Uploaded by Gjhhyj

scribd.vdownloaders.com 3m-steri-vac-8xl-service-manual

advertisement
3M™ Steri-Vac™
8XL Gas Sterilizer/Aerator
Service and Repair Manual
Health Care
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TABLE OF CONTENTS
Safety Notices ...........................................................................................................................
1
Introduction
Steri-Vac 8XL Sterilizer Modes of Operation ............................................................
Methods of Selecting Modes and Options ..................................................................
Steri-Vac 8XL Sterilizer Modes Menu Structure .......................................................
Printed Sterilization Reports .......................................................................................
4
4
5
6
Normal Operating Mode .............................................................................................
Service Mode ..............................................................................................................
Site Setup ....................................................................................................................
Modifying Site Setup ............................................................................................
Date Format, Time Format, Date, and Time .........................................................
Daylight Savings Time (DST) ..............................................................................
Owner Name .........................................................................................................
User Codes (Option) .............................................................................................
Printer Type ..........................................................................................................
Remote ..................................................................................................................
Language ................................................................................................................
Critical Options .....................................................................................................
9
11
13
14
15
17
18
19
20
21
22
23
Setup
Checkout
Service ........................................................................................................................
Calibration ............................................................................................................
Temperature Calibration (Automatic) ..................................................................
Temperature Calibration (Individually Selected) .................................................
Pressure Calibration ..............................................................................................
RH Calibration ......................................................................................................
Vacuum Test ...............................................................................................................
Switch Test .................................................................................................................
Video Test ...................................................................................................................
Print Sterilizer Setup and Trend Information .............................................................
25
26
27
28
29
30
31
35
36
38
Demo Mode ................................................................................................................
41
Demo
© 3M 1995 – 2005
i
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TABLE OF CONTENTS
Troubleshooting and Repair
Help Window ..............................................................................................................
Solenoid Test ..............................................................................................................
DIP Switch Selectable Service Mode Options ...........................................................
Final Vacuum ........................................................................................................
Emergency Aeration .............................................................................................
System Reset Mode
Reset Cycle to Idle ................................................................................................
Clear Silent Abort .................................................................................................
Reset Site Setup ....................................................................................................
Error Code Flow Description ......................................................................................
Error Code Listing ......................................................................................................
Error Code Failure Analysis .......................................................................................
43
44
47
47
47
48
49
50
51
52
56
Special Cycles
Emergency Aeration (from Normal Mode) ................................................................
Accelerate Cycle Time ................................................................................................
Simulate Low Pressure ...............................................................................................
Define Cycles ...............................................................................................................
Temperature, Preheat Option ......................................................................................
Vacuum, Humidity .......................................................................................................
Exposure, Aeration .....................................................................................................
85
86
87
88
89
90
91
Parts Listing
Parts Diagrams and Listings .......................................................................................
93
Schematics
5 Volt Power Supply ...................................................................................................
24 Volt Power Supply .................................................................................................
Overall Machine Schematic ........................................................................................
Controller ....................................................................................................................
MCU Circuit ...............................................................................................................
Process Control Output Circuit ...................................................................................
Keypad Video Output .................................................................................................
Serial Comm, DIP Switches, Printer Connection Circuit ...........................................
Process Control Input Circuit ......................................................................................
Keypad Display Interface Board .................................................................................
Sterilizer Operational Flow Diagram ..........................................................................
ii
147
148
149
151
153
155
157
159
161
163
165
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SAFETY NOTICES
Please read and understand all instructions before using the Steri-Vac sterilizer/aerator.
DANGER
The Steri-Vac sterilizer/aerator uses ethylene oxide gas to sterilize heat-and/or
moisture-sensitive devices. Ethylene oxide is flammable and toxic. Follow all
instructions and precautions carefully.
Toxicity
First Aid
Acute Inhalation
May be fatal if
inhaled in high
concentrations.
May cause
irritation of
respiratory tract,
chest tightness,
headache, nausea, vomiting, diarrhea,
light-headed feeling, dizziness,
weakness, drowsiness, cyanosis, loss
of coordination, convulsions, coma,
delay lung injury (fluid in the lungs),
immediate or delayed skin irritation or
blisters, allergic skin reaction.
If Inhaled
Move exposed person to
fresh air. Keep warm. If
person is not breathing,
call 911 or an
ambulance, then give
artificial respiration,
preferably mouth-tomouth if possible. If
breathing is difficult, give
oxygen. Call a poison
control center or doctor
immediately for further
treatment advice. Call a
physician even if no
symptoms are present.
Keep under medical
observation. Symptoms
may be delayed.
Chronic Inhalation
The Occupational Safety and Health
Administration (OSHA) classifies
ethylene oxide (EO) as a probable
human carcinogen and reproductive
hazard.
Toxicity
First Aid
Skin
Contact
Liquid
ethylene
oxide may
cause skin
irritation,
dermatitis,
and blistering.
If On Skin Or Clothing
Immediately flush skin with
plenty of water for 15 – 20
minutes while removing
contaminated clothing and
shoes. Call a poison control
center or doctor immediately for
treatment advice. Aerate, wash,
or clean contaminated clothing,
and discard leather goods.
Toxicity
First Aid
Toxicity
First Aid
Eye Contact
Splashes of
ethylene oxide
may cause
severe eye injury.
High gas
concentrations
may cause
severe eye irritation and injury.
If In Eyes
Hold eye open and rinse
slowly and gently with
water for 15 – 20
minutes. Remove contact
lenses, if present, after
the first 5 minutes, then
continue rinsing eye. Call
a poison control center or
doctor for treatment
advice.
Ingestion
A highly
unlikely
route of
exposure.
Liquid
ethylene
oxide,
upon ingestion, is caustic and
may cause severe irritation and
burns to the gastrointestinal
mucosa.
If Swallowed
Call a poison control center or
doctor immediately for
treatment advice. Have person
sip a glass of water if able to
swallow. Do not induce vomiting
unless told to do so by a poison
control center or doctor. Do not
give anything by mouth to an
unconscious person.
Flammability
Ethylene oxide is flammable in concentrations
from 3% (30,000 ppm) to 100%. Keep all
sources of ignition such as matches, lighted
cigarettes, sparks, and static discharge away
from the sterilizer and cartridges.
© 3M 1995 – 2005
Ethylene Oxide Leaks Or Spills
The following indicate Steri-Gas cartridge leakage:
•
Liquid ethylene oxide spurting or rapidly dripping from a
cartridge
•
A cartridge that feels very cold to the touch
•
Cartridge weight loss
1
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SAFETY NOTICES
3M™ Steri-Gas™ EO Cartridge 8-170 Label
2
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SAFETY NOTICES
WARNING
Hazardous Voltage
Use care when setting the DIP switches with power applied to
the Steri-Vac 8XL Sterilizer. This machine contains HIGH
VOLTAGES. Failure to use care can result in electrical shock or
injury to personnel, including death.
CAUTION
Make sure static discharge wrist ground straps are worn when
setting configuration switches or when working with the Steri-Vac
8XL Sterilizer covers removed. Static discharge will cause damage
to equipment. Failure to wear static discharge straps can result in
damage to equipment or injury to personnel.
© 3M 1995 – 2005
3
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
INTRODUCTION
Steri-Vac 8XL Sterilizer Modes of Operation
Methods of Selecting Modes and Options
The Steri-Vac 8XL Sterilizer provides four modes of
operation.
There are two methods of selecting Steri-Vac 8XL Sterilizer
modes and options. The modes are selected by setting DIP
switches, specifically SW1 switches 7 and 8. The options
available in the modes are either selected with a softkey or
by configuring SW1 switches 1-6. Softkey selections are
evident when described in the procedures. DIP switch
settings are described and a picture is provided to show the
switch positions. Figure 1 shows the relationship between
the positions of the configuration switches on SW1 or SW2
and the configuration tables used in the procedures. The
arrows correspond to the following: ↑ Closed, ↓ Open.
Figure 2 shows the Steri-Vac 8XL Sterilizer Controller Card
configuration switch, test point, and connector locations.
Normal Mode - The sterilizer is operated in Normal Mode
when performing routine sterilization operations. The
procedure on page 9 provides instructions on placing the
sterilizer in Normal Mode. For detailed instructions on
operating the sterilizer in Normal Mode see the Operator’s
Manual.
Service Mode - The Service Mode is used to:
•
Set up the sterilizer when it is initially installed, or
when there is a change to the setup information.
Critical options are set during setup.
•
Check out the sterilizer when it is initially installed or
if a problem is encountered during operation. Checkout
provides instructions for the Calibration of temperature,
pressure, and RH sensors; vacuum test, switch test, and
video test; and printing setup and trend information.
•
Troubleshoot various sterilizer problems by running
selected stages of a sterilization cycle and printing trend
reports for analyzing failures. Additionally,
troubleshooting includes error analysis and error
clearing procedures.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Service Mode
Figure 1. Example of DIP Switch Positions and
Configuration Table Listing.
Detailed Service procedures start on page 11.
Demo Mode - The Sterilizer Demo Mode provides a
graphic simulation of a sterilization cycle. This mode can
be used for familiarization or training purposes. See page
41 for instructions on entering the demo mode.
SW1
SW2
System Reset Mode - The System Reset Mode has three
options and is only available for use by a qualified service
technician. Refer to page 48 for a detailed description of
the System Reset Mode options.
Note: A service representative access code is encoded in
the software to prevent unauthorized access to some
options. The access code is provided to the service
representative upon request.
Figure 2. Steri-Vac 8XL Sterilizer Controller Card
Configuration Switch, Test Point, and Connector Layout.
4
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
INTRODUCTION
Steri-Vac 8XL Sterilizer Modes Menu
Structure
Modes (SW1 DIP switch selectable)
Normal
Service
See
Operator’s
Manual
Demo
SETUP
DATE/TIME
DST
EMERGENCY AERATION
OWNERS NAME
PRINT INFO (Setup and Trend)
USER CODES
SERVICE
Steri-Vac
performs a
graphic
simulation of
a sterilization
cycle.
System Reset
RESET CYCLE TO IDLE
CLEAR SILENT ABORT
RESET SITE SETUP
SET CYCLES
PRINTER
REMOTE
LANGUAGE
CALIBRATE
SENSOR CAL.
TEMPS
PRESSURE
RH
VACUUM TEST
DEADEND TEST
PUMP TEST
LEAK TEST (Low pressure)
LEAK TEST (High Pressure)
SWITCH TEST
SOLENOID TEST
Video test *
Final Vacuum *
Emergency Aeration *
Clear EEPROM *
© 3M 1995 - 2005
* SW1 DIP Switch Selectable
5
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
INTRODUCTION
Printed Sterilization Reports
The Steri-Vac 8XL Sterilizer can be configured with an
installed printer, an external printer, or no printer. When a
printer is connected to the sterilizer, the printer is used for
printing a graphical representation of each sterilization
cycle, for printing all of the errors encountered during the
cycle, and for printing setup and trend information.
sterilization cycle, the error is printed at the place in the
sterilization report where the error was encountered. If
report printing is interrupted because of an out-of-paper
condition, the printing will continue once new paper is
loaded and the printer initializes.
An attached printer is initialized when the sterilizer is
powered on.
Figure 3 is an example of a printed sterilization report with
a cycle temperature of 37ºC. Figure 3 also indicates the
stages of the sterilization cycle and labels the signals
printed on a report.
The sterilization cycle report is printed from the time the
cycle begins, until the Locked Aeration stage begins (see
Figure 3). If the error is discovered during the
Note: An error or caution message is only printed once.
Figure 3. Sterilization Cycle Report with a Cycle Temperature of 37ºC.
6
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
INTRODUCTION
© 3M 1995 - 2005
7
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
INTRODUCTION
This page intentionally left blank.
8
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Normal Operating Mode
Normal sterilization cycles are performed when SW1
switches 7 and 8 are configured (open) for Normal Mode
as shown in Figure 4. When the Steri-Vac 8XL Sterilizer
is powered up with SW1 configured for Normal Mode
(SW1 switches 7 and 8 open), switches 1-6 can be set in
any of the following combinations to give the options
listed below and shown in Figure 5. These options are
selected when performing various service actions.
Note: If SW1 switches 7 and 8 are not in Normal
operating mode, the function of switches 1-6 change
significantly. See Service Mode on page 11.
Selectable options for Normal Mode are:
Simulate Low
Pressure
Software ignores most pump downs
(see page 87).
Accelerate Time
Accelerates some cycle times by 120
(see page 86).
Note: Screen times are not
accelerated.
Help Window
Displays a help window containing
system information. This screen
shows temperature sensor values,
door status, handle status, timer, etc.
Half Gas Expose
Cycle
Shortens gas expose cycle for testing.
Air Injection
During
Precondition
Forces air into chamber during
humidification. This option is
designed for specialized testing and
not intended as a normal service tool.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Normal Mode
Figure 4. Normal Operating Mode.
1 2 3 4 5 6 7 8 SW1 DIP Switch #'s
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ NORMAL MODE
↑
↓ ↓ Simulate Low Pressure
↑
↓ ↓ Accelerate Time
↑
↓ ↓ Help Window
↑
↓ ↓ Half Gas Cycle
↑
↓ ↓ Air Injection
↑ ↓ ↓ Full Strip Chart
Figure 5. Normal Operating Mode Options.
Note: This option has been disabled.
Full Strip Chart
© 3M 1995 - 2005
Plots heater sensors (end, middle,
and heatsink), calibration
temperature (47.5°), and RH
(throughout precondition), in
addition to normally plotted
parameters.
9
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
To select the Normal Mode of Operation:
1. Power Off the Steri-Vac 8XL Sterilizer.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
4.
5.
6.
7.
Set SW1 configuration switches 7 and 8 to the open
position.
Set SW1 configuration switches 1 - 6 for the desired
options as shown in Figure 5 on page 9.
Set Critical Options (see page 23).
Power On the Steri-Vac 8XL Sterilizer.
On initial installation or when the sterilizer is powered
On with User Access Codes enabled the display in
Figure 6 is shown and a User Access Code has to be
entered to operate the sterilizer. After entering the
User Access Code the display in Figure 7 is shown.
Operate the sterilizer in Normal Mode as outlined in
the Operator’s Manual.
Figure 6. Normal Operating Mode Display with User
Access Codes Option Enabled.
Note: On initial installation or after the system defaults
the Supervisor’s Setup Code is 1234.
Figure 7. Normal Operating Mode Display after User
Access Codes Entered.
10
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Service Mode
Configuration and testing of the Steri-Vac 8XL Sterilizer
can be performed when SW1 switches 7 and 8 are
configured (closed) for Service Mode as shown in Figure
8. When the sterilizer is powered up with SW1 configured
for Service Mode (SW1 switches 1-6 open and switches 7
and 8 closed), the following sterilizer Setup selections are
available through displayed softkeys.
Site Setup
page 13
Vacuum Test
page 31
Switch Test
page 35
Solenoid Test
page 44
Other Service Mode options are available by configuring
SW1 switches 1-6 as show in Figure 9.
Final Vacuum
page 47
Emergency Aeration
page 47
Note: Emergency Aeration is also selectable
directly from Normal Mode by using the
procedure on page 85.
Trends
page 38
Video Test
page 36
To Select the Service Mode of Operation:
1. Power Off the Steri-Vac 8XL Sterilizer.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
4.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Service Mode
Figure 8. Service Mode.
1
↓
↓
↓
↓
↓
↑
2
↓
↓
↓
↓
↑
↓
3
↓
↓
↓
↓
↓
↓
4
↓
↓
↓
↑
↓
↓
5
↓
↓
↑
↓
↓
↓
6
↓
↑
↓
↓
↓
↓
7
↑
↑
↑
↑
↑
↑
8
↑
↑
↑
↑
↑
↑
SW1 DIP Switch #'s
Service Mode
Final Vacuum
Unused
Emergency Aeration
Trends
Video Test
Figure 9. Service Mode Options.
Set SW1 configuration switches 7 and 8 to the closed
position.
If one of the DIP switch selectable options is desired,
set SW1 configuration switches 1-6 for the desired
option as shown in Figure 9.
Set Critical Options (see page 23).
Figure 10. Service Menu Display.
© 3M 1995 - 2005
11
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
5.
6.
Power On the sterilizer.
If gas is in the chamber or the Gas Present LED on the
Controller circuit board is On (see Figure 2), the
Service mode is not entered and the message Press
Stop to Clear Abort is displayed, do one of the
following:
a.
b.
If you want to finish the current sterilization
cycle, power off the sterilizer, set SW1 switches
for Normal Mode, then power up the sterilizer
and it will complete the cycle.
If you want to end the current sterilization cycle,
press the Stop button. The sterilizer will
sequence through final vacuum, purge, and
mandatory aeration.
DANGER
Use care when setting the DIP switches
with power applied to the Steri-Vac 8XL
Sterilizer. This machine contains HIGH
VOLTAGES. Failure to observe use care
can result in electrical shock or injury to
personnel, including death.
7.
8.
12
The Service Menu displays. If options were selected
by setting SW1 configuration switches 1-6 in Step 3,
after the selected option starts, set those switches to
the open position as shown in Figure 8.
If no options were selected in Step 3, go to the
appropriate Service Mode procedure to continue.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Site Setup
The Site Setup option is menu driven. Four softkeys on the
right side of the display enable the service representative
to choose options from a selection list. When the desired
option in the selection list is highlighted, pressing Select
softkey either selects the option (or de-selects items
already selected) or displays a sub-menu of options
corresponding to the selection. Following is the list of
softkeys, which are used to perform Site Setup.
Up
Moves highlight up the selection list.
Down
Moves highlight down the selection list.
Select
Selects the current highlighted option on the
selection list. In some cases this will display
another sub-menu of options.
Done
Displays the previous menu.
Figure 11. Site Setup Display.
Note: A highlighted menu item is not selected unless the
Select softkey is pressed. If the Done softkey is pressed
before a highlighted menu item is selected, the display
returns to the previous menu and the highlighted item is
not selected.
The Site Setup option allows the service representative
to permanently setup the following options for the
Steri-Vac 8XL Sterilizer:
• Date Format, Time Format, Date, and Time
• Daylight Savings Time (DST)
• Aeration (Locked)
• Emergency Aeration
• Print Information
• User Codes Option
• Owner Name
• Service
Note: The modified options are permanently stored in the
controller when the Done softkey is pressed in the Site
Setup menu. If the sterilizer’s power is turned Off before
the Done softkey is pressed, an E2 error will occur the
next time the sterilizer’s power is turned On. All the Site
Setup information will be lost and have to be re-entered.
Site Setup Option
Default Value
Date Format
Time Format
Date
Time
Daylight Savings Time
(DST)
User Codes
Owner
Install Number
Aeration (Locked)
MM/DD/YY
23:59
August 4, 2003
00:00
OFF
Temp Cycles
Printer
Remote
Preheat
Punct. Press.
© 3M 1995 - 2005
OFF
<blank>
1
Read from Critical Options Dip
Switch, see Critical Options on
page 23.
37°C For 3 Hours
55°C For 1 Hour
Seiko
None
None
160 mBars
13
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
Modifying Site Setup
1.
2.
3.
Place the Steri-Vac 8XL Sterilizer in the Service
Mode (see page 11).
Press Site Setup softkey. The Site Setup menu is
displayed.
Perform the necessary operations from the following
procedures to enter the site setup information into the
sterilizer.
Figure 12. Service Menu Display.
Note: Make sure the following step is accomplished
before exiting Site Setup, otherwise all setup information
will be lost.
4. After all site information is entered, press the Done
softkey to store the information in the sterilizer’s
controller.
Figure 13. Site Setup Display.
14
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Date Format, Time Format, Date, and Time
1.
2.
3.
4.
5.
6.
Using the Up and Down softkeys, highlight the
DATE/TIME option on the Site Setup menu.
Press the Select softkey.
The Date Format menu is displayed. Using the Up and
Down softkeys, highlight the desired Date format.
Press the Select softkey.
The Time Format menu is displayed. Using the Up
and Down softkeys, highlight the desired Time
format.
Press the Select softkey.
© 3M 1995 - 2005
15
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
Date format, Time Format, Date, and Time Cont.
7.
Using the Month Up and Month Down softkeys, set
the current date.
Note: The day of the week is automatically updated each
time the date changes.
8. Press the Save Date softkey when current date is set.
9. Using the Hours Up softkey, set the current hour.
10. Using the Minute Up and Minute Down softkeys, set
the current minute.
11. Press the Save Time softkey when current time is set.
12. The display shows the Site Setup Menu.
Note: The date and time are displayed to the right of
DATE/TIME on the Main menu.
When the sterilization cycle is started and a printer is
available, the date and time (in their respective formats)
are printed on the final printout.
16
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Daylight Savings Time (DST)
This option adjusts the time for Daylight Savings Time
(DST). When DST is toggled from Off to On, one hour is
added to the sterilizer internal clock time. When DST is
toggled from On to Off, one hour is subtracted from the
sterilizer internal clock time.
1. Using the Up and Down softkeys, highlight the DST
option on the Site Setup menu.
2. Press the Select softkey to toggle DST from ON to
OFF or from OFF to ON.
© 3M 1995 - 2005
17
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
Owner Name
This option is for entering an owner (institutional,
corporation, supervisor) name.
1. Using the Up and Down softkeys, highlight the
OWNER NAME option on the Site Setup menu.
2. Press the Select softkey.
Note: In the following step the last (highlighted)
character is not included in the Name. Only the text that is
not highlighted is saved as the owner name. Up to two
lines of text can be entered.
3.
4.
5.
6.
7.
18
Using the Up and Down softkeys, select the first letter
of the name to be entered, then press the Select
softkey. Continue until the name is complete.
Pressing the Stop button erases the last letter entered.
When the name is completely entered, press the Done
softkey.
The display shows the OWNER name and
STERIVAC NUMBER. The sterilizer number can
now be selected. This number is used for multiple
Steri-Vac 8XL Sterilizer installations using a remote
display. All sterilizers connected to a single remote
display must have unique sterilizer installation
numbers. If this is a single installation without a
remote display, select NONE.
Using the Up and Down softkeys, change the
STERIVAC NUMBER.
Press the Select softkey.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
User Codes (Option)
If the User Codes option was purchased with this sterilizer,
the User Codes setup instructions are contained in
Steri-Vac 8XL Gas Sterilizer/Aerator, User Access Codes
Setup, Service Manual Supplement, a separate supplement
to the Service Manual. If you have purchased this option
and did not receive the Service Manual supplement,
contact your 3M representative.
© 3M 1995 - 2005
19
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
Printer Type
The Steri-Vac 8XL Sterilizer can be set up with an internal
or external printer, or no printer.
1. If the Misc. Service menu is not already displayed,
use the Up and Down softkeys to highlight the
SERVICE option on the Site Setup menu.
2. Press the Select softkey.
3.
4.
Using the Up and Down softkeys, highlight the
PRINTER option on the Misc. Service menu.
Press the Select softkey.
5.
Using the Up and Down softkeys, select the type of
printer connected to the sterilizer.
6. Press the Select softkey. The Misc. Service display
shows the selected printer.
Note: Do not select the Fujitsu printer type when a Seiko
printer is installed. This will cause the sterilizer electronics
to enter an unstable state.
20
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Remote
The Steri-Vac 8XL Sterilizer can be operated at the
display panel on the front of the unit or from a remote
display panel. Remote is used when one or more
sterilizers are connected to a remote control/display panel.
1. If the Misc. Service menu is not already displayed,
use the Up and Down softkeys to highlight the
SERVICE option on the Site Setup menu.
2. Press the Select softkey.
3.
4.
5.
6.
Using the Up and Down softkeys, highlight the
REMOTE option on the Misc. Service menu.
Press the Select softkey.
Using the Up and Down softkeys, highlight the
applicable remote display option.
Press the Select softkey. The Misc. Service display
shows the selected option.
© 3M 1995 - 2005
21
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SETUP
7. Press the Select softkey. The Language display shows
the selected options.
22
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SETUP
Critical Options
The AUX DIP switch (SW2) is a hardware setup for
system configuration (see Figure 14).
Note: The SW2 (Critical Options) switches must be set
before entering Normal Mode. If software failures occur,
this DIP switch corresponds to the system installation
settings that were active before the failure.
1
↑
2
4
↓
↓
↑
↓
↑
↓
↑
↑
5
6
• Abator Type
• Two Door System
The following are continuously monitored:
To set the AUX DIP switch SW2 switches:
1. Make sure the Steri-Vac 8XL Sterilizer is powered
Off.
2. Set SW2 configuration switches 1-6 to correspond to
the sterilizer configuration requirements (see Figure
15 for an example of switch settings).
3. Set SW2 configuration switch 8 for 3 hour locked
aeration time or no locked aeration time.
4. Set SW2 configuration switch 7 to enable or disable
the E31 Leak Test.
5. Set SW1 configuration switches for the desired mode
of operation or as directed in service or operational
procedures.
6. Power On the Steri-Vac 8XL Sterilizer.
8
↓
↓
↑
↓
↑
↓
↑
↑
Electric Pump
Access Codes
Nitrogen Injection
The following are read at Reset Site Setup
• Locked Aeration
7
↓
The following are read at power up:
•
•
•
3
SW2 DIP Switch #'s
Electric pump
Access Codes Disabled
No Abator Connected
50 CFM Abator
50 CFM Abator
w/Aeration
50 CFM Abator Lockout
Two Door System
Nitrogen Injection
No locked aeration
Locked Aeration 3 hours
Leak test off
Figure 14. Critical Options.
1
2
3
4
5
6
7
8
SW2 Dip Switch #s
Switch settings
Locked Aeration 3hrs
Leak test on
Nitrogen Injection
Single door system
50 CFM abator
Access codes disabled
Electric Pump
Figure 15. Example of selected Critical Options set on
SW2 DIP switch.
© 3M 1995 - 2005
23
Service and Repair
Steri-Vac Gas Sterilizer/Aerator Model 8XL
SETUP
This page intentionally left blank.
24
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
Service
Selecting Service on the Site Setup menu displays the
Misc. Service menu with the following options:
Option
When Selected
Set Cycles
Special cycle applications
Printer
Selected during Setup
Remote
Selected during Setup
Language
Selected during Setup
Calibrate
Performed during Checkout
To display the Misc. Service menu, do the following:
1. Using the Up and Down softkeys, highlight the
SERVICE option on the Site Setup menu.
2.
Press the Select softkey. The Misc. Service menu
displays.
© 3M 1995 - 2005
25
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
Calibration
The Calibration mode allows the permanent calibration of
the following sensors:
• Heatsink Temperature
• Display Temperature
• End Temperature
• Middle Temperature
• Chamber Pressure
• Chamber Relative Humidity
Note: The Master Access Code is required to gain access
to the Calibration menus.
1. If the Misc. Service menu is not already displayed, see
page 25 to display the Misc. Service menu.
2. Using the Up and Down softkeys, highlight the
CALIBRATE option on the Misc Service menu.
3. Press the Select softkey.
4.
Using the number softkeys, enter the Master Access
Code. After entering the correct Master Access Code
the Sensor Calibration menu appears with the
following selection list:
•
•
•
26
Temps
Pressure
RH
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
Temperature Calibration (Automatic)
Automatic Temperature Conditioning Circuit Calibration
(Auto Cal) should be performed first, then, if necessary,
the individual temperature sensors can be adjusted.
The temperature conditioning circuit contains a precision
resistor calibrated to 47.5°C. The software calibrates the
conditioning circuit (not each individual sensor).
The equations for determining the actual temperature are
adjusted according to the value read from the calibration
resistor.
1. Using the Up and Down softkeys, highlight TEMPS
on the Sensor Calibration menu.
2. Press the Select softkey.
3.
4.
Using the Up and Down softkeys, highlight the
AUTO CAL on the Temp Calibration menu.
Press the Select softkey. Auto calibration is
performed.
Notes: The Help window (page 43) can be viewed to
check calibration. Calibration (CAL) should be
47.5º ±.5º.
5.
Each temperature sensor can be calibrated
independently of the others. This is only required
if temperature problems occur, such as the middle
of the chamber is too hot or too cold as compared
with the end temperatures. Normally individual
calibration does not need to be performed.
If none of the temperature sensors require individual
calibration, press the Done softkey and the Sensor
Calibration menu is displayed. Otherwise continue
with the next procedure.
© 3M 1995 - 2005
27
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
Temperature Calibration (Individually
Selected)
Note: Make sure Automatic Temperature Calibration on
page 27 is performed before performing this procedure.
1. Using the Up and Down softkeys, highlight one of the
temperature sensors.
•
•
•
•
2.
Heatsink
Display
End
Middle
Press the Select softkey. A highlighted box is shown
at the top of the display with a message similar to:
Operate 0.0°C Cooler=xx.xxºC.
Note: In the following step, pressing the Up softkey will
increase the C reading by about .5°C. The display will
show Operate 0.5°C Warmer and the selected sensor will
operate hotter by 0.5°C. Pressing the Down softkey will
decrease the C reading by about .5°C. The display will
show Operate 0.5°C Cooler and the selected sensor will
operate cooler by 0.5°C. These changes are saved until an
Auto Cal is performed.
3. Using the Up and Down softkeys, set the sensor
calibration value as required.
4. Press the Select softkey to set the sensor calibration
value and return to the Temp Calibration menu.
5. Repeat steps 1-4 for additional temperature sensors
requiring individual calibration.
6. When temperature calibration is complete, press the
Done softkey. The Sensor Calibration menu is
displayed.
Note: If C12 occurs at the start of a cycle and calibration
is performed, only none-manually adjusted sensors are
calibrated.
28
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
Pressure Calibration
This calibration procedure calibrates the sterilizer pressure
sensor to ambient pressure with the sterilizer door open.
1. Using a calibrated pressure meter, measure the
ambient pressure near the sterilizer.
2.
If the Sensor Calibration menu is not already
displayed, see page 26 to display the menu.
3. Using the Up and Down softkeys, highlight
PRESSURE on the Sensor Calibration menu.
4. Make sure the sterilizer door is open.
Note: When performing pressure calibration the sterilizer
door must be open.
5.
Press the Select softkey. A highlighted box is
shown at the top of the display with a message
similar to: Operate 0.0 mBar Lower=xx.xx
(xx.xx = the current reading).
Note: In the following step, pressing the Up softkey will
increase the mBar reading by about 5 mBars. The display
will show Operate 5 mBar Higher and the selected sensor
will operate 5 mBar higher than normal. Pressing the
Down softkey will decrease the mBar reading by about 5
mBars. The display will show Operate 5 mBar Lower and
the selected sensor will operate 5 mBar lower than normal.
6. Using the Up and Down softkeys, set the sensor
pressure calibration value to approximately 6
mbars above the ambient pressure measured in
Step 1 above.
7. Press the Select softkey to set the pressure
calibration value. The Sensor Calibration menu is
displayed.
© 3M 1995 - 2005
29
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
RH Calibration
1.
2.
3.
If the Sensor Calibration menu is not already
displayed, see page 26 to display the menu.
Using the Up and Down softkeys, highlight RH on
the Sensor Calibration menu.
Press the Select softkey. The RH display allows the
HyCal sensor set points to be adjusted.
Note: The Steri-Vac 8XL Sterilizer should be supplied
with the HyCal Engineering Data Sheet. Near the bottom
of this supplied sheet two set points are given, 0% rh =
some voltage and 75.3% rh = some voltage. These
voltages are the set points that need to be entered for the
RH sensor.
The current set points are displayed on the screen. A small
circle (o) to the left of each set point indicates the set point
has been changed. A dash (-) indicates the set point has
not been changed. Humidity errors may occur during
normal operating mode if the values are not set.
4. Using the Up and Down softkeys, change the set
points to the required settings.
5. Press the Select softkey. After the last set point is
entered and selected, the Sensor Calibration menu is
displayed.
6. Press the Done softkey. The Misc. Service menu is
displayed.
30
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
Vacuum Test
Vacuum Test is selected from the main Service Menu and
tests the sterilizer vacuum system and chamber leakage.
This test is nearly automatic.
1. Place the Steri-Vac 8XL Sterilizer in the Service
Mode (see page 11).
2. On the Service menu press Vacuum Test softkey.
The initial vacuum test display is shown.
DANGER
Use care when setting the DIP switches
with power applied to the Steri-Vac 8XL
Sterilizer. This machine contains HIGH
VOLTAGES. Failure to observe use care
can result in electrical shock or injury to
personnel, including death.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
3.
4.
5.
Open the sterilizer door.
Press the Start Vacuum softkey.
Install the Dead End Vacuum tester
© 3M 1995 - 2005
31
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
6.
When the pressure reading stabilizes at the lowest
value, press the Stop Vacuum softkey.
Note: A Dead End pressure greater than 120 mBars
will cause a Vacuum Test failure.
7.
Remove the Dead End Vacuum tester.
Note: If the sterilizer door is already closed, the message
Close Door will not be displayed.
8. Make sure the sterilizer door is closed and press the
Start softkey.
32
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
9.
The pump and leak tests are automatic. The following
operations are performed during these tests.
Note: Pressing the Abort softkey will stop the
current test.
Pump Test
When the vacuum level is reached, Vacuum Test
automatically advances to Low Pressure Leak Test.
Note: A pump down time of more than 18 minutes
will cause a Vacuum Test failure.
Low Pressure Leak Test (10 minutes)
After 10 minutes the Vacuum Test automatically
advances to High Pressure Leak Test.
Notes: A Low Pressure Leak Test of more than 15
mBar positive pressure change causes a
Vacuum Test failure.
A negative reading indicates a decrease in
chamber pressure. A negative reading will
not cause a Vacuum Test failure.
© 3M 1995 - 2005
33
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
High Pressure Leak Test (10 minutes)
After 10 minutes the Vacuum Test displays the results
of the test, either Passed or Failed.
Note: More than 15 mBar positive pressure change
will cause a Vacuum Test failure.
10. Press the Print softkey. If a printer is installed,
connected, and setup (See Site Setup on page 13), the
Vacuum test final display (as shown to the right) will
be printed.
Note: The top right corner of the printout will show the
test results. If a fault occurred, the clock time on the report
for the test will be reversed (black with white numbers)
and the test results will show Failed. Otherwise the test
results will show Passed.
34
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
Switch Test
The Switch Test is used to verify selected Steri-Vac 8XL
Sterilizer switches change their state of operation when
actuated during this test.
1. Place the Steri-Vac 8XL Sterilizer in the Service
Mode (see page 11).
2. On the Service menu press Switch Test softkey. A
Switch Test menu is displayed listing all the sterilizer
switches that are testable.
DANGER
Use care when setting the DIP switches
with power applied to the Steri-Vac 8XL
Sterilizer. This machine contains HIGH
VOLTAGES. Failure to observe use care
can result in electrical shock or injury to
personnel, including death.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
3.
4.
To test the operation of a switch actuate the selected
switch and observe the Switch Test display. If the
switch is operating correctly, the switch name will be
highlighted on the display when the state of the switch
is changed.
To end the switch test, set SW1 configuration
switches to the Normal Operating Mode (all switches
Open, see page 9). This signals the software the
Switch Test is complete. At the end of the test a
printed copy of the Switch Test display is printed on
the sterilizer’s printer and the Service Menu appears.
Note: If other tests are to be performed, the SW1
configuration switches need to be set to the Service mode.
© 3M 1995 - 2005
Switch Test
TO QUIT, PLACE SW1 ALL OPEN
LD Handle
LD Latch
LD Open
UD Handle
UD Latch
UD Open
Comp. Air
Cartridge
Printer On
Puncture Pin
Air Flow
Water Low
Nitro Low
Abator Rdy.
Abator Fail
Stop
Key 1
Key 2
Key 3
Key 4
}
}
External
Signals
Front
Panel
Softkeys
Switch Key:
LD
UD
Comp Air
Cartridge
Load Door
Unload Door
Compressed Air
Gas cartridge position
35
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
Video Test
The Video test is a Service Mode operation to test the
ability of the Steri-Vac 8XL Sterilizer display to display
all pixels On, to display vertical lines, and to display
horizontal lines.
1. Power Off the Steri-Vac 8XL Sterilizer.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
4.
5.
6.
36
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Video Test
Figure 16. Video Test Switch Positions.
Set SW1 configuration switches 7 and 8 to the closed
position (see Figure 16).
Set SW1 configuration switch 1 closed and switches
2-6 open (see Figure 16).
Power On the Steri-Vac 8XL Sterilizer.
The first Video Test display shows with all pixels On.
Press any key to advance to the next screen.
The second Video Test display shows vertical lines.
Press any key to advance to the next screen.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
7.
8.
The third Video Test display shows horizontal and
vertical lines. Press any key to display the Service
Mode menu.
If additional Service Mode operations are to be
performed, make sure SW1 configuration switches are
placed in the correct position for the procedure to be
performed.
© 3M 1995 - 2005
37
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
Print Sterilizer Setup and Trend Information
This option prints critical options, setup/calibration
information, and Trend Analysis information. The
printout can be used to verify proper system configuration.
Note: This procedure should only be used after all other
Site Setup procedures are done. This will ensure all
printed information is complete for the current Site Setup.
1. Make sure the printer is selected and printer power is
On.
2. If the Steri-Vac 8XL Sterilizer is not in the Service
Mode, go to page 11 and place the sterilizer in the
Service Mode.
3. If the Site Setup menu is not displayed, press the Site
Setup softkey on the Service menu.
4.
5.
38
Using the Up and Down softkeys, highlight the PRT
INFO option on the Site Setup menu.
Press the Select softkey. The sterilizer printer prints
the critical options, setup/calibration, and trend
analysis information. See the following page for a
sample printout.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
CHECKOUT
Abbreviations used on the Setup and Trend Analysis
printout.
Cycle Number
CY
Cycle Temperature selected
CT
Temperature of chamber at start (C°)
ST
Pump down time (seconds)
PT
Chamber Leak (mBars)
CL
Note: A positive CL number indicates a
leak into the chamber.
Number of injections
NI
Final Humidity before puncture (%RH)
RH
Pressure at puncture (mBar)
PP
Final Vacuum pumpdown time (seconds) not
FV
displayed on screen.
Figure 17. Sterilizer Setup and Trend Information
Printout.
© 3M 1995 - 2005
39
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
CHECKOUT
This page intentionally left blank.
40
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
DEMO
Demo Mode
The Demo mode performs a graphic simulation of a
sterilization cycle. The Stop button operates as a pause
key during the Demo mode. The Demo takes about 3-4
minutes.
To place the Steri-Vac 8XL Sterilizer in the Demo
mode of operation:
1. Power Off the Steri-Vac 8XL Sterilizer.
DANGER
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Demo Mode
Figure 18. Demo Mode.
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
4.
Open SW1 configuration switches 1 - 6 as shown in
Figure 18.
Set SW1 configuration switches 7 and 8 to the
positions shown in Figure 18.
Power On the Steri-Vac 8XL Sterilizer.
Note: The only error code check during Demo mode is
for E14, Load Door Open.
5. The sterilizer performs a graphic simulation of a
sterilization cycle.
Note: Pressing the Stop button pauses the Demo mode.
To resume Demo mode press the Stop button again.
Figure 19. Normal Operating Mode Graphic Simulation
during Demo Mode.
© 3M 1995 - 2005
41
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
DEMO
This page intentionally left blank.
42
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
Help Window
The Help Window displays the current status of
Steri-Vac 8XL Sterilizer switches and settings
and can be accessed in any mode at any time.
1. Place SW1 switch 3 in the Closed position to display
the Help Window.
2. Place SW1 switch 3 in the Open position to hide the
Help Window.
RH L/RHH
LAST SW *
CAUTIONS
RH Voltage set points, Low and High, measured in
Volts. If RH Set point not entered, “NA” will be
displayed.
Shows numbers representing active switches, see
below for explanation of each switch.
Displays last four caution numbers, latest caution is
leftmost digit.
Displays last four error numbers, latest error is
leftmost digit.
Displays timers value in seconds (not always
TIMER
applicable).
Displays a E ON if the end heaters are on,
HEATERS
otherwise E OFF.
Displays a M ON if the middle heaters are on,
otherwise M OFF.
HEATSINK Displays ON if heatsink is being heated, otherwise
OFF.
STG START Display the time at which the current stage started.
STG CURR Displays the current time of this stage (for comparison
to STG START).
Abator Name Currently installed: NONE,
ABATOR
50CFMAER, 50CFM, 50CFMLOC. See Critical
Options (page 23).
50CFM: Displays YES/NA if abator is ready, and
AR/AF
NO/NA if abator is still warming.
Remote Display Status Help Window Section.
REMOTE
ERRORS
DISP
PACKET
STAT
Display packet status (OK, BAD)/RS232 Hardware
serial Status (OK, ERR.) Relates to Remote Display
Panel.
* Each number displayed after LAST SW in the Help
Window represents the following switch
CHBR
END
MID
AMB
CAL
HSK
RHT
VAC
AMB
REF
RH
CNT
STG
MODE
PUMPTIME
Chamber Temperature, measured in °C. Second
parameter is digital value as read by A/D Converter.
End Temperature, measured in °C.
Middle Temperature, measured in °C.
Ambient Temperature, measured in °C.
Calibration Resistor Temperature, measured in °C.
Heatsink Temperature, measured in °C.
N/A
Chamber Pressure, measured in mBar.
Ambient Pressure, measured in mBar.
Voltage Reference, measured in Volts.
Relative Humidity, measured in %RH.
Cycle Count.
Stage of current cycle, IDLE, START, PRECON,
GAS INJ, GAS EXP, FNL VAC, PURGE,
LOCKAER, UNLOCK.
Displays current operating mode, NORMAL,
RESET, SERVICE, or DEMO.
Electric Pump Time, Shows amount of time
remaining before the electric pump turns off (OFF
IN XXX SEC), or displays ON. Must be setup in
Critical Options.
© 3M 1995 - 2005
SW
Switch Name
Active State
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Softkey 1
Softkey 2
Softkey 3
Softkey 4
Stop Button
Load Door Closed
Load Door Handle Down
Load Door Latched
Printer off
Puncture Pin Up
Air Flow Failure
Water Low
Low Compressed Air
Cartridge Not Present
15
16
17
18
Latching Relay GAS
Unload Door Closed
Unload Door Handle Down
Unload Door Latched
Down
Down
Down
Down
Down
Closed
Closed
Closed
Off Position
Open
Closed
Open
Closed
Cartridge Plunger
out
Closed
Closed
Closed
Closed
Note:
Switches 16-18 must be set up in System Configuration
Dip Switch (SW2) Critical Options, Two Door System.
43
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
Solenoid Test
The Solenoid test allows the service representative to
activate/deactivate certain solenoids and heaters by DIP
switch control.
Note: This test is available as a system troubleshooting
tool only and is not intended as a substitute for other tests
described in this manual.
1. Place the Steri-Vac 8XL Sterilizer in the Service
Mode (see page 11).
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
1
2
3
4
5
6
7
8
Heatsink heaters
Vent solenoid
Door interlock
RH solenoid
Steam injection
Help window
Vacuum
Chamber heaters
When the Service menu appears, place all SW1 and
SW2 configuration switches to the CLOSED positions
(see Figure 20 and 21).
Note: To hide the Help screen, place SW1 switch 3 in the
Open position.
SW1 Dip Switch #s
Switch settings
Figure 20. SW1 Configuration Switch Positions for the
Solenoid Test.
1
2
3
4
5
6
7
8
SW2 Dip Switch #s
Switch settings
Not used
Report
Nitrogen injection
Printer test
Electric pump
Figure 21. SW2 Configuration Switch Positions for the
Solenoid Test.
44
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
3.
On the Service menu press Sol. Test softkey. The
Solenoid Test display is shown listing all the sterilizer
solenoids that are testable.
Note: Information shown on the left side of the display is
for hardware troubleshooting.
DANGER
Use care when setting the DIP switches
with power applied to the Steri-Vac 8XL
Sterilizer. This machine contains HIGH
VOLTAGES. Failure to observe use care
can result in electrical shock or injury to
personnel, including death.
4.
Close or open SW1 and SW2 configuration DIP
switches to test specific solenoids. The solenoids are
tested as shown in the tables (Figures 22 and 23) on
the following page.
Notes: The Printer Test option will automatically be
performed (see Figure 24).
Press the STOP button to exit this test.
5.
At the end of the Solenoid Test place SW1 and SW2
configuration switches back to their original
configuration (Service Mode and Critical Options).
© 3M 1995 - 2005
DIP SWITCHES
SW1=C0 HEX
SW2 = 03 HEX
CTL OUTPUTS
HEX
PC01 =0
PC02 = 0
HEX
PC03 = 0
HEX
CTL INPUTS
PCI1 = F7
HEX
PCI2 = 3F
HEX
PCI3 = FF
HEX
SW1 TEST OPTIONS
1 – HEATER=55C OFF
2 – VACUUM
OFF
3 – HELP SCREEN
4 – STEAM 1.0 SEC 0
5 – RH SENSOR OPEN
6 – DOORS UNLOCKED
7 – VENT
CLOSED
8 – HEATSINK=95C OFF
SW2 TEST OPTIONS
1 – ELEC PUMP
2 – PRINTER
3 – NITROGEN
OFF
OFF
4 – REPORT
45
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
SW #
Description
Purpose
1
Chamber Heaters
2
Vacuum
3
4
Help Window
Steam Injection
5
6
RH Solenoid
Door Interlock
7
Vent Solenoid
8
Heatsink Heater
Turns heaters On and Off,
controls to last set temperature
on main menu (Normal
Operating Mode)
Turns vacuum solenoid On and
Off
Displays Help Window
Opens and automatically closes
water solenoid (1 second
injection)
Opens and closes RH solenoids
Locks and unlocks load door
(and unload door if equipped,
see Critical Options page 23)
Opens and closes the vent
solenoid
Turns on heatsink heater
controls to 95 °C
Figure 22. SW1 Configuration Switch Functions During
Solenoid Test.
SW #
Description
Purpose
1
Electric Pump
2
Printer Test
3
Nitrogen Injection
4
Report
Indicates electric pump is
installed, required for vacuum
(see above)
If printer is installed and was
setup in Site Setup, a printer test
is printed (see Figure 24)
Opens and closes nitrogen
solenoid (if equipped, see
Critical Options page 23)
Allows strip chart recorder to
start, records all sensors
Figure 23. SW2 Configuration Switch Functions During
Solenoid Test.
Figure 24. Printer Test Printout.
46
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
DIP Switch Selectable Service Mode Options
Final Vacuum
Emergency Aeration
The Final Vacuum option automatically forces the system
into final vacuum, purge, locked, and unlocked aeration.
1. Power Off the Steri-Vac 8XL Sterilizer.
The DIP switch selectable Emergency Aeration option is
used by the service representative to automatically force
the system to the aeration stage. Emergency aeration runs
for the locked aeration time selected in Site Setup.
1. Power Off the Steri-Vac 8XL Sterilizer.
DANGER
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
Set SW1 configuration switches 6, 7, and 8 to the
closed position (see Figure 25).
Set SW1 configuration switches 1-5 open (see Figure
26).
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Final Vacuum
(Service Mode)
1
Figure 25. SW1 Configuration Switch Positions for
Final Vacuum.
4.
5.
6.
Power On the Steri-Vac 8XL Sterilizer.
The sterilizer automatically enters into final vacuum,
purge, locked, and unlocked aeration.
After Final Vacuum starts, place switches 6, 7, and 8
to the Open position.
2
3
4
5
6
7
8
SW1 Dip Switch #s
Emergency
Aeration
(Service Mode)
Figure 26. SW1 Configuration Switch Positions for
Emergency Aeration.
4.
5.
6.
© 3M 1995 - 2005
Set SW1 configuration switches 4, 7, and 8 to the
closed position (see Figure 26).
Set SW1 configuration switches 1-3 and 5-6 open (see
Figure 27).
Power On the Steri-Vac 8XL Sterilizer.
The sterilizer automatically enters the aeration stage,
for the locked aeration time setup in Site Setup.
After Aeration starts, place switches 4, 7, and 8 to the
Open position.
47
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
Reset Cycle to Idle
DANGER
Reset Cycle to Idle should only be used by
a trained 3M Service Representative and
only after all other attempts to clear errors
have failed. It is possible during this
procedure for the sterilizer door to unlock
regardless of what is in the chamber.
Refer to the Ethylene Oxide Safety
information at the front of this manual for
the effects of EO exposure.
After power is cycled in this procedure, the software
requests the Master access code. If the correct code is
given, the Door will UNLOCK, and the chamber will
equalize regardless of what is in the chamber.
1.
Power Off the Steri-Vac 8XL Sterilizer.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
4.
5.
6.
48
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Reset Cycle to Idle
(System Reset
Mode)
Figure 27. SW1 Configuration Switch Positions to Reset
Cycle to Idle.
Set SW1 configuration switches 7 to the closed
position (see Figure 27).
Set SW1 configuration switches 1-6, and 8 open (see
Figure 27).
Power On the sterilizer.
Using the number softkeys, enter the Master Access
Code. After entering the correct Master Access Code
one of the Normal Mode displays will appear (see
page 9).
Set SW1 configuration switches to the Normal Mode
positions (see page 9, Figures 4 and 5).
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
Clear Silent Abort
Clear Silent Abort is used to correct a Silent Abort error.
This procedure is only used to clear specified errors and
should not be performed unless directed by those error
analysis procedures.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
Note: Do not cycle power.
1.
2.
Set SW1 configuration switches 3 and 7 to the closed
position (see Figure 28).
Set SW1 configuration switches 1, 2, 4-6, and 8 open
(see Figure 28).
During Clear Silent Abort the sterilizer automatically
does the following:
•
•
•
•
•
•
3.
Starts printing the report.
Punctures gas cartridge for one minute.
Runs a time based final vacuum for 18 minutes.
Opens the vent and purges for 30 minutes.
Closes the vent and locks up.
After Clear Silent Abort, the screen clears, the
error is displayed, and the report stops printing.
If an error is still displayed after the gas is removed
from the chamber, return to the error troubleshooting
procedure to clear the error.
© 3M 1995 - 2005
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Clear Silent Abort
(System Reset
Mode)
Figure 28. SW1 Configuration Switch Positions to Clear
Silent Abort.
49
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
Reset Site Setup
Reset Site Setup is automatically performed by the
sterilizer when an E1 or E2 checksum error occurs or
when a new controller circuit board is installed. Reset Site
Setup can be forced by the service representative for
troubleshooting purposes or if one of the conditions
previously mentioned did not automatically cause a reset
to occur.
Note: When a new controller board is installed and the
sterilizer automatically performs a system reset, three long
beeps will be heard when the sterilizer is powered up.
This indicates the system parameters are set to their default
settings.
1.
Power Off the Steri-Vac 8XL Sterilizer.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
2.
3.
4.
5.
6.
50
Set SW1 configuration switches 2 and 7 to the closed
position (see Figure 29).
Set SW1 configuration switches 1, 3-6, and 8 open
(see Figure 29).
Power On the sterilizer.
Six short beeps will be heard indicating the system
parameters are set to their default settings.
A final vacuum, purge, and mandatory aeration are
performed.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Reset Site Setup
(System Reset
Mode)
Figure 29. SW1 Configuration Switch Positions to Reset
Site Setup.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
Error Code Flow Description
Each of the processes tested before or during the
sterilization cycle has a specific error number associated
with it.
The error codes are selected to indicate to the operator or
the service rep when the failure was detected (before the
cycle, before gas was injected, or after gas was injected)
and what process or component most likely failed
(examples: RAM test, initial vacuum, final vacuum, or
temperature during gas exposure).
© 3M 1995 - 2005
51
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
Error Code Listing
Error
Code
Error Message
Condition
E1
Cycle Information Checksum Invalid
Displayed when the cycle information checksum is invalid. This error
automatically assumes the worst-case scenario, gas in chamber. The final
vacuum, purge, and mandatory aeration are performed. Under most
circumstances, this error never occurs. Three medium beeps provide an
audible indication of this error.
E2
Setup Information Checksum Invalid
Displayed when the site setup information checksum is invalid. This error
automatically assumes the worst-case scenario, gas in chamber. The final
vacuum, purge, and mandatory aeration are performed. Under most
circumstances, this error never occurs. Four medium beeps provide an
audible indication of this error.
E3
Electronics Failure
This error does not allow the user access to the Steri-Vac 8XL Sterilizer. If
this electronics failure occurs, video may not be initialized. Three long
beeps provide an audible indication of
this error.
E4
Chamber Temperature Sensor
Temperature sensor range is 5°C to 65°C. If the temperature exceeds this
range at power up or at the start of a cycle, E4 is displayed.
E5
Heatsink Temperature Sensor
Temperature sensor range is 5°C to 99°C. If the temperature exceeds this
range at power up or at the start of a cycle, E5 is displayed.
E6
Vacuum Sensor Failure
Indicates a vacuum sensor failure. This is only a range check and is
designed to detect a disconnected sensor.
E7
Vacuum Sensor With Door Open Failure
Indicates a range failure while the door is open, out-of-range ambient
pressure.
E8
Wall Temp Sensor Failure, End
Temperature sensor range is 5°C to 65°C. If the temperature exceeds this
range at power up or at the start of a cycle, E8 is displayed.
E9
Wall Temp Sensor Failure, Middle
Temperature sensor range is 5°C to 65°C. If the temperature exceeds this
range at power up or at the start of a cycle, E9 is displayed.
E10
Low Water Error
If the water is low, C2 caution appears during idle. If a cycle is started
with C2 caution displayed, E10 occurs.
E11
Compressed Air Failure
If the compressed air pressure is low, C4 caution appears during idle. If a
cycle is started with C4 caution displayed, E11 occurs.
E12
No Corresponding Error Description
E13
No Gas Cartridge Inserted
No gas cartridge has been inserted or the last cartridge was not removed
and replaced with a new cartridge.
E14
Loading Door Open
Load door open after the Start softkey is pressed.
E15
Unloading Door Open
Unload door open after the Start softkey is pressed. C6 caution is
displayed until the unload door is closed.
E16
Unloading Door Unlocked
Unload door unlocked after the Start softkey is pressed.
E17
Loading Door Unlock
Load door unlocked after the Start softkey is pressed.
E18 E19
No Corresponding Error Description
52
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
Error
Code
Error Message
Condition
E20
Chamber Too Hot
The temperature of the chamber is 3°C hotter (or more) than the selected
cycle temperature.
E21
No Vacuum
This error occurs within one minute after vacuum is started for early
detection of vacuum failures.
E22
Initial Vacuum Failure
A time-out condition occurred while pulling a vacuum.
E23
Chamber Preheat Failure
A time-out during the preheating of the chamber to the cycle temperature
selected.
E24
Heatsink PreHeat Time-out
Indicates a time-out in the preheating of the heatsink to the temperature of
95°C.
E25
Chamber Over Temperature Error
Indicates a temperature control problem. The chamber is at least 3°C
hotter than the selected cycle temperature.
E26
Chamber Under Temperature Error
Indicates a temperature control problem. The chamber is at least 3°C
cooler than the selected cycle temperature.
E27
No Corresponding Error Description
E28
No Moisture Injected
Indicates problems with the water injection and RH chamber systems. If
the heatsink doesn't decrease in temperature after the third moisture
injection, this error is displayed.
E29
Low Humidity Error
Indicates too many moisture injections occur.
E30
RH Sensor Failure
RH sensor range is 1% to 99%. If the RH exceeds this range at power up
or at the start of a cycle, E30 is displayed.
E31
Chamber Leak
Indicates a leak into the chamber.
E32
Pressure Level Before Puncture
Indicates an improper pressure before end of precondition and start of gas
injection.
E33
Interlock Relay Close Error
Indicates a latching relay failure. This failure prevents the puncture
solenoid from being energized.
E34 E37
No Corresponding Error Description
E38
Nitrogen Pressure Error
Indicates nitrogen tank pressure switch failed.
E39
Nitrogen Injection Time-out Error
Indicates a nitrogen injection problem by monitoring the pressure in the
chamber.
E40
User Interruption During Precondition
Indicates Stop Button was pressed during the precondition stage.
E41 E49
No Corresponding Error Description
E50
Gas Cartridge Empty
Indicates no pressure increase in the chamber after the cartridge is
punctured in the gas injection stage.
E51
Chamber Vacuum Leakage
Indicates chamber pressure too close to ambient pressure and vacuum
cannot be maintained.
E52
Chamber Temperature Under Selected
Cycle During Gas Expose
Indicates a temperature control problem. The chamber is at least 3°C
cooler than the selected cycle temperature.
© 3M 1995 - 2005
53
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
Error
Code
Error Message
Condition
E53
Chamber Temperature Over Selected Cycle
During Gas Expose
Indicates a temperature control problem. The chamber is at least 3°C
hotter than the selected cycle temperature.
E54
Extended Power Outage
Indicates a chamber pressure problem after a power outage.
E55
Undefined Error Condition
Indicates an unknown condition error. This error occurs if
(1) A self test error code occurred during a stage other than cycle start
stage.
(2) A precondition error code occurs during gas expose stage.
(3) An unknown stage occurred.
Verifies the error occurred in the proper stage of the cycle.
E56 E59
No Corresponding Error Description
E60
User Interruption During Gas Expose
E61 E70
No Corresponding Error Description
E71
Final Vacuum Pump down Error
Indicates a time-out condition while pulling a vacuum in the final vacuum
stage.
E72
Obstructed Air Inlet
Indicates a clogged air inlet filter.
E73
Vacuum Sensor Gas Error
Indicates a vacuum sensor failure while gas is in the chamber.
E74
No Corresponding Error Description
E75
Gas Low Pressure Error
Indicates an insufficient pressure increase after the gas is injected.
E76
Interlock Relay Open Error
Interlock won't open to indicate no gas in chamber, or gas is in the chamber
and the door unlock function was called, or the cycle was complete and gas
was still in chamber. This is a safety precaution. Before access is allowed
into the chamber, the interlock relay must be open.
E77
Puncture Pin Stuck Error
Indicates the puncture pin is stuck in the up position or switch is
misaligned.
E80
Abator Error
Indicates a malfunctioning abator. Abator not ready in 120 minutes after
final vacuum starts.
54
Indicates the Stop button was pressed during the gas expose stage.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
Caution
Code
Caution Message
Condition
C1
Air Flow Caution
Loss of air flow in exhaust hood.
C2
Low Water Caution
Low water in the container.
C3
Power Interruption Caution
Loss of system power during a cycle.
C4
Compressed Air Caution
Loss of compressed air.
C5
Temperature Failed during Aeration
Caution
Improper temperature during aeration.
C6
Unload Door Not Shut Caution
The unload door is open.
C7
Door Locked Caution
The load door is locked when it should be unlocked.
C8
Printer Out of Paper Caution
No paper or printer ribbon cable is reversed on one end
C9
Printer Failure Caution
Printer problem.
C10
Abator Caution
Abator problem.
C11
Replace Cartridge Caution
Ethylene Oxide cartridge may not have been replaced.
C12
Calibration Required
Possible calibration problem. Check and adjust 5 volt supply. When the 5
volt supply is properly adjusted, the controller’s reference voltage (REF)
will be 2.54 V to 2.46 V. Calibration (CAL) should be 47.5 0C + 0.5 0C.
(See Help Window on page 43.)
© 3M 1995 - 2005
55
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
Error Code Failure Analysis
E1 - Cycle Information Checksum Invalid
The Cycle Information Checksum Invalid error is
displayed when the cycle information checksum is
invalid. This error automatically assumes the worst-case
scenario of gas in chamber. The final vacuum, purge, and
mandatory aeration will be performed. Under most
circumstances this error never occurs. Three medium
beeps provide an audible indication of this error.
Checked
• At power up
Possible Causes
•
A new device (U2, U23, U24, U27) was
installed without resetting the setup information
on the first power up
• Unstable power supply
• Power supply short
Note: When replacing components, make sure you
Reset Site Setup (page 50) to default all site setup
information.
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
56
E2 - Setup Information Checksum Invalid
The Setup Information Checksum Invalid error is
displayed when the site setup information checksum is
invalid. This error will automatically assume the worstcase scenario of gas in chamber. The final vacuum, purge,
and mandatory aeration will be performed. Under most
circumstances this error will never occur. Four medium
beeps provide an audible indication of this error.
Checked
• At power up
Possible Causes
•
New device (U2, U23, U24, U27) was installed
without resetting the setup information on the
first power up
• Unstable power supply
• Power supply short
Note: When replacing components, make sure you
Reset Site Setup (page 50) to default all site setup
information.
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E3 - Electronics Failure
E4 - Chamber Temperature Sensor
The Electronics Failure error will not allow the user
access to the sterilizer.
The valid range for this temperature sensor is 5°C to
65°C. If the temperature exceeds this range at either
power up or at the start of a cycle, E4 will be displayed.
Note: If this error occurs during power up, the video
display may not be initialized. Three long beeps provide
an audible indication of this error.
If this error occurs during precondition or purge, the
sterilizer will lock up until the error is manually cleared
by the service representative. See Reset Cycle to Idle
(page 48) for clearing procedure.
Checked
• At power up
• During precondition
• During purge
Possible Causes
Checked
• At power up
• At the start of each cycle
Possible Causes
• Disconnected sensor, faulty connection
• Chamber too hot
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
• EEPROM checksum failure U23
• RH set point saved failure
• RAM failure U24
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
© 3M 1995 - 2005
57
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E5 - Heatsink Temperature Sensor
E6 - Vacuum Sensor Failure
The valid range for this temperature sensor is 5°C to
99°C. If the temperature exceeds this range at either
power up or at the start of a cycle, E5 will be displayed.
The Vacuum Sensor Failure error indicates a vacuum
sensor failure. This is only a range check and is designed
to detect a disconnected sensor.
Checked
Checked
• At power up
• At the start of each cycle
Possible Causes:
• Disconnected sensor, faulty connection
• Chamber too hot
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
58
• At power up
• At the start of each cycle
Possible Causes:
• Disconnected sensor, faulty connection
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E7 - Vacuum Sensor with Door Open Failure
E8 - Wall Temperature Sensor Failure, End
The Vacuum Sensor with Door Open Failure error
indicates a range error while the door is open,
out-of-range ambient pressure.
The valid range for this temperature sensor is 5°C to
65°C. If the temperature exceeds this range at either
power up or at the start of a cycle, E8 will be displayed.
Checked
Checked
• At power up, with door open
• At the start of each cycle, if door is open
Possible Causes:
• At power up
• At the start of each cycle
Possible Causes:
• Disconnected sensor, faulty connection
• Sensor failure
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
• Disconnected sensor, faulty connection
• Sensor failure
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
© 3M 1995 - 2005
59
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E9 - Wall Temperature Sensor Failure,
Middle
The valid range for this temperature sensor is 5°C to
65°C. If the temperature exceeds this range at either
power up or at the start of a cycle, E9 will be displayed.
Checked
• At power up
• At the start of each cycle
Possible Causes:
• Disconnected sensor, faulty connection
• Sensor failure
Clearing
The sterilizer stops and ignores all softkeys until power is
cycled.
60
E10 - Low Water Error
If the water is low, caution C2 will appear during idle. If
a cycle is started with this caution displayed, E5 will
occur.
Checked
• At the start of each cycle
Possible Causes:
• Water low
• Disconnected sensor, faulty connection
• Sensor failure
Clearing
1. Raise the door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E11 - Compressed Air Failure
E12 - No Corresponding Error Description
The Compressed Air Failure error indicates the
compressor air pressure is low. Caution C4 will appear
during idle. If a cycle is started with this caution
displayed, E11 will occur.
Checked
• At the start of each cycle
Possible Causes:
• Compressed air disconnected
• Compressed air pressure insufficient
• Disconnected sensor, faulty connection
Clearing
1. Raise the door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
61
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E13 - No Gas Cartridge Inserted
E14 - Loading Door Open
The No Gas Cartridge Inserted error indicates no gas
cartridge has been inserted, or the last cartridge was not
removed and replaced with a new cartridge.
The Loading Door Open error indicates the load door is
open after the Start softkey is pressed.
Checked
• At the start of each cycle
Possible Causes:
• At the start of each cycle
Possible Causes:
• No gas cartridge
• Gas cartridge has not been removed
• Disconnected sensor or faulty connection
Clearing
1. Raise door handle.
2. After door opens, press Stop.
62
Checked
•
•
Load door is open
Disconnected or misaligned switch, faulty
connection
Clearing
1. If the door is open:
a. Press Stop
b. Close door
c. Restart cycle.
2. Otherwise:
a. Raise door handle.
b. After door opens, press Stop.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E15 - Unloading Door Open
E16 - Unloading Door Locked
Note: This error only occurs with the 2-door model of
the Steri-Vac 8XL Sterilizer.
The Unloading Door Open error indicates the unload door
is open after the Start softkey is pressed. Caution C6
will be displayed until the unload door is closed.
The Unloading Door Locked error indicates the unload
door is unlocked after the Start softkey is pressed.
Checked
• At the start of each cycle
Possible Causes:
•
•
Unload door is open
Disconnected or misaligned switch, or faulty
connection
Clearing
1. Close the unload door.
2. If the door is open:
a. Press Stop
b. Close door
c. Restart cycle.
3. Otherwise:
a. Raise door handle.
b. After door opens, press Stop.
© 3M 1995 - 2005
Checked
• At the start of each cycle
Possible Causes:
•
•
Unload door locking latch is stuck
Disconnected or misaligned switch, or faulty
connection
Clearing
1. If the door is open:
a. Press Stop
b. Close door
c. Restart cycle.
2. Otherwise;
a. Raise door handle.
b. After door opens, press Stop.
63
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E17 - Loading Door Unlock
The Loading Door Unlock error indicates the load door is
unlocked after the Start softkey is pressed.
E18 - E19 - No Corresponding Error
Description
Checked
• At the start of each cycle
Possible Causes:
•
•
Load door locking latch is stuck
Disconnected or misaligned switch, or faulty
connection
• Door handle not in closed position
Clearing
1. If the door is open:
a. Press Stop
b. Close door
c. Restart cycle.
2. Otherwise:
a. Raise door handle.
b. After door opens, press Stop.
64
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E20 - Chamber Too Hot
E21 - No Vacuum
The Chamber Too Hot error indicates the temperature of
the chamber is 3°C hotter (or more) than the selected
cycle temperature.
The No Vacuum error occurs within one minute after
vacuum is started for early detection of vacuum failures.
Checked
• At the start of precondition
Possible Causes:
• Chamber needs to cool down
Clearing
1. Raise door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
Checked
•
At the start of precondition (pressure must
decrease 52 mBars in 60 seconds)
Possible Causes:
• No compressed air
• Chamber leak
• Faulty vacuum sensor
Clearing
1. Raise door handle.
2. After door opens, press Stop.
65
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E22 - Initial Vacuum Failure
E23 - Chamber Preheat Failure
The Initial Vacuum Failure error indicates a time-out
condition occurred while pulling vacuum.
The Chamber Preheat Failure error indicates a time-out
condition occurred while preheating the chamber to the
cycle temperature selected.
Checked
During precondition
•
Pressure must decrease to puncture pressure 40
minutes, after temperature is reached
• Pressure must decrease to puncture pressure 10
minutes, before humidification
• Pressure must decrease to puncture pressure 40
minutes, after nitrogen injection
Possible Causes:
• Chamber leak
• Faulty vacuum sensor
• Defective vacuum pump
• Low compressor air pressure
• Restricted vent line
Clearing
1. Raise door handle.
2. After door opens, press Stop.
66
Checked
•
•
During precondition, time-out of 45 minutes
During Recovery of gas expose stage, time-out
of 45 minutes
Possible Causes:
•
•
Chamber High Limit thermostat
Solid state relay failure, SSR1 or SSR2, always
open circuited
• Calibration of end and middle temperature
sensors
• Disconnected Heater Blanket
Clearing
1. Raise door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E24 - Heatsink Preheat Time-out
E25 - Chamber Over Temperature
The Heatsink Preheat Time-out error indicates a time-out
condition occurred while preheating the heatsink to the
temperature of 95°C.
The Chamber Over Temperature error indicates a
temperature control problem. The chamber temperature
is at least 3°C hotter than the selected cycle temperature.
Checked
Checked
•
During precondition, after initial pumpdown,
time-out of 15 minutes
Possible Causes:
• Heatsink High Limit thermostat
• Solid state relay failure, SSR3
Clearing
1. Raise door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
• During precondition, humidification
• At the end of precondition
• During gas expose stage
Possible Causes:
• Chamber High Limit thermostat
• Solid state relay failure, SSR1 or SSR2 shorted
Clearing
1. Raise door handle.
2. After door opens, press Stop.
67
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E26 - Chamber Under Temperature
E27 - No Corresponding Error Description
The Chamber Under Temperature error indicates a
temperature control problem. The chamber temperature
is at least 3°C cooler than the selected cycle temperature.
Checked
• During precondition, humidification
• At the end of precondition
Possible Causes:
•
•
Chamber High Limit thermostat
Solid state relay failure, SSR1 or SSR2 always
open circuited
Clearing
1. Raise door handle.
2. After door opens, press Stop.
68
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E28 - No Moisture Injected
E29 - Low Humidity
The No Moisture Injected error indicates a problem with
water injection. If the heatsink doesn’t decrease in
temperature after the third moisture injection, this error
will be displayed.
The Low Humidity error indicates too many moisture
injections have occurred.
Checked
•
•
During precondition, humidification
Third moisture injection - checks heatsink
temperature decrease
Possible Causes:
•
•
•
Disconnected heatsink temperature sensor, faulty
connection
Heatsink temperature sensor failure
Clogged water line
Clearing
1. Raise door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
Checked
•
During precondition, humidification
Too many injections before reaching the target
RH value (default is 70%). There are typically
about 16 injections for a cool cycle and 15
injections for a warm cycle.
Possible Causes:
•
•
•
Water low
RH chamber leak
RH sensor disconnected, faulty connection, or
sensor failure
• Water solenoid failure
Clearing
1. Raise door handle.
2. After door opens, press Stop.
69
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E30 - RH Sensor Failure
E31 - Chamber Leak
The valid range for the RH sensor is 1% to 99%. The RH
Sensor Failure error indicates the RH exceeds this range
at end of initial vacuum in precondition.
The Chamber Leak error indicates a leak in the chamber.
Checked
•
•
• At end of initial vacuum in precondition
Possible Causes:
During precondition, before humidification
Chamber pressure increases more than 15 mBars
in five minutes
Possible Causes:
• Disconnected sensor, faulty connection
• Saturated sensor
• Leak in RH chamber
Clearing
1. Raise door handle.
2. After door opens, press Stop.
• Door seal
• Loose fittings
• Check valve failure
Clearing
1. Raise door handle.
2. After door opens, press Stop.
Checked
70
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E32 - Pressure Level Before Puncture
E33 - Interlock Relay Close
The Pressure Level Before Puncture error indicates an
improper pressure before end of precondition and start of
gas injection.
The Interlock Relay Close error indicates a latching relay
failure. This failure prevents the puncture solenoid from
being energized.
Checked
Checked
•
After humidification in the precondition stage
(Time-out of 15 minutes)
Possible Causes:
•
•
No compressed air
Compressed air sensor disconnected, faulty
connection
• Pressure transducer disconnected, faulty
connection, or sensor failure
Clearing
1. Raise the door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
• Before gas injection
Possible Causes:
• U6 or U13 defective
• 24-volt power supply failure
Clearing
1. Raise the door handle.
2. After door opens, press Stop.
71
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E34 - E37 - No Corresponding Error
Description
E38 - Nitrogen Pressure (Italian version only)
The nitrogen pressure is low.
Checked
• During precondition
Possible Causes:
• Nitrogen tank is empty
• Faulty pressure transducer or switch
Clearing
1. Raise the door handle.
2. After door opens, press Stop.
72
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E39 - Nitrogen Injection Time-Out (Italian
version only)
The Nitrogen Injection Time-Out error indicates a
nitrogen injection problem by monitoring the pressure in
the chamber.
Checked
•
During nitrogen injection (precondition)
(Pressure did not reach 800 mBars in three
minutes)
Possible Causes:
• Nitrogen tank is empty
Clearing
1. Raise the door handle.
2. After door opens, press Stop.
© 3M 1995 - 2005
E40 - User Interruption During Precondition
The User Interruption during precondition error indicates
the Stop button was pressed during the precondition
stage.
Checked
• During precondition
Possible Causes:
• User pressed Stop
• Faulty Stop switch
• Sudden ground spike
Clearing
1. Raise the door handle.
2. After door opens, press the Stop button.
73
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E41 - E49 - No Corresponding Error
Description
E50 - Gas Cartridge Empty
The Gas Cartridge Empty error indicates there is no
pressure increase in the chamber after the cartridge is
punctured in the gas injection stage.
Checked
•
During gas injection, pressure increased less
than 50 mBars after puncture
Possible Causes:
• Cartridge not properly seated
• Puncture pin failure
• No cartridge inserted (see E13)
• Empty cartridge
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
74
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E51 - Chamber Vacuum Leakage
The Chamber Vacuum Leakage error occurs when the
chamber pressure approaches within 80 mBar of ambient
pressure three times.
Checked
• During the gas injection and gas exposure stages
Possible Causes:
• Door seal
• Loose fittings
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
© 3M 1995 - 2005
E52 - Chamber Temperature Under Selected
Cycle During Gas Expose
The Chamber Temperature Under Selected Cycle During
Gas Expose error indicates a temperature control
problem. The chamber is at least 3°C cooler than the
selected cycle temperature.
Checked
• During gas injection and gas exposure stages
Possible Causes:
•
•
•
Chamber High Limit thermostat
Solid state relay failure, SSR1 or SSR2
Improper calibration of end and/or middle
temperature sensors
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
75
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E53 - Chamber Temperature Over Selected
Cycle During Gas Expose
The Chamber Temperature Over Selected Cycle During
Gas Expose error indicates a temperature control
problem. The chamber is at least 3°C hotter than the
selected cycle temperature.
Checked
• During gas injection and gas exposure stages
Possible Causes:
•
•
Solid state relay failure, SSR1 or SSR2
Improper calibration of end and/or middle
temperature sensors
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
76
E54 - Extended Power Outage
The Extended Power Outage error indicates a chamber
pressure problem after a power outage. Pressure inside
chamber is too close to ambient.
Checked
•
During recovery of gas injection or gas exposure
stages, after a power outage
Possible Causes:
• Air leak into chamber
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E55 - Undefined Error Condition
The Undefined Error Condition error is an unknown error
condition and occurs if:
•
A selftest error code occurred during a stage other
than cycle start stage
•
A precondition error code occurs during gas expose
stage
E56 - 59 - No Corresponding Error
Description
• An unknown stage occurred
This verifies the error occurred in the proper stage of the
cycle.
Checked
• Error manager, after another error occurs
Possible Causes:
• See E2 and E3
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
© 3M 1995 - 2005
77
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E60 - User Interruption During Gas Expose
The User Interruption During Gas Expose error indicates
the Stop button was pressed during the gas injection or
gas expose stage.
E61 - E70 - No Corresponding Error
Description
Checked
• During gas injection or gas exposure stages
Possible Causes:
• User pressed Stop
• Stop switch failure
• Sudden ground spike
Clearing
1. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
2. After door opens, press the Stop button.
78
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E71 - Final Vacuum Pump Down
E72 - Obstructed Air Inlet
The Final Vacuum Pump Down error indicates a time-out
condition occurred while pulling vacuum in the final
vacuum stage.
If an abator in installed (see Critical Options, page 23), a
time-out occurs if the pressure does not reach 250 mBars
within 60 minutes, and does not reach 160 mBars within
an additional 30 minutes.
If an abator is not installed, a time-out will occur if the
pressure does not reach 160 mBars within 90 minutes.
This error may also occur during gas expose, if the
pressure can not be lowered to 180 mBars below ambient
pressure within 2 minutes.
The Obstructed Air Inlet error indicates a clogged air inlet
filter. When the purge stage starts, the vent is opened and
the vacuum remains On. If the pressure does not reach
ambient pressure minus 400 mBars within eight minutes,
then this error is displayed.
Checked
• During final vacuum
• During gas expose
Possible Causes:
•
•
•
Compressed air disconnected
Compressed air pressure insufficient
Disconnected pressure transducer or faulty
connection
• Vacuum pump failure
• Restricted vent line
• Air leak into chamber
Clearing
1. After fixing the source of the error, press the Start
softkey.
2. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
3. After door opens, press the Stop button.
© 3M 1995 - 2005
Checked
• At the start of the purge stage
Possible Causes:
• Clogged air inlet filter
• Vent solenoid failure
• 24-volt power supply failure
Clearing
1. After fixing the source of the error, press the Start
softkey.
2. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
3. After door opens, press the Stop button.
79
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E73 - Vacuum Sensor Gas
E74 - No Corresponding Error Description
The Vacuum Sensor Gas error indicates a vacuum sensor
failure while gas is in the chamber.
Clearing
This error locks the system with the vacuum running.
1. When the error occurs, the sterilizer automatically
does the following:
a. Stops the timer and stops printing the report.
b. Closes the vent and pulls the puncture pin down.
c. Turns Off the heaters and turns On the vacuum
for 18 minutes.
Note: Do not cycle power.
2. Place SW1 switches to Clear Silent Abort (see page
49).
Note:
During Clear Silent Abort the sterilizer
automatically does the following:
•
•
•
3.
80
Starts printing the report
Punctures gas cartridge for one minute.
Runs a time-based final vacuum for 18
minutes.
• Opens the vent and purges for 30 minutes.
• Closes the vent and locks up.
a. After Clear Silent Abort starts, place SW1
configuration switches into Normal mode (see
page 9).
b. When Clear Silent Abort is complete the screen
clears and displays the error, and the report stops
printing.
To clear the error (after the gas is removed from the
chamber):
a. Turn the sterilizer power Off
b. Perform Reset Cycle to Idle (see page 48).
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E75 - Gas Low Pressure
E76 - Interlock Relay Open
The Gas Low Pressure error indicates an insufficient
pressure increase after the gas is injected.
The Interlock Relay Open error indicates the interlock
will not open to indicate no gas in chamber, or gas is in
the chamber and the door unlock function was called, or
the cycle was complete and gas was still in the chamber.
This is a safety precaution, before access is allowed into
the chamber, the interlock relay must be open.
If the pressure increase is less than 200 mBar after gas
injection stage, this error will occur and a time-base final
vacuum and purge are performed.
Checked
• During gas injection
Possible Causes:
• Partial puncture of a gas cartridge
• Partially filled gas cartridge
• Disconnected or defective pressure transducer
• Faulty pressure transducer connection
• 24-volt power supply failure
• Gas cartridge popout (if E13 is displayed)
Clearing
This error locks the system with the vacuum running.
1. When the error occurs, the sterilizer automatically
does the following:
a. Stops the timer and stops printing the report.
b. Closes the vent and pulls the puncture pin down.
c. Turns Off the heaters and turns On the vacuum
for 18 minutes.
Checked
• During aeration, if gas is in chamber
• While unlocking doors
Possible Causes:
• U13 defective
• Invalid stage - software failure
• 24-volt power supply failure
Clearing
1. After fixing the source of the error, press the Start
softkey.
2. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
3. After door opens, press the Stop button.
Note: Do not cycle power.
2. Place SW1 switches to Clear Silent Abort (see page
49).
Note:
During Clear Silent Abort the sterilizer
automatically does the following:
•
•
•
3.
Starts printing the report.
Punctures gas cartridge for one minute.
Runs a time-based final vacuum for 18
minutes.
• Opens the vent and purges for 30 minutes.
• Closes the vent and locks up.
a. After Clear Silent Abort starts, place SW1
configuration switches into Normal mode (see
page 9).
b. When Clear Silent Abort is complete, the screen
clears and displays the error, and the report stops
printing.
To clear the error (after the gas is removed from the
chamber):
a. Turn the sterilizer power Off.
b. Perform Reset Cycle to Idle (see page 48).
© 3M 1995 - 2005
81
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
E77 - Puncture Pin Stuck
The Puncture Pin Stuck error indicates the puncture pin is
stuck in the up position or puncture pin switch has failed.
E78 - E79 - No Corresponding Error
Description
Checked
• At end of gas injection and at end of purge
Possible Causes:
•
•
Note:
Defective puncture pin mechanism
Disconnected or faulty puncture pin position
switch
If E77 is printed on the report after gas injection
and purge, this indicates a puncture pin switch
failure or that the puncture pin is stuck in the
down position.
If E77 is printed on report after purge only, this
indicates the puncture pin is stuck in the up
position.
Clearing
1. After fixing the source of the error, press the Start
softkey. If Start is not displayed, cycle power.
2. A puncture pin switch test is automatically
performed.
3. After final vacuum, purge, and mandatory aeration,
turn handle and wait for door to open.
4. After door opens, press the Stop button.
82
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
TROUBLESHOOTING AND REPAIR
E80 - Abator Time-Out
The abator time-out error indicates the Abator Ready
signal was not received within 120 minutes after the final
vacuum stage starts. Caution code C10 will be displayed
and printed only if the sterilizer is set for 50 CFM Abator
or 50 CFM Abator w/ Aeration.
Checked
• At the start of final vacuum stage
Possible Causes:
• Abator failure
• Disconnected or faulty abator ready line
Clearing
This error locks the system, to clear the error:
1. Perform Clear Silent Abort (see page 49).
2.
3.
Note: Do not cycle power.
After Clear Silent Abort the sterilizer returns to
Idle.
Place SW1 configuration switches into Normal mode
(see page 9).
© 3M 1995 - 2005
83
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
TROUBLESHOOTING AND REPAIR
This page intentionally left blank.
84
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SPECIAL CYCLES
Emergency Aeration
Emergency aeration is used to aerate a load that needs
further aeration. This option can be activated by
configuring SW1 DIP switches (see page 47) or from the
Main menu during Normal Mode (the following
procedure).
DANGER
Make sure all precautions pertaining to the
handling, use, and storage of Ethylene
Oxide (EO) are observed. See the Site
Planning and Installation Guide and the
Operator’s Manual for detailed information
in the handling, use, and storage of EO.
Refer to the Ethylene Oxide Safety
information at the front of this manual for
the effects of EO exposure.
1.
2.
3.
4.
5.
6.
7.
Place the items requiring further aeration in the
sterilizer chamber.
With the sterilizer in Normal Mode, use the
temperature select softkey (1) to set the cycle
temperature. Use the Up arrow and Down arrow
softkeys (2, 3) to set the cycle time.
Close the sterilizer door.
Press the Stop button (4) twice (within one second) to
display the operator’s Site Setup menu.
1
2
3
4
Using the Up and Down softkeys, highlight the
EMGY AER on the Site Setup menu.
Press the Select softkey to toggle EMGY AER to ON.
Press the Done softkey to start the emergency aeration
cycle.
© 3M 1995 - 2005
85
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SPECIAL CYCLES
Accelerate Cycle
This SW1 DIP switch controllable option allows the
service representative to accelerate the time of selected
stages of a cycle by 120 during a sterilization cycle, in the
normal operating mode (1 hour will equal 30 seconds).
See table at right for specific elements of the cycle that are
accelerated.
The acceleration DIP switch can be actuated anytime
during the sterilization cycle, usually before the cycle
starts. If the DIP switch is set to accelerate during a cycle,
the time for the current stage will not be accelerated.
During a cycle, times are first calculated then the stages
are performed. After a stage has started, the only way to
accelerate that stage is to cycle the power. When the
sterilization cycle restarts, the acceleration option is
detected and the new time is calculated. If the accelerated
option is changed to non-accelerated during an accelerated
cycle, the accelerated time continues until the next stage
starts.
Precondition
Vacuum Test
Warm up
Not accelerated, approximately 1 minute
Not accelerated, time depends on cycle
temperature selected
Not accelerated, time depends on
puncture pressure selected and vacuum
system installed
Dwell times and RH read times are
accelerated
Accelerated, approximately 5 seconds
Not accelerated, always 50 seconds
Accelerated (1 hour equals 30 seconds)
Not accelerated, time depends on
puncture pressure selected and vacuum
system installed
Accelerated
Accelerated
Accelerated
Pump down
Humidification
Leak Test (optional)
Gas Injection
Gas Expose
Final Vacuum
Purge
Locked Aeration
Unlocked Aeration
Notes: Do not consider items loaded in the sterilizer
chamber during accelerated times sterilized.
Accelerate Cycle is a Normal Mode service operation.
The sterilizer must be set to operate or be operating in
the Normal Mode to activate this option.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
1.
Make sure the sterilizer is in Normal Mode.
2.
3.
Close SW1 configuration switch 2 (see Figure 30).
If no additional cycles are to be run with accelerated
time, set SW1 configuration switch 2 to open.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Accelerate Cycle
(Normal Mode)
Figure 30. Accelerate Cycle Time SW1 Switch Positions.
86
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SPECIAL CYCLES
Simulating Low Pressure
This SW1 DIP switch controllable option allows the
service representative to simulate most pressure related
checks during a normal sterilization cycle (see table at
right for specific elements of the cycle that are simulated).
Note: Do not consider items loaded in the sterilizer
chamber during simulated pressure cycles sterilized.
The simulated pressure DIP switch can be closed anytime
during the cycle; usually the switch is closed before the
cycle starts. If the DIP switch is set to simulate low
pressure during a cycle, the pressure for the current stage
may not be simulated.
DANGER
Make sure static discharge wrist ground
straps are worn when setting configuration
switches or when working with the SteriVac 8XL Sterilizer covers removed. Static
discharge will cause damage to equipment.
Failure to wear static discharge straps can
result in damage to equipment or injury to
personnel.
Precondition
Vacuum Test
No pressure-related checks - not
simulated
No pressure-related checks - not
simulated
Simulated
One injection
Simulated
Not simulated. If pressure level does not
increase by 200 mBars a silent abort will
occur
Not simulated
Simulated
No pressure-related checks - Not
simulated
No pressure-related checks - Not
simulated
No pressure-related checks - Not
simulated
Warm up
Pump down
Humidification
Leak Test (optional)
Gas Injection
Gas Expose
Final Vacuum
Purge
Locked Aeration
Unlocked Aeration
Note: Simulating Low Pressure is a Normal Mode
service operation. The sterilizer must be set to operate or
be operating in the Normal Mode to activate this option.
1.
Make sure the sterilizer is in Normal Mode.
2.
Close SW1 configuration switch 1 (see Figure 31).
All low-pressure stages are simulated during the
sterilization cycle.
If no additional cycles are to be run with simulated
low pressure, set SW1 configuration switch 1 to open.
3.
1
2
3
4
5
6
7
8
SW1 Dip Switch #s
Simulate low Press.
(Normal Mode)
Figure 31. Simulating Low Pressure SW1 Switch
Positions.
© 3M 1995 - 2005
87
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SPECIAL CYCLES
Define Cycles
Five different temperature cycles can be set in the
Steri-Vac 8XL Sterilizer.
CAUTION
Only the 37°C, 180-minute and the 55°C,
60-minute cycles are approved for hospital
use in the US. Additional cycles must be
validated by the end user to meet FDA
requirements. Use of unapproved cycles
can result in inadequate sterilization.
Note: Only the 37°C, 180-minute and 55°C, 60-minute
cycles are active after Defaulting System.
1. The temperature and gas cycles and times are set
through the SET CYCLES option of the Misc. Service
menu. If the sterilizer is not at the Misc. Service
menu perform the necessary steps from the following
to display the Misc. Service menu:
a. Place the sterilizer in the Service Mode - page 11.
b. Place the sterilizer in Site Setup - page 13.
c. Place the sterilizer in Misc. Service - page 25.
2.
3.
4.
Using the Up and Down softkeys, highlight the SET
CYCLES option on the Misc. Service menu.
Press the Select softkey.
Using the Up and Down softkeys, highlight the
desired temperature cycle.
Notes: A zero (0) to the left of a listed temperature
indicates that temperature cycle is not selected. A
one (1) indicates the temperature is selected.
Selecting temperature with a 1 next to it deselects that temperature and it will not display on
the Standby Screen.
5.
At least one of the Cycle Temps must be selected.
Press the Select softkey. Press YES or NO to define
this cycle. See Figure 32.
Figure 32. Define Cycle
88
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SPECIAL CYCLES
Temperature
1.
The Cycle Temperature selection list is displayed.
Using the Up and Down softkeys, select the desired
temperature for the currently selected temperature
cycle. See Figure 33.
Note: A custom temperature may be entered for any
temperature cycle.
2. Press the Select softkey. The display advances to the
Preheat selection, allowing Preheat Time to be set up.
Figure 33. Temperature
Preheat Option
The Pre-Heat option allows the load to be pre-heated for a
specified amount of time during the precondition stage.
1. Using the Up and Down softkeys, select the
PREHEAT Time desired. See Figure 34
2. Press the Select softkey. The display advances to
Vacuum.
Figure 34. Preheat
© 3M 1995 - 2005
89
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SPECIAL CYCLES
Vacuum
The puncture pressure is the pressure to which the
sterilizer pumps to during normal operation before
puncturing the gas cartridge.
Note: The standard puncture pressure for sterilizers in
the US is 160 mBar. Some countries may require a lower
puncture pressure.
1. Using the Up and Down softkeys, select the puncture
pressure desired (min. 0 mbars, max. 240 mbars). See
Figure 35.
2. Press the Select softkey. The display advances to
Humidity.
Figure 35. Vacuum
Humidity
The Relative Humidity level is preset to 70%. This level
can be changed to suit special conditions.
1. Using the Up and Down softkeys, select the
appropriate relative humidity level. See Figure 36.
2. Press the Select softkey. The display advances to
Exposure time.
Figure 36. Humidity
90
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
SPECIAL CYCLES
Exposure
The Gas Exposure Time selection is displayed.
1. Using the Up and Down softkeys, select the desired
time. See Figure 37.
2. Press the Select softkey. The display will advance to
Aeration.
Figure 37. Exposure
Aeration
A locked aeration time can be customized for the needs of
the installation. If the optional exhaust hood is not
installed, three hours of aeration are recommended and
preset.
1. Using the Up and Down softkeys, select the
appropriate aeration time. See Figure 38
2. Press the Select softkey. The display advances to the
Define Cycles and displays the selected options.
Note: A small one (1) is now displayed next to the cycle
temperature indicating that temperature cycle is selected
and will be available during normal operation.
3. If additional Cycle Temps and Gas Exposure Times
need to be selected, perform steps 4 and 5 on page 88
to select additional temperature cycles.
4. After all of the desired temperature cycles are selected
and the Cycle Temps selection list is displayed, press
the Done softkey.
Note: When the temperature selection softkey is pressed
during normal operation, only the temperature cycles
selected in this procedure will appear on the display.
© 3M 1995 - 2005
Figure 38. Aeration
91
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
SPECIAL CYCLES
This page intentionally left blank.
92
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
Illustrated Parts Breakdown (IPB)
This Illustrated Parts breakdown (IPB) section contains information necessary to locate and identify assemblies, subassemblies, and specific parts of the Model 8XL Steri-Vac.
Part Numbers
All components available on a replacement basis are listed with their 3M identification number. The parts identification list
opposite each illustration provides the description adequate to identify components. If the component is available only as a
part of the next higher assembly, the description column provides the nest higher assembly description or identification
number.
Locating and Identifying a Part
If physical appearance and general location of a part are known, refer to an exploded view assembly or sub-assembly
drawing to locate the part. The part ID numbers in this manual are from the most recent data available. If part identification
is dependent upon the machine serial number, verify the machine serial number on the serial plate located on the Frame
Chamber Assembly.
Ordering Parts
When ordering parts, include the following information:
1. Machine model number and serial number.
2. Part identification number.
3. Part Description
4. Quantity required.
For orders call 800 292-6298 between 8:00 a.m. and 4:30 p.m., CST, except Saturday, Sunday, and holidays.
Abbreviations
PWA
printed wiring assembly
Mach
Machine
Dia
diameter
© 3M 1995 - 2005
93
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
This page intentionally left blank.
94
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
Illustrated Parts Breakdown Contents
Title_________________________________________________________________________________________Page
Final Assembly ………………………………………………………………………………………………………….96
Frame-Chamber Assembly ……………………………………………………………………………………………...100
Door Lock Assembly …………………………………………………………………………………………………...104
Water Filler, Printer Assembly …………………………………………………………………………………………108
Power Supply Assembly ……………………………………………………………………………....………………..110
Services-In Assembly …………………………………………………………………………………………………..112
Manifold Assembly …………………………………………………………………………………………………….116
Display Panel Assembly ……………………………………………………………………………………………….120
Water Tank Assembly …………………………………………………………………………………………………124
Chamber Assembly ……………………………………………………………………………………………………126
Chamber-Heater Assembly ……………………………………………………………………………………………132
Heat Sink Assembly ………………………………………………………………………………………………….. 134
Door Assembly ………………….……………………………………………………………………………………..136
Local Exhaust Assembly ………………………………………………………………………………………………140
Controller Bracket Assembly ………………………………………………………………………………………….144
IMPORTANT NOTICE
3M reserves the right to sell or furnish NEW or
RECONDITIONED PARTS. New and reconditioned parts are
covered by the same 3M parts limited warranty.
© 3M 1995 - 2005
95
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 1
FINAL ASSEMBLY (ONE DOOR)
96
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
FINAL ASSEMBLY (ONE DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31 (Not shown)
Part Number
78-8078-6667-4
78-8078-5677-4
26-1003-9521-4
78-8078-3878-0
Not Used
26-1011-6138-3
78-8078-5714-5
78-8078-3986-1
78-8078-3970-5
78-8078-3864-0
26-1004-3140-7
78-8078-3876-4
78-8078-5715-2
78-8078-5863-0
78-8078-3871-5
78-8078-3998-6
78-8005-1710-0
78-8078-5999-2
26-1011-7366-9
78-8078-5678-2
78-8078-5235-1
26-1004-2619-1
26-1002-5278-7
26-1000-6654-2
78-8078-3964-8
26-1003-7072-0
26-1002-8122-4
78-8078-5861-4
78-8078-6644-3
78-8078-6626-0
26-1011-8416-1
78-8078-5807-7
© 3M 1995 - 2005
Description
FINAL ASSEMBLY
REAR PANEL - TOP
Screw, Mach, 1/4-20 X .375
PANEL, Upper Side
Qty.
REF
1
45
2
HOLE PLUG (For 11/16 Hole)
ASSEMBLY – CENTER SIDE PANEL CONTROLLER
PANEL, Lower Services
FRAME CHAMBER ASSEMBLY (See IPB Figure 2)
PANEL, Bottom
BUSHING, Strain Relief
PANEL, Lower Side
CENTER SIDE PANEL ASSEMBLY, Hinge Side
LOCAL EXHAUST ASSEMBLY, Load Side
PANEL, Top
INSULATION, Rear panel
TAPE, Cloth, 1.00 W
COVER, Hose, Exhaust Hood
FASTENER
REAR PANEL, Bottom
DECAL, 8XL
SCREW, Mach, 10-32 X .500 Phillips, Black
SCREW, Mach, 8-32 X .500, Pan Head, Phillips
NUT, Mach, 8-32, ST, Zinc
EXHAUST FLANGE
HOSE
CLAMP
CATCH, Latch
RFI SHIELD, Finger Stock
DECAL, Printer Paper Loading
SHIELD, Printer Switch Cable
Power Cord
28
1
1
REF
1
1
1
1
1
1
2
AR
1
8
1
1
3
9
8
1
1
2
1
5
1
1
1
97
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 1 (continued)
FINAL ASSEMBLY (TWO DOOR)
98
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
FINAL ASSEMBLY (TWO DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
53
55
56 (Not shown)
Part Number
78-8078-6732-6
Not Used
26-1003-9521-4
78-8078-3878-0
78-8078-6626-0
26-1011-6138-3
78-8078-5714-5
78-8078-3986-1
78-8078-5883-8
78-8078-3864-0
26-1004-3140-7
78-8078-3876-4
78-8078-5715-2
Not Used
78-8078-3871-5
78-8078-5086-8
Not Used
Not Used
26-1011-7366-9
78-8078-5894-5
78-8078-5235-1
26-1004-2619-1
26-1002-5226-6
Not Used
Not Used
26-1000-6654-2
78-8078-5164-3
78-8078-6272-3
78-8078-5861-4
78-8078-3964-8
26-1003-7072-0
78-8078-6271-5
26-1002-8122-4
78-8078-5863-0
78-8078-6273-1
26-1011-8416-1
78-8078-6644-3
78-8078-5807-7
© 3M 1995 - 2005
Description
FINAL ASSEMBLY
Qty.
REF
Screw, Mach, 1/4-20 X .375
PANEL, Upper Side
DECAL – PRINTER PAPER LOADING
HOLE PLUG (For 11/16 Hole)
ASSEMBLY - CENTER SIDE PANEL CONTROLLER
PANEL, Lower Services
FRAME CHAMBER ASSEMBLY TWO DOOR (See IPB Figure 2)
PANEL, Bottom
BUSHING, Strain Relief
PANEL, Lower Side
ASSEMBLY - CENTER SIDE PANEL, Hinge Side
42
2
1
28
1
1
REF
1
2
1
1
PANEL, Top
DECAL - OPERATING INSTRUCTIONS
1
1
FASTENER
BOTTOM PANEL, UNLOAD SIDE
DECAL, 8XL
SCREW, Mach, 10-32 X .500 Phillips, Black
SCREW, Mach, 8-32 X .375, Pan Head, Phillips
8
1
2
6
4
NUT, Mach, 8-32
DECAL – OPERATING INSTRUCTIONS, UNLOAD SIDE
CATCH - Latch, Unload Side
CATCH – Latch
EXHAUST FLANGE
HOSE
EXHAUST TEE
CLAMP, HOSE
LOCAL EXHAUST ASSEMBLY – LOAD SIDE
EXHAUST HOOD ASSEMBLY – UNLOAD SIDE
SHIELD, EMI, CYLINDRICAL, FERRITE
RFI SHIELD, FINGER STOCK
Power Cord
4
1
1
1
1
1
1
8
1
1
1
4
1
99
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 2
FRAME-CHAMBER ASSEMBLY (ONE DOOR)
100
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
FRAME-CHAMBER ASSEMBLY (ONE DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
Part Number
78-8078-3970-5
78-8078-3869-9
12-7991-1519-6
78-8078-3868-1
78-8078-3960-6
78-8078-4618-9
26-1002-5288-6
78-8078-4617-1
78-8005-6087-8
26-1005-2272-6
78-8078-3983-8
78-8078-3954-9
78-8078-3959-8
26-1011-6996-4
26-1002-3590-7
12-7991-1546-9
78-8078-3870-7
78-8078-3966-3
26-1003-6373-3
26-1002-5312-4
78-8078-3973-9
26-1004-2619-1
26-1002-5287-8
78-8078-5862-2
78-8078-6725-0
78-8078-3969-7
78-8078-3867-3
26-1002-7765-1
26-1005-4049-6
12-2376-3046-7
78-8078-5672-5
12-2376-9709-4
12-2376-9835-7
12-2376-9836-5
78-8078-4611-4
78-8078-5787-1
12-7991-1482-7
26-1000-4007-5
12-2376-9588-2
© 3M 1995 - 2005
Description
FRAME-CHAMBER ASSEMBLY
TUBE, Horizontal Tie
SCREW, Cap, ¼-20 X .500, Hex
TUBE, Horizontal Support
TUBE, P/S Bracket Support
POWER SUPPLY ASSEMBLY (see IPB Figure 5)
Screw, Mach, ¼-20 X .500, Pan Head, Phil
SERVICES-IN ASSEMBLY (see IPB Figure 6)
TIE WRAP
CASTER, Swivel Mount, 3.00 Dia. X 1.25W X 4.00 OAH
WATER TANK ASSEMBLY (see IPB Figure 9)
TANK SUPPORT
TUBE, Water Support
LEVELER, Non-swivel, 3/8-16 X 3.00
CASTER
SCREW, Mach, 5/16-18 X .750, Hex
TUBE - STEP, Horizontal Support
WATER FILLER, PRINTER ASSEMBLY (see IPB Figure 4)
SCREW, Mach, 6-32 X .375, Pan Head, Phil
SCREW, Mach, 6-32 X .500, Phil
DISPLAY PANEL ASSEMBLY (see IPB Figure 8)
SCREW, Mach, 10-32 X .500, Phil
SCREW, Mach, ¼-20 X .375, Pan, Phil, STL, Black, Zinc
BRACE, Casters
WIRING HARNESS, Low Voltage
CHAMBER ASSEMBLY (see IPB Figure 10)
FRAME, Weldment
SCREW, Mach, 8-32 X .500, Pan Head, Phil
WASHER, Lock, #8
LABEL, GROUND
SERIAL, PLATE, 8XL
DECAL, Grounding
LABEL, Installation Requirements
DECAL, Poison
WIRE HARNESS, High Voltage
SCREEN, Weldment
NUT, Mach, 5/16-18, Hex
WASHER, Plain, .375 ID X 1.125 OD X .065
WIRE, RFI
Qty.
REF
4
40
1
1
REF
7
REF
1
2
REF
1
1
4
2
16
1
REF
3
6
REF
6
3
4
1
REF
2
2
2
1
REF
1
1
1
1
1
4
12
1
101
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 2 (continued)
FRAME-CHAMBER ASSEMBLY (TWO DOOR)
102
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
FRAME-CHAMBER ASSEMBLY (TWO DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
Part Number
78-8078-5883-8
78-8078-3869-9
12-7991-1519-6
78-8078-3868-1
78-8078-3960-6
78-8078-4618-9
26-1002-5288-6
78-8078-6989-2
78-8005-6087-8
78-8078-5862-2
78-8078-3983-8
78-8078-3954-9
78-8078-3959-8
26-1011-6996-4
26-1002-3590-7
12-7991-1546-9
78-8078-3870-7
78-8078-3966-3
26-1003-6373-3
26-1002-5312-4
78-8078-3973-9
26-1004-2619-1
26-1002-5287-8
78-8078-8872-8
78-8078-6725-0
78-8078-4033-1
78-8078-3867-3
26-1002-7765-1
26-1005-4049-6
12-2376-3046-7
78-8078-6372-1
12-2376-9709-4
12-2376-9835-7
12-2376-9836-5
78-8078-4611-4
78-8078-5787-1
12-7991-1482-7
26-1000-4007-5
12-2376-9588-2
78-8078-4031-5
78-8078-5893-7
26-1005-2272-6
78-8078-5086-8
26-1003-9521-4
26-1004-1786-9
© 3M 1995 - 2005
Description
FRAME-CHAMBER ASSEMBLY TWO DOOR
TUBE, Horizontal Tie
SCREW, Cap, ¼-20 X .500, Hex
TUBE, Horizontal Support
TUBE, P/S Bracket Support
POWER SUPPLY ASSEMBLY (see IPB Figure 5)
Screw, Mach, ¼-20 X .500, Pan Head, Phil
SERVICES-IN ASSEMBLY – 2 DOOR (see IPB Figure 6)
TIE WRAP
BRACE – CASTERS
WATER TANK ASSEMBLY (see IPB Figure 9)
TANK SUPPORT
TUBE, Water Support
LEVELER, Non-swivel, 3/8-16 X 3.00
CASTER
SCREW, Mach, 5/16-18 X .625, Hex
TUBE_STEP, Horizontal Support
WATER FILLER PRINTER ASSEMBLY (see IPB Figure 4)
SCREW, Mach, 6-32 X .375, Pan Head, Phil
SCREW, Mach, 6-32 X .500, Phil
DISPLAY PANEL ASSEMBLY (see IPB Figure 8)
SCREW, Mach, 10-32 X .500, Phil
SCREW, Mach, ¼-20 X .375, Pan, Phil, STL, Black, Zinc
D-SUB ADAPTER
WIRING HARNESS, Low Voltage
CHAMBER ASSEMBLY – TWO DOOR (see IPB Figure 10)
FRAME, Weldment
SCREW, Mach, 10-32 X .500, Pan Head, Phil
WASHER, Lock, #8
LABEL, GROUND
SERIAL, PLATE – TWO DOOR
DECAL, Grounding
LABEL, Installation Requirements
DECAL, Poison
WIRE HARNESS, High Voltage
SCREEN, Weldment
NUT, Mach, 5/16-18, Hex
WASHER, Plain, .375 ID X 1.125 OD X .065
WIRE, RFI
DISPLAY PANEL SUPPORT, UNLOAD SIDE
DISPLAY PANEL ASSEMBLY – UNLOAD SIDE
CASTER, SWVL MT, 3.00 DIA X 1.25 W X 4.00 OAH
DECAL – OPERATING INSTRUCTIONS
SCREW, Mach, .25-20 X .375, PAN, PHIL
LOCTITE
Qty.
REF
4
40
1
1
REF
7
REF
1
4
REF
1
1
4
2
16
1
REF
7
12
REF
6
6
1
1
REF
2
2
2
1
1
1
1
1
1
1
4
12
1
1
1
2
1
3
AR
103
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 3
DOOR LOCK ASSEMBLY (LOAD SIDE)
104
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
DOOR LOCK ASSEMBLY (LOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
Part Number
78-8078-3967-1
26-1011-7908-8
78-8078-3505-9
26-1002-8892-2
78-8078-3502-6
26-1002-5209-2
78-8078-4758-1
78-8078-5873-9
26-1004-5304-7
26-1007-0998-4
78-8678-2261-7
78-8078-5185-8
26-1017-0602-1
© 3M 1995 - 2005
Description
DOOR LOCK ASSEMBLY, Load Side
CYLINDER
BRACKET, Switch Cylinder
NUT, Mach, 10-32
BLOCK, Locking
SCREW, Mach, 4-40 X .625, Pan Head, Phil
PLATE, Nut
SWITCH, Interlock, Modified
GASKET, Fiber
FITTING, Connector, ¼ OD X 10-32
TUBING, .25 OD X .040 Wall
JUMPER, Door Switch
SWITCH
Qty.
REF
1
1
1
1
4
2
1
1
1
AR
1
1
105
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 3 (continued)
DOOR LOCK ASSEMBLY (UNLOAD SIDE)
106
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
DOOR LOCK ASSEMBLY (UNLOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
Part Number
78-8078-4029-9
26-1011-7908-8
78-8078-4023-2
26-1002-8892-2
78-8078-4017-4
26-1002-5209-2
78-8654-4758-1
78-8078-5873-9
26-1004-5304-7
26-1007-0998-4
78-8678-2261-7
78-8078-5185-8
26-1017-0602-1
© 3M 1995 - 2005
Description
DOOR LOCK ASSEMBLY, Unload Side
CYLINDER
BRACKET, Switch Cylinder, Unload Side
NUT, Mach, 10-32
BLOCK, Locking, Unload Side
SCREW, Mach, 4-40 X .625, Pan Head, Phil
PLATE, Nut
SWITCH, Interlock, Modified
GASKET, Fiber
FITTING, Connector, ¼ OD X 10-32
TUBING, .25 OD X .040 Wall
JUMPER, Door Switch
SWITCH
Qty.
REF
1
1
1
1
4
2
1
1
1
AR
1
1
107
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 4
WATER FILLER, PRINTER ASSEMBLY
108
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
WATER FILLER, PRINTER ASSEMBLY
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
Part Number
78-8078-3966-3
78-8078-5270-8
12-7996-4559-8
78-8078-3873-1
78-8078-3862-4
12-2376-9492-7
12-2376-9414-1
78-8078-5821-8
26-1002-6285-1
78-8078-3863-2
26-1000-0188-7
78-8078-3955-6
78-8078-3861-6
78-8078-5224-5
26-1002-5218-3
26-1011-6947-7
78-8005-4273-6
78-8078-7181-5
78-8078-3988-7
78-8654-2504-1
78-8078-5339-1
78-8078-6381-2
78-8078-6626-0
© 3M 1995 - 2005
Description
WATER FILLER, PRINTER ASSEMBLY
BOARD, Printer, Altered
SCREW, Cap, 6-32 X.250 Hex
DOOR, Printer/Water
SUPPORT, Vertical
SWITCH, SPDT, 25 VAC, 6 Amp
DECAL, Printer
PRINTER
SCREW, Mach, M3 X 0.5 X 6.0, Pan Head, Phil
COVER, Printer
SCREW, Mach, 10-32 X .375, Pan Head, Phil
ADAPTER
WATER FILLER ASSEMBLY
CAP, Water
SCREW, Mach, 6-32 X .500, Pan Head Phil
FITTING, Plastic
CATCH
DECAL, Health & Safety
SPACER, Display Board
NUT, Mach, 6-32, Hex
CABLE, Controller to Printer
PANEL, Caution and Error Message
DECAL, Printer Loading
Qty.
REF
1
4
1
1
1
1
1
2
1
2
1
1
1
4
1
1
1
4
4
1
1
1
109
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 5
POWER SUPPLY ASSEMBLY
110
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
POWER SUPPLY ASSEMBLY
Item No.
0
1
2
3
4
5
6
7
Part Number
78-8078-4618-9
78-8078-3957-2
12-2376-9403-4
26-1001-7983-2
78-8078-6063-6
26-1002-5218-3
78-8412-9417-2
78-8005-2893-3
© 3M 1995 - 2005
Description
POWER SUPPLY ASSEMBLY
POWER SUPPLY BRACKET ASSEMBLY
24 VDC POWER SUPPLY ASSEMBLY
Nut, Mach, 6-32
5 VDC POWER SUPPLY ASSEMBLY
SCREW, Mach, Phil, Pan Head, 6-32 X .500
FUSE, 24 VDC Power Supply
FUSE, 5 VDC Power Supply
Qty.
REF
1
1
5
1
1
1
1
111
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 6
SERVICES-IN ASSEMBLY (ONE DOOR)
112
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
SERVICES-IN ASSEMBLY (ONE DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Part Number
78-8078-4617-1
78-8078-3958-0
26-1002-5277-9
26-1011-7365-1
26-1002-5345-4
26-1001-7983-2
26-1011-7372-7
26-1002-2929-8
78-8078-5154-4
26-1011-8072-2
26-1000-6654-2
26-1004-2053-3
78-8078-5349-0
78-8078-8392-7
26-1011-7370-1
80-6100-6110-5
26-1014-7078-4
78-8078-8615-1
© 3M 1995 - 2005
Description
SERVICES IN ASSEMBLY
BRACKET SERVICE ASSEMBLY
SCREW, Mach, 8-32 X .375, Phil
CONNECTOR, Panel
SCREW, Mach, 6-32 X .375, Phil
NUT, Mach, 6-32
SWITCH, Circuit Breaker, 15 Amp
SCREW, Mach, 10-32 X .375, Pan Head, Phil
MANIFOLD ASSEMBLY (see IPB Figure 7)
FILTER
NUT, Mach, 8-32
WASHER, Lock, #8
COVER, Services In
PLATE, Connectors
FASTENER, Lock Assembly
FASTENER, 3M, JACK SCR ASSY
CONNECTOR, In-Line Coupler
CABLE, Comm
Qty.
REF
1
8
1
2
2
1
2
REF
1
2
2
1
1
2
1
1
1
113
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 6 (continued)
SERVICES-IN ASSEMBLY (TWO DOOR)
114
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
SERVICES-IN ASSEMBLY (TWO DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Part Number
78-8078-6989-2
78-8078-3958-0
26-1002-5277-9
26-1011-7365-1
26-1002-5345-4
26-1001-7983-2
26-1011-7372-7
26-1002-2929-8
78-8078-6988-4
26-1011-8072-2
26-1000-6654-2
26-1004-2053-3
78-8078-5349-0
78-8078-8392-7
26-1011-7370-1
80-6100-6110-5
26-1014-7078-4
78-8078-8615-1
© 3M 1995 - 2005
Description
SERVICES IN ASSEMBLY - TWO DOOR
BRACKET SERVICE ASSEMBLY
SCREW, Mach, 8-32 X .375, Phil
CONNECTOR, Panel
SCREW, Mach, 6-32 X .375, Phil
NUT, Mach, 6-32
SWITCH, Circuit Breaker, 15 Amp
SCREW, Mach, 10-32 X .375, Pan Head, Phil
MANIFOLD ASSEMBLY – 8XL, 2 DOOR STERILIZER
FILTER
NUT, Mach, 8-32
WASHER, Lock, #8
COVER, Services In
PLATE, Connectors
FASTENER, Lock Assembly
FASTENER, 3M, JACK SCR ASSY
CONNECTOR, In-Line Coupler
CABLE, Comm
Qty.
REF
1
8
1
2
2
1
2
REF
1
2
2
1
1
2
1
1
1
115
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 7
MANIFOLD ASSEMBLY (ONE DOOR)
116
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
MANIFOLD ASSEMBLY (ONE DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Part Number
78-8078-5154-4
12-2376-9410-9
26-1004-2500-3
26-1004-0660-7
12-2376-3296-8
26-1005-4039-7
78-8062-9784-8
78-8078-5583-4
78-8678-2261-7
26-1005-7377-8
12-2376-9141-0
26-1005-7651-6
78-8032-5447-9
26-1011-6867-7
26-1011-6866-9
26-1000-8487-5
78-8078-7876-0
26-1004-0661-5
26-1011-6993-1
12-2376-9143-6
26-1003-2994-0
26-1004-5970-5
© 3M 1995 - 2005
Description
MANIFOLD ASSEMBLY
MANIFOLD
FITTING, 1/4 IPS X 3/8, Male-Female
FITTING, 90 Degree Street Elbow
TRANSDUCER, Vacuum Venturi
FITTING, Tube Connector, 1/8 NPT X ¼ OD
TUBE, Air
TUBE, Vacuum
TUBING, .250 OD X .040 Wall
FITTING, Tube Connector, 1/8 NPT X ¼ OD
SOLENOID VALVE, PUMP
FITTING, 90 Degree, 1/8 NPT, Male
INSTRUMENT, gauge
FITTING, Cross, 1/8 NPTF, Female
Sensor, Pressure, SPST, 1 Amp
FITTING, 1/8 NPT
REGULATOR – AIR
FITTING, Nipple, ¼ NPT X 1/8
FITTING, 1/8 NPTF X .27
SOLENOID VALVE, DOOR INTERLOCK
TAPE, .250 Wide, Teflon (not shown)
FITTING, Adapter, ¼ Female NPT X 1/8 Male
Qty.
REF
1
1
1
1
2
1
1
AR
2
1
2
1
1
1
1
1
1
1
1
AR
1
117
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 7 (continued)
MANIFOLD ASSEMBLY (TWO DOOR)
118
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
MANIFOLD ASSEMBLY (TWO DOOR)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
Part Number
78-8078-6988-4
12-2376-9410-9
26-1004-2500-3
26-1004-0660-7
12-2376-3296-8
26-1005-4039-7
78-8062-9784-8
78-8078-5583-4
78-8678-2261-7
26-1005-7377-8
12-2376-9141-0
26-1005-7651-6
78-8032-5447-9
26-1011-6867-7
26-1011-6866-9
26-1000-8487-5
78-8078-7876-0
26-1004-0661-5
26-1006-2508-1
12-2376-9143-6
26-1003-2994-0
26-1004-5970-5
12-7995-0202-1
© 3M 1995 - 2005
Description
MANIFOLD ASSEMBLY – 8XL, 2 DOOR STERILIZER
MANIFOLD
FITTING, 1/4 IPS X 3/8, Male-Female
FITTING, 90 Degree Street Elbow
TRANSDUCER, Vacuum Venturi
FITTING, Tube Connector, 1/8 NPT X ¼ OD
TUBE, Air
TUBE, Vacuum
TUBING, .250 OD X .040 Wall
FITTING, Tube Connector, 1/8 NPT X ¼ OD
SOLENOID VALVE, PUMP
FITTING, 90 Degree, 1/8 NPT, Male
INSTRUMENT, gauge
FITTING, Cross, 1/8 NPTF, Female
Sensor, Pressure, SPST, 1 Amp
FITTING, 1/8 NPT
REGULATOR - AIR
FITTING, Nipple, ¼ NPT X 1/8
FITTING, 1/8 NPT X .30
SOLENOID VALVE, DOOR INTERLOCK
TAPE, .250 Wide, Teflon (not shown)
FITTING, Adapter, ¼ Female NPT X 1/8 Male
FITTING, TEE, 1/8 NPTF X 1/8
Qty.
REF
1
1
1
1
2
1
1
AR
3
1
1
1
1
1
3
1
1
1
1
AR
1
1
119
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 8
DISPLAY PANEL ASSEMBLY (LOAD SIDE)
120
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
DISPLAY PANEL ASSEMBLY (LOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
Part Number
78-8078-3973-9
78-8078-3963-0
78-8078-7580-8
78-8078-6081-8
78-8078-4619-7
26-1001-7983-2
26-1011 7384-2
78-8078-5396-1
78-8078-3988-7
78-8078-8870-2
© 3M 1995 - 2005
Description
DISPLAY PANEL ASSEMBLY
PANEL, Display
PANEL, Membrane
LCD DISPLAY ASSEMBLY
BOARD, Keypad/Display Interface Assembly
NUT, Mach, 6-32
SPACER, Nylon
INSULATOR, LCD Screen
SPACER, Display, Board
CABLE – DISPLAY, SHIELDED
Qty.
REF
1
1
1
1
4
4
1
4
1
121
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 8 (continued)
DISPLAY PANEL ASSEMBLY (UNLOAD SIDE)
122
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
DISPLAY PANEL ASSEMBLY (UNLOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
10
Part Number
78-8078-5893-7
78-8078-4032-3
78-8078-7580-8
78-8078-3988-7
78-8078-6081-8
26-1011-7384-2
78-8078-4619-7
26-1001-7893-2
78-8078-5396-1
78-8078-7181-5
78-8078-8871-0
© 3M 1995 - 2005
Description
DISPLAY PANEL ASSEMBLY – UNLOAD SIDE
PANEL, Display, Unload Side
PANEL, Membrane
SPACER, DISPLAY BOARD
LCD DISPLAY ASSEMBLY
SPACER, 6-32 THD .50 H NYL
ASSY, KEYPAD/DISPLAY INTERFACE BOARD
NUT, Mach, 6-32
INSULATOR, LCD SCREEN
DECAL – HEALTH AND SAFETY
CABLE DISPLAY, SHIELDED
Qty.
REF
1
1
4
1
4
1
4
1
1
1
123
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 9
WATER TANK ASSEMBLY
124
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
WATER TANK ASSEMBLY
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
Part Number
78-8078-3983-8
78-8078-6287-1
26-1011-6135-9
78-8033-0846-5
26-1011-6948-5
78-8078-6068-5
26-1011-6949-3
26-1011-8040-9
26-1000-6403-5
26-1011-6947-7
26-1011-6992-3
78-8078-5148-6
© 3M 1995 - 2005
Description
WATER TANK ASSEMBLY
ASPIRATOR, ½ inch, 9 liter (2 gal.)
FITTING, Plastic, tank, ¾ inch Internal
ADHESIVE, Silicon Sealer
FITTING, Plastic, Tank, ¼ inch Internal
WATER LEVEL SENSOR ASSEMBLY
FITTING, Adapter, Plastic
CLAMP, Hose, 7/16 to 13/16 Hose
TUBING, Plastic, .500 ID X .750 OD
FITTING, Adapter, Plastic
FITTING, Connector, ¼
TUBR, Water
Qty.
REF
1
1
AR
2
1
2
5
AR
1
1
REF
125
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 10
CHAMBER ASSEMBLY (ONE DOOR)
126
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
CHAMBER ASSEMBLY (ONE DOOR)
Item No.
0
1
2
3
4
5
6
7
8
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
Part Number
78-8078-3969-7
78-8078-9381-9
26-1002-7761-0
26-1005-7377-8
12-2376-3895-7
12-2376-9420-8
78-8078-3989-5
78-8078-5148-6
26-1005-8600-2
78-8078-0332-1
12-2376-0602-0
78-8933-1006-6
78-8078-3285-8
78-8078-4045-5
12-7997-8082-5
26-1004-2053-3
78-8078-3288-2
78-8078-3503-4
12-7997-9965-0
26-1011-7276-0
26-1002-5297-7
78-8078-3144-7
78-8033-0846-5
78-8078-3316-1
26-1011-7386-7
78-8078-3380-7
78-8078-5168-4
26-1000-6654-2
78-8678-2261-7
78-8078-3324-5
78-8078-3379-9
18-3402-7075-2
12-2376-3486-5
26-1004-2078-0
78-8078-3968-9
78-8078-3381-5
12-7997-0813-1
26-1011-7860-1
12-2376-0490-0
78-8005-0972-7
70-8000-5374-9
© 3M 1995 - 2005
Description
CHAMBER ASSEMBLY-SINGLE DOOR
CONTROLLER BRACKET ASSEMBLY (see IPB Figure 15)
SCREW, Mach, 10-32 X .625, Pan Head, Phil
FITTING, 90 degree Elbow
FLEX HOSE, 15 inch, with Coupling
PRESSURE SENSOR, with Connector
HEATSINK ASSEMBLY (see IPB Figure 12)
TUBE, Water
FILTER, Mechanical
ASSEMBLY, PUNCTURE CYLINDER/SWITCH
O-RING
LUBRICANT, Silicone
DOOR CATCH
SUPPORT, Cylinder Rod
SCREW, Cap, 5/16-18 X .875, Hex Socket
WASHER, Lock, #8
PLATE, Guide
CAP-DOOR Catch
SCREW, Cap, 10-32 X .625, Socket, Hex
LATCH, Strike Plate
SCREW, Mach, 4-40 X .312, Phil
GASKET, Door
ADHESIVE, Super Silicon Sealer
CHAMBER, Heater Assembly (see IPB Figure 11)
PIN, Dowel, .2501 X 1.00
ASSEMBLY, PLUNGER
SPRING, Plunger
NUT, Mach, 8-32
TUBING, .250 OD X .040 Wall
SEAL, Cartridge
GUIDE, Cartridge
SCREW, Set, 8-32 X.125, Cup Point, Hex, Socket
BOLT, Door
Washer, Plain, .500 ID X 1.00 OD X .031
DOOR ASSEMBLY, Load Side (see IPB Figure 13)
BLOCK, Cartridge Stop
SCREW, Cap, ¼-28 X 1.00, Hex, Socket
PLUNGER
SPRING, Compression
WASHER, Plain, ½ inch
NUT, Jam, ½-13, Hex
Qty.
REF
REF
6
1
1
1
1
1
1
1
1
AR
1
1
2
1
1
1
2
1
1
2
AR
REF
1
1
1
1
AR
1
1
1
6
6
REF
1
1
2
6
12
6
127
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
CHAMBER ASSEMBLY (ONE DOOR) (Continued)
Item No.
44
45
46
47
48
49
50
51
52
53
55
56
57
58
59
60
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
128
Part Number
26-1001-2646-0
26-1011-7274-5
78-8078-5709-5
78-8078-5708-7
78-8033-0822-6
26-1011-7270-3
78-8078-4069-5
78-8078-4073-7
26-1004-0660-7
78-8078-5583-4
12-2376-9138-6
26-1005-8075-7
26-1017-0513-0
26-1005-4663-4
12-2376-3912-0
26-1003-2994-0
26-1017-0590-8
78-8078-4003-4
78-8078-4041-4
26-1002-5214-2
78-8078-2916-9
78-8078-4070-3
12-7997-5279-0
12-7995-0202-1
78-8078-4071-1
78-8032-8443-5
78-8078-4865-6
26-1002-5205-0
12-2376-3876-7
78-8078-3967-1
26-1004-2619-1
26-1005-7360-4
12-2376-9407-5
26-1002-5319-9
26-1002-5229-0
78-8078-4072-9
12-7997-7015-6
26-1011-7271-1
26-1005-6122-9
78-8078-6778-9
26-1011-8169-6
78-8003-9774-3
12-7996-2621-8
78-8007-2369-0
26-1009-7463-8
Description
NUT, Mach, ½-13
FITTING, Plug, 3/8 Male
CHAMBER INTERIOR TEMPERATURE SENSOR ASSEMBLY
TEMPERATURE SENSOR ASSEMBLY
LUBRICANT
FITTING, Adapter, ¼ Reducing
HUMIDITY SENSOR ASSEMBLY
TUBE, Humidity Sensor
FITTING, 90 degree Street Elbow, ¼
TUBE, Vacuum
SOLENOID, Valve, Vent
FITTING, ¾ Pipe, Hex Head
FITTING, ¼ Tee
FITTING, 3/8 X ¼ Reducer
VALVE, Check
TAPE
FITTING, Pipe Nipple, ¼ NPT
BACKPLATE ASSEMBLY
THERMOSTAT, Limit
SCREW, Mach, 6-32 X .250, Pan Head, Phil
BUSHING, Modified
TUBE, Tee to Chamber
FITTING, Union, ¼ Tube OD
FITTING, Tee, 1/8 NPTF
TUBE, Heatsink to Tee
FITTING, 3/8 Bushing
CARTRIDGE SENSOR ASSEMBLY
SCREW, Mach, 4-40 X.312, Pan Head, Phil
NOZZLE, Steam
DOOR LOCK ASSEMBLY (see IPB Figure 3)
SCREW, Mach, 10-32 X .500
SWITCH, Door, Interlock
BRACKET, Door Switch
SCREW, Mach, 8-32 X .500, Phil
SCREW, Mach, 8-32 X .625, Pan Head, Phil
TUBE, Chamber to Heatsink
FITTING, Connector, 1/8 NPT X ¼ Tube OD
FITTING, ½ Bushing, Reducer
WASHER, Plain, .120 ID X .280 OD X .050 Thick
VALVE, Solenoid
FITTING, Nipple
FITTING, Bushing, 1/8 NPTF X ¼
PLUG
LOCKTITE
ADHESIVE, PIPE SEALANT
Qty.
6
1
1
2
AR
1
1
1
5
1
2
1
2
2
1
AR
1
1
1
2
2
2
1
1
1
1
1
2
2
REF
2
1
1
2
1
1
3
1
2
1
1
1
1
AR
AR
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 10 (continued)
CHAMBER ASSEMBLY (TWO DOOR)
© 3M 1995 - 2005
129
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
CHAMBER ASSEMBLY (TWO DOOR)
Item No.
0
1
2
3
4
5
6
7
8
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
130
Part Number
78-8078-4033-1
78-8078-9381-9
26-1002-7761-0
26-1005-7377-8
12-2376-3895-7
12-2376-9420-8
78-8078-3989-5
78-8078-5148-6
26-1005-8600-2
78-8078-0332-1
12-2376-0602-0
78-8933-1006-6
78-8078-3285-8
78-8078-4045-5
12-7997-8082-5
26-1004-2053-3
78-8078-3288-2
78-8078-3503-4
12-7997-9965-0
26-1011-7276-0
26-1002-5297-7
78-8078-3144-7
78-8033-0846-5
78-8078-3316-1
26-1011-7386-7
78-8078-3380-7
78-8078-5168-4
26-1000-6654-2
78-8678-2261-7
78-8078-3324-5
78-8078-3379-9
18-3402-7075-2
12-2376-3486-5
26-1004-2078-0
78-8078-3968-9
78-8078-3381-5
12-7997-0813-1
26-1011-7860-1
12-2376-0490-0
78-8005-0972-7
70-8000-5374-9
Description
CHAMBER ASSEMBLY-TWO DOOR
CONTROLLER BRACKET ASSEMBLY (see IPB Figure 15)
SCREW, Mach, 10-32 X .625, Pan Head, Phil
FITTING, 90 degree Elbow
FLEX HOSE, 15 inch, with Coupling
PRESSURE SENSOR, with Connector
HEATSINK ASSEMBLY (see IPB Figure 12)
TUBE, Water
FILTER, Mechanical
ASSEMBLY, PUNCTURE CYLINDER/SWITCH
O-RING
LUBRICANT, Silicone
DOOR CATCH
SUPPORT, Cylinder Rod
SCREW, Cap, 5/16-18 X .875, Hex Socket
WASHER, Lock, #8
PLATE, Guide
CAP-DOOR, Catch
SCREW, Cap, 10-32 X .625, Socket, Hex
LATCH, Strike Plate
SCREW, Mach, 4-40 X .312, Phil
GASKET, Door
ADHESIVE, Super Silicon Sealer
CHAMBER, Heater Assembly (see IPB Figure 11)
PIN, Dowel, .2501 X 1.00
ASSEMBLY, PLUNGER
SPRING, Plunger
NUT, Mach, 8-32
TUBING, .250 OD X .040 Wall
SEAL, Cartridge
GUIDE, Cartridge
SCREW, Set, 8-32 X.125, Cup Point, Hex, Socket
BOLT, Door
Washer, Plain, .500 ID X 1.00 OD X .031
DOOR ASSEMBLY, Load Side (see IPB Figure 13)
BLOCK, Cartridge Stop
SCREW, Cap, ¼-28 X 1.00, Hex, Socket
PLUNGER
SPRING, Compression
WASHER, Plain, ½ inch
NUT, Jam, ½-13, Hex
Qty.
REF
REF
6
1
1
1
1
1
1
1
1
AR
1
1
4
1
2
1
4
1
1
2
AR
REF
1
1
1
1
AR
1
1
1
4
4
REF
1
1
2
4
8
4
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
CHAMBER ASSEMBLY (TWO DOOR) (Continued)
Item No.
44
45
46
47
48
49
50
51
52
53
55
56
57
58
59
60
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
Part Number
26-1001-2646-0
26-1011-7274-5
78-8078-5709-5
78-8078-5708-7
78-8033-0822-6
26-1011-7270-3
78-8078-4069-5
78-8078-4073-7
26-1004-0660-7
78-8078-5583-4
12-2376-9138-6
26-1005-8075-7
26-1017-0513-0
26-1005-4663-4
12-2376-3912-0
26-1003-2994-0
26-1017-0590-8
78-8078-4028-1
78-8078-4041-4
26-1002-5214-2
78-8078-2916-9
78-8078-4070-3
12-7997-5279-0
12-7995-0202-1
78-8078-4071-1
78-8032-8443-5
78-8078-4865-6
26-1002-5205-0
12-2376-3876-7
78-8078-3967-1
26-1004-2619-1
26-1005-7360-4
12-2376-9407-5
26-1002-5319-9
26-1002-5229-0
78-8078-4072-9
12-7997-7015-6
26-1011-7271-1
26-1005-6122-9
78-8078-6778-9
26-1011-8169-6
78-8003-9774-3
12-7996-2621-8
78-8007-2369-0
26-1009-7463-8
78-8078-4029-9
78-8078-4046-3
78-8078-4019-0
78-8078-4020-8
© 3M 1995 - 2005
Description
NUT, Mach, ½-13
FITTING, Plug, 3/8 Male
CHAMBER INTERIOR TEMPERATURE SENSOR ASSEMBLY
TEMPERATURE SENSOR ASSEMBLY
LUBRICANT
FITTING, Adapter, ¼ Reducing
HUMIDITY SENSOR ASSEMBLY
TUBE, Humidity Sensor
FITTING, 90 degree Street Elbow, ¼
TUBE, Vacuum
SOLENOID, Valve, Vent
FITTING, ¾ Pipe, Hex Head
FITTING, ¼ Tee
FITTING, 3/8 X ¼ Reducer
VALVE, Check
TAPE
FITTING, ¼ NPT
DOOR ASSEMBLY, UNLOAD SIDE
THERMOSTAT, Limit
SCREW, Mach, 6-32 X .250, Pan Head, Phil
BUSHING, Modified
TUBE, Tee to Chamber
FITTING, Union, ¼ Tube OD
FITTING, Tee
TUBE, Heatsink to Tee
FITTING, 3/8 Bushing
CARTRIDGE SENSOR ASSEMBLY
SCREW, Mach, 4-40 X.312, Pan Head, Phil
NOZZLE, Steam
DOOR LOCK ASSEMBLY
SCREW, Mach, 10-32 X .500
SWITCH, Door, Interlock
BRACKET, Door Switch
SCREW, Mach, 8-32 X .500, Phil
SCREW, Mach, 8-32 X .625, Pan Head, Phil
TUBE, Chamber to Heatsink
FITTING, Connector, 1/8 NPT X ¼ Tube OD
FITTING, ½ Bushing, Reducer
WASHER, Plain, .120 ID X .280 OD X .050 Thick
VALVE, Solenoid
FITTING, Nipple
FITTING, Bushing, 1/8 NPTF X ¼
PLUG
LOCKTITE
ADHESIVE, PST Pipe Sealant
DOOR LOCK ASSEMBLY, UNLOAD SIDE
SUPPORT-CYLINDER ROD, UNLOAD SIDE
CAP-DOOR CATCH, UNLOAD SIDE
DOOR CATCH, UNLOAD SIDE
Qty.
4
1
1
2
AR
1
1
1
5
1
2
1
2
2
1
AR
1
1
1
2
2
2
1
1
1
1
1
2
2
1
4
2
2
4
1
1
3
1
2
1
1
1
1
AR
AR
1
1
1
1
131
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 11
CHAMBER-HEATER ASSEMBLY
132
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
CHAMBER-HEATER ASSEMBLY
Item No.
0
1
2
3
4
5
6
7
8
9
Part Number
78-8078-3316-1
78-8078-2945-8
78-8078-3319-5
78-8078-3992-9
78-8078-3993-7
78-8078-3394-5
78-8078-3395-2
78-8005-1710-0
78-8078-3996-0
78-8033-0846-5
© 3M 1995 - 2005
Description
CHAMBER-HEATER ASSEMBLY
CHAMBER
HEATER AND BLANKET ASSEMBLY
PANEL #1, Chamber Insulation
PANEL #2, Chamber Insulation
PANEL #3, Chamber Insulation
PANEL #4, Chamber Insulation
TAPE, Glass Cloth, 1.00 Wide
INSULATION, Cartridge Chamber
ADHESIVE
Qty.
REF
REF
12
1
1
1
1
AR
1
AR
133
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 12
HEATSINK ASSEMBLY
134
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
HEATSINK ASSEMBLY
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
Part Number
78-8078-3989-5
78-8078-5791-3
78-8078-5708-7
78-8033-0822-6
12-2376-9373-9
78-8078-5223-7
26-1003-3223-3
78-8055-6013-9
26-1004-0661-5
12-2376-9138-6
26-1004-0660-7
78-8055-5906-5
78-8055-5905-7
78-8055-6012-1
12-7995-8546-3
78-8055-6025-3
26-1011-6988-1
78-8078-5148-6
78-8078-6778-9
12-7997-5507-4
26-1006-2508-1
26-1003-2994-0
78-8005-1710-0
26-1000-8487-5
© 3M 1995 - 2005
Description
HEATSINK ASSEMBLY
HEATSINK 9With Plugs)
TEMPERATURE SENSOR ASSEMBLY
LUBRICANT
ROD HEATER ASSEMBLY
THERMOSTAT, Hi Limit, Heatsink
SCREW, Mach, 4-40 X .188, Pan Head, Phil
BRACKET, Heatsink
FTTING, Nipple, ¼ NPT X 1/8
SOLENOID, Valve, Vent
FITTING, 90 degree, Street Elbow, ¼
FILTER ADAPTER, Female
FILTER
INSULATION, Heatsink
FITTING, Elbow, ¼ Tube OD X 1/8
TUBE, Cartridge to Heatsink
FITTING, Tee, 1/8 Tube X 1/8 Male
TUBE, Water
SOLENOID, Valve, 3-Way
FITTING, for .125 OD Tube
FITTING, 1/8 NPT X .30
TAPE, .250 Wide
TAPE, 1.00 Wide
FITTING, 1/8 NPT
Qty.
REF
1
1
AR
1
2
4
1
2
1
1
1
1
1
2
REF
1
REF
1
1
1
AR
AR
1
135
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 13
DOOR ASSEMBLY (LOAD SIDE)
136
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
DOOR ASSEMBLY (LOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
Part Number
78-8078-3968-9
26-1011-6157-3
78-8078-3504-2
26-1011-6136-7
78-8078-3982-0
78-8078-3289-0
80-8078-3286-6
26-1012-1550-2
78-8055-5959-5
78-8078-3292-4
78-8078-3319-5
78-8078-3287-4
78-8078-3997-8
26-1011-7387-5
26-1001-6816-5
78-8078-3294-0
12-2376-9379-6
26-1004-3140-7
26-1003-4963-3
26-1002-5277-9
12-2376-5848-4
70-8000-2823-8
70-7023-5420-3
26-1005-4049-6
26-1002-5353-8
78-8078-3290-8
78-8078-3291-6
12-7995-5429-5
26-1004-2053-3
78-8078-5086-8
78-8007-2369-0
© 3M 1995 - 2005
Description
DOOR ASSEMBLY, LOAD SIDE
SCREW, Cap, 3/8-16 X 1.5, Socket, Hex
ADAPTER, Door Lock
WASHER, Spring, Wave, .62 ID X .87 OD X .13
BUSHING, Altered, Door
WASHER
DOOR CAM
SCREW, Mach, ¼-20 X .50 HEX SKT
Y HARNESS DOOR
PLATE, Door
HEATER BLANKET ASSEMBLY
ADAPTER, Door Handle
INSULATION, Door
HANDLE, Non-Locking
SCREW, Cap, 10-32 X .750, Hex, Socket
DOOR PANEL
CORD, Door Heater
BUSHING, Strain Relief
TERMINAL
SCREW, Mach, 8-32 X .375, Pan Head Phil
HINGE, Door
SCREW, Cap, 10-32 X 1.00, Hex Socket
NUT, Mach, 10-32, Hex
WASHER, Lock, #10
SCREW, Mach, 8-32 X .625, Phil
BUSHING
STRAP
SCREW, Cap, 5/16-18 X 1.375, Hex Socket
WASHER, Lock, #8
DECAL, Operating Instructions
LOCKTITE
Qty.
REF
1
1
1
1
1
1
2
1
1
2
1
2
1
2
1
1
1
1
1
1
11
24
13
12
2
1
2
1
REF
AR
137
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 13 (continued)
DOOR ASSEMBLY (UNLOAD SIDE)
138
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
DOOR ASSEMBLY (UNLOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
Part Number
78-8078-4028-1
26-1011-6157-3
78-8078-4016-6
26-1011-6136-7
78-8078-3982-0
78-8078-3289-0
78-8078-4018-2
26-1012-1550-2
78-8055-5959-5
78-8078-4022-4
78-8078-3319-5
78-8078-3287-4
78-8078-3997-8
26-1011-7387-5
26-1001-6816-5
78-8078-4021-6
12-2376-9379-6
26-1004-3140-7
26-1003-4963-3
26-1002-5277-9
12-2376-5848-4
70-8000-2823-8
70-7023-5420-3
26-1005-4049-6
26-1002-5353-8
78-8078-3290-8
78-8078-3291-6
12-7995-5429-5
26-1004-2053-3
78-8078-5164-3
78-8007-2369-0
© 3M 1995 - 2005
Description
DOOR ASSEMBLY, UNLOAD SIDE
SCREW, Cap, 3/8-16 X 1.5, Socket, Hex
ADAPTER, Door Lock, Unload Side
WASHER, Spring, Wave, .62 ID X .87 OD X .13
BUSHING, Altered, Door
WASHER
DOOR CAM – UNLOAD SIDE
SCREW, Mach, ¼-20 X .50, HEX SKT
Y HARNESS DOOR
PLATE, Door, Unload Side
HEATER BLANKET ASSEMBLY
ADAPTER, Door Handle
INSULATION, Door
HANDLE, Non-Locking
SCREW, Cap, 10-32 X .750, Hex, Socket
DOOR PANEL – UNLOAD SIDE
CORD, Door Heater
BUSHING, Strain Relief
TERMINAL
SCREW, Mach, 8-32 X .375, Pan Head Phil
HINGE, Door
SCREW, Cap, 10-32 X 1.00, Hex Socket
NUT, Mach, 10-32, Hex
WASHER, Lock, #10
SCREW, Mach, 8-32 X .625, Phil
BUSHING
STRAP
SCREW, Cap, 5/16-18 X 1.375, Hex Socket
WASHER, Lock, #8
DECAL, Operating Instructions, Unload Side
LOCKTITE
Qty.
REF
1
1
1
1
1
1
2
1
1
2
1
2
1
2
1
1
1
1
1
1
11
24
13
12
2
1
2
1
REF
AR
139
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 14
LOCAL EXHAUST ASSEMBLY (LOAD SIDE)
140
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
LOCAL EXHAUST ASSEMBLY (LOAD SIDE)
Item No.
0
1
2
3
4
5
6
7
8
9
10
11
Part Number
78-8078-5863-0
78-8078-5870-5
12-2376-5166-1
12-2376-5167-9
26-1001-7289-4
12-2376-6210-6
78-8063-0467-7
26-1002-5214-2
12-2376-9368-9
26-1002-5268-8
78-8164-1894-7
78-8078-5861-4
© 3M 1995 - 2005
Description
LOCAL EXHAUST ASSEMBLY, Load side
PANEL, Upper Front, Local Exhaust, Load Side
LATCH, Slide
SPRING, Return
SCREW, Cap, 8-32 X .25, Hex
LABEL, Door Release
AIR FLOW SENSOR ASSEMBLY
SCREW, 6-32 X .25, Pan Head, Phil
BRACKET, Air Flow Switch
SCREW, Mach, 6-32 X 3/8, Pan Head, Phil
BUSHING
LATCH, Catch
Qty.
REF
1
1
1
1
1
1
1
1
1
1
1
141
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 14 (continued)
LOCAL EXHAUST ASSEMBLY (UNLOAD SIDE)
142
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
LOCAL EXHAUST ASSEMBLY (UNLOAD SIDE)
Item No.
0
1
2
3
4
5
Part Number
78-8078-6273-1
78-8078-6295-4
12-2376-5166-1
12-2376-5167-9
26-1001-7289-4
12-2376-6210-6
© 3M 1995 - 2005
Description
EXHAUST HOOD ASSEMBLY, Unload side
PANEL, Upper Front, Local Exhaust, Unload side
LATCH, Slide
SPRING, Return
SCREW, Cap, 8-32 X .25, Hex
LABEL, Door Release
Qty.
REF
1
1
1
1
1
143
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
IPB Figure 15
CONTROLLER BRACKET ASSEMBLY
144
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ILLUSTRATED PARTS BREAKDOWN
CONTROLLER BRACKET ASSEMBLY
Item No.
0
1
2
3
4
5
6
7
8
9
10
Part Number
78-8078-9381-9
78-8078-6517-1
78-8078-5091-8
26-1014-7185-7
26-1002-5216-7
26-1006-0179-3
26-1001-2475-4
78-8078-6725-0
26-1005-7876-9
26-1002-5208-4
78-8078-9382-7
© 3M 1995 – 2005
Description
CONTROLLER BRACKET ASSEMBLY
CONTROLLER ASSEMBLY (without PROM) (see IPB Fig 15)
PROM, Programmed, 8XL
GUIDE
SCREW, Mach, 6-32 X .375, Pan Head, Phil
RELAY, Solid State, 240 VAC 8 AMP
TIE WRAP
WIRING HARNESS, Low Voltage
WASHER, Lock
SCREW, Mach, 4-40 X .500, Pan Head Phil
BRACKET, Controller
Qty.
REF
1
1
2
7
3
AR
REF
1
2
1
145
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
This page intentionally left blank.
146
© 3M 1995 - 2005
T1
© 3M 1995 - 2005
A
AT
A
IN5401
CR3
IN4003
CR1
IN5401
CR4
E1
IN4003
CR2
+
10,000 uf
C1
16V
Q2
2N2219
+
100 uf
C2
35V
2
R2
2.2K
ILIM
R1
500 M
1
3
2N6569
Q1
IN4003
CR5
6
1
V.ADJ.
2K
2
1
0
2
R6
11
.12m
2W
R5
R5
47m
3
R7
750 m
U1
uA723
12
5
R8
3.6K
3
7
13
4
R10
47m
C3
.0033uf
100V
R9
2.2K
C4
1 uf
50V
R11
6.8m
+
S0303LS3
SCR1
5.6 V
400MV
CR6
C5
220 uf
16V
+
IN4003
CR7
R13
6.8m
R14
220 m
R12
6.8m
-S
-OUT
5V D.C.
UNREGULATED
5V D.C.
REGULATED
+OUT
+S
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
ELECTRICAL SCHEMATICS
5 Volt Power Supply
147
148
A
BT
A
R711A
CR4
C6
.1 uf
100V
56.8
VAC
IN4003
CR1
*
24 Volt Power Supply
C1
13,000 uf
50V
2.4K
R5
2N6569
Q3
1
3
R10
5.1K
R4
2.4K
500 m
R6 I.LIM
2
2N3055
Q1
IN4003
CR3
IN4448
CR6
CW
2N3055
Q2
NOT USED IN SOME ASSEMBLIES
IN4003
CR2
R9*
2.4K
R8
R19*
33m
2W
R7
2
10
R18*
33m
2W
R3
6
5
R20*
uA723
U1
11
7
12
10 uf
220 m C3
25V
R13
3
R1
6.2K
4
13
R2
6.2K
C2
470pf
1KV
C4
IN974B
CR5
100 uf
100V
+
1
3
R14
2.4K
R12
4.7K
2
R11 2K
V ADJ
CW
+
C4
330 uf
35V
IN5401
CR7
R16
2.4K
R17
6.8m
5V D.C.
REGULATED
R12
6.8m
-S
-OUT
24 VDC
+OUT
+S
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
ILLUSTRATED PARTS BREAKDOWN
© 3M 1995 - 2005
2
ThermoStat
ThermoStat
220-240 VAC Input
Mounted on Heatsink (105 C)
Heatsink Heater
Middle Chamber Heater Blankets
1
Mounted On Chamber Wall (65C)
ThermoStat
1
2
2
End Chamber Heater Blankets
High Voltage
Low Voltage
3
4
5
Power
Entry
Module
Pin2
Pin4
Pin3
Pin1
SSR3
Pin4
Pin3
Pin4
Pin3
Pin2
SSR1
Pin1
Pin2
SSR2
Pin1
AMP 583617-1
6
4
9
8
55
30
55
61
55
28
Line
RFI
Filter
Load
11
10
Power
Switch
2
1
Gnd
Out
Vcc
20
22
31
52
31
41
51
50
53
52
27
41
20
49
48
20
24
23
20
22
68
70
69
67
14
13
12
11
10
9
8
7
6
5
4
3
2
1
J4
J7
1
2
3
4
RJ11
HD24-4.8-A
Power One
International Series
24V Power Supply
+Out
+S
-S
-Out
4
J9
J2
3
72
44
43
42
54
55
39
38
57
21
56
58
34
56
35
36
33
64
47
56
46
61
28
30
77
32
63
78
26
37
25
59
29
82
Gnd
Out
Vcc
NO COM
NO COM
NO COM
NO COM
NO COM
NO COM
NO COM
NO COM
COM
NC
20
24
23
HB5 3/OVP-A
Power One
International Series
5V Power Supply
+Out
+S
-S
-Out
Pressure
Transducer
22
4
2
3
1/4 Amp
250 VAC
1
1
9
2
10
3
11
4
12
5
13
6
14
7
15
8
High Voltage
5
Abator Interface Port
Transformer
2
Red
Brown
Black
Chamber Pressure
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
A
B
C
D
E
F
H
J
K
L
M
N
P
R
S
T
U
V
W
X
Y
Z
1
J6
J8
1
2
GND 3
+15V
Signal GND
Puncture Pin Position
Abator Ready
AirFlow Detect Gnd
AirFlow Detect
AirFlow Detect; Power
H2O Detect; Gnd
H2O Detect
H2O Detect; Power
Load Door Handle
Load Door Latch
Load Door Shut
Cartridge Pos. Gnd
Cartridge Position
Cartridge Pos. Power
Compressed Air
Nitrogen Pressure
AbatorFailed
Unload Door Handle
Unload Door Latch
Unload Door Shut
Signal GND
8
15
7
14
6
13
5
12
4
11
3
10
2
9
1
2
5
DB15
J16
Extra I/O
Extra I/O
Extra 1 Sol
Extra 2 Sol
Extra 3 Sol
Frame Ground
Electric Pump Vent Sol.
Electric Pump Power
Nitrogen Solenoid
Middle Heater Output
End Heater Output
Heatsink Heater Output
Cycle Starting (Abator)
Vent Solenoid
RH Solenoids
Gas Coming (Abator)
Vacuum Pump Solenoid
Water Injection Solenoid
Load Door Solenoid
Unload Door Solenoid
Puncture Solenoid
/Printer Initialization
FLM
LU0
Backlight
LU1
GND
LU2
KEY0
LU3
KEY1
CL2
KEY2
CL1
KEY3
M
STOP
22
78
24
77
38
20
72
1
1
Transformer
8/10 Amp
250 VAC
2
Puncture Sim Sol. Output
Receive Data RXD
GND
Transmit Data TXD
FRAME GROUND
5XLe Controller Board
78-8078-6518-9
RecorderTemperature
VTC2
Heatsink Temperature
VTC4
End Temperature
VTC3
Middle Temperature
VTC2
Chamber Temperature
VTC4
GND
RH
LATCHPOS IS+5V
Vcc
GND
Vcc
+S
GND
+24V
1
J5
\3m\6519-7\reva\control.prj
External Audio
ELECTRICAL SCHEMATICS
Vcc
Audio
GND
1
2
3
AMP 583617-1
Top Layer
AMP 583617-1
Bottom Layer
Steri-Vac 8XL Gas Sterilizer/Aerator
3 Printer On
2 GND
1 /Printer Form Feed
Switch
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Printer
GND
/Printer Fault
Printer Error
Printer Busy
/Printer Ack
Printer Data7
Printer Data6
Printer Data5
Printer Data4
Printer Data3
Printer Data2
Printer Data1
Printer Data0
/Printer Data Strobe
8
15
7
14
6
13
5
12
4
11
3
10
2
9
1
Vcc
GND
GND
GND
+24V
+24V
/Printer Data Strobe
Printer Data0
Printer Data1
Printer Data2
Printer Data3
Printer Data4
Printer Data5
Printer Data6
Printer Data7
/Printer Ack
Printer Busy
Printer Error
/Printer Fault
GND
J1
5
4
3
2
1
+24V
GND
+S
Vcc
GND
STOP
M
KEY3
CL1
KEY2
CL2
KEY1
LU3
KEY0
LU2
GND
LU1
Backlight
LU0
FLM
J16
78-8078-4616-3
Interface Board
KeyPad/Display
Low Voltage
24
20
22
20
DB15
1
2
3
4
5
6
CN1
CN9
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Display/Keypad Interface Board
c:\3m\board\keypadb\keydisp.sch
24
20
20
20
22
22
82
Seiko IF5003-01B
Printer Interface Board
CN11
1
2
1
2
3
1
2
VPS
GND
PS
HS
GND
Stanley GMF32024ABTW
LCD Display Module
FLM
M
CL1
CL2
Display On
LU0
LU1
LU2
LU3
Vcc
GND
-24V
Contrast V
FG
AC_IN2
AC_IN1
FLM
M
CL1
CL2
Display On
LU0
LU1
LU2
LU3
Vcc
GND
-24V
Contrast V
FG
CN2
4
3
2
1
CN1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
CN1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
CN6
CN5
1
2
3
4
5
6
CN4
MPower
MPower
Sigma1
Sigma2
Sigma3
Sigma4
8
7
6
5
4
3
2
1
8
7
6
5
4
3
2
1
J2
LEDGND 9
Vcc 8
KEY0 7
KEY1 6
KEY2 5
KEY3 4
STOP 3
GND 2
NC 1
CN2
AC_OUT1 4
3
2
AC_OUT2 1
DIP1
DIP2
1
2
1
2
3
KeyPad
78-8078-3987-9
LED
HS
GND
VPS
GND
PS
MPower
MPower
Sigma1
Sigma2
Sigma3
Sigma4
Seiko LPT5347A-S576
Thermal Printer
1
2
3
4
5
6
KEY0
KEY1
KEY2
KEY3
STOP
149
Service and Repair
150
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
EX_IN2
EX_IN1
UDRSHUT
UDRLATCH
UDRHDL
ABATFAIL
N_PRESS
COMPAIR
CARTPOS
LDRSHUT
LDRLATCH
LDRHDL
ADDWATER
AIRFLOW
ABATRDY
PUNCTPIN
/PRTFF
PRTON
/PRTINIT
HEATSINK
AUDIO
END_HEAT
MID_HEAT
LATCHPOS
CLOSE_LR
OPEN_LR
PUNCTSOL
LD_SOL
WATER_SOL
N_SOL
EX1_SOL
VAC_SOL
DV_SOL
AV_SOL
GAS_SIG
RH_SOL
EP_POWER
EP_VALVE
UD_SOL
VENT_SOL
PUNCT_SIM
EX2_SOL
CYC_START
\3m\6519-7\revc\perph.sch
STOP
KEY3
KEY2
KEY1
KEY0
PRT_D[0..7]
TXD_CTL
RXD_CTL
TXD
RXD
D[0..7]
/PRTFF
/PRTACK
PRTBUSY
PRTERR
/PRTINIT
/PRTFAULT
/PRTSTB
/CSCS
/AUXCS
/KEYEN
/CLRKEY
/KEYCS
Perpherials
\3m\6519-7\revc\pcio.sch
ELECTRICAL SCHEMATICS
Steri-Vac 8XL Gas Sterilizer/Aerator
STOP
KEY3
KEY2
KEY1
KEY0
/PRTACK
PRTBUSY
PRTERR
/PRTINIT
/PRTFAULT
/PRTSTB
/CSCS
/AUXCS
/KEYEN
/CLRKEY
/KEYCS
LATCHPOS
TXD_CTL
RXD_CTL
TXD
RXD
EX_IN2
EX_IN1
UDRSHUT
UDRLATCH
UDRHDL
ABATFAIL
N_PRESS
COMPAIR
CARTPOS
LDRSHUT
LDRLATCH
LDRHDL
ADDWATER
AIRFLOW
ABATRDY
PUNCTPIN
/PRTFF
PRTON
HEATSINK
AUDIO
END_HEAT
MID_HEAT
CLOSE_LR
OPEN_LR
PUNCTSOL
LD_SOL
WATER_SOL
N_SOL
EX1_SOL
VAC_SOL
DV_SOL
AV_SOL
GAS_SIG
RH_SOL
EP_POWER
EP_VALVE
UD_SOL
VENT_SOL
PUNCT_SIM
EX2_SOL
CYC_START
/PRTACK
PRTBUSY
PRTERR
/PRTINIT
/PRTFAULT
/PRTSTB
/CSCS
/AUXCS
/KEYEN
/CLRKEY
/KEYCS
PRT_D[0..7]
TXD_CTL
RXD_CTL
TXD
RXD
D[0..7]
EX_IN2
EX_IN1
UDRSHUT
UDRLATCH
UDRHDL
ABATFAIL
N_PRESS
COMPAIR
CARTPOS
LDRSHUT
LDRLATCH
LDRHDL
ADDWATER
AIRFLOW
ABATRDY
PUNCTPIN
/PRTFF
PRTON
HEATSINK
AUDIO
END_HEAT
MID_HEAT
LATCHPOS
CLOSE_LR
OPEN_LR
PUNCTSOL
LD_SOL
WATER_SOL
N_SOL
EX1_SOL
VAC_SOL
DV_SOL
AV_SOL
GAS_SIG
RH_SOL
EP_POWER
EP_VALVE
UD_SOL
VENT_SOL
PUNCT_SIM
EX2_SOL
CYC_START
\3m\6519-7\revc\mcu.sch
/CLRKEY
/KEYEN
BACKLITE
DISPTMG
DISPON
DISPRS
XOUT
R//W
/FPGA_RST
E
FPGA_VIDEO
VA[11..13]
A[0..15]
LATCHPOS
CH07
EXA0
EXA1
RH
CHPRESS
EXA2
EXA3
VREF
SEL0
SEL1
SEL2
/CLRKEY
/KEYEN
BACKLITE
DISPTMG
DISPON
DISPRS
XOUT
R//W
/FPGA_RST
E
FPGA_VIDEO
VA[11..13]
A[0..15]
CH07
EXA0
EXA1
RH
CHPRESS
EXA2
EXA3
VREF
SEL0
SEL1
SEL2
D[0..7]
STOP
KEY3
KEY2
KEY1
KEY0
\3m\6519-7\revc\video.sch
/CLRKEY
/KEYEN
BACKLITE
DISPTMG
DISPON
DISPRS
XOUT
R//W
/FPGA_RST
E
FPGA_VIDEO
VA[11..13]
A[0..15]
LCTC Video Controller
\3m\6519-7\revc\analog.sch
LATCHPOS
CH07
EXA0
EXA1
RH
CHPRESS
EXA2
EXA3
VREF
SEL0
SEL1
SEL2
Process Control Inputs
78-8078-6519-7 Sheet 1 of 6
Schematic, Controller, Steri-Va
151
Service and Repair
152
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
J7
HEADER 5
Power Input
CH07
EXA0
LEAKPRES
RH
CHPRESS
EXA2
EXA3
VREF
RXD
TXD
RXD_CTL
BACKLITE
/KEYEN
DISPON
3
3
C32
0.1uF
C31
0.1uF
0.1uF
C33
0.1uF
C17
C16
0.1uF
C9
0.1uF
.01uF
C62
4.7K
R5
C3
2
1
C10
VSS
VSS
VSS
/XIRQ
/IRQ
EXTAL
NC
NC
NC
XTAL
PE0/AN0
PE1/AN1
PE2/AN2
PE3/AN3
PE4/AN4
PE5/AN5
PE6/AN6
PE7/AN7
PD0/RXD
PD1/TXD
PD2/MISO
PD3/MOSI
PD4/SCK
PD5//SS
PC0/DATA0
PC1/DATA1
PC2/DATA2
PC3/DATA3
PC4/DATA4
PC5/DATA5
PC6/DATA6
PC7/DATA7
PB0/ADDR8
PB1/ADDR9
PB2/ADDR10
PB3/ADDR11
PB4/ADDR12
PB5/ADDR13
PB6/ADDR14
PB7/ADDR15
C12
0.1uF
C34
0.1uF
C35
C28
0.1uF
C36
0.1uF
C27
0.1uF
C26
0.1uF
C20
0.1uF
0.1uF
C19
0.1uF
TP13
JP1
0.1uF
0.1uF
C18
0.1uF
C11
GND_ANALOG
Vcc
Jumper
C4
.01uF
GND
1uF
C13
0.1uF
C29
0.1uF
C21
0.1uF
0.1uF
C30
0.1uF
C14
Microcontroller
PA0/IC3
PA1/IC2
PA2/IC1
PA3/IC4/OC5/OC1
PA4/OC4/OC1
PA5/OC3/OC1
PA6/OC2/OC1
PA7/PAI/OC1
U2
MC68HC11K1CFN3
+24V
3
24
63
21
GND_SOL
1uF
C61
GND_FRAME
3
C8
C15
Vcc
R28
4.7K
54
67
20
R6
22
10M 25
MCU_EXTAL
Y1
GND_FRAME
22uF 35V
C2 3
0.1uF
0.1uF
42
41
40
39
38
37
36
35
CH07
EXA0
EXA1
RH
CHPRESS
EXA2
EXA3
VREF
GND_FRAME
22uF 35V
C60 3
C7
U18
T_CAP
U1
T_CAP
U16
T_CAP
C6
22pF
0.1uF
15M_1
15M_2
15M_3
U15
T_CAP
15M_1
U17
15M_4T_CAP
/IRQ
72
73
74
75
76
77
RXD
TXD
RXD_CTL
BACKLITE
/KEYEN
DISPON
68
55
56
57
58
59
60
61
62
D0
D1
D2
D3
D4
D5
D6
D7
MCU_XTAL
11
10
9
8
7
6
5
4
A8
A9
A10
A11
A12
A13
A14
A15
12 MHz
22pF
C5
1
84
83
82
81
80
79
78
/PRTACK
PRTBUSY
PRTERR
/PRTINIT
/PRTFAULT
/PRTFF
/PRTSTB
DISPTMG
Vcc
/RST 4
1
VCC
Vcc
2
MAX6314
IN
U11
MC34064
3
2
Vcc
45
34
43
44
70
71
/RST
R27
10K .1%
R26
1.37K .1%
Vcc
Vcc
XOUT
3
Ferrite
U49
C37
100uF
R35
4.7K
Vcc
FPGA_VIDEO2
/CLRKEY
FPGA_VIDEO1
TXD_CTL
SEL0
SEL1
SEL2
DISPRS
FPGA_DONE
4.4Volts
4.7K
R13
BANKA14
BANKA15
FPGA_VIDEO2
/CLRKEY
FPGA_VIDEO1
TXD_CTL
LATCHRES
LATCHSET
VA11
VA12
VA13
SEL0
SEL1
SEL2
DISPRS
R//W
69 /RST
65
66
12
13
14
15
16
17
18
19
33
32
31
30
29
28
27
26
53
52
51
50
49
48
47
46
A0
A1
A2
A3
A4
A5
A6
A7
Vcc
2
23
64
FPGA Reset Circuit
/RESET /MR
GND
U10
/RST
AVSS
AVDD
VRL
VRH
MODB/VSTBY
MODA//LIR
/RESET
E
XOUT
PH0/PW1
PH1/PW2
PH2/PW3
PH3/PW4
PH4/CSI0
PH5/CSGP1
PH6/CSGP2
PH7/CSPROG
PG0/XA13
PG1/XA14
PG2/XA15
PG3/XA16
PG4/XA17
PG5/XA18
PG6
PG7/R//W
PF0/ADDR0
PF1/ADDR1
PF2/ADDR2
PF3/ADDR3
PF4/ADDR4
PF5/ADDR5
PF6/ADDR6
PF7/ADDR7
VDD
VDD
VDD
LATCHPOS
ABATFAIL
N_PRESS
/CSCS
PCI2, D3
PCI2, D4
PCI2, D5
/FPGA_RST
UDRSHUT
UDRLATCH
UDRHDL
/KEYCS
/AUXCS
PCI2, D0
PCI2, D1
PCI2, D2
E
MCU Address Bus
MCU Data Bus
32
33
34
35
36
27
28
29
11
12
13
14
15
16
17
18
19
20
21
22
23
24
D3
D2
D1
D0
/RTCCS
A9
6
7
8
1
2
3
D5
D4
D7
D6
PRT_D[0..7]
Vcc
10
9
8
7
6
5
/RTCCS 20
22
/R/W
27
R//W
A0
A1
A2
A3
A4
A5
PRT_D[0..7]
GND
GCK1-I/O
I/O
I/O
I/O
M1-I/O
GND
M0-I/O
GND
I/O
I/O
I/O-TMS
I/O
I/O
I/O
I/O
I/O
GND
Vcc
I/O
I/O
I/O
I/O
I/O
I/O
I/O-TD1
I/O-TCK
GND
CE
OE
WE
A0
A1
A2
A3
A4
A5
SQW
INTA
INTB
D0
D1
D2
D3
D4
D5
D6
D7
Realtime Clock
U27
DS1286
PRT_D5
PRT_D4
/PRTACK
PRTBUSY
PRTERR
/PRTINIT
/PRTFAULT
/PRTFF
/PRTSTB
DISPTMG
1
2
3
4
5
D0
D1
D2
D3
D4
D5
D6
D7
23 NC
1
/IRQ
26 NC
11
12
13
15
16
17
18
19
PRT_D3
PRT_D2
PRT_D1
PRT_D0
BANKA15
BANKA14
ELECTRICAL SCHEMATICS
1
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
EA14
EA15
EA16
/EEOE
Vcc
118
Steri-Vac 8XL Gas Sterilizer/Aerator
GND
1
PRT_D7
PRT_D6
144
143
142
Vcc
I/O
I/O
Vcc
M2-I/O
GCK2-I/O
I/O-HDC
37
38
39
40
139
138
137
I/O
I/O
GND
I/O
I/O-LDC
GND
43
44
45
LATCHSET
LATCHRES
/R/W
EA14
EA15
EA16
134
133
132
131
130
129
128
127
126
125
124
123
122
121
I/O
I/O
I/O
I/O
I/O
I/O
Vcc
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O-INIT
Vcc
GND
I/O
I/O
I/O
I/O
I/O
I/O
48
49
50
51
52
53
54
55
56
57
58
59
60
61
GND
R//W
/EEOE
116
115
I/O
I/O
GND
I/O
I/O
64
65
66
16
22
24
31
12
11
10
9
8
7
6
5
27
26
23
25
4
28
29
3
2
GCK3-I/O
I/O
PROG
Vcc
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O-/RS
I/O
GND
Vcc
I/O
I/O
I/O
I/O
I/O-/CS0
I/O
I/O-/BUSY/RDY
I/O
GND
Vcc
CCLK
I/O-DOUT
I/O-DIN
E
76
75
74
73
81
80
79
A12
A13
A14
A15
97
96
95
94
93
92
91
90
89
88
87
86
85
84
A8
A7
A10
A11
D0
D1
D2
D3
D4
D5
D6
D7
Vcc
102
101
100
108
107
106
105
30
1
13
14
15
17
18
19
20
21
32
XC5202
U14
R//W
TMS29F512-C
VSS
/E
/G
/W
A0
VCC
A1
A2
DQ0
A3
DQ1
A4
DQ2
A5
DQ3
A6
DQ4
A7
DQ5
A8
DQ6
A9
DQ7
A10
A11
NC/A17
A12
NC
A13
A14
A15
NC/A16
External EEPROM
U23
/RAMOE
E
112
111
110
109
GCK4-I/O
I/O-/WS
GND
I/O
I/O
I/O
GND
DONE
69
70
71
72
Vcc
10
9
8
7
6
5
4
3
25
24
21
23
2
26
1
27
20
22
14
HEATSINK_CTL
EXT_AUDIO_CTL
/FPGA_RST
RH_SOL
EP_POWER
EP_VALVE
UD_SOL
VENT_SOL
PUNCT_SIM
EX1_SOL
CYC_START
GAS_SIG
EX3_SOL
WATER_SOL
N_SOL
EX2_SOL
END_HTR_CTL
MID_HTR_CTL
VAC_SOL
PUNCTSOL
LD_SOL
CLOSE_LR
OPEN_LR
PUNCTPIN
PRTON
LDRSHUT
LDRLATCH
LDRHDL
ADDWATER
AIRFLOW
ABATRDY
CARTPOS
COMPAIR
R//W
/RAMOE
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
PCO1, D0
PCO1, D1
PCO1, D2
PCO1, D3
PCO1, D4
PCO1, D5
PCO1, D6
PCO1, D7
PCO2, D0
PCO2, D1
PCO2, D2
PCO2, D3
PCO2, D4
PCO2, D5
PCO2, D6
PCO2, D7
PCO3, D0
PCO3, D1
PCO3, D2
PCO3, D3
PCO3, D4
PCO3, D5
PCO3, D6
PCO3, D7
PCI1, D6
PCI1, D7
PCI1, D0
PCI1, D1
PCI1, D2
PCI1, D3
PCI1, D4
PCI1, D5
PCI2, D6
PCI2, D7
DS1225D
Battery Backed NV-SRAM
U24
28
A0
VCC
11
A1
DQ0
12
A2
DQ1
13
A3
DQ2
15
A4
DQ3
16
A5
DQ4
17
A6
DQ5
18
A7
DQ6
19
A8
DQ7
A9
A10
A11
A12
E2/A13
NC/A14
/W
/E
/G
VSS
Vcc
D0
D1
D2
D3
D4
D5
D6
D7
Vcc
R32
470 Ohm
1
2
3
4
XC1765D
DATA
Vcc
CLK
Vpp
OE/RESET CEO
CE
GND
BX1
8
7
6
5
A[0..15]
D[0..7]
78-8078-6519-7 Sheet 2 of 6
Vcc
VA[11..13]
Schematic, Controller, Steri-Vac
DATA_BX1
A[0..15]
D[0..7]
VA[11..13]
153
Service and Repair
154
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
ELECTRICAL SCHEMATICS
14
Vcc
CYC_START
EX1_SOL
PUNCT_SIM
VENT_SOL
UD_SOL
EP_VALVE
EP_POWER
RH_SOL
GAS_SIG
EX3_SOL
U44
STAR KMB-6
HEATSINK_CTL
5
3
ULN2068/2
U8F
12
14
5
7
Electric Pump Power
Electric Pump Vent Valve
9
13
12
14
1
4
5
7
16
9
13
12
14
15
1
4
5
7
14
7
1
EXT_AUDIO_CTL 9
MID_HTR_CTL
7
12
5
END_HTR_CTL
13
4
Mid Heaer SSR Control
U8A 2
4050
U8D 10
4050
R34
LED MIDDLE HEATERS
R31
D2
470 Ohm
LED END HEATERS
R7
D1
470 Ohm
470 Ohm
EX2SOL_OUT
NITROSOL
WATERSOL
Vcc
Vcc
Vcc
CYCSTART
EX1SOL_OUT
PUNCTSIM
VENT
ELECTVV
ELECTP
RHSOL
GASCOMING
HEAT_CTRL
VACPUMP
Vcc
Vcc
15
EXTERNAL_AUDIO
MIDHEATER
ENDHEATER
HEATSINK
LED HEATSINK
D4
U8C 6
4050
U8B 4
4050
4050
14
4050
9
1
End Heater SSR Control
12
16
15
Audio Output Control
U8E
10
11
Heatsink SSR Control
11
8
ULN2068/2
2
6
U7
10
11
ULN2068/2
8
6
3
+24V
2
3
U3
16
15
ULN2068/2
10
11
Cycle Starting Control Signal
+24V
8
6
Extra Solenoid Output Control
2
3
4
5
6
7
8
9
10
2
3
Puncture Simulate Control
Vent Solenoid Control
U4
13
4
RH Solenoid Control
Unload Door Solenoid Control
9
1
Gas Comming Control Signal
16
10
15
8
6
11
2
14
16
10
Heater Cutout Control
+24V
7
Vacuum Pump Solenoid Control
3
5
Extra Solenoid Output Control
U5
4
Nitrogen Injection Solenoid Cotrol
+24V
12
1
Water Injection Solenoid Control
ULN2068/2
13
15
Load Door Solenoid Control
2
3
4
5
6
7
8
9
10
9
11
Puncture Solenoid Control
8
6
2
Open Latching Relay Control
U6
3
Close Latching Relay Control
Audio Output
MID_HTR_CTL
END_HTR_CTL
EXT_AUDIO_CTL
HEATSINK_CTL
RM2
10K X 9
1
VAC_SOL
RM4
10K X 9
1
EX2_SOL
N_SOL
WATER_SOL
LD_SOL
PUNCTSOL
OPEN_LR
CLOSE_LR
+24V
1
TP12
Set
Set
RLY DS4ESL2
7
5
3
9
11
13
15
U13
Gas Present
16
Vcc
LATCHPOS
LED D3
Set
Set
GAS PRESENT
10
12
14
8
6
4
2
1
+24V
Vcc
Puncture Solenoid Simulation Testpoint
470 Ohm
R3
HEAT_CTRL
EX2SOL_OUT
EX1SOL_OUT
PUNCTURE
UDOORSOL
LDOORSOL
WATERSOL
VACPUMP
GASCOMING
RHSOL
VENT
CYCSTART
HEATSINK
ENDHEATER
MIDHEATER
NITROSOL
ELECTP
ELECTVV
/PRTINIT
JP3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
Bottom Layer
EXTERNAL_AUDIO
Vcc
HEADER 3
1
2
3
J2
1
A
AMP 583617-1
HEADER 22 HEADER 22
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
JP2
Top Layer
ABATRDY
PUNCTPIN
AIRFLOW
ADDWATER
LDRSHUT
LDRLATCH
LDRHDL
CARTPOS
UDRSHUT
UDRLATCH
UDRHDL
ABATFAIL
N_PRESS
COMPAIR
Vcc
Vcc
Vcc
1
2
3
/PRTFF
Vcc
Vcc
Printer ON/FormFeed Switch
HEADER 3
J6
LDRSHUT
LDRLATCH
LDRHDL
ADDWATER
AIRFLOW
ABATRDY
PUNCTPIN
UDRSHUT
UDRLATCH
UDRHDL
ABATFAIL
N_PRESS
COMPAIR
CARTPOS
1
Steri-Vac 8XL Gas Sterilizer/Aerator
8
1
RM3
10K X 9
221 Ohms 1%
R4
2
3
4
5
6
7
8
9
10
RM1
10K X 9
PRTON
/PRTFF
LDRSHUT
LDRLATCH
LDRHDL
ADDWATER
AIRFLOW
ABATRDY
PUNCTPIN
UDRSHUT
UDRLATCH
UDRHDL
ABATFAIL
N_PRESS
COMPAIR
CARTPOS
78-8078-6519-7 Sheet 3 of 6
Schematic, Controller, Steri-Vac
PRTON
2
3
4
5
6
7
8
9
10
155
Service and Repair
156
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
BACKLITE
/FPGA_RST
FPGA_DONE
DISPON
DISPRS
/CLRKEY
/KEYEN
FPGA_DONE
VA[11..13]
R//W
E
FPGA_VIDEO1
FPGA_VIDEO2
XOUT
DISPTMG
R//W
E
FPGA_VIDEO1
FPGA_VIDEO2
XOUT
DISPTMG
BACKLITE
CSD2
CSD3
CSD4
CSD5
CSD6
CSD7
/CSCS
CSD0
CSD1
A4
A5
A6
A7
VA11
VA12
VA13
D1
D0
A0
A1
A2
A3
Vcc
J16
DB15
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
GND
GCK1-I/O
I/O
I/O
I/O
I/O-TD1
I/O-TCK
GND
I/O
I/O
I/O-TMS
I/O
I/O
I/O
I/O
I/O
GND
Vcc
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
GND
I/O
I/O
I/O
I/O
I/O
I/O
M1-I/O
GND
M0-I/O
M
VA[11..13]
D2
D3
D4
D5
D6
D7
LU3
A[0..15]
LU2
A[0..15]
LU1
D[0..7]
A8
A9
A10
A11
A12
A13
A14
A15
M
CL1
CL2
LU3
LU2
LU1
LU0
FLM
FLM
D[0..7]
Keypad/Video Conector
D-Sub Header With Grounding Pads
U28
T_CAP
U26
T_CAP
U25
T_CAP
U22
T_CAP
U21
T_CAP
U19
T_CAP
3
3
3
3
3
3
DISPON
DISPRS
STOP
ELECTRICAL SCHEMATICS
AUXD0
AUXD1
AUXD2
AUXD3
AUXD4
AUXD5
AUXD6
AUXD7
/AUXCS
AUXD0
AUXD1
AUXD2
AUXD3
AUXD4
AUXD5
AUXD6
AUXD7
VMA12
VMA13
VMA7
VMA8
VMA6
VMA9
Steri-Vac 8XL Gas Sterilizer/Aerator
VMA5
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
Vcc
CCLK
I/O-DOUT
I/O-DIN
I/O
I/O
I/O-/BUSY/RDY
I/O
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O-/RS
I/O
GND
Vcc
I/O
I/O
I/O
I/O
I/O-/CS0
I/O
I/O
I/O
GND
I/O
I/O
I/O
I/O
GCK3-I/O
I/O
PROG
Vcc
Vcc
I/O
I/O
I/O
I/O
I/O
I/O
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
Vcc
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
GND
I/O
I/O
I/O
I/O
I/O
GCK4-I/O
I/O-/WS
GND
I/O
KEY0
KEY1
KEY2
KEY3
STOP
Vcc
M2-I/O
GCK2-I/O
I/O-HDC
I/O
I/O
I/O
I/O-LDC
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O-INIT
Vcc
GND
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
GND
I/O
I/O
I/O
I/O
I/O
I/O
GND
DONE
KEY3
KEY2
KEY1
KEY0
/CLRKEY
/KEYEN
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
1
9
2
10
3
11
4
12
5
13
6
14
7
15
8
108
107
106
105
104
103
102
101
100
99
98
97
96
95
94
93
92
91
90
89
88
87
86
85
84
83
82
81
80
79
78
77
76
75
74
73
XC5204
U12
/KEYCS
DATA
Vcc
CLK
Vpp
OE/RESET CEO
CE
GND
BX2
Vcc
XC17128D
R30
470 Ohm
1
VCLK_STROBE 2
3
4
VD3
/VB1_CE
/VB_WE
/VB2_CE
/VB_OE
VD11
VD7
VD0
VD6
VD1
VD5
VD2
VD4
VMA0
VD15
VD8
VD14
VD9
VD13
VD10
VD12
VMA11
VMA4
VMA3
VMA10
VMA2
VMA1
8
7
6
5
Vcc
/VB_WE
/VB2_CE
/VB_OE
/VB_WE
/VB1_CE
/VB_OE
Vcc
10
9
8
7
6
5
4
3
25
24
21
23
2
26
1
27
20
22
14
VMA0
VMA1
VMA2
VMA3
VMA4
VMA5
VMA6
VMA7
VMA8
VMA9
VMA10
VMA11
VMA12
VMA13
Vcc
Vcc
10
9
8
7
6
5
4
3
25
24
21
23
2
26
1
27
20
22
14
VMA0
VMA1
VMA2
VMA3
VMA4
VMA5
VMA6
VMA7
VMA8
VMA9
VMA10
VMA11
VMA12
VMA13
1
MCM6206
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
E2/A13
NC/A14
/W
/E
/G
VSS
U42
MCM6206
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
E2/A13
NC/A14
/W
/E
/G
VSS
U41
2
3
4
5
6
7
8
9
10
RM9
10K X 9
VCC
DQ0
DQ1
DQ2
DQ3
DQ4
DQ5
DQ6
DQ7
VCC
DQ0
DQ1
DQ2
DQ3
DQ4
DQ5
DQ6
DQ7
Vcc
28
11
12
13
15
16
17
18
19
Vcc
28
11
12
13
15
16
17
18
19
VD0
VD1
VD2
VD3
VD4
VD5
VD6
VD7
VD8
VD9
VD10
VD11
VD12
VD13
VD14
VD15
78-8078-6519-7 Sheet 4 of 6
Schematic, Controller, Steri-Vac
EXT_AUDIO_CTL
MID_HTR_CTL
END_HTR_CTL
HEATSINK_CTL
157
Service and Repair
158
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
A
B
C
D
ELECTRICAL SCHEMATICS
1
RX
TXD
TXD_CTL
RXD_CTL
1
Steri-Vac 8XL Gas Sterilizer/Aerator
RX
TXD
10uF
10uF
C5
C5
1
1
1
1
1
2
4
5
6
1
R2I
R1I
T2OU
T1OU
V-
VC
V+
/SHD
2
PRTBUS
YPRTER
/R
PRTINIT
/PRTFAULT
/PRTF
/BPRTACK
/PRTST
PRT_D[0..7]
Serial Communications
MAX242CPN
GN
R2OU
R1OU
T2I
T1I
C1
C1
C2
C2
/E
U3
2
C4
0.1u
10uF
C5
PRT_D[0.
9
1
8
1
7
1
3
1
Vcc
Vcc
/PRTINIT
/PRTFAULT
/PRTFF
PRTERR
/PRTSTB
/PRTACK
PRTBUSY
TP1
1
RXD_I 1
GN
2
TXD_OU3
4
10uF
C5
J
3
1
2
3
4
5
6
7
8
PRT_D
PRT_D
PRT_D
PRT_D
PRT_D
PRT_D
PRT_D
PRT_D
SW DIP-8
SW1
1
1
1
1
1
1
1
9
Vcc
2 3 4 5 6 7 8 9 10
1
Configuration DIP Switches
4
CSD
CSD
CSD
CSD
CSD
CSD
CSD
CSD
RM8
10K X 9
Optional
Ribbon Cable Connector
Printer Connector for Fujitsu (Centronics)
4
3029282726252423222120191817161514131211109 8 7 6 5 4 3 2 1
RJ1
HEADER 30
J1
3
SW DIP-8
SW2
2
4
6
8
1
1
1
Ribbon Cable Connector
Printer Connector for Seiko
1
3
5
7
9
1
1
HEADER
J9
1
2
3
4
5
6
7
8
1
1
1
1
1
1
1
9
Vcc
5
Numb
21-SepC:\Documents and
B
Dat
File
Sheet
Drawn
78-8078-6519-7 Sheet 5 of 6
Siz
Titl
AUXD
AUXD
AUXD
AUXD
AUXD
AUXD
AUXD
AUXD
RM7
10K X 9
6
6
Schematic, Controller, Steri-Vac
2 3 4 5 6 7 8 9 10
1
Auxillary DIP Switches
5
Revisi
A
B
C
D
159
Service and Repair
160
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
1
2
3
4
5
6
7
8
9
10
11
12
13
14
SEL0
SEL1
SEL2
HEADER 14
J4
Vcc
SEL0
SEL1
SEL2
1
Temperature Sensor
T1
R2
T2
3
LATCHPOS
R29
221K 1%
Vcc
VREF
CHTEMPIN
ZONE1IN
ZONE2IN
RECTEMPIN
AMTEMPIN
HSTEMPIN
CATEMPIN
EXTEMP1
9
10
6
VREF5
4051
Multiplexed Temp Sensors
U9
13
X0
X
14
X1
15
X2
12
X3
1
X4
5
X5
2
X6
4
X7
VDD
6
INH
11
A
VSS
10
B
9
C
VEE
2.5VVoltage Reference
D10
LM236-2.5V
1
NC
2.49K 1%
R1
R25
R24
R23
R22
R21
R20
R19
R18
1K .1% 1K .1% 1K .1% 1K .1% 1K .1% 1K .1% 1K .1% 1K .1%
2.21K 1%
Calibration Resistor
VTC11
Middle Temp
1
End Temp
U47
TD5A
1
RH
1
Pressure
1
Chamber Temp
VTC4
CHTEMPIN
VTC2
ZONE1IN
VTC3
ZONE2IN
VTC4
HSTEMPIN
VTC2
RECTEMPIN
RH
1
GND
1
Leak Presure
1
VCC
Vcc
NC
2
1
Rec Temp
ELECTRICAL SCHEMATICS
7
8
16
3
VREF
VREF
Vcc
T_MUXED
VTC3
13
12
2
3
2
3
Vcc
TLC27L4
1
U33D
TLC27L4
U39A
1
R15
10K .1%
TLC27L4
R37
10K .1%
24.9K .1%
4.64K .1%
R40
TLC27L4
U40B
7
13
12
221K .1%
0.0K
U40A
1
U39B
7
TLC27L4
U39D
14
U40C
8
TLC27L4
U40D
14
1
2
3
EXA3
Vcc
EXA2
EXA0
U39C
8
+15V
EXTEMP1
VTC2
5.11K .1%
R38
P_IN
Chamber Pressure Input
HEADER 3
J8
+15V
TLC27L4
Unity Gain Amplifier
9
10
TLC27L4
Unity Gain Amplifier
Unity Gain Amplifier
9
10
2
R9
221 Ohms 1%
+Vout
Chamber Pressure Conditioning Circuit
R8
2.7K 1%
Vin
U20
LM317
TLC27L4
Non-Inverting Amplifier
R17
6
5
11.3K .1%
13
12
1
C1
0.1uF
+24V
R39
R36
6
5
R16
Adjustment Voltage
VTC4
TLC27L4
14
Unity Gain Amplifier
Vcc
U33A
VTC1
Unity Gain Amplifier
2
3
Vcc
24.9K .1%
VTC3 R14
TLC27L4
8
U33C
VTC2
TLC27L4
7
U33B
4
11
3
2
4
11
Steri-Vac 8XL Gas Sterilizer/Aerator
4
11
ADJ
3
Vcc
R12
1K .1%
C59
1uF
Lowpass Filter
LEAK_PRESURE
CH07
CHPRESS
RH
VREF
78-8078-6519-7 Sheet 6 of 6
Schematic, Controller, Steri-Vac
HEADER 12
1
2
3
4
5
6
7
8
9
10
11
12
J10
R11
2.75K .1%
R10
1K .1%
D9
IN4148
Input Voltage Protection
LEAKPRES
CH07
CHPRESS
RH
VREF
161
Service and Repair
162
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
150K Ohm
TP2
GND
1
R8
2.5M ohm
C4
0.1uF
C2
0.1uF
+24V
CT
C1
150uF
HEADER 5
5
4
3
2
1
DC/DC Converter (5V to -24V)
Power Input
J1
D-Sub Header
R3
LCD Video Header
To Controller
2
4
3
DB15
R2
LU0
FLM
LU1
LU2
LU3
CL2
CL1
M
EXT
OUT
CS
V+
VCC
2.2M Ohm
FB
REF
SHDN
U2
MAX774
1
9
2
10
3
11
4
12
5
13
6
14
7
15
8
7
1
6
5
D1
L1 1N5820
22uH
Q1
Si9435
R1
0.07 Ohm
BACKLITE
SIGNALGND
KEY0
KEY1
KEY2
KEY3
STOP
C3
100uF
C5
100uF
2
1
TP1
-24V
-24V
HEADER 2
J3
1
J16
GND
8
U3
JUMPER
JP4
DS1E-M-DC5V
JP5
10
JUMPER
Q2
2N3904
2
1
Backlight Diming Circuit
1K 5%
R7
2
1
1
12
-+
6
7
ELECTRICAL SCHEMATICS
VCC
2
1
COM
AC VOUT1
AC VOUT2
TDK DC/AC Inverter
CXA-L10A
GND
VCC
U1
1
3
4
5
VCC
JP2
JUMPER
2
Steri-Vac 8XL Gas Sterilizer/Aerator
JP1
JUMPER
2
1
HEADER 4
4
3
2
1
CN2
78-8078-4557-9
AC Output Header
For Stanley
JP3
JUMPER
HEADER 14
1
2
3
4
5
6
7
8
9
10
11
12
13
14
CN1
LCD Video Heade
KeyPad Header
HEADER 9
1
2
3
4
5
6
7
8
9
J2
KeyPad Display Interface Board
2
1
LU0
LU1
LU2
LU3
LCD_VCC
LCD_GND
-24V
CL1
CL2
FLM
Overlay Power On LED
1K
Contrast Control Circuit
R4
-24V
R5
4.3K
R6
50K
VCC
STOP
KEY3
KEY2
KEY1
KEY0
SIGNALGND
163
Service and Repair
164
ILLUSTRATED PARTS BREAKDOWN
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
Steri-Vac 8XL Gas Sterilizer/Aerator
Service and Repair
STERILIZER OPERATIONAL FLOW DIAGRAM
© 3M 1995 - 2005
System Initialization
Go to page 167
Service Mode
Go to page 167
Recovery
Go to page 167
Main Menu
Go to page 167
Precondition
Go to page 169
Gas Injection
Go to page 169
Gas Expose
Go to page 169
Final Vacuum
Go to page 169
Purge
Go to page 171
Locked Aeration
Go to page 171
Unlocked Aeration
Go to page 171
Standby Aeration
Go to page 171
Reset Cycle
Go to page 171
165
Service and Repair
Steri-Vac 8XL Gas Sterilizer/Aerator
STERILIZER OPERATIONAL FLOW DIAGRAM
This page intentionally left blank.
166
© 3M 1995 - 2005
Service Mode
© 3M 1995 - 2005
Go to
sheet 2
Pre-Condition
Main Menu
Recovery
Gas Present?
No
No
Passwords
Enabled?
Go to
Main Menu
No
Cycle In
Progress?
Go to
sheet 2
Final Vacuum
Yes
Yes
No
Valid
Password?
Yes Display
Main Menu
No
Yes
No
Yes
No
Adjust
Aeration
Time?
Yes
Display Errors
Reprogram
RH Set-points
to Default
Yes
Clear
RH Setpoints No
in EEPROM
Option?
Default Variables
Critical Options
Initialize PCO's
Initialize Keypad
Initialize Clock
Initialize Video
Initialize ADC
Start Printer
Aeration?
Go to
sheet 2
Final Vacuum
Go to
sheet 3
Locked Aeration
Final
Vacuum
Option?
E3
Yes
No
No
RAM Okay?
Yes
Aeration
Option?
Check RAM
Yes Print and Display C3
Draw Cycle Screen
Stop
No
Key Pressed?
Yes
Verify MCU
System
Intialization
STERILIZER OPERATIONAL FLOW DIAGRAM
Steri-Vac 8XL Gas Sterilizer/Aerator
No
No
Adjust
Cycle
Temp?
No
Yes
Gas Inject Yes
Gas Expose
Final Vac?
Print or
Display Trend
Info
Yes
Trend
Report
Option?
Turn On Interrupts
Initialize Printer
Calibrate RH Sensor
No
No
Yes
E3
Go to
sheet 2
Final Vacuum
Display Service
Menu
No
Setup
Yes
Info Checksum
Okay?
No
E11
Compressed Yes
Air Okay?
Wait for Chamber
Temperature
If TO(45) E52
Start Cycle?
No
P>
Yes
AmbientP - 80?
E54
Display
Screen
Pattern
Yes
Screen Test
Option?
Check Checksums
Yes
Yes
No
E1
No
E10
Water Level Yes
Okay?
No
Gas
Expose
Stage?
No
Gas
Injection
Stage?
Perform Site
Setup
Yes
Site Setup
Selected?
No
Cycle
Yes
Info Checksum
Okay?
No
No
Analog
Sensors
Okay?
E4,E5,
E6,E8,
E9,E12
Yes
Go to
sheet 2
Final Vacuum
Go to
sheet 2
Gas Expose
Go to
sheet 2
Gas Injection
Perform
Vacuum
Test
Yes
Vacuum
Test
Selected?
Start IRQ Interupt
Clear Screen
No
Yes
Yes
No
Perform
Switch Test
Yes
Switch Test
Selected?
No
C12
Printer Report
Header
Calibrate
Analog Circuit
Heaters On (cycle)
Vacuum On
Open Vent
Set Timer
Heaters On (cycle)
Vacuum On
Calibration
Required?
No
Purge?
Perform
Solenoid
Test
Yes
Solenoid
Test
Selected?
No
167
Go to
sheet 2
Pre-Condition
Go to
sheet 3
Locked Aeration
Go to
sheet 3
Purge
Go to
Recovery
Yes
Stop Switch
Pressed?
Go to
Service Mode
Service and Repair
168
STERILIZER OPERATIONAL FLOW DIAGRAM
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
Heatsink On
(cycle)
Final Vacuum
No
Heatsink On
(cycle)
Go to
sheet 3
Purge
Yes
Cycle Starting
(ON)
Yes
No
No
E28
Yes
Vacuum
Off
Vacuum
On
Wait 3.5
Sec
Wait 3.5
Sec
Yes
Yes
E30
Print Error
E77
Yes
Inject water (1sec)
Wait (60sec)
Open RH Solenoids
Vacuum On (20sec)
Read RH
Close RH Solenoids
Vacuum Off
No
No
Vacuum
On
Yes
No
50 cfm
Option?
Vacuum
On
If TO(90)
E71
Yes If TO(30)
E71
Check Pressure
Sensor
E73
No
P<
160 mBar?
No
Nitrogen
On
No
Temp
Okay?
(+/-2C)
No
Vacuum
Off
Pulse
Cartridge
Pump 3
Minutes
Yes
Check Pressure
Sensor
E73
Yes Display Delay
Errors
No
No
No
No
No
3 Minute
Timeout?
No
E25,
E26
No
Temp
Okay?
(+/-2C)
Yes
Yes
Yes
Nitrogen
Off
Partial Puncture
Error
E75
Empty Cartridge
Error
E50
Pressure
Increase after Yes
Puncture <
200mBar?
Yes
P >= 800?
Pressure
Increase after
Puncture <
50mBar?
Yes Nitrogen
On
Delay Error
E53 Too Hot
E52 Too Cool
Yes
Nitrogen
Option?
Temp
Okay?
(+/-2C)
Nitrogen
Off
Final 3 Minute Vacuum
Read RH Every 10sec
Yes
Number
No
of RH checks
< 5?
Increment # of
Long Dwells
Yes
If TO(3)
E39
No
P >= 800?
If RH < 80 Inject (1sec)
Wait (60sec)
Open RH Solenoids
Vacuum On (20sec)
Read RH
Close RH Solenoids
Vacuum Off
Delay Error
E53 Too Hot
E32 Too Cool
Yes
Delay
Errors?
Pulse
Cartridge
E73
Yes
Number
No
of Long Dwells
< 3?
No
Nitrogen
Option?
# of
Yes
P increases
> 2?
Pump until
# P increases > 2?
P=AmbientP - 80
If TO(2) Delay Error E71 Delay Error E73
Increment # P increases
No
Puncture
Switch
Okay?
Yes
RH < 60%?
Chamber Leak Test
Wait 5 Min
If P increase 15 mBars
E31
Yes
P>
Pump until
AmbientP - Yes
P=AmbientP - 80
80 mBar?
If TO(2) Error E71
Increment # P increases
If TO(60)
E71
No
P<
250 mBar?
No
P>
AmbientP 80 mBar?
Check Pressure
Sensor E73
If P Sensor Error
Delay Error E73
No
No
Rh < 2 or
RH > 98?
2nd
Injection & No No
Heatsink Temp
Decrease?
Gas
Yes Stop Puncture Cartridge
Inject Complete
Heatesink Off
(50 sec)?
E28
E29
3rd
Injection &
No RH
Increase?
Yes
No
Pumpdown to
Target P,
If TO(40) E22
Close Vent
Heatsink On (95C)
Yes
Maximum
Number of
Injections?
Preheat Option
Wait 15 Min
Wait for Chamber
& Heatsink Temp
If TO(45) E23
Send Gas
Coming Signal
for 20 Sec
Go to
Final Vacuum
Yes Wait for Abator
50 cfm
Ready Signal
Abator Option?
If TO (120) C10
No
Gas
Expose
Complete?
Close Gas Relay
Puncture Cartridge
Open RH Solenoids
Vacuum On (20sec)
Read RH
Close RH Solenoids
Vacuum Off
Heaters On (cycle)
Heatsink On (cycle)
Open Vent, Vacuum On
Open RH Solenoids
Test Vacuum for 1 min
If P>AmbientP-52mBars
E21
International Yes Heatsink Cooldown
Version?
If TO (20) E24
No
50scfm
Abator
Option?
Gas Expose
Gas Injection
PreCondition
STERILIZER OPERATIONAL FLOW DIAGRAM
Steri-Vac 8XL Gas Sterilizer/Aerator
Go to
sheet 3
Purge
Go to
Gas Expose
Go to
Final Vacuum
Go to
Gas Injection
169
Service and Repair
170
STERILIZER OPERATIONAL FLOW DIAGRAM
Service and Repair
This page intentionally left blank.
© 3M 1995 - 2005
Steri-Vac 8XL Gas Sterilizer/Aerator
© 3M 1995 - 2005
Reset Cycle
Heaters Off
Stop Timer
5 Second Beep
Standby
Aeration
Stop Report
Stop Timer
Heaters Off
Heatsink Off
Vacuum Off
No
Nitrogen Off
Clear Cautions
No
Remove
Gas Buildup?
D
C
Vacuum On
(2 Minutes)
No
Latching
Relay
Closed?
Yes
E
No
Yes
E76
No
No
50 cfm
Abator with
Aeration?
No
Open Vent
Close RH Valves
Stop Puncture
Unlock Load Door
Stage = Standby
Clear Errors
Stop
Timer
Adjust
Vent
No
Handle
Yes
Turned or Door
Open?
Yes Cycle Starting
(Off)
No
No
B
Pulse
Cartridge
Open Vent
Open Vent
No
Vacuum Off
Open Vent
Stop Timer
No
Yes
B
C
No
Yes
No
A
No
Puncture
Pin Stuck?
No
Handle
Turned
Down?
No
Door
Open?
No
Yes
No
Door
Open?
No
Standby
Aeration?
No
Handle
Turned
Down?
Yes
Open Latching Relay
Stop Report
Yes
Locked
Aeration?
Yes
Yes
Check Aeration
Heater Caution
Adjust Vent
8
Yes
Minute Purge
Done?
Valid
Password?
Stop
Key Pressed Yes
& Setup Code
Entered?
Passords
Enabled?
No
7
Yes
Minute Purge
Done?
7
Yes
Minute Vacuum
Done?
Locked
Yes
Aeration Time
Complete?
Close Vent
Open Vent
PurgeCnt
< 3?
Chamber
Yes Restart
Temp within
Timer
Range?
Print Report
Tailer
No
Handle
Yes
Turned or Door
Open?
Vacuum On
Check Aeration
Heater Caution
Adjust Vent
No
PurgeCnt = 0
Heatsink On
Chamber
Yes Heaters On
Temp within
Vacuum On
Range?
Open RH Solenoid
Unlocked
Yes
Aeration Time
Complete?
Vacuum Off
(3 Minutes)
Yes
Heatsink On
Heaters On
Vacuum On
Open RH Solenoid
Start Timer
Heaters On (cycle)
Initial
Yes Check Obstructed
8 Minute Purge
Filter
E72
Done?
Yes Cycle Starting
(Off)
Start Timer
Unlock Unload Door
No
50 cfm
Abator w/o
Aeration?
Open Vent
Vacuum On
Heatsink On (cycle)
UnLocked
Aeration
Locked Aeration
Purge
STERILIZER OPERATIONAL FLOW DIAGRAM
Steri-Vac 8XL Gas Sterilizer/Aerator
Yes
E
Go to
Standby Aeratio
Go to
Unlocked
Aeration
Go to
Locked Aeration
Yes
C
D
A
Go to
sheet 1
Main Menu
Print Door Opened
Vacuum On
Open RH Solenoids
Yes
No
Stop
Key
Pressed?
E77
171
Service and Repair
This page intentionally left blank.
3
Health Care
3M Center Building, 275-4E-01
St. Paul, MN 55144-1000
1-800-228-3957
3M Canada Inc.
P.O. Box 5757
London, Ontario, Canada N6A4T1
1-800-563-2921
Printed on recycled paper
Printed in U.S.A.
© 3M 1995 - 2005 78-6970-6954-6 Rev. E
Distribution Limited To Current 3M Approved Service Providers
Distribution Limited To Current 3M Approved Service Providers
Distribution Limited To Current 3M Approved Service Providers
Distribution Limited To Current 3M Approved Service Providers
Distribution Limited To Current 3M Approved Service Providers
Equipment
Bulletin
48
Subject:
Service - Final Checks Process
Date:
August 5, 2002
Machines Affected:
4XL, 5XL, and 8XL
Condition:
N/A
Cause:
N/A
Solution:
This Bulletin is written to describe the Final Checks and
procedures that must always be performed on the sterilizer
before the sterilizer is put back into use.
Check #1:
Temperature Calibration
Use either the 49C method or the 2231 Ohm resistor method.
Temperature Calibration Procedure Using the 49C Method
(Preferred method on all sterilizers)
1. Put the Fluke thermocouple under the TCO on a 5XL or near the chamber
temperature sensor on a 4XL. Use silver tape to attach thermocouple, if needed.
Use Thermal Grease on the Thermocouple / Chamber interface to insure good
heat transfer. DO NOT place thermocouple under the Chamber High Limit
Thermostat on a 4XL.
2. Put the unit in cycle and allow the display to go to 51 C.
3. Then turn power off, set DIP SW1-1 and SW1-3 to the up position, and allow the
temperature (on the fluke) to go to exactly 49 C. At that point, turn the unit on
and it will calibrate to within a couple of tenths of a degree of true accuracy. (The
door latch will activate indicating acceptance of the new calibration.)
Health Care/3M
Page 1 of 4
Distribution Limited To Current 3M Approved Service Providers
4. Turn power off and set DIP Switch SW1-1 and SW1-3, down.
5. Turn power on. Unit will come up in a C3, and the unit will continue in cycle.
6. Observe the display and Fluke to insure the temperatures match. If OK, remove
the thermocouple and press stop button to terminate cycle. If temperature on
Fluke and on the display do not match, repeat steps 1 – 6.
IMPORTANT NOTE: If machine will not accept the calibration (vacuum pump comes
on instead of latch activation), the machine is too far out of calibration to accept the 49 C.
To correct, observe the display when the Fluke reaches 49 C. If above 51 C, calibrate the
temperature in 2 C increments until between 51 C and 49 C. Then make final calibration
to 49 C. If the temperature of the chamber reaches 51 C, and the Fluke is below 49 C, the
2231 Ohm resistor method must be used to calibrate the temperature.
Temperature Calibration Procedure Using the 2231 Ohm Resistor
(Recommended when had to change the controller board on 4XL and on old
5XL, just to get close to correct temperature. Then use above method to get
accurate calibration.)
1. Take the 2231 ohm variable resistor from your tool case and insert the ends into
the multi-meter (common and Ohms), set to read resistance. Adjust resistor until
the multi-meter reads 2231.
2. Measure the temperature of the chamber by placing the thermocouple near the
location of the chamber temperature sensor. Attach Thermocouple with Silver
Tape if needed. Add Thermal Grease to thermocouple / chamber interface to
insure good heat transfer. Start a warm cycle, and allow the temperature to
stabilize at 55C.
3. Note the difference between the actual chamber temperature measured by the
thermocouple and the displayed temperature. This is the amount you want to
offset. (For example, suppose the actual temperature is measured as 53C, 2
degrees lower than displayed. If more than 2C offset, see NOTE below.)
4. Disconnect the chamber sensor at the connector and put the 2231 ohm resistor
leads into the connector end that goes to the controller board, not the connector
end that goes to the sensor. The display will now read 48C.
5. Adjust the resistor until the display reads the offset temperature correction. (For
this example the actual is 2 degrees lower (See #3), so the resistor should be
adjusted until the display reads 50C).
Health Care/3M
Page 2 of 4
Distribution Limited To Current 3M Approved Service Providers
6. Turn the power off. Place DIP switches 1 and 3 Up and turn the power on again.
This now calibrates the temperature.
7. Turn the power off again, remove the variable resistor and reconnect the chamber
temperature sensor.
8. Place DIP switches 1 and 3 down again and turn the power on. The calibration
procedure is complete. Recheck the chamber temperature to see that the actual
temperature is the same as displayed.
9. If the temperature is now close, within 1-2 degrees C, use the 49C method
described in the first section to precisely calibrate the temperature.
IMPORTANT NOTE: If the actual temperature is more than 2C offset from the
displayed temperature, the calibration can not be accomplished in one calibration. The
above calibration procedure will have to be done more than once, adjusting the
temperature 2C at one time. THE CALIBRATION WILL NOT BE ACCOMPLISHED
IF MORE THAN 2C OFFSET IS ATTEMPTED.
Check #2:
Proper Default Temperature on Display Panel
1. Cycle Power 1 time
a. Check return to default cycle.
b. If not at customer requested default, i.e. warm cycle or cold cycle, change
default.
When power is cycled on a 4XL or old 5XL, the unit will default to the
temperature it was set on during the last in cycle power interruption. This
can be a problem if the service rep interrupts power on a cold cycle and
the customer runs only warm cycles. To correct this, simply cycle power
after servicing to make sure the unit defaults to warm (or cold if the
customer wants to default to the cold cycle). If it doesn’t, you can select
the desired default cycle, start the cycle and turn the power switch off and
back on. Press Stop to get an E40, open the door and press Stop again.
Check for proper default by cycling power again.
Health Care/3M
Page 3 of 4
Distribution Limited To Current 3M Approved Service Providers
Check #3:
After Regulator Re-build or Replacement
Adjust air pressure up 5 psi to account for drop across new diaphragm. Use caution
with questionable air quality.
Check #4:
Run Last Cycle (5XLe and 8XL only)
Run last cycle information and check printout for any abnormalities.
Check #5:
Run Accelerated Cycle
Run Accelerated cycle. Make sure cartridge is punctured properly.
Check #6:
Final Power Test
After side panel is back on, cycle power and start a cycle. Allow machine to run for
30 seconds, check printout and make sure machine is going into cycle. If OK, press
Stop, and clear code.
Check #7:
Personnel Discussion
Discuss any abnormal information with appropriate hospital (facility) employee
(Biomed, Maintenance Manager, etc.)
Health Care/3M
Page 4 of 4
Distribution Limited To Current 3M Approved Service Providers
Equipment
Bulletin
Subject:
Date:
47
4XL, 5XL and 8XL EO Gas cartridges
May 18, 2004
Machines Affected: All 4XL, 5XL and 8XL sterilizers
Condition:
Damage to the Labels on the Gas cartridges.
Cause:
Solution:
Liquid EO escapes around the cartridge seal and melts the Label.
Perform the following procedure.
Note: This issue may occur if liquid ethylene oxide leaks around the cartridge guide seal and damages
the shrink-wrapped label surrounding the Steri-Gas cartridge. Consequences are discoloration and / or
deterioration of the shrink-wrapped label and staining of the sterilizer. There is no safety risk, as the
Ethylene Oxide is released only in the closed Steri-Vac chamber. After sterilization, the gas is
removed by the normal aeration cycle. The functionality of your 3M Steri-Vac Sterilizer will not be
compromised. Standard sterilization monitoring procedures will help to ensure an adequate sterilization
cycle. Future lots of Steri-Gas cartridges will not have shrink-wrapped labels.
1. Loosen the setscrew or screws that secure the cartridge guide in the cartridge well of the
Steri-Vac chamber.
2. Remove the cartridge guide and the cartridge seal. Discard the old seal.
3. Install the new seal in the cartridge well. Use part number 12-2376-2845-3 for the 4XL and
5XL and part number 78-8078-3324-5 for the 8XL.
4. Be certain that the seal is properly oriented. ( see Fig. 1)
Fig. 1
5. Check the 8XL cartridge holder for proper spring tension. They are located at the top of
the cartridge holder. They should hold the cartridge securely in place.
Health Care/3M
Page 1 of 1
Distribution Limited To Current 3M Approved Service Providers
Equipment
Bulletin
46A
Subject:
4XL,5XL and 8XL Chamber Leaks
Date:
December 3, 2004
Machines Affected:
All 4XL, 5XL and 8XL sterilizers
Condition:
Chamber vacuum leaks.
Cause:
Solution:
The chamber does maintain the proper vacuumduring testing
Perform the following procedure.
STERI-VAC STERILIZER CHAMBER
VACUUM LEAK ISOLATION
The Steri-Vac sterilizer chambers are required to maintain a vacuum level through out the exposure portion of
the cycle. The permissible level of vacuum loss during exposure depends on the model of sterilizer. The 4XL
and 5XL sterilizers are allowed to have a 40 millibar leak in 10 minutes. The 8XL sterilizer is allowed to have a
15 millibar leak in 10 minutes. The primary source of chamber leaks is the chamber check valve. Clean or
replace this component first. Then perform a leak test. To conduct the leak tests refer to the leak test procedures
described in the respective Service Manuals. A kit of plugs, part number 78-8069-7680-5, is available to assist
in isolating the vacuum leak.
Compressed air is used to perform several functions during the operation of the sterilizer. Failures of the sealing
surfaces in the compressed air components can allow compressed air to leak into the chamber and appear to be
a vacuum leak in the new 5XL and 8XL The 4XL and old 5XL will not experience leaks from compressed air
components although they may leak ambient air into the chamber. Isolate the source of a leak in the chamber by
eliminating the compressed air components as a possible source. The following procedure will eliminate the
compressed air components.
1.
Perform the leak test as described in the Service Manual. Perform the “Dead End Vacuum Test”.
If you cannot obtain a dead end vacuum of less than 100 millibars, the leak source could be
located in the fittings on the chamber side of the chamber check valve or at the fittings around the
pressure sensor and puncture solenoid valve. Note the leak rate of the dead end vacuum tester. If
there is a rapid loss of vacuum when the test is stopped, the leak may be located in one of these
areas.
2.
Continue with the chamber leak test. If a leak is detected, re-run the test but during the 10 minute
dwell time, disconnect or turn off the compressed air and bleed the air from the compressed air
line for the new 5XL or 8XL.
Health Care/3M
Page 1 of 5
Distribution Limited To Current 3M Approved Service Providers
3.
If the leak continues at the same rate the leak is not caused by the compressed air components.
4.
If the leak has stopped or been significantly reduced, one or more of the compressed air
components is faulty.
5.
The puncture solenoid valve, which is used to switch between compressed air and chamber
vacuum for the operation of the puncture cylinder, can leak compressed air into the chamber on
the new 5XL and the 8XL. The solenoid valve plunger, seat or dirt could cause the leak.
6.
Remove the puncture solenoid valve and the 1/8 inch NPT nipple from the “T” fitting and plug the
hole with 1/8 inch pipe plug, part number 26-1014-7121-2. Check the “T” fitting for cracks. (see
Fig. 1) If cracks are visible, replace the “T” fitting.
CRACK
Fig. 1
7.
Perform the leak test again. If the leak rate is reduced replace the solenoid valve assembly. If the
leak is not reduced continue the procedure.
8.
The puncture cylinder can leak compressed air into the chamber on the new 5XL or 8XL. A plug,
part number 78-8083-0419-6, is available to eliminate the puncture cylinder. Install the plug and
o-ring.
9.
Perform the leak test again. If the leak is reduced or eliminated replace the puncture cylinder. If
the leak is not reduced continue the procedure.
10. The 4XL and old 5XL sterilizers will not experience a compressed air leak from these
components. However, there could be an ambient air leak through these components. Following
steps 6, 7, 8 and 9 can isolate leaks in these components. If the leak is not reduce or eliminated the
leak is in another area of the sterilizer.
11. On the new and old 5XL and 8XL disconnect the tubing that connects the cartridge well to
heatsink and use a ¼ inch tube compression plug, part number 78-8005-7373-1 to seal the
cartridge well port.
12. Remove the tubing from the heatsink to the chamber, one port for the 5XL and 2 ports for the
8XL. Plug the chamber port(s) with the ¼ inch tube compression plugs, part number 78-80057373-1.
13. Run the leak test again. If the leak is reduced or eliminated a leak exists in the heatsink assembly. .
If the leak is not reduce or eliminated the leak is in another area of the sterilizer.
14. If the leak is in the heatsink assembly, examine the components for cracks or bad seals around the
compression fitting ferrules. The ferrules should not turn on the tubing. Also check the vent
solenoid valve and the moisture solenoid valves and tubing.
15. On the 4XL sterilizer replace the 6 o-rings on the heatsink coupling, check the vent solenoid valve
and the tubing to the moisture solenoid valve for bad seals around the ferrules.
Note: There are several ports through the chamber wall. Some of these ports have brass bushings that can crack
if the components connected to them are over tightened.
Health Care/3M
Page 2 of 5
Distribution Limited To Current 3M Approved Service Providers
16. If the leak continues, examine the bushings that go through the chamber walls. Look for hairline
cracks in the brass. (see Fig. 2) If there is a doubt replace the bushing with an appropriately sized
plug, which is usually a 3/8 inch NPT plug, part number 26-1011-7274-5.
CRACK
Fig.2
17. The new 5XL and 8XL have humidity sensing components that could also cause a chamber
vacuum leak. Isolate these components using a 3/8 NPT plug, 26-1011-7274-5, at the chamber
connection and a ¼ inch NPT plug, 12-7992-5759-2, at the check valve connection.
18. Perform another leak test. If the leak is reduced or eliminated, the leak is in the humidity sensing
components. Examine the flex tubing for any cracks. Replace the tubing if there is a doubt. Be
certain that the Exhaust ports (EXH) for the 5XL and the Atmosphere port (ATM) for the 8XL
solenoid valves are connected to the chamber side of the fittings not the sensor side of the fittings.
19. Reinstall the humidity sensing components and perform another leak test. If the leak continues,
the door gaskets are the last repairable source. The old and new 5XL and the 8XL have gaskets on
the back panel of the chamber as well as the load door side.
20. Examine the load door gasket for any cuts or tears. Over time, continuing vacuums in the chamber
flattens the gaskets. Once the lips of the gasket are flattened to the level of the rim around the
chamber, the gasket may not seal consistently.
21. If the gasket is flattened, replace the door gasket using proper procedures described in Technical
Bulletin 74 and Equipment Bulletin 45.
22. The back plate gasket can be tested without removing the back plate. Remove the outside cabinet
panel from the sterilizer.
23. Use a thin (0.010-0.020 inch) plastic or steel ruler or feeler gauge to test for good contact of the
gasket to the back plate.
24. Draw a deep vacuum in the chamber. Observe the gasket contact to the back chamber plate. Look
at the corners to be certain that the outer lip of the gasket is making contact with the back plate.
25. Slide the ruler or feeler gauge along the edge of the gasket that is against the back plate. Use only
slight pressure against the gasket. When an area of poor contact is found the ruler or feeler gauge
will slip between the gasket and back plate and there will be a whistling or screeching noise as air
leaks past the gasket. The corners are most often the problem spots. (see Fig. 3)
Health Care/3M
Page 3 of 5
Distribution Limited To Current 3M Approved Service Providers
Fig. 3
26. If no noise is heard the gasket is probably OK. However, if all other tests have failed to solve the
problem, replace the gasket and perform another leak test.
27. Go through this entire procedure at least twice before considering the chamber to be defective
Note: Be certain to use Teflon tape on all pipe thread fittings and plugs to insure a good seal. Do not over
tighten pipe thread fittings as the brass fittings will stretch or crack causing additional leaks. By using good
workmanship practices and observation techniques, most sources of leaks can be found in a reasonable time.
Health Care/3M
Page 4 of 5
Distribution Limited To Current 3M Approved Service Providers
The plugs listed below are included in the plug kit part number 78-8069-7680-5.
26-1011-6993-1
26-1014-7121-2
78-8005-7373-1
12-7992-5759-2
26-1011-7274-5
78-8083-0419-6
Health Care/3M
Page 5 of 5
Distribution Limited To Current 3M Approved Service Providers
37A
Equipment
Bulletin
Subject:
4XL, 5XL and 8XL PMA Procedure
Date:
July 10, 2004
Machines Affected:
All Sterilizers.
Condition:
None
Cause:
None
Solution:
Perform routine maintenance every 6 Months.
PMA Procedure
Service Procedure
Note: This procedure should be performed at 6-month interval
1. Compressed Air.
Disassemble the air line filters,
clean and replace the 0.3
micron filter element (261011-8085-7), 0.1 micron filter
element (26-1011-8084-7) and
O-rings.
2. Check the Vent Line
A) Checks that all vent line
connections between the
sterilizer and the hospital
installed vent line are tight.
B) Check for kinks or any
restrictions that may hinder
Health Care/3M
proper flow through the vent
line.
C) Check to be sure the venting
specifications are being met.
3. Moisture System
A) Replace the moisture
solenoid valve plunger and
O-rings.
(12-2376-8096-7)
B) Replace the moisture
Cylinder. (78-8063-0814-0)
(4XL and 5XL prior to
serial number 720xxx or
820xxx)
Page 1 of 4
Distribution Limited To Current 3M Approved Service Providers
4. Air Venturi Assembly
A) Disassemble the air venturi
solenoid valve. Clean the
plunger and seat. Replace if
necessary. (12-2376-9141-0)
B) Rebuild the air regulator with
the kit.
(78-8078-6376-2)
C) Clean the venturi with a pipe
cleaner and alcohol.
5. Door Latch Assembly
A) Disassemble the door latch
solenoid valve. Clean the stem
and seat. Replace the plunger
and O-rings, (78-8078-5880-4)
B) Check the door latch
mechanism for free operation.
7. Disassemble the puncture
solenoid valve. Replace the
plunger and the O-rings.
(12-2376-8096-7)
8. Remove the water tank
assembly from the 4XL and
5XL.
A) Drain the water from the
tank.
B) Rinse with clean water.
C) Replace the clear tubing.
(26-1003-5452-6)
NOTE: The 8XL water should not
be removed and cleaned unless
there is a noticeable odor at the
water fill port or there are repeated
water injection problems.
9. Puncture Assembly
C) (8XL only) Check the brass
striker for wear. Replace if
necessary (78-8078-3288-2)
D) (8XL only) Check the allen bolt
that holds the door handle
assembly for tightness.
6. Vent Assembly
A) Replace the vent bacterial filter.
(78-8055-5905-7) The arrow
points toward the solenoid
valve.
Health Care/3M
A) Remove the cartridge guide
and seal with a 5/64-inch
allen wrench. Clean the
cartridge well.
B) Replace the cartridge seal.
4XL, 5XL (12-2376-28453), 8XL (78-8078-3324-5)
Position the cartridge seal
with proper orientation in the
cartridge well. Apply antiseize grease to the cartridge
guide set screw and fasten
the guide in place.
Page 2 of 4
Distribution Limited To Current 3M Approved Service Providers
C) Check the puncture cylinder for
tightness. Tighten the cylinder
hand tight. Do not over tighten.
(8XL, 12-2376-9138-6, 5XL,
12-2376-9140-2)
16. Functional test.
10. Reinstall the water tank assembly.
A) Fill the tank with clean distilled
water.
B) Purge any air bubbles from the
line.
11. Clean the doorplate, door gasket
and the chamber with warm soapy
water. Replace the door gasket if
the seal is damaged or if it is
compressed to the point that the
door will not seal.
4XL (12-2376-6704-8)
5XL (78-8055-5777-0)
8XL (78-8078-3144-7)
12. Replace the chamber check valve.
(12-2376-3912-0)
13. Check and adjust the 5-volt power
supply at the controller board. 5
volts DC + 0.1 volt DC. (For the
8XL and 5XL after 720xxx and
820xxx, check the Help Window
REF, 2.500 + 0.04.)
14. (4XL only) Check the Heatsink
assembly screws for tightness. Do
not over torque.
15. (8XL and 5XL after 720xxx and
820xxx) Disassemble the humidity
solenoid valves. Clean the plunger
and seat. Replace if necessary.
Health Care/3M
A.) Run the Display/Switch test as
described in the Service
Manual, Section VIII. (4XL,
5XL prior to serial number
720xxx or 820xxx)
B.) Run the Video/Switch test as
described in the Service
Manual, Checkout section.
(8XL and 5XL after 720xxx
and 820xxx)
17. Check the Water Calibration as
described in the Service
Manual, Section VIII. (4XL,
5XL prior to serial number
720xxx or 820xxx)
18. Run the Vacuum Test as
described in the Service
Manual, Section VIII. (4XL
5XL prior to serial number
720xxx or 820xxx) Service
Manual, Checkout section.
(8XL and 5XL after 720xxx
and 820xxx)
19. Check the puncture operation
with an empty cartridge. Run a
normal cycle. The sterilizer will
abort 1 minute after puncture.
Remove the empty cartridge
and be certain that the puncture
pin is down.
Page 3 of 4
Distribution Limited To Current 3M Approved Service Providers
20. (8XL only) Check to be certain all
set up information is correct. Run a
Trend Printout as described in the
Service Manual, Checkout section.
21. (8XL and 5XL after 720xxx and
820xxx) If any sensors have been
replaced, check the calibration as
described in the Service Manual,
Checkout section.
Health Care/3M
22. Insure that all the service
switches are in the normal position
by starting a cycle and check for
proper initial vacuum. Cancel the
cycle with the STOP switch.
23. Seal the sterilizer with the
wire seal as directed in the Field
Service Reference Manual.
Page 4 of 4
Distribution Limited To Current 3M Approved Service Providers
Distribution Limited To Current 3M Approved Service Providers
Download