Single centre experience of flow diverter treatment of complex intracranial aneurysms
from South India: Intermediate and long-term outcomes
Santhosh K Kannath1, Aneesh Mohimen1, Kapilamoorthy T Raman1, Mathew Abraham2, Suresh Nair2, Jayadevan E Rajan1,
1 Department of Imaging Sciences and Interventional Radiology, Neurointervention Center, Sree Chitra Tirunal Institute of Medical Sciences and
Technology, Trivandrum, Kerala, India
2
Department of Neurosurgery, Neurointervention Center, Sree Chitra Tirunal Institute of Medical Sciences and Technology, Trivandrum, Kerala,
India
Correspondence Address:
Dr. Jayadevan E Rajan
Neurointervention Center, Department of Imaging Sciences and Interventional Radiology, Sree Chitra Tirunal Institute of Medical Sciences and
Technology, Trivandrum, Kerala
India
Abstract
Background and Purpose: To report the intermediate and long-term clinical and angiographic outcomes of the
treatment of complex intracranial aneurysms with flow diverter (FD) stents. Setting: A tertiary care centre from
south India. Materials and Methods: Patients treated with FD stents were retrospectively analyzed. The clinical
demographics, technical success, angiographic, and long-term outcomes were recorded. Results: A total of 13
patients underwent FD treatment, in whom 11 procedures were successful. The cohort included large or giant
intracranial aneurysms and recurrent aneurysms following conventional endovascular treatment. Major morbidity was
observed in 1 patient, who developed basal ganglia bleed that needed evacuation. Minor complications were seen in
36% of patients without clinical sequelae. Significant obliteration of aneurysm was noted on 1 month computed
tomography angiogram in >80% of the patients. Angiographic complete obliteration was noted in 89% of the patients
at 6 months. Cranial nerve deficits were noted in 2 patients that improved on subsequent follow up. There was no
mortality observed in this cohort. Conclusion: FD treatment of complex cerebral aneurysms was associated with
favorable clinical and angiographic outcomes in the intermediate and long-term follow up. Minor complications were
common, which needed to be effectively managed to prevent major catastrophic events. The steep learning curve
influenced the technical success of the procedure.
How to cite this article:
Kannath SK, Mohimen A, Raman KT, Abraham M, Nair S, Rajan JE. Single centre experience of flow diverter treatment of complex
intracranial aneurysms from South India: Intermediate and long-term outcomes.Neurol India 2019;67:797-802
How to cite this URL:
Kannath SK, Mohimen A, Raman KT, Abraham M, Nair S, Rajan JE. Single centre experience of flow diverter treatment of complex
intracranial aneurysms from South India: Intermediate and long-term outcomes. Neurol India [serial online] 2019 [cited 2022 Nov 28
];67:797-802
Available from: https://www.neurologyindia.com/text.asp?2019/67/3/797/263195
Full Text
Flow diverter (FD) stents have revolutionized the treatment of complex cerebral aneurysms, and several studies and
meta-analysis have demonstrated the efficacy and safety of these new devices in the management of aneurysms that
are not amenable to conventional endovascular or surgical therapy.[1],[2],[3],[4] The concept of flow diversion
includes decoupling of inflow jet and centralized diversion of flow vectors resulting in stagnation of blood flow within
the aneurysm and eventual thrombosis and exclusion of the aneurysm from the circulation. These stents also act as a
scaffold for the neoendothelialization, leading to durable and sustained occlusion. The experience of FD stent in our
country is limited with only one study of small series reported in the literature.[5] In this study, we aim to report a
single institutional experience on the treatment of complex aneurysms with flow diverters with emphasis on the
immediate and long-term clinical and angiographic outcomes.
Materials and Methods
Study subjects
All patients who had undergone FD stent placement as treatment for complex intracranial aneurysms at this institute
were included in this retrospective study. Clinical documents of the patients, imaging records, preprocedural
management regime, periprocedural notes, and subsequent outpatient follow-up details were analyzed. FD treatment
was generally considered for aneurysms that are giant, fusiform, or recurrent and not amenable to conventional
endovascular therapy.
Preprocedural antiplatelet protocol
After obtaining baseline platelet function tests using light aggregometry test (LTA), the patients were started on dual
antiplatelets such as aspirin 150 mg and clopidogrel 75 mg, and the platelet function tests were repeated after 7 days
to assess the degree of suppression. If the suppression was found to be unsatisfactory, prasugrel was initiated at 10
mg and embolization was planned after 5 days. Six patients required prasugrel following the observation of
clopidogrel resistance (46%). Aspirin resistance was not observed in our cohort.
Embolisation procedure
All the procedures were performed under general anesthesia and systemic heparinization was achieved to maintain
the activating clotting time (ACT) between 250 and 300 s. The target vessel was accessed using a 6F guide catheter
and coaxial systems were considered when the arteries were tortuous. The artery distal to the aneurysm was
accessed by Marksman catheter (for pipeline embolization device, PED, Covidien, US) or Headway 27 (for flow
restoration embolization device, FRED, Microvention, Tustin, US). The device diameter was matched to the largest
arterial diameter which was usually the proximal artery. Multiple overlapping devices were considered if the neck
could not be covered adequately with a single stent. Three-dimensional (3D) computed tomography (CT) was
routinely obtained to assess the wall apposition and opening of the FD stent. If the stent mal-apposition was found,
the segment was angioplastied using compliant balloons such as Scepter C (Microvention, Tustin, US). The arterial
access sites were later closed using Perclose vascular closure device (Abbot vascular, Redwood City, California, USA).
The patients were routinely discharged after 3–4 days of observation.
Follow up
The follow-up protocol included regular clinic visits at 1, 3, 6, and 12 months and yearly thereafter. A CT angiogram
was routinely obtained at a 1-month follow-up visit and digital subtraction angiography was performed after 6
months. Further imaging studies were decided based on the angiographic findings. All the patients were continued on
dual antiplatelet therapy for at least a year, and on lifelong aspirin thereafter, if the stent remained patent and
features of in-stent stenosis were absent.
Results
A total of 13 patients with complex intracranial aneurysms have been considered for FD stents at our institute thus
far. The demographic characteristics, clinical features, and aneurysm characteristics are presented in [Table 1]. Of the
13 patients, 3 patients underwent prior treatment of the aneurysm; 2 for treatment failure following balloon-assisted
coiling and 1 patient after regrowth of aneurysm following surgical wrapping. One patient had bilateral cavernous
aneurysms and had undergone surgery and parent vessel trapping for the contralateral side; and, another patient
had been treated for direct carticocavernous fistula of the ipsilateral side 16 years ago. The rest of the patients had
no previous history of surgical or endovascular treatment.{Table 1}
Technical results
The procedure was successfully completed in 11 patients (84.6%), and in 2 patients, embolization was abandoned
due to the inaccessibility of delivery microcatheter beyond the aneurysm into the distal parent artery. PED was
deployed in 8 patients and FRED in 3 patients. A total of 12 PEDs were deployed, 1 each in 6 patients and 3 in two
patients. No significant technical issues were noted in the patients treated with a single PED. In patients treated with
multiple PEDs, the loss of distal arterial access after deployment of the first or second device resulted in prolongation
of treatment duration. Among the FRED cohort, inadequate opening of the device and inability to introduce the device
through the recommended catheter resulted in the use of additional FRED in two patients.
Periprocedural complications
Major morbidity was noted in 1 patient (9%) who developed a large hematoma in the basal ganglia immediately after
the procedure. This patient underwent emergent craniectomy and evacuation of hematoma. He had significant
weakness of left upper and lower limb and his modified Rankin score at the time of discharge was 4. Transient
worsening of the 3rd cranial nerve palsy was noted in 2 patients, which regressed gradually with conservative
management. Two patients had postprocedural retroperitoneal hematoma that was managed conservatively. One
patient developed self-limiting hematuria. The overall incidence of minor complications was 45% (5 patients). No
thromboembolic complications were noted.
Early clinical and imaging outcomes
One-month CT imaging study was available for 90% of the patients that revealed complete exclusion of the aneurysm
in 54% of patients and near complete occlusion (<10%) in 28% of patients. Significant residual filling of aneurysm
(>20%) was noted in 1 patient.
Intermediate and long-term clinical outcomes
The clinical follow-up ranged from 3 to 30 months. Glasgow outcome score remained 0 in 90% of the patients during
the follow-up period. One patient had residual 3rd nerve and V1 nerve palsy on a long-term follow up. This patient
along with another patient developed trigeminal neuralgia at 6 months that was alleviated with medical treatment.
Preprocedural cranial nerve deficits and visual field defects noted in patients showed improvement on follow up visits.
Periprocedural events and long-term outcomes of FD treatment are shown in [Table 2].{Table 2}
Angiographic outcome
An angiographic follow-up (>6 months) was available for 9 patients. Complete or near-complete obliteration of the
aneurysm was noted in 89% of the patients, and in 1 patient, mild residual filling of the aneurysm (~10%) was
observed. There was no in-stent stenosis or recanalization of the aneurysm. Progressive aneurysmal obliteration was
noted in 25% of patients in the follow-up period. Asymptomatic occlusion of covered branch artery was observed only
in 1 patient. Representative cases of FD treatment of cerebral aneurysms are demonstrated in [Figure 1] and [Figure
2].{Figure 1}{Figure 2}
Discussion
FD stents represent a new treatment paradigm where the arterial segment harbouring the aneurysm is reconstructed
using a tightly-woven mesh stent with a high metal and pore density. The metal density varies between 35–50%
among the various stents and porosity varies between 45 and 70%.[3] There are currently four flow diverters
available in our country; PED, SILK (Balt Extrusion Technology, Montmorency, France), FRED, and Surpass (Stryker
Neurovascular, Fremont, CA, USA). The stents are delivered through a large lumen microcatheter (0.021–0.027 inch)
placed beyond the aneurysm in the distal parent artery, and the stent is deployed by a combination of unsheathing
and forward push of the microcatheter to ensure a good opening of the stent as well as wall apposition. SILK, FRED,
newer generation PED, Pipeline Flex, and Surpass allow partial-to-complete resheathability after deployment. Though
the FD was initially considered for complex, difficult-to-treat aneurysms, recently it is being increasingly considered
for simple uncomplicated aneurysms, blister aneurysms, or dissecting aneurysms.[6],[7],[8]
Several reports demonstrate the technical feasibility and safety of FDs in the treatment of complex intracranial
aneurysms. Technical problems or failures were noted in 5% of PED procedures in a pooled analysis.[4] Though most
of the studies report an overall technical success of more than 95–99% across the various FDs, device malfunction or
migration is an important concern for most of the FDs. While incomplete expansion necessitating additional balloon
apposition was observed in 12% of PEDs, device misdeployment leading to additional interventions such as stenting
or balloon dilatation or consequent parent artery occlusions were observed in 12% of SILK stents.[9],[10] Similarly,
an imprecise deployment, guide wire perforation, or intrastent clot formation were reported with Surpass stent;
malapposition was an important concern leading to additional manoeuvres in 19.4% of the study population.[11]
Significant technical issues were not reported for FRED.[12],[13] The immediate angiographic changes observed
within the aneurysm following FD include intrasaccular stagnation of contrast; however, the aneurysm progressively
thromboses in the follow-up period, achieving occlusion rates similar to that obtained following conventional
endovascular therapy.[2] The aneurysm obliteration rate achieved was high among the different FDs, which varied
between 73 and 83% at 6-month of follow up.[2],[4],[11],[12],[13],[14] The aneurysms also demonstrated higher
occlusion rates on long-term follow-up with significant reduction in the mass effect.[1] The treatment with FD is not
without complications; the mortality rates vary between 0 and 4.9% and the morbidity rates were reported to be
between 4 and 12% for different FDs.[2],[4],[11],[12] The periprocedural complications and mortality rates were
observed to be higher with SILK stents.[14] The long-term impact of a FD on parent vessel is not clearly delineated
and there are conflicting reports of high as well low incidence of in-stent stenosis after FD treatment. These lesions
are often asymptomatic; however, they may warrant a long-term clinical and angiographic surveillance to characterize
the progression and plan therapeutic modulation or further intervention.[15],[16]
The present study is the largest study from our country reporting the feasibility, immediate, and long-term outcomes
of FD treatment for complex intracranial aneurysms. One important disadvantage of our study is that the number of
subjects included in the analysis was low compared to other reports. This observation was primarily due to the fact
that the number of subjects ultimately willing for FD treatment among the referred patients was very less due to the
huge treatment costs involved, especially when multiple FDs were contemplated. In our study, the inclusion criteria
were homogeneous and all the patients underwent antiplatelet function tests to document the degree of platelet
activity suppression, and the aggressive usage of a potent antiplatelet drug such as prasugrel was considered only
when the response was inadequate. As point of care assays, such as multiplate assay or verifynow assay, are not
widely available in our country, we relied on the gold standard conventional light aggregometry tests to assess the
degree of suppression of platelet function.
In our series, procedural failure was noted in 2 patients, in whom the microcatheter could not be negotiated beyond
the aneurysm into the distal parent artery. The authors noted a few important observations regarding the 2 FD stents
used in this study. PED was found to be more sturdy while deployment with minimal proximal or distal migration,
which helped in accurate placement of the stent across the aneurysm. Although the distal end of the FRED is flared to
improve wall apposition, significant distal migration was noted during the deployment and while manoeuvring the
stent to improve stent expansion. The longer available lengths of the FRED and retrievability become important when
longer coverages are needed or catheter instability is contemplated. However, these observations stem from the
authors' modest experience and might reflect the steep learning curve associated with these procedures. The
permanent morbidity and mortality rates of the present series and other previously published larger studies are
compared in [Table 3].[1],[2],[4],[9],[10],[11],[12],[13],[14],[17],[18],[19],[20],[21],[22],[23],[24],[25],[26]
Thromboembolic complications are reported in 6.8% patients undergoing PED placement; however, we did not
observe any such events in our cohort. This may be related to the rigorous adherence to the protocol of performing
platelet function tests in all the patients considered for FD.[23] The periprocedural self-limiting non-neurological
hemorrhagic complications and vascular injuries were relatively high in this series (36%). Six minor periprocedural
events were observed in the present study, of which, 4 were directly related to the embolization and two were related
to the clinical consequence of the FD therapy. Aggravation or new onset oculomotor nerve palsy or trigeminal nerve
involvement in the postoperative period is an expected complication of giant cavernous internal carotid artery
aneurysms due to progressive thrombosis of the aneurysmal sac. Minor complications have rarely been a focus in
many series; however, Park et al.,[24] reported 28.6% of temporary complications with PED, of which 12.6% were
related to the procedure alone. Retroperitoneal hematoma is a dreaded complication with disastrous consequences,
and its early recognition is important to avoid a potential catastrophe. We routinely perform hemoglobin estimation 4
and 12 h after the procedure for its early recognition and to initiate further investigations such as ultrasound or CT
scanning, and blood transfusion. We believe that the probable cause of a retroperitoneal hematoma in uneventful
arterial access is slow extravasation from the puncture site or posterior wall of the artery, and hence, the artery is
regularly compressed for a while after securing the sheath to minimize the ongoing extravasation.{Table 3}
Significant obliteration of the aneurysm >90% was noted in 82% patients at a 1-month follow-up and progression to
complete occlusion at an intermediate follow-up was seen in 25% of patients. Asymptomatic branch vessel occlusion
was seen in 20% patients in our cohort, which is consistent with the reported literature. Our results show that the FD
treatment of giant aneurysms is technically feasible with minimal mortality and acceptable morbidity. The majority of
morbidities were transient and did not have any clinical impact when expeditiously managed. Although the procedural
complexity compared to the conventional endovascular technique is less, the deployment of FD is technically
challenging and has a steep learning curve, which will ultimately determine the outcome of the treatment.[25],[26]
Long-term angiographic data showed that the aneurysm obliteration was durable and there were no delayed stentrelated complications.
Conclusion
FD treatment of complex cerebral aneurysms was associated with favorable clinical and angiographic outcomes in the
intermediate and long-term follow up. Minor complications were common which needed to be effectively managed to
prevent major catastrophic events. The steep learning curve influenced the technical success of the procedure.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
References
1
Piano M, Valvassori M, Quilici L, Pero G, Boccardi E. Midterm and long-term follow-up of cerebral aneurysms
treated with flow diverter devices: A single-center experience. J Neurosurg 2013;118:408-16.
2
Brinjikji W, Murad MH, Lanzino G, Cloft HJ, Kallmes DF. Endovascular treatment of intracranial aneurysms with
flow diverters: A meta-analysis. Stroke 2013;44:442-7.
3
Alderazi YJ, Shastri D, Kass-Hout T, Prestigiacomo CJ, Gandhi CD. Flow diverters for intracranial aneurysms.
4
Stroke Res Treat 2014;2014:415653.
Leung GK, Tsang AC, Lui WM. Pipeline embolization device for intracranial aneurysm: A systematic review. Clin
5
Neuroradiol 2012;22:295-303.
Cherian MP, Yadav MK, Mehta P, Vijayan K, Arulselvan A, Jayabalan S. First Indian single center experience with
pipeline embolization device for complex intracranial aneurysms. Neurol India 2014;62:618-24.
6
Chalouhi N, Starke RM, Yang S, Bovenzi CD, Tjoumakaris S, Hasan D, et al. Extending the indications of flow
diversion to small, unruptured, saccular aneurysms of the anterior circulation. Stroke 2014;45:54-8.
7
de Barros Faria M, Castro RN, Lundquist J, Scrivano E, Ceratto R, Ferrario A, et al. The role of the pipeline
embolization device for the treatment of dissecting intracranial aneurysms. AJNR Am J Neuroradiol
8
9
2011;32:2192-5.
Aydin K, Arat A, Sencer S, Hakyemez B, Barburoglu M, Sencer A, et al. Treatment of ruptured blood blister-like
aneurysms with flow diverter SILK stents. J Neurointerv Surg 2015;7:202-9.
Fischer S, Vajda Z, Aguilar Perez M, Schmid E, Hopf N, Bäzner H, et al. Pipeline embolization device (PED) for
neurovascular reconstruction: Initial experience in the treatment of 101 intracranial aneurysms and dissections.
10
Neuroradiology 2012;54:369-82.
Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al. Flow-diverter silk stent for the treatment of
11
intracranial aneurysms: 1-year follow-up in a multicenter study. AJNR Am J Neuroradiol 2012;33:1150-5.
Wakhloo AK, Lylyk P, de Vries J, Taschner C, Lundquist J, Biondi A, et al. Surpass flow diverter in the treatment
12
13
14
15
of intracranial aneurysms: A prospective multicenter study. AJNR Am J Neuroradiol 2015;35:98-107.
Kocer N, Islak C, Kizilkilic O, Kocak B, Saglam M, Tureci E. Flow Re-direction endoluminal device in treatment of
cerebral aneurysms: Initial experience with short-term follow-up results. J Neurosurg 2014;120:1158-71.
Möhlenbruch MA, Herweh C, Jestaedt L, Stampfl S, Schönenberger S, Ringleb PA, et al. The FRED flow-diverter
stent for intracranial aneurysms: Clinical study to assess safety and efficacy. AJNR Am J Neuroradiol
2015;36:1155-61.
Murthy SB, Shah S, Shastri A, Venkatasubba Rao CP, Bershad EM, Suarez JI. The SILK flow diverter in the
treatment of intracranial aneurysms. J Clin Neurosci 2014;21:203-6.
Cohen JE, Gomori JM, Moscovici S, Leker RR, Itshayek E. Delayed complications after flow-diverter stenting:
Reactive in-stent stenosis and creeping stents. J Clin Neurosci 2014;21;1116-22.
16
John S, Bain M, Hui F, Hussain MS, Masaryk T, Rasmussen P, et al. Long-term follow-up of in-stent stenosis after
pipeline flow diversion treatment of intracranial aneurysms. Neurosurgery 2016;78:862-7.
17
Velioglu M, Kizilkilic O, Selcuk H, Kocak B, Tureci E, Islak C, et al. Early and midterm results of complex cerebral
18
aneurysms treated with Silk stent. Neuroradiology 2012;54:1355-65.
Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M. Early experience in the treatment of intra-cranial
aneurysms by endovascular flow diversion: A multicentre prospective study. PLOS One 2010;5:pii: e12492.
19
Saatci I, Yavuz K, Ozer C, Geyik S, Cekirge HS, Treatment of intracranial aneurysms using the pipeline flowdiverter embolization device: A single-center experience with long-term follow-up results. AJNR Am J
Neuroradiol. 2012;33:1436-46.
20
Yu SC, Kwok CK, Cheng PW, Chan KY, Lau SS, Lui WM, et al. Intracranial aneurysms: Midterm outcome of
pipeline embolization device—A prospective study in 143 patients with 178 aneurysms. Radiology
2012;265:893-901.
21
Briganti F, Napoli M, Tortora F, Solari D, Bergui M, Boccardi E, et al. Italian multicenter experience with flowdiverter devices for intracranial unruptured aneurysm treatment with periprocedural complications—A
22
retrospective data analysis. Neuroradiology 2012;54:1145-52.
Fischer S, Aguilar-Pérez M, Henkes E, Kurre W, Ganslandt O, Bäzner H, et al. Initial experience with p64: A
novel mechanically detachable flow diverter for the treatment of intracranial saccular sidewall aneurysms. AJNR
23
Am J Neuroradiol 2015;36:2082-9.
Tan LA, Keigher KM, Munich SA, Moftakhar R, Lopes DK. Thromboembolic complications with pipeline
embolization device placement: Impact of procedure time, number of stents and pre-procedure P2Y12 reaction
unit (PRU) value. J Neurointerv Surg 2015;7:217-21.
24
Park MS, Albuquerque FC, Nanaszko M, Sanborn MR, Moon K, Abla AA, et al. Critical assessment of
complications associated with use of the pipeline embolization device. J Neurointerv Surg 2015;7:652-9.
25
Siddiqui AH, Abla AA, Kan P, Dumont TM, Jahshan S, Britz GW, et al. Panacea or problem: Flow diverters in the
treatment of symptomatic large or giant fusiform vertebrobasilar aneurysms. J Neurosurg 2012;116:1258-66.
26
Lubicz B, Van der Elst O, Collignon L, Mine B, Alghamdi F. Silk flow-diverter stent for the treatment of
intracranial aneurysms: A series of 58 patients with emphasis on long-term results. AJNR Am J Neuroradiol
2015;36:542-6.
Monday, November 28, 2022
Site Map | Home | Contact Us | Feedback | Copyright and Disclaimer
Download