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
8.2.5 Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
8.2.6 X-ray Tube Unit Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
8.2.7 Minimum Inherent Filtration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
8.2.8 DR System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.2.9 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15
8.2.10Custom Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
8.3 Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-17
M503-E028
MobileDaRt Evolution OPERATION MANUAL
Contents-7
1
1
1
1
Contents
8.3.1 General Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-17
8.4 X-Ray Tube Unit Reference Axis and Focal Spot Position . . . 8-18
8.5 Radiography Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
8.5.1 Normal Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
Large Focus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
Small Focus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-20
8.5.2 Low-Battery Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
Large Focus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
Small Focus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-22
8.6 Anatomical Program Number . . . . . . . . . . . . . . . . . . . . . . . . . 8-23
8.6.1 Anatomical Program Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-23
8.7 Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-34
Name Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-34
Warning Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36
Indication Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37
Other Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37
8.7.1 Labeling For X-Ray Tube Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-38
Name Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-38
Warning Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-39
Indication Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-39
Other Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-39
8.8 Statement of Compliance [For Europe] . . . . . . . . . . . . . . . . . . 8-40
8.8.1 Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-40
8.8.2 Company's Quality System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-40
8.8.3 International Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-41
8.9 Statement of Compliance with Standards . . . . . . . . . . . . . . . . 8-42
8.10 Manufacturer Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-43
Appendix
Check List for Trouble . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . App-2
Safety Check Registry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . App-3
Index
Contents-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
1
Chapter
Overview
1
J Contents
1.1
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2
Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.4
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . .1-4
1.5
Classification of Equipment . . . . . . . . . . . . . . . . . . . . . .1-8
1.6
Operator Profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
1.7
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
1 Overview
1.1
Applications
The MobileDaRt Evolution is a general purpose mobile digital X-ray system, which can be moved
freely throughout a hospital to directly obtain X-ray images of various areas of the body.
Do NOT use the system for any other purpose than indicated above.
Do NOT connect the system to any other equipment, either by electrical
or mechanical means, or modify the system.
This system does not support long view radiography as described in the Canon CXDI Wireless User's
Manual.
1.2
Principle
This system uses digital technology to capture, display, and manipulate images. A portable design
and the ability to operate on rechargeable batteries enable a single staffer to transport the system to
any location within a building. The system is generally used for bedside X-ray radiography and
radiography during surgery.
Upon conducting X-ray radiography, X-rays are irradiated from the X-ray tube assembly and
detected by the X-ray flat panel detector after passing through the patient's body. The captured image
is then recorded to the system's internal memory device and a reference image is displayed on the
touch screen monitor. Images can be subject to a variety of image processing.
Images may also be captured without the X-ray flat panel detector by using an X-ray film cassette or
cassette with a photo-stimulable phosphor plate to detect the X-rays that pass through a patient's
body. In addition, captured images can be transferred outside the system via an external network.
1-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
1.3 Features
1.3
1
Features
‹ Smooth and Quiet Movement
1
Power-assisted technology allows moving the system easily and quietly.
‹ Compact Design
Excellent forward visibility allows moving freely through even tight spaces.
Depending on the height of column, there are the tall column type (Type T) and the short column
type (Type S).
‹ Easy Positioning
1
1
The system can be freely and quickly maneuvered into position using the Inch-Mover buttons to
move the main body and to swivel the column to move the X-ray tube unit.
‹ Immediate Response
1
A reference image appears on the display in approximately 3 seconds, allowing on-site image
confirmation.
‹ High Throughput
Improves workflow by eliminating the need to prepare a cassette, develop images, or read CR plates,
and using the hospital internal network to send image data to an imager or PACS.
‹ High Frequency Inverter
1
1
Using a high-frequency inverter which generates high voltage (with a maximum frequency of
60 kHz) provides efficient X-ray generation with low-ripple distortion.
‹ Standard Anatomical Programs
1
Anatomical programs (APR) are installed as standard, to easily allow setting radiography
parameters.
‹ Cordless System
1
The built-in battery enables the obtaining of X-ray images without plugging in the system.
‹ Status Indicator Lamp
The status indicator lamp illuminates or blinks in response to X-ray exposure or system
abnormalities. This allows a visual confirmation of the system status.
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
1-3
1 Overview
1.4
Environmental Conditions
To obtain proper performance, be sure to use the system under the specified environmental
conditions.
Charge the battery under conditions described in "1.4.1 Operation
Environment" P.1-4, not "1.4.2 Transportation and Storage Environment" P.15.
1.4.1
Operation Environment
Use the system under the environmental conditions listed below:
The installation of a dedicated air-conditioner in the examination room is recommended if the
building air-conditioning cannot meet the necessary environmental conditions.
Do NOT use the system in an oxygen-rich environment.
The use of the system in an oxygen-rich environment may cause fatal or
serious injuries or damage to the system due to easy ignition.
Even under the prescribed conditions, avoid rapid changes of
temperature or humidity.
Condensation may occur and cause failure. Also, rust or corrosion may occur
inside the system.
1-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
1.4 Environmental Conditions
Item
1
Specifications
General
The system which is used in hospitals, corridor, patient room etc.
Atmosphere
No explosive or corrosive gases
Ambient temperature
10 to 35 °C
Relative humidity
30 to 75 % (with no condensation)
Atmospheric pressure
800 to 1060 hPa
Environment luminosity
150 to 500 lx
Condition of visibility
Ambient brightness is 100 lx min., 800 lx max.
Frequency of use
15000 exposures / year
1
1
1
While the Canon CXDI Wireless User's Manual states that the access point and IR data
communication unit cannot be used in patient environments, this system (which includes an access
point and IR data communication unit) is medical equipment and can therefore be used in patient
environments.
1.4.2
Transportation and Storage Environment
Item
1
1
Specifications
Ambient Temperature
-10 to 40 °C
Relative Humidity
30 to 60 % (with no condensation)
Atmospheric Pressure
700 to 1060 hPa
1
1
1.4.3
Power Supply
Item
Phase
Single phase AC
Frequency
50/60 Hz
Standard Voltages
100, 110, 120, 200, 220, 230, 240 V
(Set to one of the voltages when installing)
Voltage Fluctuation Range
±10 % of nominal voltage
Supply Capacity
1 kVA
Power Supply Impedance
1
Specifications
1
1
100, 110, 120 V:
Max. 1.0 Ω
200, 220, 230, 240 V:
Max. 4.0 Ω
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
1-5
1 Overview
Be sure to use the power supply specified in the operation manual.
Using a power supply other than the one specified may cause equipment
malfunction or serious accidents such as fire, smoke emission, or explosions.
1.4.4
Grounding
Max. ground resistance: 100 Ω
Ground Terminal:
Additional Ground Terminal: Max. ground resistance: 100 Ω
Be sure to connect the equipment only to a (commercial) power outlet
with a ground terminal.
If the outlet does not have a ground terminal, electric shock may occur.
1.4.5
Wireless Communications
The use of a wireless FPD enables wireless communication between the wireless module built into
the imaging unit and the access point.
Wireless communication is performed based on the IEEE 802.11n (2.4 GHz/5 GHz) LAN standard.
During installation, your Shimadzu service representative will change the frequency band from
2.4 GHz/5 GHz and select a specific frequency (channel) for use.
^ Reference For further details, refer to the Canon CXDI manual.
In addition, the wireless FPD supports a wired connection by using the optional wired connection kit.
^ Reference "5.13 Wired Connection Kit (for the CXDI-70C Wireless/80C Wireless)" P.5-28
CXDI-70C Wireless/80C Wireless
Item
Frequency Range
2412-2472 MHz
IEEE 802.11 n (HT20)
5180-5240 MHz
5745-5825 MHz
1-6
Modulation
OFDM (Orthogonal
Frequency Division
Multiplexing)
MobileDaRt Evolution OPERATION MANUAL
Output Watts
10.3 dBm
16.8 dBm
16.8 dBm
M503-E028
1.4 Environmental Conditions
Item
Frequency Range
2422-2462 MHz
IEEE 802.11 n (HT40)
5190-5230 MHz
5755-5795 MHz
Modulation
OFDM (Orthogonal
Frequency Division
Multiplexing)
1
Output Watts
9.3 dBm
16.8 dBm
1
16.8 dBm
Access Point
Item
Frequency Range
2412-2472 MHz
IEEE 802.11 n (HT20)
5180-5240 MHz
5745-5825 MHz
2422-2462 MHz
IEEE 802.11 n (HT40)
5190-5230 MHz
5755-5795 MHz
Modulation
OFDM (Orthogonal
Frequency Division
Multiplexing)
OFDM (Orthogonal
Frequency Division
Multiplexing)
1
Output Watts
20 dBm
20 dBm
1
20 dBm
20 dBm
1
20 dBm
20 dBm
The maximum power level depends on the channel and the regulation of each region or country.
1
1
1
1
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
1-7
1 Overview
1.5
Classification of Equipment
This system is classified as follows, based on safety standards for electrical medical equipment.
J Protection Method Against Electric Shock
Class I equipment (when connected by the power supply plug)
Internally-powered equipment (when powered by the battery)
J Classification of Applied Parts
Equipment including Type B Applied Parts
J Operation Mode
Continuous operation with intermittent loading
J Degree of Protection Against Liquid Ingress
Ordinary equipment
J For Use in an Oxygen-rich Environment
Do NOT use the system in an oxygen-rich environment.
The use in an oxygen-rich environment may cause fatal or serious injuries or
damage to the system due to easy ignition.
J For Use in Flammable Atmosphere
Do NOT use the system or system in the presence of flammable
anesthetics gas.
It may cause an explosion.
J Classification of Installation Type
Mobile equipment
1-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
1.6 Operator Profile
1.6
1
Operator Profile
Item
1
Details
Age
Age that person can obtain the license of Radiologic Technologist or a
license equal to it.
Sex
No limitation
Nationality
No limitation
1
Radiological Technologist or person who has a license equal to it.
Education
1
The capacity to read and understand the operation manual.
Knowledge
Radiological Technologist or person who has a license equal to it.
Language
Can read and understand English.
1
Necessary.
Experience
Every operator needs to take training for operating the system before
using it.
Permissible impairments
Others
Minimum of 0.7 visibility (corrected vision with lenses acceptable).
No serious physical impairments in general health or extremities.
Minimum height of 140 cm to ensure clear visibility ahead of system
during transport.
1
1
1
1
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
1-9
1 Overview
1.7
Symbols
The symbols used on the system are shown below:
Symbol
Location
Meaning
On key switch
Power ON
On key switch
Power OFF
On name plate
Alternating current
Inside the equipment, where protective
earth conductor in power cord is
connected
Protective earth ground
On name plate
Refer to the operation manual
On warning and caution labels
Observe described items, or refer to the
operation manual
On name plate
Refer to the operation manual
On name plate
Year and month of manufacture
On name plate
Manufacturer
On name plate
Serial number
On access point
RF transmitter
Inch-Mover Buttons
On collimator
1-10
^ Reference
"3.3.2 Travel Mode and Inch-Mover
Buttons" P.3-10
MobileDaRt Evolution OPERATION MANUAL
M503-E028
1.7 Symbols
Symbol
Location
1
Meaning
Arm lock release button
On guide handle
^ Reference
"3.4.1 Locking and Releasing the Arm"
P.3-15
1
Emergency brake release button
On left of grip bar
^ Reference
"6.3 Releasing the Brake in Emergency"
P.6-5
1
1
1
1
1
1
1
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
1-11
1 Overview
This page is intentionally left blank.
1-12
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2
Chapter
Part Names and
Descriptions
2
J Chapter Contents
2.1
External Appearance . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2
Grip Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3
Power Supply Panel . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.4
X-Ray Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.5
Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
2.6
Hand Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
2.7
Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
2.8
Apron Hanger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
2.9
Disposable Wipes Holder . . . . . . . . . . . . . . . . . . . . . .2-15
2.10 Imaging Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
2.11 Storage Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-22
2.12 Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-24
2.13 Main Circuit Breaker and Additional Protective Grounding
Wire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-25
2.14 LAN Connector for External Connection . . . . . . . . . . .2-27
2 Part Names and Descriptions
2.1
External Appearance
X-ray Tube Unit
Column
Arm (P.2-13)
Arm Joint (P.2-13)
Collimator (P.2-10)
Power Supply Panel and Touch Panel Monitor
(P.2-4)
X-ray Control Panel (P.2-6)
Grip Bar (P.2-3)
Hand Switch
(P.2-12)
Main Circuit Breaker
(P.2-25)
Imaging Unit (P.2-16)
Storage Case (P.2-22)
2-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.2 Grip Bar
2.2
2
Grip Bar
This bar is used to move the system. Squeezing the brake release bar releases the brake and allows
pushing or pulling the system in the desired direction.
2
Grip Bar
2
2
2
Brake Release Bar
2
Do NOT touch the brake release bar unless intending to move the
system.
2
Doing so will release the brake, allowing the system to move unexpectedly.
2
J Emergency Brake Release Button
When an accident occurs and the power-assist does not function, the brake can be released, allowing
users to manually move the system.
^ Reference "6.3 Releasing the Brake in Emergency" P.6-5
2
Emergency Brake Release Button
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-3
2 Part Names and Descriptions
2.3
Power Supply Panel
1
2
3
6
7
8
4
5
9
No.
Name
Function
1
Status Indicator Lamp
2
Key Switch
3
Battery Indicator
Normally indicates the amount of energy remaining in the battery, but
also indicates charging status during battery recharging.
4
Power Line Connection
Indicator Lamp
Illuminates when connected to a power outlet.
5
Charging Lamp
6
Emergency Stop Button
7
DR Button
8
USB Connector
Indicates system status by blinking or changing color.
Turning the key switches system power ON
The key can be removed only in the OFF
or OFF
.
position.
Blinks when battery is being charged.
Illuminates when charging is complete.
Used to stop travel movement in an emergency situation.
^ Reference "3.3.4 Emergency Stop Button" P.3-13
Used to start the DR system.
2-4
For DR system operating instructions, refer to the Canon CXDI
manual.
Used to connect external storage media (such as a hard drive or USB
memory unit).
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.3 Power Supply Panel
No.
Name
Function
The system is equipped with a color touch panel. The DR system can
be operated by touching the panel.
For DR system operating instructions, refer to the Canon CXDI
manual.
9
Touch Panel Monitor
2
2
2
• Do NOT place or hang objects on the
touch panel.
Doing so could damage the touch panel.
2
• Do NOT press on the panel with a
sharp object, subject it to strong
shocks, or scratch the panel.
Doing so could damage the touch panel.
2
• Ingesting liquid leaked from the touch
panel can be toxic.
2
J Status Indicator Lamp
The status indicator lamp changes color in the following sequence, according to system status.
2
2
2
If the status indictor lamp is red, the code is displayed on the mAs
display on the X-ray control panel. If the status indicator is red, please
carefully read "6 Troubleshooting" P.6-1 and take the appropriate
measures.
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-5
2 Part Names and Descriptions
2.4
X-Ray Control Panel
1
2
3
4
5
6
16
7
17
8
18
15 14
No.
Description
13
Display/
Button
12
11
10
9
Function
1
Radiography Ready
Indicator Lamp
Illuminates when ready for radiography.
2
X-ray Irradiation
Indicator Lamp
Illuminates whenever X-rays are being emitted.
3
Standby Indicator
Lamp
Illuminates when the system is ready to prepare for X-ray
radiation.
4
Warning Indicator
Lamp
Illuminates when the system does not function correctly.
5
Malfunction Indicator
Lamp
Illuminates if system malfunctions.
6
Travelling Warning
Sound
Enables you to set the volume of the sound emitted during
X-ray exposure or when radiography preparation is finished.
Cassette Mode:
Left LED illuminated
7
FPD/Cassette
Radiography Mode
Button
FPD Mode:
Right LED illuminated or blinked
8
Focus Mode Button
9
Collimator Lamp
Button
Changes focus of the X-ray tube. Toggles between large and
small focuses with each press, where illuminated LED
indicates small focus.
Switches collimator lamp ON or OFF. Also used in
combination with the key switch for obstruction detector
canceling function.
^ Reference "3.3.3 Obstruction Detector" P.3-12
2-6
10
Remote Controller
Button
Switches between infrared remote controller enabled (LED
ON) and disabled (LED OFF) modes.
11
DR Long-Term
Exposure Button
LED illuminates when DR long-term exposure mode is
selected.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.4 X-Ray Control Panel
Display/
Button
2
No.
Description
Function
12
mAs Display /
Current-Time Product
Buttons
Displays mAs setting. Pressing buttons sets the current-time
product.
13
kV Display / Tube
Voltage Buttons
Displays kV setting. Pressing buttons sets the tube voltage.
14
User Settings Button
^ Reference "4.1 User Settings" P.4-2
15
Registration Button
Registers anatomical programs and user settings
2
2
2
Selects imaging of the head region (head, face, nose, ear,
neck, etc.).
2
Selects imaging of the chest region (collarbone, ribs,
shoulder blade, thoracic spine, etc.).
Selects imaging of the abdomen region (abdomen, lumbar,
etc.).
16
Imaging Region
Selection Buttons
Selects imaging of the waist region (pelvis, coccyx, fetus,
femur, etc.).
Selects imaging of the leg region (knee, shin, etc.).
2
2
Selects imaging of the foot region (heel, foot, toes, etc.).
Selects imaging of the arm region (upper arm, elbow, wrist,
etc.).
Selects imaging of the hand region (hand, fingers, etc.).
2
Selects frontal projection.
17
Projection Direction
Selection Buttons
Selects lateral projection.
2
Selects oblique projection.
Selects Thin.
18
Body Thickness
Selection Buttons
2
2
Selects Normal.
Selects Thick.
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-7
2 Part Names and Descriptions
1
2
3
4
5
6
13
14
15
12 11
No.
Description
Display/
Button
10
9
8
7
Function
1
Radiography Ready
Indicator Lamp
Illuminates when ready for radiography.
2
X-ray Irradiation
Indicator Lamp
Illuminates whenever X-rays are being emitted.
3
Standby Indicator
Lamp
Illuminates when the system is ready to prepare for X-ray
radiation.
4
Warning Indicator
Lamp
Illuminates when the system does not function correctly.
5
Malfunction Indicator
Lamp
Illuminates if system malfunctions.
6
Travelling Warning
Sound
Enables you to set the volume of the sound emitted during
X-ray exposure or when radiography preparation is finished.
Collimator Lamp
Button
Switches collimator lamp ON or OFF. Also used in
combination with the key switch for obstruction detector
canceling function.
7
^ Reference "3.3.3 Obstruction Detector" P.3-12
2-8
8
Remote Controller
Button
Switches between infrared remote controller enabled (LED
ON) and disabled (LED OFF) modes.
9
mAs Display /
Current-Time Product
Buttons
Displays mAs setting. Pressing buttons sets the current-time
product.
10
kV Display / Tube
Voltage Buttons
Displays kV setting. Pressing buttons sets the tube voltage.
11
User Settings Button
^ Reference "4.1 User Settings" P.4-2
12
Registration Button
Registers anatomical programs and user settings
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.4 X-Ray Control Panel
No.
Description
Display/
Button
2
Function
Selects imaging of the head region (head, face, nose, ear,
neck, etc.).
2
Selects imaging of the chest region (collarbone, ribs,
shoulder blade, thoracic spine, etc.).
Selects imaging of the abdomen region (abdomen, lumbar,
etc.).
13
Imaging Region
Selection Buttons
Selects imaging of the waist region (pelvis, coccyx, fetus,
femur, etc.).
2
2
Selects imaging of the leg region (knee, shin, etc.).
Selects imaging of the foot region (heel, foot, toes, etc.).
Selects imaging of the arm region (upper arm, elbow, wrist,
etc.).
2
Selects imaging of the hand region (hand, fingers, etc.).
2
Selects frontal projection.
14
Projection Direction
Selection Buttons
Selects lateral projection.
2
Selects oblique projection.
Selects Thin.
15
Body Thickness
Selection Buttons
2
Selects Normal.
Selects Thick.
2
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-9
2 Part Names and Descriptions
2.5
Collimator
1
2
3
4
5
6
7
No.
1
2
Description
Function
Status Indicator Lamp
Indicates system status by blinking or changing color.
Arm Lock Release Buttons
Pressing the arm lock release button unlocks the column rotation,
vertical X-ray tube unit movement, and arm extension/contraction.
(All Free button)
3
Inch-Mover Buttons
When the arm is unlocked, pressing the inch-mover buttons will move
the system forward or backward in small increments (about 5 cm/s).
When the arm is locked, the inch-mover buttons are non-functional.
4
Irradiation Field Adjustment
Knobs
Adjusts the size of the irradiation field.
5
Tape Measure
Used to measure SID (distance between focal spot and cassette or
imaging unit).
6
Collimator Lamp Button
Used to switch the collimator lamp ON or OFF. It has the same
function as the
(collimator lamp button) on the X-ray control
panel.
When the arm is locked, switches the system travel alert sound ON or
OFF.
7
2-10
Added Filters Mount
Mounts an added filter.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.5 Collimator
J Status Indicator Lamp
2
The status indicator lamp changes color in the following sequence, according to system status.
2
2
If the status indictor lamp is red, the code is displayed on the mAs
display on the X-ray control panel. If the status indicator is red, please
carefully read "6 Troubleshooting" P.6-1 and take the appropriate
measures.
2
2
2
2
2
2
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-11
2 Part Names and Descriptions
2.6
Hand Switch
1
No.
1
Name
Radiography Preparation
Button
(pushed to first level)
2
Function
When held down, the radiography ready indicator lamp illuminates
green to indicate that the system has finished preparation for
radiography. Pressing and immediately releasing the button allows
switching the collimator lamp ON and OFF.
^ Reference "3.4.3 Setting the Irradiation Field" P.3-19
2
2-12
X-Ray Exposure Button
(pushed to second level)
When this button is pressed while the radiography ready indicator is lit
green, the X-ray irradiation indicator lamp illuminates yellow and
radiography is performed.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.7 Arm
2.7
2
Arm
2
2
1
2
2
2
3
2
No.
Name
Arm Lock Release Button
Function
(All Free button)
Pressing the arm lock release button unlocks the column rotation,
vertical X-ray tube unit movement, and arm extension/contraction.
2
Arm Joint
Mechanism that holds the X-ray tube unit.
3
Arm Lock Release Lever
This lever has the same function as the arm lock release button. It
allows unlocking the arm by lifting up the X-ray tube unit while
holding the arm lock release lever down.
1
2
2
2
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-13
2 Part Names and Descriptions
2.8
Apron Hanger
This hanger is used for hanging X-ray protective aprons.
The apron hangers are attached in two locations
Apron Hanger
Apron Hanger
Installation
Locations
• Use the apron hanger only for hanging aprons.
• Do NOT hang objects over 7 kg on each apron hanger.
• Do NOT hang an apron on the apron hanger when the arm is not
locked.
• Move the system carefully if an apron is on the apron hanger,
especially when moving the system over any changes in floor level
or on slopes.
2-14
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.9 Disposable Wipes Holder
2.9
2
Disposable Wipes Holder
Cylindrical wet tissue containers can be held, with a radius between 120 mm to 150 mm and a height
between 190 mm to 230 mm.
2
2
2
Disposable Wipes
Holder
2
2
The maximum weight capacity is 5 kg.
2
2
2
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-15
2 Part Names and Descriptions
2.10 Imaging Unit
The imaging unit is an applied part that contacts the patient's body.
2.10.1 CXDI-50G/50C/55G/55C/60G/60C
^ Reference
For further details, refer to the Canon CXDI manual.
CXDI-50G/50C
CXDI-55G/55C
CXDI-60G/60C
2-16
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.10 Imaging Unit
2
2
Treat this precision imaging unit carefully.
Do NOT drop or hit the imaging unit or expose it to large shocks, all of which
could damage the unit or cause it to malfunction.
The imaging unit is equipped with four shock detectors. These detectors are
normally white, but change color if exposed to excessive shock. If even one
shock detector changes color, there is a chance the imaging unit is damaged
or broken. If a shock detector changes color, any damage incurred is not
covered by warranty.
2
2
Shock Detectors
CXDI-50G/50C
2
2
2
Shock Detectors
CXDI-55G/55C
2
2
Shock Detectors
CXDI-60G/60C
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-17
2 Part Names and Descriptions
• Do NOT pull the cable of the imaging unit.
If the cable catches on something, do not pull on the imaging unit. Doing
so can result in cable damage, fire, electric shock, unit damage, or
malfunction.
• Do NOT apply any load to the imaging unit that exceed its weight
limit.
Doing so can damage or break the internal sensor of the imaging unit.
<Load limits>
Uniform loads: 150 kg over the entire imaging surface
Point loads: 100 kg / 40 mm dia. area
• Be sure to use the imaging unit on a level floor or stand.
If force is applied to the imaging unit at an angle, the internal sensor of the
imaging unit can be damaged or broken.
• When handling the imaging unit, be careful not to run over the cable
of the imaging unit with the wheels.
Doing so can damage the cable, cause a fire or electric shock, or damage
the unit.
J Usage of CXDI-55G/55C/60G/60C
•
With CXDI-55G/55C/60G/60C, it is possible to turn the imaging unit ON/OFF using the
remote switch.
When performing radiography, press the remote switch to turn it ON.
Remote Switch
2-18
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.10 Imaging Unit
•
Remove the sensor cable relay connector when straightening the cable to eliminate the
torsion, or when the positioning of the sensor unit is not easy with the sensor cable relay
connector attached.
When moving the system, store the sensor cable relay connector in the connector holder.
Connector Holder
Sensor Cable
Relay Connector
2
2
2
2
2
2
• When transporting the CXDI-55G/55C/60G/60C, do NOT allow the
sensor cable relay connector to drag on the floor or ground or bump
against anybody or any outside objects or personal property. It is
recommended that the user hold the sensor cable relay connector
while transporting the CXDI-55G/55C/60G/60C.
Failure to do so can cause the sensor cable relay connector to get caught
by, or bump against somebody which may result in injury, or cause
damage to personal property, outside objects and/or the sensor cable
relay connector itself.
• When connecting or disconnecting the sensor cable relay connector,
set the power switch of the remote switch to OFF and confirm that
the POWER lamp on both the sensor unit and the remote switch are
off.
The ERROR lamp will light if the sensor cable relay connector is
connected or disconnected while the POWER lamp is lit.
For details about connecting and disconnecting the sensor cable relay
connector, refer to the CXDI Series Operation Manual.
• Be sure to hold the sensor cable relay connector by its grip in order
to the sensor cable relay connect or disconnect it.
Holding the cable to connect or disconnect the sensor cable relay
connector may break the wires inside the cable.
• Do NOT drop the sensor cable relay connector when connecting or
disconnecting it.
You may injure yourself or cause damage to outside objects, personal
property, or the sensor cable relay connector.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2
2
2
2
2
2
2
2-19
2 Part Names and Descriptions
2.10.2 CXDI-70C Wireless/80C Wireless
^ Reference
For further details, refer to the Canon CXDI manual.
CXDI-80C Wireless
CXDI-70C Wireless
• Treat this precision imaging unit carefully.
Do NOT drop or hit the imaging unit or expose it to large shocks, all of
which could damage the unit or cause it to malfunction.
• Do NOT apply any load to the imaging unit that exceed its weight
limit.
Doing so can damage or break the internal sensor of the imaging unit.
<Load limits>
Uniform loads: 150 kg over the entire imaging surface
Point loads: 100 kg / 40 mm dia. area
• Be sure to use the imaging unit on a level floor or stand.
If force is applied to the imaging unit at an angle, the internal sensor of the
imaging unit can be damaged or broken.
• Do NOT pull on the cable when using a wired connection (option).
Do NOT pull the imaging unit when the cable is caught on something. This
may damage the cable and lead to fire or electric shock and system
damage or failure.
• Be careful not to run over the cable to the image unit when using a
wired connection (option).
This may damage the cable and lead to fire or electric shock and system
damage or failure.
^ Reference
2-20
"5.13 Wired Connection Kit (for the CXDI-70C Wireless/80C
Wireless)" P.5-28
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.10 Imaging Unit
J Usage of CXDI-70C Wireless/80C Wireless
A wireless connection is established between the wireless module built into the imaging unit and the
access point built into the system when using the CXDI-70C Wireless/80C Wireless.
The CXDI-70C Wireless/80C Wireless are used to link (recognize/connect) the imaging unit to the
system by performing communication between the light receiver of the IR data communication unit
and the imaging unit using infrared rays.
2
2
2
2
Light Receiver
of the IR Data
Communication
Unit
Access Point
Access Point LED
Display
2
2
The access point LED display indicates the following system statuses.
System Status
2
Green
The system is operating normally (wireless clients are not available).
2
Blue
The system is operating normally (wireless clients are available).
LED Display
OFF
The power supply is OFF.
Blinking (Green)
The system is starting up.
ON
2
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-21
2 Part Names and Descriptions
2.11Storage Case
2.11.1 CXDI-50G/50C/55G/55C/60G/60C
J One Imaging Unit
There are front and rear slots in the storage case.
Rear Slot
Front Slot
CXDI-50G/50C/55G/55C
CXDI-60G/60C
•
The front slot is used to store the imaging unit. The imaging unit can be stored even with the
grid unit attached.
•
The rear slot is used to store the grid unit and cassette. If only cassettes are used, up to three
can be inserted.
In the rear slot, the stored grid unit can be attached to and detached from the imaging unit.
Imaging Unit
The grid unit is
attached to
the imaging unit.
Grid Unit
Rear Slot
2-22
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.11 Storage Case
2
Generating X-rays with a cassette left in the storage case may expose the film in the
cassette.
2
2
J Two Imaging Unit
^ Reference "5.12 Second FPD Kit" P.5-24
2
2.11.2 CXDI-70C Wireless/80C Wireless
Front and rear slots and a battery storage case are provided for the CXDI-70C Wireless
/80C Wireless.
2
Storage Case For Extra
Battery Pack
Rear Slot
Front Slot
2
2
2
•
The front slot is used to store the imaging unit. The imaging unit can be stored with the handle
unit attached.
•
The rear slot is used to store the handle unit, grid frame, and cassettes. When only storing
cassettes, this slot can hold up to three.
2
2
Generating X-rays with a cassette left in the storage case may expose the film in the
cassette.
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-23
2 Part Names and Descriptions
2.12Power Cable
Use when the battery charging.
Move the system to a nearby outlet, pull out the power plug from the front of the system and insert it
into the power outlet.
When pulling out the power cable, the cable is held within the range to where click sound from the
cable reel can be heard. The cable is rewound in the range where no sound from the cable reel can be
heard.
Power Cable
• When pulling out the power cable, hold the plug portion and do NOT
apply excessive force to the cable.
Excessive force may damage the cable or internal cable reel.
• When rewinding the cable, hold the plug until it is completely
retracted.
2-24
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.13 Main Circuit Breaker and Additional Protective Grounding Wire
2.13Main Circuit Breaker and Additional Protective
2
Grounding Wire
2
2
Additional Protective
Grounding Wire
Main Circuit Breaker
ON
2
OFF
2
2.13.1 Main Circuit Breaker
The main circuit breaker opens or closes the connection between the battery and internal circuits.
2
If an overcurrent flows from the battery, the main circuit breaker detects the overcurrent and
automatically switches OFF.
2
If the main circuit breaker switches OFF frequently, stop using the
system and contact your Shimadzu service representative.
2
2
2
•
Keep the main circuit breaker switched OFF when not using the system for any
long period of time.
•
The system can not be used when the main circuit breaker is turned OFF.
•
If the main circuit breaker is switched OFF while the key switch is turned ON, the
connection between the battery and the internal circuits is closed, and the system
may remain ON for some time due to the residual charge within the internal
circuits.
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-25
2 Part Names and Descriptions
2.13.2 Additional Protective Grounding Wire
While charging, Shimadzu recommends connecting the additional protective
grounding wire to a grounding terminal with a grounding resistance of 100 Ω or less.
2-26
MobileDaRt Evolution OPERATION MANUAL
M503-E028
2.14 LAN Connector for External Connection
2.14 LAN Connector for External Connection
2
The external connection LAN connector is used for sending obtained image data to external
computers or printers.
Insert the cable into the LAN connector, connect the cable to the external device and transfer data in
DICOM format.
2
2
2
2
LAN Connector
2
2
2
Do NOT connect the LAN connector to external peripherals (such as
printers) not specified by Shimadzu or any devices that do NOT meet
safety standards.
Doing so could cause unit failure, malfunction, smoke, overheating, electrical
shock, and loss or damage of data.
2
2
2
2
2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
2-27
2 Part Names and Descriptions
This page is intentionally left blank.
2-28
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3
Chapter
Basic Operation
3
J Chapter Contents
3.1
Starting and Shutting Down the System . . . . . . . . . . . .3-2
3.2
Warm-up the X-Ray Tube Unit . . . . . . . . . . . . . . . . . . .3-6
3.3
Moving the System . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.4
Arm Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
3.5
Setting the Radiography Parameters . . . . . . . . . . . . .3-21
3.6
X-Ray Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . .3-24
3.7
Charging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . .3-35
3.8
Imaging Unit Storage (CXDI-70C Wireless/80C Wireless)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-38
3 Basic Operation
3.1
3.1.1
Starting and Shutting Down the System
Starting the System
To start the system, turn the key switch to the ON
position.
The system confirms its functions before the remaining battery charge is indicated on the battery
indicator. After the battery charge status has been indicated, the system can be moved, the arm lock
released, X-ray radiography operations performed, or the DR system started. When the
(standby
indicator lamp) is illuminated, radiography can be performed. The following limits are placed on Xray radiography depending on the remaining battery charge display.
No.
Battery Indicator
1
~
2
3
X-ray Radiography Availability
Radiography is possible.
Low Battery - only low-battery radiography is possible.
Blinking
^ Reference "3.6.6 Low-Battery Radiography" P.3-33
Radiography is not possible. Immediately charge the battery.
^ Reference "3.7 Charging the Battery" P.3-35
Indicator segment is illuminated if shown in black J and off if shown in white F above.
3-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.1 Starting and Shutting Down the System
3
•
The
(standby indicator lamp) will illuminate almost immediately if restarting
the system directly after shutting it down; It may take as long as one minute to
illuminate if restarting after the system has not been performed over a long period
of time.
•
If the key switch is turned ON
with buttons or levers left ON, code d4 to d11
appears on the mAs display. Release the switch or lever.
•
If the key switch is turned ON
with the hand switch still depressed, code F64
or F65 appears on the mAs display. Release the hand switch.
•
Please note that the battery voltage rises temporarily immediately after the key
switch is turned ON
. As a result, the battery indicator will indicate more than
the actual remaining charge. After the system is used for a while, the battery
display will return to its actual remaining charge.
•
3
3
3
3
If the key switch is turned ON
with the emergency stop button still depressed,
code F6d appears on the mAs display.
If the emergency stop button is reset after the key switch is turned ON
with
the emergency stop button in the pressed state, code d13 will appear and the
system will lock tight. In this case, restart the system while the emergency stop
button is in the reset state.
* It is possible to change the APR indicated when starting the system.
3
3
3
See "4.1 User Settings" P.4-2.
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-3
3 Basic Operation
3.1.2
Shutting Down the System
To shut down the system, turn the key switch to the OFF
3-4
position.
•
When leaving the system unattended, be sure to turn the key switch to OFF
remove the key, and keep it in an appropriate place.
•
Even with key switch in the OFF
position, the DR system will not be shut
down. Be sure to shut down the DR system when leaving the system unattended.
•
When not using the system for a long period of time, turn the main circuit breaker
to OFF.
MobileDaRt Evolution OPERATION MANUAL
,
M503-E028
3.1 Starting and Shutting Down the System
J Auto-Power-OFF Function
When the key switch is ON
, the auto-power-off function shuts the system off if not used for a
long time. The auto-power-off function default setting is 15 minutes. When the auto-power-off
function starts up, [at OFF] is displayed on the X-ray control panel until the system is shut off. When
the power plug is connected, [at OFF] will not disappear even if the system is shut off.
To reset from the auto-power-off condition, turn the key switch OFF
turn it back ON
.
, wait three seconds, then
3
3
3
* The time setting can be changed if necessary.
See "4.1 User Settings" P.4-2.
3
•
When the auto-power-off function is started with the connection of the power plug,
disconnect the power plug to restart the system.
•
When the system is used by connecting the power plug, deactivate the autopower-off function or set the auto-power-off time to 30 minutes.
•
Even if the auto-power-off function shuts the system off, the battery will continue
to consume a small amount of power. When the system is not in use, turn the key
switch to OFF
.
•
If the battery indicator displays no illuminated portion, the auto-power-off function
will automatically start in 5 minutes to prevent wearing out the battery.
•
The auto-power-off function does not turn the DR system OFF.
•
If the system is left with the DR system still running, the battery will continue to
consume a small amount of power. Therefore, always turn the DR system OFF
when the system is not in use.
3
3
3
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-5
3 Basic Operation
3.2
Warm-up the X-Ray Tube Unit
X-ray tube units must be warmed up before they can be used. Always age the X-ray tube unit prior to
using the system.
Always age the X-ray tube unit before starting radiography. In
particular, if X-ray tube voltage is set to over 100 kV, warm-up should be
performed on a daily basis.
Performing radiography without warm-up the X-ray tube unit may cause an
electrical discharge.
3.2.1
Warm-up Procedure
Usage Condition
Warm-up Description
Warm-up 1
When using the X-ray tube unit for the
first time that day
When the X-ray tube voltage to be used
is 100 kV or higher
When the X-ray tube voltage to be used
is 100 kV or higher, after normally
using a voltage of 80 kV or less
3-6
Generate X-rays, starting with the parameters indicated for Step 1
below, then successively generate X-rays indicated for each step,
until reaching the X-ray tube voltage to be actually used. If anything
abnormal occurs, go back two steps and repeat the process.
Warm-up 2
Start generating X-rays at Step 3 and successively increase steps
until reaching the X-ray tube voltage to be actually used. If anything
abnormal occurs, go back two steps and repeat the process.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.2 Warm-up the X-Ray Tube Unit
Warm-up Procedure
3
Step
Tube Voltage (kV)
Current-Time Product
(mAs)
Number of Times
Pause (seconds)
1
60
10
2
40
2
70
10
2
40
3
80
10
2
40
4
90
10
2
40
5
100
10
2
40
6
110
10
2
40
7
115
10
2
40
8
120
10
2
40
9
125
10
2
40
10
130
10
2
40
11
133
10
2
40
3
3
3
3
3
3
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-7
3 Basic Operation
3.3
Moving the System
• To move the system, hold the grip bar securely.
Do NOT operate the grip bar using only one hand, or underhanded.
Doing so can result in not only operation errors and system damage, but
may also cause serious accident leading to injury or death.
• Do NOT lay your hand on the bar retaining cover when holding the
grip bar.
If you hold the grip bar while laying your hand on the cover, a force may
apply toward the rear, causing the system to suddenly accelerate and
move backward.
Bar retaining cover
3-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.3 Moving the System
3
3
• When driving the system where people are present, near equipment,
or close to a wall, drive slowly to be able to stop immediately and
maintain safety in the surroundings.
3
• Do NOT touch the brake release bar unless intending to move the
system.
Doing so will release the brake, allowing the system to move
unexpectedly.
3
^ Reference "2.2 Grip Bar" P.2-3
• If the battery indicator is not illuminated, immediately charge the
battery.
3
• If the system moves abnormally, immediately stop using the system
and contact your Shimadzu service representative.
3
3
• The operator may experience static electric shock when moving the
system if their body has become statically charged, depending on
conditions such as humidity and type of clothing.
• When the [PAS] or [Acc] value for power-assist settings has been
changed, not only forward but also backward movement will be influenced.
To maintain the same backward movement, adjust the [bAc] value
accordingly.
3.3.1
Imaging Unit Storage
3
3
3
Stow the imaging unit securely in the storage case before moving the system.
^ Reference "2.11 Storage Case" P.2-22
3
Lock the CXDI-70C Wireless/80C Wireless detector into the storage case.
^ Reference "3.8 Imaging Unit Storage (CXDI-70C Wireless/80C Wireless)" P.3-38
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-9
3 Basic Operation
3.3.2
Travel Mode and Inch-Mover Buttons
• When moving the system, make sure that there are no obstacles
ahead.
The obstacle detector may not be able to detect light weight obstacles in
front of it, such as IV stands or cables from other equipment. Hitting an IV
stand or running over cables from other equipment could be very
dangerous to patients.
• Do NOT move the system over any steps or bumps greater than 2 cm,
or slopes steeper than 7 degrees. Always stop the system on a level
surface and avoid sloped surfaces.
Failure to do so may cause the system to roll down the slope or tip over.
• Do NOT operate this system outdoors.
This system is designed for indoor use only. Using this system outdoors
may lead to system failure.
The system can be moved using the normal travel mode, low-speed travel mode, or inch-mover
buttons.
By squeezing the brake release bar on the lower surface of the grip bar and pushing on the bar, the
system moves in the direction it is pushed. In addition, the system's power-assist technology allows
changing the travel speed according to how much force is applied.
J Normal Travel Mode
The system is in the normal travel mode when the arm is locked.
^ Reference "3.4.1 Locking and Releasing the Arm" P.3-15
Maximum speed is about 5 km/h.
* Travel characteristics can be changed.
See "4.1 User Settings" P.4-2.
J Low-Speed Travel Mode
If the arm is not locked, the system will automatically select the low-speed travel mode.
^ Reference "3.4.1 Locking and Releasing the Arm" P.3-15
3-10
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.3 Moving the System
The traveling speed of the system will be limited to about half the normal speed and the degree of
power-assist is limited to about half of its normal level.
3
3
• Do Not move the system over bumps or other floor-level changes in
low-speed travel mode.
The system may tip over if the arm is not locked.
3
• Do Not move the system over slopes steeper than 5 degrees.
3
J Inch-Mover Buttons
Holding down the inch-mover buttons with the arm unlocked allows the movement of the system
forward or backward more slowly than in the low-speed travel mode. Maximum speed is about 5 cm/
s.
The inch-mover buttons can be operated continuously for about 10 seconds.
3
To continue using the buttons, press them again.
Motor power is lower than in the low-speed travel mode. If the system cannot be moved due to thickpiled carpet, or other resistance, use the grip bar.
3
The system does not move if both inch-mover buttons are pressed simultaneously, or pressing one of
the buttons while the other is already pressed.
3
Inch-Mover Button
(moves forward)
Inch-Mover Button
(moves backward)
3
3
3
If the system does not stop due to inch-mover button failure, stop the
system by pressing the emergency stop button on the power supply
panel.
3
^ Reference "3.3.4 Emergency Stop Button" P.3-13
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-11
3 Basic Operation
3.3.3
Obstruction Detector
The obstruction detector is located at the lower front part of the system. If the obstruction detector
hits an obstacle, the system will stop and the unit can only be moved backwards.
* The obstruction detector can be set to emit a sound if activated.
See "4.1 User Settings" P.4-2.
Obstruction Detector
J Obstruction Detector Cancel Function
If the obstruction detector is damaged, the obstruction detection function may remain ON (activated).
In this case, the obstruction detector cancel function can be used to temporarily cancel the
obstruction detector function, allowing the system to be moved.
• The obstruction detector cancel function is only a temporary
measure. If the obstruction detector does not function normally,
contact your Shimadzu service representative.
• While the obstruction detector cancel function is enabled, the
detector will not emit any sound, even if activated. Because there is
no auditory warning, be sure to pay careful attention to the area in
the periphery of the system while moving.
• After moving to a safe location, disable the obstruction detector
cancel function.
1
2
Turn the key switch to OFF
.
Turn the key switch ON
while pressing the
button) on the X-ray control panel.
(collimator lamp
* The obstruction detector cancel function is disabled when the key switch is turned to the OFF
position. To enable this function again, repeat Step 2 above.
3-12
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.3 Moving the System
3.3.4
Emergency Stop Button
3
To stop system movement in an emergency, press the emergency stop button on the power supply
panel.
3
Radiography is not possible when the emergency stop button is depressed.
3
Emergency Stop
Button
3
3
3
• If you press the emergency stop button due to abnormal behavior of the
system, stop all operations and contact your Shimadzu service
representative.
3
• When the emergency stop button is pressed, turn the key switch to OFF
, and move the system to a safe location using the emergency brake
release function.
3
^ Reference "6.3 Releasing the Brake in Emergency" P.6-5
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-13
3 Basic Operation
J Releasing the Emergency Stop
To release the emergency stop button, if you have pressed it accidentally or pressed it for testing, turn
the key switch to OFF
and turn the red portion of the emergency stop button in the direction of
the arrow.
Emergency Stop
Button
Do NOT turn the sliver-colored portion of the emergency stop button.
3-14
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.4 Arm Operation
3.4
3.4.1
Arm Operation
3
Locking and Releasing the Arm
3
The arm is in the locked position when the arm and the main unit are coupled at the arm joint section.
During radiography, the arm lock is released to allow the movement of the X-ray tube unit to the
imaging position.
3
3
Do NOT release the arm lock if the unit is on a slope greater than 5
degrees.
The system may tip over if the arm is rotated, extended/contracted, or moved
up or down on a slope greater than 5 degrees.
3
3
J Releasing the Arm Lock
Releasing the arm lock enables rotation, extension/contraction, and up/down movements of the arm.
In addition, radiography parameters are displayed on the X-ray control panel and radiography is
enabled. In this state, the system will be in the low-speed travel mode, where maximum travel speed
is limited to about half of the normal travel speed.
3
The lock can be released in one of the following three ways:
No.
Location
3
Meaning
1
Arm Lock Release
Lever
Lift up the X-ray tube unit while holding down the arm lock release lever
located at the arm joint section.
2
Arm Lock Release
Button
With pressing the arm lock release button at the guide handle grip, force the
arm slightly downward, and then lift up the X-ray tube unit.
3
Arm Lock Release
Button
With pressing the arm lock release button at the collimator control handle,
force the arm slightly downward, and then lift up the X-ray tube unit.
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-15
3 Basic Operation
A
1
Arm Lock
Release
Lever
View From A
3
Arm Lock
Release
Button
2Arm Lock Release Button
J Setting the Arm Lock
Once the arm is locked, the system enters the normal travel mode and maximum speed is about
5 km/h.
Arm Lock
Release Button
Locking Pin
Locking Pin
Hole
1
Push the X-ray tube unit as far as possible toward the column until it
makes contact with the stopper, while pressing the arm lock release
button.
* The unit may feel heavy just before reaching the stopper. Be sure to push it all the way in.
3-16
2
3
Return the tilt of the X-ray tube unit and collimator to the center position.
4
Insert the locking pin, located on the lower surface of the arm, into the
locking pin hole in the arm joint section.
Slowly push the X-ray tube unit down while pressing the arm lock release
button.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.4 Arm Operation
5
After locking the arm, a sound is emitted and the status indicator lamp
switches OFF.
3
3
• When lowering the X-ray tube unit, ensure that the front of the
collimator does not hit the control panel buttons or the panel itself.
3
• Do NOT place your hand on the column cover.
By doing so, a part will protrude from a hole in the cover when the arm
moves up and down.
3.4.2
3
3
Positioning the X-Ray Tube Unit
3
3
Do NOT rotate the X-ray tube unit more than ±180 degrees around the
arm axis.
Doing so may damage the cables.
The rotated and tilted position of the X-ray tube unit is held by frictional force (no button operations
are needed to change the tilt of the X-ray tube unit).
When positioning the X-ray tube unit, use the angle indicator label as a reference for rotating the unit
around the arm axis.
Angle Markings
Pointer
3
3
3
X-ray Tube Unit
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-17
3 Basic Operation
•
Use the angle indicator label as a reference for rotating the X-ray tube unit around the tube
axis. Turn the knob to secure it in place or release it.
X-ray Tube Unit
Angle Markings
Pointer
Knob
The knob has a stopper mechanism. DO NOT loosen the knob farther than the
stopper. The knob will fall off.
•
3-18
Click stops are provided every 90 degrees as a tactile guide for positioning the X-ray tube unit
exactly both vertically and horizontally.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.4 Arm Operation
3.4.3
Setting the Irradiation Field
3
J Adjusting the Irradiation Field
When setting the X-ray irradiation field, operate the irradiation field adjusting knobs on the front of
the collimator. Adjust the size of the irradiation field by referring to the following two guides:
•
The tape measure attached to the side of the collimator
•
The angle markings printed on the front of the collimator
3
3
3
Angle Markings
Tape Measure
3
3
Irradiation Field Adjusting Knob
In addition, loosening the knob on the top of the collimator allows the collimator to be rotated around
the X-ray axis.
3
3
Knob
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-19
3 Basic Operation
J Irradiation Field Reference Light
Illuminating this light allows confirming the X-ray irradiation field.
The cross-hair intersection in the light field indicates the center of the X-ray irradiation field.
The collimator lamp buttons are located on the front of the collimator, the X-ray control panel, and
remote control (optional). Pressing any of these buttons switches on the collimator lamp. The
collimator lamp is automatically turned off after about 30 seconds. It is also turned off by pressing
the button once more.
* The collimator lamp can also be turned ON or OFF by pressing the radiography hand switch to the first
level and then immediately releasing it.
See "4.1 User Settings" P.4-2.
If radiography preparation is performed while the collimator lamp is illuminated, the
collimator lamp turns OFF. When radiography preparation is finished, the collimator
lamp turns ON again.
^ Reference If the collimator lamp does not turn ON again even after radiography preparation is
finished, see "4.1 User Settings" P.4-2.
3-20
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.5 Setting the Radiography Parameters
3.5
3.5.1
Setting the Radiography Parameters
3
Setting the Radiography Parameters
3
Radiography parameters on this system are set using two controls ⎯ X-ray tube voltage and currenttime product. Based on the radiography tube current-time product (mAs) setting, the system
automatically determines the maximum X-ray tube current (mA) and the shortest exposure time(s)
available within the rated limits of the system and X-ray tube unit. Radiography parameters can be
changed using the methods described below.
The radiography parameter settings permitted on the MobileDaRt Evolution are constrained by Xray tube voltage (kV), current-time product, and radiography mode.
^ Reference For more details regarding radiography parameter settings,
3
3
See "8.2 Specifications" P.8-9.
3.5.2
Setting the Radiography Parameters Using the Anatomical
3
Programs (APR)
3
J Selecting an Anatomical Program for DR Radiography Mode
Selecting radiography mode in the DR system selects the corresponding anatomical program and
changes the radiography settings to those registered for the program. The corresponding LEDs are
illuminated.
3
^ Reference For further details on operating the DR system, refer to the Canon CXDI manual.
3
•
Anatomical programs cannot be selected from the X-ray control panel during DR
radiography mode.
•
Pressing the
(FPD/cassette radiography mode button) will not switch the
mode from cassette radiography to DR radiography.
Select a radiography mode from the DR system to switch the mode from the
cassette radiography to the DR radiography. Once done, the corresponding
anatomical program is selected and the radiography parameters are changed to
those registered to that program.
•
When the
(user settings button) is pressed to switch to the change settings
mode, and then the
(user settings button) is pressed again to return to the
normal mode, the anatomical program changes. From that point, select the
radiography mode using the DR system once again.
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-21
3 Basic Operation
J Selecting an Anatomical Program for Cassette Radiography Mode
Use the X-ray control panel buttons to select the registered radiography parameters. The LED light
for selected buttons will be illuminated.
1
Select the imaging region.
Button
2
Imaging
Region
Button
Imaging
Region
Head
Waist
Arm
Chest
Leg
Hand
Abdomen
Foot
Select the projection direction.
Button
Projection
Direction
Button
Front
3
Imaging
Region
Button
Projection
Direction
Button
Lateral
Projection
Direction
Oblique
Select the body thickness correction setting.
Button
Body
Thickness
Thin
Button
Body
Thickness
Button
Normal
Body
Thickness
Thick
Examples of selecting anatomical programs
Example of frontal projection of the abdomen of a patient with a standard body type.
3-22
1
Select the imaging region: Press the
2
Select the projection direction: Press the
3
Select body thickness correction: Press the
(abdomen).
MobileDaRt Evolution OPERATION MANUAL
(frontal projection).
(normal).
M503-E028
3.5 Setting the Radiography Parameters
3.5.3
Manually Setting the Radiography Parameters
3
3
Settings can be changed manually only using the X-ray control panel.
3
Radiography parameters for the DR system cannot be changed manually.
When manually setting the radiography parameters, refer to the methods described
below.
If settings registered in the anatomical program are changed, the imaging region,
projection direction, and body thickness correction selection buttons will blink.
3
3
J Setting the X-ray Tube Voltage (kV)
The setting value is shown on the kV display.
•
Press the [+] button to increase X-ray tube voltage setting in 1 kV increments.
•
Press the [-] button to decrease X-ray tube voltage setting in 1 kV decrements.
•
Hold down the [+] or [-] button to change the setting more quickly.
3
3
J Setting the Current-Time Product (mAs)
The set value is shown on the mAs display.
•
Press the [+] button to increase the product in 1-step increments.
•
Press the [-] button to decrease the product in 1-step decrements.
•
Hold down the [+] or [-] button to change the setting more quickly.
3
3
J Changing the Focal Spot (only for High Power Type (32 kW))
Use the
(focus mode button) to change between large and small focuses.
•
When LED is not illuminated, a large focus is selected.
•
When LED is illuminated, a small focus is selected.
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-23
3 Basic Operation
3.6
X-Ray Radiography
•
X-ray tube units must be warmed up before they can be used. Always age the Xray tube unit before radiography.
^ Reference "3.2 Warm-up the X-Ray Tube Unit" P.3-6
•
Inadequate warm-up may result in the X-ray tube unit discharging during
radiography.
^ Reference "3.4.1 Locking and Releasing the Arm" P.3-15
•
Release the arm lock during radiography.
Radiography is not possible when the system is being moved. To perform
radiography, release the grip bar and inch-mover buttons.
•
Confirm the DR system READY indicator before attempting radiography.
Attempting radiography before the DR system is ready will result in error F68 and
prevent radiography.
•
If the DR system exposure time exceeds 1000 msec, the DR system ready time is
extended to about 30 seconds.
•
Using the collimator, adjust the X-ray irradiation field to the required range, and
then start radiography.
^ Reference "3.4.3 Setting the Irradiation Field" P.3-19
3.6.1
DR Radiography Preparation (for the CXDI-70C Wireless/80C
Wireless)
Link (recognize/connect) the CXDI-70C Wireless/80C Wireless and the imaging unit prior to use.
^ Reference Refer to the Canon CXDI Wireless User's Manual for details.
1
2
3
3-24
Turn ON power to the imaging unit.
Make the DR system recognize the imaging unit.
A connection between the wireless module built into the imaging unit and
the access point built into the system is established automatically by the
communication circuit.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.6 X-Ray Radiography
3
The linking operation does not need to be performed every time when the system is
used in the same radiography environment.
3.6.2
3
3
DR Radiography Procedure (Changing to the DR
Radiography Mode)
3
3
•
•
•
•
1
When using the DR system for a long period of time, connect the power cable to a
power outlet and switch from battery power to outlet power in order to reduce
battery power consumption.
If the system screen is blank after pressing the DR button, and once it is
confirmed that the power cable is not connected to the power outlet, turn both the
key switch and the main circuit breaker to the OFF position, and restart the
system.
Be sure to stop the DR system after use is complete.
If the system is left as is with the DR system operated and the power plug
disconnected from the outlet, the battery will run out. Failure to check may have
an adverse effect on battery life and overall performance.
On rare occasions, DR system will automatically restart during the system startup. In this case, DR system is ready to use after the reboot is complete.
If the message "The external LAN port is not available!" is displayed on the
screen, shut down and restart the DR system. If this message appears frequently,
contact your Shimadzu service representative.
3
3
3
3
3
Press the DR button on the upper cover.
The DR system starts up.
2
Turn ON power to the imaging unit when using the CXDI-70C Wireless/
80C Wireless.
Check that imaging unit is recognized and the communication circuit is connected.
3
3
^ Reference Refer to the Canon CXDI Wireless User's Manual for details.
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-25
3 Basic Operation
3
Select radiography mode.
* • Pressing the
(FPD/cassette radiography mode button) will not switch the mode from
cassette radiography to DR radiography.
Select a radiography mode from the DR system to switch the mode from cassette
radiography to DR radiography. Once done, the corresponding anatomical program is
selected and the radiography parameters are changed to those registered in that program
(except for a part of the radiography methods by CXDI-60G/60C).
• When an APR with long-term exposure is selected, the
button) will be illuminated on the control panel.
4
(DR long-term exposure
Press the hand switch button to the first level (radiography preparation).
Radiography preparation begins. The status indicator lamp will blink green.
* • Radiography preparation cannot start unless the
(standby indicator lamp) is
illuminated. Before pressing the hand switch, confirm that the indicator is illuminated.
• During radiography preparation, the collimator lamp will turn off automatically.
Pressing the Hand Switch
to the First Level
After radiography preparation has finished, the
(radiography ready indicator lamp)
will be illuminated and the status indicator lamp is illuminated green and the collimator
lamp is illuminated again.
* • If an X-ray is not radiated within 30 seconds after the radiography preparation is finished,
the preparation will be automatically cancelled. To prepare for radiography again, release
the hand switch from the state pressed to the first level.
• To illuminate the collimator lamp again after the radiography preparation is finished, see
"4.1 User Settings" P.4-2.
• Confirm that the DR system is READY before attempting radiography. Attempting
radiography before the DR system is ready will result in error F68 and prevent radiography.
• If error F68 appears, cancel the error by confirming the READY indicator on the DR system
touch panel monitor and then pressing the FPD radiography selection button.
• Long-term exposure extends the DR system READY time to about 30 seconds.
3-26
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.6 X-Ray Radiography
5
After confirming the DR system READY indicator, press the hand switch
button to the second level (X-ray exposure).
3
Pressing the button starts X-ray exposure.
During X-ray exposure, the
(X-ray irradiation indicator lamp) is illuminated, a sound
is emitted, and the status indicator lamp will be illuminated yellow.
3
* The sound volume during X-ray exposure can be made louder.
See "4.1 User Settings" P.4-2.
3
Pressing the Hand Switch
to the Second Level
3
3
3
After X-ray exposure, the [kV] and [mAs] displays will blink for 5 seconds.
The status indicator lamp will blink until the battery has recharged.
* • During radiography the hand switch (first and second levels) must be continuously and fully
depressed until the exposure finishes. If the switch is released before completion of the
exposure, the exposure will be terminated prematurely. In this case, the film may not be
clear due to insufficient X-ray exposure.
• After performing radiography, another image cannot be taken until the battery is recharged.
Wait until the
(standby indicator lamp) is illuminated before performing any operation.
Depending on the radiography parameters, this may require 1 minute or more.
• When repeating radiography preparation, release the hand switch (first and second levels)
once.
3
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-27
3 Basic Operation
6
Confirm the reference image.
After exposure, a reference image will appear on the touch panel monitor. Use the monitor
to check for blurs caused by patient movement.
•
The touch panel monitor is not designed for diagnostic use. Displayed images are
for reference use only.
The kV and mAs displays may differ from the radiography parameters displayed
on the touch panel monitor.
•
The system can save a maximum of 3500 images. If this number is exceeded, the
system automatically deletes excess images, starting from the oldest image (all
images taken in an examination are deleted together).
7
After DR radiography is completed, press the DR shutdown button on the
touch panel monitor.
The DR system shuts down.
•
If a DR system malfunction occurs, such as not responding, after confirming that
the power cable is not connected to a power outlet, turn the key switch and the
main circuit breaker to OFF, and stop the DR system.
•
If the DR system is not shut down using the correct procedure, an error will be
displayed on the DR system monitor when the DR system is started up next time.
For further details, refer to the Canon CXDI manual.
8
Press and hold the POWER button (approx. 5 seconds) to turn off the
detector when using the CXDI-70C Wireless/80C Wireless.
Power of the detector are turned off if all the LED lamps are off.
3-28
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.6 X-Ray Radiography
3.6.3
Transferring Images to an External Printer or PACS System
3
3
Do NOT connect any DICOM printer, workstation, PACS, or other
peripheral device that does not meet safety standards, to the external
LAN connector.
3
Doing so could cause unit failure, malfunction, smoke, overheating, electric
shock, or loss or damage of data.
3
Images stored to the system can be output to a DICOM printer, workstation, or PACS.
1
Insert the cable to the external LAN connector at the bottom of the grip bar
and connect it to the printer or network.
2
Output the image from the DR system according to the Canon CXDI
manual.
3
3
^ Reference For further details about the printer or other peripheral units,
refer to the corresponding instruction manual.
3
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-29
3 Basic Operation
3.6.4
Saving Images to External Media
Do NOT connect any peripheral device to the USB connector that does
not meet IEC 60950-1, or any other safety standards equivalent to IEC
60950-1.
Doing so could cause unit failure, malfunction, smoke, overheating, electric
shock, or data loss or damage.
1
Connect a hard drive, USB memory device, or other external storage
media to the USB connector on the upper cover.
2
Save the image to the external media as instructed in the Canon CXDI
manual.
3
Press the [Eject Media] button on the DR system [Examination Image List]
to remove the media.
4
Connect the media to a printer, workstation, or other unit and output the
image.
^ Reference For further details about external storage media (such as USB memory device) or
peripherals (such as printers), refer to the manual for that media/device.
3-30
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.6 X-Ray Radiography
3.6.5
Cassette Radiography Procedure
1
3
Switch to the cassette radiography mode on the X-ray control panel.
To switch to the cassette radiography mode, hold down the
mode button) for about 2 seconds.
(cassette radiography
2
Select an anatomical program from the X-ray control panel, or set
radiography parameters from the current-time product and tube voltage
buttons.
3
Press the hand switch button to the first level (radiography preparation).
3
3
3
Radiography preparation begins and the status indicator lamp will blink green.
* • If the
(standby indicator lamp) is not illuminated, radiography preparation cannot start.
Before pressing the hand switch, confirm that the indicator is illuminated.
3
• During radiography preparation, the collimator lamp will turn OFF automatically.
Pressing the Hand Switch
to the First Level
3
3
After radiography preparation has finished, the
(radiography ready indicator lamp)
will be illuminated and the status indicator lamp is illuminated green and the collimator
lamp is illuminated again.
* • If an X-ray is not radiated within 30 seconds after the radiography preparation is finished,
3
3
the preparation will be automatically cancelled. To prepare for radiography again, release
the hand switch from the state pressed to the first level.
• To prevent the collimator lamp from turning ON again after the radiography preparation is
finished, see "4.1 User Settings" P.4-2.
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-31
3 Basic Operation
4
When the
(radiography ready indicator lamp) is illuminated, press the
hand switch button to the second level (X-ray exposure).
Pressing the button starts X-ray exposure.
During X-ray exposure, the
(X-ray irradiation indicator lamp) is illuminated, a sound
is emitted, and the status indicator lamp will be illuminated yellow.
* The sound volume during X-ray exposure can be made louder.
See "4.1 User Settings" P.4-2.
Pressing the Hand Switch
to the Second Level
After X-ray exposure, the [kV] and [mAs] displays will blink for 5 seconds.
The status indicator lamp will blink until the battery has recharged.
* • After performing radiography, another image cannot be taken until the battery is recharged.
Wait until the
(standby indicator lamp) is illuminated before performing any operation.
Depending on the radiography parameters, this may require 1 minute or more.
• During radiography the hand switch (first and second levels) must be continuously and fully
depressed until the exposure finishes. If the switch is released before completion of the
exposure, the exposure will be terminated prematurely. In this case, the film may not be
clear due to insufficient X-ray exposure.
• When repeating radiography preparation, release the hand switch (first and second levels)
once.
3-32
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.6 X-Ray Radiography
3.6.6
Low-Battery Radiography
No.
Battery Indicator
1
~
2
3
X-ray Radiography Availability
3
Radiography is possible.
Low Battery ⎯ only low-battery radiography is possible.
Blinking
^ Reference "3.6.6 Low-Battery Radiography" P.3-33
Radiography is not possible. Immediately charge the battery
3
3
^ Reference "3.7 Charging the Battery" P.3-35
3
J Current-Time Product Limits for Low Battery Voltage
Low-battery radiography limits DR radiography parameters and exposure may take longer than
normal.
3
^ Reference "8.2.3 High-Voltage Generator" P.8-10
3
3
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-33
3 Basic Operation
3.6.7
Additional Radiography Function
If the system is used continuously while the battery is low, radiography will not be possible.
After the battery indicator stops blinking, if additional radiography is needed, about ten more times is
possible (when radiography condition is general chest radiography).
1
2
Turn the key switch OFF
.
Turn the key switch ON
button).
in combination with the
When the LED is blinking on the battery indicator
(user settings
, low-battery radiography is
possible.
3-34
•
Charge the battery as soon as possible if the necessary radiography is finished.
•
The number of the possible radiography depends on the battery condition.
•
Radiography will not be possible when the battery is depleted.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.7 Charging the Battery
3.7
3
Charging the Battery
3
3
• Charge the battery in a well-ventilated room.
• Do NOT cover the system during the charging of the battery.
3
• When pulling out the power cable, hold the plug portion and do NOT
apply excessive force to the cable.
Excessive force may damage the cable or internal cable reel.
• When rewinding the cable, hold the plug until it is completely
retracted.
3
3
3
•
3
Charge the battery under the environmental conditions specified in "1.4
Environmental Conditions" P.1-4. Charging the battery under conditions
exceeding these limits may significantly reduce battery life and overall
performance.
3
•
Attach an additional protective ground wire with maximum 100 Ω ground
resistance, during the charging of the battery.
•
To maximize battery life and performance, recharge the battery after one or more
LEDs on the battery indicator go off. Charging a fully charged battery may have an
adverse affect on battery life and performance.
•
System travel (movement) is not possible while the power cable is connected to a
power outlet.
•
Radiography is not possible while charging.
When the key switch is set to ON
, charging stops and X-ray radiography can
be performed.
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-35
3 Basic Operation
3.7.1
Charging After Installation
To stabilize battery performance, it is recommended that, after installation, the system is used until
no battery indicator LEDs remain illuminated before recharging. Repeat this cycle several times.
3.7.2
Charging Procedure
1
2
Turn ON the main circuit breaker.
Turn the key switch OFF
.
•
Recharging is not possible with the key switch ON.
•
When turning the key switch ON during the charging of the battery, wait more than
2 seconds until turning the key switch OFF again.
3
Pull the power plug out from the front of the system and insert it into a
power outlet.
Power Plug
The
(power line connection indicator lamp) is illuminated. After about 10 seconds
the battery begins charging. During charging, the
(charging lamp) is illuminated
and the battery indicator indicates the charging status.
Charging
Lamp
Battery Indicator
Blinking
Charging Status
Charging in progress.
~
indicates battery is at
least 90 % charged.
Illuminated
Fully charged.
Indicator segment is illuminated if shown in black J and off if shown in white F above.
3-36
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.7 Charging the Battery
When the battery is fully charged, charging automatically stops and the
(charging
lamp) is illuminated. It normally takes approximately five hours to charge a battery from
no indication on the Battery Indicator to charged, to a level of 90 %, (the Battery Indicator
being fully illuminated). When the battery is charged for more than eight hours it is at
100 % capacity (10 hours max.).
When turning ON the key switch, the
3
3
(charging lamp) will turn OFF.
3
•
During charging, the battery indicator shows the charging status, not the amount
charged.
•
Please note that the battery voltage rises temporarily immediately after charging.
As a result, the battery indicator will indicate more than the actual remaining
charge. After the system is used for a short period, the battery display will display
the actual remaining charge.
If charging is interrupted, the battery indicator will initially be lit up completely.
After a short period, the number of lights may decrease suddenly, but this does
not indicate a malfunction.
•
4
Charging will take a long time if performed with the DR system operating. When
charging the battery, stop the DR system after its operation is finished.
3
3
3
3
After charging, unplug and stow the power cable.
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-37
3 Basic Operation
3.8
Imaging Unit Storage (CXDI-70C Wireless/80C
Wireless)
Lock the CXDI-70C Wireless/80C Wireless to the system to prevent theft when not in use.
Although a wire for theft prevention is provided as part of the system, a padlock is
not. Please provide your own padlock.
3-38
1
Attach the wire to the rear slot.
2
3
Insert the CXDI-70C Wireless/80C Wireless into the front slot.
Loop the end of the wire through the front slot.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
3.8 Imaging Unit Storage (CXDI-70C Wireless/80C Wireless)
4
Lock with a padlock.
3
3
3
3
3
3
3
3
3
3
3
3
3
M503-E028
MobileDaRt Evolution OPERATION MANUAL
3-39
3 Basic Operation
This page is intentionally left blank.
3-40
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4
Chapter
Advanced Operation
4
J Chapter Contents
4.1
User Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.2
Changing the Operation Sound Volume . . . . . . . . . . .4-12
4.3
Changing Anatomical Program Parameters . . . . . . . .4-14
4 Advanced Operation
4.1
User Settings
Item
kV Display
Description
Ref.
Personal ID Setting
Sets and changes personal ID numbers.
P.4-4
Power-Assist Setting
Adjusts the operational characteristics of moving the
system.
P.4-4
Acceleration Setting
Adjusts the acceleration characteristics.
P.4-5
Steer Control Setting
Adjusts the turning characteristics.
P.4-5
Backward PowerAssist Setting
Adjusts the backing characteristics.
P.4-6
Auto-Power-Off Time
Setting
Changes the setting time of the auto-power-off function.
P.4-6
Sets volume of sound emitted from control panel buttons.
Buzzer Setting
Collimator Lamp
Control Setting
P.4-7
Sets volume of warning sound emitted when obstruction
detector is activated.
Enables or disables using hand switch to control radiation
field reference light.
Specifies radiation field reference light after radiography
preparation is finished.
P.4-7
APR Settings at
Startup
Specifies APR radiography parameters specified at startup.
P.4-8
Operation Sound
Setting
Sets the sound tone and volume of the sound emitted
during X-ray exposure or system travel, or when
radiography preparation is finished.
P.4-9
• The grade ability changes in accordance with user settings data (PAS,
Acc, and bAc) for the characteristics of moving the system. The grade
ability decreases when a set value is small. The grade ability increases
when a set value is large.
• When the [PAS] or [Acc] value for power-assist settings has been
changed, not only forward but also backward movement will be influenced.
To maintain the same backward movement, adjust the [bAc] value
accordingly.
4-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.1 User Settings
4.1.1
Entering the User Settings Mode
4
After holding down the
(user settings button) for about 1 second, [usr ope] appears on the Xray control panel momentarily and the display returns to the user settings mode.
4
After changing to the user settings mode, [Pid] first appears on the kV display.
Toggle through the setting items by pressing kV [+] and kV [-] buttons.
Normal Mode
Press the
(user settings button) for
about 1 second.
4
Press the
or
Press the
(registration button).
4
(user settings button).
User Settings Mode
Press the kV [-] button.
4
4
Press the kV [+] button.
4
4
4
If keyless entry is set, after changing to the user settings mode, [PAS] first
appears on the display. [Pid] does not appear on the display.
4
J Registering Changed Data
Press the
(registration button).
This saves the changes to data, displays [rec ord] on the X-ray control panel, and returns to the
normal mode.
4
J Not Registering Changed Data
Press the
4
(user settings button).
This returns to the normal mode without saving changes.
4
* It is possible to register changed data for multiple items at once.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-3
4 Advanced Operation
4.1.2
Personal ID Setting [Pid]
Personal ID number is displayed on the mAs display.
The personal ID number can be set using any 10 digits, from 0 to 9.
Press the mAs [+] and mAs [-] buttons to change personal ID number.
Other settings can be registered for each personal ID.
4.1.3
Power-Assist Setting [PAS]
The power-assist setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
Level of Power Assist
100
Easier to Move
95
K
:
K
75
K
70
Normal
65
L
:
L
35
L
30
Harder to Move
Power-assist is set to [70] when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
4-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.1 User Settings
4.1.4
Acceleration Setting [Acc]
4
The acceleration setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
Level of Acceleration Setting
100
Accelerates more quickly
95
K
:
K
65
K
60
Normal
55
L
:
L
35
L
30
Accelerates more slowly
4
4
4
4
4
Acceleration is set to 60 when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
4.1.5
4
Steer Control Setting [stc]
4
The steering setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
Effect on Steering
100
Easier to turn (normal)
95
K
90
K
:
K
:
L
55
L
50
Easier to go straight
4
4
4
4
Steering control is set to [100] when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
4
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-5
4 Advanced Operation
4.1.6
Backward Power-Assist Setting [bAC]
The backward power-assist setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
Level of Power-Assist for Backward Movement
100
Easier to move backward
95
K
:
K
85
K
80
Normal
75
L
:
L
55
L
50
Harder to move backward
Backward power-assist is set to [80] when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
4.1.7
Auto-Power-Off Time Setting [APO]
The auto-power-off setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
Auto-Power-Off Delay
--
Auto-power-off not set
05
5 minutes
10
10 minutes
15
15 minutes (normal)
20
20 minutes
30
30 minutes
Auto-power-off is set to 15 minutes when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
* The auto-power-off function does not turn the DR system OFF.
See "3.1.2 Shutting Down the System" P.3-4.
4-6
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.1 User Settings
4.1.8
Buzzer Setting [bon]
4
The sound setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
4
Sound Setting
A sound is emitted when X-ray control panel buttons are pressed.
A sound is emitted if the obstruction detector is activated (unless the
obstruction detection feature is cancelled).
4
No sound is emitted when kV [+], [-] or mA [+], [-] buttons are held
down to change radiography parameters rapidly.
No sound is emitted when the obstruction detector is activated.
No sound is emitted when buttons on the X-ray control panel are
pressed (except when registering anatomical programs or changing
settings).
4
4
No sound is emitted when the obstruction detector is activated.
Sound is set to [on] when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
4.1.9
Collimator Lamp Control Setting [LAP]
4
Collimator lamp control setting is displayed on the mAs display.
Press the mA [+] and mA [-] buttons to change settings in the follow sequence:
[O-C]⇔[N-N]⇔[E-N]⇔[O-N]⇔[N-C]⇔[E-C]⇔[O-C]
Left Side of mAs Display
4
Collimator Lamp Setting
After radiography preparation is finished, the collimator lamp is
illuminated until X-ray exposure starts, or for up to approximately 10
seconds, regardless of whether the collimator lamp is turned ON or
OFF when the radiography preparation starts.
The collimator lamp turns OFF during radiography preparation (stop
radiography preparation to turn ON the collimator lamp).
If the collimator lamp is ON, it turns OFF when radiography
preparation starts, then turns back ON after preparation is finished and
remains ON until X-ray exposure starts, or up to a maximum of about
10 seconds.
Right Side of mAs Display
4
Collimator Lamp Setting
4
4
4
Enables using the hand switch to turn the collimator lamp ON/OFF.
Disables using the hand switch to turn the collimator lamp ON/OFF.
4
Collimator lamp control is set to [ O - C ] when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-7
4
4 Advanced Operation
4.1.10 APR Settings at Startup [aPr]
The APR startup setting is displayed on the mAs display.
Press the mAs [+] and mAs [-] buttons to change the setting.
mAs Display
APR Settings at Startup
0
Remembers and restores radiography parameters from just before the
power was shut OFF (if APR radiography parameters were changed
manually, the changed parameters are restored).
1
Restores the parameters for the APR that was used just before the
power was shut OFF (if APR radiography parameters were changed
manually, parameters registered for that APR are restored, rather than
the changed parameters).
2
Displays the parameters for the preselected APR. Select the APR
buttons to display at system startup on the X-ray control panel
(imaging region, projection direction, and body thickness correction).
The corresponding LEDs illuminate. The APR parameters specified
are displayed at system startup.
The APR startup setting is set to [ 0 ] when delivered.
If nothing else needs setting, press the
(registration button) to register the settings. [rec ord]
appears on the X-ray control panel and the user settings mode returns to normal mode.
4-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.1 User Settings
4.1.11 Operation Sound Setting [Snd]
4
On the X-ray control panel, the 3rd digit of the mAs display displays the sound pattern setting No.
The 2nd digit shows the sound tone No. and the 1st digit shows the volume value.
4
mAs display, 3rd digit (sound pattern)
mAs display, 2nd digit (sound tone)
mAs display, 1st digit (sound volume)
Operation Sound Volume
Settings
4
4
4
Registration
System Travel Alert Sound
Setting at Power-Up
Pattern Selection
Operation Sound Pattern
Tone Settings
Checking the Operation Sound
Pattern Tone and Volume
4
4
J Pattern Selection
Use the mAs [-] switch to select the sound pattern setting No.
3rd Digit in mAs Display
(Operation Sound Number)
4
Pattern
1
The sound emitted during X-ray exposure.
2
The sound emitted when radiography preparation is finished.
3
The sound emitted during system travel to alert people of approaching
mobile unit.
4
4
J Setting the Pattern Tone
Use the mAs [+] button to set the tone of the sound pattern.
4
Select from 7 sound tones.
4
4
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-9
4 Advanced Operation
J Setting the Sound Volume
Set the sound volume.
The
(sound volume) setting shown by the 1st digit of the mAs display increases each time the
sound button is pressed.
1st Digit in mAs Display
(Volume Setting Number)
Sound Volume Setting
0
No sound (during X-ray exposure a minimum level of operation
sound is emitted)
1
Minimum volume
2
K
3
K
4
Normal
5
L
6
High
Sound volume is set to "4" when shipped.
Even if the X-ray exposure sound is set to mute, a minimum level of sound is still
emitted to warn those in the vicinity that X-ray radiation is present.
J Setting the sound emitted during X-ray exposure.
4-10
1
Press the mAs [-] button until "1" is displayed in the third digit of the mAs
display.
2
3
4
Press the mAs [+] button to set the sound tone.
Press the
(sound button) to set the sound volume.
Press the
(registration button) to register the settings.
[rec ord] appears on the X-ray control panel and the user settings mode
returns to normal mode.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.1 User Settings
J Setting the sound emitted when radiography preparation is finished.
1
Press the mAs [-] button until "2" is displayed in the third digit of the mAs
display.
2
3
4
Press the mAs [+] button to set the sound tone.
Press the
4
(sound button) to set the sound volume.
Press the
(registration button) to register the settings.
[rec ord] appears on the X-ray control panel and the user settings mode
returns to normal mode.
4
4
J Setting the sound emitted during system travel
1
Press the mAs [-] button until "3" is displayed in the third digit of the mAs
display.
2
3
4
Press the mAs [+] button to set the sound tone.
Press the
4
(sound button) to set the sound volume.
Press the
(remote controller button) and determine whether or not to
emit a system travel alert sound for the setting at power-up.
When the remote controller button is illuminated, holding the brake release
lever will emit a system travel alert sound.
4
4
4
* • When the remote control button is not illuminated and the 1st digit of the mAs display is
blinking, no system travel alert sound will be emitted if you hold the brake release lever.
• To turn the system travel alert sound ON or OFF temporarily without changing the operation
sound setting, lock the arm and hold down the collimator lamp button 1 second or longer.
See"4.2 Changing the Operation Sound Volume" P.4-12
5
Press the
(registration button) to register the settings.
[rec ord] appears on the X-ray control panel and the user settings mode
returns to normal mode.
4
4
4
4
When the system is turned OFF, the system travel alert sound ON/OFF will return to
that made at the operation sound setting [Snd].
4
J Checking the Sound Pattern
Press the collimator lamp button on the X-ray control panel to confirm the sound tone and volume.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-11
4
4 Advanced Operation
4.2
Changing the Operation Sound Volume
The operation sound volume can be easily changed by operating the sound button on the X-ray
control panel or the collimator lamp button.
4.2.1
Changing the Sound Volume Emitted During X-Ray
Exposure or When Radiography Preparation Is Finished
Press the sound button on the X-ray control panel to change the sound volume emitted during X-ray
exposure or when radiography preparation is finished, as shown below.
Sound volume set at the operation
sound setting [Snd]
(LED on the sound button illuminates)
Minimum Volume
(LED on the sound button blinks)
Mute
(LED on the sound button is OFF)
•
Even if the X-ray exposure operation sound is set to mute, a minimum level of
operation sound is still emitted to warn those in the vicinity that X-rays radiation is
present.
•
When the system is turned OFF, the operation sound volume will be reset to that
made at the operation sound setting [Snd].
^ Reference "4.1.11 Operation Sound Setting [Snd]" P.4-9
4-12
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.2 Changing the Operation Sound Volume
4.2.2
Using the Collimator Lamp Button to Adjust Volume
4
When the collimator lamp button is held down 1 second or longer with the arm locked, the system
travel alert sound will be turned ON or OFF as shown below.
4
Sound volume set at the operation
sound setting [Snd]
4
4
Mute
4
If the operation sound volume is set using the sound button on the X-ray control
panel or the collimator lamp button, the system travel alert sound volume will be reset
to that made at the operation sound setting [Snd] when the system is turned OFF.
^ Reference "4.1.11 Operation Sound Setting [Snd]" P.4-9
4
4
4
4
4
4
4
4
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-13
4 Advanced Operation
4.3
4.3.1
Changing Anatomical Program Parameters
Registering Anatomical Programs
In the anatomical program, a total of 72 types (8 imaging portions x 3 directions x 3 body
thicknesses) can be registered for FPD and cassette radiography respectively.
The following radiography parameters can be registered in the anatomical programs.
•
X-ray tube voltage, X-ray current-time product and focal spot size
^ Reference "2.4 X-Ray Control Panel" P.2-6
J Registration Procedure for DR Radiography
•
When registering the anatomical programs for DR radiography, make sure the
anatomical code registered for the DR radiography mode matches the anatomical
program number registered for DR radiography.
•
For further details about editing the DR radiography mode, refer to the Canon
CXDI manual.
For anatomical codes specified in DR radiography modes, see "8.6 Anatomical
Program Number" P.8-23.
•
If using anatomical programs for DR radiography, register radiography parameters
that result in exposure times less than the maximum permitted by the DR system.
•
The maximum exposure time for the DR system is initially set to 1000 msec. It is
possible, however, to extend this to 3000 msec by switching to the DR long-term
exposure mode. If the DR system exposure time exceeds 1000 msec, the DR
system READY time is extended to about 30 seconds.
Settings on DR System
1
Specify radiography parameters for the DR radiography mode.
^ Reference Refer to the Cannon CXDI manual for further details about editing the DR
radiography mode.
4-14
2
Specify an anatomical program number that matches the anatomical code
specified for the DR radiography mode.
3
Register specified radiography parameters.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.3 Changing Anatomical Program Parameters
4
Select the DR radiography mode.
Check that the anatomical program that corresponds to that on the X-ray control panel has
been selected.
4
Settings on Main Unit
1
4
On the X-ray control panel, specify the radiography parameters and focal
size to be registered.
Specify the radiography parameters and focal size specified for DR radiography mode to
the anatomical program.
4
^ Reference "3.5.3 Manually Setting the Radiography Parameters" P.3-23
•
If radiography parameters cannot be set, it means that the exposure time exceeds
1000 msec. Change the anatomical program to the DR long-term exposure mode and
specify the radiography parameters.
^ Reference "Procedure for Changing to DR Long-Term Exposure Mode" P.4-15
•
If reducing exposure time from more than 1000 msec to less than 1000 msec, change
the anatomical program from DR long-term exposure mode to DR radiography mode
before registering radiography parameters.
^ Reference "Changing from DR Long-Term Exposure Mode to DR Radiography Mode"
P.4-16
2
Press the
4
4
4
(registration button) for about 3 seconds.
A sound is emitted and the
(registration button) blinks.
Keep pressing the
(registration button) until [aPr set] appears on the X-ray control
panel, which indicates that the anatomical program has been registered.
During DR long-term exposure mode, the
illuminated.
(DR long-term exposure button) will be
4
4
4
When registering anatomical programs, do not release the registration button until
[aPr set] appears on the X-ray control panel.
4
Procedure for Changing to DR Long-Term Exposure Mode
1
Press the
(DR long-term exposure button).
The
(DR long-term exposure button) blinks.
2
Press the
(registration button).
A sound is emitted and the
(registration button) blinks.
4
4
4
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-15
4 Advanced Operation
3
Hold down the
(registration button) for about 3 seconds until [aPr set] appears
on the X-ray control panel to register the DR long-term exposure mode.
When the DR long-term exposure mode is registered, the
(DR long-term
exposure button) will be illuminated.
4
Specify the radiography parameters to be registered to the DR long-term exposure
mode.
Changing from DR Long-Term Exposure Mode to DR Radiography Mode
1
Set radiography parameters to below 1000 msec.
2
Press the
(DR long-term exposure button) when it is illuminated.
The
(DR long-term exposure button) turns OFF.
3
Make sure that the imaging region, projection direction, body thickness correction, and
radiography parameters are specified as intended.
J Registration Procedure for Cassette Radiography
1
2
Change to the cassette radiography mode.
Select the anatomical program to be registered.
Select the imaging region, projection direction, and body thickness correction.
3
Specify the radiography parameters and focus size to be registered.
Make sure that the imaging region, projection direction, body thickness correction,
radiography parameters, and focal spot size are specified as intended.
4
Press the
(registration button) for about 3 seconds.
A sound is emitted and the
(registration button) blinks.
Keep pressing the
(registration button) until [aPr set] appears on the X-ray control
panel, which indicates the anatomical program has been registered.
When registering anatomical programs, do not release the registration button until
[aPr set] appears on the X-ray control panel.
4-16
MobileDaRt Evolution OPERATION MANUAL
M503-E028
4.3 Changing Anatomical Program Parameters
4.3.2
Deleting Unused Anatomical Programs
4
Anatomical programs not normally used, can be deleted.
Once an anatomical program is deleted it can no longer be used for radiography, even if it is selected.
If a deleted anatomical program is selected, [aPr set] appears on the X-ray control panel.
J Deleting Unused Anatomical Programs
1
4
4
Select the anatomical program to delete.
Select the unused imaging region, projection direction, and body thickness correction.
2
Hold down the
about 3 seconds.
(registration button) and press mAs [+] button for
A sound is emitted and the
4
(registration button) blinks.
Keep pressing the
(registration button) and mAs [+] button until [del ete] appears on
the X-ray control panel, which indicates the anatomical program has been deleted.
4
[_ _ _ _ _ _] appears on the X-ray control panel.
4
•
When deleting an anatomical program, do not release the registration button and
mAs [+] button until [del ete] appears on the X-ray control panel.
4
•
After deleting an anatomical program, change the correspondence between the
anatomical program number and the anatomical code specified for DR
radiography.
4
4
4
4
4
4
M503-E028
MobileDaRt Evolution OPERATION MANUAL
4-17
4 Advanced Operation
J How to Use the Deleted Anatomical Program
If a deleted anatomical program is selected, [_ _ _ _ _ _] will appear on the X-ray control panel and
radiography is not possible.
However, the following procedure allows using a deleted anatomical program temporarily.
1
Select the deleted anatomical program.
Select the deleted anatomical program to be used. [_ _ _ _ _ _] appears on the X-ray
control panel.
2
Hold down kV [+] or kV [-] button, or mAs [+] or mAs [-] button for about 3
seconds.
A sound is emitted.
Continue pressing the button until [_ _ _ _ _ _] disappears from the X-ray control panel
and radiography parameters are displayed.
Radiography parameters in the X-ray control panel will be displayed and allow the use of
the deleted anatomical program for radiography.
4-18
•
To enable using a deleted anatomical program, keep holding down kV [+] or kV [-]
button, or mAs [+] or mAs [-] button until the radiography parameters appear on
the X-ray control panel.
•
This procedure only temporarily enables the use of deleted anatomical program.
To register the radiography parameters used, register them as an anatomical
program.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5
Chapter
Optional Features
5
J Chapter Contents
5.1
Remote Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.2
Grid Unit and Grid Frame . . . . . . . . . . . . . . . . . . . . . . .5-6
5.3
Protective Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
5.4
Dose Area Product Meter . . . . . . . . . . . . . . . . . . . . . .5-10
5.5
Dose Calculation Unit . . . . . . . . . . . . . . . . . . . . . . . . .5-14
5.6
Distance Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
5.7
Keyless Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17
5.8
Changing the Grip Bar Height . . . . . . . . . . . . . . . . . . .5-19
5.9
Luminous Hand Switch . . . . . . . . . . . . . . . . . . . . . . . .5-20
5.10 Additional Hand Switch . . . . . . . . . . . . . . . . . . . . . . . .5-21
5.11 Wireless LAN Option . . . . . . . . . . . . . . . . . . . . . . . . . .5-22
5.12 Second FPD Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-24
5.13 Wired Connection Kit (for the CXDI-70C Wireless/80C
Wireless) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-28
5.14 Handle Unit (for the CXDI-70C Wireless/80C Wireless)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-31
5.15 External Monitor Interface . . . . . . . . . . . . . . . . . . . . . .5-32
5.16 Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-35
5.17 FPD Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-37
5 Optional Features
5.1
Remote Controller
Using the infrared remote controller allows performing radiography and switching the collimator
lamp ON or OFF remotely.
Make sure no TV or other remote controllers are being operated in the room before
using the remote controller for radiography.
Make sure the remote control receiver is not exposed to bright lights, such as
incandescent lamps.
4
1
2
3
No.
5-2
Name
Function
1
Radiography Preparation
Button
Starts radiography preparation.
2
X-ray Exposure Button
Starts X-ray exposure when pressed in combination with the
radiography preparation button.
3
Collimator Lamp Button
Turns collimator lamp ON/OFF.
4
Transmission Status
Indicator
A red LED is illuminated when the remote controller is transmitting a
command to the main unit.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.1 Remote Controller
5.1.1
Using the Remote Controller
5
5
If the red LED transmission status indicator is illuminated even if no
remote controller button is pressed, this indicates a fault with the
remote controller. Stop using the remote controller, remove the battery,
and contact your Shimadzu service representative.
1
Press the
(remote controller button) on the X-ray control panel to
illuminate the LED.
2
Press the radiography preparation button while aiming the remote
controller at the remote control receiver.
5
5
5
The system status indicator lamp will blink green.
^ Reference "Infrared Remote Control Receiver" P.5-4
3
5
Keep pressing the radiography preparation button until the
(radiography ready indicator lamp) is illuminated on the X-ray control
panel.
When the
is finished.
(radiography ready indicator lamp) is illuminated, radiography preparation
The system status indicator lamp is illuminated green.
* If using the DR system for radiography, confirm the DR system READY indicator before
5
5
attempting radiography. Attempting radiography before the DR system is ready will result in
error F68 and prevent radiography.
4
5
Press the X-ray exposure button while pressing the radiography
preparation button.
This irradiates the subject with X-rays and the system status indicator lamp is illuminated
yellow.
Once the X-ray exposure using the remote controller has finished, the remote controller is
disabled for safety reasons and the LED of the
(remote controller button) is turned
OFF.
To continue using the remote controller for radiography, press the
controller button) so the LED is illuminated.
(remote
* • When the
(remote controller button) is illuminated, using the hand switch for
radiography will turn OFF the LED and disable the remote controller.
• Once radiography preparation starts, keep aiming the remote controller transmitter at the
receiver until the X-ray exposure is finished.
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-3
5 Optional Features
5.1.2
Operable Range of the Remote Controller
The remote controller is operable within a radius of 5 meters and range of 270 degrees around the
remote control receiver.
Operable Range of the Remote Controller
Infrared Remote
Control Receiver
5m
270°
Infrared Remote Control Receiver
Receiver
The operable range of the remote controller may be affected by surrounding
conditions.
5-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.1 Remote Controller
5.1.3
Replacing Batteries
5
If the remote controller transmission status indicator is not illuminated when pressing the remote
controller button, replace the batteries.
1
Slide the battery cover on the back of the remote controller in the direction
of the arrow, and remove the cover.
2
3
Exchange the old batteries with new AA batteries (2 cells).
5
5
Re-install the battery cover on the remote controller.
5
5
When not using the remote controller for an extended period, remove
the batteries from the controller.
Otherwise, a battery leak may cause remote controller failure.
5
5
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-5
5 Optional Features
5.2
Grid Unit and Grid Frame
Using the grid and grid frames allows clearer images to be obtained through the reduction of
scattered X-ray influences.
The grid unit is available for the imaging unit CXDI-50G/50C/55G/55C/60G/60C, and also the grid
frame is for the imaging unit CXDI-70C Wireless/80C Wireless.
Remove the grid, however, when it is necessary to reduce the absorbed dose, such as during infant
radiography.
5.2.1
Grid Unit (CXDI-50G/50C/55G/55C/60G/60C)
The optional grid unit can be attached to the imaging unit.
When the grid unit is not being used, store it in the rear slot of the storage case.
^ Reference For detailed information about attaching and detaching the grid unit, refer to the Canon CXDI
manual.
CXDI-50G/50C
CXDI-55G/55C
CXDI-60G/60C
5-6
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.2 Grid Unit and Grid Frame
5.2.2
Grid Frame (for the CXDI-70C Wireless/80C Wireless)
5
The grid frame facilitates attachment and retention of the grid on the imaging unit.
Insert the grid frame into the rear slot of the storage case when not in use.
5
^ Reference For detailed information about attaching and detaching the grid frame, refer to the Canon
CXDI manual.
CXDI-70C Wireless
5
5
5
CXDI-80C Wireless
5
5
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-7
5 Optional Features
5.3
Protective Screen
Operator exposure can be reduced by using the protective screen.
Item
5.3.1
Specification
Lead Impregnated Acrylic
Lead equivalent min. 0.3 mm Pb
Dimensions (during usage)
W46 cm × H34 cm × 0.8 cm thick
Using the Shield
1
Before radiography, stand the protective screen in place by pulling the
handle toward yourself.
Handle
2
After radiography, fold the protective screen down by pulling on the
handle.
Handle
5-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.3 Protective Screen
5
5
• Do NOT rotate the column, raise/lower or store the arm with the
protective screen in the standing position.
Doing so could cause the arm or column to hit and damage the screen.
5
• Handle the protective screen carefully.
Accidentally tipping the protective screen over may damage it.
• Do NOT move the system with the protective screen left in the
standing position.
Tipping the screen over could damage the screen, cover, or other parts.
5
• Be sure to store the protective screen in the folded position.
Storing the screen unfolded may cause the arm to hit the collimator when
stowing the arm.
5
5
5
Using the protective screen limits the optional remote controller's operating range.
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-9
5 Optional Features
5.4
Dose Area Product Meter
The dose area product meter is attached to the front of the collimator and displays the dose-area
product on the dose-area display.
After turning the power ON, wait 3 minutes for the dose area product meter to stabilize.
Dose Area Product Meter
Display Unit
Do NOT subject the dose area product meter to mechanical forces.
Doing so could damage or break the dose area product meter.
5-10
•
Calibrate the dose area product meter periodically (Every two years).
For calibration, contact your Shimadzu service representative.
•
The dose area product meter cannot be attached concurrently with "5.5 Dose Calculation
Unit" P.5-14 or "5.6 Distance Indicator" P.5-16.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.4 Dose Area Product Meter
5.4.1
Specifications
5
J VacuDAP 1570015
•
Dose area product accuracy (compliance with IEC 60580) : 1 m Gycm2
•
Dose area product effective measuring range (lower limit in compliant with IEC 60580):
10 to 999,999 m Gycm2
5
J VacuDAP 1580015
Mode
High Resolution mode
High Rate mode
Dose area product accuracy
(compliance with IEC 60580)
0.1 m Gycm2
1 m Gycm2
Dose area product effective measuring range
(lower limit in compliant with IEC 60580)
1 to 99,999 m Gycm2
10 to 99,999 m Gycm2
•
5
5
5
Measurement mode range
For radiography in large focus, the measurement mode automatically switches. For
radiography in small focus, the measurement mode will not switch.
5
The ranges of each measurement mode in radiography condition are as follows:
•
High Rate mode colored deep gray
•
High Resolution mode colored light gray
5
(kV)
5
5
5
5
(mAs)
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-11
5 Optional Features
5.4.2
Using the Dose Area Product Meter
^ Reference To check the accuracy of the dose area product meter and for maintenance information, refer to
the dose area product meter instruction manual.
1
2
Position the X-ray tube unit at the imaging position.
Perform exposure.
The dose-area product is displayed on the DR system and the dose-area display, and is
recorded in the DICOM image header.
Units of dose-area product are "mGy ⋅ cm2" on the display and "dGy ⋅ cm2" in the
DICOM image header.
3
5.4.3
The display is reset if radiography parameters change or radiography
preparation starts.
Cleaning the Dose Area Product Meter
To clean the dose area product meter, wipe the surface with cloth soaked in pH-neutral cleaner and
which is thoroughly wrung out.
Do NOT pour water on the dose area product meter or use cleaning
agents containing solvents (such as benzene or alcohol) or organic
solvents.
Doing so could break or damage the meter.
5-12
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.4 Dose Area Product Meter
5
•
Make sure no moisture enters the dose area product meter during cleaning. After
cleaning, dry it thoroughly before use. Clean the buttons by wiping it with a cloth
soaked in a disinfectant.
•
The dose-area product is displayed on the DR system and the dose-area display,
and is recorded in the DICOM image header.
• Units of dose-area product are "mGy ⋅ cm2" on the display and "dGy ⋅ cm2" in
the DICOM image header.
5
5
5
5
5
5
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-13
5 Optional Features
5.5
Dose Calculation Unit
The dose calculation unit calculates and displays the X-ray radiation dosage applied to the surface of
the patient’s skin during radiography. The calculation is based on the radiography parameter settings
and the distance between the surface of the patient and X-ray tube unit focal spot which is measured
by an ultrasonic ranging unit attached to the collimator.
Dose Calculation Unit
Display Unit
Calculation Method/
Dose (mGy) =
C (kV) × mAs × 1
(focal spot - patient surface distance)2
C (kV): Correction factor of X-ray tube voltage
C (kV) depends on factors such as radiation quality and system used and is calibrated during
installation.
mAs: Current-time product
5-14
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.5 Dose Calculation Unit
5.5.1
Using the Dose Calculation Unit
1
5
Position the X-ray tube unit and specify the radiography parameters in the
same manner as normal X-ray radiography.
Based on the radiography parameter settings and the distance between the surface of the
patient’s skin and the X-ray tube unit focal spot, the dose calculation unit calculates and
displays the X-ray radiation dosage given to the patient during radiography.
5
5
•
The calculated values displayed by this dose calculation unit are reference values
only. Use them only as a guide.
The calculated values differ from actual dose by up to about 20 %∗1 due to factors
such as distance measurement error, collimator aperture, and X-ray
backscattering. Therefore, the accuracy of displayed values is not guaranteed.
Furthermore, the calculated and actual values can also differ by over 20 %∗1
depending on the imaging region.
*1: Error is with respect to the dosimeter measurement value used in calibration. Due to individual differences in
precision between dosimeters, error values may vary depending on the dosimeter used.
•
Wearing a shirt or other clothing may prevent an accurate measurement of
distance.
•
When measuring distance, do not enter the field of measurement (within a 20 cm
radius from the ranging unit when the field of measurement is at a distance of 1 m
from the focal spot.). Ultrasonic waves are not always radiated in a circular cone
shape. Be sure that persons or objects other than the patient do not enter into the
vicinity of the ultrasonic ranging unit as much as possible. If people or objects
other than the patient enter the field of measurement, distance may be measured
to this secondary person or object rather than to the patient.
•
When distance cannot be measured, [_ _ _ _ _ _] is displayed.
•
The dose calculation unit cannot be used concurrently with "5.4 Dose Area
Product Meter" P.5-10.
5
5
5
5
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-15
5 Optional Features
5.6
Distance Indicator
The distance indicator displays the distance from the focal spot to the surface of the patient. The
distance is measured by the dose calculation unit.
Distance Indicator
Dose Calculation Unit
5-16
•
The display values are reference values only. Use them only as a guide.
Measurement accuracy : ±3 %
(when measuring distances to concrete or similar hard objects, with no persons or
items in vicinity of ranging unit)
Measurable range : 61 to 266 cm
•
Wearing a shirt or other clothing may prevent an accurate measurement of
distance.
•
When measuring distance, do not enter the field of measurement (within a 20 cm
radius from the ranging unit when the field of measurement is at a distance of 1 m
from the focal spot.). Ultrasonic waves are not always radiated in a circular cone
shape. Be sure that persons or objects other than the patient do not enter into the
vicinity of the ultrasonic ranging unit as much as possible. If people or objects
other than the patient enter the field of measurement, distance may be measured
to this secondary person or object rather than to the patient.
•
When distance cannot be measured, [_ _ _ _ _ _] is displayed.
•
To use the distance indicator, attach also "5.5 Dose Calculation Unit" P.5-14.
•
The distance indicator cannot be attached concurrently with "5.4 Dose Area
Product Meter" P.5-10.
•
When performing oblique or other types of radiography, displayed distance may
be shorter than actual distance.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.7 Keyless Entry
5.7
5
Keyless Entry
The system can be managed using a password instead of a key switch.
5
The following operations are restricted if a password is not entered.
•
Electric travel
•
X-ray exposure
•
Start up of DR system
5
5
Secure the password appropriately and renew it periodically.
Do NOT leave a written copy of the password for others to see, and do NOT
write it on the system.
5.7.1
5
Using the Keyless Entry
1
Turn the power switch on the power supply panel ON
.
Turning system power ON requires password entry. [P ] will be displayed in the kV
displays on the X-ray control panel until the password is entered.
2
5
Use the numeric keypad on the X-ray control panel to enter the correct 4digit password, then press the registration button.
5
5
Numeric Keypad
5
5
5
Numeric Delete Button
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-17
5 Optional Features
3
After the correct password is entered, electric travel, radiography and the
DR system are enabled.
•
The power is automatically switched OFF if an incorrect password is entered three
times.
After the third time an incorrect password is entered, the power is immediately
switched OFF whenever an incorrect password is entered.
•
If more or less than 4 digits are entered and the registration button is pressed, a
setting error occurs and [P ] [Err] appears in the kV and mAs displays. Try
entering the password again.
•
Up to 10 passwords can be registered.
User setting parameters can be specified and stored for each password.
•
If the keyless entry is set, after changing to the user settings mode, [PAS] first
appears on the display. [Pid] does not appears on the display.
If keyless entry is specified, the password must be entered each time the system
is started up.
The password is not checked in the following situations:
• The system is restarted within a short time (5 seconds or less) after turning
the power switch OFF
.
• The power switch is turned OFF
and ON
to reset an error.
^ Reference "6.2 Malfunction Indicator Lamp" P.6-4
5-18
•
To use the system with a different personal ID, turn the power switch OFF
and wait for at least 5 seconds before turning the power switch ON
again.
•
Charging the battery is possible only by connecting the power plug. Password
entry is not necessary.
•
Charging is not possible with the power switch ON
•
When the auto-power-off function start with a connection of the power plug,
disconnect the power plug to restart the system.
.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.8 Changing the Grip Bar Height
5.8
5
Changing the Grip Bar Height
The grip bar height can be changed to 950 mm, 970 mm, or 1000 mm from the floor surface. Select
the desired height during installation (default bar height is 910 mm).
5
5
After changing the grip bar height, be careful not to bump against
anybody or any peripheral articles when moving the system.
Since the grip position is far from the main unit, the ease of rotation changes.
•
950 mm
5
5
5
•
5
970 mm
5
5
•
5
1000 mm
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-19
5 Optional Features
5.9
Luminous Hand Switch
The hand switch is illuminated to indicate when the radiography preparation is finished, or that
radiography is in progress.
5.9.1
Using the Luminous Hand Switch
1
Press the hand switch button to the first level.
Illuminated in
Pale Green
2
Confirm that the hand switch is illuminated.
When the radiography preparation is finished, the hand switch is illuminated green.
Illuminated Green
3
Press the hand switch button to the second level.
X-ray exposure begins.
During exposure, the hand switch is illuminated yellow.
Press down on hand switch button continuously until the exposure is finished.
Illuminated Yellow
5-20
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.10 Additional Hand Switch
5.10Additional Hand Switch
5
An illuminated hand switch can be added to the front of the system.
5
5
5
5
5
5
5
During X-ray exposure, use only one of the hand switches.
Though 2 hand switches can be operated at the same time, use only one
hand switch during X-ray radiography in order to avoid unintended exposure.
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-21
5 Optional Features
5.11Wireless LAN Option
By attaching a commercial wireless LAN adapter (USB connection-type), images stored on the
system can be transferred through a wireless LAN network to a DICOM printer, workstation, PACS,
or other external device.
Wireless LAN
Adapter
Wireless LAN
Adapter
CXDI-50G/50C/55G/55C/60G/60C
CXDI-70C Wireless/80C Wireless
5.11.1 Using the Wireless LAN Adapter
1
2
5-22
Connect to the wireless LAN network.
Output images using the DR system, as described in the Canon CXDI
manual.
•
Refer to the manual that comes with the wireless LAN adapter for further
information.
•
When connected to both wired and wireless LAN networks, the wired network
takes priority.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.11 Wireless LAN Option
5.11.2 Wireless LAN Adapter Specifications
Item
5
Description
5
Connection
USB 2.0
Standard
IEEE 802.11g, IEEE 802.11b,
IEEE 802.11a (W52), IEEE 802.11a (W53), IEEE 802.11a (W56)
IEEE 802.11g, IEEE 802.11b
2.4 GHz range
IEEE 802.11a (W52)
5.2 GHz range
IEEE 802.11a (W53)
5.3 GHz range
IEEE 802.11a (W56)
5.6 GHz range
IEEE 802.11g, IEEE 802.11b
2412 to 2484 MHz
IEEE 802.11a (W52)
5150 to 5250 MHz
IEEE 802.11a (W53)
5250 to 5350 MHz
IEEE 802.11a (W56)
5470 to 5725 MHz
5
Frequency Range
5
5
Bandwidth
Other
No auto-install function
Recommended Product
NEC PA-WL300NU-AG (functionality verified by Shimadzu)
5
5
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-23
5 Optional Features
5.12Second FPD Kit
Using the second FPD kit, the CXDI-55G/55C/60G/60C are allowed to attached/detached two
sensor units by means of the sensor cable relay connector. The CXDI-70C Wireless/80C Wireless
are allowed to change over by touch screen operations.
5.12.1 For the CXDI-55G/55C/60G/60C
J Using the Sensor Unit Attach/Detach Function
1
2
3
Turn OFF the remote switch.
4
Turn ON the remote switch.
•
Disconnect the sensor cable relay connector from the sensor unit.
Connect the sensor cable relay the sensor cable relay connector to the
sensor unit.
^ Reference
For details about connecting and disconnecting the sensor cable relay
connector, refer to the CXDI Series Operation Manual.
Remove the sensor cable relay connector when replacing the sensor unit, when straightening
the cable to eliminate the torsion, or when the positioning of the sensor unit is not easy with
the sensor cable relay connector attached.
When moving the system, store the sensor cable relay connector in the connector holder.
Connector Holder
5-24
MobileDaRt Evolution OPERATION MANUAL
Sensor Cable
Relay Connector
M503-E028
5.12 Second FPD Kit
•
Store the sensor cable relay connector of the unused sensor unit in the connector holder.
Connector Holder
5
5
5
5
5
• When transporting the CXDI-55G/55C/60G/60C, do NOT allow the
sensor cable relay connector to drag on the floor or ground or bump
against anybody or any outside objects or personal property. It is
recommended that the user hold the sensor cable relay connector
while transporting the CXDI-55G/55C/60G/60C.
Failure to do so can cause the sensor cable relay connector to get caught
by, or bump against somebody which may result in injury, or cause
damage to personal property, outside objects and/or the sensor cable
relay connector itself.
• When connecting or disconnecting the sensor cable relay connector,
set the power switch of the remote switch to OFF and confirm that
the POWER lamp on both the sensor unit and the remote switch are
off.
The ERROR lamp will light if the sensor cable relay connector is
connected or disconnected while the POWER lamp is lit.
For details about connecting and disconnecting the sensor cable relay
connector, refer to the CXDI Series Operation Manual.
5
5
5
5
5
• Be sure to hold the sensor cable relay connector by its grip in order
to connect or disconnect it.
Holding the cable to connect or disconnect the sensor cable relay
connector may break the wires inside the cable.
5
• Do NOT drop the sensor cable relay connector when connecting or
disconnecting it.
You may injure yourself or cause damage to outside objects, personal
property, or the sensor cable relay connector.
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-25
5 Optional Features
J Storage case
There are front, middle and rear slots in the storage case for Second FPD kit.
Rear Slot
Middle Slot
Front Slot
•
The front slot is used to store the CXDI-55G/55C imaging unit. The imaging unit can be
stored even with the grid unit attached.
•
The middle slot is used to store the CXDI-60G/60C imaging unit. The imaging unit can be
stored even with the grid unit attached.
•
The rear slot is used to store the grid unit and cassette.
In the rear slot, the stored grid unit can be attached to and detached from the imaging unit.
Generating X-rays with a cassette left in the storage case may expose the
film in the cassette.
5-26
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.12 Second FPD Kit
5.12.2 For the CXDI-70C Wireless/80C Wireless
5
J Using the Sensor Unit Attach/Detach Function
No special operation is required for switching the imaging unit. Select the desired imaging unit via
the CXDI control software NE and perform radiography.
5
^ Reference Refer to the Canon CXDI Wireless User's Manual for details.
J Storage case
5
There are front, middle and rear slots in the storage case for Second FPD kit.
5
Rear Slot
Middle Slot
Front Slot
5
5
•
At installation time, the user can designate which imaging unit to store in each slot. The
imaging unit can be stored even with the handle unit attached.
•
The rear slot is used to store the handle unit and cassette.
5
5
Generating X-rays with a cassette left in the storage case may expose the
film in the cassette.
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-27
5 Optional Features
5.13Wired Connection Kit
(for the CXDI-70C Wireless/80C Wireless)
The CXDI-70C Wireless/80C Wireless can be utilized with a wired connection by using the wired
connection kit.
As the cable incorporates both communications and power supply, this connection method can be
used as a backup in cases when the communication state of a wireless connection is unstable or the
amount of remaining charge in the battery is insufficient.
^ Reference Refer to the Canon CXDI Wireless User's Manual for details on how to attach the handle unit.
Always attach the battery pack to the imaging unit regardless of whether you are
using a wired or wireless connection.
5.13.1 Switching from a Wireless to Wired Connection
1
Change the wired/wireless selection switch to "Wired".
The access point display light goes out.
Wired/Wireless Selection Switch
Access Point
2
5-28
Attach the handle unit to the imaging unit.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.13 Wired Connection Kit (for the CXDI-70C Wireless/80C Wireless)
3
Take the sensor cable out of the storage case.
5
5
Storage Case
5
5
4
5
Connect the sensor cable to the imaging unit.
5
Check the connection.
The wired connection is automatically recognized by the CXDI control system.
5
The remainder of the procedure is the same as that for a wireless connection.
5
Always use the handle unit when using a wired connection.
Failure to use the handle unit may result in damage to the connector part of
the imaging unit.
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-29
5 Optional Features
5.13.2 Switching from a Wired to Wireless Connection
1
Change the wired/wireless selection switch to "Wireless".
The access point display light turns on.
Wired/Wireless Selection Switch
Access Point
2
3
Disconnect the sensor cable from the imaging unit.
Store the sensor cable in the storage case.
Storage Case
4
Check the connection.
The wireless connection is automatically recognized by the CXDI control system.
5-30
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.14 Handle Unit (for the CXDI-70C Wireless/80C Wireless)
5.14Handle Unit
5
(for the CXDI-70C Wireless/80C Wireless)
5
Attachment to the imaging unit improves the portability of the imaging unit.
^ Reference Refer to the Canon CXDI Wireless User's Manual for details.
5
CXDI-70C Wireless
5
5
5
CXDI-80C Wireless
5
5
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-31
5 Optional Features
5.15External Monitor Interface
Use the external monitor interface to connect an external monitor.
Only electrical devices conforming to IEC60950-1 or equivalent safety
standards can be connected to the DVI connector as external monitors.
Ignoring this precaution may cause system failure, malfunction, smoke
emission, heat generation or electric shock.
Do NOT connect external monitors that are not configured for connection.
Only external monitors that have been configured for connection in advance can be
connected to the external monitor interface.Connecting an external monitor that has
not been configured for connection will cause both the main unit monitor and external
monitor to display incorrectly. If you wish to add a new external monitor, contact your
Shimadzu service representative with the connection configuration.
5-32
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.15 External Monitor Interface
5.15.1 Using the External Monitor Interface
1
5
Connect the DVI connector cable of the external monitor to the external
monitor interface. (The figure below shows the case of wireless FPD.)
5
5
5
5
External Monitor
Interface
2
Check that the same image is output on both the external monitor and
touch screen monitor.
5
5
5
Touch screen operations cannot be performed on the external monitor.
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-33
5 Optional Features
5.15.2 Specifications
•
Output terminal :DVI-D
•
Output resolution :
1024 x 768 (when using CXDI Control Software NE)
640 x 480 (when using CXDI Control Software RD)
•
Display method: Duplicate display
•
Recommended external monitor : TOTOKU ELECTRIC CO. Ltd CCL208
•
Recommended cable length: 5 m max.
When using CXDI Control Software NE, check that the external monitor and
cable are connected before starting the DR system.
Images will not be displayed on the external monitor if the cable is connected
after the DR system starts up.
If the cable is disconnected and then reconnected after startup, images will
be displayed again.
5-34
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.16 Barcode Reader
5.16Barcode Reader
5
The barcode reader allows the study order associated with the read "Patient ID" or "Accession
Number" to be automatically selected from the study list obtained from the MWM worklist.
5
5
Perform radiography after checking that the correct study order is
selected.
5
5
Whether the patient ID or Accession Number is read from the barcode is
configured during installation. (Only the patient ID can be read when using
CXDI Control Software RD.)
5
5
5.16.1 Using the Barcode Reader
1
Display the study list on the DR system.
2
Read the patient ID or Accession Number from the barcode.
5
^ Reference Refer to the Canon CXDI Wireless User's Manual for details.
The study orders are narrowed down and redisplayed. (Only the patient ID can be read
when using CXDI Control Software RD.)
5
5
5
The target study order is selected.
^ Reference Refer to the Canon CXDI User's Manual for details.
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-35
5 Optional Features
Secure the barcode reader with the belt when moving the system.
5-36
MobileDaRt Evolution OPERATION MANUAL
M503-E028
5.17 FPD Bag
5.17FPD Bag
5
This bag allows the imaging unit to slide smoothly.
5
Put the bag over the imaging unit.
5
5
CXDI-55G/55C
5
CXDI-60G/60C
5
5
CXDI-70G Wireless/70C Wireless
5
CXDI-80C Wireless
5
5
5
5
5
M503-E028
MobileDaRt Evolution OPERATION MANUAL
5-37
5 Optional Features
This page is intentionally left blank.
5-38
MobileDaRt Evolution OPERATION MANUAL
M503-E028
6
Chapter
Troubleshooting
6
J Chapter Contents
6.1
Warning Indicator Lamp . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.2
Malfunction Indicator Lamp . . . . . . . . . . . . . . . . . . . . . .6-4
6.3
Releasing the Brake in Emergency . . . . . . . . . . . . . . . .6-5
6.4
Notification Function . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
6.5
Storing the System for a Long Period . . . . . . . . . . . . .6-10
6 Troubleshooting
6.1
Warning Indicator Lamp
The
(warning indicator lamp) is illuminated and radiography is disabled under the conditions
described below. The warning code appears on the mAs display for about 3 seconds.
Warning Indicator Lamp
mAs Display
Code
bbb
Description
Radiography not possible due to low battery
voltage.
d0,
Travel system error
d13 to 15
d1 to d3
Motor system error
Solution
Charge the battery.
Immediately stop using the system and contact
your Shimadzu service representative.
If the unit must be moved, do not restore power
to the system. See "6.3 Releasing the Brake in
Emergency" P.6-5.
Immediately stop using the system and contact
your Shimadzu service representative.
Release the brake release bar.
d4
Brake release bar was activated during startup.
d5, d6
Inch-mover button was activated during
startup.
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
Release the inch-mover button.
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
Release the grip bar.
d7, d8
Grip bar was activated during startup.
d9
Obstruction detector was activated during
startup.
d10
An arm lock release button was activated
during startup.
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
Remove the input to the obstruction detector.
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
Release the arm lock release button.
6-2
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
6.1 Warning Indicator Lamp
Code
Description
6
Solution
Release the arm lock release lever.
d11
Arm lock release lever was activated during
startup.
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
6
Stop traveling.
d12
E
EEE
Increased temperature was detected in control
circuit.
Indicates the replacement period for the CPU
memory backup battery.
Warning clears after the control circuit cools
and the temperature falls.
The system can be moved until the lighting of
warning indicator is canceled
although radiography is available.
Although normal travel and radiography are
possible, the CPU memory backup battery must
be replaced.
6
6
Contact your Shimadzu service representative.
E5
Indicates the replacement period for the
emergency brake release battery.
6
Although normal travel and radiography are
possible, the brake cannot be released in an
emergency.
The emergency brake release battery must be
replaced.
Contact your Shimadzu service representative.
F64
F65
Invalid hand switch operation.
F68
X-ray exposure was attempted before DR
radiography preparation was finished.
F76
Invalid anatomical code (invalid value) was
sent from DR system.
F6d
Emergency stop button was pressed.
Warning clears after the hand switch button is
released.
6
Wait until DR radiography preparation is
finished before starting radiography.
Warning clears when the FPD radiography
mode button is pressed.
Make the anatomical code correspond with the
anatomical program number.
Warning clears when a valid value is sent from
the DR system.
Resolve error.
H1
Under specified radiography parameters
exceeds maximum anode heat content.
6
6
6
Warning clears when the emergency stop button
is turned in the direction of the arrow and reset.
Reduce radiography parameters, or stop
radiography until the warning indicator lamp
turns OFF.
Warning clears when X-ray tube unit cools
sufficiently.
H2
Maximum anode heat content has reached.
Stop radiography until the warning indicator
lamp turns OFF.
H4
X-ray tube unit temperature is too high.
Warning clears when X-ray tube unit cools
sufficiently.
6
6
6
^ Reference For further details on DR system errors, refer to the Canon CXDI manual.
6
M503-E028
MobileDaRt Evolution OPERATION MANUAL
6-3
6 Troubleshooting
6.2
Malfunction Indicator Lamp
If the self-diagnostic function detects any malfunction in the system, the
(malfunction indicator
lamp) is illuminated, and radiography is disabled. The malfunction code appears on the mAs display.
If the malfunction indicator lamp is illuminated, switch OFF the power after turning the key switch
to OFF
.
If the malfunction indicator lamp remains illuminated after turning the
power OFF and ON, it indicates that the malfunction still remains. When
this occurs, do NOT operate the system.
If the malfunction indicator lamp turns OFF, but is illuminated
repeatedly, do NOT operate the system. Immediately turn the power
OFF and contact your Shimadzu service representative.
Code
Description
F01, F02
Supply voltage exceeds normal range
F03
Internal power supply circuit error
X-ray tube voltage control error
F1*1
X-ray exposure is not cancelled if F15 or F16 appear, however, these codes indicate that a tube
voltage control error is detected. If F15 or F16 appears frequently, battery performance may
have deteriorated.
X-ray Tube Unit Discharge
F14
The error may be reset by warm-up the X-ray tube unit.
^ Reference "3.2 Warm-up the X-Ray Tube Unit" P.3-6
F2*1
X-ray tube current control error
F3*1
Filament heating control error
F4*1
Capacitor charging control error
F5*1
Battery or charging error
The plug might be connected to the outlet with the main breaker OFF.
F5C
Check the status of the main breaker.
^ Reference "3.7 Charging the Battery" P.3-35
F61
Starter error
F62
Imaging unit error
F6E, F6F
Radiography error
*1: Indicates a number from 0 to 9 or a letter from a to f.
^ Reference For DR system error information, refer to the Canon CXDI manual.
6-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
6.3 Releasing the Brake in Emergency
6.3
6
Releasing the Brake in Emergency
This system is equipped with a function to release the brake in an emergency, such as a case when
the system malfunctions or the battery runs out.
6
6
• Only use the brake release function in emergencies.
6
• Be especially careful when using the brake release function on
sloped surfaces.
6.3.1
6
Activating the Brake Release Function
1
2
Turn OFF
the key switch on the power supply panel.
6
Press the emergency stop button.
This activates the brake release function.
* When the brake release function is activated, the brake release lever on the grip bar does not
function. The key switch is also turned OFF, so power-assist does not function either.
6.3.2
Releasing the Brake in Emergencies
6
6
While pressing the emergency brake release button located to the left of the grip bar, the brake is
released to allow moving the system by hand.
6
Emergency Brake Release Button
6
6
6
6
M503-E028
MobileDaRt Evolution OPERATION MANUAL
6-5
6 Troubleshooting
6.4
Notification Function
If the warning indicator illuminates when turning ON
the key switch and the kV display indicates
any value from "U01" to "U05", it is time to replace a specific part of the system.
Contact your Shimadzu service representative.
KV Display
kV Display
Part
Warning Indicator Lamp
Information on Replacement Time Display
U01
Battery
Displayed after 2 years from installation.
U02
CPU Backup Battery
Displayed after 2 years from installation.
U03
Emergency Brake
Release Battery
Displayed after 2 years from installation.
U04
X-Ray Tube Unit
Displayed after 30,000 exposures from installation.
U05
Motor
Displayed after the system travels 7,500 km from installation.
When the warning indicator illuminates and the kV display indicates any value from
"U01" to "U05", only brake release and arm operations are available.
Pressing the
6-6
(registration button) allows normal operation.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
6.4 Notification Function
6.4.1
Confirming the Usage Period, Count and Distance
6
Information for Each Part
Turn the key switch to ON
while holding the
(remote controller button) down, to display
the usage period, count, and distance information for each part in the kV and mAs displays on the Xray control panel. Toggle between information for each part by pressing the kV [+] and [-] buttons, or
the imaging region selection buttons, to display the information. This causes the lamp to blink for the
region button corresponding to the selected replacement part. When the kV display, mAs display,
and warning indicator blink, it is time to replace a specific part of the system. Contact your Shimadzu
service representative.
6
6
6
6
6
Remote Control Button
Imaging Region
Selection
Button
Part
6
Information Displayed in kV and mAs
Displays
Battery
Number of months used
CPU Backup Battery
Number of months used
Emergency Brake Release Battery
Number of months used
X-Ray Tube Unit
Times X-rays were exposed
Motor
Distance traveled
6
6
6
To return to normal operating mode after confirming information, temporarily turn the power OFF
and restart the system.
6
6
When the kV/mAs displays indicate the duration period, number of times, and
distance of each part, and when the warning indicator blinks, only brake release and
arm operations are available.
M503-E028
MobileDaRt Evolution OPERATION MANUAL
6-7
6
6 Troubleshooting
6.4.2
Temporarily Disabling the Replacement Time Display
Function
When the replacement time is displayed, contact your Shimadzu service
representative to replace the parts within the temporarily disabled
period.
Do NOT continue to use the system by temporarily disabling the replacement
time indication repeatedly.
1
Turn the key switch ON while pressing the
(remote controller button)
on the X-ray control panel. Check the data on usage, number of cycles,
and distance for each part.
2
Press the mAs [+] and [-] buttons.
This displays [clr] [ no] in the kV and mAs displays on the X-ray control panel.
mAs [+] and [-] Button
3
Remote Control Button
Press the mAs [+] and [-] buttons.
This displays [clr] [YES] in the kV and mAs displays on the X-ray control panel.
6-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
6.4 Notification Function
4
With [clr] [YES] shown on the kV and mAs displays, hold the
(registration button) down for at least 3 seconds.
6
This can temporarily disable the replacement time display function.
6
6
6
Registration Button
Even if the replacement time display function is temporarily disabled, after a given period of time
has elapsed, a warning sound is emitted and a replacement time and warning indicator are displayed
on the X-ray control panel when the key switch is turned ON .
6
The period for which the replacement time display function can be temporarily disabled differs
depending on the part.
Part
Period Notification Function Can Be Temporarily Disabled
Battery
1 month
CPU Backup Battery
1 month
Emergency Brake Release Battery
1 month
X-Ray Tube Unit
3000 exposures
Motor
100 km travel distance
6
6
6
6
6
While the replacement time display function is temporarily disabled, the above
cancellation procedure cannot be performed until the replacement time display
function is enabled again.
6
6
6
M503-E028
MobileDaRt Evolution OPERATION MANUAL
6-9
6 Troubleshooting
6.5
Storing the System for a Long Period
Avoid leaving the system unused for extended periods.
If the system is not to be used for an extended period, follow the procedures below to
reduce battery wear. This does not, however, extend battery life. Shimadzu does not
guarantee the battery life even if the following procedures are implemented.
6.5.1
Storage Procedure
1
2
3
Fully charge the battery before storing the system.
After charging, turn the main circuit breaker OFF.
While storing the system, keep the ambient temperature between -10 °C
and 40 °C.
It is recommended to keep the temperature below 30 °C.
4
Recharge battery once every 6 months during storage (given that ambient
temperature is 25 °C).
Double this frequency for every 10 °C increase in ambient temperature above 25 °C.
5
6-10
After storing the system for an extended period, run the system until the
battery indicator displays no illuminated portion, then charge the battery
and repeat this cycle several times to restore battery performance.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
7
Chapter
Maintenance and
Inspection
7
J Chapter Contents
7.1
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.2
Cleaning the System . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
7.3
Periodic Replacement Parts . . . . . . . . . . . . . . . . . . . .7-10
7 Maintenance and Inspection
7.1
7.1.1
Inspection
Daily Inspection
Before using the system, be sure to inspect the following items.
If any abnormality is found during the maintenance check, stop using
the system and contact your Shimadzu service representative.
When a user performs the maintenance check for himself or herself, please
keep in mind his or her own safety. Shimadzu shall not be liable for any
damage resulting from the checks other than those by Shimadzu or
Shimadzu representative.
X-ray Tube Unit
Column
Arm
Arm Joint
Collimator
Power Supply Panel and Touch Panel Monitor
Cable
X-ray Control Panel
Grip Bar
Hand Switch
Power Plug
Obstruction
Detector
Grounding Cable
Imaging Unit
Wheel
7-2
Storage Case
MobileDaRt Evolution OPERATION MANUAL
M503-E028
7.1 Inspection
No.
Item
7
Inspection Point
1
Appearance
•
Any cracks, dents, or other damage to the cover?
2
Grip Bar
•
Is the grip bar to loose?
•
Any abnormal sounds during rotation?
•
Does the column stop if the arm lock release button or lever is released
during rotation?
•
Any signs of physical misalignment?
3
4
Column
7
•
Is the arm joint connected without undue stress?
•
Any abnormal sounds, looseness, or resistance during up/down, forward/
reverse, or extension/contraction movement?
•
Does the arm stop if the arm lock release button or lever is released during
up/down, forward/reverse, or extension/contraction movement?
•
Is the arm to loose when extended to maximum limit?
•
Is the rail attached below inner arm too loose?
Arm
7
7
7
Arm Joint
7
Inner Arm
7
•
Is the top plate loose?
•
Is the arm catch too loose?
7
Screws
5
Arm Joint
Top Plate
7
Arm Catch
7
6
X-ray Tube Unit
•
7
Is X-ray tube unit holder loose?
7
7
M503-E028
MobileDaRt Evolution OPERATION MANUAL
7-3
7 Maintenance and Inspection
No.
Item
Inspection Point
•
Is the irradiation field made narrow?
•
Any loose parts in the vicinity of the collimator?
A
View From A
Screw B
7
Screw A
Collimator
Screws
8
9
X-ray Control Panel
•
Any flickering or abnormal display?
•
Are the front wheel attachments too loose?
•
Any physical damage or distortions to the front wheels?
•
Any abnormal vibrations or sounds when the drive wheels rotate?
Wheels
Screws
10
7-4
Obstruction Detector
•
Does it appear normal?
•
Any loose parts?
•
Does the switch function normally?
MobileDaRt Evolution OPERATION MANUAL
M503-E028
7.1 Inspection
No.
Item
•
11
7
Inspection Point
Does it appear normal?
•
Is the grounding cable touching the floor?
•
Is any debris attached to the cable?
7
7
Grounding Cable
Grounding
Cable
7
Remove dust that has adhered to the belt.
•
Is the power plug case damaged?
•
Are any of the plug prongs bent or deformed?
•
Is the cover damaged or frayed?
•
Anything abnormal about the power outlet used?
7
7
Power Plug
12
Power Plug
7
Prongs Not Deformed
13
Cables
•
Is the cable cover of cable reel damaged or frayed?
•
Any resistance or unusual noise when reeling the cable in or out?
•
Any twisting or physical damage of the outer insulation?
•
14
Wire Rope
Is the wire rope
7
No Damage to Cover
7
broken?*1
•
Any strands frayed or damaged?
•
Any corrosion?
7
^ Reference "7.1.2 Periodic Inspection" P.7-6
15
16
Storage Case
Other
•
Is there any dust on the fan vents?
•
Do the fan vents exhausted normally?
•
Any unusual smells or abnormal sounds when the power is turned ON or
while the battery is being charged?
•
Any abnormal sound or vibration coming from the X-ray tube unit?
•
Any warning or malfunction indication appears?
7
7
*1: If even a single strand of wire rope is broken, stop using the system and contact your Shimadzu service representative
immediately.
7
M503-E028
MobileDaRt Evolution OPERATION MANUAL
7-5
7 Maintenance and Inspection
J Inspection for Warning and Caution Labels
Periodically (once a year) inspect the labels attached on the system.
If any label is peeled or unreadable by stain or scratch, contact your Shimadzu service representative
for replacement of a new one.
^ Reference "Warning and Caution Labels" P.xxiv
7.1.2 Periodic Inspection
Even if nothing unusual is detected during daily inspections, contact your Shimadzu service
representative to inspect the following items:
Be sure to perform periodic inspection (every 6 months).
Failure to do this may cause serious accidents or significantly shorten the
lifetime of the system.
Periodic inspections mainly check the system performance and the internal mechanisms. The
inspections require good knowledge of the internal mechanisms and can also be dangerous. Contact
your Shimadzu service representative to request a periodic inspection. It is recommended to conduct
periodic inspections every 6 months. A fee is charged for periodic inspections after expiry of the
warranty periods.
J Periodic Inspection (by Shimadzu Service Representative)
No.
1
Column
2
Battery
3
4
5
7-6
Item
Inspection Point
•
Confirm that all fastener bolts are tight
•
Any abnormalities in appearance?
•
Voltage correct?
Power Supply
•
Is the output voltage normal?
X-ray Output
•
Confirm X-ray tube voltage, current, and exposure time.
Wire Rope*1
6
X-ray Tube Rotation
7
Arm Catch
•
Any frayed or broken strands?
•
Is the rope secured normally?
•
Is the rope lubricated?
•
Any deformation or corrosion?
•
Are rotating parts lubricated?
•
Confirm arm stows securely. Is the catch secure?
•
Is the arm catch too loose?
•
Confirm storage signal
MobileDaRt Evolution OPERATION MANUAL
M503-E028
7.1 Inspection
No.
Item
8
Arm
9
Obstruction Detector
10
Grounding Cable
11
Cable
12
Power Plug
7
Inspection Point
•
Any resistance or abnormal sound during up/down movement?
•
Is appearance normal?
•
Confirm normal operation
•
Is appearance normal?
•
Are any foreign objects attached?
•
Any twisting or damage to the cable insulation?
•
Is the casing of power plug damaged?
•
Any gap between the plug and stopper?
•
Are any prongs bent or deformed?
•
Is the cable cover of cable reel damaged or frayed?
•
Any resistance or unusual noise when reeling in or out the cable?
•
Does irradiation field adjusting knob move smoothly?
7
7
7
13
Collimator
*1
Even if wire ropes appear normal during daily inspection, have your Shimadzu service representative replace the
wire rope 7 years after installation.
Shimadzu Part No:
503-64844 (for Type T)
2 cables
503-64844-03 (for Type S)
2 cables
7
7
7
7
7
7
7
7
7
M503-E028
MobileDaRt Evolution OPERATION MANUAL
7-7
7 Maintenance and Inspection
7.2
Cleaning the System
Be sure to turn the system power OFF before cleaning and disinfecting
the system.
Otherwise, a malfunction may occur in the system, or the system may
operate in an unintended way.
Thoroughly ventilate the room before turning ON the power after disinfection
work is complete.
Be sure to clean and disinfect the system.
Cleaning and disinfection is very important to ensure that the system can be
used hygienically and safely. Strictly follow the methods prescribed.
Do NOT directly apply rubbing alcohol or water, or spray it onto the
system.
Wipe the surface of the system with a cloth soaked with rubbing alcohol.
If rubbing alcohol gets inside the system, it can cause failure or accidents.
To clean the dose area product meter, wipe the surface with cloth soaked in
pH-neutral cleaner and which is thoroughly wrung out.
Do NOT use an organic solvent.
Organic solvents may change the surface color. If an organic solvent adheres
to the surface, wipe it out immediately.
7-8
MobileDaRt Evolution OPERATION MANUAL
M503-E028
7.2 Cleaning the System
7
7
Do NOT use the following disinfectants:
If any of the following disinfectants are applied, the system performance and
safety cannot be guaranteed.
7
• Chlorine-based disinfectants
• Disinfectants that corrode metals, plastics, rubber, or paint
7
• Disinfectants unsuitable for metals, plastics, rubber, or paint
• Spray-gas type disinfectants
• Volatile disinfectants
7
• Disinfectants that may enter the system
7
7
Use disinfectants at a minimum.
Repeated disinfection over a long time may lead to discoloring and cracking
on the system surface, and deterioration of rubber and plastic. If any
abnormality is found on the system after disinfection, stop using the system
immediately. Contact your Shimadzu service representative for repair.
7
7
7
•
Never use paint thinners or other organic solvents, other than rubbing alcohol for
disinfection.
•
Film-developing or fixing solutions will cause discoloration.
Wipe off immediately with a damp cloth.
•
Make sure no moisture enters the dose area product meter during cleaning. After
cleaning, dry it thoroughly before use.
7
7
7
M503-E028
MobileDaRt Evolution OPERATION MANUAL
7-9
7 Maintenance and Inspection
7.3
Periodic Replacement Parts
Certain parts should be replaced periodically to ensure the system continues to function properly.
When replacing consumable parts, use the genuine parts with the following part number.
These periodic replacement parts should be installed by your Shimadzu service representative.
• If anything unusual is found regarding wire ropes during daily
inspections, contact your Shimadzu service representative.
• Battery life varies widely depending on operating conditions. If the
number of uses per charge decreases significantly, it may be due to
the expiration of battery life. Have the battery replaced by your
Shimadzu service representative.
As a general guideline, even if nothing unusual is noticed, replace
the battery every 2 years.
Part Name
7-10
Part No.
Recommended
Replacement Period
Fuse
See next page.
1 year
Halogen Lamp
062-65012-02
1 year
Silicon Packing (for high-voltage generator)
532-23029
1 year
Silicon Packing and Silicon Grease (for X-ray tube unit)
511-15061-13
1 year
Contact Pin
511-15061-11
1 year
Battery (replacement requires 20 units)
074-71014-01
2 years
Cable Reel (Power Supply)
503-75758
2 years
Grounding Cable
503-57552
2 years
CPU Backup Battery
074-73306-10
2 years
Battery, Alkaline LR61
074-70020-06
2 years
Relay, G7Z-4A-11Z DC24
065-60052-01
2 years
Relay, G9EA-1-B DC24
065-60053-01
5 years
O-Ring, 4D SS-060
036-11143-21
5 years
SW, Selector HW1K-2B40
064-90066-01
5 years
SW, AVLW32211D-R
064-60785-97
5 years
Cable Cover
511-18106
5 years
MobileDaRt Evolution OPERATION MANUAL
M503-E028
7.3 Periodic Replacement Parts
Part Name
Recommended
Replacement Period
Part No.
Cable Cover
511-18106-01
5 years
Wire Rope (for Type T)
503-64844
7 years
Wire Rope (for Type S)
503-64844-03
7 years
Motor
511-77099-01
7500 km
X-Ray Tube Unit
532-24780
7
30,000 exposures
J Fuse Rating List
Unit / Board
7
7
Component Name
DR Unit
F21
Inverter Unit Board*1
F1
Current
Rating
Voltage
Rating
2A
250 V
072-01664-26
150 A
600 V
072-06033-10
160 A
660 V
072-06037-11
Part No.
MUX Charge-32K Board
F1
0.5 A
250 V
072-01664-15
MUX Charge-04B Board
F1
5A
250 V
072-01664-33
F1, F2, F10
2A
250 V
072-01664-26
6.25 A
250 V
072-01664-34
F4, F5
10 A
250 V
072-01659-83
F6
0.5 A
250 V
072-01664-15
F7, F8
5A
250 V
072-01664-33
F9
10 A
250 V
072-01665-34
F1
6.25 A
250 V
072-01664-34
F3
7
7
7
7
MUX Power Board
MUX LC1BF Board
7
7
*1: For the inverter unit board, select a current rating of either 150 A or 160 A.
7
7
7
7
M503-E028
MobileDaRt Evolution OPERATION MANUAL
7-11
7 Maintenance and Inspection
This page is intentionally left blank.
7-12
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8
Chapter
Specifications
8
J Chapter Contents
8.1
Environmental Conditions of EMC (Electromagnetic
Compatibility) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
8.2
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9
8.3
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . .8-17
8.4
X-Ray Tube Unit Reference Axis and Focal Spot Position
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-18
8.5
Radiography Parameters . . . . . . . . . . . . . . . . . . . . . .8-19
8.6
Anatomical Program Number . . . . . . . . . . . . . . . . . . .8-23
8.7
Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-34
8.8
Statement of Compliance [For Europe] . . . . . . . . . . . .8-40
8.9
Statement of Compliance with Standards . . . . . . . . . .8-42
8.10 Manufacturer Information . . . . . . . . . . . . . . . . . . . . . .8-43
8 Specifications
8.1
Environmental Conditions of EMC
(Electromagnetic Compatibility)
The system satisfies the EMC (Electromagnetic Compatibility) standard below:
• IEC 60601-1-2:2007
Pay attention to the electromagnetic circumstances at the installation
site.
The system may be affected by the electromagnetic environment at the
installation site. Also, the installation of the system may affect other existing
equipment.
J Classification of EMI in Accordance with EN/IEC60601-1-2: 2007
Group 1, Class A
The system uses radio-frequency energy only for its internal function and is not intended to deliver
energy to the patient. But little leakage radio-frequency energy does harm to highsensitive
equipment.
The system main power line in the clinical site should be connected to the domestic power sources
which are separated from the public main network.
For replacement parts of internal components, make sure to apply the
cables supplied by Shimadzu.
The use of non-cable devices, accessories, or cables other than those sold
by Shimadzu as replacement parts for the internal components may result in
increased emissions or decreased immunity of the system.
J Performance to Be EMC Immunity Tested (Essential Performance)
Essential performances of this system are as followings;
8-2
•
Positioning the X-ray irradiation field
•
Setting radiography parameters
•
Irradiating imaging region with X-rays
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.1 Environmental Conditions of EMC (Electromagnetic Compatibility)
J Cable List
8
Cable
Length
Shield
4.0 m
No
Shimadzu Corporation / 503-75758
4.5 m
Yes
-
FPD Cable (CXDI-70C Wireless/80C
Wireless)
3.0 m
Yes
-
Hand Switch Cable
4.5 m
No
Shimadzu Corporation / 511-64136
Luminous Hand Switch Cable
4.5 m
No
Shimadzu Corporation / 511-64129
Additional Protective Grounding Wire
4.2 m
No
Shimadzu Corporation / 501-90150-02
LAN Cable
3.0 m
No
-
DVI Cable
5.0 m
Yes
-
USB Cable (Barcode Reader)
1.5 m
Yes
Shimadzu Corporation / 511-90683-01
Cable Name
Power Cable
FPD Cable
(CXDI-50G/50C/55G/55C/60G/60C)
Note (Manufacturer/Part No.)
8
8
8
8
8
8
The cables listed above are the parts specified to be compliant with the
standards. These parts are not provided to the system.
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-3
8 Specifications
J Option List
Option Type
Note (Manufacturer)
Remote Controller
Shimadzu Corporation
Dose Area Product Meter
-
Dose Calculation Unit
Shimadzu Corporation
Distance Indicator
Shimadzu Corporation
Luminous Hand Switch
Shimadzu Corporation
Additional Hand Switch
Shimadzu Corporation
Wireless LAN
Shimadzu Corporation
FPD Wired Connection Kit
Shimadzu Corporation
External Monitor Interface
Shimadzu Corporation
Barcode reader
Shimadzu Corporation
The option listed above are the parts specified to be compliant with the
standards. These parts are not provided to the system.
8-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.1 Environmental Conditions of EMC (Electromagnetic Compatibility)
J Guidance and Manufacturer's Declaration - Electromagnetic Emissions
8
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The MobileDaRt Evolution is intended for use in the electromagnetic environment specified below. The
customer or the user of the MobileDaRt Evolution should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
EN 55011/
CISPR 11
Group 1
RF emissions
EN 55011/
CISPR 11
Class A
Harmonic emissions
EN 61000-3-2/
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
EN 61000-3-3/
IEC 61000-3-3
Class A
8
Electromagnetic Environment - Guidance
The MobileDaRt Evolution use RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
8
8
The MobileDaRt Evolution is suitable for use in
all establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
8
8
Complies
8
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-5
8 Specifications
J Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MobileDaRt Evolution is intended for use in the electromagnetic environment specified below. The
customer or the user of the MobileDaRt Evolution should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
EN/IEC 60601
Test Level
Compliance Level
Floors should be wood,
concrete or ceramic tile.
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Electrical fast transient /
burst
±2 kV
±2 kV
for power supply lines
EN 61000-4-4/
IEC 61000-4-4
±1 kV
EN 61000-4-2/
IEC 61000-4-2
±1 kV
for input/output lines
If the floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
±1 kV line(s) to line(s) ±1 kV line(s) to line(s)
that of a typical commercial or
±2 kV line(s) to earth ±2 kV line(s) to earth hospital environment.
EN 61000-4-5/
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11/
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
for power supply lines
for input/output lines
Surge
Electromagnetic
Environment - Guidance
5 % UT
(>95 % dip in UT)
for 0.5 cycle
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 5 sec
3 A/m
3 A/m
EN 61000-4-8/
IEC 61000-4-8
Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of the MobileDaRt
Evolution require continued
operation during power mains
interruptions, it is
recommended that the
MobileDaRt Evolution should
be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE
•
8-6
UT is the a.c. mains voltage prior to application of the test level.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.1 Environmental Conditions of EMC (Electromagnetic Compatibility)
J Guidance and Manufacturer's Declaration - Electromagnetic Immunity
8
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MobileDaRt Evolution is intended for use in the electromagnetic environment specified below. The
customer or the user of the MobileDaRt Evolution should assure that it is used in such an environment.
Immunity Test
EN/
IEC 60601
Test Level
Compliance
Level
8
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the MobileDaRt
Evolution, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
EN 61000-4-6/
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to
80 MHz
d = 1.2 √P
Radiated RF
EN 61000-4-3/
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to
2.5 GHz
d = 1.2 √P 80 MHz to 800 MHz
8
8
8
8
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
8
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey*1, should be less than
the compliance level in each frequency range*2.
Interference may occur in the vicinity of equipment
marked with the following symbol:
8
8
8
NOTE
•
At 80 MHz and 800 MHz, the higher frequency range applies.
•
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
*1: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the MobileDaRt Evolution is used exceeds the
applicable RF compliance level above, the MobileDaRt Evolution should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
MobileDaRt Evolution.
8
8
*2: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-7
8 Specifications
J Recommended Separation Distances between Portable and Mobile
RF Communications Equipment and the EQUIPMENT or SYSTEM
Recommended Separation Distances between Portable
and Mobile RF Communications Equipment and the MobileDaRt Evolution
The MobileDaRt Evolution is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled.
The customer or the user of the MobileDaRt Evolution can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the MobileDaRt Evolution as recommended below, according to the maximum output power of the
communications equipment.
Separation Distance According to Frequency of Transmitter (m)
Rated Maximum
Output Power of
Transmitter
(W)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2 √P
d = 1.2 √P
d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
NOTE
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
8-8
•
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
•
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.2 Specifications
8.2
8
Specifications
Two types of the equipment are available, each with different column heights.
Type
8.2.1
Type S
1780 mm
Type T
1930 mm
8
Operating Environment
8
Item
8.2.2
8
Column Height
Specification
Atmosphere
No explosive or corrosive gases
Ambient Temperature
10 to 35 °C
Relative Humidity
30 to 75 % (with no condensation)
Atmospheric Pressure
800 hPa to 1060 hPa
Environment Luminosity
150 to 500 lx
8
8
8
Power Supply
Item
When Powered
by Battery
8
Specification
Power Supply
Internal battery
Battery Type
Sealed lead storage battery (12 V × 20 cells)
Supply Voltage
Single phase 100, 110, 120, 200, 220, 230, 240 V AC
Supply Frequency
50/60 Hz
8
8
Power Supply Rating 1 kVA
When Charging
Battery
Supply Impedance
Single phase 100, 110, 120 V AC: 1.0 Ω max.
Single phase 200, 220, 230, 240 V AC: 4.0 Ω max.
Ground Resistance
8
Ground terminal: 100 Ω max.
Additional ground terminal: 100 Ω max.
Power Cable Length
4m
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-9
8 Specifications
8.2.3
High-Voltage Generator
Item
Specification
Max. Electric Power
32 kW (100 kV, 320 mA, 20 ms / 80 kV, 400 mA, 20 ms)
16 kW (100 kV, 160 mA, 0.1 s)
Nominal Electric Power
(product of max current flow possible for 0.1 seconds and tube
voltage, at 100 kV nominal tube voltage)
Ratings
Nominal maximum tube voltage and
maximum tube current where nominal
maximum tube voltage output is possible:
133 kV, 200 mA
Maximum tube current and maximum tube
voltage where maximum tube current
output is possible:
400 mA, 80 kV
Combination of tube voltage and tube
current that outputs maximum electrical
power:
100 kV, 320 mA /
80 kV, 400 mA
Minimum current-time product:
0.32 mAs
Maximum current-time product:
320 mAs
Nominal minimum irradiation time: 1 msec
Current-Time Product
Settings Possible at
Given Cassette
Radiography Voltage*1
Current-Time Product
Settings Possible at
Given (Low Battery
mode) Cassette
Radiography Voltage*1
Current-Time Product
Settings Possible at
Given DR Long-Term
Exposure Mode
Voltage*1
8-10
Large
Focus
Small
Focus
40 to 90 kV:
0.32 to 320 mAs
101 to 110 kV: 0.32 to 250 mAs
0.32 to 280 mAs
111 to 120 kV: 0.32 to 220 mAs
121 to 133 kV: 0.32 to 200 mAs
40 to 90 kV:
0.32 to 320 mAs
101 to 110 kV: 0.32 to 250 mAs
91 to 100 kV:
0.32 to 280 mAs
111 to 120 kV: 0.32 to 220 mAs
121 to 133 kV: 0.32 to 200 mAs
Common
with
40 to 90 kV:
0.32 to 320 mAs
Large
101 to 110 kV: 0.32 to 250 mAs
Focus
and
121 to 133 kV: 0.32 to 200 mAs
Small
Focus
Large
Focus
91 to 100 kV:
91 to 100 kV:
0.32 to 280 mAs
111 to 120 kV: 0.32 to 220 mAs
40 to 65 kV:
0.32 to 320 mAs
66 to 80 kV:
0.32 to 280 mAs
81 to 85 kV:
0.32 to 250 mAs
86 to 100 kV:
0.32 to 220 mAs
101 to 105 kV: 0.32 to 200 mAs
106 to 125 kV: 0.32 to 180 mAs
126 to 133 kV: 0.32 to 140 mAs
Small
Focus
40 to 65 kV:
0.32 to 320 mAs
66 to 80 kV:
81 to 100 kV:
0.32 to 220 mAs
101 to 125 kV: 0.32 to 180 mAs
0.32 to 280 mAs
126 to 133 kV: 0.32 to 140 mAs
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.2 Specifications
Item
Current-Time Product
Settings Possible at
Given DR Radiography
Voltage*1
Current-Time Product
Settings Possible at
Given (Low Battery
mode) DR Long-Term
Exposure Mode
Voltage*1
Current-Time Product
Settings Possible at
Given (Low Battery
mode) DR Radiography
Voltage*1
Specification
Large
Focus
40 to 50 kV:
0.32 to 200 mAs
51 to 60 kV:
0.32 to 160 mAs
61 to 80 kV:
0.32 to 125 mAs
81 to 100 kV:
0.32 to 100 mAs
101 to 125 kV: 0.32 to 80 mAs
126 to 130 kV: 0.32 to 63 mAs
131 to 133 kV: 0.32 to 50 mAs
Small
Focus
Common
with
Large
Focus
and
Small
Focus
Large
Focus
40 to 60 kV:
0.32 to 160 mAs
61 to 80 kV:
81 to 100 kV:
0.32 to 100 mAs
101 to 125 kV: 0.32 to 80 mAs
8
0.32 to 125 mAs
8
126 to 130 kV: 0.32 to 63 mAs
131 to 133 kV: 0.32 to 50 mAs
40 to 65 kV:
0.32 to 320 mAs
66 to 80 kV:
81 to 100 kV:
0.32 to 220 mAs
101 to 120 kV: 0.32 to 180 mAs
8
8
0.32 to 280 mAs
121 to 133 kV: 0.32 to 140 mAs
40 to 50 kV:
0.32 to 200 mAs
51 to 55 kV:
0.32 to 160 mAs
56 to 65 kV:
0.32 to 125 mAs
66 to 80 kV:
0.32 to 100 mAs
81 to 100 kV:
0.32 to 80 mAs
101 to 120 kV: 0.32 to 63 mAs
121 to 133 kV: 0.32 to 50 mAs
Small
Focus
Tube Voltage Setting Range and
Display
Current-Time Product Setting Range
and Display*1
40 to 55 kV:
0.32 to 160 mAs
56 to 65 kV:
0.32 to 125 mAs
66 to 80 kV:
0.32 to 100 mAs
81 to 100 kV:
0.32 to 80 mAs
101 to 120 kV: 0.32 to 63 mAs
Display:
8
Digital
Setting Range: 0.32, 0.36, 0.40, 0.45, 0.50, 0.56, 0.63, 0.71, 0.80,
0.90, 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8, 3.2,
3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1, 8.0, 9.0, 10, 11, 12, 14,
16, 18, 20, 22, 25, 28, 32, 36, 40, 45, 50, 56, 63, 71,
80, 90, 100, 110, 125, 140, 160, 180, 200, 220, 250,
280, 320 mAs
Digital
A total of 144 types of anatomical programs can be registered, using 8
imaging portions, 3 directions, 3 body thicknesses, and 2 radiography
mode (8 × 3 × 3 × 2 = 144).
The following radiography parameters can be registered in anatomical
programs.
•
8
121 to 133 kV: 0.32 to 50 mAs
Setting Range: 40 kV to 133 kV, in 1 kV increments
Display:
Anatomical Programs
8
8
8
8
X-ray tube voltage, X-ray current-time product and focal spot size
Up to 216 anatomical programs only when second FPD kit
installation.
8
(8 regions × 3 orientations × 3 body thicknesses × 3 = 216)
*1:The various conditions are as follows (conform to IEC-standards):
Tube voltage (within ±10 %), Tube current (within ± 20 %)
mAs within ± (10 % + 0.2 mAs), Time within ± (10 % + 1 ms)
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-11
8 Specifications
8.2.4
X-ray Tube Unit
Item
Model
0.7/1.3U163C-36
Nominal Focal Spot Size
0.7/1.3 mm
Target Angle
16 degrees
Nominal Max.
Tube Voltage
X-ray Tube
Unit (tube and
housing)
X-ray Tube
(tube only)
8.2.5
Specification
Radiography
133 kV
Max. Heat Content
750 kJ (1060 kHU)
Max. Continuous
Heat Dissipation
Rate
120 W (170 HU/s)
Max. Anode Heat
Content
210 kJ (300 kHU)
Max. Anode Heat
Dissipation Rate
800 W (1130 HU/s)
Max. Continuous
Heat Dissipation
Rate
210 W (300 HU/s)
Surface Temperature on Touchable
Surface of X-ray Tube Unit
Maximum 60°C
(Compliant to IEC60601-2-28 for painted X-ray tube unit surface)
Mass
12.8 kg
Collimator
Item
Model
Field
Specification
R-20C
Shape
Rectangular
Max. Field
430 mm × 430 mm at SID 1 m
Min. Field
0 mm × 0 mm (leaves closed)
Average Illumination 160 lx min.
Illuminance Ratio
3 min.
Precision
2 % of SID
Center Indicator
Cross hairs
Type Of Lamp
Halogen lamp (JM12 V 100 W 10 H)
Illumination Period
30 seconds max., with automatic off timer
Illuminated Field
8-12
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.2 Specifications
Item
Field Size
Indicator
8.2.6
SID Display
1, 1.5, 2 m
Dimensions Display
20 (8), 23 (9), 25 (10), 28 (11), 30 (12), 36 (14), 43 (17) cm (in.)
Leaf Movement
Manual
Focal Length (distance between focal
spot and collimator attachment surface,
when installed)
56 mm
External Dimensions
202 mm × 211 mm × 200 mm
8
Specification
8
Type S: 600 mm to 1860 mm
Focal Spot Height
8
8
X-ray Tube Unit Holder
Item
8.2.7
8
Specification
Type T: 600 mm to 2010 mm
Arm Length
635 mm to 1200 mm
Column Rotation Range
±270 degrees
X-ray Tube Rotation Angle
±180 degrees
X-ray Tube Axial Rotation
Forward 90 degrees, backward 20 degrees
8
8
Minimum Inherent Filtration
Item
8
Specification
Total
2.5 mm Al at 70 kV
X-ray Tube Unit
1.5 mm Al equivalent at 70 kV
Collimator
1.0 mm Al equivalent at 70 kV
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-13
8 Specifications
8.2.8
DR System
Item
Specification
Application
General X-ray radiography
FPD
CXDI-50G/50C, CXDI-55G/55C, CXDI-60G/60C,
CXDI-70C Wireless/80C Wireless
Scintillator
Pixel Size
CXDI-50G/55G/60G : GOS
CXDI-50C/55C/60C/70C Wireless/80C Wireless : CsI
CXDI-50G/50C/55G/55C/60G/60C : 160 μm × 160 μm
CXDI-70C Wireless/80C Wireless : 125 μm × 125 μm
CXDI-50G/50C/55G/55C: 2208 × 2688 (about 5.9 megapixels)
Effective Number Of Pixels
CXDI-60G/60C : 1464 × 1776 (about 2.6 megapixels)
CXDI-70C Wireless : 2800 × 3408 (about 9.5 megapixels)
CXDI-80C Wireless : 2192 × 2800 (about 6.1 megapixels)
Automatic irradiation field recognition
Effective Irradiation Range
CXDI-50G/50C/55G/55C/70C Wireless : 35 cm × 43 cm max.
CXDI-60G/60C : 23 cm × 28 cm max.
CXDI-80C Wireless : 27.4 cm × 35 cm max.
DR
Dynamic Range
About 4-digit level
Gradation
12 bit (4096 gradations)
Max. Exposure Time
3000 msec
CXDI-50G/50C : 4.8 kg
CXDI-55G/55C : 3.4 kg (without cable)
Mass
CXDI-60G/60C : 2.5 kg (without cable)
CXDI-70C Wireless : 3.4 kg
CXDI-80C Wireless : 2.3 kg
Mechanical Strength
Partial load: 100 kg, given 40 mm diameter
Uniform load: 150 kg over all surfaces
CXDI-50G/50C : W491 mm × D477 mm × H23 mm
CXDI-55G/55C : W480 mm × D481 mm × H15 mm
External Dimensions
CXDI-60G/60C : W344 mm × D380 mm × H22.5 mm
CXDI-70C Wireless : W384 mm × D460 mm × H15 mm
CXDI-80C Wireless : W307 mm × D460 mm × H15 mm
Display
Unit
Long View Radiography
Unavailable
CXDI Sensor Port
100BASE-TX (Different to the hardware requirements described
in the Canon CXDI Wireless User's Manual.)
Size
15 inches
Wireless communications are based on the IEEE 802.11 n LAN standard (2.4 GHz/5 GHz).
^ Reference For further details, refer to the Canon CXDI manual.
8-14
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.2 Specifications
The wireless FPD can also be connected in a wired configuration using the cable provided in the FPD
wired connection kit.
8
^ Reference "5.13 Wired Connection Kit (for the CXDI-70C Wireless/80C Wireless)" P.5-28
8
CXDI-70C Wireless/80C Wireless
Item
Frequency Range
2412-2472 MHz
IEEE 802.11 n (HT20)
5180-5240 MHz
5745-5825 MHz
2422-2462 MHz
IEEE 802.11 n (HT40)
5190-5230 MHz
5755-5795 MHz
Modulation
OFDM (Orthogonal
Frequency Division
Multiplexing)
OFDM (Orthogonal
Frequency Division
Multiplexing)
Output Watts
8
10.3 dBm
16.8 dBm
16.8 dBm
8
9.3 dBm
16.8 dBm
16.8 dBm
8
Access Point
Item
Frequency Range
2412-2472 MHz
IEEE 802.11 n (HT20)
5180-5240 MHz
5745-5825 MHz
2422-2462 MHz
IEEE 802.11 n (HT40)
5190-5230 MHz
5755-5795 MHz
Modulation
OFDM (Orthogonal
Frequency Division
Multiplexing)
OFDM (Orthogonal
Frequency Division
Multiplexing)
Output Watts
8
20 dBm
20 dBm
8
20 dBm
20 dBm
20 dBm
8
20 dBm
The maximum power level depends on the channels and regulations specific to each region or country.
8.2.9
8
Options
Item
Specification
8
Using the infrared remote controller allows performing
radiography and switching the collimator lamp ON or OFF
remotely.
Remote Controller
Using the grid and grid frames, the more clear image is acquired
with reducing the influence of scattered X-ray.
M503-E028
Grid Unit and Grid Frame
The grid unit is available for the imaging unit CXDI-50G/50C/
55G/55C/60G/60C, and also the grid frame is for the imaging unit
CXDI-70C Wireless/80C Wireless.
Protective Screen (folding)
Operator exposure can be reduced by using the protective screen.
Dose Area Product Meter
Indicates the dose-area product (cannot be installed together with
the dose calculation unit or the distance indicator).
MobileDaRt Evolution OPERATION MANUAL
8-15
8
8
8
8 Specifications
Item
Specification
Dose Calculation Unit
Calculates and displays the radiation dosage of the X-ray applied
to surface of the patient (cannot be installed together with dose
area product meter).
Distance Indicator
Shows the distance measured by the dose calculation unit
(cannot be installed together with dose area product meter).
Keyless Entry
The system can be managed using a password instead of a key
switch.
Changing the Grip Bar Height
The grip bar height can be changed to 950 mm, 970 mm, or 1000
mm from the floor surface. Select the desired height during
installation (default bar height is 910 mm).
Luminous Hand Switch
The hand switch is illuminated to indicate when the radiography
preparation is finished, or that radiography is in progress.
Additional Hand Switch
Allows adding an illuminated hand switch to front of system.
Wireless LAN
With connecting the commercial wireless LAN adaptor, the
wireless LAN function to the network in the hospital is available.
Second FPD kit
Using the second FPD kit, the CXDI-55G/55C/60G/60C are
allowed to attached/detached two sensor units by means of the
sensor cable relay connector. The CXDI-70C Wireless/80C
Wireless are allowed to change over by touch screen operations.
FPD Wired Connection Kit
(CXDI-70C Wireless/80C Wireless)
Allows the CXDI-70C Wireless/80C Wireless to be used for wired
connection.
Handle Unit (CXDI-70C Wireless/80C
Wireless)
Attachment to the imaging unit improves the portability of the
imaging unit.
External Monitor Interface
This interface allows external monitor connection. Clone image of
the touch panel will be displayed on the external monitor.
Barcode reader
The barcode reader allows the study order associated with the read
"Patient ID" or "Accession Number" to be automatically selected
from the study list obtained from the MWM worklist. (Only the
patient ID can be read when using CXDI Control Software RD.)
FPD bag
This bag allows the imaging unit to slide smoothly.
Decoration label
Decoration label for exterior body
8.2.10 Custom Specifications
Item
Focal Spot Lift
(only for Type S)
Specification
Lifts focal spot up by 150 mm (focal spot height 750 mm to
2010 mm).
Note: Custom specifications cannot be changed after delivery.
8-16
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.3 Dimensional Drawings
8.3
8.3.1
Dimensional Drawings
8
General Dimensions
8
Item
Specification
Total Width
580 mm
Total Length
1220 mm
8
8
Type S: 1780 mm
Height of Column
Type T: 1930 mm
Mass
420 kg
Maximum Travel Speed
Approx. 5 km/h (may vary depending on condition)
8
8
8
8
8
∗Values shown in (
) are for Type S.
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-17
8 Specifications
8.4
X-Ray Tube Unit Reference Axis and Focal
Spot Position
X-ray Tube Unit Focal Spot Position (Top View)
Focal Spot∗1
Reference Axis
∗1: Focal spot is indicated by a mark on the X-ray tube unit surface.
Unit: mm
X-ray Tube Unit Focal Spot Position (Side View)
Focal Spot
Reference Axis
8-18
MobileDaRt Evolution OPERATION MANUAL
Unit: mm
M503-E028
8.5 Radiography Parameters
8.5
8.5.1
Radiography Parameters
8
Normal Use
8
J Large Focus
Tube
CurrentTime
Product
(mAs)
8
Tube Voltage (kV)
133
125
100
80
60
40
0.32
0.0016
0.0012
0.0010
0.0010
0.0010
0.0012
0.5
0.0025
0.0020
0.0016
0.0012
0.0016
0.0020
1
0.0050
0.0040
0.0032
0.0025
0.0032
0.0040
2
0.0100
0.0080
0.0063
0.0050
0.0063
0.0080
5
0.040
0.020
0.016
0.012
0.016
0.020
8
0.080
0.063
0.040
0.020
0.025
0.032
10
0.100
0.080
0.063
0.040
0.032
0.040
16
0.320
0.200
0.100
0.080
0.063
0.063
20
0.400
0.250
0.200
0.100
0.080
0.080
50
1.00
0.63
0.50
0.40
0.32
0.25
100
2.00
1.25
1.00
0.80
0.63
0.50
200
4.00
3.20
2.50
2.00
1.25
1.00
320
-
-
-
3.20
2.50
1.60
8
(Unit: sec)
8
8
8
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-19
8 Specifications
J Small Focus
Tube
CurrentTime
Product
(mAs)
Tube Voltage (kV)
133
125
100
80
60
40
0.32
0.0032
0.0025
0.0020
0.0016
0.0016
0.0020
0.5
0.0050
0.0040
0.0032
0.0025
0.0025
0.0032
1
0.0100
0.0080
0.0063
0.0050
0.0050
0.0063
2
0.0200
0.0160
0.0120
0.0100
0.0100
0.0120
5
0.063
0.050
0.040
0.032
0.025
0.032
8
0.100
0.100
0.063
0.050
0.050
0.050
10
0.125
0.125
0.080
0.063
0.063
0.063
16
0.320
0.200
0.125
0.100
0.100
0.100
20
0.400
0.250
0.200
0.125
0.125
0.125
50
1.00
0.63
0.50
0.40
0.32
0.32
100
2.00
1.60
1.00
0.80
0.63
0.63
200
4.00
3.20
2.50
2.00
1.25
1.25
320
-
-
-
3.20
2.50
2.00
(Unit: sec)
8-20
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.5 Radiography Parameters
8.5.2
Low-Battery Mode
8
J Large Focus
Tube
CurrentTime
Product
(mAs)
8
Tube Voltage (kV)
133
125
100
80
60
40
0.32
0.0016
0.0012
0.0010
0.0010
0.0010
0.0012
0.5
0.0025
0.0020
0.0016
0.0012
0.0016
0.0020
1
0.0050
0.0040
0.0032
0.0025
0.0032
0.0040
2
0.0100
0.0080
0.0063
0.0050
0.0063
0.0080
5
0.040
0.020
0.016
0.012
0.016
0.020
8
0.160
0.080
0.040
0.020
0.025
0.032
10
0.200
0.160
0.080
0.040
0.032
0.040
16
0.320
0.320
0.200
0.160
0.100
0.063
20
0.400
0.400
0.250
0.200
0.160
0.080
50
1.00
1.00
0.63
0.50
0.40
0.25
100
2.00
2.00
1.25
1.00
0.80
0.50
200
4.00
4.00
2.50
2.00
1.60
1.00
320
-
-
-
3.20
2.50
1.60
8
(Unit: sec)
8
8
8
8
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-21
8 Specifications
J Small Focus
Tube
CurrentTime
Product
(mAs)
Tube Voltage (kV)
133
125
100
80
60
40
0.32
0.0032
0.0025
0.0020
0.0016
0.0016
0.0020
0.5
0.0050
0.0040
0.0032
0.0025
0.0025
0.0032
1
0.0100
0.0080
0.0063
0.0050
0.0050
0.0063
2
0.0200
0.0160
0.0120
0.0100
0.0100
0.0120
5
0.100
0.100
0.063
0.040
0.025
0.032
8
0.160
0.160
0.100
0.080
0.063
0.050
10
0.200
0.200
0.125
0.100
0.080
0.063
16
0.320
0.320
0.200
0.160
0.125
0.100
20
0.400
0.400
0.250
0.200
0.160
0.125
50
1.00
1.00
0.63
0.50
0.40
0.32
100
2.00
2.00
1.25
1.00
0.80
0.63
200
4.00
4.00
2.50
2.00
1.60
1.25
320
-
-
-
3.20
2.50
2.00
(Unit: sec)
8-22
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.6 Anatomical Program Number
8.6
8.6.1
Anatomical Program Number
8
Anatomical Program Number
8
The factory-set anatomical program numbers (APR numbers) and their radiography parameters are
indicated below.
APR for Chest and leg are set 32 kw electric power, using the large focus, to minimize image blur
from body movement. Exposure time may be extended and tube current decreases in some
radiography settings when choosing the small focus.
Electric power* of High mAs settings like APR for Abdomen and Waist have no difference between
the large and small focus.
8
8
APR Numbers 1 to 72 are number of CXDI-50G/50C/55G/55C's APR Number. APR Numbers 73 to
144 are number of CXDI-60G/60C's APR Number.
If APR Number 1 to 72 selected,
illuminated.
FPD/Cassette Radiography Mode Button's Right LED is
If APR Number 73 to 144 selected,
blinked.
8
FPD/Cassette Radiography Mode Button's Right LED is
8
* Electric power mean kV times mAs.
CXDI-50G/55G
M503-E028
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
1
Head
Frontal
Thin
64
14
Small
2
Head
Frontal
Normal
64
14
Small
3
Head
Frontal
Thick
66
16
Small
4
Head
Lateral
Thin
56
14
Small
5
Head
Lateral
Normal
58
14
Small
6
Head
Lateral
Thick
61
14
Small
7
Head
Oblique
Thin
56
18
Small
8
Head
Oblique
Normal
59
18
Small
9
Head
Oblique
Thick
62
18
Small
10
Chest
Frontal
Thin
79
2.8
Large
11
Chest
Frontal
Normal
79
3.2
Large
12
Chest
Frontal
Thick
82
3.2
Large
13
Chest
Lateral
Thin
88
3.2
Large
14
Chest
Lateral
Normal
91
3.6
Large
15
Chest
Lateral
Thick
95
4
Large
16
Chest
Oblique
Thin
87
3.2
Large
MobileDaRt Evolution OPERATION MANUAL
8-23
8
8
8
8
8
8
8
8 Specifications
8-24
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
17
Chest
Oblique
Normal
90
3.2
Large
18
Chest
Oblique
Thick
93
3.2
Large
19
Abdomen
Frontal
Thin
74
56
Small
20
Abdomen
Frontal
Normal
77
63
Small
21
Abdomen
Frontal
Thick
80
71
Small
22
Abdomen
Lateral
Thin
86
63
Small
23
Abdomen
Lateral
Normal
90
63
Small
24
Abdomen
Lateral
Thick
95
63
Small
25
Abdomen
Oblique
Thin
78
80
Small
26
Abdomen
Oblique
Normal
83
80
Small
27
Abdomen
Oblique
Thick
89
80
Small
28
Waist
Frontal
Thin
68
22
Small
29
Waist
Frontal
Normal
68
25
Small
30
Waist
Frontal
Thick
71
25
Small
31
Waist
Lateral
Thin
82
22
Small
32
Waist
Lateral
Normal
84
25
Small
33
Waist
Lateral
Thick
86
28
Small
34
Waist
Oblique
Thin
71
22
Small
35
Waist
Oblique
Normal
74
25
Small
36
Waist
Oblique
Thick
76
28
Small
37
Leg
Frontal
Thin
63
7.1
Large
38
Leg
Frontal
Normal
63
8
Large
39
Leg
Frontal
Thick
66
8
Large
40
Leg
Lateral
Thin
65
7.1
Large
41
Leg
Lateral
Normal
68
7.1
Large
42
Leg
Lateral
Thick
71
7.1
Large
43
Leg
Oblique
Thin
65
8
Large
44
Leg
Oblique
Normal
65
9
Large
45
Leg
Oblique
Thick
65
10
Large
46
Foot
Frontal
Thin
58
2.2
Small
47
Foot
Frontal
Normal
58
2.5
Small
48
Foot
Frontal
Thick
61
2.5
Small
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.6 Anatomical Program Number
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
49
Foot
Lateral
Thin
57
2
Small
50
Foot
Lateral
Normal
58
2
Small
51
Foot
Lateral
Thick
61
2
Small
52
Foot
Oblique
Thin
55
2.5
Small
53
Foot
Oblique
Normal
58
2.5
Small
54
Foot
Oblique
Thick
61
2.8
Small
55
Arm
Frontal
Thin
58
2.8
Small
56
Arm
Frontal
Normal
58
3.2
Small
57
Arm
Frontal
Thick
61
3.2
Small
58
Arm
Lateral
Thin
61
2.8
Small
59
Arm
Lateral
Normal
63
3.2
Small
60
Arm
Lateral
Thick
65
3.6
Small
61
Arm
Oblique
Thin
60
3.2
Small
62
Arm
Oblique
Normal
60
3.6
Small
63
Arm
Oblique
Thick
60
4
Small
64
Hand
Frontal
Thin
53
0.9
Small
65
Hand
Frontal
Normal
53
1
Small
66
Hand
Frontal
Thick
56
1
Small
67
Hand
Lateral
Thin
56
0.9
Small
68
Hand
Lateral
Normal
58
0.9
Small
69
Hand
Lateral
Thick
60
0.9
Small
70
Hand
Oblique
Thin
56
1
Small
71
Hand
Oblique
Normal
57
1
Small
72
Hand
Oblique
Thick
57
1.1
Small
CXDI-50C/55C/70C Wireless
M503-E028
8
8
8
8
8
8
8
8
8
8
8
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
1
Head
Frontal
Thin
64
7.1
Small
2
Head
Frontal
Normal
64
8
Small
3
Head
Frontal
Thick
66
8
Small
4
Head
Lateral
Thin
56
7.1
Small
MobileDaRt Evolution OPERATION MANUAL
8-25
8
8
8 Specifications
8-26
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
5
Head
Lateral
Normal
58
7.1
Small
6
Head
Lateral
Thick
61
7.1
Small
7
Head
Oblique
Thin
56
9
Small
8
Head
Oblique
Normal
59
9
Small
9
Head
Oblique
Thick
62
9
Small
10
Chest
Frontal
Thin
79
1.4
Large
11
Chest
Frontal
Normal
79
1.6
Large
12
Chest
Frontal
Thick
82
1.6
Large
13
Chest
Lateral
Thin
88
1.6
Large
14
Chest
Lateral
Normal
91
1.8
Large
15
Chest
Lateral
Thick
95
2
Large
16
Chest
Oblique
Thin
87
1.6
Large
17
Chest
Oblique
Normal
90
1.6
Large
18
Chest
Oblique
Thick
93
1.6
Large
19
Abdomen
Frontal
Thin
74
28
Small
20
Abdomen
Frontal
Normal
77
32
Small
21
Abdomen
Frontal
Thick
80
36
Small
22
Abdomen
Lateral
Thin
86
32
Small
23
Abdomen
Lateral
Normal
90
32
Small
24
Abdomen
Lateral
Thick
95
32
Small
25
Abdomen
Oblique
Thin
78
40
Small
26
Abdomen
Oblique
Normal
83
40
Small
27
Abdomen
Oblique
Thick
89
40
Small
28
Waist
Frontal
Thin
68
11
Small
29
Waist
Frontal
Normal
68
12
Small
30
Waist
Frontal
Thick
71
12
Small
31
Waist
Lateral
Thin
82
11
Small
32
Waist
Lateral
Normal
84
12
Small
33
Waist
Lateral
Thick
86
14
Small
34
Waist
Oblique
Thin
71
11
Small
35
Waist
Oblique
Normal
74
12
Small
36
Waist
Oblique
Thick
76
14
Small
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.6 Anatomical Program Number
M503-E028
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
37
Leg
Frontal
Thin
63
3.6
Large
38
Leg
Frontal
Normal
63
4
Large
39
Leg
Frontal
Thick
66
4
Large
40
Leg
Lateral
Thin
65
3.6
Large
41
Leg
Lateral
Normal
68
3.6
Large
42
Leg
Lateral
Thick
71
3.6
Large
43
Leg
Oblique
Thin
65
4
Large
44
Leg
Oblique
Normal
65
4.5
Large
45
Leg
Oblique
Thick
65
5
Large
46
Foot
Frontal
Thin
58
1.1
Small
47
Foot
Frontal
Normal
58
1.2
Small
48
Foot
Frontal
Thick
61
1.2
Small
49
Foot
Lateral
Thin
57
1
Small
50
Foot
Lateral
Normal
58
1
Small
51
Foot
Lateral
Thick
61
1
Small
52
Foot
Oblique
Thin
55
1.2
Small
53
Foot
Oblique
Normal
58
1.2
Small
54
Foot
Oblique
Thick
61
1.4
Small
55
Arm
Frontal
Thin
58
1.4
Small
56
Arm
Frontal
Normal
58
1.6
Small
57
Arm
Frontal
Thick
61
1.6
Small
58
Arm
Lateral
Thin
61
1.4
Small
59
Arm
Lateral
Normal
63
1.6
Small
60
Arm
Lateral
Thick
65
1.8
Small
61
Arm
Oblique
Thin
60
1.6
Small
62
Arm
Oblique
Normal
60
1.8
Small
63
Arm
Oblique
Thick
60
2
Small
64
Hand
Frontal
Thin
53
0.45
Small
65
Hand
Frontal
Normal
53
0.5
Small
66
Hand
Frontal
Thick
56
0.5
Small
67
Hand
Lateral
Thin
56
0.45
Small
68
Hand
Lateral
Normal
58
0.45
Small
MobileDaRt Evolution OPERATION MANUAL
8-27
8
8
8
8
8
8
8
8
8
8
8
8
8
8 Specifications
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
69
Hand
Lateral
Thick
60
0.45
Small
70
Hand
Oblique
Thin
56
0.5
Small
71
Hand
Oblique
Normal
57
0.5
Small
72
Hand
Oblique
Thick
57
0.56
Small
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
73
Head
Frontal
Thin
64
14
Small
74
Head
Frontal
Normal
64
16
Small
75
Head
Frontal
Thick
66
16
Small
76
Head
Lateral
Thin
56
14
Small
77
Head
Lateral
Normal
58
14
Small
78
Head
Lateral
Thick
61
14
Small
79
Head
Oblique
Thin
56
18
Small
80
Head
Oblique
Normal
59
18
Small
81
Head
Oblique
Thick
62
18
Small
82*1
Chest
52
2.2
Large
83*1
Chest
55
2.5
Large
84*1
Chest
57
2.8
Large
85*1
Chest
59
2.8
Large
86*1
Chest
60
2.8
Large
87
Chest
Lateral
Thick
95
4
Large
88
Chest
Oblique
Thin
87
3.2
Large
89
Chest
Oblique
Normal
90
3.2
Large
90
Chest
Oblique
Thick
93
3.2
Large
91*1
Abdomen
50
2.2
Small
92*1
Abdomen
52
2.5
Small
CXDI-60G
8-28
Infant
(Frontal/Thin)
1 and 2-year-old
(Frontal/Normal)
3 to 6-year-old
(Frontal/Thick)
7 to 9-year-old
(Lateral/Thin)
10 to 12-year-old
(Lateral/Normal)
Infant
(Frontal/Thin)
1 and 2-year-old
(Frontal/Normal)
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.6 Anatomical Program Number
M503-E028
APR
Number
Imaging
Region
Projection
Direction
93*1
Abdomen
94*1
Abdomen
95*1
Abdomen
96
Abdomen
Lateral
97
Abdomen
98
Body
Thickness
kV
mAs
Focus Size
55
2.8
Small
58
3.2
Small
58
4
Small
Thick
95
63
Small
Oblique
Thin
78
80
Small
Abdomen
Oblique
Normal
83
90
Small
99
Abdomen
Oblique
Thick
89
100
Small
100
Waist
Frontal
Thin
68
22
Small
101
Waist
Frontal
Normal
68
25
Small
102
Waist
Frontal
Thick
71
25
Small
103
Waist
Lateral
Thin
82
22
Small
104
Waist
Lateral
Normal
84
25
Small
105
Waist
Lateral
Thick
86
28
Small
106
Waist
Oblique
Thin
71
22
Small
107
Waist
Oblique
Normal
74
25
Small
108
Waist
Oblique
Thick
76
28
Small
109
Leg
Frontal
Thin
63
7.1
Large
110
Leg
Frontal
Normal
63
8
Large
111
Leg
Frontal
Thick
66
8
Large
112
Leg
Lateral
Thin
65
7.1
Large
113
Leg
Lateral
Normal
68
7.1
Large
114
Leg
Lateral
Thick
71
7.1
Large
115
Leg
Oblique
Thin
65
8
Large
116
Leg
Oblique
Normal
65
9
Large
117
Leg
Oblique
Thick
65
10
Large
118
Foot
Frontal
Thin
58
2.2
Small
119
Foot
Frontal
Normal
58
2.2
Small
120
Foot
Frontal
Thick
61
2.5
Small
121
Foot
Lateral
Thin
57
2
Small
122
Foot
Lateral
Normal
58
2
Small
3 to 6-year-old
(Frontal/Thick)
7 to 9-year-old
(Lateral/Thin)
10 to 12-year-old
(Lateral/Normal)
MobileDaRt Evolution OPERATION MANUAL
8-29
8
8
8
8
8
8
8
8
8
8
8
8
8
8 Specifications
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
123
Foot
Lateral
Thick
61
2
Small
124
Foot
Oblique
Thin
55
2.5
Small
125
Foot
Oblique
Normal
58
2.5
Small
126
Foot
Oblique
Thick
61
2.8
Small
127
Arm
Frontal
Thin
58
2.8
Small
128
Arm
Frontal
Normal
58
3.2
Small
129
Arm
Frontal
Thick
61
3.2
Small
130
Arm
Lateral
Thin
61
2.8
Small
131
Arm
Lateral
Normal
63
3.2
Small
132
Arm
Lateral
Thick
65
3.6
Small
133
Arm
Oblique
Thin
60
3.2
Small
134
Arm
Oblique
Normal
60
3.6
Small
135
Arm
Oblique
Thick
60
4
Small
136
Hand
Frontal
Thin
53
0.9
Small
137
Hand
Frontal
Normal
53
1
Small
138
Hand
Frontal
Thick
56
1
Small
139
Hand
Lateral
Thin
56
0.9
Small
140
Hand
Lateral
Normal
58
0.9
Small
141
Hand
Lateral
Thick
60
0.9
Small
142
Hand
Oblique
Thin
56
1
Small
143
Hand
Oblique
Normal
57
1
Small
144
Hand
Oblique
Thick
57
1.1
Small
*1: The imaging region, projection direction and body thickness displayed on the DR system are not linked to each LED's
illumination status on the X-ray control panel.
CXDI-60C/80C Wireless
8-30
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
73
Head
Frontal
Thin
64
7.1
Small
74
Head
Frontal
Normal
64
8
Small
75
Head
Frontal
Thick
66
8
Small
76
Head
Lateral
Thin
56
7.1
Small
77
Head
Lateral
Normal
58
7.1
Small
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.6 Anatomical Program Number
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
78
Head
Lateral
Thick
61
7.1
Small
79
Head
Oblique
Thin
56
9
Small
80
Head
Oblique
Normal
59
9
Small
81
Head
Oblique
Thick
62
9
Small
82*1
Chest
52
1.1
Large
83*1
Chest
55
1.2
Large
57
1.4
Large
59
1.4
Large
60
1.4
Large
*1
M503-E028
Infant
(Frontal/Thin)
1 and 2-year-old
(Frontal/Normal)
3 to 6-year-old
84
Chest
85*1
Chest
86*1
Chest
87
Chest
Lateral
Thick
95
2
Large
88
Chest
Oblique
Thin
87
1.6
Large
89
Chest
Oblique
Normal
90
1.6
Large
90
Chest
Oblique
Thick
93
1.6
Large
91*1
Abdomen
50
1.1
Small
92*1
Abdomen
52
1.2
Small
93*1
Abdomen
55
1.4
Small
94*1
Abdomen
58
1.6
Small
95*1
Abdomen
58
2
Small
96
Abdomen
Lateral
Thick
95
32
Small
97
Abdomen
Oblique
Thin
78
40
Small
98
Abdomen
Oblique
Normal
83
45
Small
99
Abdomen
Oblique
Thick
89
50
Small
100
Waist
Frontal
Thin
68
11
Small
101
Waist
Frontal
Normal
68
12
Small
102
Waist
Frontal
Thick
71
12
Small
(Frontal/Thick)
7 to 9-year-old
(Lateral/Thin)
10 to 12-year-old
(Lateral/Normal)
Infant
(Frontal/Thin)
1 and 2-year-old
(Frontal/Normal)
3 to 6-year-old
(Frontal/Thick)
7 to 9-year-old
(Lateral/Thin)
10 to 12-year-old
(Lateral/Normal)
MobileDaRt Evolution OPERATION MANUAL
8-31
8
8
8
8
8
8
8
8
8
8
8
8
8
8 Specifications
8-32
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
103
Waist
Lateral
Thin
82
11
Small
104
Waist
Lateral
Normal
84
12
Small
105
Waist
Lateral
Thick
86
14
Small
106
Waist
Oblique
Thin
71
11
Small
107
Waist
Oblique
Normal
74
12
Small
108
Waist
Oblique
Thick
76
14
Small
109
Leg
Frontal
Thin
63
3.6
Large
110
Leg
Frontal
Normal
63
4
Large
111
Leg
Frontal
Thick
66
4
Large
112
Leg
Lateral
Thin
65
3.6
Large
113
Leg
Lateral
Normal
68
3.6
Large
114
Leg
Lateral
Thick
71
3.6
Large
115
Leg
Oblique
Thin
65
4
Large
116
Leg
Oblique
Normal
65
4.5
Large
117
Leg
Oblique
Thick
65
5
Large
118
Foot
Frontal
Thin
58
1.1
Small
119
Foot
Frontal
Normal
58
1.1
Small
120
Foot
Frontal
Thick
61
1.2
Small
121
Foot
Lateral
Thin
57
1
Small
122
Foot
Lateral
Normal
58
1
Small
123
Foot
Lateral
Thick
61
1
Small
124
Foot
Oblique
Thin
55
1.2
Small
125
Foot
Oblique
Normal
58
1.2
Small
126
Foot
Oblique
Thick
61
1.4
Small
127
Arm
Frontal
Thin
58
1.4
Small
128
Arm
Frontal
Normal
58
1.6
Small
129
Arm
Frontal
Thick
61
1.6
Small
130
Arm
Lateral
Thin
61
1.4
Small
131
Arm
Lateral
Normal
63
1.6
Small
132
Arm
Lateral
Thick
65
1.8
Small
133
Arm
Oblique
Thin
60
1.6
Small
134
Arm
Oblique
Normal
60
1.8
Small
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.6 Anatomical Program Number
APR
Number
Imaging
Region
Projection
Direction
Body
Thickness
kV
mAs
Focus Size
135
Arm
Oblique
Thick
60
2
Small
136
Hand
Frontal
Thin
53
0.45
Small
137
Hand
Frontal
Normal
53
0.5
Small
138
Hand
Frontal
Thick
56
0.5
Small
139
Hand
Lateral
Thin
56
0.45
Small
140
Hand
Lateral
Normal
58
0.45
Small
141
Hand
Lateral
Thick
60
0.45
Small
142
Hand
Oblique
Thin
56
0.5
Small
143
Hand
Oblique
Normal
57
0.5
Small
144
Hand
Oblique
Thick
57
0.56
Small
8
8
8
8
8
*1: The imaging region, projection direction and body thickness displayed on the DR system are not linked to each LED's
illumination status on the X-ray control panel.
8
8
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-33
8 Specifications
8.7
Labels
Locations where the labels are attached are shown as follows.
6
eg
280i
e
1345bf
ij
a
h
7
e
d
c
9
J Name Plates
1Name plate of MobileDaRt Evolution
8-34
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.7 Labels
8
2Name plate of collimator R-20C
8
8
8
3Identification label of X-ray control unit MUX-200D-XC (UL Version Only)
8
8
8
4Identification label of high voltage transformer D150MH-40 (UL Version Only)
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-35
8 Specifications
J Warning Labels
5Caution Label A
8Caution Label D
9Caution Label E
0Warning Label A
6Caution Label B
aWarning Label B (UL Version Only)
7Caution Label C
bCaution Label of Breaker
cCaution Label
8-36
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.7 Labels
J Indication Labels
dSmall Movement Switch Label
8
eBrake Release Switch Label
8
fAdditional Earthing Wire Label
8
gBrake Release Switch Label
8
8
hRF Transmitter Label
8
J Other Labels
8
iUL Label ASSY (UL Version Only)
8
8
jCE Label ASSY (CE Version Only)
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-37
8 Specifications
8.7.1
Labeling For X-Ray Tube Unit
5
2
4
16
17
3
J Name Plates
1Name plate of X-ray tube assembly
Name plate of X-ray tube assembly ( For UL)
8-38
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.7 Labels
J Warning Labels
2Warning Label (High Voltage)
8
3Label around a radiation gate
8
8
J Indication Labels
4Indication Label "+" (PLUS)
8
5Indication Label "-" (MINUS)
8
J Other Labels
6UL Label ASSY
8
8
8
7CE Label ASSY
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-39
8 Specifications
8.8
8.8.1
Statement of Compliance [For Europe]
Regulatory Information
For Europe:
The system complies with the requirement of the Medical Device Directive 93/42/EEC
Equipment Name:
MOBILE X-RAY SYSTEM
Model Name:
MOBILE X-RAY SYSTEM MobileDaRt Evolution
Parts Number:
503-77777-XX
Manufacturer:
SHIMADZU CORPORATION
Medical Systems Division
Address:
1, NISHINOKYO-KUWABARACHO,
NAKAGYO-KU, KYOTO, 604-8511, JAPAN
Authorized
SHIMADZU EUROPA GmbH
Representative in EU:
Address:
8.8.2
Albert-Hahn-Strasse 6-10, 47269 Duisburg, F.R. Germany
Company's Quality System
The company's Quality System is satisfied with Annex II, Article 3 for 93/42/EEC as amended by
2007/47/EC, which is certified by TUV Rheinland LGA Products GmbH; Tillystrasse 2, D-90431
Nurnberg, Germany (Notified under No. 0197) as Registration No.: HD 60029841 0001
8-40
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.8 Statement of Compliance [For Europe]
8.8.3
International Standards
8
This system conforms the following international standards.
•
IEC 60601-1:2005 / EN 60601-1:2006
•
IEC 60601-1-2:2007 / EN 60601-1-2:2007
•
IEC 60601-1-3:2008 / EN 60601-1-3:2008
•
IEC 60601-1-6:2010 / EN 60601-1-6:2010
•
IEC 60601-2-54:2009 / EN 60601-2-54:2009
•
IEC 60627:2001 / EN 60627:2001
•
ISO 10993-1:2009 / EN ISO 10993-1:2009
•
ISO 14971:2007, Corrected version / EN ISO 14971:2009
•
IEC 62220-1:2003 / EN 62220-1:2004
•
IEC 62304:2006 / EN 62304:2006
•
IEC 62366:2007 / EN 62366:2008
•
EN 980:2008
•
EN 1041:2008
8
8
8
8
8
8
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-41
8 Specifications
8.9
8-42
Statement of Compliance with Standards
•
X-RAY EQUIPMENT for RADIOGRAPHY ... MOBILE X-RAY SYSTEM MobileDaRt
Evolution IEC 60601-2-54:2009
•
X-RAY EQUIPMENT for RADIOGRAPHY ... MOBILE X-RAY SYSTEM MobileArt
Evolution IEC 60601-2-54:2009
•
X-RAY EQUIPMENT for RADIOGRAPHY ... MOBILE X-RAY SYSTEM MobileDaRt
Evolution EN 60601-2-54:2009
•
X-RAY EQUIPMENT for RADIOGRAPHY ... MOBILE X-RAY SYSTEM MobileArt
Evolution EN 60601-2-54:2009
MobileDaRt Evolution OPERATION MANUAL
M503-E028
8.10 Manufacturer Information
8.10Manufacturer Information
Manufacturer:
SHIMADZU CORPORATION
Medical Systems Division
Address:
1, NISHINOKYO-KUWABARACHO,
NAKAGYO-KU, KYOTO, 604-8511, JAPAN
8
8
8
8
8
8
8
8
8
8
8
8
8
M503-E028
MobileDaRt Evolution OPERATION MANUAL
8-43
8 Specifications
This page is intentionally left blank.
8-44
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Appendix
9
Appendix
Check List for Trouble
Fill following blanks and contact your Shimadzu service representative.
Hospital :
Phone :
FAX :
Serial Number:
Date of Installation:
Item
Check
Someone injured
When did the trouble occur?
Is there a sign of the trouble?
Is there a power failure or thunderbolt when the trouble
occurred?
The device has been watered?
The device has been given an impact?
The number of patients a day with the device?
When is the last periodic inspection?
App-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Appendix
9
Safety Check Registry
9
Name
Phone
9
Extension number
Hospital
FAX
Address
9
Room number
Name
Equipment
Serial number
9
Date of installation
Affiliation
Instructor
Phone
9
FAX
Date
Description
This device's
maintenance and
operation staff
Instructor
Signature
Signature
9
9
9
9
9
•
•
M503-E028
At the time of installation or when changing the designated operator, your Shimadzu service
representative will explain the safety instructions described in this manual to the maintenance
and operation staff, and record the information.
This manual is important. We request all maintenance personnel to keep it in a safe place
close to the device.
MobileDaRt Evolution OPERATION MANUAL
App-3
9
9
Appendix
This page is intentionally left blank.
App-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Index
Index
A
Column ................................................................... 2-2
Acc......................................................................... 4-5
Acceleration setting.................................................. 4-2
Additional hand switch ........................................... 5-21
Additional protective grounding wire....................... 2-26
Additional radiography function .............................. 3-34
All Free button ............................................. 2-10, 2-13
Anatomical code .................................................... 4-14
Anatomical program............................... 3-21, 1-3, 4-14
Anatomical program number.......................... 4-14, 8-23
APO ....................................................................... 4-6
APR ..............................................1-3, 3-21, 4-8, 8-23
aPr.......................................................................... 4-8
APR settings at startup ............................................. 4-2
CPU backup battery.................................................. 6-6
Current-time product...............................................3-21
Current-time product buttons ..............................2-7, 2-8
CXDI-50C .............................................................2-16
CXDI-50G.............................................................2-16
CXDI-55C .............................................................2-16
CXDI-55G.............................................................2-16
CXDI-60C .............................................................2-16
CXDI-60G.............................................................2-16
CXDI-70C Wireless................................................ 2-20
CXDI-80C Wireless................................................ 2-20
D
Apron hanger......................................................... 2-14
DICOM image header ............................................. 5-12
Arm ............................................................... 2-2, 2-13
Disposable wipes holder.......................................... 2-15
Arm joint ....................................................... 2-2, 2-13
Distance indicator................................................... 5-16
Arm lock release button.......................................... 2-13
Dose area product meter.......................................... 5-10
Arm lock release buttons ........................................ 2-10
Dose calculation unit .............................................. 5-14
Arm lock release lever............................................ 2-13
DR button ................................................................ 2-4
at OFF .................................................................... 3-5
DR long-term exposure button ...................................2-6
Auto-power-OFF function ........................................ 3-5
DR long-term exposure mode .................................. 4-15
Auto-power-off time setting...................................... 4-2
DR radiography mode............................................. 3-21
DR system ....................................................3-21, 3-25
B
bAC........................................................................ 4-6
E
Backward power-assist setting................................... 4-2
Emergency brake release battery ................................6-6
Battery.................................................................. 3-35
Emergency brake release button..........................2-3, 6-5
Battery indicator ...................................................... 2-4
Emergency stop button.....................................2-4, 3-13
Body thickness selection buttons ........................ 2-7, 2-9
External connection LAN connector.........................2-27
bon ......................................................................... 4-7
Brake release bar ..................................................... 2-3
Buzzer setting.......................................................... 4-2
F
Focus mode button ................................................... 2-6
FPD bag ................................................................5-37
C
FPD/Cassette radiography mode button ...................... 2-6
Cassette radiography mode ............................ 3-21, 3-22
Changing the grip bar height ................................... 5-19
Charging ............................................................... 3-36
Charging lamp ......................................................... 2-4
Check List for Trouble ............................................. 9-2
Collimator...................................................... 2-2, 2-10
Collimator lamp button....................2-6, 2-8, 2-10, 4-13
Front slot ................................... 2-23, 2-22, 5-26, 5-27
G
Grid unit ..................................................................5-6
Grip bar ............................................................2-2, 2-3
Grounding cable .......................................................7-5
Collimator lamp control setting ................................. 4-2
M503-E028
MobileDaRt Evolution OPERATION MANUAL
Index-1
Index
H
P
Hand switch.................................................... 2-2, 2-12
PACS ....................................................................3-29
PAS.........................................................................4-4
I
Password................................................................5-17
Imaging region selection buttons......................... 2-7, 2-9
Imaging unit ................................................... 2-2, 2-16
Inch-mover buttons........................................ 2-10, 3-11
Irradiation field ...................................................... 3-19
Irradiation field adjustment knobs ............................ 2-10
Periodic replacement parts .......................................7-10
Personal ID setting....................................................4-2
Pid...........................................................................4-4
Positioning .............................................................3-17
Power cable............................................................2-24
Power line connection indicator lamp .........................2-4
Power supply panel ........................................... 2-2, 2-4
K
Power-assist setting...................................................4-2
Key switch............................................................... 2-4
Projection direction selection buttons.................. 2-7, 2-9
Keyless entry ......................................................... 5-17
Protective screen.......................................................5-8
kV display ........................................................ 2-7, 2-8
L
R
Radiography parameters ................................ 3-21, 8-19
LAP ........................................................................ 4-7
Radiography preparation button ...............................2-12
Large focus............................................................ 8-19
Radiography ready indicator lamp ...................... 2-6, 2-8
Locking the arm ..................................................... 3-15
READY .................................................................3-24
Low-battery mode .................................................. 3-33
Rear slot .................................... 2-23, 2-22, 5-26, 5-27
Low-speed travel mode........................................... 3-10
Registration button............................................ 2-7, 2-8
Luminous hand switch ............................................ 5-20
Releasing the brake in emergency...............................6-5
Remaining battery charge ..........................................3-2
M
Remote control receiver.............................................5-3
Main circuit breaker ........................................ 2-2, 2-25
Malfunction indicator lamp .........................2-6, 2-8, 6-4
mAs ...................................................................... 3-21
mAs display...................................................... 2-7, 2-8
Middle slot ................................................... 5-26, 5-27
Motor...................................................................... 6-6
Remote controller .....................................................5-2
Remote controller button ................................... 2-6, 2-8
Remote switch ........................................................2-19
Replacement time display ..........................................6-6
S
Safety check registry .................................................9-3
N
Second FPD kit.......................................................5-24
Normal travel mode................................................ 3-10
Notification function................................................. 6-6
Numeric Delete Button ........................................... 5-17
Numeric keypad ..................................................... 5-17
Sensor cable relay connector ....................................2-19
Small focus ............................................................8-20
Snd..........................................................................4-9
sound emitted during system travel.............................4-9
sound emitted during X-ray exposure..........................4-9
sound emitted when radiography preparation
O
is finished ................................................................4-9
Obstruction detector ............................................... 3-12
Standby indicator lamp...................................... 2-6, 2-8
Obstruction detector canceling function.................... 3-12
Status indicator lamp......................................... 1-3, 2-4
Operation sound ..................................................... 4-12
stc ...........................................................................4-5
Operation sound setting ............................................ 4-2
Steer control setting ..................................................4-2
Storage case ................................................... 2-2, 2-22
Index-2
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Index
1
Storage case for extra battery pack........................... 2-23
1
T
Tape measure ............................................... 2-10, 3-19
Top plate................................................................. 7-3
1
Touch panel monitor ......................................... 2-2, 2-5
Transmission status indicator .................................... 5-2
Travelling warning sound .................................. 2-6, 2-8
Tube voltage.......................................................... 3-21
1
Tube voltage buttons ......................................... 2-7, 2-8
U
USB connector ........................................................ 2-4
1
User settings button........................................... 2-7, 2-8
User settings mode ................................................... 4-3
U01 ........................................................................ 6-6
U02 ........................................................................ 6-6
1
U03 ........................................................................ 6-6
U04 ........................................................................ 6-6
U05 ........................................................................ 6-6
1
W
Warm-up................................................................. 3-6
Warning indicator lamp .............................. 2-6, 2-8, 6-2
1
Wire rope ................................................................ 7-5
Wireless LAN option ............................................. 5-22
X
1
X-ray control panel ........................................... 2-2, 2-6
X-ray exposure button ............................................ 2-12
X-ray irradiation indicator lamp ......................... 2-6, 2-8
X-ray tube unit .......................................... 2-2, 3-6, 6-6
1
1
1
1
M503-E028
MobileDaRt Evolution OPERATION MANUAL
Index-3
Index
This page is intentionally left blank.
Index-4
MobileDaRt Evolution OPERATION MANUAL
M503-E028
Download