Service Manual Eclipse ABR and OAE Part number 80701116 / New no 8011775 Rev 1 Valid from 2012/09 – EP software version 4.0 Contents Contents ...................................................................................................................................................................... 1 Technical Specifications ............................................................................................................................................ 3 Eclipse Hardware Specifications ........................................................................................................................ 3 Specifications EP15/EP25 .................................................................................................................................. 4 Specifications TEOAE25 .................................................................................................................................... 6 Specifications DPOAE20 .................................................................................................................................... 6 Specifications ASSR........................................................................................................................................... 7 Specifications ABRIS.......................................................................................................................................... 8 Detachable Parts ...................................................................................................................................... 8 Accessories ................................................................................................................................................................. 9 Consumables .................................................................................................................................................... 12 Spare Parts ....................................................................................................................................................... 16 Schematics ................................................................................................................................................................ 17 ERA Board Block Diagram ............................................................................................................................... 17 Main Board Component Index Potentiometers and Measuring Points location .................................... 18 Advanced ERA Switch Board Block Diagram .................................................................................................. 19 Switch Board Component Index............................................................................................................. 20 Assembly – Spare Parts Index ................................................................................................................................ 21 Power Supply Description ................................................................................................................................ 22 Electrical Assembling .............................................................................................................................................. 23 Instrument Sketch............................................................................................................................................. 23 Rear Panel Connections................................................................................................................................... 25 Preamplifier ............................................................................................................................................................... 26 Block diagram ................................................................................................................................................... 26 OAE Probe Complete ....................................................................................................................................... 27 Calibration Procedure .............................................................................................................................................. 28 Service, adjustment and repair ......................................................................................................................... 28 Hardware Calibration ........................................................................................................................................ 28 EPx5/ABRIS Calibration ................................................................................................................................... 29 Stimuli Attenuator Calibration. ................................................................................................................ 29 Calibration of the Attenuator in Channel 1. ............................................................................................ 29 Control of the fixed Attenuator steps in the Attenuator Channel 1. ........................................................ 30 Control of the fixed Attenuator steps in the Attenuator Channel 2. ........................................................ 31 EPx5 Input section ................................................................................................................................. 32 Input Amplifier Control. ........................................................................................................................... 32 Control of the input filter. ........................................................................................................................ 33 EPx5 Stimuli Level ................................................................................................................................. 34 EPx5 Tone burst, Click, Chirp & NB Chirp calibration ........................................................................... 36 Select the Settings Tab .......................................................................................................................... 39 TEOAE 25 calibration ....................................................................................................................................... 40 Stimuli Attenuator Calibration. ................................................................................................................ 40 Calibration of the Attenuator in Channel 1. ............................................................................................ 41 Control of the fixed Attenuator steps in the Attenuator Channel 1. ........................................................ 41 Calibration of Attenuator in Channel 2. .................................................................................................. 42 Control of the fixed Attenuator steps in the Attenuator Channel 2. ........................................................ 43 OAE input amplifier. ............................................................................................................................... 43 TEOAE stimuli calibration. ..................................................................................................................... 44 Select the Settings Tab .......................................................................................................................... 45 DPOAE 20 calibration....................................................................................................................................... 46 Stimuli Attenuator Calibration................................................................................................................. 46 Calibration of the Attenuator in Channel 1. ............................................................................................ 46 Control of the fixed Attenuator steps in the Attenuator Channel 1. ........................................................ 47 Control of the fixed Attenuator steps in the Attenuator Channel 2......................................................... 48 OAE input amplifier. ............................................................................................................................... 49 DPOAE20 Microphone frequency response calibration......................................................................... 50 Select the Settings Tab. ......................................................................................................................... 51 Eclipse ABR and OAE Service Manual 1 ASSR calibration. ............................................................................................................................................. 52 Stimuli Attenuator Calibration. ................................................................................................................ 52 Calibration of the Attenuator in Channel 1. ............................................................................................ 52 Control of the fixed Attenuator steps in the Attenuator Channel 1. ........................................................ 53 Control of the fixed Attenuator steps in the Attenuator Channel 2. ........................................................ 54 ASSR Input Section................................................................................................................................ 55 Input Amplifier Control. ........................................................................................................................... 55 Control of the input filter. ........................................................................................................................ 56 ASSR Frequencies calibration. .............................................................................................................. 57 Select the Settings Tab .......................................................................................................................... 58 Trouble Shooting in TEOAE25, EP15 and EP25 Modules .................................................................................... 59 Trouble Shooting in ASSR Module ......................................................................................................................... 59 Trouble Shooting in the DPOAE Module ................................................................................................................ 60 Appendix U Update news ......................................................................................................................................... 61 Eclipse ABR and OAE Service Manual 2 Technical Specifications Eclipse Hardware Specifications Medical CE-mark: The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II of the Medical Device Directive 93/42/EEC. Approval of the quality system is made by TÜV – identification no. 0123. Standards: Safety: IEC 60601-1:2005 ANSI/AAMI ES60601-1:2005, Class I, Applied parts Type BF CAN/CAS-C22.2 No. 60601-1:08 EMC: IEC 60601-1-2 Test signal IEC 60645-3 Operation environment: Temperature: 15 – 35 °C Rel. Humidity: 30 – 90% Storing/handling: Temperatures below 0°C and above 50°C may cause permanent damage on the instrument and its accessories. Warm up time: 3 minutes at room temperature (20 °C). Designed to withstand standard transport methods, provided that original packing material and methods are used. Transport EPA Preamplifier: Two channels (standard) EPA4 Cable Collector (4 electrodes). Standard 50cm. Option:5cm or 290cm One channel (optional) EPA3 Cable collector (3electrodes) 50cm Gain: Frequency response: Noise: 80 dB/ 60dB (-20dB VEMP mode) 0.5Hz - 5kHz 4 nV/√Hz 0.22 µV rms (0 – 3 kHz) Minimum >118 dB. Typical 130dB <100Hz Typically 25dB improvement over previous available designs CMR Ratio: Radio Frequency Immunity Max input offset voltage: 2.5V Input impedance: 10 MΩ PC requirements: 1 GHz Pentium 3 class CPU (or better) with at least 1 GB ram, DVDROM drive, USB connector. Display: minimum 1024x768 and at least 40 GB free disk space. System RAM: 1 GB Supported Systems: Windows XP (SP2 required) , Windows Vista or Windows 7 (32bit) Data I/O USB Power Supply: Input Volts: Consumption: Construction: Dimensions: Weight: 100 – 240V, 50/60 Hz 60VA Metal cabinet (L x W x H) 28 x 32 x 5,5 cm 3,7 kg / 8 lbs excluding accessories Eclipse ABR and OAE Service Manual 3 Specifications EP15/EP25 Stimulus: Stimulus Stimulus rate: Click: Tone Burst Frequency: Tone Burst Stimulation Time: Burst Envelopes/windows: NB CE-Chirp CE-Chirp Click 200Hz-10kHz Polarity: Level: Masking: Masking Level: Transducer: Traditional Click, Burst, NB CE-Chirp, CE-Chirp, Click 200Hz10kHz. 0.3 to 80.1 stimuli per second in steps of 0.1 Hz Minimize interference add a random time every 10 stimuli. External trigger in, max speed 60 Hz. 100 µs electrical (IEC 60645-3 2007) 500, 750, 1000, 1500, 2000, 3000 and 4000 Hz. Stimulation up to 12 ms EP15 Stimulation up to 780 ms depending on the measurement type on EP25 Barlett, Blackmann, Gaussian, Hamming, Hanning, Rectangle and Manual (Rise/Fall and Plateau) 500, 1000, 2000 and 4000 Hz. 1 octave bandwidth. 175Hz-11200HZ Band limit click from 200 Hz to 10kHz. Condensation, Rarefaction, Alternating. 20 – 130 dB peSPL, (-10 – 100 dB nHL) in 1 dB steps. Air conduction White noise. Calibrated to SPL. 0 - -40 dB relative to stimulus level. Ear Tone ABR insert phone, calibrated on an IEC60318-4 coupler.(IEC711) Optionally a TDH39 with independent calibration can be supplied. calibrated on an IEC60318-1 Optionally a B71 bone conductor can be supplied. calibrated on an IEC 60318-6 Recording: Analysis Time: Pre recording time: Number of stimuli A/D Resolution Artefact Reject System: Gain: Dots per Trace: Analog High Pass Filter DSP Input low pass filter Display High Pass Filter: Display Low Pass Filter EP15: 0-30ms EP25: 0-980ms depending on the measurement type. EP15: -2.5 ms to + 2.5 ms. EP25 -150 ms to +150 ms depending on the measurement type. 20 to 32000 accepted recording. 16 bit. Standard peak voltage based system + two additional advanced rejection algorithms, in EABR disable rejection a possibility 10- 320 µV (74-104 dB.) Auto or Manual selection. 100 - 3200µV ( 54-84 dB) in VEMP mode (-20dB on preamplifier) 450 displayed. 0.5, 1.0, 3.3, 10, 33, 100 Hz/6dB octave 10, 100 Hz 12 dB octave 100, 300, 750, 1K, 1.5K, 2K, 3K, 4K, 5K, 7.5kHz 102 taps FIR filter 50,100, 150, 300, 500 Hz 6dB/octave IIR filter 0.83 500Hz 6 dB/octave depending on the measurement type. IIR Filter 1k, 1.5k, 2k 3k, 4k, 5k, 8k, 12kHz 33 taps FIR filter. 17 12kHz depending on the measurement type, 33 taps FIR filter. Controlled parameters: Data collection: Stimuli Rate, Number of stimuli, Polarity, Click, Tone Burst (Frequency, no. of sine waves, window), Stimulus intensity, Number of curves per intensity, Intensity (Ascending, Descending), Soft attenuator, Stimulus ear, Transducer, Masking level, Preliminary filter setting, Recording onset, Automatic next intensity (Wave repro level on screen), General Display Gain, Single Curve Display Gain, Baseline, Latency norm, Report templates, Print out, Manual stimulus to familiarisation, Talk Forward, Talk Back Monitor. Stop criteria, Protocol setup. Impedance test, Waveform buffer (A/B, Contra, Ipsi-Contra, A-B = Noise), Curve (Hide, Fixate, Merge, Delete), Online EEG, Waveforms storage in unlimited storage database. Eclipse ABR and OAE Service Manual 4 Reference Threshold survey The standardized calibration values (ISO 389-6 2007) for the TDH 39, ER-3A and B-71 transducers are given in Table I (text font Bold). The values for the interleaved frequencies 750, 1500 and 3000 Hz and the proposed tone burst BC-values (text font italics) are estimated. Click Tone bursts (2-1-2) repetition rate 20 Hz 750 Hz 1000 Hz 1500 Hz 2000 Hz 3000 Hz 4000 Hz 23.0 19.0 18.5 21.0 25.0 27.0 27.5 35.5 23.5 21.0 21.5 26.0 28.5 30.0 32.5 51.5 69.5 61.0 56.0 51.5 47.5 46.0 52.0 100 µs 500 Hz TDH 39 31.0 ER-3A B-71 Table I Reference calibration values in dB peSPL and dB peVFL A test report by PTB (Fedtke & Hensel, 2008) recommends peRETSPL values for the ER-3A to be used to calibrate the broad-band chirp and the four frequency specific chirps (NB-chirps) by Interacoustics 1. The corresponding values for the TDH 39 are estimated, and interim BC-values for the B-71 are estimated. A final set of calibration values for the TDH 39, ER-3A and B-71 transducers are proposed in Table II. Chirp NB-chirps repetition rate 20 Hz 500 Hz 1000 Hz 2000 Hz 4000 Hz TDH 39 27.5 26.0 21.5 27.0 30.0 ER-3A 32.0 26.5 24.5 30.5 35.0 B-71 51.5 75.0 61.5 50.0 55.5 Table II Reference calibration values in dB peSPL and dB peVFL 1 For NB-chirp calibration it is recommended that the pure tone frequency corresponds to the center frequency of the NB-chirp. Eclipse ABR and OAE Service Manual 5 Specifications TEOAE25 Stimulus: Stimulus rate: Transducer: Level: Polarity: Click: 12-20 ms dependent of stop time. OAP30 50 – 90 dB peSPL,1 dB steps. Linear or not linear click. 100 µs Analysis Time: Stop record time 12-20 ms. adjustable in 0.1ms step Start record time 0-12ms. adjustable in 0.1ms step 25-32000 selectable in step of 25 measuring 16 bit 25-55 dB RMS 500 to 6000Hz 5ms Recording: Number of measuring A/D Resolution Reject : Frequency response Probe Check Controlled parameters: Recording Begin, Recording end = Stimuli Rate, Stimuli (Linear not linear), Stimulus intensity, Stimulus ear, Wave repro level on screen, Display options (FFT, Octave, S/N ratio), Automatic stop criteria for screening, General Display Gain, Report templates, Print layout. Probe test, Waveform buffer (Total, A/B, A-B = Noise), Waveforms storage in unlimited storage database. Data collection: Specifications DPOAE20 The DPOAE system is an OAE instrument with input/outputs for OAP30 (Interacoustics OAE Probe). It can produce linear clicks for probe check and dual pure tone stimuli for measurement. Stimulus: Stimulus types: Stimulus Freq: Click rate Click wide Transducer: Level: Click stimuli and pure tone. 370 - 8000 Hz 25 ms 100 μS OAP30 30 – 75 dB SPL,1 dB steps. Analysis Time: 2 sec. - unlimited. adjustable in 1 sec. step below 2 min. and 1 min. steps up to 30 min (which is max limited recording time). One set of data is collected at 137 ms 16 bit Adjustable (-30 to 30 dB RMS) or off 240 to 8000Hz 25ms/40Hz 80 dB Recording: A/D Resolution Reject : Frequency response Probe Check Controlled parameters: Record begin / end, Stimuli frequency, stimuli level, frequency relation, Level relation, Test time, S/N ratio, Rejection level, Stimuli tolerance, DP close up graph on / off, Report template. Data collection: Probe test 2x256 samples (one for each transducer) DP test 3x4096 samples (three buffers in memory) Unlimited database memory Eclipse ABR and OAE Service Manual 6 Specifications ASSR Stimulus: Stim. rate: Transducer: Channels Level: Tone Burst Frequency: Bandwith Masking 40 or 90 Hz Ear Tone ABR insert phone, TDH39 headphones (Option) 2 0 – 100 dB nHL in 5 dB steps. 500, 1000, 2000, and 4000 Hz, both ears same time. 1 octave ± ½ octave – 3 dB White noise 0 – 100 dB nHL Recording: A/D Resolution Artefact Reject System: Gain: channels Accuracy method Gain Anti aliasing filter 6 minutes to detect a ASSR signal – can be extended up to 15 minutes 16 bit. Standard voltage based system 74 – 104 dB. Auto or Manual selection. 2, with separate detection algorithm 1% or 5%. Manual 5, 10, 20, 40, 80, 160, 320 uV input Analog 5kHz, 24 dB / octave USB Stimuli level and frequency Session status Expert or Basic View Customer selectable correction factors available General Display Gain , applicable during testing. Display Gain, applicable during testing. Independent control of up to 8 simultaneous stimuli (max 4 per ear) Independent start, stop control for each of the 8 stimuli Stimulus level control for each of the 8 stimuli False pass probability 1 or 5% Test protocols included for children and adult USB 1.1 Analysis Time: Display Display Gain: Controlled parameter Data I/O LBK15 is not valid for functional check of ASSR and ABRIS Reference Threshold survey Interacoustics use this value as reference threshold sound pressure level for NB broad-band chirp as normal hearing level (nHL) the calibration execution in peak-peak SPL calibration. Found as study Cebulla and Stürzebecher. NB-chirps repetition rate 90 Hz TDH 39 ER-3A 500 Hz 30.0 26.0 1000 Hz 25.0 23.0 2000 Hz 20.5 23.0 4000 Hz 22.0 25.0 Table III. Reference calibration values in dB peSPL NB-chirps repetition rate 40 Hz TDH 39 ER-3A 500 Hz 32.5 28.5 1000 Hz 27.5 25.5 2000 Hz 23.0 25.5 4000 Hz 24.5 27.5 Table IV. Reference calibration values in dB peSPL Eclipse ABR and OAE Service Manual 7 Specifications ABRIS Stimulus: Stim. rate: Transducer: Channels Level: Click: 93 Hz Ear Tone ABR insert phone, TDH39 headphones (Option) 1 30 – 35 - 40 dB nHL.(RETSPL = 30dB) 100 µs electrical (IEC 60645-3 2007) Analysis Time: A/D Resolution Artefact Reject System: Gain: channels Algorithmical sensitivity Specificity Anti aliasing filter Typical 20 seconds per ear. 16 bit. Standard voltage based system 10- 80 µV (86-104 dB) 2, with separate detection algorithm 99.9%. Typical 97% Analog 5kHz, 24 dB / octave in preamplifier USB Raw EEG curve or EEG level bar AEP wave form Algorithmical sensitivity over test time USB 1.1 Recording: Display Data I/O Detachable Parts • EPA PreAmplifier, EPA3 collector box or EP4 Collector box • Power Cable • ETB Standard Electrode cable kit • ETU Universal Electrode cable kit • ETR Electrode reuseable cable kit • Tube of Skin Preparation Gel • Set of Single Use Electrodes • EarTone ABR including Eartips • Operation Manual • Service Manual • CE Manual Eclipse ABR and OAE Service Manual 8 Accessories IA Part Number Description 80001101 TDH39 shielded HB7 headband New no. 8030457 80020004 Earphone cable IEC-M21 symmetric mono New no. 8010822 80041801 TDH39 telephone 10R Shielded New no. 8010865 With cushion 80042201 TDH39 Earphone shielded New no. 8010869 80053101 HB7 black with 2 yokes New no. 8010945 80044802 DD51 cushion New no 8010894 80001901 EARtone ABR Earphones 50 ohm New no. 8010808 80000701 B71 Bone Conductor Complete 30° plug New no. 8010800 80220501 B71 Bone Conductor only New no. 8011096 80220701 P3333 Headband for B71 New no. 8011098 Eclipse ABR and OAE Service Manual 9 IA Part Number Description 80220202 IEC-B11 conductor cable New no. 8011094 80001001 DT48 A Headset New no. 8010803 106700101 New no. 8013641 Preamplifier EPA4 Cable connector 500 mm 106700401 New no. 8013644 Preamplifier EPA4 Cable connector 500mm and 50mm 106700201 Preamplifier EPA4 US – without VEMP New no. 8013642 Cable connector 500mm 106700501 Preamplifier EPA4 US – without VEMP New no. 8013645 Cable connector 500mm and 50mm 106700301 Preamplifier EPA3 Cable collector 500 mm New no. 8013643 80451502 Accessory kit EP15 New no. 8011376 80451402 Accessory kit EP25 New no. 8011375 82416801 TM Electrode Starter Kit New no. 8013209 Eclipse ABR and OAE Service Manual 10 IA Part Number Description 80428401 ETU universal electrode cable kit New no. 8011344 80428601 ETR Electrode cable reusable kit New no. 8011345 80428801 ETB standard electrode cable kit New no. 8011346 80428201 ETT Electrode cable kit New no. 8011343 80429201 ETS Electrode spring kit New no. 8011349 80429501 ETSE Tap surface electrode kit New no. 8011352 New no.8500410 Pinch clip cables 80311101 OAE Probe complete New no. 8011175 82418601 Optical USB 1.1 isolation MED New no. 8013227 80470101 Mains cable E/F Smart plug New no. 8011384 80470301 Mains cable B North America 2.5m New no. 8011386 Smart plug Transparent Mains cables country specific Mains cables country specific 804700xx New no. 8011383 DK 804702xx New no. 8011385 UK 80470401 New no. 8011387 IT Eclipse ABR and OAE Service Manual 11 IA Part Number Description 804705xx New no. 8011388 AU 804706xx New no. 8011389 CH 804707xx New no. 8011390 AR 804718xx New no. 8011401 BR 80407702 Cable USB 2m New no. 8011241 80408901 LBK15 loop Back Interacoustic New no. 8011252 Consumables IA Part Number Description 81402104 Assortment box BET 25 New no. 8012838 TEOAE/DPOAE Including probe cleaning tool 81402402 Assortment box BET 60 New no. 8012841 SPL AUD 80429101 Probe cleaning tool New no 8011348 Assembled / open 341357xx New no 8002028 341358xx New no 8002029 34135901 New no 8002030 Eartip flanged 3-5mm red Eartip flanged 4-7mm blue Eartip flanged 5-8mm yellow 341347xx New no 8002018 Eartip 7mm mushroom blue 341348xx New no 8002019 Eartip 8mm mushroom yellow 341349xx New no 8002020 Eartip 9mm mushroom green Eclipse ABR and OAE Service Manual 12 IA Part Number Description 341350xx New no 8002021 Eartip 10mm mushroom red 341351xx New no 8002022 Eartip 11mm mushroom blue 341352xx New no 8002023 Eartip 12mm mushroom yellow 341353xx New no 8002024 Eartip 13mm mushroom green 341354xx New no 8002025 Eartip 14mm mushroom red 341355xx Eartip 15mm mushroom blue New no 8002026 341356xx New no 8002027 Eartip 19mm mushroom yellow 81426201 Probe tip kit OAE Ø3,3 8 pcs New no. 8012960 For OAE Probe 80452101 Accessory kit TEOAE25/DPOAE20 containing: New no. 8011380 81402104 Assortment box BET 25 81426201 Probe tip kit OAE Sanibel no. 850090 Standard insert foam tips 3A Medium yellow 50 pcs Sanibel box of 750 pcs. No. 8500120 Sanibel no. 8500110 Pediatric insert foam tips 3B Small beige 50 pcs Sanibel box of 750 pcs. No. 8500130 Sanibel no. 8500100 Jumbo insert foam tips 3C Large yellow 24 pcs 81402001 TTE25 Tip Trode Electrode New no. 8012837 Eclipse ABR and OAE Service Manual 13 IA Part Number Description Sanibel 8500440 Gold Tip Trode, 10mm 80495801 Disposable tap electrodes 36 pcs Sanibel no 8011491 ETSE electrode 80495901 Disposable snap electrodes 36 pcs Sanibel no 8011492 ETB electrode Sanibel no. 80495601 TM Electrode for ECochG Box of 10 pcs. 81402701 Infant ear tips INE 3,5mm Bag of 10 pcs. New no. 8012843 81402801 Infant ear tips INE 4,0mm Bag of 10 pcs. New no. 8012844 Sanibel no.8500140 Tube nipples for Eartone 3A 81400301 SPG15 Skin Preparation Gel Sanibel no 8500010 Sanibel: NuPrep gel 4oz/114g tube 81400501 Sonavelle ® Electrode Gel *not for US Sanibel no 8500380 Sanibel: Sonavelle gel 81400801 Alcohol Pads 100 pcs. Sanibel no 8500050 Sanibel: Alcohol pads Eclipse ABR and OAE Service Manual 14 IA Part Number Description 34815301 Gauze swabs 8 layer 75x75mm Package of 100 pcs. Sanibel no 8500040 Sanibel: Gauze swabs 81400201 PEG15 Electrode Pads Sanibel no 8500060 Sanibel: Pregel foam snap elec (PEG15) 81400401 Ten-20 Electrode Gel Sanibel no 8500070 Sanibel: Ten-20 Electrode paste Eclipse ABR and OAE Service Manual 15 Spare Parts Position Part index IA Part number. Description 36308101 Potentiometer 8 Power supply 60W 16 73907701 Power switch 1858.1103 23 New no. 8009908 Switch 10A ON-NONE-OFF 75102301 8 pol DIN connector 25 PCA Main board 1101 28 PCA Advanced switch board 29 3 pol power inlet 52 3 pin stereo jack connector 55 6 pin stereo jack connector 56 New no. 8003653 50001803 New no. 8006026 New no. 8010093 51004312 New no. 8006145 51005304 New no. 8006157 75831901 New no. 8010383 75000301 New no. 8010036 75000401 New no. 8010037 Eclipse ABR and OAE Service Manual 16 Schematics ERA Board Block Diagram ERA Main ERA Board Board block block diagram diagram Reset Reset (pc7) (pc7) Talk TalkForward Forward Right Right MP6 MP6 MP2 MP2 120 120dB dB dynamic dynamic Attenuator Attenuator CH1 CH1 FL0 FL0 FL2 FL2FL1 FL1FL0 FL0 Left Leftinin DSP DSP22 ADSP2188 ADSP2188 Right Rightinin Audio Audio Codec Codec PCM PCM3001 3001 Sport Sport00 Sport Sport11 clk clk pc1 pc1 FL1 FL1 Extend Extend ranges ranges +20 +20dB dB CH1 CH1 20 20dB dBFix Fix Attenuator Attenuator CH1 CH1 Output Output switch switch CH1 CH1 pa0-6 pa0-6 MP5 MP5 MP3 MP3 120 120dB dB dynamic dynamic Attenuator Attenuator CH2 CH2 Bone Bone FL2 FL2 Extend Extend ranges ranges +20 +20dB dB CH2 CH2 20 20dB dBFix Fix Attenuator Attenuator CH2 CH2 Output Output switch switch CH2 CH2 pb0-6 pb0-6 AUX AUX Out Out SP1 SP1 Dual Dual DAC DAC AD AD7249 7249 30kHz 30kHz Sample Sample Left Left SP2 SP2 To ToEclipse Eclipse instrument instrument Switch Switch board board 15,36 15,36 MHz MHz pa7 pa7 DSP1 DSP1-DSP -DSP22Control Control pb7 pb7 pc3 pc3 To ToEclipse Eclipse instrument instrument Switch Switch board board USB_B USB_B Address Address Data DataBus Bus Host HostData DataBus Bus Infineon Infineon C165 C165UTAH UTAH MP10 MP10 CO1.1 CO1.1 Left_input Left_input Left Left Filter FilterBank Bank Left Left Input InputAtt. Att. To ToEclipse Eclipse instrument instrument Switch Switchboard board CO1.2 CO1.2 Right_input Right_input clk clk pc4-6 pc4-6 MP13 MP13 Right Right Filter FilterBank Bank Host HostAddress Address Bus Bus pc2 pc2 ADC Sport ADC Sport00 AD AD1877 1877 Trigger TriggerOut Out DSP1 DSP1 ADSP2188 ADSP2188 Trigger TriggerInIn Right Right Input InputAtt. Att. FLASH FLASH 11/ /88MB MB RAM RAM 256K 256K/ /1MB 1MB DSP1 DSP1FL0-2 FL0-2 Project ProjectName Name Document DocumentName Name Author Author Current Currentrevision revision Revision Revisiondate date File FileLocation Location ERA ERABoard BoardPlatform Platform ERA ERABoard BoardBlock Blockdiagram diagram IbIbPedersen Pedersen 1.0 1.0 2005/08/25 2005/08/25 C:\Project\ERA C:\Project\ERAPlatform PlatformLine\Hardware\ Line\Hardware\ ERA ERAPlatform\Design Platform\DesignHistory HistoryFiles\Design Files\Design and andImplementation\PCB Implementation\PCBDesign Design Eclipse ABR and OAE Service Manual 17 Main Board Component Index Potentiometers and Measuring Points location D01009002 Eclipse ABR and OAE Service Manual 18 Trigger Out 400Hz Mic. gain 30dB 0dB Probe for OAE Probe Mic 4 orden 12KHz lowpassfilter 40dB 2 orden 350Hz highpassfilter 2 orden 450Hz highpassfilter MP7 Echo Out MP6 Sp1 100Hz Mic. gain Stimulus out 400Hz From DSP board MP2 Sp2 4 orden 100Hz highpassfilter Talk Back Stimulus out 100Hz Monitor Gain adj L-V ISO Amp. MP4 Left_in on DSP Board L-V +5V ISO +12V ISO GNG ERA Pre. amp. Socket 1.5W DC/DC Input select ISO -12V MP1 Gain adj R-V GND VEMP Right_in on DSP Board R-V ISO Amp. GND Pa7 pc3 Pb7 MP5 Date Constr. GND Revision Title Constr. 101108 Appr. PSC - TLL - Sheet 1 of 1 Block diagram Switch board Adv. 101 Partno. Drawingno. NP = Not Present D02024301 Eclipse ABR and OAE Service Manual 19 Switch Board Component Index D01000103 Eclipse ABR and OAE Service Manual 20 Assembly – Spare Parts Index 56 55 25 8 16 28 29 23 52 For part numbers see spare part list on page 16 Eclipse ABR and OAE Service Manual 21 Power Supply Description Power Supply Unit 60W XP p/n ECM60UT31>2449 input 100-240VAC Interacoustics Production Code 50001803 Input Output J1 Connector J1 J1 Pin 1 2 Signal Line Neutral Cable No./ Wire Colour 14 / Blue 14 / Brown J2 J2 J2 J2 J2 J2 1 2 3 4 5 6 +5V, 2,5A +5V, 2,5A GND GND -12V, 1A +12V, 3A 9 / Yellow 9./ Orange 9 / Black 9 / Black NC 9 / Red J2 Eclipse ABR and OAE Service Manual 22 Electrical Assembling Instrument Sketch I06001705 Note! Cable 5 wires must be held together with cable binder. See picture on page 2 Details for rear panel assembling see drawing I060018xx Power switch Power inlet Cable 10 BLU Cable 10 Cable 5 1 CO3 CO10 CO5 Cable 12 Connected to Top profile BLK Cable 1 CO4 Cable 2 Bottom BRN BLU Cable 5 to DIN (Preamplifier) YEL CO13 CO2 BLU Power Supply PCA Cable 14 IC1 RED BRN 5 GRN 3 7 2 8 4 ORG 1 6 BLU Cable 13 BLU CO6 1 CO7 GRN YEL ORG RED BRN V1 CO8 1 RED WHT CO11 BLK Cable 8 RED Cable 3 RED BRN CO2 ORG BLK BLK RED CO12 1 BRN CO5 Cable 9 1 Cable 6 CO4 BRN BRN Cable 2 Cable 14 to bottom tags Cable 11 VIO Top BLU Sign.1 YEL/GRN Sign.2 Pat Resp ODU PCB Talk Back SW CO9 CO8 CO7 CO3 CO6 BRN 1 CO1 Cable 9 Cable 16 Wires for LED are fixed to solder tag and rotary knob with cable binder Note! 6 cable binders are mounted as close as possible Eclipse ABR and OAE Service Manual 23 Rear panel – DIN connector Strip wires tight Eclipse ABR and OAE Service Manual 24 Rear Panel Connections Top Talk Back Cable 7 Cable 15 Plug Power inlet Cable 10 Top BLU BRN 2 Cable 17+ Cable 19 Cable 18 Bottom BRN BLU Power supply PCA cable 14 VIO Pat.Resp. BLUE Signal 1 (RED) Signal 2 (BLUE) Power switch Cable 5 Talk Forw. Talk Back 5 36 Bottom Switch board adv. CO8 Jumper SW Signal 1 (RED) Signal 2 (BLUE) Cable 15 2x Signal 1 (RED) 2x SW Trigger In/out 2x Signal 2 (BLUE) Cable 17 + Cable 18 Signal (Blue) Talk Forw. SW Cable 5 DIN socket (Preampl.) YEL RED BRN 5 GRN 3 7 2 8 4 1 6 Mainboard CO5 BLU VIO BLUE Jumper Mainboard CO2 SW Cable 17 Cable 7 Signal 2 (BLUE) SW Signal 1 (RED) Signal (Blue) Cable 19 SW Signal (BLUE) Cable 18 Connectors View from soldering side Switch board adv. CO5 Date Constr./ Draft Construction 06.11.03 BLC EN Revision 21.10.08 BLC TLL Title Appr. Sheet 1 of 1 Scale 1:1 A3 ECO: 2008-0182 Eclipse Instrument rear panel Assembly wires and cables Part no. 9101 Drawing no. I06001803 QM0100401 - A4 landscape instruction, 15.03.06/SJ Eclipse ABR and OAE Service Manual 25 Offset detect L 0dB Right 0dB 0dB Amplifier / Impedance switch Vertex HF input filter & Clamp circuit Diff Amp 30dB DRL 5KHz LP Filter Left out Left V R 30dB 180º Add Diff Amp 30dB Signal switch Left Square wawe gen. 5KHz LP Filter Right out Right Gain select 30dB / 50dB Impedance measurement circuit Date Constr. Appr. 101108 PSC TLL Sheet Revision 220909 PSC TLL 1 of 1 Constr. Title Block Diagram Preamplifier 067 Partno. Drawingno. NP = Not Present D02024202 Eclipse ABR and OAE Service Manual 26 OAE Probe Complete Item no see accessories list page 9 Disassembling the connector Cable Connections pin6 YEL pin1 RED pin2 BLU pin5 BLK pin3 SW pin7 GRN pin4 WHT OAE Cable length 3m Eclipse ABR and OAE Service Manual 27 Calibration Procedure Service, adjustment and repair This instrument can be serviced, adjusted and repaired without loosening the validity of the CE-marking provided the measuring equipment used is fulfilling the demands below, the adjustment procedures are followed and the personnel is having the necessary qualifications approved by Interacoustics. Equipment for adjustment In order to assure correct calibration of the audiometric equipment, it is important that the necessary measuring equipment is present and that it is reliable and stable. Minimum requirements for measuring equipment: Measuring amplifier with input for condenser microphone or a sound level meter, fulfilling the demands of IEC 651 Type 1. Acoustical coupler according to IEC60318-4 (IEC 711) Occluded ear simulator, IEC60318-1 artificial ear, IEC60318-6 artificial mastoid . Alternatively a 2cc coupler or 6ccm can be used, provided that it fulfils the demands of IEC 60318-5 (IEC 126) or IEC60318-3 (IEC303) and that the corresponding corrections are applied. Furthermore it is recommended to have an acoustic calibrator for the control of the complete measuring chain. General purpose frequency counter. General purpose digital multimeter. General purpose signal generator. General purpose oscilloscope Audio spectrum analyzer or distortion meter Hardware Calibration There are 2 print boards to be calibrated. The Advanced Switch Board and the ERA Board, Switch off the main power and dismount the Top profile. Switch on power. The calibration is separated in parts: the EPx5, TEOEAE 25, DPOAE 20, ASSR part. There are four different software modules to this hardware and some of the calibrations are identical for the different modules, it is not necessary to calibrate the same hardware part more than one time. Eclipse ABR and OAE Service Manual 28 EPx5/ABRIS Calibration Switch on power and activated the one of follow calibration program. The calibration program for the four different software modules can be found on the following paths: The EPx5 Version 4.xx Calibration program can be found in “C:\Program Files\Interacoustics\Eclipse\EPx5 4.00\EPx5 Calibration.exe” Activate the icon for the desired Calibration Application and type in the password if necessary. Stimuli Attenuator Calibration. The Tab called Stimuli attenuator, uses to calibrated to output attenuator. All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Calibration of the Attenuator in Channel 1. Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 1 is selected in “Channel to calibrate” area on the display. Dynamic Attenuator 0 dB 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP6. 2. Adjust P6 to minimum Total Harmonic Distortion. 3. Adjust P3 until the AC voltmeter shows 5.3VRMS (+14.5 dBV) ± 0.1 dB. 4. Press the – 60-dB button in the “Dynamic Attenuator” section. Dynamic Attenuator -60 dB 1. Adjust the P2 until the AC voltmeter shows 5.3mVRMS (-45.5 dBV) ± 0.1 dB. 2. Press the – 120 dB key in the “Dynamic Attenuator” section. Eclipse ABR and OAE Service Manual 29 Dynamic Attenuator -120 dB Notice: This Point can only be performed by using an AC voltmeter with 1000 Hz band pass filter or a spectrum analyzer. 1. Move the slider in the bottom of the “Dynamic Attenuator” section until the AC voltmeter or spectrum analyzer shows 5.3µVRMS (-105.5 dBV) ± 0.5 dB. 2. Select the 0 dB button in the “Dynamic Attenuator” section and scroll down to “Fix attenuator test” section. Control of the fixed Attenuator steps in the Attenuator Channel 1. Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP6. 2. Selected the -5 dB button in the “Fixed Attenuator” section. Fix Attenuator -5 dB 1. Verify that the AC voltmeter shows 2.99VRMS (9.5 dBV) ± 0.5 dB. 2. Select the -10 dB button in the “Fixed attenuator” section. Fix Attenuator -10 dB 1. Verify that the AC voltmeter shows 1.68VRMS (4.5 dBV) ± 0.5 dB. 2. Select the -15 dB button in the “Fixed attenuator” section. Fix Attenuator -15 dB 1. Verify that the AC voltmeter shows 944 mVRMS (-0.5 dBV) ± 0.5 dB. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Verify that the AC voltmeter shows 531 mVRMS (-5.5 dBV) ± 0.5 dB. 2. Select the Channel 2 button in the “Channel to calibrate” section. Calibration of Attenuator in Channel 2. All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Verify Channel 2 is selected in “Channel to calibrate” area on the display. Eclipse ABR and OAE Service Manual 30 Dynamic Attenuator 0 dB 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP5. 2. Adjust P5 to minimum THD. 3. Adjust P4 until the AC voltmeter shows 5.3VRMS (+14.5 dBV) ± 0.1 dB. 4. Press the – 60-dB key in the “Dynamic Attenuator” section. Dynamic Attenuator -60 dB 1. Adjust the P1 until the AC voltmeter shows 5.3mVRMS (-45.5 dBV) ± 0.1 dB. 2. Press the – 120 dB key in the “Dynamic Attenuator area” section. Dynamic Attenuator 120 dB Notice: This Point can only be performed by using an AC voltmeter with 1000 Hz band pass filter or a spectrum analyzer. 1. Move the slider in the bottom of the “Dynamic Attenuator” section until the AC voltmeter or spectrum analyzer shows 5.3µVRMS (-105.5 dBV) ± 0.5 dB. 2. Select the 0 dB button in the “Dynamic Attenuator” section and scroll down to Fix attenuator test section. Control of the fixed Attenuator steps in the Attenuator Channel 2. Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP5. 2. Selected the -5 dB buttom in the “Fixed Attenuator” section. Fix Attenuator -5 dB 1. Verify that the AC voltmeter shows 2.99VRMS (9.5 dBV) ± 0.5 dB. 2. Select the -10 dB button in the “Fixed attenuator” section. Fix Attenuator -10 dB 1. Verify that the AC voltmeter shows 1.68VRMS (4.5 dBV) ± 0.5 dB. 2. Select the -15 dB button in the “Fixed attenuator” section. Fix Attenuator -15 dB 1. Verify that the AC voltmeter shows 944 mVRMS (-0.5 dBV) ± 0.5 dB. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Verify that the AC voltmeter shows 531 mVRMS (-5.5 dBV) ± 0.5 dB. Eclipse ABR and OAE Service Manual 31 EPx5 Input section Select the EPx5 calibration program and select the Input Attenuator tab. All the measurements and adjustments are made on the Switch Board board. All the measure point’s (MP) and potentiometers (P) refers to the Advanced Switch Board Component Index page 20 All measurements are made with reference to ground MP1. 1. Connect a sine wave generator pre selected to 1000 Hz and 1.00 Volt rms. to the Left-Vertex input pin on the preamplifier input socket (pin 5) and ground to (pin 2). 2. Adjust P8 until the EPx5 display shows 100 µVolt ± 1µVolt on the left channel in position 80 dB. 3. Connect a sine wave generator pre selected to 1000 Hz 1.00 Volt Rms. to the Right-Vertex input pin on the preamplifier-input socket (pin 4) and generator ground to (pin 2). 4. Adjust P9 until the EPx5 display shows 100 µVolt ± 1µVolt on the right channel in position 80 dB. Input Amplifier Control. 104 dB Amplification. 1. Connect a sine wave generator pre-selected to 1000 Hz, 70.7 mV RMS (100mV peak) to the preamplifier socket with 8 pins, in the Eclipse. Make a short circuit between pin 4 and pin 5, and connect these two pins to the sine wave signal. The ground of the sine wave generator, shall be connected to preamplifier socket, pin 2. 2. Verify that the Left and Right input level meter in the bottom of the display shows 0.0 dB ± 0.5dB. 3. Select the 98 dB button in the “Amplification” section. 98 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -6.0 dB ± 0.5 dB. 2. Select the 92 dB button on the “Amplification” section. 92 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -12.0 dB ± 0.5 dB. 2. Select the 86 dB button on the “Amplification” section. 86 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -18.0 dB ± 0.5 dB. 2. Select the 80 dB button on the “Amplification” section. Eclipse ABR and OAE Service Manual 32 80 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -24.0 dB ± 0.5 dB. 2. Select the 74 dB button on the “Amplification” section. 74 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -30.0 dB ± 0.5 dB. 2. Select the 68 dB button on the “Amplification” section. 68 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -36.0 dB ± 0.5 dB. 2. Select the 62 dB button on the Fixed “Amplification” section. 62 dB Amplification. 1. Verify that the Left and Right input level meter in the bottom of the display shows -42.0 dB ± 0.5 dB. Control of the input filter. Select 62 dB amplification in Input Att. tap. 1. Verify the input filter. Keep the same connection of the sine wave generator as during the “Input Att. Amplification test”. 2. Verify that the left and right level meter in the bottom of the display shows -42dB ± 0.5dB. 3. Change the sine wave generator frequency to the selected filter frequency. 4. Verify that the left and right input level meter in the bottom of the display shows –45dB ± 2dB. 5. Repeat point 3 4 for all 8 filter. 6. Connect an EPA preamplifier model, to the 8-pin connector on the Eclipse, with a VEMP -20dB push button supplied. 7. Connect the EPA preamplifier via the EPA4 cable adaptor to the LBK15 loop back box. 8. Connect the 6.3mm jack plug on the LBK15 to a sine wave generator pre-selected to 1000Hz 70.7mV RMS (100mVp). 9. When done, select “Turn VEMP mode on, in Preamp”. 1. Verify that the Left and Right input level meter in the bottom of the display shows -20 dB ±0.5 dB, and that the blue -20 dB LED will turn on, on the preamplifier. Eclipse ABR and OAE Service Manual 33 EPx5 Stimuli Level Activate the EPx5 calibration program and select the Stimuli Level tab (Click Tradit, WN, TF, Monitor Tone). Calibration of the stimulus level is done in peak-to-peak equivalent sound pressure peSPL or vibratory force level peVFL, see figure. peak-peak Select Right/Left ABR3A for calibration 1. Connect the EAR ABR 3A insert phone to the Occluded ear simulator, which is connected to a sound level meter (SLM), and an oscilloscope. 2. Adjust the slider for an output in the coupler corresponding to 100 dB SPL on the SLM. Note the peak-peak voltage V1 on the oscilloscope display. 3. Select the Click Traditional button in the “Stimuli type” section. ABR3A Click Traditional 1. Adjust the slider until the oscilloscope shows the same peak-peak voltage as the noted V1. 2. Select the White noise button (WN) in the “Stimulus type” section. ABR3A WN 1. Adjust the slider for an output in the coupler corresponding to 100 dB SPL on the SLM. 2. Select the Talk Forward in the Stimulus type” section. ABR3A Talk Forward 1. Connect a Talk Forward microphone and talk at normal level in the microphone and adjust the slider for an output in the coupler corresponding to 100 dB SPL on the SLM. 2. Select the Monitor Tone in the “Stimulus type” section. Eclipse ABR and OAE Service Manual 34 ABR3A Monitor Tone 1. Adjust the slider for an output in the coupler corresponding to 100 dB SPL on the SLM. 2. Select the next transducer that needs to be calibrated. Select Right/Left TDH 39 for calibration. 1. Connect the TDH 39 phone to the artificial ear, which is connected to a sound level meter (SLM), and an oscilloscope. 2. Adjust the slider for an output in the artificial ear corresponding to 100 dB SPL on the SLM. Note the peakpeak voltage V1 on the oscilloscope display. TDH 39 Click Traditional 1. Adjust the slider until the oscilloscope shows the same peak-peak voltage as the noted V1. 2. Select the White noise button (WN) in the “Stimulus type” section. TDH 39 WN 1. Adjust the slider for an output in the artificial ear corresponding to 100 dB SPL on the SLM. 2. Select the Talk Forward in the Stimulus type” section. TDH 39 Talk Forward 1. Connect a Talk Forward microphone and talk at normal level in the microphone and adjust the slider for an output in the artificial ear corresponding to 100 dB SPL on the SLM. 2. Select the Monitor Tone in the “Stimulus type” section. TDH 39 Monitor Tone 1. Adjust the slider for an output in the artificial ear corresponding to 100 dB SPL on the SLM. 2. Select the next transducer that needs to be calibrated. Select Bone for calibration. 1. Connect the Bone conductor to the artificial mastoid, which is connected to a sound level meter (SLM), and an oscilloscope. 2. Adjust the slider for an output in the artificial mastoid corresponding to 80 dB VFL on the SLM. Note the peak-peak voltage V1 on the oscilloscope display. Bone Click Traditional 1. Adjust the slider until the oscilloscope shows the same peak-peak voltage as the noted V1. 2. Select the White noise button (WN) in the “Stimulus type” section. Bone WN 1. Adjust the slider for an output in the artificial mastoid corresponding to 80 dB VFL on the SLM. 2. Select the Talk Forward in the Stimulus type” section. Bone Talk Forward 1. Connect a Talk Forward microphone and talk at normal level in the microphone and adjust the slider for an output in the artificial mastoid corresponding to 80 dB VFL on the SLM. 2. Select the Monitor Tone in the “Stimulus type” section. Bone Monitor Tone 1. Adjust the slider for an output in the artificial mastoid corresponding to 80 dB VFL on the SLM. 2. Select the next transducer that needs to be calibrated. Eclipse ABR and OAE Service Manual 35 EPx5 Tone burst, Click, Chirp & NB Chirp calibration Select the Tone burst, Click, Chirp &NB Chirp tab in the EPx5 Calibration program. Calibration of the stimulus level is done in peak-to-peak equivalent sound pressure peSPL or vibratory force level peVFL. It’s also possible to set the distance between pe SPL and nHL for the selected Tone Burst. Tone Ref. 1kHz ref signal Example on the acoustics measure signal on a EAR ABR 3A insert phone “TDH 39 phone to the artificial ear” “Bone conductor to the artificial mastoid” 1. Connect the “EAR ABR 3A insert phone to the Occluded ear simulator”, which is connected to a sound level meter (SLM), and an oscilloscope. 2. Select the frequency that needs to be calibrated. “artificial ear corresponding to 100 dB SPL” “mastoid corresponding to 80 dB VFL” 3. Adjust the slider for an output in the “coupler corresponding to 100 dB SPL” on the SLM. Note the peakpeak voltage V2 on the oscilloscope display. 4. Select Burst in the “Stimuli” section, and “Level (peSPL)” to 100 dB. Burst 500Hz 1000Hz 2000Hz 4000Hz Example on the acoustics measure signal on a EAR ABR 3A insert phone 1. Verify that the same frequency is selected as during “Tone ref.” calibration before starting the calibration of the Tone Burst. 2. Adjust the slider until the oscilloscope shows the same peak-peak value as V2 from Tone Ref calibration. 3. If the current frequency for the calibration is 500, 1000, 2000 or 4000 Hz, select NB CE-Chirp® in the “Stimuli” section – otherwise select “Tone Ref” again. Eclipse ABR and OAE Service Manual 36 NB CE-Chirp® 500Hz 1000Hz 2000Hz 4000Hz Example on the acoustics measure signal on a EAR ABR 3A insert phone 1. Verify that the same frequency is selected as under Tone ref. calibrations before starting to calibrate the NB CE-Chirp. 2. Adjust the slider until the oscilloscope shows the same peak-peak value as V2 from Tone Ref calibration. 3. If the current frequency for the calibration is 1000 Hz, select Click 200Hz-10kHz in the “Stimuli” section – otherwise select “Tone Ref.” again. Click 200Hz-10kHz Click 200Hz-10kHz Example on the acoustics measure signal on a EAR ABR 3A insert phone 1. This calibration expects that the Tone ref. at 1000Hz is calibrated. 2. Adjust the slider until the oscilloscope shows the same peak-peak value as V2 from Tone Ref when calibrated at 1000Hz. 3. Select CE-Chirp® in the “Stimuli” section. CE-Chirp® CE-Chirp Example on the acoustics measure signal on a EAR ABR 3A insert phone 1. This calibration expects that the Tone ref. at 1000Hz is calibrated. 2. Adjust the slider until the oscilloscope shows the same peak-peak value as V2 from Tone Ref when calibrated at 1000Hz. 3. Select Tone Ref. in “Stimuli” section. peSPL nHL Important this section is only to uses if you wanted to find special peRETSPL or changes Interacoustics standard value see General Technical specifications section survey of reference and max hearing level. In this section uses to find you own peRETSPL or adjust the peRETSPL value to standard value for the different stimulus type Burst, NB CE-Chirp®, Click 200Hz-10kHz, CE-Chirp® and depending of test type. Important if there changes on the peRETSPL that will changes stimulus level for measuring in nHL dB and give other latency time for the measuring. Eclipse ABR and OAE Service Manual 37 The standardized calibration values (ISO 389-6 2007) for the TDH 39, ER-3A and B-71 transducers are given in Table I. The values for the interleaved frequencies 750, 1500 and 3000 Hz and the proposed tone burst BC-values (shown in italics) are estimated. Click TDH 39 ER-3A B-71 100 µs 200-10kHz 31.0 Tone bursts (2-1-2) repetition rate 20 Hz 500 Hz 750 Hz 1000 Hz 1500 Hz 2000 Hz 3000 Hz 4000 Hz 23.0 19.0 18.5 21.0 25.0 27.0 27.5 35.5 23.5 21.0 21.5 26.0 28.5 30.0 32.5 51.5 69.5 61.0 56.0 51.5 47.5 46.0 52.0 Table I Reference calibration values in dB peSPL and dB peVFL A test report by PTB (Fedtke & Hensel, 2008) recommends peRETSPL values for the ER-3A to be used to calibrate the broad-band chirp and the four frequency specific chirps (NB-chirps) by Interacoustics 2. The corresponding values for the TDH 39 are estimated, and interim BC-values for the B-71 are estimated. A final set of calibration values for the TDH 39, ER-3A and B-71 transducers are proposed in Table II. NB-chirps repetition rate 20 Hz Chirp TDH 39 ER-3A B-71 27.5 32.0 500 Hz 26.0 26.5 1000 Hz 21.5 24.5 2000 Hz 27.0 30.5 4000 Hz 30.0 35.0 51.5 75.0 61.5 50.0 55.5 Table II Reference calibration values in dB peSPL and dB peVFL Burst 0dB 1. This function is used to find the peRETSPL (peSPL nHL) correction figure for the Transducer selected Tone Burst frequency. There is only one stored correction figure per frequency. 2. Select the “Measure type” and “Transducer type” for which the peRETSPL correction figure is sought. 3. Select the desired “Tone Burst” frequency, Sine waves, Window, Rise/Fall,Plateau and Stimuli frequency. 4. Let a person with normal hearing listen to the output transducer, in a suitably quiet test environment. 5. Adjust the slider until the threshold is found. 6. Select the next “Tone Burst” frequency and repeat point 3-6 until all peRETSPL for all frequencies are found. 7. Select the next measure type or/and the next stimuli type for which the peRETSPL correction figure is required. 2 For NB-chirp calibration it is recommended that the pure tone frequency corresponds to the center frequency of the NB-chirp. Eclipse ABR and OAE Service Manual 38 NB CE-Chirp® 0dB 1. This function is used to find the peRETSPL (peSPL nHL) correction figure for the Transducer selected NB Chirp® frequency. There is only one stored correction figure per frequency. 2. Select the “Measure type” and “Transducer type” for which the peRETSPL correction figure is sought. 3. Select the desired “Tone Burst” frequency 500-1000-2000-4000Hz. 4. Let a person with normal hearing listen to the output transducer, in a suitably quiet test environment. 5. Adjust the slider until the threshold is found. 6. Select the next Tone Burst frequency and repeat point 3-6 until all peRETSPL for all frequencies are found. 7. Select the next measure type or/and the next stimuli type for which the peRETSPL correction figure is required. Click 200Hz-10kHz 0 dB 1. This function is used to find the peRETSPL (peSPL nHL) correction figure for the Click 200Hz-10kHz. 2. Select the “Measure type” and “Transducer type” for which the peRETSPL correction figure is sought. 3. Let a person with normal hearing listen to the output transducer, in a suitably quiet test environment. 4. Adjust the slider until the threshold is found. 5. Select the next measure type or/and the next stimuli type for which the peRETSPL correction figure is required. CE-Chirp® 0 dB 1. This function is used to find the peRETSPL (peSPL nHL) correction figure r for the CE-Chirp® 2. Select the “Measure type” and “Transducer type” for which the peRETSPL correction figure is sought. 3. Let a person with normal hearing listen to the output transducer, in a suitably quiet test environment. 4. Adjust the slider until the threshold is found. 5. Select the next measure type or/and the next stimuli type for which the peRETSPL correction figure is required. Select the Settings Tab Version info. There is information about software, firmware and hardware. Calibration program version nr. Firmware version nr. Hardware version nr. Eclipse ABR and OAE Service Manual 39 Password: In Settings it’s possible to change/delete the password, import and export calibration. To change Password: 1. Type the old password if passwords are select. 2. Type the new password. 3. Confirm the new password. 4. After select OK the new settings will be stored. Level pefix: There can select between peSPL and nHL intensity scale in new protocol setup in main program. Click type: There can select between Traditional 100 µS or a band limits click as stimulus in new protocol setup in main program. Level measure method: There can select between peak to baseline or peak to trough measuring level main program latency measuring. Calibration: The Export – Import feature can be used to backup the calibration value. TEOAE 25 calibration Switch on power and activated the one of follow calibration program. The calibration program for the four different software modules can be found on the following paths: The TEOAE 25 Calibration can be found in “C:\IABASE2000\TEOAE\TEOAE25 Calibration.exe” Activate the icon for the desired Calibration Application and type in the password if necessary. Stimuli Attenuator Calibration. The Tab called Stimuli attenuator, uses to calibrated to output attenuator. Stimuli Att. calibration is included in all software modules, it is only necessary to calibrate the stimuli Att. once. There can be differences in the calibration procedure in the individual software modules. The difference in between the calibration procedure holds for the normal calibration of tolerances, so the method is applied is optional. Eclipse ABR and OAE Service Manual 40 All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Calibration of the Attenuator in Channel 1. Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 1 is selected in “Channel to calibrate” area on the display. This calibration is not described in the TEOAE calibration software 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP6. 2. Adjust P6 to minimum THD. Dynamic Attenuator 0 dB 1. Connect an AC Voltmeter to MP2 and notice the output V1. 2. Connect the voltmeter to MP6. 3. Adjust P3 until the voltmeter shows exactly 20 dB ± 0.1 dB more than V1. Notice the output V2. 4. Press the – 60-dB button in the dynamic attenuator area. Dynamic Attenuator -60 dB 1. Adjust the P2 until the AC voltmeter shows exactly 60 dB ± 0.1 dB less than V2. 2. Press the – 120 dB key in the dynamic attenuator area on the display. Dynamic Attenuator -120 dB Notice: This Point can only be done using an AC voltmeter with 1000 Hz band pass filter. 1. Move slider in the bottom of the dynamic attenuator area until the AC voltmeter, with 1000 Hz filter, shows exactly 120 dB ± 0.5 dB less than V2. 2. Select the 0 dB button in the dynamic attenuator area. Control of the fixed Attenuator steps in the Attenuator Channel 1. Scroll down to description for fixed attenuator test: Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP6 and notice the Output V3. 2. Selected the -5 dB button in the “Fixed Attenuator” area. Fix Attenuator -5 dB 1. Control that the AC voltmeter output is 5 dB ± 0.5 dB less than V3. 2. Select the -10 dB button in the “Fixed attenuator” area. Fix Attenuator -10 dB 1. Control that the AC voltmeter output is 10 dB ± 0.5 dB less than V3. 2. Select the -15 dB button in the “Fixed attenuator” section. Eclipse ABR and OAE Service Manual 41 Fix Attenuator -15 dB 1. Control that the AC voltmeter output is 15 dB ± 0.5 dB less than V3. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Control that the AC voltmeter output is 20 dB ± 0.5 dB less than V3. 2. Select the Channel 2 button in the “Channel to calibrate” area. Calibration of Attenuator in Channel 2. All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 1 is selected in “Channel to calibrate” area on the display. This calibration is not described in the TEOAE calibration software 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP5. 2. Adjust P5 to minimum THD. Dynamic Attenuator 0 dB 1. Connect an AC Voltmeter to MP3 and notice the output V4. 2. Connect the voltmeter to MP5. 3. Adjust P4 until the voltmeter shows exactly 20 dB ± 0.1 dB more than V4. Notice the output V5. 4. Press the – 60-dB button in the dynamic attenuator area. Dynamic Attenuator -60 dB 1. Adjust the P1 until the AC voltmeter shows exactly 60 dB ± 0.1 dB less than V5. 2. Press the – 120 dB key in the dynamic attenuator area on the display. Eclipse ABR and OAE Service Manual 42 Dynamic Attenuator -120 dB Notice: This Point can only be performed by using an AC voltmeter with 1000 Hz band pass filter. 1. Move slider in the bottom of the dynamic attenuator area until the AC voltmeter, with 1000 Hz filter, shows exactly 120 dB ± 0.5 dB less than V5. 2. Select the 0 dB button in the dynamic attenuator area. Control of the fixed Attenuator steps in the Attenuator Channel 2. Scroll down to description for fixed attenuator test: Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP5 and notice the Output V6. 2. Selected the -5 dB button in the “Fixed Attenuator” area. Fix Attenuator -5 dB 1. Control that the AC voltmeter output is 5 dB ± 0.5 dB less than V6. 2. Select the -10 dB button in the “Fixed attenuator” area. Fix Attenuator -10 dB 1. Control that the AC voltmeter output is 10 dB ± 0.5 dB less than V6. 2. Select the -15 dB button in the “Fixed attenuator” section. Fix Attenuator -15 dB 1. Control that the AC voltmeter output is 15 dB ± 0.5 dB less than V6. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Control that the AC voltmeter output is 20 dB ± 0.5 dB less than V6. OAE input amplifier. Select the Input Attenuator tab in the Calibration program and activate the 400 Hz filter. All the measurements and adjustments are made on the Advanced Switch Board “In TEOAE Software call OAE preamplifier board”. All the measure points’ (MP) and potentiometers (P) refers to the Advanced Switch Board Component Index page 20. All measurements are made with reference to ground MP1. Eclipse ABR and OAE Service Manual 43 400Hz filter 1. Connect the OAP probe to the artificial ear (IEC711). 2. Adjust the Test signal level slider until the sound level meter show 94 dB SPL. 3. Adjust the P6 on the OAE preamplifier board, until the stimulus RMS level meter show 94,0 dB SPL (1.00 Pa). 4. Control the Stimulus peak level meter show 97,0 dB ± 0.25 dB 5. Control the Echo peak level meter show 97,0 dB 0 + 0.5 dB 6. Control the Echo RMS level meter show 94,0 dB 0 + 0,5 dB 7. Select the input filter to 100 Hz. Activate the 100Hz filter 1. Adjust the P5 on the OAE preamplifier board, until the stimulus RMS level meter show 94,0 dB SPL (1.00 Pa). 2. Control the Stimulus peak level meter show 97,0 dB ± 0.25 dB 3. Control the Echo peak level meter show 97,0 dB 0 + 0.5 dB 4. Control the Echo RMS level meter show 94,0 dB 0 + 0,5 dB 5. Select Offset button. Activate the Offset 1. Disconnect the OAP probe from the socket. 2. Connect a DC millivoltmeter to MP7 with reference to GND MP1. 3. Adjust P7 until DC millivoltmeter show 0,0 mV TEOAE stimuli calibration. Select the stimuli Level to make this part of calibration. Calibration of stimulus level is done in peak equivalent sound pressure peSPL Eclipse ABR and OAE Service Manual 44 Peak 1. Connect the OAE probe to the artifical ear, which connected to a sound level meter (SLM). 2. Selected 1 kHz click ref button. 3. Adjust the slider until the SLM show 80 dB SPL. 4. Connected an oscilloscope to the output of the SLM, and notice the peak value V1. 5. Selected the click button and control the linear button is select. 6. Adjust the slider until the oscilloscope peak value is the same as V1. 7. Selected SP2 and repeat point 26. Select the Settings Tab Setting is identical for all 4 software modules. In Settings it’s possible to change/delete the password, import and export calibration. To change Password: 1. Type the old password if passwords are select. 2. Type the new password. 3. Confirm the new password. 4. After select OK the new settings will be stored. The Export – Import feature can be used to backup the calibration value. Eclipse ABR and OAE Service Manual 45 DPOAE 20 calibration. Switch on power and activated the one of follow calibration program. The calibration program for the four different software modules can be found on the following paths: The DPOAE 20 Calibration can be found in “C:\IABASE2000\DPOAE\Calibration\DPOAE20 Calibration.exe” Activate the icon for the desired Calibration Application and type in the password if necessary. Stimuli Attenuator Calibration. The Tab called Stimuli attenuator, uses to calibrated to output attenuator. Stimuli Att. calibration is included in all software modules, it is only necessary to calibrate the stimuli Att. once. There can be differences in the calibration procedure in the individual software modules. The difference in between the calibration procedure holds for the normal calibration of tolerances, so the method is applied is optional. All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Calibration of the Attenuator in Channel 1. Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 1 is selected in “Channel to calibrate” area on the display. This calibration is not described in the DPOAE calibration software 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP6. 2. Adjust P6 to minimum THD. Dynamic Attenuator 0 dB 1. Connect an AC Voltmeter to MP2 and notice the output V1. 2. Connect the voltmeter to MP6. 3. Adjust P3 until the voltmeter shows exactly 20 dB ± 0.1 dB more than V1. Notice the output V2. 4. Press the – 60-dB button in the dynamic attenuator area. Eclipse ABR and OAE Service Manual 46 Dynamic Attenuator -60 dB 1. Adjust the P2 until the AC voltmeter shows exactly 60 dB ± 0.1 dB less than V2. 2. Press the – 120 dB key in the dynamic attenuator area on the display. Dynamic Attenuator -120 dB Notice: This Point can only be done using an AC voltmeter with 1000 Hz band pass filter. 1. Move slider in the bottom of the dynamic attenuator area until the AC voltmeter, with 1000 Hz filter, shows exactly 120 dB ± 0.5 dB less than V2. 2. Select the 0 dB button in the dynamic attenuator area. Control of the fixed Attenuator steps in the Attenuator Channel 1. Scroll down to description for fixed attenuator test: Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP6 and notice the Output V3. 2. Selected the -5 dB button in the “Fixed Attenuator” area. Fix Attenuator -5 dB 1. Control that the AC voltmeter output is 5 dB ± 0.5 dB less than V3. 2. Select the -10 dB button in the “Fixed attenuator” area. Fix Attenuator -10 dB 1. Control that the AC voltmeter output is 10 dB ± 0.5 dB less than V3. 2. Select the -15 dB button in the “Fixed attenuator” section. Fix Attenuator -15 dB 1. Control that the AC voltmeter output is 15 dB ± 0.5 dB less than V3. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Control that the AC voltmeter output is 20 dB ± 0.5 dB less than V3. 2. Select the Channel 2 button in the “Channel to calibrate” area. Eclipse ABR and OAE Service Manual 47 All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 2 is selected in “Channel to calibrate” area on the display. This calibration is not described in the DPOAE calibration software 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP5. 2. Adjust P5 to minimum THD. Dynamic Attenuator 0 dB 1. Connect an AC Voltmeter to MP3 and notice the output V4. 2. Connect the voltmeter to MP5. 3. Adjust P4 until the voltmeter shows exactly 20 dB ± 0.1 dB more than V4. Notice the output V5. 4. Press the – 60-dB button in the dynamic attenuator area. Dynamic Attenuator -60 dB 1. Adjust the P1 until the AC voltmeter shows exactly 60 dB ± 0.1 dB less than V5. 2. Press the – 120 dB key in the dynamic attenuator area on the display. Dynamic Attenuator -120 dB Notice: This Point can only be performed by using an AC voltmeter with 1000 Hz band pass filter. 1. Move slider in the bottom of the dynamic attenuator area until the AC voltmeter, with 1000 Hz filter, shows exactly 120 dB ± 0.5 dB less than V5. 2. Select the 0 dB button in the dynamic attenuator area. Control of the fixed Attenuator steps in the Attenuator Channel 2. Scroll down to description for fixed attenuator test: Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP5 and notice the Output V6. 2. Selected the -5 dB button in the “Fixed Attenuator” area. Fix Attenuator -5 dB 1. Control that the AC voltmeter output is 5 dB ± 0.5 dB less than V6. 2. Select the -10 dB button in the “Fixed attenuator” area. Fix Attenuator -10 dB 1. Control that the AC voltmeter output is 10 dB ± 0.5 dB less than V6. 2. Select the -15 dB button in the “Fixed attenuator” section. Fix Attenuator -15 dB 1. Control that the AC voltmeter output is 15 dB ± 0.5 dB less than V6. 2. Select the -20 dB button in the “Fixed attenuator” section. Eclipse ABR and OAE Service Manual 48 Fix Attenuator -20 dB 1. Control that the AC voltmeter output is 20 dB ± 0.5 dB less than V6. OAE input amplifier. Select the Input Attenuator tab in the Calibration program and activate the 400 Hz filter. All the measurements and adjustments are made on the Advanced Switch Board “In TEOAE Software call OAE preamplifier board”. All the measure points’ (MP) and potentiometers (P) refers to the Advanced Switch Board Component Index page 20. All measurements are made with reference to ground MP1. 400Hz filter 1. Connect the OAP probe to the artificial ear (IEC711). 2. Adjust the Test signal level slider until the sound level meter show 94 dB SPL. 3. Adjust the P6 on the OAE preamplifier board, until the stimulus RMS level meter show 94,0 dB SPL (1.00 Pa). 4. Control the Stimulus peak level meter show 97,0 dB ± 0.25 dB 5. Control the Echo peak level meter show 97,0 dB 0 + 0.5 dB 6. Control the Echo RMS level meter show 94,0 dB 0 + 0,5 dB 7. Select the input filter to 100 Hz. Activate the 100Hz filter 1. Adjust the P5 on the OAE preamplifier board, until the stimulus RMS level meter show 94,0 dB SPL (1.0 Pa). 2. Control the Stimulus peak level meter show 97,0 dB ± 0.25 dB 3. Control the Echo peak level meter show 97,0 dB 0 + 0.5 dB 4. Control the Echo RMS level meter show 94,0 dB 0 + 0,5 dB 5. Select Offset button. Eclipse ABR and OAE Service Manual 49 Activate the Offset 1. Disconnect the OAP probe from the socket. 2. Connect a DC millivoltmeter to MP7 with reference to GND MP1. 3. Adjust P7 until DC millivoltmeter show 0,0 mV DPOAE20 Microphone frequency response calibration. Select the DPOAE20 calibration program to make this part of calibration. Measure the difference between the Ear simulator and the OAE probe microphone frequency response for the following frequencies: 370, 500, 1000, 2000, 3000, 4000, 5000, 6000, 7000, 8000 Hz. 1. Connecte the OAE probe to the artifical ear, which is connected to a sound level meter (SLM): 2. Select a 370 in the list. Freq. 370Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 500 Hz in the list. Freq. 500Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 1000 Hz in the list. Freq. 1000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 2000 Hz in the list. Freq. 2000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 3000 Hz in the list. Freq. 3000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 4000 Hz in the list. Freq. 4000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 5000 Hz in the list. Eclipse ABR and OAE Service Manual 50 Freq. 5000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 6000 Hz in the list. Freq. 6000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 7000 Hz in the list. Freq. 7000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Select 8000 Hz in the list. Freq. 8000Hz. 1. Adjust SP1 Level Slider until SLM shows 80 dB. 2. Calibration is done. Select the Settings Tab. In Settings it’s possible to change/delete the password, import and export calibration. To change Password: 1. Type the old password if passwords are select. 2. Type the new password. 3. Confirm the new password. 4. After select OK the new settings will be stored. The Export – Import feature can be used to backup the calibration value. Eclipse ABR and OAE Service Manual 51 ASSR calibration. Switch on power and activated the one of follow calibration program. The calibration program for the four different software modules can be found on the following paths: The Interacoustics ASSR Calibration module can be found in “C:\Program Files\ Interacoustics\ Eclipse\ Interacoustics ASSR 1.2.3 Build 42\ASSR Calibration.exe” on a English windows. “Program Files” are language dependent. Activate the icon for the desired Calibration Application and type in the password if necessary. Stimuli Attenuator Calibration. The Tab called Stimuli attenuator, uses to calibrated to output attenuator. Stimuli Att. calibration is included in all software modules, it is only necessary to calibrate the stimuli Att. once. There can be differences in the calibration procedure in the individual software modules. The difference in between the calibration procedure holds for the normal calibration of tolerances, so the method is applied is optional. All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Calibration of the Attenuator in Channel 1. Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 1 is selected in “Channel to calibrate” area on the display. This calibration is not described in the ASSR calibration software 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP6. 2. Adjust P6 to minimum THD. Dynamic Attenuator 0 dB 1. Connect an AC Voltmeter to MP2 and notice the output V1. 2. Connect the voltmeter to MP6. 3. Adjust P3 until the voltmeter shows exactly 20 dB ± 0.1 dB more than V1. Notice the output V2. 4. Press the – 60-dB button in the dynamic attenuator area. Eclipse ABR and OAE Service Manual 52 Dynamic Attenuator -60 dB 1. Adjust the P2 until the AC voltmeter shows exactly 60 dB ± 0.1 dB less than V2. 2. Press the – 120 dB key in the dynamic attenuator area on the display. Dynamic Attenuator -120 dB Notice: This Point can only be done using an AC voltmeter with 1000 Hz band pass filter. 1. Move slider in the bottom of the dynamic attenuator area until the AC voltmeter, with 1000 Hz filter, shows exactly 120 dB ± 0.5 dB less than V2. 2. Select the 0 dB button in the dynamic attenuator area. Control of the fixed Attenuator steps in the Attenuator Channel 1. Scroll down to description for fixed attenuator test: Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP6 and notice the Output V3. 2. Selected the -5 dB button in the “Fixed Attenuator” area. Fix Attenuator -5 dB 1. Control that the AC voltmeter output is 5 dB ± 0.5 dB less than V3. 2. Select the -10 dB button in the “Fixed attenuator” area. Fix Attenuator -10 dB 1. Control that the AC voltmeter output is 10 dB ± 0.5 dB less than V3. 2. Select the -15 dB button in the “Fixed attenuator” section. Fix Attenuator -15 dB 1. Control that the AC voltmeter output is 15 dB ± 0.5 dB less than V3. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Control that the AC voltmeter output is 20 dB ± 0.5 dB less than V3. 2. Select the Channel 2 button in the “Channel to calibrate” area. Eclipse ABR and OAE Service Manual 53 All the measure points’ (MP) and potentiometers (P) refers to the ERA Board Component Index page 18 All measurements are made with reference to GND (MP4). Select the Stimuli Attenuator tab in the Calibration program. Verify Channel 2 is selected in “Channel to calibrate” area on the display. This calibration is not described in the ASSR calibration software 1. Connect a spectrum analyzer / distortion meter and an AC voltmeter to MP5. 2. Adjust P5 to minimum THD. Dynamic Attenuator 0 dB 1. Connect an AC Voltmeter to MP3 and notice the output V4. 2. Connect the voltmeter to MP5. 3. Adjust P4 until the voltmeter shows exactly 20 dB ± 0.1 dB more than V4. Notice the output V5. 4. Press the – 60-dB button in the dynamic attenuator area. Dynamic Attenuator -60 dB 1. Adjust the P1 until the AC voltmeter shows exactly 60 dB ± 0.1 dB less than V5. 2. Press the – 120 dB key in the dynamic attenuator area on the display. Dynamic Attenuator -120 dB Notice: This Point can only be done using an AC voltmeter with 1000 Hz band pass filter. 1. Move slider in the bottom of the dynamic attenuator area until the AC voltmeter, with 1000 Hz filter, shows exactly 120 dB ± 0.5 dB less than V5. 2. Select the 0 dB button in the dynamic attenuator area. Control of the fixed Attenuator steps in the Attenuator Channel 2. Scroll down to description for fixed attenuator test: Fix Attenuator 0 dB 1. Connect an AC voltmeter to MP5 and notice the Output V6. 2. Selected the -5 dB button in the “Fixed Attenuator” area. Fix Attenuator -5 dB 1. Control that the AC voltmeter output is 5 dB ± 0.5 dB less than V6. 2. Select the -10 dB button in the “Fixed attenuator” area. Fix Attenuator -10 dB 1. Control that the AC voltmeter output is 10 dB ± 0.5 dB less than V6. 2. Select the -15 dB button in the “Fixed attenuator” section. Eclipse ABR and OAE Service Manual 54 Fix Attenuator -15 dB 1. Control that the AC voltmeter output is 15 dB ± 0.5 dB less than V6. 2. Select the -20 dB button in the “Fixed attenuator” section. Fix Attenuator -20 dB 1. Control that the AC voltmeter output is 20 dB ± 0.5 dB less than V6. ASSR Input Section Select the Input Attenuator tab. All the measurements and adjustments are made on the Switch Board board. All the measure point’s (MP) and potentiometers (P) refers to the Advanced Switch Board Component Index page 20 All measurements are made with reference to ground MP1. Input Amplifier Control. 104 dB Amplification. 1. Connect a sine wave generator pre-selected to 1000 Hz and 70,7 mV rms (100mV peak) to the 8pins preamplifier input socket on the EP15( Eclipse). Short circuit pin 5 and pin 4 and connect this point to the sine wave generator. The ground wire shall be connected to pin 2. 2. Control the Left and Right level meter in the bottom of the display shows 0.0 dB ± 0.5 dB. 3. Select the 98 dB button on the Fixed attenuator area. 98 dB Amplification. 3. Control the Left and Right level meter in the bottom of the display shows -6.0 dB ± 0.5 dB. 4. Select the 92 dB button on the Fixed attenuator area. 92 dB Amplification. 3. Control the Left and Right level meter in the bottom of the display shows -12.0 dB ± 0.5 dB. 4. Select the 86 dB button on the Fixed attenuator area. Eclipse ABR and OAE Service Manual 55 86 dB Amplification. 3. Control the Left and Right level meter in the bottom of the display shows -18.0 dB ± 0.5 dB. 4. Select the 80 dB button on the Fixed attenuator area. 80 dB Amplification. 3. Control the Left and Right level meter in the bottom of the display shows -24.0 dB ± 0.5 dB. 4. Select the 74 dB button on the Fixed attenuator area. 74 dB Amplification. 3. Control the Left and Right level meter in the bottom of the display shows -30.0 dB ± 0.5 dB. 4. Select the 68 dB button on the Fixed attenuator area. 68 dB Amplification. 3. Control the Left and Right level meter in the bottom of the display shows -36.0 dB ± 0.5 dB. 4. Select the 62 dB button on the Fixed attenuator area. 62 dB Amplification. 2. Control the Left and Right level meter in the bottom of the display shows -42.0 dB ± 0.5 dB. 3. Select the 92 dB button on the Fixed attenuator area. Control of the input filter. 1. Select 62 dB amplification. 2. Connect a sine wave generator pre selected to 1000 Hz and 70.7 mV rms. (100 mV peak) to the 8 pins preamplifier input socket on the rear plate. Short-circuit pin 5 and pin 4 and connect this point to the signal of the sine wave generator. The ground wire shall be connected to pin2. 3. Control that the left and right level meter in the bottom of the display shows -42dB ± 0.5dB. 4. Select the sine wave generator frequency to same frequency as the filter settings frequency. 5. Control that the left and right level meter in the bottom of the display shows –43dB -48 dB. 6. Select next filter setting frequency and repeat point 2 6. Eclipse ABR and OAE Service Manual 56 ASSR Frequencies calibration. Select the ASSR Calibration program to make this part of calibration. peak-peak Calibration of stimulus level is done in peak equivalent sound pressure peSPL 1kHz Ref. 1. Connect the ABR 3A phone to an artifical ear, which is connected to a sound level meter (SLM). 2. Adjust the slider until the SLM shows 100 dB SPL. 3. Connect an oscilloscope to the output of the SLM, and notice the peak to peak value V1. 4. Select the ASSR Frequencies button in the stimuli type area, and 500 Hz in frequency area, 1 Octave in Bandwidth area and 90 Hz in the base frequency area. 500Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 750 Hz button in the frequency area. 750Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 1000 Hz button in the frequency area. 1000Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 1500 Hz button in the frequency area. 1500Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. Eclipse ABR and OAE Service Manual 57 2. Select the 2000 Hz button in the frequency area. 2000Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 3000 Hz button in the frequency area. 3000Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 4000 Hz button in the frequency area. 4000Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 6000 Hz button in the frequency area. 6000Hz Frequency 1. Adjust the slider until oscilloscope peak to peak value is the same as V1. 2. Select the 500 Hz button in the frequency area, 40 Hz button in base frequency area and 1 Octave. Repeat 500Hz 6000 Hz 40 Hz 1 Octave calibration. WN 1. Adjust slider until the SLM shows 100 dB SPL. 2. Select the 1 kHz ref, and next transducer to calibrate. Repeat 1kHz ref, ASSR Frequencies and WN calibration for all transducer there are support. Select the Settings Tab In Settings it’s possible to change/delete the password, import and export calibration. To change Password: 1. Type the old password if passwords are select. 2. Type the new password. 3. Confirm the new password. 4. After select OK the new settings will be stored. The Export – Import feature can be used to backup the calibration value. Eclipse ABR and OAE Service Manual 58 Trouble Shooting in TEOAE25, EP15 and EP25 Modules Error message/behavior List of auto tests is disabled and empty Cause Hardware disconnected No license Hardware turned off DSP board needs to be calibrated!!! The DSP is busy What to do Connected hardware. Go into the Help and select About box and press "License". The Dialog shown below will appear. Send the DSP Serial no to Interacoustics A/S,Assens and you will receive a "New License Key" number that must be typed in Turn on hardware. Contact the distributor for a calibration. Unknown error Close module, restart hardware and start module again. Trouble Shooting in ASSR Module Error message/behavior Hardware not detected DSP board needs to be calibrated!!! DSP board needs to be calibrated!!! DSP board are calibrated for EPx5. Do you want to reuse these settings, and accept a possible error of up to 2dB? (Settings will be stored in hardware!) No License Cause Hardware disconnected What to do Connect hardware and turn on. Hardware turned off Hardware not calibrated Turn on hardware. Hardware not calibrated No license Contact the distributor for a calibration. Answer Yes to accept and use precalibration values from EPx5. A possible error of up to 2dB does exist. Answer No, to use instrument as not calibrated. The next time the program is started, program will prompt again with this question. Go into the Help and select About box and press "License". The Dialog shown below will appear. Send the DSP Serial no to Interacoustics A/S, Assens and you will receive a "New License Key" number that must be typed in Eclipse ABR and OAE Service Manual 59 Trouble Shooting in the DPOAE Module Error message/behavior Start button disabled Cause Hardware disconnected No license Hardware turned off DSP board needs to be calibrated!!! Unable to correct level Nothing happens when start button is pushed or message : Error booting DSP All other errors What to do Connected hardware. Go into the Help and select About box and press "License". The Dialog shown below will appear. Send the DSP Serial no to Interacoustics A/S,Assens and you will receive a "New License Key" number that must be typed in Turn on hardware. Contact the distributor for a calibration. Precaution to prevent too loud output – probably caused by a misfit or clotted probe. Bad connection to hardware Check probe and probefit and run test again. Turn hardware off and on again after a few seconds and try again Contact Interacoustics A/S for support Eclipse ABR and OAE Service Manual 60 Appendix U Update news Following modifications are made for Manual number 80701116 09/ 2012 Spare parts updated Change of headband New USB isolation Introduction of Sanibel consumable part numbers Introduction of new numbering system Following modifications are made from serial 898082 Manual number 80701115 02/ 2012 New Main Board 51004312 PCA update for heavy oscillation New Ear eartips and Ear tip box New OAE Probe with probe tip New Snap and tap electrodes New Headset Following modifications are made from serial 856099 Manual number 80701114 06/ 2011 New Main board 51004311 Compatible to previous version PCA update for optimizing usage of components with various brands New power switch 10A ON-NONE-OFF New optical isolation CO25 Optical USB Following modifications are made from serial 802804 Manual number 80701113 10/ 2010 Not compatible to previous versions New Mainboard 51004310 Implementation of trigger extension Improvement of attenuator adjustment Optimizing layout New Assembly sketch Optimizing for EMC and safety tests Wiring updates Eclipse ABR and OAE Service Manual 61 Following modifications are made for Manual number 80701112 02 / 2010: New software version 4.0 Improved VEMP Speed optimization of the data communication between PC and Eclipse. New Preamplifier board PCB 69030905 PCA 51017307 from serial 773475 Detection of electrode fall of introduced When an electrode drops off, the preamplifier will deliver a very high signal to the Eclipse, which will trigger the reject system on the EPx5 module, and then gives the operator a rejection warning (only works, if “Disable rejection” in EPx5 isn’t set) New Main board PCA 51004308 and trigger board PCA 51020501 from serial 769946 Improvement of respond to shorter trigger pulses Following modifications are made for instruments from serial 749308 - PreAmplifier tracking no. PL0800738001 Manual number 80701111 01/2009 Instrument design: Top profile colour and instrument name design change New Preamplifier and Collector boxes EPA PreAmplifier, EPA3 Collector box, EPA4 Collector box New design and improved RFI rejection PreAmplifier PCA 51017305 New Advanced switch board 51005304 Redesign for better insulation mode rejection and UL-approved components in isolation path. Eclipse ABR and OAE Service Manual 62