Service Manual HAMILTON-C2 0197 REF 160001 Software version 2.2.x 624165/07 | August 2018 Service Manual HAMILTON-C2 August 2018 624165/07 © 2018 Hamilton Medical AG. All rights reserved. Printed in Switzerland. No part of this publication may be reproduced, stored in a database or retrieval system, or transmitted in any form or by any means, electronic, mechanical, or by photocopying, recording, or otherwise, without prior written permission of Hamilton Medical AG. This document may be revised, replaced, or made obsolete by other documents by Hamilton Medical AG at any time and without notice. Ensure that you have the most current applicable version of this document; if in doubt, contact the technical support department of Hamilton Medical AG, Switzerland. While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment. Nothing in this document shall limit or restrict in any way Hamilton Medical AG’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Hamilton Medical AG has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including software) described herein. Manufacturer Hamilton Medical AG Via Crusch 8, CH-7402 Bonaduz, Switzerland Phone: (+41) 58 610 10 20 Fax: (+41) 58 610 00 20 info@hamilton-medical.com www.hamilton-medical.com Distributor in USA Hamilton Medical, Inc. 4990 Energy Way, P.O. Box 30008 Reno, NV 89520 Phone: (775) 858-3200 Toll-free: (800) 426-6331 Fax: (775) 856-5621 marketing@hamilton-medical.net The equipment must be operated, serviced, or upgraded only by trained professionals. Hamilton Medical AG’s sole responsibility with respect to the equipment and its use is as stated in the limited warranty provided in the device Operator’s Manual. Hamilton Medical AG shall not be liable for any loss, cost, expense, inconvenience, or damage that may arise out of misuse of the product, or if non-Hamilton Medical AG parts were used when replacing parts, or if serial numbers were amended, deleted, or removed. If returning parts to Hamilton Medical AG, be sure to use the standard Hamilton Medical returned goods authorization (RGA) procedure. Disposal of parts shall follow all local, state, and federal regulation with respect to environmental protection. Hamilton Medical AG will make available, on request, circuit diagrams, component parts lists, descriptions, calibration instructions, or other information that will assist appropriately trained personnel to repair those parts of the equipment designated by Hamilton Medical AG to be repairable. For all proprietary as well as third-party trademarks used by Hamilton Medical AG, see www.hamilton-medical.com/ trademarks. Product and/or company names marked with a § symbol may be the trademarks and/or registered trademarks of their respective owners, including but not limited to Aerogen§, Nihon Kohden§, Masimo§, Masimo SET§, Masimo rainbow SET§, Respironics§. 4 English | 624165/07 Table of Contents Table of Contents Chapter 1 Chapter 2 Chapter 3 Conventions.................................................................................................................. 13 1.1 Notes, Cautions and Warnings ............................................................................................................ 14 1.2 Typographic Conventions .................................................................................................................... 14 1.3 Expressions.......................................................................................................................................... 15 1.4 Foreword ............................................................................................................................................ 16 HAMILTON-C2 Overview ............................................................................................... 19 2.1 Front Components Overview ............................................................................................................... 20 2.2 Rear Components Overview ................................................................................................................ 21 2.3 Interaction Panel Internal Components Overview ................................................................................. 22 2.4 Ventilation Unit Internal Components Overview................................................................................... 23 2.4.1 Top Section ................................................................................................................................ 24 2.4.2 Middle Section ........................................................................................................................... 25 2.4.3 Bottom Section........................................................................................................................... 26 Pneumatics: Overview and Theory of Operation............................................................. 27 3.1 Overview ............................................................................................................................................. 28 3.2 Blower Module Overview..................................................................................................................... 29 3.3 Oxygen Mixer Block Assembly Overview .............................................................................................. 30 3.4 Inspiratory Valve Overview................................................................................................................... 31 3.5 Ambient Valve Overview...................................................................................................................... 32 3.6 Expiratory Valve Overview.................................................................................................................... 33 3.7 Patient Flow Sensor Overview .............................................................................................................. 34 3.8 Oxygen Sensor Overview ..................................................................................................................... 35 3.9 Principal Gas Flow Overview ................................................................................................................ 36 3.9.1 Components that Manage the Principal Gas Flow....................................................................... 37 3.9.2 Principal Gas Flow ...................................................................................................................... 38 3.9.3 Components for the Ambient State Gas Flow ............................................................................. 39 3.9.4 Components for the Over-pressure Relief Gas Flow .................................................................... 40 3.9.5 Components for the Nebulizer Gas Flow..................................................................................... 41 3.10 Chapter 4 Ventilation Control Principle ................................................................................................................ 42 3.10.1 Pressure Sensor Assembly ........................................................................................................... 43 3.10.2 Flow Sensors used for Flow Measurement Overview ................................................................... 44 3.10.3 Flow Restrictors used for Flow Reduction Overview..................................................................... 45 3.11 Gas Rinse Flows................................................................................................................................... 46 3.12 Measurements of Pneumatic Sensors................................................................................................... 47 3.13 Components that Measure Gas Flow and Pressure............................................................................... 48 Electronics: Component Functions Overview.................................................................. 49 4.1 Overview ............................................................................................................................................. 50 4.2 Interaction Panel Electronics Components............................................................................................ 51 4.3 Ventilation Unit Electronics Components ............................................................................................. 52 4.3.1 Mainboard MSP160200.............................................................................................................. 52 Hamilton Medical | HAMILTON-C2 Service Manual 5 Table of Contents Chapter 5 5.2 4.3.4 Voltage Distribution and Test Points ........................................................................................... 56 4.3.5 Power Supply Components Overview (MSP160600 Revision 02) ................................................. 58 Rechargeable Lithium Ion Battery......................................................................................................... 60 5.1.1 Introduction ............................................................................................................................... 60 5.1.2 Replacing the Battery.................................................................................................................. 62 5.1.3 Storage Hints.............................................................................................................................. 62 5.1.4 Backup Battery ........................................................................................................................... 62 5.1.5 Battery Life Expectancy / Li-Ion Battery Maintenance................................................................... 62 5.1.6 Lithium Ion Battery Maintenance ................................................................................................ 63 Battery Charger / Calibrator................................................................................................................. 64 5.2.1 Battery Charger / Calibrator Installation ...................................................................................... 65 5.2.2 Battery Charging ........................................................................................................................ 65 5.2.3 Battery Recharge Time................................................................................................................ 65 5.4 Battery Re-calibration .......................................................................................................................... 68 5.4.1 General Information ................................................................................................................... 69 5.4.2 Re-calibration Time..................................................................................................................... 70 5.4.3 Re-calibration Description ........................................................................................................... 71 5.4.4 Rechargeable Lithium Ion Battery Capacity ................................................................................. 72 Lithium Ion Battery Handling Precautions............................................................................................. 73 Preventive Maintenance and Testing.............................................................................. 75 Overview ............................................................................................................................................. 76 6.1.1 Introduction ............................................................................................................................... 76 6.1.2 Checking the Software Level ...................................................................................................... 76 6.1.3 Items Required for Preventive Maintenance and Testing.............................................................. 77 6.1.4 Overview of Preventive Maintenance and Testing ....................................................................... 77 6.2 Hospital Preventive Maintenance ......................................................................................................... 78 6.3 Engineer Preventive Maintenance ........................................................................................................ 79 Electrical Safety ............................................................................................................. 81 7.1 Electrical Safety Overview .................................................................................................................... 82 7.2 Electrical Safety Tests........................................................................................................................... 82 7.2.1 Preparation for Tests................................................................................................................... 82 7.2.2 Perform the Tests ....................................................................................................................... 83 7.2.3 Device Type ................................................................................................................................ 83 7.2.4 Internal Resistance Checks.......................................................................................................... 84 7.2.5 Electrical Safety Tests - IEC 62353............................................................................................... 87 Service Software............................................................................................................ 91 8.1 6 Mainboard Components............................................................................................................. 54 Evaluation of the Battery State of Health ............................................................................................. 66 6.1 Chapter 8 4.3.3 5.3 5.5 Chapter 7 Mainboard MSP160382.............................................................................................................. 53 Lithium Ion Battery ........................................................................................................ 59 5.1 Chapter 6 4.3.2 Introduction ........................................................................................................................................ 92 English | 624165/07 Table of Contents 8.2 Functions of the Service Software ........................................................................................................ 93 8.3 Structure of the Service Software......................................................................................................... 94 8.4 Starting the Service Software............................................................................................................... 95 8.5 Service Software Screen Layout ........................................................................................................... 96 8.6 Making Screenshots ............................................................................................................................ 97 8.7 Ventilator Info ..................................................................................................................................... 98 8.8 Technical State .................................................................................................................................... 99 8.8.1 Hardware Version Tab ................................................................................................................ 100 8.8.2 Software Version Tab ................................................................................................................. 101 8.8.3 Modify Tab................................................................................................................................. 102 8.8.4 Transfer Tab ............................................................................................................................... 104 8.9 Service Timer ....................................................................................................................................... 105 8.9.1 Service Timer Tab ....................................................................................................................... 105 8.9.2 Blower Timer Tab ....................................................................................................................... 107 8.10 Real Time Clock (RTC).......................................................................................................................... 108 8.11 Battery Data ........................................................................................................................................ 109 8.12 Tests Overview .................................................................................................................................... 110 8.13 Tests / Calibration Screen..................................................................................................................... 111 8.14 Adjustments / Calibration .................................................................................................................... 112 8.14.1 Calibration Tab........................................................................................................................... 113 8.14.2 Show Tab ................................................................................................................................... 134 8.14.3 Default Settings Tab ................................................................................................................... 140 8.15 Component Tests ................................................................................................................................ 142 8.15.1 Electronics Tab ........................................................................................................................... 145 8.15.2 User Interface Tab ...................................................................................................................... 161 8.15.3 Pneumatics 1 Tab ....................................................................................................................... 163 8.15.4 Pneumatics 2 Tab ....................................................................................................................... 178 8.16 System Test ......................................................................................................................................... 194 8.16.1 Pressure...................................................................................................................................... 196 8.16.2 Flow ........................................................................................................................................... 198 8.16.3 O2 Mixer .................................................................................................................................... 200 8.16.4 Tightness.................................................................................................................................... 202 8.16.5 Alarming .................................................................................................................................... 204 8.17 Sensor Data......................................................................................................................................... 207 8.17.1 Sensor 1 ..................................................................................................................................... 208 8.17.2 Sensor 2 ..................................................................................................................................... 209 8.17.3 Sensor 3 ..................................................................................................................................... 210 8.17.4 Sensor Status.............................................................................................................................. 211 8.18 Log/Config Files................................................................................................................................... 212 8.18.1 Event Log ................................................................................................................................... 213 8.18.2 Service Log ................................................................................................................................. 214 Hamilton Medical | HAMILTON-C2 Service Manual 7 Table of Contents 8.18.3 8.19 Chapter 9 Export (Instrument Report and Events) ........................................................................................ 215 Software Updates................................................................................................................................ 218 8.19.1 Software Update from Version < 2.2.0 ....................................................................................... 218 8.19.2 Software Update from Version 2.x to 2.2.5 or Higher ................................................................. 221 8.20 Pre-operational Checks........................................................................................................................ 226 8.21 General Tests....................................................................................................................................... 227 8.22 CO2 Sensor Accuracy Check................................................................................................................ 230 8.23 Nurse Call Functional Check ................................................................................................................ 240 8.24 Final Tests............................................................................................................................................ 242 8.25 Set the Service Timer ........................................................................................................................... 242 8.26 Check the Battery Data........................................................................................................................ 242 Technical Faults .............................................................................................................243 9.1 Technical Faults Overview .................................................................................................................... 244 9.2 Troubleshooting .................................................................................................................................. 246 Chapter 10 Components Removal/Assembly....................................................................................247 10.1 Notes on Maintenance and Replacement............................................................................................. 248 10.2 Securing the Power Cord..................................................................................................................... 249 10.3 Interaction Panel Components Removal/Assembly ............................................................................... 250 10.3.1 Ventilation Unit Top Cover Removal/Assembly ............................................................................ 250 10.3.2 Interaction Panel Small Rear Cover Removal/Assembly ................................................................ 251 10.3.3 Interaction Panel Removal/Assembly ........................................................................................... 252 10.3.4 Interaction Panel Large Rear Cover Removal/Assembly ................................................................ 253 10.3.5 Backlight Driver Board Removal/Assembly................................................................................... 254 10.3.6 Front Panel Board Removal/Assembly.......................................................................................... 255 10.3.7 LCD Display Removal/Assembly................................................................................................... 258 10.3.8 LCD Display Backlight Removal/Assembly.................................................................................... 259 10.3.9 P&T Control Knob Encoder Removal/Assembly............................................................................ 261 10.4 Ventilation Unit Components Removal/Assembly ................................................................................. 262 10.4.1 Backup Battery Pack Removal/Assembly...................................................................................... 262 10.4.2 Filter Cover Removal/Assembly ................................................................................................... 265 10.4.3 HEPA Filter Removal/Assembly .................................................................................................... 266 10.4.4 High Pressure Oxygen Connector Removal/Assembly .................................................................. 267 10.4.5 Front and Rear Covers Removal/Assembly................................................................................... 268 10.4.6 ESM Module Removal/Assembly ................................................................................................. 269 10.4.7 Ventilation Unit Mainboard Removal/Assembly ........................................................................... 270 10.4.8 Top Foam Removal/Assembly...................................................................................................... 271 10.4.9 Cooling Fan Removal/Assembly .................................................................................................. 273 10.4.10 Blower Module Removal/Assembly ............................................................................................. 274 10.4.11 Power Supply Removal/Assembly ................................................................................................ 275 10.4.12 Middle Foam Removal/Assembly................................................................................................. 276 10.4.13 Pressure Sensor Assembly Removal/Assembly.............................................................................. 278 8 English | 624165/07 Table of Contents 10.4.14 Inspiratory Valve Removal/Assembly............................................................................................ 280 10.4.15 Qvent Flow Sensor Removal/Assembly ........................................................................................ 282 10.4.16 Ambient Valve Removal/Assembly .............................................................................................. 284 10.4.17 Oxygen Sensor Block Removal/Assembly..................................................................................... 286 10.4.18 Rinse Flow Block Removal/Assembly ........................................................................................... 288 10.4.19 Expiratory Valve Removal/Assembly ............................................................................................ 291 10.4.20 Battery Compartment Removal/Assembly ................................................................................... 292 10.4.21 Oxygen Mixer Block Assembly Removal/Assembly ....................................................................... 294 10.4.22 Nebulizer Valve Removal/Assembly ............................................................................................. 296 10.4.23 Bottom Foam Removal/Assembly ................................................................................................ 297 10.5 Ventilation Unit Components, Tubings and Cables Assembly ............................................................... 298 10.5.1 Bottom Foam Assembly .............................................................................................................. 298 10.5.2 Mains Power Cable Assembly ..................................................................................................... 299 10.5.3 DC Input Power Cable Assembly................................................................................................. 300 10.5.4 Oxygen Mixer Block Assembly Mounting .................................................................................... 301 10.5.5 Oxygen Mixer Valve Cable Assembly........................................................................................... 302 10.5.6 Nebulizer Cable Assembly........................................................................................................... 303 10.5.7 Battery Compartment Assembly ................................................................................................. 304 10.5.8 Expiratory Valve Assembly .......................................................................................................... 305 10.5.9 Rinse Flow Block and Tubing Assembly ....................................................................................... 306 10.5.10 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly............................................. 309 10.5.11 Oxygen Sensor Block and Tubing Assembly ................................................................................ 312 10.5.12 Pressure Sensor Module Assembly .............................................................................................. 313 10.5.13 Middle Foam Assembly............................................................................................................... 314 10.5.14 Power Supply Assembly .............................................................................................................. 315 10.5.15 Cooling Fan Assembly ................................................................................................................ 316 10.5.16 Blower Module Assembly ........................................................................................................... 317 10.5.17 Top Foam and Mainboard Assembly ........................................................................................... 319 Chapter 11 Maintenance Tools and Test Equipment ........................................................................323 11.1 Overview ............................................................................................................................................. 324 11.2 Standard Tools .................................................................................................................................... 324 11.3 Special Tools........................................................................................................................................ 325 11.3.1 Digital Voltmeter ........................................................................................................................ 325 11.3.2 Electrical Safety Tester ................................................................................................................ 325 11.3.3 ESD (ElectroStatic Discharge) Protection...................................................................................... 326 11.3.4 Two Pin Cylindrical Service Tool .................................................................................................. 327 11.4 Test Equipment ................................................................................................................................... 328 11.5 Calibration of Test Equipment ............................................................................................................. 332 Chapter 12 Spare Parts ....................................................................................................................333 12.1 Introduction to Major Components ..................................................................................................... 334 12.2 Trolley PN 160170 ............................................................................................................................... 335 Hamilton Medical | HAMILTON-C2 Service Manual 9 Table of Contents 12.3 Trolley PN 160150 ............................................................................................................................... 336 12.4 Major Components of the Interaction Panel......................................................................................... 337 12.5 Major Components of the Ventilation Unit .......................................................................................... 338 12.5.1 Ventilation Unit Covers............................................................................................................... 339 12.5.2 Ventilation Unit Top Section ....................................................................................................... 340 12.5.3 Ventilation Unit Middle Section .................................................................................................. 341 12.5.4 Ventilation Unit Bottom Section.................................................................................................. 342 12.6 Interaction Panel Parts ......................................................................................................................... 343 12.6.1 Interaction Panel Covers ............................................................................................................. 343 12.6.2 Backlight Driver Board ................................................................................................................ 344 12.6.3 Front Panel Board ....................................................................................................................... 345 12.6.4 LCD Display and Touch Screen.................................................................................................... 346 12.6.5 P&T Control Knob Encoder ......................................................................................................... 347 12.7 Ventilation Unit Parts........................................................................................................................... 348 12.7.1 Top Cover .................................................................................................................................. 348 12.7.2 Front Cover ................................................................................................................................ 349 12.7.3 Rear Cover ................................................................................................................................. 350 12.7.4 Ventilation Unit Base .................................................................................................................. 351 12.7.5 Backup Battery Pack Compartment............................................................................................. 352 12.7.6 Expiratory Valve Assembly .......................................................................................................... 353 12.7.7 Mixer Block Assembly ................................................................................................................. 354 12.8 Detailed Parts Descriptions................................................................................................................... 355 12.8.1 Tubing and Fittings..................................................................................................................... 355 12.8.2 Clamps and Fasteners................................................................................................................. 357 12.8.3 Stickers and Labels ..................................................................................................................... 358 12.8.4 Pneumatic Parts and Assemblies ................................................................................................. 359 12.8.5 Flow Sensors .............................................................................................................................. 361 12.8.6 Assembled Components............................................................................................................. 362 12.8.7 Colored Metal Rings and Controls .............................................................................................. 363 12.8.8 Rubber Seals and Grommets....................................................................................................... 363 12.8.9 Metal Brackets and Frame Components...................................................................................... 364 12.8.10 Electrical/Electronic Cables.......................................................................................................... 366 12.8.11 Electronic Printed Circuit Boards ................................................................................................. 369 12.8.12 Electrical/Electronic Devices......................................................................................................... 372 12.8.13 External Covers and External Hardware....................................................................................... 374 12.8.14 Screws........................................................................................................................................ 376 12.8.15 Standoffs.................................................................................................................................... 379 12.8.16 Washers ..................................................................................................................................... 380 12.8.17 Screw Nuts ................................................................................................................................. 380 Chapter 13 Schematics ....................................................................................................................381 13.1 10 Block Diagram ..................................................................................................................................... 382 English | 624165/07 Table of Contents 13.2 Wiring Diagram ................................................................................................................................... 383 Chapter 14 Software Revisions, Features and Compatibility .............................................................385 14.1 Introduction ........................................................................................................................................ 386 14.2 Software Version Information .............................................................................................................. 386 Chapter 15 Hardware Revisions, Features and Compatibility ............................................................397 15.1 Introduction ........................................................................................................................................ 398 15.2 Component History and Improvements................................................................................................ 398 15.2.1 Interaction Panel......................................................................................................................... 398 15.2.2 Ventilation Unit .......................................................................................................................... 402 Chapter 16 Glossary ........................................................................................................................415 16.1 Definitions of Expression...................................................................................................................... 416 16.2 Alarm Codes ....................................................................................................................................... 421 16.2.1 Patient Alarms - 100000 Alarm Codes ........................................................................................ 421 16.2.2 Technical Alarms - 200000 Alarm Codes .................................................................................... 423 16.2.3 Technical Failure Ending in Safety Mode - 300000 Alarm Codes ................................................. 429 16.2.4 Technical Failure Ending in Ambient Mode - 400000 Alarm Codes ............................................. 431 16.2.5 500000 Alarm Codes ................................................................................................................. 433 Chapter 17 Test Report....................................................................................................................435 17.1 Summary of Test Report ...................................................................................................................... 436 Chapter 18 Environmental Requirements .........................................................................................439 18.1 Environmental Requirements ............................................................................................................... 440 Chapter 19 Document History .........................................................................................................441 19.1 Document History................................................................................................................................ 442 Hamilton Medical | HAMILTON-C2 Service Manual 11 Table of Contents 12 English | 624165/07 1 Conventions 1.1 Notes, Cautions and Warnings ............................................................................................................. 14 1.2 Typographic Conventions..................................................................................................................... 14 1.3 Expressions .......................................................................................................................................... 15 1.4 Foreword ............................................................................................................................................. 16 13 1 Conventions 1.1 Notes, Cautions and Warnings WARNING A WARNING alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the use or misuse of the device. CAUTION A CAUTION alerts the user to the possibility of a problem with the device associated with its use or misuse, such as device malfunction, device failure, damage to the device, or damage to other property. NOTICE A NOTE emphasizes information of particular importance. 1.2 Typographic Conventions Table 1-1. Typographic Conventions Effect Example Function Bold Configuration Marks text quoted directly from the ventilator screen. Bold TRIGGER Marks text that is quoted directly from: • The ventilator case • A touch key • A printed circuit board Number sequence 1. Step one in a sequence 2. Step two in a sequence Letter sequence A identifies part A B identifies part B Blue text 1.2 14 Organizes the performance of actions into steps. Used in photos and illustrations to identify the topic being discussed and relates to a specific text. If you are using a PDF file to view this, you can hyperlink to the reference by clicking on these items. English | 624165/07 1 Expressions 1.3 Expressions Table 1-2. Expressions Expression Example Explanation Activate Activate LED ON/OFF. Using the P&T Control Knob, you must first select the LED ON/OFF Button on the ventilator's screen, and then press the P&T Control Knob. The button on the screen changes its appearance, so that it looks “pressed”. It now performs its function (turning the LED ON in this case). Sometimes you are told to “activate and set” a field. In this case you first activate the field, and then turn the P&T Control Knob to set a value. Deactivate Deactivate LED ON/OFF. With LED ON/OFF still selected and activated, you must press the P&T Control Knob again. The button on the screen changes its appearance, so that it looks “unpressed”. It stops performing its function (turning the LED OFF in this case). Select Select LED ON/OFF. Touch the Touch Screen to select the LED ON/OFF Button. De-select De-select LED ON/OFF. Touch the Touch Screen to deselect the LED ON/OFF Button. Pressure Patient pressure is 80 mbar. Pressure refers to the amount of pressure above ambient pressure. If the patient pressure (Ppat) is 80 mbar, it means the pressure is 80 mbar above the ambient (room) pressure. Software version 0.6.0 The ventilator contains a memory device that holds software identified by a version number. Update This kit or software enables an upgrade to implement a new function. An update is an improvement to an existing function. An update normally involves only software. A software update is indicated by an increment of the last digit of the three digit software version number (e.g. 1.2.1 to 1.2.2). Upgrade This kit or software enables an upgrade to implement a new function. An upgrade is the addition of new functions to a device. There are three ways to perform an upgrade: • Add a hardware item that offers additional functions. • Upgrade to a higher software revision. Depending on how important is the software upgrade, it is either marked by an increment on the first or on the second digit (e.g. 1.2.0 to 1.3.0 or 1.2.0 to 2.0.0). • Upgrade to a higher type of software. * With this kit, you can update or upgrade from software version. Hamilton Medical | HAMILTON-C2 Service Manual Unless otherwise stated, a syntax variable (“wild card”) indicates the use of any alphanumeric character. 15 1 Conventions 1.4 Foreword WARNING To prevent possible patient injury, disconnect the patient from the ventilator before you start service or maintenance. Service the HAMILTON-C2 only as described in this manual, using only parts approved or supplied by Hamilton Medical. Incorrect parts, components or assemblies could result in patient injury. See available Spare Parts Section 12. Hamilton Medical does not offer any warranty on incorrectly used parts. The Service Manual is for: Engineers who have successfully completed a Hamilton Medical service training course for the HAMILTON-C2. Training courses are held regularly in Bonaduz, Switzerland, at Hamilton Medical’s headquarters, and at other locations throughout the world. For more information, visit the Hamilton Medical AG Partner-net. NOTICE If you have questions about testing or any part of this manual, contact Hamilton Medical AG Technical Support. The HAMILTON-C2 Service Manual contains: The architecture and components of the HAMILTON-C2. In addition, information on testing, troubleshooting and repairing the HAMILTON-C2. Appendices. The HAMILTON-C2 Service Manual does not contain: Information about operating the HAMILTON-C2. See the HAMILTON-C2 Operator’s Manual for operating instructions. This HAMILTON-C2 Service Manual covers: The HAMILTON-C2 Software version 1.1.1 and higher versions. 16 English | 624165/07 1 Foreword The HAMILTON-C2 Service Manual Structure: Section Function Your Responsibility HAMILTON-C2 Overview Section 2 This section explains the theory behind the HAMILTON-C2. You should fully understand this section. Pneumatics: Overview and Theory of Operation Section 3 This section explains each component and the gas flows, flow measurements, and pressure measurements in the pneumatic circuits. You should be able to name and explain the functions of all the major components. Electronics: Component Functions Overview Section 4 This section explains the basic functions of the printed circuit boards, Printed circuit boards are not repaired in the field. You should be able to identify all circuit boards, and know where they are positioned in the HAMILTON-C2. Lithium Ion Battery Section 5 This section explains the use, care and maintenance of the Lithium Ion Battery Pack. You have to understand the safety concerns and hazards, and know how to perform charging and calibration of the Lithium Ion Battery Pack. Preventive Maintenance and Testing Section 6 This section gives a schedule for maintenance. You have to be familiar with the maintenance schedule for the HAMILTON-C2. Hospital Preventive Maintenance Section 6.2 This section gives the Hospital Preventive Maintenance details. You should be able to perform this maintenance and ascertain if this maintenance is being performed regularly. Engineer Preventive Maintenance Section 6.3 This section gives Engineer Preventive Maintenance details. You have to be able to perform all the tasks in this section. Electrical Safety Overview Section 7.1 This section lists further test's you must perform on the HAMILTON-C2 before you start the Service Software test's. You have to be able to use this section to test the HAMILTON-C2. Service Software Section 8 This section explains how you perform the test's that are built into the software of the HAMILTON-C2. You have to know how to perform all the appropriate test's. Technical Faults Section 9.1 This section gives an overview of the alarm indications. You should know how to interpret technical faults to isolated faulty components. Components Removal/Assembly Section 10 This section explains how to remove and assemble each major component. You have to be able to use this section to make repairs. Maintenance Tools and Test Equipment Section 11 This section lists the equipment you require to work on the HAMILTON-C2. Check this section to make sure you have the correct tools and test equipment. Spare Parts Section 12 Information resource. You only require this section when you must order spare parts. Schematics Section 13 This section includes many of the schematics produced by Hamilton Medical for internal use. You are sometimes directed to this section when reading in other parts of the manual. Software Revisions, Features and Compatibility Section 14.1 This section informs about the software versions that were introduced over time. You only require this section when the software must be updated or when the update is recommended. Hardware Revisions, Features and Compatibility Section 15 This section informs about the hardware components that have changed over time and about its compatibility. You only require this section when parts need to be replaced. Glossary Section 16 Glossary contains Definitions of Expressions and Alarm Codes Helps you to understand the ventilator and the alarms. Test Report Section 17.1 Test Report pages for the Service Software section. Complete the report when using the service software tests. Hamilton Medical | HAMILTON-C2 Service Manual 17 1 18 Conventions English | 624165/07 2 HAMILTON-C2 Overview 2.1 Front Components Overview ................................................................................................................ 20 2.2 Rear Components Overview ................................................................................................................. 21 2.3 Interaction Panel Internal Components Overview.................................................................................. 22 2.4 Ventilation Unit Internal Components Overview ................................................................................... 23 19 2 HAMILTON-C2 Overview 2.1 Front Components Overview Figure 2-1. Front Components Overview A Alarm Lamp G Front Panel Keys B USB Socket H P&T (Press and Turn) Control Knob C 10.4” TFT Display with Touch Screen and Backlight I Patient Breathing Circuit Connection to the Patient D Nebulizer Connection J Patient Breathing Circuit Connection from the Patient E Oxygen Sensor and Cover (Oxygen Sensor not shown) K Quick Lock to remove HAMILTON-C2 from the Trolley F Patient Flow Sensor Connections L Shelf Stand 20 English | 624165/07 Rear Components Overview 2 2.2 Rear Components Overview Figure 2-2. Rear Components Overview A Ethernet Connection H DC Input B RS232 Interface Connection I Lift Handle C Option Slot for Communication Board J Interaction Panel D Dust Air Filter and HEPA Filter Element K Cooling Fan E High Pressure Oxygen DISS or NIST Connection L Battery Compartment F Low Pressure Oxygen Connection M Ventilation Unit Filter Cover G Mains Power Inlet N Cable Holder Hamilton Medical | HAMILTON-C2 Service Manual 21 2 HAMILTON-C2 Overview 2.3 Interaction Panel Internal Components Overview Figure 2-3. Interaction Panel Internal Components Front Overview A Alarm Lamp LED’s C Front Panel Board B 10.4” TFT Display with Backlight D P&T (Press and Turn) Control Knob Encoder G Front Panel Board Figure 2-4. Interaction Panel Internal Components Rear Overview 22 E Loudspeaker F Backlight Inverter Board (Obsolete) English | 624165/07 Ventilation Unit Internal Components Overview 2 2.4 Ventilation Unit Internal Components Overview Figure 2-5. Ventilation Unit Internal Components Overview The Ventilation Unit is divided into 3 sections: A Top Section B Middle Section Hamilton Medical | HAMILTON-C2 Service Manual C Bottom Section 23 2 HAMILTON-C2 Overview 2.4.1 Top Section Figure 2-6. Top Section A Mainboard C Options Slot for Communication Board B ESM (Embedded System Module) with Shield D Top Foam Section Jumper position for the HAMILTON-C2 with Mainboard (MSP160382): For the HAMILTON-C2 use the two left pins. Figure 2-7. Jumper position for the HAMILTON-C2 24 English | 624165/07 Middle Section 2 2.4.2 Middle Section Figure 2-8. Ventilation Unit Internal Components Middle Section Front Overview A Blower Module B Middle Foam Section Figure 2-9. Ventilation Unit Internal Components Middle Section Rear Overview C Heat Exchanger F HEPA Filter D Power Supply G Cooling Fan E HEPA Filter Latch Hamilton Medical | HAMILTON-C2 Service Manual 25 2 HAMILTON-C2 Overview 2.4.3 Bottom Section Figure 2-10. Ventilation Unit Internal Components Bottom Section Front Overview A Pressure Sensor Assembly F Oxygen Sensor B Autozero Valves G QO2 Flow Sensor C Qvent Flow Sensor H Expiratory Valve D Inspiratory Valve I Bottom Foam Section E Ambient Valve Figure 2-11. Ventilation Unit Internal Components Bottom Section Rear Overview J Battery Module K Oxygen Mixer Block 26 L Battery Release Latches English | 624165/07 3 Pneumatics: Overview and Theory of Operation 3.1 Overview.............................................................................................................................................. 28 3.2 Blower Module Overview ..................................................................................................................... 29 3.3 Oxygen Mixer Block Assembly Overview ............................................................................................... 30 3.4 Inspiratory Valve Overview ................................................................................................................... 31 3.5 Ambient Valve Overview ...................................................................................................................... 32 3.6 Expiratory Valve Overview .................................................................................................................... 33 3.7 Patient Flow Sensor Overview............................................................................................................... 34 3.8 Oxygen Sensor Overview ...................................................................................................................... 35 3.9 Principal Gas Flow Overview ................................................................................................................. 36 3.10 Ventilation Control Principle ................................................................................................................. 42 3.11 Gas Rinse Flows ................................................................................................................................... 46 3.12 Measurements of Pneumatic Sensors ................................................................................................... 47 3.13 Components that Measure Gas Flow and Pressure ............................................................................... 48 27 3 Pneumatics: Overview and Theory of Operation 3.1 Overview WARNING Repair of parts, components or assemblies is not permitted by Hamilton Medical AG, as incorrectly repaired parts, components or assemblies could result in patient injury. See the available Spare Parts Section 12 This section introduces all the major pneumatic components in the ventilator. In addition, the gas flows and theory of operation are explained. NOTICE The figures in this section show exploded and transparent views of the ventilator components. The exploded diagrams may not always show components in their correct positions. Figure 3-1. Pneumatic System Drawing Air LowPressure Inlet O2 LowPressure Inlet O2 HighPressure Inlet Instrument Borders Filter Air Filter O2 Filter Air HEPA Sensor QO2 Sensor Pfilter Mixer valve O2 Noise dampening Nebulizer valve Sensor TBlower Sensor Pambient Blower Noise dampening Restrictor Nebulizer Sensor Tinst Heat exchanger Sensor Pvent_control Sensor Paw Sensor Pflowsensor Inspiratory valve Filter Rinseflow Sensor Qvent Autozero valve Sensor Pvent_monitor Oxygen sensor Autozero valve Autozero valve Ambient valve Standard rinse flow Internal expiratory valve Standard rinse flow Nebulizer proximal flow proximal pressure 28 Expiratory valve cover to patient English | 624165/07 Blower Module Overview 3 3.2 Blower Module Overview The Blower Module supplies the air and oxygen mixture for patient inspiration. Figure 3-2. Blower Module Overview A Blower Module C Dust Air Filter (not shown) B HEPA Filter Element D Blower and Motor Hamilton Medical | HAMILTON-C2 Service Manual 29 3 Pneumatics: Overview and Theory of Operation 3.3 Oxygen Mixer Block Assembly Overview The Oxygen Mixer Block Assembly controls the flow of oxygen into the Blower Assembly. Figure 3-3. Oxygen Mixer Block Assembly Overview A Shaped Tube from the QO2 Flow Sensor to the Blower Module E High Pressure Oxygen Control Valve B QO2 Flow Sensor F Outlet to supply Oxygen to the Front Panel Nebulizer Connection C Oxygen Mixer Block Assembly G High pressure Oxygen DISS or NIST Connection D Nebulizer Valve H Low pressure Oxygen Connection 30 English | 624165/07 Inspiratory Valve Overview 3 3.4 Inspiratory Valve Overview The Inspiratory Valve precisely controls the volume and pressure of the air/oxygen mixture from the Blower Module to the Patient Breathing Circuit. Figure 3-4. Inspiratory Valve Overview A The Air/Oxygen Mixture from the Blower Module D Pressure Spring B Inspiratory Valve E Inspiratory Valve Voice Coil Assembly C Inspiratory Valve Sealing Membrane F Qvent Flow Sensor Hamilton Medical | HAMILTON-C2 Service Manual 31 3 Pneumatics: Overview and Theory of Operation 3.5 Ambient Valve Overview The Ambient Valve enables the patient to inhale ambient air, if there is a fatal failure on the ventilator (Ambient Mode). It is opened if there is a fault on the ventilator, allowing the patient to inspire ambient air. Figure 3-5. Ambient Valve Overview A The Air/Oxygen Mixture from the Inspiratory Valve to the Patient Breathing Circuit D Ambient Valve Solenoid B Ambient Valve Assembly E Outlet connection to the Patient Breathing Circuit C Ambient Valve Sealing Membrane 32 English | 624165/07 Expiratory Valve Overview 3 3.6 Expiratory Valve Overview WARNING Never attach a Spirometer or any other device or tube to the exhaust port of the Expiratory Valve. This will cause the ventilator to lose full control of PEEP/CPAP. The Expiratory Valve enables gas to escape from the Patient Breathing Circuit in a controlled manner, allowing the patient to exhale. The Expiratory Valve: • Maintains PEEP/CPAP if required - to do this, the action of the Expiratory Valve is closely synchronized with the action of the Inspiratory Valve • In Ambient State, the Expiratory Valve is open Figure 3-6. Expiratory Valve Overview A Expiratory Valve Assembly E Expired gas from the Patient Breathing Circuit B Plunger F Expiratory Valve Voice Coil Assembly C Silicone Membrane G Expiratory Valve Exhaust D Inlet connection from the Patient Breathing Circuit Hamilton Medical | HAMILTON-C2 Service Manual 33 3 Pneumatics: Overview and Theory of Operation 3.7 Patient Flow Sensor Overview The HAMILTON-C2 accurately measures flow, volume, and pressure in the patient’s airway with the Patient Flow Sensor. The area of the orifice changes depending on the flow rate. It opens progressively as the flow increases, creating a pressure drop across the orifice. The pressure difference is measured by a high-precision differential pressure sensor (Pflowsensor) located on the Pressure Sensor Assembly inside the ventilator. The pressure difference varies with flow (relationship determined during flow sensor calibration), so the patient’s flow is determined from the pressure drop. The HAMILTON-C2 calculates volume from the flow measurements. The Patient Flow Sensor is highly accurate even in the presence of secretions, moisture, and nebulized medications. The HAMILTON-C2 continuously flushes the sensing tubes with mixed gases (rinse flow) to prevent blockage. Figure 3-7. Patient Flow Sensor Overview A Blue Tube - Proximal side measured Pflowsensor Pressure and Paw Pressure D Two-way Gas Flow connection on the Distal side of the Patient Flow Sensor B Clear Tube - Distal side measures Pflowsensor Pressure E Two-way Gas Flow connection on the Proximal side of the Patient Flow Sensor C Variable Orifice Membrane NOTICE Ventilation is not totally dependent on the Patient Flow Sensor. If the Patient Flow Sensor malfunctions, patient ventilation continues. The ventilator has an internal flow measurement at the Inspiratory Valve for Inspiratory Flow. 34 English | 624165/07 Oxygen Sensor Overview 3 3.8 Oxygen Sensor Overview WARNING The ventilator should never be used for ventilating a patient without some means of monitoring the oxygen content in the gas mixture delivered to the patient. The ventilator can not be operated without an Oxygen Sensor installed. The Oxygen Sensor (A, see Figure 3-8 Oxygen Sensor Overview) is attached to the front left side of the ventilator behind the Oxygen Sensor Cover (B, see Figure 3-8 Oxygen Sensor Overview). It is used to monitor the oxygen concentration in the gases delivered to the patient. The Oxygen Sensor performs only a monitoring function, and can be disabled by the user, if required. Figure 3-8. Oxygen Sensor Overview A Oxygen Sensor B Oxygen Sensor Cover NOTICE Hamilton Medical Oxygen Sensors are available from Hamilton Medical AG only. The Hamilton Medical Oxygen Sensor produces a voltage between 11 and 13mV at 21% O2 which changes with oxygen concentration. Each unit comprises a teflon-bonded gold cathode and a lead anode, submerged in a liquid electrolyte solution. When oxygen diffuses through the flouropolymer membrane, the electrochemical reduction of oxygen on the cathode and the corresponding oxidation of the anode generates an electrical current that is proportional to the concentration of oxygen. Hamilton Medical | HAMILTON-C2 Service Manual 35 3 Pneumatics: Overview and Theory of Operation 3.9 Principal Gas Flow Overview The ventilator regulates the flow of gases at three points, as represented in the Principal Gas Flow Diagram (see Figure 3-9 Principal Gas Flow Diagram). The “Principal Gas Flow” is the flow of air and oxygen into the ventilator, out to the patient and directly at the Expiratory Valve back to the room. Room air (A) is aspirated, mixed with compressed oxygen (C) from an external high or low pressure supply (B) and compressed in the Blower Module (D). This flow of the air/oxygen mixture is managed by components that regulate the gases at three points: • Room air and oxygen mixture in the Blower Module (D regulation point 1). • Room air and oxygen mixture through the Inspiratory Valve and Patient Flow Sensor (E - regulation point 2). • Expired gas from the patient through the Expiratory Valve (F - regulation point 3). Expired gases from the patient are allowed to escape into the room. The following sections provides details of the components that manage the “Principal Gas Flow”. Figure 3-9. Principal Gas Flow Diagram NOTICE High-pressure-oxygen: Maximum Pressure 600 kPa / Maximum Flow 200 l/min Low-pressure-oxygen: Maximum Pressure 600 kPa / Maximum Flow 15 l/min 36 English | 624165/07 3 Components that Manage the Principal Gas Flow 3.9.1 Components that Manage the Principal Gas Flow The main components that control the flow of gases through the ventilator are shown in the Figure 3-10 Components that Manage the Principal Gas Flow . The components have the following functions: Figure 3-10. Components that Manage the Principal Gas Flow A Room Air Room Air. B Oxygen Oxygen from an external high or low pressure source. C Mixer Block Assembly Measures and Regulates the flow of the high pressure oxygen (B, see Figure 3-10 Components that Manage the Principal Gas Flow) into the Blower Module. D Blower Module Compresses the air and oxygen mixture with a Turbine Blower. E Inspiratory Valve and Qvent Flow Sensor Measures and Regulates the flow of the air/oxygen mixture to the patient and the pressure during inspiration. It regulates the Base Flow during expiration. F Patient Breathing Circuit Supplies air/oxygen gas mixture to the patient, and removes exhalation gases from the patient. G Patient Flow Sensor The Patient Flow Sensor monitors the flow of gases to and from the patient’s airway. H Patient The patient breathes in and out (inhales and exhales) with the assistance of the ventilator. I Expiratory Valve Regulates the flow of expiratory gases from the patient. The Expiratory Valve is nearly closed during inspiration and regulates PEEP during expiration. Hamilton Medical | HAMILTON-C2 Service Manual 37 3 Pneumatics: Overview and Theory of Operation 3.9.2 Principal Gas Flow The following flow occurs in the “Principal Gas Flow”: Figure 3-11. Principal Gas Flow A Room Air F From the Turbine Blower to the Inspiratory Valve B High or Low Pressure Oxygen Supply G Passes the Ambient Valve C HEPA Filter Assembly H Air/Oxygen Mixture into the Patient Breathing Circuit D Air and oxygen mixed in the Blower Module I Expired gas to the Expiratory Valve E Air/Oxygen Mixture flows through the Turbine Blower 38 English | 624165/07 Components for the Ambient State Gas Flow 3 3.9.3 Components for the Ambient State Gas Flow If there is a total breakdown of ventilation, the ventilator goes into Ambient State. In the Ambient State, the Ambient Valve is unpowered, allowing the patient to inspire ambient air through the Ambient Valve Sealing Membrane. In the Ambient State, the Expiratory Valve is also unpowered, enabling gas to escape from the Patient Breathing Circuit in a controlled manner, allowing the patient to exhale into the room. The Expiratory Valve Sealing Membrane has a one-way flow from the Patient Breathing Circuit. The one-way function of the Expiratory Valve make sure the patient does not inhale his exhaled CO2 (rebreathing). Figure 3-12. Components for the Ambient State Gas Flow A Ambient Valve Sealing Membrane E Expiratory Valve B Ambient air from the room F Expiratory Valve Membrane C Ambient Valve Solenoid G Expired gas to the room D Into the Patient Breathing Circuit H From the Patient Breathing Circuit Hamilton Medical | HAMILTON-C2 Service Manual 39 3 Pneumatics: Overview and Theory of Operation 3.9.4 Components for the Over-pressure Relief Gas Flow If the pressure from the Inspiratory Valve for the Patient Breathing Circuit is too high (15 mbar above Pmax), the Ambient Valve is driven to an open position to relieve the excess pressure to the room. It also functions as an additional backup to the pressure limit set by the user for the Patient Breathing Circuit. Figure 3-13. Components for the Over-pressure Relief Gas Flow A Over-pressure from the Inspiratory Valve D Ambient Valve Solenoid B Ambient Valve Sealing Membrane E Over-pressure from the Patient Breathing Circuit C Excess pressure to the room 40 English | 624165/07 Components for the Nebulizer Gas Flow 3 3.9.5 Components for the Nebulizer Gas Flow The ventilator provides oxygen with sufficient pressure to drive the Nebulizer Jar. The oxygen flow is reduced to approximately 8 liters/minute by means of an internal flow restrictor in the Oxygen Mixer Block Assembly. A valve attached to the Oxygen Mixer Block Assembly switches the flow of oxygen to the Nebulizer Jar through the output connection at the Front Panel. The pneumatic nebulizer is inactive when low pressure oxygen is used. Figure 3-14. Components for the Nebulizer Gas Flow A Nebulizer Front Panel connection C Nebulizer Valve B Tubing from Oxygen Mixer Block Assembly to the Nebulizer Front Panel connection D High Pressure Oxygen Inlet Hamilton Medical | HAMILTON-C2 Service Manual 41 3 Pneumatics: Overview and Theory of Operation 3.10 Ventilation Control Principle The HAMILTON-C2 is a pressure controlled ventilator. This means that the applied volume is controlled by a specific calculated pressure and not by a fixed, permanent volume only. The HAMILTON-C2 does not contain a tank. The required pressure is generated by a turbine, known as a Blower Module, which turns at approximately 35'000 RPM during normal ventilation. This pressure source provides a pressure that is always at least 10 mbar above the required ventilation pressure, e.g. PEEP + Pcontrol + 10 mbar = Pblower. To get the pressure in the Patient Breathing Circuit, the outlet pressure of the Inspiratory Valve is measured 1000 times per second, so the Expiratory Valve can react within a fraction of a second. On the other hand, the Expiratory Valve manages the outflow of the gas. It applies a known force to the membrane which has a direct effect on the gas in the Patient Breathing Circuit (back pressure). It never completely closes the outlet of the Patient Breathing Circuit during ventilation, so the membrane is floating all the time. The correct ventilation pressure is given by the close collaboration of the Inspiratory Valve, that applies the pressure into the Patient Breathing Circuit and the Expiratory Valve that stabilizes it by a controlled release of the gas. 42 English | 624165/07 3 Pressure Sensor Assembly 3.10.1 Pressure Sensor Assembly Figure 3-15. Pressure Sensor Assembly A Pressure Sensor Assembly E • Provides components for pressure measurements B Pvent_monitor Pressure Sensor • A measurement of the pressure in the Patient Breathing Circuit as measured at the Proximal Flow Sensor, in the proximal chamber attached to the blue (patient side) pressure-sensing tube F • A sensor which measures the pressure at the patient outlet, used for monitoring C Pvent_control Pressure Sensor Pflowsensor Pressure Sensor • A measurement of the pressure difference between the proximal and distal chambers of the Patient Proximal Flow Sensor Hamilton Medical | HAMILTON-C2 Service Manual Pvent_monitor Autozero Valve • Switches the Pvent_monitor air way pressure measured at the Patient Flow Sensor to allow for offset re-calculation to compensate for sensor drift for Pvent_control and Pvent_monitor G • A sensor which measures the pressure at the patient outlet, used for controlling D Paw Pressure Sensor Distal Autozero Valve • Switches the Pflowsensor Pressure Sensor to Ambient Air to allow for offset re-calculation to compensate for sensor drift H Proximal Autozero Valve • Switches the Pflowsensor and Paw Pressure Sensor to Ambient Air to allow for offset re-calculation to compensate for sensor drift 43 3 Pneumatics: Overview and Theory of Operation 3.10.2 Flow Sensors used for Flow Measurement Overview These Flow Sensors are used to measure the inspiration and expiration flow to/from the patient. Figure 3-16. Flow Sensors used for Flow Measurement Overview A The QO2 Flow Sensor measures the flow of oxygen into the Blower Module. It works with the Qvent Flow Sensor to control the Air/Oxygen Mixture. B The Qvent Flow Sensor measures the flow of the Air/Oxygen Mixture into the Patient Breathing Circuit 44 C Patient Flow Sensor with a variable orifice. The variable orifice works as a very flexible resistance to make flow/pressure relation proportional. English | 624165/07 Flow Restrictors used for Flow Reduction Overview 3 3.10.3 Flow Restrictors used for Flow Reduction Overview Figure 3-17. Flow Restrictors used for Flow Reduction Overview A A Restrictive “Orifice” for the Oxygen Flow to the Nebulizer Valve B Two precisely matched “Orifice” Restrictors for the Patient Flow Sensor Rinse Flow Hamilton Medical | HAMILTON-C2 Service Manual C A Restrictive “Screen” for the Patient Flow Sensor Rinse Flow supply from the Inspiratory Valve 45 3 Pneumatics: Overview and Theory of Operation 3.11 Gas Rinse Flows The rinse flow is provided on the blower side of the Inspiratory Block Assembly. A constant small flow of gas (approximately 11.5 ml/min at 35 mbar blower pressure) is provided through both tubes to the Patient Flow Sensor. The difference between the Proximal and Distal Rinse Flows are ±10%. The rinse flow eliminates the possibility of contamination from the patient, condensing humidity in the measuring tubes and cross-contamination to other patients. Figure 3-18. Components that generate the Patient Flow Sensor Rinse Flow A Gas from the Blower Module E Pflowsensor Sensor connections B Inspiratory Valve outlet to QVent Flow Sensor F Rinse Flow Assembly (including flow restrictor) C Tube outlet to the Oxygen Sensor Mounting Block G Front panel connections for the Patient Proximal Flow Sensor D Tube to the Rinse Flow Assembly 46 English | 624165/07 Measurements of Pneumatic Sensors 3 3.12 Measurements of Pneumatic Sensors The following table lists all the Pneumatic Sensors in the Ventilator: Table 3-1. Pneumatic Sensors in the Ventilator Name Location Comment Pvent_control Pressure Sensor Assembly Measures pressure at the Inspiratory Valve outlet for controlling Pvent_monitor Pressure Sensor Assembly Measures pressure at the Inspiratory Valve outlet for monitoring Paw Pressure Sensor Assembly Measures pressure at the Proximal Side of the Patient Flow Sensor Pflowsensor Pressure Sensor Assembly Measures differential pressure across the Patient Flow Sensor Pambient Mainboard Measures ambient pressure or the room pressure Pfilter Blower Module Measures pressure after the filter; used to indicate if the HEPA Filter needs to be replaced Qvent Ventilation Flow Sensor Measures the air/oxygen flow in the Patient Breathing Circuit. QO2 Oxygen Flow Sensor Measures the oxygen flow into the Blower Module Hamilton Medical | HAMILTON-C2 Service Manual 47 3 Pneumatics: Overview and Theory of Operation 3.13 Components that Measure Gas Flow and Pressure The Qvent Flow Sensor measures the flow of Air/Oxygen in the Patient Breathing Circuit. The Pvent_control measurement is used to control the Inspiratory Valve. The Pvent_monitor Sensor measures the pressure in the Patient Breathing Circuit. During normal ventilation, the Flow Sensor Autozero Valves open at calculated intervals to expose the Pflowsensor Pressure Sensor to ambient air for the Autozero offset compensation. Figure 3-19. Components that Measure Gas Flow and Pressure A Qvent Flow Sensor D • Measures the flow of Air/Oxygen to the Patient Breathing Circuit B Pvent_monitor Pressure Sensor • Monitor the pressure in the Patient Breathing Circuit C Pflowsensor Pressure Sensor • Measures the pressure difference between the two chambers of the Patient Flow Sensor, and enables the flow of gas to and from the patient to be calculated E Paw Pressure Sensor • Measures the pressure on the proximal side of the Patient Flow Sensor Pvent_control Pressure Sensor • Used to control the Inspiratory Valve NOTICE During the Autozero procedure, which occurs in a fraction of a second, the patient is ventilated normally. Since parameters such as expired tidal volume, resistance, compliance, and PEEP cannot be measured or monitored, the ventilator uses the most recently measured values until the procedure is complete. 48 English | 624165/07 4 Electronics: Component Functions Overview 4.1 Overview.............................................................................................................................................. 50 4.2 Interaction Panel Electronics Components ............................................................................................ 51 4.3 Ventilation Unit Electronics Components .............................................................................................. 52 49 4 Electronics: Component Functions Overview 4.1 Overview WARNING Service the ventilator only as described in this manual, using only parts approved or supplied by Hamilton Medical AG. Incorrectly repaired parts, components or assemblies could result in patient injury. See available Spare Parts Section 12. This section introduces the major electronic components in the ventilator. Where you require more information, cross-references direct you to other parts of this manual. This section does not include a theory of operation, because engineers do not require a detailed knowledge of board-level electronics to service and maintain the ventilator. All electronic failures are dealt with by replacing complete circuit boards. Any repairs at a lower level than the spare parts listed in the Spare Parts Section 12 is not permitted NOTICE Always send defective Printed Circuit Boards to Hamilton Medical AG with a completed Returned Goods ID Tag. Refer to PN 699138 ky2help User Guide, available as a download from the Hamilton Medical AG Partner-net. 50 English | 624165/07 Interaction Panel Electronics Components 4 4.2 Interaction Panel Electronics Components Figure 4-1. Interaction Panel Electronics Components A Front Panel Board E Loudspeaker and Microphone • Provides 7 Hard Keys for the operator with LED Indicators • Audible indication of alarms • Interfaces the P&T Control Knob Encoder • Microphone: • LED Colors available: – The Microphone monitors the operation of the Loudspeaker – Red – If the Loudspeaker does not function, audible alarm indication is transferred to the Buzzer – Green – Yellow • Voltage inputs: – 12 VDC for the Key and LED Communications Board – 5 VDC for the Backlight Inverter Board – 3.3 VDC for the P&T Control Knob Encoder and USB B Backlight Inverter Board (Obsolete) F • Converts 5 VDC (Input Voltage) to 1100 VAC for the Display Backlight • Used to: – Download Software from a USB Stick – Export Data to a USB Stick Refer to the Hardware Revisions, Features and Compatibility > Backlight Driver Board Section 15.2.1.3 C Alarm Lamp LED's • Indicates Alarm Conditions: – Yellow - Medium and Low Priority Alarms – Red - High Priority Alarms and Technical Faults USB Socket G P&T Encoder • Provides additional controls for interaction with various screen functions – -16 Encoder positions from center – +16 Encoder positions from center – Switch activation when the P&T Control Knob is depressed D Graphic LCD TFT VGA 10.4" NLT Display with Front Panel and LED Backlight • Graphical User Interface (GUI) Hamilton Medical | HAMILTON-C2 Service Manual 51 4 Electronics: Component Functions Overview 4.3 Ventilation Unit Electronics Components 4.3.1 Mainboard MSP160200 Figure 4-2. Mainboard MSP160200 A Mainboard MSP160200 O J11 − FFC Cable to Expiratory Valve (MSP160240) B ESM Module P P12 − Cable to Ambient Valve (MSP160290) • Embedded System Microprocessor Module • Power PC MPC5200 Processor C J1 − Ethernet Connection (for internal use only) Q P13 − Cable to Nebulizer Valve (PN 160400) D J2 − RS232 Connection R J14 − FFC Cable (PN 160355) to Pressure Sensor Assembly (MSP160300) E P25 − Cable from Power Supply for Fan Supply S P16 − FDC Cable (PN 160357) to Display F P3 − Cable (PN 160539) from Power Supply 24 VDC (MSP160600) T J17 − FFC Cable (PN 160356) to Key Panel for P&T and Key Panel G J4 − FFC Cable (PN 160351) to Battery Data U J18 − FFC Cable (PN 160356) to Key Panel for Backlight, Speaker and Microphone H P5 − Cable (PN 160372) to DC Input V P19 − Cable (PN 160387) to Qvent Flow Sensor (TSI, MSP399123 or Sensirion, MSP161657) I P6 − Cable (PN 160370) to Battery Power W P20 − Cable (PN 160354) to Oxygen Sensor (PN 396200) J P60 − Options Slot for Communication Board (for CO2, (SpO2 not implemented in HAMILTON-C2) and Nurse Call Options) X J21 − FFC Cable (PN 160353) to Blower Module Temperature Sensor K P7 − Cable to O2 Proportional Valve Y J22 − FFC Cable to Inspiratory Valve (MSP160230) L J9 − FFC Cable (PN 160347) to Filter Pressure Board Z P23 − Cable to Blower Module (MSP160250) M J8 − FFC Cable (PN 160359) to Pressure Sensor Board Binary Valves N P10 − Cable (PN 160387) to QO2 Flow Sensor (TSI, MSP399124 or Sensirion, MSP161657) 52 AA P28 − Cable to Fan 12V (PN 160346) English | 624165/07 4 Mainboard MSP160382 4.3.2 Mainboard MSP160382 Figure 4-3. Mainboard MSP160382 A Mainboard MSP160382 O P12 − Cable to Ambient Valve (MSP160290) B ESM Module P P13 − Cable to Nebulizer Valve (PN 160400) • Embedded System Microprocessor Module • Power PC MPC5123 Processor C J1 − Ethernet Connection (for internal use only) Q J14 − FFC Cable (PN 160355) to Pressure Sensor Assembly (MSP160300) D J2 − RS232 Connection R J23 − Not used for the HAMILTON-C2 E P3 − Cable (PN 160539) from Power Supply 24VDC (MSP160600) S P16 − FDC Cable (PN 160386) to Display F J4 − FFC Cable (PN 160351) to Battery Data T J17 − FFC Cable (PN 160356) to Key Panel for P&T Encoder and Key Panel G P5 − Cable (PN 160372) to DC Input U J18 − FFC Cable (PN 160356) to Key Panel for Backlight, Speaker and Microphone H P6 − Cable (PN 160370) to Battery Power V P19 − Cable (PN 160387) to Qvent Flow Sensor (TSI, MSP399123 or Sensirion, MSP161657) I P60 − Options Slot for Communication Board (for CO2, SpO2 (Not Implemented in HAMILTON-C2) and Nurse Call Options) W P20 − Cable (PN 160354) to Oxygen Sensor (PN 396200) J P7 − Cable to O2 Proportional Valve X J21 − FFC Cable (PN 160353) to Blower Module Temperature Sensor Board K J9 − FFC Cable (PN 160347) to Filter Pressure Board Y J22 − FFC Cable to Inspiratory Valve (MSP160230) L J8 − FFC Cable (PN 160359) to Pressure Sensor Board Binary Valves Z J23 − Cable to Blower Module (MSP160250) M P10 − Cable (PN 160387) to QO2 Flow Sensor (TSI, MSP399124 or Sensirion, MSP161657) N J11 − FFC Cable to Expiratory Valve (MSP160240) Hamilton Medical | HAMILTON-C2 Service Manual AA P28 − Cable to Fan 12V (PN 160346) 53 4 Electronics: Component Functions Overview 4.3.3 Mainboard Components Table 4-1. Mainboard Components Mainboard Serves as the Carrier Board for the Processor Board, Power Management System and a Connector Board for all Sensors, Valves and Periphery Components Controls • Blower • ON/OFF Key • Hard Key and Lamps • Key Panel • TFT Display • Ambient Valve • Expiratory Valve • Mixer Valves • Nebulizer Valve • Emergency Status Indicator Interfaces with the following Ventilation Unit boards: • Power Supply • Pressure Sensor Assembly • Filter Board • Smart Batteries Sensor Monitoring: • Blower Temperature • Instrument Temperature • Pvent_control • Pvent_monitor • Pflowsensor • Pambient • Qvent Flow Sensor • QO2 Flow Sensor • Paw • Oxygen Sensor Input Voltage supplied from the Power Supply: • 24 VDC ±5% 6.6A/150W (Range 22.8V...25.2V) Output Voltages created by Switching Regulators on the Ventilation Unit Mainboard: 24 VDC • Blower 12 VDC • Autozero Valves • Inspiratory Valve • Expiratory Valve • O2 Inlet Proportional Valve • Ambient Valve • Nebulizer Valve • Alarm Lamps (IP) • Speaker • Cooling Fan 5 VDC • Qvent Flow Sensor • QO2 Flow Sensor • DC/AC Backlight Inverter or LED Backlight Driver • USB Port 54 English | 624165/07 Mainboard Components Output Voltages created by Switching Regulators on the Ventilation Unit Mainboard: 3.3 VDC 4 • Qvent Flow Sensor • QO2 Flow Sensor • Pressure Sensors • P&T Control Knob • ESM Module • Blower Temperature Sensor • HEPA Filter Temperature Sensor Hamilton Medical | HAMILTON-C2 Service Manual 55 4 Electronics: Component Functions Overview 4.3.4 Voltage Distribution and Test Points Mainboard Voltage Inputs: Table 4-2. Mainboard Voltage Inputs: Signal Name Voltage Level Voltage Range Current Watts +24V_PS (from power supply) 24 VDC 22.8 - 25.2 VDC 7A at 22.8 V; 5.7 at 25.2 150W +24V_DcIn (from 20 V 100W DC Input) 12-24 VDC 11.0 - 27.0 VDC 9.1A at 11 V; 5A at 20V 100W 7.5A at >20 V; 5.6A at 27V (Battery starts charging from DC Input if voltage is above 20 V) 150W Battery 1 (Primary Battery) 14.4 VDC 12.0 - 16.8 VDC 6A at 16.8 V; 8.3A at 12V 100W Battery 2 (Optional Battery) 14.4 VDC 12.0 - 16.8 VDC 6A at 16.8 V; 8.3A at 12V 100W Mainboard Voltage Outputs and Test Points: Table 4-3. Voltage Test Points P44 Signal Name Voltage Level Voltage Range GND Pin Pin Signal Name o o GND Voltage Level Voltage Range +5V_Usb 5.0 VDC 4.75 - 5.25 V o o +24V_DcIn 12-24 VDC 11.0 - 27.0 VDC +5V 5.0 VDC 4.60 - 5.40 V o o +24V_PS 24 VDC 22.8 - 25.2 VDC +3V3 3.3 VDC 3.10 - 3.60 V o o +24V_In +3V3_BackUp 3.3 VDC 3.00 - 3.60 V o o +31V_Blower +3V3_Stby 3.3 VDC 3.00 - 4.20 V o o +24V 26.80 VDC 24 - 29 V +3V3_Ooc 3.3 VDC 3.00 - 4.20 V o o +14V4_Bat2 14.4 VDC 12.0 - 16.8 VDC +3V_Ref 3.0 VDC 2.91 - 3.09 V o o +14V4_Bat1 14.4 VDC 12.0 - 16.8 VDC +2V5_Ref 2.5 VDC 2.475 - 2.525 V o o +12V 12 VDC 9.00 - 13.2 V o 1 GND GND P44 56 English | 624165/07 Voltage Distribution and Test Points 4 Table 4-4. Voltage Test Points P42/P41 Pin Signal Name Voltage Level o GND o Pventmonitor 0.472 o O2 o Voltage Range Pin Signal Name Voltage Level Voltage Range o GND 0.372 - 0.572 o Iexpvalve 1.25 0.227 (when the oxygen sensor is disconnected) 0.223 - 0.231 o Iinspvalve not used Pflowsensor 1.25 1.15 - 1.35 o Iblower not used o Paw 0.472 0.372 - 0.572 o Pventcontrol 0.472 0.372 - 0.572 o SoundLevel 1.15 (if silent) 1.11 - 1.19 o TO2 1.25 (measured at 25°C, when the oxygen sensor is disconnected) 1.15 - 1.35 o n.c. o QO2 (TSI) 0.25 0.05 - 0.45 QO2 (Sensirion) 0.7 0.5 - 0.9 o Tvent 1.25 1.15 - 1.35 o Qvent (TSI) 0.25 0.05 - 0.45 Qvent (Sensirion) 0.7 0.5 - 0..9 o Tblower 0.5 V o ExtInstrument not used o GND P42 Hamilton Medical | HAMILTON-C2 Service Manual + 10mV/ºC 1 1.05 - 1.45 GND P41 57 4 Electronics: Component Functions Overview 4.3.5 Power Supply Components Overview (MSP160600 Revision 02) Figure 4-4. Power Supply Overview A Power Supply C • Provides power conversion from the Mains Power Power Outlet Connector • Cable from the Power Supply to the Mainboard P25 Connector and Mainboard P3 Connector • Input Power: – 100 to 240 VAC – 50 to 60 Hz • Output Voltage: – +24 VDC ±5% B Power Inlet Connector • Cable from AC Power Inlet Power Supply Voltage Input and Output: Table 4-5. Power Supply Voltage Input and Output Signal Name Mains Input Voltage Voltage Level [V] 100 - 240 VAC Voltage Range [V] 90 - 264 VAC 50 - 60 Hz Output Voltage 58 +24 VDC 22.8 - 25.2 English | 624165/07 5 Lithium Ion Battery 5.1 Rechargeable Lithium Ion Battery ......................................................................................................... 60 5.2 Battery Charger / Calibrator ................................................................................................................. 64 5.3 Evaluation of the Battery State of Health .............................................................................................. 66 5.4 Battery Re-calibration ........................................................................................................................... 68 5.5 Lithium Ion Battery Handling Precautions ............................................................................................. 73 59 5 Lithium Ion Battery 5.1 Rechargeable Lithium Ion Battery WARNING It is mandatory that the HAMILTON-C2 is operated with at least one battery installed. NOTICE • At high operating temperature, the battery does not start charging. For example after transport and high temperature. • The backup battery is intended for short-term use only. It is not intended to be a primary power source. • Hamilton Medical recommends that the ventilators batteries be fully charged before you ventilate a patient. If the batteries are not fully charged and the AC power fails, always pay close attention to the level of battery charge. The Backup Battery provides backup power to the HAMILTON-C2. 5.1.1 Introduction Two backup batteries, one standard and one optional, are used the HAMILTON-C2 when the primary power source is low or unavailable. When the primary power source (either AC mains power or a DC power supply) fails, the ventilator automatically switches to backup battery operation with no interruption in ventilation. An alarm sounds to signal the switchover. To reset the alarm, silence the alarm to confirm notification of the power system change. If the optional battery (battery 2) is available and adequately charged, the ventilator switches to this battery first. When battery 2 is depleted or unavailable, the ventilator switches to the standard battery (battery 1). The batteries power the ventilator until the primary power source is adequate or the battery is depleted. As a further safeguard, the HAMILTON-C2 provides a low battery alarm. It also has a capacitor-driven backup buzzer that sounds continuously for at least 2 minutes when battery power is completely lost. The ventilator recharges the batteries whenever the ventilator is connected to either AC or >20VDC, with or without the ventilator power switched on. The battery charge indicator (see Figure 5-1 Power Source Symbols and Battery Charge Indicator) lights to show that the batteries are being charged. The power source symbols in the bottom right-hand corner of the screen show the available power sources. A frame around a symbol indicates the current ventilator power source. Green indicates the level of battery charge. 60 English | 624165/07 Introduction 5 Figure 5-1. Power Source Symbols and Battery Charge Indicator Hamilton Medical | HAMILTON-C2 Service Manual 61 5 Lithium Ion Battery 5.1.2 Replacing the Battery WARNING When a battery has reached its end-of-life, the capacity is reduced and the battery weakens. The use of a ventilator with weak batteries can cause the device to report incorrect fuel gauge information, leading to a possible unexpected ventilation stop if a power failure occurs. The batteries are hot-swappable; that is, they can be replaced while the ventilator is operating. Replace with a newly charged battery. For battery replacement refer to the Backup Battery Pack Removal/Assembly Section 10.4.1. 5.1.3 Storage Hints • Never store a battery in a discharged state. • Recharge batteries every 6 months. • Store below 25°C. The ideal range is between 5°C and 21°C. • If the storage temperature exceeds 25°C during this 6-month period, the shelf life for the batteries is reduced. Be sure to recharge these batteries periodically. • Storage place should be free from vibration, dust, direct sunlight, moisture and corrosive gases. 5.1.4 Backup Battery CAUTION It is mandatory that the HAMILTON-C2 is operated with at least one battery installed. NOTICE - The backup battery is intended for short-term use only. It is not intended to be a primary power source. - Hamilton Medical recommends that the ventilators batteries be fully charged before you ventilate a patient. If the batteries are not fully charged and the AC power fails, always pay close attention to the level of battery charge. 5.1.5 Battery Life Expectancy / Li-Ion Battery Maintenance • The lifetime of a battery is defined by the capacity loss. A Li-ion batteries capacity is lowered due to the number of charge cycles, as well as elapsed time. Given normal storage and use, you can expect the battery to deliver 75% or more of its initial capacity after 300 charge/discharge cycles or within 2.5 years of its manufacture date. • Hamilton Medical recommends replacing a battery when the capacity is reduced to 80% of its designed capacity. (HAMILTON-C2: 6500 mAh). • Cycle numbers (cy.) and full charge capacity (fc.) are provided on PAGE No. 1102 (See the Lithium Ion Battery Maintenance Section 5.1.6) in the Service software (SW ≥ 1.1.X). • The battery’s manufacture date is stamped on top of the battery connector (IEYYWWVV). The date is encoded as follows: In the example IE110221, the year is 2011 (11), the week is (02), and the software version is (2.1). • The HAMILTON-C2 software version > 1.1.x checks the battery capacity and does not accept weak batteries. Batteries with a capacity loss of more than 25% generate the error message Battery calibration required. 62 English | 624165/07 5 Lithium Ion Battery Maintenance 5.1.6 Lithium Ion Battery Maintenance 5.1.6.1 Maintenance Hints • When the battery is used in a device, a calibration cycle on the battery charger (PN 369136) is recommended once per year during preventive maintenance. • Periodical use of the batteries can extend the battery life time and allows the battery a calibration cycle each time it is recharged. • When a ventilator is not in use, connect it to mains power for 1 hour every week. Make sure to disconnect it afterwards. 5.1.6.2 Warranty • Batteries are consumable parts. Therefore they are excluded from the standard warranty - excepting out-of-box failures! 5.1.6.3 Disposal • Regulations vary country by country. Dispose of the batteries in accordance with local regulations. NOTICE When the battery reaches its end-of-life, the capacity is reduced and the battery becomes unreliable. The use of a ventilator with such batteries may cause wrong fuel gauge information leading to an unexpected ventilation stop when a power fail occurs! Hamilton Medical | HAMILTON-C2 Service Manual 63 5 Lithium Ion Battery 5.2 Battery Charger / Calibrator WARNING • Do not expose the charger or power supply to water or conductive liquids, this is not a sealed case. • Do not open the charger or power supply case. There are no user serviceable parts inside. • Do not cover the fan exhaust or obstruct the airflow, as this will cause overheating. Use only the manufacturer’s 24V 2.5A power supply and observe terminal polarity • Place the charger in a cool spot, away from external heat sources. CAUTION During re-calibration, the battery connector and base of the charger may become warm. The Battery Charger / Calibrator (PN 369104, see Figure 5-2 Battery Charger / Calibrator, please make sure you are using a charger with Revision 07 or higher) is a standalone desktop battery charger with the added ability to re-calibrate the fuel gauge on smart batteries. It works with all Hamilton Medical smart batteries. Figure 5-2. Battery Charger / Calibrator Package contents: • One desktop Battery Charger / Calibrator (PN 369104) • Three plastic spacers to modify the battery recess to suit your battery size • One 24V 2.5A AC:DC Power Supply, universal mains input • One mains cable – European chargers (CH5000E) are packed with a European 2-pin Mains Power Cord – UK chargers (CH5000U) are packed with a UK 3-pin Mains Power Cord – North American chargers (CH5000A) are packed with a US 3-pin Mains Power Cord • Instruction manual NOTICE Chargers shipped with medical grade power supplies carry the suffix “M” – eg. CH5000AM. 64 English | 624165/07 Battery Charger / Calibrator Installation 5 5.2.1 Battery Charger / Calibrator Installation 1. Place the charger unit on a flat, level surface away from sources of heat and moisture. 2. Plug the DC connector from the power supply into the connector socket on the back of the charger. 3. Please apply the supplied cable for connecting the battery charger to the AC mains supply. 5.2.2 Battery Charging 1. Place the battery into the battery bay. 2. Make sure that the 5-way connector is fully seated. NOTICE The charger will automatically begin to charge the battery. The LEDs in the status window will provide status information as detailed in the table. Table 5-1. Battery Lights Signal Name Voltage Level Green flashing Charging Green solid Fully charged Blue flashing In calibration mode Blue solid Calibration complete Red flashing Fuel gauge calibration required Red solid Error 5.2.3 Battery Recharge Time The times given below are for a full charge from 0% to 100% state of charge. Table 5-2. Battery Recharge Time Battery Chemistry Li Ion Battery Model MSP369106 Hamilton Medical | HAMILTON-C2 Service Manual Typical Recharge Time 4.5 hours 65 5 Lithium Ion Battery 5.3 Evaluation of the Battery State of Health Knowing the Battery State of Health (SoH) is very important to guarantee a safe operation of the ventilator running on battery power. With the following information, you will be able to evaluate the Battery State of Health. With this awareness, you can decide when a battery needs to be replaced. You normally perform these checks once a year during Preventive Maintenance. Read the four battery parameters as follows: Table 5-3. Evaluation of the Battery State of Health Parameter Where to look for: How to access and how to read: Full Charge Capacity FCC [mAh] See the technical state or read the value in the exported and printed instrument report. Go into service software, technical state, the value is shown on the bottom. Calendar age AGE [Y] See printed on battery. The battery’s date of manufacture is stamped on top of the battery connector (IEyywwvv). A coding of e.g. IE144514 means Year 2014, Week 45, Version 1.4, SN 67, 6.8Ah Cycles CYC [1] See the technical state or read the value in the exported and printed instrument report. Go into service software, technical state, value shown on the bottom. Error ERR (%) See the technical state or read the value in the exported and printed instrument report. Go into service software, technical state, the value is shown on the bottom. Figure 5-3. Battery State of Health 66 English | 624165/07 Evaluation of the Battery State of Health 5 Compare the values of the battery with the following chart: If one or more parameters is out of the green range, perform the recommended action. Figure 5-4. 4 Key Values to Evaluate the Battery State of Health Hamilton Medical | HAMILTON-C2 Service Manual 67 5 Lithium Ion Battery 5.4 Battery Re-calibration The battery calibration is done automatically. If the battery fuel gauge needs re-calibrating, the red LED will flash when you put the battery into the battery charger. This indicator provides feedback on the accuracy of the fuel gauge and avoids unnecessary battery calibration cycles. You have the option to calibrate the fuel gauge and charge the battery or to only charge the battery. This option is given because a recalibration cycle is longer than a charge cycle. To recalibrate the battery, press the calibrate button on the front of the charger. The blue calibration LED will flash to indicate that the battery is undergoing the re-calibration cycle. There may be a short delay before the calibration begins. During calibration the discharge resistors will heat up and the fan will operate to maintain the temperature within acceptable limits. At the end of the procedure the blue calibration LED will illuminate indicating a fully charged, fully calibrated battery. NOTICE No action is necessary if only a recharge is required. The charger will automatically begin to charge the battery. The most common cause of calibration failure is overheating of the battery during discharge. Keep the charger away from direct sunlight or heat sources. CAUTION During recalibration, the battery connector and base of the charger may become warm. 68 English | 624165/07 5 General Information 5.4.1 General Information If the battery is in need of fuel gauge re-calibration, the red LED will flash upon insertion of the battery into the battery charger. This indicator provides feedback to the user on the accuracy of the fuel gauge and avoids unnecessary battery calibration cycles. The user has the option to calibrate the fuel gauge and charge the battery, or only charge the battery. This option is given because a re-calibration cycle is longer than a charge cycle. For re-calibration of the battery, please press the "calibration button" on the front of the charger. NOTICE No additional action is required for re-charging the battery. The process starts automatically. The blue calibration LED will flash to indicate the re-calibration cycle. There may be a short delay before the calibration begins. During calibration the discharge resistors will heat up and the fan will operate to maintain temperature within acceptable limits. At the end of this procedure the blue LED will stay constant indicating a fully charged, fully calibrated battery. CAUTION The most common cause of calibration failure is overheating of the battery during discharge. Please keep the charger away from direct sunlight or heat sources. Hamilton Medical | HAMILTON-C2 Service Manual 69 5 Lithium Ion Battery 5.4.2 Re-calibration Time The re-calibration cycle begins by discharging the residual capacity. Then a calibration charge is delivered to the battery. This is followed by a calibration discharge. Finally the battery is given a regular charge. A calibration cycle will be faster if the battery is fully discharged to begin with. Re-calibration time (5-4) is governed by the battery voltage and capacity. Larger batteries, and low voltage batteries will take longer to re-calibrate. Calibration is initiated each time the re-calibration button is pressed, so it is not recommended to press the re-calibration button part way through the re-calibration cycle. Calibration begins each time the re-calibration button is pressed. Do not press the re-calibration button part way through the re-calibration cycle. The re-calibration cycle is as follows: 1. Discharge of any residual capacity. 2. A calibration charge is delivered to the battery. 3. Discharge of calibration capacity. 4. The battery is given a regular charge. Table 5-4. Battery Re-Calibration Times Battery Li-ion 70 Battery Model MSP369106 Minimum Recalibration Time 14.6 hours Maximum Recalibration Time 19.2 hours English | 624165/07 5 Re-calibration Description 5.4.3 Re-calibration Description Impedance tracking fuel gauges retain accuracy longer than coulomb-counters and can even self re-calibrate in use as long as there are periods of inactivity in the cycle so that the impedance measurement can be made more accurately. If no periods of rest are present in the cycle then the Impedance tracking fuel gauge will accumulate the Max-Error at a rate of 1% every 20 cycles. Impedance-Tracking fuel gauge re-calibration is achieved by charging the battery, allowing it to rest, discharging it allowing it to rest again. This algorithm can only be performed on the external charger/calibrator CH5000 (see the Re-calibration Time Section 5.4.2). So re-calibration is used to re-set the fuel gauge to match the actual capacity in the battery. In this way, even as the battery ages and things change, the accuracy and reliability of the fuel gauge can be retained throughout the life of the battery. Hamilton Medical | HAMILTON-C2 Service Manual 71 5 Lithium Ion Battery 5.4.4 Rechargeable Lithium Ion Battery Capacity Figure 5-5. Rechargeable Lithium Ion Battery Capacity The battery can directly display the capacity information. The battery capacity is displayed as the RSoC (Relative State of Charge). Each LED segment represents 25 percent of the full charge capacity. The LED pattern definition is given in the table below. The LED’s illuminate for 4 seconds following switch activation. If the battery voltage is too low, there will be no LED indication. Table 5-5. Battery Capacity Information Capacity LED Indicators # 1 2 3 Note 4 At or below 10% x Blinks 10% - 25% x Lit for 4 seconds. 26% - 50% x x 51% - 75% x x x 76% - 100% x x x Lit for 4 seconds. Lit for 4 seconds. x Lit for 4 seconds. Refer to the HAMILTON-C2 Operator's Manual > Electrical Specifications 72 English | 624165/07 5 Lithium Ion Battery Handling Precautions 5.5 Lithium Ion Battery Handling Precautions WARNING • Before using lithium ion batteries for the first time, carefully study this section, including all Cautions and Warnings. • Keep this section for future reference. WARNING • Do not use the battery for a purpose other than those specified. Otherwise, its performance will be compromised and/or its service life will be shortened. Depending on the equipment in which the battery is used, excessively high current can flow through the battery, possibly damaging it and leading to acid leakage, overheating, smoke emission, bursting and/ or ignition. • Do not use the battery in combination with primary batteries (such as dry-cell batteries) or batteries of different capacities or brands. Otherwise, the battery can be overdischarged during use, or overcharged during recharging, abnormal chemical reactions may occur, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • Do not use or leave the battery near a heat source such as a fire or a heater (+80ºC or higher). If the resin separator should be damaged due to overheating, internal short-circuiting may occur to the battery, possibly leading to acid leakage, smoke emission, bursting and/or ignition. • Do not put the battery into a microwave oven or pressurized container. Rapid heating or disrupted sealing can lead to acid leakage, overheating, smoke emission, bursting and/or ignition. • Do not discard the battery into fire or heat it. Otherwise, its insulation can melt down, its gas release vent or safety features will be damaged and/or its electrolyte can ignite, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition on it. • Do not immerse the battery in liquid, or allow it to get wet. Otherwise, the protective features in it can be damaged, it can be charged with extremely high current and voltage, abnormal chemical reactions may occur in it, possibly leading to acid leakage, smoke emission, bursting and/or ignition. • Do not pierce the battery with a nail or other sharp objects, strike it with a hammer, or step on it. Otherwise, the battery will become damaged and deformed, internal short-circuiting can occur, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • Do not strike or throw the battery. The impact might cause leakage, overheating, smoke emission, bursting and/or ignition. Also, if the protective feature in it becomes damaged, it could become charged with an extremely high current and voltage, abnormal chemical reactions can occur, which can lead to acid leakage, overheating smoke emission, bursting and/or ignition. • Do not disassemble or modify the battery. The battery is equipped with built-in safety/protection features. Should these features be disabled, the battery can leak acid, overheat, emit smoke, burst and/or ignite. • If the battery leaks and the electrolyte gets into the eyes, do not rub them. Instead, rinse the eyes with clean running water and immediately seek medical attention. Otherwise, eye injury may result. • Do not use an apparently damaged or deformed battery. Otherwise, acid leakage, overheating, smoke emission, bursting and/or ignition of the battery may occur. • Consider consolidating these three into: Do not expose the battery to fire, flames, or excessive heat such as is generated by soldering, welding, radiators and so on. If the battery gives off a bad odor, melts, becomes discolored or deformed, leaks electrolyte fluid, appears in any way abnormal at any time, remove it from the equipment as the battery can leak acid (electrolyte fluid), emit smoke, burst and/or ignite. • Do not reverse the positive (+) and negative (-) terminals. Otherwise, during recharging, the battery will be reversecharged, abnormal chemical reactions then may occur, or excessively high current can flow during discharging, leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • The positive (+) and negative (-) terminals are arranged in a particular orientation. Do not force the connection if you cannot easily connect the battery terminals to the battery charger or other equipment. Confirm that the terminals are correctly oriented. Reversing the terminals will result in reverse-charging, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • Do not connect the positive (+) and negative (-) terminals with a metal object such as wire. Do not transport or store the battery together with metal objects such as necklaces, hair pins, etc. Otherwise, short-circuiting will occur, over-current will flow, causing the battery to leak acid, overheat, emit smoke, burst and/or ignite or the metal object such as wire, necklace or hair pin can generate heat. • Do not connect the battery to an electrical outlet, vehicle cigarette lighter, etc. When subjected to large voltage, overcurrent can flow on the battery, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • Do not recharge the battery near fire or above 45°C. Otherwise, hot temperatures can trigger its built-in protective features, inhibiting recharging or can damage the built-in protective features, causing it to be charged with an extremely Hamilton Medical | HAMILTON-C2 Service Manual 73 5 Lithium Ion Battery high current and voltage and, as a result, abnormal chemical reactions can occur in it, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • To recharge the battery, use the battery charger specifically designed for the purpose and observe the recharging conditions, refer to 5.2. A recharging operation under non-conforming recharging conditions (higher temperature and larger voltage/current than specified, modified battery charger, etc.) can cause the battery to be overcharged, or charged with extremely high current. An abnormal chemical reaction can occur in the battery, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • If recharging operation fails to complete even when a specified recharging time has elapsed, immediately stop further recharging. Otherwise, acid leakage, overheating, smoke emission, bursting and/or ignition can occur. CAUTION • If you find rust, a bad odor, overheating and/or other irregularities when using the battery for the first time, return it to Hamilton Medical AG. • If acid leaking from the battery comes into contact with your skin or clothing, immediately wash it away with running water. Otherwise, skin inflammation can occur. • The battery incorporates built-in safety devices. Do not use it in a location where static electricity (greater than the manufacturer’s guarantee) may be present. Otherwise, the safety devices can be damaged, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition. • Do not use or subject the battery to intense sun or above 65°C. Otherwise, acid leakage, overheating and/or smoke emission can occur. Also, its guaranteed performance will be lost and/or its service life will be shortened. • The guaranteed recharging temperature range is 0°C to +45ºC. A recharging operation outside this temperature range can lead to acid leakage and/or overheating of the battery and may cause damage to it. • Store the battery in a location where children cannot reach it. Also, make sure that no child takes the battery out of the battery charger or equipment. 74 English | 624165/07 6 Preventive Maintenance and Testing 6.1 Overview.............................................................................................................................................. 76 6.2 Hospital Preventive Maintenance .......................................................................................................... 78 6.3 Engineer Preventive Maintenance ......................................................................................................... 79 75 6 Preventive Maintenance and Testing 6.1 Overview WARNING − To prevent disease transmission, you must use personal protective equipment when handling contaminated bacterial filters or other patient accessories. Refer to the HAMILTON-C2 Operator’s Manual for instructions on sterilizing patient system parts. − Device is potentially contaminated − The device must be cleaned and disinfected to prevent the spread of infections and germs CAUTION Make sure to take full ESD (ElectroStatic Discharge) precautions before handling any electronic parts, or before opening the HAMILTON-C2. For more information refer to: − Electrical Safety Overview Section 7.1 − Standard Tools Section 11.2 − Special Tools Section 11.3 − Test Equipment Section 11.4 − You must complete a service training course for the HAMILTON-C2 with Hamilton Medical before undertaking the maintenance and testing procedures described in this manual. 6.1.1 Introduction All Preventive Maintenance and Testing must be performed: • After replacing any component • Once a year or once every 5000 operating hours, whichever comes first To perform Preventive Maintenance, perform all the steps shown in the Overview of Preventive Maintenance and Testing Section 6.1.4. 6.1.2 Checking the Software Level In general, Hamilton Medical AG recommends updating to the latest Software Version available. See the Hamilton Medical AG Partner-net. 76 English | 624165/07 Items Required for Preventive Maintenance and Testing 6 6.1.3 Items Required for Preventive Maintenance and Testing Table 6-1. Items Required for Preventive Maintenance and Testing Step Items Required, or Possibly Required Hospital Preventive Maintenance Section 6.2 • The HAMILTON-C2 Operator’s Manual or local-language equivalent Engineer Preventive Maintenance Section 6.3 • HEPA Filter (PN 160216) • Filter Set (containing 2 Dust Air Filter and 1 Fan Filter (PN 160215)) • Blower Module (MSP160250) • Lithium Ion Battery Pack(s) (MSP369106) • Battery Charger / Calibrator (PN 369104, Rev 07 or higher) • Oxygen Sensor (PN 396200) • HPO Inlet Filter Kit (PN 160497) • Complete Patient Breathing Circuit (Adult, PN 260086): – Test Lung with ET-Tube (PN 151815) and Adapter (PN 281420) – Inspiratory Filter (PN 279204) – Short Silicone Tube (PN 260100) NOTICE! For more details of parts, refer to the Spare Parts Section 12. Electrical Safety Tests Section 7.2 • Test equipment for the HAMILTON-C2; See the Hamilton Medical AG Partnernet > Technical Support > Recommended Spare Parts List > Recommended Test Equipment Service Software Section 8 NOTICE! For more details of Test Equipment, refer to the Test Equipment Section 11.4. 6.1.4 Overview of Preventive Maintenance and Testing Work methodically through the sections shown in the Overview of Preventive Maintenance and Testing Table 6-2 below. Maintenance and testing is not complete until all steps are performed successfully. If you make a replacement, you must go back to “Step 3” in the Overview of Preventive Maintenance and Testing Table 6-2 below. Table 6-2. Overview of Preventive Maintenance and Testing Step Task Where Found Time Required 1. Perform (or confirm it has been performed) the Hospital Preventive Maintenance Hospital Preventive Maintenance Section 6.2 5-15 min. 2. Perform the Engineer Preventive Maintenance Engineer Preventive Maintenance Section 6.3 10 min. 3. Perform parts replacements as necessary Components Removal/ Assembly Section 10 N/A 4. Perform the Electrical Safety Tests Electrical Safety Tests Section 7.2 10 min. 5. Perform the Service Software Checks Service Software Section 8 40 min. 6. Finish the testing by completing the tasks documented in the Tests, Calibrations and Utilities Section of the HAMILTON-C2 Operator’s Manual The HAMILTON-C2 Operator’s Manual or local-language equivalent 10 min. Hamilton Medical | HAMILTON-C2 Service Manual 77 6 Preventive Maintenance and Testing 6.2 Hospital Preventive Maintenance The Hospital Preventive Maintenance Table 6-3 shows the maintenance tasks that hospital staff must perform. It is copied from the English Version of the Operator’s Manual. Examine each ventilator for which you are responsible, and satisfy yourself that hospital staff are regularly performing these tasks. If necessary: • Perform the tasks yourself • Train staff how to perform these tasks Table 6-3. Hospital Preventive Maintenance Interval Between patients and according to hospital policy Part/Accessory Procedure Patient Breathing Circuit (including Mask, Inspiratory Filter, Flow Sensor, Nebulizer Jar, Expiratory Valve and Membrane) Replace with sterilized or new single use parts. Run the Tightness Test and the Flow Sensor Calibration as shown in the HAMILTON-C2 Operator’s Manual. Entire ventilator Run the pre-operational check as shown in the HAMILTON-C2 Operator’s Manual. Every 2 days or according to hospital policy Patient Breathing Circuit Empty any water from breathing tubes or water traps. Inspect parts for damage. Replace as necessary. Every month (or more often, if required) Air Intake Dust Filter and Fan Filter Set (5 pieces) (rear panel) Check for dust and lint. If needed, clean or replace as shown in the HAMILTON-C2 Operator’s Manual. Every 6 months (while the ventilator is in storage) Battery Recharge battery by plugging the Ventilator into AC Power for at least 4 hours. 78 English | 624165/07 6 Engineer Preventive Maintenance 6.3 Engineer Preventive Maintenance WARNING This section is not a stand-alone, independent part of the manual. Perform the tasks detailed here only as a part of, and as instructed. Refer to the Preventive Maintenance and Testing Overview Section 6.1 and Hospital Preventive Maintenance Section 6.2. Perform the Engineer Preventive Maintenance according to the table below: Table 6-4. Engineer Preventive Maintenance Schedule Interval Yearly or every 5000 hours, whichever comes first, or as necessary Part/Accessory Procedure Ventilator Perform service-related Preventive Maintenance. Calibrations and Tests Perform all calibrations and tests in the Service Software. (Refer to the Service Software Introduction Section 8.1). Pre-Operational Checks Perform the Pre-operational Checks as shown in the HAMILTON-C2 Operator’s Manual > Running the Pre-operational Check. Alarm Tests Perform the alarm checks as shown in the HAMILTON-C2 Operator’s Manual > Alarm Tests. HEPA Filter Replace as described in the HAMILTON-C2 Operator's Manual > Servicing the Air Intake and Fan Filters. HPO Inlet Filter Replace as described in the High Pressure Oxygen Connector Removal/Assembly Section 10.4.4. Oxygen Sensor Replace if depleted as shown in the HAMILTON-C2 Operator’s Manual Section > Replacing the Oxygen Sensor. Lithium Ion Battery Replace the Lithium Ion Battery (Refer to Replacing the Battery Section 5.1.2). Every 5 years (30,000 Hours) LCD Display Backlight Replace the LCD Display Backlight as described (Refer to LCD Backlight Removal/Assembly Section 10.3.8). When ebt or blower timer is ≥ 100%. Blower Module Replace the Blower Module when the Blower Timer has reached 100%(dynamic lifetime surveillance, typically after 8 years). Also replace the Blower Module when the prediction of Blower Timer in the next 6 months reaches 100% (Refer to Blower Module Removal/Assembly Section 10.4.10). Battery to be checked. Replace battery when (whichever comes first): • MSP369106 when capacity < 5280 mAh • Battery age > 3 Years • Cycles > 400 Hamilton Medical | HAMILTON-C2 Service Manual 79 6 80 Preventive Maintenance and Testing English | 624165/07 7 Electrical Safety 7.1 Electrical Safety Overview ..................................................................................................................... 82 7.2 Electrical Safety Tests ........................................................................................................................... 82 81 7 Electrical Safety 7.1 Electrical Safety Overview WARNING Electrical Safety Tests detailed in this section must be performed as part of, or as instructed in the Overview of Preventive Maintenance and Testing Section 6.1.4. In addition, to comply with IEC 62353, the Electrical Safety Test's must be performed after: − The Power Supply is replaced − The Mainboard is replaced − Removing any ground contact from the Ventilator − When performing Preventive Maintenance − Any repair − Installation Hamilton Medical AG performs a set of Electrical Safety Tests, as specified in IEC 60601-1, on all the ventilator and compressor units that it manufactures. Hamilton Medical AG performs these tests automatically, using the RIGEL 288 Safety Analyzer or a device like Metron QA90. As stated in the warning above, it is a legal necessity that, after performing the Preventive Maintenance, after a repair or after an adjustment described above, the Electrical Safety Tests are performed. 7.2 Electrical Safety Tests 7.2.1 Preparation for Tests • Multimeter • Safety Analyzer • Electrical Safety Test Cable Figure 7-1. RIGEL 288 Safety Analyzer 82 English | 624165/07 Perform the Tests 7 7.2.2 Perform the Tests The Electrical Safety Tests you must perform are explained in this section. If you have an automated safety device such as the RIGEL 288 Safety Analyzer used by Hamilton Medical, perform the automated tests in addition to the tests shown in this section. 7.2.3 Device Type For the IEC 62353 Regulations concerning medical devices, the HAMILTON-C2 is a: • Class 2 • Type B Device • Type BF if the Communication Board for Capnography is installed Hamilton Medical | HAMILTON-C2 Service Manual 83 7 Electrical Safety 7.2.4 Internal Resistance Checks For these tests, a Digital Voltmeter is required (refer to Digital Voltmeter Section 11.3.1) Check the resistance between the Earth Ground on the HAMILTON-C2 with the Test Connector, and other components as shown in Table 7-1 Electrical Tests. Table 7-1. Electrical Tests Check Minimal Acceptable Resistance Figure A. Ground (earth) pin to: Communications Interface Connector > 2.4MΩ See Figure 7-2 B. Ground (earth) pin to: Interaction Panel > 2.4MΩ See Figure 7-3 C. Ground (earth) pin to: High Pressure Oxygen > 2.4MΩ See Figure 7-4 D. Ground (earth) pin to: Low Pressure Oxygen > 2.4MΩ See Figure 7-5 Figure 7-2. Earth Ground to the Communication Interface Connector Resistance 84 English | 624165/07 Internal Resistance Checks 7 Figure 7-3. Earth Ground to the Interaction Panel Resistance Figure 7-4. Earth Ground to the High Pressure Oxygen Resistance Hamilton Medical | HAMILTON-C2 Service Manual 85 7 Electrical Safety Figure 7-5. Earth Ground to the Low Pressure Oxygen Resistance 86 English | 624165/07 Electrical Safety Tests - IEC 62353 7 7.2.5 Electrical Safety Tests - IEC 62353 It is a legal necessity, after a repair or adjustment that includes replacing the Power Supply, Mainboard or removing any of the internal earth connectors, perform an Electrical Safety Test on the ventilator. The Electrical Safety Test is performed with suitable equipment such as the RIGEL 288 Safety Analyzer (See Figure 7-6 RIGEL 288 Safety Analyzer). Details of these Automated Electrical Safety Tests depend on the test equipment used, it is impossible to offer any detailed descriptions. Figure 7-6. RIGEL 288 Safety Analyzer Specifications Table 7-2. IEC 62353 Specifications for Class II Type B and BF Devices Current in mA (RMS - Root Mean Square) Setup Equipment leakage (direct method)* Patient leakage current (direct method) AC* Type B Setup for Type B Device (no Communication Board Installed) Section 7.2.5.1 Type BF Setup for Type BF Device (with Communication Board for Capnography) Section 7.2.5.2 0.1 mA 0.1 mA – 5 mA * refer to IEC62353 All measured values must be documented using the measurement process and are considered reference values. Should the obtained values measured during the next maintenance cycle be within 90 to 100% of the permissible values, the reference values are to be used to evaluate the ventilators electrical safety. Definition Applied Parts of the medical device, which are designed to come into physical contact with the patient or parts that are likely to be brought into contact with the patient. Type B Applied Parts Type B Applied Parts are those parts, which are usually Earth referenced. Type B is used for applied parts that are generally not conductive and can be immediately released from the patient. Type BF Applied Parts Type BF Applied Parts are generally for devices that have conductive contact with the patient, or having medium or long term contact with the patient. For example, using capnostat and/or SPO2 finger probe on a patient. This applied part complies with a higher degree of protection against electric shock than Type B applied part. Hamilton Medical | HAMILTON-C2 Service Manual 87 7 Electrical Safety 7.2.5.1 Setup for Type B Device (no Communication Board Installed) NOTICE For detailed information regarding the use of your Electrical Safety Tester, please check the Users Manual for the device. 1. For setup for the RIGEL 288 Safety Analyzer Tests, attach the Ground Cable (E) from the USB Port (G) of the HAMILTONC2 to the 4mm Earth Bond Probe Socket of the RIGEL 288 Safety Analyzer (see Figure 7-7). 2. Connect the AC Mains Cable from the HAMILTON-C2 to the AC Connection at the front of the RIGEL 288 Safety Analyzer (see Figure 7-7). 3. Turn on the HAMILTON-C2 (Standby mode) for the electrical safety tests. Figure 7-7. Setup for Type B Device (no Communication Board Installed) A AC Power Cable Inlet (RIGEL 288 Safety Analyzer) E Electrical Safety Test Cable (PN 161618) B RIGEL 288 Safety Analyzer (Handheld Medical Electrical Safety Tester) F Power Cord (HAMILTON-C2) C 4mm Earth Bond Probe Socket (green) (RIGEL 288 Safety Analyzer) G USB Port D EUT Socket (RIGEL 288 Safety Analyzer) 88 English | 624165/07 7 Setup for Type BF Device (with Communication Board for Capnography) 7.2.5.2 Setup for Type BF Device (with Communication Board for Capnography) NOTICE For detailed information regarding the use of your Electrical Safety Tester, please check the Users Manual for the device. 1. Connect the RIGEL 288 Safety Analyzer to the AC Power Supply. 2. Connect the RIGEL 288 Safety Analyzer and the HAMILTON-C2 to each other as shown below (see Figure 7-8 Setup for Type BF Device (with Communication Board for Capnography). 3. Turn on the HAMILTON-C2 (Standby mode) for the electrical safety tests. Figure 7-8. Setup for Type BF Device (with Installed Communication Board for Capnography) A AC Power Cable Inlet (RIGEL 288 Safety Analyzer) G Communication Board B RIGEL 288 Safety Analyzer (Handheld Medical Electrical Safety Tester) H Power Plug C 4 mm Earth Bond Probe Socket (green) (RIGEL 288 Safety Analyzer) I USB Port D EUT Socket (RIGEL 288 Safety Analyzer) J CO2 Sensor Electrical Safety Test Cable (PN 159171) E Power Cord (HAMILTON-C2) K SPO2 (Not Implemented in HAMILTON-C2) Sensor Electrical Safety Test Cable (PN 159681) F Electrical Safety Test Cable (PN 161618) L AP-Box 331A700 (RIGEL 288 Safety Analyzer) Hamilton Medical | HAMILTON-C2 Service Manual 89 7 90 Electrical Safety English | 624165/07 8 Service Software 8.1 Introduction ......................................................................................................................................... 92 8.2 Functions of the Service Software......................................................................................................... 93 8.3 Structure of the Service Software ......................................................................................................... 94 8.4 Starting the Service Software ............................................................................................................... 95 8.5 Service Software Screen Layout ............................................................................................................ 96 8.6 Making Screenshots ............................................................................................................................. 97 8.7 Ventilator Info...................................................................................................................................... 98 8.8 Technical State..................................................................................................................................... 99 8.9 Service Timer........................................................................................................................................105 8.10 Real Time Clock (RTC) ..........................................................................................................................108 8.11 Battery Data.........................................................................................................................................109 8.12 Tests Overview .....................................................................................................................................110 8.13 Tests / Calibration Screen .....................................................................................................................111 8.14 Adjustments / Calibration.....................................................................................................................112 8.15 Component Tests.................................................................................................................................142 8.16 System Test..........................................................................................................................................194 8.17 Sensor Data .........................................................................................................................................207 8.18 Log/Config Files ...................................................................................................................................212 8.19 Software Updates ................................................................................................................................218 8.20 Pre-operational Checks ........................................................................................................................226 8.21 General Tests .......................................................................................................................................227 8.22 CO2 Sensor Accuracy Check ................................................................................................................230 8.23 Nurse Call Functional Check.................................................................................................................240 8.24 Final Tests ............................................................................................................................................242 8.25 Set the Service Timer ............................................................................................................................242 8.26 Check the Battery Data ........................................................................................................................242 91 8 Service Software 8.1 Introduction WARNING − Read the Preventive Maintenance and Testing Overview Section 6.1, before performing any of the tests in this section. − If one of the tests indicates that you must replace a part, do so immediately and update the Service Software Modify Tab (see the Modify Tab Section 8.8.3) and then repeat the complete series of tests. See contents in this section. CAUTION To prevent patient or ventilator contamination, always use a bacterial filter between the HAMILTON-C2 and the Inspiratory Limb of the Patient Breathing Circuit. − For troubleshooting see knowledge base − Record all results on the HAMILTON-C2 Summary of Test Report (see the Summary of Test Report Section 17.1) − Confirm that the Technical State is updated and that the device has been restarted after Technical State modifications (see the Modify Tab Section 8.8.3). Always use an Adult Breathing Circuit, if a circuit is required during the tests. NOTICE The HAMILTON-C2 needs a warm-up period. Make sure it was running for at least 20 minutes in the Ventilation Mode. This section describes each of the units containing the HAMILTON-C2 Service Software. Before starting, be sure that you are familiar with Typographic Conventions Section 1.2 and Expressions Section 1.3. The HAMILTON-C2 Summary of Test Report is the standard form to be used and must be completed each time the Service Software is performed. If you do not have a suitable form, you can photocopy and use the form named HAMILTON-C2 Summary of Test Report (see the Summary of Test Report Section 17.1) at the back of this manual. 92 English | 624165/07 Functions of the Service Software 8 8.2 Functions of the Service Software Units in the Service Software perform the following functions: • Enables display information (concerning revisions and versions of the HAMILTON-C2 Hardware and Software) • Enables checks on the HAMILTON-C2 Hardware and Software • Enables calibration of the HAMILTON-C2 Hardware • Enables viewing and exporting of the Event Log and Service Log • Enables software upgrades Hamilton Medical | HAMILTON-C2 Service Manual 93 8 Service Software 8.3 Structure of the Service Software Figure 8-1. Service Software Structure 94 English | 624165/07 8 Starting the Service Software 8.4 Starting the Service Software NOTICE To start the Service Software, you must place the HAMILTON-C2 into the Service Software Mode. 1. Connect the HAMILTON-C2 to mains power. 2. Press the ON (A) Button located on the HAMILTON-C2 Interaction Panel and then press and hold the 100% O2 (B) and Manual Breath (B) Buttons at the same time. The 2 buttons can be released as soon as the Start Service Software message appears below the progress bar. If a Mainboard Revision 15 or higher is installed, the 100% O2 and Manual Breath Buttons can be pressed shortly. Press and hold the 100% O2 (B) and Manual Breath (B) Buttons at the same time. Figure 8-2. Starting the Service Software 3. After the Service Software starts, the Main Service Software Screen is displayed. Figure 8-3. The HAMILTON-C2 Main Service Software Screen 4. To exit the Service Software, switch OFF the HAMILTON-C2. Hamilton Medical | HAMILTON-C2 Service Manual 95 8 Service Software 8.5 Service Software Screen Layout Figure 8-4. Service Software Screen Layout Table 8-1. Service Software Screen Layout 1. Header Window 3c. Optional Menu Tab Layer 3 2. Test Window including Dialogs 4. Alarm Window 3a. Menu Tab Layer 1 5. Energy State Window 3b. Optional Menu Tab Layer 2 96 English | 624165/07 Making Screenshots 8 8.6 Making Screenshots 1. Insert the USB Memory Stick (B, PN 396207) into the HAMILTON-C2 USB Slot and press the Print Screen Button (A) for 1 second until the corresponding LED lights up. Figure 8-5. Making Screenshots The print screen function saves a JPG file of the current ventilator screen to a USB Memory Stick (B). The USB Memory Stick can be removed when the Print Screen Key LED is no longer lit. The filename takes this format: • screenshot_yyyymmdd_HHMMss.jpg where: • yyyy is the year • mm is the month • dd is the day • HH is the hour (in 24-hour format) • MM is the minute • ss is the second NOTICE Specification for the USB memory stick: File system: FAT or FAT32 Unpartitioned memory No operating system or security software installed Hamilton Medical | HAMILTON-C2 Service Manual 97 8 Service Software 8.7 Ventilator Info 1. From the Main Service Software Screen, touch the Ventil. Info Button. Figure 8-6. Main Service Software Screen 2. On the Ventilator Info Screen are the: • Technical State Button • Service Timer Button • Real Time Clock (RTC) Button • Battery Data Button • Back Button (go back to the Main Menu) Figure 8-7. The Ventilator Info Screen 98 English | 624165/07 Technical State 8 8.8 Technical State Touch the Technical State Button to enter the technical state section. Figure 8-8. Ventilator Info Screen Hamilton Medical | HAMILTON-C2 Service Manual 99 8 Service Software 8.8.1 Hardware Version Tab 1. Touch the HW Version Tab (if not already selected). The HW Version Tab displays the Device Name, Part Number, Revision, Serial Number and Timing Information. 2. The information of the following components will be updated automatically from the HAMILTON-C2 during start-up: • Battery 1 • Battery 2 • Flow Sensor AIR (Qvent) • Flow Sensor O2 (QO2) • Oxygen Sensor • Communication Board Figure 8-9. Hardware Version Tab 3. In case the technical state is corrupt, or after upgrading from SW version < 2.1.2, the Clean-up Button will appear. 4. In case the technical state cannot be read out from the EEprom of a component, the unit will fail in self-test and alarm with “Technical state failed”. Other Technical faults will appear as after effect. 5. Record the Hardware Version Information on the HAMILTON-C2 Summary of Test Report Section 17.1. 100 English | 624165/07 8 Software Version Tab 8.8.2 Software Version Tab 1. Touch the SW Version Tab. The Software Version Tab displays the device names and revisions of the operating software. Figure 8-10. Software Version Tab 2. Record the Software Version Information on the HAMILTON-C2 Summary of Test Report Section 17.1. Hamilton Medical | HAMILTON-C2 Service Manual 101 8 Service Software 8.8.3 Modify Tab NOTICE Always update the technical state when an assembly as listed below has been replaced. Restart the device prior to performing the service software tests and calibrations. The Modify Tab allows updating information when a part has been replaced. 1. Touch the Modify Tab. Figure 8-11. Modify Tab Step 1 2. Select the hardware component you want to modify. 3. To change the serial number or revision number, touch the corresponding button. Figure 8-12. Modify Tab Step 2 102 English | 624165/07 8 Modify Tab 4. Enter the number. Figure 8-13. Modify Tab Step 3 5. Proceed with Confirm. When finished, touch the Save Button. 6. To change the Part Number, touch the Part No. Button and select the correct number. 7. Proceed with Confirm. When finished, touch the Save Button. Figure 8-14. Modify Tab Step 4 NOTICE Make sure to select the correct Part Number. Cross check with the label of the Spare Part / Installation Guide of the MSP Spare Part. Hamilton Medical | HAMILTON-C2 Service Manual 103 8 Service Software 8.8.4 Transfer Tab The Transfer Tab contains 2 buttons: Export and Import To allow you to export/import the technical state of the device from a USB Memory Stick. Figure 8-15. Transfer Tab By exporting, the service software creates a directory, if it does not exist, on the USB Memory Stick. The naming convention of this directory is: INSTRUMENT-sn#device_serial_number: /myUSBDevice/C2-sn-1010/ the file which is exported is a .csv (comma separated file) file containing the device technical state. It has the same naming convention as the directory created on the USB Memory Stick: e.g.: C2-sn-1010.csv For importing values to the device, be sure that there is a directory following the previous given naming convention and a .csv file having the proper name according to the device serial number. The technical state can be exported and imported to/from a USB Memory Stick. Ensure the USB Memory Stick is connected to the ventilator, then touch the corresponding button. NOTICE Insert USB Memory Stick is visible, as long as there is no memory stick inserted. NOTICE The format of the file is .csv (comma separated values) and can be edited with a text editor (e.g. Notepad, Wordpad). Do not use Excel to modify the .csv file. 104 English | 624165/07 8 Service Timer 8.9 Service Timer 8.9.1 Service Timer Tab 1. Touch the Service timer Tab. 2. On the Service timer Tab, the total Operating hours are displayed and the Service timer hours are displayed since the last time the Service timer was previously reset. Figure 8-16. Service Timer Tab 3. Record the Operating hours and the Service timer hours on the HAMILTON-C2 Summary of Test Report Section 17.1. 4. After successful service, reset the service counter by touching the Reset Button. To enable the reset function, select the Alarm limit Timer first. 5. The Alarm limit Timer Button allows the set number of hours between service intervals to be changed. Figure 8-17. Setting the Alarm Limit Hamilton Medical | HAMILTON-C2 Service Manual 105 8 Service Software 6. Touch the Alarm limit Timer Button or rotate the P&T Control Knob until the Alarm Limit Rotary Screen Button is highlighted, then press the P&T Control Knob. 7. The number of hours can be changed by rotating the P&T Control Knob. 8. Touch the Alarm limit Timer Button again or press the P&T Control Knob to save the new value. Figure 8-18. Alarm Limit Reset Button 9. Touch Reset. 10. After resetting the service timer, the current operating hours will be added to the set alarm limit hours. 106 English | 624165/07 8 Blower Timer Tab 8.9.2 Blower Timer Tab NOTICE The expected operation time (100%) is calculated based on the operating hours, temperature and rotation speed of the blower. When 100% is reached, the alarm Blower service required appears and the Blower Module needs to be replaced. Replace the Blower Module when the blower timer ≥ 100%. Also replace the Blower Module when the Prediction Blower Timer in the Next 6 Months reaches 100%. See prediction calculation formula in the table below. After replacement of the Blower Module, enter the Serial Number and Revision Number of the replacement blower in the Technical State (see the Modify Tab 8.8.3). This will reset the Blower timer. Table 8-2. Prediction of the blower timer in 6 months: Where: ebt = expected blower timer in 6 months [%] bt = current blower timer [%] y = number of years since the installation of the blower [Year(s)] 1. Touch the Blower timer Tab. 2. On the Blower timer Tab, the total blower time hours are displayed in percentage. Figure 8-19. Blower Timer Tab 3. Record the Blower timer % on the HAMILTON-C2 Summary of Test Report Section 17.1. Hamilton Medical | HAMILTON-C2 Service Manual 107 8 Service Software 8.10 Real Time Clock (RTC) 1. Check the current Date and Time: Figure 8-20. Current Date and Time Reset Step 1 2. Touch the Screen Button(s) that need to be changed, or rotate the P&T Control Knob until the desired button is highlighted, then press the P&T Control Knob. 3. Change the number displayed by rotating the P&T Control Knob. 4. Touch the Screen Button again or press the P&T Control Knob to keep the new value. 5. When all selections are complete, touch the Set Button. Figure 8-21. Current Date and Time Reset Step 2 6. To change to the Main Menu, touch the Back Button. 108 English | 624165/07 8 Battery Data 8.11 Battery Data The Battery Data Button displays all of the data provided for the batteries (see the Evaluation of the State of Health Section 5.3). Figure 8-22. Battery Data Hamilton Medical | HAMILTON-C2 Service Manual 109 8 Service Software 8.12 Tests Overview Table 8-3. Adjustments / Calibration Tests Screenshots Page No Touch Screen Page No 2321 Touch Screen Section 8.14.1.1 Inspiratory Valve Page No 2342 Inspiratory Valve Section 8.14.1.2 Pressure Page No 2341 Pressure Section 8.14.1.3 Expiratory Valve Page No 2343 Expiratory Valve Section 8.14.1.4 Oxygen Sensor Page No 2346 Oxygen Sensor Section 8.14.1.5 Flow Sensor Page No 2347 Flow Sensor Section 8.14.1.6 Table 8-4. Components Tests Tests Screenshots Page No Alarm System Page No 2102 Alarm System Section 8.15.1.1 Alarm Monitor 1 Page No 2113 Alarm Monitor 1 Section 8.15.1.2 Alarm Monitor 2 Page No 2114 Alarm Monitor 2 Section 8.15.1.3 User Interface Tab Page No 2115 User Interface Tab Section 8.15.2 Binary Valve Page No 2106 Binary Valve Section 8.15.3.1 Autozero Page No 2109 Autozero Section 8.15.3.2 Blower Flow Page No 2104 Blower Flow Section 8.15.3.3 Blower Pressure Page No 2105 Blower Pressure Section 8.15.3.4 Expiratory Valve Page No 2111 Expiratory Valve Section 8.15.3.5 Inspiratory Valve Page No 2107 Inspiratory Valve Section 8.15.3.6 O2 Input Page No 2112 O2 Input Section 8.15.4.1 Nebulizer Valve Page No 2116 Nebulizer Valve Section 8.15.4.2 Ambient Valve Page No 2108 Ambient Valve Section 8.15.4.3 Proximal Flow Page No 2110 Proximal Flow Section 8.15.4.4 Air Entry Page No 2117 Air Entry Section 8.15.4.5 Table 8-5. System Tests Tests Screenshots Page No Pressure Page No 2201 Pressure Section 8.16.1 Flow Page No 2202 Flow Section 8.16.2 O2 Mixer Page No 2203 O2 Mixer Section 8.16.3 Tightness Page No 2204 Tightness Section 8.16.4 Alarming Page No 2205 Alarming Section 8.16.5 110 English | 624165/07 Tests / Calibration Screen 8 8.13 Tests / Calibration Screen 1. From the Main Service Software Screen, touch the Tests / Calibration Button. Figure 8-23. Main Service Software Screen 2. On the Tests / Calibration Screen are the: • Adjustments / Calibration Button • Component Test Button • System Test Button • Sensor Data Button • Back Button Figure 8-24. Tests / Calibration Screen Hamilton Medical | HAMILTON-C2 Service Manual 111 8 Service Software 8.14 Adjustments / Calibration 1. From the Main Service Software Screen, touch the Tests / Calibration Button. Figure 8-25. Main Service Software Screen 2. Touch the Adjustment / Calibration Button. Figure 8-26. Adjustment / Calibration Screen 112 English | 624165/07 Calibration Tab 8 8.14.1 Calibration Tab Touch the Calibration Tab. Figure 8-27. Calibration Tab Hamilton Medical | HAMILTON-C2 Service Manual 113 8 Service Software Table 8-6. Adjustments / Calibration Adjustments / Calibration Adj/Calib > Calibration > Touch Screen (Page No 2321) Description Component Adjusted / Calibrated Allows calibrating the coordinates of the touch controller (4–point calibration). • Touch Screen Allows aligning the gain and offset for voltage-offset of the inspiratory valve. • Inspiratory Valve Touch Screen Section 8.14.1.1 Adj/Calib > Calibration > Inspiratory Valve (Page No 2321) Inspiratory Valve Section 8.14.1.2 Sets Offset 1 (low pressure) at 20 mbar blower pressure and Offset 2 (high pressure) at 60 mbar blower pressure when required Qvent Flow is reached. Adj/Calib > Calibration > Pressure (Page Allows adjusting the pressure sensor gain using an external pressure measurement as reference. No 2341) Pressure Section 8.14.1.3 Adj/Calib > Calibration > Exp. Valve (Page No 2343) Expiratory Valve Section 8.14.1.4 Pressure Sensor: • Pvent_control • Pvent_monitor and Paw Before calibration starts, an autozero of Pvent_control, Pvent_monitor, Paw, Pflow-sensor and IexpValve is performed. If the autozero procedure fails, a failure will be displayed and calibration cannot be started. • Expiratory Valve During calibration, gain and offset for the valve characteristics of the expiratory valve will be aligned. Adj/Calib > Calibration > O2 Cell (Page The Oxygen Sensor calibration is separated into two parts: No 2346) • Offset calibration. Calibrates the offset voltage of the amplifier circuit. Oxygen Sensor Calibration Section 8.14.1.5 Adj/Calib > Calibration > Flow Sensor (Page No 2347) Patient Flow Sensor Calibration Section 8.14.1.6 114 • Oxygen Sensor • Gain calibration. During this 2-min calibration of the Oxygen Sensor, the ventilator delivers an increased oxygen concentration (if oxygen is connected in the high pressure mode) or 21% oxygen (if oxygen is connected in the low pressure mode or disconnected). It tests the Oxygen Sensor and resets the calibration points specific to the Oxygen Sensor in use. This calibration checks and resets the calibration point specific to the flow sensor in use. • Patient Flow Sensor The Patient Flow Sensor is calibrated in both directions. English | 624165/07 Touch Screen 8 8.14.1.1 Touch Screen Allows calibrating the coordinates of the touch controller (4–point calibration). 1. Touch the Touch screen Button Figure 8-28. Touch Screen Calibration Step 1 2. Touch the Start Button to begin the Touch Screen Calibration. Figure 8-29. Touch Screen Calibration Step 2 Hamilton Medical | HAMILTON-C2 Service Manual 115 8 Service Software 3. The test begins automatically indicated by Touch Screen Calibration Running on the screen. Figure 8-30. Touch Screen Calibration Step 3 4. You are instructed to touch the cross located on the top left of the screen with your finger. 5. Next, you are instructed to touch the cross located on the top right, then bottom right and then bottom left of the screen with your finger. 6. Next, you are instructed to touch the Test Button in the middle of the screen with your finger. Figure 8-31. Touch Screen Calibration Step 4 116 English | 624165/07 Touch Screen 8 7. The Touch Screen Calibration is complete. Touch Apply to accept the new Touch Screen Calibration or Cancel. Figure 8-32. Touch Screen Calibration Step 5 8. The touch screen calibration is complete indicated by Touch screen calibration values saved. on the screen. Figure 8-33. Touch Screen Calibration Step 6 Hamilton Medical | HAMILTON-C2 Service Manual 117 8 Service Software 8.14.1.2 Inspiratory Valve Allows aligning the gain and offset for voltage-offset of the Inspiratory Valve. Sets Offset 1 (low pressure) at 20 mbar blower pressure and Offset 2 (high pressure) at 60 mbar blower pressure when required Qvent flow is reached. NOTICE − Confirm that the Technical State is updated with the current Inspiratory Valve serial number AND revision number, refer to Modify Tab 8.8.3. − The HAMILTON-C2 needs a warm-up period. Make sure it was running for at least 20 minutes in the Ventilation Software. The Inspiratory Valve has a flow-voltage-hysteresis, in that the valve opening characteristic is different from the valve closing characteristic. The valve needs to be calibrated with either the opening process or the closing process. Based on the design and pressure control mechanism, the Inspiratory valve is calibrated with the closing characteristic. The required flow adjustment depends on the Inspiratory Valve revision: • MSP160230/00 –> 12 l/min (Qvent) • MSP160230/01 and higher –> 18 l/min (Qvent) The Technical State needs to contain the correct Inspiratory Valve revision to ensure proper control function Inspiratory valve calibration: 1. Touch the Insp. Valve Button. 2. Wait approximately 10 seconds until the flow has stabilized. Figure 8-34. Inspiratory Valve Adjustment / Calibration Step 1 3. Disconnect the inspiration tube as shown. 118 English | 624165/07 8 Inspiratory Valve 4. Touch the Start Button. Figure 8-35. Inspiratory Valve Adjustment / Calibration Step 2 Figure 8-36. Inspiratory Valve Adjustment / Calibration Step 3 5. The Inspiratory Valve Calibration is complete indicated by Inspiration valve calibration successful displayed on the screen. Hamilton Medical | HAMILTON-C2 Service Manual 119 8 Service Software 8.14.1.3 Pressure Allows adjusting the Pressure Sensor Gain using an external pressure measurement as a reference. 1. Touch the Pressure Button. Figure 8-37. Pressure Adjustment / Calibration Step 1 2. Attach a Pressure Connector to the Patient Connection. 3. Attach the Tube System to the Pressure Connector and the Expiratory Connection. 4. Attach an External Pressure Gauge to the Pressure Connector. 5. Close the Flow Sensor outlet. 6. Touch the Start Button. Figure 8-38. Pressure Adjustment / Calibration Step 2 7. As soon as you touch the Start Button, the Ventilator does an autozero procedure to prevent wrong calibrations due to remaining pressure inside the Patient Breathing Circuit. 120 English | 624165/07 Pressure 8 8. The Values on the screen should equal the value on the Pressure Gauge. 9. If adjustment is necessary, adjust the Gain by using the P&T Control Knob. Make the adjustment with the P&T Control Knob, then, press the P&T Control Knob for the change in Gain to be applied. Figure 8-39. Pressure Adjustment / Calibration Step 3 10. After completion, touch the Save Button to save the changes. Figure 8-40. Pressure Adjustment / Calibration Step 4 NOTICE If the test cannot be performed, check the whole setup for leaks. Furthermore, there should be almost no flow from the expiration outlet. If you perform this test more than once, please close, then re-open the Pressure Calibration Window. Hamilton Medical | HAMILTON-C2 Service Manual 121 8 Service Software 8.14.1.4 Expiratory Valve Before calibration starts, an autozero of Pvent_control, Pvent_monitor, Paw, Pflow-sensor and IexpValve is performed. If the autozero procedure fails, a failure will be displayed and calibration cannot be started. During calibration, gain and offset for the valve characteristics of the Expiratory Valve will be aligned. 1. Touch the Exp. Valve Button. Figure 8-41. Expiratory Valve Adjustment / Calibration Step 1 2. Connect a 30cm tube with a filter (PN 279204) as shown on the screen. 3. Touch the Start Button. Figure 8-42. Expiratory Valve Adjustment / Calibration Step 2 4. The Ventilator performs an autozero procedure to prevent wrong calibrations due to remaining pressure inside the limb. 122 English | 624165/07 Expiratory Valve 8 5. The calibration runs automatically indicated by the Pvent_monitor and Iexp. values changing during the calibration process. Figure 8-43. Expiratory Valve Adjustment / Calibration Step 3 6. The successful finish of the Expiratory Valve will be indicated with following message: Calibration successfully finished. NOTICE If the test takes more than 3 minutes or if calibration is not OK, replace the membrane and perform the test again. Hamilton Medical | HAMILTON-C2 Service Manual 123 8 Service Software 8.14.1.5 Oxygen Sensor The Oxygen Sensor calibration is separated into two parts: • Offset calibration. Calibrates the offset voltage of the amplifier circuit. • Gain calibration. During this 2-min calibration of the Oxygen Sensor, the ventilator delivers an increased oxygen concentration (if oxygen is connected in the high pressure mode) or 21% oxygen (if oxygen is connected in the low pressure mode or disconnected). It tests the Oxygen Sensor and resets the calibration points specific to the Oxygen Sensor in use. 8.14.1.5.1 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 1-5 NOTICE With Mainboard (MSP160200) Revisions 1-5, perform the test as shown below. With Mainboard (MSP160200) Revisions 6 and higher and Mainboard (MSP160382), perform the Oxygen Sensor Calibration as shown in the Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher Section 8.14.1.5.2. 1. Connect the HAMILTON-C2 to high pressure O2. 2. Touch the O2 cell Button. Figure 8-44. Oxygen Sensor Adjustment / Calibration Step 1 3. Disconnect the Oxygen Sensor Cable from the Oxygen Sensor and connect the Oxygen Sensor Calibration Tool (PN 160367). 124 English | 624165/07 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 1-5 8 4. Touch the Start Button. Figure 8-45. Osygen Sensor Adjustment / Calibration Step 2 5. The Offset Calibration runs automatically until Offset calibration completed appears on the screen. 6. Connect the Oxygen Sensor Cable to the Oxygen Sensor. 7. Touch the Start Button to begin the Gain Calibration. Figure 8-46. Oxygen Sensor Adjustment / Calibration Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 125 8 Service Software 8. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... displayed on the screen. Figure 8-47. Oxygen Sensor Adjustment / Calibration Step 4 9. The calibration is complete when OK is displayed on the screen. Figure 8-48. Oxygen Sensor Adjustment / Calibration Step 5 126 English | 624165/07 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher 8 8.14.1.5.2 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher NOTICE For Mainboard (MSP160200) Revisions 6 and higher and Mainboard (MSP160382) Revisions 1-4. 1. Connect the HAMILTON-C2 to high pressure O2. 2. Touch the O2 Cell button. Figure 8-49. Oxygen Sensor Adjustment / Calibration Step 1 3. Disconnect the Oxygen Sensor Cable from the Oxygen Sensor. Make sure the Oxygen Sensor connector is freely suspended. 4. Touch the Start button. Figure 8-50. Oxygen Sensor Adjustment / Calibration Step 2 5. The Offset Calibration runs automatically until Offset calibration completed appears on the screen. Hamilton Medical | HAMILTON-C2 Service Manual 127 8 Service Software 6. Connect the Oxygen Sensor Cable to the Oxygen Sensor. 7. Touch the Start Button to begin the Gain Calibration. Figure 8-51. Oxygen Sensor Adjustment / Calibration Step 3 8. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... displayed on the screen. Figure 8-52. Oxygen Sensor Adjustment / Calibration Step 4 128 English | 624165/07 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher 8 9. The calibration is complete when OK is displayed on the screen. Figure 8-53. Oxygen Sensor Adjustment / Calibration Step 5 Hamilton Medical | HAMILTON-C2 Service Manual 129 8 Service Software 8.14.1.5.3 Oxygen Sensor Calibration with Mainboard MSP160382 Revisions 5 and Higher NOTICE For Mainboard (MSP160382) Revisions 5 and higher: The calibration is complete when OK is displayed on the screen. 1. Connect the HAMILTON-C2 to high pressure O2. 2. Touch the O2 Cell Button. 3. Touch the Start Button. Figure 8-54. Oxygen Sensor Adjustment / Calibration 4. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... indicated on the screen. 5. The calibration is complete when OK is displayed on the screen. 130 English | 624165/07 8 Flow Sensor 8.14.1.6 Flow Sensor This calibration checks and resets the calibration point specific to the Patient Flow Sensor in use. The Patient Flow Sensor is calibrated in both directions. 1. Touch the Flow sensor Button. Figure 8-55. Patient Flow Sensor Calibration 2. Connect the tubing system and turn the Patient Flow Sensor. The Patient Flow Sensor has to be inverted for the Patient Flow Sensor Calibration and an adapter is required. Figure 8-56. Setup for the Patient Flow Sensor Calibration Hamilton Medical | HAMILTON-C2 Service Manual 131 8 Service Software 3. Touch the Start Button. Figure 8-57. Patient Flow Sensor Calibration Step 1 4. The Patient Flow Sensor Calibration is in process. Figure 8-58. Patient Flow Sensor Calibration Step 2 132 English | 624165/07 8 Flow Sensor 5. Turn the Patient Flow Sensor. 6. Touch the Next Button to continue. Figure 8-59. Patient Flow Sensor Calibration Step 3 7. The successful finish of the Patient Flow Sensor Calibration will be indicated with following message: Flow sensor calibration successful. Figure 8-60. Patient Flow Sensor Calibration Step 4 Hamilton Medical | HAMILTON-C2 Service Manual 133 8 Service Software 8.14.2 Show Tab 1. Touch the Adjustment / Calibration Button. Figure 8-61. Adjustment / Calibration Screen 2. Touch the Show Tab. Figure 8-62. Show Tab 134 English | 624165/07 Pressure 8 8.14.2.1 Pressure 1. Touch the Pressure Button. Figure 8-63. Pressure Screen 2. The screen displays the Pvent_control, Pvent_monitor and Paw Pressure Sensors Offset and Gain values. Hamilton Medical | HAMILTON-C2 Service Manual 135 8 Service Software 8.14.2.2 Flow Sensor 1. Touch the Flow sensor Button. Figure 8-64. Flow Sensor Screen 2. Displays the Inspiratory Pressure and Expiratory Pressure at different flows. NOTICE Not required on the HAMILTON-C2 Summary of Test Report. 136 English | 624165/07 Valves 8 8.14.2.3 Valves 1. Touch the Valves Button. Figure 8-65. Valves Screen 2. Displays the Expiratory Valve Offset and Gain values. Hamilton Medical | HAMILTON-C2 Service Manual 137 8 Service Software 8.14.2.4 Oxygen Sensor 1. Touch the O2 cell Button. Figure 8-66. Oxygen Sensor Screen 2. Displays the Oxygen Sensor Offset and Gain values. 138 English | 624165/07 Rinsp 8 8.14.2.5 Rinsp 1. Touch the Rinsp Button Figure 8-67. Rinsp Screen 2. Displays the Inspiratory Resistance values. Hamilton Medical | HAMILTON-C2 Service Manual 139 8 Service Software 8.14.3 Default Settings Tab CAUTION If calibration is not possible due to a corrupted calibration file, touch the Default set. Tab. Therefore, all steps of the Service Software have to be run again. 1. Touch the Default settings tab. Figure 8-68. Default Settings Tab 2. The Default Set. Tab provides a method to reset or set the values to a pre-determined Factory Default Setting. You can choose between setting the default values of only the Flow Sensor or of All settings. 3. To reset the values, touch the Set Button. Figure 8-69. Factory Settings Reset Step 1 140 English | 624165/07 Default Settings Tab 8 4. When the Set Button is selected, choose between Flow sensor and All settings. Figure 8-70. Factory Settings Reset Step 2 5. If you choose All settings, a Warning is displayed. Figure 8-71. Factory Settings Reset Step 3 6. You must touch the Confirm Button to reset to the Default Settings. Hamilton Medical | HAMILTON-C2 Service Manual 141 8 Service Software 8.15 Component Tests 1. From the Main Service Software Screen, touch the Tests / Calibration Button. Figure 8-72. Main Service Software Screen 2. Touch the Comp test Button. Figure 8-73. Components Test Screen 142 English | 624165/07 Component Tests 8 Table 8-7. Electronics Tab Component Tests Description Adj/Calib > Calibration > Alarm System • Touch Screen Alarm Section 8.15.1.1 This test allows activating the components related to the alarm system, it also tests the loudness levels. The test result needs to be confirmed manually. Adj/Calib > Calibration > Alarm Monitor (Steps 1 — 10) • Alarm Silence Button and LED 1 (Page No 2113) These tests diagnose the alarm monitoring system components. • Ambient Mode (Step 11 — 13) • Fan Testing the Ambient Mode. This test consists of setting the blower to achieve a certain pressure and a certain flow for 5 seconds, afterwards the instrument will be switched to ambient mode causing the blower to stop. • Inspiratory Valve (Page No 2102) Alarm Monitor 1 Section 8.15.1.2 Component Tested • Alarm lamp lights red or yellow, depending • Buzzer • Ambient Valve The Ambient Mode has activated the alarm silence LED and the buzzer. The test result needs to be confirmed manually. (Step 14 — 17) This test checks the alarm in case of fan failure. It requires stopping the fan. Adj/Calib > Calibration > Alarm Monitor 2 (Page No 2114) Alarm Monitor 2 Section 8.15.1.3 This test activates the software watchdog. The test result has to be confirmed manually. Afterwards the ventilator needs to be restarted. • Watchdog (software tasks) Table 8-8. User Interface Tab Component Tests Description Adj/Calib > Calibration > User Interface With this test, the User Interface related components can be activated in order to check the interaction of the Hard Keys with the GUI. Hard Key combinations can also be tested. Dimming the screen and alarm lamp is tested using the Day/Night Button. The P&T Control Knob has 16 steps. The test is to see that all 16 steps register with the ventilator. Tab (Page No 2115) User Interface Tab Section 8.15.2 Component Tested • P&T Control Knob • Hard Keys + LED’s • Backlight (day/night brightness) Table 8-9. Pneumatics 1 Tab Component Tests Adj/Calib > Calibration > Binary Valve (Page No 2106) Binary Valve Section 8.15.3.1 Adj/Calib > Calibration > Autozero (Page No 2109) Description Component Tested This test checks the Autozero Valves during operation and autozeroing sequence. A constant pressure is applied. Both status, “running” and“autozero”,are tested for the following valves: Pvent_monitor, PFlowsensor. • Autozero Valves The test repeats the Autozero sequence 5 times under a defined pressure. • Autozero of: Autozero Section 8.15.3.2 • Pvent_monitor Pflowsensor – Paw Pressure Sensor – Qaw Proximal Flow Sensor – Pvent_monitor – Pvent_control Adj/Calib > Calibration > Blower Flow (Page No 2104) Blower Flow Section 8.15.3.3 Adj/Calib > Calibration > Blower Pressure (Page No 2105) Blower Pressure Section 8.15.3.4 Several flows (ml/s) will be set as target to the blower. The achieved blower speed in rpm measured by the Hall Sensor should be within the defined tolerance. • Blower Several pressures are set as target to the blower. The achieved pressure measured by the Pvent_monitor Pressure Sensor should be within the given tolerance. • Blower Hamilton Medical | HAMILTON-C2 Service Manual 143 8 Service Software Component Tests Adj/Calib > Calibration > Expiratory Valve (Page No 2111) Expiratory Valve Section 8.15.3.5 Adj/Calib > Calibration > Inspiratory Valve (Page No 2107) Inspiratory Valve Section 8.15.3.6 Description Component Tested This test checks the Expiratory Valve. Several pressures are set as target for a constant flow Qinsp to the blower. The pressures are measured by Paw and have to be within the defined tolerance • Expiratory Valve This test checks the Inspiratory Valve tightness and flow control with Qinsp 0, 3, 18 and 150 l/min as well as pressure measurement Pvent_monitor with Pinsp of 2, 10, 25 and 40 mbar. • Inspiratory Valve Table 8-10. Pneumatics 2 Tab Component Tests Adj/Calib > Calibration > O2 Input (Page No 2112) O2 Input Section 8.15.4.1 Adj/Calib > Calibration > Nebulizer Valve (Page No 2116) Nebulizer Valve Section 8.15.4.2 Adj/Calib > Calibration > Ambient Valve (Page No 2108) Ambient Valve Section 8.15.4.3 Adj/Calib > Calibration > Proximal Flow Description For this test, the ventilator needs to be connected to high pressure oxygen. Several flows will be set as a target for the O2 Mixer Valve. The oxygen flow is measured by the internal QO2 Flow Sensor and has to be within a certain tolerance. In a second test the leakage is tested. • O2 Mixer Assembly For this test, the ventilator needs to be connected to high pressure oxygen. The Patient Flow Sensor is used to measure the generated flow when the Nebulizer Valve is in a closed or opened state. The test results successfully if the measured Qaw matches to the defined tolerances. • Nebulizer Valve This test checks the ambient valve with a blower pressure of 20 and 60 mbar in off, closed and active opened state. The result is measured with Pvent_monitor Pressure Sensor. • Ambient Valve This test consists of three parts: • Rinse Flow (Page No 2110) Proximal Flow Section 8.15.4.4 Component Tested - Rinse Flow Test: This test requires checking the rinse flow manually by immersing the 2 flow tubes in a glass of water as depicted. The number of bubbles has to be approximately equal on each outlet. • Paw Pressure Sensor • Qaw Proximal Flow Sensor - Proximal Pressure: This test allows checking the proximal pressure of the Patient Flow Sensor. Two different pressures are applied consecutively by the blower. The achieved pressure is measured by the Paw Pressure Sensor and has to be within the tolerance. - Proximal Flow: A certain flow is set as target to the blower, controlled by the internal Qvent Flow Sensor. The flow through the proximal side of the Patient Flow Sensor is measured by the internal differential Qaw Proximal Flow Sensor and has to be within a certain tolerance. Adj/Calib > Calibration > Air Entry (Page No 2117) This test allows checking the Pfilter Pressure Sensor with and without an obstructed filter. • HEPA Filter • Pfilter Pressure Sensor Air Entry Section 8.15.4.5 144 English | 624165/07 8 Electronics Tab 8.15.1 Electronics Tab NOTICE This loudness setting also sets the loudness for the System Test, Alarming. To check the alarm sound on High-, Medium-, Low-Priority Alarms with the loudness range 1–10, set the loudness here, then perform the System Test, Alarming (see Alarming Section 8.16.5). 1. From the Main Service Software Screen, touch the Tests / Calibration Button. Figure 8-74. Main Service Software Screen 2. Touch the Comp test Button. 3. Touch the Electronics Tab. Figure 8-75. Components Test Screen Hamilton Medical | HAMILTON-C2 Service Manual 145 8 Service Software 8.15.1.1 Alarm System This test allows activating the components related to the alarm system, it also tests the loudness levels. The test result needs to be confirmed manually. 1. Touch the Alarm System Button. Figure 8-76. Alarm System Tests Step 1 2. Touch the Speaker Start Button. Figure 8-77. Alarm System Tests Step 2 3. Observe that the speaker makes an audible sound. 146 English | 624165/07 8 Alarm System 4. The speaker volume can be adjusted with the screen adjustment knob. Check that the volume changes to minimum and maximum. Figure 8-78. Alarm System Tests Step 3 5. Indicate on the screen if the speaker is operating properly by touching OK or NOT OK. Figure 8-79. Alarm System Tests Step 4 Hamilton Medical | HAMILTON-C2 Service Manual 147 8 Service Software 6. When the OK Button is touched, the speaker will switch OFF, the Speaker Start Button will switch to Start and OK will be indicated on the screen. Figure 8-80. Alarm System Tests Step 5 7. Touch the Yellow Lamp Start Button. Figure 8-81. Alarm System Tests Step 6 8. Observe that the yellow lamp is switched ON. 148 English | 624165/07 Alarm System 8 9. Indicate on the screen if the Yellow lamp is operating properly by touching OK or NOT OK. Figure 8-82. Alarm System Tests Step 7 10. When the OK Button is touched, the Yellow Lamp will switch OFF, the Yellow Lamp Start Button will switch to Start and OK will be indicated on the screen. Figure 8-83. Alarm System Tests Step 8 Hamilton Medical | HAMILTON-C2 Service Manual 149 8 Service Software 11. Touch the Red Lamp Start Button. Figure 8-84. Alarm System Tests Step 9 12. Observe that the red lamp is switched ON, 13. Indicate on the screen if the red lamp is operating properly by touching OK or Not OK. Figure 8-85. Alarm System Tests Step 10 150 English | 624165/07 Alarm System 8 14. When the OK Button is touched, the Red Lamp will switch OFF, the Red Lamp Start Button will switch to Start and OK will be indicated on the screen. Figure 8-86. Alarm System Tests Step 11 15. After completion of the Alarm System Tests, the results are indicated on the screen. Hamilton Medical | HAMILTON-C2 Service Manual 151 8 Service Software 8.15.1.2 Alarm Monitor 1 • Procedure Steps 1 — 10 – These tests diagnose the alarm monitoring system components. • Procedure Steps 11 — 13 – Testing the Ambient Mode. This test consists of setting the blower to achieve a certain pressure and a certain flow for 5 seconds, afterwards the ventilator will be switched to Ambient Mode causing the blower to stop. – The Ambient Mode will activate the Alarm Silence LED and the Buzzer. The test result needs to be confirmed manually. • Procedure Steps 14 — 17 – This test checks the alarm in case of fan failure. It requires stopping the fan. 1. Touch the Alarm Monitor 1 Button. Figure 8-87. Alarm Monitor 1 Screen 152 English | 624165/07 Alarm Monitor 1 8 2. Touch the Start Button. Figure 8-88. Alarm Monitor 1 Tests Start 3. Indicate on the screen if the Alarm Light and the Alarm Silence LED are blinking by touching OK or NOT OK. NOTICE With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will blink. Figure 8-89. Alarm Monitor 1 Tests Step 1 4. Record the results on the HAMILTON-C2 Summary of Test Report Section Summary of Test Report. Hamilton Medical | HAMILTON-C2 Service Manual 153 8 Service Software 5. Next, indicate on the screen if the Alarm Light and the Alarm Silence LED are ON by touching OK or NOT OK. NOTICE With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will be ON. Figure 8-90. Alarm Monitor 1 Tests Step 2 6. Next, touch the Alarm Silence Button. 7. Indicate on the screen if the Alarm Light is ON by touching OK or NOT OK. NOTICE With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will be ON. Figure 8-91. Alarm Monitor 1 Tests Step 3 154 English | 624165/07 Alarm Monitor 1 8 8. Next, indicate on the screen if the Alarm Light is OFF by touching OK or NOT OK. Figure 8-92. Alarm Monitor 1 Tests Step 4 9. Next, touch the Alarm Silence Button. 10. Indicate on the screen if the Alarm Light is OFF by touching OK or NOT OK. Figure 8-93. Alarm Monitor 1 Tests Step 5 Hamilton Medical | HAMILTON-C2 Service Manual 155 8 Service Software 11. Touch the Alarm Silence Button and wait 5 seconds. Figure 8-94. Alarm Monitor 1 Tests Step 6 12. Indicate on the screen if the Expiratory Valve Plunger can be moved by hand by touching OK or NOT OK. Figure 8-95. Alarm Monitor 1 Tests Step 7 156 English | 624165/07 Alarm Monitor 1 8 13. Indicate on the screen if the Alarm Silence LED is ON and the Buzzer sounds by touching OK or NOT OK. Figure 8-96. Alarm Monitor 1 Tests Step 8 WARNING Be careful when stopping the cooling fan by hand. 14. Next, remove the Rear Cover and Fan Filter, then physically STOP the cooling fan by hand until the alarm is shown on the screen. Figure 8-97. Alarm Monitor 1 Tests Step 9 Hamilton Medical | HAMILTON-C2 Service Manual 157 8 Service Software 15. After the Fan Failure Alarm is displayed, release the fan. Figure 8-98. Alarm Monitor 1 Tests Step 10 16. Verify that the message Test completed is displayed on the screen. Figure 8-99. Alarm Monitor 1 Tests Step 11 17. Re-install the Fan Filter and Rear Cover. 158 English | 624165/07 8 Alarm Monitor 2 8.15.1.3 Alarm Monitor 2 This test activates the software watchdog. The test result has to be confirmed manually. Afterwards, the ventilator needs to be restarted. 1. Touch the Alarm Monitor 2 Button. Figure 8-100. Alarm Monitor 2 Tests 2. Touch the Start Button to activate the Watchdog. Figure 8-101. Alarm Monitor 2 Tests Step 1 • The Alarm LED should blink • The Buzzer should sound Hamilton Medical | HAMILTON-C2 Service Manual 159 8 Service Software 3. Touch the OK or NOT OK Button and reboot the ventilator into the Service Software. Figure 8-102. Alarm Monitor 2 Tests Step 2 4. After completion, the HAMILTON-C2 must be switched OFF, then ON again. 160 English | 624165/07 User Interface Tab 8 8.15.2 User Interface Tab With this test, the user interface related components can be activated in order to check the interaction of the hard keys with the GUI. Hard key combinations can also be tested. Dimming the screen and alarm lamp is tested using the day/night button. The P&T Control Knob Encoder has 16 steps. The test is to see that all 16 steps register with the ventilator. 1. From the Main Service Software Screen, touch the Tests / Calibration Button. 2. Touch the Comp test Tab. 3. Touch the User I/F Tab. Figure 8-103. User Interface Tab Screen 4. Test the P&T Control Knob Encoder by selecting the P&T Control Knob on the screen and rotating the knob. Observe there are 16 steps in 1 full turn of the P&T Control Knob. Figure 8-104. User Interface Tests Step 1 Hamilton Medical | HAMILTON-C2 Service Manual 161 8 Service Software 5. Press each button on the Interaction Panel and observe the reaction to the button pressed on the screen. 6. Touch the Night/Day Button to switch between the two display brightness levels. 7. Confirm with OK/NOT OK. Figure 8-105. User Interface Tests Step 2 162 English | 624165/07 Pneumatics 1 Tab 8 8.15.3 Pneumatics 1 Tab 1. From the Main Service Software Screen, touch the Tests / Calibration Button. 2. Touch the Comp test Tab. 3. Touch the Pneumatics 1 Tab. Figure 8-106. Pneumatics 1 Tab 4. Connect a complete Patient Breathing Circuit before starting the tests. Hamilton Medical | HAMILTON-C2 Service Manual 163 8 Service Software 8.15.3.1 Binary Valve This test checks the Autozero Valves during operation and autozeroing sequence. A constant pressure is applied. Both status, “running” and“autozero”,are tested for the following valves: Pvent_monitor, PFlowsensor. 1. Touch the Binary valve Button. Figure 8-107. Binary Valve Tests 2. Connect the tubing as shown with a 7 mm ET Tube to the Flow Sensor. 3. Touch the Start Button to begin the Binary Valve Test. Figure 8-108. Binary Valve Tests Step 1 164 English | 624165/07 Binary Valve 8 4. The test runs automatically indicated by Component test binary valves running... displayed on the screen. Figure 8-109. Binary Valve Tests Step 2 5. The Component test binary valves is complete when OK is indicated on the screen. This means that the internal sensor checks are ok. Figure 8-110. Binary Valve Tests Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 165 8 Service Software 8.15.3.2 Autozero The test repeats the Autozero sequence 5 times under a defined pressure. 1. Touch the Autozero Button. Figure 8-111. Autozero Valves Tests Screen 2. Touch the Start Button. Figure 8-112. Autozero Valves Tests Step 1 166 English | 624165/07 Autozero 8 3. The Paw Pressure Sensor and Qaw Proximal Flow Sensor will Autozero. Figure 8-113. Autozero Valves Tests Step 2 4. The Paw Pressure Sensor and Qaw Proximal Flow Sensor Autozero adjustment is complete when OK is indicated on the screen. Figure 8-114. Autozero Valves Tests Step 3 5. The Autozero results will be recorded in the Service Log. Hamilton Medical | HAMILTON-C2 Service Manual 167 8 Service Software 8.15.3.3 Blower Flow Several flows (ml/s) will be set as target to the blower. The achieved blower speed in rpm measured by the Hall Sensor should be within the defined tolerance. 1. Touch the Blower flow Button. Figure 8-115. Blower Flow Tests Screen 2. Disconnect the Inspiration Tube. 3. Touch the Start Button. Figure 8-116. Blower Flow Tests Step 1 168 English | 624165/07 Blower Flow 8 4. The test runs automatically indicated by Component test blower flow running... displayed on the screen. Figure 8-117. Blower Flow Tests Step 2 5. The test is complete when Component test blower flow completed. is displayed on the screen and the results are displayed with OK or NOT OK. Figure 8-118. Blower Flow Tests Step 3 NOTICE The Blower Pressure values are in mbar and the Flow values are in ml/minute. Hamilton Medical | HAMILTON-C2 Service Manual 169 8 Service Software 8.15.3.4 Blower Pressure Several pressures are set as a target for the blower. The achieved pressure measured by the Pvent_monitor Pressure Sensor should be within the given tolerance. 1. Touch the Blower pres. Button. Figure 8-119. Blower Pressure Tests 2. Disconnect the Inspiration Tube and seal the Patient Outlet with a stopper. 3. Touch the Start button. Figure 8-120. Blower Pressure Tests Step 1 170 English | 624165/07 Blower Pressure 8 4. The test runs automatically indicated by Component test blower pressure running... displayed on the screen. Figure 8-121. Blower Pressure Tests Step 2 5. The test is complete when Component test blower flow completed. is displayed on the screen and the results are displayed with OK or NOT OK. Figure 8-122. Blower Pressure Tests Step 3 NOTICE The Blower Pressure values are in mbar. Hamilton Medical | HAMILTON-C2 Service Manual 171 8 Service Software 8.15.3.5 Expiratory Valve NOTICE During this test, the Technical Event: 231001 may appear. This failure message can be ignored. This test checks the Expiratory Valve. Several pressures are set as target for a constant flow Qinsp to the blower. The pressures are measured by Paw and have to be within the defined tolerance. 1. Touch the Exp. valve Button. Figure 8-123. Expiratory Valve Tests Screen 2. Connect a complete Patient Breathing Circuit 3. Touch the Start Button to begin the Expiratory Valve Leakage and Pressure Tests. Figure 8-124. Expiratory Valve Tests Step 1 172 English | 624165/07 8 Expiratory Valve 4. Test runs automatically indicated by Expiration valve pressure test in progress... displayed on the screen. Figure 8-125. Expiratory Valve Tests Step 2 5. The test is complete when Expiration valve pressure test completed is displayed on the screen and the result is displayed with OK or Not OK. 6. If the pressure test or leakage test fails, the values which are out of tolerance are displayed at the bottom of the page. Figure 8-126. Expiratory Valve Tests Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 173 8 Service Software 8.15.3.6 Inspiratory Valve This test checks the Inspiratory Valve tightness and flow control with Qinsp 0, 3, 18 and 150 l/min as well as pressure measurement Pvent_monitor with Pinsp of 2, 10, 25 and 40 mbar. 1. Touch the Insp. valve Button. Figure 8-127. Inspiratory Valve Tests Screen 2. Connect an Adult 22 mm Tube (PN 260100) and Filter (PN 279211) to the ventilator and attach the Flow Analyzer (as shown in the picture on the screen). 3. Touch the Start Button. Figure 8-128. Inspiratory Valve Tests Step 1 4. The test runs automatically indicated by Leakage test in progress... displayed on the screen. 174 English | 624165/07 Inspiratory Valve 8 5. The Inspiratory Valve Leakage Test is complete when OK is indicated on the screen. If the pressure tests fails, the out of tolerance pressure is displayed. Figure 8-129. Inspiratory Valve Tests Step 2 6. Touch the Next Button to begin the Flow Control Tests. Figure 8-130. Inspiratory Valve Tests Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 175 8 Service Software 7. Pass through the Flow Control Test with the values of 0 l/min., 3 l/min., 18 l/min. and 150 l/min. Only verify the 18 l/min. value with the external Flow Analyzer. It must be within a tolerance of ± 8% (16.5 - 19.5 l/min). 8. The test runs automatically indicated by Flow control in progress on the screen. Figure 8-131. Inspiratory Valve Tests Step 4 9. The Inspiratory Valve Flow Control Test is complete when OK is indicated on the screen. 10. Connect the tubing as shown with a 7mm ET Tube. 11. Touch the Start Button to continue to the Inspiratory Valve Pressure Control Test. Figure 8-132. Inspiratory Valve Tests Step 5 176 English | 624165/07 Inspiratory Valve 8 12. The test runs automatically indicated by Pressure control in progress on the screen. Figure 8-133. Inspiratory Valve Tests Step 6 13. The Inspiratory Valve Pressure Control Test is complete when OK is indicated on the screen. Figure 8-134. Inspiratory Valve Tests Step 7 Hamilton Medical | HAMILTON-C2 Service Manual 177 8 Service Software 8.15.4 Pneumatics 2 Tab 1. From the Main Service Software Screen, touch the Tests / Calibration Button. 2. Touch the Comp test Tab. 3. Touch the Pneumatics 2 Tab. Figure 8-135. Pneumatics 2 Tab 178 English | 624165/07 O2 Input 8 8.15.4.1 O2 Input For this test, the ventilator needs to be connected to high pressure oxygen. Several flows will be set as a target for the O2 Mixer Valve. The oxygen flow is measured by the internal QO2 Flow Sensor and has to be within a certain tolerance. In a second test the leakage is tested. 1. Touch the O2 input Button. Figure 8-136. O2 Input Tests Screen 2. Connect the ventilator to high pressure oxygen (Minimum 2.8 bar to Maximum 6.0 bar (41 - 86 psi)). 3. Disconnect the Inspiration Tube. 4. Touch the Start Button to begin the O2 Input Flow and Leakage Tests. Figure 8-137. O2 Input Tests Step 1 Hamilton Medical | HAMILTON-C2 Service Manual 179 8 Service Software 5. The test runs automatically indicated by O2 valve flow test in progress... displayed on the screen. Figure 8-138. O2 Input Tests Step 2 6. The O2 Valve Flow Test is complete when OK is indicated on the screen. 7. The next test runs automatically indicated by O2 valve leakage test in progress... displayed on the screen. Figure 8-139. O2 Input Tests Step 3 180 English | 624165/07 O2 Input 8 8. The O2 Valve Leakage Test is complete when OK is indicated on the screen. Figure 8-140. O2 Input Tests Step 4 Hamilton Medical | HAMILTON-C2 Service Manual 181 8 Service Software 8.15.4.2 Nebulizer Valve For this test, the ventilator needs to be connected to high pressure oxygen. The Patient Flow Sensor is used to measure the generated flow when the Nebulizer Valve is in a closed or opened state. The test results are successful if the measured Qaw matches the defined tolerances. 1. Touch the Neb. valve Button. Figure 8-141. Nebulizer Valve Tests Screen 2. Connect the tube system as shown. 3. Touch the Start Button to start running the Nebulizer Valve Test with high inlet oxygen pressure Minimum 2.8 bar to Maximum 6.0 bar (41 - 86 psi). Figure 8-142. Nebulizer Valve Tests Step 1 182 English | 624165/07 Nebulizer Valve 8 4. The adjustment is complete when OK is indicated on the screen. Figure 8-143. Nebulizer Valve Tests Step 2 Hamilton Medical | HAMILTON-C2 Service Manual 183 8 Service Software 8.15.4.3 Ambient Valve NOTICE In software version 2.0.6 and earlier, during this test, the Technical Event: 231001 may appear. This failure message can be ignored. This test checks the Ambient Valve with a blower pressure of 20 and 60 mbar in off, closed and active opened state. The result is measured with the Pvent_monitor Pressure Sensor. 1. Touch the Ambi valve Button. Figure 8-144. Ambient Valve Tests Screen 2. Connect the tube system as shown and close the Flow Sensor Outlet. 3. Touch the Start Button. Figure 8-145. Ambient Valve Tests Step 1 184 English | 624165/07 Ambient Valve 8 4. The test runs automatically indicated by Test ambient valve power off in progress displayed on the screen. 5. The Ambient Valve Power-Off Test is complete when OK is indicated on the screen. 6. The next test runs automatically indicated by Test ambient valve Closed in progress displayed on the screen. Figure 8-146. Ambient Valve Tests Step 2, Possible Technical Event (only with SW 2.0.6 or lower) 7. The Ambient Valve Closed Test is complete when OK is indicated on the screen. 8. The next test runs automatically indicated by Test ambient valve Active open in progress displayed on the screen. 9. The Ambient Valve Active Open Test is complete when OK is indicated on the screen. Figure 8-147. Ambient Valve Tests Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 185 8 Service Software 8.15.4.4 Proximal Flow This test consists of three parts: • Rinse Flow Test: – This test requires checking the rinse flow manually by immersing the 2 flow tubes in a glass of water as depicted. The number of bubbles has to be approximately equal on each outlet. • Proximal Pressure: – This test allows checking the proximal pressure of the Patient Flow Sensor. Two different pressures are applied consecutively by the blower. The achieved pressure is measured by the Paw Pressure Sensor and has to be within the tolerance. • Proximal Flow: – A certain flow is set as a target to the blower, controlled by the internal Qvent Flow Sensor. The flow through the proximal side of the Patient Flow Sensor is measured by the internal differential Qaw Proximal Flow Sensor and has to be within a certain tolerance. 1. Touch the Prox. test Button. Figure 8-148. Proximal Flow Tests Screen 186 English | 624165/07 Proximal Flow 8 2. Touch the Start Button. Figure 8-149. Proximal Flow Tests Step 1 3. Attach 2 identical length tubes to the Flow Sensor connectors, immerse the tubes in a glass of water. Verify that the bubbles appear at both tube outlets and that the number of bubbles are approximately equal on each outlet. 4. Indicate on the screen if the Rinse Flow Test passed by pressing OK or NOT OK. Figure 8-150. Proximal Flow Tests Step 2 Hamilton Medical | HAMILTON-C2 Service Manual 187 8 Service Software 5. The Rinse Flow Test is complete when OK is displayed on the screen. Figure 8-151. Proximal Flow Tests Step 3 6. Connect the Adult Tube System to the ventilator with a 7 mm ET Tube. 7. Touch the Start Button to begin the Proximal Pressure Test. 8. The test runs automatically indicated by Proximal pressure in progress... displayed on the screen. 9. The Proximal Pressure Test is complete when OK is indicated on the screen. Figure 8-152. Proximal Flow Tests Step 4 10. Remove the 7 mm ET Tube from the Adult Tube System. 188 English | 624165/07 Proximal Flow 8 11. Touch the Start button to begin the Proximal Flow Test. Figure 8-153. Proximal Flow Tests Step 5 12. The test runs automatically indicated by Proximal flow in progress... displayed on the screen. Figure 8-154. Proximal Flow Tests Step 6 Hamilton Medical | HAMILTON-C2 Service Manual 189 8 Service Software 13. The Proximal Flow Test is complete when OK is indicated on the screen. Figure 8-155. Proximal Flow Tests Step 7 190 English | 624165/07 Air Entry 8 8.15.4.5 Air Entry NOTICE If the test is NOT OK, restart the HAMILTON-C2 Service Software and perform the test again. This test allows checking the Filter Pressure Sensor Pfilter with and without an obstructed filter. 1. Touch the Air entry Button. Figure 8-156. Air Entry Tests Screen 2. Disconnect the Inspiration Tube. 3. Touch the Start Button. Figure 8-157. Air Entry Tests Step 1 Hamilton Medical | HAMILTON-C2 Service Manual 191 8 Service Software 4. The default HEPA Filter Test is completed if OK is stated behind the description. Figure 8-158. Air Entry Tests Step 2 5. Obstruct the HEPA Filter with a piece of paper (A) with a small hole. Figure 8-159. Obstruction in the Air Entry 192 English | 624165/07 Air Entry 8 6. Connect the obstructed HEPA Filter and touch the Next Button. The pressure must be higher than 2 mbar as shown. 7. The test is complete when OK is indicated on the screen. Figure 8-160. Air Entry Tests Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 193 8 Service Software 8.16 System Test 1. From the Main Service Software Screen, touch the Tests / Calibration Button. Figure 8-161. Main Service Software Screen 2. Touch the System test Button. Figure 8-162. System Test Screen 194 English | 624165/07 System Test 8 Table 8-11. System Test Adjustment/ Calibration Description System Test > Pressure (Page No This test requires three different pressures to be applied and to verify the values (Pvent_control, Pvent_monitor and Paw) are within the given tolerance range (shown in a green or red background). • Pvent_control This test requires a flow to be applied and to verify the internal Qvent Flow Sensor and the Qaw Proximal Flow Sensor values are within the given tolerance range (shown in a green or red background). • Qvent Flow Sensor This test requires a flow at different oxygen concentrations to be applied to verify the internal QO2 Flow Sensor, Qvent Flow Sensor, the Patient Flow Sensor and the Oxygen Sensor values are within the given tolerance range (shown in a green or red background). • O2 Mixer Assembly This test checks for leakage in the Patient Breathing Circuit and determines the circuit's compliance compensation factor. The ventilator is pressurized to a certain pressure. The Patient Breathing Circuit is considered tight if this pressure can be maintained. If there is a leak, the pressure drops in proportion to the size of the leak. • Patient Breathing Circuit This test allows activation of the 3 different alarm priorities and combinations by pressing the corresponding button and to verify the correct alarm lamp and sound has been generated. • Alarm Priorities 2201) Pressure Section 8.16.1 System Test > Flow (Page No 2202) Flow Section 8.16.2 System Test > O2 Mixer (Page No 2203) O2 Mixer Section 8.16.3 System Test > Tightness (Page No 2204) Tightness Section 8.16.4 System Test > Alarming (Page No 2205) Alarming Section 8.16.5 Hamilton Medical | HAMILTON-C2 Service Manual Component Adjusted/ Calibrated • Pvent_monitor • Paw • Qaw Proximal Flow Sensor • QO2 Flow Sensor • Oxygen Sensor 195 8 Service Software 8.16.1 Pressure This test requires three different pressures to be applied and to verify the values (Pvent_control, Pvent_monitor and Paw) are within the given tolerance range (shown in a green or red background). 1. Touch the Pressure Button. Figure 8-163. Pressure System Tests Screen 2. Attach the tube system as shown in the picture. 3. Touch the ON Button. Figure 8-164. Pressure System Tests Step 1 196 English | 624165/07 Pressure 8 4. Set the Pinsp to 5 mbar, 25 mbar and 50 mbar. 5. Select the Pinsp Control with the P&T Control Knob. Press the P&T Control Knob and rotate the setting to the values listed above. Press the P&T Control Knob to activate the setting. Figure 8-165. Pressure System Tests Step 2 6. All values should read 5 mbar ±1.0, 25 mbar ±1.2 and 50 mbar ±2.5. 7. The test is complete when the Status OK is shown. Figure 8-166. Pressure System Tests Step 3 8. Touch the Off Button. Hamilton Medical | HAMILTON-C2 Service Manual 197 8 Service Software 8.16.2 Flow This test requires a flow to be applied and to verify the internal Qvent Flow Sensor and the Qaw Proximal Flow Sensor values are within the given tolerance range (shown in a green or red background). 1. Touch the Flow Button. Figure 8-167. Qvent Flow Sensor Tests Screen 2. Set the Insp. Flow to 18 l/min. 3. Touch the On Button. Figure 8-168. Qvent Flow Sensor Tests Step 1 198 English | 624165/07 Flow 8 4. The test is complete when the Status OK is shown. Figure 8-169. Qvent Flow Sensor Tests Step 2 5. Touch the Off Button. Hamilton Medical | HAMILTON-C2 Service Manual 199 8 Service Software 8.16.3 O2 Mixer This test requires a flow at different oxygen concentrations to be applied to verify the internal QO2 Flow Sensor, Qvent Flow Sensor, the Patient Flow Sensor and the Oxygen Sensor values are within the given tolerance range (shown in a green or red background). 1. Connect the ventilator to high pressure oxygen. 2. Touch the O2 mixer Button. Figure 8-170. O2 Mixer Tests Screen 3. Set the Insp. flow to 10 l/min. 4. Touch the On Button. Figure 8-171. O2 Mixer Tests Step 1 5. Check for the following FiO2 settings: 21%, 90% and 61%. 200 English | 624165/07 O2 Mixer 8 6. The test is complete when the Status OK is shown. Figure 8-172. O2 Mixer Tests Step 2 7. Touch the Off Button. Hamilton Medical | HAMILTON-C2 Service Manual 201 8 Service Software 8.16.4 Tightness NOTICE During this test the Technical Event: 231001 may appear. You can ignore this fault message. This test checks for leakage in the Patient Breathing Circuit and determines the circuit's compliance compensation factor. The ventilator is pressurized to a certain pressure. The Patient Breathing Circuit is considered tight if this pressure can be maintained. If there is a leak, the pressure drops in proportion to the size of the leak. 1. Touch the Tightness Button. Figure 8-173. Tightness Tests Screen 2. Touch the Start Button. Figure 8-174. Tightness Tests Step 1 202 English | 624165/07 Tightness 8 3. The test runs automatically indicated by Tightness tubing test in progress... indicated on the screen. Figure 8-175. Tightness Tests Step 2 4. The Tightness Test is complete when OK is indicated on the screen. Figure 8-176. Tightness Tests Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 203 8 Service Software 8.16.5 Alarming NOTICE During this test, Technical Events and Technical Faults will be displayed. This is a needed behavior and does not need any action. This test allows activation of the 3 different alarm priorities and combinations by pressing the corresponding button and to verify the correct alarm lamp and sound has been generated. 1. Touch the Alarming Button. Figure 8-177. Alarming Tests Screen 204 English | 624165/07 Alarming 8 2. Touch the Alarm High, Alarm Medium and Alarm Low Buttons to generate alarm(s). 3. Check if the correct Alarm Lamp and Alarm Sounds are generated. 4. Check the Alarm Priority by combining the Alarms. 5. Touch the Alarm High Button. 6. Select the Alarm High and Alarm Medium Buttons and verify if the alarm priority is working correctly. Figure 8-178. Alarming Screen Tests Step 1 7. Touch the Alarm Medium Button. Figure 8-179. Alarming Screen Tests Step 2 Hamilton Medical | HAMILTON-C2 Service Manual 205 8 Service Software 8. Touch the Alarm Low Button. Figure 8-180. Alarming Screen Tests Step 3 9. Touch OK / NOT OK to confirm / cancel the test. Figure 8-181. Alarming Screen Tests Step 4 206 English | 624165/07 Sensor Data 8 8.17 Sensor Data NOTICE Only for internal use. 1. From the Main Service Software Screen, touch the Tests / Calibration Button. Figure 8-182. Main Service Software Screen 2. Touch the Sensor data Tab Figure 8-183. Sensor Data Tab Hamilton Medical | HAMILTON-C2 Service Manual 207 8 Service Software 8.17.1 Sensor 1 Touch the Sensor 1 Tab to display the sensors values and status. Figure 8-184. Sensor 1 208 English | 624165/07 Sensor 2 8 8.17.2 Sensor 2 Touch the Sensor 2 Tab to display the sensors values and status. Figure 8-185. Sensor 2 Hamilton Medical | HAMILTON-C2 Service Manual 209 8 Service Software 8.17.3 Sensor 3 Touch the Sensor 3 Tab to display the sensors value and status. Figure 8-186. Sensor 3 210 English | 624165/07 8 Sensor Status 8.17.4 Sensor Status The displayed sensors status (hex) is a value which consists of a combination of several status (bits). The table below shows possible sensors status values. Not all status values are applicable for every sensor. Figure 8-187. Sensor Status Hamilton Medical | HAMILTON-C2 Service Manual 211 8 Service Software 8.18 Log/Config Files NOTICE How to export Event Log, Service Log, Config File and Instrument Report. 1. From the Main Service Software Screen, touch the Log/Config files Button. Figure 8-188. The Main Service Software Screen 2. On the Log/Config Files Screen are the: • Event Log Button • Service Log Button • Export Button (Instrument Report and Events Log Files) • Back Button Figure 8-189. The Log / Config Files Screen 212 English | 624165/07 8 Event Log 8.18.1 Event Log 1. Touch the Event log Button to open the Event Log. Figure 8-190. Viewing the Event Log 2. The Event Log displays the Technical Events which have occurred with the date, time, device affected and a description of the Technical Event. 3. To view another section of the Event Log listing, rotate the P&T Control Knob to select the slider on the right side of the screen. 4. Press the P&T Control Knob to activate the slider, then rotate the P&T Control Knob to move the slider up or down. Hamilton Medical | HAMILTON-C2 Service Manual 213 8 Service Software 8.18.2 Service Log 1. Touch the Service log Button to open the Service Log. Figure 8-191. Service Log Step 1 2. The Service Log displays the test and calibration results with the date, time and a description of the results. Figure 8-192. Service Log Step 2 3. To view another section of the Service Log listing, rotate the P&T Control Knob to select the slider on the right side of the screen. 4. Press the P&T Control Knob to activate the slider, then rotate the P&T Control Knob to move the slider up or down. 214 English | 624165/07 Export (Instrument Report and Events) 8 8.18.3 Export (Instrument Report and Events) 1. The Export Tab allows exporting the Service Log / Instrument Report to a USB Memory Stick. 2. Insert a USB Memory Stick into the USB Port on the side of the Interaction Panel. Figure 8-193. Export Tab Screen Hamilton Medical | HAMILTON-C2 Service Manual 215 8 Service Software 8.18.3.1 Export Events Log Files 1. Touch the Events Tab. Figure 8-194. Export Event Log Step 1 2. Touch the Start Button. Figure 8-195. Export Event Log Step 2 3. The export is performed automatically. 4. The Event File is exported in a folder named as c2-sn##### (##### device serial number). In this folder is a file named year-month-day_hour-minute-second_serviceLog_enGB.txt NOTICE If you have to send log files to Hamilton Medical for further investigation, please send all files which are created in the specific folder (error.log, eventLog.txt, serviceLog.txt, etc.). 216 English | 624165/07 Export Instrument Report 8 8.18.3.2 Export Instrument Report 1. Touch the Instrument Report Tab. Figure 8-196. Export Instrument Report Step 1 2. Touch the Start Button. Figure 8-197. Export Instrument Report Step 2 4. The export is performed automatically. 5. The Instrument report is exported in a folder named as c2-sn##### (##### device serial number). In this folder is a file named year-month-day_hour-minute-secondInstrumentRepport.txt. Hamilton Medical | HAMILTON-C2 Service Manual 217 8 Service Software 8.19 Software Updates 8.19.1 Software Update from Version < 2.2.0 WARNING After a software update, the ventilator must be restarted and the complete Service Software Calibration and Test's must be performed. Software downgrade is not allowed/possible, because this will corrupt the unit and the ESM Module will need to be exchanged. To update the HAMILTON-C2 from Software Version 1.x.x to Software Version 2.2.0 or higher, a migration step is required. For further information see the HAMILTON-C2 Software Upgrade Guide (PN 624200). 1. The following files are required: • C2_V2.2.0-Migration.tar This is required as a migration step from 1.x.x to 2.2.0 since the software structure has been changed. • C2_EM01A_V2.2.5.tar or higher This is the latest Software to be installed after the Migration Software. For the latest software check the Partner-net. 2. From the Main Service Software Screen, touch the Software update Button. Figure 8-198. Software Update Step 1 3. The Software Update Screen allows update or upgrade of the HAMILTON-C2 Software. 218 English | 624165/07 Software Update from Version < 2.2.0 8 4. Touch the SW update Button. Figure 8-199. Software Update Step 2 5. Insert the USB Memory Stick containing the migration and upgrade version as listed in the table above. If a correct file is available on the USB Memory Stick, the message appears indicating the update version, otherwise, the message No Update tarball file is available... will be displayed. In case there is more than one software update stored on the USB Memory Stick, please ensure to choose the correct version. 6. If the USB Memory Stick content is not displayed directly, refresh this page by closing and entering this page again. Figure 8-200. Software Update from Version 1.x.x to 2.2.0 7. Select the software file to be installed (C2_V2.2.0-Migration.tar). 8. Touch the Start Button to start the software update which consists of extracting, verifying and installing. 9. A progress bar is shown indicating the current status of the software update. 10. When the process is complete, Done is shown on the screen. Hamilton Medical | HAMILTON-C2 Service Manual 219 8 Service Software 11. Restart the HAMILTON-C2. The unit starts up automatically in the Service Software. 12. For further update of the HAMILTON-C2 Software, see the Software Update from Version 2.x to 2.2.5 or Higher Section 8.19.2. 220 English | 624165/07 Software Update from Version 2.x to 2.2.5 or Higher 8 8.19.2 Software Update from Version 2.x to 2.2.5 or Higher WARNING After a software update, the ventilator must be restarted and the complete Service Software Calibration and Test's must be performed. NOTICE Software downgrade is not allowed/possible, because this will corrupt the unit and the ESM Module will need to be exchanged. 1. From the Main Service Software Screen, touch the Software update Button. Figure 8-201. Software Update Step 1 2. The Software Update Screen allows update or upgrade of the HAMILTON-C2 Software. Hamilton Medical | HAMILTON-C2 Service Manual 221 8 Service Software 3. Touch the SW update Button. Figure 8-202. Software Update Step 2 4. If the USB Memory Stick is not inserted into the USB Connector or no software is available on the USB Memory Stick, the message No Update tarball file is available... is displayed. Figure 8-203. Software Update Step 3 222 English | 624165/07 Software Update from Version 2.x to 2.2.5 or Higher 8 5. If the correct file is available on the USB Memory Stick, the message appears indicating the update version. Figure 8-204. Software Update Step 4 Figure 8-205. Software Update Step 5 Hamilton Medical | HAMILTON-C2 Service Manual 223 8 Service Software 6. To begin the software update, touch the Start Button. With software version 2.2.0 or higher, the software which is to be updated can be selected from the scroll down menu (shown on the picture below). Figure 8-206. Software Update, Step 6 7. A progress bar is shown indicating the progress of the update. Figure 8-207. Software Update, Step 7 224 English | 624165/07 Software Update from Version 2.x to 2.2.5 or Higher 8 8. After the software update, the message update successfully finished is displayed on the screen. Figure 8-208. Software Update, Step 8 9. Restart the device and do not terminate the device during the first start up after performing a software update! 10. Perform the complete Calibration and Tests as described in the Service Software Section 8 Hamilton Medical | HAMILTON-C2 Service Manual 225 8 Service Software 8.20 Pre-operational Checks Refer to the HAMILTON-C2 Operator‘s Manual. Perform the pre-operational checks: • Tightness Test • Patient Flow Sensor Calibration • Oxygen Sensor Calibration • Alarm tests: – High Pressure Alarm – Low Minute Alarm – Low Oxygen Alarm / Oxygen Supply Failed Alarm – Disconnection on Patient Side – Loss of External Power – Exhalation Obstructed – Apnea 226 English | 624165/07 8 General Tests 8.21 General Tests NOTICE All devices connected to the HAMILTON-C2 must be for medical use and meet the requirements of standard IEC 60601-1. 1. Connect the HAMILTON-C2 to the AC Mains Power and to the DC Power Supply. 2. Switch on the HAMILTON-C2 and stay in Stand-By Mode. 3. Go to Alarm -> Loudness and set Loudness to 10. 4. Perform the following tests: A. AC — DC Test • Disconnect the AC Mains Power. Verify that the unit continues to run on the DC Power Supply and that the DC symbol is displayed on the screen. B. DC — Battery Test • Disconnect the DC Power Supply. Verify that the unit continues to run on internal Battery Power and that the battery symbol is displayed on the screen. Verify that the alarm speaker sounds correctly for at least one minute. C. AC — Battery Test • If no DC Power Supply is available, disconnect the AC Mains Power and verify that the HAMILTON-C2 switched to internal Battery Power. Verify that the alarm speaker sounds correctly for at least one minute. D. Power Loss Test • If no DC Power Supply is available, verify that the HAMILTON-C2 switches to internal Battery Power as soon as the AC Mains Power is removed. • Remove the internal batteries. Verify that the unit alarms. Connect the internal batteries again and verify that the unit starts up again. CAUTION The buzzer is powered by the gold caps (super capacitors) on the Mainboard. Make sure the unit is switched on (Stand-by mode or Service Mode) for at least 20 minutes before the test in order that the gold caps are fully charged. Replace the Mainboard (MSP160382) if the buzzer sounds for less than 2 minutes. Hamilton Medical | HAMILTON-C2 Service Manual 227 8 Service Software E. Fan Check • In order to check the fan for proper assembly, place a piece of paper over the fan inlet. If the paper sticks while the ventilator is on, the fan blows into the ventilator correctly. Figure 8-209. Fan Check F. Buzzer Test • To make sure that the buzzer works as specified, it has to be tested during the yearly Preventive Maintenance. – Start the device in the ventilation software. – Disconnect the AC power cable. – Remove both internal batteries. – Verify that the buzzer sounds for at least 2 minutes. • If the buzzer sounds for less than 2 minutes, replace the Mainboard (MSP160382). G. Communication Interface RS232 Test • Connect a Patient Monitoring System. Verify proper communication with the Patient Monitoring System. • Alternatively you can check the signal availability with Datalogger Software: – DATALOGGER Software (download on Partner-net). – DATALOGGER Manual (download on Partner-net). – A Hamilton Ventilator with activated RS232 Interface or Communication Board. – A Windows XP or Windows 7 computer with RS232 or USB Interface – To connect HAMILTON-C2 use Interface Cable (PN 160366) 228 English | 624165/07 8 General Tests 1. Run HAMILTON-Datalogger.exe (available as download from the Hamilton Medical Partner-net). The Welcome window appears. 2. Click > ENTER Figure 8-210. Datalogger 3. The software automatically recognizes the connected device’s software: • For HAMILTON-C1/T1/C2/C3 ventilators, the Datalogger displays HAMILTON-C2. 4. To check whether the connection is working properly, start ventilation. If the Datalogger is working properly, the display shows ventilation values as well as the device serial number (S/N at Device information). Also check the Errors Counter on the right lower corner. NOTICE In addition, for all connections you may order USB to RS232 Converter (PN 396187, in case your computer has only an USB Interface). Important: The USB to RS232 Converter needs an additional driver to work properly. It is enclosed on the CD in the package. Hamilton Medical | HAMILTON-C2 Service Manual 229 8 Service Software 8.22 CO2 Sensor Accuracy Check Follow these steps to check the accuracy of the CO2 Monitor, if installed. In addition to the HAMILTON-C2 ventilator with CO2 Monitor installed, this check requires: • Patient Breathing Circuit (not required : Humidifier, Water Traps or Inspiratory Filter) • Test Lung • CO2 Sensor (PN 281718) • CO2 Adult Airway Adapter, Single Use, 10/box (PN 281719) • CO2 Adapters (3x), (one Box contains 25 pieces) (PN 281803) • Barometric Pressure Gauge (mbar) or TSI • CO2 Calibration Gas Cylinder or equivalent (CO2 5% ± 0.03; O2 0% ± 0.03; N2 Balance) (PN 159173) • CO2 Gas Regulator (PN 159172) 1. Enter the Service Software -> Tests/calibration > Sensor data > Sensor 3. 2. Note the displayed barometric pressure (Pressure sensor Pambient) and verify that it is within ± 20 mbar of the actual barometric pressure, measured with a second gauge or device (e.g. the TSI Flow Meter). Figure 8-211. Pressure Sensor Pambient 230 English | 624165/07 CO2 Sensor Accuracy Check 8 Figure 8-212. TSI Flow Meter 3. Restart the ventilator in operational mode. 4. Enter the Configuration Menu: Utilities > Configuration. Figure 8-213. Utilities Configuration Hamilton Medical | HAMILTON-C2 Service Manual 231 8 Service Software 5. Activate the CO2 Hardware Option: Utilities > Configuration > Options > HW options Figure 8-214. HW Options 6. Connect the CO2 Sensor with the CO2 Communication Board. 7. Activate the CO2 Sensor and deactivate the Oxygen Sensor: System > Sensors on/off. Set oxygen (FiO2) = 21%. 8. Wait 3 minutes to allow the CO2 Sensor to reach the working temperature. Figure 8-215. Sensor Update 232 English | 624165/07 CO2 Sensor Accuracy Check 8 9. Connect the Patient Breathing Circuit; Test Lung, CO2 Sensor, CO2 Adapters, CO2 Calibration Gas and CO2 Gas Regulator as shown. Figure 8-216. CO2 Sensor Accuracy Check Illustration Hamilton Medical | HAMILTON-C2 Service Manual 233 8 Service Software 10. Select the System Tab, Tests & calib, then CO2 Sensor to start the CO2 Sensor Calibration. Figure 8-217. CO2 Sensor Calibration 11 Start ventilation with standard settings. Select the PCO2 curve for displaying on the screen (see Figure 8-218 CO2 Sensor Accuracy Check). 12. Open the CO2 Regulator slightly so that the display shows a stable CO2 curve (see Figure 8-218 CO2 Sensor Accuracy Check). Figure 8-218. CO2 Sensor Accuracy Check NOTICE Only a small regulated gas flow is required to achieve a stable curve. To achieve a stable CO2 reading, ensure that the CO2 Sensor and CO2 Adapters are upright. 13. To get the PCO2 value in mmHg, press the Freeze Button (e.g. 34mmHg as shown). 234 English | 624165/07 8 CO2 Sensor Accuracy Check Example : if the Ambient Temperature is 20 °C, Barometric Pressure is 900 mbar, and displayed PCO2 reading is 34 mmHg, the displayed PCO2 Value is at the low end, but still within the range. Figure 8-219. Graph Based on Ambient Temperature 15. See the following graphs based on ambient temperature: • Ambient temperature 20 °C (17.5 to 22.5 °C): (see Figure 8-220) • Ambient temperature 25 °C (22.5 to 27.5 °C): (see Figure 8-221) • Ambient temperature 30 °C (27.5 to 32.5 °C): (see Figure 8-222) • Ambient temperature 35 °C (32.5 to 37.5 °C): (see Figure 8-223) Hamilton Medical | HAMILTON-C2 Service Manual 235 8 Service Software Figure 8-220. Acceptable Range for PCO2 at Atmospheric Pressure (17.5 to 22.5 °C) 236 English | 624165/07 CO2 Sensor Accuracy Check 8 Figure 8-221. Acceptable Range for PCO2 at Atmospheric Pressure (22.5 to 27.5 °C) Hamilton Medical | HAMILTON-C2 Service Manual 237 8 Service Software Figure 8-222. Acceptable Range for PCO2 at Atmospheric Pressure (27.5 to 32.5 °C) 238 English | 624165/07 CO2 Sensor Accuracy Check 8 Figure 8-223. Acceptable Range for PCO2 at Atmospheric Pressure (32.5 to 37.5 °C) Hamilton Medical | HAMILTON-C2 Service Manual 239 8 Service Software 8.23 Nurse Call Functional Check 1 Connect the nurse call connector (PN160166) to the HAMILTON-C2 Communication Board. 2 Check the three relays with a multimeter by measuring the resistance during ventilation, alarm off and alarm on between the following wires: Figure 8-224. Nurse Call Cable and Connector Table 8-12. Nurse Call Cable and Connector Callouts Connector Description Color 1 Alarm return normally closed brown 2 Alarm return normally open white 3 Alarm common green 4 I:E Contact 2 yellow 5 I:E Contact 1 grey 6 — — Table 8-13. Nurse Call Cable and Connector Actions Status green — white green — brown yellow — grey Alarm on closed open depending Alarm off open close depending Inspiration depending depending closed Expiration depending depending open 240 English | 624165/07 Nurse Call Functional Check 8 3. Check the three relays with a multimeter by measuring the resistance during ventilation, alarm off and alarm on between the following wires: • Enter Service software > Tests/Calibration > System test > Alarming (See Figure 8-225 Alarm On/Off Relay Check). • Activate/deactivate alarm (e.g. Alr. High). Figure 8-225. Alarm On/Off Relay Check 4. I:E relay check: • Restart the HAMILTON-C2 in operational mode • Start ventilation • Check I:E contact with the multimeter Hamilton Medical | HAMILTON-C2 Service Manual 241 8 Service Software 8.24 Final Tests 1. Install a USB Memory Stick into the HAMILTON-C2 USB Port. 2. Switch on the unit to get into Service Software mode. 3. Select the Download Menu and then Instr Report. 4. Start the Instr Report download. 5. Verify from the USB Memory Stick, that the download is complete. 6. Finish this unit test by saving the Instrument Report with your documentation. 8.25 Set the Service Timer Set the Service Timer. If the Reset Button is greyed out, the Alarm limit Knob needs to be activated. After doing so, the reset function is enabled. 8.26 Check the Battery Data NOTICE Depending on how frequent you use the ventilator, you can replace the battery if the State of Health (SoH) is < 40%. Refer to the Evaluation of the Battery State of Health Section 5.3 for details on how to read the battery data. 1. Check the Battery State of Health (SoH). 2. Replace the battery if: – the State of Health (SoH) is < 20% – the cell voltages are permanently asymmetrical, that is, the difference between cells voltage is more than 100 mV. 3. Calibrate the battery if the Error is ≥ 5%. 4. Record the battery data on the Summary of Test Report Section 17.1. 242 English | 624165/07 9 Technical Faults 9.1 Technical Faults Overview.....................................................................................................................244 9.2 Troubleshooting...................................................................................................................................246 243 9 Technical Faults 9.1 Technical Faults Overview NOTICE If more than one alarm is active, the associated alarm messages alternate in the message bar. Alarm Indications Table 9-1. Alarm Indications in the Ventilator Alarm Type Message Bar Alarm Lamp Audio Action Required High-priority alarm Red, with alarm message Red A sequence of 5 beeps, repeated until the alarm is reset. If the audible alarm is not silenced during the first minute, the continuous-tone buzzer also sounds. The patient’s safety is compromised. The patient needs immediate attention. Medium-priority alarm Yellow, with alarm message Yellow A sequence of 3 beeps, repeated periodically. If the audible alarm is not silenced during the first minute, the continuous-tone buzzer also sounds. The patient needs prompt attention. Low-priority alarm Yellow, with alarm message Yellow Two sequences of beeps. This is not repeated. Operator awareness is required. Technical fault Red, with Safety ventilation: xxxxxx or Technical fault: xxxxxx Red Same as for high-priority alarm, if technically possible. At the minimum a continuous buzzer tone. The buzzer cannot be silenced. The ventilator enters safety mode, or, if it cannot safely ventilate, the ambient state. Provide alternative ventilation. Turn off the ventilator. Have the ventilator serviced. Figure 9-1. Alarm Class Table 9-2. Alarm Class Alarm Codes Alarm Class Patient Alarms - 100000 Alarm Codes Section 16.2.1 Patient Alarms Technical Alarms - 200000 Alarm Codes Section 16.2.2 Technical Events Technical Failure Ending in Safety Mode - 300000 Alarm Codes Section 16.2.3 Technical Failures resulting in the ventilator switching to the Safety Mode Technical Failure Ending in Ambient Mode - 400000 Alarm Codes Section 16.2.4 Technical Failures resulting in the ventilator switching to the Ambient Mode 500000 Alarm Codes Section 16.2.5 System Failures; invisible TF for user, no action required. For internal use. 244 English | 624165/07 Technical Faults Overview 9 • Patient Alarms: (Alarm code 100000) Alarms related to : Pressure, Flow, Volume, Apnea – High Priority Alarms – Medium Priority Alarms – Low Priority Alarms • Technical Alarms: (Alarm code 200000) Alarms related to: Temperature, Battery, Valves, Sensors – High Priority Alarms – Medium Priority Alarms – Low Priority Alarms Technical Faults Technical Faults end up in one of the following states: • Fatal -> Safety Mode: (Alarm code 300000) – In case of some technical failure the ventilator switches to SAFETY Mode. This gives the user time for corrective actions, such as organizing a replacement ventilator. The turbine runs constant to create Pinsp. The expiratory valve switches system pressure levels between PEEP and inspiratory pressure. Patient monitoring is nonfunctional during safety ventilation. You must turn off the ventilator power to exit safety ventilation. • Fatal -> Ambient Mode: (Alarm code 400000) – If the technical fault alarm is serious enough to possibly compromise safe ventilation, the ventilator enters the ambient state. The inspiratory channel and expiratory valve are opened, letting the patient breathe room air unassisted. You must switch off ventilator power to exit the ambient state. • Invisible Technical Faults (Alarm code 500000) / Technical Note (since < 2.2.0) – These are classed as “Invisible Technical Faults” and are not shown on the LCD Display during operation. They refer to exception handling performed by ventilator during operation, and are for Hamilton Medical internal use. Do not consult Hamilton Medical AG Technical Support concerning these technical faults. They have no significance for hospital workers or Hamilton-Medical Trained Field Service Engineers. – For detail description go to Glossary - Alarm Codes Section 16.2. Hamilton Medical | HAMILTON-C2 Service Manual 245 9 Technical Faults 9.2 Troubleshooting For troubleshooting, refer to document number 624485 "Troubleshooting Platform C" available for download from the Hamilton Medical Partner-net. Follow suggested troubleshooting steps in order as described. Follow these steps when troubleshooting: 1. Confirm that the test configuration is correct. 2. When replacing components, repeat the applicable test following each replacement. If the test passes successfully, repeat all test's in order (see the Service Software Section 8). 3. If a problem cannot be resolved, enter a Customer Event Report (CER) on the Hamilton Medical AG on the Online Help Desk (https://hamilton.ky2help.com). Export the Event Log and Instrument Report for failure analysis. Check regularly the knowledge base of the ky2help for the latest information and troubleshooting. 246 English | 624165/07 10 Components Removal/Assembly 10.1 Notes on Maintenance and Replacement .............................................................................................248 10.2 Securing the Power Cord .....................................................................................................................249 10.3 Interaction Panel Components Removal/Assembly ................................................................................250 10.4 Ventilation Unit Components Removal/Assembly..................................................................................262 10.5 Ventilation Unit Components, Tubings and Cables Assembly................................................................298 247 10 Components Removal/Assembly 10.1 Notes on Maintenance and Replacement WARNING You must perform the repairs detailed in this section only as instructed in the Preventive Maintenance and Testing Overview Section 6.1, Hospital Preventive Maintenance Section 6.2 and Engineer Preventive Maintenance Section 6.3. Service the HAMILTON-C2 only as described in this manual, using only parts approved or supplied by Hamilton Medical AG. Incorrectly repaired parts, components or assemblies could result in patient injury. See available Spare Parts in the Spare Parts Section 12. − Always switch the HAMILTON-C2 OFF and disconnect the HAMILTON-C2 from the external Mains Power and remove the battery, before opening the cover. − Never use any kind of lubrication on any part of the HAMILTON-C2. − Some parts may be hot after operational use (Blower Module, Rinse Flow Valve, Power Supply, etc.). CAUTION After performing Maintenance or Replacement of a component or module, perform the necessary Service Software checks, calibrations and Safety Tests to ensure the HAMILTON-C2 is performing properly. Hamilton Medical AG does not permit repairs to parts that are supplied as an assembly. For example: − You have to replace the Blower Module as a complete assembly − You have to replace the Expiratory Valve as a complete assembly − You have to replace the Mixer Block as complete assembly − You have to replace Hamilton Medical AG printed circuit boards only Make sure to take full ESD (electrostatic discharge) precautions before opening the HAMILTON-C2. Refer to the ESD (Electrostatic Discharge ) Protection Section 11.3.3. When the HAMILTON-C2 is switched ON, even when not connected to the external Mains Power, the Battery Pack(s) supply power. Therefore, a short circuit is possible when the HAMILTON-C2 is switched ON. NOTICE Always send defective parts, components or assemblies to Hamilton Medical AG with a completed Return Goods Authorization (RGA) request. Before making any repairs, remove from the HAMILTON-C2: − External Mains Power Supply and Batteries − Oxygen Supply − Patient Breathing Circuit (if not needed) − Flow Sensor Tubing (if not needed) Update the Technical State if necessary after exchanging the parts and restart the ventilator. This must be done before performing the Service Software. 248 English | 624165/07 Securing the Power Cord 10 10.2 Securing the Power Cord NOTICE To prevent unintentional disconnection of the Power Cord, make sure it is well seated into the ventilator’s socket and secured with the Power Cord Security Clamp. The HAMILTON-C2 does not require protective earth grounding, because it is a class II device, as classified according to IEC 60601-1. Securing the Power Cord with the suitable Security Clamp. Secure the Power Cord with the Security Clamp as shown in the picture below. Two different Security Clamps are available. Use the suitable Security Clamp as listed in the table below. Figure 10-1. Securing the Power Cord Table 10-1. Security Clamp Part Number 361061 Description Image Cable Security Clamp D=5.0 for Power Cords: • PN 355199 GB • PN 355200 EU 361097 Cable Security Clamp D=6.2 for Power Cords: • PN 355198 USA • PN 355308 CN Replacing the Power Cord To replace the Power Cord, release the screw (PN 420724) of the Security Clamp. When reconnecting the Power Cord, make sure it is well seated into the ventilator’s socket and secured with the Power Cord Security Clamp. Hamilton Medical | HAMILTON-C2 Service Manual 249 10 Components Removal/Assembly 10.3 Interaction Panel Components Removal/Assembly 10.3.1 Ventilation Unit Top Cover Removal/Assembly NOTICE The Interaction Panel must be tilted completely to the front to remove the Top Cover. To remove the Ventilation Unit Top Cover: 1. Lift the Handle of the Ventilation Unit. 2. Remove 6 Torx screws from the Top Cover of the Ventilation Unit: • A: 2 Torx screws (PN 420682) • B: 4 Torx screws (PN 420642) 3. Remove the Ventilation Unit Top Cover (C). Figure 10-2. Ventilation Unit Top Cover Removal 4. Assemble in the reverse order of removal. 250 English | 624165/07 Interaction Panel Small Rear Cover Removal/Assembly 10 10.3.2 Interaction Panel Small Rear Cover Removal/Assembly WARNING Torx screws (A) cannot be any longer than the specified size (M3x6), otherwise, damage can be caused to the Interaction Panel Board. To remove the Interaction Panel Small Rear Cover (PN 160402): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10-2). 2. Remove 4 Torx screws (A, PN 420642) from the Small Rear Cover. 3. Remove the Small Rear Cover (B). Figure 10-3. Interaction Panel Small Cover Removal 4. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 251 10 Components Removal/Assembly 10.3.3 Interaction Panel Removal/Assembly To remove the Interaction Panel: 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 2. Remove the Security Strap (C) from the 1 Flexible Flat Cable and disconnect it from the Ventilation Unit Mainboard. 3. Disconnect the other 2 Flexible Flat Cables (D) from the Interaction Panel Board. NOTICE The 2 Flexible Flat Cables (D) can more easily be disconnected from the Interaction Panel Board, but can also be disconnected from the Ventilation Unit Mainboard. The ferrite cores for these 2 cables are secured to the top of the front cover, and are not intended to be removed with the cables. 4. Remove 4 Torx screws (B, PN 420727) from the Hinges of the Interaction Panel. 5. Remove the Interaction Panel (A) from the Ventilation Unit. 6. Place aside in a safe location. Figure 10-4. Interaction Panel Removal 6. Assemble in the reverse order of removal. 7. Update the Technical State (refer to Modify Tab Section 8.8.3). 252 English | 624165/07 Interaction Panel Large Rear Cover Removal/Assembly 10 10.3.4 Interaction Panel Large Rear Cover Removal/Assembly To remove the Interaction Panel Large Rear Cover (PN 160326): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Cover Rear Cover Removal/Assembly Section 10.3.2). 3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 3. Remove 5 Torx screws (A, PN 420642) from the Large Rear Cover. 4. Remove the Large Rear Cover (B). Figure 10-5. Interaction Panel Large Rear Cover Removal 5. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 253 10 Components Removal/Assembly 10.3.5 Backlight Driver Board Removal/Assembly WARNING When replacing the GRAPHIC LCD TFT VGA 10.4” NEC Display (Obsolete, PN 380031), order the Display Upgrade Kit (MSP380046). The DC/AC Backlight Inverter Boards (Obsolete, PN 396197 or PN 396314) are not compatible with the latest GRAPHIC LCD TFT VGA 10.4" NLT Display. Use the Backlight Driver Board (PN 396314) if the Front Panel Boards is below Revision 08. Front Panel Boards at Revision 08 and higher have integrated Backlight Driver Components (see the Front Panel Board Removal/Assembly Section 10.3.6 for configuration instructions). To remove the DC/AC Backlight Inverter Boards (Obsolete, PN 396197 or PN 396314) or the Backlight Driver Board (PN 396314): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section 10.3.4). 5. Disconnect 2 Cables (A) from the Backlight Driver Board (B). 6. Remove 2 Torx screws (C, PN 420692) from the Backlight Driver Board. 7. Remove the Backlight Driver Board (B). Figure 10-6. Backlight Inverter Board Removal 8. Assemble in the reverse order of removal. 254 English | 624165/07 Front Panel Board Removal/Assembly 10 10.3.6 Front Panel Board Removal/Assembly WARNING The Backlight LED Driver Components are integrated into Front Panel Boards with Revision 08 and higher, compatible with the latest GRAPHIC LCD TFT VGA 10.4" NLT Display. The installation of an External Backlight Driver Board is not necessary when you configure Jumper P14 to pins 1 and 2 on the new Front Panel Board. An Extension Cable (PN 161569) is used to bridge the connections for the removed Backlight Driver Board (see instructions below). CAUTION Update the Technical State (refer to Modify Tab Section 8.8.3). To Remove the Front Panel Board (MSP160196): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section 10.3.4). 5. Disconnect 1 cable (A) from the Backlight Driver Board and 4 cables (B) from the Interaction Panel Board. Figure 10-7. Interaction Panel Board Removal Step 1 Hamilton Medical | HAMILTON-C2 Service Manual 255 10 Components Removal/Assembly 6. Remove 5 Torx screws (C, PN 420724) and 2 Torx screws (D, PN 420641) from the Interaction Panel Board. 7. Remove the Interaction Panel Board (E). Figure 10-8. Interaction Panel Board Removal Step 2 8. Note the location of Jumper P14 on the Front Panel Board. Figure 10-9. Location of Jumper Pins on the Front Panel Board 256 English | 624165/07 Front Panel Board Removal/Assembly 10 9. Connect Jumper P14 to Pins 1 and 2 on the new Front Panel Board if no external Backlight Driver Board is installed. Connect the Extension Cable (PN 161569) from the Backlight Connector to P13 on the Front Panel Board. Pins 1 and 2 if no External Backlight Driver Board is Installed Figure 10-10. Pins 1 and 2 if no External Backlight Driver Board is Installed 10. Connect Jumper P14 to Pins 2 and 3 when using an external Backlight Driver Board with the new Front Panel Board. 11. Connect the Backlight Driver Board between the Backlight Cable and P7 Connector on the Front Panel Board. Figure 10-11. Pins 2 and 3 if using an External Backlight Driver Board 12. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 257 10 Components Removal/Assembly 10.3.7 LCD Display Removal/Assembly WARNING When replacing the GRAPHIC LCD TFT VGA 10.4” NEC Display (Obsolete, PN 380027 or PN 380031), order the Display Upgrade Kit (MSP380046). To remove the LCD Display (PN 380031 or PN 380046): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section 10.3.4). 5. Remove the Front Panel Board (refer to the Front Panel Board Removal/Assembly Section 10.3.6). 6. Remove 2 Torx screws (A, PN 420641), 1 washer (B, PN 409105) and 2 Hex Standoff screws (C, PN 257038). 7. Remove the LCD Display (D). Figure 10-12. LCD Display Removal 8. Assemble in the reverse order of removal. 258 English | 624165/07 LCD Display Backlight Removal/Assembly 10 10.3.8 LCD Display Backlight Removal/Assembly To remove the LCD Display Backlight (PN 380044 / SN 1676 > PN 380030): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section 10.3.4). 5. Push down the small black lever (A) and pull out the LCD Display Backlight (B). Figure 10-13. LCD Display Backlight Removal Step 1 6. Remove the LCD Display Backlight (B). Hamilton Medical | HAMILTON-C2 Service Manual 259 10 Components Removal/Assembly Figure 10-14. LCD Display Backlight Removal Step 2 7. Insert a new LCD Display Backlight. 8. Assemble in the reverse order of removal. 260 English | 624165/07 P&T Control Knob Encoder Removal/Assembly 10 10.3.9 P&T Control Knob Encoder Removal/Assembly CAUTION The P&T Control Knob must be positioned to allow free movement. Attach so there is a 3 mm gap between the back of the Knob and the face of the Interaction Panel. Test the Knob actions for both Press and Turn. CAUTION Do not remove the P&T Control Knob for maintenance. Removal of the P&T Control Knob can damage the P&T Control Knob Encoder. Only remove if necessary to replace a non-functioning P&T Control Knob Encoder. To remove the P&T (Press and Turn) Knob Encoder (PN 372036): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section 10.3.4). 5. Disconnect the Cable from the Front Panel Board. 6. Remove the P&T Control Knob (A, PN 159449) from the front of the Interaction Panel. 7. Remove the Hex Nut (B) and Washer (C) from the P&T Control Knob Encoder (D). 8. Remove the P&T Control Knob Encoder (D) and Large Washer (E). Figure 10-15. P&amp;T Control Knob Encoder Removal 9. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 261 10 Components Removal/Assembly 10.4 Ventilation Unit Components Removal/Assembly 10.4.1 Backup Battery Pack Removal/Assembly CAUTION It is mandatory that the ventilator is operated with at least one battery installed. NOTICE − The Backup Battery Compartment provides space for 2 Backup Battery Packs. The primary Battery Pack is positioned into the left compartment. An Optional 2nd Battery Pack is positioned into the right compartment. − A Battery Lock is used to secure the Battery Release Latch. The Battery Release Latch must be in the latched position for the Battery Door to close. To remove Backup Battery Pack (MSP369106): 1. Open the Backup Battery Door by depressing the Door Latch (A). Figure 10-16. Backup Battery Pack Removal Step 1 262 English | 624165/07 Backup Battery Pack Removal/Assembly 10 2. Open the Battery Lock (B) by rotating the Battery Lock counter-clockwise with a screwdriver. Figure 10-17. Backup Battery Pack Removal Step 2 3. Raise the Battery Release Latch (C) to release the Battery Pack (D) from the Battery Compartment. 4. Remove the Battery Pack (D). Figure 10-18. Backup Battery Pack Removal Step 3 Hamilton Medical | HAMILTON-C2 Service Manual 263 10 Components Removal/Assembly 5. If a second Battery Pack is used, release the Battery Release Latch (E) and remove the Battery Pack (F) from the Battery Compartment. Figure 10-19. Backup Battery Pack Removal Step 4 6. Assemble in the reverse order of removal. 264 English | 624165/07 Filter Cover Removal/Assembly 10 10.4.2 Filter Cover Removal/Assembly To remove the Filter Cover (PN 160343): 1. Detach from the top and swivel (A) away from the rear of the Ventilation Unit. 2. Remove the Filter Cover (B). Figure 10-20. Filter Cover Removal 3. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 265 10 Components Removal/Assembly 10.4.3 HEPA Filter Removal/Assembly To remove the HEPA Filter (PN 160216): 1. Remove the Filter Cover Panel (refer to the Filter Cover Panel Removal/Assembly Section 10.4.2). 2. Raise the HEPA Filter Latch (A, PN 160234). 3. Use the Handle (B) to pull the HEPA Filter from the holder. 4. Remove the HEPA Filter (C). Figure 10-21. HEPA Filter Removal 5. Assemble in the reverse order of removal. 266 English | 624165/07 High Pressure Oxygen Connector Removal/Assembly 10 10.4.4 High Pressure Oxygen Connector Removal/Assembly To remove the High Pressure Oxygen Connector, DISS (PN 160470) or NIST (PN 160471): 1. Remove 2 Torx screws (B, PN 420734) mounting the High Pressure Oxygen Connector (A). 2. Remove the High Pressure Oxygen Connector (A). Figure 10-22. High Pressure Oxygen Connector Removal 3. Remove the O2 Inlet Filter. 4. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 267 10 Components Removal/Assembly 10.4.5 Front and Rear Covers Removal/Assembly NOTICE It is not necessary to remove the Interaction Panel from the Front Cover. To remove the Front (MSP160480) and Rear Covers (MSP160481): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove 2 Phillips screws (B, PN 420730) to remove the Communication Board Cover (A). 5. Remove 4 Torx screws (C, PN 420667) from the back of the Rear Cover (D). 6. Remove the Rear Cover (D). 7. Remove 1 Torx screw (E, PN 420667) from inside the Battery Compartment. 8. Remove the Expiratory Valve Cover. 9. Depress the Trolley Release Front Latch to remove the Front Cover (F). Figure 10-23. Front and Rear Covers Removal 10. Assemble in the reverse order of removal. 268 English | 624165/07 ESM Module Removal/Assembly 10 10.4.6 ESM Module Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). NOTICE Export the Technical State (Instrument Report) before removing the Ventilation Unit ESM (refer to the Export Section 8.18.3) because, after removing the Ventilation Unit ESM, the Technical State will be lost. To remove the ESM Module (MSP160206): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove 4 screws (A, PN 420864). 3. Remove the top cover of the ESM Module Shield Housing. 4. Remove 4 standoffs (PN 257100) holding the ESM Module (B), 5. With both hands, disconnect the ESM Module (B) from 2 connector sockets (C) on the Mainboard and carefully lift the ESM Module. 6. Remove the ESM Module (B). Figure 10-24. ESM Module Removal 7. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 269 10 Components Removal/Assembly 10.4.7 Ventilation Unit Mainboard Removal/Assembly CAUTION Take care not to damage the Top Foam Section Latches used to secure the Ventilation Unit Mainboard. NOTICE Export the Technical State (Instrument Report) before removing the Ventilation Unit Mainboard (refer to the Export Section 8.18.3) because, after removing the Ventilation Unit Mainboard the Technical State will be lost. CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). Import the Technical State you have exported before and change the revision and serial number of the replacement Mainboard. To remove the Ventilation Unit Mainboard (MSP160200 or MSP160382): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Disconnect all cabling connections from the Ventilation Unit Mainboard. 6. Remove the Ventilation Unit Mainboard (A) from the Top Foam Section. 7. Remove the ESM Holder (PN 257080) and LCD Cable Lock Holder (PN 362051). Figure 10-25. Ventilation Unit Mainboard Removal 8. Assemble in the reverse order of removal. 9. After entering the new Revision and Serial Number of the Mainboard, you will be asked to enter the Blower Timer and Operating Hours of the unit. To do this, please delete the description in the input field (Blower % and Device Hour) and enter the corresponding value in numbers. For example: 5 for 5% Blower Timer / 100 for 100 Operating Hours. 270 English | 624165/07 Top Foam Removal/Assembly 10 10.4.8 Top Foam Removal/Assembly To remove the Top Foam (PN 160239): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Disconnect all cables connected to the Ventilation Unit Mainboard. Figure 10-26. Top Foam Removal Step 1 Hamilton Medical | HAMILTON-C2 Service Manual 271 10 Components Removal/Assembly 6. Lift the Top Foam Section (A) from the Ventilation Unit. Figure 10-27. Top Foam Removal Step 2 7. Assemble in the reverse order of removal. 272 English | 624165/07 Cooling Fan Removal/Assembly 10 10.4.9 Cooling Fan Removal/Assembly CAUTION Be careful not to bend the FFC Pins when re-assembling the Blower Module. NOTICE The air flow of the Fan is directed into the device. When the power is on, hold a piece of paper against the Fan opening. If the Fan is installed correctly the paper, will be drawn against the Fan. To remove the Cooling Fan which includes the small board (PN 160346): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Lift the Cooling Fan (A) from the Middle Foam Section (B) of the Ventilation Unit. Figure 10-28. Cooling Fan Removal 7. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 273 10 Components Removal/Assembly 10.4.10 Blower Module Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). NOTICE If Software > 2.2.3 is installed, you have to set the Blower Timer manually, after replacement of the Blower Module To remove the Blower Module (MSP160250): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Lift the Blower Module (A) from the Middle Foam Section (B) of the Ventilation Unit. Figure 10-29. Blower Module Removal 7. Assemble in the reverse order of removal. 274 English | 624165/07 Power Supply Removal/Assembly 10 10.4.11 Power Supply Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). To remove the Power Supply (MSP160600): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Lift the Power Supply (A) from the Middle Foam Section (B) of the Ventilation Unit and move away from the back to have access to the AC Mains Power Cable (C). 7. Disconnect the AC Mains Power Cable (C). 8. Remove the Power Supply (A). Figure 10-30. Power Supply Removal 9. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 275 10 Components Removal/Assembly 10.4.12 Middle Foam Removal/Assembly CAUTION Be sure that all cables are free from the Middle Foam Section. To remove the Middle Foam (PN 160238): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Unscrew the Tube Flange (A) used to fit the Blower Module (refer to the Two Pin Cylindrical Service Tool, PN 500314 (see the Two Pin Cylindrical Service Tool Section 11.3.4)). 10. Lift the Middle Foam Section (B) from the Ventilation Unit. Figure 10-31. Middle Foam Removal Step 1 276 English | 624165/07 Middle Foam Removal/Assembly 10 Figure 10-32. Middle Foam Removal Step 2 11. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 277 10 Components Removal/Assembly 10.4.13 Pressure Sensor Assembly Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). To remove the Pressure Sensor Assembly (MSP160300): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the Tubing (B) from the left side of the Pressure Sensor Assembly (A). Figure 10-33. Pressure Sensor Assembly Removal Step 1 278 English | 624165/07 Pressure Sensor Assembly Removal/Assembly 10 NOTICE The tubes are color coded. 11. Lift the Pressure Sensor Assembly (A) slightly to remove the 2 Tubings (B) from the bottom of the Pressure Sensor Assembly. Figure 10-34. Pressure Sensor Assembly Removal Step 2 12. Remove the Pressure Sensor Assembly (A). 13. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 279 10 Components Removal/Assembly 10.4.14 Inspiratory Valve Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). To remove the Inspiratory Valve (MSP160230): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the Pressure Sensor Assembly (refer to the Pressure Sensor Assembly Removal/Assembly Section 10.4.13). 11. Disconnect 3 Tubings (A) from the Inspiratory Valve, Qvent Flow Sensor and Ambient Valve. Figure 10-35. Inspiratory Valve Removal Step 1 280 English | 624165/07 Inspiratory Valve Removal/Assembly 10 12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the Ventilation Unit. Figure 10-36. Inspiratory Valve Removal Step 2 13. Disconnect the Inspiratory Valve (B) from the Qvent Flow Sensor (C). 14. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 281 10 Components Removal/Assembly 10.4.15 Qvent Flow Sensor Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). To remove the Qvent Flow Sensor (MSP161657 (Sensirion) or MSP399124 (TSI)): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the Pressure Sensor Assembly (refer to the Pressure Sensor Assembly Removal/Assembly Section 10.4.13). 11. Disconnect 3 Tubings (A) from the Inspiratory Valve, Qvent Flow Sensor and Ambient Valve. Figure 10-37. Inspiratory Valve Removal Step 1 282 English | 624165/07 Qvent Flow Sensor Removal/Assembly 10 12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the Ventilation Unit. Figure 10-38. Inspiratory Valve Removal Step 2 13. Disconnect the Inspiratory Valve (B) from the Qvent Flow Sensor (C). 14. Not shown; Remove a metal bracket (PN 160741) attached with 2x screws (PN 420636) secures the Qvent Flow Sensor to the Ambient Valve. 15. Disconnect the Qvent Flow Sensor (C) from the Ambient Valve (D). 16. Remove the Qvent Flow Sensor (C). 17. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 283 10 Components Removal/Assembly 10.4.16 Ambient Valve Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). To remove the Ambient Valve (MSP160290): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2. 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the Pressure Sensor Assembly (refer to the Pressure Sensor Assembly Removal/Assembly Section 10.4.13). 11. Disconnect all Tubings (A) from the Inspiratory Valve, Qvent Flow Sensor and Ambient Valve. Figure 10-39. Inspiratory Valve Removal Step 1 284 English | 624165/07 Ambient Valve Removal/Assembly 10 12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the Ventilation Unit (). Figure 10-40. Inspiratory Valve Removal Step 2 13. Not shown; Remove a metal bracket (PN 160741) attached with 2x screws (PN 420636) secure the Qvent Flow Sensor to the Ambient Valve. 14. Disconnect the Qvent Flow Sensor (C) from the Ambient Valve (D). 15. Remove the Ambient Valve (D). 16. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 285 10 Components Removal/Assembly 10.4.17 Oxygen Sensor Block Removal/Assembly To remove the Oxygen Sensor Block (PN 160100): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Disconnect 2 tubings (A and B) from the Oxygen Sensor Block. Figure 10-41. Oxygen Sensor Block Removal Step 1 286 English | 624165/07 Oxygen Sensor Block Removal/Assembly 10 11. Lift the Oxygen Sensor Block (C) from the Bottom Foam. 12. Remove the Oxygen Sensor Block (C). Figure 10-42. Oxygen Sensor Block Removal Step 2 13. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 287 10 Components Removal/Assembly 10.4.18 Rinse Flow Block Removal/Assembly To remove the Rinse Flow Block (MSP160498): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the top tubing (A) from the Oxygen Sensor Block (refer to the Oxygen Sensor Block Removal/Assembly Section 10.4.17). 11. Remove the Rinse Flow Tube (B) from the Inspiratory Valve. Figure 10-43. Rinse Flow Block Removal Step 1 288 English | 624165/07 Rinse Flow Block Removal/Assembly 10 12. Tubings (C and E) should be already disconnected from the Pressure Sensor Board. 13. Remove the Nebulizer Tube (D). 14. Disconnect the Qvent Flow Sensor cable and remove from the mount on the Rinse Flow Block. Figure 10-44. Rinse Flow Block Removal Step 2 15. Lift the Rinse Flow Block (F) from the Bottom Foam. Hamilton Medical | HAMILTON-C2 Service Manual 289 10 Components Removal/Assembly Figure 10-45. Rinse Flow Block Removal Step 3 16. Assemble in the reverse order of removal. 290 English | 624165/07 Expiratory Valve Removal/Assembly 10 10.4.19 Expiratory Valve Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). To remove the Expiratory Valve (MSP160240): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Lift the Expiratory Valve Assembly (A) from the Bottom Foam Section of the Ventilation Unit. Figure 10-46. Expiratory Valve Removal 11. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 291 10 Components Removal/Assembly 10.4.20 Battery Compartment Removal/Assembly To remove the Battery Compartment (PN 160302): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly 10.4.12). 10. Remove 2 Torx screws (C, PN 420699) to remove the Cable Clamp (B, PN 160405) from the rear of the Battery Compartment (A). Figure 10-47. Battery Compartment Removal Step 1 292 English | 624165/07 Battery Compartment Removal/Assembly 10 6. Lift the Battery Compartment (A) from the Bottom Foam Section of the Ventilation Unit. Figure 10-48. Battery Compartment Removal Step 2 8. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 293 10 Components Removal/Assembly 10.4.21 Oxygen Mixer Block Assembly Removal/Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). Also update the Technical State of the QO2 Flow Sensor. To remove the Oxygen Mixer Block Assembly: • MSP160608 (with Sensirion QO2 Flow Sensor, MSP161657) or • MSP160226 (with TSI QO2 Flow Sensor, MSP399123) 1. Remove the Top Cover from the Ventilation Unit (refer to the Top Cover Ventilation Unit Removal/Assembly 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly 10.4.12). 10. Remove the Nebulizer Tubing. 11. Remove 3 Torx screws (C, PN 420657) attaching the Oxygen Mixer Block Assembly (A) to the Base Plate (B). 12. Remove the Oxygen Mixer Block Assembly (A). Figure 10-49. Oxygen Mixer Block Assembly Removal 294 English | 624165/07 Oxygen Mixer Block Assembly Removal/Assembly 10 13. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 295 10 Components Removal/Assembly 10.4.22 Nebulizer Valve Removal/Assembly To remove the Nebulizer Valve (PN 160400) from the Mixer Block: 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the Oxygen Mixer Block Assembly (refer to the Oxygen Mixer Block Assembly Removal/Assembly Section 10.4.21). 11. Remove 2 Phillips screws (A) from the Nebulizer Valve (B). 12. Remove the Nebulizer Valve (B). Figure 10-50. Nebulizer Valve Removal 13. Assemble in the reverse order of removal. 296 English | 624165/07 Bottom Foam Removal/Assembly 10 10.4.23 Bottom Foam Removal/Assembly To remove the Bottom Foam (PN 160237): 1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10.3.1). 2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section 10.3.2). 3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section 10.4.5). 5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8). 6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9). 7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10). 8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11). 9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12). 10. Remove the Oxygen Mixer Block Assembly (refer to the Oxygen Mixer Block Assembly Removal/Assembly Section 10.4.21). 11. Remove the Bottom Foam (A) from the Ventilation Unit Base Frame. Figure 10-51. Bottom Foam Removal 12. Assemble in the reverse order of removal. Hamilton Medical | HAMILTON-C2 Service Manual 297 10 Components Removal/Assembly 10.5 Ventilation Unit Components, Tubings and Cables Assembly 10.5.1 Bottom Foam Assembly 1. The Bottom Foam (A, PN 160237) is positioned onto the Chassis Frame (B). Figure 10-52. Bottom Foam Assembly 298 English | 624165/07 Mains Power Cable Assembly 10 10.5.2 Mains Power Cable Assembly 1. The Mains Power Cable (C, PN 160348) is connected to the Mains Power Connector (A), with one side connected to the 2A Circuit Breaker (D, PN 378009), then routed through a Ferrite Core (B) and positioned on the side of the Mixer Block Assembly (C). 2. It will then be connected to the Power Supply when assembled. Figure 10-53. Mains Power Cable Assembly Hamilton Medical | HAMILTON-C2 Service Manual 299 10 Components Removal/Assembly 10.5.3 DC Input Power Cable Assembly 1. The DC Input Power Cable (B, PN 160372) from the DC Power Connector (A) is positioned on the side of the Mixer Block Assembly (B) and will be attached to the Battery Compartment (C) when assembled. 2. It will then be connected to the Ventilation Unit Mainboard when assembled. Figure 10-54. DC Input Power Cable Assembly 300 English | 624165/07 Oxygen Mixer Block Assembly Mounting 10 10.5.4 Oxygen Mixer Block Assembly Mounting 1. The Oxygen Mixer Block Assembly (A) is mounted to the Chassis (B) with 3 Torx screws (C, 2x PN 420667 and 1x PN 420727). Figure 10-55. Oxygen Mixer Block Assembly Mounting Hamilton Medical | HAMILTON-C2 Service Manual 301 10 Components Removal/Assembly 10.5.5 Oxygen Mixer Valve Cable Assembly 1. The Oxygen Mixer Valve Cable (D) (part of the Oxygen Mixer Valve, no part number) from the Oxygen Mixer Valve (A) is positioned over the Low Pressure Oxygen Input Tubing (B) and behind the 2A Circuit Breaker (C), then placed to the side of the Mixer Block Assembly (D). Figure 10-56. Oxygen Mixer Valve Cable Assembly 2. The Cable will route on the side of the Middle and Top Foam Sections when assembled. 3. It will then be connected to the Ventilation Unit Mainboard when assembled. 302 English | 624165/07 Nebulizer Cable Assembly 10 10.5.6 Nebulizer Cable Assembly 1. The Nebulizer Cable (A) (part of the Nebulizer Valve, no part number) from the Nebulizer Valve is positioined in a channel in the Bottom Foam Section (B) and routed to the right side of the Ventilator (C). Figure 10-57. Nebulizer Cable Assembly Top View 2. The Cable will route on the side of the Middle and Top Foam Sections when assembled. 3. It will then be connected to the Ventilation Unit Mainboard when assembled. Hamilton Medical | HAMILTON-C2 Service Manual 303 10 Components Removal/Assembly 10.5.7 Battery Compartment Assembly 1. The Battery Compartment (A) is positioned at the rear of the Bottom Foam Section. 2. The Battery Power Cable (B, PN 160350) and the Battery Data FFC (C, PN 160351) are routed to the rear of the Battery Compartment (D). They are secured along with the DC Input Power Cable (E, PN 160372) to the rear of the Battery Compartment with the Cable Clamp (F, PN 160405) and 2 Torx screws (G, PN 420699). Figure 10-58. Battery Compartment Assembly 3. It will then be connected to the Ventilation Unit Mainboard when assembled. 304 English | 624165/07 Expiratory Valve Assembly 10 10.5.8 Expiratory Valve Assembly 1. The Expiratory Valve (A, MSP160240) is positioned with 3 slots (C) into the Bottom Foam Section (B). Figure 10-59. Expiratory Valve Assembly Step 1 2. The Expiratory Valve FFC (D) (part of the Expiratory Valve, no part number) is routed under a channel in the Bottom Foam Section to the side. Figure 10-60. Expiratory Valve Assembly Step 2 3. The Cable will route on the side of the Bottom, Middle and Top Foam Sections when assembled. 4. It will then be connected to the Ventilation Unit Mainboard when assembled. Hamilton Medical | HAMILTON-C2 Service Manual 305 10 Components Removal/Assembly 10.5.9 Rinse Flow Block and Tubing Assembly 1. Insert the Rinse Flow Block (A, MSP160498) into the Bottom Foam Section. Figure 10-61. Rinse Flow Block and Tubing Assembly Step 1 2. Connect the Proximal Flow Sensor Tubing (B, from Tubing Set MSP160476) and the Distal Flow Sensor Tubing (D, from Tubing Set MSP160476) to the Rinse Flow Block. 3. Connect the Nebulizer Tubing (C, PN 160409) from the Mixer Block Fitting and the Nebulizer Connector to the Rinse Flow Block. 5. The other tubing ends (B and D) will connect to the Pressure Sensor Assembly when assembled. 306 English | 624165/07 Rinse Flow Block and Tubing Assembly 10 Figure 10-62. Rinse Flow Block and Tubing Assembly Step 2 Hamilton Medical | HAMILTON-C2 Service Manual 307 10 Components Removal/Assembly 4. If removed, connect the Oxygen Sensor Tubing (E). 3. Connect the tubing (F) to the Inspiratory Valve which contains a Flow Restrictor. Figure 10-63. Rinse Flow Block and Tubing Assembly Step 3 308 English | 624165/07 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly 10 10.5.10 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly CAUTION Update the Technical State (refer to the Modify Tab 8.8.3). NOTICE The Ambient Valve cable (F) is part of the Ambient Valve Assembly and has no part number. The Inspiratory Valve FFC (G) is part of the Inspiratory Valve and has no part number. 1. Pre-assemble the Inspiratory Valve (A, MSP160230), Qvent Flow Sensor (B, MSP161657 (Sensirion) or MSP399124 (TSI)), Ambient Valve (C, MSP160290), Shaped Tube (D, PN 160223) and Patient Connection Tube (E, PN 160295). 2. Not shown; a metal bracket (PN 160741) attached with 2x screws (PN 420636) secures the Qvent Flow Sensor to the Ambient Valve. Figure 10-64. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 1 3. Insert the complete assembly (I) into the Bottom Foam Section (J). Hamilton Medical | HAMILTON-C2 Service Manual 309 10 Components Removal/Assembly Figure 10-65. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 2 4. Connect the tubing (K) from the Front Panel Connector Block to the Inspiratory Valve Assembly. 5. Place the Qvent Flow Sensor cable into the Cable Holder (L) located on top of the Front Panel Connector Block. 6. The cable will route on the side of the Middle and Top Foam Sections when assembled. 7. It will then be connected to the Ventilation Unit Mainboard when assembled. 310 English | 624165/07 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly 10 8. The Inspiratory Valve FFC (M) (part of the Inspiratory Valve, no part number) is routed under a channel in the Bottom Foam Section to the side. Figure 10-66. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 3 9. The cable will route on the side of the Bottom, Middle and Top Foam Sections when assembled. 10. It will then be connected to the Ventilation Unit Mainboard when assembled. 11. The cable from the Ambient Valve Assembly (part of the Ambient Valve, no part number) is routed through a channel (N) in the Bottom Foam Section, over the Shaped Tube (O) and through another channel (P) to the side. Figure 10-67. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 4 12. The cable will route on the side of the Middle and Top Foam Sections when assembled. 13. It will then be connected to the Ventilation Unit Mainboard when assembled. Hamilton Medical | HAMILTON-C2 Service Manual 311 10 Components Removal/Assembly 10.5.11 Oxygen Sensor Block and Tubing Assembly 1. Insert the Oxygen Sensor Block (A, PN 160298) into the Bottom Foam Section (B). Figure 10-68. Oxygen Sensor Block and Tubing Assembly Step 1 2. Connect the Oxygen Sensor Cable (D, PN 160354) to the Oxygen Sensor (C, PN 396200). 3. The Cable will route on the side of the Middle and Top Foam Sections when assembled. 4. It will then be connected to the Ventilation Unit Mainboard when assembled. 5. Connect the tubing (E, PN 160475) from the side of the Oxygen Sensor Block to the Inspiratory Valve. Figure 10-69. Oxygen Sensor Block and Tubing Assembly Step 2 6. Connect the tubing (F, PN 160475) from the top of the Oxygen Sensor Block, through the T-Connector (H, PN 279865), through tubing (I, PN 160475) (routed through a channel in the Bottom Foam Section) to the top of the Ambient Valve (J). 7. The other side of the T-Connector (H) is connected to a tubing (G, PN 160475) and will be connected to the Pressure Sensor Assembly when assembled. 312 English | 624165/07 Pressure Sensor Module Assembly 10 10.5.12 Pressure Sensor Module Assembly 1. Insert the Pressure Sensor Assembly (B, MSP160300) above the Ambient Valve. 2. Connect the 2 Pflowsensor tubings (A) to the bottom of the Pressure Sensor Assembly (Tubes are color coded). 3. Connect the tubing (C) from the Y-Connector to the side of the Pressure Sensor Assembly. 4. Position the Pressure Sensor Assembly FFC (D, PN 160355) along the top of the Shaped Tubes. 5. The cable will route on the side of the Middle and Top Foam Sections when assembled. 6. The cable will then be connected to the Ventilation Unit Mainboard when assembled. Figure 10-70. Pressure Sensor Module Assembly Hamilton Medical | HAMILTON-C2 Service Manual 313 10 Components Removal/Assembly 10.5.13 Middle Foam Assembly 1. Position the Middle Foam (A, PN 160238) onto the Bottom Foam. Figure 10-71. Middle Foam Assembly Step 1 2. Assemble the Tube Flange (C, PN 160287) and screw onto the Shaped Tubes from the QO2 Flow Sensor. NOTICE Do not tighten the Tube Flange too tight. Figure 10-72. Middle Foam Assembly Step 2 314 English | 624165/07 Power Supply Assembly 10 10.5.14 Power Supply Assembly 1. Insert the Power Supply (A, MSP160600) into the rear of the of the Middle Foam Section (B). 2. Connect the Mains Power Cable (C) to the Power Supply. 3. Connect the Power Supply Cable (D, PN 160539) to the Power Supply. 4. The cables will route on the side of the Middle and Top Foam Sections when assembled. 5. The cables will then be connected to the Ventilation Unit Mainboard when assembled. NOTICE To prevent noise in operation by vibrations, look out for correct implementation of the Power Supply. Figure 10-73. Power Supply Assembly Hamilton Medical | HAMILTON-C2 Service Manual 315 10 Components Removal/Assembly 10.5.15 Cooling Fan Assembly 1. Insert the Cooling Fan (A, PN 160346) into the rear of the Middle Foam Section. NOTICE Observe the air flow direction of the Fan. Flow direction is into the ventilator. 2. Position the Fan 12V Cable to the side (B). 3. The cable will route on the side of the Top Foam Section when assembled. 4. The cable will then be connected to the Ventilation Unit Mainboard when assembled. Figure 10-74. Cooling Fan Assembly 316 English | 624165/07 Blower Module Assembly 10 10.5.16 Blower Module Assembly CAUTION Before installing the Blower Module, ensure that the P-Filter cable will not be damaged. CAUTION Update the Technical State (refer to the Modify Tab Section Modify Tab). NOTICE To prevent noise in operation by vibrations, look out for correct implementation of the Blower Module. 1. Insert the Blower Module (A, MSP160250) into the Middle Foam Section (B). Figure 10-75. Blower Module Assembly Step 1 Hamilton Medical | HAMILTON-C2 Service Manual 317 10 Components Removal/Assembly 2. Position the Blower Cable (C) (part of the Blower Module, no part number) and the Temperature Sensor FFC (D, PN 160353) to the side. Figure 10-76. Blower Module Assembly Step 2 3. The cables will route on the side of the Top Foam Section when assembled. 4. The cables will then be connected to the Ventilation Unit Mainboard when assembled. 5. Position the Filter Pressure Sensor Cable (E) into the cutout of the Middle Foam Section. Figure 10-77. Blower Module Assembly Step 3 6. The Filter Pressure Sensor Cable will route on the side of the Top Foam Section when assembled. 7. The cable will then be connected to the Ventilation Unit Mainboard when assembled. 318 English | 624165/07 Top Foam and Mainboard Assembly 10 10.5.17 Top Foam and Mainboard Assembly CAUTION Update the Technical State (refer to the Modify Tab Section 8.8.3). 1. Position the Top Foam (A, PN 160239) onto the Middle Foam (B). Figure 10-78. Top Foam and Mainboard Assembly Step 1 2. Insert the Mainboard (C, MSP160200 or MSP160382) onto the Top Foam. Figure 10-79. Top Foam and Mainboard Assembly (Front View) Step 2 Hamilton Medical | HAMILTON-C2 Service Manual 319 10 Components Removal/Assembly 3. Route the Fan 12V Cable (D) into the side of the Top Foam and plug into Connector P28 on the Mainboard. Figure 10-80. Top Foam and Mainboard Assembly (Left Side View) Step 3 4. Route the Blower Cable (E) into the side of the Top Foam and plug into Connector P23 on the Mainboard. 5. Route the Temperature Sensor FFC (F) into the side of the Top Foam and plug into Connector J21 on the Mainboard. 6. Route the Inspiratory Valve FFC (G) into the Bottom, Middle and Top Foam and plug into Connector J22 on the Mainboard. 7. Route the Oxygen Sensor Cable (H) into the front of the Middle and Top Foam and plug into Connector P20 on the Mainboard. Figure 10-81. Top Foam and Mainboard Assembly (Front View) Step 4 8. Route the Qvent Flow Sensor Cable (I) into the front of the Middle and Top Foam and plug into Connector P19 on the Mainboard. 9. Route the Pressure Sensor Board FFC (J) into the front of the Middle and Top Foam and plug into Connector J17 on the Mainboard. 320 English | 624165/07 Top Foam and Mainboard Assembly 10 10. Route the Nebulizer Valve Cable (K) into the side of the Middle and Top Foam and plug into Connector P13 on the Mainboard. Figure 10-82. Top Foam and Mainboard Assembly (Right Side View) Step 5 11. Route the Ambient Valve Cable (L) into the side of the Middle and Top Foam and plug into Connector P12 on the Mainboard. 12. Route the Expiratory Valve FFC (M) into the side of the Bottom, Middle and Top Foam and plug into Connector J11 on the Mainboard. 13. Route the O2 Valve Cable (N) into the side of the Middle and Top Foam and plug into Connector P7 on the Mainboard. 14. Route the Filter Pressure Sensor FFC (O) into the side of the Middle and Top Foam and plug into Connector J9 on the Mainboard. 15. Route the Binary Valve FFC (P) from the Pressure Sensor Assembly into the side of the Middle and Top Foam and plug into Connector J8 on the Mainboard. 16. Route the O2 Flow Sensor Cable (Q) into the side of the Middle and Top Foam and plug into Connector P10 on the Mainboard. Hamilton Medical | HAMILTON-C2 Service Manual 321 10 Components Removal/Assembly 17. Route the Battery Power Cable (R) into the rear of the Middle and Top Foam and plug into Connector P6 on the Mainboard. Figure 10-83. Top Foam and Mainboard Assembly (Rear View) Step 6 18. Route the DC Input Cable (S) into the rear of the Middle and Top Foam and plug into Connector P5 on the Mainboard. 19. Route the Battery Data FFC (T) into the rear of the Middle and Top Foam and plug into Connector J4 on the Mainboard. 20. Route the Fan Supply Cable (U) into the rear of the Middle and Top Foam and plug into Connector P25 on the Mainboard. 21. Route the Power Supply Cable (V) into the rear of the Middle and Top Foam and plug into Connector P3 on the Mainboard. 22. Assemble the Front and Rear Covers (refer to the 10.4.5). 23. Assemble the Interaction Panel (refer to the Interaction Panel Removal/Assembly Section 10.3.3). 24. Assemble the Ventilation Unit Top Cover (refer to the 10.3.1). 322 English | 624165/07 11 Maintenance Tools and Test Equipment 11.1 Overview..............................................................................................................................................324 11.2 Standard Tools .....................................................................................................................................324 11.3 Special Tools ........................................................................................................................................325 11.4 Test Equipment ....................................................................................................................................328 11.5 Calibration of Test Equipment ..............................................................................................................332 323 11 Maintenance Tools and Test Equipment 11.1 Overview Standard Tools, Special Tools, ElectroStatic Discharge (ESD) protection and Test Equipment detailed in the following sections are required to carry out: • The Preventive Maintenance procedures in the Engineer Preventive Maintenance Section 6.3 • The Tests Functions in the Service Software Section 8 • The Component Replacements in the Components Removal/Assembly Section 10 11.2 Standard Tools To perform basic maintenance on equipment from Hamilton Medical AG, you require a range of: • Screwdrivers (Torx, flat and cross-head) • Metric Spanners (wrenches) • Metric Hex (Allen) Keys (wrenches) 324 English | 624165/07 Special Tools 11 11.3 Special Tools 11.3.1 Digital Voltmeter To perform the tests (Internal Resistance Checks, refer to Internal Resistance Checks Section 7.2.4) a Digital Voltmeter (DVM) is required for measuring voltage (to a tolerance of 0.5%) or resistance (to a tolerance of 1 (less than 1 Ohm). Figure 11-1. Digital Voltmeter 11.3.2 Electrical Safety Tester Electrical Safety Testing is required according to IEC-62353. A RIGEL 288, or similar equipment is required. Figure 11-2. Rigel 288 Safety Analyzer Hamilton Medical | HAMILTON-C2 Service Manual 325 11 Maintenance Tools and Test Equipment 11.3.3 ESD (ElectroStatic Discharge) Protection ESD (Electrostatic Discharge) equipment must be used to prevent damage to sensitive electronic circuits. Typically, this comprises: • ESD grounding (earthing) cable and wrist band connected to the ventilator for use when working inside the ventilator • 11-3 shows a wrist band and connecting cable complete with the crocodile clip that must be attached to the work surface Figure 11-3. ESD Wrist Strap and Cable (PN 500009) Figure 11-4. ESD Service Kit (PN 500077) 326 English | 624165/07 Two Pin Cylindrical Service Tool 11 11.3.4 Two Pin Cylindrical Service Tool The Two Pin Cylindrical Service Tool (PN 500314) is required for the removal of the oxygen tube adapter ring before you can remove the middle foam section. For details, see the Middle Foam Removal/Assembly Section 10.4.12. Figure 11-5. Two Pin Cylindrical Service Tool (PN 500314) Hamilton Medical | HAMILTON-C2 Service Manual 327 11 Maintenance Tools and Test Equipment 11.4 Test Equipment The following test equipment is required, as well as the tools listed above, to complete the test's and adjustments included in the Service Software Section 8. Pictures and details of most of these items are in the product catalog. Table 11-1. Test Equipment Part Number 500058 Description Photo Pressure Measurement Kit • A complete WIKA gauge set can be obtained from Hamilton Medical AG Pressure gauge with the following specifications: • Range: 0 – 400 mbar accuracy: 0.5% 500084 TSI Flow Meter Kit Contains: • 500308 TSI Flow meter • 500085 TSI-Flow meter Battery box • 500086 TSI-Flow meter Soft carrying case • 279204 Bacteria filter 260100 Silicone tube 30cm 22F 500330 Hand Pump - Digital Voltmeter (With red and black crocodile clips) 260206 Coaxial Adult Breathing Circuit 260241 Neonatal Breathing Circuit 328 English | 624165/07 Test Equipment Part Number Description 281637 Flow Sensor Adult/Pediatric (single use) 155500 Flow sensor Neonatal (single use) 279962 Adapter for Flow Sensor Calibration Neonatal 279928 Adapter for Flow Sensor Calibration Adult/Pediatric 281717 Stopper (used for test configurations) 500300 Pressure Connector 279812 Clamp, Flow Restrictor 500077 ESD Wrist Band Hamilton Medical | HAMILTON-C2 Service Manual 11 Photo 329 11 Maintenance Tools and Test Equipment Part Number Description 151815 Adult Demo Lung with 7mm ET Tube 281420 Connector 22M/15F-22M/15F 279913 Connector 15M/4M 7249057 2 Meter Silicone Tube 4mm ID, 7mm OD 279963 Patient Filter HMEF, Adult/Pediatric 161618 Electrical Safety Test Cable for USB Port 159171 Electrical Safety Test Cable CO2 Option - Safety Analyzer (Electric Safety Tester) 330 Photo English | 624165/07 Test Equipment Part Number Description - Safety Analyzer Hand Held Device (Electric Safety Tester) 396207 USB Memory Stick 11 Photo Specification for the USB stick: • File System: FAT or FAT32 • Unpartitioned Memory • No Operating System or Security Software Installed 369104 External Battery Charger 159172 CO2 Gas Regulator For CO2 Sensor Accuracy Check (if CO2 monitor option is installed) 159173 CO2 Calibration Gas For CO2 Sensor Accuracy Check (if CO2 monitor option is installed). 281803 Adapter (three pieces required for CO2 Accuracy Check) Hamilton Medical | HAMILTON-C2 Service Manual 331 11 Maintenance Tools and Test Equipment 11.5 Calibration of Test Equipment Some test equipment must be tested and calibrated periodically. Hamilton Medical recommends the following schedule: Table 11-2. Calibration of Test Equipment Item Schedule Action Pressure Gauge As recommended by the manufacturer, or at least once per year. Send the pressure gauge back to the manufacturer for testing. (for example, www.wika.com for calibration information). Digital Voltmeter (DVM) As recommended by the manufacturer. As recommended by manufacturer. Flow Analyzer As recommended by the manufacturer. As recommended by manufacturer. 332 English | 624165/07 12 Spare Parts 12.1 Introduction to Major Components ......................................................................................................334 12.2 Trolley PN 160170................................................................................................................................335 12.3 Trolley PN 160150................................................................................................................................336 12.4 Major Components of the Interaction Panel .........................................................................................337 12.5 Major Components of the Ventilation Unit ...........................................................................................338 12.6 Interaction Panel Parts..........................................................................................................................343 12.7 Ventilation Unit Parts ...........................................................................................................................348 12.8 Detailed Parts Descriptions ...................................................................................................................355 333 12 Spare Parts 12.1 Introduction to Major Components This section lists replacement parts that are available for the HAMILTON-C2. The first section contains diagrams that enable you to locate major components. Later sections offer much more detailed listings. NOTICE − For information about consumables (such as tubing) and complete assemblies (such as the Patient Tubing Support Arm), see the Hamilton Medical Product Catalog (PN 689060). It can be located on the Hamilton Medical AG Partner-net. − Some photographs shown in later sections may not be displayed to scale. − Some Spare Parts have a prefix MSP (Medical Spare Part). If you exchange a MSP Part, you usually have to update the Technical State of the device with the new Serial Number and Revision Number. Figure 12-1. HAMILTON-C2 A Ventilation Unit (PN 160001) See the Ventilation Unit Parts Section 12.7 334 B Interaction Panel (PN 160325) See the Interaction Panel Parts Section 12.6 English | 624165/07 Trolley PN 160170 12 12.2 Trolley PN 160170 Figure 12-2. Trolley PN 160170 A Hose Clamp (PN 160982) C Clip Tube Holder (PN 160156) B Support Arm HAMILTON-C2 (PN 160153) D HAMILTON-C2 Trolley Wheel (MSP282637) Hamilton Medical | HAMILTON-C2 Service Manual 335 12 Spare Parts 12.3 Trolley PN 160150 Figure 12-3. Trolley PN 160150 A Head Blue (MSP160606) D Support Arm HAMILTON-C2 (PN 160153) B Clip Tube Holder (PN 160156) E Head White (MSP160607) C Tube Clamp (PN 160982) F HAMILTON-C2 Trolley Handle Complete (MSP160621) 336 English | 624165/07 Major Components of the Interaction Panel 12 12.4 Major Components of the Interaction Panel Figure 12-4. Major Components of the Interaction Panel LCD Display Kit (MSP380046) D Backlight Inverter Board (Obsolete) See the Backlight Driver Board Section 12.6.2 A LCD Display (PN 380046) E See the LCD Display and Touch Screen Section 12.6.4 B Display Front (MSP160362) See the Interaction Panel Covers Section 12.6.1 F See the LCD Display and Touch Screen Section 12.6.4 C P&T Control Knob Complete (MSP160328) See the P&T Control Knob Section 12.6.5 Hamilton Medical | HAMILTON-C2 Service Manual Interaction Panel Small Rear Cover (PN 160402) Interaction Panel Large Rear Cover (PN 160326) See the Interaction Panel Covers Section 12.6.1 G Front Panel Board (MSP160196) See the Front Panel Board Section 12.6.3 337 12 Spare Parts 12.5 Major Components of the Ventilation Unit Figure 12-5. Major Components of the Ventilation Unit A Ventilation Unit Covers C See the Ventilation Unit Covers Section 12.5.1 B Ventilation Unit Top Section See the Ventilation Unit Top Section 12.5.2 338 Ventilation Unit Middle Section See the Ventilation Unit Middle Section 12.5.3 D Ventilation Unit Bottom Section See the Ventilation Unit Bottom Section 12.5.4 English | 624165/07 Ventilation Unit Covers 12 12.5.1 Ventilation Unit Covers Figure 12-6. Ventilation Unit Covers A Ventilation Unit Filter Cover (PN 160343) D See the External Covers and External Hardware Section External Covers and External Hardware B Complete Rear Cover (MSP160481) See the External Covers and External Hardware Section External Covers and External Hardware E See the External Covers and External Hardware Section External Covers and External Hardware C Battery Cover (PN 160506) See the External Covers and External Hardware Section External Covers and External Hardware Hamilton Medical | HAMILTON-C2 Service Manual Ventilator Unit Top Cover (PN 160601) Front Cover (MSP160480) See the External Covers and External Hardware Section External Covers and External Hardware F Ventilation Unit Base See the Metal Brackets and Frame Components Section Metal Brackets and Frame Components 339 12 Spare Parts 12.5.2 Ventilation Unit Top Section Figure 12-7. Ventilation Unit Top Section A Mainboard-C2 (MSP160382) See the Electronic Printed Circuit Boards Section 12.8.11 B C Top Foam (PN 160239) See the Metal Brackets and Frame Components Section 12.8.9 Embedded System Module (ESM) (MSP160206) See the Electronic Printed Circuit Boards Section 12.8.11 340 English | 624165/07 Ventilation Unit Middle Section 12 12.5.3 Ventilation Unit Middle Section Figure 12-8. Ventilation Unit Middle Section A Fan 12 V (PN 160346) D See the Electrical/Electronic Devices Section 12.8.12 B Blower Module (MSP160250) See the Pneumatic Parts and Assemblies Section 12.8.4 C Hepa Filter Assembly (PN 160216) See the Pneumatic Parts and Assemblies Section 12.8.4 E Service Power Supply (MSP160600) See the Electronic Printed Circuit Boards Section 12.8.11 Middle Foam (PN 160238) See the Metal Brackets and Frame Components Section 12.8.9 Hamilton Medical | HAMILTON-C2 Service Manual 341 12 Spare Parts 12.5.4 Ventilation Unit Bottom Section Figure 12-9. Ventilation Unit Bottom Section A Battery Compartment (PN 160302) G See the Backup Battery Pack Compartment Section 12.7.5 B Inspiratory Valve Complete (MSP160230) See the Electronic Printed Circuit Boards Section 12.8.11 H See the Assembled Components Section 12.8.6 C Qvent Flow Sensor (MSP161657, Sensirion) or (MSP399124, TSI) Pressure Sensor Assembly (MSP160300) Mixer Block Assembly (MSP160608 or MSP160226) See the Mixer Block Assembly Section 12.7.7 I QO2 Flow Sensor See the Pneumatic Parts and Assemblies Section 12.8.4 See the Flow Restrictors and Flow Sensors Section 12.8.5 D Oxygen Sensor HAMILTON-C2 (coded) (PN 396200) J See the Electrical/Electronic Devices Section 12.8.12 E Rinse Flow Assembly (MSP160498) See the Assembled Components Section 12.8.6 F Expiratory Valve Assembly (MSP160240) See the Expiratory Valve Assembly Section 12.7.6 K Ambient Valve Complete (MSP160290) See the Assembled Components Section 12.8.6 Bottom Foam (PN 160237) See the Metal Brackets and Frame Components Section 12.8.9 342 English | 624165/07 Interaction Panel Parts 12 12.6 Interaction Panel Parts 12.6.1 Interaction Panel Covers Complete Interaction Panel MSP160325 Figure 12-10. Interaction Panel Covers A Interaction Panel Large Rear Cover (PN 160326) D See the External Covers and External Hardware Section 12.8.13 B 5x Screws (PN 420642) See the Screws Section 12.8.14 C Interaction Panel Small Rear Cover (PN 160402) See the External Covers and External Hardware Section 12.8.13 E 4x Screws (PN 420642) See the Screws Section 12.8.14 Alarm Lamp (PN 160329) See the External Covers and External Hardware Section 12.8.13 Hamilton Medical | HAMILTON-C2 Service Manual 343 12 Spare Parts 12.6.2 Backlight Driver Board Figure 12-11. Backlight Driver Board A DC/AC Inverter Board (PN 396197 (Obsolete) or PN 396229 (Obsolete) D 2x Screws (PN 420663) See the Screws Section 12.8.14 Backlight LED Driver Board (PN 396314, no longer available) See the Electronic Printed Circuit Boards Section 12.8.11 B C 2x Screws (PN 420692) E Cable (PN 160358) to the DC/AC Inverter Board See the Screws Section 12.8.14 See the Electrical/Electronic Cables Section 12.8.10 2x Standoffs (PN 257076) Bypass Cable (PN 161569) for Front Panel Board Revision 08 or higher for the LED Backlights See the Standoffs Section 12.8.15 See the Electrical/Electronic Cables Section 12.8.10 NOTICE Please check compatibility for HAMILTON-C2 Serial Number <1676, see Interaction Panel Section 15.2.1 344 English | 624165/07 Front Panel Board 12 12.6.3 Front Panel Board Figure 12-12. Front Panel Board A Front Panel Board (MSP160196) E See the Electronic Printed Circuit Boards Section 12.8.11 B 5x Screws (PN 420724) FFC Cable to Key Panel (PN 160356) See the Electrical/Electronic Cables Section 12.8.10 F Speaker Mount (PN 160529) G Loudspeaker (PN 161532) See the Screws Section 12.8.14 C 1x Screw (PN 420641) See the Screws Section 12.8.14 D See the Electrical/Electronics Devices Section 12.8.12 Bypass Cable (PN 161569) to LED Backlights See the Electrical/Electronic Cables Section 12.8.10 NOTICE Please see Interaction Panel Section 15.2.1 for Display compatibility. Hamilton Medical | HAMILTON-C2 Service Manual 345 12 Spare Parts 12.6.4 LCD Display and Touch Screen Figure 12-13. LCD Display and Touch Screen A Display Front (MSP160362) I See the Electronic Printed Circuit Boards Section 12.8.11 B Display Mounting Bracket (PN 160404) See the Metal Brackets and Frame Components Section 12.8.9 J See the Metal Brackets and Frame Components Section 12.8.9 C 4x Screws (PN 420724) Cable Holder (PN 160488) K Graphic LCD TFT VGA 10.4” NEC (PN 380027 or 380031 (Obsolete)) USB Cover (PN 341548) See the Clamps and Fasteners Section 12.8.2 L See the Clamps and Fasteners Section 12.8.2 E FDC Cable to Display (PN 160357) See the Electrical/Electronic Cables Section 12.8.10 See the Screws Section 12.8.14 D Display Gasket (PN 160341) Symbol Insert See the Electronic Printed Circuit Boards Section 12.8.11 M 2x Ferrite Sleeves (PN 393113) N 2x Cable Clips (PN 393114) Graphic LCD TFT VGA 10.4" NLT (PN 380046, new) See the Electronic Printed Circuit Boards Section 12.8.11 F 3x Screws (PN 420641) see the Screws Section 12.8.14 G 2x Washers (PN 409105) See the Washers Section 12.8.16 H 1x Standoff (PN 257038) ESD Isolation Sticker on Monitor (PN 160467) See the Standoffs Section 12.8.15 NOTICE Refer to KB ID 3315 if you have to exchange an old version of LCD Display (PN 380027 or PN380031), or an old version of DC/AC Inverter Board (PN 396197 or PN 396229) Please see Interaction Panel Section 15.2.1 for Display compatibility. 346 English | 624165/07 P&T Control Knob Encoder 12 12.6.5 P&T Control Knob Encoder Figure 12-14. P&amp;T Control Knob Encoder A P&T Control Knob complete (MSP160328) See the Colored Metal Rings and Controls Section 12.8.7 B C Washer (PN 409908) See the Washers Section 12.8.16 P&T Control Knob Encoder (PN 372036) See the Electrical/Electronic Devices Section 12.8.12 Hamilton Medical | HAMILTON-C2 Service Manual 347 12 Spare Parts 12.7 Ventilation Unit Parts 12.7.1 Top Cover Figure 12-15. Ventilation Unit Top Cover A Ventilator Unit Top Cover (PN 160601) See the External Covers and External Hardware Section 12.8.13 B C 2x Screws (PN 420682) See the Screws Section 12.8.14 4x Screws (PN 420642) See the Screws Section 12.8.14 348 English | 624165/07 Front Cover 12 12.7.2 Front Cover Figure 12-16. Ventilation Unit Front Cover A Ventilation Unit Front Cover with Label (MSP160480) G See the External Covers and External Hardware Section 12.8.13 B 1x Screw (PN 420667) See the Metal Brackets and Frame Components Section 12.8.9 H See the Screws Section 12.8.14 C Interaction Panel Hinge Type A Left (PN 281781) 2x Screws (PN 420727) I 2x Screws (PN 420657) J 2x Screw Nuts (PN 420159) See the Screws Section 12.8.17 Hamilton Medical | HAMILTON-C2 Service Manual 2x Screw Nuts (PN 421059) See the Nuts Section 12.8.17 K See the Screws Section 12.8.14 F 2x Screws (PN 420657) See the Screws Section 12.8.14 See the Screws Section 12.8.14 E 2x Screws (PN 420727) See the Screws Section 12.8.14 See the Metal Brackets and Frame Components Section 12.8.9 D Interaction Panel Hinge Type B Right (PN 281782) Hamilton Medical Label (PN 160429) See the Stickers and Labels Section 12.8.3 L Oxygen Sensor Cover (PN 160401) See the External Covers and External Hardware Section 12.8.13 349 12 Spare Parts 12.7.3 Rear Cover Figure 12-17. Ventilation Unit Rear Cover Complete Rear Cover MSP160481 A P-Clamp / Screw (PN 361061 / PN 420724 ) G See the Clamps and Fasteners Section 12.8.2 B Stickers (PN 160020) See the Clamps and Fasteners Section 12.8.2 H See the Stickers and Labels Section 12.8.3 C Stickers (PN 160020) Blank Cover for Communication Board (PN 160431) I 2x Screws (PN 420730) J Sticker (PN 255358) See the Stickers and Labels Section 12.8.3 350 4x Screws (PN 420667) See the Screws Section 12.8.14 K See the Screws Section 12.8.14 F Ventilator Unit Rear Cover complete (MSP160481) See the External Covers and External Hardware Section 12.8.13 See the External Covers and External Hardware Section 12.8.13 E RS 232 Cover (PN 341326) See the Clamps and Fasteners Section 12.8.2 See the Stickers and Labels Section 12.8.3 D Ethernet Cover (PN 160509) Cable Support (PN 160408) See the External Covers and External Hardware Section 12.8.13 L 3x Screws (PN 420684) See the Screws Section 12.8.14 English | 624165/07 Ventilation Unit Base 12 12.7.4 Ventilation Unit Base Figure 12-18. Ventilation Unit Base A Base Plate (MSP160228) D See the Metal Brackets and Frame Components Section 12.8.9 B 4x Screws (PN 420657) See the Metal Brackets and Frame Components Section 12.8.9 E See the Screws Section 12.8.14 C Side Braces (Left PN 160426 / Right 160427) See the Metal Brackets and Frame Components Section 12.8.9 Hamilton Medical | HAMILTON-C2 Service Manual Base Frame (PN 160483) Release Handle (PN 160403) See the Metal Brackets and Frame Components Section 12.8.9 F 3x Screws (PN 420724) See the Screws Section 12.8.14 351 12 Spare Parts 12.7.5 Backup Battery Pack Compartment Figure 12-19. Backup Battery Pack Compartment A Battery Compartment (PN 160302) See the Electrical/Electronic Devices Section 12.8.12 B Battery LI-ION (MSP369106) See the Electrical/Electronic Devices Section 12.8.12 NOTICE 160302/01 since SN 1500 Revision 01 has been implemented with ECO M749. SN 1500 and higher are equipped with an air slot to get the hot air out of the casing. 352 English | 624165/07 Expiratory Valve Assembly 12 12.7.6 Expiratory Valve Assembly Figure 12-20. Expiratory Valve Assembly A Expiratory Valve Complete (MSP160240) See the Assembled Components Section 12.8.6 B C Expiratory Valve Silicone Membrane (PN 160500) See the Rubber Seals and Grommets Section 12.8.8 Expiratory Valve Seal with Membrane (PN 160245) See the Assembled Components Section 12.8.6 Hamilton Medical | HAMILTON-C2 Service Manual 353 12 Spare Parts 12.7.7 Mixer Block Assembly Figure 12-21. Mixer Block Assembly A O2 Mixer Assembly Complete G (MSP160226 with TSI QO2 Flow Sensor) O2 - NIST Connector (PN 160471) See the Tubings and Fittings Section 12.8.1 (MSP160608 with Sensirion QO2 Flow Sensor) See the Pneumatic Parts and Assemblies Section 12.8.4 B 3x Screws (PN 420657) H See the Screws Section 12.8.14 C Mixer Block Cover Plate (PN 160459) 2x Screws (PN 420734) See the Screws Section 12.8.14 I Nebulizer Valve (PN 160400) See the Pneumatic Parts and Assemblies Section 12.8.4 D 2x Screws (PN 420664) 12.8.14 J 2 Amp Circuit Breaker (PN 378009) See the Electrical/Electronic Devices Section 12.8.12 E O2 - DISS Connector (PN 160470) See the Tubings and Fittings Section 12.8.1 F 2x Screws (PN 420734) See the Screws Section 12.8.14 Polyurethane Tubing O2 blue 4x6 (PN 160474) (see Tubing and Fittings Section 12.8.1). 354 English | 624165/07 Detailed Parts Descriptions 12 12.8 Detailed Parts Descriptions 12.8.1 Tubing and Fittings Table 12-1. Tubing and Fittings Part Number Description 160223 FORMED TUBE PATIENT CIRCUIT 160285 FORMED TUBE O2 HD 160288 CONNECTING ADAPTER 160248 CONNECTING ADAPTER 160287 TUBE RING NUT FITTING 160295 PATIENT CONNECTION Hamilton Medical | HAMILTON-C2 Service Manual Photo 355 12 Spare Parts Part Number 160298 Description Photo OXYGEN SENSOR MOUNTING BLOCK TUBING SET (PN 160475, if the currently installed Oxygen Sensor Mounting Block is made of aluminum) (Since > SN 2341 the OXYGEN SENSOR MOUNTING BLOCK is made out of plastic) 160470 O2 DISS CONNECTOR 160471 O2 NIST CONNECTOR 160474 POLYURETHANE TUBING (O2 BLUE 4x6) 160475 TUBING SET (O2) MSP160498 TUBING SET (Nebulizer Rinse Flow) 356 English | 624165/07 Clamps and Fasteners 12 12.8.2 Clamps and Fasteners Table 12-2. Clamps and Fasteners Part Number Description 361000 TIE WRAP L=100 B=2.5MM 361016 CABLE HOLDER 20x20x5MM 361061 P-CLAMP D=5.0 PA 6 341458 USB COVER 341326 RS 232 COVER 160509 ETHERNET COVER 160488 CABLE HOLDER Hamilton Medical | HAMILTON-C2 Service Manual Photo 357 12 Spare Parts 12.8.3 Stickers and Labels Table 12-3. Stickers and Labels Part Number Description 160020 STICKERS 160429 HAMILTON MEDICAL LABEL 255358 STICKER 161998 MR UNSAFE STICKER 161919 RoHS STICKER for Asia (white/green) 159696 RoHS STICKER for Asia (white/orange) 358 Photo English | 624165/07 Pneumatic Parts and Assemblies 12 12.8.4 Pneumatic Parts and Assemblies Table 12-4. Pneumatic Parts and Assemblies Part Number Description 160216 HEPA FILTER ASSEMBLY MSP160226 O2 MIXER ASSEMBLY with TSI FLOW SENSOR (complete) Photo • Use Cable (PN 160387) from Oxygen Flow Sensor to Mainboard MSP160382 • Use Cable (PN 160373) from Oxygen Flow Sensor to Mainboard MSP160200 Includes: • Mixer Assembly Complete • TSI O2 Flow Sensor with Cable • Nebulizer Valve with Cable • Required Screws • Cable Tie Does not include: • 2A Circuit Breaker • O2 Connector • DC Cable NOTICE! Only compatible when the TSI Qvent Flow Sensor MSP399123 is installed. MSP160608 O2 MIXER ASSEMBLY with SENSIRION FlOW SENSOR (complete) NOTICE! Installed in the C2 from SN 9216 • Use Cable (PN 160387) from Oxygen Flow Sensor to Mainboard MSP160382 • Use Cable (PN 160373) from Oxygen Flow Sensor to Mainboard MSP160200 Includes: • Mixer Assembly Complete • Sensirion O2 Flow Sensor with Cable • Nebulizer Valve with Cable • Required Screws • Cable Tie Does not include: • 2A Circuit Breaker • O2 Connector • DC Cable NOTICE! Only compatible when the Sensirion Qvent Flow Sensor MSP161657 is installed. Hamilton Medical | HAMILTON-C2 Service Manual 359 12 Spare Parts Part Number Description MSP160250 BLOWER MODULE 160497 HPO INLET SERVICE KIT 160400 NEBULIZER VALVE MSP160262 SOUND INSULATION MAT 360 Photo English | 624165/07 Flow Sensors 12 12.8.5 Flow Sensors CAUTION Always use Flow Sensors from the same brand in one ventilator (Sensirion or TSI). Table 12-5. Flow Sensors Part Number MSP161657 Description Photo SENSIRION FLOW SENSOR AIR (Qvent) Does not include: • Cable (PN 160387) from Qvent Flow Sensor to Mainboard (MSP160382) • Cable (PN 160373) from Qvent Flow Sensor to Mainboard (MSP160200) • Tie Wrap (PN 361000) • Cable Holder (PN 361016) MSP399123 TSI FLOW SENSOR AIR (Qvent) Does not include: • Cable (PN 160387) from Qvent Flow Sensor to Mainboard (MSP160382) • Cable (PN 160373) from Qvent Flow Sensor to Mainboard (MSP160200) • Tie Wrap (PN 361000) • Cable Holder (PN 361016) Hamilton Medical | HAMILTON-C2 Service Manual 361 12 Spare Parts 12.8.6 Assembled Components Table 12-6. Assembled Components Part Number Description MSP160230 INSPIRATORY VALVE COMPLETE MSP160240 EXPIRATORY VALVE COMPLETE 160245 EXPIRATORY VALVE SEAL WITH MEMBRANE MSP160290 AMBIENT VALVE COMPLETE MSP160498 RINSE FLOW ASSEMBLY 362 Photo English | 624165/07 Colored Metal Rings and Controls 12 12.8.7 Colored Metal Rings and Controls Table 12-7. Colored Metal Rings and Controls Part Number MSP160328 Description Photo P&T CONTROL KNOB COMPLETE CONTAINS THE P&T KNOB PN 159449 12.8.8 Rubber Seals and Grommets Table 12-8. Rubber Seals and Grommets Part Number 160500 Description Photo EXPIRATORY VALVE SILICONE MEMBRANE 5 PIECES Hamilton Medical | HAMILTON-C2 Service Manual 363 12 Spare Parts 12.8.9 Metal Brackets and Frame Components Table 12-9. Metal Brackets and Frame Components Part Number Description MSP160228 BASE PLATE 160237 BOTTOM FOAM 160238 MIDDLE FOAM 160239 TOP FOAM 160341 DISPLAY GASKET 160403 RELEASE HANDLE 160404 DISPLAY MOUNTING BRACKET 364 Photo English | 624165/07 Metal Brackets and Frame Components Part Number Description 160426 LEFT SIDE BRACE 160427 RIGHT SIDE BRACE 160483 BASE FRAME 281781 INTERACTION PANEL HINGE TYPE A 281782 INTERACTION PANEL HINGE TYPE B Hamilton Medical | HAMILTON-C2 Service Manual 12 Photo 365 12 Spare Parts 12.8.10 Electrical/Electronic Cables Table 12-10. Electrical/Electronic Cables Part Number Description 160347 FFC CABLE TO FILTER PRESSURE BOARD 160348 CABLE TO AC INPUT AND CIRCUIT BREAKER 160349 CABLE AC INPUT TO CIRCUIT BREAKER 160370 CABLE TO BATTERY POWER 160351 FFC CABLE TO BATTERY DATA 160539 CABLE TO POWER SUPPLY 366 Photo English | 624165/07 Electrical/Electronic Cables Part Number Description 160354 CABLE TO OXYGEN SENSOR 160355 FFC CABLE TO PRESSURE SENSOR BOARD 160356 FFC CABLE TO KEY PANEL 160357 FDC CABLE TO DISPLAY 160358 CABLE TO BACKLIGHT DC/AC INVERTER BOARD (PN 396197 old or PN 396229 new) 161569 CABLE TO GRAPHIC LCD TFT VGA 10.4” NLT DISPLAY BACKLIGHT LED's 12 Photo (Bypass for the Backlight Driver Board) 160359 FFC CABLE TO BINARY VALVES Hamilton Medical | HAMILTON-C2 Service Manual 367 12 Spare Parts Part Number Description 160372 CABLE DC INPUT 160373 CABLE FROM OXYGEN FLOW SENSOR TO MAINBOARD (MSP160200) 160387 CABLE FROM OXYGEN FLOW SENSOR TO MAINBOARD (MSP160382) 355198 USA POWER CABLE 2 POL 3MT C7 G (United States) 355199 GB POWER CABLE 2 POL 3MT C7 G (Great Britain) 355200 EU POWER CABLE 2 POL 3MT C7 G (European) 355308 CHINA POWER CABLE 160187 CAR ADAPTER 368 Photo English | 624165/07 Electronic Printed Circuit Boards 12 12.8.11 Electronic Printed Circuit Boards Table 12-11. Electronic Printed Circuit Boards Part Number MSP160196 Description Photo FRONT PANEL BOARD (Revision 08 and higher with built-in Backlight Driver Components for GRAPHIC LCD TFT VGA 10.4” NLT Display Backlight LED's) For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for compatibility. MSP160200 MAINBOARD-C2 Includes: • Display Cable Holder (PN 160457) • Distance Holder 6mm (PN 362051) MSP160382 MAINBOARD-C3 (used in the Hamilton-C2) Includes: • Display Cable Holder (PN 160457) • Distance Holder 6mm (PN 362051) • 2x Cables to Flow Sensor (QO2 and Qvent) (PN 160387) MSP160206 EMBEDDED SYSTEM MODULE MSP160578 ESM SHIELDING MSP160300 PRESSURE SENSOR ASSEMBLY Hamilton Medical | HAMILTON-C2 Service Manual 369 12 Spare Parts Part Number MSP160362 Description Photo DISPLAY FRONT See the LCD Display and Touch Screen Section 12.6.4 Includes: • Touch Screen • Key Panel • Display Gasket (PN 160341) • Symbol Insert (PN 160376) Does not include: • P&T Control Knob • LCD Display MSP160600 SERVICE POWER SUPPLY 380027 GRAPHIC LCD TFT VGA 10.4” NEC NOTICE! Obsolete For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for compatibility. 380031 GRAPHIC LCD TFT VGA 10.4” NEC NOTICE! Obsolete For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for compatibility. 380046 GRAPHIC LCD TFT VGA 10.4” NLT For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for compatibility. 396197 DC/AC BACKLIGHT INVERTER BOARD 396229 NOTICE! Obsolete For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for compatibility. 396314 BACKLIGHT LED DRIVER BOARD NOTICE! No longer available If replaced, order Front Panel Board (MSP160196) For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for compatibility. 370 English | 624165/07 Electronic Printed Circuit Boards Part Number Description 160140 EXTENDED COMMUNICATION BOARD CO2, SPO2 (Not implemented in HAMILTON-C2) AND NURSE CALL 160143 EXTENDED COMMUNICATION BOARD NURSE CALL Hamilton Medical | HAMILTON-C2 Service Manual 12 Photo 371 12 Spare Parts 12.8.12 Electrical/Electronic Devices Table 12-12. Electrical/Electronic Devices Part Number 160302 Description Photo BATTERY COMPARTMENT Includes: • PN 160370 Cable to Battery Power • PN 160351 FFC Cable to Battery Data 160346 FAN 12V 161532 REPLACEMENT SPEAKER 340530 AC ELECTRICAL INLET PLUG SNAP-IN CLASS II MSP369106 BATTERY LI-ION 14.4V/6.8Ah 372036 P&T CONTROL KNOB ENCODER 378009 2 AMP CIRCUIT BREAKER 372 English | 624165/07 Electrical/Electronic Devices Part Number Description 380044 BACKLIGHT FOR LCD 396200 OXYGEN SENSOR (Hamilton Medical AG) Hamilton Medical | HAMILTON-C2 Service Manual 12 Photo 373 12 Spare Parts 12.8.13 External Covers and External Hardware Table 12-13. External Covers and External Hardware Part Number Description 160326 INTERACTION PANEL LARGE REAR COVER 160329 ALARM LAMP COVER 160343 VENTILATOR UNIT FILTER COVER 160506 BATTERY COVER 160401 OXYGEN SENSOR COVER 160402 INTERACTION PANEL SMALL REAR COVER 160408 CABLE SUPPORT 374 Photo English | 624165/07 External Covers and External Hardware Part Number Description 160431 BLANK COVER FOR COMMUNICATIONS BOARD MSP160480 VENTILATOR UNIT FRONT COVER with Label MSP160481 VENTILATOR UNIT REAR COVER COMPLETE (see 12.7.3) 160601 VENTILATOR UNIT TOP COVER 160155 TUBING HOLDER DOUBLE 160156 CLIP FOR TUBING HOLDER Hamilton Medical | HAMILTON-C2 Service Manual 12 Photo 375 12 Spare Parts 12.8.14 Screws Table 12-14. Screws Part Number Description 420623 TORX ROUNDHEAD SCREW W/I-6-ROUND M4x16 420638 TORX ROUNDHEAD SCREW I-6-R M2.5x6 A4 420641 TORX ROUNDHEAD SCREW I-6-R M3x5 A4 420642 TORX ROUNDHEAD SCREW I-6-R M3x6 A4 420643 TORX ROUNDHEAD SCREW I-6-R M3x8 A4 420654 TORX CYL-SCREW I-6-R NK M3x4 A2 376 Photo English | 624165/07 Screws Part Number Description 420657 TORX CYL-SCREW I-6-R NK M4x10 A2 420659 TORX C-SUNK SCREW I-6-R M3x6 A4 420663 TORX C-SUNK SCREW I-6-R M2x4 A4 420664 TORX C-SUNK SCREW I-6-R M3x8 420667 TORX CYL-SCREW I-6-R M4x8 A2 420669 TORX ROUNDHEAD SCREW I-6-R M4x30 A4 420671 TORX ROUNDHEAD SCREW I-6-R M2x3 A2 Hamilton Medical | HAMILTON-C2 Service Manual 12 Photo 377 12 Spare Parts Part Number Description 420682 TORX ROUNDHEAD SCREW I-6-R M4x12 A4 420684 TORX CYL-SCREW I-6-R ENK M3x6 A2 420692 TORX ROUNDHEAD SCREW I-6-R M2x4 A4 420699 TORX FILLISTER HEAD SCREW 22x5 D1=2.2, L=5, 6 IP 420710 TORX FILLISTER HEAD SCREW 25x8 D1=2.5, L=8 (Galvanized blue) 420720 TORX FILLISTER HEAD SCREW I-6R M3x10 (Galvanized blue steel) 420724 TORX FILLISTER HEAD SCREW 30x8 378 Photo English | 624165/07 Standoffs Part Number Description 420727 TORX SCREW I-6-R M4x16 420730 PHILLIPS HEAD SCREW M3x20 DIN7985 PA6.6 420734 TORX CYL-SCREW I-6-R NK M3x10 12 Photo 12.8.15 Standoffs Table 12-15. Standoffs Part Number Description 257038 STANDOFF SCREW M3x8 257076 STANDOFF M2x5 Hamilton Medical | HAMILTON-C2 Service Manual Photo 379 12 Spare Parts 12.8.16 Washers Table 12-16. Washers Part Number 409105 WASHER M3 DIN125A 409264 WASHER M 2 DIN125A 409908 WASHER M10/1.2 BN735 411001 STAR WASHER M3 DIN6798 Description Photo Description Photo 12.8.17 Screw Nuts Table 12-17. Screw Nuts Part Number 421059 380 SCREW SQUARE NUT M4 English | 624165/07 13 Schematics 13.1 Block Diagram......................................................................................................................................382 13.2 Wiring Diagram....................................................................................................................................383 381 13 Schematics 13.1 Block Diagram Table 13-1. Block diagram Diagram description HAMILTON-C2 block diagram Number/Revision BD614227/00 Figure 13-1. HAMILTON-C2 block diagram 382 English | 624165/07 Wiring Diagram 13 13.2 Wiring Diagram Table 13-2. Wiring diagram Diagram description HAMILTON-C2 wiring diagram Number/Revision WD616063/06 Figure 13-2. HAMILTON-C2 wiring diagram Hamilton Medical | HAMILTON-C2 Service Manual 383 13 384 Schematics English | 624165/07 14 Software Revisions, Features and Compatibility 14.1 Introduction .........................................................................................................................................386 14.2 Software Version Information...............................................................................................................386 385 14 Software Revisions, Features and Compatibility 14.1 Introduction NOTICE For actual information and latest software, log in to Hamilton Medical AG Partner-net. There are the latest information about upgrading and updating a HAMILTON-C2. This section gives an overview of all published HAMILTON-C2 Software. It includes information about new features introduced with software upgrades and software updates information about compatibility between software of different versions. The improvement list is not meant to be complete and covers only the most relevant changes. 14.2 Software Version Information WARNING The (*) indicates an obsolete software version. Mandatory update must be performed. Table 14-1. Software Version 2.2.5 Software Version Date Released 2.2.5 July 2018 Improvements / Corrections The following improvements were implemented: • Battery SoH added to the Instrument Report – The Battery State of Health (SoH) information is added to the instrument report in the new software • FPGA update – A new internal Field Programmable Gate Array (FPGA) clock distribution improves the reliability of the integrated circuit boards • Blower time can be updated – The blower time can be set again as it was before • Extend log files CPU temperature and POST will be logged in the new software • Wrong 'positive oxygen supply failed' alarm solved Table 14-2. Software Version 2.2.4 Software Version Date Released 2.2.4 Not Implemented Improvements / Corrections Table 14-3. Software Version 2.2.3 Software Version Date Released 2.2.3 June 2015 Improvements / Corrections The following bugs were solved: • Low Oxygen Alarm despite of correct FiO2 • Trend Pmean in DuoPap/APRV • Mismatch-IBW • Alarm criteria disconnection on Ventilator side improved The following improvements were implemented: • Firmware Version of the internal flow sensor (Sensirion only) now displayed in the Instrument Report NOTICE! For more details see Knowledge Base ID 3247. 386 English | 624165/07 Software Version Information 14 Table 14-4. Software Version 2.2.2 Software Version Date Released 2.2.2 Not Implemented Hamilton Medical | HAMILTON-C2 Service Manual Improvements / Corrections 387 14 Software Revisions, Features and Compatibility Table 14-5. Software Version 2.2.1 388 English | 624165/07 Software Version Information Software Version Date Released 2.2.1 January 2015 14 Improvements / Corrections The following bugs were solved: • False positive TF 431008 during start-up • False positive TF 431005 during ventilation • TF341001, TF383001, TF385002 (settings not accepted) • TF346044, TF346043, TF385002 and TF485001 (10s after Standby) • Increased rate of sigh in SIMV+ • Oxygen Sensor defective alarm improved • Pinsp Minimum set to 5mbar (Japan only) The following improvements were implemented: • New display languages added: Croatian, Finnish, Romanian • Once activated, CO2 monitoring remains activated until manually deactivated • Changed the method for calibrating the infant flow sensor; a calibration adapter is now used • Alarm loudness setting moved to Settings window and Alarm limit settings and ranges table changed • Default values changed for the following parameters: ETS, I:E, Rate, TI, Vt/kg • P-ramp limitation by TI added • Alarm tables have been updated • HPO/LPO selection moved to Utilities window, available in Standby • Ability to configure a minimum alarm volume (loudness) for the device • Battery low • Loss of external power • Performance limited by altitude • Battery replacement required (new) • High PEEP (new) • High oxygen • Inspiratory volume limitation (now disabled in noninvasive modes) • CO2 sensor warmup • Function Key not operational • Invalid Communication Board • Replace Oxygen Sensor • Suctioning maneuver • Technical error • Technical state failed • Touch not functional • Unknown part number • Release valve defective Replaced the numerical code with text and added troubleshooting information to the Operator’s Manual for the following technical alarms: • Battery communication error (244011) • Battery totally discharged (444001) • Blower fault (431001) • Buzzer defective (243004) • Loudspeaker defective (243003) • Options not found (249001) • Self test failed (485002) • Blower lifetime NOTICE! For more details see Knowledge Base ID 3153. Hamilton Medical | HAMILTON-C2 Service Manual 389 14 Software Revisions, Features and Compatibility Table 14-6. Software Version 2.2.0 * Software Version Date Released 2.2.0 * December 2014 Improvements / Corrections The following bugs were solved: • False positive TF 431008 during start-up • False positive TF 431005 during ventilation • TF341001, 383001, 385002 (settings not accepted) • TF346044, TF346043, TF385002 and TF485001 (10s after Standby) • Oxygen Sensor defective alarm improved The following improvements were implemented: • New display languages added: Croatian, Finnish, Romanian • Once activated, CO2 monitoring remains activated until manually deactivated • Changed the method for calibrating the infant flow sensor; a calibration adapter is now used • Alarm loudness setting moved to Settings window and Alarm limit settings and ranges table changed • Several alarms have been added or updated • Replaced the numerical code with text and added troubleshooting information to the Operator’s Manual for several technical alarms • Default values changed for the following parameters: ETS, I:E, Rate, TI, Vt/kg • P-ramp limitation by TI added • Alarm tables have been updated • HPO/LPO selection moved to Utilities window, available in Standby • Ability to configure a minimum alarm volume (loudness) for the device NOTICE! For more details see Knowledge Base ID 3153. Table 14-7. Software Version 2.1.2 * Software Version Date Released 2.1.2 * May 2013 Improvements / Corrections The following bugs were solved: The following improvements were implemented: • AUTO function on Vthigh and Vt low only active during startup • The Vt high and low alarm have to be set manually by the user during ventilation – Note: The alarm message "VT high" is replaced by "Breath aborted, Vt high". • The parameters for the loudspeaker have been optimized • The new software resolves the situation in which the ventilation and the alarm system continue to operate while the display is frozen • False positive oxygen supply missing alarms suppressed 390 English | 624165/07 Software Version Information 14 Table 14-8. Software Version 2.1.1 * Software Version Date Released 2.1.1 * June 2012 Improvements / Corrections The following bugs were solved: • TF 431016 problem solved • TF 331001 triggered instead of TF 431016, additionally, occurrence has been minimized • Oxygen Sensor calibration procedure improved (faulty Gain values should no longer occur) • Oxygen Sensor alarms modified (Oxygen Sensor calibration needed instead of Oxygen Sensor defective if the Oxygen Sensor voltage is slightly out of range) • TF 341909 problem solved (TF occurred if I:E calculator detected a problem) • Technical state: - The Power Supply (MSP160600) can now be saved (SW 2.1.0 or lower only allowed to save the Power Supply (PN 396199)) Table 14-9. Software Version 2.1.0 * Software Version Date Released 2.1.0 * November 2011 Improvements / Corrections The following bugs were solved: Ventilation mode: • The ambient and the safety mode is not displayed any more as regular TF xxx’xxx In case of a TF leading to ambient mode "Ventilation cancelled" will be displayed. • In case of a TF leading to safety mode "Safety ventilation" will be displayed • Trend time has been extended to 72 hours • Suctioning tool is now displayed in the options list • Occurrence of watchdog failures has been reduced again • Known technical faults during suctioning maneuver do not occur anymore • GUI "black holes" problem solved • RTC time will not be lost during ventilation • Technical event: 243005 (loudspeaker sounds continual) will only be checked during start-up • Technical event: 232037 (Unknown alarm) during the CO2 calibration will not appear anymore • The sequence of the component test has been improved • Service software: – Technical state now displays all hardware revisions in one window – Pressure calibration procedure is adapted to the new expiratory valve cover with the metal inlay: With the expiratory valve cover (revision 02) the pressure inside the breathing circuit can no longer be adjusted to 50 mbar (51 cmH2O), because the new covers are significantly tighter than the previous version. Therefore the pressure will be calibrated by comparing the external reference (pressure gauge) with the internal measurement displayed on screen (Pvent_control). Furthermore the saving of the gain value (adjustable with turn knob on screen) will be disabled if the difference between the Pvent_control and Pvent_monitor resp. Pressure sensor Paw value is more than 3 mbar. – Technical fault and event during performing the service software will be suppressed – Pneumatics 1 -> Exp. valve test: The test limiting values could be enlarged – The sequence of the component test has been improved Hamilton Medical | HAMILTON-C2 Service Manual 391 14 Software Revisions, Features and Compatibility Table 14-10. Software Version 2.0.7 * Software Version Date Released 2.0.7 * March 2011 Improvements / Corrections The following bugs were solved: • Watchdog Alarm Routine: With the new software version, the probability of watchdog failures is significantly decreased • Battery over temperature: A plausibility check has been introduced to avoid false positive alarms about hot batteries • DuoPAP: The alarm limit has been adapted to avoid false positive alarm "Disconnection on patient side" • "Exhalation port occluded" in NIV: The alarm has been eliminated. Consequently the LiteCircuit has been withdrawn from the product line. • CO2 curve transmission over RS232: The new unit for CO2 curve transmission over RS232 is mmHg instead of cmH2O • Language files: The following language files have been updated: Spanish, Portuguese, Chinese • Intersurgical LiteCircuit: With the release of the software version 2.0.7 for the HAMILTON-C2 we will withdraw the Intersurgical LiteCircuits from our product lines for the HAMILTON-C2, Hamilton-C1 and Raphael. The use of the LiteCircuit bears the potential risk, that the whisper valve gets blocked and CO2 is not effectively removed. The affected part numbers are: – PN 151817 LiteCircuit Standard – PN 151819 LiteCircuit 580 Table 14-11. Software Version 2.0.6 * Software Version Date Released 2.0.6 * January 2011 Improvements / Corrections The following bugs were solved: • Russian label (Day/Night) corrected The following improvements were implemented: • With SW version 2.0.6 the HAMILTON-C2 is able to communicate with the Capsuletec PDMS system Table 14-12. Software Version 2.0.5 * Software Version Date Released 2.0.5 * December 2010 Improvements / Corrections The following bugs were solved: • The software version 2.0.5 increases the oscillation stability of the HAMILTONC2. This issue becomes visible only under certain circumstances, i.e. cold temperatures, beginning of ventilation, coaxial breathing circuit without inspirations filter and our test lung. The effect can be observed especially while using DuoPAP or APRV. Table 14-13. Software Version 2.0.4 * Software Version Date Released 2.0.4 * December 2010 Improvements / Corrections The following bugs were solved: • The software version 2.0.4 includes improvements of DuoPAP and APRV regarding trigger sensitivity and pressure support on Phigh. • The software version 2.0.3 is still considered to be safe. However, we would advice to updated systems in the field with software version 2.0.4 for customers ventilating their patients intensively with DuoPAP or APRV. 392 English | 624165/07 Software Version Information 14 Table 14-14. Software Version 2.0.3 * Software Version Date Released 2.0.3 * November 2010 Improvements / Corrections The following bugs were solved: • Tightness test: Issue: With SW 2.0.2, in some cases the HAMILTON-C2 faces issues to complete the tightness test successfully Solution: Install SW 2.0.3. We improved the algorithm significantly and made the test more reliable. • Suctioning Tool: Issue: With SW 2.0.2, with activated sigh function in nonadaptive modes, the HAMILTON-C2 can in certain, rare cases fall into the TF 341001 during the suctioning maneuver. Solution: Install SW 2.0.3: Sighs are suppressed during suctioning. The following improvements were implemented: • Labeling: We used the chance and made improvements on the software labels. We adapted the following languages: French, Danish, Hungarian, Polish, Slovak, Serbian and Chinese. NOTICE! The update from previous software versions to SW 2.0.3 is not mandatory, but recommended. Table 14-15. Software Version 2.0.2 * Software Version Date Released 2.0.2 * October 2010 Improvements / Corrections The following bugs were solved: • In the German version SIMV+ must be replaced by DuoPAP+ • In pressure calibration window when calibrating to 50 mbar, additionally, it is stated to calibrate to 51cmH2O when cmH2O is used in the external pressure gauge instead of mbar Table 14-16. Software Version 2.0.1 * Software Version Date Released 2.0.1 * September 2010 Improvements / Corrections The following bugs were solved: • No false positive alarm "Oxygen supply failed" in the low FiO2 ranges (22% 30%) during the first breaths • Increased stability of O2 valve leak detection (TF 231008) • Support of new batteries in the technical file. The shown part number in the technical state is read out from the battery itself. • Correction of alarm rules for blower speed supervision (TF 231009) • Reduction of false positive alarm "Exhalation port occluded" during NIV with closed system • Correction of consequences of internal communication error (TF 344001) (high priority alarm instead of ambient mode) • Correction of issues regarding connectivity with Dräger and Philips monitors PDMS protocol sequences adapted to the latest and coming driver versions Hamilton Medical | HAMILTON-C2 Service Manual 393 14 Software Revisions, Features and Compatibility Table 14-17. Software Version 2.0.0 * Software Version Date Released 2.0.0 * October 2011 Improvements / Corrections The following bugs were solved: • Quick-start-up settings: Configure 3 quick-start-up settings with the corresponding mode, control parameters, alarm limits and ventilation start-up settings. This allows you to start within a second the appropriate ventilation treatment. • Suctioning tool: The suctioning tool recognizes a disconnection of the ventilator during oxygen enrichment. Ventilation is immediately stopped and resumed when a re-connection is detected. For a clean and safe suctioning procedure. • Dimmable screen: Change the intensity of the backlight in the night to let the patient rest. • Automatic alarm settings: Adjusts the alarms automatically around the actual value. • Automatic apnea backup: The automatic apnea backup helps to adjust apnea backup parameters correctly corresponding to the ideal body weight of the patient. • Improved IntelliTrig: IntelliTrig allows the HAMILTON-C2 to compensate leakages and adapt trigger sensitivity. With the new implemented algorithm, we did a further step to increase the reactivity of the system to leaks. Table 14-18. Software Version 1.1.4 * Software Version Date Released 1.1.4 * October 2011 Improvements / Corrections The following bugs were solved: • The software solves a problem met during extreme stress test in house with an external simulation system • A decrease of the blower speed after stand-by prevents the device from false positive "External power loss" alarms • TF 243005: loudspeaker sounds continually: the alarm rule has been adapted • TF 231008/TF231013: During start-up in LPO mode the false positive TF's 231008/231013 doesn't pop up anymore. • No false positive TF 485001 during the switch off procedure any more • Increased O2 mixer stability • several other, minor adjustments NOTICE! The software 1.1.4 (1.0.5 for Japan) MUST be installed to all devices in the field according to the information you received in the partner letter April 6th, 2010. Table 14-19. Software Version 1.1.3 * Software Version Date Released 1.1.3 * December 2009 Improvements / Corrections The following bugs were solved: • The technical alarm TF 232007 is displayed as “Check Flow Sensor tubing” alarm on screen during ventilation in order to give the user a better understanding of the situation. In the event and service log the technical alarm TF 232007 is displayed as Check Flow Sensor tubing alarm. • Autozero procedure: Due to internal investigation the autozero interval takes place more frequently to improve the flow measurement performance. • Event Log: All technical faults generated while performing the Service Software tests will no longer be registered in the user event log as numerical codes. 394 English | 624165/07 Software Version Information 14 Table 14-20. Software Version 1.1.2 * Software Version Date Released 1.1.2 * August 2009 Improvements / Corrections The following improvements were implemented: • Technical Fault Handling • Flow sensor calibration • Auto triggering behavior • Power Management (battery power handling) • New Languagefiles; DuoPAP+ to PSIMV+ • Language adaption based on International Standards Table 14-21. Software Version 1.1.1 * Software Version Date Released 1.1.1 * May 2009 Improvements / Corrections The following bugs were solved: • Adjusting alarm limit blower service required The following improvements were implemented: • Internal tightness test applicable • System test O2 mixer implemented • System test flow implemented Table 14-22. Software Version 1.1.0 * Software Version Date Released 1.1.0 * March 2009 Improvements / Corrections The following improvements were implemented: • DuoPAP / APRV • Trend • Loops • Asian languages • Ferrites in Interaction Panel not necessary Table 14-23. Software Version 1.0.2 * Software Version Date Released 1.0.2 * December 2008 Improvements / Corrections The following improvements were implemented: • Reducing of maximum power consumption during starting the device • Adjusting limits of the buzzer control • New settings of O2 control low alarm • Changing identification algorithm of the Oxygen Sensor • Increased the divergence of flow sensor calibration Table 14-24. Software Version 1.0.1 * Software Version Date Released 1.0.1 * October 2008 Improvements / Corrections The following improvements were implemented: • SW 1.0.1, ready for software upgrade with USB memory stick Hamilton Medical | HAMILTON-C2 Service Manual 395 14 396 Software Revisions, Features and Compatibility English | 624165/07 15 Hardware Revisions, Features and Compatibility 15.1 Introduction .........................................................................................................................................398 15.2 Component History and Improvements ................................................................................................398 397 15 Hardware Revisions, Features and Compatibility 15.1 Introduction This section brings together information found in other parts of the service manual concerning hardware components that have changed over time. The section also provides additional information including. Hamilton ventilator serial numbers and the dates of changes. 15.2 Component History and Improvements 15.2.1 Interaction Panel WARNING If the Interaction Panel Display (PN 380027 or PN 380031) or the Backlight DC/AC Inverter (PN 396197 or PN 396229) needs replacement and the Front Panel Board is Revision 01 to 07, please order the following parts: 1x Display (MSP380046) 1x Backlight LED Driver Board (PN 396314) 1x Cable (PN 161569) 2x Ferrite Plus (PN 393133) 2x Ferrite Clip (PN 393134) WARNING If the Interaction Panel Display (PN 380027 or PN 380031) or the Backlight DC/AC Inverter (PN 396197 or PN 396229) needs replacement and the Front Panel Board is Revision 08 or higher, please order the following parts: 1x Display (MSP380046) 1x Backlight Extension Cable (PN 161569) 1x Cable (PN 161569) 2x Ferrite Plus (PN 393133) 2x Ferrite Clip (PN 393134) Set the Front Panel Board Jumper P14 to pins 1 and 2. 398 English | 624165/07 Interaction Panel Display 15 15.2.1.1 Interaction Panel Display Table 15-1. Interaction Panel Display (PN 380027, PN 380031 and PN 380046) Revision History 00 PN 380027 LCD TFT VGA 10.4” NEC Graphic Display, 1st generation with Florescent Backlights (PN 380030) Image Used in HAMILTON-C2 from SN1001 to SN1675 NOTICE! Obsolete 00 PN 380031 LCD TFT VGA 10.4” NEC Graphic Display 2nd generation with Florescent Backlights (PN 380044) Used in HAMILTON-C2 from SN1676 to SN11339 NOTICE! Obsolete 00 PN 380046 LCD TFT VGA 10.4” NLT Graphic Display With LED Backlights (no replacement defined) Used in HAMILTON-C2 from SN11340 01 PN 380046 LCD TFT VGA 10.4” NLT Graphic Display With LED Backlights (no replacement defined) Hamilton Medical | HAMILTON-C2 Service Manual 399 15 Hardware Revisions, Features and Compatibility 15.2.1.2 Front Panel Board Table 15-2. Front Panel Board (MSP160196) Revision History Image 00 01 Initial release. 02 New layout, connector handling. 03 Speaker loudness. Remove Foil Cover Sticker from the Interaction Panel Cover (if still attached) after Rev > 03 installation. 04 Improvement of the TF 243003 problem and loudspeaker loudness and sound quality. 05 Improvement of loudspeaker lifetime. Speaker no longer soldered, but with connector. 06 Without central red backup alarm LED. 07 Not implemented. 08 Integrated Backlight LED Components for the Graphic LCD TFT VGA 10.4” NLT Display (PN 380046) with LED backlights. With this Front Panel Board, the external Backlight LED Driver Board (PN 396314) is no longer required. An Extension Cable (PN 161569) is required as a bypass for the Backlight Driver Board. CAUTION! Integrated Backlight LED Components on the Front Panel Board are not compatible with the Graphic LCD TFT VGA 10.4” NEC (Obsolete, PN 380027 or PN 380031). CAUTION! If the Front Panel Board is replaced and the Graphic LCD TFT VGA 10.4” NEC (Obsolete, PN 380027 or PN 380031) is being used, set the Jumper P14 to pins 2 and 3 and use the DC/ AC Backlight Inverter Board (PN 396229), if available. 400 English | 624165/07 Backlight Driver Board 15 15.2.1.3 Backlight Driver Board Table 15-3. Backlight DC/AC Inverter Board (PN 396197 and PN 396229) and Backlight LED Driver Board (PN 396314) Revision History 00 PN 396197 DC/AC Backlight Inverter Board for the Graphic LCD TFT VGA 10.4” NEC Display Fluorescent Backlight Lamps. Image NOTICE! Obsolete 00 PN 396229 DC/AC Backlight Inverter Board for the Graphic LCD TFT VGA 10.4” NEC Display Fluorescent Backlight Lamps. NOTICE! Obsolete 00 PN 396314 Backlight LED Driver Board for the Graphic LCD TFT VGA 10.4” NLT Display Backlight LED's. 01 PN 396314 Backlight LED Driver Board for the Graphic LCD TFT VGA 10.4” NLT Display Backlight LED's. NOTICE! No longer available Hamilton Medical | HAMILTON-C2 Service Manual 401 15 Hardware Revisions, Features and Compatibility 15.2.2 Ventilation Unit 15.2.2.1 Mainboard (MSP160200) Table 15-4. Mainboard (MSP160200) Revision History Image 00 01 02 03 First official release. Used in HAMILTON-C2 from SN1001 to SN1134. 04 05 06 New layout and new firmware. Oxygen Sensor calibration without external tool. Use in HAMILTON-C2 from SN1266 to SN2140. 07 New layout. Blower end stage robustness. Used in HAMILTON-C2 from SN2141 to SN2422. 08 New firmware. Used in HAMILTON-C2 from SN 2423 to SN 2784. 09 Qvent Flow Sensor precision improved. EMC immunity to 20 v/m increased. Used in HAMILTON-C2 from SN2785 to SN3050. 10 Not implemented. 11 Soft boot problem of battery 2 solved. Used in HAMILTON-C2 from SN3051 to SN3369. 12 Alarm monitor improvements. Used in HAMILTON-C2 from SN3370 to SN4455. 13 EMC improvements and buzzer adjustability. Neonatal implemented. Used in HAMILTON-C2 from SN4456. 14 Exchange of discontinued parts. Used in HAMILTON-C2 from SN5900 15 EMI improvement to conductive ESD. Switching off problem solved. Easy to enter service software by storing the keys. NOTICE The Oxygen Sensor Calibration Tool (PN 160367) is no longer required for Mainboard Revisions 06 and higher. 402 English | 624165/07 Mainboard (MSP160382) 15 15.2.2.2 Mainboard (MSP160382) NOTICE The Mainboard MSP160382 was previously used in the HAMILTON-C3 only. In the HAMILTON-C2, it is assembled in devices with SN > 10’000. It is compatible from older devices as long as software version 2.2.0 or higher is installed. Table 15-5. Mainboard (MSP160382) Revision History Image 00 01 02 03 04 05 First official release in HAMILTON-C2. Used in HAMILTON-C2 from > SN10’000. No disconnection during O2 offset calibration needed. Different connector for the oxygen and air flow sensors. Additional fan supply from power supply is no longer needed. Alarm lamp blinking according new standard of IEC 60601-1. Jumper to select the device (HAMILTON-C2 or HAMILTON-C3). 06 New PoL voltage converters for improved EMI behavior. Oxygen Sensor calibration without disconnection. RS232 monitoring interface bit rate up to 38.2 kBaud for Hamilton block protocol. 07 Not implemented. 08 Increased ESD immunity against contact discharge to fulfill the new standard for electromagnetic compatibility (EMC) IEC 60601-1-2 (2014) 4th edition. Improved polarity protection on DC Input. New gold caps with improved lifetime. 09 10 New blower end stage due to discontinued parts. CAUTION! Due to an incompatibility with older software versions, the use of a Mainboard (MSP160382) with revision 10 or higher requires the following software to be installed on the instrument: HAMILTON-C2 SW 2.2.5 or higher. NOTICE If Mainboard (MSP160382) replaces Mainboard (MSP160200), make sure to roll up the excess Fan Supply Cable. Hamilton Medical | HAMILTON-C2 Service Manual 403 15 Hardware Revisions, Features and Compatibility 15.2.2.3 ESM Module Table 15-6. ESM Module (MSP160206) Revision History Image 00 01 First official release. 02 New layout. New Firmware. SW 1.0.1. Ready for SW Upgrade with USB memory stick. 03 Not used components dropped. 04 Not used components dropped. 05 SW 1.0.5. No more loss of Technical State due to SW upgrade, not used components dropped. 06 EMC improvements. 07 Pre-installed latest software 2.x.x. 08 Layout correction. 09 Pre-installed SW 2.2.1. 10 Pre-installed SW 2.2.3. 11 Clock generators replaced due to discontinued parts. BGA components underfilled to improve mechanical robustness. 15.2.2.4 ESM Shielding Table 15-7. ESM Shielding (MSP160578) Revision History Image 00 01 First official release. 02 Improved EMC Used in HAMILTON-C2 from SN6976. 404 English | 624165/07 Flow Sensors 15 15.2.2.5 Flow Sensors CAUTION Always use Flow Sensors from the same brand in one ventilator (Sensirion or TSI). Table 15-8. TSI Qvent Flow Sensor (MSP399123) Revision History Image 00 01 02 03 Table 15-9. TSI O2 Flow Sensor (MSP399124) Revision History Image 00 01 02 03 Table 15-10. Sensirion Qvent and QO2 Flow Sensors (MSP161657) Revision History Image 00 01 Hamilton Medical | HAMILTON-C2 Service Manual 405 15 Hardware Revisions, Features and Compatibility 15.2.2.6 Oxygen Mixer Block Assembly Table 15-11. Oxygen Mixer Block Assembly (MSP160226 or MSP160608) Revision 00 History Image First official release With TSI QO2 Flow Sensor 01 Features: • Plastic Disk (A) • Without Turbulence Filter (A) • Manufactured with the small O2 connector seat (B) (see Mounting Plate Section 15.2.2.7) • Manufactured without O2 Inlet Filter (C) (see O2 Inlet Filter Section 15.2.2.8) In HAMILTON-C2 from SN1001 to SN1271. 02 Features: • Sinter disk (A) • Equipped with turbulence screen (B) • Manufactured with the small O2 connector seat (B) from SN1590 to SN2154 (see Mounting Plate Section 15.2.2.7) • Manufactured without O2 Inlet Filter (C) (see O2 Inlet Filter Section 15.2.2.8) In HAMILTON-C2 from SN1272 to SN1589. 03 Features: • The QO2 Flow Sensor (MSP161657) is part of the mixer assembly (A, MSP160226) • Manufactured with the small O2 connector seat (A1) from SN1590 to SN1625 • Manufactured with the large O2 connector seat (B) from SN1626 • Orifice (C) • Turbulence screen (D) In HAMILTON-C2 from SN1590 to SN2154. 04 Adaption of the manufacturing process. In HAMILTON-C2 from SN 2155. 05 Introduction of an Injection Molded Connection Plate. In HAMILTON-C2 from SN 5587. 406 English | 624165/07 Oxygen Mixer Block Assembly Revision 06 History 15 Image Additional Cable (PN 160387) for Mainboard (MSP160387). CAUTION! When replacing the Oxygen Mixer Block Assembly, within the same ventilator, match the Oxygen Mixer Block Assembly to have the same brand QO2 Flow Sensor as the Qvent Flow Sensor (TSI or Sensirion). Oxygen Mixer Block Assembly (MSP160226) with TSI QO2 Flow Sensor (MSP399124). Oxygen Mixer Block Assembly MSP160608) with Sensirion QO2 Flow Sensor (MSP161657). Hamilton Medical | HAMILTON-C2 Service Manual 407 15 Hardware Revisions, Features and Compatibility 15.2.2.7 Mounting Plate Table 15-12. Mounting Plate (PN 160458) Revision 00 History Image First official release. The mounting plate with the small O2 Connector seat (left image) is installed in all HAMILTON-C2 from SN1001 to SN1625. 01 408 The mounting plate with the large O2 Connector seat (right image) is installed in all HAMILTON-C2 from SN1626. English | 624165/07 High Pressure O2 Inlet Filter 15 15.2.2.8 High Pressure O2 Inlet Filter Table 15-13. High Pressure O2 Inlet Filter (PN 160491) Revision 00 History Image To prevent the O2 Proportional Valve (PN 160441) from obstruction caused by particles entering the device, the High Pressure O2 Inlet Filter (PN 160491) has been implemented. The High Pressure O2 Inlet Filter (PN 160491) is part of the DISS (PN 160470) or NIST (PN 160471) Connector. For devices which do not have a High Pressure O2 Inlet Filter installed, we strongly recommend installing the filter to avoid device malfunction. DISS Connector with Mounting Adapter NIST Connector with Mounting Adapter For this purpose, we provide the following solutions: 1. HAMILTON-C2 with Oxygen Mixer Block Assembly (MSP160226 Revision 00 and Revision 01), the following items need to be ordered for the update: High Pressure O2 Inlet Filter – Mounting Kit (PN 160496, includes Mounting Adapter (PN 160494)) – O2 DISS Connector (PN 160470) or – O2 NIST Connector (PN 160471) 2. HAMILTON-C2 with Oxygen Mixer Block Assembly (MSP160226 Revision 02), the following items need to be ordered for the update: – O2 DISS Connector (PN 160470) or – O2 NIST Connector (PN 160471) The High Pressure O2 Inlet Filter (PN 160491) is part of the yearly Preventive Maintenance and can be ordered separately or order the HPO Inlet Filter Service Kit (PN 160497). NOTICE A number of devices with Mixer Block Assembly (MSP160226) REV 02 (SN 1590 - 1625) are not equipped with the large O2 connector seat. For these devices, it is necessary to order the Mounting Kit (PN 160496) together with the DISS or NIST connector. Hamilton Medical | HAMILTON-C2 Service Manual 409 15 Hardware Revisions, Features and Compatibility 15.2.2.9 Pressure Sensor Module Table 15-14. Pressure Sensor Module (MSP160300) Revision 410 History 00 First official release. 01 Design of the Holding Plate improved. Image English | 624165/07 Blower Module Complete 15 15.2.2.10 Blower Module Complete Table 15-15. Blower Module (MSP160250) Revision History 00 First official Release. 01 Noise reduction and Blower Turbine sealed up. 02 Assembly process improved. 03 Vibration damper included ECO M900. 04 Heat sink improvement. 05 Replacement of filter pressure board introduced. 06 Improved EMI behavior. Image 15.2.2.11 Ambient Valve Complete Table 15-16. Ambient Valve Complete (MSP160290) Revision History Image 00 01 02 03 04 15.2.2.12 Inspiratory Valve Complete Table 15-17. Inspiratory Valve Complete (MSP160230) Revision History 00 First official release in HAMILTON-C2 until SN 3004. 01 Housing and flow characteristic improved in HAMILTON-C2 since SN 3005. 02 Stick-slip effect improved in HAMILTON-C2 since SN 4233. Image Neonatal implemented. 15.2.2.13 Expiratory Valve Complete Table 15-18. Expiratory Valve Complete (MSP160240) Revision History Image 00 01 Neonatal implemented. Hamilton Medical | HAMILTON-C2 Service Manual 411 15 Hardware Revisions, Features and Compatibility 15.2.2.14 Oxygen Sensor Table 15-19. Oxygen Sensor (PN 396200) Revision History Image 00 01 First official release in HAMILTON-C2 from SN 1000. 02 Temperature compensation improved July 2013. 15.2.2.15 Oxygen Sensor Cover Table 15-20. Oxygen Sensor Cover (PN 160401) Revision History 00 First official release. 01 Improved ultraviolet rays resistance. 02 Oxygen Sensor Cover improved. Image 15.2.2.16 Battery Compartment Table 15-21. Battery Compartment (PN 160302) Revision History 00 First release. 01 To prevent an increase of battery temperatures > 54°C 5, vent holes (A) are drilled on the top of the Battery Compartment. Image 15.2.2.17 Battery Door Table 15-22. Battery Door PN (160344, old) or (PN 160506, new) Revision 00 01 History Image PN 160344 Old Version NOTICE! Obsolete Old Version 00 PN 160506 New Version If the broken Battery Cover is the Old Version (PN 160344), order the complete Rear Cover (MSP160481), (see the Rear Cover Section 15.2.2.20). New Version 412 English | 624165/07 Top Cover Handle 15 15.2.2.18 Top Cover Handle Table 15-23. Top Cover Handle (PN 160342) Revision History Image 00 01 To prevent damage of the device, the cylinder bolt (A) was extended and locked against removing with a retaining ring (B) inside. The top cover can be replaced with PN 160601. This improvement is implemented from HAMILTONC2 SN1271. 15.2.2.19 Front Cover Table 15-24. Front Cover (MSP160480) Revision History 00 First official release. 01 Improved ultraviolet rays resistance. 02 Oxygen Sensor Cover improved. Image 15.2.2.20 Rear Cover Table 15-25. Rear Cover (MSP160481) Revision History 00 First official release. 01 Improved ultraviolet rays resistance. 02 Several further improvements like Battery Door. Hamilton Medical | HAMILTON-C2 Service Manual Image 413 15 414 Hardware Revisions, Features and Compatibility English | 624165/07 16 Glossary 16.1 Definitions of Expression ......................................................................................................................416 16.2 Alarm Codes ........................................................................................................................................421 415 16 Glossary 16.1 Definitions of Expression This glossary offers definitions of expressions not included, or extended definitions of expressions briefly included, in the glossary of the HAMILTON-C2 Operator’s Manual. It should be used together with the glossary in this guide. Table 16-1. Definitions of Expression Activate An action on the Touch Screen or the P&T Control Knob to choose a function or action. Air Source used in the HAMILTON-C2 Alarm Buffer An area of memory containing details of the twenty most recent alarms. The most recent six of the twenty alarms can be accessed by activating the alarm symbol at the bottom left of the screen when the HAMILTONC2 is in normal operating mode. Alarm Lamp Indicates alarm conditions • Red - High Priority Alarms and Technical Faults • Yellow - Medium and Low Priority Alarms Alarm Status Indicator An area at the bottom of the Touch Screen of the HAMILTON-C2 which displays the alarm status and power status. Ambient State A state that the HAMILTON-C2 uses when it cannot function normally because of an internal or external fault. The Ambient State is often associated with Technical Faults, but can also be caused by such things as air or oxygen supply. It is always accompanied by a high-priority patient alarm, sounded by the loudspeaker or buzzer. Ambient State Gas Flow If the HAMILTON-C2 is in the Ambient State, the patient breathes unassisted. Analog - Digital Conversion (ADC) Converts an analog signal to a digital signal. Audible Alarm An alarm sounded when there is a Technical Fault or other error condition. An audible alarm is generated. A buzzer is used as a backup if the audible alarm through the loudspeaker does not function. Autorinse See Rinse Flow Autozero A method to automatically adjust for electronic drift of a device due to temperature and environmental conditions. Backlight Lights used to help illuminate the screen. Basic Input Output System (BIOS) The basic input output system for the central processing unit (CPU). This is held on an EPROM or EEPROM mounted on the ESM. Basket An accessory located on the back of the HAMILTON-C2 trolley. Battery Charger Circuits located in the driver poard that supply power to the primary and optional battery packs when mains power is available. • Primary battery pack - 14.4 VDC maximum charging voltage • Optional battery pack - 14.4 VDC maximum charging voltage Both battery packs can also be charged using an external charger. Breathing Circuit See Patient Breathing Circuit Buzzer A piezoelectric device mounted on the Control Board, used by the HAMILTON-C2 to sound some alarms. It functions independently of the loudspeaker and the power supply, and typically indicates a high-priority technical faults that causes the HAMILTON-C2 to go into the ambient state. Despite its name, the buzzer makes a high frequency sound. NOTICE! Despite its name, the buzzer makes a high frequency sound. Communication Board A communication interface for the CO2, SpO2 and Nurse Call functions. Communication Interface The Ethernet Port on the rear of the HAMILTON-C2. Configuration Menu A screen used to select ventilation and other operation parameters of the HAMILTON-C2. Control Knob Used to select an option on the LCD screen by rotating and can be used to activate a selection on the LCD screen by pressing. 416 English | 624165/07 Definitions of Expression 16 De-activate An action on the Touch Screen or the P&T Control Knob to clear the selection of a function or action. De-select An action on the Touch Screen or the P&T Control Knob to clear the selection of a function or action. Digital - Analog Converter (DAC) Converts a digital signal to an analog signal. DISS Diameter index safety standard, a standard connector used for connection of oxygen to the HAMILTON-C2, designed so the oxygen supply cannot be connected incorrectly. dP Flow Sensor Measures the differential pressure (pressure difference) across the flow sensor. The measurement is performed by a pressure sensor inside the HAMILTON-C2, and is used to calculate airway gas flow. EEPROM Electrically Erasable Programmable Read Only Memory Electrical Safety Tests A set of electrical tests used to check the safety of a device. Electrostatic discharge (ESD) Electrostatic Discharge Emergency Buzzer Alarm An alarm that sounds when a Technical Faults occurs that is serious enough to cause the HAMILTON-C2 to switch to Ambient Mode. In Ambient Mode, all valves switch to their unpowered position, and the HAMILTON-C2 is completely passive. NOTICE! The buzzer makes a high frequency sound. It functions independently of the loudspeaker. ESM Board Embedded system module. A PCB containing the processor Event Log A record of most activity in the HAMILTON-C2. This includes user actions and internal activity such as: • Calibration Results • Alarms • Technical Faults • Controls Settings • Configuration, Serial Numbers, Revision Numbers • Switch-on Times The Event Log always holds a maximum of 1,000 events. Switching OFF and ON the HAMILTON-C2 does not delete the storage of the Event Log. You can access a subset of the event log suited for clinical use by activating the Event Log symbol in the monitoring menu in normal operating mode. Expiration The act of the patient breathing out. Expiratory Valve A valve controlling pressure in the Patient Breathing Circuit, enabling the patient to exhale and the HAMILTON-C2 to maintain PEEP. Expiratory Valve Block Connection The plastic connector used to attach the Patient Breathing Circuit on the side of the HAMILTON-C2. Fan Filter A filter used to capture dust from the Cooling Fan. Flow Restrictor A device that restricts flow of a gas. Front Panel Keys Control buttons at the front of the Front Panel. Front Panel Part of the HAMILTON-C2 containing the user interface for interaction with the ventilator; LCD Display and Hard Keys. Galvanic Oxygen Sensor See Oxygen Sensor Gas Inlet Connection for the inlet of oxygen to the HAMILTON-C2. Gold Caps One provides power for emergency buzzer alarm for 2 minutes and one for the real time clock for 3 months when no power source is provided from mains power or battery. Ground Refers to the 0 electrical potential of a device. Ground (GND) Ground HAMILTON-C2 Product name HEPA High efficiency particle air filter Hamilton Medical | HAMILTON-C2 Service Manual 417 16 Glossary Hot Swappable A term used to indicate a device can be disconnected without removing power. HPO High Pressure Oxygen Inspiration The act of the patient breathing in. Internal Temperature The HAMILTON-C2 contains devices to measure the internal operating temperatures. Light Emitting Diode (LED) Light Emitting Diode LPO Low Pressure Oxygen Loudspeaker A speaker used to indicate alarm conditions from the ventilation unit. Main Power Switch Powers the HAMILTON-C2 ON and OFF. Metron EST Tester Automated device used to perform the electrical safety test. Microphone In the HAMILTON-C2, a microphone is positioned near the loudspeaker as a feedback circuit to determine that the loudspeaker is operating properly. Nebulizer Connection Connection on the side of the HAMILTON-C2 to attach tubing for the Nebulizer output. Nebulizer Valve Valve used to control the flow of oxygen to the Nebulizer Jar. NIST Non-interchangeable screw thread, a standard connector used for connection of oxygen to the HAMILTONC2, designed so the oxygen supply cannot be connected incorrectly. Option Board See Communication Board O2 Cell For O2 Cell, O2 Sensor or Oxygen Cell, see the Oxygen Sensor. Check Valve Used to restrict gas flow to only one direction. Orifice Flow Restrictor A device with a fixed opening which restricts the flow of a gas. Orifice Tube Used in the test configuration to perform adjustments in the service software. Oxygen Sensor A small, replaceable, plastic unit used by the HAMILTON-C2 to measure oxygen concentration. The Oxygen Sensor reacts to the presence of oxygen, producing a voltage output in proportion to the oxygen concentration. The Oxygen Sensor must be replaced after a period of service, when it can no longer be calibrated. Typically, about one year. Oxygen Sensor Calibration A procedure which supplies a controlled flow of air, oxygen and air/oxygen mixture to the Oxygen Sensor for calibration. Pambient A sensor which measures the ambient pressure or the room pressure. Patient Alarms An alarm indicating that there is a problem or potential problem in ventilating the patient.There are three levels of patient alarms: high, medium and low. They are indicated by beeps on the loudspeaker, messages on the display and the alarm lamp. Patient Breathing Circuit A Patient Breathing Circuit carries the air/oxygen mixture to the patient and carries the expired (exhaust) air from the patient. It may include Tubings, Mask, Inspiratory Filter, Flow Sensor, Nebulizer Jar, Expiratory Valve Cover and Membrane. Paw A sensor which measures the pressure at the proximal side of the Patient Flow Sensor. Pfilter A sensor which measures the pressure after the HEPA Filter; used to indicate if the HEPA filter needs to be replaced. Pflowsensor A sensor which measures the differential pressure across the patient airway flow sensor. Portable Tank Oxygen tanks used for supply in a portable environment when the facility source is not available. Press and Turn Control Knob (P&T) A button on the front of the front panel used to select and activate screen functions. Press and Turn Encoder (P&T Encoder) An electrical device which provides input signals for control of screen functions. Pressure Gauge A mechanical device used to measure gas pressure. Pressure Regulator A device that regulates the pressure and restricts the flow of a gas. Pressure Sensor Assembly Provides components for pressure measurements used to monitor ventilation. 418 English | 624165/07 Definitions of Expression 16 Preventive Maintenance (PM) A term used for periodic maintenance of a device with specific planned tasks and items to check, adjust and replace. Principal Gas Flow The main gas flow through the HAMILTON-C2 from the HEPA Filter for air and the oxygen inlets to the patient, and then from the patient through the expiratory valve. Pvent_control A sensor which measures the pressure at the patient outlet, used for controlling. Pvent_monitor A sensor which measures the pressure at the patient outlet, used for monitoring. QO2 Flow Sensor Used to measure the oxygen flow into the Blower Module. Qvent Flow Sensor Used to measure the air/oxygen flow into the Patient Breathing Circuit. Rinse Flow A very small, continuous flow of gas through both the blue (patient side; proximal) and clear (ventilator side; distal) tubes to the Patient Flow Sensor. The flow minimizes the possibility of tube blockage, and hinders the potential migration of bacteria and viruses from the patient’s expired gases through the tubes, towards the pressure sensors inside the ventilator. Safety Mode An emergency state that ensures a basic minute ventilation while giving the user time for corrective actions in case of some technical fault alarms. The default inspiratory pressure is maintained, the expiratory valve opens as needed to switch system pressure levels between PEEP and inspiratory pressure, and patient sensing is nonfunctional. Select An action on the touchscreen or the P&T control knob to choose a function or action. Service Software The software used to perform tests, adjustments and calibrations of the HAMILTON-C2. Also enables viewing and exporting data from the Event Log. The HAMILTON-C2 cannot be used for patient ventilation when in service software mode. NOTICE! The HAMILTON-C2 cannot be used for patient ventilation when in Service Software Mode. Service Software Mode The operational status of the HAMILTON-C2 when in the service software. Smart Battery Pack Li-Ion battery, hot swappable which monitors and communicates the battery condition. SoH The State of Health (SoH) is a "measurement" that reflects the general condition of a battery and its ability to deliver the specified performance compared with a fresh battery. It takes into account such factors as charge acceptance, internal resistance, voltage and self-discharge. It is a measure of the long term capability of the battery and gives an "indication" not an absolute measurement, of how much of the available "lifetime energy throughput" of the battery has been consumed, and how much is remaining. Using the automotive analogy, it can be compared to the "odometer"display function which indicates the number of miles traveled since the vehicle was new. T-Fitting A type of tubing connector in the shape of a T. Technical Events A very minor fault or event recorded by the HAMILTON-C2 in the Event Log for use only by software developers. Technical Faults An alarm condition indicating a major malfunction of the HAMILTON-C2. (This contrasts with an alarm, that indicates a problem with the status of a patient.) Technical Faults are intended to alert users and engineers of the need for intervention, and are recorded in the Event Log. Test Configuration (TC) A particular configuration of devices and tubings use to perform the tests/calibrations in the service software. Touch Screen An interaction method where the LCD display screen can be touched with a finger to activate or select a function. Trolley A part of the HAMILTON-C2 on which the ventilator is mounted for transport within the customer location. Update An update is an improvement to an existing function. An update normally involves only software. A software update is generally a revision number increment in a digit after the decimal point: for example, from 3.2 to 3.3. Upgrade An upgrade is the addition of new functions to a device. There are three ways to perform an upgrade: • Add a hardware item that offers additional functions • Upgrade to a higher software revision indicated by a higher value before the decimal point: for example 01.03 to 02.00 • Upgrade to a higher type of software Hamilton Medical | HAMILTON-C2 Service Manual 419 16 Glossary Variable Orifice Membrane A variable flow restrictor as used in the Patient Flow Sensor. Y-Fitting A type of tubing connector in the shape of a Y. 420 English | 624165/07 Alarm Codes 16 16.2 Alarm Codes 16.2.1 Patient Alarms - 100000 Alarm Codes Table 16-2. Patient Alarms - 100000 Alarm Codes SID Number CSystemConfiguration ID enGB Text Value 132001 paBM_pawPressureLow Loss of PEEP 132002 paBM_pawTubing Check flow sensor tubing 132003 paBM_qawFlowSensorTubing Check flow sensor 132004 paBM_qawFlowSensorMissing Check flow sensor 132005 paBM_O2SensorDefect Oxygen sensor defective 132006 paBM_O2SensorMissing Oxygen sensor n/a 132007 paBM_O2SensorInvalid Oxygen sensor not system compatible 132008 paBM_O2SensorCalibrationNeeded Oxygen sensor calibration needed 132010 paBM_CO2SensorMissing CO2 sensor disconnected 132011 paBM_CO2SensorDefect CO2 sensor defect 132012 paBM_CO2SensorOverTemperature CO2 sensor over temperature 132013 paBM_CO2SensorWarmUp CO2 sensor warmup 132014 paBM_wrongFlowsensor Wrong flow sensor 132015 paBM_FlowSensorCalibrationNeeded Flow sensor calibration needed Breath Monitoring Breath Pattern Generation 132016 paBM_CO2CheckAirwayAdapter Check CO2 airway adapter 132017 paBM_CO2CheckSamplingLine Check CO2 sampling line 133001 paBPG_apnea n/a 133002 paBPG_volumeLimitReached Inspiratory volume limitation 133003 paBPG_pressureLimitReached High pressure 133004 paBPG_CheckPatientInterface Ventilation Mode Control 141001 paVMC_pressureLimitation Pressure limitation 141002 paVMC_VThigh Vt high 141003 paVMC_VTlow Vt low 141004 paVMC_expMinVolHigh High minute volume 141005 paVMC_expMinVolLow Low minute volume 141006 paVMC_fTotalHigh High frequency 141007 paVMC_fTotalLow Low frequency 141008 paVMC_oxygenHigh High oxygen 141009 paVMC_oxygenLow Low oxygen 141010 paVMC_apnea Apnea 141011 paVMC_apneaVentilationEnd Apnea ventilation ended 141012 paVMC_apneaVentilation Apnea ventilation 141013 paVMC_highPressureDuringSigh High pressure during sigh 141014 paVMC_turnFlowSensor Turn the flow sensor Hamilton Medical | HAMILTON-C2 Service Manual 421 16 Glossary SID Number CSystemConfiguration ID enGB Text Value 141016 paVMC_disconnectionVentilator Disconnection on ventilator side 141017 paVMC_disconnectionPatient Disconnection on patient side 141018 paVMC_exhalationObstructed Exhalation obstructed 141019 paVMC_IRV IRV 141020 paVMC_ASVunableToReachTarget ASV: Cannot meet target 141021 paVMC_ASVplimitChanged Pressure limit has changed 141023 paVMC_sensorFailMode External flow sensor failed 141024 paVMC_pressureLow Low pressure 141025 paVMC_PetCO2High PetCO2 high 141026 paVMC_PetCO2Low PetCO2 low 141027 paVMC_suctioningManoeuvre Suctioning maneuver 141028 paVMC_performanceLimitedByHighAltitude Performance limited by high altitude 141051 paVMC_PEEPHigh High PEEP 141052 paVMC_PEEPLow Low PEEP 422 English | 624165/07 Technical Alarms - 200000 Alarm Codes 16 16.2.2 Technical Alarms - 200000 Alarm Codes Table 16-3. Technical Alarms - 200000 Alarm Codes Error No. Error Description enGB Text Value 231001 taGD_pressureControllerPressureLow Technical event 231002 taGD_pressureControllerPressureHigh Technical event 231003 taGD_flowControllerFlowLow Technical event 231004 taGD_flowControllerFlowHigh Technical event 231005 taGD_inspirationValveLeak Technical event 231006 taGD_O2ControllerFlowLow Oxygen supply failed 231007 taGD_O2ControllerFlowHigh Technical event 231008 taGD_O2ValveLeak Technical event 231009 taGD_blowerControllerSpeedLow Technical event 231010 taGD_blowerControllerSpeedHigh Technical event 231011 taGD_ventOutputTemperatureHigh Vent outlet temperature high 231012 taGD_qventFlowSensorDefect Technical event 231013 taGD_qO2FlowSensorDefect Technical event 231014 taGD_ambientValveError Technical event 231015 taGD_autozeroPventControlOK n/a 231016 taGD_autozeroPventControlOutOfRange n/a 231017 taGD_blowerServiceRequired Blower service required 231018 taGD_selftestWithError Technical event 231019 taGD_iInspValveSensorDefect Technical event 231020 taGD_regulatorDataLogOn Technical event 231021 taGD_o2PresenceCheckFailed n/a 231022 taGD_pExpValveSensorDefect Technical event 231023 taGD_invalidFlowSensor Technical event 231026 taGD_expValveCoverInvalid 231028 taGD_pressureControllerFlowHigh n/a 231029 taGD_autozeroPExpValveOK n/a 231030 taGD_autozeroPExpValveError n/a 231031 taGD_expValveCoverUniversal n/a 231032 taGD_expirationValveDisconnected n/a 231033 taGD_blowerControllerSpeedLimited n/a 231040 taGD_calibratePVentControlOK n/a 231041 taGD_calibratePVentControlError n/a 231044 taGD_calibrateExpValveOK 231045 taGD_calibrateExpValveError 231046 taGD_calibrateIExpValveOK n/a 231047 taGD_calibrateIExpValveError n/a 231099 taGD_communicationCheck n/a Gas Delivery Hamilton Medical | HAMILTON-C2 Service Manual 423 16 Glossary Error No. Error Description enGB Text Value Breath Monitoring (SERVICE SOFTWARE) 232001 taBM_pressureSensorTolerance n/a 232002 taBM_pventMonitorSensorDefect Technical event 232003 taBM_pawSensorDefect Technical event 232004 taBM_pressureNotReleased Pressure not released 232005 taBM_blowerHot Technical event 232006 taBM_blowerTemperatureSensorDefect Technical event 232007 taBM_qawFlowSensorDefect Check flow sensor tubing 232008 taBM_pambientSensorDefect Technical event 232009 taBM_pventAutozeroValveError n/a 232010 taBM_pventAutozeroValveMissing n/a 232011 taBM_proximalAutozeroValveError n/a 232012 taBM_proximalAutozeroValveMissing n/a 232013 taBM_distalAutozeroValveError n/a 232014 taBM_distalAutozeroValveMissing n/a 232015 taBM_proximalDistalAutozeroValveMissing n/a 232016 taBM_proximalDistalAutozeroValveError n/a 232017 taBM_autozeroPventMonitorOK n/a 232018 taBM_autozeroPventMonitorOutOfRange n/a 232019 taBM_autozeroPventControlOK n/a 232020 taBM_autozeroPventControlOutOfRange n/a 232022 taBM_autozeroPawOutOfRange n/a 232024 taBM_autozeroQawOutOfRange n/a 232025 taBM_autozeroPawQawOK n/a 232026 taBM_autozeroPawQawOutOfRange n/a 232027 taBM_instrumentTemperatureHigh Device temperature high 232028 taBM_pFilterPressureHigh Technical event 232029 taBM_tinstSensorDefect Technical event 232030 taBM_clockError Technical event 232031 taBM_fiO2CalibrationError n/a 232032 taBM_fiO2CalibrationOK n/a 232034 taBM_O2SensorError Replace Oxygen Sensor 232035 taBM_pfilterSensorDefect Technical event 232036 taBM_co2CalibrationError n/a 232037 taBM_co2CalibrationOK n/a 232038 taBM_co2CalibrationNeeded CO2 calibration needed 232040 taBM_disconnectionQuick n/a 232041 taBM_co2CalibrationStartError n/a 232042 taBM_co2CalibrationStartOK n/a 232043 taBM_reconnectionQuick n/a 232046 taBM_autozeroExpValveOK n/a 424 English | 624165/07 Technical Alarms - 200000 Alarm Codes Error No. Error Description enGB Text Value 232047 taBM_autozeroExpValveError n/a 232048 taBM_autozeroProxValveLeak n/a 232049 taBM_autozeroDistValveLeak n/a 232050 taBM_pventMonitorCalibrationError n/a 232051 taBM_pventMonitorCalibrationOK n/a 232052 taBM_pawCalibrationError n/a 232053 taBM_pawCalibrationOK n/a 232054 taBM_qawCalibrationError n/a 232055 taBM_qawCalibrationOK n/a 232056 taBM_pAmbientPfilterMismatch Technical event 232099 taBM_communicationCheck n/a 233001 taBPG_autozeroPventMonitorFail Technical event 233002 taBPG_autozeroPventControlFail Technical event 233003 taBPG_autozeroPawFail Technical event 233004 taBPG_autozeroQawFail Technical event 233005 taBPG_pressureSensorTolerance Technical event 233006 taBPG_nebulizerValveError Technical event 233020 taBPG_autozeroPventOK n/a 233021 taBPG_autozeroPawQawOK n/a 233022 taBPG_autozeroVentValveOK n/a 233023 taBPG_autozeroVentValveMissing n/a 233024 taBPG_autozeroVentValveError n/a 233025 taBPG_autozeroProxValveOK n/a 233026 taBPG_autozeroProxValveMissing n/a 233027 taBPG_autozeroProxValveError n/a 233028 taBPG_autozeroDistValveOK n/a 233029 taBPG_autozeroDistValveMissing n/a 233030 taBPG_autozeroDistValveError n/a 233031 taBPG_autozeroPventMonitorTimeout n/a 233032 taBPG_autozeroPventControlTimeout n/a 233033 taBPG_autozeroPawQawTimeout n/a 233034 taBPG_autozeroPventMonitorOutOfRange n/a 233035 taBPG_autozeroPventControlOutOfRange n/a 233036 taBPG_autozeroPawOutOfRange n/a 233037 taBPG_autozeroQawOutOfRange n/a 233038 taBPG_autozeroProxDistValveMissing n/a 233039 taBPG_autozeroProxDistValveError n/a 233040 taBPG_autozeroExpValveOK n/a 233041 taBPG_autozeroExpValveError n/a 233042 taBPG_autozeroExpValveTimeout n/a 233043 taBPG_autozeroExpValveFail n/a Hamilton Medical | HAMILTON-C2 Service Manual 16 425 16 Glossary Error No. Error Description enGB Text Value 233044 taBPG_autozeroProxValveLeak n/a 233045 taBPG_autozeroDistValveLeak n/a Ventilation Mode Control 234099 taSND_communicationCheck n/a 241001 taVMC_replaceHEPAfilter Replace HEPA filter 241002 taVMC_tightnessTestReleaseValveDefect Alarming 243001 taALR_alarmSilenceError Technical event 243002 taALR_alarmUnknown Technical event 243003 taALR_loudspeakerDefect Loudspeaker defective 243004 taALR_buzzerDefectAtStartup Buzzer defective 243006 taALR_developSound Technical event 243007 taALR_rtcReset Realtime clock failure 244001 taPM_externalPowerLoss Loss of external power 244002 taPM_batteryCalibrationRequiredBat1 Battery 1: Calibration required 244003 taPM_batteryCalibrationRequiredBat2 Battery 2: Calibration required 244004 taPM_batteryTemperatureHighBat1 Battery 1: Temperature high 244005 taPM_batteryTemperatureHighBat2 Battery 2: Temperature high 244006 taPM_batteryPowerLowPriorityHigh Battery low 244007 taPM_batteryPowerLoss Battery power loss 244008 taPM_battery1WrongBattery Battery 1: Wrong battery 244009 taPM_battery2WrongBattery Battery 2: Wrong battery 244010 taPM_batteryPowerLowPriorityLow Battery low 244011 taPM_batterySystemManagerBusError Battery communication error 244012 taPM_battery1PermanentFailure Battery 1: Defective 244013 taPM_battery2PermanentFailure Battery 2: Defective 244014 taPM_batteryTaskLifeSignTimeout Technical event 244015 taPM_batteryPowerLowPriorityMedium Battery low 244016 taPM_batteryReplacementRequiredBat1 Battery 1: Replacement required 244017 taPM_batteryReplacementRequiredBat2 Battery 2: Replacement required 244099 taPM_communicationCheck n/a 246001 taLLS_serviceNeeded Preventive maintenance required 246002 taLLS_cpuTemperatureHigh Technical event 246003 taLLS_fanError Fan failure 246004 taLLS_processorOverload Technical event 246005 taLLS_alarmMonitorDefect Technical event 246006 taLLS_eepromDefaults Technical event 246007 taLLS_eepromWriteFailed Technical event 246008 taLLS_cpuTemperatureDefect Technical event Power Management Low Level Security 426 English | 624165/07 Technical Alarms - 200000 Alarm Codes Error No. Error Description enGB Text Value 246009 taLLS_devWatchdogDisabled Technical event 246010 taLLS_hardwarepmeterError Technical event 246012 taLLS_postEthernetError Technical event 246013 taLLS_postI2CError Technical event 246014 taLLS_postBootloaderError Technical event 246016 taLLS_postTouchError Touch not functional 246017 taLLS_externalDisplayError Technical event 249001 taCFG_OhOptionFileNotRead Options not found 249002 taCFG_OhOptionFileNotWritten Technical event 249003 taCFG_OhOptionFileWrongVersion Technical event 249004 taCFG_OhOptionFileDefaultGenerated Technical event 249010 taCFG_DscDeviceConfigFileError Technical event 249011 taCFG_DscSetupConfigFileError Technical event 249012 taCFG_DscLastSettingConfigFileError Technical event 249020 taCFG_InvalidOptionBoard Invalid Communication Board 249021 taCFG_CheckHardwareCompatibility Invalid Communication Board 249099 taCFG_communicationCheck n/a 255001 taIVS_PercentageMinVolHighFailed n/a 255002 taIVS_PercentageMinVolHighLimitWrong n/a 255003 taIVS_PercentageMinVolLowFailed n/a 255004 taIVS_PercentageMinVolLowLimitWrong n/a 255005 taIVS_PeepHighLimitWrong n/a 255006 taIVS_PeepHighFailed n/a 255007 taIVS_PeepLowLimitWrong n/a 255008 taIVS_PeepLowFailed n/a 255009 taIVS_OxygenLowFailed n/a 255010 taIVS_OxygenHighFailed n/a 255011 taIVS_PercentageMinVolChangePosFailed n/a 255012 taIVS_PercentageMinVolChangeNegFailed n/a 255013 taIVS_PeepChangePosFailed n/a 255014 taIVS_PeepChangeNegFailed n/a 255015 taIVS_OxygenChangePosFailed n/a 255016 taIVS_OxygenChangeNegFailed n/a 255017 taIVS_RecruitmentRunningTime n/a 255018 taIVS_SimulationSuperviseFailed n/a 255019 taIVS_PercentageMinVolHighFineFailed 255020 taIVS_PercentageMinVolHighLimitFineWrong 255021 taIVS_PercentageMinVolLowFineFailed 255022 taIVS_PercentageMinVolLowLimitFineWrong 255023 taIVS_CtrlOszillationFiO2 255024 taIVS_CtrlOszillationPEEP Hamilton Medical | HAMILTON-C2 Service Manual 16 427 16 Glossary Error No. Error Description enGB Text Value 255025 taIVS_CtrlOszillationPercMinVol 255026 taIVS_DataMismatch Technical event 255027 taIVS_PetCO2FrozenFailed Technical event 255028 taIVS_SpO2FrozenFailed 255029 taIVS_fSpontFrozenFailed Technical event 283001 taVGUI_StartupFailed Technical event 283003 taVGUI_languageNotLoaded Technical event 283004 taVGUI_deviceConfigFileError Technical event Ventilation GUI Technical event 283005 taVGUI_setupConfigFileError Technical event 283007 taVGUI_lastSettingError Technical event 283008 taVGUI_checkSetting Check settings 283009 taVGUI_selftestFailed Self test failed 284001 taSGUI_StartupFailed n/a 284002 taSGUI_alarmServiceHigh Technical event 284003 taSGUI_alarmServiceMedium Technical event 284004 taSGUI_alarmServiceLow Technical event 284005 taSGUI_FlowSensorCalibrationNeeded Technical event 284006 taSGUI_RestartDeviceToEnableTestsCalib Technical event 285001 taAGL_alarmLampsErrorDefect Technical event 285002 taAGL_alarmLampsWarningDefect Technical event 285003 taAGL_BacklightDefect Technical event 285004 taAGL_HardkeyDefect Function key not operational Service GUI 428 English | 624165/07 Technical Failure Ending in Safety Mode - 300000 Alarm Codes 16 16.2.3 Technical Failure Ending in Safety Mode - 300000 Alarm Codes Table 16-4. Technical Failure Ending in Safety Mode - 300000 Alarm Codes Error No. Error Description Safety ventilation Gas Delivery 331001 tfsGD_pventPressureSensorDefect Safety ventilation Safety ventilation Breath Monitoring 332001 enGB Text Value tfsBM_qawFlowSensorError Safety ventilation Ventilation Mode Control Safety ventilation 341001 tfsVMC_breathSettingsNotAccepted Safety ventilation 341002 tfsVMC_returnedBreathSettingsIncorrect Safety ventilation 341003 tfsVMC_adaptiveSettingsInvalid Safety ventilation 341004 tfsVMC_breathSettingsTimeout Safety ventilation 341005 tfsVMC_controllerSettingsInvalid Safety ventilation 341006 tfsVMC_returnedAlarmLimitsIncorrect Safety ventilation 341007 tfsVMC_modeTransitionNotDefined Safety ventilation 343001 tfsALR_communicationTimeout Safety ventilation 346002 tfsLLS_watchdogFailedALR Safety ventilation 346003 tfsLLS_watchdogFailedALR_MFmeasure Safety ventilation 346004 tfsLLS_watchdogFailedESL Safety ventilation 346005 tfsLLS_watchdogFailedESL_MMILog Safety ventilation 346006 tfsLLS_watchdogFailedESL_ControlLog Safety ventilation 346007 tfsLLS_watchdogFailedESL_BreathLog Safety ventilation 346008 tfsLLS_watchdogFailedESL_EventSDRReader Safety ventilation 346009 tfsLLS_watchdogFailedESL_ServiceSDRReader Safety ventilation 346010 tfsLLS_watchdogFailedESL_EventSDRWriter Safety ventilation 346011 tfsLLS_watchdogFailedESL_ServiceSDRWriter Safety ventilation 346012 tfsLLS_watchdogFailedESL_MFmeasure Safety ventilation 346013 tfsLLS_watchdogFailedGUIL_Touch Safety ventilation 346014 tfsLLS_watchdogFailedGUIL_PTKnob Safety ventilation 346015 tfsLLS_watchdogFailedGUIL_Hardkeys Safety ventilation 346016 tfsLLS_watchdogFailedGUIL_ScreenShot Safety ventilation 346017 tfsLLS_watchdogFailedLM Safety ventilation 346019 tfsLLS_watchdogFailedLLS_HWException Safety ventilation 346020 tfsLLS_watchdogFailedLLS_CommonServer Safety ventilation 346022 tfsLLS_watchdogFailedLLS_AlarmStatus Safety ventilation 346023 tfsLLS_watchdogFailedLLS_MFmeasure Safety ventilation 346024 tfsLLS_watchdogFailedPM Safety ventilation 346025 tfsLLS_watchdogFailedPM_MFmeasure Safety ventilation 346026 tfsLLS_watchdogFailedQSPI Safety ventilation 346027 tfsLLS_watchdogFailedQSPI_MMI_Server Safety ventilation 346028 tfsLLS_watchdogFailedQSPI_BM_Server Safety ventilation Hamilton Medical | HAMILTON-C2 Service Manual 429 16 Glossary Error No. Error Description enGB Text Value 346029 tfsLLS_watchdogFailedQSPI_BPG_Server Safety ventilation 346030 tfsLLS_watchdogFailedQSPI_LLS_Server Safety ventilation 346031 tfsLLS_watchdogFailedQSPI_QuadSPI Safety ventilation 346032 tfsLLS_watchdogFailedQSPI_MFmeasure Safety ventilation 346033 tfsLLS_watchdogFailedRTC_RealTimeClock Safety ventilation 346034 tfsLLS_watchdogFailedRTC_AlarmClock Safety ventilation 346035 tfsLLS_watchdogFailedSC Safety ventilation 346036 tfsLLS_watchdogFailedSGUI Safety ventilation 346037 tfsLLS_watchdogFailedSTU Safety ventilation 346038 tfsLLS_watchdogFailedVMC Safety ventilation 346039 tfsLLS_watchdogFailedVMC_VentAlarming Safety ventilation 346040 tfsLLS_watchdogFailedVMC_VentControl Safety ventilation 346041 tfsLLS_watchdogFailedVMC_VentMonitoring Safety ventilation 346042 tfsLLS_watchdogFailedVMC_MFmeasure Safety ventilation 346043 tfsLLS_watchdogFailedVGUI Safety ventilation 346044 tfsLLS_watchdogFailedVGUI_ModeControl Safety ventilation 346045 tfsLLS_watchdogFailedVGUI_MFmeasure Safety ventilation 346046 tfsLLS_watchdogFailedLLS_eepromWrite Safety ventilation 346047 tfsLLS_watchdogFailedBM_AlarmingSlow Safety ventilation 346048 tfsLLS_watchdogFailedSND_SoundControl Safety ventilation 346049 tfsLLS_watchdogFailedSND_Sound Safety ventilation 346050 tfsLLS_watchdogFailedVT Safety ventilation 346051 tfsLLS_watchdogFailedCFG_Configuration Safety ventilation 346052 tfsLLS_watchdogFailedEXM_ComBase Safety ventilation 346053 tfsLLS_watchdogFailedIVS n/a 346054 tfsLLS_safetyFailureDetected Safety ventilation 346055 tfsLLS_watchdogFailedPD_BlackBoxLogging Safety ventilation 383001 tfsVGUI_settingsNotAccepted Safety ventilation 383002 tfsVGUI_returnedSettingsIncorrect Safety ventilation 383003 tfsVGUI_settingsValidation Safety ventilation 383004 tfsVGUI_monitoringChannelObservationFailed Safety ventilation 383005 tfsVGUI_VMCTimeout Safety ventilation 383006 tfsVGUI_returnedDeviceSettingsIncorrect Safety ventilation 383007 tfsVGUI_trendingChannelObservationFailed Safety ventilation Ventilation GUI Application GUI Library 385001 tfsAGL_alarmingChannelObservationFailed Safety ventilation 385002 tfsAGL_safetyFailureDetected Safety ventilation 385003 tfsAGL_PMChannelObservationFailed Safety ventilation tfsGUIL_bitmapNotLoaded Safety ventilation GUI Library 386001 430 English | 624165/07 Technical Failure Ending in Ambient Mode - 400000 Alarm Codes 16 16.2.4 Technical Failure Ending in Ambient Mode - 400000 Alarm Codes Table 16-5. Technical Failure Ending in Ambient Mode - 400000 Alarm Codes Error No. Error Description enGB Text Value 431001 tfaGD_blowerFault Blower fault 431002 tfaGD_blowerDisconnected Technical fault 431004 tfaGD_inspirationValveOverCurrent Technical fault 431006 tfaGD_expirationValveOverCurrent Technical fault 431007 tfaGD_ventOutputOverTemperature Technical fault 431008 tfaGD_qventFlowSensorError Technical fault 431009 tfaGD_qO2FlowSensorError Technical fault 431010 tfaGD_controlREGSPITimeout Technical fault 431011 tfaGD_monitorREGSPITimeout Technical fault 431012 tfaGD_flowSensorMeasurementImprecise Technical fault 431013 tfaGD_calibrationReadFailed Technical fault 431014 tfaGD_iExpValveSensorDefect Technical fault 431015 tfaGD_emergencyOffFailed Technical fault 431017 tfaGD_inspirationValveDisconnected Technical fault 432001 tfaBM_blowerOverTemperature Technical fault 432002 tfaBM_instrumentOverTemperature Technical fault 432003 tfaBM_pressureNotReleasedAmbient n/a 433001 tfaBPG_breathMonitoringTickTimeout Technical fault 433002 tfaBPG_controlTickTimingError n/a 443001 tfaALR_watchdogFailedLLS Technical fault 444001 tfaPM_batteriesTotalDischarge Battery totally discharged 444004 tfaPM_voltageOutOfTolerance Technical fault 444005 tfaPM_shutdownFailed Technical fault 446001 tfaLLS_cpuTemperatureCritical Technical fault 446002 tfaLLS_safetyFailed Technical fault 446003 tfaLLS_watchdogFailedGD Technical fault 446004 tfaLLS_watchdogFailedGD_ValveRegulator Technical fault 446005 tfaLLS_watchdogFailedGD_BlowerRegulator Technical fault 446006 tfaLLS_watchdogFailedGD_Monitoring Technical fault 446007 tfaLLS_watchdogFailedGD_ValveLog Technical fault 446008 tfaLLS_watchdogFailedGD_BlowerLog Technical fault 446009 tfaLLS_watchdogFailedGD_MFmeasure Technical fault 446010 tfaLLS_watchdogFailedGD_GPIOFaultInput Technical fault 446011 tfaLLS_watchdogFailedBM Technical fault 446012 tfaLLS_watchdogFailedBM_GasDeliveryCom Technical fault 446013 tfaLLS_watchdogFailedBM_Hardwarecontrol Technical fault 446014 tfaLLS_watchdogFailedBM_BPGclientReceiver Technical fault 446015 tfaLLS_watchdogFailedBM_BPGserverReceiver Technical fault Hamilton Medical | HAMILTON-C2 Service Manual 431 16 Glossary Error No. Error Description enGB Text Value 446016 tfaLLS_watchdogFailedBM_BreathDataCalc Technical fault 446017 tfaLLS_watchdogFailedBM_MFmeasure Technical fault 446018 tfaLLS_watchdogFailedBPG Technical fault 446019 tfaLLS_watchdogFailedBPG_GDclientReceiver Technical fault 446020 tfaLLS_watchdogFailedBPG_MFmeasure Technical fault 446021 tfaLLS_exceptionHappened Technical fault 446022 tfaLLS_voltageError Technical fault 446023 tfaLLS_watchdogFailedLLS_Workload Technical fault 446024 tfaLLS_Am3v3Error Technical fault 446025 tfaLLS_AmAdcError Technical fault 446026 tfaLLS_AmVrefError Technical fault 446028 tfaLLS_clockError Technical fault 446029 tfsLLS_ambientFailureDetected Technical fault 446030 tfaLLS_AmWatchdogOccurred Technical fault 446031 tfaLLS_restartVentilationFailed Technical fault 481001 tfaSTU_firstRunMigrationError Technical fault 481002 tfaSTU_crcError Technical fault 481003 tfaSTU_unknownPartNumber Unknown part number 481004 tfaSTU_technicalStateError Technical state failed 483001 tfaVGUI_StartupBasicStartupFailure Technical fault 483002 tfaVGUI_StartupScreenFailure Technical fault 483003 tfaVGUI_StartupLoadGuiBasicsFailure Technical fault 483004 tfaVGUI_StartupSelftestPrologFailure Technical fault 483005 tfaVGUI_StartupSelftestFailure Technical fault 483006 tfaVGUI_StartupSelftestEpilogFailure Technical fault 483007 tfaVGUI_StartupLoadingBitmapsFailure Technical fault 483008 tfaVGUI_StartupInitializationFailure Technical fault 484001 tfaSGUI_StartupBasicStartupFailure Technical fault 484002 tfaSGUI_StartupScreenFailure Technical fault 484003 tfaSGUI_StartupLoadGuiBasicsFailure Technical fault 484004 tfaSGUI_StartupSelftestPrologFailure Technical fault 484005 tfaSGUI_StartupSelftestFailure Technical fault 484007 tfaSGUI_StartupLoadingBitmapsFailure Technical fault 484008 tfaSGUI_StartupInitializationFailure Technical fault 485001 tfaAGL_ambientFailureDetected Technical fault 432 English | 624165/07 500000 Alarm Codes 16 16.2.5 500000 Alarm Codes These are classed as “invisible technical faults” and are not shown on the LCD Display during operation. They refer to exception handling performed by HAMILTON-C2 during operation, and are for Hamilton Medical internal use. Do not consult Hamilton Medical Technical Support concerning these technical faults. They have no significance for hospital workers or Hamilton-Medical Trained Field Service Engineers. Hamilton Medical | HAMILTON-C2 Service Manual 433 16 434 Glossary English | 624165/07 17 Test Report 17.1 Summary of Test Report.......................................................................................................................436 435 17 Test Report 17.1 Summary of Test Report NOTICE For details on the tests, checks, calibrations, and maintenance procedures, refer to the relevant Chapters/Sections of the HAMILTON-C2 Service Manual (PN 624165) and the HAMILTON-C2 Operator's Manual (PN 624131). Table 17-1. Unit Information HAMILTON-C2 Serial Number: SN __ __ __ __ __ __ Inventory Number: __ __ __ __ __ __ __ Software Version: __ __ __ __ __ __ __ Location of Installation: ________________________________ Name of Customer: ________________________________ Date (YYYY/MM/DD): __ __ __ __ /__ __ /__ __ Table 17-2. Component Replacement Task Yes No Not Applicable HAMILTON-C2 Manual Reference Oxygen Sensor replaced? See the Operator's Manual Section 10.3.3. Air Intake Dust Filter cleaned or replaced? See the Operator's Manual Section 10.3.1. Fan Filter cleaned or replaced? See the Operator's Manual Section 10.3.1. O2 Inlet Filter replaced? See the Operator's Manual Section 13.3. HEPA Filter replaced? See the Operator's Manual Section 10.3.1. Batteries calibrated (% error > 5%)? — See the Service Manual Section 5. Batteries replaced if State of Health (SoH) < 20% — See the Service Manual Section 5. Blower Module replaced (ebt ≥ 100%)? — See the Service Manual Section 8.9.2 and Section 10.4.10. Table 17-3. General Maintenance Task All marks, symbols, and labels related to safety are complete? Yes No Not Applicable — Trolley: are the trolley wheels and brakes functioning correctly and all the fixing screws tightened? — AC Battery Test successfully performed? — DC Battery Test successfully performed? 436 See the Operator's Manual Section 1.4. — Visually inspect the unit for damage and clean the unit if necessary. Power (battery) loss simulated: Ambient Mode test successfully performed? HAMILTON-C2 Manual Reference — See the Service Manual Section 8.21. See the Service Manual Section 8.21. — See the Service Manual Section 8.21. English | 624165/07 Summary of Test Report Task Yes No Not Applicable Buzzer Sound Test: Does the buzzer sounds for at least 2 minutes? — 17 HAMILTON-C2 Manual Reference See the Service Manual Section 8.21. RS232 Test performed successfully? See the Service Manual Section 8.21. Nurse Call functional test performed successfully? See the Service Manual Section 8.23. CO2 accuracy in range? See the Service Manual Section 8.22. Visually check that the battery charge indicator on the IP display is green. — — Pre-operational checks performed in accordance with the procedures in the Operator's Manual. — See the Operator's Manual Section 3.2. Alarm tests performed in accordance with the procedures in the Operator's Manual? — See the Operator's Manual Section 3.5. Electrical safety tests performed successfully? — See the Service Manual Section 7. — See the Service Manual Section 8.24. — See the Service Manual Section 8.24. Instrument report download performed and archived? — Event Log download performed and archived? Table 17-4. Service Software Task Yes No Not Applicable Ventilator Info: Technical State, Service, and RTC (Real Time HAMILTON-C2 Manual Reference — See the Service Manual Section 8.7. — See the Service Manual Section 8.14. — See the Service Manual Section 8.15. — See the Service Manual Section 8.16. Clock) checked and each status is OK? Adjustments / Calibration: Touch Screen, Insp. Valve, Pres- sure, Exp. Valve, Oxygen Sensor, Flow Sensor calibrated and each status is OK? Component Tests: Electronics, User I/F, Pneumatics 1, and Pneumatics 2 checked and each status is OK? System Tests: Pressure, Flow, O2 mixer, Tightness, Alarming checked and each status is OK? Table 17-5. Record of Test Equipment Test Equipment Type/Model Serial Number Electrical Safety Analyzer __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __ Flow Analyzer __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __ Pressure Gauge __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __ Remarks Table 17-6. Declaration of Tests and Calibration Declaration of Tests and Calibration: Date (YYYY/MM/DD): Name: Signature: __ __ __ __/__ __/__ __ ___________________________________ __________________________ Hamilton Medical | HAMILTON-C2 Service Manual 437 17 438 Test Report English | 624165/07 18 Environmental Requirements 18.1 Environmental Requirements ................................................................................................................440 439 18 Environmental Requirements 18.1 Environmental Requirements Table 18-1. Environmental Requirements Temperature Operating: 5°C to 40°C (41°F to 104°F) Storage: -20°C to 60°C (-4°F to 140°F), in original packaging -15°C to 60°C (5°F to 140°F) otherwise Altitude -650m to 4000m (-2132ft. to 13,120 ft.) above sea level Relative Humidity Operating / Storage 10% to 95%, non-condensing Water Protection IP21 440 English | 624165/07 19 Document History 19.1 Document History ................................................................................................................................442 441 19 Document History 19.1 Document History Table 19-1. Document History 00 October 2008 Release of the HAMILTON-C2 Service Manual 01 July, 2009 Release of the HAMILTON-C2 Service Manual. Complete rework and new update. 02 November, 2009 Release of the HAMILTON-C2 Service Manual. New update. 03 November, 2010 Updating of software versions. 04 Redesign and updated information of Service Manual. 05 Update for speaker and schematics. 06 February, 2015 New spare parts, software installation, updated pictures. 07 August, 2018 Complete rework and update. Implementation into Schema ST4. 442 English | 624165/07 Hamilton Medical AG Via Crusch 8, 7402 Bonaduz, Switzerland ( +41 58 610 10 20 info@hamilton-medical.com www.hamilton-medical.com