Uploaded by Jorge Vieira

Brivo OEC850 SM

advertisement
GE Healthcare
Brivo OEC 850 Mobile C-Arm X-ray Product
Service Manual
5358881-1EN
Rev. 2
© 2010
©General Electric Company
All Rights Reserved
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Revision History
Revision
Date (Month & Year)
Description of Change
1
Feb. 2010
Initial release
2
Apr. 2010
Update information
IMPORTANT SAVE THESE INSTRUCTIONS. . PLEASE READ THIS MANUAL BEFORE USING Brivo OEC 850 Mobile C-Arm
X-ray Product.
This manual may not be reproduced, in whole or in part, without the written permission of GE Healthcare.
Other product and company names mentioned herein are the property of their respective owners.
The contents of this document are accurate at the time of publication. However, changes in design and
additional features can, at any time, be incorporated in the hardware and software and may not be reflected in
this version of the document. Contact GE Healthcare Technical Support for clarification, if discrepancies arise.
The text of this manual was originally written, approved and published by the manufacturer in English.
GE HUALUN Medical Systems Co. Ltd., a General Electric company, going to market as GE Healthcare.
GE HUALUN Medical Systems Co. Ltd
No. 1 Yongchang North Road
Beijing Economic & Technological Development Area
Beijing, P.R. China 100176
Tel: 8610-58068888
Fax: 8610-67881850
i
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
REGULATORY REQUIREMENTS
This product complies with the regulatory requirements of the following:
Council Directive 93/42/EEC concerning medical devices: the CE 0459
CE label affixed to the product testifies compliance to all requirements of the Directive.
The location of the CE 0459 label is described in this manual.
European Representative:
GE Medical Systems S.C.S.
Quality Assurance Manager
283 rue de la Minière
78530 BUC France
Tel: +33 1 30 70 40 40
International Electrotechnical Commission (IEC), international standards organization, when applicable.
ii
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
ПРЕДУПРЕЖДЕНИЕ Това упътване за работа е налично само на английски език.
(BG)
• Ако доставчикът на услугата на клиента изиска друг език, задължение на клиента
е да осигури превод.
• Не използвайте оборудването, преди да сте се консултирали и разбрали
упътването за работа.
• Неспазването на това предупреждение може да доведе до нараняване на
доставчика на услугата, оператора или пациентa в резултат на токов удар,
механична или друга опасност.
警告
本维修手册仅提供英文版本。
(ZH-CN)
• 如果客户的维修服务人员需要非英文版本,则客户需自行提供翻译服务。
•
未详细阅读和完全理解本维修手册之前,不得进行维修。
•
警告
(ZH-HK)
警告
(ZH-TW)
UPOZORENJE
(HR)
VÝSTRAHA
(CS)
ADVARSEL
(DA)
WAARSCHUWING
(NL)
忽略本警告可能对维修服务人员、操作人员或患者造成电击、机械伤害或其他形式的伤
害。
本服務手冊僅提供英文版本。
•
倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服務。
•
除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
•
不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其他的危險。
本維修手冊僅有英文版。
•
若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
•
請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危險而受傷。
Ovaj servisni priručnik dostupan je na engleskom jeziku.
• Ako davatelj usluge klijenta treba neki drugi jezik, klijent je dužan osigurati prijevod.
• Ne pokušavajte servisirati opremu ako niste u potpunosti pročitali i razumjeli ovaj
servisni priručnik.
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge, operatera ili
pacijenta uslijed strujnog udara, mehaničkih ili drugih rizika.
Tento provozní návod existuje pouze v anglickém jazyce.
•
V případě, že externí služba zákazníkům potřebuje návod v jiném jazyce, je
zajištění překladu do odpovídajícího jazyka úkolem zákazníka.
•
Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod
a pochopili jeho obsah.
•
V případě nedodržování této výstrahy může dojít k poranění pracovníka
prodejního servisu, obslužného personálu nebo pacientů vlivem elektrického proudu,
respektive vlivem mechanických či jiných rizik.
Denne servicemanual findes kun på engelsk.
• Hvis en kundes tekniker har brug for et andet sprog end engelsk, er det kundens
ansvar at sørge for oversættelse.
• Forsøg ikke at servicere udstyret uden at læse og forstå denne servicemanual.
• Manglende overholdelse af denne advarsel kan medføre skade på grund af elektrisk
stød, mekanisk eller anden fare for teknikeren, operatøren eller patienten.
Deze onderhoudshandleiding is enkel in het Engels verkrijgbaar.
• Als het onderhoudspersoneel een andere taal vereist, dan is de klant verantwoordelijk
voor de vertaling ervan.
• Probeer de apparatuur niet te onderhouden alvorens deze onderhoudshandleiding
werd geraadpleegd en begrepen is.
• Indien deze waarschuwing niet wordt opgevolgd, zou het onderhoudspersoneel, de
operator of een patiënt gewond kunnen raken als gevolg van een elektrische schok,
mechanische of andere gevaren.
iii
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
WARNING
(EN)
HOIATUS
(ET)
VAROITUS
(FI)
ATTENTION
(FR)
WARNUNG
(DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
iv
This service manual is available in English only.
• If a customer's service provider requires a language other than english, it is the
customer's responsibility to provide translation services.
• Do not attempt to service the equipment unless this service manual has been
consulted
and is understood.
• Failure to heed this warning may result in injury to the service provider, operator or
patient from electric shock, mechanical or other hazards.
See teenindusjuhend on saadaval ainult inglise keeles
• Kui klienditeeninduse osutaja nõuab juhendit inglise keelest erinevas keeles, vastutab
klient tõlketeenuse osutamise eest.
• Ärge üritage seadmeid teenindada enne eelnevalt käesoleva teenindusjuhendiga
tutvumist ja sellest aru saamist.
• Käesoleva hoiatuse eiramine võib põhjustada teenuseosutaja, operaatori või patsiendi
vigastamist elektrilöögi, mehaanilise või muu ohu tagajärjel.
Tämä huolto-ohje on saatavilla vain englanniksi.
• Jos asiakkaan huoltohenkilöstö vaatii muuta kuin englanninkielistä materiaalia,
tarvittavan käännöksen hankkiminen on asiakkaan vastuulla.
• Älä yritä korjata laitteistoa ennen kuin olet varmasti lukenut ja ymmärtänyt tämän
huolto-ohjeen.
• Mikäli tätä varoitusta ei noudateta, seurauksena voi olla huoltohenkilöstön, laitteiston
käyttäjän tai potilaan vahingoittuminen sähköiskun, mekaanisen vian tai muun
vaaratilanteen vuoksi.
Ce manuel d’installation et de maintenance est disponible uniquement en anglais.
• Si le technicien d'un client a besoin de ce manuel dans une langue autre que l'anglais,
il incombe au client de le faire traduire.
• Ne pas tenter d'intervenir sur les équipements tant que ce manuel d’installation et de
maintenance n'a pas été consulté et compris.
• Le non-respect de cet avertissement peut entraîner chez le technicien, l'opérateur ou
le patient des blessures dues à des dangers électriques, mécaniques ou autres.
Diese Serviceanleitung existiert nur in englischer Sprache.
• Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es Aufgabe des
Kunden für eine entsprechende Übersetzung zu sorgen.
• Versuchen Sie nicht diese Anlage zu warten, ohne diese Serviceanleitung gelesen
und verstanden zu haben.
• Wird diese Warnung nicht beachtet, so kann es zu Verletzungen des
Kundendiensttechnikers, des Bedieners oder des Patienten durch Stromschläge,
mechanische oder sonstige Gefahren kommen.
Το παρόν εγχειρίδιο σέρβις διατίθεται μόνο στα αγγλικά.
• Εάν ο τεχνικός σέρβις ενός πελάτη απαιτεί το παρόν εγχειρίδιο σε γλώσσα εκτός των
αγγλικών, αποτελεί ευθύνη του πελάτη να παρέχει τις υπηρεσίες μετάφρασης.
• Μην επιχειρήσετε την εκτέλεση εργασιών σέρβις στον εξοπλισμό αν δεν έχετε
συμβουλευτεί και κατανοήσει το παρόν εγχειρίδιο σέρβις.
• Αν δεν προσέξετε την προειδοποίηση αυτή, ενδέχεται να προκληθεί τραυματισμός
στον τεχνικό σέρβις, στο χειριστή ή στον ασθενή από ηλεκτροπληξία, μηχανικούς ή
άλλους κινδύνους.
Ezen karbantartási kézikönyv kizárólag angol nyelven érhető el.
• Ha a vevő szolgáltatója angoltól eltérő nyelvre tart igényt, akkor a vevő felelőssége a
fordítás elkészíttetése.
• Ne próbálja elkezdeni használni a berendezést, amíg a karbantartási kézikönyvben
leírtakat nem értelmezték.
• Ezen figyelmeztetés figyelmen kívül hagyása a szolgáltató, működtető vagy a beteg
áramütés, mechanikai vagy egyéb veszélyhelyzet miatti sérülését eredményezheti.
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
AÐVÖRUN
(IS)
AVVERTENZA
(IT)
(JA)
Þessi þjónustuhandbók er aðeins fáanleg á ensku.
• Ef að þjónustuveitandi viðskiptamanns þarfnast annas tungumáls en ensku, er það
skylda viðskiptamanns að skaffa tungumálaþjónustu.
• Reynið ekki að afgreiða tækið nema að þessi þjónustuhandbók hefur verið skoðuð og
skilin.
• Brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda, stjórnanda eða
sjúklings frá raflosti, vélrænu eða öðrum áhættum.
Il presente manuale di manutenzione è disponibile soltanto in lingua inglese.
• Se un addetto alla manutenzione richiede il manuale in una lingua diversa, il cliente è
tenuto a provvedere direttamente alla traduzione.
• Procedere alla manutenzione dell'apparecchiatura solo dopo aver consultato il
presente manuale ed averne compreso il contenuto.
• Il mancato rispetto della presente avvertenza potrebbe causare lesioni all'addetto alla
manutenzione, all'operatore o ai pazienti provocate da scosse elettriche, urti
meccanici o altri rischi.
このサービスマニュアルには英語版しかありません。
•サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業者の
責任で行うものとさせていただきます。
•このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでください。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、感
電や機械的又はその他の危険により負傷する可能性があります。
경고
(KO)
본 서비스 매뉴얼은 영어로만 이용하실 수 있습니다.
• 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는
것은고객의 책임입니다.
•
본 서비스 매뉴얼을 참조하여 숙지하지 않은 이상 해당 장비를 수리하려고 시도하지
마십시오.
•
RĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
본 경고 사항에 유의하지 않으면 전기 쇼크, 기계적 위험, 또는 기타 위험으로 인해
서비스 제공자, 사용자 또는 환자에게 부상을 입힐 수 있습니다.
Šī apkopes rokasgrāmata ir pieejama tikai angļu valodā.
• Ja klienta apkopes sniedzējam nepieciešama informācija citā valodā, klienta
pienākums ir nodrošināt tulkojumu.
• Neveiciet aprīkojuma apkopi bez apkopes rokasgrāmatas izlasīšanas un saprašanas.
• Šī brīdinājuma neievērošanas rezultātā var rasties elektriskās strāvas trieciena,
mehānisku vai citu faktoru izraisītu traumu risks apkopes sniedzējam, operatoram vai
pacientam.
Šis eksploatavimo vadovas yra tik anglų kalba.
• Jei kliento paslaugų tiekėjas reikalauja vadovo kita kalba – ne anglų, suteikti vertimo
paslaugas privalo klientas.
• Nemėginkite atlikti įrangos techninės priežiūros, jei neperskaitėte ar nesupratote šio
eksploatavimo vadovo.
Jei nepaisysite šio įspėjimo, galimi paslaugų tiekėjo, operatoriaus ar paciento sužalojimai
dėl elektros šoko, mechaninių ar kitų pavojų.
v
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
ADVARSEL
(NO)
OSTRZEŻENIE
(PL)
AVISO
(PT-BR)
ATENÇÃO
(PT-PT)
ATENŢIE
(RO)
ОСТОРОЖНО!
(RU)
UPOZORENJE
(SR)
vi
Denne servicehåndboken finnes bare på engelsk.
• Hvis kundens serviceleverandør har bruk for et annet språk, er det kundens ansvar å
sørge for oversettelse.
• Ikke forsøk å reparere utstyret uten at denne servicehåndboken er lest og forstått.
• Manglende hensyn til denne advarselen kan føre til at serviceleverandøren,
operatøren eller pasienten skades på grunn av elektrisk støt, mekaniske eller andre
farer.
Niniejszy podręcznik serwisowy dostępny jest jedynie w języku angielskim.
• Jeśli serwisant klienta wymaga języka innego niż angielski, zapewnienie usługi
tłumaczenia jest obowiązkiem klienta.
• Nie próbować serwisować urządzenia bez zapoznania się z niniejszym podręcznikiem
serwisowym i zrozumienia go.
• Niezastosowanie się do tego ostrzeżenia może doprowadzić do obrażeń serwisanta,
operatora lub pacjenta w wyniku porażenia prądem elektrycznym, zagrożenia
mechanicznego bądź innego.
Este manual de assistência técnica encontra-se disponível unicamente em inglês.
• Se outro serviço de assistência técnica solicitar a tradução deste manual, caberá ao
cliente fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual de
assistência técnica.
• A não observância deste aviso pode ocasionar ferimentos no técnico, operador ou
paciente decorrentes de choques elétricos, mecânicos ou outros.
Este manual de assistência técnica só se encontra disponível em inglês.
• Se qualquer outro serviço de assistência técnica solicitar este manual noutro idioma,
é da responsabilidade do cliente fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual de
assistência técnica.
• O não cumprimento deste aviso pode colocar em perigo a segurança do técnico, do
operador ou do paciente devido a choques eléctricos, mecânicos ou outros.
Acest manual de service este disponibil doar în limba engleză.
• Dacă un furnizor de servicii pentru clienţi necesită o altă limbă decât cea engleză,
este de datoria clientului să furnizeze o traducere.
• Nu încercaţi să reparaţi echipamentul decât ulterior consultării şi înţelegerii acestui
manual de service.
• Ignorarea acestui avertisment ar putea duce la rănirea depanatorului, operatorului sau
pacientului în urma pericolelor de electrocutare, mecanice sau de altă natură.
Данное руководство по техническому обслуживанию представлено только на
английском языке.
•
Если сервисному персоналу клиента необходимо руководство не на
английском, а на каком-то другом языке, клиенту следует самостоятельно
обеспечить перевод.
•
Перед техническим обслуживанием оборудования обязательно обратитесь
к данному руководству и поймите изложенные в нем сведения.
•
Несоблюдение требований данного предупреждения может привести к тому,
что специалист по техобслуживанию, оператор или пациент получит удар
электрическим током, механическую травму или другое повреждение.
Ovo servisno uputstvo je dostupno samo na engleskom jeziku.
• Ako klijentov serviser zahteva neki drugi jezik, klijent je dužan da obezbedi
prevodilačke usluge.
• Ne pokušavajte da opravite uređaj ako niste pročitali i razumeli ovo servisno uputstvo.
• Zanemarivanje ovog upozorenja može dovesti do povređivanja servisera, rukovaoca
ili pacijenta usled strujnog udara ili mehaničkih i drugih opasnosti.
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
UPOZORNENIE
(SK)
ATENCION
(ES)
VARNING
(SV)
OPOZORILO
(SL)
DİKKAT
(TR)
Tento návod na obsluhu je k dispozícii len v angličtine.
• Ak zákazníkov poskytovateľ služieb vyžaduje iný jazyk ako angličtinu, poskytnutie
prekladateľských služieb je zodpovednosťou zákazníka.
• Nepokúšajte sa o obsluhu zariadenia, kým si neprečítate návod na obluhu a
neporozumiete mu.
• Zanedbanie tohto upozornenia môže spôsobiť zranenie poskytovateľa služieb,
obsluhujúcej osoby alebo pacienta elektrickým prúdom, mechanické alebo iné
ohrozenie.
Este manual de servicio sólo existe en inglés.
• Si el encargado de mantenimiento de un cliente necesita un idioma que no sea el
inglés, el cliente deberá encargarse de la traducción del manual.
• No se deberá dar servicio técnico al equipo, sin haber consultado y comprendido este
manual de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor de
servicios, el operador o el paciente sufran lesiones provocadas por causas eléctricas,
mecánicas o de otra naturaleza.
Den här servicehandboken finns bara tillgänglig på engelska. .
• Om en kunds servicetekniker har behov av ett annat språk än engelska, ansvarar
kunden för att tillhandahålla översättningstjänster.
• Försök inte utföra service på utrustningen om du inte har läst och förstår den här
servicehandboken.
• Om du inte tar hänsyn till den här varningen kan det resultera i skador på
serviceteknikern, operatören eller patienten till följd av elektriska stötar, mekaniska
faror eller andra faror.
Ta servisni priročnik je na voljo samo v angleškem jeziku.
• Če ponudnik storitve stranke potrebuje priročnik v drugem jeziku, mora stranka
zagotoviti prevod.
• Ne poskušajte servisirati opreme, če tega priročnika niste v celoti prebrali in razumeli.
• Če tega opozorila ne upoštevate, se lahko zaradi električnega udara, mehanskih ali
drugih nevarnosti poškoduje ponudnik storitev, operater ali bolnik.
Bu servis kılavuzunun sadece ingilizcesi mevcuttur.
• Eğer müşteri teknisyeni bu kılavuzu ingilizce dışında bir başka lisandan talep ederse,
bunu tercüme ettirmek müşteriye düşer.
• Servis kılavuzunu okuyup anlamadan ekipmanlara müdahale etmeyiniz.
• Bu uyarıya uyulmaması, elektrik, mekanik veya diğer tehlikelerden dolayı teknisyen,
operatör veya hastanın yaralanmasına yol açabilir.
vii
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Contents
Chapter1. Introduction and Safety ........................................................................................................................................... 1-1
1.1.
Introduction ................................................................................................................................................................... 1-2
1.1.1.
Objective .................................................................................................................................................................... 1-2
1.1.2.
Scope ........................................................................................................................................................................... 1-2
1.1.3.
Service Philosophy ................................................................................................................................................ 1-2
1.1.4.
Qualifications........................................................................................................................................................... 1-2
1.1.5.
Unauthorized Modifications ............................................................................................................................. 1-2
1.1.6.
Manual Organization and Use......................................................................................................................... 1-2
1.2.
Safety ................................................................................................................................................................................ 1-3
1.2.1.
Safety Hazard Alerts............................................................................................................................................. 1-3
1.2.2.
Ingress of Fluids...................................................................................................................................................... 1-4
1.2.3.
Electrical Shock....................................................................................................................................................... 1-4
1.2.4.
Explosion.................................................................................................................................................................... 1-5
1.2.5.
Thermal Harm ......................................................................................................................................................... 1-5
1.2.6.
Equipment Stability............................................................................................................................................... 1-5
1.2.7.
X-Radiation ............................................................................................................................................................... 1-6
1.2.8.
Motorized Mechanical Movement................................................................................................................. 1-6
1.2.9.
Safety Interlock....................................................................................................................................................... 1-7
1.2.10.
Labels .......................................................................................................................................................................... 1-7
1.2.11.
Locations of the Labels ...............................................................................................................................1-16
1.2.12.
Symbols ....................................................................................................................................................................1-18
1.2.13.
Electromagnetic Compatibility Statement ..............................................................................................1-20
Chapter2.
System Overview ................................................................................................................................................... 2-1
2.1.
System Descriptions............................................................................................................................................. 2-2
2.1.1.
System overview .............................................................................................................................................. 2-2
2.1.2.
Component Locator........................................................................................................................................ 2-3
2.2.
System block ................................................................................................................................................................. 2-7
2.3.
System Schematics .................................................................................................................................................... 2-8
Chapter3.
Installation ................................................................................................................................................................ 3-1
3.1.
Overview .................................................................................................................................................................... 3-2
3.2.
Pre-Installation........................................................................................................................................................ 3-2
viii
3.2.1.
Dimensions and Weights of System.................................................................................................. 3-2
3.2.2.
Site Requirement ........................................................................................................................................ 3-2
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.2.3.
Power Supply Requirement....................................................................................................................3-2
3.2.4.
Environments ................................................................................................................................................3-3
3.2.5.
Tools and Test Equipment.......................................................................................................................3-4
3.3.
Unpacking..................................................................................................................................................................3-5
3.4.
Inspection................................................................................................................................................................ 3-10
3.4.1.
Label Inspection........................................................................................................................................ 3-10
3.4.2.
External Inspection.................................................................................................................................. 3-11
3.4.3.
Internal Inspection................................................................................................................................... 3-12
3.5.
Setup ......................................................................................................................................................................... 3-13
3.5.1.
Power Supply Adaption ......................................................................................................................... 3-13
3.5.2.
UPS .................................................................................................................................................................. 3-15
3.5.3.
Power on....................................................................................................................................................... 3-16
3.5.4.
System Configuration............................................................................................................................. 3-17
3.6.
Functional Checks............................................................................................................................................... 3-18
3.6.1.
C-Arm Checks ............................................................................................................................................ 3-18
3.6.2.
Workstation Checks ................................................................................................................................ 3-23
3.6.3.
Auto Technique Tracking...................................................................................................................... 3-29
3.6.4.
Image Resolution ..................................................................................................................................... 3-29
3.6.5.
Dose Checks ............................................................................................................................................... 3-29
3.7.
Options Installation ............................................................................................................................................ 3-31
3.7.1.
Laser Aimer ................................................................................................................................................. 3-31
3.7.2.
Video distributor ....................................................................................................................................... 3-34
3.7.3.
Printers .......................................................................................................................................................... 3-35
3.7.4.
Film cassette holder................................................................................................................................ 3-37
3.7.5.
DAP.................................................................................................................................................................. 3-37
3.7.6.
Skin Spacer.................................................................................................................................................. 3-42
3.7.7.
110V Power Cable Kit ............................................................................................................................. 3-42
3.7.8.
Two-pedal Footswitch ........................................................................................................................... 3-42
3.7.9.
EU standard Plug...................................................................................................................................... 3-43
Chapter4.
Subsystem .................................................................................................................................................................4-1
4.1.
Power Distribution and Control.............................................................................................................................4-2
4.2.
MCB.....................................................................................................................................................................................4-5
4.3.
Display Board.................................................................................................................................................................4-8
4.4.
kV control and IGBT assembly...............................................................................................................................4-9
ix
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4.5.
Filament Control Board ..........................................................................................................................................4-11
4.6.
Monoblock ....................................................................................................................................................................4-12
4.7.
X-ray/key Switches and Emergency Switch.................................................................................................4-14
4.8.
Image Chain.................................................................................................................................................................4-16
4.8.1.
Overview.............................................................................................................................................................4-16
4.8.2.
Collimator Control..........................................................................................................................................4-16
4.8.3.
Image Intensifier.............................................................................................................................................4-18
4.8.4.
CCD/CCU ............................................................................................................................................................4-19
4.8.5.
Image Grabber Card.....................................................................................................................................4-22
4.8.6.
Monitor ................................................................................................................................................................4-22
4.9.
Computer and Peripheral......................................................................................................................................4-23
4.9.1.
Computer ...........................................................................................................................................................4-23
4.9.2.
DVD and USB....................................................................................................................................................4-24
4.9.3.
Keyboard and Mouse ...................................................................................................................................4-25
Chapter5.
Calibration................................................................................................................................................................. 5-1
5.1.
Overview.......................................................................................................................................................................... 5-2
5.2.
kV Measurement and Adjustment ...................................................................................................................... 5-3
5.2.1.
kV Measurement and Accuracy Check ................................................................................................. 5-3
5.2.2.
kV Adjustment.................................................................................................................................................... 5-3
5.3.
mA Measurement and Adjustment.................................................................................................................. 5-6
5.3.1.
mA Measurement and Accuracy Check ............................................................................................... 5-6
5.3.2.
mA calibration ................................................................................................................................................... 5-9
5.4.
x
mAs Measurement and Calibration .................................................................................................................5-12
5.4.1.
mAs Measurment and Accuracy Check..............................................................................................5-12
5.4.2.
mAs Calibration...............................................................................................................................................5-12
5.5.
X-Ray Generator Warm..........................................................................................................................................5-13
5.6.
Beam Alignment ........................................................................................................................................................5-14
5.6.1.
Beam Alignment Setup................................................................................................................................5-14
5.6.2.
Camera Orientation ......................................................................................................................................5-17
5.6.3.
Camera Centering .........................................................................................................................................5-18
5.6.4.
Collimator Centering.....................................................................................................................................5-21
5.6.5.
I.I. Field Size .......................................................................................................................................................5-23
5.6.6.
Limiting Field Size (Collimator Iris Field Size) .....................................................................................5-24
5.6.7.
ABS Tracking Check and Adjustment ...................................................................................................5-25
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
ABS Tracking Check ............................................................................................................................................................ 5-25
Entrance Dose Rate Check and Camera Iris Adjustment................................................................................. 5-25
5.6.8.
Limiting Resolution........................................................................................................................................ 5-27
5.6.9.
Shading Correction ....................................................................................................................................... 5-30
5.7.
DAP and Dose Calibration .................................................................................................................................... 5-33
5.8.
Monitor Brightness and Contrast Adjustment ............................................................................................ 5-33
5.8.1.
Keys assignment and operation LED ................................................................................................... 5-33
5.8.2.
Lock and unlock the OSD menu ............................................................................................................. 5-33
5.8.3.
Key functions in the OSD menu .............................................................................................................. 5-33
5.8.4.
Brightness and Contrast Adjustment................................................................................................... 5-34
Chapter6.
Software......................................................................................................................................................................6-1
6.1.
Overview.....................................................................................................................................................................6-2
6.2.
Service Tool ...............................................................................................................................................................6-3
6.2.1.
Service ID Setting..............................................................................................................................................6-4
6.2.2.
Multi-language Setting...................................................................................................................................6-5
6.2.3.
Time Zone Setting.............................................................................................................................................6-5
6.2.4.
Get System Information.................................................................................................................................6-6
6.2.5.
Exposure Management .................................................................................................................................6-7
6.2.6.
Database Backup and Restore ............................................................................................................... 6-15
6.2.7.
Set Imaging Defaults.................................................................................................................................... 6-16
6.2.8.
Consistency Check ........................................................................................................................................ 6-17
6.2.9.
Image Import/Export ................................................................................................................................... 6-18
6.2.10.
Log Viewer......................................................................................................................................................... 6-21
6.2.11.
Config File Export ........................................................................................................................................... 6-23
6.2.12.
Restore Default ............................................................................................................................................... 6-24
6.3.
Software Installation and Upgrade ............................................................................................................ 6-25
6.3.1.
Software Installation .................................................................................................................................... 6-25
6.3.2.
Application Software Upgrade................................................................................................................ 6-38
7.Chapter7. Troubleshooting......................................................................................................................................................7-1
7.1.
Overview ..........................................................................................................................................................................7-2
7.2.
Error Recovery Steps..................................................................................................................................................7-2
7.3.
Messages .........................................................................................................................................................................7-2
7.4.
Troubleshooting Guide........................................................................................................................................... 7-19
Chapter8.
Replacement.............................................................................................................................................................8-1
xi
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.1.
FRU list ........................................................................................................................................................................ 8-2
8.2.
Locations of FRU .................................................................................................................................................... 8-7
8.3.
Replacement..........................................................................................................................................................8-10
8.3.1.
Covers..................................................................................................................................................................8-10
8.3.2.
Brake, Handle and Wheel ..........................................................................................................................8-15
8.3.3.
Generator...........................................................................................................................................................8-19
8.3.4.
Collimator(with MP1 board).......................................................................................................................8-20
8.3.5.
PCBs......................................................................................................................................................................8-22
8.3.6.
Parts in Computer and Peripheral .........................................................................................................8-28
8.3.7.
Exposure Indicator ........................................................................................................................................8-31
8.3.8.
Monitors..............................................................................................................................................................8-32
8.3.9.
CCD Camera with CCU ................................................................................................................................8-33
8.3.10.
Image Intensifier with power supply ....................................................................................................8-38
8.3.11.
UPS ........................................................................................................................................................................8-39
8.3.12.
Misc Parts...........................................................................................................................................................8-40
8.3.13.
Linear Actuator ...............................................................................................................................................8-48
8.3.14.
Switch ..................................................................................................................................................................8-51
8.3.15.
Others ..................................................................................................................................................................8-52
Chapter9.
9.1.
xii
Periodic Maintenance.......................................................................................................................................... 9-1
Overview.......................................................................................................................................................................... 9-2
9.1.1.
Safety ..................................................................................................................................................................... 9-2
9.1.2.
Tools and Test Equipment............................................................................................................................ 9-3
9.2.
PM Checklist................................................................................................................................................................... 9-4
9.3.
Administrative Information .................................................................................................................................... 9-6
9.4.
Safety Inspection......................................................................................................................................................... 9-6
9.4.1.
Antistatic Drag Wires ..................................................................................................................................... 9-6
9.4.2.
Electrical Plug and power cord.................................................................................................................. 9-6
9.4.3.
Interconnect cable and Connector ......................................................................................................... 9-6
9.4.4.
Footswitch cable and pins........................................................................................................................... 9-7
9.4.5.
Ground Continuity............................................................................................................................................ 9-7
9.5.
PC Clean........................................................................................................................................................................... 9-7
9.6.
System Function checks .......................................................................................................................................... 9-7
9.6.1.
Perform Power On Checks ................................................................................................................................ 9-7
9.6.2.
C-Arm Checks .................................................................................................................................................... 9-8
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9.6.3.
Workstation Checks...................................................................................................................................... 9-10
9.6.4.
Dose Rate .......................................................................................................................................................... 9-11
9.7.
DAP Calibration.......................................................................................................................................................... 9-11
9.8.
Radiographic Film Beam Alignment Test...................................................................................................... 9-13
9.9.
X-ray Generator Accuracy Tests ....................................................................................................................... 9-13
9.9.1.
kVp Accuracy ................................................................................................................................................... 9-13
9.9.2.
mA Accuracy.................................................................................................................................................... 9-14
9.9.3.
mAs Accuracy.................................................................................................................................................. 9-14
9.10.
Reporting ...................................................................................................................................................................... 9-14
9.10.1.
Install Equipment Covers ........................................................................................................................... 9-14
9.10.2.
Complete PM Comments ........................................................................................................................... 9-14
Chapter 10. Technical Reference..........................................................................................................................................10-1
10.1. Overview .................................................................................................................................………………………………………..10-2
10.2.
System Specifications ....................................................................................................................................... 10-2
10.2.1.
Safety Classification................................................................................................................................ 10-2
10.2.2.
Environmental Requirements............................................................................................................. 10-2
10.2.3.
Power Requirements .............................................................................................................................. 10-2
10.2.4.
Circuit Breaker ........................................................................................................................................... 10-3
10.2.5.
Earth leakage current ............................................................................................................................ 10-3
10.2.6.
Physical Specifications .......................................................................................................................... 10-4
10.3.
Specification of Key Components ............................................................................................................... 10-6
10.3.1.
X-Ray Generator ....................................................................................................................................... 10-6
10.3.2.
X-Ray tube ................................................................................................................................................... 10-7
10.3.3.
X-ray tube assembly............................................................................................................................... 10-8
10.3.4.
Collimator..................................................................................................................................................... 10-9
10.3.5.
Others ......................................................................................................................................................... 10-10
10.4.
Tube Rating Chart ............................................................................................................................................ 10-11
10.4.1.
Filament Emission Characteristics................................................................................................ 10-11
10.4.2.
Single Load Ratings.............................................................................................................................. 10-13
10.4.3.
Thermal Characteristic ....................................................................................................................... 10-15
10.5.
Radiation Safety Specification ................................................................................................................... 10-16
10.5.1.
Scatter Data............................................................................................................................................. 10-16
10.5.2.
Scatter radiation specification........................................................................................................ 10-17
10.6.
Material Safety Data Sheets............................................................................................................................. 10-18
xiii
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.7.
Appendix
xiv
Material Recycling.................................................................................................................................................. 10-18
Chapter1.
Introduction and Safety
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1.1. Introduction
1.1.1. Objective
Provide Brivo OEC 850 Mobile C-Arm X-ray Product service documentation that is consistently organized,
easy to access and is designed to aggressively expedite fault isolation and decrease product down-time.
1.1.2. Scope
This manual’s contents document the necessary information for service.
1.1.3. Service Philosophy
Brivo OEC 850 Mobile C-Arm X-ray Products use precision components and utilize complex
manufacturing techniques. Component-level repairs that maintain compliance with these requirements
are sometimes not possible in the field. The system is therefore designed to be field serviceable at the
modular or Field Replaceable Unit (FRU) level.
1.1.4. Qualifications
Reading this manual does not qualify an untrained person to service the equipment. This publication
assumes the reader is a skilled and fully qualified electronic technician who has experience servicing
medical imaging systems. Factory trained service personnel will benefit the most from this material.
1.1.5. Unauthorized Modifications
GE HUALUN Medical Systems, Inc. certifies that when assembled according to the manufacturer's
instructions, this equipment meets US Federal Performance standard 21 CFR Subchapter J and
applicable international standards. Unauthorized modifications to the equipment may impact adherence
to these standards and make the equipment unsafe to operate.
The owner is responsible for verifying continued compliance with all applicable regulations and
standards. Consult local, state, federal and/or international agencies regarding specific requirements
and regulations applicable to the use of this type of medical electronic equipment.
1.1.6. Manual Organization and Use
The information documented here has been analyzed and divided into several sections. This manual
contains the following types of information:
1-2
•
Introduction and Safety-Information of the system and safety.
•
System Overview-General description of the system.
•
Installation Procedure –Information of pre-installation, unpacking and installation check.
Introduction and Safety
•
Subsystem-The subsystems have been analyzed and divided into functional sections. Each
functional section contains major assembly and control locations, theory, functional tests and
other information relevant at the subsystem level.
•
Adjustment and Calibration-Procedures for tuning and calibration relevant at the subsystem
and system level.
•
Software Maintenance- Provides procedures for software maintenance and service mode
utilities.
•
Troubleshooting- Provides error code, error messages, and diagnostic programs that expedite
the diagnostic process.
•
Replacement- Removal and installation procedures for fuses, covers and field replaceable units.
•
Periodic Maintenance Procedure-Procedures for periodic maintenance.
•
Technique Reference-Technique Reference of the system.
•
Appendix-Schematics and information of some key parts.
1.2. Safety
Do not attempt to service or maintain this equipment before reading this entire section.
1.2.1. Safety Hazard Alerts
There are three hazard classifications used in this document. They are:
WARNING
WARNING indicates a potentially hazardous situation that, if not avoided, could result in
death or serious injury.
CAUTION
CAUTION indicates a potentially hazardous situation that, if not avoided, may result in
moderate to minor injury, equipment damage or loss of data.
NOTE
NOTE is given in situation requiring special attention.
1-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1.2.2. Ingress of Fluids
Excessive amounts of fluids such as antiseptics, cleaning solutions or bodily fluids may damage internal
components if they are allowed inside the equipment.
WARNING
The equipment is not rated for water-tight operation. Do not allow any liquids to leak into
the equipment where they can cause electric shock, fires, and short circuits. If leakage
does occur, ensure that all internal circuitry is completely dry before attempting to
operate the equipment.
WARNING
Never store or operate the system in the presence of conductive fluids like water or
saline solution unless the system is adequately protected by an approved bagging or
draping system.
1.2.3. Electrical Shock
If the equipment must be serviced with the covers removed, observe the following precautions:
WARNING
WARNING
WARNING
WARNING
WARNING
1-4
Equipment that contains high power electrical components must be serviced only by
personnel familiar with proper safety procedures for working near these components.
Disconnect AC power (and battery packs if used) before discharging static electricity in
components such as electrolytic capacitors and high voltage cables. Failure to heed this
message will result in death or severe personal injury.
Allow adequate time for static charges to discharge through bleeder resistors. Use a
high wattage resistor when discharging circuits to avoid burns.
Have someone watch while you work near high voltages. This person must remain
clear of all circuitry and be prepared to turn off the power to the system and render aid
in an emergency.
High voltage cables, electrolytic capacitors, and CRTs can retain a dangerous static
charge for long periods after power has been removed. Some devices can acquire a
charge spontaneously without direct contact with other circuitry. Do not touch these
components unless power has been completely removed and they have been
discharged.
CAUTION
Never operate the system unless all safety ground wires (green wire with a yellow
stripe) and related components are fastened in place. Dangerous electrical shock and
improper equipment operation can otherwise result.
CAUTION
Remove all metal rings and watchbands before working on system circuitry. Skin burns
and damage from involuntary muscle contractions can result if metal jewelry shorts
electrical circuits.
Introduction and Safety
1.2.4. Explosion
If an abnormal condition occurs, such as the room fills with flammable gas, steps must be taken to
prevent the gas from contacting components within the equipment. Follow these guidelines:
1.
Do not turn the system on or off.
2.
Do not un-plug or plug-in system or peripheral equipment power cords.
3.
Do not operate light switches or other electrical equipment.
4.
Evacuate all personnel immediately.
5.
Notify the hospital administration and/or fire department.
6.
Ventilate the room to clear the air of all flammable vapor or gas.
7.
Identify and remove the source of the flammable liquid, vapor, or gas.
1.2.5. Thermal Harm
Before you remove the cover of the system for maintenance or service. Please pay attention to the points
with high temperature, for example, the current limiters, internal power distribute circuit etc. Otherwise it
may harm your hand or body.
1.2.6. Equipment Stability
Brivo OEC 850 Mobile C-Arm X-ray Product is mounted on wheels. And if it is moved or operated
improperly it could roll out of control. Follow these guidelines:
Two people should maintain control of the equipment when moving up or down an incline.
Place all mechanical assemblies in their most compact (transport) position and lock brake handles prior
to moving the equipment.
Use the handles designed for moving the equipment and mechanical assemblies.
Never attempt to move the system up or down steps.
Do not operate the equipment on unlevel floors.
Do not lock the wheel brakes and leave the equipment unattended on unlevel floors.
Always apply the wheel locks when the system is in its final position.
Do not move the equipment if the castors or wheels are not functioning properly.
Mechanical shocks to the equipment while disk drives are accessing information may cause damage to
the disk drive and loss of data.
1-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1.2.7. X-Radiation
CAUTION
This equipment either produces or is used in the vicinity of ionizing radiation. Observe
proper safety practices during operation. Use lead aprons, eye protection, thyroid
protection, and similar devices to protect yourself and others.
CAUTION
Where local Entrance Exposure Rates (EER) are different than the GE HUALUN rate
specified, the EER should be adjusted to the lower of the two.
1.2.8. Motorized Mechanical Movement
Brivo OEC 850 Mobile C-Arm X-ray Product has motorized mechanical assemblies. Please follow these
guidelines:
Always observe mechanical assemblies when operating the motor to avoid pinching or collision with a
person or object.
Use care when working around equipment to avoid unintentional motor actuation. Do not carelessly
place objects on the equipment or bump or lean against the equipment.
Observe and prevent articles of clothing from getting caught in moving parts.
1-6
Introduction and Safety
1.2.9. Safety Interlock
WARNING
Never bypass, jumper, or otherwise disable any equipment safety feature.
1.2.10. Labels
NOTE
Item
1
2
3
4
Even though this manual is supplied only in English, some labels may have a foreign
language equivalent that will appear on your system.
Label
Quantity
Description
1
This label indicates the position of
the C-Arm Wig-Wag movement
handle. When the handle is
rotated to dashed position, the
Wig-Wag brake is released. When
the handle is in horizontal
position, the Wig-Wag brake is
locked.
1
This label indicates the position of
the C-Arm Wig-Wag movement
handle. When the handle is
rotated to dashed position, the
Wig-Wag brake is released. When
the handle is in horizontal
position, the Wig-Wag brake is
locked.
1
This label indicates the position of
the C-Arm rotation handle. When
the handle is rotated to the
dashed position, the brake is
released. When the handle is in
horizontal position, the brake is
locked.
1
This label indicates the position of
the C-Arm rotation handle. When
the handle is rotated to the
dashed position, the brake is
released. When the handle is in
horizontal position, the brake is
locked.
1-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1
This label indicates the position of
the C-Arm orbital movement
handle. When the handle is
rotated to the dashed position,
the brake is released. When the
handle is in horizontal position,
the brake is locked.
1
This label indicates the position of
the C-Arm orbital movement
handle. When the handle is
rotated to the dashed position,
the brake is released. When the
handle is in horizontal position,
the brake is locked.
7
1
This label indicates the position of
the C-Arm horizontal movement
handle. When the handle is
rotated to the horizontal position,
the brake is released. When the
handle is in vertical position, the
brake is locked.
8
1
9
1
5
6
1-8
This label indicates that the
Mobile
C-Arm’s
mechanical
subassemblies position should be
placed in their most compact
position before transporting or
moving the system.
Introduction and Safety
10
1
This label on the control panel of
the
C-Arm
indicates
the
procedure of shutting down and
restart to avoid data missing.
11
1
Potential equalization terminal
label.
12
4
This label means that there is
accessible moving parts and keep
hands/fingers away.
1
This label near 2050 lateral
rotation limit switch reminds
operator to release the switch to
continue rotating.
13
14
1
This label indicates that you
should not touch the connector
and the patient at the same time.
The
connector
must
be
connected to the appointed
device only.
This port is only used for user to
store patient image by USB stick.
Any other equipment is forbidden
to be connected strictly.
1-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1
This label on the footswitch
means that the degree of
protection of the footswitch is IP
X8.
The electrical mechanism
within the footswitch is protected
from the effects of continuous
immersion in water.
16
1
This symbol indicates the location
of Emergency Stop Switch on the
Brivo OEC 850 equipment.
17
1
System rating plate indicates
manufacture information and
input power requirements.
18
1
This is the label of generator
rating plate.
19
1
This is the label of collimator.
Remove the tube head cover. This
label is on the collimator.
15
1-10
IPX8
Introduction and Safety
1
This is the label of laser warning.
(The label will only be required if
there is laser aimer.)
1
This label warns the operator or
the patient not to look at the laser
beams directly. (The label will only
be required if there is laser aimer.)
2
This
label
indicates
the
equipment was tested by a
Notified Body and was found to
be in compliance with the
requirements of all relevant
directives and standards in effect
within the European Union at the
time of manufacture.
23
2
This label indicates that waste
electronic
and
electrical
equipment must not be disposed
of as unsorted municipal waste
and must be collected separately.
Please contact an authorized
representative
of
the
manufacturer for information
concerning the decommissioning
of your equipment.
24
2
This is the label of branding.
20
21
22
0459
1-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1
This symbol indicates the system
was tested by CCC authorized
body and was found to be in
compliance
with
the
requirements of all relevant
directives and standards at the
time of manufacture.
1
This label indicates the system
was tested by ITS authorized
body and was found to be in
compliance
with
the
requirements of all relevant
directives and standards at the
time of manufacture.
27
1
This label can be found on the XRay generator. It indicates the
location of the X-Ray source and
the controls used to produce
ionizing X-Radiation. Use
appropriate precautions and
protective equipment when using
the system and at all times when
X-Rays are present.
28
1
USB warning label
29
1
C-Arm weight label
25
26
1-12
S
Introduction and Safety
30
1
Workstation weight label
This label is near the capacitor to
warn that you shall discharge the
capacitor before touching it.
31
32
33
34
35
2
The label attached to UPS
indicates that the users are not
allowed to replace battery of UPS.
The UPS is charged even though
the system was not connected to
power supply. Do not touch UPS
at random.
1
This label indicates the key
switch. Turn the key switch
located on the C-arm interface
panel to enable X-rays and
motorized mechanical
movement.
1
This is the label of functional
earth ground connection. It is
attached near the COM pin of
PS1.
1
This label is on the right side of
the two-pedal footswitch. Twopedal footswitch is an optional
part. Press the right switch to
save an image displayed on the
left monitor.
1
This label is on the left side of the
two-pedal footswitch. Two-pedal
footswitch is an optional part.
Press the left switch to start
fluoroscopy in the mode selected.
1-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
36
1
This is the label of J1 connector
37
2
This is the label of Mark A.
38
1
This is the label of Mark E. It’s on
the tube.
39
1
This is the label of fuse
1-14
Introduction and Safety
40
41
42
43
44
1
This is the label of J3 connector.
Remove the workstation back
cover. The label is near the
connector on the bottom.
1
This warning label is on the
control panel. It is only on the
system in America. Attach the
label if the system is installed in
America.
1
This label indicates the system
complies with the US Federal
Performance standard 21 CFR
Subchapter J and applicable
international standards. It is only
on the system in America. Attach
the label on the top cover of Carm console if the system is
installed in America.
1
This label indicates the system
complies with 21 CFR 801.109.
Attach the label on the base of
the C-arm if the system is
installed in America.
1
This label is attached on UPS shut
down button to avoid misusing.
Only when UPS can’t shut down
automatically after system power
off process complete, the user
should press the UPS shut down
button to shut off UPS manually.
1-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1.2.11. Locations of the Labels
12
42
1-16
Introduction and Safety
44
14
30
1-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1.2.12. Symbols
Be familiar with the following symbols that may appear on equipment and schematics so you can safely
maintain and operate the system:
Dangerous Voltage
This symbol identifies areas where hazardous voltages may be present. Use appropriate
safety precautions.
Protective Earth Ground
This symbol identifies the system’s main protective earth ground terminal, which connects to
the facility earth ground through the grounding prong on the AC power plug. This connection
must be in place at all times for safe system operation.
Earth Ground Connection
This symbol identifies an earth ground connection that is necessary to maintain grounding
throughout the system.
X-Ray Source
This X-Ray Source symbol is used to identify controls that will produce ionizing X-radiation
when activated. Always use appropriate safety precautions when generating X-rays.
This X-Ray Source symbol is used to identify controls that will produce ionizing X-radiation
when activated. Always use appropriate safety precautions when generating X-rays.
Potential Equalization Terminal
This symbol identifies system terminals which, when connected together, are at the same
potential (not necessarily earth ground). Green/yellow wires normally connect potential
equalization terminals together.
1-18
Introduction and Safety
Attention
Attention, see accompanying documentation for information. You can refer to this manual for
the operation instructions.
Type B
This symbol indicates the equipment is protected against electric shock by a protective earth
ground connection.
Ionization Radiation
This symbol indicates that there is ionization radiation.
1-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1.2.13. Electromagnetic Compatibility Statement
This equipment may generate and use radio frequency energy. The equipment must be installed and
used according to the manufacturer’s instructions in order to avoid radio frequency interference. If this
equipment generates or receives interference do the following to correct the problem:
•
Verify that the equipment is the cause by turning the system on and off.
•
In the event of unintended motor actuation, immediately remove power to the equipment.
•
In the event of unintended X-ray actuation, immediately remove power to the equipment.
•
Reorient the equipment until the interference stops.
•
Relocate the equipment with respect to other equipment in the room.
•
Plug the equipment into a different outlet so that the equipment and the receiver are on different
branch circuits.
•
Use only input/output (I/O) cables supplied by GE HUALUN Medical Systems Co. Ltd.
This product conforms with IEC60601-1-2: 2001 Amd. 1: 2004 Ed.2.1 EMC standard for medical devices.
NOTE
This equipment generates, uses, and can radiate radio frequency energy. The
equipment may cause or be subject to radio frequency interference with other medical
and non–medical devices and radio communications. To provide
reasonable
protection against such interference, the Brivo OEC 850 system complies with
emissions limits for a Group 1, Class A Medical Devices and has applicable immunity
level as stated in IEC60601-1-2: 2001 Amd.1: 2004 Ed2.1.
However, there is no guarantee that interference will not occur in a particular
installation. Special precautions and other information regarding EMC provided in the
accompanying documents of this equipment shall be observed during installation and
operation of this equipment.
NOTE
If this equipment is found to cause interference (which may be determined by
switching the equipment on and off), the user (or qualified service personnel) should
attempt to correct the problem by one or more of the following measure(s):
1. Reorient or relocate the affected device(s).
2. Increase the separating space between the equipments and the affected device.
3. Power the equipment from a source different from that of the affected device.
4. Consult the point of purchase or service representative for further suggestions.
WARNING
1-20
Use of accessories, transducers, cables and other parts other than those specified by
the manufacturer of this equipment may result in increased emissions or decreased
immunity of the equipment. The manufacturer is not responsible for any interference
caused either by the use of interconnect cables other than those recommended, or
unauthorized changes or modifications to this equipment. Unauthorized changes or
modifications could void the user’s authority to operate the equipment.
Introduction and Safety
To comply with the regulations applicable to an electromagnetic interface for a Group 1,
Class A Medical Device, and to minimize interference risks, the following requirements
shall be applied:
NOTE
1. All interconnect cables to peripheral devices must be shielded and properly grounded.
The use of cables that are not properly shielded and grounded may result in the
equipment causing radio frequency interference in violation of the European Union
Medical Device directive and FCC regulations.
2. All of those recommended guidance regarding electromagnetic environment shall be
followed.
WARNING
Do not use devices that intentionally transmit RF signals (Cellular Phones, Transceivers,
or Radio Controlled Products) in the vicinity of this equipment as it may cause
performance outside the published specifications. Keep the power to these types of
devices turned off when near the equipment.
The medical staff in charge of this equipment is required to instruct technicians, patients,
and others.
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The Brivo OEC 850 Mobile C-Arm X-ray Product is intended to use in the specified electromagnetic
environment. The purchaser or user of the Brivo OEC 850 Mobile C-Arm X-ray Product should assure
that it is used in an electromagnetic environment as described below:
Emissions Test
RF Emissions
Compliance
Electromagnetic Environment Guidance
Group1
The Brivo OEC 850 Mobile C-Arm X-ray Product uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
CISPR11
RF Emissions
Class A
CISPR11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Not
applicable
The Brivo OEC 850 Mobile C-Arm X-ray Product is
suitable for use in all establishments other than
domestic and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Not
applicable
IEC 61000-3-3
1-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Guidance and manufacturer’s declaration - Electromagnetic Immunity
The Brivo OEC 850 Mobile C-Arm X-ray Product is intended to use in the specified electromagnetic
environment. The purchaser or user of the Brivo OEC 850 Mobile C-Arm X-ray Product should assure that it
is used in an electromagnetic environment as described below:
Immunity Test
IEC 60601
Compliance Level
Test Level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient Pulse
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power supply
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV for input/output
lines
±1 kV differential mode
voltage
±1kV differential mode
voltage
±2 kV common mode
voltage
±2 kV common mode
voltage
< 5 % UT ,for 0.5
cycle
0 % UT for 5 sec
(> 95 % dip in UT)
(> 95 % dip in UT)
40 % UT ,for 5 cycles
(60 % dip in UT)
70 % UT ,for 25
cycles
(30 % dip in UT)
< 5 % UT ,for 5 s
Electromagnetic
Environment Guidance
Floors are wood, concrete, or
ceramic tile, or floors are
covered with synthetic material
and the relative humidity is at
least 30 %.
Mains power quality is that of a
typical commercial and/or
hospital environment
Mains power quality is that of a
typical commercial and/or
hospital environment.
Mains power quality is that of a
typical commercial and/or
hospital environment. If the
user of the Brivo OEC 850
Mobile C-Arm X-ray Product
requires continued operation
during power mains
interruptions, it is
recommended that the Brivo
OEC 850 Mobile C-Arm X-ray
Product be powered from an
uninterruptible power supply or
a battery.
(> 95 % dip in UT)
Power frequency
magnetic field
(50/60Hz)
3 A/m
3 A/m
IEC 61000-4-8
Note:
UT means AC voltage before the test.
These are guidelines. Actual conditions may vary.
1-22
Power frequency magnetic
fields are at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
Introduction and Safety
Guidance and manufacturer’s declaration - Electromagnetic Immunity
The Brivo OEC 850 Mobile C-Arm X-ray Product is intended to use in the specified electromagnetic
environment. The purchaser or user of the Brivo OEC 850 Mobile C-Arm X-ray Product should assure that it is
used in an electromagnetic environment as described below:
IEC 60601
Test Level
Conducted RF
3 V rms
IEC 61000-4-6
150 kHz to
Compliance
Level
[V1 =] 3 V
80 MHz
Electromagnetic Environment
Portable and mobile RF communications
equipment are used no closer to any part of the
[EQUIPMENT and/or SYSTEM], including cables,
than the recommended separation distance
calculated from the equation appropriate for the
frequency of the transmitter.
Recommended separation distance
d =[
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
[E1=] 3 V/m
d =[
2.5 GHz
3,5
] P
3
7
d =[ ] P
3
3,5
] P
3
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Note: P is the power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey a,
are less than the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with the following symbol:
a: Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength exceeds the RF compliance level above, observe the Brivo OEC 850 Mobile C-Arm X-ray Product to verify normal
operation in each use location. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the [EQUIPMENT and/or SYSTEM].
b: Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note3: These are guidelines. Actual conditions may vary.
1-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
The Recommended Separation Distances are listed in the next table.
Recommended Separation Distances for Portable and Mobile RF Communications
Equipment and the Brivo OEC 850 system
The Brivo OEC 850 Mobile C-Arm X-ray Product is intended to use in the electromagnetic environment
where the RF interference is controlled. According to the rating power of the communication
equipment, the purchaser or user of the Brivo OEC 850 Mobile C-Arm X-ray Product can prevent the
Electromagnetic interference by the recommended separation distance below:
Separation Distance/m
Rated Power of
Transmitter/W
150KHz to 80 MHz
d =[
3,5
] P
3
80 MHz to 800 MHz
d =[
3,5
] P
3
800 MHz to 2,5 GHz
7
d =[ ] P
3
0.12
0.12
0.23
0.38
0.38
0.73
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
0.01
0.1
1
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the equation
in the corresponding column, where P is the power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Note 3: These are guidelines. Actual conditions may vary.
1-24
Chapter2. System Overview
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
2.1. System Descriptions
2.1.1. System overview
Brivo OEC 850 Mobile C-Arm X-ray Product is intended to provide fluoroscopic and film imaging of the
patient in diagnostic and surgical applications. Clinical applications may include, but are not limited to,
surgery, critical care and emergency room procedures. Such users will use this system on a daily basis.
The system may be used for other imaging applications at the user’s discretion.
WARNING
Only properly trained persons can operate this equipment.
Brivo OEC 850 Mobile C-Arm X-ray Product includes:
• Mobile C-Arm
Includes high frequency X-Ray generator, image intensifier, CCD camera and console.
•
Mobile WorkStation
Includes dual LCD monitors, 1 computer, External storage media (DVD/USB), etc.
•
Options
Thermal Printer
DAP
Laser aimer device on tube and image intensifier
Two-pedal footswitch
EU standard Plug
30cm skin spacer
Film cassette holder
Video Distributor for External DVI video output
110V power kit
The C-Arms derive their name from the rigid, C-shaped weldment that supports the Image Intensifier,
collimator, CCD Camera, and X-ray tube. These components together are commonly called the “C”, and
the entire assembly from the casters up is known as the “C-Arm”.
The X-Ray tube generates an X-ray beam that is shaped and oriented by the collimator. After passing
through patient anatomy, the beam is directed onto a film cassette to produce a radiographic film
image, or through the image intensifier and CCD camera, to convert the X-ray image to a video signal
that is transmitted to the Workstation for display, processing, and storage.
2-2
System Overview
2.1.2. Component Locator
1
8
2
9
10
3
11
4
12
5
13
6
1
14
7
15
16
18
19
17
2-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1. Image intensifier and CCD camera
2. Lateral rotation movement brake handle
3. Orbital movement brake handle
4. Skin spacer
5. X-ray tube head (X-ray tube and collimator)
6. Front wheel
7. Rear wheel
8. Wig-wag movement brake handle
9. Steering handle
10. Horizontal movement brake handle
11. Control panel
12. Hand switch
13. Brake release pedal
14. Brake pedal
15. Driving handle
16. Fast stop switch
17. Female sockets (interconnect cable and footswitch)
18. Incoming circuit breaker
19. Key switch and power on indicator
2-4
System Overview
4
1
2
5
6
7
3
8
8
10
10
11
9
9
2-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1. LCD Monitors
2. Keyboard
3. USB connector
4. Exposure indicator
5. Mouse
6. Cabinet for printer
7. DVD driver
8. Front wheels
9. Back wheels
10. Handles
11. Interface for DICOM and Video distributor
2-6
System Overview
2.2. System block
2-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
2.3. System Schematics
2-8
System Overview
2-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
2-10
System Overview
2-11
Chapter3.
Installation
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.1. Overview
This chapter described the installation procedure of Brivo OEC 850 Mobile C-Arm X-ray Product.
3.2. Pre-Installation
3.2.1. Dimensions and Weights of System
Dimensions:
C-Arm: 181cm × 76cm × 175cm (71.3’’ × 30’’ × 69’’)
Workstation: 62.8cm×89cm×167cm (24.7’’×35’’×65.8’’)
Weights:
C-Arm: 245±10kg
Workstation: 140±10kg
3.2.2. Site Requirement
Scan room (L × W × H)
≥5.0 m × 5.5 m ×2.8 m
Entrance (W× H)
≥1.2 m × 2.0 m
Aisle (W)
≥2m
WARNING
Enough room should be provided for the movement, installation and operation of the
system.
WARNING
The floor should be hard enough to support the system. It should also be smooth and
horizontal.
3.2.3. Power Supply Requirement
Voltage:
AC 100V/110V/120V/220V/230V/240V
Frequency:
50Hz/60Hz±3Hz
Rated Momentary Line Current (shorter than 5s):
3-2
•
25A (220V/230V/240V)
•
50A (100V/110V/120V)
Installation
Rated Continuous Line Current:
•
10A (220V/230V/240V)
•
20A (100V/110V/120V)
NOTE
16A power supply is required in China. And 20 A is required in Japan.
Suitable sockets are required both in Scan room and Operating room for maintenance. Power supply
should comply with requirement of the system power supply requirements. The sockets should be 3
cores with protective earth, which complies with the requirement of the applicable electrical codes. Other
equipment electrically connected with this apparatus share the same grounding connection.
WARNING
You should check the local power supply first. If it is not 220 V/230 V/240 V (50 Hz), you
need change the power supply cables and change the connections of the system. Refer to
the Power Supply Adaptation section for details.
3.2.4. Environments
Operating Conditions:
Temperature:
+10°C~+40°C (+50F~+104F)
Humidity:
30~75% Relative, non-condensing
Atmospheric Pressure:
700hPa~1060hPa
Transportation and Storage:
Temperature:
-20°C~+55°C (-4°F~+131°F)
Humidity:
10~95% Relative, non-condensing
Atmospheric Pressure:
500hPa~1060hPa
3-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.2.5. Tools and Test Equipment
Standard Tools and Test Equipment are required to install the system. In addition to standard service
tools and test equipments, the following special tools and standard test equipment are required to
complete this procedure.
Standard Tool
Item
Tool
1
A series of screw driver
2
Wrench
3
Dosimeter
4
10’’×12’’ Film
5
Blank CD/DVD disk
6
Spatial Resolution Tool
7
Lead Apron, thicker than 3mm
8
Pinchers
9
Scissors
10
Stopwatch
11
Network cable
12
Printer paper or film
13
Ruler longer than 12 inches (30cm)
14
Digital volt-ohm meter or equivalent
Special Tool
Item
3-4
Tool
GE part number
1
9 inch Film cassette holder (if required)
5304073
2
9 inch target for Laser Aimer (if required)
00-878223-02
3
Three copper sheets, each 1mm thick
00-877682-01
Installation
3.3. Unpacking
NOTE
There are 2 packages for Brivo OEC 850 Mobile C-Arm X-ray Product. One is for C-Arm (L X
W X H=1900mmX900mmX1970mm. And the other is WorkStation (L X W X
H=800mmX960mmX1820mm).
NOTE
Before unpacking, inspect the shipping enclosure /boxes for damage and report to the
shipper if available.
The unpacking of the system should be done before moving it into the operation room, in order to check
if there’s damage or non-apparent damage occurred during transportation.
The packages should be examined minutely during installation to avoid missing parts with small
dimensions or envelopes containing screws, washers, etc. Once damage is observed, notify service of the
transportation and insurance company immediately.
If non-apparent damage due to transportation was found during unpacking or installation, take the
same action in 15 days after delivery. The transportation corporation will not pay for non-apparent
damage generally unless making an application for inspection in 15 days after delivery.
If damage is discovered, contact the related company immediately, with specifying the type of
apparatus, the serial number and the order number if possible, and describing the nature of the damage.
Please follow the steps below to unpack the Brivo OEC 850 Mobile C-Arm X-ray Product.
C-Arm:
1. Place the C-Arm and workstation packages on the flat and leave enough space around.
2. Cut off the binding tie and remove the packing of the C-Arm. Remove the wrapper on the C-Arm.
3. Lock the cross-arm, wig-wag, lateral rotation and orbital movement brakes. Get the accessory
package, crowbar, stow-woods and ramp from the C-arm package. Put the wood ramp behind
of the C-arm.
NOTE
The ramp and crowbar will also be used to workstation unpacking.
3-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Ramp
Accessory
Crowbar and
stow-woods
4. Loosen the screws and remove the wood support of image intensifier.
Wood
support
5. Remove the baffle of the tube.
3-6
Installation
6. Loosen the screws and remove the three fastening metal strips of the wheels.
7. Pry up one side of the C-arm with the crowbar and put the stow-wood under each rear wheel.
Loose all the screws and remove the wood fastening assembly on each rear wheels of the Carm.
8. Hold the handles and move the C-arm backward to draw out backwards the image intensifier
and tube until the front wheels located on the front wood fastening assembly.
NOTE
Take care to protect the Image intensifier and gird when move the C-arm.
3-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9. Release the lateral rotation brake. Rotate the C-arm to horizontal location and lock the brake.
Remove the stow-woods under the rear wheels. Hold the right steering handle in a vertical
position to keep the rear wheels in a straight direction. Then carefully move the C-Arm backward
along the ramp to the site you desire.
10. Loosen the bolt and remove the backward-forward fastening. Loosen the horizontal movement
brake handle and push the C-Arm forward. Lock the brake handle.
11. Check parts according to packing list (refer to the packing list with the system for details).
WARNING
3-8
Before moving the system, please make sure the 4 brake handles: the C-Arm orbital
rotation brake, the lateral rotation brake, the Wig-Wag brake, and the horizontal crossarm brake are locked. Otherwise, equipment damage or personal injury may result.
Installation
Workstation:
1. Cut off the binding tie and remove the packing of the workstation. Take out the ramp on standby.
And remove the wrapper on the workstation.
2. Loose the screws of the fastening assembly cover around the four wheels. Remove all the
screws on the wood support. Use the crowbar to pry and remove the wood support.
3. Put the ramp in position. Move the workstation along the ramp.
3-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.4. Inspection
3.4.1. Label Inspection
After the system installation, check and verify that all labels in section 1.2.10 should be in position.
NOTE
Labels below should be in local language according to the regulation requirement. Find
the labels in local language in the language option and C-arm package. Attach the labels
to the correct position.
(There’s only English and Chinese version)
(Only exists when there’s laser aimer option)
3-10
Installation
3.4.2. External Inspection
1. Cabling and Connector inspection
Check the interconnect cable for damage or other problems. Verify that connector on cable mates well
with the connector on C-Arm and that there are no bent or missing pins. Make sure the strain relief
attaches securely to both the interconnect cable and connector.
2. C-Arm Inspection
Inspect each C-arm component listed under the heading C-arm Inspection. Select one check box for
each item in the list. The choices are:
•
No problem Found
•
Loose or missing parts
•
Broken or damaged parts
•
Chipped or scratched paint
The items to inspect are grouped according to the larger visible, easily accessible assemblies and/or
modular parts of the system. Each inspection of an assembly item should include its visible component
parts and sub-assemblies.
C-Arm Inspection
Physical Location
No problem
found
Loose or Missing
parts
Broken or
Damaged Parts
Chipped or
Scratched paint
Image Intensifier
C-Arm
X-ray Tube
Control Panel
Handles
Cabinet
Wheels
3-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3. Workstation Inspection
Inspect each workstation component listed under the heading workstation inspection. Select one check
box for each item in the list.
WorkStation Inspection
Physical Location
No problem
found
Loose or missing
parts
Broken or
damaged parts
Chipped or
scratched paint
Monitor and Arm
Keyboard and
Mouse
Covers
Frame
Handles
Wheels
3.4.3. Internal Inspection
NOTE
WARNING
Observe electrostatic discharge safety procedures.
Removing any covers exposes dangerous electrical voltages on several components.
When possible, disconnect system from AC line before removing any covers. When it is
necessary to leave the power applied, use extreme care to avoid death or injury from
electrical shock.
(1)
Remove workstation cover. Inspect the internal labels and parts. Check if there are loose parts.
(2)
Remove the C-Arm’s covers. Check all cables and PCBs inside C-arm. Look for loose PCBs and
connectors and reset.
3-12
Installation
3.5. Setup
3.5.1. Power Supply Adaption
The system is prefabricated for 220V power. If your power supply is 220V, go to the section 3.5.2. If your
power supply is 110V, please follow steps below to change power supply:
1. Make sure that C-arm covers have been removed. Unplug the cable connected to circuit breaker
and earth. Loosen the screws of the circuit breaker. Remove the 220V power supply cable and
breaker.
2. Connect adaptive plug to 110V power cable.
Select or buy the relative plugs according to your local regulatory requirements and sockets style.
They must be reliably grounded.
If the system is used in Canada and Australian and USA, please select and buy the relative plugs
according to the list below. The selected plugs shall be listed or recognized type plugs. The
manufacture doesn’t provide these types of plugs.
Countries
Rated Current
Comments
Voltage
Frequency
Canada
20 A
Hospital Grade Plug
120 V
60 Hz
Australian
16 A
Transparent
230 V
50 Hz
USA
20 A
"Hospital grade" or
"Hospital only" type plug
120 V
60 Hz
NOTE
The plug should be installed by GE service representative.
3. Replace 220V breaker with 110V breaker. Connect 110 V power supply cable to breaker and
earth.
3-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4. Change the bridge connectors of the transformer to connect 9-10 and 11-12 of J1.
The system can also adapt power supply of AC230V, AC240V and AC120V by
different connections of the transformer. Before install the system, please check the
power supply voltage range and adjust the connector.
Transformer taps
Wire
240 V
230 V
220 V
120 V
110 V
Wire N (blue)
1
1
1
1
1
Wire L1 (Brn)
4
3
3
4
3
Wire N1 (blue)
5
5
5
5
5
Wire L(Brn)
8
8
7
8
7
10-11
10-11
10-11
9-10
9-10
11-12
11-12
120VAC
1
2
3
4
5
6
7
8
9
10
11
12
3-14
220VAC
1
2
3
4
5
6
7
8
9
10
11
12
12
11
10
9
8
5
2
3
4
5
110VAC
L
N1
4
2
1
1
J1
3
L1
N
6
7
8
Jumper
6
Voltage
7
NOTE
230VAC
1
2
3
4
5
6
7
8
9
10
11
12
240VAC
1
2
3
4
5
6
7
8
9
10
11
12
1
2
3
4
5
6
7
8
9
10
11
12
Installation
5. Adjust jumper on PDU board.
6. Plug the power supply plug into the socket and verify the system can power on normally.
7. Install all the covers of the C-arm.
3.5.2. UPS
Follow these steps to setup the UPS.
1. Make sure that the rear cover and both lower side covers of workstation have been removed.
2. Plug the UPS battery connector.
3. Connect the interconnect cable to C-arm. Plug the power cord into appropriate AC receptacle.
Voltage
increase
Power on
indicator
Voltage
Decrease
Test
Overload
On Battery Indicator
Replace
battery
Off
3-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4. Press and hold UPS “test” button for several seconds to initiate a self-test. The workstation will
start up and enter BIOS. There will be a message to inform that there’s no AC power and need to
shut down.
5. Press “off” button to turn off the UPS.
6. Install all the covers of the workstation.
The UPS receives AC power whenever the system power switch is on. The battery recharges if
necessary. If the AC mains power fails, the UPS supplies power to workstation using its internal
battery.
The online indicator goes out and the battery indicator comes on. UPS sounds “bi-bi-bi-bi” every four
seconds.
Do not connect any other devices to the UPS receptacles.
WARNING
UPS battery should be disconnected before transportation. Battery can be left in the
UPS.
3.5.3. Power on
CAUTION
When you place a recently unpacked system into use, allow it to stand in its new
environment long enough for any condensed moisture to evaporate. A decrease in
temperature or a humidity increase can cause water vapor in the air to condense,
sometimes on electric circuits. You may damage the equipment if you apply power
before this condensation dries.
1. Take out foot switch from the accessory package. Connect the Footswitch to C-arm.
2. Press the power on switch on control panel.
WARNING
Before power on the system, please check and make sure that the power frequency and
voltage of the system are compatible with the power supply. Or you must change the
power supply and frequency settings of the system.
Refer to Section 3.5.1 for details.
3. Verify that the workstation and C-arm complete software boot and initialization with no errors
reported on their displays.
The auto fluoroscopy mode is selected and the power on indicator lights. The field size is in its maximum.
The left monitor of workstation will display GE logo screen. The right monitor will enter into the PATIENT
INFORMATION Screen automatically.
3-16
Installation
4. Press vertical lift button and verify that vertical lift column is operational.
3.5.4. System Configuration
Hold down Ctrl +Shift+Alt key and press ”E”, ”V”, ”R” keys orderly and quickly to invoke the service tool.
1. Input the service ID of the system. Refer to section 6.2.1.
2. Set up local language and time zone. Refer to section 6.2.2 and 6.2.3.
3. Choose the exposure option and dose group by country though Exposure Management screen.
Refer to section 6.2.5.
4. Choose Fluoro Alarm Control through exposure management screen. Refer to section 6.2.5.
5. Return main screen of service tool and click “Shutdown” button to shut down the system.
CAUTION
The configuration must be complete in case the local regulation is not complied.
3-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.6. Functional Checks
3.6.1. C-Arm Checks
Mechanical Movement
Perform the following mechanical inspections and checks.
Wig-Wag Movement
1. Release the Wig-Wag brake.
2. Move the mechanics through the entire operational range. Verify Wig-Wag for ease of movement
without excessive play.
3. Verify that tightening the Wig-Wag brake prevents Wig Wag motion.
Horizontal Movement
1. Unlock Horizontal bake and fully extend and retract the Horizontal Cross-Arm.
2. Verify that the Cross-Arm moves smoothly and quietly through the bearing block without
excessive play.
3. Place the Cross-arm brake in the lock position.
4. Verify that the cross-arm does not move.
CAUTION
Make sure cross arm is fully retracted and brake is set before moving C-arm. Equipment
damage or personal injury may otherwise result.
C-arm Orbital Rotation
CAUTION
When you release the orbital movement brake, place your other hand on the C-arm so
that you have control of the C-arm’s movement.
1. Release the orbital brake.
2. Verify that C-arm orbits smoothly in its cradle without excessive play.
3. Verify that the entire orbital scale is readable and in good condition.
4. Set the brake handle to lock position and verify that C-arm locks firmly in position.
Lateral Rotation
1. Release lateral rotation brake.
2. Verify C-arm rotates smoothly in both directions.
3. Lock the brake and verify the brake locks C-arm firmly in position.
WARNING
3-18
Before transporting the system, please make sure the 4 brake handles: the C-Arm
orbital rotation brake, the lateral rotation brake, the Wig-Wag brake, and the horizontal
cross-arm brake are locked. Otherwise, equipment damage or personal injury may
result.
Installation
Vertical Column test
1. Power on the system. Set the keyswitch in ON position (
). Press and hold the up vertical
switch. The maximum distance is about 40cm. Verify that vertical column stops when it
column
reaches the limit and that operation is smooth and quiet.
2. Press and hold down vertical column switch. When the movement stops, press two keys
at the same time and C-arm will be lowered to the lowest position.
Wheels
1. Push C-arm slowly and verify that the wheels contact the floor and that they roll freely.
2. Make sure there are no worn or damaged areas on the wheel tread.
3. Verify that there is no wobble or binding associated with wheels.
Brakes
1. Step on the rear wheel pedal brake. Verify that the rear wheels lock.
2. Step on wheel pedal release brake. Verify that brakes release and that both rear wheels roll and
steer freely.
Steering
1. Verify that both rear casters respond appropriately to movement of the steering handle.
2. Verify that you can steer the rear casters 90 degrees to the left and 90 degrees to the right.
3. Verify that the casters track together when you steer them.
4. Place the mobile C-arm’s rear wheel brake in release position. Push the C-arm slowly and verify
that C-arm can turn smoothly.
3-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
C-Arm Controls
The following procedure produces X-rays. Use the appropriate personal protective
equipment to protect yourself from X-ray exposure.
WARNING
, mA/mAs
kVp
Press the kVp and mA/mAs keys and verify that the technique can be adjusted manually.
Pulse Fluoroscopy
Select Pulsed normal fluoro mode via the C-arm control panel. The X-Ray Exposure indicator flashes
when taking an exposure.
Manual HLF
+
Select manual HLF mode. Take exposure and verify that a HLF exposure is taken. Verify that audible tone
beeps at normal rate.
Auto HLF
+
+
Place an object in front of Image Intensifier. Select Auto HLF mode to take exposure. The exposure will
stop automatically after 20 seconds if you selected “Exposure Limit 20sec” in “Exposure Management”
screen of service tool.
Digital Spot
Select digital spot via control panel. Take a digital spot shot and verify that fluoro exposure automatically
terminates and image automatically saves to hard drive.
Image Orientation
Using the line pair tool as a phantom, take exposure in normal fluoro mode and verify the following
functions:
•
Image Rotation key
rotates the image clockwise and counter-clockwise.
•
Image Reverse key
inverts and reinverts the image.
3-20
Installation
Collimator
Press each of the following collimator switches while taking exposure and verify each of the following
collimator actions.
•
Collimator iris opens and closes.
•
Collimator leaves open and close.
•
Collimator leaves rotate.
Field size
repeatedly, verifying ability to select
Place something on the Image Intensifier. Press field size key
ability to select any of the three available sizes: NORM, MAG1 and MAG2.
Recursive Factor
to select 2, 4, 8, 16 respectively to take an exposure. Verify the amount of
Press recursive factor key
filtering can be increased and decreased. The noise in the image should be decreased as the Recursive
Factor is increased.
Motion Artifact Reduction
Take exposure and rotate the image by pressing image rotation key. Observe the motion artifact of the
image. Press motion artifact reduction key
and verify that the indicator lights.
Alarm Reset
Press and hold Alarm Reset key
arm control panel resets to zero.
for approximately 2 seconds. Verify Fluoro time displayed on the C-
Radiographic Mode
Select Radiography mode and verify that the indicator lights.
3-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
C-arm key/X-ray switch test
Perform the following test to verify that the system’s X-ray and key switches are operating satisfactorily.
WARNING
The following procedure produces X-rays. Use the appropriate personal protective
equipment to protect yourself from X-ray exposure.
Key switch
). Press Handswitch and verify that no X-rays
1. Place the X-ray keyswitch in STANDBY position (
are generated. Press Vertical lift button and verify that vertical lift column is not operational.
). Press handswitch and verify that no X-rays are
2. Place X-ray keyswitch in XRAYOFF position (
generated. Press vertical lift button and verify the vertical lift column is operational.
3. Place X-ray keyswitch in ON position (
). Press handswitch, footswitch and vertical lift button
to verify X-ray and vertical lift column are enable.
Emergency Switch
Perform the following test to verify the emergency switch.
1. Press the emergency switch during raising C-arm.
Emergency
Switch
2. Verify that vertical column operation and exposure are disabled.
3. Release the emergency switch. Verify that the vertical column operation and exposure are
available.
3-22
Installation
3.6.2. Workstation Checks
Mechanical Movement
1. Push the workstation and verify that the wheels roll smoothly without vibration and that they are
securely attached to the wheels.
2. Verify the operation of the brakes of workstation.
3. Check movement of the monitor arm for ease of movement without excessive play. Make sure
that the monitors do not move after being positioned.
Workstation Operation
) key and verify this function exchanges images between the left monitor and
1. Press SWAP (
the right monitor.
) key. And then press
2. Press the SAVE (
in the Image Directory screen.
key. Verify that the left monitor image is previewed
3. Verify Patient Information function activates the Patient Information screen. Enter patient
information (
), display scheduled exams, display resume exams screen.
4. Verify Image Directory function (
) allows you to enter Image Directory screen.
5. Verify the SETUP function (
) allows access to the Setup screen. Press “Security” button and
select “Require Users to Login” in the screen.
6. Verify ZOOM function (
magnify the ROI.
7. Verify NEGATE (
left monitor.
) displays a square region of interest (ROI) box and that you can
) function reverses the light and dark values on the image displayed on the
8. Verify the CONTRAST function (
contrast can be increased and decreased.
) displays and indicator bar and that the amount of
9. Verify the BRIGHTNESS function (
brightness can be increased and decreased.
10. Verify that Auto Contrast/Brightness (
contrast and brightness.
) displays an indicator bar and that amount of
) automatically selects the optimum values of
) displays an indicator bar and that amount of
11. Verify ENGE ENHANCE function (
enhancement level increased and decreased.
3-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
12. Press ALT key and the LOGOUT key (
) at the same time. Verify that it can log the user out of
the workstation applications. The default password is “123456”. Enter “Security” screen and you
can disable the logout function or change password.
CD/DVD check
1. Place a blank CD or DVD in the in the CD/DVD drive and wait for the green LED on the front of the
drive to turn off.
) on the keyboard to open the Image Directory. Make sure the
2. Press the Image Directory key (
Image Directory has some saved images. Otherwise, then save some images by taking test shots.
3. Select the “Copy To…” button on the “Image Directory” screen. The “Copy To… “ screen lists the
available print and storage devices. Select CD/DVD and click the “OK” button to confirm your
selection.
4. Click “Option…” button. Select “Include Viewer” in copy option and click “OK” button.
5. Select one or more images from the Image Directory.
6. Click “Copy” button on the Image Directory screen to copy the queued image.
7. Insert the CD/DVD to other computer. The media viewer will run automatically. Use the viewer to
confirm that the image was written to the CD/DVD successfully.
3-24
Installation
DICOM Check
NOTE
You can choose the verification items from DICOM worklist server, storage server and
printer server, according to the customer’s network situation and equipment.
1. Remove the cable cover from the back cover of workstation.
2. Plug the network cable into the network connector on the computer.
3. Install the cable cover.
4. Press the Setup
key on the workstation keyboard.
3-25
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5. Click “Network Config…”button. Input the “Station Name” and “AE Title”. The default AE Title is
“OEC850”. You can change it according to the DICOM service’s requirement. Click “Config…”button
and fill in the network information. You can obtain the information from network administrator.
6. Select “DICOM Worklist…”button from the workstation setup screen.
Click “Add…” button and fill in the information of the DICOM worklist server. Obtain the information from
network administrator. Click “OK” button to return to DICOM worklist screen. Click “Verify” button. The
message “ The DICOM server has verified successfully.” will display if the server has connected. Click “OK”
button and then select the Exit button to close the DICOM Worklist screen.
3-26
Installation
7. Select “DICOM Storage…” from the workstation setup screen.
Click “Add…” button and fill in the information of the DICOM storage server. Obtain the information from
network administrator. Click “OK” button to return to DICOM storage server list screen. Click “Verify”
button. The message “ The DICOM server has verified successfully.” will display if the server has
connected. Click “OK” button and then select the “Exit” button to close the DICOM Storage server list
screen.
8. Click “DICOM printer…” button on setup screen.
3-27
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Click “Add…” button and fill in the information of the DICOM printer server. Obtain the information
from network administrator. Click “OK” button to return to DICOM printer server list screen. Click
“Verify” button. The message “ The DICOM server has verified successfully.” will display if the server
has connected. Click “OK” button and then select the “Exit” button to close the DICOM printer server
list screen.
9. Select “Scheduled Exams…” button on Patient Information screen. Choose the worklist server and
click the “schedule filter” button. Setup the query filter and click OK.
Click “Update schedule” button to perform a DICOM query. Verify that the search results will be
displayed in the scheduled exams screen.
10. Press Image Directory
key on the workstation keyboard. Select DICOM printer server from the
Copy To screen. Click the preview image and then click the queue box. Click “Print” button. Verify that
the images can be printed successfully after the process complete.
11. Click “Copy To…” button. Select the DICOM storage server. Select one or more preview images you
want to archive. Press and hold CTRL key on the keyboard to select multiple images from the
previews in the directory. Click “Copy” button. Verify that the images are copied to the storage server
successfully after the copy progress complete.
3-28
Installation
3.6.3. Auto Technique Tracking
Select Auto Fluoroscopy mode (
), and the NORM field size to take an exposure. Put the copper filter
on the tube head. Verify that the kVp changes accurately as copper filters are inserted into the field in the
following ranges:
Copper mm
kVp range
1mm
2mm
3mm
64±4kVp
74±4kVp
82±4kVp
If the kVp is not in the range, refer to 5.7.1 for calibration.
3.6.4. Image Resolution
Place a special resolution card in the front of the image intensifier. The angle between the horizontal axis
of the image and the test card must be 45°. Select Manual Fluoroscopy mode (
) and appropriate
kV value. Take an exposure, and stop when image is stable. The image is frozen on the left monitor. Verify
the resolution in lp/mm is equal to, or greater than, the values listed below. Repeat and select any of
three field sizes (NORM, MAG1 and MAG2) to verify the resolution.
NORM
Field Size
Resolution (lp/mm) ≥ 2.2
MAG1
MAG2
≥ 3.0
≥ 3.5
If the resolution check fails, refer to section 5.6.8 for Camera Focus calibration.
3.6.5. Dose Checks
Normal fluoro dose rate check
1. Position the dosimeter in the center of the X-ray beam 30cm above the grid. Cover the image
intensifier with lead thicker than 3mm for protection of the image intensifier.
2. Set kVp as 110kVp, increase mA to maximum in manual mode. Measure the dose rate while
making exposure. Limits vary by the dose group, which was selected in software configuration
section by country.
Continuous fluoro
≤ 50 mGy/min (Dose group: Normal 50 and HLF 50)
≤ 88 mGy/min (Dose group: Normal 88 and HLF 100)
Pulse fluoro
≤ 50 mGy/min (Dose group: Normal 50 and HLF 50)
≤ 88 mGy/min (Dose group: Normal 88 and HLF 100)
3-29
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
HLF dose rate check
1. Position the dosimeter in the center of the X-ray beam 30cm above the grid. Cover the image
intensifier with lead thicker than 3mm for protection of the image intensifier.
2. Set kVp as 110kVp, increase mA to maximum in manual mode. Measure the dose rate while
making exposure. Limits vary by the dose group, which was selected in software configuration
section by country.
Continuous fluoro
≤ 50 mGy/min (Dose group: Normal 50 and HLF 50)
≤ 100 mGy/min (Dose group: Normal 88 and HLF 100)
Pulse fluoro
≤ 50 mGy/min (Dose group: Normal 50 and HLF 50)
≤ 100 mGy/min (Dose group: Normal 88 and HLF 100)
Digital Spot Dose rate check
1. Position the dosimeter in the center of the X-ray beam 30cm before the center of the entrance
surface of the image intensifier. Cover the image intensifier with lead thicker than 3mm for
protection of the image intensifier.
2. Set parameters as 110kVp, 12mA in digital spot mode. Measure dose of each pulse while taking
exposure.
Dose rate =Dose value/exposure time
The dose rate should be less than 82.4 mGy/min.
NOTE
3-30
After the installation, please fill in the product locator card for PLC parts below and mail
the cards to local service operation office.
Item
Part Number
Part Description
1
5325576
Generator, monoblock, EVR
2
5330059
Assembly, I.I with power supply, EVR (Cert.
Comp.)
3
5336403
ASM COLLIMATOR, EVR
Installation
3.7. Options Installation
3.7.1. Laser Aimer
Laser Aimer on Tube
1. Laser Aimer, which attaches to tube, is incorporated with the tube cover. Remove the standard
tube cover. Position the tube cover with laser aimer and plug the power cable. Tighten the secure
screws of the tube cover. Remove the laser aimer cover. Power on the system.
Power
cable
2. Install the beam alignment tool to Image Intensifier.
3. Press the laser aimer’s green ON switch to generate the laser beam.
4. Loose the holding inside so that you can adjust the beam by the adjuster.
Holding
5. Adjust the adjusters of the laser until the intersection of the laser beam is in the center circle of
the beam alignment tool. (Use the circle to adjust the line to be vertical with the other and the
screw to adjust the position of the line.)
Circle adjuster
3-31
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6. Fasten the holding and then install the laser aimer cover.
7. Functional check: Press the laser aimer’s ON switch and verify that the intersection of the laser
beam is in the center circle of the beam alignment tool and the two laser lines are vertical.
Laser Aimer on Image Intensifier
Perform the following procedure to install and operate the laser aimer on the Image Intensifier. The Carm can be placed in any orientation that is convenient for attaching the laser aimer. There are no
alignment marks or notches on the laser aimer that need to match any points on the image intensifier.
When the laser aimer is attached to the C-arm, releasing the orbital or rotational brake
could cause the C-arm to move. Make sure to hold the C-arm to control its movement
CAUTION
whenever a brake is released.
1. Place the laser aimer’s attachment handle in the open position.
2. Move the laser aimer flush with the Image Intensifier, but slightly off center.
3. Center the laser aimer with the relation to the Image Intensifier. As you center the laser aimer, the
attachment hooks catch on the edge of the image intensifier.
4. Turn the attachment handle inward until the handle is completely closed and the laser aimer is
securely attached to the image intensifier. Insert the metal bolt to the hole of the handle to lock.
3-32
Installation
CAUTION
Do not grasp the Laser Aimer's handle to position the C-arm or to move the system.
Pulling the handle may cause the Laser Aimer to release from the image intensifier and
fall.
5. Use the image intensifier handle to position the C-arm or to move the system to the desired location.
6. After attaching the laser aimer, place the C-arm in the position it will be used during the procedure.
7. Turn the laser aimer ON and adjust the beam alignment knobs until the laser beam is centered on
the intersection of the cross-hair on the tube cover.
CAUTION
If the laser is ON, do not look directly into the laser aimer’s aperture.
8. Functional check: Using fluoroscopy, verify that the cross-hair in the fluoro image, at the center of the
area of interest.
NOTE
The location where the laser beam marks the patient corresponds to the center of the
area of interest, indicated by the cross-hair in the fluoro image.
Perform procedure below to remove the laser cover from the laser aimer:
1. Switch OFF the power to the laser aimer.
2. Grasp the two knobs on the laser cover.
3. Press down and rotate the cover (approximately 10 degrees counterclockwise) until the knobs align
with the notches on the housing.
4. Lift off the cover and resume laser aimer operation.
3-33
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.7.2. Video distributor
Perform the following procedures to install the video distributor.
1. Power off the system. Remove the rear cover of the workstation.
2. Put the video distributor on the backboard. Tighten the four screws to secure the video
distributor.
3. Unplug the video cable of left monitor from computer and plug it to the DVI-I connector of the
video distributor.
4. Plug the video cable of video distributor to the computer’s video connector DVI-1, which was
connected with the left monitor’s cable before.
5. Plug power cable to video distributor power input port.
3-34
Installation
6. Functional check: Power on the system. Check that the left monitor will display GE logo screen.
Plug the left monitor video cable and connect to output2, output 3 and output4 respectively.
Verify that the left monitor will display GE logo screen in each case. Then connect the video cable
to DVI-I port.
7. Install the rear cover of the workstation.
3.7.3. Printers
Brivo OEC 850 offers four types of printers as option.
•
UP-990AD Thermal Printer
•
UP-D897 Printer
•
UP-970AD Thermal Printer
•
UP-DF500 DICOM Printer
Follow the steps below to install the printers except for the UP-DF500 DICOM printer.
1. Power off the system. Open the rear cover of the workstation.
2. Plug one end of the video cable to the video port of the printer. Plug the power supply cable into
the socket on the printer. Put the cables through the hole of the workstation. Put the printer in the
cabinet.
3. Plug the video cable into the USB port on computer.
3-35
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4. Cut off the other end of the power supply cable. Strip off 3 wire ends. Connect the 3 wires to the
terminal blocks at the bottom of the workstation (connect the brown line to connector 1, 2 or 3,
blue line to 6,7 or 8 and the green-yellow line to earth). Put the power supply cable along the
wiring harness to get a good layout.
5. Functional check: Power on the system. Verify that an image can be printed to the printer and the
printer works with both paper and film.
Perform the following steps to install UP-DF500 DICOM printer.
1. Plug one end of the network cable to the printer.
2. Remove the cable cover of the workstation.
3. Plug the other end of the network cable to the network interface port.
4. Connect the power cable plug of printer to AC receptacle.
5. Functional check: Power on the system. Verify that an image can be printed to the printer and the
printer works with both paper and film.
3-36
Installation
3.7.4. Film cassette holder
Please follow the steps below to install and remove the cassette holder:
1. Place the cassette holder over the face of the image intensifier with the handle opened out.
2. Rotate the cassette holder handle to securely attach the cassette holder to the image intensifier.
WARNING
Verify that the cassette holder is securely attached to the image intensifier. Unsecured
cassette holders may fall, injuring patients or personnel.
3. Insert a film cassette into the cassette holder and center it.
NOTE
WARNING
The cassette holder uses friction to hold the cassette.
Verify that cassette is held securely within the cassette holder. Unsecured film cassettes
may fall, injuring patients or personnel.
4. Remove the film cassette by pushing the cassette out of the cassette holder.
5. To remove the cassette holder, open the cassette holder handle and press firmly against the side of
the cassette holder.
NOTE
The cassette holder is designed to snuggle attach to the image intensifier lip when the
handle is engaged. It may be necessary to strike the cassette holder with the palm of
your hand to dislodge the cassette holder from the image intensifier.
3.7.5. DAP
Perform the following procedure to install the DAP.
1. Loose the screws of tube cover and remove the tube cover.
2. Install the three studs on the top plate of collimator.
3. Fasten the DAP chamber to the adjustment set. Fasten the chamber set to the three studs with
screws.
3-37
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4. Place the DAP module on the metal block and tighten the 2 screws as below:
DAP Module
Chamber
5. Use socket to connect DAP cable and cable of chamber.
6. Don’t install the tube cover first. After connected all the cables. Put the circular adjustment tool
on the chamber. Ensure the circles of the iron sheet and DAP chamber are tangent.
7. Adjust precision of the DAP chamber’s position according to the procedure below:
(1) Coarse tuning
Boot up Brivo OEC 850 system. Take exposure at low kV (about 45 kV) in normal fluoro mode.
Image below will display on the screen. Release the screws that secured DAP. Adjust the
adjustable screws of the DAP chamber set until the circular image is on the center of the
screen.
3-38
Installation
Image of the
adjustment tool
(2) Fine tuning (Ensure the iris works normally before this procedure)
Switch to the manual fluoro mode. Take exposure at high or middle kV (about 70kV).
Meanwhile, make the iris in the field and adjust kV value to ensure the edges of the iris and
circular tool are clear enough.
Adjust the screws of the chamber set until the center of the circular image and center of the
iris are superposed. Close the iris to make the image edge of the iris is close to the edge of
circular image. Maintain the center of the circular image and center of the iris are
superposed. The error of the center’s position should not exceed 2mm.
Edge of
circular tool
Edge of the
iris
8. Login the service mode. Click “Exposure management…”button on the main screen. Choose
“enable” item for the DAP option. Save the setting and shut down the system.
3-39
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9. Screw down the screws that secure the DAP chamber set. Remove the circular adjustment tool.
Install the tube cover.
DAP Test and Calibration
Set Dosimeter as accumulative/maintenance. Put the detector of dosimeter on the center of the grid on
image intensifier. Power on the system and create a new exam. Ensure the DAP and Dose value on the
right monitor are cleared. Set exposure parameters as 80kV, 10 mAs in radiography mode. Install the
cassette holder on image intensifier. Insert a film to cassette holder and press hand switch to take
exposure. Record the DAP show value on right monitor and dose value on dosimeter. Get the film and
test the diameter of the image. Calculate the circle area of the image on film.
DAP calculated value= Dosimeter reading×circular area on film
Compare the DAP calculated value with DAP show value on the right monitor. The error should be less
than 20%.
ERROR=
DAPCalculatedvalue − DAPshowvalue
DAPCalculatedvalue
× 100%
1. Units should be consistent during comparing.
NOTE
2. The test should be done after the tube cover installation complete.
If the error above is too large, adjust DAP calibration coefficient to calibrate: Login service tool. Click
“Exposure Management” button on main screen. Change the DAP Calibration Coefficient. DAP show
value on the workstation right monitor is in direct proportion with the calibration coefficient.
Set the coefficient and test DAP value again until the error is less than 20%.
Dose Test and Calibration
Install the beam alignment tool. Set Dosimeter as accumulative/maintenance. Put the detector of
dosimeter on the axis of image intensifier and focus, 30 cm away from grid center. Clear the dosimeter.
Create a new exam on workstation and ensure the DAP and dose show value on the workstation are
cleared.
Set kV/mA value as 110kV/4mA in manual fluoro mode. Open the collimator completely. Take exposure
for 3-5 seconds. Record Dose show value on right monitor and dosimeter measured value. Repeat three
times and get the average of the dose show value on right monitor and the average of measured value
on dosimeter. Calculate the error between show value average and measured value average.
NOTE
After each test, reset the dosimeter and create a new exam to clear the DAP and dose
show value on workstation.
ERROR=
DoseMeasuredvalue − Doseshowvalue
DoseMeasuredvalue
× 100%
If the error is more than 5%, adjust Dose Calibration Coefficient: Login service tool. Enter “Exposure
Management” screen. Change Dose Calibration Coefficient. Dose/Dose rate show value on the
workstation right monitor is in direct proportion with the calibration coefficient.
Dose Accuracy Test
According to the collimator field size settings, adjust the collimator opening size by beam alignment tool.
In each opening size test, create a new exam on the workstation. Set the exposure parameters as 45kV,
3-40
Installation
1.3mA in manual fluoro. Adjust collimator field size. Record dose show value on right monitor of the
workstation as D0. Reset the dosimeter. Take exposure at high and lower kV value for 3-5 seconds.
Record the dose show value on the right monitor and measured dose value on dosimeter. Calculate dose
value on the right monitor of the workstation.
@60kV, 1.3mA: Dose value on workstation=D1-D0 (D1: Dose show value on right monitor after exposure
@60kV, 1.3mA)
@110kV, 4mA: Dose value on workstation=D2-D1 (D2: Dose show value on right monitor after exposure
@110kV, 4mA)
Calculate the error between the dose value on workstation and measured value on dosimeter.
ERROR=
DoseMeasuredvalue − Dosevalue(onworkstation)
DoseMeasuredvalue
× 100%
The error should be less than 30% in every state.
NOTE
After each test, dosimeter must be cleared. Create new exam to ensure DAP and dose
show value on workstation be cleared.
Collimator field size
45mm-43mm
110-120mm
140-150mm
Largest size
NOTE
kV/mA
Dose value
(mGy)
Dose measured
value on dosimeter
(mGy)
Error
60kv/1.3mA
110kv/4mA
60kv/1.3mA
110kv/4mA
60kv/1.3mA
110kv/4mA
60kv/1.3mA
110kv/4mA
The field size in the form should be measured by field size measure tool.
3-41
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3.7.6. Skin Spacer
Install the skin spacer as the following steps:
1. If there isn’t laser aimer on tube, go to step 2 directly. Otherwise, put the skin spacer on the X-ray
tube head. Make sure the skin spacer’s quadrate hole superpose with the hole of tube head cover
as below.
2. Find two screw holes on the skin spacer. Remove the skin spacer and then loose two screws on
the tube head cover, which are corresponding to the two screw holes on skin spacer.
3. Install the skin spacer on the cover and tighten the two screws to secure skin spacer to tube
head cover.
3.7.7. 110V Power Cable Kit
110V power cable kit includes 110V power cable and breaker for 110V power supply.
Refer to section 3.3.1 for replacement of breaker and cable.
3.7.8. Two-pedal Footswitch
Brivo OEC 850 Mobile C-Arm X-ray Product provides a two-pedal footswitch as an option.
Plug the cable of two-pedal footswitch to the socket on the right side of the C-Arm interface panel. Verify
that the connector locks in place.
3-42
Installation
Functional check: Power on the system. Press the left switch to start fluoroscopy in the mode selected.
Press right switch to save an image displayed on the left monitor.
3.7.9. EU standard Plug
Brivo OEC 850 provides a EU standard Plug. If the system is used in some countries of Europe, where the
plug is applicable, change the power supply plug to the EU standard plug.
3-43
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3-44
Chapter4.
Subsystem
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4.1. Power Distribution and Control
Power to main transformer is applied by pressing the Power ON key on the membrane control panel.
Relay JR1 in PDU board energizes, then the relay JR1 and timer circuit in PDU board begin to work. The
transformer will be powered by R1 and R2. After about 1.5 second, the R1 and R2 are shorted by the
relay, which is controlled by timer circuit. The R1 and R2 can resolve the power on surge current problem.
Relay JR1 in PDU board energizes, then the relay JR1 and timer circuit in PDU board begin to work.
Pressing the Power Off key on control panel will power off the equipment.
In interlock loop, the interconnect cable, display board, collimator control board (MP1), MCB, filament
control board (MP4), and PDU Board compose the safety loop. If the interconnect cable between C-Arm
and the Workstation is removed or any PCBs in C-Arm loose from slot due to collision, the relay JR2 will
not be energized so that the system can't be powered on. There is a relay in workstation, which is
powered by PC +12V. The contactor of the relay is connected to power off key, if the PC isn't powered off,
the power off key on C-Arm is open and the system will not be powered off. This function is used to
protect PC from unintended power off.
4-2
Subsystem
Power Distribution_Mainframe
To
workstation
IGBT Fan
Boot Up
Transformer
Line Filter
TUBE
AC24
AC220
IGBT
AC/DC
Display BD
(Boot up Section)
DC24
DC+-15
MP4
(Filament Drive)
DC5
Main
Transformer
Seal BD
Display BD
(Display Section)
DC15/5
AC220
Membrane
Panel
DC3.3
DC3.3
PDU
Board
AC220
AC220
DC+-15/5
PS1
+-15/
+5VDC
Laser Aimer
(Tube)
Laser Aimer(I.I)
Battery
DC+-15/5
AC22
MCB
PS2
+5/+24VDC
MP1
DC8
Collimator
DC12
DAP
DC24
Image
Intensifier
AC24
DC24
Mother BD
CCU
DC24
Circuit
Breaker
Line
Filter
Up/Down
Driver
24VDC
DC 24
CCD
Camera
X-ray On Indicator
(WKS)
Up/Down
Actuator
DC+-15
Drive
Transformer
Mul Volt.
KV Control BD
4-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Power Distribution_Workstation
AC220
Monitor1
AC220
Monitor2
J3
PC
Display adapter
DC+-12/5
AC220
Interconnect
Cable
AC220
Line
Filter
AC220
UPS
PC
ATX PS
Frame Grabber
AC220
Main BD
SATA HD
UPS
Control BD
AC220
AC220
DC24
4-4
DC+-12/5
DVD-RW
DC12
Protection
Relay
Video distributor
USB Printer
(not including
DICOM Printer)
X-Ray
indicator
USB
Subsystem
4.2. MCB
The MCB (Mainframe Control Board) is located in C-Arm mainframe and is the main control unit in
mainframe. It is responsible for five functions.
-
X-ray control: KV, mA and exposure control interfacing with collimator board, filament board, and KV
control board
-
Image chain control: control CCU and image intensifier
-
User interface: key input and LED display interfacing with display board
-
System monitor: check various error and warning
-
Communication with workstation
The structure of MCB interface is illustrated in the following diagram.
4-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
For detailed circuit schematics of MCB, please refer to Appendix. MCB consists of 10 function blocks,
which are illuminated as follows.
MCB functional Block
MCU function
-
Read the registers of peripheral controller to get the status of sub-systems, such as getting
the alarm status, etc.
-
Write the registers of peripheral controller to control the status of sub-systems, such as PIO
(Programmable Input Output), light indicator, buzzer, etc.
-
Sample watchdog bite signal from supervisor circuit to check peripheral is still working
-
Check all switch signals to respond user’s input
-
Control power down sequence
-
Communicate with Control Panel board to get the key input and control the LED output
-
A/D convert all analog signal, such as shutter location, temperature of oil tank, mA feedback,
KV feedback, IBS, etc.
-
Communicate with DAC to control mA and KV output
-
Communication with Workstation
-
Communication with DAP
-
Communication with peripheral controller to report the working status of MCU
• Peripheral controller function
4-6
-
Sample the status of input pins, and record them into status registers polled by MCU
-
Set output pins according to the control registers accessed by MCU
-
Generate interrupt signal of alarm to MCU in order to request the processing
Subsystem
-
Generate exposure control signals
-
Check switch stuck error
-
Control mAS control signals
• Operational amplifier circuit function
-
Amplify or adjust input analog signals, such as kV feedback, mA feedback, temperature,
shutter location sample signal, and output those signals to the ADC unit (Analog Digital
Converter)
-
Amplify or adjust analog signals output from DAC (Digital Analog Converter), such as kV
setting, mA setting, mA preheat, and output those signals to other sub-systems
-
mA PI control
• Transceiver
-
Transceiver circuit is responsible for regulating voltage level and supplying driving current.
• Optical coupling
-
Optical coupling circuit is responsible for changing voltage level and isolating the ground of
circuit at each sides of optical coupler.
• DAC
-
DAC (Digital Analog Converter) circuit is responsible for output analog signal, such as KV
setting, mA setting and mA preheat.
• VFC
-
VFC (Voltage Frequency Converter) circuit is responsible for converting mA signal to
frequency value in order to implement mAs function.
• Communication
-
RS485 protocol is used for MCU to communicate with workstation or DAP.
-
SPI protocol is used for MCU to communicate with Control Panel.
• Indication
-
Indication circuit includes LED and buzzer, which are controlled by MCU and peripheral
controller.
• Miscellaneous hardware
-
Other circuits include power supply, supervisor circuit, watchdog circuit, clock generator, and
security control, etc.
The external physical interfaces of MCB are defined in the table below:
Connector
J1
Name
MCB digital
signal connector
Type
DN41612 C
Type Male
Gender
Pin
J2
MCB analog
signal connector
DN41612 Half
C Type Male
Pin
Description
All digital signals from/to other sub-system, such as
collimator board, filament board, sealing board, CCU, KV
control board etc.
All analog signals from/to other sub-system, such as
collimator board, filament board, sealing board, CCU, KV
control board etc.
4-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4.3. Display Board
Display board and membrane kit compose a whole control panel of C-Arm. Control panel receives user
input command and sends signal to MCB, PDU and motor drive board. For detailed circuit schematics of
display board, please refer to Appendix.
PWR_ON UP/DOWN
Membrane ControlB
KV_UP
mA_DOWN
DX1~DX4
DX7
PDU
DX8
Motor Driver
Display Board
DX6
MX1
MCB
SPI
Mother Board
Signal from MCB to display board:
Name
Description
Attribution Scale
SPI_CLK
Clock of SPI bus
Digital
0/+15V
SPI_SIMO
Data from master to
slave
Digital
0/+15V
SPI_/CS1
Chip select 1 for
MAX9654
Digital
0/+15V 0V=enable
SPI_/CS2
Chip select 2 for
MAX7301
Digital
0/+15V 0V=enable
YELLOW_LED
Enable signal for
exposure LED on
Control Panel
Digital
0/+15V +15V=enable
POWER_OFF_COMMAND One termination of
power off relay
Passive
-
POWER_OFF_RETURN
Passive
-
4-8
The other termination
of power off relay
Comment
POWER_OFF_COMMAND &
POWER_OFF_RETURN
connected means power off
command
Subsystem
Signal from display board to MCB:
Name
Description
Attribution Scale
SPI_SOMI
Data from slave to master
Digital
0/24V
DB_INT
Interrupt of Display Board
Digital
0/24V
Comment
0V=enable
4.4. kV control and IGBT assembly
X-ray generator structure is illustrated in below diagram. For detailed circuit schematics of kV control
and IGBT board, please refer to Appendix. It mainly consists of two parts: inverter and monoblock. The
inverter consists of rectifier bridge, DC bus capacitor, driver power transformer, kV control board, IGBT
inverter board. This section only involves inverter. The detailed information of monoblock will be
introduced in section 4.3.5.
AC
Main
transformer
DC bus
capacitor
Rectifier
bridge
24V
Generator
Driver power
transformer
kV control
board
MCB
L, C
filter
IGBT
inverter
board
Monoblock
Mother board
The inverter provides 40kHz AC power to monoblock. It can control kV value to follow the kV set value.
There are two parts in the inverter: power converter and controller.
Power converter consists of 1 AC-DC converter and 1 DC-AC converter. AC-DC converter includes rectifier
bridge D and DC bus capacitor C1, C2. DC-AC converter includes fuse F, IGBT devices Q1, Q2, Q3, Q4 and
resonant inductor L and resonant capacitor C. Power converter input AC 220V voltage and output 40kHz
variable voltage to monoblock. The serial resonant frequency of L and C is nearly 40kHz.
AC-DC
DC-AC
F
Q1
D
Q3
L
C1 C2
C
AC
AC
Q2
Q4
4-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
The controller mainly consists of kV control board and driver power transformer. The power transformer
provides 4 channel isolation powers to kV control board. Every channel output voltage is 20V AC.
The kV control board will provide the following function and performance.
•
Send out 4 channel PWM (Pulse Width Modulation) signal for IGBT inverter board
•
Provide insulated driver signal for IGBT
•
KV feedback control circuit
•
KV feedback value output
•
Exposure control
•
KV value set
Protect circuit and error signal output
Signals from MCB to KV control board are listed as follows.
Name
Description
SAFETY_EN
Safety
line
EXPOSURE_COMMAND
EXPOSURE_COMMAND
Attribution Scale
for Digital
Exposure command
0/+15V
Comment
+15V=enable
0V=disable
Digital
0/+15V
+15V=enable
0V=disable
KV_SET_INPUT
KV set value to KV control Analog
board
0-6V
Unit is 20KV/V
Signals from KV control board to MCB are listed as follows.
Name
Description
Attribution Scale
Comment
KV_LOW_ALM
KV too low alarm, lower than 35KV
Digital
0/+15V
0V=alarm
KV_HIGH_ALM
KV too High alarm, higher than
120KV
Digital
0/+15V
0V=alarm
OVERCURRENT_ALM
Inverter over current alarm
Digital
0/+15V
0V=alarm
IGBT_ERROR_ALM
IGBT error alarm
Digital
0/+15V
0V=alarm
DC_ERROR_ALM
DC Power Bus Error alarm
Digital
0/+15V
0V=alarm
KV_UNBALANCE_ALM
KV unbalance in two bridges of
inverter.
Digital
0/+15V
0V=alarm
KV_FEEDBACK_OUTPUT
KV sample value from KV control
board
Analog
0-6V
Unit
20KV/V
4-10
Subsystem
4.5. Filament Control Board
Filament control board (MP4) can be separated to two parts simply, one is power unit and other is control
unit. The power unit offers the energy to the filament transformer primary coil and the control unit
control the energy input to the transformer by control the output voltage of the power unit. The theory is
illustrated in the following chart. For detailed circuit schematics of filament control board, please refer to
Appendix.
In the power unit, the DC 24V is fed through two parallel power transistors V6 & V7 and drives the pushpull complementary amplifier consisting of V8, V9, V10 and V11. The V6, V7, V13 and V12 are used to
adjust the voltage of the filament transformer primary. A 600Hz pulse as the V8, V9, V10 and V11 drive
signal is from MCB. V1, V4, R4 and R12 compose over current protecting circuit. When the current
through R4//R12 is too big, a signal will lock then trigger the alarm module.
In the control unit, A3A acts as an analog adder and drivers the V13 and V12 through A3B. Adjusting
POT1 may control the filament base heating current.
In Fluoroscopy mode, gate A4B is on, and MAC is fed to A3A. MAC is the Fluoro mA reference signal and
controls the tube current in Fluoro mode.
In the system, the Radiography current is constant 20mA. If the Radiography mode been selected, IDRGHC is active, gate A4B is off, gate D5C is on, a 10V voltage inputs to A3A positive port for Radiography
preheating.
When the exposure commend signal “EXPRCM-HC” is effective, gate A4D & D5D are on. A 10V voltage
adds to A3B for Radiography quick heating and KVC signal adds to the negative port of A3B through
amplifier A2B for KV compensation.
The relay K1 is used to select the small focal spot of the X-ray tube in Fluoro and the large focal spot in
Radiography In the Radiography mode, IDRG is high, K1 is active, and the large focal spot works.
Signals from filament control board to MCB:
Name
Description
Attribution Scale
Comment
FILAL-HC
Over current alarm of +24V power
supply in filament board
Digital
+15V=alarm
0/+15V
Signals from MCB to filament control board:
Name
Description
Attribution
Scale
Comment
EXPRCM-HC
Exposure command in
radiographic function
Digital
0/+15V
+15V=enable
ALRSCM-HC
System general alarm
Digital
0/+15V
+15V=alarm
Trigger by anyone of several alarms
IDRG-HC_1
Radiographic function
selected
Digital
0/+15V
+15V=enable
GENRESHC
Generator reset signal
during system boot up
Digital
0/+15V
+15V=enable
Persist High for 300mS
approximately after power on
4-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1200Hz
Clock of filament board
Digital
0/+15V
Used for filament drive signal
(600Hz) generation
MAC
Fluoro mA control
signal
Analog
3.05.0V
-
KVC
KV control signal
Analog
4-12V
Unit V/10KV
Analog signals (KVC/MAC), which are from DAC (Digital Analog Converter) circuit of MCB to MP4, are
amplified or adjusted by operational amplifier circuit. All the digital signals between MCB and MP4 are
transformed their voltage levels between +3.3V and +15V through optical couple circuit.
4.6. Monoblock
Monoblock can generate X-ray. It mainly consists of high voltage generator and tube insert. All high
voltage components are put into oil box, which can enhance the insulation performance.
Monoblock generates 110kV inside the oil tank. The input voltage changes from 0V to 240V peak value.
High voltage is got from high frequency and high voltage transformer T1. The primary coil is 10T and the
secondary coil is 2500T. So the maximum voltage of secondary side is 60kV peak value at 40kHz.
T1
D1
C1
R1
+kV
D2
{
C2
R2
-mA
kV
Inverter
Tube
D3
C3
D4
C4
R3
-kV
R4
T2
{
filament
Inverter
T3
Monoblock Schematic Diagram
High voltage circuit comprises rectifier circuit and voltage doubler. High voltage circuit composes of high
voltage diode D1-D4, high voltage capacitor C1-C4. If output voltage from T1 is 60kV peak value the high
voltage circuit output 120kV to the tube.
The controller needs to get kV value and mA value from monoblock. KV sense circuit comprises high
voltage resistor R1, R4 and sense resistor R2, R3. The high voltage resistor value is 400M ohm, sense
resistor is 20k ohm If the voltage on the tube is 120kV the +kV output 3V and –kV output –3V. mA value is
measured by serial resistor between –mA and ground. So the mA voltage value is negative.
4-12
Subsystem
The tube is used to generate X-ray. This tube is a kind of double focus with fixed copper anode tube. It is
suitable for oil insulation generator.
Analog signals KV_GND and MA_CAT of sealing board are the anode and cathode of mA current signal
respectively.
There are two temperature measurement circuits in sealing board. One’s output is analog signal
OIL_TEMP.
The other’s output is connected/open status between THERMAL_SWITCH_A and THERMAL_SWITCH_B. If
the temperature of oil tank is higher than a preset threshold, the relationship between
THERMAL_SWITCH_A and THERMAL_SWITCH_B changes from connected to open.
Signals from sealing board to MCB:
Name
Description
Attribution Scale Comment
OIL_TEMP
Output voltage of
temperature sensor
Analog
0-5V
-
OIL_TEMP_GND
Reference of OIL_TEMP
Ground
0V
-
Passive
-
THERMAL_SWITCH_A &
THERMAL_SWITCH_B
connected means that
temperature is lower than
threshold.
THERMAL_SWITCH_A One termination of
temperature switch
THERMAL_SWITCH_A &
THERMAL_SWITCH_B
disconnected means that
temperature is higher than
threshold.
THERMAL_SWITCH_B The other termination of
temperature switch
Passive
-
MA_CAT
The cathode of mA sample
Analog
-
-
KV_GND
The anode of mA sample
Analog
-
-
Technical Specifications:
Nominal high voltage
110 kV
Nominal input power
2.2 kW
Maximum anode heat content
53 kJ
Target material
Tungsten
Target angle
10°
Small focal spot
0.6×1.4 mm
Large focal spot
1.4 mm
4-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Filtration equiv.
0.55 mm Al
Anode
Stationary
Typical filament specification
4.2 V, 3.5A
Maximum filament current
4.6A
Weight
0.65 kg
Nominal heat capacity
76kHU
4.7. X-ray/key Switches and Emergency Switch
Key switch is used to control the unauthorized usage of X-ray exposure and vertical lift motor.
Signals from Key switch to MCB:
Name
Description
Attribution Scale
XRAYOFF
The termination which
indicates lift column
enable and x-ray
disable
Passive
Open/Close XRAYOFF & COMMON is
connected means XRAYOFF
switch has pressed, which
represents lift column function
has been selected and
exposure enable has not been
selected.
ON
The termination which
indicates lift column
and x-ray are both
enable
Passive
Open/Close ON & COMMON is connected
means ON switch has pressed,
which represent both lift
column and exposure enable
function have been selected.
Passive
-
COMMON Common termination
Comment
-
X-rays can be enabled via the Footswitch or the Handswitch. Only Fluoro can be initiated from footswitch
while handswitch can also initiate radiography. Signals from footswitch and handswitch to MCB are
listed in the following charts. Detailed interconnect information, please refer to chapter 2.3.
Signals from Handswitch to MCB:
4-14
Name
Description
Attribution
Scale
Comment
HAND_SWITCH
Hand switch termination
Passive
Open/Clos
e
HAND_SWITCH &
REF_HAND_SWT is
connected means
HAND_SWITCH has
pressed
HAND_SECURIT
Y
Hand security
termination
Passive
Open/Clos
e
HAND_SECURITY &
REF_HAND_SWT is
connected means
HAND_SECURITY has
Subsystem
pressed
REF_HAND_SW
T
Common termination
Passive
-
-
Signals from Footswitch to MCB:
Name
Description
Attribution Scale
Comment
FOOT-SWITCH
Foot switch
termination
Passive
Open/Close FOOT_SWITCH & REF_FOOT_SWT is
connected means FOOT_SWITCH
has pressed
FOOT-SECURITY
Foot security
termination
Passive
Open/Close FOOT_SECURITY & REF_FOOT_SWT
is connected means
FOOT_SECURITY has pressed
REF_FOOT_SWT
Common
termination
Passive
-
RESERVE_SWT
Reserved foot
switch
termination
Passive
Open/Close RESERVE_SWT & REF_FOOT_SWT is
connected means RESERVE_SWT
has pressed
RESERVE_SECURITY Reserved foot
security
termination
Passive
Open/Close RESERVE_SECURITY &
REF_FOOT_SWT is connected
means RESERVE_SECURITY has
pressed
-
Emergency switch, located on the control panel of C-arm mainframe, stops all motorized motion
immediately, interrupts any X-rays in progress, and prevents further X-rays until you reset the system. At
any time, if the emergency switch is pushed down, the X-ray exposure shall be stopped immediately.
System shall trigger the related error message.
4-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4.8. Image Chain
4.8.1. Overview
The following diagram the image chain pipeline of Brivo850.
X-ray
Generation
X-ray
Gen
X-ray
Collimator
X-rays
Collimat X-rays
or
X-ray Detection
Grid
X-rays
I.I
Image Acquisition
Light
Camera
head
Analog
Video
CCU
Image Processing
Analog
Video
Frame
grabber
Digital Video
IP
CPU &
display
board
Image Display
Digital
Image
Left
19" LCD
Realtime
display
VLI (kV+)
(kV, mA)
Exposure
Manage
ABS Servo
Exposure
Manage
Collimator collimates the X-ray beam to the appropriate field size and user’s setting. The x-ray beam
passes through Grid to reduce scattered X-ray photons. Image Intensifier converts this X-ray into lights
for each field size selection, and sends the light to CCD Camera head. CCD Camera head captures the
light image and sends the analog video to CCU. The CCU performs preliminary image processing, and
outputs analog signal to Frame Grabber Card. Frame Grabber card in the workstation reads the analog
signal, register into digital video, and the CPU performs real-time post-processing (Brightness, Contrast,
Edge Enhance, Zoom, and Negate) to be sent to image display. CCU also analyzes the analog input
signal, and outputs an indicator kv+ to MCB; higher kv+ when the X-ray dose is higher than desired, and
lower kv+ when the X-ray dose is lower than desired. With AutoTechnique enabled, MCB S/W converts
this signal into digital signal, and send command to X-ray Generator to increase or decrease kVp/mA
depending on the kv+ value.
4.8.2. Collimator Control
The collimator is an electromechanical assembly designed to allow the operator to restrict the size and
shape of the X-ray beam. The unit has a lead iris and shutters and is mounted directly over the X-ray
tube beam aperture. The collimator use three independent motors to control the following degrees of
freedom:
•
The diameter of the iris aperture
•
The gap spacing between the semi-transparent copper leaves
•
The rotational position/orientation of the copper leaves
Collimator specification:
•
Length×Width×Height: 290 mm×120 mm×100 mm
•
Weight: 2kg
4-16
Subsystem
•
8 leafs Motorized iris collimator
•
Shutter rotate angle: 180 degree
The collimator is controlled by control panel through MCB and collimator control board (MP1).
The function of the collimator control board is to control the iris and the Semi-transparent leaves of
collimator. The circuit of MP1 is in Appendix. Motor M1 opens or closes the shutter. Depressing the switch
“OPEN” or “CLOSE” on the console panel energizes the motor drives the petals for desired width.
Analog signals (SWPO, SRPO, FSPO and VREF), which are transmitted from MP1 to MCB, show the location
information of iris. They are amplified by operational amplifier circuit, and then sampled by ADC circuit of
MCU. The output control signal of MCB is transmitted to MP1 through optical couple circuit. LEFT, RIGHT,
OPEN, CLOSE, IRD and IRI are control command from control panel. F1 and F2 are three-horizon control
of image intensifier. IDRG-HC is exposure control command.
Signals from MCB to MP1:
Name
Description
Attribution
Scale
Comment
Right
Shutter turn clockwise
Digital
0V/open
Left
Shutter turn counter-clockwise
Digital
0V/open
0V=enable,
Mutex
signals,
which
means
those two
signals
should not
enable at
the same
time.
Open
Shutter open
Digital
0V/open
0V=enable
Close
Shutter close
Digital
0V/open
Mutex
signals
IRD
Iris close
Digital
0V/open
0V=enable
IRI
Iris open
Digital
0V/open
Mutex
signals
F1
The LSB (Least Significant Bit) of three Digital
fields code, which indicates the MAG
size.
0V/+15V
“F2F1”=”00
”
means
that MAG
size set to
9’’
“F2F1”=”01
”
means
that MAG
size set to
4.5’’
F2
The MSB (Most Significant Bit) of three Digital
fields code
0V/+15V
“F2F1”=”10
”
means
that MAG
size set to
6’’
4-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
“F2F1”=”11
”
is
invalidation
37.5Hz
Clock of MP1
Digital
0V/+15V
The clock
doesn’t
need
a
precise
frequency
IDRG-HC_2
Radiographic function selected
Digital
0V/+15V
+15V=enab
le
Signals from MP1 to MCB are listed as follows.
Name
Description
Attribution Scale
Comment
FSPO
Feedback voltage from Iris position rheostat.
Analog
05.6V
Reserved for preview
SRPO
Feedback voltage from shutter rotation position
rheostat.
Analog
05.6V
Reserved for preview
SWPO
Feedback voltage from shutter width position
rheostat.
Analog
05.6V
Reserved for preview
4.8.3. Image Intensifier
Parameter
Code
Description
Setup Criteria
Image tube size
A
Adjust the imaging field size
Meet imaging field size requirement
Image tube focus
B&C
Adjust the image tube focusing
Meet spatial resolution requirement
The code position can refer to the following Fig of I.I power supply,
4-18
Subsystem
4.8.4. CCD/CCU
N30 PWR2
NDR ADV
J
1
2
1 2
J6
1 2
J8
J7
J4
R1
01
Parameter
CODE
SHADING ENABLE
J4
AUTO WTH
J5
AUTO BLK
J6
KV CONTROL
MODALITY
J7-J8
SHADING
(VS-VP-HS-HP)
AUTO KV
WINDOW
R1
00
R8
4
R6
1
R5
2
R4
7
R4
8
R7 R1
9 06
DESCRIPTION
It is used to enable shading regulations
(J4 ON: position 1-2, J4 OFF: position 2-3)
It is used to enable the Auto White Level.
(J5 ON: position 1-2, J5 OFF: position 2-3)
It is used to enable the Auto black Level.
(J6 ON: position 1-2, J6 OFF: position 2-3)
Setup Criteria
J4 ON: position 1-2
J5 OFF: position 2-3
J6 OFF: position 2-3
Controls the KV control modality.
J7: position 2-3
(J7-J8 UP/DOWN mode: position 1-2, J7-J8 ANALOG mode:
position 2-3)
J8: position 2-3
R61-R52R47-R48
Set the shading gain to compensate picture non-uniformity
because of Image Intensifier.
Adjust to get uniform image with
1mm copper on the tube.
R106
Increase or decrease the sensibility of the automatic KV
control. This feature is very useful in according to adapt the
CCD camera N33 for any kind of X-ray generator
Anti-clockwise maximum limit (all
the way to the left)
4-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
BLACK LEVEL
R101
Ad just black level pedestal (BOTTOM BLACK). Regulate the
pedestal to 30mV respect referring video level
20~30mV
AUTO KV LEVEL
R79
Set the value of AUTO KV
Adjust to get –80~0mV when
there is no x-ray
IRIS
On camera
head
Adjust Iris open/close position to balance image brightness
and X-ray dose.
Adjust iris to get 3.8~4.2mR/min
of dose rate at the II using the
stable technique in the ABS mode
with 2mm Copper filter on the
tube.
FOCUS
On camera
head
Adjust the camera focusing
Meet limiting resolution
requirement
CCU Connector Legend
DB9
V4
CH
B
V3
CH
A
24V
DB25
DB25 Connector Legend
4-20
Name
Description
CHA
Output video channel A
CHB
Output video channel B
V3
Sync output
V4
Output video scan conversion (option)
DB9
CCD head connector (FEMALE)
DB25
Commands interface connector (FEMALE)
24V
Power Supply 24Vdc
Subsystem
+KV
-KV
ND FILTER
AUTO OFF
NOT USED
NOT USED
NOT USED
ROI
INVH
FREEZE-A
LIH
X16 FILTER
X8 FILTER
1
14
2
GND
15
ROT
3
16
ROT
4
5
6
7
8
9
10
11
12
13
17
18
19
20
21
22
23
24
25
HALF GAIN
NOT USED
NOT USED
NOT USED
DRF
INVV
FREEZE-B
X4 FILTER
X0 FILTER
4-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4.8.5. Image Grabber Card
Specification:
Specification
A/D conversion
10 bits
Brightness/Contrast Has
(offset/gain) control
Pixel jetter
0.5ns
Acquisite video:
1049lines /60Hz
Has
Support Linux OS
Support
Supplier
JoinHope
Model
OK_M30B
Setup points: No setup points available.
4.8.6. Monitor
Parameter
Code
Description
Setup Criteria
Brightness
N/A
Image display brightness: 500/600/700
Default: 500
Display curve
N/A
Image display curve:Dicom/CIE/
Gamma1.8/Gamma2.0/Gamma2.2
Defult: Gamma2.2
4-22
Subsystem
4.9. Computer and Peripheral
4.9.1. Computer
The computer mounted in the workstation takes over most of the communication with external devices
and is the main device of workstation. The computer runs Linux system working with image grabber card
and display card.
The computer has 2 1GB memory banks and a 250GB hard disk with standard 3.5’ SATA interface. The
partitioned SATA hard drive contains Linux system and system software. The Serial ATA interface
provides several advantages over the traditional (parallel) ATA interface. The primary advantages include:
•
Easy installation and configuration with true plug-and-play connectivity. It is not necessary to set
any jumpers or other configuration options.
•
Thinner and more flexible cabling for improved enclosure airflow and ease of installation.
•
Scalability to higher performance levels.
The Ethernet port can be used to connect DICOM printer. The system operator can use the Ethernet
connector to interface the workstation to a DICOM server through the facility’s computer network. Once
the Workstation is connected to the network, the operator can query the local server to determine if the
desired images are available for download, and then download selected images to the Workstation. The
Workstation can also upload images over the same network connection. Detailed instructions on DICOM
operation are available in the Operator Manual.
Signals from MCB to workstation:
Name Description
Attribution
Scale
Wk_Y
One of two differential signals from MCB to Workstation
Differential signal 1.5-5V
Wk_Z
The other one of two differential signals from MCB to workstation Differential signal 1.5-5V
Signals from workstation to MCB:
Name
Description
Attribution
Scale
Wk_A
One of two differential
signals from workstation to
MCB
Differential
signal
1.55v
Wk_B
The other one of two
differential signals from
workstation to MCB
Differential
signal
1.55v
Comment
4-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
POWER_SECURITY_A One termination of relay
which indicates whether
power of workstation is
on/off
Passive
-
POWER_SECURITY_A &
POWER_SECURITY_B
connected means that
workstation is powered on
POWER_SECURITY_A &
POWER_SECURITY_B
disconnected means that
workstation is powered off
POWER_SECURITY_B The other termination of
relay which indicates
whether power of
workstation is on/off
Passive
-
4.9.2. DVD and USB
A CD/DVD RW drive that archives images runs to disk and is also used to update workstation software.
The DVD RW drive is SATA interface and supports the following writing medias:
•
DVD+R media.
•
DVD+RW
•
DVD-R
•
DVD-RW
•
CD-R
•
CD-RW
The drive doesn’t have CD-DA audio circuitry. The drive unit accepts a standard CD or DVD disc using a
powered tray for both loading and unloading. This drive unit has an optical pickup head, servo
electronics to maintain correct focus, tracking. feed position, radial tilt and spindle speed, analog
electronics to recover conventional CD-DA audio stereo sound, digital electronics to recover the recorded
data and provide error correction in Mode 1 and Mode 2 Form 1 to the maximum capabilities of the CDROM ECC, and a Serial ATA interface to the host computer.
The DVD is mounted in the front of workstation and connected to computer chassis backplane with a
SATA interface and power cable.
As the interconnect diagram in chapter2.3 shows, the computer provides 4 USB 2.0 ports. USB1 supports
the USB printer. USB3 connects to external USB port, which can be used to support external USB device.
The remaining 2 USB ports are spare. Each USB device is hot swappable, meaning you can plug it in or
disconnect it at any time.
4-24
Subsystem
4.9.3. Keyboard and Mouse
The keyboard contains a simple switch matrix for alphanumeric data. The keyboard connects to
computer though PS2 interface. The following illustration shows the workstation keyboard.
The mouse is an ambidextrous three-button optical wheel mouse that plugs into a PS2 port.
4-25
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4-26
Chapter5.
Calibration
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.1. Overview
The Calibration described herein should be performed each time you replace a major system
component. Perform these tests in sequence that appears in this section if necessary.
The following tools and test equipment are required to complete the calibration.
Standard Tool
Item
Tool
1
Oscilloscope Tektronix 2236 or equivalent
2
Digital Volt-Ohm-Meter or equivalent
3
Spatial Resolution Tool
4
Lead Apron
5
kV meter
6
Dosimeter
7
Ampere meter or equivalent
8
Ruler longer than 12 inches (30cm)
9
Erasable marker (whiteboard marker)
10
mAs meter
11
Small (specify size) of screw driver to adjust camera iris, focus
12
Transparent adhesive tape, with cutter (or need scissor or knife to
cut the tape)
Special Tool
Item
Tool
GE part number
1
Three copper sheets, each 1mm thick
00-877682-01
2
9-inch BAFT Tool
010-878105-02
3
PWA, MCB extend board, EVR
5326478
WARNING
The system will produce X-Rays when implementing the following procedures. Take
appropriate measures and precautions to protect yourself and others from X-Ray
exposure.
CAUTION
You must obey the following steps to guarantee the safe operation of the C-Arm. Improper
adjustment may cause premature failure of major system components.
5-2
Calibration
5.2. kV Measurement and Adjustment
5.2.1. kV Measurement and Accuracy Check
The X-ray generator is composed of high voltage transformer, high voltage rectifier and X-ray tube in one
pressurized oil tank. KV value can be measured by kV meter. Position the kV meter in the center of the Xray beam 30 cm above the focal spot.
Take exposure at kV values below in manual normal fluoro mode. The error should be within ±5%
ranges (meter accuracy is not less than ±2%.). Otherwise, perform the kV adjustment procedure in
section 5.2.2 until all the kV values meet the requirement above.
Exposure mode
KVp value
Minimum (kVp)
Maximum (kVp)
Normal Fluoro
66kVp
62.7
69.3
99 kVp
94.05
103.95
110 kVp
104.5
115.5
66 kVp
62.7
69.3
99 kVp
94.05
103.95
110 kVp
104.5
115.5
55 kVp (1mAs)
52.25
57.25
55 kVp (2mAs)
52.25
57.25
110 kVp (2mAs)
104.5
115.5
Digital Spot
Radiography
5.2.2. kV Adjustment
1. Remove the C-arm console cover. Check that the connection of IGBT board cables.
2. Set to 70kV, 2.4mA. Remove the cover of PCB box. Adjust POT6 on MCB board until TP5 on MCB is
3.5V±0.01V.
3. Power on the system. Check DS9, DS10 and DS3 indicators on kV control board should be light.
4. Rotate R40 and R27 on kV control board clockwise to the end.
5. Test the frequency of TP4 on kV control board with Oscilloscope. Adjust R13 to make the
frequency of TP4 is 80kHz±1kHz. (TP6 and TP9 are signal ground.)
6. Test the voltage of TP16 and TP17 on kV control board with Digital Volt-Ohm-Meter. The voltage
should be 3.15±0.05V. Otherwise, adjust R109 and R110.
7. Take exposure at 70kV, 2.4mA manual normal fluoro. Test wave of TP15 on kV control board with
Digital Volt-Ohm-Meter. Adjust R50 on kV control board to make the voltage is 3.5V±0.05V.
Check that DS1 should be light when exposure.
Refer to the drawing below for the test points, indicators and adjustors.
5-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
MCB
5-4
Calibration
kV control board
5-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.3. mA Measurement and Adjustment
5.3.1. mA Measurement and Accuracy Check
1. Remove the cover of the C-arm console and the cover of PCB box. Extend MCB board with MCB
extend board.
2. Remove the J46 jumper on MCB extend board. Connect the probes of ampere meter to TP222 (+)
and TP210 (-) on MCB extend board.
3. Power on and take exposure. The mA value can be measured.
The measured value will include mA real value and mA leakage from tube insert. So that:
Exposure mA_real =mA_measure –kV/800
Take exposure at mA values below in manual normal fluoro mode. Calculate the Exposure mA_real
value. The error of Exposure mA_real value should be within ±9% ranges (meter accuracy is not less
5-6
Calibration
than ±1%.). Otherwise, mA calibration procedure should be performed until all the mA values meet
the requirement in the list.
Exposure mode
mA value Setting
Minimum (mA)
Maximum (mA)
Normal Fluoro
0.1mA@45kVp
0.091
0.109
0.8mA@55kVp
0.728
0.872
1.8mA@60kVp
1.638
1.962
3.5mA@80kVp
3.185
3.815
4mA@110kVp
3.64
4.36
0.2mA@45kVp
0.182
0.218
1.5mA@50kVp
1.365
1.635
3.8mA@60kVp
3.458
4.142
7.0mA@80kVp
6.37
7.63
10.0mA@60kVp
9.10
10.9
12mA@55kVp
10.92
13.08
HLF
4. Measure the voltage of TP7 on MCB board with oscilloscope.
Exposure mA_real =Voltage(TP7)/400-kV/800
Calculate the Exposure mA_real value. The error of Exposure mA_real value should be within ±9%
ranges (meter accuracy is not less than ±1%.). Otherwise, mA calibration procedure should be
performed until all the mA values meet the requirement in the list.
Digital Spot
1mA@45kVp
0.91
1.09
4mA@60kVp
3.6
4.36
8mA@80kVp
7.28
8.72
12mA@110kVp
10.92
13.08
16mA@80kVp
14.56
17.44
5-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5-8
Calibration
5.3.2. mA calibration
Refer to the drawings at the end of this section for the location of the points.
1. Power on the system. Adjust POT4 on MCB board to make the voltage of TP17 is 0V.
2. Adjust POT1 on MCB board to make the voltage of TP16 is 0V.
3. Make J5 on MCB board short. Hold down Ctrl +Shift+Alt key and press ”E”, ”V”, ”R” keys orderly
and quickly to invoke the service tool. Login service tool and click “Exposure management”
button. Click “Calibration” button.
Select “Adjust mA” in the calibration screen. “CA1” will display on control panel.
4. Take exposure and adjust POT4 on MP4 board to make the measured mA value is 2.1mA.
Clockwise rotation will increase the current. And anticlockwise rotation will decrease the current.
5. Remove jumper on J5 of MCB board. Take exposure and adjust POT 5 on MCB board to make the
measurement current is 2.1mA. Clockwise rotation will increase the current. And anticlockwise
rotation will decrease the current.
6. Make J5 short. Select “Adjust preheat table” in calibration screen of service mode. “CA2”will
display on control panel. Press footswitch or hand switch and hold on until “END” is displayed on
control panel LED field, which would last for about 20 minutes.
If the footswitch or hand switch is released early, you should press the switch again until the
control panel LED displays “END”.
7. Remove the jumper on J5. Reboot the system. Take exposure at 80kV, 0.2mA. Check the
measured current value should be 0.3mA. Otherwise, adjust POT4 on MCB board. Clockwise
rotation will increase the current. And anticlockwise rotation will decrease the current.
8. Take exposure at 80kV, 2mA. Measure TP7 on MCB board with oscilloscope. The wave should be
orthogonal and similar to the picture below:
5-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Otherwise, adjust POT4 on MP4 board.
5-10
Calibration
5-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.4. mAs Measurement and Calibration
5.4.1. mAs Measurment and Accuracy Check
Connect the probes of ampere meter to TP222 (+) and TP210 (-) on MCB extend board. Connect the mAs
meter to ampere meter in serial.
Take exposure by the parameters below. The error of mAs should be within ±8% (meter accuracy is not
less than 2%).
kV, mAs
Minimum (mAs)
Maximum (mAs)
40kV, 80mAs
73.6
86.4
70kV, 2mAs
1.84
2.16
110kV, 1mAs
0.92
1.08
5.4.2. mAs Calibration
1. Rotate POT2 and POT3 on MP4 board anticlockwise to the end.
2. Take exposure at 80kVp, 20mAs. Adjust POT1 on MP4 board until the measured mA value is 6.5
±0.5mA.
3. Take exposure at 110kVp, 20mAs. Adjust POT2 on MP4 board until the measured mA value is
20.0±1mA.
4. Take exposure at 40kVp, 20mAs. Adjust POT3 on MP4 board until the measured mA value is 20.0
±1mA.
5. Take exposure at 80kVp, 20mAs. Adjust POT2 on MCB board until the measured mAs value is
20.0mAs±0.1mAs.
NOTE
5-12
After the calibration, connect the jumper to J46 on MCB extend board.
Calibration
5.5. X-Ray Generator Warm
This procedure is used to improve the life of the X-Ray generator. It is only performed after the Generator
is not powered for more than 3 months.
1. Power on the system
2. Close the collimator IRIS and place a lead shield in the X-Ray beam to protect the Image
Intensifier and CCD camera.
3. Select the Manual Fluoroscopy mode. Increase the kVp value to 60 kVp. Take an exposure for 3
minutes.
4. Wait for 5 minutes after exposure.
5. Increase the kVp value to 80kVp and take an exposure for 3 minutes.
6. Wait for 5 minutes.
7.
Power off the system.
5-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.6. Beam Alignment
5.6.1. Beam Alignment Setup
Workstation Setup
1. Power on the system, and wait until the system boot up completes.
2. Press Ctrl+Alt+Shift, and type “e”,”v”,”r” to enter Service Mode.
3. Select User as “Service”, and type password as “@dminB50”, and click “Login” button.
4. Select Imaging Defaults.
5. Disable IP Circular Mask, disable CCU Circular Mask, enable Display Marks, and press OK.
6. Press Shutdown, and select “Yes” to confirm.
7. Wait for the system to shutdown, and reboot the system.
NOTE
NOTE
Enabling Display Marks displays 3 iso-centric white circles, and one center crossing.
Disabling IP Circular Blanking and CCU Circular Blanking shows image outside the
normal display field.
With CCU Circular Blanking disabled, Recursive Factor button will not be effective, and
the image will show much noise.
With IP Circular Blanking disabled, Image Processing buttons (Brightness, Contrast,
Negate, Edge sharpening) will be effective only within the middle white circle region.
5-14
Calibration
BAFT Installation
1. Put the 9-inch BAFT (Beam Alignment Tool) on top of the Image Intensifier.
2. Orient the tool so that the number 1 and 2 on the BAFT tool is away from the mainframe, number
3 and 4 is on the Mainframe side. (See the picture below)
3. Orient the tool so that the tool sides are aligned with the mounting surface of the Image
Intensifier.
Aligned
in
4.
Set to Mag2 mode, and fully open the collimator, and to make a shot.
5. Check if you can see a full circular image. If camera is not centered, image does not show full
circular image. If collimator is not centered, it will shadow one side of the image boundary. If the
camera or collimator is not in reasonably calibrated position, roughly calibrate them first. Refer to
Section 5.6.3 for Camera Centering, and Section 5.6.4 for Collimator centering.
<Camera is not centered>
<Collimator shadows the image boundary>
5-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6. Align the BAFT tool to the circular optical field in the image as follows:
a. Loosen the securing screws (4 screws for each horizontal and vertical adjustment).
b. Adjust position adjustment screws (1 screw for each horizontal and vertical adjustment)
on the BAFT tool.
c. Make x-ray shot, read the BAFT tool’s tick marks on circular optical field boundary.
d. Check if BAFT tool’s tick marks reading is the same in all angles.
e. Tighten the securing screws, after the make x-ray shot, and double check the BAFT
alignment.
8 Securing Screws
2 Alignment Screws
5-16
Calibration
5.6.2. Camera Orientation
Camera Orientation Check
1. Create a new exam.
2. Set to Mag2 mode, and make an exposure.
3. Verify if the central square of the BAFT tool image is aligned with the vertical and horizontal lines
of the display screen. Visual accuracy is sufficient.
4. Verify if the BAFT tool number marking 1,2,3,4 displays as shown below.
5. Verify if the center square of BAFT tool image is aligned with the vertical and horizontal lines of
the display screen (to the original position).
Camera Home Positioning Check
1. Take an image in a rotated camera angle.
2. Create a new exam.
3. Verify if the camera rotated to the home position.
Camera Orientation Calibration
1. Refer to Camera Replacement Instruction, and check if Camera is assembled in the right angle.
2. There is no calibration available. If the camera is installed in the right angle, this is a defective
camera. Replace the camera.
5-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.6.3. Camera Centering
Camera Centering Verification
1. During a MAG 2 Fluoro
exposure, use a Dry-Erase
marker to make a dot on the
left monitor at the center of
the BAFT.
2. Rotate the Camera to 90°,
180° and 270° and mark the
BAFT center at each camera
rotation angle. Draw a line
between the 0° mark and the
180° mark. Draw a line
between the 90° mark and
the 270° mark. The
intersection of the two lines is
the Camera Rotation Center.
3. Observe the gap between the
Rotation center and BAFT
tool center. This is the decentering deviation.
4. The de-centering deviation should be within the range of the values below. (Note: Each ring in the
Bull’s eve is 1mm wide with 1mm spacing between rings.)
Mag Mode
MAG2
Decentering (mm)
≤3
5. If the requirement is not met, adjust the Camera Centering.
5-18
Calibration
Camera Centering Calibration
1. To have access to the Camera Securing Screws, perform the following.
a. Power off the system.
b. Position the C-arm to lateral position with Image Intensifier on the left and X-ray tube on
the right side when looking from the main frame.
c. Remove the I.I. back cover.
d. Disassemble the Camera cable plug-in port of the EMI shield.
e. Power on the system.
2. Loosen the 3 Camera Securing Screws. (Do not unscrew 6 EMI Shield Securing Screw).
EMI Shield
Securing Screw
Camera
Securing Screw
EMI Shield
Camera cable
plug-in port is
disassembled
3. Make sure that 4 Camera Alignment Screws are not too loose. If they are loose, tighten all 4
screws with modest torque.
Camera
Alignment
Screws (there
are 4 screws
with 90 degrees
apart)
5-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
4. Set to Mag2 mode, and make a shot.
5. Refer to the BAFT image, and use the 4 Camera alignment screws to adjust the camera center.
Make sure to unscrew one screw first, and then tighten the screw on the opposite side.
Camera Centering Example
a.
b.
c.
a. BAFT image center is not aligned with Rotation Center Marking. Need to Shift BAFT
center to the SE (bottom right) direction.
b. Shift the camera to the same direction to shift the BAFT image center, i.e., S.E. (South
East, or bottom right) direction.
c. Make a shot, and confirm the alignment.
6. Perform Camera Centering Verification. Repeat as many time as necessary.
7. When the camera centering verification meets the specification, tighten the 3 Camera Securing
Screws.
8. Remove the rotation center marking on display monitor.
Cleanup
a. Erase the Camera Rotation Center marking on the display monitor.
b. When you follow the normal beam alignment procedure, at this stage, you do not need to
reassemble the EMI shield and Camera Cable Port, and I.I. back cover. You may still need to
access the camera head for Camera focus and Camera Iris adjustment.
Recover the workstation setup.
1. Power On the workstation, and wait until the system boot up completes.
2. Press Ctrl+Alt+Shift, and type “e”,”v”,”r” to enter Service Mode.
3. Select User as “Service”, and type password as “@dminB50”, and click “Login” button.
4. Select Imaging Defaults.
5. Enable IP Circular Mask, enable CCU Circular Mask, disable Display Marks, and press OK.
5-20
Calibration
6. Press Shutdown, and select “Yes” to confirm.
7. Wait for the system to shutdown, and reboot the system.
5.6.4. Collimator Centering
Collimator Iris Centering Check / Calibration
1. Position the C-arm to AP position with
Image Intensifier below the X-ray tube.
2. Create a new exam to make default
settings.
3. Make Normal Fluoro exposures. Close the
iris a bit to show the iris near the image
boundary.
4. If not aligned, open the X-ray tube cover,
and adjust the collimator iris X and Y-axis
alignment screws to center the collimator
on the BAFT. (Reference the BAFT image to
determine the direction to shift the
collimator iris)
5. After Iris center is adjusted, make sure to
tighten the collimator’s X and Y-axis screws.
6.
Make another exposure to verify that the
collimator iris is still centered.
5-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Shutter Centering Check / Calibration
1. Close the collimator shutter to the smallest.
2. Select manual Fluoroscopy mode and take
exposures.
3. Rotate the shutter and stop at vertical position
during live exposure.
4. Open the collimator shutters a little to see the
BAFT center square.
5. Check if the center of the shutter is aligned with
the BAFT center square.
6. If not aligned, loosen the securing screws, and
adjust collimator shutter vertical adjustment
screw. (Reference the BAFT image to determine
the direction to shift)
7. Rotate the shutter and stop at horizontal position
during live exposure.
8. Check if the center of the shutter is aligned with
the BAFT center square.
9. If not aligned, loosen the securing screws, and
adjust two collimator shutter horizontal
adjustment screws. (Reference the BAFT image
to determine the direction to shift)
10. After shutter centering is adjusted, make sure to
tighten the 2 securing screws.
11. Make another exposure to verify that the
collimator shutter is still centered.
Cleanup
1. When collimator iris/shutter center is adjusted, make sure to tighten all 2 Collimator Iris securing
screws, and 2 Collimator shutter securing screws.
2. Assemble the X-ray tube cover.
3. Remove the BAFT tool from the Image Intensifier
5-22
Calibration
5.6.5. I.I. Field Size
I.I. Field Size Verification
1. Select relative field size, open the collimator to the maximum, and take an exposure.
2. Read the BAFT tick mark on the boundary of the image. The size of the image should be within
the range of the values below.
Mag Mode
NORM
MAG1
MAG2
Displayed Image
Diameter (mm)
215 ±8
157 ±8
119 ±8
3. If the requirement is not met, you must perform I.I. Field Size Calibration.
I.I. Field Size Calibration
I. I. Size adjustment is done in the factory. Adjust with much care only if necessary.
NOTE
Always use non-metal flathead screwdriver to adjust the I.I. potentiometer to avoid short
circuit.
1. Open the I. I. Potentiometer cover, and carefully adjust the I. I. Size trimmer.
2. Make an exposure, and check the resolution again. Repeat as many times as necessary.
3. Close the I.I. Potentiometer cover.
A. Size
B. Focus
C. Edge
5-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.6.6. Limiting Field Size (Collimator Iris Field Size)
Limiting Field Size Verification
1. Create a new exam.
2. Load a film cassette in the BAFT.
3. Make an exposure at 50kVp@1.3mAs in Normal, MAG 1, MAG 2, and then with the collimator iris
completely closed. Take the four exposures on one sheet of film. Note: to select the Film MAG
Mode, do the following: Select Manual - Fluoro, then Normal, MAG1 or MAG 2 and then Film
mode.
4. Remove the Film Cassette and develop the film.
5. Measure, using the ruler, the maximum diameter of the collimator iris as projected on the film
when the collimator iris is fully closed. Ensure that it is less than 50.0mm.
6. Measure, using the ruler, the Normal field size diameter of the collimator iris as projected on the
film at 45° and 135° angle.
7. Set the system to Normal field size, and take an exposure.
8. Measure, using the BAFT tick marks on the display monitor, the diameter of the circular image at
45° and 135° angle.
9. Calculate the diameter difference between film and display monitor at each 45° and 135° angle.
10. The difference should be within the range of the values below.
Mag Mode
NORM
MAG1
MAG2
Diameter difference
at 45° and 135° angle
≤ 10 mm
≤ 10 mm
≤ 10 mm
11. Calculate the sum of diameter difference at 45° and 135° angle.
12. The difference should be within the rnage of the values below.
Mag Mode
NORM
MAG1
MAG2
Combined diameter
difference
≤ 16 mm
≤ 16 mm
≤ 16 mm
13. Repeat step 6 to step 12 for MAG1 and MAG2 field size.
14. If the requirements are not met, you must perform Limiting Field Size (Collimator Iris Field Size)
Calibration.
Limiting Field Size (Collimator Iris Field Size) Calibration
Refer to Collimator Replacement instruction to adjust default opening for collimator iris.
5-24
Calibration
5.6.7. ABS Tracking Check and Adjustment
ABS Tracking Check
1. Position the C-arm to AP position with I.I. above the X-ray tube.
2. Select Auto Fluoroscopy mode, and the NORM field size to take an exposure. Put the 1mm,
2mm, and 3mm copper filter on the tube head.
3. Verify that the kVp stabilizes in the specified range below for each copper filter thickness:
Copper Filter Thickness
kVp range
1mm
2mm
3mm
64±4kV
74±4kV
82±4kV
4. If the ABS Tracking check fails, proceed to Entrance Dose Rate Check and Camera Iris
Adjustment.
Entrance Dose Rate Check and Camera Iris Adjustment
NOTE
Camera Iris adjustment is done in the factory. Adjust with much care only if necessary.
1. Place the dosimeter probe on the I.I. grid center, and secure it with transparent tape.
2. Use 2mm copper to cover the beam path on the X-ray tube, and secure it with tape.
3. Position the C-arm to lateral position.
4. Make an exposure in Auto Fluoroscopy, and check if the dose rate is within 4.0 ± 0.3mR/min
(35.04 ± 2.63 μGy/min or 2102 ± 157 μGy/sec).
5. If failed, open the Camera head cover, and carefully adjust the camera iris trimmer.
6. Make an exposure to make sure the entrance dose rate 4.0mR/min (35.04 μGy/min or 2102
μGy/sec) as close as possible. Repeat as many time as necessary.
5-25
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Camera Iris
CCW: Close Iris to increase dose rate
CW: Open Iris to decrease dose rate
Camera Focus
7. If failed after many trials, check the kVp and mA calibration, and come back to Camera Iris
Adjustment.
8. When passed, remove the dosimeter probe from the I.I., and copper from the X-ray tube, and
perform ABS Tracking check.
Cleanup
When you follow the normal beam alignment procedure, at this stage, you do not need to
reassemble the Camera Head Cover. You may still need to access the camera head for Camera focus
adjustment.
5-26
Calibration
5.6.8. Limiting Resolution
Limiting Resolution Setup
1. Position the C-arm to AP position with I.I. below the X-ray tube.
2. Create a new exam to make default settings.
3. Place a Spatial Resolution tool on the Image Intensifier, and secure it with transparent tape. The
angle between the horizontal axis of the image and the Spatial Resolution tool shall be around 45
degrees.
~45°
4. Select Auto Fluoroscopy mode. Take an exposure, and stop when image is stable. Adjust kVp and
mA if necessary.
5. Read the limiting resolution on the display image. It is the maximum number (lp/mm) where half
of the lines are distinguishable.
6. Verify if the resolution is within the specified value range below.
7. Repeat step 2 to step 6 for MAG1 mode.
8. Repeat step 2 to step 6 for MAG2 mode.
Field Size
NORM
MAG1
MAG2
Resolution (lp/mm)
≥ 2.2
≥ 3.0
≥ 3.5
9. If the resolution check fails, adjust the Camera Focus.
5-27
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Camera Focus Calibration
NOTE
Camera Focus adjustment is done in the factory. Adjust with much care only if
necessary.
1. To have access to CCD Camera Focus Trimmer, perform the following.
a. Turn off the power.
b. Position the C-arm to lateral position.
c. Disassemble plug-in port of Camera cable, and disconnect the Camera cable.
d. Disassemble EMI shield by unscrewing 6 screws. (Do not remove 3 Camera securing
screws)
e. Open the CCD Camera Head Cover.
f.
Reconnect the Camera Cable.
g. Power on the system.
2. Use a small flat head screwdriver to carefully adjust the camera focus trimmer.
3. Make an exposure, and check the resolution again. Repeat as many times as necessary.
4. If the resolution check still fails after many trials, proceed to I.I. Focus adjustment.
Camera Iris
Camera Focus
Slowly turn CW or CCW
5-28
Calibration
Image Intensifier Focus Calibration
I.I. Focus adjustment is done in the factory. Adjust with much care only if necessary.
Always use non-metal flathead screwdriver to adjust the I.I. potentiometer to avoid short
circuit.
NOTE
1. Open the I. I. Potentiometer cover, and carefully adjust the I. I. Focus trimmer.
2. Make an exposure, and check the resolution again. Repeat as many times as necessary.
3. Put the I.I. Potentiometer cover back.
A. Size
B. Focus
C. Edge
Cleanup
1. After passing, remove the resolution tool from the I.I. Grid.
2. Make sure to re-assembly of all components as the following:
a. Power off the system.
b. Assemble Camera Head Cover.
c. Disconnect Camera Head Cable.
d. Assemble EMI shield by screwing 6 securing screws.
e. Connect Camera Head Cable, and make sure to tighten the securing screw.
f.
Assemble the Camera Head Cable Port in the EMI Shield.
g. Assemble the I.I. cover.
h. Power on the system.
5-29
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5.6.9. Shading Correction
Shading Check
1. Put 1mm Copper on the X-ray tube and secure it with tape.
2. Set to Normal Mag mode, Auto Technique On.
3. Take a Shot.
4. Turn on the Automatic Brightness& Contrast.
5. Keep the brightness level and manually increase Contrast Level to see well on Brightness
uniformity. Adjust the brightness level if necessary.
6. At contrast level 67, pass if you see the whole circular region.
For example, at contrast 67, the 1st image shows the whole circular region, and the 2nd image
dose not.
7. At contrast level 85 or higher, pass if you do not see the central region darker than the outside.
For example, at contrast 90, the 1st image shows brighter center, and 2nd image shows darker
center. Too much shading makes darker center.
8. If you fail in step 5 or 6, proceed to Shading Correction.
5-30
Calibration
Shading Correction
1. Power off the system, and wait for
complete shutdown.
2. Open Mainframe Back Cover to access t
he CCU.
3 CCU
Securing
Screws
3. Disconnect the cables from CCU.
4. Unscrew 3 CCU securing screws, and shift
the CCU to the left side and pull out the
CCU.
5. Unscrew 8 CCU Cover screws, and open
the CCU Cover.
6. Assemble the CCU to the Mainframe with
the CCU cover open.
7. Connect the cables to CCU. (Power,
Camera Cable DB9, Camera Cable DB25,
Video, Frame Sync).
Power
Camera Cable
DB25
Video
8. Power on the system.
Camera
Cable DB9
Frame
Sync
<Warning> The following steps are strongly recommended to reduce the Radiation Hazard to the FSE.
a. Position the C-arm so that the X-ray tube is directing away from the Mainframe. This is
important to minimize the radiation hazard of the FES.
b. Make sure to wear Radiation Protective Garment.
c. If available, locate the mobile Radiation Protective Shield between X-ray tube and
Mainframe.
9. Put 1mm copper on the X-ray tube, and secure it with tape.
10. Make an exposure with Auto Technique on.
11. Turn on the Automatic Brightness& Contrast.
12. Keep the brightness level and manually increase Contrast level to see well on Brightness
uniformity. Adjust the brightness level if necessary.
13. Manually increase contrast to 85.
5-31
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
14. During live x-ray exposures, adjust trimmer R61, R52, R47, and R48 to adjust brightness
uniformity. Keep contrast level 85 and adjust brightness if necessary to see the shading effect.
R61
R52
R47
R48
R61
R52
R47
R48
Vertical Level
Vertical Position
Horizontal Position
Horizontal Level
15. Adjust contrast level 85 or higher, and make sure that the center region is not darker than the
image boundary.
Cleanup
1. Power off the system, and wait for complete shutdown.
2. Disconnect the cables from CCU.
c. Disassemble the CCU from the Mainframe.
d. Assemble CCU cover, and tighten 8 cover screws.
e. Assemble the CCU to the Mainframe, and tighten 3 CCU securing screws to the Mainframe.
f.
Connect the cables to CCU (Power, Camera Cable DB9, Camera Cable DB25, Video, Frame Sync;
reference the picture in step 8 above).
g. Re-assemble the Mainframe Back Cover.
h. Power on the system.
5-32
Calibration
5.7. DAP and Dose Calibration
Refer to section 3.7.5 for DAP and dose calibration.
5.8. Monitor Brightness and Contrast Adjustment
5.8.1. Keys assignment and operation LED
A “dynamic help for keypad function” is available for each menu. It explains the role of each
Key on the OSD menu window, which is currently active.
5.8.2. Lock and unlock the OSD menu
Action
Lock OSD menu (except hotkeys
if enabled)
Unlock OSD menu (except
hotkeys if enabled)
Key sequence
Press key Exit and Up Up Up in 3 seconds,
when keys are unlocked.
Press key Exit and Up Up Up in 3 seconds,
when keys are locked
5.8.3. Key functions in the OSD menu
Key(s)
Situation
Action
Menu
Always
Jump to next line
Up
Slide controller
Increase value
Selection point
To previous selection
Command
"Enter key"
Slide controller
Decrease value
Down
5-33
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Exit
Selection point
Except "Exit OSD" menu
In "Exit OSD" menu
To subsequent selection
One menu level upwards (settings are
retained)
Return to main menu (settings are retained)
5.8.4. Brightness and Contrast Adjustment
Adjust monitor’s
Contrast
Adjust monitor’s
Backlight.
NOTE
Main Menu
The monitor is factory calibrated. For DVI-D input we suggest don’t change the contrast
value, any change of this will change the optimized LUT settings and will lose some of
the image information. We suggest to only adjust the backlight of the monitor if the
luminance is not acceptable.
Function
Adjustment range
Description
Adjustment of contrast
Brightness
Contrast
0…255
This allows the brighter area to be seen more
distinctly. The center point is in 128 position.
Note: for DVI-D signals the Contrast setting is
optimized. Manual changes are not recommended.
It is used to adjust the black level of the monitor. The
center point is in 128 position.
Brightness
0…255
This allows the darker area to be seen more
distinctly.
Note: for DVI-D signals the brightness setting is
optimized. Manual changes are not recommended.
Backlight
5-34
0…255
Adjustment brightness of LCD backlight to adapt
total brightness for room illumination
Calibration
SETTING1
Settings
SETTING2
SETTING3
While choose backlight it will recall factory default
contrast, brightness, backlight.
User
Accept changes
Quit OSD
menu
Reject changes
OSD is
locked
Quick OSD menu and either save or reject the
changes. If you have reached this menu
unintentionally, you can return to the main menu
using the EXIT key.
If this message is displayed, you do not have the
authority to carry out the changes in the OSD menu.
Please contact your servicing partner in this case.
5-35
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5-36
Chapter6. Software
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.1. Overview
Brivo OEC 850 Mobile C-Arm X-ray Product provides service tool in software for the service engineer.
Using the service tool, service engineer can:
•
Set service ID, language and time zone for system
•
Get system information
•
Exposure management
•
Database Backup and Restore
•
Set Imaging Defaults
•
Consistency Check
•
Import/Export image
•
Export Config file
•
Restore Defaults
•
Search and export system log
This chapter will describe how to use the service tool and the installation and upgrade procedure for
software.
6-2
Software
6.2. Service Tool
Hold down Ctrl +Shift+Alt key and press ”E”, ”V”, ”R” keys orderly and quickly to invoke the service tool,
when Brivo OEC 850 software is running in user interface. The service tool login screen will display.
Select Service as the user. Input the password: @dminB50. Click “Login” button to enter the main screen
of service tool.
Click “Shut Down” button to shut the down the system. Click “Logout” button to return to service tool
login screen.
6-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.2.1. Service ID Setting
Click “Service ID Setting… “ button on the main screen to set or update service ID for the system. Current
service ID is displayed on the screen automatically. Fill in the new service ID in the textbox. Then press
“OK” button to update.
The service ID will be left blank before leaving factory. Field engineer must set Service ID during
installation and make sure it’s the same as the SID in service record system (GIB). This ID can’t be
changed after the first setting, if it is not necessary.
6-4
Software
6.2.2. Multi-language Setting
Click “Multi-Language Setting…” button on main screen to select the local language by country.
6.2.3. Time Zone Setting
Click “ Time Zone…” button on main screen to set time zone for system.
6-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.2.4. Get System Information
Press “System information…” on main screen to get the system information, including:
6-6
•
Version of workstation software
•
Version of MCB software
•
Number of saved exams in system
•
Number of saved images in system
•
Free disk space
•
Total disk space
Software
6.2.5. Exposure Management
Select “Exposure Management…” on the main screen.
The exposure management screen is displayed.
•
Dose Group Setting
Input Voltage 100V AC: Enable or disable “AC100V power” according to the AC power used. If AC
100V power is enabled, the Dose Group will be set as Normal 50 and HLF 50 automatically.
NOTE
The following section describes the instruction how to set Dose Group and X-ray Control
according to the following Table “Exposure Management Matrix”. If the instruction
contradicts the corresponding regulatory requirements of certain countries or regions (for
example, some regulatory requirements are updated), change the setting according to the
regulatory requirements.
Select Dose Group option by country to set the dose limitation in normal (standard) and HLF
exposure. The available options are:
Normal 88 and HLF 100
Normal 50 and HLF 50
Normal 50 and HLF 100
Refer to the following table “Exposure Management Matrix” to set Dose Group.
The skin dose rate in the Dose Group represents the value for conditions of free-in-air irradiation at the
patient skin: 30cm along the beam axis from the image intensifier surface.
6-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1. Two dose groups in nomral (standard) fluoro mode
-Normal50: The skin dose rate does not exceed 50mGy/min in standard fluoro mode, such as
Australia, Japan, New Zealand, etc.
-Normal88: The skin dose rate does not exceed 88mGy/min in standard fluoro mode, such as
USA, Germany, etc.
2. Two dose groups in HLF mode:
-HLF50: The skin dose rate does not exceed 50 mGy/min in HLF mode, such as Australia, New
Zealand, etc.
-HLF100: The skin dose rate does not exceed 100 mGy/min in HLF mode, such as Japan, Italy, UK,
etc.
For the country, such as USA, if the skin dose rate shall not exceed 176 mGy/min in HLF mode from its
regulatory requirements, the system will use HLF100 in HLF mode because the maximum skin dose rate
of the X-ray tube cannot exceed 100 mGy/min.
3. For some country, its dose limitation should be the combination of the above normal and HLF groups.
For example,
-Australia: Normal50 and HLF50
-China: Normal88 and HLF100
4. For countries or regions that have a higher standard fluoro dose (at the patient skin) rate limit than
USA, the dose limits are set to the same value as for USA.
5. For countries or regions that have a higher HLF dose (at the patient skin) rate limit than USA, the dose
limits are set to the same value as for USA.
6. For countries or regions that do not specify standard fluoro dose at patient skin, the dose limits are set
to the same value as for USA.
7. For countries or regions that do not specify HLF dose at patient skin, the dose limits are set to the
same value as for USA.
•
HLF scene setting:
Accessible Only In Auto: Enable or disable “HLF mode only accessible in auto mode”. Some
countries or regions, such as certain provincial locations in Australia, may require enabling
it.
Exposure Limit 20 Sec: Enable or disable “HLF terminates after 20 seconds of continuous
exposure”. Some countries or regions, such as Australia or Switzerland, require enabling it.
Auto-return After Exposure: Enable or disable “ The system automatically returns to normal
mode after HLF termination”. Some countries or regions, such as certain provincial
locations in Australia, may require enabling it.
Auto-return After 5min Without X-Ray: Enable or disable “ The system returns from HLF
mode to normal mode if not used for 5 minutes”. Some countries or regions, such as
certain provincial locations in Australia, may require enabling it.
Refer to the following table “Exposure Management Matrix” to set HLF scene setting.
6-8
Software
•
Fluoro Alarm Control:
Enable or disable X-ray termination function after 9’30’’ fluoro. If this function is enable,
when the accumulated fluoroscopy time reaches 9’30’’, X-rays will be terminated
automatically.
Refer to the following table “Exposure Management Matrix” to set Fluoro Alarm Control.
•
Select DAP option to enable or disable DAP function.
NOTE
•
Only when the system is equipped with DAP meter, can DAP option be enabled.
Calibration:
Fill in DAP calibration coefficient. The value should be from 0.0 to 10.0.
Fill in Dose calibration coefficient. The value should be from 0.0 to 1.0.
These two coefficients are used to DAP and Dose calibration. Refer to chapter 3.
Click OK button to save the setting. Select Cancel button to leave the screen.
Table : Exposure Management Matrix
Countries /
Regions
Auto-return
Auto-return After 5min Terminate XDose Group Accessible
Exposure
After
Without X- Ray After
Option
Only In Auto Limit 20 Sec Exposure
9'30" Fluoro
Ray
Algeria
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Argentina
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Australia
Enable for
certain
Normal 50
provincial
and HLF 100 locations
Enable
Enable for
certain
provincial
locations
Enable for
certain
provincial
locations
Enable
Austria
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Bangladesh
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Belgium
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Bolivia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Bosnia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
6-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Countries /
Regions
Auto-return
Auto-return After 5min Terminate XDose Group Accessible
Exposure
After
Without X- Ray After
Option
Only In Auto Limit 20 Sec Exposure
Ray
9'30" Fluoro
Brazil
Normal
50
and HLF 100 Disable
Disable
Disable
Disable
Enable
Brunei
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Bulgaria
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Cambodia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Canada
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Chile
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
China (P.R.)
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Columbia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Costa Rica
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Cyprus
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Czech
Republic
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Denmark
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Dominican
Republic
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Egypt
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Estonia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Finland
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
France
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
6-10
Software
Countries /
Regions
Auto-return
Auto-return After 5min Terminate XDose Group Accessible
Exposure
After
Without X- Ray After
Option
Only In Auto Limit 20 Sec Exposure
Ray
9'30" Fluoro
Georgia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Germany
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Greece
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Guatemala
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Honduras
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Hongkong
/Macau
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Hungary
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Iceland
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
India
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Indonesia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Iraq
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Ireland
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Israel
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Italy
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Jamaica
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Japan
Normal 50
and HLF 50
Disable
Disable
Disable
Disable
Disable
Jordan
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
6-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Countries /
Regions
Auto-return
Auto-return After 5min Terminate XDose Group Accessible
Exposure
After
Without X- Ray After
Option
Only In Auto Limit 20 Sec Exposure
Ray
9'30" Fluoro
Kuwait
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Laos
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Latvia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Lebanon
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Libya
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Lithuania
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
Liechtenstein and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
Luxembourg and HLF 100 Disable
Disable
Disable
Disable
Enable
Macedonia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Malaysia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Malta
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Mexico
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Normal 88
Montenegro and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal
50
New Zealand and HLF 100 Disable
Enable
Disable
Disable
Enable
Normal 88
Netherlands and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Morocco
Norway
6-12
Software
Countries /
Regions
Auto-return
Auto-return After 5min Terminate XDose Group Accessible
Exposure
After
Without X- Ray After
Option
Only In Auto Limit 20 Sec Exposure
Ray
9'30" Fluoro
Oman
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Pakistan
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Panama
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Paraguay
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Peru
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Philippines
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Poland
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Portugal
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Puerto Rico
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Romania
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Russia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
Saudi Arabia and HLF 100 Disable
Disable
Disable
Disable
Enable
Singapore
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Slovenia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
South Africa and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
South Korea and HLF 100 Disable
Disable
Disable
Disable
Enable
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Spain
6-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Countries /
Regions
Auto-return
Auto-return After 5min Terminate XDose Group Accessible
Exposure
After
Without X- Ray After
Option
Only In Auto Limit 20 Sec Exposure
Ray
9'30" Fluoro
Sweden
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Switzerland
Normal
88
and HLF 100 Disable
Enable
Disable
Disable
Enable
Taiwan
(Republic of
China)
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Thailand
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Trinidad &
Tobago
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Tunisia
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Turkey
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
United Arab
Emirates
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
USA
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
United
Kingdom
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Uruguay
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
Venezuela
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Disable
Vietnam
Normal 88
and HLF 100 Disable
Disable
Disable
Disable
Enable
6-14
Software
6.2.6. Database Backup and Restore
Click “Database Backup/Restore…” button to display the screen.
You can backup the current database file on this screen. The database file will save patient information
and images. Click “Start backup…” button to save the current database file. A double confirm screen will
pop up. Click “Yes” button to start backup. The process will complete until “Database backup finished.”
message display.
Click “Start Restore…”. The backup time of the file will be displayed in the name of the backup file. You
can choose a database according to the time. Click Restore button to restore the file. After double
confirm, the restore will complete until “Database restore finished.” message display.
6-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.2.7. Set Imaging Defaults
The service tool allows to set imaging parameters’ defaults. Click “Image Defaults…” on the main screen.
NOTE
Changes on Image Brightness, Image Contrast, Black/Wihte Margin and Sharpen
power are not recommended because these values have been optimized for the image
quality to compromise any anatolomical imaging. Any changes will affect the system
image quality.
•
Image Brightness: Pressing up or down arrow or dragging the block allows you to adjust image
brightness default from 0 to 100.
•
Image Contrast: Pressing up or down arrow or dragging the block allows you to adjust image
contrast default from 0 to 100.
•
Black/White Margin: This adjustment shall be made very carefully only when necessary, as it may
affect the overall image quality. Pressing up or down arrow allows you to adjust image margin
from 0 to 50%. Increasing White Margin value may enhance contrast in bright regions, and
increasing Black Margin value may enhance contrast in dark regions. Increasing White Margin
may decrease overall image brightness. Increasing Black Margin may increase overall image
brightness.
•
Sharpen Power: Select default value of sharpening power from 0-10.
•
Enable or disable IP circular mask: it is used to assist beam alignment calibration. Make sure it is
set back to the default setting “Enable” after beam alignment.
•
Enable or disable CCU circular mask: it is used to assist beam alignment calibration. Make sure it
is set back to the default setting “Enable” after beam alignment.
6-16
Software
•
Enable or disable Display Marks: it is used to assist beam alignment calibration. Make sure it is
set back to the default setting “Disable” after beam alignment.
6.2.8. Consistency Check
The consistency check function is to check the consistency of patient information with database. Click
“Consistency Check…” button on main screen to display the screen.
Click “Check” button on the screen. And then click “Yes” button in confirmation screen to start the check
process. The result will display on the frame after the process complete.
6-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.2.9. Image Import/Export
Service tool allows importing images from USB disk to local disk or exporting images from local disk to
USB device.
1. Click “Image Import/Export…” on the main screen. The “Image Import& Export” screen will display.
2. If you want to import images from USB to local disk, click Import button on the screen. The screen will
show local directory used to save the images in “Copy To” column.
6-18
Software
(1) Insert USB disk to USB interface. Click “Copy From” button to select USB devices to copy from.
Choose the device in the USB list and click “OK”.
NOTE
Only those images, which are exported from Brivo OEC850 system, can be imported
back.
(2) Select the images to copy.
Select All: Used to select all images in the gallery for copying.
Up/Down Arrow: Used to scroll through all preview images. The arrow buttons are only active if
more than sixteen images are retrieved.
(3) Click “Copy” button. Click “Yes” button in confirmation screen to execute a copy operation.
Import images to local disk. The process will complete with a message: “ Image Import finished.”.
6-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3. If you want to export images from local disk to USB, click export button on the screen. The screen will
show local saved directory to copy from.
(1) Click “Copy To” button. Select USB device as target destination to copy. Choose the device in the
USB list and click OK.
(2) Select the images to copy.
Select All: Used to select all images in the gallery for copying.
Up/Down Arrow: Used to scroll through all preview images. The arrow buttons are only active if
more than sixteen images are retrieved.
(3) Click “Copy” button. Then click “Yes” in confirmation screen to execute a copy operation. Copy
selected images to the designated destination. The process will complete when the message
“Image export finished.” displayed.
NOTE
6-20
The exported images can’t be viewed in other computer and only can be imported to
Brivo OEC850 systems.
Software
6.2.10. Log Viewer
Log viewer screen allows service engineer to review the system log including the event of software and
exposure and detailed information of error, warning and information messages.
Click “Log Viewer…” button on main screen.
1
2
3
4
5
6
1. Search Options: Enable to set the conditions to search.
Date: Used to select the period to show the log. You can search the logs created in the selected
period. The available options are: all date, day, week, month, half year or one year.
Type: Select the log type including error, warning, information and event.
Sub-system: Search the log information according to software sub-system that the event or failure
happened on.
Search for: Input the key words of the log in the textbox to search.
Error code: Input the error code displayed on the monitor and control panel to search the log.
Module: Search the log information according to module that the event or failure happened on. The
available options are: computer, software, X-ray exposure and all.
2. Search
Click the button to start the search operation.
6-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3. Page Number
Display the total and current page number of the log. Click the arrows to turn the pages.
4. Log Information
Display the search results of the log. This field will show the error code, Type, sub-system, module,
create date and log text of the log.
5. Reset
Click “Reset” button to clear all selected search option and restore to defaults.
6. Export
Click “Export” button to export log file to USB disk. The export complete with a message “Log export
finished.”.
6-22
Software
6.2.11. Config File Export
The config file export screen allows service engineer to export system configuration files:
IPConfig.xml: This file contains the parameter of image processing, including the setting values in
“Image Defaults” screen.
ServiceTool.xml: This file contains all the setting values in service tool except for image related.
Click “Config File Export…” on main screen.
Click “Target Path” button to select the destination.
Click “Export” button to start the export process. When the process complete, a message “ Export config
files finished.” will display.
6-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.2.12. Restore Default
Click “Restore Default…” on main screen.
From this screen, you can:
Delete all database records and all images.
WARNING
Click “Delete” button will remove all patient information and all images.
Clear all system logs. Click the “Clear” button to cleanup all system logs.
Restore the information in ServiceTool.xml to defaults. It will restore all setting parameters in service tool
to defaults, except for those in “Image Defaults” screen.
Restore the information in IPConfig.xml to defaults. It will restore parameters in “Image Defaults” screen
to defaults.
Restore the information in WKSSetup.xml to defaults. It will restore all parameters in “Setup” screen of
workstation user interface.
6-24
Software
6.3. Software Installation and Upgrade
6.3.1. Software Installation
There are two software installation disks for Brivo OEC 850 C-Arm X-ray Product. One is the Operating
System (OS) installation CD. The other is the Application CD. The OS installation CD should be installed
prior to Application during the initial installation.
NOTE
System will sounds beeps during the whole software installation procedure.
Install Operating System
CAUTION
You must press “Del” key immediately during system booting in step 1 and step8.
1. Turn on the system, during the memory testing, press “Del” key on the keyboard to get to BIOS.
6-25
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
2. Type in “EVR850BJ”, then press “Enter” key.
3.
6-26
When the BIOS menu shows, use the arrow down button to select the “Advanced BIOS
Features” and then press “Enter”.
Software
4. Select “First Boot Device” and press “Enter”. Then choose the “CDROM” and press “Enter”.
6-27
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
5. Press the “
” key and then press “Enter”. This will save your settings, and the system will
reboot. (The system is now set to boot from CD ROM first.)
6-28
Software
6. When the screen display “DISK BOOT FAILURE, INSERT SYSTEM DISK AND PRESS ENTER”, insert the
OS installation CD into the CD/DVD drive of the WS, then press “Enter”.
7. Once you get to the prompt, type in “SG_EVR”. The OS software will then load on the system.
6-29
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8. Once the software installation is complete, CD will be rejected automatically. Remove the CD and
close the CD tray. Click the “reboot” button.
9. During the memory testing, press “Del” key on the keyboard to get to BIOS.
10. Type in “EVR850BJ” , then press “Enter” key.
11. When the BIOS menu shows, use the arrow down button and “Enter” key to select the “Advanced
BIOS Features” and then press “Enter”.
6-30
Software
12. Select “First Boot Device” and press “Enter”. Then choose the “HDD-0” and press “Enter”.
“key and then press “Enter”. This will save your settings, and the system will
13. Press the “
reboot. (The system is now set to boot from hard disk first)
6-31
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
14. System will enter GDM desktop with a terminal pop up.
6-32
Software
Install Application Software
1. Insert the Application CD to CD/DVD drive of the workstation. Click “Run Command” in the warning
message box.
2.The welcome screen will display. And the serial number of the system from the manufacture will be
displayed automatically.
6-33
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Select “OK” to start the Stage 1 process.
6-34
Software
If the user data is detected on the workstation, the welcome screen will ask you to choose whether retain
user data or not.
If you want to retain user data, make Option 1 selected and enter OK. Select “Yes” to confirm your
selection and start the Stage 1 process. Select “No” to return back the previous welcome screen.
CAUTION
Select do not retain the existing user data will delete all existing data and it cannot be
recovered.
6-35
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3. After stage1 process complete, enter “OK” to reboot the workstation.
The Stage2 process will start automatically after reboot.
6-36
Software
If “Retain existing user data” option is selected restoring user data screen will display after the
application packages installed.
4. At the end of stage2, press “OK” to shut down the system.
Functional test: Power on the system. The left monitor of workstation will display GE logo screen. The
right monitor will enter into the PATIENT INFORMATION Screen automatically. Perform workstation
operation check in section3.6.2.
6-37
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6.3.2. Application Software Upgrade
1. Login service tool. Click “Upgrade” button on main screen.
Preparation process screen will display.
2. After the preparation process the workstation will reboot.
NOTE
6-38
Error code will display on the control panel board when the system enter reboot process.
Software
3. Login the system with “root” account and the password is “install”.
4. System will enter GDM desktop with a terminal pop up.
5. Insert the Application CD to CD/DVD drive of the workstation. Click “Run Command” in the warning
message box.
6-39
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6. The welcome screen will display. And the serial number of the system from the manufacture will be
displayed automatically.
Click “OK” button and the system will ask you to choose whether retain user data or not.
If you want to retain user data, make Option 1 selected and enter OK. Select “No” to return back the
previous welcome screen. Select “Yes” to confirm your selection and start the Stage 1 process.
6-40
Software
7. The stage1 of installation will start. The rest processes will be the same as the installation process
including stage1 and stage 2.
6-41
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9. After stage1 process complete, enter “OK” to reboot the workstation.
10. The Stage2 process will start automatically after reboot.
6-42
Software
If “Retain existing user data” option is selected restoring user data screen will display after the
application packages installed.
11. At the end of stage2, press “OK” to shut down the system.
Due to C-arm can’t shut down automatically after upgrade process completes, you should press shut
down button on control panel manually.
6-43
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Functional test: Power on the system. The left monitor of workstation will display GE logo screen. The
right monitor will enter into the PATIENT INFORMATION Screen automatically. Perform workstation
operation check in section3.6.2.
6-44
7.
Chapter7. Troubleshooting
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
7.1. Overview
This chapter describes messages that appear on the right workstation monitor during system operation.
Messages may indicate any of the following:
•
Information
•
Warning Messages
• Error Messages
When the error message appears, error code will also displayed on the kV, mA display field on control
panel.
7.2. Error Recovery Steps
Perform the following error recovery procedures if you encounter problems during start-up or operation:
1. Some messages can be cleared by pressing OK button or taking exposure, but will periodically redisplay while the condition exists.
2. If an error message occurs, power off the system. Wait one minute, and then press the power on key.
If this fails to restore normal operation then proceed with step 3.
3. Press the power off key, disconnect the power cord, and call for service. Refer to chapter 1 for
communication center telephone numbers. Do not continue using the system.
System may fail at any time. Ignoring error and warning messages may result in
equipment damage and personal injury.
WARNING
7.3. Messages
Error
Code
00001
00002
00003
7-2
Type
Error
Error
Error
Message
"Please turn the system off, wait 1
minute, and then restart the system.
Comments
Servant activation fails.
If this message persists, call for service.
Code:00001"
"Please turn the system off, wait 1
minute, and then restart the system.
Memory allocation fails.
If this message persists, call for service.
Code:00002"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:00003"
Binding name service fails.
Troubleshooting
00004
00005
00006
00007
00008
00009
00010
00011
00012
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Invoking interface fails.
If this message persists, call for service.
Code:00004"
"Please turn the system off, wait 1
minute, and then restart the system.
Canceling thread fails.
If this message persists, call for service.
Code:00005"
"Please turn the system off, wait 1
minute, and then restart the system.
CORBA notification service connection
fails.
If this message persists, call for service.
Code:00006"
"Please turn the system off, wait 1
minute, and then restart the system.
CORBA notification service
disconnection fails.
If this message persists, call for service.
Code:00007"
"Please turn the system off, wait 1
minute, and then restart the system.
ORB creation fails.
If this message persists, call for service.
Code:00008"
"Please turn the system off, wait 1
minute, and then restart the system.
Database operation fails.
If this message persists, call for service.
Code:00009"
"Please turn the system off, wait 1
minute, and then restart the system.
Servant deactivation fails.
If this message persists, call for service.
Code:00010"
"Please turn the system off, wait 1
minute, and then restart the system.
ORB destruction fails.
If this message persists, call for service.
Code:00011"
"Please turn the system off, wait 1
minute, and then restart the system.
ORB initialization fails.
If this message persists, call for service.
Code:00012"
7-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
00013
00014
00015
00016
00017
00018
00019
00020
00021
7-4
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Joining thread fails.
If this message persists, call for service.
Code:00013"
"Please turn the system off, wait 1
minute, and then restart the system.
Loading multi-language fails.
If this message persists, call for service.
Code:00014"
"Please turn the system off, wait 1
minute, and then restart the system.
Resource file crashes.
If this message persists, call for service.
Code:00015"
"Please turn the system off, wait 1
minute, and then restart the system.
CORBA name service resolution fails.
If this message persists, call for service.
Code:00016"
"Please turn the system off, wait 1
minute, and then restart the system.
CORBA ORB execution fails.
If this message persists, call for service.
Code:00017"
"Please turn the system off, wait 1
minute, and then restart the system.
Setting UI to front fails.
If this message persists, call for service.
Code:00018"
"Please turn the system off, wait 1
minute, and then restart the system.
Starting thread fails.
If this message persists, call for service.
Code:00019"
"Please turn the system off, wait 1
minute, and then restart the system.
Accessing system resource fails.
If this message persists, call for service.
Code:00020"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:00021"
Unbinding CORBA name service fails.
Troubleshooting
00022
00023
00101
00102
00103
00104
00105
00106
00107
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Java module fails to wait for notify.
If this message persists, call for service.
Code:00022"
“Please turn the system off, wait 1
minute, and then restart the system.
UI activation fails.
If this message persists, call for service.
Code:00023”
"Please turn the system off, wait 1
minute, and then restart the system.
Serial port closedown fails.
If this message persists, call for service.
Code:00101"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication disconnected.
If this message persists, call for service.
Code:00102"
"Please turn the system off, wait 1
minute, and then restart the system.
Incompatible RCC version.
If this message persists, call for service.
Code:00103"
"Please turn the system off, wait 1
minute, and then restart the system.
Opening serial port fails.
If this message persists, call for service.
Code:00104"
"Please turn the system off, wait 1
minute, and then restart the system.
Reading serial port fails.
If this message persists, call for service.
Code:00105"
"Please turn the system off, wait 1
minute, and then restart the system.
Sending message fails.
If this message persists, call for service.
Code:00106"
"Please turn the system off, wait 1
minute, and then restart the system.
Setting serial port fails.
If this message persists, call for service.
Code:00107"
7-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
00108
00109
00501
00601
00602
00603
00604
00605
00606
7-6
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Connection setup fails.
If this message persists, call for service.
Code:00108"
"Please turn the system off, wait 1
minute, and then restart the system.
Writing serial port fails.
If this message persists, call for service.
Code:00109"
Please turn the system off, wait 1
minute, and then restart the system.
Loading library fails.
If this message persists, call for service.
Code:00501
Please turn the system off, wait 1
minute, and then restart the system.
Capturing single image fails.
If this message persists, call for service.
Code:00601
Please turn the system off, wait 1
minute, and then restart the system.
Choosing two-buffered visual fails.
If this message persists, call for service.
Code:00602
Please turn the system off, wait 1
minute, and then restart the system.
GLX context creation fails.
If this message persists, call for service.
Code:00603
"Please turn the system off, wait 1
minute, and then restart the system.
Window creation fails.
If this message persists, call for service.
Code:00604"
"Please turn the system off, wait 1
minute, and then restart the system.
Servant deactivation fails.
If this message persists, call for service.
Code:00605"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:00606"
Font resource error.
Troubleshooting
00607
00608
00609
00610
00611
00612
00613
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Grabber card buffer error.
If this message persists, call for service.
Code:00607"
"Please turn the system off, wait 1
minute, and then restart the system.
Image process parameter error.
If this message persists, call for service.
Code:00608"
"Please turn the system off, wait 1
minute, and then restart the system.
Grabber card initialization fails.
If this message persists, call for service.
Code:00609"
"Please turn the system off, wait 1
minute, and then restart the system.
Loading grabber card config file fails.
If this message persists, call for service.
Code:00610"
"Please turn the system off, wait 1
minute, and then restart the system.
Loading IP config file fails.
If this message persists, call for service.
Code:00611"
"Please turn the system off, wait 1
minute, and then restart the system.
Making GLX current fails.
If this message persists, call for service.
Code:00612"
"Please turn the system off, wait 1
minute, and then restart the system.
Timer creation fails.
If this message persists, call for service.
Code:00613"
"Please turn the system off, wait 1
minute, and then restart the system.
00614
00615
Error
Error
Grabber card is missing. Check the
connection of grabber card in the
computer. If the connection is OK,
If this message persists, call for service.
replace the grabber card.
Code:00614"
"Please turn the system off, wait 1
minute, and then restart the system.
Grabber card driver is missing.
If this message persists, call for service.
Code:00615"
7-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
00616
00901
00902
00903
01101
01102
01103
01104
01105
7-8
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Opening default display fails.
If this message persists, call for service.
Code:00616"
"Please turn the system off, wait 1
minute, and then restart the system.
CORBA communication error.
If this message persists, call for service.
Code:00901"
"Please turn the system off, wait 1
minute, and then restart the system.
Image data processing error.
If this message persists, call for service.
Code:00902"
"Please turn the system off, wait 1
minute, and then restart the system.
Main screen startup fails.
If this message persists, call for service.
Code:00903"
"Please turn the system off, wait 1
minute, and then restart the system.
Exam Management activation fails.
If this message persists, call for service.
Code:01101"
"Please turn the system off, wait 1
minute, and then restart the system.
WorkStation Setup customization fails.
If this message persists, call for service.
Code:01102"
"Please turn the system off, wait 1
minute, and then restart the system.
Loading domain fails.
If this message persists, call for service.
Code:01103"
"Please turn the system off, wait 1
minute, and then restart the system.
Locking fails.
If this message persists, call for service.
Code:01104"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01105"
Login error occurs.
Troubleshooting
01106
01107
01108
Error
Error
Error
01109
Error
01201
Error
01202
01203
01204
01301
01302
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
MCB version mismatches.
If this message persists, call for service.
Code:01106"
"Please turn the system off, wait 1
minute, and then restart the system.
UI status error occurs.
If this message persists, call for service.
Code:01107"
"Please turn the system off, wait 1
minute, and then restart the system.
Waiting firmware fails.
If this message persists, call for service.
Code:01108"
"Please turn the system off, wait 1
minute, and then restart the system.
Software unlocking fails.
If this message persists, call for service.
Code:01109"
" Need to restart the system after
deleting all information. Code:01201"
Need to restart the system after
deleting all information.
"Please turn the system off, wait 1
minute, and then restart the system.
Checking MCB compatibility fails.
If this message persists, call for service.
Code:01202"
"Please turn the system off, wait 1
minute, and then restart the system.
Configuration file resume fails.
If this message persists, call for service.
Code:01203"
"Please turn the system off, wait 1
minute, and then restart the system.
WKS Setup file IO error
If this message persists, call for service.
Code:01204"
"Please turn the system off, wait 1
minute, and then restart the system.
ADC sample time out.
If this message persists, call for service.
Code:01301"
"Please turn the system off, wait 1
minute, and then restart the system.
Collimator stuck.
If this message persists, call for service.
Code:01302"
7-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
01303
01304
01305
01306
01307
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01303"
"Please turn the system off, wait 1
minute, and then restart the system.
Error
Control panel initialization fails.
Change control panel or MCB.
If this message persists, call for service.
Code:01304"
"Please turn the system off, wait 1
minute, and then restart the system.
CPLD version incompatible.
If this message persists, call for service.
Code:01305"
"Please turn the system off, wait 1
minute, and then restart the system.
CPLD watch dog error.
If this message persists, call for service.
Code:01306"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01307"
"Please turn the system off, wait 1
minute, and then restart the system.
01308
Command buffer overflow.
DC power bus error. Check the DC bus
fuse. Check if the input power voltage
exceeds the range.
Filament over current.
Check signal FILAL_HC. If the value is
If this message persists, call for service. 0V, change MCB. If the value is +15V
approximately, change MP4. If the
Code:01308"
value is still +15V, change monoblock
or filament.
"Please turn the system off, wait 1
minute, and then restart the system.
01309
Error
Foot switch stuck. Verify that the
footswitch was pressed correctly and
check the connection of the
If this message persists, call for service.
footswitch. Otherwise, replace the
Code:01309"
footswitch or MCB.
"Please turn the system off, wait 1
minute, and then restart the system.
01310
01311
7-10
Error
Error
Hand switch stuck. Verify that the
hand switch was pressed correctly
and check the connection of the hand
If this message persists, call for service.
switch. Otherwise, replace the hand
Code:01310"
switch or MCB.
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01311"
IGBT error occurs. Check and replace
IGBT inverter module.
Troubleshooting
01312
01313
01314
01315
01316
01317
01318
01319
01320
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Inverter over current.
If this message persists, call for service.
Code:01312"
"Please turn the system off, wait 1
minute, and then restart the system.
kV value is higher than 120KV.
If this message persists, call for service.
Code:01313"
"Please turn the system off, wait 1
minute, and then restart the system.
kV value is lower than 35KV.
If this message persists, call for service.
Code:01314"
"Please turn the system off, wait 1
minute, and then restart the system.
Bridges of inverter unbalance. Change
monoblock.
If this message persists, call for service.
Code:01315"
"Please turn the system off, wait 1
minute, and then restart the system.
SPI2 with control panel error
If this message persists, call for service.
Code:01316"
"Please turn the system off, wait 1
minute, and then restart the system.
SPI9 error
If this message persists, call for service.
Code:01317"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01318"
"Please turn the system off, wait 1
minute, and then restart the system.
Monitor tube temperature by
hardware. The temperature is over
60(+/-) centigrade.
Communication UART with WKS
command buffer overflow.
If this message persists, call for service.
Code:01319"
"Please turn the system off, wait 1
minute, and then restart the system.
UART1 transmit error.
If this message persists, call for service.
Code:01320"
7-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
01321
01322
01323
01324
01325
01326
01327
01328
01329
7-12
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
Communication with WKS faults.
If this message persists, call for service.
Code:01321"
"Please turn the system off, wait 1
minute, and then restart the system.
WKS version incompatible.
If this message persists, call for service.
Code:01322"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication with WKS RCC time
out.
If this message persists, call for service.
Code:01323"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication with WKS received
buffer overflow.
If this message persists, call for service.
Code:01324"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication with WKS transmit
time out.
If this message persists, call for service.
Code:01325"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication with WKS transmit
buffer overflow.
If this message persists, call for service.
Code:01326"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication UART with WKS time
out.
If this message persists, call for service.
Code:01327"
"Please turn the system off, wait 1
minute, and then restart the system.
Communication UART with WKS
writing fails.
If this message persists, call for service.
Code:01328"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01329"
WKS power off.
Troubleshooting
01330
01331
01332
01333
01334
01401
01402
01403
01404
Error
Error
Error
Error
Error
Error
Error
Error
Error
"Please turn the system off, wait 1
minute, and then restart the system.
When there is no exposure, current
deviation value is over 0.05mA.
If this message persists, call for service.
Code:01330"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01331"
"Please turn the system off, wait 1
minute, and then restart the system.
If this message persists, call for service.
Code:01332"
"Please turn the system off, wait 1
minute, and then restart the system.
Monitor tube temperature by software.
The temperature is over 60 centigrade
at 6 times continuously.
No calibrated mA table for preheating
the filament is found in flash. Create a
calibrated mA table first.
Writing flash fails.
If this message persists, call for service.
Code:01333"
"Please turn the system off, wait 1
minute, and then restart the system.
Can not find WKS communication in 4
minutes after MCB boot up.
If this message persists, call for service.
Code:01334"
"Please turn the system off, wait 1
minute, and then restart the system.
Serial port closedown fails.
If this message persists, call for service.
Code:01401"
"Please turn the system off, wait 1
minute, and then restart the system.
Opening serial port fails.
If this message persists, call for service.
Code:01402"
"Please turn the system off, wait 1
minute, and then restart the system.
Reading serial port fails.
If this message persists, call for service.
Code:01403"
"Please turn the system off, wait 1
minute, and then restart the system.
Setting serial port fails.
If this message persists, call for service.
Code:01404"
7-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
"Please turn the system off, wait 1
minute, and then restart the system.
Writing serial port fails
01405
Error
01406
Error
AC power is lost. The system will shut
down in 30 seconds. Code: 01406
AC power lost. Check power supply.
10002
Warning
A Software Exception has occurred.
Code: 10002
Adjusting priority fails.
10003
Warning
Loading multi-language fails. English
will be used as default. Code: 10003
Loading multi-language fails.
10201
Warning
DICOM operation fails. Code: 10201
DICOM association abortion fails.
10202
Warning
DICOM operation fails. Code: 10202
DICOM association request fails.
10203
Warning
DICOM operation fails. Code: 10203
Association is rejected by DICOM
server.
10204
Warning
DICOM operation fails. Code: 10204
Connection to DICOM server fails.
10205
Warning
DICOM operation fails. Code: 10205
DICOM interchange file creation fails.
10206
Warning
DICOM operation fails. Code: 10206
DICOM file or temp file creation fails.
10207
Warning
DICOM operation fails. Code: 10207
Temp folder creation fails.
10208
Warning
DICOM operation fails. Code: 10208
A bad DICOM file is accessed.
10209
Warning
DICOM operation fails. Code: 10209
DICOM file does not exist.
10210
Warning
DICOM operation fails. Code: 10210
System association destruction fails.
10211
Warning
DICOM operation fails. Code: 10211
System disconnection from server
fails.
10212
Warning
DICOM verify fails. Code: 10212
DICOM verify fails.
10213
Warning
DICOM server configuration error.
Code: 10213
DICOM server Parameters setting
error.
10214
Warning
DICOM operation fails. Code: 10214
DICOM application initialization fails.
10215
Warning
DICOM print fails. Code: 10215
DICOM print fails.
10216
Warning
DICOM worklist query fails. Code:
10216
DICOM worklist query fails.
10217
Warning
DICOM operation fails. Code: 10217
DICOM association release fails.
10218
Warning
DICOM operation fails. Code: 10218
DICOM resource file is lost.
10219
Warning
DICOM operation fails. Code: 10219
DICOM application exception occurs.
10220
Warning
DICOM storage fails. Code: 10220
DICOM storage fails.
10221
Warning
DICOM operation fails. Code: 10221
DICOM Service is not supported.
10222
Warning
DICOM operation fails. Code: 10222
DICOM CD Viewer files are lost.
7-14
If this message persists, call for service.
Code:01405"
Troubleshooting
10401
A Software Exception has occurred.
Code: 10401
DICOM worklist query data parse error.
Displayed when a software exception
occurred. Click OK button to close the
message. You can continue to take
exposure.
A Software Exception has occurred.
Code: 10501
Closing file fails. Displayed when a
software exception occurred. Click OK
button to close the message. You can
continue to take exposure.
Warning
10501
Warning
10502
Warning
Deleting image fails. Code: 10502
Deleting image file fails.
10503
Warning
A software exception has occurred.
Code: 10503
Reading config file fails.
10504
Warning
Reading image fails. Code: 10504
Reading image file fails.
10505
Warning
A software exception has occurred.
Code: 10505
Writing config file fails.
10506
Warning
A software exception has occurred.
Code: 10506
Writing image file writing fails.
A Software Exception has occurred.
Code: 10601
Grabber capture thread's priority error
occurs. Displayed when a software
exception occurred. Click OK button to
close the message. You can continue
to take exposure.
A Software Exception has occurred.
Code: 10602
Grabber capture thread's scheduler
error occurs. Displayed when a
software exception occurred. Click OK
button to close the message. You can
continue to take exposure.
A Software Exception has occurred.
Code: 10603
OpenGL thread setting fails. Displayed
when a software exception occurred.
Click OK button to close the message.
You can continue to take exposure.
10601
10602
Warning
Warning
10603
Warning
10801
Warning
Accessing device fails. Code: 10801
Accessing device fails
10802
Warning
The target device is not found or has
been removed from the USB port.
Code: 10802
The destination directory is not found
or has been removed from the usb
port.
There is no space for the files. Code:
10803
The space is unsupported to store the
files. The current target disk doesn’t
have enough space to complete the
copy operation. Insert a new disk and
retry the operation.
The printer initialization fails. Code:
10804
The printer initialization fails. Check the
printer and refer to operating
instructions of the printer for more
10803
Warning
10804
Warning
7-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
information.
10901
10902
10903
10904
10905
Warning
Warning
A Software Exception has occurred.
Code: 10901
There is no such view. Displayed when
a software exception occurred. Click
OK button to close the message. You
can continue to take exposure.
A Software Exception has occurred.
Code: 10902
Resource files are damaged. Displayed
when a software exception occurred.
Click OK button to close the message.
You can continue to take exposure.
A Software Exception has occurred.
Code: 10903
There is no database primary key, can
not change DAP. Displayed when a
software exception occurred. Click OK
button to close the message. You can
continue to take exposure.
A Software Exception has occurred.
Code: 10904
There is no database primary key, can
not change dose. Displayed when a
software exception occurred. Click OK
button to close the message. You can
continue to take exposure.
A Software Exception has occurred.
Code: 10905
There is no database primary key, can
not change dose exposure Time.
Displayed when a software exception
occurred. Click OK button to close the
message. You can continue to take
exposure.
A Software Exception has occurred.
Code: 11201
Displayed when a software exception
occurred. Click OK button to close the
message. You can continue to take
exposure.
Warning
Warning
Warning
11201
Warning
11202
Warning
One format should be selected at least.
Code: 11202
Displayed when the format is not
selected during DICOM print.
11203
Warning
Maximum must be greater than
minimum. Code: 11203
Maximum must be greater than
minimum.
11204
Warning
Same name DICOM server exists. Code:
11204
Same name DICOM server exists.
11205
Warning
A software exception has occurred.
Code: 11205
Configuration file loading fails.
11206
Warning
All text fields except gateway must be
input. Code: 11206
All text fields except gateway must be
input.
11207
Warning
AE Title must be input. Code: 11207
AE Title is missing.
11208
Warning
The length of password is not correct,
4-10 is acceptable. Code: 11208
The length of password is not correct,
4-10 is acceptable.
7-16
Troubleshooting
11209
Warning
New password mismatches with the
retyped one. Code: 11209
New password mismatches with the
retyped.
11210
Warning
Old password is invalid. Code: 11210
Old password is invalid.
IP or gateway is invalid. Code: 11211
11211
Warning
IP or gateway is invalid. Obtain the
correct information from network
administrator.
Port Number is invalid. Code: 11212
Port number is invalid. Obtain the
correct information from network
administrator.
11212
Warning
11213
Warning
Network configuration fails. Code:
11213
Network configuration fails.
11214
Warning
There is invalid gateway setting in
DICOM printer servers. Code: 11214
There is invalid gateway setting in
DICOM printer servers.
11215
Warning
There is invalid gateway setting in
DICOM storage servers. Code: 11215
There is invalid gateway setting in
DICOM storage servers.
11216
Warning
There is invalid gateway setting in
DICOM worklist servers. Code: 11216
There is invalid gateway setting in
DICOM worklist servers.
DAP not work normally. Code: 11301
11301
Warning
DAP does not work normally. Click OK
button to close the message. You can
continue to take exposure.
Emergency switch is pressed. Code:
11302
Emergency switch is pressed.
Displayed during exposure. Release
the emergency switch and retry the
operation.
Since hand switch has been pressed,
exposure will be refused. Please
release hand switch first. Code: 11303
Hand switch is pressed during power
on. Release the switch and try again.
Key switch off. Code: 11304
Key switch is off. Displayed during
exposure. Rotate the key switch to on
position and retry the operation.
11302
Warning
11303
Warning
11304
Warning
11305
Warning
kVp is not accurate. Please call service
for calibration. Code: 11305
kVp is not accurate. Calibrate kVp
according to chapter 5.
11306
Warning
mA is not accurate. Please call service
for calibration. Code: 11306
mA is not accurate. Calibrate mA
according to chapter 5.
11307
Warning
Tube temperature warm. Code: 11307
Tube temperature warm.
11308
Warninng
Since foot switch has been pressed,
exposure will be refused. Please
release foot switch first. Code: 11308
Foot switch is pressed during power
on. Release the switch and try again.
11401
Warning
UPS is abnormal. Code: 11401
Check UPS and the connected cable.
7-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
11402
Warning
UPS's battery needs to be replaced.
Code: 11402
Replace UPS battery.
11403
Warning
Can not detect UPS output. There is a
potential risk of data loss. Code: 11403
Check UPS and connected cables.
11404
Warning
UPS connection fails. There is a
potential risk of data loss. Code: 11404
Check UPS and connected cables.
20001
Information
The device is read only. Code: 20001
The device is read only.
20002
Information
No disk is found in the disk drive. Code:
20002
No disk is found in the disk drive.
20003
Information
No device is found. Code: 20003
No device is found.
20201
Information
DICOM server has verified successfully.
Code: 20201
DICOM verify is successful.
20202
Information
DICOM operation fails. Code: 20202
DICOM application is busy.
20501
Information
Disk is full. Code: 20501
Disk is full.
20801
Information
Device is busy. Code: 20801
Device is in busy.
20802
Information
A Software Exception has occurred.
Code: 20802
Message queue creation in MM
module fails.
20803
Information
There is no paper in the printer. Code:
20803
There is no paper in the local thermal
printer.
20804
Information
The printer's door is open. Code: 20804
The local thermal printer door is open.
20805
Information
The printer is under analog mode, can
not print through USB port. Code:
20805
Change the printer setting as digital
mode.
20806
Information
The paper type does not match printer
setting. Code: 20806
Paper type does not match printer
setting. Replace the paper.
20807
Information
The printer paper scrolling fails. Code:
20807
Paper scrolls in wrong direction. Open
the paper box and load paper
properly.
20808
Information
The paper is stuck in the printer. Code:
20808
Open the paper box and take out the
paper.
20809
Information
The local printer is disconnected. Code:
20809
Please check the connection of the
printer.
The printer type is not supported.
Code: 20810
System driver can not support this
printer. Replace the printer. Available
types are: UP-990AD, UP-D897, and
UP-DF500.
The printer's thermal accessory is
abnormal. Code: 20811
The temperature of printer is too high.
Please wait a moment and then
continue.
20810
Information
20811
Information
7-18
Troubleshooting
20812
Information
The paper entrance is wrong. Code:
20812
Paper entrance error. Put the paper
into printer again.
7.4. Troubleshooting Guide
Power on System
Boot up Failed?
No
Yes
Boot up Failed
X_ray Exposure Failed?
No
Yes
X_ray Exposure Failed
Image display failed?
No
Yes
Image display failed
Power off failed?
Yes
Power Off failed
No
Normal System Off
7-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
7-20
Troubleshooting
7-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
7-22
Troubleshooting
7-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
7-24
Troubleshooting
7-25
Chapter8.
Replacement
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.1. FRU list
Item
Part number
Description
Quantity
1
5326480
PWA, COLLIMATOR CONTROL BOARD, EVR
1
2
5326484
PWA, MAINFRAME CONTROLLOR BOARD, EVR
1
3
5326475
PWA, FILAMENT CONTROL BOARD, EVR
1
4
5339933
ASM, IGBT INVERTER MODULE,EVR
1
5
5326487
PWA, kV CONTROL BOARD, EVR
1
6
5326482
PWA, MOTHER BOARD, EVR
1
7
5326476
PWA, UP_DOWN DRIVE BOARD, EVR
1
8
5326479
PWA, POWER DISTRIBUTION BOARD, EVR
1
9
5330056
Component, Image Frame Grabber, EVR
1
10
5325576
Generator, monoblock, EVR
1
11
5336403
ASM, COLLIMATOR, EVR
1
12
5328978
Workstation, Chassis Power, EVR
1
13
5328979
Workstation, Memory Bank, EVR
2
14
5328980
Workstation, mainboard assembly with CPU,EVR
1
15
5328981
Workstation, Display Card, EVR
1
16
5328982
Workstation, Hard Disk, EVR
1
17
5328983
Workstation, DVD-RW, EVR
1
18
5328985
Workstation, Mouse and Keyboard, EVR
1
19
5331655
Cable, Interconnect, EVR
1
20
5331597
MISC, MAIN TRANSFORMER, EVR
1
21
5331610
MISC, WORK STATION RELAY, EVR
1
22
5333060
Cable, 110V Power, EVR
1
8-2
Replacement
Item
Part number
Description
Quantity
23
5331651
Cable, 220V Power, EVR
1
24
5331599
MISC, DC POWER SUPPLY +24V AND +5V, EVR
1
25
5331598
MISC, DC POWER SUPPLY +15V, -15V AND +5V, EVR
1
26
5331606
MISC, BREAKER FOR 110V, EVR
1
27
5331605
MISC, BREAKER FOR 220V, EVR
1
28
5331596
MISC, START UP TRANSFORMER, EVR
1
29
5329283
MISC, DRIVE POWER TRANSFORMER, EVR
1
30
5336404
FRU, FUSE KIT, EVR
1
31
5329281
MISC, RESONANT CAPACITOR, EVR
1
32
5329275
MISC, RECTIFIER BRIDGE, EVR
1
33
5331600
MISC, C_ARM POWER FILTER, EVR
1
34
5331604
MISC, POWER ON INDICATOR, EVR
1
35
5331607
MISC, WORK STATION POWER FILTER
1
36
5330055
Component, LCD monitor, EVR
1
37
5330061
Component, I.I Power Supply, EVR
1
38
5330063
Assembly, CCD Camera with CCU, EVR
1
39
5330059
Assembly, I.I with power supply, EVR (Cert. Comp.)
1
40
5330066
Component, Grid, EVR
1
41
5333080
FRU, LINEAR ACTUATOR KIT, EVR
1
42
5333084
FRU, MAINFRAME STEERING HANDLE KIT, EVR
1
43
5333085
FRU, CROSS-ARM BRAKE KIT,EVR
1
44
5333086
FRU, WIG-WAG BRAKE HANDLE KIT, EVR
1
45
5333087
FRU, ORBITAL AND ROTATIONAL BRAKE HANDLES KIT, EVR
1
46
5343599
ASM, UNIVERSAL TRUCKLE WITHOUT BRAKE, EVR
1
8-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Item
Part number
Description
Quantity
47
5343601
ASM, UNIVERSAL TRUCKLE WITH BRAKE, EVR
1
48
5336263
ASM, ONE PEDAL FOOTSWITCH, EVR
1
49
5331609
MISC, HAND SWITCH ASSY, EVR
1
50
5330849
MISC, EMERGENCY SWITCH, EVR
1
51
5325779
ASM, FRONT WHEEL, EVR
1
52
5333094
PLASTIC, SKIN SPACER 30CM,EVR
1
53
2418430
Component, UP-D897 Printer, EVR
1
54
5333092
FRU, STEERING WHEEL PLATE KIT, EVR
1
55
5331638
MISC, EXPOSURE INDICATOR, EVR
1
56
5330067
Component, UP-990AD Thermal Printer, EVR
1
57
5333095
FRU, TUBE COVER KIT, EVR
1
58
5364175
ASM, LASERAIMER KIT IN TUBE, EVR
1
59
5329278
MISC, DC BUS CAPACITOR, EVR
1
60
5329279
MISC, DC BUS FUSE, EVR
1
61
5333096
FRU, C-ARM HANDLES KIT, EVR
1
62
5333088
FRU, WORKSTATION FRONT COVER KIT, EVR
1
63
5333099
FRU, WORKSTATION SIDE AND BACK COVERS KIT, EVR
1
64
5333100
FRU, MAINFRAME BODY COVERS ASSEMBLY, EVR
1
65
5333120
FRU, WORKSTATION TOP COVERS KIT, EVR
1
66
5333101
FRU, WORKSTATION HANDLE, EVR
1
67
5332898
Component, Video distributor for external DVI video output, EVR
1
68
5333061
Cable, Video, EVR
1
69
5326483
PWA, EXTENDED BOARD, EVR
1
70
5326478
PWA, MCB EXTENDED BOARD, EVR
1
8-4
Replacement
Item
Part number
Description
Quantity
71
5332880
ASM, CONTROL PANEL, EVR
1
72
5241505
DAP chamber, Dose_DAP measure, EverView7500
1
73
00-878105-02
9" BEAM ALIGHMENT TOOL ASSEMBLY
1
5330473-01
ASM, LASER AIMER, 9 INCH ENGLISH WITH ALL LANGUAGES
SUPPLEMENT
1
00-878223-02
9" Target for Laser Aimer (round clear plastic with lead crosshairs in
middle)
1
5344737
ASM, DOUBLE FOOT SWITCH, EVR
5325571
Plastic, 9 inch II cover, EVR
5366466
Misc, UPS, EVR
5370278
PWA, UPS control board, EVR
74
75
76
77
78
79
NOTE
1
1
1
1
For the 8 parts as below, field engineer must ship the failed parts to Surgery engineering in
China for autopsy in the first year after M3.
Contact:
Zhang Lihua
GE HUALUN Medical Systems Co. Ltd
No. 1 Yongchang North Road,
Beijing Economic & Technological Development Area
Beijing, P.R. China 100176
Tel: 8610-58068888
Item
Part Number
Part Description
1
5326484
PWA, MAINFRAME CONTROLLOR BOARD, EVR
2
5326487
PWA, KV CONTROL BOARD, EVR
3
5330056
Component, Image Frame Grabber, EVR
4
5325576
Generator, monoblock, EVR
5
5328981
Workstation, Display Card, EVR
6
5330055
Component, LCD monitor, EVR
8-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
7
5330063
Assembly, CCD Camera with CCU, EVR
8
5332880
ASM, CONTROL PANEL, EVR
After the replacement of the three parts below, please fill in the product locator card and
mail it to local service operation office.
Item
Part Number
Part Description
1
5325576
Generator, monoblock, EVR
2
5330059
Assembly, I.I with power supply, EVR (Cert.
Comp.)
3
5336403
ASM COLLIMATOR EVR
The FRU list contains five special tools to be used in calibration or troubleshooting.
Item
1
2
3
4
5
8-6
Part Number
Part Description
5333061
Cable, Video, EVR
5326483
PWA, EXTENDED BOARD, EVR
5326478
PWA, MCB EXTENDED BOARD, EVR
00-878105-02
9" BEAM ALIGHMENT TOOL ASSEMBLY
00-878223-02
9" Target for Laser Aimer (round clear plastic
with lead crosshairs in middle)
Replacement
8.2. Locations of FRU
8-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8-8
Replacement
36
65
18
17
66
63(Right side cover)
63(Left side cover)
62
19
47
46
55
78
63 (Rear cover)
8-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3. Replacement
WARNING
Electrical circuits inside the equipment use voltages that are capable of Causing
serious injury or death from electrical shock. Use appropriate precautions.
Please use specified fuse.
WARNING
If the system has power applied, power off the machine and unplug the AC power cord
from the wall outlet.
8.3.1. Covers
WARNING
Never allow anyone to use the system when any C-arm cover is removed. Use extreme
caution if you intend to troubleshoot the equipment with the covers off and power
applied. Failure to heed this warning may result death, personal injury, or equipment
damage.
C-Arm covers
You must remove the rear cover prior to the front cover.
You can remove the Cross-arm bearing cover, X-ray tube cover and Image Intensifier cover in any order.
NOTE
Set the brakes before attempting to remove C-Arm covers.
Follow these steps to remove the C-Arm covers:
1. Disconnect AC power plug from AC outlet.
2. Loosen four screws that secure rear cover to C-Arm.
3. Lift rear cover away from C-Arm.
4. Using the screwdriver through the hole to remove the four screws are located in the front cover.
Using great care to avoid electric shock.
5. Loosen two screws that secure front cover to T-base.
6. Slowly move front cover away from C-Arm.
NOTE
8-10
Install the covers in reverse sequence (Front cover prior to rear cover). The torque to
screw the C-Arm covers is 2.5Nm.
Replacement
Workstation covers
You must remove these workstation covers in the following sequence:
1. Rear Cover
2. Left and Right lower side covers
3. Front Cover
4. Left and right upper side covers
Follow these steps to remove the workstation covers:
1. Loosen eight screws of the rear cover. Remove the rear cover.
2. Loose the screws. And remove both side lower covers.
8-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3. Remove the 2 screws inside the front cover and 2 screws in the upside of the printer cabinet as
below. Remove the front cover.
4. Loose the screw and disassemble two workstation handles.
5. Remove the screws of each side and remove the left and right upper side covers.
Install the covers in reverse sequence.
NOTE
The torque requirement to fasten the screws that secure the rear and side covers is
2.5Nm.
The torque requirement to fasten the screws that secure the front cover is 5.8Nm.
8-12
Replacement
Upper Covers
NOTE
Rear cover must be removed before upper covers can be removed.
1. Loose the screws and remove the monitor support cover.
2. Remove the screws that secure the rear top cover. Remove the rear top cover.
8-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
3. Loose the screws on the sheet metal. Disconnect the cables connected to monitor. Pull out the
cables through the hole of the sheet metal. Then you can remove the sheet metal.
4. Remove all screws that secure the front top cover. And remove the front top cover.
Replace the new top cover kits and install them in reverse sequence. The torque of the screws that
secure the top covers is 2.5Nm.
8-14
Replacement
8.3.2. Brake, Handle and Wheel
Lateral Rotation Brake, Orbital Movement brake and Wig-Wag brake have the same replacement
procedure:
1. Make sure the brake is set.
2. Loosen the screw that secures the handle to C-arm.
3. Remove the handle.
4. Install the new handle and fasten the screw.
5. Release the brake. Move the C-arm and verify that you can lock C-arm by placing the brake in
lock position.
Horizontal Movement Brake
1. Make sure the horizontal movement brake is set.
2. Loosen 2 screws from horizontal brake handle and Remove the handle.
3. Remove the screws of the cross arm cover and remove the cover.
4. Loosen the screws of the plate cover of the brake and remove the cover.
5. Replace the sheet metal.
8-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6. Reassemble the covers of the cross arm.
7. Replace the new horizontal movement handle.
8. Place the handle in unlock position. Verify that you can push or pull the cross arm and lock the
cross arm by placing the brake handle in lock position.
C-Arm Steering Handle
1. Remove the 2 screws of the steering handle.
2. Remove the handle.
3. Install the new handle and fasten the screws.
4. Verify that the steering handle can turn the rear wheel.
Front wheel of the C-arm
Pry up the side of the C-arm carefully and keep the balance of the C-arm. Loose the screws and
remove the wheel according to the bomb drawing below:
Replace the new wheel. Move the C-arm and verify that the C-arm can move freely.
8-16
Replacement
Rear wheel of the C-arm
The FRU part for the rear wheel provides the 2 plates and cable pushers (P/N:5333092).
1. Pry up the side of the C-arm carefully and keep the balance of the C-arm. Remove the spring ring
on the center of the wheel.
2. Remove the plate of the wheel. Loose the screws and then remove the cable pusher.
3. Replace the new parts and reassemble the all the parts.
4. Move the C-arm and verify that the C-arm can move freely.
8-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Universal truckle of workstation
The workstation has two kinds of the truckle: the universal truckle with brake (P/N: 5343601) and the
universal without brake (P/N: 5343599). The FRU part contained the wheel and cable pusher.
1. Pry up the side of the C-arm carefully and keep the balance of the workstation.
2.
Loose the nut and remove the truckle. Then replace a new one.
Universal truckle with brake (P/N: 5343601)
Universal without brake (P/N: 5343599)
3. Move the workstation and verify that the workstation can move freely.
8-18
Replacement
8.3.3. Generator
1. Remove the tube cover.
2. Unplug the cables connected to collimator. Unscrew the four screws that secured the collimator
and remove it.
3. Disconnect the cables. Loosen the screws and replace the generator.
Torque to
10.1Nm during
installation
4. Reassemble the collimator. Start the system.
5. Perform calibration procedure in chapter5.
6. Shut down the system.
7. Cover the X-ray tube assembly. Fasten the screws with the torque to 0.56Nm.
If the X-ray tube has been produced more than 3 months, it should be warm first.
8-19
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.4. Collimator(with MP1 board)
1. Remove the tube cover. Loosen screws and replace the collimator.
2. Remove the rear cover of the console and the PCB box cover.
3. Take off the collimator control board (MP1 board) and insert a new one.
4. Power on the system; press the semi-transparent shutter close for more than 10s.
5. Check and verify the movement of the leaf is smooth.
6. Select fluoroscopy mode to take exposure. Verify the shutter’s rotation is at the center of the
blanking circle. If not, loosen the screws that secure the shutter and adjustment the shutter.
7. Fasten locating screws.
8. Extend the MP1 board with extend board (5326483). Select the radioscopy mode. Adjust the
resistor POT303 on the collimator board. Open the collimator leafs to maximum position. The
voltage measured on the two poles of the drive motor should be less than 3V.
9. Power off.
10. Connect Iris (see below Fig for reference), and install it on its position but don’t screw down.
11. Take off the bridge gear. Close the IRIS manually, and then power on the system.
8-20
Replacement
Take off the
Bridge Gear
12. Press the IRIS close Icon for more than 10s, and then adjust the resistor on the IRIS until the
motor doesn’t rotate. Keep the motor resting and install the bridge well.
13. Select the manual fluoroscope mode and take exposure. Verify the IRIS field is at the center of the
blanking circle. Screw down the fastening nuts.
14. Open the IRIS to maximum and take exposure. Adjust the POT302 on the collimator control board
to make the X-ray field in tangent with blanking circle.
15. Place the Beam Alignment Fixture Tool (9 inch) in front of the Image Intensifier, close the IRIS to
minimum, take an exposure to measure the minimum field size is less than 45mm, if not repeat
the step14 and step 16.
16. Select MAG1 field mode, take exposure and adjust the POT2 to make the X-ray field in tangent
with blanking field.
17. Select MAG2 field mode, take exposure and adjust the POT1 to make the X-ray field in tangent
with blanking field.
18. Select the radioscopy mode, adjust the resistor POT304 on the collimator board let the IRIS open
to maximum position but the voltage measured on the two poles of the drive motor less than 3V.
19. Perform I.I. field size check and calibration (5.6.5), limiting field size check (5.6.6), ABS tracking
procedure (5.6.7), resolution check (5.6.8), DAP and dose calibration (3.7.5).
20. Power off and assemble all the covers.
8-21
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.5. PCBs
Collimator control board (MP1 board)
Refer to chapter 8.3.4 for collimator control board replacement and calibration.
Mainframe Control Board (MCB board)
1. Remove the rear cover of C-Arm console.
2. Unscrew the screws of PCB box cover and remove it.
3. Remove the MCB board and insert a new one.
4. Perform kV (section 5.2), mA (section 5.3) and mAs (section 5.4) check and calibration procedure in
chapter 5.
5. Perform ABS Tracking check and adjustment (section 5.6.7) procedure in chapter 5.
6. Perform DAP and dose calibration in section 3.7.5.
Filament Control Board (MP4 board)
1. Remove the rear cover of C-Arm console.
2. Unscrew the screws of PCB box cover and remove it.
8-22
Replacement
3. Extend the new MP4 board with extend board (5326483). Perform mA check and calibration
(section 5.3), mAs check and calibration (section 5.4) procedure.
4. Remove the extend board. Insert the new MP4 board and lock in place.
5. Perform ABS Tracking check and adjustment (section 5.6.9) procedure in chapter 5.
6. Perform DAP and dose calibration in section 3.7.5.
7. Install the rear cover of C-arm console.
Mother board
1. Remove the rear cover of C-Arm console.
2. Unscrew the screws of PCB box cover and remove it. Remove the MP1, MCB and MP4 board.
3. Unscrew the screws and remove the PCB box.
4. Unplug all cables connected to the motherboard and then remove the metal cover.
5. Loose the screws and replace a new mother board.
6. Reassembly the metal cover. Connect the cables according to the label of the cable and the mark
on metal cover.
7. Install the PCB box and insert the MP1, MCB and MP4 board. Install the PCB box cover and C-arm
console cover.
8. Power on the system and take exposure. Verify that the system works normally.
8-23
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
kV control board
1. Remove the rear cover of C-Arm console.
2. Unplug all cables connected to kV control board (KJ1, KJ2, KJ3, KJ4, KJ5, KJ6, KJ7).
3. Unscrew the four screws that secured the board. Replace the new kV control board. Connect all
cables.
7. Perform kV (section 5.2), mA (section 5.3) and mAs (section 5.4) check and calibration procedure in
chapter 5.
8. Perform ABS Tracking check and adjustment (section 5.6.7) procedure in chapter 5.
9. Perform DAP and dose calibration in section 3.7.5.
IGBT inverter module
1. Remove the rear cover of the C-arm console.
2. Loose the four screws and remove the shield cover.
3. Disconnect all cables connected to IGBT board and the cables to the fan.
8-24
Replacement
Remain the screws on the board as below after disconnect the cable. Be careful not to break the
pins.
4. Loose the screws that secured the IGBT module and replace a new one.
5. Connect all cables to IGBT module according to the labels on the cable and the mark on board.
(Note: C2 to M1,C1 to P1,C4 to M2,C3 to P2,IGBT+ to P2, IGBT- to N2, yellow line to T1,
black line to T2)
6. Reassembly the shield cover and the rear cover of console.
7. Power on the system. Take exposure and verify the system works normally.
Power Distribution Board (PDU)
1. Power off the system and disconnect C-arm’s AC plug from the AC receptacle.
2. Remove the covers of console.
3. Disconnect the cabling (PJ1, PJ3, PJ4, PJ5, PJ6, P1) from the PCB board.
4. Remove screws of PDU board.
Note: The nylon pillar
washer may fall off.
5. Remove the PDU board.
8-25
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
6. Put the insulation washer and new PDU board on position and tighten the screws.
7. Plug the wires (PJ1, PJ3, PJ4, PJ5, PJ6, P1) in the corresponding socket.
8. The default power supply is AC 220V power input. If your power supply is 110V, adjust the
jumpers (PJ9 and PJ10) according to wiring diagram. Refer to chapter 3.3 for details.
9. Reinstall the covers of console and restart system to confirm system can power on and take
exposure.Check all functions can work normally.
Up-down Drive Board
1. Power off the system and disconnect C-arm’s AC plug from the AC receptacle.
2. Remove the covers of console.
3. Disconnect the cabling (J1, J2, J3) of up down drive board.
8-26
Replacement
4. Remove four screws that secure Up-down drive board to chassis. Notice that there are four pillar
washers behind the board.
5. Remove old up-down drive board.
6. Install replacement up down drive PCB. Plug the cables to corresponding connectors (J1, J2, J3).
7. Reinstall the covers. Restart system and press lift and down vertical column switches to verify.
UPS control board
1. Remove the rear cover of the workstation.
2. Disconnect all cables connected to UPS control board.
3. Loose the screws that secured the board. Remove the board and replace a new one.
4. Functional check: Power on the system. After the booting up process complete, unplug the AC
power cord. Verify that the UPS can supply power to workstation until the system can shut down
completely.
Control panel (including membrane, display board and support board)
1. Remove the rear cover of the C-arm. Unscrew the three screws and open the package box.
2. Disconnect all cables connected to control panel.
3. Loose the screws that secure the control panel and remove the metal shield strips. Then remove
the control panel.
4. Replace a new control panel and connect the cables. Install the C-arm cover.
5. Functional check: Power on the system. Pressing the buttons and verify that relative indicator can
light.
8-27
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.6. Parts in Computer and Peripheral
Perform the procedures below to replace the parts of computer.
1. Remove the rear cover of the workstation. Unplug the cables from the computer.
2. Loosen the screws and disassemble the network connector from the computer.
3. Loosen the two screws of the computer cover. Lift up and remove the cover.
4. Remove the baffle in the computer.
5. You can replace the FRU parts (such as the hard disk, display card, Memory Bank, mother board,
image grabber card and Chassis Power) in the computer.
8-28
Replacement
Chassis Power
Image Grabber card
Mother board
Display Card
Memory Bank
Hard disk
NOTE
Cut off the binding tie before removing the memory bank. And replace a new one to
secure the memory bank after the replacement.
Grabber card shall not be installed in the slot beside Display card.
There are two hard disks in the computer. One is data disk, and the other is system disk. During
replacement of the hard disk, please identify the disk by label on the disk. The labels on one end of the
SATA data cables, which marked “SYSTEM” or “DATA”, are used to connect the cable to the corresponding
disk. The other ends of the SATA data cables are connected to SATA ports on the mother board. Connect
system SATA cable to SATA1, data SATA cable to SATA2 and DVD ROM SATA cable to SATA4. After
replacement of either hard disk, you should reinstall both OS and Application software. Refer to
section6.3.1.
SATA
system
SATA
data
SATA ports
8-29
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Functional check: Power on the system. After login the Patient information screen, enter the patient
information and verify that you can enter normally. Create a new exam and take exposure. Verify the
system can work normally and the image will display on the left monitor.
If you have replaced the SYSTEM hard disk and backup the user data, you should also verify that the
original user data and images are remain in the system.
Peripheral
According to the drawing attached to the computer to install the mouse (connect to port2), keyboard
(connect to port3) and DVD RW (connect to 16 and 17).
After replacing the mouse, power on the system and move the mouse. Verify that the mouse can
move freely.
After replacing the keyboard, power on the system. Use the keyboard to enter words to Patient
information screen to verify the keyboard.
After replacing the DVD RW, power on the system. Perform CD/DVD check in section 3.6.2.
8-30
Replacement
8.3.7. Exposure Indicator
Perform the procedures below to replace exposure indicator:
1. Unscrew the 2 screws on the covering strip. Remove the strip.
2. Unplug the cable connected to the indicator. And then remove the four screws secure the
support sheet metal of the indicator.
3. Loose the three screws and then you can remove the exposure indicator and replace a new one.
4. Power on the system and take exposure. Verify that the exposure indicator lights during
exposure.
8-31
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.8. Monitors
Perform the procedures below to replace the monitors.
1. Loosen the screw and remove the monitor receptacle access cover. Pull out the two cables.
Covering strip
2. Remove the covering strip of the exposure indicator wiring.
3. Loosen the screws of the block that secure the cables. And then loosen the screws of the sheet
metal and remove it.
Remove the sheet metal
4. Remove the sheet metal of the exposure indicator support.
5. Unscrew the four screws on the monitor and remove the monitor.
6. Install the new monitor and reset all parts and cables.
7. Power on the system. The left monitor of workstation will display GE logo screen. The right
monitor will enter into the PATIENT INFORMATION Screen automatically. Perform the brightness
and contrast adjustment in section5.8 if necessary.
8-32
Replacement
The video cable (P/N: 5333061) is used for troubleshooting. Disconnect the DVI cable from the left
monitor. Connect the CCU V4 and left monitor VGA interface with the video cable. Verify if there is
image displayed on left monitor.
8.3.9. CCD Camera with CCU
Dangerous voltage levels are present when covers are removed from the C-arm.
WARNING
CAUTION
To prevent ESD damage to both the old and the new CCD Camera and CCU, make sure
that you are grounded with a properly functioning write strap before handling any
ESD sensitive electronic components.
NOTE
CCD and CCU is one set (P/N:5330063). If one component fails, you have to replace both
CCU and CCD together.
Step 1. Replace CCD Camera
1. Power off the system.
2. Position the C-arm for easy access.
3. Remove the I.I. back cover.
4. Disassemble Camera cable plug-in port.
5. Disconnect the Camera Head Cable from the Camera Head by unscrewing 2 Camera Cable
securing screw.
6. Remove the EMI Shield by unscrewing 6 EMI Shield Securing screw.
7. Loosen 4 Camera alignment screws.
EMI Shield
Securing Screw
Camera
Securing Screw
Camera Head
Cable
Camera cable
plug-in port
disassembled
EMI Shield
8-33
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Camera
Alignment
Screws (there
are 4 screws
with 90 degrees
apart)
8. Unwrap and prepare new camera to be assembled immediately in the next step.
9. Remove 3 Camera Securing Screws, and remove the old Camera from the I.I.
NOTE
Be careful not to touch the camera lens or I.I. output lens.
10. Assemble the new Camera on the I.I, and tighten 3 Camera Securing screws.
11. Tighten 4 camera adjustment screws with modest torque.
12. Connect the Camera Head Cable.
Step2. Replace CCU
1. Power off the system. Plug out the power cable.
2. Remove the Front and Rear Covers of the Mainframe.
3. Disconnect all the cables on the CCU (V3-FrameSync, CHB-Video, DB9, DB25, 24V Power).
4. Unscrew the 3 CCU Securing Screws.
3 CCU
Securing
Screws
5. Slide the CCU to the left, and pull out the CCU to remove from the Mainframe.
6. Remove the CCU Securing Bracket from the old CCU, and assemble it to the new CCU.
8-34
Replacement
7. Open the new CCU cover.
CCU Securing
Bracket
CCU Cover
opened
8. Check the jumper settings on the CCU.
N30 PWR20
NDR ADV0
1
1
2
2
3 J53 J6
J8
1 2
J7
1 2
R61
R52
R47
R48
J4
R10
1
R10
0
Parameter
CODE
SHADING ENABLE
J4
AUTO WTH
J5
AUTO BLK
J6
KV CONTROL
MODALITY
J7-J8
R84
R79 R10
6
1
2
3
DESCRIPTION
It is used to enable shading regulations
(J4 ON: position 1-2, J4 OFF: position 2-3)
It is used to enable the Auto White Level.
(J5 ON: position 1-2, J5 OFF: position 2-3)
It is used to enable the Auto black Level.
(J6 ON: position 1-2, J6 OFF: position 2-3)
Setup Criteria
J4 ON: position 1-2
J5 OFF: position 2-3
J6 OFF: position 2-3
Controls the KV control modality.
J7: position 2-3
(J7-J8 UP/DOWN mode: position 1-2, J7-J8 ANALOG mode:
position 2-3)
J8: position 2-3
9. Assemble the new CCU (with CCU cover open) back to the mainframe.
10. Connect the cables back to CCU (V3-FrameSync, CHB-Video, DB9, DB25, 24V Power).
8-35
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
DB25
24V Power
CHA
DB9 CameraHead
V3 FrameSync
CHB
V4
11. Plug in the system power cable, and power on the system.
Step 3. Calibrate the CCU
A. Black level setting
Parameter
CODE
DESCRIPTION
Setup Criteria
BLACK LEVEL
R101
Ad just black level pedestal (BOTTOM BLACK). Regulate the
pedestal to 30mV respect referring video level
40~60mV from CHA
a. Disconnect DB 25 connector from the CCU.
b. Measure Video Output voltage in Channel A (CHA) with Oscilloscope. This is DC signal. Use vertical
precision 40msec/div.
c. Adjust the R101 potentiometer in the CCU, so that the Video Output voltage becomes 40~60mV
from CHA.
B. AutoKV Level and Window
Parameter
CODE
DESCRIPTION
Setup Criteria
AUTO KV
WINDOW
R106
Increase or decrease the sensibility of the automatic KV
control. This feature is very useful in according to adapt the
CCD camera N33 for any kind of X-ray generator
Anti-clockwise maximum limit (all
the way to the left)
AUTO KV LEVEL
R79
Set the value of AUTO KV
Adjust to get –80~0mV when
there is no x-ray
1. Disconnect DB 25 connector from the CCU.
2. Measure the kV+ signal at Pin1 of DB25.
8-36
Replacement
+KV
-KV
ND FILTER
AUTO OFF
NOT USED
NOT USED
NOT USED
ROI
INVH
FREEZE-A
LIH
X16 FILTER
X8 FILTER
1
1
2
3
1
1
4
5
6
7
8
9
10
11
12
13
GND
ROT
ROT
1
HALF GAIN
1
NOT USED
1
2
2
2
2
2
2
NOT USED
NOT USED
DRF
INVV
FREEZE-B
X4 FILTER
X0 FILTER
3. Adjust R106 (AutoKV Window) potentiometer all the way to the left (anti-clock-wise).
4. Adjust R79 (AutoKV Level) potentiometer, so that the KV+ signal (average) becomes –80~0 mV.
NOTE
From here, follow the whole procedures in Section 5.6. Beam Alignment Procedure.
8-37
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.10. Image Intensifier with power supply
Perform the procedures below to replace the image intensifier with power supply (P/N: 5330059).
1. Position the C-arm for easy access. Remove the I.I. back cover. Remove CCD camera according to
the procedure in section 8.3.9.
2. Loose the four screws that secured the image intensifier and remove the image intensifier.
3. Replace a new I.I. with power supply. Reassemble the I.I and camera.
4. Perform section 5.6 Beam alignment procedure.
NOTE
There’s another FRU for I.I. power supply (P/N: 5330061). During the replacement of the
power supply, replace the new power supply after step2.
1. Disconnect all cables connected to the power supply. Loose the screws on I.I. back
cover.
2. Replace the new power supply. Connect the cables and reassembly the I.I. cover.
3. Perform section 5.6 beam alignment procedure.
8-38
Replacement
8.3.11. UPS
1. Remove the rear and front covers of the workstation.
2. Loose the screws that secured UPS assembly.
3. Disconnect the power cable connected to UPS. Use the handle to remove the UPS.
4. Replace a new UPS and reassemble the parts. Connect the cable.
5. Plug the UPS battery connector.
6. Connect the interconnect cable to C-arm. Plug the power cord into appropriate AC receptacle.
Voltage
increase
Power on
indicator
Voltage
Decrease
Test
Overload
On Battery Indicator
Replace
battery
Off
7. Press and hold UPS “test” button for several seconds to initiate a self-test. The workstation will
start up and enter BIOS. There will be a message to inform that there’s no AC power and need to
shut down.
8. Press “off” button to turn off the UPS.
9. Power on the system. Verify that the system can power on normally. Power off the system. Verify
that the system can shut down with no error.
10. Install all the covers of the workstation.
8-39
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.12. Misc Parts
Rectifier Bridge
1. Remove the rear cover of the C-arm. Remove the kV control board according to 8.3.5.
2. Disconnect the cables connected to the rectifier bridge.
3. Loose the screw and replace a new part.
4. Plug the cables to the relative jacks.
AC
_
+
AC
5. Reassemble all the parts.
6. Power on the system and take exposure. Verify that the system can work normally.
Drive Power transformer
1. Remove the rear cover of the C-arm. Remove the kV control board according to 8.3.5.
2. Cut off the binding tie. Disconnect the wires of the transformer, which are connected to J2.
3. Loose the screws and replace a new transformer. Connect the wires and bind the wires with the
binding tie to make a good layout.
4. Reassemble all the parts.
5. Power on the system and take exposure. Verify that the system can work normally.
8-40
Replacement
Main Transformer
1. Remove the rear cover of the C-arm. Unscrew the three screws and open the package box.
2. Cut off the binding tie. Disconnect the wires of the transformer, which are connected to J1 and
J2.
3. Loose the secure screws and remove the transformer.
4. Replace a new one. Connect the wires and bind them with binding tie to make a good wire
layout.
5. Power on the system and take exposure. Verify that the workstation and C-arm complete
software boot and initialization with no errors reported on their displays. Select the film mode,
take exposure with 80mAs, 110kVp. Verify that the system can work normally.
Start Up transformer
1. Remove the rear cover of the C-arm. Unscrew the three screws and open the package box.
2. Cut off the binding tie. Disconnect the wires of the transformer, which are connected to PDU
board PJ4 and ground.
3. Loose the secure screws and remove the transformer.
4. Replace a new one. Connect the wires and make a good wire layout.
5. Power on the system. Verify that the workstation and C-arm complete software boot and
initialization with no errors reported on their displays.
8-41
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
DC Power Supply
1. Remove the rear cover of the C-arm. Unscrew the three screws and open the package box. The
parts are on the back of the package.
DC power supply
+15V, -15V and
+5V (P/N:
5331598)
DC power
supply +24V
and +5V (P/N:
5331599)
2. Disconnect all cables connected to the power supply. Loose the screws and replace a new part.
3. Connect the cables according to the labels. Reassemble all parts.
4. Power on the system. Verify that the workstation and C-arm complete software boot and
initialization with no errors reported on their displays.
DC bus capacitor
1. Remove the rear cover of the C-arm. Unscrew the three screws and open the package box. The
part is on the back of the package.
There are two capacitors here but only one in FRU kit. Please identify the trouble for the part to
determine the quantity of FRU parts.
DC bus
capacitor
CAUTION
Do not touch the capacitors unless power has been completely removed and they have
been discharged.
2. Disconnect the cables. Unscrew the secured screw and remove the capacitor.
3. Replace a new one. Reassemble all the parts.
4. Power on the system and take exposure. Verify that the system can work normally.
8-42
Replacement
Resonant Capacitor and DC bus fuse
1. Remove the rear cover of the C-arm. Unscrew the three screws and open the package box.
2. Disconnect the cables connected to the resonant capacitor (P/N: 5329281) or DC bus fuse (P/N:
5329279). Loose the screws and replace a new one.
Resonant
capacitor
DC bus fuse
3. Power on the system and take exposure. Verify that the system can work normally.
Breaker
Brivo OEC850 provides 110V breaker (P/N: 5331606) and 220V breaker (P/N: 5331605).
1. Remove the rear cover and front cover of the C-arm. Unplug the cable connected to the circuit
breaker.
2. Loose the screws and remove the breaker. Then replace a new breaker.
3. Connect the cables to the breaker and reassemble the parts.
4. Power on the system. Verify that the workstation and C-arm complete software boot and
initialization with no errors reported on their displays.
8-43
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
C-arm Power filter
There are two power filters in the C-arm console.
One is behind the kV control board. You should remove the kV control board first.
The other is on the base near the breaker. You should remove the rear and front covers of the C-arm
console first.
Connect the cable according to the schematic on the filter:
After the replacement, power on the system and take exposure. Verify that the system can work
normally.
8-44
Replacement
Power on Indicator
1. Remove the rear and front covers of the C-arm console.
2. Unplug the wires connected to the power on indicator.
3. Remove the indicator and replace a new one.
4. Connect the AC cord to the socket. Verify that the power on indicator will light.
Workstation Power filter
1. Remove the workstation rear cover. Remove the UPS control board first.
Workstation
power filter
2. Unplug the cable connected to the filter.
3. Loose the screws and remove the filter. Replace a new part and connect the cable according to
the schematic on the filter.
4. Reassemble the parts. Power on the system. Verify that the workstation and C-arm complete
software boot and initialization with no errors reported on their displays.
8-45
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Workstation Replay
Loose the screws and remove the mounting. Loose the screws and unplug the wires connected to the
relay mounting. Replace a new one.
Power on the system. Verify that the workstation and C-arm complete software boot and initialization
with no errors reported on their displays. Shut down the system and verify the system can shut down
normally.
Interconnect cable
Unscrew the sleeve of the interconnect cable. Cut off the binding tie. Unplug the wires of the cable from
the socket. And then draw the cable out.
Replace a new cable. Insert the wires and make a good wire layout.
Power on the system. Verify that the workstation and C-arm complete software boot and initialization
with no errors reported on their displays. Shut down the system and verify the system can shut down
normally.
8-46
Replacement
Grid
Loose the screws and remove the grid. Replace a new one. Notice the circle mark on the grid for the
direction.
Fuse kit
The kit contains 10 kinds of fuse. You can choose the fuse to replace according to the requirement.
Ref.
FUSE5, FUSE6
Part
number
Description
5331602
Can be used
5330588
ASM, Install plate
assebmly, EVR
5330588
ASM, Install plate
assebmly, EVR
Misc, fuse model 6.3A, EVR
FUSE3, FUSE4
F5,F9
F11
F3,F12
F6,F7,F10
F8,F13,F14
F1,F2,F4
5331601
MISC, FUSE MODEL 1A, EVR
5326482
PWA, mother baord, EVR
5347003
FUSE,Circuit Protection,fast-acting
Blow, Long Time Lag,
5X20mm,250V 0.5A
5326482
PWA, mother baord, EVR
5347002
FUSE,Circuit Protection,fast-acting
Blow, Long Time Lag,
5X20mm,250V 1A
5326482
PWA, mother baord, EVR
5346999
FUSE,Circuit Protection,Fastacting Blow, Long Time Lag,
5X20mm,125V 5A
5326482
PWA, mother baord, EVR
5347001
FUSE,Circuit Protection,fast-acting
Blow, Long Time Lag,
5X20mm,125V 4A
5326482
PWA, mother baord, EVR
5347000
FUSE,Circuit Protection,fast-acting
Blow, Long Time Lag,
5X20mm,250V 2A
5326482
PWA, mother baord, EVR
5346998
FUSE,Circuit Protection,Slow
Blow, Long Time Lag,
5X20mm,250V 10A
5326479
PWA, power distribution
board, EVR
5326476
PWA, up-down drive
board
FL_L,FL_N,FL_L1,FL_N1
5337466
FUSE1,FUSE2
5337461
5mm x 20mm Fuses 250V .25A
Time Lag _Slo-Blo
5mm x 20mm Fuses 250V 3.15A
Time Lag_Slo-Blo
8-47
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8.3.13. Linear Actuator
1. Remove the C-arm console covers. Rotate the C-arm to the horizontal position.
2. Loosen 2 screws from horizontal brake handle and remove the handle.
3. Remove the screws of the cross arm cover and remove the cover.
4. Loose the screws and remove the rear plate. Remove the pin in the shaft.
8-48
Replacement
5. Draw out the C-arm slowly. Use some things to block up the C-arm. Loose the screws that secure
the cable and pull the cable long enough for the toppled C-arm console. And then topple the Carm console.
6. Remove the cover under the linear actuator. And then remove the pin.
8-49
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
7. Unscrew the nut on the top end of the axis. Open the package box and remove the side cover.
Screw the M4 screw to the hole and then draw the pin out.
8. Unplug the J3 from the up down drive board. Cut off the 2 wires. Cut off the binding tie and then
draw out the cable from the hole.
9. Draw out the linear actuator from the bottom.
8-50
Replacement
10. Replace a new linear actuator and reassemble the parts. Plug the wire into the connector. And
then insert the connector to J3 of the up down drive board.
11. Insert the pin with M4 screw into the hole and check if the pin is in the axis of the linear actuator.
Unscrew the M4 screw and tighten the nut on the hole. Lock the pin on the bottom of the linear
actuator and put the cover back.
12. Lift the C-arm and reassemble all the parts.
13. Connect the interconnect cable and power on the system. Press the up and down key on control
panel and verify that the C-arm can rise and drop freely.
8.3.14. Switch
Hand switch
1. Remove the rear and front cover of the C-arm console. Disconnect the cable from the hand
switch.
2. Loose the screw that secured the hand switch cable.
3. Remove the hand switch and replace a new one.
4. Power on the system. Press the hand switch to take exposure. Verify that the image will display
on the left monitor.
8-51
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
One pedal Footswitch
Insert the footswitch to the connector. Power on the system. Press the footswitch to take exposure.
Verify that the image will display on the left monitor.
Emergency Switch
1. Remove the rear and front cover of the C-arm console. Disconnect the cable connected to
mounting of the emergency switch.
2. Loose the screws and remove the base and switch. Then replace a new one.
3. Power on the system. Press the emergency switch, vertical column operation will cease and Xrays will be disabled. Rotate as arrow indication to release the emergency switch. You can
continue to use the system.
8.3.15. Others
Please refer to chapter3 for the replacement of these FRU parts.
Part Number
Part Description
Section Number
5333060
Cable, 110V Power, EVR
3.5.1
5331651
Cable, 220V Power, EVR
3.5.1
5333094
PLASTIC, SKIN SPACER 30CM,EVR
3.7.6
2418430
Component, UP-D897 Printer, EVR
3.7.3
5330067
Component, UP-990AD Thermal Printer, EVR
3.7.3
5364175
ASM, LASERAIMER KIT IN TUBE, EVR
3.7.1
5332898
Component, Video distributor for external
DVI video output, EVR
3.7.2
8-52
Replacement
5330473-01
ASM, LASER AIMER, 9 INCH ENGLISH WITH
ALL LANGUAGES SUPPLEMENT
3.7.1
5344737
ASM, DOUBLE FOOT SWITCH,EVR
3.7.8
5241505
DAP chamber and adjustment tool
3.7.5
8-53
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
8-54
Chapter9.
Periodic Maintenance
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9.1.
Overview
This procedure describes how to conduct performance checks and maintenance procedures. Use the
procedure in conjunction with PM checklist in section 9.2. Follow the procedures in the order they appear
here and record all required information on the checklist.
Serious injury and property damage can result from incorrectly performed service procedures. Observe
all operating and safety procedures in this document.
WARNING
9.1.1.
Procedures should be performed by service personnel specifically trained by GE
HUALUN Medical Systems, Inc. to calibrate the Brivo OEC 850 C-Arm and Workstation.
Safety
Ingress of Fluids
WARNING
The system must never be operated or stored in locations where conductive fluids, like
water and saline solution, might spill on the equipment.
Always unplug the AC power cable from the wall outlet before cleaning the equipment. Do not allow
water, soap, or other liquids to drip into the equipment and cause short circuits, electric shock and fire
hazards.
Unauthorized Modifications
Under no circumstances should the safety interlock in the system be bypassed, jumped, or otherwise
disabled.
NOTE
All GE Healthcare systems comply with International Electro Technical Commission safety
standard IEC-60601. Do not connect any external device to the system that does not
meet the requirements of IEC-60601. Only devices provided or approved by GE HUALUN
Medical Systems, Inc. should be connected to the system.
Motorized Mechanical Operation
The Vertical Column is motorized and should be observed continuously during operation to avoid
pinching or colliding with people and object.
If covers are removed, use extreme care when operating motorized features. Do not carelessly place
objects on the equipment or bump or lean against the equipment. Do not wear loose clothing that may
be caught by mechanical mechanisms. Keep hands and fingers clear of motorized equipment while the
motors are in operation.
Electrical Shock
WARNING
9-2
This system can generate lethal voltages. Practice safe electrical testing
procedures when performing periodic maintenance procedures.
Periodic Maintenance
After taking X-rays, the high voltage cables can retain a lethal charge, even if you
power down the system. Even though the contacts are recessed, high voltages can
arc several inches from their terminals if the cables are removed from the
connector wells. Use caution when disconnecting high voltage cables even when
the power is turned off. If you remove these connections, immediately discharge
their contacts against the high voltage tank case (ground).
WARNING
X-radiation Exposure
The X-ray tube assembly produces X-radiation when energized. Never operate this device without X-ray
shielding in place. Use lead shielding and draping to protect personnel.
Lead aprons, radiation monitors, and appropriate radiation shields must be utilized
for the protection of all personnel in the vicinity while performing tests. Living
human anatomy should never be used as a phantom or demonstration aid. Do not
produce radiation in hallways.
WARNING
Many of the PCB's in this system contain components, which are sensitive to ElectroStatic Discharge (ESD). Observe ESD safety procedures.
CAUTION
9.1.2.
Tools and Test Equipment
In addition to standard service tools and test equipment, the following special tools and test equipments
are required to complete the PM procedure.
Standard Tool
Item
Tool
1
Oscilloscope Tektronix 2236 or equivalent
2
Digital Volt-Ohm-Meter
3
Dosimeter
4
Spatial Resolution Tool
5
10’’×12’’ Film
6
kVp meter
7
Lead Apron
Special Tool
Item
Tool
GE part number
1
PWA, MCB extend board, EVR
5326478
2
UNIVERSAL BEAM ALIGHMENT TOOL
ASSEMBLY
00-878105-02
3
9’ target for Laser Aimer
00-878223-02
4
Three copper sheets, each 1mm thick
00-877682-01
5
PWA, extend board, EVR
5326483
9-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9.2.
PM Checklist
Brivo OEC 850 PM Checklist
Administrative Information
Medical Facility Information
Hospital or Clinic Name:
Address:
State:
Postal Code/Zip:
Contact Phone:
Field Service Engineer Information
FSE name:
Employee SSO:
SO Number:
Installation Date:
Region ID number:
City:
Country:
Contact Person:
Department:
System Configuration
Workstation Software Version:
Serial Number:
MCB Software Version:
Service ID:
Safety Inspection
Antistatic Drag-wires
Electrical Plug and power cord
Footswitch cable and pins
Ground Continuity
Interconnect Cable and Connector
PC Clean
Yes
No
System Function Checks
Power On Checks
Pass
Fail
C-Arm Checks
C-Arm Controls:
Pass
Fail
Key Switch:
Pass
Fail
Mechanical Movement:
Pass
Auto Technique Tracking
Copper Filters
Specification
(kVp)
Measured
(kVp)
Fail
Image Resolution
Adjusted
Mag Mode
(Field Size)
Specification
(LP/mm)
Measured
(LP/mm)
Adjusted
1
64 ±4
Yes
No
NORM (9”)
≤ 2.2
Yes
No
2
74± 4
Yes
No
MAG1 (6”)
≤ 3.0
Yes
No
3
82 ± 4
Yes
No
MAG2 (4.5”)
≤ 3.5
Yes
No
Workstation Checks
Pass
Fail
Dose Checks
Dose rate
Measured as found:
_______mR/min
No adjustment necessary
(≤4.0±0.3 mR/min)
DAP calibration
Radiographic Film Beam Alignment
9-4
Measured if adjusted:
_______ mR/min
Periodic Maintenance
Mag Mode
NORM(215 ±8)
MAG1(157 ±8)
MAG2 (119 ±8)
Displayed Image
Diameter (mm)
X-ray Generator Accuracy Tests
kV Accuracy (± 5%)
mA Accuracy (± 9%)
Normal Fluoro: 66kVp:_______, 99kVp: ________,
110kVp:________
Digital spot: 66kVp:_______, 99kVp: ________, 110kVp:________
Radiography:
55 kVp@1mAs :______, 55 kVp@2mAs: ______,
110 kVp@2mAs: ______
Normal Fluoro:
0.1mA@45kVp:________, 0.8mA@55kVp:_______,
1.8mA@60kVp:_______, 3.5mA@80kVp:________,
4mA@110kVp:________
HLF:
0.2mA@45kVp:_______, 1.5mA@50kVp:_______, 3.8mA@60kVp:_______
7.0mA@80kVp:_______, 10.0mA@60kVp:______, 12mA@55kVp:________
Digital Spot:
1mA@45kVp:_______, 4mA@60kVp:________, 8mA@80kVp:________,
12mA@110kVp:________,16mA@80kVp:_______
40kV, 80mAs:______
mAs Accuracy(±8%)
70kV, 2mAs: _______
110kV, 1mAs:_______
Reporting
Install Equipment Covers
Reporting complete
Yes
Yes
No
No
Calibrated Tools List
Tool Name
Tool Number
Calibration Date
Comments
FSE and Customer sign and date this section when the PM has been completed.
Date
Field Service Signature
Customer Signature
9-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9.3.
Administrative Information
Medical Facility Information:
Enter all information about the medical facility where the system is installed.
Field Service Engineer Information:
Enter Field Service Engineer name, Region ID number and Employee ID number.
System Configuration:
Complete each field in the System Configuration section. This section identifies specific information about
the system being maintained.
9.4.
Safety Inspection
Follow this procedure to inspect system components. Look for loose, damaged, or missing parts. Check
each component for proper operation.
WARNING
Electrical circuits inside the equipment use voltages that are capable of
causing serious injury or death from electrical shock. Use appropriate
precautions.
If the system has power applied, power off the system and unplug the C-Arm’s AC power cord from AC
outlet. Remove C-arm and workstation covers as needed.
9.4.1.
Antistatic Drag Wires
Check the system’s antistatic drag wires. There is one suspended below the workstation and one
underneath the C-Arm. Both wires must securely attach to the system and not be damaged or
encumbered by debris.
9.4.2.
Electrical Plug and power cord
Inspect the C-Arm’s AC power cord and AC line plug for damage and missing parts. Make sure all wiring
insulation is intact. Ensure that there are no broken or frayed wires. Make sure grounding pin is present
on AC line plug, and that all conductors on plug are secure and in good condition.
WARNING
9.4.3.
Some devices such as high voltage cables and electrolytic capacitors can retain a
dangerous static charge for long periods after power has been removed. Do not touch
these components unless power has been completely removed and they have
discharged.
Interconnect cable and Connector
Check the interconnect cable for damage or other problems. Verify that connector on cable mates well
with the connector on C-Arm and that there are no bent or missing pins. Make sure the strain relief
attaches securely to both the interconnect cable and connector.
9-6
Periodic Maintenance
9.4.4.
Footswitch cable and pins
Inspect footswitch cable and verify that the cable, connector and pins are not damaged.
9.4.5.
Ground Continuity
Perform the ground continuity test described below.
WARNING
1.
Do not perform this continuity test with the AC plug connected to a power
receptacle.
Verify continuity between the following:
From
Pin 4 of the Footswitch Connector
2.
To
Screw on the bottom of the
Footswitch
Spec
< 0.5 Ω
Connect the Interconnect Cable to the C-Arm and verify continuity between the following:
From
To
Spec
Screw on the top left corner of the WorkStation
AC Power Plug Ground Pin
< 0.5 Ω
Screw on the tube cover
AC Power Plug Ground Pin
< 0.5 Ω
Screw on the bottom of the Footswitch
AC Power Plug Ground Pin
< 0.5 Ω
Record only the highest resistance on the PM check list.
9.5.
PC Clean
1. Remove the workstation cover.
2. Unplug the cable connected to computer. Take the computer out of the workstation.
3. Remove PC cover
4. Clean the PC outside and inside.
5. Recover the computer cover and put it back into the position. Plug the cable to PC. Recover the
workstation cover.
9.6.
System Function checks
Check the following areas of the C-Arm and workstation for proper operation.
9.6.1.
WARNING
Perform Power On Checks
Observe safety procedures when operating motorized features.
Perform the following steps with the system powered up and normally booted.
9-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
1. Connect the integrated cable in the C-Arm connector and lock it. Connect the Footswitch
cable to the C-Arm.
2. Plug the AC power cord into the receptacle.
3. Press the system on key located on the C-Arm control panel. Verify that the workstation and
C-Arm complete their software boot and initialization with no errors reported on the
workstation and C-Arm displays.
Mechanical Movement Test
Refer to section 3.6.1 for mechanical inspections and checks.
9.6.2.
C-Arm Checks
C-Arm Controls
Perform the following C-Arm control checks:
WARNING
Some of the following procedures produce X-rays. Take appropriate precautions.
1. Take X-ray exposure and verify the following Image Orientation functions:
A. Image rotation key
rotates the image clockwise and counter-clockwise.
B. Image Flip key inverts and re-inverts left monitor image vertically.
C. Image Flop key inverts and re-inverts left monitor image horizontally.
2. Press the field size key
repeatedly, verifying ability to select any of the three available field
sizes: NORM, MAG1 and MAG2.
3. Press each of the following collimator switches while X-ray exposure. Verify each of the following
collimator functions:
A. Collimator iris opens/closes.
B. Collimator semi-transparent leaves open and close.
C. Collimator semi-transparent leaves rotate.
4. Press kVp and mA keys and verify that the technique can be adjusted manually.
5. Press and hold ALARM RESET key
for approximately 2 seconds. Verify fluoro time displayed
on the C-arm control panel resets to zero.
6. Check HLF mode as follows:
A. Cover the Image Intensifier with the lead aprons.
9-8
Periodic Maintenance
B. Take a normal fluoro shot in order allow the kV level to stabilize.
C. Select HLF mode via control panel and take a high-level fluoro shot.
D.
Verify that the audible tone beeps at twice the normal rate. Verify that the mA is driven
higher than the normal fluoro at the same kV value.
E. Leave lead aprons in place for next step.
7. Check HLF Pulse Mode as follows:
A. Select PULSE mode via the C-arm control panel. Select a pulse rate of 4 PPS.
B. Take a high-level fluoro shot.
C.
Verify that the audible tone beeps at twice the normal rate and the X-ray indicator
flashes at 2 times/3 seconds. Verify that the mA is driven higher than normal Fluoro.
D. Leave lead aprons in place for next step.
8. Assure correct Digital Spot function as follows:
via control panel.
A. Select DIGITAL SPOT
B. Take a digital spot shot. Verify fluoro exposure automatically terminates and image
automatically saves to hard drive.
C. Remove the lead aprons from the Image Intensifier.
9. Auto Technique Tracking
A. Select Auto Fluoro mode and the NORM field size via control panel.
B. Open the collimator Iris and leaves completely.
C. Place copper filters on the collimator and take exposure. Verify that the kVp tracks within
the range indicated in table below.
Copper Filter
kVp range
1mm
2mm
3mm
64kVp±4kVp
74kVp±4kVp
82kVp±4kVp
D. Remove all copper filters.
If the kVp is not in the range, refer to 5.71 for calibration.
10. Image Resolution
A. Using the Converging line pairs/millimeter (lp/mm) tool on the Image Intensifier. Verify the
resolution in lp/mm is equal to, or greater than the following values:
Mag Mode
(Field Size)
NORM (9’’)
Resolution (lp/mm) ≥ 2.2
MAG1 (6’’)
MAG2 (4.5’’)
≥ 3.0
≥ 3.5
B. Record test results on PM checklist.
If the resolution check fails, refer to section 5.6.8 for Camera Focus calibration.
11. Verify Film mode as follows:
9-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
A. Select Film mode
.
B. Take a 50 kVp 80mAs shot and verify that the audible tone beeps are heard after the
exposure and no errors are displayed.
X-ray Key/Switch Functional Test
Perform the following test to verify that the system’s X-ray and key switches are operating satisfactorily.
WARNING
This procedure produces X-rays. Take appropriate precautions.
1. Press Place the X-ray keyswitch in STANDBY position (
). Press handswitch and verify that no Xrays are generated. Press Vertical lift button and verify that vertical lift column is not operational.
). Press handswitch and verify that no X-rays are
2. Place X-ray keyswitch in XRAYOFF position (
generated. Press vertical lift button and verify the vertical lift column is operational.
). Press handswitch, footswitch and vertical lift button to
3. Place X-ray keyswitch in ON position (
verify X-ray and vertical lift column are enable.
9.6.3. Workstation Checks
1. Press SWAP (
) key and verify this function exchanges images between the left monitor and
the right monitor.
) key and press
2. Press the SAVE (
Image Directory screen.
key. Verify that the left monitor image is previewed in the
) activates the Patient Information screen. Enter patient
3. Verify Patient Information function (
information. Display scheduled exams, display resume exams screen.
4. Verify Image Directory (
dose information.
5. Verify the SETUP (
) function allows you to review, print and archive stored images and
) function allows access to the Setup screen.
6. Verify ZOOM (
) function displays a square region of interest (ROI) box and that you can
magnify the ROI.
7. Verify NEGATE (
left monitor.
) function reverses the light and dark values on the image displayed on the
) function displays and indicator bar and that the amount of
8. Verify the CONTRAST (
contrast can be increased and decreased.
9-10
Periodic Maintenance
9. Verify the BRIGHTNESS (
) function displays an indicator bar and that amount of
brightness can be increased and decreased.
10. Verify that Auto Contrast/Brightness (
contrast and brightness.
) automatically selects the optimum values of
) function displays an indicator bar and that amount of
11. Verify ENGE ENHANCE (
enhancement level increased and decreased.
12. Press ALT key and the LOGOUT key (
) at the same time. Verify that it can log the user out of
the workstation applications. The default password is “123456”.
9.6.4. Dose Rate
Entrance exposure is the amount of radiation available at the entrance point of patient anatomy.
Entrance exposure is measured 9" above the surface of the Image Intensifier in the center of the X-ray
beam. This measurement is used to represent the average thickness of patient anatomy.
1. Using a dosimeter for the measurements. Place the dosimeter probe on the I.I. grid center. Use
2mm copper to cover the beam path on the X-ray tube. Take exposure in Auto fluoro mode. Dose
rate should be within 4.0±0.3 mR/min.
2. Indicate if adjustments were made on the Checklist.
9.7.
DAP Calibration
DAP Test and Calibration
Set Dosimeter as accumulative/maintenance. Put the detector of dosimeter on the center of the grid on
image intensifier. Power on the system and create a new exam. Ensure the DAP and Dose value on the
right monitor are cleared. Set exposure parameters as 80kV, 10 mAs in radiography mode. Install the
cassette holder on image intensifier. Insert a film to cassette holder and press hand switch to take
exposure. Record the DAP show value on right monitor and dose value on dosimeter. Get the film and
test the diameter of the image. Calculate the circle area of the image on film.
DAP calculated value= Dosimeter reading×circular area on film
Compare the DAP calculated value with DAP show value on the right monitor. The error should be less
than 20%.
ERROR=
DAPCalculatedvalue − DAPshowvalue
DAPCalculatedvalue
× 100%
1. Units should be consistent during comparing.
NOTE
2. The test should be done after the tube cover installation complete.
9-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
If the error above is too large, adjust DAP calibration coefficient to calibrate: Login service tool. Click
“Exposure Management” button on main screen. Change the DAP Calibration Coefficient. DAP show
value on the workstation right monitor is in direct proportion with the calibration coefficient.
Set the coefficient and test DAP value again until the error is less than 20%.
Dose Test and Calibration
Install the beam alignment tool. Set Dosimeter as accumulative/maintenance. Put the detector of
dosimeter on the axis of image intensifier and focus, 30 cm away from grid center. Clear the dosimeter.
Create a new exam on workstation and ensure the DAP and dose show value on the workstation are
cleared.
Set kV/mA value as 110kV/4mA in manual fluoro mode. Open the collimator completely. Take exposure
for 3-5 seconds. Record Dose show value on right monitor and dosimeter measured value. Repeat three
times and get the average of the dose show value on right monitor and the average of measured value
on dosimeter. Calculate the error between show value average and measured value average.
NOTE
After each test, reset the dosimeter and create a new exam to clear the DAP and dose
show value on workstation.
ERROR=
DoseMeasuredvalue − Doseshowvalue
DoseMeasuredvalue
× 100%
If the error is more than 5%, adjust Dose Calibration Coefficient: Login service tool. Enter “Exposure
Management” screen. Change Dose Calibration Coefficient. Dose/Dose rate show value on the
workstation right monitor is in direct proportion with the calibration coefficient.
Dose Accuracy Test
According to the collimator field size settings, adjust the collimator opening size by beam alignment tool.
In each opening size test, create a new exam on the workstation. Set the exposure parameters as 45kV,
1.3mA in manual fluoro. Adjust collimator field size. Record dose show value on right monitor of the
workstation as D0. Reset the dosimeter. Take exposure at high and lower kV value for 3-5 seconds.
Record the dose show value on the right monitor and measured dose value on dosimeter. Calculate dose
value on the right monitor of the workstation.
@60kV, 1.3mA: Dose value on workstation=D1-D0 (D1: Dose show value on right monitor after exposure
@60kV, 1.3mA)
@110kV, 4mA: Dose value on workstation=D2-D1 (D2: Dose show value on right monitor after exposure
@110kV, 4mA)
Calculate the error between the dose value on workstation and measured value on dosimeter.
ERROR=
DoseMeasuredvalue − Dosevalue(onworkstation)
DoseMeasuredvalue
× 100%
The error should be less than 30% in every state.
NOTE
9-12
After each test, dosimeter must be cleared. Create new exam to ensure DAP and dose
show value on workstation be cleared.
Periodic Maintenance
Collimator field size
45mm-43mm
110-120mm
140-150mm
Largest size
NOTE
9.8.
kV/mA
Dose value
(mGy)
Dose measured
value on dosimeter
(mGy)
Error
60kv/1.3mA
110kv/4mA
60kv/1.3mA
110kv/4mA
60kv/1.3mA
110kv/4mA
60kv/1.3mA
110kv/4mA
The field size in the form should be measured by field size measure tool.
Radiographic Film Beam Alignment Test
Field Size Verification
1. Select relative field size, open the collimator to the maximum, and take an exposure.
2. Read the BAFT tick mark on the boundary of the image. The size of the image should be within
the range of the values below.
Mag Mode
NORM
MAG1
MAG2
Displayed Image
Diameter (mm)
215 ±8
157 ±8
119 ±8
If the requirement is not met, you must perform I.I. Field Size Calibration in 5.6.5.
9.9.
X-ray Generator Accuracy Tests
These tests measure the kVp accuracy of the system. Complete the X-ray Generator Accuracy section.
9.9.1. kVp Accuracy
Refer to section5.2.1 for kVp accuracy check.
9-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
9.9.2. mA Accuracy
Refer to section5.3.1 for kVp accuracy check.
9.9.3. mAs Accuracy
Refer to section5.4.1 for kVp accuracy check.
9.10. Reporting
9.10.1. Install Equipment Covers
Install all equipment covers you removed earlier. Make sure all air filters are clean and properly attached
to the covers.
9.10.2. Complete PM Comments
Write any comments important to the PM procedure in the Comments section of the PM Checklist.
9-14
10.
Chapter 10. Technical Reference
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.1. Overview
The policy of GE Healthcare is one of continual product development and improvement.
For this reason, GE Healthcare reserves the right to change the operating characteristics and
specifications of newer products at any time, without prior notice, and without incurring any obligation
relating to previously manufactured items.
The specifications listed are limited to general performance and physical data. Specifications of optional
equipment provided by other manufacturers are given in the applicable manuals provided with those
options.
10.2. System Specifications
10.2.1. Safety Classification
Class I, Type B, IEC60601-1
Ordinary protection against ingress of water
Degree of Protection:
•
System: IPX0
•
Footswitch: IPX8
10.2.2. Environmental Requirements
Operating Conditions:
Temperature:
+10°C~+40°C (+50F~+104F)
Humidity:
30%~75% Relative, non-condensing
Atmospheric Pressure:
700hPa~1060hPa
Transportation and Storage:
Temperature:
-20°C~+55°C (-4F~+131F)
Humidity:
10~95% Relative, non-condensing
Atmospheric Pressure:
500hPa~1060hPa
10.2.3. Power Requirements
Voltage:
AC 100V/110V/120V/220V/230V/240V
Frequency:
50Hz/60Hz±3Hz
Rated Momentary Line Current (shorter than 5s):
•
10-2
25A (220V/230V/240V)
Technical Reference
• 50A (100V/110V/120V)
Rated Continuous Line Current:
•
10A (220V/230V/240V)
•
20A (100V/110V/120V)
Maximum impedance: 0.6 Ω/0.3 Ω
NOTE
16A power supply is required in China. And 20 A is required in Japan.
10.2.4. Circuit Breaker
Breaker for 110V:
Rated Voltage
AC 110 V, AC 120 V +/-10%
Frequency
50 Hz /60 Hz
Rated Current
25A (AC)
Breaker for 220V:
Rated Voltage
AC 220 V
Frequency
50 Hz /60 Hz
Rated Current
12A (AC)
+/-10%
10.2.5. Earth leakage current
Normal condition
0.5mA
Single fault condition
0.5mA
10-3
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.2.6. Physical Specifications
C-Arm
10-4
SID
39.2’’ (100cm)
Horizontal travel
8’’ (20cm)
Vertical travel
16’’ (40cm)
Orbital rotation
120°(90°/30°over scan)
Lateral rotation
±205°
Wig-Wag
25°(±12.5°)
Weight
245±10kg
Technical Reference
Workstation
Weight: 140±10kg
Option Weight
DAP
0.22±0.02kg
Skin Spacer
0.17±0.02kg
Laser Aimer (tube)
0.65±0.06kg
Laser Aimer (Image Intensifier)
2.16±0.2kg
Film Cassette holder
1.4±0.1kg
Video Distributor
2.3±0.2kg
Two-Pedal Footswitch
4±0.4kg
10-5
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.3. Specification of Key Components
10.3.1. X-Ray Generator
Nominal electric power
2.0kw (100kV, 20mA, 0.1s)
Maximum electric power
Normal Fluoro
440W
HLF
660W
Digital Spot
1320W
Radiography
2.2 kW
High frequency generator
40kHz
Voltage
40kV~110kV, step value: 1kV
Normal Fluoro
When the X-ray tube current reaches its
highest value of 4mA, the obtainable highest
X-ray tube voltage is 110kV.
HLF
When the X-ray tube current reaches its
highest value of 12mA, the obtainable highest
X-ray tube voltage is 55kV.
Digital Spot
When the X-ray tube current reaches its
highest value of 16mA, the obtainable highest
X-ray tube voltage is 82kV.
Radiography:
Current: 20mA, stationary; the obtainable
highest X-ray tube voltage is 110kV.
Current
Step value: 0.1mA
Normal Fluoro
0.1 mA~4 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 4mA.
HLF
0.2 mA~12 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 6mA.
Digital Spot
0.2mA~16 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 12mA.
Radiography
1mAs~80mAs; current: 20mA, stationary; the
obtainable highest X-ray tube voltage is 110kV.
Radiography
Voltage
10-6
40~110kVp
Technical Reference
Value of mAs
1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13,
16, 20, 25, 32, 40, 50, 63, 80 mAs
Pulse Fluoro
1/2/4/8pps
The lowest Current Time Product or the
combination of Loading Factors
40kVp, 1mAs
10.3.2. X-Ray tube
(mm)
Nominal high voltage
110 kV
Nominal input power
2.2 kW
Maximum anode heat capacity
53 kJ
Target material
Tungsten
Target angle
10°
The reference axis
From the center of the spot to the center
of the image intensifier
Small focal spot
0.6×1.4 mm
10-7
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Large focal spot
1.4 mm
Anode
Stationary
Typical filament specification
4.2 V, 3.5A
Maximum filament current
4.6A
Weight
0.65 kg
Anode cooling rate
37 kHU/min
Anode heat capacity
76kHU
Manufacturer
LOHMANN
MODEL
110/3DF
Maximum voltage between anode and
ground
55kV
Maximum voltage between cathode and
ground
-55kV
Maximum filament voltage
24V AC
X-ray radiation angle
10 degree
Tolerance:
Tube voltage: ± 7%
The tube current: ± 10%
The current time product: ± 10%
10.3.3. X-ray tube assembly
Maximum continuous heat dissipation
177W
Tube assembly dimension (length×Width×Height)
327mm ×132 mm×238 mm
Tube assembly weight
17.5 kg
Maximum heat capacity
800,000HU
Maximum cooling rate
18,000HU/min
Cooling method
Nature cooling
Nominal loading condition
70kVp, 2.0mA
Maximum symmetrical radiation field on the X-axis
100cm from the spot:
Fluoro: 235 mm (diagonal)
Radiography: 240mm (diameter)
10-8
Technical Reference
To replace the X-ray tube assembly:
1. Open the cover of the X-ray tube assembly.
2. Demount the collimator.
3. Demount the X-ray tube assembly and replace with new ones.
4. Start the system, and choose manual fluoro mode.
5. Adjust the fluoro current, voltage and the radiography mAs, voltage.
6. Shut down the system.
7. Cover the X-ray tube assembly.
If the X-ray tube has been produced more than 3 months, it should be warm first.
10.3.4. Collimator
Input power: DC 6V, 0.9W
Electrical connection mode: through DB9.
Mechanical connection mode: cast aluminum bracket, screw fitting
Fluoroscopy
Nominal diameter circle: 23/15/11cm (9”/6”/4.5”)
X-Ray beam continuously adjustable to an area of less than 5×5 cm at the image receptor
Radiography
Nominal diameter circle: 23cm (9”)
10-9
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.3.5. Others
Semi-transparent shutters
Rotation
180°
Operation
Controlled on panel
Grid
Type
Circular
Material
Carbon fiber
Lines/cm
60
Ratio
10:1
Focus distance
100 cm
Image Intensifier
Type
3 Fields
2 Fields
1 Fields
Input field size
9’’(23 cm)
6’’(15cm)
4.5’’(11cm)
Central resolution
≥51Lp/cm
≥57Lp/cm
≥67Lp/cm
Monitor
Type
Monochrome LCD
Size
19’’
Image Storage
Capacity
80,000 images
Size
1K x 1K
Digital Processor
Function
10-10
Image Processing, Recursive Noise Reduction,
Motion Artifact Reduction, Real-time Edge
Enhancement, Auto Brightness/Contrast,
Brightness/Contrast Adjustment, Negate, Last
Image Hold, Zoom/Roam, Image
Rotation/Reversal/Invert and Swap
Technical Reference
10.4. Tube Rating Chart
The following charts and data describe tube characteristics when operated with three-phase full-wave
rectification, a reasonable approximation to the high frequency generator with minimal ripple.
10.4.1. Filament Emission Characteristics
Small Focal Spot
10-11
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Large Focal Spot
10-12
Technical Reference
10.4.2. Single Load Ratings
Small Focal Spot
Exposure time in seconds
10-13
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
Large Focal Spot
Exposure time in seconds
10-14
Technical Reference
10.4.3. Thermal Characteristic
Anode Heating and cooling characteristics:
Tube Housing heating and cooling characteristic:
Joules
800 000
300W
200W
640 000
480 000
320 000
160 000
0
0
25
50
Time (minutes)
75
100
10-15
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.5. Radiation Safety Specification
Operators are strongly urged to be familiar with the current recommendations from the International
Commission on Radiological Protection. (ICPR)
Take radiation protection of the patient and operator into consideration seriously.
Scatter data in this manual is only for reference.
10.5.1. Scatter Data
Filter in the x-ray tube
assembly
2.55mm Al
Additional filter
1mmAl
Total filter
3.55mmAl
Leakage radiation
Dose area product rate is less than 1.0mGy/h, averaged
over any area of 100cm2 of which no principal linear
dimension exceeds 20cm at 1m from the focal spot
under conditions of loading. (110kVp, 4mA)
SID (focal spot to image
receptor distance)
100cm
WARNING
NOTE
10-16
Patients must be kept as far as possible from the focus, with minimum distance of 20cm.
Make sure not to operate when a plastic cover in x-ray assembly is disassembled, which
properly make the distance meet the above requirements.
The operating table matched this equipment should be made of small attenuation
equivalent material, otherwise, X-ray dose may be increased.
Technical Reference
10.5.2. Scatter radiation specification
10-17
Brivo OEC 850 Mobile C-Arm X-ray Product Service Manual
10.6. Material Safety Data Sheets
Manufacturer’s Material Safety Data Sheets are available from the manufacture upon request. Contact
the following manufacturers with regard to the materials listed:
Oil
Sinkiang Kelamayi Lubricating Oil Plant
Address: Jinlong Town Youlong Road No. 8, Kelamayi City , Sinkiang Uigur Munisipality
Post Code: 834003
Tel: 0990-6831426
10.7. Material Recycling
Parts or accessories at end–of–life:
The disposing of parts and accessories must be in accordance with national regulations or local
regulations for waste processing. All materials and components that could have risks to the environment
must be removed from the end–of–life parts and accessories (For examples: transformer oil, etc...).
Please consult your local GEMS representative before discarding these products.
Packing materials:
The materials used to pack our equipment are recycled. They must be collected and processed in
accordance with the local regulations.
10-18
11.
Appendix
Brivo OEC 850 Mobile C-Arm X-ray Product
List
Appexdix-2
Name
Part Number
Revision
Total Pages
Filament Board (MP4)
5326475
1
1
UP/Down Driver Board
5326476
4
1
Power Distribution Board
5326479
1
1
Mother Board
5326482
1
4
Extended Board
5326483
1
2
Collimator Control Board (MP1)
5326480
1
4
Mainframe Control Board
5326484
4
32
Display Board
5326486
2
5
Keyboard Board
5326481
2
2
KV Control Board
5326487
2
9
IGBT Inverter board
5326477
3
2
Seal board
5330746
2
2
UPS control board
5370278
1
2
MCB extend board
5326478
1
7
Circuit Schematics
Appendix- 3
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-4
Circuit Schematics
Appendix- 5
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-6
Circuit Schematics
Appendix- 7
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-8
Circuit Schematics
Appendix- 9
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-10
Circuit Schematics
Appendix- 11
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-12
Circuit Schematics
Appendix- 13
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-14
Circuit Schematics
Appendix- 15
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-16
Circuit Schematics
Appendix- 17
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-18
Circuit Schematics
Appendix- 19
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-20
Circuit Schematics
Appendix- 21
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-22
Circuit Schematics
Appendix- 23
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-24
Circuit Schematics
Appendix- 25
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-26
Circuit Schematics
Appendix- 27
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-28
Circuit Schematics
Appendix- 29
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-30
Circuit Schematics
Appendix- 31
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-32
Circuit Schematics
Appendix- 33
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-34
Circuit Schematics
Appendix- 35
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-36
Circuit Schematics
Appendix- 37
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-38
Circuit Schematics
Appendix- 39
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-40
Circuit Schematics
Appendix- 41
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-42
Circuit Schematics
Appendix- 43
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-44
Circuit Schematics
Appendix- 45
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-46
Circuit Schematics
Appendix- 47
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-48
Circuit Schematics
Appendix- 49
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-50
Circuit Schematics
Appendix- 51
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-52
Circuit Schematics
Appendix- 53
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-54
Circuit Schematics
Appendix- 55
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-56
Circuit Schematics
Appendix- 57
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-58
Circuit Schematics
Appendix- 59
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-60
Circuit Schematics
Appendix- 61
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-62
Circuit Schematics
Appendix- 63
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-64
Circuit Schematics
Appendix- 65
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-66
Circuit Schematics
Appendix- 67
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-68
Circuit Schematics
Appendix- 69
Brivo OEC 850 Mobile C-Arm X-ray Product
Appexdix-70
Download