Smart Glucose Monitoring System using MEMS Glucose Sensor A MINI PROJECT REPORT Submitted by BLESSING BENZER.B.S (3122224004002) M.E.VLSI DESIGN DEPARTMENT OF ELECTRONICS AND COMMUNICATION ENGINEERING SSN COLLEGE OF ENGINEERING CHENNAI 1 ABSTRACT Glucose molecules are detected in blood using a MEMS sensor and the extent of glucose present is analyzed. This design consists of a strip coated with the opposite dehydrating agent of the glucose and an electrode which is polarized when the battery is switched on. When a blood drop is placed on the strip, the glucose reacts with the dehydrating agent and forms chemical by-products and electrons, which are driven to the electrode which is polarized by the battery power. Virtualization of the concept on Comsol Multiphysics tool is used to establish a relation between current and glucose concentration based on the obtained result. 2 TABLE OF CONTENTS CHAPTER NO. TITLE PAGE NO. ABSTRACT 2 1 INTRODUCTION 4 2 ELECTROCHEMICAL 5 PRINCIPLE 3 WORKING OF SENSOR 6 4 MODELING CONCEPT 7 5 SIMULATION OUTPUT 8 5 RESULT AND DISCUSSION 9 6 CONCLUSION 10 7 REFERENCE 11 3 1.INTRODUCTION Glucose sensing is one of the most widespread and commercially successful uses of electroanalysis. In an electrochemical glucose sensor, the concentration of glucose in a sample is measured using amperometry; that is, the measurement of an electric current. An applied voltage causes the oxidation of glucose, and the current due to this oxidation is measured at the electrode. In a well designed glucose sensor, there is a linear relationship between the glucose concentration and the current, enabling a calibrated measurement. Typically, the oxidation of glucose does not occur directly at the working electrode where current is measured. Instead, the reaction is accomplished by a chemical oxidant and accelerated by a biological enzyme such as glucose oxidase (GOx), which makes the sensor specific to glucose and independent of the concentration of other oxidizable species that may be present in the analyte solution. The reduced form of the oxidant, after its reaction with glucose, can be re-oxidized directly at the electrode. In nature, the oxidant is oxygen, but this suffers from slow kinetics and the rate of oxidation is perturbed by the oxygen concentration dissolved from the atmosphere into the analyte solution. Instead an inorganic oxidant with fast electrode kinetics, such as the hexacyanoferrate (III) anion (commonly, “ferricyanide”), is suitable for use in a glucose sensor, since the measured current is made independent of oxygen concentration and is not limited by slow electrode kinetics (Ref. 1). This example demonstrates a steady-state analysis of the current drawn in a unit cell of solution above an interdigitated electrode, where the counter electrode reacts ferricyanide to ferrocyanide. The linearity of the response of the sensor is demonstrated for a typical range of glucose concentrations. Model Definition The model contains a single 2D domain. 4 2.ELECTROCHEMICAL PRINCIPLE In this method we are sensing the glucose levels by using electro analysis. In this method we design a gluco-strip with certain geometry. When a certain voltage is provided to this strip and add an oxidizing agent to the glucose then a certain number of electrons are released due to the oxidation of glucose. Based on the amount of electrons evolved the current obtained is measured. Now depending on the current measurement glucose extent in the blood is predicted. Glucose + ferri + e- ↔ products + Ferro Here we use ferricyanide as an oxidizing agent to accelerate the reaction. The voltage required for this strip is provided by the battery cell provided in the device that is used to measure the strip. When the device is switched on and the strip is inserted into it then the voltage from the battery is provided to the strip and initiates the oxidization. Actually when the voltage is given then the strip part is polarized and separated into two different electrodes one as anode and other as cathode. This design is done in the electrochemistry module of the model builder in COMSOL MULTIPHYSICS. When the oxidation is done all the electrons are collected at one electrode and the other electrode is empty. Then by using thedevice we measure the current at the electrode in amperes and analyze the glucose concentration based on the amount of amperes of current evolved. MEMS innovation holds the possibility to permit coordinated implantable sensors for metabolic observing. Continuous glucose monitoring (CGM) sensors based on affinity detection are desirable for long-term and stable glucose model The rate of this reaction (milliMol/m3 or (mMol/L)) is given by a Michaelis–Menten rate law as 5 Here, the parameter Vmax is the maximum rate of the enzyme-catalyzed reaction, depending on the quantity of enzyme available, here Km is Michaelis-Menten coefficient. At large glucose concentration, the rate becomes independent of the glucose concentration and solely depends on the enzyme kinetics. The current density for this reaction is given by the electroanalytical Butler–Volmer equation for an oxidation: in which k0 is the heterogeneous rate constant of the reaction, αc is the cathodic transfer coefficient, and η is the over potential at the working electrode. 3.WORKING OF GLUCOSE SENSOR The glucose sensor determines the concentration of glucose in the solution.Most glucose sensor are based on electrochemical technology, they use electrochemical test strips to perform the measurement. A small drop of the solution to tested is placed on a disposable test strip that the glucose sensor uses for the glucose measurement.Glucose sensor consist of two types of electrode Figure 1:Glucose Sensor 6 Reference electrode: Held at a constant voltage with respect to the working electrode to push the desired chemical reactions. Counter electrode: Supplies current to the working electrode. Most of the glucose meter designs use only two electrodes, reference electrode and working electrode. 4.MODELLING CONCEPT The model contains a single 2D domain representing a 100 μm-wide unit cell of solution above an interdigitated electrode (Figure 1). The real geometry is a periodic repetition of this unit cell in the x direction. The cell and electrode are assumed to extend sufficiently far out-of-plane of the model that the 2D approximation is suitable. At the top of the unit cell is a bulk boundary where the concentrations are assumed to equal those in the bulk solution of the analyte. At the bottom of the unit cell, the y = 0 axis is divided by four points into separate electrode and insulator boundaries. The anode (working electrode) is at the center of the cell in the range 37.5 μm < x < 62.5 μm. The 3 | GLUCOSE SENSOR unit cell contains half of each of the two neighboring cathodes (counter electrodes) in the ranges x < 12.5 μm and x > 87.5 μm. Between the anode and cathode surfaces, a solid insulating material is present. Figure2:Modeling of Glucose Sensor 7 This is the rectangular strip designed in the Comsol Multiphysics tool. This modeling is done in the electro analysis part by selecting a rectangular strip with certain dimension. Then provide the necessary parameters and required equations. Later design for the cathode and anode part at the bottom of the rectangular strip. The orange-red part indicates the electrode cloud region.And an insulation is created between cathode and anode regions. The width of cathode ,anode and insulation are varied for different models 5.SIMULATION OUTPUT Figure 3:Module 100μm*1000μm dimension Figure 4:Module 50μm*750μm dimension Figure 5:Module 50μm*1500μm dimension Figure 6:Module 50μm*1500μm dimension 8 Figure 6:Module 160μm*1000μm *10μm dimension 6. RESULTS AND DISCUSSION Table 1 9 From the above result analysis extracted into the table Figure 3 and 4 are linear up to 6 mMol/L only, from 6-10 mMol/L it doesn’t show the linearity. Similarly figure 6 is linear up to 4mMol/L and later on from 4-10mMol/L it doesn’t obey linearity. When there is no perfect linear relationship between glucose concentration and current exactly we cannot asses the glucose extent of a person. But in figure 5 module has the best range of current and completely linear.so glucose concentration can be more accurately estimated from this module since it is perfectly linear. Though figure 5 also has perfect linearity its current range is smaller than that of figure 5 module. The more the range the more the accuracy in prediction of the glucose 6.CONCLUSION The enzymes responsible for the oxidation of glucose molecules are studied and finally designed different models for the glucose-strip to respond and determine different glucose levels and chosen the best model among them to have a perfect linearity relationship between current in amperes and concentration of glucose. By this we can estimate the glucose levels of a person either before or after fasting based on the ampere reading that we get from the current reading. 10 7.REFERENCE •Shameem, S., Krishna, D. V., Kumar, K. S., & Avinash, K. (2018). Design Of MEMS Sensor For The Detection Of Diabetes. 2018 3rd International Conference on Inventive Computation Technologies (ICICT). doi:10.1109/icict43934.2018.9034268 •https://ieeexplore.ieee.org/abstract/document/9034268 11