Production and Use of X-rays
X-rays are a form of electromagnetic radiation. X rays are produced when high-speed electrons
hit metal targets.
A simplified diagram of a modern form of X-ray tube is shown:
Electrons are emitted from the heated cathode by thermionic effect. The electrons are accelerated
through a large potential difference before bombarding a metal anode.
The X-rays produced leave the tube via a ‘window’.
Whenever a charged particle is accelerated, electromagnetic radiation is emitted. The greater the
acceleration, the shorter is the wavelength of the emitted radiation. When high-speed electrons
strike a metal target, large accelerations occur and the radiation produced is in the X-ray
region of the electromagnetic spectrum. Since the electrons have a continuous distribution of
accelerations, a continuous distribution of wavelengths of X-rays is produced.
A typical spectrum of the X-rays produced is shown:
1
The spectrum consists of two components. There is a continuous distribution of wavelengths
with a sharp cut-off at short wavelength and also a series of high-intensity spikes that are
characteristic of the target material.
There is a minimum wavelength (a cut-off wavelength) where the whole of the energy of the
electron is converted into the energy of one photon.
That is,
β„Žπ‘
=
πœ†
where e is the charge on the electron that has moved through a potential difference V, h is the
Planck
constant, c is the speed of light and λ is the wavelength of the emitted X-ray photon.
The sharp peaks observed correspond to the emission line spectrum of the atoms of the target.
The electrons that bombard the target, excite orbital electrons in the lower energy levels and the
subsequent de-excitation of electrons gives rise to the line spectrum.
Heating of the metal anode: Since the majority of the energy of the electrons in transferred to
thermal energy in the metal anode, the anode is either water-cooled or is made to spin rapidly so
that the target area is increased. The anode is held at earth potential.
Intensity and hardness of X-ray beam
Intensity: (the energy per second per unit area incident normally on a surface)
ο‚· Determined by the rate of arrival of electrons at target metal (known as tube current)
ο‚· Tube current is controlled by the heater current of the cathode.
ο‚· The greater the heater current, the hotter the filament and hence the greater the rate of
emission of thermo-electrons.
ο‚· A more intense X-ray produces an image quicker
Hardness: (penetration of x-ray)
ο‚· Controlled by the accelerating voltage between cathode and anode.
ο‚· More penetrating X-rays have higher photon energies and thus a larger accelerating
potential is required.
ο‚· A harder X-ray has more penetrating power
Note: From the spectrum it can be seen that longer wavelength X rays (‘softer’ X-rays) are
always also produced. Indeed some X-ray photons are of such low energy that they would not be
able to pass through the patient. These ‘soft’ X-rays would contribute to the total radiation dose
without any useful purpose. Consequently, an aluminium filter is frequently fitted across the
window of the X-ray tube to absorb the ‘soft’ X-ray photons.
X-ray Imaging
X-ray radiation affects photographic plates in much the same way as visible light. A
photographic plate, once exposed, will appear blackened after development. The degree of
blackening is dependent on the total X-ray exposure.
X-ray beams are used to obtain ‘shadow’ pictures of the inside of the body to assist in the
diagnosis or treatment of illness. If a picture is required of bones, this is relatively simple since
the absorption by bone of X-ray photons is considerably greater than the absorption by
2
surrounding muscles and tissues. X-ray pictures of other parts of the body may be obtained if
there is sufficient difference between the absorption properties of the organ under review and the
surrounding tissues.
The quality of the shadow picture (the image) produced on the photographic plate depends on
its sharpness and contrast.
X-ray Quality
Sharpness:
It is the ease with which the edges of structures can be determined.
The sharpness of X-ray can be improved by:
ο‚· Reducing the area of the target anode
ο‚·
ο‚·
Reducing aperture size (window) produced by overlapping metal plates, through which
the X-ray beam passes after leaving the tube: reduces beam width
using a lead grid in front of the photographic film to absorb scattered X-ray photons and
reduces partial image
Contrast:
It is the visual difference between the areas of blackening and light(the degress of blackening of
structures)
Contrast can be improved by:
ο‚· Using a contrast medium:
For example, the stomach may be examined by giving the patient a drink containing barium
sulphate. Similarly, to outline blood vessels, a contrast medium that absorbs strongly the Xradiation would be injected into the bloodstream.
3
ο‚·
ο‚·
ο‚·
ο‚·
Increasing the Exposure time
Using harder X-Rays: increases penetration power
Reducing the scattering of X-Ray beam
backing the photographic film with a fluorescent material.
Attenuation of X-rays in matter
Consider a mono-energetic x-ray beam (single wavelength) passing through a given thickness x
of a medium.
When a beam of X-rays passes through matter, some X-ray photons are absorbed and some are
scattered. The intensity of the transmitted beam is said to be attenuated because its intensity
decreases the further it travels through the matter.
For a beam of X-rays that passes through an ‘absorber’ of thickness x, the transmitted intensity I
is given by
𝐼 = 𝐼0 𝑒 −πœ‡π‘₯
where 𝐼0 is the incident intensity of the beam
𝐼 is the emergent x-ray
π‘₯ is the thickness of the medium
and  is the attenuation coefficient (sometimes called the
absorption coefficient) of the absorber substance. The unit of  is
m−1.
The variation with thickness x of an absorber of the percentage transmission of a parallel beam of
X-ray radiation
4
The thickness of the medium required to reduce the transmitted intensity to one half of its initial
value is a constant and is known as the half-value thickness 𝑋1 or HVT. The half-value thickness
2
𝑋1 is related to the linear absorption coefficient μ by the expression
2
𝑋1 × μ = ln2.
2
Derivation :
𝐼 = 𝐼0 𝑒 −πœ‡π‘₯
𝐼0
= 𝐼0 𝑒 −πœ‡π‘₯
2
1
ln = ln 𝑒 −πœ‡π‘₯
2
𝑙𝑛1 − 𝑙𝑛2 = −πœ‡π‘₯
ln 2 = πœ‡π‘₯
ln 2 = πœ‡ × π‘‹1
2
Computed Tomography (CAT/CT Scan)
The image produced on an X-ray plate is a flat image and does not give the depth of the tissue.
Tomography is a technique by which an image of a slice or plane of an object may be obtained.
A series of x-ray images are obtained from different angles through a slice but in the same plane.
Computer techniques combine these images to give an image of the slice. This techniques is
called computed tomography or CT scanning.
Combining successive slices will give a 3D image that can be rotated and viewed from any
angle.
5
The basic principle of CT scanning
The aim of CT scanning is to produce an image of a slice through the cube. The cube is divided
into a series of small units called voxels.
Each voxel will absorb the X-ray beam to a different extent. The intensity transmitted through
each voxel alone can be given a number, referred to as pixel.
The various pixels are built up from measurements of the x ray intensity along different
directions through the slice.
Suppose a section consists of four voxels with intensities as shown:
The number on each voxel is the pixel intensity that is to be reproduced.
1.
If a beam of X-rays is directed from the left, then detectors will give readings of 5 and 9
2.
The X-ray tube and detectors are now rotated through 45°
6
3.
The procedure is repeated after rotating the X-ray tube and the detectors through a further
45°.
4.
The final images are taken after rotating the X-ray tube and the detectors through a
further 45°.
7
5.
The final pattern of pixels
In order to obtain the original pattern of pixels, two operations must be performed.
1. The ‘background’ intensity must be removed. The ‘background’ intensity is the total of each
set of detector readings. In this case, 14 is deducted from each pixel.
2. After deduction of the ‘background’, the result must be divided by three to allow for the
duplication of the views of the section. (since it was rotated 3 times)
8
June 2012 P 42 Q 10
(a) An aluminium block is placed near to a small source of X-ray radiation, as shown in
Fig. 10.1.
X-rays from the source are detected at point A and at point B.
State two reasons why the intensity of the X-ray beam at point B is not as great as the intensity
at point A.
1. .............................................................................................................................................
........................................................................................................................................................
2. .............................................................................................................................................
..........................................................................................................................................[2]
(b) A cross-section through a model of a finger is shown in Fig. 10.2.
The thickness of the model is 2.4 cm and that of the bone in the model is 1.1 cm. Parallel
beams of X-rays are incident on the model in the directions AB and CD, as shown in Fig. 10.2.
9
Data for the linear attenuation (absorption) coefficient πœ‡ο€ for the bone and the soft tissue in the
model are given in Fig. 10.3.
πœ‡ / π‘π‘š−1
Bone
Soft tissue
Calculate the ratio
3.0
0.27
𝑖𝑛𝑑𝑒𝑛𝑠𝑖𝑑𝑦 π‘œπ‘“ π‘₯ − π‘Ÿπ‘Žπ‘¦ π‘π‘’π‘Žπ‘š 𝑖𝑛𝑐𝑖𝑑𝑒𝑛𝑑 π‘œπ‘› π‘‘β„Žπ‘’ π‘šπ‘œπ‘‘π‘’π‘™
𝑖𝑛𝑑𝑒𝑛𝑠𝑖𝑑𝑦 π‘œπ‘“ π‘₯ − π‘Ÿπ‘Žπ‘¦ π‘’π‘šπ‘’π‘Ÿπ‘”π‘’π‘›π‘‘ π‘“π‘Ÿπ‘œπ‘š π‘‘β„Žπ‘’ π‘šπ‘œπ‘‘π‘’π‘™
for
(i)
the beam AB,
(ii)
the beam CD.
ratio = .................................................. [2]
ratio = .................................................. [2]
(c) Use your answers in (b) to suggest why, for this model, an X-ray image with good contrast
may be obtained.
.........................................................................................................................................................
................................................................................................................................................... [1]
10
Answer:
(c) much greater absorption in bone than in soft tissue