Uploaded by chitranshu chandel

533586510-HAMILTON-C2-Service-Manual-en-624165-07

advertisement
Service Manual
HAMILTON-C2
0197
REF 160001
Software version 2.2.x
624165/07 | August 2018
Service Manual
HAMILTON-C2
August 2018
624165/07
© 2018 Hamilton Medical AG. All rights reserved. Printed in
Switzerland.
No part of this publication may be reproduced, stored in a
database or retrieval system, or transmitted in any form or
by any means, electronic, mechanical, or by photocopying,
recording, or otherwise, without prior written permission of
Hamilton Medical AG.
This document may be revised, replaced, or made obsolete
by other documents by Hamilton Medical AG at any time
and without notice. Ensure that you have the most current
applicable version of this document; if in doubt, contact the
technical support department of Hamilton Medical AG,
Switzerland. While the information set forth herein is
believed to be accurate, it is not a substitute for the exercise
of professional judgment.
Nothing in this document shall limit or restrict in any way
Hamilton Medical AG’s right to revise or otherwise change
or modify the equipment (including its software) described
herein, without notice. In the absence of an express, written
agreement to the contrary, Hamilton Medical AG has no
obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including
software) described herein.
Manufacturer
Hamilton Medical AG
Via Crusch 8, CH-7402 Bonaduz,
Switzerland
Phone: (+41) 58 610 10 20
Fax: (+41) 58 610 00 20
info@hamilton-medical.com
www.hamilton-medical.com
Distributor in USA
Hamilton Medical, Inc.
4990 Energy Way, P.O. Box 30008
Reno, NV 89520
Phone: (775) 858-3200
Toll-free: (800) 426-6331
Fax: (775) 856-5621
marketing@hamilton-medical.net
The equipment must be operated, serviced, or upgraded
only by trained professionals. Hamilton Medical AG’s sole
responsibility with respect to the equipment and its use is as
stated in the limited warranty provided in the device Operator’s Manual.
Hamilton Medical AG shall not be liable for any loss, cost,
expense, inconvenience, or damage that may arise out of
misuse of the product, or if non-Hamilton Medical AG parts
were used when replacing parts, or if serial numbers were
amended, deleted, or removed.
If returning parts to Hamilton Medical AG, be sure to use
the standard Hamilton Medical returned goods authorization (RGA) procedure. Disposal of parts shall follow all local,
state, and federal regulation with respect to environmental
protection.
Hamilton Medical AG will make available, on request, circuit
diagrams, component parts lists, descriptions, calibration
instructions, or other information that will assist appropriately trained personnel to repair those parts of the equipment designated by Hamilton Medical AG to be repairable.
For all proprietary as well as third-party trademarks used by
Hamilton Medical AG, see www.hamilton-medical.com/
trademarks. Product and/or company names marked with a
§
symbol may be the trademarks and/or registered trademarks of their respective owners, including but not limited
to Aerogen§, Nihon Kohden§, Masimo§, Masimo SET§, Masimo rainbow SET§, Respironics§.
4
English | 624165/07
Table of Contents
Table of Contents
Chapter 1
Chapter 2
Chapter 3
Conventions.................................................................................................................. 13
1.1
Notes, Cautions and Warnings ............................................................................................................ 14
1.2
Typographic Conventions .................................................................................................................... 14
1.3
Expressions.......................................................................................................................................... 15
1.4
Foreword ............................................................................................................................................ 16
HAMILTON-C2 Overview ............................................................................................... 19
2.1
Front Components Overview ............................................................................................................... 20
2.2
Rear Components Overview ................................................................................................................ 21
2.3
Interaction Panel Internal Components Overview ................................................................................. 22
2.4
Ventilation Unit Internal Components Overview................................................................................... 23
2.4.1
Top Section ................................................................................................................................ 24
2.4.2
Middle Section ........................................................................................................................... 25
2.4.3
Bottom Section........................................................................................................................... 26
Pneumatics: Overview and Theory of Operation............................................................. 27
3.1
Overview ............................................................................................................................................. 28
3.2
Blower Module Overview..................................................................................................................... 29
3.3
Oxygen Mixer Block Assembly Overview .............................................................................................. 30
3.4
Inspiratory Valve Overview................................................................................................................... 31
3.5
Ambient Valve Overview...................................................................................................................... 32
3.6
Expiratory Valve Overview.................................................................................................................... 33
3.7
Patient Flow Sensor Overview .............................................................................................................. 34
3.8
Oxygen Sensor Overview ..................................................................................................................... 35
3.9
Principal Gas Flow Overview ................................................................................................................ 36
3.9.1
Components that Manage the Principal Gas Flow....................................................................... 37
3.9.2
Principal Gas Flow ...................................................................................................................... 38
3.9.3
Components for the Ambient State Gas Flow ............................................................................. 39
3.9.4
Components for the Over-pressure Relief Gas Flow .................................................................... 40
3.9.5
Components for the Nebulizer Gas Flow..................................................................................... 41
3.10
Chapter 4
Ventilation Control Principle ................................................................................................................ 42
3.10.1
Pressure Sensor Assembly ........................................................................................................... 43
3.10.2
Flow Sensors used for Flow Measurement Overview ................................................................... 44
3.10.3
Flow Restrictors used for Flow Reduction Overview..................................................................... 45
3.11
Gas Rinse Flows................................................................................................................................... 46
3.12
Measurements of Pneumatic Sensors................................................................................................... 47
3.13
Components that Measure Gas Flow and Pressure............................................................................... 48
Electronics: Component Functions Overview.................................................................. 49
4.1
Overview ............................................................................................................................................. 50
4.2
Interaction Panel Electronics Components............................................................................................ 51
4.3
Ventilation Unit Electronics Components ............................................................................................. 52
4.3.1
Mainboard MSP160200.............................................................................................................. 52
Hamilton Medical | HAMILTON-C2 Service Manual
5
Table of Contents
Chapter 5
5.2
4.3.4
Voltage Distribution and Test Points ........................................................................................... 56
4.3.5
Power Supply Components Overview (MSP160600 Revision 02) ................................................. 58
Rechargeable Lithium Ion Battery......................................................................................................... 60
5.1.1
Introduction ............................................................................................................................... 60
5.1.2
Replacing the Battery.................................................................................................................. 62
5.1.3
Storage Hints.............................................................................................................................. 62
5.1.4
Backup Battery ........................................................................................................................... 62
5.1.5
Battery Life Expectancy / Li-Ion Battery Maintenance................................................................... 62
5.1.6
Lithium Ion Battery Maintenance ................................................................................................ 63
Battery Charger / Calibrator................................................................................................................. 64
5.2.1
Battery Charger / Calibrator Installation ...................................................................................... 65
5.2.2
Battery Charging ........................................................................................................................ 65
5.2.3
Battery Recharge Time................................................................................................................ 65
5.4
Battery Re-calibration .......................................................................................................................... 68
5.4.1
General Information ................................................................................................................... 69
5.4.2
Re-calibration Time..................................................................................................................... 70
5.4.3
Re-calibration Description ........................................................................................................... 71
5.4.4
Rechargeable Lithium Ion Battery Capacity ................................................................................. 72
Lithium Ion Battery Handling Precautions............................................................................................. 73
Preventive Maintenance and Testing.............................................................................. 75
Overview ............................................................................................................................................. 76
6.1.1
Introduction ............................................................................................................................... 76
6.1.2
Checking the Software Level ...................................................................................................... 76
6.1.3
Items Required for Preventive Maintenance and Testing.............................................................. 77
6.1.4
Overview of Preventive Maintenance and Testing ....................................................................... 77
6.2
Hospital Preventive Maintenance ......................................................................................................... 78
6.3
Engineer Preventive Maintenance ........................................................................................................ 79
Electrical Safety ............................................................................................................. 81
7.1
Electrical Safety Overview .................................................................................................................... 82
7.2
Electrical Safety Tests........................................................................................................................... 82
7.2.1
Preparation for Tests................................................................................................................... 82
7.2.2
Perform the Tests ....................................................................................................................... 83
7.2.3
Device Type ................................................................................................................................ 83
7.2.4
Internal Resistance Checks.......................................................................................................... 84
7.2.5
Electrical Safety Tests - IEC 62353............................................................................................... 87
Service Software............................................................................................................ 91
8.1
6
Mainboard Components............................................................................................................. 54
Evaluation of the Battery State of Health ............................................................................................. 66
6.1
Chapter 8
4.3.3
5.3
5.5
Chapter 7
Mainboard MSP160382.............................................................................................................. 53
Lithium Ion Battery ........................................................................................................ 59
5.1
Chapter 6
4.3.2
Introduction ........................................................................................................................................ 92
English | 624165/07
Table of Contents
8.2
Functions of the Service Software ........................................................................................................ 93
8.3
Structure of the Service Software......................................................................................................... 94
8.4
Starting the Service Software............................................................................................................... 95
8.5
Service Software Screen Layout ........................................................................................................... 96
8.6
Making Screenshots ............................................................................................................................ 97
8.7
Ventilator Info ..................................................................................................................................... 98
8.8
Technical State .................................................................................................................................... 99
8.8.1
Hardware Version Tab ................................................................................................................ 100
8.8.2
Software Version Tab ................................................................................................................. 101
8.8.3
Modify Tab................................................................................................................................. 102
8.8.4
Transfer Tab ............................................................................................................................... 104
8.9
Service Timer ....................................................................................................................................... 105
8.9.1
Service Timer Tab ....................................................................................................................... 105
8.9.2
Blower Timer Tab ....................................................................................................................... 107
8.10
Real Time Clock (RTC).......................................................................................................................... 108
8.11
Battery Data ........................................................................................................................................ 109
8.12
Tests Overview .................................................................................................................................... 110
8.13
Tests / Calibration Screen..................................................................................................................... 111
8.14
Adjustments / Calibration .................................................................................................................... 112
8.14.1
Calibration Tab........................................................................................................................... 113
8.14.2
Show Tab ................................................................................................................................... 134
8.14.3
Default Settings Tab ................................................................................................................... 140
8.15
Component Tests ................................................................................................................................ 142
8.15.1
Electronics Tab ........................................................................................................................... 145
8.15.2
User Interface Tab ...................................................................................................................... 161
8.15.3
Pneumatics 1 Tab ....................................................................................................................... 163
8.15.4
Pneumatics 2 Tab ....................................................................................................................... 178
8.16
System Test ......................................................................................................................................... 194
8.16.1
Pressure...................................................................................................................................... 196
8.16.2
Flow ........................................................................................................................................... 198
8.16.3
O2 Mixer .................................................................................................................................... 200
8.16.4
Tightness.................................................................................................................................... 202
8.16.5
Alarming .................................................................................................................................... 204
8.17
Sensor Data......................................................................................................................................... 207
8.17.1
Sensor 1 ..................................................................................................................................... 208
8.17.2
Sensor 2 ..................................................................................................................................... 209
8.17.3
Sensor 3 ..................................................................................................................................... 210
8.17.4
Sensor Status.............................................................................................................................. 211
8.18
Log/Config Files................................................................................................................................... 212
8.18.1
Event Log ................................................................................................................................... 213
8.18.2
Service Log ................................................................................................................................. 214
Hamilton Medical | HAMILTON-C2 Service Manual
7
Table of Contents
8.18.3
8.19
Chapter 9
Export (Instrument Report and Events) ........................................................................................ 215
Software Updates................................................................................................................................ 218
8.19.1
Software Update from Version < 2.2.0 ....................................................................................... 218
8.19.2
Software Update from Version 2.x to 2.2.5 or Higher ................................................................. 221
8.20
Pre-operational Checks........................................................................................................................ 226
8.21
General Tests....................................................................................................................................... 227
8.22
CO2 Sensor Accuracy Check................................................................................................................ 230
8.23
Nurse Call Functional Check ................................................................................................................ 240
8.24
Final Tests............................................................................................................................................ 242
8.25
Set the Service Timer ........................................................................................................................... 242
8.26
Check the Battery Data........................................................................................................................ 242
Technical Faults .............................................................................................................243
9.1
Technical Faults Overview .................................................................................................................... 244
9.2
Troubleshooting .................................................................................................................................. 246
Chapter 10 Components Removal/Assembly....................................................................................247
10.1
Notes on Maintenance and Replacement............................................................................................. 248
10.2
Securing the Power Cord..................................................................................................................... 249
10.3
Interaction Panel Components Removal/Assembly ............................................................................... 250
10.3.1
Ventilation Unit Top Cover Removal/Assembly ............................................................................ 250
10.3.2
Interaction Panel Small Rear Cover Removal/Assembly ................................................................ 251
10.3.3
Interaction Panel Removal/Assembly ........................................................................................... 252
10.3.4
Interaction Panel Large Rear Cover Removal/Assembly ................................................................ 253
10.3.5
Backlight Driver Board Removal/Assembly................................................................................... 254
10.3.6
Front Panel Board Removal/Assembly.......................................................................................... 255
10.3.7
LCD Display Removal/Assembly................................................................................................... 258
10.3.8
LCD Display Backlight Removal/Assembly.................................................................................... 259
10.3.9
P&T Control Knob Encoder Removal/Assembly............................................................................ 261
10.4
Ventilation Unit Components Removal/Assembly ................................................................................. 262
10.4.1
Backup Battery Pack Removal/Assembly...................................................................................... 262
10.4.2
Filter Cover Removal/Assembly ................................................................................................... 265
10.4.3
HEPA Filter Removal/Assembly .................................................................................................... 266
10.4.4
High Pressure Oxygen Connector Removal/Assembly .................................................................. 267
10.4.5
Front and Rear Covers Removal/Assembly................................................................................... 268
10.4.6
ESM Module Removal/Assembly ................................................................................................. 269
10.4.7
Ventilation Unit Mainboard Removal/Assembly ........................................................................... 270
10.4.8
Top Foam Removal/Assembly...................................................................................................... 271
10.4.9
Cooling Fan Removal/Assembly .................................................................................................. 273
10.4.10 Blower Module Removal/Assembly ............................................................................................. 274
10.4.11 Power Supply Removal/Assembly ................................................................................................ 275
10.4.12 Middle Foam Removal/Assembly................................................................................................. 276
10.4.13 Pressure Sensor Assembly Removal/Assembly.............................................................................. 278
8
English | 624165/07
Table of Contents
10.4.14 Inspiratory Valve Removal/Assembly............................................................................................ 280
10.4.15 Qvent Flow Sensor Removal/Assembly ........................................................................................ 282
10.4.16 Ambient Valve Removal/Assembly .............................................................................................. 284
10.4.17 Oxygen Sensor Block Removal/Assembly..................................................................................... 286
10.4.18 Rinse Flow Block Removal/Assembly ........................................................................................... 288
10.4.19 Expiratory Valve Removal/Assembly ............................................................................................ 291
10.4.20 Battery Compartment Removal/Assembly ................................................................................... 292
10.4.21 Oxygen Mixer Block Assembly Removal/Assembly ....................................................................... 294
10.4.22 Nebulizer Valve Removal/Assembly ............................................................................................. 296
10.4.23 Bottom Foam Removal/Assembly ................................................................................................ 297
10.5
Ventilation Unit Components, Tubings and Cables Assembly ............................................................... 298
10.5.1
Bottom Foam Assembly .............................................................................................................. 298
10.5.2
Mains Power Cable Assembly ..................................................................................................... 299
10.5.3
DC Input Power Cable Assembly................................................................................................. 300
10.5.4
Oxygen Mixer Block Assembly Mounting .................................................................................... 301
10.5.5
Oxygen Mixer Valve Cable Assembly........................................................................................... 302
10.5.6
Nebulizer Cable Assembly........................................................................................................... 303
10.5.7
Battery Compartment Assembly ................................................................................................. 304
10.5.8
Expiratory Valve Assembly .......................................................................................................... 305
10.5.9
Rinse Flow Block and Tubing Assembly ....................................................................................... 306
10.5.10 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly............................................. 309
10.5.11 Oxygen Sensor Block and Tubing Assembly ................................................................................ 312
10.5.12 Pressure Sensor Module Assembly .............................................................................................. 313
10.5.13 Middle Foam Assembly............................................................................................................... 314
10.5.14 Power Supply Assembly .............................................................................................................. 315
10.5.15 Cooling Fan Assembly ................................................................................................................ 316
10.5.16 Blower Module Assembly ........................................................................................................... 317
10.5.17 Top Foam and Mainboard Assembly ........................................................................................... 319
Chapter 11 Maintenance Tools and Test Equipment ........................................................................323
11.1
Overview ............................................................................................................................................. 324
11.2
Standard Tools .................................................................................................................................... 324
11.3
Special Tools........................................................................................................................................ 325
11.3.1
Digital Voltmeter ........................................................................................................................ 325
11.3.2
Electrical Safety Tester ................................................................................................................ 325
11.3.3
ESD (ElectroStatic Discharge) Protection...................................................................................... 326
11.3.4
Two Pin Cylindrical Service Tool .................................................................................................. 327
11.4
Test Equipment ................................................................................................................................... 328
11.5
Calibration of Test Equipment ............................................................................................................. 332
Chapter 12 Spare Parts ....................................................................................................................333
12.1
Introduction to Major Components ..................................................................................................... 334
12.2
Trolley PN 160170 ............................................................................................................................... 335
Hamilton Medical | HAMILTON-C2 Service Manual
9
Table of Contents
12.3
Trolley PN 160150 ............................................................................................................................... 336
12.4
Major Components of the Interaction Panel......................................................................................... 337
12.5
Major Components of the Ventilation Unit .......................................................................................... 338
12.5.1
Ventilation Unit Covers............................................................................................................... 339
12.5.2
Ventilation Unit Top Section ....................................................................................................... 340
12.5.3
Ventilation Unit Middle Section .................................................................................................. 341
12.5.4
Ventilation Unit Bottom Section.................................................................................................. 342
12.6
Interaction Panel Parts ......................................................................................................................... 343
12.6.1
Interaction Panel Covers ............................................................................................................. 343
12.6.2
Backlight Driver Board ................................................................................................................ 344
12.6.3
Front Panel Board ....................................................................................................................... 345
12.6.4
LCD Display and Touch Screen.................................................................................................... 346
12.6.5
P&T Control Knob Encoder ......................................................................................................... 347
12.7
Ventilation Unit Parts........................................................................................................................... 348
12.7.1
Top Cover .................................................................................................................................. 348
12.7.2
Front Cover ................................................................................................................................ 349
12.7.3
Rear Cover ................................................................................................................................. 350
12.7.4
Ventilation Unit Base .................................................................................................................. 351
12.7.5
Backup Battery Pack Compartment............................................................................................. 352
12.7.6
Expiratory Valve Assembly .......................................................................................................... 353
12.7.7
Mixer Block Assembly ................................................................................................................. 354
12.8
Detailed Parts Descriptions................................................................................................................... 355
12.8.1
Tubing and Fittings..................................................................................................................... 355
12.8.2
Clamps and Fasteners................................................................................................................. 357
12.8.3
Stickers and Labels ..................................................................................................................... 358
12.8.4
Pneumatic Parts and Assemblies ................................................................................................. 359
12.8.5
Flow Sensors .............................................................................................................................. 361
12.8.6
Assembled Components............................................................................................................. 362
12.8.7
Colored Metal Rings and Controls .............................................................................................. 363
12.8.8
Rubber Seals and Grommets....................................................................................................... 363
12.8.9
Metal Brackets and Frame Components...................................................................................... 364
12.8.10 Electrical/Electronic Cables.......................................................................................................... 366
12.8.11 Electronic Printed Circuit Boards ................................................................................................. 369
12.8.12 Electrical/Electronic Devices......................................................................................................... 372
12.8.13 External Covers and External Hardware....................................................................................... 374
12.8.14 Screws........................................................................................................................................ 376
12.8.15 Standoffs.................................................................................................................................... 379
12.8.16 Washers ..................................................................................................................................... 380
12.8.17 Screw Nuts ................................................................................................................................. 380
Chapter 13 Schematics ....................................................................................................................381
13.1
10
Block Diagram ..................................................................................................................................... 382
English | 624165/07
Table of Contents
13.2
Wiring Diagram ................................................................................................................................... 383
Chapter 14 Software Revisions, Features and Compatibility .............................................................385
14.1
Introduction ........................................................................................................................................ 386
14.2
Software Version Information .............................................................................................................. 386
Chapter 15 Hardware Revisions, Features and Compatibility ............................................................397
15.1
Introduction ........................................................................................................................................ 398
15.2
Component History and Improvements................................................................................................ 398
15.2.1
Interaction Panel......................................................................................................................... 398
15.2.2
Ventilation Unit .......................................................................................................................... 402
Chapter 16 Glossary ........................................................................................................................415
16.1
Definitions of Expression...................................................................................................................... 416
16.2
Alarm Codes ....................................................................................................................................... 421
16.2.1
Patient Alarms - 100000 Alarm Codes ........................................................................................ 421
16.2.2
Technical Alarms - 200000 Alarm Codes .................................................................................... 423
16.2.3
Technical Failure Ending in Safety Mode - 300000 Alarm Codes ................................................. 429
16.2.4
Technical Failure Ending in Ambient Mode - 400000 Alarm Codes ............................................. 431
16.2.5
500000 Alarm Codes ................................................................................................................. 433
Chapter 17 Test Report....................................................................................................................435
17.1
Summary of Test Report ...................................................................................................................... 436
Chapter 18 Environmental Requirements .........................................................................................439
18.1
Environmental Requirements ............................................................................................................... 440
Chapter 19 Document History .........................................................................................................441
19.1
Document History................................................................................................................................ 442
Hamilton Medical | HAMILTON-C2 Service Manual
11
Table of Contents
12
English | 624165/07
1
Conventions
1.1
Notes, Cautions and Warnings ............................................................................................................. 14
1.2
Typographic Conventions..................................................................................................................... 14
1.3
Expressions .......................................................................................................................................... 15
1.4
Foreword ............................................................................................................................................. 16
13
1
Conventions
1.1 Notes, Cautions and Warnings
WARNING
A WARNING alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the use or
misuse of the device.
CAUTION
A CAUTION alerts the user to the possibility of a problem with the device associated with its use or misuse, such as device
malfunction, device failure, damage to the device, or damage to other property.
NOTICE
A NOTE emphasizes information of particular importance.
1.2 Typographic Conventions
Table 1-1. Typographic Conventions
Effect
Example
Function
Bold
Configuration
Marks text quoted directly from the ventilator screen.
Bold
TRIGGER
Marks text that is quoted directly from:
• The ventilator case
• A touch key
• A printed circuit board
Number sequence
1. Step one in a sequence
2. Step two in a sequence
Letter sequence
A identifies part A
B identifies part B
Blue text 1.2
14
Organizes the performance of actions into
steps.
Used in photos and illustrations to identify
the topic being discussed and relates to a
specific text.
If you are using a PDF file to view this, you
can hyperlink to the reference by clicking on
these items.
English | 624165/07
1
Expressions
1.3 Expressions
Table 1-2. Expressions
Expression
Example
Explanation
Activate
Activate LED ON/OFF.
Using the P&T Control Knob, you must first
select the LED ON/OFF Button on the ventilator's screen, and then press the P&T Control Knob. The button on the screen
changes its appearance, so that it looks
“pressed”. It now performs its function
(turning the LED ON in this case). Sometimes you are told to “activate and set” a
field. In this case you first activate the field,
and then turn the P&T Control Knob to set a
value.
Deactivate
Deactivate LED ON/OFF.
With LED ON/OFF still selected and activated, you must press the P&T Control Knob
again. The button on the screen changes its
appearance, so that it looks “unpressed”. It
stops performing its function (turning the
LED OFF in this case).
Select
Select LED ON/OFF.
Touch the Touch Screen to select the LED
ON/OFF Button.
De-select
De-select LED ON/OFF.
Touch the Touch Screen to deselect the LED
ON/OFF Button.
Pressure
Patient pressure is 80 mbar.
Pressure refers to the amount of pressure
above ambient pressure. If the patient pressure (Ppat) is 80 mbar, it means the pressure
is 80 mbar above the ambient (room) pressure.
Software version
0.6.0
The ventilator contains a memory device
that holds software identified by a version
number.
Update
This kit or software enables an upgrade to
implement a new function.
An update is an improvement to an existing
function. An update normally involves only
software. A software update is indicated by
an increment of the last digit of the three
digit software version number (e.g. 1.2.1 to
1.2.2).
Upgrade
This kit or software enables an upgrade to
implement a new function.
An upgrade is the addition of new functions
to a device. There are three ways to perform
an upgrade:
• Add a hardware item that offers additional functions.
• Upgrade to a higher software revision.
Depending on how important is the software upgrade, it is either marked by an
increment on the first or on the second
digit (e.g. 1.2.0 to 1.3.0 or 1.2.0 to
2.0.0).
• Upgrade to a higher type of software.
*
With this kit, you can update or upgrade
from software version.
Hamilton Medical | HAMILTON-C2 Service Manual
Unless otherwise stated, a syntax variable
(“wild card”) indicates the use of any alphanumeric character.
15
1
Conventions
1.4 Foreword
WARNING
To prevent possible patient injury, disconnect the patient from the ventilator before you start service or maintenance.
Service the HAMILTON-C2 only as described in this manual, using only parts approved or supplied by Hamilton Medical.
Incorrect parts, components or assemblies could result in patient injury. See available Spare Parts Section 12.
Hamilton Medical does not offer any warranty on incorrectly used parts.
The Service Manual is for:
Engineers who have successfully completed a Hamilton Medical service training course for the HAMILTON-C2.
Training courses are held regularly in Bonaduz, Switzerland, at Hamilton Medical’s headquarters, and at other locations
throughout the world. For more information, visit the Hamilton Medical AG Partner-net.
NOTICE
If you have questions about testing or any part of this manual, contact Hamilton Medical AG Technical Support.
The HAMILTON-C2 Service Manual contains:
The architecture and components of the HAMILTON-C2.
In addition, information on testing, troubleshooting and repairing the HAMILTON-C2. Appendices.
The HAMILTON-C2 Service Manual does not contain:
Information about operating the HAMILTON-C2. See the HAMILTON-C2 Operator’s Manual for operating instructions.
This HAMILTON-C2 Service Manual covers:
The HAMILTON-C2 Software version 1.1.1 and higher versions.
16
English | 624165/07
1
Foreword
The HAMILTON-C2 Service Manual Structure:
Section
Function
Your Responsibility
HAMILTON-C2 Overview Section 2
This section explains the theory behind the
HAMILTON-C2.
You should fully understand this section.
Pneumatics: Overview and Theory of
Operation Section 3
This section explains each component and the
gas flows, flow measurements, and pressure
measurements in the pneumatic circuits.
You should be able to name and explain the
functions of all the major components.
Electronics: Component Functions
Overview Section 4
This section explains the basic functions of the
printed circuit boards, Printed circuit boards
are not repaired in the field.
You should be able to identify all circuit
boards, and know where they are positioned
in the HAMILTON-C2.
Lithium Ion Battery Section 5
This section explains the use, care and maintenance of the Lithium Ion Battery Pack.
You have to understand the safety concerns
and hazards, and know how to perform
charging and calibration of the Lithium Ion
Battery Pack.
Preventive Maintenance and Testing
Section 6
This section gives a schedule for maintenance.
You have to be familiar with the maintenance
schedule for the HAMILTON-C2.
Hospital Preventive Maintenance Section 6.2
This section gives the Hospital Preventive
Maintenance details.
You should be able to perform this maintenance and ascertain if this maintenance is
being performed regularly.
Engineer Preventive Maintenance Section 6.3
This section gives Engineer Preventive Maintenance details.
You have to be able to perform all the tasks in
this section.
Electrical Safety Overview Section 7.1
This section lists further test's you must perform on the HAMILTON-C2 before you start
the Service Software test's.
You have to be able to use this section to test
the HAMILTON-C2.
Service Software Section 8
This section explains how you perform the
test's that are built into the software of the
HAMILTON-C2.
You have to know how to perform all the
appropriate test's.
Technical Faults Section 9.1
This section gives an overview of the alarm
indications.
You should know how to interpret technical
faults to isolated faulty components.
Components Removal/Assembly Section 10
This section explains how to remove and
assemble each major component.
You have to be able to use this section to
make repairs.
Maintenance Tools and Test Equipment
Section 11
This section lists the equipment you require to
work on the HAMILTON-C2.
Check this section to make sure you have the
correct tools and test equipment.
Spare Parts Section 12
Information resource.
You only require this section when you must
order spare parts.
Schematics Section 13
This section includes many of the schematics
produced by Hamilton Medical for internal
use.
You are sometimes directed to this section
when reading in other parts of the manual.
Software Revisions, Features and Compatibility Section 14.1
This section informs about the software versions that were introduced over time.
You only require this section when the software must be updated or when the update is
recommended.
Hardware Revisions, Features and Compatibility Section 15
This section informs about the hardware components that have changed over time and
about its compatibility.
You only require this section when parts need
to be replaced.
Glossary Section 16
Glossary contains Definitions of Expressions
and Alarm Codes
Helps you to understand the ventilator and the
alarms.
Test Report Section 17.1
Test Report pages for the Service Software
section.
Complete the report when using the service
software tests.
Hamilton Medical | HAMILTON-C2 Service Manual
17
1
18
Conventions
English | 624165/07
2
HAMILTON-C2 Overview
2.1
Front Components Overview ................................................................................................................ 20
2.2
Rear Components Overview ................................................................................................................. 21
2.3
Interaction Panel Internal Components Overview.................................................................................. 22
2.4
Ventilation Unit Internal Components Overview ................................................................................... 23
19
2
HAMILTON-C2 Overview
2.1 Front Components Overview
Figure 2-1. Front Components Overview
A
Alarm Lamp
G
Front Panel Keys
B
USB Socket
H
P&T (Press and Turn) Control Knob
C
10.4” TFT Display with Touch Screen and Backlight
I
Patient Breathing Circuit Connection to the Patient
D
Nebulizer Connection
J
Patient Breathing Circuit Connection from the Patient
E
Oxygen Sensor and Cover (Oxygen Sensor not shown)
K
Quick Lock to remove HAMILTON-C2 from the Trolley
F
Patient Flow Sensor Connections
L
Shelf Stand
20
English | 624165/07
Rear Components Overview
2
2.2 Rear Components Overview
Figure 2-2. Rear Components Overview
A
Ethernet Connection
H
DC Input
B
RS232 Interface Connection
I
Lift Handle
C
Option Slot for Communication Board
J
Interaction Panel
D
Dust Air Filter and HEPA Filter Element
K
Cooling Fan
E
High Pressure Oxygen DISS or NIST Connection
L
Battery Compartment
F
Low Pressure Oxygen Connection
M
Ventilation Unit Filter Cover
G
Mains Power Inlet
N
Cable Holder
Hamilton Medical | HAMILTON-C2 Service Manual
21
2
HAMILTON-C2 Overview
2.3 Interaction Panel Internal Components Overview
Figure 2-3. Interaction Panel Internal Components Front Overview
A
Alarm Lamp LED’s
C
Front Panel Board
B
10.4” TFT Display with Backlight
D
P&T (Press and Turn) Control Knob Encoder
G
Front Panel Board
Figure 2-4. Interaction Panel Internal Components Rear Overview
22
E
Loudspeaker
F
Backlight Inverter Board (Obsolete)
English | 624165/07
Ventilation Unit Internal Components Overview
2
2.4 Ventilation Unit Internal Components Overview
Figure 2-5. Ventilation Unit Internal Components Overview
The Ventilation Unit is divided into 3 sections:
A
Top Section
B
Middle Section
Hamilton Medical | HAMILTON-C2 Service Manual
C
Bottom Section
23
2
HAMILTON-C2 Overview
2.4.1 Top Section
Figure 2-6. Top Section
A
Mainboard
C
Options Slot for Communication Board
B
ESM (Embedded System Module) with Shield
D
Top Foam Section
Jumper position for the HAMILTON-C2 with Mainboard (MSP160382):
For the HAMILTON-C2 use the two left pins.
Figure 2-7. Jumper position for the HAMILTON-C2
24
English | 624165/07
Middle Section
2
2.4.2 Middle Section
Figure 2-8. Ventilation Unit Internal Components Middle Section Front Overview
A
Blower Module
B
Middle Foam Section
Figure 2-9. Ventilation Unit Internal Components Middle Section Rear Overview
C
Heat Exchanger
F
HEPA Filter
D
Power Supply
G
Cooling Fan
E
HEPA Filter Latch
Hamilton Medical | HAMILTON-C2 Service Manual
25
2
HAMILTON-C2 Overview
2.4.3 Bottom Section
Figure 2-10. Ventilation Unit Internal Components Bottom Section Front Overview
A
Pressure Sensor Assembly
F
Oxygen Sensor
B
Autozero Valves
G
QO2 Flow Sensor
C
Qvent Flow Sensor
H
Expiratory Valve
D
Inspiratory Valve
I
Bottom Foam Section
E
Ambient Valve
Figure 2-11. Ventilation Unit Internal Components Bottom Section Rear Overview
J
Battery Module
K
Oxygen Mixer Block
26
L
Battery Release Latches
English | 624165/07
3
Pneumatics: Overview and Theory of
Operation
3.1
Overview.............................................................................................................................................. 28
3.2
Blower Module Overview ..................................................................................................................... 29
3.3
Oxygen Mixer Block Assembly Overview ............................................................................................... 30
3.4
Inspiratory Valve Overview ................................................................................................................... 31
3.5
Ambient Valve Overview ...................................................................................................................... 32
3.6
Expiratory Valve Overview .................................................................................................................... 33
3.7
Patient Flow Sensor Overview............................................................................................................... 34
3.8
Oxygen Sensor Overview ...................................................................................................................... 35
3.9
Principal Gas Flow Overview ................................................................................................................. 36
3.10
Ventilation Control Principle ................................................................................................................. 42
3.11
Gas Rinse Flows ................................................................................................................................... 46
3.12
Measurements of Pneumatic Sensors ................................................................................................... 47
3.13
Components that Measure Gas Flow and Pressure ............................................................................... 48
27
3
Pneumatics: Overview and Theory of Operation
3.1 Overview
WARNING
Repair of parts, components or assemblies is not permitted by Hamilton Medical AG, as incorrectly repaired parts, components or assemblies could result in patient injury. See the available Spare Parts Section 12
This section introduces all the major pneumatic components in the ventilator. In addition, the gas flows and theory of operation are explained.
NOTICE
The figures in this section show exploded and transparent views of the ventilator components. The exploded diagrams may
not always show components in their correct positions.
Figure 3-1. Pneumatic System Drawing
Air LowPressure Inlet
O2 LowPressure Inlet
O2 HighPressure Inlet
Instrument Borders
Filter Air
Filter O2
Filter Air
HEPA
Sensor QO2
Sensor Pfilter
Mixer valve O2
Noise dampening
Nebulizer
valve
Sensor TBlower
Sensor Pambient
Blower
Noise dampening
Restrictor
Nebulizer
Sensor Tinst
Heat exchanger
Sensor
Pvent_control
Sensor
Paw
Sensor
Pflowsensor
Inspiratory valve
Filter
Rinseflow
Sensor Qvent
Autozero
valve
Sensor
Pvent_monitor
Oxygen sensor
Autozero
valve
Autozero
valve
Ambient valve
Standard
rinse flow
Internal expiratory
valve
Standard
rinse flow
Nebulizer
proximal flow
proximal pressure
28
Expiratory
valve cover
to patient
English | 624165/07
Blower Module Overview
3
3.2 Blower Module Overview
The Blower Module supplies the air and oxygen mixture for patient inspiration.
Figure 3-2. Blower Module Overview
A
Blower Module
C
Dust Air Filter (not shown)
B
HEPA Filter Element
D
Blower and Motor
Hamilton Medical | HAMILTON-C2 Service Manual
29
3
Pneumatics: Overview and Theory of Operation
3.3 Oxygen Mixer Block Assembly Overview
The Oxygen Mixer Block Assembly controls the flow of oxygen into the Blower Assembly.
Figure 3-3. Oxygen Mixer Block Assembly Overview
A
Shaped Tube from the QO2 Flow Sensor to the Blower Module
E
High Pressure Oxygen Control Valve
B
QO2 Flow Sensor
F
Outlet to supply Oxygen to the Front Panel Nebulizer Connection
C
Oxygen Mixer Block Assembly
G
High pressure Oxygen DISS or NIST Connection
D
Nebulizer Valve
H
Low pressure Oxygen Connection
30
English | 624165/07
Inspiratory Valve Overview
3
3.4 Inspiratory Valve Overview
The Inspiratory Valve precisely controls the volume and pressure of the air/oxygen mixture from the Blower Module to the
Patient Breathing Circuit.
Figure 3-4. Inspiratory Valve Overview
A
The Air/Oxygen Mixture from the Blower Module
D
Pressure Spring
B
Inspiratory Valve
E
Inspiratory Valve Voice Coil Assembly
C
Inspiratory Valve Sealing Membrane
F
Qvent Flow Sensor
Hamilton Medical | HAMILTON-C2 Service Manual
31
3
Pneumatics: Overview and Theory of Operation
3.5 Ambient Valve Overview
The Ambient Valve enables the patient to inhale ambient air, if there is a fatal failure on the ventilator (Ambient Mode). It is
opened if there is a fault on the ventilator, allowing the patient to inspire ambient air.
Figure 3-5. Ambient Valve Overview
A
The Air/Oxygen Mixture from the Inspiratory Valve to the
Patient Breathing Circuit
D
Ambient Valve Solenoid
B
Ambient Valve Assembly
E
Outlet connection to the Patient Breathing Circuit
C
Ambient Valve Sealing Membrane
32
English | 624165/07
Expiratory Valve Overview
3
3.6 Expiratory Valve Overview
WARNING
Never attach a Spirometer or any other device or tube to the exhaust port of the Expiratory Valve. This will cause the ventilator to lose full control of PEEP/CPAP.
The Expiratory Valve enables gas to escape from the Patient Breathing Circuit in a controlled manner, allowing the patient to
exhale.
The Expiratory Valve:
• Maintains PEEP/CPAP if required - to do this, the action of the Expiratory Valve is closely synchronized with the action of
the Inspiratory Valve
• In Ambient State, the Expiratory Valve is open
Figure 3-6. Expiratory Valve Overview
A
Expiratory Valve Assembly
E
Expired gas from the Patient Breathing Circuit
B
Plunger
F
Expiratory Valve Voice Coil Assembly
C
Silicone Membrane
G
Expiratory Valve Exhaust
D
Inlet connection from the Patient Breathing Circuit
Hamilton Medical | HAMILTON-C2 Service Manual
33
3
Pneumatics: Overview and Theory of Operation
3.7 Patient Flow Sensor Overview
The HAMILTON-C2 accurately measures flow, volume, and pressure in the patient’s airway with the Patient Flow Sensor.
The area of the orifice changes depending on the flow rate. It opens progressively as the flow increases, creating a pressure
drop across the orifice. The pressure difference is measured by a high-precision differential pressure sensor (Pflowsensor)
located on the Pressure Sensor Assembly inside the ventilator. The pressure difference varies with flow (relationship determined during flow sensor calibration), so the patient’s flow is determined from the pressure drop. The HAMILTON-C2 calculates volume from the flow measurements. The Patient Flow Sensor is highly accurate even in the presence of secretions,
moisture, and nebulized medications. The HAMILTON-C2 continuously flushes the sensing tubes with mixed gases (rinse
flow) to prevent blockage.
Figure 3-7. Patient Flow Sensor Overview
A
Blue Tube - Proximal side measured Pflowsensor Pressure
and Paw Pressure
D
Two-way Gas Flow connection on the Distal side of the
Patient Flow Sensor
B
Clear Tube - Distal side measures Pflowsensor Pressure
E
Two-way Gas Flow connection on the Proximal side of the
Patient Flow Sensor
C
Variable Orifice Membrane
NOTICE
Ventilation is not totally dependent on the Patient Flow Sensor. If the Patient Flow Sensor malfunctions, patient ventilation
continues. The ventilator has an internal flow measurement at the Inspiratory Valve for Inspiratory Flow.
34
English | 624165/07
Oxygen Sensor Overview
3
3.8 Oxygen Sensor Overview
WARNING
The ventilator should never be used for ventilating a patient without some means of monitoring the oxygen content in the
gas mixture delivered to the patient. The ventilator can not be operated without an Oxygen Sensor installed.
The Oxygen Sensor (A, see Figure 3-8 Oxygen Sensor Overview) is attached to the front left side of the ventilator behind the
Oxygen Sensor Cover (B, see Figure 3-8 Oxygen Sensor Overview). It is used to monitor the oxygen concentration in the gases
delivered to the patient. The Oxygen Sensor performs only a monitoring function, and can be disabled by the user, if
required.
Figure 3-8. Oxygen Sensor Overview
A
Oxygen Sensor
B
Oxygen Sensor Cover
NOTICE
Hamilton Medical Oxygen Sensors are available from Hamilton Medical AG only.
The Hamilton Medical Oxygen Sensor produces a voltage between 11 and 13mV at 21% O2 which changes with oxygen
concentration. Each unit comprises a teflon-bonded gold cathode and a lead anode, submerged in a liquid electrolyte solution. When oxygen diffuses through the flouropolymer membrane, the electrochemical reduction of oxygen on the cathode
and the corresponding oxidation of the anode generates an electrical current that is proportional to the concentration of oxygen.
Hamilton Medical | HAMILTON-C2 Service Manual
35
3
Pneumatics: Overview and Theory of Operation
3.9 Principal Gas Flow Overview
The ventilator regulates the flow of gases at three points, as represented in the Principal Gas Flow Diagram (see Figure 3-9
Principal Gas Flow Diagram).
The “Principal Gas Flow” is the flow of air and oxygen into the ventilator, out to the patient and directly at the Expiratory
Valve back to the room. Room air (A) is aspirated, mixed with compressed oxygen (C) from an external high or low pressure
supply (B) and compressed in the Blower Module (D). This flow of the air/oxygen mixture is managed by components that
regulate the gases at three points:
• Room air and oxygen mixture in the Blower Module (D regulation point 1).
• Room air and oxygen mixture through the Inspiratory Valve and Patient Flow Sensor (E - regulation point 2).
• Expired gas from the patient through the Expiratory Valve (F - regulation point 3).
Expired gases from the patient are allowed to escape into the room.
The following sections provides details of the components that manage the “Principal Gas Flow”.
Figure 3-9. Principal Gas Flow Diagram
NOTICE
High-pressure-oxygen: Maximum Pressure 600 kPa / Maximum Flow 200 l/min
Low-pressure-oxygen: Maximum Pressure 600 kPa / Maximum Flow 15 l/min
36
English | 624165/07
3
Components that Manage the Principal Gas Flow
3.9.1 Components that Manage the Principal Gas Flow
The main components that control the flow of gases through the ventilator are shown in the Figure 3-10 Components that
Manage the Principal Gas Flow .
The components have the following functions:
Figure 3-10. Components that Manage the Principal Gas Flow
A
Room Air
Room Air.
B
Oxygen
Oxygen from an external high or low pressure source.
C
Mixer Block Assembly
Measures and Regulates the flow of the high pressure oxygen (B, see Figure 3-10 Components that Manage the Principal Gas Flow) into the Blower Module.
D
Blower Module
Compresses the air and oxygen mixture with a Turbine Blower.
E
Inspiratory Valve and Qvent Flow Sensor
Measures and Regulates the flow of the air/oxygen mixture to the patient and the pressure during inspiration. It regulates the Base Flow during expiration.
F
Patient Breathing Circuit
Supplies air/oxygen gas mixture to the patient, and removes exhalation gases from the
patient.
G
Patient Flow Sensor
The Patient Flow Sensor monitors the flow of gases to and from the patient’s airway.
H
Patient
The patient breathes in and out (inhales and exhales) with the assistance of the ventilator.
I
Expiratory Valve
Regulates the flow of expiratory gases from the patient. The Expiratory Valve is nearly
closed during inspiration and regulates PEEP during expiration.
Hamilton Medical | HAMILTON-C2 Service Manual
37
3
Pneumatics: Overview and Theory of Operation
3.9.2 Principal Gas Flow
The following flow occurs in the “Principal Gas Flow”:
Figure 3-11. Principal Gas Flow
A
Room Air
F
From the Turbine Blower to the Inspiratory Valve
B
High or Low Pressure Oxygen Supply
G
Passes the Ambient Valve
C
HEPA Filter Assembly
H
Air/Oxygen Mixture into the Patient Breathing Circuit
D
Air and oxygen mixed in the Blower Module
I
Expired gas to the Expiratory Valve
E
Air/Oxygen Mixture flows through the Turbine Blower
38
English | 624165/07
Components for the Ambient State Gas Flow
3
3.9.3 Components for the Ambient State Gas Flow
If there is a total breakdown of ventilation, the ventilator goes into Ambient State.
In the Ambient State, the Ambient Valve is unpowered, allowing the patient to inspire ambient air through the Ambient
Valve Sealing Membrane.
In the Ambient State, the Expiratory Valve is also unpowered, enabling gas to escape from the Patient Breathing Circuit in a
controlled manner, allowing the patient to exhale into the room. The Expiratory Valve Sealing Membrane has a one-way flow
from the Patient Breathing Circuit.
The one-way function of the Expiratory Valve make sure the patient does not inhale his exhaled CO2 (rebreathing).
Figure 3-12. Components for the Ambient State Gas Flow
A
Ambient Valve Sealing Membrane
E
Expiratory Valve
B
Ambient air from the room
F
Expiratory Valve Membrane
C
Ambient Valve Solenoid
G
Expired gas to the room
D
Into the Patient Breathing Circuit
H
From the Patient Breathing Circuit
Hamilton Medical | HAMILTON-C2 Service Manual
39
3
Pneumatics: Overview and Theory of Operation
3.9.4 Components for the Over-pressure Relief Gas Flow
If the pressure from the Inspiratory Valve for the Patient Breathing Circuit is too high (15 mbar above Pmax), the Ambient
Valve is driven to an open position to relieve the excess pressure to the room.
It also functions as an additional backup to the pressure limit set by the user for the Patient Breathing Circuit.
Figure 3-13. Components for the Over-pressure Relief Gas Flow
A
Over-pressure from the Inspiratory Valve
D
Ambient Valve Solenoid
B
Ambient Valve Sealing Membrane
E
Over-pressure from the Patient Breathing Circuit
C
Excess pressure to the room
40
English | 624165/07
Components for the Nebulizer Gas Flow
3
3.9.5 Components for the Nebulizer Gas Flow
The ventilator provides oxygen with sufficient pressure to drive the Nebulizer Jar. The oxygen flow is reduced to approximately 8 liters/minute by means of an internal flow restrictor in the Oxygen Mixer Block Assembly. A valve attached to the
Oxygen Mixer Block Assembly switches the flow of oxygen to the Nebulizer Jar through the output connection at the Front
Panel. The pneumatic nebulizer is inactive when low pressure oxygen is used.
Figure 3-14. Components for the Nebulizer Gas Flow
A
Nebulizer Front Panel connection
C
Nebulizer Valve
B
Tubing from Oxygen Mixer Block Assembly to the Nebulizer
Front Panel connection
D
High Pressure Oxygen Inlet
Hamilton Medical | HAMILTON-C2 Service Manual
41
3
Pneumatics: Overview and Theory of Operation
3.10 Ventilation Control Principle
The HAMILTON-C2 is a pressure controlled ventilator. This means that the applied volume is controlled by a specific calculated pressure and not by a fixed, permanent volume only. The HAMILTON-C2 does not contain a tank. The required pressure
is generated by a turbine, known as a Blower Module, which turns at approximately 35'000 RPM during normal ventilation.
This pressure source provides a pressure that is always at least 10 mbar above the required ventilation pressure, e.g. PEEP +
Pcontrol + 10 mbar = Pblower. To get the pressure in the Patient Breathing Circuit, the outlet pressure of the Inspiratory
Valve is measured 1000 times per second, so the Expiratory Valve can react within a fraction of a second.
On the other hand, the Expiratory Valve manages the outflow of the gas. It applies a known force to the membrane which
has a direct effect on the gas in the Patient Breathing Circuit (back pressure). It never completely closes the outlet of the
Patient Breathing Circuit during ventilation, so the membrane is floating all the time.
The correct ventilation pressure is given by the close collaboration of the Inspiratory Valve, that applies the pressure into the
Patient Breathing Circuit and the Expiratory Valve that stabilizes it by a controlled release of the gas.
42
English | 624165/07
3
Pressure Sensor Assembly
3.10.1 Pressure Sensor Assembly
Figure 3-15. Pressure Sensor Assembly
A
Pressure Sensor Assembly
E
• Provides components for pressure measurements
B
Pvent_monitor Pressure Sensor
• A measurement of the pressure in the Patient Breathing
Circuit as measured at the Proximal Flow Sensor, in the
proximal chamber attached to the blue (patient side) pressure-sensing tube
F
• A sensor which measures the pressure at the patient outlet, used for monitoring
C
Pvent_control Pressure Sensor
Pflowsensor Pressure Sensor
• A measurement of the pressure difference between the
proximal and distal chambers of the Patient Proximal Flow
Sensor
Hamilton Medical | HAMILTON-C2 Service Manual
Pvent_monitor Autozero Valve
• Switches the Pvent_monitor air way pressure measured at
the Patient Flow Sensor to allow for offset re-calculation
to compensate for sensor drift for Pvent_control and
Pvent_monitor
G
• A sensor which measures the pressure at the patient outlet, used for controlling
D
Paw Pressure Sensor
Distal Autozero Valve
• Switches the Pflowsensor Pressure Sensor to Ambient Air
to allow for offset re-calculation to compensate for sensor
drift
H
Proximal Autozero Valve
• Switches the Pflowsensor and Paw Pressure Sensor to
Ambient Air to allow for offset re-calculation to compensate for sensor drift
43
3
Pneumatics: Overview and Theory of Operation
3.10.2 Flow Sensors used for Flow Measurement Overview
These Flow Sensors are used to measure the inspiration and expiration flow to/from the patient.
Figure 3-16. Flow Sensors used for Flow Measurement Overview
A
The QO2 Flow Sensor measures the flow of oxygen into the
Blower Module. It works with the Qvent Flow Sensor to control the Air/Oxygen Mixture.
B
The Qvent Flow Sensor measures the flow of the Air/Oxygen
Mixture into the Patient Breathing Circuit
44
C
Patient Flow Sensor with a variable orifice. The variable orifice works as a very flexible resistance to make flow/pressure
relation proportional.
English | 624165/07
Flow Restrictors used for Flow Reduction Overview
3
3.10.3 Flow Restrictors used for Flow Reduction Overview
Figure 3-17. Flow Restrictors used for Flow Reduction Overview
A
A Restrictive “Orifice” for the Oxygen Flow to the Nebulizer
Valve
B
Two precisely matched “Orifice” Restrictors for the Patient
Flow Sensor Rinse Flow
Hamilton Medical | HAMILTON-C2 Service Manual
C
A Restrictive “Screen” for the Patient Flow Sensor Rinse Flow
supply from the Inspiratory Valve
45
3
Pneumatics: Overview and Theory of Operation
3.11 Gas Rinse Flows
The rinse flow is provided on the blower side of the Inspiratory Block Assembly. A constant small flow of gas (approximately
11.5 ml/min at 35 mbar blower pressure) is provided through both tubes to the Patient Flow Sensor. The difference between
the Proximal and Distal Rinse Flows are ±10%. The rinse flow eliminates the possibility of contamination from the patient,
condensing humidity in the measuring tubes and cross-contamination to other patients.
Figure 3-18. Components that generate the Patient Flow Sensor Rinse Flow
A
Gas from the Blower Module
E
Pflowsensor Sensor connections
B
Inspiratory Valve outlet to QVent Flow Sensor
F
Rinse Flow Assembly (including flow restrictor)
C
Tube outlet to the Oxygen Sensor Mounting Block
G
Front panel connections for the Patient Proximal Flow Sensor
D
Tube to the Rinse Flow Assembly
46
English | 624165/07
Measurements of Pneumatic Sensors
3
3.12 Measurements of Pneumatic Sensors
The following table lists all the Pneumatic Sensors in the Ventilator:
Table 3-1. Pneumatic Sensors in the Ventilator
Name
Location
Comment
Pvent_control
Pressure Sensor Assembly
Measures pressure at the Inspiratory Valve outlet for controlling
Pvent_monitor
Pressure Sensor Assembly
Measures pressure at the Inspiratory Valve outlet for monitoring
Paw
Pressure Sensor Assembly
Measures pressure at the Proximal Side of the Patient Flow Sensor
Pflowsensor
Pressure Sensor Assembly
Measures differential pressure across the Patient Flow Sensor
Pambient
Mainboard
Measures ambient pressure or the room pressure
Pfilter
Blower Module
Measures pressure after the filter; used to indicate if the HEPA Filter
needs to be replaced
Qvent
Ventilation Flow Sensor
Measures the air/oxygen flow in the Patient Breathing Circuit.
QO2
Oxygen Flow Sensor
Measures the oxygen flow into the Blower Module
Hamilton Medical | HAMILTON-C2 Service Manual
47
3
Pneumatics: Overview and Theory of Operation
3.13 Components that Measure Gas Flow and Pressure
The Qvent Flow Sensor measures the flow of Air/Oxygen in the Patient Breathing Circuit. The Pvent_control measurement is
used to control the Inspiratory Valve. The Pvent_monitor Sensor measures the pressure in the Patient Breathing Circuit.
During normal ventilation, the Flow Sensor Autozero Valves open at calculated intervals to expose the Pflowsensor Pressure
Sensor to ambient air for the Autozero offset compensation.
Figure 3-19. Components that Measure Gas Flow and Pressure
A
Qvent Flow Sensor
D
• Measures the flow of Air/Oxygen to the Patient Breathing
Circuit
B
Pvent_monitor Pressure Sensor
• Monitor the pressure in the Patient Breathing Circuit
C
Pflowsensor Pressure Sensor
• Measures the pressure difference between the two chambers of the Patient Flow Sensor, and enables the flow of
gas to and from the patient to be calculated
E
Paw Pressure Sensor
• Measures the pressure on the proximal side of the Patient
Flow Sensor
Pvent_control Pressure Sensor
• Used to control the Inspiratory Valve
NOTICE
During the Autozero procedure, which occurs in a fraction of a second, the patient is ventilated normally. Since parameters
such as expired tidal volume, resistance, compliance, and PEEP cannot be measured or monitored, the ventilator uses the
most recently measured values until the procedure is complete.
48
English | 624165/07
4
Electronics: Component Functions Overview
4.1
Overview.............................................................................................................................................. 50
4.2
Interaction Panel Electronics Components ............................................................................................ 51
4.3
Ventilation Unit Electronics Components .............................................................................................. 52
49
4
Electronics: Component Functions Overview
4.1 Overview
WARNING
Service the ventilator only as described in this manual, using only parts approved or supplied by Hamilton Medical AG.
Incorrectly repaired parts, components or assemblies could result in patient injury. See available Spare Parts Section 12.
This section introduces the major electronic components in the ventilator. Where you require more information, cross-references direct you to other parts of this manual.
This section does not include a theory of operation, because engineers do not require a detailed knowledge of board-level
electronics to service and maintain the ventilator. All electronic failures are dealt with by replacing complete circuit boards.
Any repairs at a lower level than the spare parts listed in the Spare Parts Section 12 is not permitted
NOTICE
Always send defective Printed Circuit Boards to Hamilton Medical AG with a completed Returned Goods ID Tag.
Refer to PN 699138 ky2help User Guide, available as a download from the Hamilton Medical AG Partner-net.
50
English | 624165/07
Interaction Panel Electronics Components
4
4.2 Interaction Panel Electronics Components
Figure 4-1. Interaction Panel Electronics Components
A
Front Panel Board
E
Loudspeaker and Microphone
• Provides 7 Hard Keys for the operator with LED Indicators
• Audible indication of alarms
• Interfaces the P&T Control Knob Encoder
• Microphone:
• LED Colors available:
– The Microphone monitors the operation of the
Loudspeaker
– Red
– If the Loudspeaker does not function, audible alarm
indication is transferred to the Buzzer
– Green
– Yellow
• Voltage inputs:
– 12 VDC for the Key and LED Communications Board
– 5 VDC for the Backlight Inverter Board
– 3.3 VDC for the P&T Control Knob Encoder and USB
B
Backlight Inverter Board (Obsolete)
F
• Converts 5 VDC (Input Voltage) to 1100 VAC for the Display Backlight
• Used to:
– Download Software from a USB Stick
– Export Data to a USB Stick
Refer to the Hardware Revisions, Features and Compatibility
> Backlight Driver Board Section 15.2.1.3
C
Alarm Lamp LED's
• Indicates Alarm Conditions:
– Yellow - Medium and Low Priority Alarms
– Red - High Priority Alarms and Technical Faults
USB Socket
G
P&T Encoder
• Provides additional controls for interaction with various
screen functions
– -16 Encoder positions from center
– +16 Encoder positions from center
– Switch activation when the P&T Control Knob is
depressed
D
Graphic LCD TFT VGA 10.4" NLT Display with Front Panel
and LED Backlight
• Graphical User Interface (GUI)
Hamilton Medical | HAMILTON-C2 Service Manual
51
4
Electronics: Component Functions Overview
4.3 Ventilation Unit Electronics Components
4.3.1 Mainboard MSP160200
Figure 4-2. Mainboard MSP160200
A
Mainboard MSP160200
O
J11 − FFC Cable to Expiratory Valve (MSP160240)
B
ESM Module
P
P12 − Cable to Ambient Valve (MSP160290)
• Embedded System Microprocessor Module
• Power PC MPC5200 Processor
C
J1 − Ethernet Connection (for internal use only)
Q
P13 − Cable to Nebulizer Valve (PN 160400)
D
J2 − RS232 Connection
R
J14 − FFC Cable (PN 160355) to Pressure Sensor Assembly
(MSP160300)
E
P25 − Cable from Power Supply for Fan Supply
S
P16 − FDC Cable (PN 160357) to Display
F
P3 − Cable (PN 160539) from Power Supply 24 VDC
(MSP160600)
T
J17 − FFC Cable (PN 160356) to Key Panel for P&T and Key
Panel
G
J4 − FFC Cable (PN 160351) to Battery Data
U
J18 − FFC Cable (PN 160356) to Key Panel for Backlight,
Speaker and Microphone
H
P5 − Cable (PN 160372) to DC Input
V
P19 − Cable (PN 160387) to Qvent Flow Sensor (TSI,
MSP399123 or Sensirion, MSP161657)
I
P6 − Cable (PN 160370) to Battery Power
W
P20 − Cable (PN 160354) to Oxygen Sensor (PN 396200)
J
P60 − Options Slot for Communication Board (for CO2,
(SpO2 not implemented in HAMILTON-C2) and Nurse Call
Options)
X
J21 − FFC Cable (PN 160353) to Blower Module Temperature Sensor
K
P7 − Cable to O2 Proportional Valve
Y
J22 − FFC Cable to Inspiratory Valve (MSP160230)
L
J9 − FFC Cable (PN 160347) to Filter Pressure Board
Z
P23 − Cable to Blower Module (MSP160250)
M
J8 − FFC Cable (PN 160359) to Pressure Sensor Board Binary
Valves
N
P10 − Cable (PN 160387) to QO2 Flow Sensor (TSI,
MSP399124 or Sensirion, MSP161657)
52
AA
P28 − Cable to Fan 12V (PN 160346)
English | 624165/07
4
Mainboard MSP160382
4.3.2 Mainboard MSP160382
Figure 4-3. Mainboard MSP160382
A
Mainboard MSP160382
O
P12 − Cable to Ambient Valve (MSP160290)
B
ESM Module
P
P13 − Cable to Nebulizer Valve (PN 160400)
• Embedded System Microprocessor Module
• Power PC MPC5123 Processor
C
J1 − Ethernet Connection (for internal use only)
Q
J14 − FFC Cable (PN 160355) to Pressure Sensor Assembly (MSP160300)
D
J2 − RS232 Connection
R
J23 − Not used for the HAMILTON-C2
E
P3 − Cable (PN 160539) from Power Supply 24VDC
(MSP160600)
S
P16 − FDC Cable (PN 160386) to Display
F
J4 − FFC Cable (PN 160351) to Battery Data
T
J17 − FFC Cable (PN 160356) to Key Panel for P&T
Encoder and Key Panel
G
P5 − Cable (PN 160372) to DC Input
U
J18 − FFC Cable (PN 160356) to Key Panel for Backlight,
Speaker and Microphone
H
P6 − Cable (PN 160370) to Battery Power
V
P19 − Cable (PN 160387) to Qvent Flow Sensor (TSI,
MSP399123 or Sensirion, MSP161657)
I
P60 − Options Slot for Communication Board (for CO2,
SpO2 (Not Implemented in HAMILTON-C2) and Nurse
Call Options)
W
P20 − Cable (PN 160354) to Oxygen Sensor (PN 396200)
J
P7 − Cable to O2 Proportional Valve
X
J21 − FFC Cable (PN 160353) to Blower Module Temperature Sensor Board
K
J9 − FFC Cable (PN 160347) to Filter Pressure Board
Y
J22 − FFC Cable to Inspiratory Valve (MSP160230)
L
J8 − FFC Cable (PN 160359) to Pressure Sensor Board
Binary Valves
Z
J23 − Cable to Blower Module (MSP160250)
M
P10 − Cable (PN 160387) to QO2 Flow Sensor (TSI,
MSP399124 or Sensirion, MSP161657)
N
J11 − FFC Cable to Expiratory Valve (MSP160240)
Hamilton Medical | HAMILTON-C2 Service Manual
AA
P28 − Cable to Fan 12V (PN 160346)
53
4
Electronics: Component Functions Overview
4.3.3 Mainboard Components
Table 4-1. Mainboard Components
Mainboard
Serves as the Carrier Board for the Processor Board, Power Management System
and a Connector Board for all Sensors, Valves and Periphery Components
Controls
• Blower
• ON/OFF Key
• Hard Key and Lamps
• Key Panel
• TFT Display
• Ambient Valve
• Expiratory Valve
• Mixer Valves
• Nebulizer Valve
• Emergency Status Indicator
Interfaces with the following Ventilation Unit boards:
• Power Supply
• Pressure Sensor Assembly
• Filter Board
• Smart Batteries
Sensor Monitoring:
• Blower Temperature
• Instrument Temperature
• Pvent_control
• Pvent_monitor
• Pflowsensor
• Pambient
• Qvent Flow Sensor
• QO2 Flow Sensor
• Paw
• Oxygen Sensor
Input Voltage supplied from the Power Supply:
• 24 VDC ±5% 6.6A/150W (Range 22.8V...25.2V)
Output Voltages created by Switching Regulators on
the Ventilation Unit Mainboard:
24 VDC
• Blower
12 VDC
• Autozero Valves
• Inspiratory Valve
• Expiratory Valve
• O2 Inlet Proportional Valve
• Ambient Valve
• Nebulizer Valve
• Alarm Lamps (IP)
• Speaker
• Cooling Fan
5 VDC
• Qvent Flow Sensor
• QO2 Flow Sensor
• DC/AC Backlight Inverter or LED Backlight Driver
• USB Port
54
English | 624165/07
Mainboard Components
Output Voltages created by Switching Regulators on
the Ventilation Unit Mainboard:
3.3 VDC
4
• Qvent Flow Sensor
• QO2 Flow Sensor
• Pressure Sensors
• P&T Control Knob
• ESM Module
• Blower Temperature Sensor
• HEPA Filter Temperature Sensor
Hamilton Medical | HAMILTON-C2 Service Manual
55
4
Electronics: Component Functions Overview
4.3.4 Voltage Distribution and Test Points
Mainboard Voltage Inputs:
Table 4-2. Mainboard Voltage Inputs:
Signal Name
Voltage Level
Voltage Range
Current
Watts
+24V_PS (from power
supply)
24 VDC
22.8 - 25.2 VDC
7A at 22.8 V; 5.7 at 25.2
150W
+24V_DcIn (from 20 V
100W DC Input)
12-24 VDC
11.0 - 27.0 VDC
9.1A at 11 V; 5A at 20V
100W
7.5A at >20 V; 5.6A at
27V (Battery starts charging from DC Input if voltage is above 20 V)
150W
Battery 1 (Primary Battery)
14.4 VDC
12.0 - 16.8 VDC
6A at 16.8 V; 8.3A at
12V
100W
Battery 2 (Optional Battery)
14.4 VDC
12.0 - 16.8 VDC
6A at 16.8 V; 8.3A at
12V
100W
Mainboard Voltage Outputs and Test Points:
Table 4-3. Voltage Test Points P44
Signal Name
Voltage Level
Voltage Range
GND
Pin
Pin
Signal Name
o
o
GND
Voltage Level
Voltage Range
+5V_Usb
5.0 VDC
4.75 - 5.25 V
o
o
+24V_DcIn
12-24 VDC
11.0 - 27.0 VDC
+5V
5.0 VDC
4.60 - 5.40 V
o
o
+24V_PS
24 VDC
22.8 - 25.2 VDC
+3V3
3.3 VDC
3.10 - 3.60 V
o
o
+24V_In
+3V3_BackUp
3.3 VDC
3.00 - 3.60 V
o
o
+31V_Blower
+3V3_Stby
3.3 VDC
3.00 - 4.20 V
o
o
+24V
26.80 VDC
24 - 29 V
+3V3_Ooc
3.3 VDC
3.00 - 4.20 V
o
o
+14V4_Bat2
14.4 VDC
12.0 - 16.8 VDC
+3V_Ref
3.0 VDC
2.91 - 3.09 V
o
o
+14V4_Bat1
14.4 VDC
12.0 - 16.8 VDC
+2V5_Ref
2.5 VDC
2.475 - 2.525 V
o
o
+12V
12 VDC
9.00 - 13.2 V
o
1
GND
GND
P44
56
English | 624165/07
Voltage Distribution and Test Points
4
Table 4-4. Voltage Test Points P42/P41
Pin
Signal Name
Voltage Level
o
GND
o
Pventmonitor
0.472
o
O2
o
Voltage Range
Pin
Signal Name
Voltage Level
Voltage Range
o
GND
0.372 - 0.572
o
Iexpvalve
1.25
0.227 (when the
oxygen sensor is
disconnected)
0.223 - 0.231
o
Iinspvalve
not used
Pflowsensor
1.25
1.15 - 1.35
o
Iblower
not used
o
Paw
0.472
0.372 - 0.572
o
Pventcontrol
0.472
0.372 - 0.572
o
SoundLevel
1.15 (if silent)
1.11 - 1.19
o
TO2
1.25 (measured at
25°C, when the
oxygen sensor is
disconnected)
1.15 - 1.35
o
n.c.
o
QO2 (TSI)
0.25
0.05 - 0.45
QO2 (Sensirion)
0.7
0.5 - 0.9
o
Tvent
1.25
1.15 - 1.35
o
Qvent (TSI)
0.25
0.05 - 0.45
Qvent (Sensirion)
0.7
0.5 - 0..9
o
Tblower
0.5 V
o
ExtInstrument
not used
o
GND
P42
Hamilton Medical | HAMILTON-C2 Service Manual
+ 10mV/ºC
1
1.05 - 1.45
GND
P41
57
4
Electronics: Component Functions Overview
4.3.5 Power Supply Components Overview (MSP160600 Revision 02)
Figure 4-4. Power Supply Overview
A
Power Supply
C
• Provides power conversion from the Mains Power
Power Outlet Connector
• Cable from the Power Supply to the Mainboard P25 Connector and Mainboard P3 Connector
• Input Power:
– 100 to 240 VAC
– 50 to 60 Hz
• Output Voltage:
– +24 VDC ±5%
B
Power Inlet Connector
• Cable from AC Power Inlet
Power Supply Voltage Input and Output:
Table 4-5. Power Supply Voltage Input and Output
Signal Name
Mains Input Voltage
Voltage Level [V]
100 - 240 VAC
Voltage Range [V]
90 - 264 VAC
50 - 60 Hz
Output Voltage
58
+24 VDC
22.8 - 25.2
English | 624165/07
5
Lithium Ion Battery
5.1
Rechargeable Lithium Ion Battery ......................................................................................................... 60
5.2
Battery Charger / Calibrator ................................................................................................................. 64
5.3
Evaluation of the Battery State of Health .............................................................................................. 66
5.4
Battery Re-calibration ........................................................................................................................... 68
5.5
Lithium Ion Battery Handling Precautions ............................................................................................. 73
59
5
Lithium Ion Battery
5.1 Rechargeable Lithium Ion Battery
WARNING
It is mandatory that the HAMILTON-C2 is operated with at least one battery installed.
NOTICE
• At high operating temperature, the battery does not start charging. For example after transport and high temperature.
• The backup battery is intended for short-term use only. It is not intended to be a primary power source.
• Hamilton Medical recommends that the ventilators batteries be fully charged before you ventilate a patient. If the batteries are not fully charged and the AC power fails, always pay close attention to the level of battery charge.
The Backup Battery provides backup power to the HAMILTON-C2.
5.1.1 Introduction
Two backup batteries, one standard and one optional, are used the HAMILTON-C2 when the primary power source is low or
unavailable.
When the primary power source (either AC mains power or a DC power supply) fails, the ventilator automatically switches to
backup battery operation with no interruption in ventilation. An alarm sounds to signal the switchover. To reset the alarm,
silence the alarm to confirm notification of the power system change.
If the optional battery (battery 2) is available and adequately charged, the ventilator switches to this battery first. When battery 2 is depleted or unavailable, the ventilator switches to the standard battery (battery 1).
The batteries power the ventilator until the primary power source is adequate or the battery is depleted.
As a further safeguard, the HAMILTON-C2 provides a low battery alarm. It also has a capacitor-driven backup buzzer that
sounds continuously for at least 2 minutes when battery power is completely lost.
The ventilator recharges the batteries whenever the ventilator is connected to either AC or >20VDC, with or without the
ventilator power switched on. The battery charge indicator (see Figure 5-1 Power Source Symbols and Battery Charge Indicator) lights to show that the batteries are being charged.
The power source symbols in the bottom right-hand corner of the screen show the available power sources. A frame around
a symbol indicates the current ventilator power source. Green indicates the level of battery charge.
60
English | 624165/07
Introduction
5
Figure 5-1. Power Source Symbols and Battery Charge Indicator
Hamilton Medical | HAMILTON-C2 Service Manual
61
5
Lithium Ion Battery
5.1.2 Replacing the Battery
WARNING
When a battery has reached its end-of-life, the capacity is reduced and the battery weakens. The use of a ventilator with
weak batteries can cause the device to report incorrect fuel gauge information, leading to a possible unexpected ventilation
stop if a power failure occurs.
The batteries are hot-swappable; that is, they can be replaced while the ventilator is operating. Replace with a newly charged
battery. For battery replacement refer to the Backup Battery Pack Removal/Assembly Section 10.4.1.
5.1.3 Storage Hints
• Never store a battery in a discharged state.
• Recharge batteries every 6 months.
• Store below 25°C. The ideal range is between 5°C and 21°C.
• If the storage temperature exceeds 25°C during this 6-month period, the shelf life for the batteries is reduced. Be sure to
recharge these batteries periodically.
• Storage place should be free from vibration, dust, direct sunlight, moisture and corrosive gases.
5.1.4 Backup Battery
CAUTION
It is mandatory that the HAMILTON-C2 is operated with at least one battery installed.
NOTICE
- The backup battery is intended for short-term use only. It is not intended to be a primary power source.
- Hamilton Medical recommends that the ventilators batteries be fully charged before you ventilate a patient. If the batteries
are not fully charged and the AC power fails, always pay close attention to the level of battery charge.
5.1.5 Battery Life Expectancy / Li-Ion Battery Maintenance
• The lifetime of a battery is defined by the capacity loss. A Li-ion batteries capacity is lowered due to the number of charge
cycles, as well as elapsed time. Given normal storage and use, you can expect the battery to deliver 75% or more of its initial capacity after 300 charge/discharge cycles or within 2.5 years of its manufacture date.
• Hamilton Medical recommends replacing a battery when the capacity is reduced to 80% of its designed capacity.
(HAMILTON-C2: 6500 mAh).
• Cycle numbers (cy.) and full charge capacity (fc.) are provided on PAGE No. 1102 (See the Lithium Ion Battery Maintenance Section 5.1.6) in the Service software (SW ≥ 1.1.X).
• The battery’s manufacture date is stamped on top of the battery connector (IEYYWWVV). The date is encoded as follows:
In the example IE110221, the year is 2011 (11), the week is (02), and the software version is (2.1).
• The HAMILTON-C2 software version > 1.1.x checks the battery capacity and does not accept weak batteries. Batteries
with a capacity loss of more than 25% generate the error message Battery calibration required.
62
English | 624165/07
5
Lithium Ion Battery Maintenance
5.1.6 Lithium Ion Battery Maintenance
5.1.6.1 Maintenance Hints
• When the battery is used in a device, a calibration cycle on the battery charger (PN 369136) is recommended once per
year during preventive maintenance.
• Periodical use of the batteries can extend the battery life time and allows the battery a calibration cycle each time it is
recharged.
• When a ventilator is not in use, connect it to mains power for 1 hour every week. Make sure to disconnect it afterwards.
5.1.6.2 Warranty
• Batteries are consumable parts. Therefore they are excluded from the standard warranty - excepting out-of-box failures!
5.1.6.3 Disposal
• Regulations vary country by country. Dispose of the batteries in accordance with local regulations.
NOTICE
When the battery reaches its end-of-life, the capacity is reduced and the battery becomes unreliable. The use of a ventilator
with such batteries may cause wrong fuel gauge information leading to an unexpected ventilation stop when a power fail
occurs!
Hamilton Medical | HAMILTON-C2 Service Manual
63
5
Lithium Ion Battery
5.2 Battery Charger / Calibrator
WARNING
• Do not expose the charger or power supply to water or conductive liquids, this is not a sealed case.
• Do not open the charger or power supply case. There are no user serviceable parts inside.
• Do not cover the fan exhaust or obstruct the airflow, as this will cause overheating. Use only the manufacturer’s 24V
2.5A power supply and observe terminal polarity
• Place the charger in a cool spot, away from external heat sources.
CAUTION
During re-calibration, the battery connector and base of the charger may become warm.
The Battery Charger / Calibrator (PN 369104, see Figure 5-2 Battery Charger / Calibrator, please make sure you are using a
charger with Revision 07 or higher) is a standalone desktop battery charger with the added ability to re-calibrate the fuel
gauge on smart batteries. It works with all Hamilton Medical smart batteries.
Figure 5-2. Battery Charger / Calibrator
Package contents:
• One desktop Battery Charger / Calibrator (PN 369104)
• Three plastic spacers to modify the battery recess to suit your battery size
• One 24V 2.5A AC:DC Power Supply, universal mains input
• One mains cable
– European chargers (CH5000E) are packed with a European 2-pin Mains Power Cord
– UK chargers (CH5000U) are packed with a UK 3-pin Mains Power Cord
– North American chargers (CH5000A) are packed with a US 3-pin Mains Power Cord
• Instruction manual
NOTICE
Chargers shipped with medical grade power supplies carry the suffix “M” – eg. CH5000AM.
64
English | 624165/07
Battery Charger / Calibrator Installation
5
5.2.1 Battery Charger / Calibrator Installation
1. Place the charger unit on a flat, level surface away from sources of heat and moisture.
2. Plug the DC connector from the power supply into the connector socket on the back of the charger.
3. Please apply the supplied cable for connecting the battery charger to the AC mains supply.
5.2.2 Battery Charging
1. Place the battery into the battery bay.
2. Make sure that the 5-way connector is fully seated.
NOTICE
The charger will automatically begin to charge the battery.
The LEDs in the status window will provide status information as detailed in the table.
Table 5-1. Battery Lights
Signal Name
Voltage Level
Green flashing
Charging
Green solid
Fully charged
Blue flashing
In calibration mode
Blue solid
Calibration complete
Red flashing
Fuel gauge calibration required
Red solid
Error
5.2.3 Battery Recharge Time
The times given below are for a full charge from 0% to 100% state of charge.
Table 5-2. Battery Recharge Time
Battery Chemistry
Li Ion
Battery Model
MSP369106
Hamilton Medical | HAMILTON-C2 Service Manual
Typical Recharge Time
4.5 hours
65
5
Lithium Ion Battery
5.3 Evaluation of the Battery State of Health
Knowing the Battery State of Health (SoH) is very important to guarantee a safe operation of the ventilator running on battery power. With the following information, you will be able to evaluate the Battery State of Health. With this awareness, you
can decide when a battery needs to be replaced.
You normally perform these checks once a year during Preventive Maintenance.
Read the four battery parameters as follows:
Table 5-3. Evaluation of the Battery State of Health
Parameter
Where to look for:
How to access and how to read:
Full Charge Capacity FCC [mAh]
See the technical state or read the
value in the exported and printed
instrument report.
Go into service software, technical state, the value is
shown on the bottom.
Calendar age AGE [Y]
See printed on battery.
The battery’s date of manufacture is stamped on top
of the battery connector (IEyywwvv). A coding of e.g.
IE144514 means Year 2014, Week 45, Version 1.4,
SN 67, 6.8Ah
Cycles CYC [1]
See the technical state or read the
value in the exported and printed
instrument report.
Go into service software, technical state, value shown
on the bottom.
Error ERR (%)
See the technical state or read the
value in the exported and printed
instrument report.
Go into service software, technical state, the value is
shown on the bottom.
Figure 5-3. Battery State of Health
66
English | 624165/07
Evaluation of the Battery State of Health
5
Compare the values of the battery with the following chart:
If one or more parameters is out of the green range, perform the recommended action.
Figure 5-4. 4 Key Values to Evaluate the Battery State of Health
Hamilton Medical | HAMILTON-C2 Service Manual
67
5
Lithium Ion Battery
5.4 Battery Re-calibration
The battery calibration is done automatically.
If the battery fuel gauge needs re-calibrating, the red LED will flash when you put the battery into the battery charger. This
indicator provides feedback on the accuracy of the fuel gauge and avoids unnecessary battery calibration cycles.
You have the option to calibrate the fuel gauge and charge the battery or to only charge the battery. This option is given
because a recalibration cycle is longer than a charge cycle.
To recalibrate the battery, press the calibrate button on the front of the charger.
The blue calibration LED will flash to indicate that the battery is undergoing the re-calibration cycle. There may be a short
delay before the calibration begins. During calibration the discharge resistors will heat up and the fan will operate to maintain the temperature within acceptable limits.
At the end of the procedure the blue calibration LED will illuminate indicating a fully charged, fully calibrated battery.
NOTICE
No action is necessary if only a recharge is required. The charger will automatically begin to charge the battery.
The most common cause of calibration failure is overheating of the battery during discharge. Keep the charger away from
direct sunlight or heat sources.
CAUTION
During recalibration, the battery connector and base of the charger may become warm.
68
English | 624165/07
5
General Information
5.4.1 General Information
If the battery is in need of fuel gauge re-calibration, the red LED will flash upon insertion of the battery into the battery
charger. This indicator provides feedback to the user on the accuracy of the fuel gauge and avoids unnecessary battery calibration cycles.
The user has the option to calibrate the fuel gauge and charge the battery, or only charge the battery. This option is given
because a re-calibration cycle is longer than a charge cycle.
For re-calibration of the battery, please press the "calibration button" on the front of the charger.
NOTICE
No additional action is required for re-charging the battery. The process starts automatically.
The blue calibration LED will flash to indicate the re-calibration cycle. There may be a short delay before the calibration
begins. During calibration the discharge resistors will heat up and the fan will operate to maintain temperature within acceptable limits.
At the end of this procedure the blue LED will stay constant indicating a fully charged, fully calibrated battery.
CAUTION
The most common cause of calibration failure is overheating of the battery during discharge. Please keep the charger away
from direct sunlight or heat sources.
Hamilton Medical | HAMILTON-C2 Service Manual
69
5
Lithium Ion Battery
5.4.2 Re-calibration Time
The re-calibration cycle begins by discharging the residual capacity. Then a calibration charge is delivered to the battery. This
is followed by a calibration discharge. Finally the battery is given a regular charge. A calibration cycle will be faster if the battery is fully discharged to begin with.
Re-calibration time (5-4) is governed by the battery voltage and capacity. Larger batteries, and low voltage batteries will take
longer to re-calibrate. Calibration is initiated each time the re-calibration button is pressed, so it is not recommended to press
the re-calibration button part way through the re-calibration cycle.
Calibration begins each time the re-calibration button is pressed. Do not press the re-calibration button part way through the
re-calibration cycle.
The re-calibration cycle is as follows:
1.
Discharge of any residual capacity.
2.
A calibration charge is delivered to the battery.
3.
Discharge of calibration capacity.
4.
The battery is given a regular charge.
Table 5-4. Battery Re-Calibration Times
Battery
Li-ion
70
Battery Model
MSP369106
Minimum Recalibration Time
14.6 hours
Maximum Recalibration Time
19.2 hours
English | 624165/07
5
Re-calibration Description
5.4.3 Re-calibration Description
Impedance tracking fuel gauges retain accuracy longer than coulomb-counters and can even self re-calibrate in use as long as
there are periods of inactivity in the cycle so that the impedance measurement can be made more accurately.
If no periods of rest are present in the cycle then the Impedance tracking fuel gauge will accumulate the Max-Error at a rate
of 1% every 20 cycles.
Impedance-Tracking fuel gauge re-calibration is achieved by charging the battery, allowing it to rest, discharging it allowing it
to rest again. This algorithm can only be performed on the external charger/calibrator CH5000 (see the Re-calibration Time
Section 5.4.2).
So re-calibration is used to re-set the fuel gauge to match the actual capacity in the battery. In this way, even as the battery
ages and things change, the accuracy and reliability of the fuel gauge can be retained throughout the life of the battery.
Hamilton Medical | HAMILTON-C2 Service Manual
71
5
Lithium Ion Battery
5.4.4 Rechargeable Lithium Ion Battery Capacity
Figure 5-5. Rechargeable Lithium Ion Battery Capacity
The battery can directly display the capacity information. The battery capacity is displayed as the RSoC (Relative State of
Charge). Each LED segment represents 25 percent of the full charge capacity. The LED pattern definition is given in the table
below. The LED’s illuminate for 4 seconds following switch activation. If the battery voltage is too low, there will be no LED
indication.
Table 5-5. Battery Capacity Information
Capacity
LED Indicators #
1
2
3
Note
4
At or below 10%
x
Blinks
10% - 25%
x
Lit for 4 seconds.
26% - 50%
x
x
51% - 75%
x
x
x
76% - 100%
x
x
x
Lit for 4 seconds.
Lit for 4 seconds.
x
Lit for 4 seconds.
Refer to the HAMILTON-C2 Operator's Manual > Electrical Specifications
72
English | 624165/07
5
Lithium Ion Battery Handling Precautions
5.5 Lithium Ion Battery Handling Precautions
WARNING
• Before using lithium ion batteries for the first time, carefully study this section, including all Cautions and Warnings.
• Keep this section for future reference.
WARNING
• Do not use the battery for a purpose other than those specified. Otherwise, its performance will be compromised and/or
its service life will be shortened. Depending on the equipment in which the battery is used, excessively high current can
flow through the battery, possibly damaging it and leading to acid leakage, overheating, smoke emission, bursting and/
or ignition.
• Do not use the battery in combination with primary batteries (such as dry-cell batteries) or batteries of different capacities or brands. Otherwise, the battery can be overdischarged during use, or overcharged during recharging, abnormal
chemical reactions may occur, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition.
• Do not use or leave the battery near a heat source such as a fire or a heater (+80ºC or higher). If the resin separator
should be damaged due to overheating, internal short-circuiting may occur to the battery, possibly leading to acid leakage, smoke emission, bursting and/or ignition.
• Do not put the battery into a microwave oven or pressurized container. Rapid heating or disrupted sealing can lead to
acid leakage, overheating, smoke emission, bursting and/or ignition.
• Do not discard the battery into fire or heat it. Otherwise, its insulation can melt down, its gas release vent or safety features will be damaged and/or its electrolyte can ignite, possibly leading to acid leakage, overheating, smoke emission,
bursting and/or ignition on it.
• Do not immerse the battery in liquid, or allow it to get wet. Otherwise, the protective features in it can be damaged, it
can be charged with extremely high current and voltage, abnormal chemical reactions may occur in it, possibly leading to
acid leakage, smoke emission, bursting and/or ignition.
• Do not pierce the battery with a nail or other sharp objects, strike it with a hammer, or step on it. Otherwise, the battery
will become damaged and deformed, internal short-circuiting can occur, possibly leading to acid leakage, overheating,
smoke emission, bursting and/or ignition.
• Do not strike or throw the battery. The impact might cause leakage, overheating, smoke emission, bursting and/or ignition. Also, if the protective feature in it becomes damaged, it could become charged with an extremely high current and
voltage, abnormal chemical reactions can occur, which can lead to acid leakage, overheating smoke emission, bursting
and/or ignition.
• Do not disassemble or modify the battery. The battery is equipped with built-in safety/protection features. Should these
features be disabled, the battery can leak acid, overheat, emit smoke, burst and/or ignite.
• If the battery leaks and the electrolyte gets into the eyes, do not rub them. Instead, rinse the eyes with clean running
water and immediately seek medical attention. Otherwise, eye injury may result.
• Do not use an apparently damaged or deformed battery. Otherwise, acid leakage, overheating, smoke emission, bursting
and/or ignition of the battery may occur.
• Consider consolidating these three into: Do not expose the battery to fire, flames, or excessive heat such as is generated
by soldering, welding, radiators and so on. If the battery gives off a bad odor, melts, becomes discolored or deformed,
leaks electrolyte fluid, appears in any way abnormal at any time, remove it from the equipment as the battery can leak
acid (electrolyte fluid), emit smoke, burst and/or ignite.
• Do not reverse the positive (+) and negative (-) terminals. Otherwise, during recharging, the battery will be reversecharged, abnormal chemical reactions then may occur, or excessively high current can flow during discharging, leading
to acid leakage, overheating, smoke emission, bursting and/or ignition.
• The positive (+) and negative (-) terminals are arranged in a particular orientation. Do not force the connection if you
cannot easily connect the battery terminals to the battery charger or other equipment. Confirm that the terminals are
correctly oriented. Reversing the terminals will result in reverse-charging, possibly leading to acid leakage, overheating,
smoke emission, bursting and/or ignition.
• Do not connect the positive (+) and negative (-) terminals with a metal object such as wire. Do not transport or store the
battery together with metal objects such as necklaces, hair pins, etc. Otherwise, short-circuiting will occur, over-current
will flow, causing the battery to leak acid, overheat, emit smoke, burst and/or ignite or the metal object such as wire,
necklace or hair pin can generate heat.
• Do not connect the battery to an electrical outlet, vehicle cigarette lighter, etc. When subjected to large voltage, overcurrent can flow on the battery, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition.
• Do not recharge the battery near fire or above 45°C. Otherwise, hot temperatures can trigger its built-in protective features, inhibiting recharging or can damage the built-in protective features, causing it to be charged with an extremely
Hamilton Medical | HAMILTON-C2 Service Manual
73
5
Lithium Ion Battery
high current and voltage and, as a result, abnormal chemical reactions can occur in it, possibly leading to acid leakage,
overheating, smoke emission, bursting and/or ignition.
• To recharge the battery, use the battery charger specifically designed for the purpose and observe the recharging conditions, refer to 5.2. A recharging operation under non-conforming recharging conditions (higher temperature and larger
voltage/current than specified, modified battery charger, etc.) can cause the battery to be overcharged, or charged with
extremely high current. An abnormal chemical reaction can occur in the battery, possibly leading to acid leakage, overheating, smoke emission, bursting and/or ignition.
• If recharging operation fails to complete even when a specified recharging time has elapsed, immediately stop further
recharging. Otherwise, acid leakage, overheating, smoke emission, bursting and/or ignition can occur.
CAUTION
• If you find rust, a bad odor, overheating and/or other irregularities when using the battery for the first time, return it to
Hamilton Medical AG.
• If acid leaking from the battery comes into contact with your skin or clothing, immediately wash it away with running
water. Otherwise, skin inflammation can occur.
• The battery incorporates built-in safety devices. Do not use it in a location where static electricity (greater than the manufacturer’s guarantee) may be present. Otherwise, the safety devices can be damaged, possibly leading to acid leakage,
overheating, smoke emission, bursting and/or ignition.
• Do not use or subject the battery to intense sun or above 65°C. Otherwise, acid leakage, overheating and/or smoke
emission can occur. Also, its guaranteed performance will be lost and/or its service life will be shortened.
• The guaranteed recharging temperature range is 0°C to +45ºC. A recharging operation outside this temperature range
can lead to acid leakage and/or overheating of the battery and may cause damage to it.
• Store the battery in a location where children cannot reach it. Also, make sure that no child takes the battery out of the
battery charger or equipment.
74
English | 624165/07
6
Preventive Maintenance and Testing
6.1
Overview.............................................................................................................................................. 76
6.2
Hospital Preventive Maintenance .......................................................................................................... 78
6.3
Engineer Preventive Maintenance ......................................................................................................... 79
75
6
Preventive Maintenance and Testing
6.1 Overview
WARNING
− To prevent disease transmission, you must use personal protective equipment when handling contaminated bacterial filters or other patient accessories. Refer to the HAMILTON-C2 Operator’s Manual for instructions on sterilizing patient system
parts.
− Device is potentially contaminated
− The device must be cleaned and disinfected to prevent the spread of infections and germs
CAUTION
Make sure to take full ESD (ElectroStatic Discharge) precautions before handling any electronic parts, or before opening the
HAMILTON-C2. For more information refer to:
− Electrical Safety Overview Section 7.1
− Standard Tools Section 11.2
− Special Tools Section 11.3
− Test Equipment Section 11.4
− You must complete a service training course for the HAMILTON-C2 with Hamilton Medical before undertaking the maintenance and testing procedures described in this manual.
6.1.1 Introduction
All Preventive Maintenance and Testing must be performed:
• After replacing any component
• Once a year or once every 5000 operating hours, whichever comes first
To perform Preventive Maintenance, perform all the steps shown in the Overview of Preventive Maintenance and Testing Section 6.1.4.
6.1.2 Checking the Software Level
In general, Hamilton Medical AG recommends updating to the latest Software Version available. See the Hamilton Medical
AG Partner-net.
76
English | 624165/07
Items Required for Preventive Maintenance and Testing
6
6.1.3 Items Required for Preventive Maintenance and Testing
Table 6-1. Items Required for Preventive Maintenance and Testing
Step
Items Required, or Possibly Required
Hospital Preventive Maintenance Section 6.2
• The HAMILTON-C2 Operator’s Manual or local-language equivalent
Engineer Preventive Maintenance Section 6.3
• HEPA Filter (PN 160216)
• Filter Set (containing 2 Dust Air Filter and 1 Fan Filter (PN 160215))
• Blower Module (MSP160250)
• Lithium Ion Battery Pack(s) (MSP369106)
• Battery Charger / Calibrator (PN 369104, Rev 07 or higher)
• Oxygen Sensor (PN 396200)
• HPO Inlet Filter Kit (PN 160497)
• Complete Patient Breathing Circuit (Adult, PN 260086):
– Test Lung with ET-Tube (PN 151815) and Adapter (PN 281420)
– Inspiratory Filter (PN 279204)
– Short Silicone Tube (PN 260100)
NOTICE! For more details of parts, refer to the Spare Parts Section
12.
Electrical Safety Tests Section 7.2
• Test equipment for the HAMILTON-C2; See the Hamilton Medical AG Partnernet > Technical Support > Recommended Spare Parts List > Recommended Test
Equipment
Service Software Section 8
NOTICE! For more details of Test Equipment, refer to the Test Equipment Section
11.4.
6.1.4 Overview of Preventive Maintenance and Testing
Work methodically through the sections shown in the Overview of Preventive Maintenance and Testing Table 6-2 below.
Maintenance and testing is not complete until all steps are performed successfully.
If you make a replacement, you must go back to “Step 3” in the Overview of Preventive Maintenance and Testing Table 6-2
below.
Table 6-2. Overview of Preventive Maintenance and Testing
Step
Task
Where Found
Time Required
1.
Perform (or confirm it has been performed) the Hospital Preventive Maintenance
Hospital Preventive Maintenance Section 6.2
5-15 min.
2.
Perform the Engineer Preventive Maintenance
Engineer Preventive Maintenance Section 6.3
10 min.
3.
Perform parts replacements as necessary
Components Removal/ Assembly Section 10
N/A
4.
Perform the Electrical Safety Tests
Electrical Safety Tests Section 7.2
10 min.
5.
Perform the Service Software Checks
Service Software Section 8
40 min.
6.
Finish the testing by completing the tasks documented in the Tests, Calibrations and Utilities Section
of the HAMILTON-C2 Operator’s Manual
The HAMILTON-C2 Operator’s Manual
or local-language equivalent
10 min.
Hamilton Medical | HAMILTON-C2 Service Manual
77
6
Preventive Maintenance and Testing
6.2 Hospital Preventive Maintenance
The Hospital Preventive Maintenance Table 6-3 shows the maintenance tasks that hospital staff must perform. It is copied
from the English Version of the Operator’s Manual.
Examine each ventilator for which you are responsible, and satisfy yourself that hospital staff are regularly performing these
tasks.
If necessary:
• Perform the tasks yourself
• Train staff how to perform these tasks
Table 6-3. Hospital Preventive Maintenance
Interval
Between patients and according to hospital policy
Part/Accessory
Procedure
Patient Breathing Circuit (including
Mask, Inspiratory Filter, Flow Sensor,
Nebulizer Jar, Expiratory Valve and
Membrane)
Replace with sterilized or new single use parts. Run
the Tightness Test and the Flow Sensor Calibration as
shown in the HAMILTON-C2 Operator’s Manual.
Entire ventilator
Run the pre-operational check as shown in the
HAMILTON-C2 Operator’s Manual.
Every 2 days or according to hospital
policy
Patient Breathing Circuit
Empty any water from breathing tubes or water
traps. Inspect parts for damage. Replace as necessary.
Every month (or more often, if
required)
Air Intake Dust Filter and Fan Filter Set
(5 pieces) (rear panel)
Check for dust and lint. If needed, clean or replace as
shown in the HAMILTON-C2 Operator’s Manual.
Every 6 months (while the ventilator is
in storage)
Battery
Recharge battery by plugging the Ventilator into AC
Power for at least 4 hours.
78
English | 624165/07
6
Engineer Preventive Maintenance
6.3 Engineer Preventive Maintenance
WARNING
This section is not a stand-alone, independent part of the manual. Perform the tasks detailed here only as a part of, and as
instructed. Refer to the Preventive Maintenance and Testing Overview Section 6.1 and Hospital Preventive Maintenance Section 6.2.
Perform the Engineer Preventive Maintenance according to the table below:
Table 6-4. Engineer Preventive Maintenance Schedule
Interval
Yearly or every 5000 hours,
whichever comes first, or as necessary
Part/Accessory
Procedure
Ventilator
Perform service-related Preventive Maintenance.
Calibrations and Tests
Perform all calibrations and tests in the Service Software. (Refer to
the Service Software Introduction Section 8.1).
Pre-Operational Checks
Perform the Pre-operational Checks as shown in the HAMILTON-C2
Operator’s Manual > Running the Pre-operational Check.
Alarm Tests
Perform the alarm checks as shown in the HAMILTON-C2 Operator’s Manual > Alarm Tests.
HEPA Filter
Replace as described in the HAMILTON-C2 Operator's Manual >
Servicing the Air Intake and Fan Filters.
HPO Inlet Filter
Replace as described in the High Pressure Oxygen Connector
Removal/Assembly Section 10.4.4.
Oxygen Sensor
Replace if depleted as shown in the HAMILTON-C2 Operator’s
Manual Section > Replacing the Oxygen Sensor.
Lithium Ion Battery
Replace the Lithium Ion Battery (Refer to Replacing the Battery Section 5.1.2).
Every 5 years (30,000 Hours)
LCD Display Backlight
Replace the LCD Display Backlight as described (Refer to LCD Backlight Removal/Assembly Section 10.3.8).
When ebt or blower timer is ≥
100%.
Blower Module
Replace the Blower Module when the Blower Timer has reached
100%(dynamic lifetime surveillance, typically after 8 years). Also
replace the Blower Module when the prediction of Blower Timer in
the next 6 months reaches 100% (Refer to Blower Module
Removal/Assembly Section 10.4.10).
Battery to be checked. Replace
battery when (whichever comes
first):
• MSP369106 when capacity <
5280 mAh
• Battery age > 3 Years
• Cycles > 400
Hamilton Medical | HAMILTON-C2 Service Manual
79
6
80
Preventive Maintenance and Testing
English | 624165/07
7
Electrical Safety
7.1
Electrical Safety Overview ..................................................................................................................... 82
7.2
Electrical Safety Tests ........................................................................................................................... 82
81
7
Electrical Safety
7.1 Electrical Safety Overview
WARNING
Electrical Safety Tests detailed in this section must be performed as part of, or as instructed in the Overview of Preventive
Maintenance and Testing Section 6.1.4.
In addition, to comply with IEC 62353, the Electrical Safety Test's must be performed after:
− The Power Supply is replaced
− The Mainboard is replaced
− Removing any ground contact from the Ventilator
− When performing Preventive Maintenance
− Any repair
− Installation
Hamilton Medical AG performs a set of Electrical Safety Tests, as specified in IEC 60601-1, on all the ventilator and compressor units that it manufactures. Hamilton Medical AG performs these tests automatically, using the RIGEL 288 Safety Analyzer
or a device like Metron QA90.
As stated in the warning above, it is a legal necessity that, after performing the Preventive Maintenance, after a repair or
after an adjustment described above, the Electrical Safety Tests are performed.
7.2 Electrical Safety Tests
7.2.1 Preparation for Tests
• Multimeter
• Safety Analyzer
• Electrical Safety Test Cable
Figure 7-1. RIGEL 288 Safety Analyzer
82
English | 624165/07
Perform the Tests
7
7.2.2 Perform the Tests
The Electrical Safety Tests you must perform are explained in this section. If you have an automated safety device such as the
RIGEL 288 Safety Analyzer used by Hamilton Medical, perform the automated tests in addition to the tests shown in this section.
7.2.3 Device Type
For the IEC 62353 Regulations concerning medical devices, the HAMILTON-C2 is a:
• Class 2
• Type B Device
• Type BF if the Communication Board for Capnography is installed
Hamilton Medical | HAMILTON-C2 Service Manual
83
7
Electrical Safety
7.2.4 Internal Resistance Checks
For these tests, a Digital Voltmeter is required (refer to Digital Voltmeter Section 11.3.1)
Check the resistance between the Earth Ground on the HAMILTON-C2 with the Test Connector, and other components as
shown in Table 7-1 Electrical Tests.
Table 7-1. Electrical Tests
Check
Minimal Acceptable
Resistance
Figure
A. Ground (earth) pin to:
Communications Interface Connector
> 2.4MΩ
See Figure 7-2
B. Ground (earth) pin to:
Interaction Panel
> 2.4MΩ
See Figure 7-3
C. Ground (earth) pin to:
High Pressure Oxygen
> 2.4MΩ
See Figure 7-4
D. Ground (earth) pin to:
Low Pressure Oxygen
> 2.4MΩ
See Figure 7-5
Figure 7-2. Earth Ground to the Communication Interface Connector Resistance
84
English | 624165/07
Internal Resistance Checks
7
Figure 7-3. Earth Ground to the Interaction Panel Resistance
Figure 7-4. Earth Ground to the High Pressure Oxygen Resistance
Hamilton Medical | HAMILTON-C2 Service Manual
85
7
Electrical Safety
Figure 7-5. Earth Ground to the Low Pressure Oxygen Resistance
86
English | 624165/07
Electrical Safety Tests - IEC 62353
7
7.2.5 Electrical Safety Tests - IEC 62353
It is a legal necessity, after a repair or adjustment that includes replacing the Power Supply, Mainboard or removing any of
the internal earth connectors, perform an Electrical Safety Test on the ventilator. The Electrical Safety Test is performed with
suitable equipment such as the RIGEL 288 Safety Analyzer (See Figure 7-6 RIGEL 288 Safety Analyzer). Details of these Automated Electrical Safety Tests depend on the test equipment used, it is impossible to offer any detailed descriptions.
Figure 7-6. RIGEL 288 Safety Analyzer
Specifications
Table 7-2. IEC 62353 Specifications for Class II Type B and BF Devices
Current in mA (RMS - Root Mean Square)
Setup
Equipment leakage (direct method)*
Patient leakage current (direct method) AC*
Type B
Setup for Type B Device (no Communication Board Installed) Section 7.2.5.1
Type BF
Setup for Type BF Device (with Communication Board for Capnography)
Section 7.2.5.2
0.1 mA
0.1 mA
–
5 mA
* refer to IEC62353
All measured values must be documented using the measurement process and are considered reference values. Should the
obtained values measured during the next maintenance cycle be within 90 to 100% of the permissible values, the reference
values are to be used to evaluate the ventilators electrical safety.
Definition
Applied Parts of the medical device, which are designed to come into physical contact with the patient or parts that are likely
to be brought into contact with the patient.
Type B Applied Parts
Type B Applied Parts are those parts, which are usually Earth referenced. Type B is used for applied parts that are generally
not conductive and can be immediately released from the patient.
Type BF Applied Parts
Type BF Applied Parts are generally for devices that have conductive contact with the patient, or having medium or long term
contact with the patient. For example, using capnostat and/or SPO2 finger probe on a patient. This applied part complies
with a higher degree of protection against electric shock than Type B applied part.
Hamilton Medical | HAMILTON-C2 Service Manual
87
7
Electrical Safety
7.2.5.1 Setup for Type B Device (no Communication Board Installed)
NOTICE
For detailed information regarding the use of your Electrical Safety Tester, please check the Users Manual for the device.
1. For setup for the RIGEL 288 Safety Analyzer Tests, attach the Ground Cable (E) from the USB Port (G) of the HAMILTONC2 to the 4mm Earth Bond Probe Socket of the RIGEL 288 Safety Analyzer (see Figure 7-7).
2. Connect the AC Mains Cable from the HAMILTON-C2 to the AC Connection at the front of the RIGEL 288 Safety Analyzer
(see Figure 7-7).
3. Turn on the HAMILTON-C2 (Standby mode) for the electrical safety tests.
Figure 7-7. Setup for Type B Device (no Communication Board Installed)
A
AC Power Cable Inlet (RIGEL 288 Safety Analyzer)
E
Electrical Safety Test Cable (PN 161618)
B
RIGEL 288 Safety Analyzer (Handheld Medical Electrical
Safety Tester)
F
Power Cord (HAMILTON-C2)
C
4mm Earth Bond Probe Socket (green) (RIGEL 288 Safety
Analyzer)
G
USB Port
D
EUT Socket (RIGEL 288 Safety Analyzer)
88
English | 624165/07
7
Setup for Type BF Device (with Communication Board for Capnography)
7.2.5.2 Setup for Type BF Device (with Communication Board for Capnography)
NOTICE
For detailed information regarding the use of your Electrical Safety Tester, please check the Users Manual for the device.
1. Connect the RIGEL 288 Safety Analyzer to the AC Power Supply.
2. Connect the RIGEL 288 Safety Analyzer and the HAMILTON-C2 to each other as shown below (see Figure 7-8 Setup for
Type BF Device (with Communication Board for Capnography).
3. Turn on the HAMILTON-C2 (Standby mode) for the electrical safety tests.
Figure 7-8. Setup for Type BF Device (with Installed Communication Board for Capnography)
A
AC Power Cable Inlet (RIGEL 288 Safety Analyzer)
G
Communication Board
B
RIGEL 288 Safety Analyzer (Handheld Medical Electrical
Safety Tester)
H
Power Plug
C
4 mm Earth Bond Probe Socket (green) (RIGEL 288 Safety
Analyzer)
I
USB Port
D
EUT Socket (RIGEL 288 Safety Analyzer)
J
CO2 Sensor Electrical Safety Test Cable (PN 159171)
E
Power Cord (HAMILTON-C2)
K
SPO2 (Not Implemented in HAMILTON-C2) Sensor Electrical
Safety Test Cable (PN 159681)
F
Electrical Safety Test Cable (PN 161618)
L
AP-Box 331A700 (RIGEL 288 Safety Analyzer)
Hamilton Medical | HAMILTON-C2 Service Manual
89
7
90
Electrical Safety
English | 624165/07
8
Service Software
8.1
Introduction ......................................................................................................................................... 92
8.2
Functions of the Service Software......................................................................................................... 93
8.3
Structure of the Service Software ......................................................................................................... 94
8.4
Starting the Service Software ............................................................................................................... 95
8.5
Service Software Screen Layout ............................................................................................................ 96
8.6
Making Screenshots ............................................................................................................................. 97
8.7
Ventilator Info...................................................................................................................................... 98
8.8
Technical State..................................................................................................................................... 99
8.9
Service Timer........................................................................................................................................105
8.10
Real Time Clock (RTC) ..........................................................................................................................108
8.11
Battery Data.........................................................................................................................................109
8.12
Tests Overview .....................................................................................................................................110
8.13
Tests / Calibration Screen .....................................................................................................................111
8.14
Adjustments / Calibration.....................................................................................................................112
8.15
Component Tests.................................................................................................................................142
8.16
System Test..........................................................................................................................................194
8.17
Sensor Data .........................................................................................................................................207
8.18
Log/Config Files ...................................................................................................................................212
8.19
Software Updates ................................................................................................................................218
8.20
Pre-operational Checks ........................................................................................................................226
8.21
General Tests .......................................................................................................................................227
8.22
CO2 Sensor Accuracy Check ................................................................................................................230
8.23
Nurse Call Functional Check.................................................................................................................240
8.24
Final Tests ............................................................................................................................................242
8.25
Set the Service Timer ............................................................................................................................242
8.26
Check the Battery Data ........................................................................................................................242
91
8
Service Software
8.1 Introduction
WARNING
− Read the Preventive Maintenance and Testing Overview Section 6.1, before performing any of the tests in this section.
− If one of the tests indicates that you must replace a part, do so immediately and update the Service Software Modify Tab
(see the Modify Tab Section 8.8.3) and then repeat the complete series of tests. See contents in this section.
CAUTION
To prevent patient or ventilator contamination, always use a bacterial filter between the HAMILTON-C2 and the Inspiratory
Limb of the Patient Breathing Circuit.
− For troubleshooting see knowledge base
− Record all results on the HAMILTON-C2 Summary of Test Report (see the Summary of Test Report Section 17.1)
− Confirm that the Technical State is updated and that the device has been restarted after Technical State modifications
(see the Modify Tab Section 8.8.3). Always use an Adult Breathing Circuit, if a circuit is required during the tests.
NOTICE
The HAMILTON-C2 needs a warm-up period. Make sure it was running for at least 20 minutes in the Ventilation Mode.
This section describes each of the units containing the HAMILTON-C2 Service Software.
Before starting, be sure that you are familiar with Typographic Conventions Section 1.2 and Expressions Section 1.3.
The HAMILTON-C2 Summary of Test Report is the standard form to be used and must be completed each time the Service
Software is performed. If you do not have a suitable form, you can photocopy and use the form named HAMILTON-C2 Summary of Test Report (see the Summary of Test Report Section 17.1) at the back of this manual.
92
English | 624165/07
Functions of the Service Software
8
8.2 Functions of the Service Software
Units in the Service Software perform the following functions:
• Enables display information (concerning revisions and versions of the HAMILTON-C2 Hardware and Software)
• Enables checks on the HAMILTON-C2 Hardware and Software
• Enables calibration of the HAMILTON-C2 Hardware
• Enables viewing and exporting of the Event Log and Service Log
• Enables software upgrades
Hamilton Medical | HAMILTON-C2 Service Manual
93
8
Service Software
8.3 Structure of the Service Software
Figure 8-1. Service Software Structure
94
English | 624165/07
8
Starting the Service Software
8.4 Starting the Service Software
NOTICE
To start the Service Software, you must place the HAMILTON-C2 into the Service Software Mode.
1. Connect the HAMILTON-C2 to mains power.
2. Press the ON (A) Button located on the HAMILTON-C2 Interaction Panel and then press and hold the 100% O2 (B) and
Manual Breath (B) Buttons at the same time. The 2 buttons can be released as soon as the Start Service Software message
appears below the progress bar.
If a Mainboard Revision 15 or higher is installed, the 100% O2 and Manual Breath Buttons can be pressed shortly. Press and
hold the 100% O2 (B) and Manual Breath (B) Buttons at the same time.
Figure 8-2. Starting the Service Software
3. After the Service Software starts, the Main Service Software Screen is displayed.
Figure 8-3. The HAMILTON-C2 Main Service Software Screen
4. To exit the Service Software, switch OFF the HAMILTON-C2.
Hamilton Medical | HAMILTON-C2 Service Manual
95
8
Service Software
8.5 Service Software Screen Layout
Figure 8-4. Service Software Screen Layout
Table 8-1. Service Software Screen Layout
1.
Header Window
3c.
Optional Menu Tab Layer 3
2.
Test Window including Dialogs
4.
Alarm Window
3a.
Menu Tab Layer 1
5.
Energy State Window
3b.
Optional Menu Tab Layer 2
96
English | 624165/07
Making Screenshots
8
8.6 Making Screenshots
1. Insert the USB Memory Stick (B, PN 396207) into the HAMILTON-C2 USB Slot and press the Print Screen Button (A) for 1
second until the corresponding LED lights up.
Figure 8-5. Making Screenshots
The print screen function saves a JPG file of the current ventilator screen to a USB Memory Stick (B).
The USB Memory Stick can be removed when the Print Screen Key LED is no longer lit.
The filename takes this format:
• screenshot_yyyymmdd_HHMMss.jpg
where:
• yyyy is the year
• mm is the month
• dd is the day
• HH is the hour (in 24-hour format)
• MM is the minute
• ss is the second
NOTICE
Specification for the USB memory stick:
File system: FAT or FAT32
Unpartitioned memory
No operating system or security software installed
Hamilton Medical | HAMILTON-C2 Service Manual
97
8
Service Software
8.7 Ventilator Info
1. From the Main Service Software Screen, touch the Ventil. Info Button.
Figure 8-6. Main Service Software Screen
2. On the Ventilator Info Screen are the:
• Technical State Button
• Service Timer Button
• Real Time Clock (RTC) Button
• Battery Data Button
• Back Button (go back to the Main Menu)
Figure 8-7. The Ventilator Info Screen
98
English | 624165/07
Technical State
8
8.8 Technical State
Touch the Technical State Button to enter the technical state section.
Figure 8-8. Ventilator Info Screen
Hamilton Medical | HAMILTON-C2 Service Manual
99
8
Service Software
8.8.1 Hardware Version Tab
1. Touch the HW Version Tab (if not already selected). The HW Version Tab displays the Device Name, Part Number, Revision,
Serial Number and Timing Information.
2. The information of the following components will be updated automatically from the HAMILTON-C2 during start-up:
• Battery 1
• Battery 2
• Flow Sensor AIR (Qvent)
• Flow Sensor O2 (QO2)
• Oxygen Sensor
• Communication Board
Figure 8-9. Hardware Version Tab
3. In case the technical state is corrupt, or after upgrading from SW version < 2.1.2, the Clean-up Button will appear.
4. In case the technical state cannot be read out from the EEprom of a component, the unit will fail in self-test and alarm
with “Technical state failed”. Other Technical faults will appear as after effect.
5. Record the Hardware Version Information on the HAMILTON-C2 Summary of Test Report Section 17.1.
100
English | 624165/07
8
Software Version Tab
8.8.2 Software Version Tab
1. Touch the SW Version Tab. The Software Version Tab displays the device names and revisions of the operating software.
Figure 8-10. Software Version Tab
2. Record the Software Version Information on the HAMILTON-C2 Summary of Test Report Section 17.1.
Hamilton Medical | HAMILTON-C2 Service Manual
101
8
Service Software
8.8.3 Modify Tab
NOTICE
Always update the technical state when an assembly as listed below has been replaced.
Restart the device prior to performing the service software tests and calibrations.
The Modify Tab allows updating information when a part has been replaced.
1. Touch the Modify Tab.
Figure 8-11. Modify Tab Step 1
2. Select the hardware component you want to modify.
3. To change the serial number or revision number, touch the corresponding button.
Figure 8-12. Modify Tab Step 2
102
English | 624165/07
8
Modify Tab
4. Enter the number.
Figure 8-13. Modify Tab Step 3
5. Proceed with Confirm. When finished, touch the Save Button.
6. To change the Part Number, touch the Part No. Button and select the correct number.
7. Proceed with Confirm. When finished, touch the Save Button.
Figure 8-14. Modify Tab Step 4
NOTICE
Make sure to select the correct Part Number. Cross check with the label of the Spare Part / Installation Guide of the MSP
Spare Part.
Hamilton Medical | HAMILTON-C2 Service Manual
103
8
Service Software
8.8.4 Transfer Tab
The Transfer Tab contains 2 buttons:
Export and Import
To allow you to export/import the technical state of the device from a USB Memory Stick.
Figure 8-15. Transfer Tab
By exporting, the service software creates a directory, if it does not exist, on the USB Memory Stick.
The naming convention of this directory is:
INSTRUMENT-sn#device_serial_number: /myUSBDevice/C2-sn-1010/ the file which is exported is a .csv (comma separated file)
file containing the device technical state.
It has the same naming convention as the directory created on the USB Memory Stick:
e.g.: C2-sn-1010.csv
For importing values to the device, be sure that there is a directory following the previous given naming convention and
a .csv file having the proper name according to the device serial number.
The technical state can be exported and imported to/from a USB Memory Stick. Ensure the USB Memory Stick is connected to
the ventilator, then touch the corresponding button.
NOTICE
Insert USB Memory Stick is visible, as long as there is no memory stick inserted.
NOTICE
The format of the file is .csv (comma separated values) and can be edited with a text editor (e.g. Notepad, Wordpad). Do
not use Excel to modify the .csv file.
104
English | 624165/07
8
Service Timer
8.9 Service Timer
8.9.1 Service Timer Tab
1. Touch the Service timer Tab.
2. On the Service timer Tab, the total Operating hours are displayed and the Service timer hours are displayed since the last
time the Service timer was previously reset.
Figure 8-16. Service Timer Tab
3. Record the Operating hours and the Service timer hours on the HAMILTON-C2 Summary of Test Report Section 17.1.
4. After successful service, reset the service counter by touching the Reset Button. To enable the reset function, select the
Alarm limit Timer first.
5. The Alarm limit Timer Button allows the set number of hours between service intervals to be changed.
Figure 8-17. Setting the Alarm Limit
Hamilton Medical | HAMILTON-C2 Service Manual
105
8
Service Software
6. Touch the Alarm limit Timer Button or rotate the P&T Control Knob until the Alarm Limit Rotary Screen Button is highlighted, then press the P&T Control Knob.
7. The number of hours can be changed by rotating the P&T Control Knob.
8. Touch the Alarm limit Timer Button again or press the P&T Control Knob to save the new value.
Figure 8-18. Alarm Limit Reset Button
9. Touch Reset.
10. After resetting the service timer, the current operating hours will be added to the set alarm limit hours.
106
English | 624165/07
8
Blower Timer Tab
8.9.2 Blower Timer Tab
NOTICE
The expected operation time (100%) is calculated based on the operating hours, temperature and rotation speed of the
blower. When 100% is reached, the alarm Blower service required appears and the Blower Module needs to be replaced.
Replace the Blower Module when the blower timer ≥ 100%. Also replace the Blower Module when the Prediction Blower
Timer in the Next 6 Months reaches 100%. See prediction calculation formula in the table below.
After replacement of the Blower Module, enter the Serial Number and Revision Number of the replacement blower in the
Technical State (see the Modify Tab 8.8.3). This will reset the Blower timer.
Table 8-2. Prediction of the blower timer in 6 months:
Where:
ebt = expected blower timer in 6 months [%]
bt = current blower timer [%]
y = number of years since the installation of the blower [Year(s)]
1. Touch the Blower timer Tab.
2. On the Blower timer Tab, the total blower time hours are displayed in percentage.
Figure 8-19. Blower Timer Tab
3. Record the Blower timer % on the HAMILTON-C2 Summary of Test Report Section 17.1.
Hamilton Medical | HAMILTON-C2 Service Manual
107
8
Service Software
8.10 Real Time Clock (RTC)
1. Check the current Date and Time:
Figure 8-20. Current Date and Time Reset Step 1
2. Touch the Screen Button(s) that need to be changed, or rotate the P&T Control Knob until the desired button is highlighted, then press the P&T Control Knob.
3. Change the number displayed by rotating the P&T Control Knob.
4. Touch the Screen Button again or press the P&T Control Knob to keep the new value.
5. When all selections are complete, touch the Set Button.
Figure 8-21. Current Date and Time Reset Step 2
6. To change to the Main Menu, touch the Back Button.
108
English | 624165/07
8
Battery Data
8.11 Battery Data
The Battery Data Button displays all of the data provided for the batteries (see the Evaluation of the State of Health Section
5.3).
Figure 8-22. Battery Data
Hamilton Medical | HAMILTON-C2 Service Manual
109
8
Service Software
8.12 Tests Overview
Table 8-3. Adjustments / Calibration
Tests
Screenshots
Page No
Touch Screen
Page No 2321
Touch Screen Section 8.14.1.1
Inspiratory Valve
Page No 2342
Inspiratory Valve Section 8.14.1.2
Pressure
Page No 2341
Pressure Section 8.14.1.3
Expiratory Valve
Page No 2343
Expiratory Valve Section 8.14.1.4
Oxygen Sensor
Page No 2346
Oxygen Sensor Section 8.14.1.5
Flow Sensor
Page No 2347
Flow Sensor Section 8.14.1.6
Table 8-4. Components Tests
Tests
Screenshots
Page No
Alarm System
Page No 2102
Alarm System Section 8.15.1.1
Alarm Monitor 1
Page No 2113
Alarm Monitor 1 Section 8.15.1.2
Alarm Monitor 2
Page No 2114
Alarm Monitor 2 Section 8.15.1.3
User Interface Tab
Page No 2115
User Interface Tab Section 8.15.2
Binary Valve
Page No 2106
Binary Valve Section 8.15.3.1
Autozero
Page No 2109
Autozero Section 8.15.3.2
Blower Flow
Page No 2104
Blower Flow Section 8.15.3.3
Blower Pressure
Page No 2105
Blower Pressure Section 8.15.3.4
Expiratory Valve
Page No 2111
Expiratory Valve Section 8.15.3.5
Inspiratory Valve
Page No 2107
Inspiratory Valve Section 8.15.3.6
O2 Input
Page No 2112
O2 Input Section 8.15.4.1
Nebulizer Valve
Page No 2116
Nebulizer Valve Section 8.15.4.2
Ambient Valve
Page No 2108
Ambient Valve Section 8.15.4.3
Proximal Flow
Page No 2110
Proximal Flow Section 8.15.4.4
Air Entry
Page No 2117
Air Entry Section 8.15.4.5
Table 8-5. System Tests
Tests
Screenshots
Page No
Pressure
Page No 2201
Pressure Section 8.16.1
Flow
Page No 2202
Flow Section 8.16.2
O2 Mixer
Page No 2203
O2 Mixer Section 8.16.3
Tightness
Page No 2204
Tightness Section 8.16.4
Alarming
Page No 2205
Alarming Section 8.16.5
110
English | 624165/07
Tests / Calibration Screen
8
8.13 Tests / Calibration Screen
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
Figure 8-23. Main Service Software Screen
2. On the Tests / Calibration Screen are the:
• Adjustments / Calibration Button
• Component Test Button
• System Test Button
• Sensor Data Button
• Back Button
Figure 8-24. Tests / Calibration Screen
Hamilton Medical | HAMILTON-C2 Service Manual
111
8
Service Software
8.14 Adjustments / Calibration
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
Figure 8-25. Main Service Software Screen
2. Touch the Adjustment / Calibration Button.
Figure 8-26. Adjustment / Calibration Screen
112
English | 624165/07
Calibration Tab
8
8.14.1 Calibration Tab
Touch the Calibration Tab.
Figure 8-27. Calibration Tab
Hamilton Medical | HAMILTON-C2 Service Manual
113
8
Service Software
Table 8-6. Adjustments / Calibration
Adjustments / Calibration
Adj/Calib > Calibration > Touch Screen
(Page No 2321)
Description
Component Adjusted /
Calibrated
Allows calibrating the coordinates of the touch controller (4–point
calibration).
• Touch Screen
Allows aligning the gain and offset for voltage-offset of the inspiratory valve.
• Inspiratory Valve
Touch Screen Section 8.14.1.1
Adj/Calib > Calibration > Inspiratory
Valve (Page No 2321)
Inspiratory Valve Section 8.14.1.2
Sets Offset 1 (low pressure) at 20 mbar blower pressure and Offset
2 (high pressure) at 60 mbar blower pressure when required Qvent
Flow is reached.
Adj/Calib > Calibration > Pressure (Page
Allows adjusting the pressure sensor gain using an external pressure
measurement as reference.
No 2341)
Pressure Section 8.14.1.3
Adj/Calib > Calibration > Exp. Valve
(Page No 2343)
Expiratory Valve Section 8.14.1.4
Pressure Sensor:
• Pvent_control
• Pvent_monitor and
Paw
Before calibration starts, an autozero of Pvent_control, Pvent_monitor, Paw, Pflow-sensor and IexpValve is performed. If the autozero
procedure fails, a failure will be displayed and calibration cannot be
started.
• Expiratory Valve
During calibration, gain and offset for the valve characteristics of
the expiratory valve will be aligned.
Adj/Calib > Calibration > O2 Cell (Page
The Oxygen Sensor calibration is separated into two parts:
No 2346)
• Offset calibration. Calibrates the offset voltage of the amplifier
circuit.
Oxygen Sensor Calibration Section
8.14.1.5
Adj/Calib > Calibration > Flow Sensor
(Page No 2347)
Patient Flow Sensor Calibration Section
8.14.1.6
114
• Oxygen Sensor
• Gain calibration. During this 2-min calibration of the Oxygen Sensor, the ventilator delivers an increased oxygen concentration (if
oxygen is connected in the high pressure mode) or 21% oxygen
(if oxygen is connected in the low pressure mode or disconnected). It tests the Oxygen Sensor and resets the calibration
points specific to the Oxygen Sensor in use.
This calibration checks and resets the calibration point specific to
the flow sensor in use.
• Patient Flow Sensor
The Patient Flow Sensor is calibrated in both directions.
English | 624165/07
Touch Screen
8
8.14.1.1 Touch Screen
Allows calibrating the coordinates of the touch controller (4–point calibration).
1. Touch the Touch screen Button
Figure 8-28. Touch Screen Calibration Step 1
2. Touch the Start Button to begin the Touch Screen Calibration.
Figure 8-29. Touch Screen Calibration Step 2
Hamilton Medical | HAMILTON-C2 Service Manual
115
8
Service Software
3. The test begins automatically indicated by Touch Screen Calibration Running on the screen.
Figure 8-30. Touch Screen Calibration Step 3
4. You are instructed to touch the cross located on the top left of the screen with your finger.
5. Next, you are instructed to touch the cross located on the top right, then bottom right and then bottom left of the screen
with your finger.
6. Next, you are instructed to touch the Test Button in the middle of the screen with your finger.
Figure 8-31. Touch Screen Calibration Step 4
116
English | 624165/07
Touch Screen
8
7. The Touch Screen Calibration is complete. Touch Apply to accept the new Touch Screen Calibration or Cancel.
Figure 8-32. Touch Screen Calibration Step 5
8. The touch screen calibration is complete indicated by Touch screen calibration values saved. on the screen.
Figure 8-33. Touch Screen Calibration Step 6
Hamilton Medical | HAMILTON-C2 Service Manual
117
8
Service Software
8.14.1.2 Inspiratory Valve
Allows aligning the gain and offset for voltage-offset of the Inspiratory Valve.
Sets Offset 1 (low pressure) at 20 mbar blower pressure and Offset 2 (high pressure) at 60 mbar blower pressure when
required Qvent flow is reached.
NOTICE
− Confirm that the Technical State is updated with the current Inspiratory Valve serial number AND revision number, refer
to Modify Tab 8.8.3.
− The HAMILTON-C2 needs a warm-up period. Make sure it was running for at least 20 minutes in the Ventilation Software.
The Inspiratory Valve has a flow-voltage-hysteresis, in that the valve opening characteristic is different from the valve closing
characteristic. The valve needs to be calibrated with either the opening process or the closing process. Based on the design
and pressure control mechanism, the Inspiratory valve is calibrated with the closing characteristic.
The required flow adjustment depends on the Inspiratory Valve revision:
• MSP160230/00 –> 12 l/min (Qvent)
• MSP160230/01 and higher –> 18 l/min (Qvent)
The Technical State needs to contain the correct Inspiratory Valve revision to ensure proper control function
Inspiratory valve calibration:
1. Touch the Insp. Valve Button.
2. Wait approximately 10 seconds until the flow has stabilized.
Figure 8-34. Inspiratory Valve Adjustment / Calibration Step 1
3. Disconnect the inspiration tube as shown.
118
English | 624165/07
8
Inspiratory Valve
4. Touch the Start Button.
Figure 8-35. Inspiratory Valve Adjustment / Calibration Step 2
Figure 8-36. Inspiratory Valve Adjustment / Calibration Step 3
5. The Inspiratory Valve Calibration is complete indicated by Inspiration valve calibration successful displayed on the screen.
Hamilton Medical | HAMILTON-C2 Service Manual
119
8
Service Software
8.14.1.3 Pressure
Allows adjusting the Pressure Sensor Gain using an external pressure measurement as a reference.
1. Touch the Pressure Button.
Figure 8-37. Pressure Adjustment / Calibration Step 1
2. Attach a Pressure Connector to the Patient Connection.
3. Attach the Tube System to the Pressure Connector and the Expiratory Connection.
4. Attach an External Pressure Gauge to the Pressure Connector.
5. Close the Flow Sensor outlet.
6. Touch the Start Button.
Figure 8-38. Pressure Adjustment / Calibration Step 2
7. As soon as you touch the Start Button, the Ventilator does an autozero procedure to prevent wrong calibrations due to
remaining pressure inside the Patient Breathing Circuit.
120
English | 624165/07
Pressure
8
8. The Values on the screen should equal the value on the Pressure Gauge.
9. If adjustment is necessary, adjust the Gain by using the P&T Control Knob. Make the adjustment with the P&T Control
Knob, then, press the P&T Control Knob for the change in Gain to be applied.
Figure 8-39. Pressure Adjustment / Calibration Step 3
10. After completion, touch the Save Button to save the changes.
Figure 8-40. Pressure Adjustment / Calibration Step 4
NOTICE
If the test cannot be performed, check the whole setup for leaks. Furthermore, there should be almost no flow from the
expiration outlet. If you perform this test more than once, please close, then re-open the Pressure Calibration Window.
Hamilton Medical | HAMILTON-C2 Service Manual
121
8
Service Software
8.14.1.4 Expiratory Valve
Before calibration starts, an autozero of Pvent_control, Pvent_monitor, Paw, Pflow-sensor and IexpValve is performed. If the
autozero procedure fails, a failure will be displayed and calibration cannot be started.
During calibration, gain and offset for the valve characteristics of the Expiratory Valve will be aligned.
1. Touch the Exp. Valve Button.
Figure 8-41. Expiratory Valve Adjustment / Calibration Step 1
2. Connect a 30cm tube with a filter (PN 279204) as shown on the screen.
3. Touch the Start Button.
Figure 8-42. Expiratory Valve Adjustment / Calibration Step 2
4. The Ventilator performs an autozero procedure to prevent wrong calibrations due to remaining pressure inside the limb.
122
English | 624165/07
Expiratory Valve
8
5. The calibration runs automatically indicated by the Pvent_monitor and Iexp. values changing during the calibration process.
Figure 8-43. Expiratory Valve Adjustment / Calibration Step 3
6. The successful finish of the Expiratory Valve will be indicated with following message: Calibration successfully finished.
NOTICE
If the test takes more than 3 minutes or if calibration is not OK, replace the membrane and perform the test again.
Hamilton Medical | HAMILTON-C2 Service Manual
123
8
Service Software
8.14.1.5 Oxygen Sensor
The Oxygen Sensor calibration is separated into two parts:
• Offset calibration. Calibrates the offset voltage of the amplifier circuit.
• Gain calibration. During this 2-min calibration of the Oxygen Sensor, the ventilator delivers an increased oxygen concentration (if oxygen is connected in the high pressure mode) or 21% oxygen (if oxygen is connected in the low pressure mode
or disconnected). It tests the Oxygen Sensor and resets the calibration points specific to the Oxygen Sensor in use.
8.14.1.5.1 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 1-5
NOTICE
With Mainboard (MSP160200) Revisions 1-5, perform the test as shown below.
With Mainboard (MSP160200) Revisions 6 and higher and Mainboard (MSP160382), perform the Oxygen Sensor Calibration as shown in the Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher Section 8.14.1.5.2.
1. Connect the HAMILTON-C2 to high pressure O2.
2. Touch the O2 cell Button.
Figure 8-44. Oxygen Sensor Adjustment / Calibration Step 1
3. Disconnect the Oxygen Sensor Cable from the Oxygen Sensor and connect the Oxygen Sensor Calibration Tool (PN
160367).
124
English | 624165/07
Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 1-5
8
4. Touch the Start Button.
Figure 8-45. Osygen Sensor Adjustment / Calibration Step 2
5. The Offset Calibration runs automatically until Offset calibration completed appears on the screen.
6. Connect the Oxygen Sensor Cable to the Oxygen Sensor.
7. Touch the Start Button to begin the Gain Calibration.
Figure 8-46. Oxygen Sensor Adjustment / Calibration Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
125
8
Service Software
8. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... displayed on the screen.
Figure 8-47. Oxygen Sensor Adjustment / Calibration Step 4
9. The calibration is complete when OK is displayed on the screen.
Figure 8-48. Oxygen Sensor Adjustment / Calibration Step 5
126
English | 624165/07
Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher
8
8.14.1.5.2 Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher
NOTICE
For Mainboard (MSP160200) Revisions 6 and higher and Mainboard (MSP160382) Revisions 1-4.
1. Connect the HAMILTON-C2 to high pressure O2.
2. Touch the O2 Cell button.
Figure 8-49. Oxygen Sensor Adjustment / Calibration Step 1
3. Disconnect the Oxygen Sensor Cable from the Oxygen Sensor. Make sure the Oxygen Sensor connector is freely suspended.
4. Touch the Start button.
Figure 8-50. Oxygen Sensor Adjustment / Calibration Step 2
5. The Offset Calibration runs automatically until Offset calibration completed appears on the screen.
Hamilton Medical | HAMILTON-C2 Service Manual
127
8
Service Software
6. Connect the Oxygen Sensor Cable to the Oxygen Sensor.
7. Touch the Start Button to begin the Gain Calibration.
Figure 8-51. Oxygen Sensor Adjustment / Calibration Step 3
8. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... displayed on the screen.
Figure 8-52. Oxygen Sensor Adjustment / Calibration Step 4
128
English | 624165/07
Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher
8
9. The calibration is complete when OK is displayed on the screen.
Figure 8-53. Oxygen Sensor Adjustment / Calibration Step 5
Hamilton Medical | HAMILTON-C2 Service Manual
129
8
Service Software
8.14.1.5.3 Oxygen Sensor Calibration with Mainboard MSP160382 Revisions 5 and Higher
NOTICE
For Mainboard (MSP160382) Revisions 5 and higher: The calibration is complete when OK is displayed on the screen.
1. Connect the HAMILTON-C2 to high pressure O2.
2. Touch the O2 Cell Button.
3. Touch the Start Button.
Figure 8-54. Oxygen Sensor Adjustment / Calibration
4. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... indicated on the screen.
5. The calibration is complete when OK is displayed on the screen.
130
English | 624165/07
8
Flow Sensor
8.14.1.6 Flow Sensor
This calibration checks and resets the calibration point specific to the Patient Flow Sensor in use.
The Patient Flow Sensor is calibrated in both directions.
1. Touch the Flow sensor Button.
Figure 8-55. Patient Flow Sensor Calibration
2. Connect the tubing system and turn the Patient Flow Sensor. The Patient Flow Sensor has to be inverted for the Patient
Flow Sensor Calibration and an adapter is required.
Figure 8-56. Setup for the Patient Flow Sensor Calibration
Hamilton Medical | HAMILTON-C2 Service Manual
131
8
Service Software
3. Touch the Start Button.
Figure 8-57. Patient Flow Sensor Calibration Step 1
4. The Patient Flow Sensor Calibration is in process.
Figure 8-58. Patient Flow Sensor Calibration Step 2
132
English | 624165/07
8
Flow Sensor
5. Turn the Patient Flow Sensor.
6. Touch the Next Button to continue.
Figure 8-59. Patient Flow Sensor Calibration Step 3
7. The successful finish of the Patient Flow Sensor Calibration will be indicated with following message: Flow sensor calibration successful.
Figure 8-60. Patient Flow Sensor Calibration Step 4
Hamilton Medical | HAMILTON-C2 Service Manual
133
8
Service Software
8.14.2 Show Tab
1. Touch the Adjustment / Calibration Button.
Figure 8-61. Adjustment / Calibration Screen
2. Touch the Show Tab.
Figure 8-62. Show Tab
134
English | 624165/07
Pressure
8
8.14.2.1 Pressure
1. Touch the Pressure Button.
Figure 8-63. Pressure Screen
2. The screen displays the Pvent_control, Pvent_monitor and Paw Pressure Sensors Offset and Gain values.
Hamilton Medical | HAMILTON-C2 Service Manual
135
8
Service Software
8.14.2.2 Flow Sensor
1. Touch the Flow sensor Button.
Figure 8-64. Flow Sensor Screen
2. Displays the Inspiratory Pressure and Expiratory Pressure at different flows.
NOTICE
Not required on the HAMILTON-C2 Summary of Test Report.
136
English | 624165/07
Valves
8
8.14.2.3 Valves
1. Touch the Valves Button.
Figure 8-65. Valves Screen
2. Displays the Expiratory Valve Offset and Gain values.
Hamilton Medical | HAMILTON-C2 Service Manual
137
8
Service Software
8.14.2.4 Oxygen Sensor
1. Touch the O2 cell Button.
Figure 8-66. Oxygen Sensor Screen
2. Displays the Oxygen Sensor Offset and Gain values.
138
English | 624165/07
Rinsp
8
8.14.2.5 Rinsp
1. Touch the Rinsp Button
Figure 8-67. Rinsp Screen
2. Displays the Inspiratory Resistance values.
Hamilton Medical | HAMILTON-C2 Service Manual
139
8
Service Software
8.14.3 Default Settings Tab
CAUTION
If calibration is not possible due to a corrupted calibration file, touch the Default set. Tab. Therefore, all steps of the Service
Software have to be run again.
1. Touch the Default settings tab.
Figure 8-68. Default Settings Tab
2. The Default Set. Tab provides a method to reset or set the values to a pre-determined Factory Default Setting. You can
choose between setting the default values of only the Flow Sensor or of All settings.
3. To reset the values, touch the Set Button.
Figure 8-69. Factory Settings Reset Step 1
140
English | 624165/07
Default Settings Tab
8
4. When the Set Button is selected, choose between Flow sensor and All settings.
Figure 8-70. Factory Settings Reset Step 2
5. If you choose All settings, a Warning is displayed.
Figure 8-71. Factory Settings Reset Step 3
6. You must touch the Confirm Button to reset to the Default Settings.
Hamilton Medical | HAMILTON-C2 Service Manual
141
8
Service Software
8.15 Component Tests
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
Figure 8-72. Main Service Software Screen
2. Touch the Comp test Button.
Figure 8-73. Components Test Screen
142
English | 624165/07
Component Tests
8
Table 8-7. Electronics Tab
Component Tests
Description
Adj/Calib > Calibration > Alarm System
• Touch Screen
Alarm Section 8.15.1.1
This test allows activating the components related to
the alarm system, it also tests the loudness levels. The
test result needs to be confirmed manually.
Adj/Calib > Calibration > Alarm Monitor
(Steps 1 — 10)
• Alarm Silence Button and LED
1 (Page No 2113)
These tests diagnose the alarm monitoring system
components.
• Ambient Mode
(Step 11 — 13)
• Fan
Testing the Ambient Mode. This test consists of setting the blower to achieve a certain pressure and a
certain flow for 5 seconds, afterwards the instrument
will be switched to ambient mode causing the blower
to stop.
• Inspiratory Valve
(Page No 2102)
Alarm Monitor 1 Section 8.15.1.2
Component Tested
• Alarm lamp lights red or yellow,
depending
• Buzzer
• Ambient Valve
The Ambient Mode has activated the alarm silence
LED and the buzzer. The test result needs to be confirmed manually.
(Step 14 — 17)
This test checks the alarm in case of fan failure. It
requires stopping the fan.
Adj/Calib > Calibration > Alarm Monitor
2 (Page No 2114)
Alarm Monitor 2 Section 8.15.1.3
This test activates the software watchdog. The test
result has to be confirmed manually. Afterwards the
ventilator needs to be restarted.
• Watchdog (software tasks)
Table 8-8. User Interface Tab
Component Tests
Description
Adj/Calib > Calibration > User Interface
With this test, the User Interface related components
can be activated in order to check the interaction of
the Hard Keys with the GUI. Hard Key combinations
can also be tested. Dimming the screen and alarm
lamp is tested using the Day/Night Button. The P&T
Control Knob has 16 steps. The test is to see that all
16 steps register with the ventilator.
Tab (Page No 2115)
User Interface Tab Section 8.15.2
Component Tested
• P&T Control Knob
• Hard Keys + LED’s
• Backlight (day/night brightness)
Table 8-9. Pneumatics 1 Tab
Component Tests
Adj/Calib > Calibration > Binary Valve
(Page No 2106)
Binary Valve Section 8.15.3.1
Adj/Calib > Calibration > Autozero
(Page No 2109)
Description
Component Tested
This test checks the Autozero Valves during operation
and autozeroing sequence. A constant pressure is
applied. Both status, “running” and“autozero”,are
tested for the following valves: Pvent_monitor,
PFlowsensor.
• Autozero Valves
The test repeats the Autozero sequence 5 times
under a defined pressure.
• Autozero of:
Autozero Section 8.15.3.2
• Pvent_monitor Pflowsensor
– Paw Pressure Sensor
– Qaw Proximal Flow Sensor
– Pvent_monitor
– Pvent_control
Adj/Calib > Calibration > Blower Flow
(Page No 2104)
Blower Flow Section 8.15.3.3
Adj/Calib > Calibration > Blower Pressure (Page No 2105)
Blower Pressure Section 8.15.3.4
Several flows (ml/s) will be set as target to the blower.
The achieved blower speed in rpm measured by the
Hall Sensor should be within the defined tolerance.
• Blower
Several pressures are set as target to the blower. The
achieved pressure measured by the Pvent_monitor
Pressure Sensor should be within the given tolerance.
• Blower
Hamilton Medical | HAMILTON-C2 Service Manual
143
8
Service Software
Component Tests
Adj/Calib > Calibration > Expiratory
Valve (Page No 2111)
Expiratory Valve Section 8.15.3.5
Adj/Calib > Calibration > Inspiratory
Valve (Page No 2107)
Inspiratory Valve Section 8.15.3.6
Description
Component Tested
This test checks the Expiratory Valve. Several pressures are set as target for a constant flow Qinsp to
the blower. The pressures are measured by Paw and
have to be within the defined tolerance
• Expiratory Valve
This test checks the Inspiratory Valve tightness and
flow control with Qinsp 0, 3, 18 and 150 l/min as
well as pressure measurement Pvent_monitor with
Pinsp of 2, 10, 25 and 40 mbar.
• Inspiratory Valve
Table 8-10. Pneumatics 2 Tab
Component Tests
Adj/Calib > Calibration > O2 Input (Page
No 2112)
O2 Input Section 8.15.4.1
Adj/Calib > Calibration > Nebulizer
Valve (Page No 2116)
Nebulizer Valve Section 8.15.4.2
Adj/Calib > Calibration > Ambient Valve
(Page No 2108)
Ambient Valve Section 8.15.4.3
Adj/Calib > Calibration > Proximal Flow
Description
For this test, the ventilator needs to be connected to
high pressure oxygen. Several flows will be set as a
target for the O2 Mixer Valve. The oxygen flow is
measured by the internal QO2 Flow Sensor and has
to be within a certain tolerance. In a second test the
leakage is tested.
• O2 Mixer Assembly
For this test, the ventilator needs to be connected to
high pressure oxygen. The Patient Flow Sensor is used
to measure the generated flow when the Nebulizer
Valve is in a closed or opened state. The test results
successfully if the measured Qaw matches to the
defined tolerances.
• Nebulizer Valve
This test checks the ambient valve with a blower pressure of 20 and 60 mbar in off, closed and active
opened state. The result is measured with
Pvent_monitor Pressure Sensor.
• Ambient Valve
This test consists of three parts:
• Rinse Flow
(Page No 2110)
Proximal Flow Section 8.15.4.4
Component Tested
- Rinse Flow Test:
This test requires checking the rinse flow manually by
immersing the 2 flow tubes in a glass of water as
depicted. The number of bubbles has to be approximately equal on each outlet.
• Paw Pressure Sensor
• Qaw Proximal Flow Sensor
- Proximal Pressure:
This test allows checking the proximal pressure of the
Patient Flow Sensor. Two different pressures are
applied consecutively by the blower. The achieved
pressure is measured by the Paw Pressure Sensor and
has to be within the tolerance.
- Proximal Flow:
A certain flow is set as target to the blower, controlled by the internal Qvent Flow Sensor. The flow
through the proximal side of the Patient Flow Sensor
is measured by the internal differential Qaw Proximal
Flow Sensor and has to be within a certain tolerance.
Adj/Calib > Calibration > Air Entry
(Page No 2117)
This test allows checking the Pfilter Pressure Sensor
with and without an obstructed filter.
• HEPA Filter
• Pfilter Pressure Sensor
Air Entry Section 8.15.4.5
144
English | 624165/07
8
Electronics Tab
8.15.1 Electronics Tab
NOTICE
This loudness setting also sets the loudness for the System Test, Alarming. To check the alarm sound on High-, Medium-,
Low-Priority Alarms with the loudness range 1–10, set the loudness here, then perform the System Test, Alarming (see
Alarming Section 8.16.5).
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
Figure 8-74. Main Service Software Screen
2. Touch the Comp test Button.
3. Touch the Electronics Tab.
Figure 8-75. Components Test Screen
Hamilton Medical | HAMILTON-C2 Service Manual
145
8
Service Software
8.15.1.1 Alarm System
This test allows activating the components related to the alarm system, it also tests the loudness levels. The test result needs
to be confirmed manually.
1. Touch the Alarm System Button.
Figure 8-76. Alarm System Tests Step 1
2. Touch the Speaker Start Button.
Figure 8-77. Alarm System Tests Step 2
3. Observe that the speaker makes an audible sound.
146
English | 624165/07
8
Alarm System
4. The speaker volume can be adjusted with the screen adjustment knob. Check that the volume changes to minimum and
maximum.
Figure 8-78. Alarm System Tests Step 3
5. Indicate on the screen if the speaker is operating properly by touching OK or NOT OK.
Figure 8-79. Alarm System Tests Step 4
Hamilton Medical | HAMILTON-C2 Service Manual
147
8
Service Software
6. When the OK Button is touched, the speaker will switch OFF, the Speaker Start Button will switch to Start and OK will be
indicated on the screen.
Figure 8-80. Alarm System Tests Step 5
7. Touch the Yellow Lamp Start Button.
Figure 8-81. Alarm System Tests Step 6
8. Observe that the yellow lamp is switched ON.
148
English | 624165/07
Alarm System
8
9. Indicate on the screen if the Yellow lamp is operating properly by touching OK or NOT OK.
Figure 8-82. Alarm System Tests Step 7
10. When the OK Button is touched, the Yellow Lamp will switch OFF, the Yellow Lamp Start Button will switch to Start and
OK will be indicated on the screen.
Figure 8-83. Alarm System Tests Step 8
Hamilton Medical | HAMILTON-C2 Service Manual
149
8
Service Software
11. Touch the Red Lamp Start Button.
Figure 8-84. Alarm System Tests Step 9
12. Observe that the red lamp is switched ON,
13. Indicate on the screen if the red lamp is operating properly by touching OK or Not OK.
Figure 8-85. Alarm System Tests Step 10
150
English | 624165/07
Alarm System
8
14. When the OK Button is touched, the Red Lamp will switch OFF, the Red Lamp Start Button will switch to Start and OK will
be indicated on the screen.
Figure 8-86. Alarm System Tests Step 11
15. After completion of the Alarm System Tests, the results are indicated on the screen.
Hamilton Medical | HAMILTON-C2 Service Manual
151
8
Service Software
8.15.1.2 Alarm Monitor 1
• Procedure Steps 1 — 10
– These tests diagnose the alarm monitoring system components.
• Procedure Steps 11 — 13
– Testing the Ambient Mode. This test consists of setting the blower to achieve a certain pressure and a certain flow for 5
seconds, afterwards the ventilator will be switched to Ambient Mode causing the blower to stop.
– The Ambient Mode will activate the Alarm Silence LED and the Buzzer. The test result needs to be confirmed manually.
• Procedure Steps 14 — 17
– This test checks the alarm in case of fan failure. It requires stopping the fan.
1. Touch the Alarm Monitor 1 Button.
Figure 8-87. Alarm Monitor 1 Screen
152
English | 624165/07
Alarm Monitor 1
8
2. Touch the Start Button.
Figure 8-88. Alarm Monitor 1 Tests Start
3. Indicate on the screen if the Alarm Light and the Alarm Silence LED are blinking by touching OK or NOT OK.
NOTICE
With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will blink.
Figure 8-89. Alarm Monitor 1 Tests Step 1
4. Record the results on the HAMILTON-C2 Summary of Test Report Section Summary of Test Report.
Hamilton Medical | HAMILTON-C2 Service Manual
153
8
Service Software
5. Next, indicate on the screen if the Alarm Light and the Alarm Silence LED are ON by touching OK or NOT OK.
NOTICE
With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will be ON.
Figure 8-90. Alarm Monitor 1 Tests Step 2
6. Next, touch the Alarm Silence Button.
7. Indicate on the screen if the Alarm Light is ON by touching OK or NOT OK.
NOTICE
With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will be ON.
Figure 8-91. Alarm Monitor 1 Tests Step 3
154
English | 624165/07
Alarm Monitor 1
8
8. Next, indicate on the screen if the Alarm Light is OFF by touching OK or NOT OK.
Figure 8-92. Alarm Monitor 1 Tests Step 4
9. Next, touch the Alarm Silence Button.
10. Indicate on the screen if the Alarm Light is OFF by touching OK or NOT OK.
Figure 8-93. Alarm Monitor 1 Tests Step 5
Hamilton Medical | HAMILTON-C2 Service Manual
155
8
Service Software
11. Touch the Alarm Silence Button and wait 5 seconds.
Figure 8-94. Alarm Monitor 1 Tests Step 6
12. Indicate on the screen if the Expiratory Valve Plunger can be moved by hand by touching OK or NOT OK.
Figure 8-95. Alarm Monitor 1 Tests Step 7
156
English | 624165/07
Alarm Monitor 1
8
13. Indicate on the screen if the Alarm Silence LED is ON and the Buzzer sounds by touching OK or NOT OK.
Figure 8-96. Alarm Monitor 1 Tests Step 8
WARNING
Be careful when stopping the cooling fan by hand.
14. Next, remove the Rear Cover and Fan Filter, then physically STOP the cooling fan by hand until the alarm is shown on the
screen.
Figure 8-97. Alarm Monitor 1 Tests Step 9
Hamilton Medical | HAMILTON-C2 Service Manual
157
8
Service Software
15. After the Fan Failure Alarm is displayed, release the fan.
Figure 8-98. Alarm Monitor 1 Tests Step 10
16. Verify that the message Test completed is displayed on the screen.
Figure 8-99. Alarm Monitor 1 Tests Step 11
17. Re-install the Fan Filter and Rear Cover.
158
English | 624165/07
8
Alarm Monitor 2
8.15.1.3 Alarm Monitor 2
This test activates the software watchdog. The test result has to be confirmed manually. Afterwards, the ventilator needs to
be restarted.
1. Touch the Alarm Monitor 2 Button.
Figure 8-100. Alarm Monitor 2 Tests
2. Touch the Start Button to activate the Watchdog.
Figure 8-101. Alarm Monitor 2 Tests Step 1
• The Alarm LED should blink
• The Buzzer should sound
Hamilton Medical | HAMILTON-C2 Service Manual
159
8
Service Software
3. Touch the OK or NOT OK Button and reboot the ventilator into the Service Software.
Figure 8-102. Alarm Monitor 2 Tests Step 2
4. After completion, the HAMILTON-C2 must be switched OFF, then ON again.
160
English | 624165/07
User Interface Tab
8
8.15.2 User Interface Tab
With this test, the user interface related components can be activated in order to check the interaction of the hard keys with
the GUI. Hard key combinations can also be tested. Dimming the screen and alarm lamp is tested using the day/night button.
The P&T Control Knob Encoder has 16 steps. The test is to see that all 16 steps register with the ventilator.
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
2. Touch the Comp test Tab.
3. Touch the User I/F Tab.
Figure 8-103. User Interface Tab Screen
4. Test the P&T Control Knob Encoder by selecting the P&T Control Knob on the screen and rotating the knob. Observe there
are 16 steps in 1 full turn of the P&T Control Knob.
Figure 8-104. User Interface Tests Step 1
Hamilton Medical | HAMILTON-C2 Service Manual
161
8
Service Software
5. Press each button on the Interaction Panel and observe the reaction to the button pressed on the screen.
6. Touch the Night/Day Button to switch between the two display brightness levels.
7. Confirm with OK/NOT OK.
Figure 8-105. User Interface Tests Step 2
162
English | 624165/07
Pneumatics 1 Tab
8
8.15.3 Pneumatics 1 Tab
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
2. Touch the Comp test Tab.
3. Touch the Pneumatics 1 Tab.
Figure 8-106. Pneumatics 1 Tab
4. Connect a complete Patient Breathing Circuit before starting the tests.
Hamilton Medical | HAMILTON-C2 Service Manual
163
8
Service Software
8.15.3.1 Binary Valve
This test checks the Autozero Valves during operation and autozeroing sequence. A constant pressure is applied. Both status,
“running” and“autozero”,are tested for the following valves: Pvent_monitor, PFlowsensor.
1. Touch the Binary valve Button.
Figure 8-107. Binary Valve Tests
2. Connect the tubing as shown with a 7 mm ET Tube to the Flow Sensor.
3. Touch the Start Button to begin the Binary Valve Test.
Figure 8-108. Binary Valve Tests Step 1
164
English | 624165/07
Binary Valve
8
4. The test runs automatically indicated by Component test binary valves running... displayed on the screen.
Figure 8-109. Binary Valve Tests Step 2
5. The Component test binary valves is complete when OK is indicated on the screen. This means that the internal sensor
checks are ok.
Figure 8-110. Binary Valve Tests Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
165
8
Service Software
8.15.3.2 Autozero
The test repeats the Autozero sequence 5 times under a defined pressure.
1. Touch the Autozero Button.
Figure 8-111. Autozero Valves Tests Screen
2. Touch the Start Button.
Figure 8-112. Autozero Valves Tests Step 1
166
English | 624165/07
Autozero
8
3. The Paw Pressure Sensor and Qaw Proximal Flow Sensor will Autozero.
Figure 8-113. Autozero Valves Tests Step 2
4. The Paw Pressure Sensor and Qaw Proximal Flow Sensor Autozero adjustment is complete when OK is indicated on the
screen.
Figure 8-114. Autozero Valves Tests Step 3
5. The Autozero results will be recorded in the Service Log.
Hamilton Medical | HAMILTON-C2 Service Manual
167
8
Service Software
8.15.3.3 Blower Flow
Several flows (ml/s) will be set as target to the blower. The achieved blower speed in rpm measured by the Hall Sensor should
be within the defined tolerance.
1. Touch the Blower flow Button.
Figure 8-115. Blower Flow Tests Screen
2. Disconnect the Inspiration Tube.
3. Touch the Start Button.
Figure 8-116. Blower Flow Tests Step 1
168
English | 624165/07
Blower Flow
8
4. The test runs automatically indicated by Component test blower flow running... displayed on the screen.
Figure 8-117. Blower Flow Tests Step 2
5. The test is complete when Component test blower flow completed. is displayed on the screen and the results are displayed
with OK or NOT OK.
Figure 8-118. Blower Flow Tests Step 3
NOTICE
The Blower Pressure values are in mbar and the Flow values are in ml/minute.
Hamilton Medical | HAMILTON-C2 Service Manual
169
8
Service Software
8.15.3.4 Blower Pressure
Several pressures are set as a target for the blower. The achieved pressure measured by the Pvent_monitor Pressure Sensor
should be within the given tolerance.
1. Touch the Blower pres. Button.
Figure 8-119. Blower Pressure Tests
2. Disconnect the Inspiration Tube and seal the Patient Outlet with a stopper.
3. Touch the Start button.
Figure 8-120. Blower Pressure Tests Step 1
170
English | 624165/07
Blower Pressure
8
4. The test runs automatically indicated by Component test blower pressure running... displayed on the screen.
Figure 8-121. Blower Pressure Tests Step 2
5. The test is complete when Component test blower flow completed. is displayed on the screen and the results are displayed
with OK or NOT OK.
Figure 8-122. Blower Pressure Tests Step 3
NOTICE
The Blower Pressure values are in mbar.
Hamilton Medical | HAMILTON-C2 Service Manual
171
8
Service Software
8.15.3.5 Expiratory Valve
NOTICE
During this test, the Technical Event: 231001 may appear. This failure message can be ignored.
This test checks the Expiratory Valve. Several pressures are set as target for a constant flow Qinsp to the blower. The pressures are measured by Paw and have to be within the defined tolerance.
1. Touch the Exp. valve Button.
Figure 8-123. Expiratory Valve Tests Screen
2. Connect a complete Patient Breathing Circuit
3. Touch the Start Button to begin the Expiratory Valve Leakage and Pressure Tests.
Figure 8-124. Expiratory Valve Tests Step 1
172
English | 624165/07
8
Expiratory Valve
4. Test runs automatically indicated by Expiration valve pressure test in progress... displayed on the screen.
Figure 8-125. Expiratory Valve Tests Step 2
5. The test is complete when Expiration valve pressure test completed is displayed on the screen and the result is displayed
with OK or Not OK.
6. If the pressure test or leakage test fails, the values which are out of tolerance are displayed at the bottom of the page.
Figure 8-126. Expiratory Valve Tests Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
173
8
Service Software
8.15.3.6 Inspiratory Valve
This test checks the Inspiratory Valve tightness and flow control with Qinsp 0, 3, 18 and 150 l/min as well as pressure measurement Pvent_monitor with Pinsp of 2, 10, 25 and 40 mbar.
1. Touch the Insp. valve Button.
Figure 8-127. Inspiratory Valve Tests Screen
2. Connect an Adult 22 mm Tube (PN 260100) and Filter (PN 279211) to the ventilator and attach the Flow Analyzer (as
shown in the picture on the screen).
3. Touch the Start Button.
Figure 8-128. Inspiratory Valve Tests Step 1
4. The test runs automatically indicated by Leakage test in progress... displayed on the screen.
174
English | 624165/07
Inspiratory Valve
8
5. The Inspiratory Valve Leakage Test is complete when OK is indicated on the screen. If the pressure tests fails, the out of tolerance pressure is displayed.
Figure 8-129. Inspiratory Valve Tests Step 2
6. Touch the Next Button to begin the Flow Control Tests.
Figure 8-130. Inspiratory Valve Tests Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
175
8
Service Software
7. Pass through the Flow Control Test with the values of 0 l/min., 3 l/min., 18 l/min. and 150 l/min. Only verify the 18 l/min.
value with the external Flow Analyzer. It must be within a tolerance of ± 8% (16.5 - 19.5 l/min).
8. The test runs automatically indicated by Flow control in progress on the screen.
Figure 8-131. Inspiratory Valve Tests Step 4
9. The Inspiratory Valve Flow Control Test is complete when OK is indicated on the screen.
10. Connect the tubing as shown with a 7mm ET Tube.
11. Touch the Start Button to continue to the Inspiratory Valve Pressure Control Test.
Figure 8-132. Inspiratory Valve Tests Step 5
176
English | 624165/07
Inspiratory Valve
8
12. The test runs automatically indicated by Pressure control in progress on the screen.
Figure 8-133. Inspiratory Valve Tests Step 6
13. The Inspiratory Valve Pressure Control Test is complete when OK is indicated on the screen.
Figure 8-134. Inspiratory Valve Tests Step 7
Hamilton Medical | HAMILTON-C2 Service Manual
177
8
Service Software
8.15.4 Pneumatics 2 Tab
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
2. Touch the Comp test Tab.
3. Touch the Pneumatics 2 Tab.
Figure 8-135. Pneumatics 2 Tab
178
English | 624165/07
O2 Input
8
8.15.4.1 O2 Input
For this test, the ventilator needs to be connected to high pressure oxygen. Several flows will be set as a target for the O2
Mixer Valve. The oxygen flow is measured by the internal QO2 Flow Sensor and has to be within a certain tolerance. In a
second test the leakage is tested.
1. Touch the O2 input Button.
Figure 8-136. O2 Input Tests Screen
2. Connect the ventilator to high pressure oxygen (Minimum 2.8 bar to Maximum 6.0 bar (41 - 86 psi)).
3. Disconnect the Inspiration Tube.
4. Touch the Start Button to begin the O2 Input Flow and Leakage Tests.
Figure 8-137. O2 Input Tests Step 1
Hamilton Medical | HAMILTON-C2 Service Manual
179
8
Service Software
5. The test runs automatically indicated by O2 valve flow test in progress... displayed on the screen.
Figure 8-138. O2 Input Tests Step 2
6. The O2 Valve Flow Test is complete when OK is indicated on the screen.
7. The next test runs automatically indicated by O2 valve leakage test in progress... displayed on the screen.
Figure 8-139. O2 Input Tests Step 3
180
English | 624165/07
O2 Input
8
8. The O2 Valve Leakage Test is complete when OK is indicated on the screen.
Figure 8-140. O2 Input Tests Step 4
Hamilton Medical | HAMILTON-C2 Service Manual
181
8
Service Software
8.15.4.2 Nebulizer Valve
For this test, the ventilator needs to be connected to high pressure oxygen. The Patient Flow Sensor is used to measure the
generated flow when the Nebulizer Valve is in a closed or opened state. The test results are successful if the measured Qaw
matches the defined tolerances.
1. Touch the Neb. valve Button.
Figure 8-141. Nebulizer Valve Tests Screen
2. Connect the tube system as shown.
3. Touch the Start Button to start running the Nebulizer Valve Test with high inlet oxygen pressure Minimum 2.8 bar to Maximum 6.0 bar (41 - 86 psi).
Figure 8-142. Nebulizer Valve Tests Step 1
182
English | 624165/07
Nebulizer Valve
8
4. The adjustment is complete when OK is indicated on the screen.
Figure 8-143. Nebulizer Valve Tests Step 2
Hamilton Medical | HAMILTON-C2 Service Manual
183
8
Service Software
8.15.4.3 Ambient Valve
NOTICE
In software version 2.0.6 and earlier, during this test, the Technical Event: 231001 may appear. This failure message can be
ignored.
This test checks the Ambient Valve with a blower pressure of 20 and 60 mbar in off, closed and active opened state. The
result is measured with the Pvent_monitor Pressure Sensor.
1. Touch the Ambi valve Button.
Figure 8-144. Ambient Valve Tests Screen
2. Connect the tube system as shown and close the Flow Sensor Outlet.
3. Touch the Start Button.
Figure 8-145. Ambient Valve Tests Step 1
184
English | 624165/07
Ambient Valve
8
4. The test runs automatically indicated by Test ambient valve power off in progress displayed on the screen.
5. The Ambient Valve Power-Off Test is complete when OK is indicated on the screen.
6. The next test runs automatically indicated by Test ambient valve Closed in progress displayed on the screen.
Figure 8-146. Ambient Valve Tests Step 2, Possible Technical Event (only with SW 2.0.6 or lower)
7. The Ambient Valve Closed Test is complete when OK is indicated on the screen.
8. The next test runs automatically indicated by Test ambient valve Active open in progress displayed on the screen.
9. The Ambient Valve Active Open Test is complete when OK is indicated on the screen.
Figure 8-147. Ambient Valve Tests Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
185
8
Service Software
8.15.4.4 Proximal Flow
This test consists of three parts:
• Rinse Flow Test:
– This test requires checking the rinse flow manually by immersing the 2 flow tubes in a glass of water as depicted. The
number of bubbles has to be approximately equal on each outlet.
• Proximal Pressure:
– This test allows checking the proximal pressure of the Patient Flow Sensor. Two different pressures are applied consecutively by the blower. The achieved pressure is measured by the Paw Pressure Sensor and has to be within the tolerance.
• Proximal Flow:
– A certain flow is set as a target to the blower, controlled by the internal Qvent Flow Sensor. The flow through the proximal side of the Patient Flow Sensor is measured by the internal differential Qaw Proximal Flow Sensor and has to be
within a certain tolerance.
1. Touch the Prox. test Button.
Figure 8-148. Proximal Flow Tests Screen
186
English | 624165/07
Proximal Flow
8
2. Touch the Start Button.
Figure 8-149. Proximal Flow Tests Step 1
3. Attach 2 identical length tubes to the Flow Sensor connectors, immerse the tubes in a glass of water. Verify that the bubbles appear at both tube outlets and that the number of bubbles are approximately equal on each outlet.
4. Indicate on the screen if the Rinse Flow Test passed by pressing OK or NOT OK.
Figure 8-150. Proximal Flow Tests Step 2
Hamilton Medical | HAMILTON-C2 Service Manual
187
8
Service Software
5. The Rinse Flow Test is complete when OK is displayed on the screen.
Figure 8-151. Proximal Flow Tests Step 3
6. Connect the Adult Tube System to the ventilator with a 7 mm ET Tube.
7. Touch the Start Button to begin the Proximal Pressure Test.
8. The test runs automatically indicated by Proximal pressure in progress... displayed on the screen.
9. The Proximal Pressure Test is complete when OK is indicated on the screen.
Figure 8-152. Proximal Flow Tests Step 4
10. Remove the 7 mm ET Tube from the Adult Tube System.
188
English | 624165/07
Proximal Flow
8
11. Touch the Start button to begin the Proximal Flow Test.
Figure 8-153. Proximal Flow Tests Step 5
12. The test runs automatically indicated by Proximal flow in progress... displayed on the screen.
Figure 8-154. Proximal Flow Tests Step 6
Hamilton Medical | HAMILTON-C2 Service Manual
189
8
Service Software
13. The Proximal Flow Test is complete when OK is indicated on the screen.
Figure 8-155. Proximal Flow Tests Step 7
190
English | 624165/07
Air Entry
8
8.15.4.5 Air Entry
NOTICE
If the test is NOT OK, restart the HAMILTON-C2 Service Software and perform the test again.
This test allows checking the Filter Pressure Sensor Pfilter with and without an obstructed filter.
1. Touch the Air entry Button.
Figure 8-156. Air Entry Tests Screen
2. Disconnect the Inspiration Tube.
3. Touch the Start Button.
Figure 8-157. Air Entry Tests Step 1
Hamilton Medical | HAMILTON-C2 Service Manual
191
8
Service Software
4. The default HEPA Filter Test is completed if OK is stated behind the description.
Figure 8-158. Air Entry Tests Step 2
5. Obstruct the HEPA Filter with a piece of paper (A) with a small hole.
Figure 8-159. Obstruction in the Air Entry
192
English | 624165/07
Air Entry
8
6. Connect the obstructed HEPA Filter and touch the Next Button. The pressure must be higher than 2 mbar as shown.
7. The test is complete when OK is indicated on the screen.
Figure 8-160. Air Entry Tests Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
193
8
Service Software
8.16 System Test
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
Figure 8-161. Main Service Software Screen
2. Touch the System test Button.
Figure 8-162. System Test Screen
194
English | 624165/07
System Test
8
Table 8-11. System Test
Adjustment/ Calibration
Description
System Test > Pressure (Page No
This test requires three different pressures to be applied and to verify the values (Pvent_control, Pvent_monitor and Paw) are within
the given tolerance range (shown in a green or red background).
• Pvent_control
This test requires a flow to be applied and to verify the internal
Qvent Flow Sensor and the Qaw Proximal Flow Sensor values are
within the given tolerance range (shown in a green or red background).
• Qvent Flow Sensor
This test requires a flow at different oxygen concentrations to be
applied to verify the internal QO2 Flow Sensor, Qvent Flow Sensor,
the Patient Flow Sensor and the Oxygen Sensor values are within
the given tolerance range (shown in a green or red background).
• O2 Mixer Assembly
This test checks for leakage in the Patient Breathing Circuit and
determines the circuit's compliance compensation factor. The ventilator is pressurized to a certain pressure. The Patient Breathing Circuit is considered tight if this pressure can be maintained. If there is
a leak, the pressure drops in proportion to the size of the leak.
• Patient Breathing Circuit
This test allows activation of the 3 different alarm priorities and
combinations by pressing the corresponding button and to verify
the correct alarm lamp and sound has been generated.
• Alarm Priorities
2201)
Pressure Section 8.16.1
System Test > Flow (Page No
2202)
Flow Section 8.16.2
System Test > O2 Mixer (Page No
2203)
O2 Mixer Section 8.16.3
System Test > Tightness (Page
No 2204)
Tightness Section 8.16.4
System Test > Alarming (Page No
2205)
Alarming Section 8.16.5
Hamilton Medical | HAMILTON-C2 Service Manual
Component Adjusted/
Calibrated
• Pvent_monitor
• Paw
• Qaw Proximal Flow Sensor
• QO2 Flow Sensor
• Oxygen Sensor
195
8
Service Software
8.16.1 Pressure
This test requires three different pressures to be applied and to verify the values (Pvent_control, Pvent_monitor and Paw) are
within the given tolerance range (shown in a green or red background).
1. Touch the Pressure Button.
Figure 8-163. Pressure System Tests Screen
2. Attach the tube system as shown in the picture.
3. Touch the ON Button.
Figure 8-164. Pressure System Tests Step 1
196
English | 624165/07
Pressure
8
4. Set the Pinsp to 5 mbar, 25 mbar and 50 mbar.
5. Select the Pinsp Control with the P&T Control Knob. Press the P&T Control Knob and rotate the setting to the values listed
above. Press the P&T Control Knob to activate the setting.
Figure 8-165. Pressure System Tests Step 2
6. All values should read 5 mbar ±1.0, 25 mbar ±1.2 and 50 mbar ±2.5.
7. The test is complete when the Status OK is shown.
Figure 8-166. Pressure System Tests Step 3
8. Touch the Off Button.
Hamilton Medical | HAMILTON-C2 Service Manual
197
8
Service Software
8.16.2 Flow
This test requires a flow to be applied and to verify the internal Qvent Flow Sensor and the Qaw Proximal Flow Sensor values
are within the given tolerance range (shown in a green or red background).
1. Touch the Flow Button.
Figure 8-167. Qvent Flow Sensor Tests Screen
2. Set the Insp. Flow to 18 l/min.
3. Touch the On Button.
Figure 8-168. Qvent Flow Sensor Tests Step 1
198
English | 624165/07
Flow
8
4. The test is complete when the Status OK is shown.
Figure 8-169. Qvent Flow Sensor Tests Step 2
5. Touch the Off Button.
Hamilton Medical | HAMILTON-C2 Service Manual
199
8
Service Software
8.16.3 O2 Mixer
This test requires a flow at different oxygen concentrations to be applied to verify the internal QO2 Flow Sensor, Qvent Flow
Sensor, the Patient Flow Sensor and the Oxygen Sensor values are within the given tolerance range (shown in a green or red
background).
1. Connect the ventilator to high pressure oxygen.
2. Touch the O2 mixer Button.
Figure 8-170. O2 Mixer Tests Screen
3. Set the Insp. flow to 10 l/min.
4. Touch the On Button.
Figure 8-171. O2 Mixer Tests Step 1
5. Check for the following FiO2 settings: 21%, 90% and 61%.
200
English | 624165/07
O2 Mixer
8
6. The test is complete when the Status OK is shown.
Figure 8-172. O2 Mixer Tests Step 2
7. Touch the Off Button.
Hamilton Medical | HAMILTON-C2 Service Manual
201
8
Service Software
8.16.4 Tightness
NOTICE
During this test the Technical Event: 231001 may appear. You can ignore this fault message.
This test checks for leakage in the Patient Breathing Circuit and determines the circuit's compliance compensation factor. The
ventilator is pressurized to a certain pressure. The Patient Breathing Circuit is considered tight if this pressure can be maintained. If there is a leak, the pressure drops in proportion to the size of the leak.
1. Touch the Tightness Button.
Figure 8-173. Tightness Tests Screen
2. Touch the Start Button.
Figure 8-174. Tightness Tests Step 1
202
English | 624165/07
Tightness
8
3. The test runs automatically indicated by Tightness tubing test in progress... indicated on the screen.
Figure 8-175. Tightness Tests Step 2
4. The Tightness Test is complete when OK is indicated on the screen.
Figure 8-176. Tightness Tests Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
203
8
Service Software
8.16.5 Alarming
NOTICE
During this test, Technical Events and Technical Faults will be displayed. This is a needed behavior and does not need any
action.
This test allows activation of the 3 different alarm priorities and combinations by pressing the corresponding button and to
verify the correct alarm lamp and sound has been generated.
1. Touch the Alarming Button.
Figure 8-177. Alarming Tests Screen
204
English | 624165/07
Alarming
8
2. Touch the Alarm High, Alarm Medium and Alarm Low Buttons to generate alarm(s).
3. Check if the correct Alarm Lamp and Alarm Sounds are generated.
4. Check the Alarm Priority by combining the Alarms.
5. Touch the Alarm High Button.
6. Select the Alarm High and Alarm Medium Buttons and verify if the alarm priority is working correctly.
Figure 8-178. Alarming Screen Tests Step 1
7. Touch the Alarm Medium Button.
Figure 8-179. Alarming Screen Tests Step 2
Hamilton Medical | HAMILTON-C2 Service Manual
205
8
Service Software
8. Touch the Alarm Low Button.
Figure 8-180. Alarming Screen Tests Step 3
9. Touch OK / NOT OK to confirm / cancel the test.
Figure 8-181. Alarming Screen Tests Step 4
206
English | 624165/07
Sensor Data
8
8.17 Sensor Data
NOTICE
Only for internal use.
1. From the Main Service Software Screen, touch the Tests / Calibration Button.
Figure 8-182. Main Service Software Screen
2. Touch the Sensor data Tab
Figure 8-183. Sensor Data Tab
Hamilton Medical | HAMILTON-C2 Service Manual
207
8
Service Software
8.17.1 Sensor 1
Touch the Sensor 1 Tab to display the sensors values and status.
Figure 8-184. Sensor 1
208
English | 624165/07
Sensor 2
8
8.17.2 Sensor 2
Touch the Sensor 2 Tab to display the sensors values and status.
Figure 8-185. Sensor 2
Hamilton Medical | HAMILTON-C2 Service Manual
209
8
Service Software
8.17.3 Sensor 3
Touch the Sensor 3 Tab to display the sensors value and status.
Figure 8-186. Sensor 3
210
English | 624165/07
8
Sensor Status
8.17.4 Sensor Status
The displayed sensors status (hex) is a value which consists of a combination of several status (bits). The table below shows
possible sensors status values. Not all status values are applicable for every sensor.
Figure 8-187. Sensor Status
Hamilton Medical | HAMILTON-C2 Service Manual
211
8
Service Software
8.18 Log/Config Files
NOTICE
How to export Event Log, Service Log, Config File and Instrument Report.
1. From the Main Service Software Screen, touch the Log/Config files Button.
Figure 8-188. The Main Service Software Screen
2. On the Log/Config Files Screen are the:
• Event Log Button
• Service Log Button
• Export Button (Instrument Report and Events Log Files)
• Back Button
Figure 8-189. The Log / Config Files Screen
212
English | 624165/07
8
Event Log
8.18.1 Event Log
1. Touch the Event log Button to open the Event Log.
Figure 8-190. Viewing the Event Log
2. The Event Log displays the Technical Events which have occurred with the date, time, device affected and a description of
the Technical Event.
3. To view another section of the Event Log listing, rotate the P&T Control Knob to select the slider on the right side of the
screen.
4. Press the P&T Control Knob to activate the slider, then rotate the P&T Control Knob to move the slider up or down.
Hamilton Medical | HAMILTON-C2 Service Manual
213
8
Service Software
8.18.2 Service Log
1. Touch the Service log Button to open the Service Log.
Figure 8-191. Service Log Step 1
2. The Service Log displays the test and calibration results with the date, time and a description of the results.
Figure 8-192. Service Log Step 2
3. To view another section of the Service Log listing, rotate the P&T Control Knob to select the slider on the right side of the
screen.
4. Press the P&T Control Knob to activate the slider, then rotate the P&T Control Knob to move the slider up or down.
214
English | 624165/07
Export (Instrument Report and Events)
8
8.18.3 Export (Instrument Report and Events)
1. The Export Tab allows exporting the Service Log / Instrument Report to a USB Memory Stick.
2. Insert a USB Memory Stick into the USB Port on the side of the Interaction Panel.
Figure 8-193. Export Tab Screen
Hamilton Medical | HAMILTON-C2 Service Manual
215
8
Service Software
8.18.3.1 Export Events Log Files
1. Touch the Events Tab.
Figure 8-194. Export Event Log Step 1
2. Touch the Start Button.
Figure 8-195. Export Event Log Step 2
3. The export is performed automatically.
4. The Event File is exported in a folder named as c2-sn##### (##### device serial number). In this folder is a file named
year-month-day_hour-minute-second_serviceLog_enGB.txt
NOTICE
If you have to send log files to Hamilton Medical for further investigation, please send all files which are created in the specific folder (error.log, eventLog.txt, serviceLog.txt, etc.).
216
English | 624165/07
Export Instrument Report
8
8.18.3.2 Export Instrument Report
1. Touch the Instrument Report Tab.
Figure 8-196. Export Instrument Report Step 1
2. Touch the Start Button.
Figure 8-197. Export Instrument Report Step 2
4. The export is performed automatically.
5. The Instrument report is exported in a folder named as c2-sn##### (##### device serial number). In this folder is a file
named year-month-day_hour-minute-secondInstrumentRepport.txt.
Hamilton Medical | HAMILTON-C2 Service Manual
217
8
Service Software
8.19 Software Updates
8.19.1 Software Update from Version < 2.2.0
WARNING
After a software update, the ventilator must be restarted and the complete Service Software Calibration and Test's must be
performed.
Software downgrade is not allowed/possible, because this will corrupt the unit and the ESM Module will need to be
exchanged.
To update the HAMILTON-C2 from Software Version 1.x.x to Software Version 2.2.0 or higher, a migration step is required.
For further information see the HAMILTON-C2 Software Upgrade Guide (PN 624200).
1. The following files are required:
• C2_V2.2.0-Migration.tar
This is required as a migration step from 1.x.x to 2.2.0 since the software structure has been
changed.
• C2_EM01A_V2.2.5.tar or higher
This is the latest Software to be installed after the Migration Software. For the latest software
check the Partner-net.
2. From the Main Service Software Screen, touch the Software update Button.
Figure 8-198. Software Update Step 1
3. The Software Update Screen allows update or upgrade of the HAMILTON-C2 Software.
218
English | 624165/07
Software Update from Version < 2.2.0
8
4. Touch the SW update Button.
Figure 8-199. Software Update Step 2
5. Insert the USB Memory Stick containing the migration and upgrade version as listed in the table above.
If a correct file is available on the USB Memory Stick, the message appears indicating the update version, otherwise, the message No Update tarball file is available... will be displayed. In case there is more than one software update stored on the USB
Memory Stick, please ensure to choose the correct version.
6. If the USB Memory Stick content is not displayed directly, refresh this page by closing and entering this page again.
Figure 8-200. Software Update from Version 1.x.x to 2.2.0
7. Select the software file to be installed (C2_V2.2.0-Migration.tar).
8. Touch the Start Button to start the software update which consists of extracting, verifying and installing.
9. A progress bar is shown indicating the current status of the software update.
10. When the process is complete, Done is shown on the screen.
Hamilton Medical | HAMILTON-C2 Service Manual
219
8
Service Software
11. Restart the HAMILTON-C2. The unit starts up automatically in the Service Software.
12. For further update of the HAMILTON-C2 Software, see the Software Update from Version 2.x to 2.2.5 or Higher Section
8.19.2.
220
English | 624165/07
Software Update from Version 2.x to 2.2.5 or Higher
8
8.19.2 Software Update from Version 2.x to 2.2.5 or Higher
WARNING
After a software update, the ventilator must be restarted and the complete Service Software Calibration and Test's must be
performed.
NOTICE
Software downgrade is not allowed/possible, because this will corrupt the unit and the ESM Module will need to be
exchanged.
1. From the Main Service Software Screen, touch the Software update Button.
Figure 8-201. Software Update Step 1
2. The Software Update Screen allows update or upgrade of the HAMILTON-C2 Software.
Hamilton Medical | HAMILTON-C2 Service Manual
221
8
Service Software
3. Touch the SW update Button.
Figure 8-202. Software Update Step 2
4. If the USB Memory Stick is not inserted into the USB Connector or no software is available on the USB Memory Stick, the
message No Update tarball file is available... is displayed.
Figure 8-203. Software Update Step 3
222
English | 624165/07
Software Update from Version 2.x to 2.2.5 or Higher
8
5. If the correct file is available on the USB Memory Stick, the message appears indicating the update version.
Figure 8-204. Software Update Step 4
Figure 8-205. Software Update Step 5
Hamilton Medical | HAMILTON-C2 Service Manual
223
8
Service Software
6. To begin the software update, touch the Start Button. With software version 2.2.0 or higher, the software which is to be
updated can be selected from the scroll down menu (shown on the picture below).
Figure 8-206. Software Update, Step 6
7. A progress bar is shown indicating the progress of the update.
Figure 8-207. Software Update, Step 7
224
English | 624165/07
Software Update from Version 2.x to 2.2.5 or Higher
8
8. After the software update, the message update successfully finished is displayed on the screen.
Figure 8-208. Software Update, Step 8
9. Restart the device and do not terminate the device during the first start up after performing a software update!
10. Perform the complete Calibration and Tests as described in the Service Software Section 8
Hamilton Medical | HAMILTON-C2 Service Manual
225
8
Service Software
8.20 Pre-operational Checks
Refer to the HAMILTON-C2 Operator‘s Manual.
Perform the pre-operational checks:
• Tightness Test
• Patient Flow Sensor Calibration
• Oxygen Sensor Calibration
• Alarm tests:
– High Pressure Alarm
– Low Minute Alarm
– Low Oxygen Alarm / Oxygen Supply Failed Alarm
– Disconnection on Patient Side
– Loss of External Power
– Exhalation Obstructed
– Apnea
226
English | 624165/07
8
General Tests
8.21 General Tests
NOTICE
All devices connected to the HAMILTON-C2 must be for medical use and meet the requirements of standard IEC 60601-1.
1.
Connect the HAMILTON-C2 to the AC Mains Power and to the DC Power Supply.
2.
Switch on the HAMILTON-C2 and stay in Stand-By Mode.
3.
Go to Alarm -> Loudness and set Loudness to 10.
4.
Perform the following tests:
A. AC — DC Test
• Disconnect the AC Mains Power. Verify that the unit continues to run on the DC Power Supply and that the DC symbol is
displayed on the screen.
B. DC — Battery Test
• Disconnect the DC Power Supply. Verify that the unit continues to run on internal Battery Power and that the battery symbol is displayed on the screen. Verify that the alarm speaker sounds correctly for at least one minute.
C. AC — Battery Test
• If no DC Power Supply is available, disconnect the AC Mains Power and verify that the HAMILTON-C2 switched to internal
Battery Power. Verify that the alarm speaker sounds correctly for at least one minute.
D. Power Loss Test
• If no DC Power Supply is available, verify that the HAMILTON-C2 switches to internal Battery Power as soon as the AC
Mains Power is removed.
• Remove the internal batteries. Verify that the unit alarms. Connect the internal batteries again and verify that the unit
starts up again.
CAUTION
The buzzer is powered by the gold caps (super capacitors) on the Mainboard. Make sure the unit is switched on (Stand-by
mode or Service Mode) for at least 20 minutes before the test in order that the gold caps are fully charged.
Replace the Mainboard (MSP160382) if the buzzer sounds for less than 2 minutes.
Hamilton Medical | HAMILTON-C2 Service Manual
227
8
Service Software
E. Fan Check
• In order to check the fan for proper assembly, place a piece of paper over the fan inlet. If the paper sticks while the ventilator is on, the fan blows into the ventilator correctly.
Figure 8-209. Fan Check
F. Buzzer Test
• To make sure that the buzzer works as specified, it has to be tested during the yearly Preventive Maintenance.
– Start the device in the ventilation software.
– Disconnect the AC power cable.
– Remove both internal batteries.
– Verify that the buzzer sounds for at least 2 minutes.
• If the buzzer sounds for less than 2 minutes, replace the Mainboard (MSP160382).
G. Communication Interface RS232 Test
• Connect a Patient Monitoring System. Verify proper communication with the Patient Monitoring System.
• Alternatively you can check the signal availability with Datalogger Software:
– DATALOGGER Software (download on Partner-net).
– DATALOGGER Manual (download on Partner-net).
– A Hamilton Ventilator with activated RS232 Interface or Communication Board.
– A Windows XP or Windows 7 computer with RS232 or USB Interface
– To connect HAMILTON-C2 use Interface Cable (PN 160366)
228
English | 624165/07
8
General Tests
1. Run HAMILTON-Datalogger.exe (available as download from the Hamilton Medical Partner-net).
The Welcome window appears.
2. Click > ENTER
Figure 8-210. Datalogger
3. The software automatically recognizes the connected device’s software:
• For HAMILTON-C1/T1/C2/C3 ventilators, the Datalogger displays HAMILTON-C2.
4. To check whether the connection is working properly, start ventilation.
If the Datalogger is working properly, the display shows ventilation values as well as the device serial number (S/N at Device
information).
Also check the Errors Counter on the right lower corner.
NOTICE
In addition, for all connections you may order USB to RS232 Converter (PN 396187, in case your computer has only an USB
Interface).
Important: The USB to RS232 Converter needs an additional driver to work properly. It is enclosed on the CD in the package.
Hamilton Medical | HAMILTON-C2 Service Manual
229
8
Service Software
8.22 CO2 Sensor Accuracy Check
Follow these steps to check the accuracy of the CO2 Monitor, if installed.
In addition to the HAMILTON-C2 ventilator with CO2 Monitor installed, this check requires:
• Patient Breathing Circuit (not required : Humidifier, Water Traps or Inspiratory Filter)
• Test Lung
• CO2 Sensor (PN 281718)
• CO2 Adult Airway Adapter, Single Use, 10/box (PN 281719)
• CO2 Adapters (3x), (one Box contains 25 pieces) (PN 281803)
• Barometric Pressure Gauge (mbar) or TSI
• CO2 Calibration Gas Cylinder or equivalent (CO2 5% ± 0.03; O2 0% ± 0.03; N2 Balance) (PN 159173)
• CO2 Gas Regulator (PN 159172)
1. Enter the Service Software -> Tests/calibration > Sensor data > Sensor 3.
2. Note the displayed barometric pressure (Pressure sensor Pambient) and verify that it is within ± 20 mbar of the actual barometric pressure, measured with a second gauge or device (e.g. the TSI Flow Meter).
Figure 8-211. Pressure Sensor Pambient
230
English | 624165/07
CO2 Sensor Accuracy Check
8
Figure 8-212. TSI Flow Meter
3. Restart the ventilator in operational mode.
4. Enter the Configuration Menu: Utilities > Configuration.
Figure 8-213. Utilities Configuration
Hamilton Medical | HAMILTON-C2 Service Manual
231
8
Service Software
5. Activate the CO2 Hardware Option: Utilities > Configuration > Options > HW options
Figure 8-214. HW Options
6. Connect the CO2 Sensor with the CO2 Communication Board.
7. Activate the CO2 Sensor and deactivate the Oxygen Sensor: System > Sensors on/off. Set oxygen (FiO2) = 21%.
8. Wait 3 minutes to allow the CO2 Sensor to reach the working temperature.
Figure 8-215. Sensor Update
232
English | 624165/07
CO2 Sensor Accuracy Check
8
9. Connect the Patient Breathing Circuit; Test Lung, CO2 Sensor, CO2 Adapters, CO2 Calibration Gas and CO2 Gas Regulator as shown.
Figure 8-216. CO2 Sensor Accuracy Check Illustration
Hamilton Medical | HAMILTON-C2 Service Manual
233
8
Service Software
10. Select the System Tab, Tests & calib, then CO2 Sensor to start the CO2 Sensor Calibration.
Figure 8-217. CO2 Sensor Calibration
11 Start ventilation with standard settings. Select the PCO2 curve for displaying on the screen (see Figure 8-218 CO2 Sensor
Accuracy Check).
12. Open the CO2 Regulator slightly so that the display shows a stable CO2 curve (see Figure 8-218 CO2 Sensor Accuracy
Check).
Figure 8-218. CO2 Sensor Accuracy Check
NOTICE
Only a small regulated gas flow is required to achieve a stable curve. To achieve a stable CO2 reading, ensure that the CO2
Sensor and CO2 Adapters are upright.
13. To get the PCO2 value in mmHg, press the Freeze Button (e.g. 34mmHg as shown).
234
English | 624165/07
8
CO2 Sensor Accuracy Check
Example : if the Ambient Temperature is 20 °C, Barometric Pressure is 900 mbar, and displayed PCO2 reading is 34 mmHg,
the displayed PCO2 Value is at the low end, but still within the range.
Figure 8-219. Graph Based on Ambient Temperature
15. See the following graphs based on ambient temperature:
• Ambient temperature 20 °C (17.5 to 22.5 °C): (see Figure 8-220)
• Ambient temperature 25 °C (22.5 to 27.5 °C): (see Figure 8-221)
• Ambient temperature 30 °C (27.5 to 32.5 °C): (see Figure 8-222)
• Ambient temperature 35 °C (32.5 to 37.5 °C): (see Figure 8-223)
Hamilton Medical | HAMILTON-C2 Service Manual
235
8
Service Software
Figure 8-220. Acceptable Range for PCO2 at Atmospheric Pressure (17.5 to 22.5 °C)
236
English | 624165/07
CO2 Sensor Accuracy Check
8
Figure 8-221. Acceptable Range for PCO2 at Atmospheric Pressure (22.5 to 27.5 °C)
Hamilton Medical | HAMILTON-C2 Service Manual
237
8
Service Software
Figure 8-222. Acceptable Range for PCO2 at Atmospheric Pressure (27.5 to 32.5 °C)
238
English | 624165/07
CO2 Sensor Accuracy Check
8
Figure 8-223. Acceptable Range for PCO2 at Atmospheric Pressure (32.5 to 37.5 °C)
Hamilton Medical | HAMILTON-C2 Service Manual
239
8
Service Software
8.23 Nurse Call Functional Check
1 Connect the nurse call connector (PN160166) to the HAMILTON-C2 Communication Board.
2 Check the three relays with a multimeter by measuring the resistance during ventilation, alarm off and alarm on between
the following wires:
Figure 8-224. Nurse Call Cable and Connector
Table 8-12. Nurse Call Cable and Connector Callouts
Connector
Description
Color
1
Alarm return normally closed
brown
2
Alarm return normally open
white
3
Alarm common
green
4
I:E Contact 2
yellow
5
I:E Contact 1
grey
6
—
—
Table 8-13. Nurse Call Cable and Connector Actions
Status
green — white
green — brown
yellow — grey
Alarm on
closed
open
depending
Alarm off
open
close
depending
Inspiration
depending
depending
closed
Expiration
depending
depending
open
240
English | 624165/07
Nurse Call Functional Check
8
3. Check the three relays with a multimeter by measuring the resistance during ventilation, alarm off and alarm on between
the following wires:
• Enter Service software > Tests/Calibration > System test > Alarming (See Figure 8-225 Alarm On/Off Relay Check).
• Activate/deactivate alarm (e.g. Alr. High).
Figure 8-225. Alarm On/Off Relay Check
4. I:E relay check:
• Restart the HAMILTON-C2 in operational mode
• Start ventilation
• Check I:E contact with the multimeter
Hamilton Medical | HAMILTON-C2 Service Manual
241
8
Service Software
8.24 Final Tests
1. Install a USB Memory Stick into the HAMILTON-C2 USB Port.
2. Switch on the unit to get into Service Software mode.
3. Select the Download Menu and then Instr Report.
4. Start the Instr Report download.
5. Verify from the USB Memory Stick, that the download is complete.
6. Finish this unit test by saving the Instrument Report with your documentation.
8.25 Set the Service Timer
Set the Service Timer. If the Reset Button is greyed out, the Alarm limit Knob needs to be activated. After doing so, the reset
function is enabled.
8.26 Check the Battery Data
NOTICE
Depending on how frequent you use the ventilator, you can replace the battery if the State of Health (SoH) is < 40%.
Refer to the Evaluation of the Battery State of Health Section 5.3 for details on how to read the battery data.
1.
Check the Battery State of Health (SoH).
2.
Replace the battery if:
– the State of Health (SoH) is < 20%
– the cell voltages are permanently asymmetrical, that is, the difference between cells voltage is more than 100 mV.
3.
Calibrate the battery if the Error is ≥ 5%.
4.
Record the battery data on the Summary of Test Report Section 17.1.
242
English | 624165/07
9
Technical Faults
9.1
Technical Faults Overview.....................................................................................................................244
9.2
Troubleshooting...................................................................................................................................246
243
9
Technical Faults
9.1 Technical Faults Overview
NOTICE
If more than one alarm is active, the associated alarm messages alternate in the message bar.
Alarm Indications
Table 9-1. Alarm Indications in the Ventilator
Alarm Type
Message Bar
Alarm
Lamp
Audio
Action Required
High-priority
alarm
Red, with alarm
message
Red
A sequence of 5 beeps, repeated until
the alarm is reset. If the audible alarm is
not silenced during the first minute, the
continuous-tone buzzer also sounds.
The patient’s safety is compromised.
The patient needs immediate attention.
Medium-priority
alarm
Yellow, with
alarm message
Yellow
A sequence of 3 beeps, repeated periodically. If the audible alarm is not
silenced during the first minute, the
continuous-tone buzzer also sounds.
The patient needs prompt attention.
Low-priority alarm
Yellow, with
alarm message
Yellow
Two sequences of beeps. This is not
repeated.
Operator awareness is required.
Technical fault
Red, with Safety
ventilation: xxxxxx
or Technical fault:
xxxxxx
Red
Same as for high-priority alarm, if technically possible. At the minimum a continuous buzzer tone. The buzzer cannot
be silenced.
The ventilator enters safety mode, or, if
it cannot safely ventilate, the ambient
state. Provide alternative ventilation.
Turn off the ventilator. Have the ventilator serviced.
Figure 9-1. Alarm Class
Table 9-2. Alarm Class
Alarm Codes
Alarm Class
Patient Alarms - 100000 Alarm Codes Section 16.2.1
Patient Alarms
Technical Alarms - 200000 Alarm Codes Section 16.2.2
Technical Events
Technical Failure Ending in Safety Mode - 300000 Alarm Codes Section 16.2.3
Technical Failures resulting in the ventilator switching to the Safety
Mode
Technical Failure Ending in Ambient Mode - 400000 Alarm Codes
Section 16.2.4
Technical Failures resulting in the ventilator switching to the Ambient Mode
500000 Alarm Codes Section 16.2.5
System Failures; invisible TF for user, no action required. For internal
use.
244
English | 624165/07
Technical Faults Overview
9
• Patient Alarms: (Alarm code 100000) Alarms related to : Pressure, Flow, Volume, Apnea
– High Priority Alarms
– Medium Priority Alarms
– Low Priority Alarms
• Technical Alarms: (Alarm code 200000) Alarms related to: Temperature, Battery, Valves, Sensors
– High Priority Alarms
– Medium Priority Alarms
– Low Priority Alarms
Technical Faults Technical Faults end up in one of the following states:
• Fatal -> Safety Mode: (Alarm code 300000)
– In case of some technical failure the ventilator switches to SAFETY Mode. This gives the user time for corrective actions,
such as organizing a replacement ventilator. The turbine runs constant to create Pinsp. The expiratory valve switches
system pressure levels between PEEP and inspiratory pressure. Patient monitoring is nonfunctional during safety ventilation. You must turn off the ventilator power to exit safety ventilation.
• Fatal -> Ambient Mode: (Alarm code 400000)
– If the technical fault alarm is serious enough to possibly compromise safe ventilation, the ventilator enters the ambient
state. The inspiratory channel and expiratory valve are opened, letting the patient breathe room air unassisted. You
must switch off ventilator power to exit the ambient state.
• Invisible Technical Faults (Alarm code 500000) / Technical Note (since < 2.2.0)
– These are classed as “Invisible Technical Faults” and are not shown on the LCD Display during operation. They refer to
exception handling performed by ventilator during operation, and are for Hamilton Medical internal use. Do not consult
Hamilton Medical AG Technical Support concerning these technical faults. They have no significance for hospital workers or Hamilton-Medical Trained Field Service Engineers.
– For detail description go to Glossary - Alarm Codes Section 16.2.
Hamilton Medical | HAMILTON-C2 Service Manual
245
9
Technical Faults
9.2 Troubleshooting
For troubleshooting, refer to document number 624485 "Troubleshooting Platform C" available for download from the
Hamilton Medical Partner-net.
Follow suggested troubleshooting steps in order as described.
Follow these steps when troubleshooting:
1.
Confirm that the test configuration is correct.
2.
When replacing components, repeat the applicable test following each replacement. If the test passes successfully,
repeat all test's in order (see the Service Software Section 8).
3.
If a problem cannot be resolved, enter a Customer Event Report (CER) on the Hamilton Medical AG on the Online Help
Desk (https://hamilton.ky2help.com). Export the Event Log and Instrument Report for failure analysis. Check regularly the
knowledge base of the ky2help for the latest information and troubleshooting.
246
English | 624165/07
10
Components Removal/Assembly
10.1
Notes on Maintenance and Replacement .............................................................................................248
10.2
Securing the Power Cord .....................................................................................................................249
10.3
Interaction Panel Components Removal/Assembly ................................................................................250
10.4
Ventilation Unit Components Removal/Assembly..................................................................................262
10.5
Ventilation Unit Components, Tubings and Cables Assembly................................................................298
247
10
Components Removal/Assembly
10.1 Notes on Maintenance and Replacement
WARNING
You must perform the repairs detailed in this section only as instructed in the Preventive Maintenance and Testing Overview
Section 6.1, Hospital Preventive Maintenance Section 6.2 and Engineer Preventive Maintenance Section 6.3.
Service the HAMILTON-C2 only as described in this manual, using only parts approved or supplied by Hamilton Medical AG.
Incorrectly repaired parts, components or assemblies could result in patient injury. See available Spare Parts in the Spare
Parts Section 12.
− Always switch the HAMILTON-C2 OFF and disconnect the HAMILTON-C2 from the external Mains Power and remove the
battery, before opening the cover.
− Never use any kind of lubrication on any part of the HAMILTON-C2.
− Some parts may be hot after operational use (Blower Module, Rinse Flow Valve, Power Supply, etc.).
CAUTION
After performing Maintenance or Replacement of a component or module, perform the necessary Service Software checks,
calibrations and Safety Tests to ensure the HAMILTON-C2 is performing properly.
Hamilton Medical AG does not permit repairs to parts that are supplied as an assembly. For example:
− You have to replace the Blower Module as a complete assembly
− You have to replace the Expiratory Valve as a complete assembly
− You have to replace the Mixer Block as complete assembly
− You have to replace Hamilton Medical AG printed circuit boards only
Make sure to take full ESD (electrostatic discharge) precautions before opening the HAMILTON-C2. Refer to the ESD (Electrostatic Discharge ) Protection Section 11.3.3.
When the HAMILTON-C2 is switched ON, even when not connected to the external Mains Power, the Battery Pack(s) supply
power. Therefore, a short circuit is possible when the HAMILTON-C2 is switched ON.
NOTICE
Always send defective parts, components or assemblies to Hamilton Medical AG with a completed Return Goods Authorization (RGA) request.
Before making any repairs, remove from the HAMILTON-C2:
− External Mains Power Supply and Batteries
− Oxygen Supply
− Patient Breathing Circuit (if not needed)
− Flow Sensor Tubing (if not needed)
Update the Technical State if necessary after exchanging the parts and restart the ventilator. This must be done before performing the Service Software.
248
English | 624165/07
Securing the Power Cord
10
10.2 Securing the Power Cord
NOTICE
To prevent unintentional disconnection of the Power Cord, make sure it is well seated into the ventilator’s socket and
secured with the Power Cord Security Clamp.
The HAMILTON-C2 does not require protective earth grounding, because it is a class II device, as classified according to IEC
60601-1.
Securing the Power Cord with the suitable Security Clamp.
Secure the Power Cord with the Security Clamp as shown in the picture below.
Two different Security Clamps are available. Use the suitable Security Clamp as listed in the table below.
Figure 10-1. Securing the Power Cord
Table 10-1. Security Clamp
Part Number
361061
Description
Image
Cable Security Clamp D=5.0 for Power
Cords:
• PN 355199 GB
• PN 355200 EU
361097
Cable Security Clamp D=6.2 for Power
Cords:
• PN 355198 USA
• PN 355308 CN
Replacing the Power Cord
To replace the Power Cord, release the screw (PN 420724) of the Security Clamp.
When reconnecting the Power Cord, make sure it is well seated into the ventilator’s socket and secured with the Power Cord
Security Clamp.
Hamilton Medical | HAMILTON-C2 Service Manual
249
10
Components Removal/Assembly
10.3 Interaction Panel Components Removal/Assembly
10.3.1 Ventilation Unit Top Cover Removal/Assembly
NOTICE
The Interaction Panel must be tilted completely to the front to remove the Top Cover.
To remove the Ventilation Unit Top Cover:
1. Lift the Handle of the Ventilation Unit.
2. Remove 6 Torx screws from the Top Cover of the Ventilation Unit:
• A: 2 Torx screws (PN 420682)
• B: 4 Torx screws (PN 420642)
3. Remove the Ventilation Unit Top Cover (C).
Figure 10-2. Ventilation Unit Top Cover Removal
4. Assemble in the reverse order of removal.
250
English | 624165/07
Interaction Panel Small Rear Cover Removal/Assembly
10
10.3.2 Interaction Panel Small Rear Cover Removal/Assembly
WARNING
Torx screws (A) cannot be any longer than the specified size (M3x6), otherwise, damage can be caused to the Interaction
Panel Board.
To remove the Interaction Panel Small Rear Cover (PN 160402):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section 10-2).
2. Remove 4 Torx screws (A, PN 420642) from the Small Rear Cover.
3. Remove the Small Rear Cover (B).
Figure 10-3. Interaction Panel Small Cover Removal
4. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
251
10
Components Removal/Assembly
10.3.3 Interaction Panel Removal/Assembly
To remove the Interaction Panel:
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
2. Remove the Security Strap (C) from the 1 Flexible Flat Cable and disconnect it from the Ventilation Unit Mainboard.
3. Disconnect the other 2 Flexible Flat Cables (D) from the Interaction Panel Board.
NOTICE
The 2 Flexible Flat Cables (D) can more easily be disconnected from the Interaction Panel Board, but can also be disconnected from the Ventilation Unit Mainboard. The ferrite cores for these 2 cables are secured to the top of the front cover,
and are not intended to be removed with the cables.
4. Remove 4 Torx screws (B, PN 420727) from the Hinges of the Interaction Panel.
5. Remove the Interaction Panel (A) from the Ventilation Unit.
6. Place aside in a safe location.
Figure 10-4. Interaction Panel Removal
6. Assemble in the reverse order of removal.
7. Update the Technical State (refer to Modify Tab Section 8.8.3).
252
English | 624165/07
Interaction Panel Large Rear Cover Removal/Assembly
10
10.3.4 Interaction Panel Large Rear Cover Removal/Assembly
To remove the Interaction Panel Large Rear Cover (PN 160326):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Cover Rear Cover Removal/Assembly
Section 10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
3. Remove 5 Torx screws (A, PN 420642) from the Large Rear Cover.
4. Remove the Large Rear Cover (B).
Figure 10-5. Interaction Panel Large Rear Cover Removal
5. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
253
10
Components Removal/Assembly
10.3.5 Backlight Driver Board Removal/Assembly
WARNING
When replacing the GRAPHIC LCD TFT VGA 10.4” NEC Display (Obsolete, PN 380031), order the Display Upgrade Kit
(MSP380046). The DC/AC Backlight Inverter Boards (Obsolete, PN 396197 or PN 396314) are not compatible with the latest
GRAPHIC LCD TFT VGA 10.4" NLT Display. Use the Backlight Driver Board (PN 396314) if the Front Panel Boards is below
Revision 08. Front Panel Boards at Revision 08 and higher have integrated Backlight Driver Components (see the Front Panel
Board Removal/Assembly Section 10.3.6 for configuration instructions).
To remove the DC/AC Backlight Inverter Boards (Obsolete, PN 396197 or PN 396314) or the Backlight Driver Board (PN
396314):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section
10.3.4).
5. Disconnect 2 Cables (A) from the Backlight Driver Board (B).
6. Remove 2 Torx screws (C, PN 420692) from the Backlight Driver Board.
7. Remove the Backlight Driver Board (B).
Figure 10-6. Backlight Inverter Board Removal
8. Assemble in the reverse order of removal.
254
English | 624165/07
Front Panel Board Removal/Assembly
10
10.3.6 Front Panel Board Removal/Assembly
WARNING
The Backlight LED Driver Components are integrated into Front Panel Boards with Revision 08 and higher, compatible with
the latest GRAPHIC LCD TFT VGA 10.4" NLT Display. The installation of an External Backlight Driver Board is not necessary
when you configure Jumper P14 to pins 1 and 2 on the new Front Panel Board. An Extension Cable (PN 161569) is used to
bridge the connections for the removed Backlight Driver Board (see instructions below).
CAUTION
Update the Technical State (refer to Modify Tab Section 8.8.3).
To Remove the Front Panel Board (MSP160196):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section
10.3.4).
5. Disconnect 1 cable (A) from the Backlight Driver Board and 4 cables (B) from the Interaction Panel Board.
Figure 10-7. Interaction Panel Board Removal Step 1
Hamilton Medical | HAMILTON-C2 Service Manual
255
10
Components Removal/Assembly
6. Remove 5 Torx screws (C, PN 420724) and 2 Torx screws (D, PN 420641) from the Interaction Panel Board.
7. Remove the Interaction Panel Board (E).
Figure 10-8. Interaction Panel Board Removal Step 2
8. Note the location of Jumper P14 on the Front Panel Board.
Figure 10-9. Location of Jumper Pins on the Front Panel Board
256
English | 624165/07
Front Panel Board Removal/Assembly
10
9. Connect Jumper P14 to Pins 1 and 2 on the new Front Panel Board if no external Backlight Driver Board is installed. Connect the Extension Cable (PN 161569) from the Backlight Connector to P13 on the Front Panel Board.
Pins 1 and 2 if no External Backlight Driver Board is Installed
Figure 10-10. Pins 1 and 2 if no External Backlight Driver Board is Installed
10. Connect Jumper P14 to Pins 2 and 3 when using an external Backlight Driver Board with the new Front Panel Board.
11. Connect the Backlight Driver Board between the Backlight Cable and P7 Connector on the Front Panel Board.
Figure 10-11. Pins 2 and 3 if using an External Backlight Driver Board
12. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
257
10
Components Removal/Assembly
10.3.7 LCD Display Removal/Assembly
WARNING
When replacing the GRAPHIC LCD TFT VGA 10.4” NEC Display (Obsolete, PN 380027 or PN 380031), order the Display
Upgrade Kit (MSP380046).
To remove the LCD Display (PN 380031 or PN 380046):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section
10.3.4).
5. Remove the Front Panel Board (refer to the Front Panel Board Removal/Assembly Section 10.3.6).
6. Remove 2 Torx screws (A, PN 420641), 1 washer (B, PN 409105) and 2 Hex Standoff screws (C, PN 257038).
7. Remove the LCD Display (D).
Figure 10-12. LCD Display Removal
8. Assemble in the reverse order of removal.
258
English | 624165/07
LCD Display Backlight Removal/Assembly
10
10.3.8 LCD Display Backlight Removal/Assembly
To remove the LCD Display Backlight (PN 380044 / SN 1676 > PN 380030):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section
10.3.4).
5. Push down the small black lever (A) and pull out the LCD Display Backlight (B).
Figure 10-13. LCD Display Backlight Removal Step 1
6. Remove the LCD Display Backlight (B).
Hamilton Medical | HAMILTON-C2 Service Manual
259
10
Components Removal/Assembly
Figure 10-14. LCD Display Backlight Removal Step 2
7. Insert a new LCD Display Backlight.
8. Assemble in the reverse order of removal.
260
English | 624165/07
P&T Control Knob Encoder Removal/Assembly
10
10.3.9 P&T Control Knob Encoder Removal/Assembly
CAUTION
The P&T Control Knob must be positioned to allow free movement. Attach so there is a 3 mm gap between the back of the
Knob and the face of the Interaction Panel. Test the Knob actions for both Press and Turn.
CAUTION
Do not remove the P&T Control Knob for maintenance. Removal of the P&T Control Knob can damage the P&T Control
Knob Encoder. Only remove if necessary to replace a non-functioning P&T Control Knob Encoder.
To remove the P&T (Press and Turn) Knob Encoder (PN 372036):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section
10.3.4).
5. Disconnect the Cable from the Front Panel Board.
6. Remove the P&T Control Knob (A, PN 159449) from the front of the Interaction Panel.
7. Remove the Hex Nut (B) and Washer (C) from the P&T Control Knob Encoder (D).
8. Remove the P&T Control Knob Encoder (D) and Large Washer (E).
Figure 10-15. P&T Control Knob Encoder Removal
9. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
261
10
Components Removal/Assembly
10.4 Ventilation Unit Components Removal/Assembly
10.4.1 Backup Battery Pack Removal/Assembly
CAUTION
It is mandatory that the ventilator is operated with at least one battery installed.
NOTICE
− The Backup Battery Compartment provides space for 2 Backup Battery Packs. The primary Battery Pack is positioned into
the left compartment. An Optional 2nd Battery Pack is positioned into the right compartment.
− A Battery Lock is used to secure the Battery Release Latch. The Battery Release Latch must be in the latched position for
the Battery Door to close.
To remove Backup Battery Pack (MSP369106):
1. Open the Backup Battery Door by depressing the Door Latch (A).
Figure 10-16. Backup Battery Pack Removal Step 1
262
English | 624165/07
Backup Battery Pack Removal/Assembly
10
2. Open the Battery Lock (B) by rotating the Battery Lock counter-clockwise with a screwdriver.
Figure 10-17. Backup Battery Pack Removal Step 2
3. Raise the Battery Release Latch (C) to release the Battery Pack (D) from the Battery Compartment.
4. Remove the Battery Pack (D).
Figure 10-18. Backup Battery Pack Removal Step 3
Hamilton Medical | HAMILTON-C2 Service Manual
263
10
Components Removal/Assembly
5. If a second Battery Pack is used, release the Battery Release Latch (E) and remove the Battery Pack (F) from the Battery
Compartment.
Figure 10-19. Backup Battery Pack Removal Step 4
6. Assemble in the reverse order of removal.
264
English | 624165/07
Filter Cover Removal/Assembly
10
10.4.2 Filter Cover Removal/Assembly
To remove the Filter Cover (PN 160343):
1. Detach from the top and swivel (A) away from the rear of the Ventilation Unit.
2. Remove the Filter Cover (B).
Figure 10-20. Filter Cover Removal
3. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
265
10
Components Removal/Assembly
10.4.3 HEPA Filter Removal/Assembly
To remove the HEPA Filter (PN 160216):
1. Remove the Filter Cover Panel (refer to the Filter Cover Panel Removal/Assembly Section 10.4.2).
2. Raise the HEPA Filter Latch (A, PN 160234).
3. Use the Handle (B) to pull the HEPA Filter from the holder.
4. Remove the HEPA Filter (C).
Figure 10-21. HEPA Filter Removal
5. Assemble in the reverse order of removal.
266
English | 624165/07
High Pressure Oxygen Connector Removal/Assembly
10
10.4.4 High Pressure Oxygen Connector Removal/Assembly
To remove the High Pressure Oxygen Connector, DISS (PN 160470) or NIST (PN 160471):
1. Remove 2 Torx screws (B, PN 420734) mounting the High Pressure Oxygen Connector (A).
2. Remove the High Pressure Oxygen Connector (A).
Figure 10-22. High Pressure Oxygen Connector Removal
3. Remove the O2 Inlet Filter.
4. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
267
10
Components Removal/Assembly
10.4.5 Front and Rear Covers Removal/Assembly
NOTICE
It is not necessary to remove the Interaction Panel from the Front Cover.
To remove the Front (MSP160480) and Rear Covers (MSP160481):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove 2 Phillips screws (B, PN 420730) to remove the Communication Board Cover (A).
5. Remove 4 Torx screws (C, PN 420667) from the back of the Rear Cover (D).
6. Remove the Rear Cover (D).
7. Remove 1 Torx screw (E, PN 420667) from inside the Battery Compartment.
8. Remove the Expiratory Valve Cover.
9. Depress the Trolley Release Front Latch to remove the Front Cover (F).
Figure 10-23. Front and Rear Covers Removal
10. Assemble in the reverse order of removal.
268
English | 624165/07
ESM Module Removal/Assembly
10
10.4.6 ESM Module Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
Export the Technical State (Instrument Report) before removing the Ventilation Unit ESM (refer to the Export Section 8.18.3)
because, after removing the Ventilation Unit ESM, the Technical State will be lost.
To remove the ESM Module (MSP160206):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove 4 screws (A, PN 420864).
3. Remove the top cover of the ESM Module Shield Housing.
4. Remove 4 standoffs (PN 257100) holding the ESM Module (B),
5. With both hands, disconnect the ESM Module (B) from 2 connector sockets (C) on the Mainboard and carefully lift the
ESM Module.
6. Remove the ESM Module (B).
Figure 10-24. ESM Module Removal
7. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
269
10
Components Removal/Assembly
10.4.7 Ventilation Unit Mainboard Removal/Assembly
CAUTION
Take care not to damage the Top Foam Section Latches used to secure the Ventilation Unit Mainboard.
NOTICE
Export the Technical State (Instrument Report) before removing the Ventilation Unit Mainboard (refer to the Export Section
8.18.3) because, after removing the Ventilation Unit Mainboard the Technical State will be lost.
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3). Import the Technical State you have exported before and
change the revision and serial number of the replacement Mainboard.
To remove the Ventilation Unit Mainboard (MSP160200 or MSP160382):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Disconnect all cabling connections from the Ventilation Unit Mainboard.
6. Remove the Ventilation Unit Mainboard (A) from the Top Foam Section.
7. Remove the ESM Holder (PN 257080) and LCD Cable Lock Holder (PN 362051).
Figure 10-25. Ventilation Unit Mainboard Removal
8. Assemble in the reverse order of removal.
9. After entering the new Revision and Serial Number of the Mainboard, you will be asked to enter the Blower Timer and
Operating Hours of the unit. To do this, please delete the description in the input field (Blower % and Device Hour) and enter
the corresponding value in numbers. For example: 5 for 5% Blower Timer / 100 for 100 Operating Hours.
270
English | 624165/07
Top Foam Removal/Assembly
10
10.4.8 Top Foam Removal/Assembly
To remove the Top Foam (PN 160239):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Disconnect all cables connected to the Ventilation Unit Mainboard.
Figure 10-26. Top Foam Removal Step 1
Hamilton Medical | HAMILTON-C2 Service Manual
271
10
Components Removal/Assembly
6. Lift the Top Foam Section (A) from the Ventilation Unit.
Figure 10-27. Top Foam Removal Step 2
7. Assemble in the reverse order of removal.
272
English | 624165/07
Cooling Fan Removal/Assembly
10
10.4.9 Cooling Fan Removal/Assembly
CAUTION
Be careful not to bend the FFC Pins when re-assembling the Blower Module.
NOTICE
The air flow of the Fan is directed into the device. When the power is on, hold a piece of paper against the Fan opening. If
the Fan is installed correctly the paper, will be drawn against the Fan.
To remove the Cooling Fan which includes the small board (PN 160346):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Lift the Cooling Fan (A) from the Middle Foam Section (B) of the Ventilation Unit.
Figure 10-28. Cooling Fan Removal
7. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
273
10
Components Removal/Assembly
10.4.10 Blower Module Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
If Software > 2.2.3 is installed, you have to set the Blower Timer manually, after replacement of the Blower Module
To remove the Blower Module (MSP160250):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Lift the Blower Module (A) from the Middle Foam Section (B) of the Ventilation Unit.
Figure 10-29. Blower Module Removal
7. Assemble in the reverse order of removal.
274
English | 624165/07
Power Supply Removal/Assembly
10
10.4.11 Power Supply Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
To remove the Power Supply (MSP160600):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Lift the Power Supply (A) from the Middle Foam Section (B) of the Ventilation Unit and move away from the back to have
access to the AC Mains Power Cable (C).
7. Disconnect the AC Mains Power Cable (C).
8. Remove the Power Supply (A).
Figure 10-30. Power Supply Removal
9. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
275
10
Components Removal/Assembly
10.4.12 Middle Foam Removal/Assembly
CAUTION
Be sure that all cables are free from the Middle Foam Section.
To remove the Middle Foam (PN 160238):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Unscrew the Tube Flange (A) used to fit the Blower Module (refer to the Two Pin Cylindrical Service Tool, PN 500314 (see
the Two Pin Cylindrical Service Tool Section 11.3.4)).
10. Lift the Middle Foam Section (B) from the Ventilation Unit.
Figure 10-31. Middle Foam Removal Step 1
276
English | 624165/07
Middle Foam Removal/Assembly
10
Figure 10-32. Middle Foam Removal Step 2
11. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
277
10
Components Removal/Assembly
10.4.13 Pressure Sensor Assembly Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
To remove the Pressure Sensor Assembly (MSP160300):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the Tubing (B) from the left side of the Pressure Sensor Assembly (A).
Figure 10-33. Pressure Sensor Assembly Removal Step 1
278
English | 624165/07
Pressure Sensor Assembly Removal/Assembly
10
NOTICE
The tubes are color coded.
11. Lift the Pressure Sensor Assembly (A) slightly to remove the 2 Tubings (B) from the bottom of the Pressure Sensor Assembly.
Figure 10-34. Pressure Sensor Assembly Removal Step 2
12. Remove the Pressure Sensor Assembly (A).
13. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
279
10
Components Removal/Assembly
10.4.14 Inspiratory Valve Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
To remove the Inspiratory Valve (MSP160230):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the Pressure Sensor Assembly (refer to the Pressure Sensor Assembly Removal/Assembly Section 10.4.13).
11. Disconnect 3 Tubings (A) from the Inspiratory Valve, Qvent Flow Sensor and Ambient Valve.
Figure 10-35. Inspiratory Valve Removal Step 1
280
English | 624165/07
Inspiratory Valve Removal/Assembly
10
12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the
Ventilation Unit.
Figure 10-36. Inspiratory Valve Removal Step 2
13. Disconnect the Inspiratory Valve (B) from the Qvent Flow Sensor (C).
14. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
281
10
Components Removal/Assembly
10.4.15 Qvent Flow Sensor Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
To remove the Qvent Flow Sensor (MSP161657 (Sensirion) or MSP399124 (TSI)):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the Pressure Sensor Assembly (refer to the Pressure Sensor Assembly Removal/Assembly Section 10.4.13).
11. Disconnect 3 Tubings (A) from the Inspiratory Valve, Qvent Flow Sensor and Ambient Valve.
Figure 10-37. Inspiratory Valve Removal Step 1
282
English | 624165/07
Qvent Flow Sensor Removal/Assembly
10
12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the
Ventilation Unit.
Figure 10-38. Inspiratory Valve Removal Step 2
13. Disconnect the Inspiratory Valve (B) from the Qvent Flow Sensor (C).
14. Not shown; Remove a metal bracket (PN 160741) attached with 2x screws (PN 420636) secures the Qvent Flow Sensor to
the Ambient Valve.
15. Disconnect the Qvent Flow Sensor (C) from the Ambient Valve (D).
16. Remove the Qvent Flow Sensor (C).
17. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
283
10
Components Removal/Assembly
10.4.16 Ambient Valve Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
To remove the Ambient Valve (MSP160290):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2.
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the Pressure Sensor Assembly (refer to the Pressure Sensor Assembly Removal/Assembly Section 10.4.13).
11. Disconnect all Tubings (A) from the Inspiratory Valve, Qvent Flow Sensor and Ambient Valve.
Figure 10-39. Inspiratory Valve Removal Step 1
284
English | 624165/07
Ambient Valve Removal/Assembly
10
12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the
Ventilation Unit ().
Figure 10-40. Inspiratory Valve Removal Step 2
13. Not shown; Remove a metal bracket (PN 160741) attached with 2x screws (PN 420636) secure the Qvent Flow Sensor to
the Ambient Valve.
14. Disconnect the Qvent Flow Sensor (C) from the Ambient Valve (D).
15. Remove the Ambient Valve (D).
16. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
285
10
Components Removal/Assembly
10.4.17 Oxygen Sensor Block Removal/Assembly
To remove the Oxygen Sensor Block (PN 160100):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Disconnect 2 tubings (A and B) from the Oxygen Sensor Block.
Figure 10-41. Oxygen Sensor Block Removal Step 1
286
English | 624165/07
Oxygen Sensor Block Removal/Assembly
10
11. Lift the Oxygen Sensor Block (C) from the Bottom Foam.
12. Remove the Oxygen Sensor Block (C).
Figure 10-42. Oxygen Sensor Block Removal Step 2
13. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
287
10
Components Removal/Assembly
10.4.18 Rinse Flow Block Removal/Assembly
To remove the Rinse Flow Block (MSP160498):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the top tubing (A) from the Oxygen Sensor Block (refer to the Oxygen Sensor Block Removal/Assembly Section
10.4.17).
11. Remove the Rinse Flow Tube (B) from the Inspiratory Valve.
Figure 10-43. Rinse Flow Block Removal Step 1
288
English | 624165/07
Rinse Flow Block Removal/Assembly
10
12. Tubings (C and E) should be already disconnected from the Pressure Sensor Board.
13. Remove the Nebulizer Tube (D).
14. Disconnect the Qvent Flow Sensor cable and remove from the mount on the Rinse Flow Block.
Figure 10-44. Rinse Flow Block Removal Step 2
15. Lift the Rinse Flow Block (F) from the Bottom Foam.
Hamilton Medical | HAMILTON-C2 Service Manual
289
10
Components Removal/Assembly
Figure 10-45. Rinse Flow Block Removal Step 3
16. Assemble in the reverse order of removal.
290
English | 624165/07
Expiratory Valve Removal/Assembly
10
10.4.19 Expiratory Valve Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
To remove the Expiratory Valve (MSP160240):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Lift the Expiratory Valve Assembly (A) from the Bottom Foam Section of the Ventilation Unit.
Figure 10-46. Expiratory Valve Removal
11. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
291
10
Components Removal/Assembly
10.4.20 Battery Compartment Removal/Assembly
To remove the Battery Compartment (PN 160302):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly 10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly 10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly 10.4.12).
10. Remove 2 Torx screws (C, PN 420699) to remove the Cable Clamp (B, PN 160405) from the rear of the Battery Compartment (A).
Figure 10-47. Battery Compartment Removal Step 1
292
English | 624165/07
Battery Compartment Removal/Assembly
10
6. Lift the Battery Compartment (A) from the Bottom Foam Section of the Ventilation Unit.
Figure 10-48. Battery Compartment Removal Step 2
8. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
293
10
Components Removal/Assembly
10.4.21 Oxygen Mixer Block Assembly Removal/Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3). Also update the Technical State of the QO2 Flow Sensor.
To remove the Oxygen Mixer Block Assembly:
• MSP160608 (with Sensirion QO2 Flow Sensor, MSP161657)
or
• MSP160226 (with TSI QO2 Flow Sensor, MSP399123)
1. Remove the Top Cover from the Ventilation Unit (refer to the Top Cover Ventilation Unit Removal/Assembly 10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly 10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly 10.4.12).
10. Remove the Nebulizer Tubing.
11. Remove 3 Torx screws (C, PN 420657) attaching the Oxygen Mixer Block Assembly (A) to the Base Plate (B).
12. Remove the Oxygen Mixer Block Assembly (A).
Figure 10-49. Oxygen Mixer Block Assembly Removal
294
English | 624165/07
Oxygen Mixer Block Assembly Removal/Assembly
10
13. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
295
10
Components Removal/Assembly
10.4.22 Nebulizer Valve Removal/Assembly
To remove the Nebulizer Valve (PN 160400) from the Mixer Block:
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the Oxygen Mixer Block Assembly (refer to the Oxygen Mixer Block Assembly Removal/Assembly Section
10.4.21).
11. Remove 2 Phillips screws (A) from the Nebulizer Valve (B).
12. Remove the Nebulizer Valve (B).
Figure 10-50. Nebulizer Valve Removal
13. Assemble in the reverse order of removal.
296
English | 624165/07
Bottom Foam Removal/Assembly
10
10.4.23 Bottom Foam Removal/Assembly
To remove the Bottom Foam (PN 160237):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly Section
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9).
7. Remove the Blower Module (refer to the Blower Module Removal/Assembly Section 10.4.10).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12).
10. Remove the Oxygen Mixer Block Assembly (refer to the Oxygen Mixer Block Assembly Removal/Assembly Section
10.4.21).
11. Remove the Bottom Foam (A) from the Ventilation Unit Base Frame.
Figure 10-51. Bottom Foam Removal
12. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual
297
10
Components Removal/Assembly
10.5 Ventilation Unit Components, Tubings and Cables Assembly
10.5.1 Bottom Foam Assembly
1. The Bottom Foam (A, PN 160237) is positioned onto the Chassis Frame (B).
Figure 10-52. Bottom Foam Assembly
298
English | 624165/07
Mains Power Cable Assembly
10
10.5.2 Mains Power Cable Assembly
1. The Mains Power Cable (C, PN 160348) is connected to the Mains Power Connector (A), with one side connected to the
2A Circuit Breaker (D, PN 378009), then routed through a Ferrite Core (B) and positioned on the side of the Mixer Block
Assembly (C).
2. It will then be connected to the Power Supply when assembled.
Figure 10-53. Mains Power Cable Assembly
Hamilton Medical | HAMILTON-C2 Service Manual
299
10
Components Removal/Assembly
10.5.3 DC Input Power Cable Assembly
1. The DC Input Power Cable (B, PN 160372) from the DC Power Connector (A) is positioned on the side of the Mixer Block
Assembly (B) and will be attached to the Battery Compartment (C) when assembled.
2. It will then be connected to the Ventilation Unit Mainboard when assembled.
Figure 10-54. DC Input Power Cable Assembly
300
English | 624165/07
Oxygen Mixer Block Assembly Mounting
10
10.5.4 Oxygen Mixer Block Assembly Mounting
1. The Oxygen Mixer Block Assembly (A) is mounted to the Chassis (B) with 3 Torx screws (C, 2x PN 420667 and 1x PN
420727).
Figure 10-55. Oxygen Mixer Block Assembly Mounting
Hamilton Medical | HAMILTON-C2 Service Manual
301
10
Components Removal/Assembly
10.5.5 Oxygen Mixer Valve Cable Assembly
1. The Oxygen Mixer Valve Cable (D) (part of the Oxygen Mixer Valve, no part number) from the Oxygen Mixer Valve (A) is
positioned over the Low Pressure Oxygen Input Tubing (B) and behind the 2A Circuit Breaker (C), then placed to the side of
the Mixer Block Assembly (D).
Figure 10-56. Oxygen Mixer Valve Cable Assembly
2. The Cable will route on the side of the Middle and Top Foam Sections when assembled.
3. It will then be connected to the Ventilation Unit Mainboard when assembled.
302
English | 624165/07
Nebulizer Cable Assembly
10
10.5.6 Nebulizer Cable Assembly
1. The Nebulizer Cable (A) (part of the Nebulizer Valve, no part number) from the Nebulizer Valve is positioined in a channel
in the Bottom Foam Section (B) and routed to the right side of the Ventilator (C).
Figure 10-57. Nebulizer Cable Assembly Top View
2. The Cable will route on the side of the Middle and Top Foam Sections when assembled.
3. It will then be connected to the Ventilation Unit Mainboard when assembled.
Hamilton Medical | HAMILTON-C2 Service Manual
303
10
Components Removal/Assembly
10.5.7 Battery Compartment Assembly
1. The Battery Compartment (A) is positioned at the rear of the Bottom Foam Section.
2. The Battery Power Cable (B, PN 160350) and the Battery Data FFC (C, PN 160351) are routed to the rear of the Battery
Compartment (D). They are secured along with the DC Input Power Cable (E, PN 160372) to the rear of the Battery Compartment with the Cable Clamp (F, PN 160405) and 2 Torx screws (G, PN 420699).
Figure 10-58. Battery Compartment Assembly
3. It will then be connected to the Ventilation Unit Mainboard when assembled.
304
English | 624165/07
Expiratory Valve Assembly
10
10.5.8 Expiratory Valve Assembly
1. The Expiratory Valve (A, MSP160240) is positioned with 3 slots (C) into the Bottom Foam Section (B).
Figure 10-59. Expiratory Valve Assembly Step 1
2. The Expiratory Valve FFC (D) (part of the Expiratory Valve, no part number) is routed under a channel in the Bottom Foam
Section to the side.
Figure 10-60. Expiratory Valve Assembly Step 2
3. The Cable will route on the side of the Bottom, Middle and Top Foam Sections when assembled.
4. It will then be connected to the Ventilation Unit Mainboard when assembled.
Hamilton Medical | HAMILTON-C2 Service Manual
305
10
Components Removal/Assembly
10.5.9 Rinse Flow Block and Tubing Assembly
1. Insert the Rinse Flow Block (A, MSP160498) into the Bottom Foam Section.
Figure 10-61. Rinse Flow Block and Tubing Assembly Step 1
2. Connect the Proximal Flow Sensor Tubing (B, from Tubing Set MSP160476) and the Distal Flow Sensor Tubing (D, from
Tubing Set MSP160476) to the Rinse Flow Block.
3. Connect the Nebulizer Tubing (C, PN 160409) from the Mixer Block Fitting and the Nebulizer Connector to the Rinse Flow
Block.
5. The other tubing ends (B and D) will connect to the Pressure Sensor Assembly when assembled.
306
English | 624165/07
Rinse Flow Block and Tubing Assembly
10
Figure 10-62. Rinse Flow Block and Tubing Assembly Step 2
Hamilton Medical | HAMILTON-C2 Service Manual
307
10
Components Removal/Assembly
4. If removed, connect the Oxygen Sensor Tubing (E).
3. Connect the tubing (F) to the Inspiratory Valve which contains a Flow Restrictor.
Figure 10-63. Rinse Flow Block and Tubing Assembly Step 3
308
English | 624165/07
Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly
10
10.5.10 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
The Ambient Valve cable (F) is part of the Ambient Valve Assembly and has no part number. The Inspiratory Valve FFC (G) is
part of the Inspiratory Valve and has no part number.
1. Pre-assemble the Inspiratory Valve (A, MSP160230), Qvent Flow Sensor (B, MSP161657 (Sensirion) or MSP399124 (TSI)),
Ambient Valve (C, MSP160290), Shaped Tube (D, PN 160223) and Patient Connection Tube (E, PN 160295).
2. Not shown; a metal bracket (PN 160741) attached with 2x screws (PN 420636) secures the Qvent Flow Sensor to the
Ambient Valve.
Figure 10-64. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 1
3. Insert the complete assembly (I) into the Bottom Foam Section (J).
Hamilton Medical | HAMILTON-C2 Service Manual
309
10
Components Removal/Assembly
Figure 10-65. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 2
4. Connect the tubing (K) from the Front Panel Connector Block to the Inspiratory Valve Assembly.
5. Place the Qvent Flow Sensor cable into the Cable Holder (L) located on top of the Front Panel Connector Block.
6. The cable will route on the side of the Middle and Top Foam Sections when assembled.
7. It will then be connected to the Ventilation Unit Mainboard when assembled.
310
English | 624165/07
Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly
10
8. The Inspiratory Valve FFC (M) (part of the Inspiratory Valve, no part number) is routed under a channel in the Bottom Foam
Section to the side.
Figure 10-66. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 3
9. The cable will route on the side of the Bottom, Middle and Top Foam Sections when assembled.
10. It will then be connected to the Ventilation Unit Mainboard when assembled.
11. The cable from the Ambient Valve Assembly (part of the Ambient Valve, no part number) is routed through a channel (N)
in the Bottom Foam Section, over the Shaped Tube (O) and through another channel (P) to the side.
Figure 10-67. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 4
12. The cable will route on the side of the Middle and Top Foam Sections when assembled.
13. It will then be connected to the Ventilation Unit Mainboard when assembled.
Hamilton Medical | HAMILTON-C2 Service Manual
311
10
Components Removal/Assembly
10.5.11 Oxygen Sensor Block and Tubing Assembly
1. Insert the Oxygen Sensor Block (A, PN 160298) into the Bottom Foam Section (B).
Figure 10-68. Oxygen Sensor Block and Tubing Assembly Step 1
2. Connect the Oxygen Sensor Cable (D, PN 160354) to the Oxygen Sensor (C, PN 396200).
3. The Cable will route on the side of the Middle and Top Foam Sections when assembled.
4. It will then be connected to the Ventilation Unit Mainboard when assembled.
5. Connect the tubing (E, PN 160475) from the side of the Oxygen Sensor Block to the Inspiratory Valve.
Figure 10-69. Oxygen Sensor Block and Tubing Assembly Step 2
6. Connect the tubing (F, PN 160475) from the top of the Oxygen Sensor Block, through the T-Connector (H, PN 279865),
through tubing (I, PN 160475) (routed through a channel in the Bottom Foam Section) to the top of the Ambient Valve (J).
7. The other side of the T-Connector (H) is connected to a tubing (G, PN 160475) and will be connected to the Pressure Sensor Assembly when assembled.
312
English | 624165/07
Pressure Sensor Module Assembly
10
10.5.12 Pressure Sensor Module Assembly
1. Insert the Pressure Sensor Assembly (B, MSP160300) above the Ambient Valve.
2. Connect the 2 Pflowsensor tubings (A) to the bottom of the Pressure Sensor Assembly (Tubes are color coded).
3. Connect the tubing (C) from the Y-Connector to the side of the Pressure Sensor Assembly.
4. Position the Pressure Sensor Assembly FFC (D, PN 160355) along the top of the Shaped Tubes.
5. The cable will route on the side of the Middle and Top Foam Sections when assembled.
6. The cable will then be connected to the Ventilation Unit Mainboard when assembled.
Figure 10-70. Pressure Sensor Module Assembly
Hamilton Medical | HAMILTON-C2 Service Manual
313
10
Components Removal/Assembly
10.5.13 Middle Foam Assembly
1. Position the Middle Foam (A, PN 160238) onto the Bottom Foam.
Figure 10-71. Middle Foam Assembly Step 1
2. Assemble the Tube Flange (C, PN 160287) and screw onto the Shaped Tubes from the QO2 Flow Sensor.
NOTICE
Do not tighten the Tube Flange too tight.
Figure 10-72. Middle Foam Assembly Step 2
314
English | 624165/07
Power Supply Assembly
10
10.5.14 Power Supply Assembly
1. Insert the Power Supply (A, MSP160600) into the rear of the of the Middle Foam Section (B).
2. Connect the Mains Power Cable (C) to the Power Supply.
3. Connect the Power Supply Cable (D, PN 160539) to the Power Supply.
4. The cables will route on the side of the Middle and Top Foam Sections when assembled.
5. The cables will then be connected to the Ventilation Unit Mainboard when assembled.
NOTICE
To prevent noise in operation by vibrations, look out for correct implementation of the Power Supply.
Figure 10-73. Power Supply Assembly
Hamilton Medical | HAMILTON-C2 Service Manual
315
10
Components Removal/Assembly
10.5.15 Cooling Fan Assembly
1. Insert the Cooling Fan (A, PN 160346) into the rear of the Middle Foam Section.
NOTICE
Observe the air flow direction of the Fan. Flow direction is into the ventilator.
2. Position the Fan 12V Cable to the side (B).
3. The cable will route on the side of the Top Foam Section when assembled.
4. The cable will then be connected to the Ventilation Unit Mainboard when assembled.
Figure 10-74. Cooling Fan Assembly
316
English | 624165/07
Blower Module Assembly
10
10.5.16 Blower Module Assembly
CAUTION
Before installing the Blower Module, ensure that the P-Filter cable will not be damaged.
CAUTION
Update the Technical State (refer to the Modify Tab Section Modify Tab).
NOTICE
To prevent noise in operation by vibrations, look out for correct implementation of the Blower Module.
1. Insert the Blower Module (A, MSP160250) into the Middle Foam Section (B).
Figure 10-75. Blower Module Assembly Step 1
Hamilton Medical | HAMILTON-C2 Service Manual
317
10
Components Removal/Assembly
2. Position the Blower Cable (C) (part of the Blower Module, no part number) and the Temperature Sensor FFC (D, PN
160353) to the side.
Figure 10-76. Blower Module Assembly Step 2
3. The cables will route on the side of the Top Foam Section when assembled.
4. The cables will then be connected to the Ventilation Unit Mainboard when assembled.
5. Position the Filter Pressure Sensor Cable (E) into the cutout of the Middle Foam Section.
Figure 10-77. Blower Module Assembly Step 3
6. The Filter Pressure Sensor Cable will route on the side of the Top Foam Section when assembled.
7. The cable will then be connected to the Ventilation Unit Mainboard when assembled.
318
English | 624165/07
Top Foam and Mainboard Assembly
10
10.5.17 Top Foam and Mainboard Assembly
CAUTION
Update the Technical State (refer to the Modify Tab Section 8.8.3).
1. Position the Top Foam (A, PN 160239) onto the Middle Foam (B).
Figure 10-78. Top Foam and Mainboard Assembly Step 1
2. Insert the Mainboard (C, MSP160200 or MSP160382) onto the Top Foam.
Figure 10-79. Top Foam and Mainboard Assembly (Front View) Step 2
Hamilton Medical | HAMILTON-C2 Service Manual
319
10
Components Removal/Assembly
3. Route the Fan 12V Cable (D) into the side of the Top Foam and plug into Connector P28 on the Mainboard.
Figure 10-80. Top Foam and Mainboard Assembly (Left Side View) Step 3
4. Route the Blower Cable (E) into the side of the Top Foam and plug into Connector P23 on the Mainboard.
5. Route the Temperature Sensor FFC (F) into the side of the Top Foam and plug into Connector J21 on the Mainboard.
6. Route the Inspiratory Valve FFC (G) into the Bottom, Middle and Top Foam and plug into Connector J22 on the Mainboard.
7. Route the Oxygen Sensor Cable (H) into the front of the Middle and Top Foam and plug into Connector P20 on the Mainboard.
Figure 10-81. Top Foam and Mainboard Assembly (Front View) Step 4
8. Route the Qvent Flow Sensor Cable (I) into the front of the Middle and Top Foam and plug into Connector P19 on the
Mainboard.
9. Route the Pressure Sensor Board FFC (J) into the front of the Middle and Top Foam and plug into Connector J17 on the
Mainboard.
320
English | 624165/07
Top Foam and Mainboard Assembly
10
10. Route the Nebulizer Valve Cable (K) into the side of the Middle and Top Foam and plug into Connector P13 on the Mainboard.
Figure 10-82. Top Foam and Mainboard Assembly (Right Side View) Step 5
11. Route the Ambient Valve Cable (L) into the side of the Middle and Top Foam and plug into Connector P12 on the Mainboard.
12. Route the Expiratory Valve FFC (M) into the side of the Bottom, Middle and Top Foam and plug into Connector J11 on
the Mainboard.
13. Route the O2 Valve Cable (N) into the side of the Middle and Top Foam and plug into Connector P7 on the Mainboard.
14. Route the Filter Pressure Sensor FFC (O) into the side of the Middle and Top Foam and plug into Connector J9 on the
Mainboard.
15. Route the Binary Valve FFC (P) from the Pressure Sensor Assembly into the side of the Middle and Top Foam and plug
into Connector J8 on the Mainboard.
16. Route the O2 Flow Sensor Cable (Q) into the side of the Middle and Top Foam and plug into Connector P10 on the
Mainboard.
Hamilton Medical | HAMILTON-C2 Service Manual
321
10
Components Removal/Assembly
17. Route the Battery Power Cable (R) into the rear of the Middle and Top Foam and plug into Connector P6 on the Mainboard.
Figure 10-83. Top Foam and Mainboard Assembly (Rear View) Step 6
18. Route the DC Input Cable (S) into the rear of the Middle and Top Foam and plug into Connector P5 on the Mainboard.
19. Route the Battery Data FFC (T) into the rear of the Middle and Top Foam and plug into Connector J4 on the Mainboard.
20. Route the Fan Supply Cable (U) into the rear of the Middle and Top Foam and plug into Connector P25 on the Mainboard.
21. Route the Power Supply Cable (V) into the rear of the Middle and Top Foam and plug into Connector P3 on the Mainboard.
22. Assemble the Front and Rear Covers (refer to the 10.4.5).
23. Assemble the Interaction Panel (refer to the Interaction Panel Removal/Assembly Section 10.3.3).
24. Assemble the Ventilation Unit Top Cover (refer to the 10.3.1).
322
English | 624165/07
11
Maintenance Tools and Test Equipment
11.1
Overview..............................................................................................................................................324
11.2
Standard Tools .....................................................................................................................................324
11.3
Special Tools ........................................................................................................................................325
11.4
Test Equipment ....................................................................................................................................328
11.5
Calibration of Test Equipment ..............................................................................................................332
323
11
Maintenance Tools and Test Equipment
11.1 Overview
Standard Tools, Special Tools, ElectroStatic Discharge (ESD) protection and Test Equipment detailed in the following sections
are required to carry out:
• The Preventive Maintenance procedures in the Engineer Preventive Maintenance Section 6.3
• The Tests Functions in the Service Software Section 8
• The Component Replacements in the Components Removal/Assembly Section 10
11.2 Standard Tools
To perform basic maintenance on equipment from Hamilton Medical AG, you require a range of:
• Screwdrivers (Torx, flat and cross-head)
• Metric Spanners (wrenches)
• Metric Hex (Allen) Keys (wrenches)
324
English | 624165/07
Special Tools
11
11.3 Special Tools
11.3.1 Digital Voltmeter
To perform the tests (Internal Resistance Checks, refer to Internal Resistance Checks Section 7.2.4) a Digital Voltmeter (DVM)
is required for measuring voltage (to a tolerance of 0.5%) or resistance (to a tolerance of 1 (less than 1 Ohm).
Figure 11-1. Digital Voltmeter
11.3.2 Electrical Safety Tester
Electrical Safety Testing is required according to IEC-62353. A RIGEL 288, or similar equipment is required.
Figure 11-2. Rigel 288 Safety Analyzer
Hamilton Medical | HAMILTON-C2 Service Manual
325
11
Maintenance Tools and Test Equipment
11.3.3 ESD (ElectroStatic Discharge) Protection
ESD (Electrostatic Discharge) equipment must be used to prevent damage to sensitive electronic circuits. Typically, this
comprises:
• ESD grounding (earthing) cable and wrist band connected to the ventilator for use when working inside the ventilator
• 11-3 shows a wrist band and connecting cable complete with the crocodile clip that must be attached to the work surface
Figure 11-3. ESD Wrist Strap and Cable (PN 500009)
Figure 11-4. ESD Service Kit (PN 500077)
326
English | 624165/07
Two Pin Cylindrical Service Tool
11
11.3.4 Two Pin Cylindrical Service Tool
The Two Pin Cylindrical Service Tool (PN 500314) is required for the removal of the oxygen tube adapter ring before you can
remove the middle foam section. For details, see the Middle Foam Removal/Assembly Section 10.4.12.
Figure 11-5. Two Pin Cylindrical Service Tool (PN 500314)
Hamilton Medical | HAMILTON-C2 Service Manual
327
11
Maintenance Tools and Test Equipment
11.4 Test Equipment
The following test equipment is required, as well as the tools listed above, to complete the test's and adjustments included in
the Service Software Section 8.
Pictures and details of most of these items are in the product catalog.
Table 11-1. Test Equipment
Part Number
500058
Description
Photo
Pressure Measurement Kit
• A complete WIKA gauge set can be
obtained from Hamilton Medical AG
Pressure gauge with the following
specifications:
• Range: 0 – 400 mbar accuracy: 0.5%
500084
TSI Flow Meter Kit
Contains:
• 500308 TSI Flow meter
• 500085 TSI-Flow meter Battery box
• 500086 TSI-Flow meter Soft carrying
case
• 279204 Bacteria filter 260100 Silicone tube 30cm 22F
500330
Hand Pump
-
Digital Voltmeter
(With red and black crocodile clips)
260206
Coaxial Adult Breathing Circuit
260241
Neonatal Breathing Circuit
328
English | 624165/07
Test Equipment
Part Number
Description
281637
Flow Sensor Adult/Pediatric (single use)
155500
Flow sensor Neonatal (single use)
279962
Adapter for Flow Sensor Calibration
Neonatal
279928
Adapter for Flow Sensor Calibration
Adult/Pediatric
281717
Stopper (used for test configurations)
500300
Pressure Connector
279812
Clamp, Flow Restrictor
500077
ESD Wrist Band
Hamilton Medical | HAMILTON-C2 Service Manual
11
Photo
329
11
Maintenance Tools and Test Equipment
Part Number
Description
151815
Adult Demo Lung with 7mm ET Tube
281420
Connector 22M/15F-22M/15F
279913
Connector 15M/4M
7249057
2 Meter Silicone Tube 4mm ID, 7mm
OD
279963
Patient Filter HMEF, Adult/Pediatric
161618
Electrical Safety Test Cable for USB Port
159171
Electrical Safety Test Cable CO2 Option
-
Safety Analyzer (Electric Safety Tester)
330
Photo
English | 624165/07
Test Equipment
Part Number
Description
-
Safety Analyzer Hand Held Device (Electric Safety Tester)
396207
USB Memory Stick
11
Photo
Specification for the USB stick:
• File System: FAT or FAT32
• Unpartitioned Memory
• No Operating System or Security
Software Installed
369104
External Battery Charger
159172
CO2 Gas Regulator
For CO2 Sensor Accuracy Check (if
CO2 monitor option is installed)
159173
CO2 Calibration Gas
For CO2 Sensor Accuracy Check (if
CO2 monitor option is installed).
281803
Adapter (three pieces required for CO2
Accuracy Check)
Hamilton Medical | HAMILTON-C2 Service Manual
331
11
Maintenance Tools and Test Equipment
11.5 Calibration of Test Equipment
Some test equipment must be tested and calibrated periodically. Hamilton Medical recommends the following schedule:
Table 11-2. Calibration of Test Equipment
Item
Schedule
Action
Pressure Gauge
As recommended by the manufacturer,
or at least once per year.
Send the pressure gauge back to the manufacturer
for testing. (for example, www.wika.com for calibration information).
Digital Voltmeter (DVM)
As recommended by the manufacturer.
As recommended by manufacturer.
Flow Analyzer
As recommended by the manufacturer.
As recommended by manufacturer.
332
English | 624165/07
12
Spare Parts
12.1
Introduction to Major Components ......................................................................................................334
12.2
Trolley PN 160170................................................................................................................................335
12.3
Trolley PN 160150................................................................................................................................336
12.4
Major Components of the Interaction Panel .........................................................................................337
12.5
Major Components of the Ventilation Unit ...........................................................................................338
12.6
Interaction Panel Parts..........................................................................................................................343
12.7
Ventilation Unit Parts ...........................................................................................................................348
12.8
Detailed Parts Descriptions ...................................................................................................................355
333
12
Spare Parts
12.1 Introduction to Major Components
This section lists replacement parts that are available for the HAMILTON-C2. The first section contains diagrams that enable
you to locate major components. Later sections offer much more detailed listings.
NOTICE
− For information about consumables (such as tubing) and complete assemblies (such as the Patient Tubing Support Arm),
see the Hamilton Medical Product Catalog (PN 689060). It can be located on the Hamilton Medical AG Partner-net.
− Some photographs shown in later sections may not be displayed to scale.
− Some Spare Parts have a prefix MSP (Medical Spare Part). If you exchange a MSP Part, you usually have to update the
Technical State of the device with the new Serial Number and Revision Number.
Figure 12-1. HAMILTON-C2
A
Ventilation Unit (PN 160001)
See the Ventilation Unit Parts Section 12.7
334
B
Interaction Panel (PN 160325)
See the Interaction Panel Parts Section 12.6
English | 624165/07
Trolley PN 160170
12
12.2 Trolley PN 160170
Figure 12-2. Trolley PN 160170
A
Hose Clamp (PN 160982)
C
Clip Tube Holder (PN 160156)
B
Support Arm HAMILTON-C2 (PN 160153)
D
HAMILTON-C2 Trolley Wheel (MSP282637)
Hamilton Medical | HAMILTON-C2 Service Manual
335
12
Spare Parts
12.3 Trolley PN 160150
Figure 12-3. Trolley PN 160150
A
Head Blue (MSP160606)
D
Support Arm HAMILTON-C2 (PN 160153)
B
Clip Tube Holder (PN 160156)
E
Head White (MSP160607)
C
Tube Clamp (PN 160982)
F
HAMILTON-C2 Trolley Handle Complete (MSP160621)
336
English | 624165/07
Major Components of the Interaction Panel
12
12.4 Major Components of the Interaction Panel
Figure 12-4. Major Components of the Interaction Panel
LCD Display Kit (MSP380046)
D
Backlight Inverter Board (Obsolete)
See the Backlight Driver Board Section 12.6.2
A
LCD Display (PN 380046)
E
See the LCD Display and Touch Screen Section 12.6.4
B
Display Front (MSP160362)
See the Interaction Panel Covers Section 12.6.1
F
See the LCD Display and Touch Screen Section 12.6.4
C
P&T Control Knob Complete (MSP160328)
See the P&T Control Knob Section 12.6.5
Hamilton Medical | HAMILTON-C2 Service Manual
Interaction Panel Small Rear Cover (PN 160402)
Interaction Panel Large Rear Cover (PN 160326)
See the Interaction Panel Covers Section 12.6.1
G
Front Panel Board (MSP160196)
See the Front Panel Board Section 12.6.3
337
12
Spare Parts
12.5 Major Components of the Ventilation Unit
Figure 12-5. Major Components of the Ventilation Unit
A
Ventilation Unit Covers
C
See the Ventilation Unit Covers Section 12.5.1
B
Ventilation Unit Top Section
See the Ventilation Unit Top Section 12.5.2
338
Ventilation Unit Middle Section
See the Ventilation Unit Middle Section 12.5.3
D
Ventilation Unit Bottom Section
See the Ventilation Unit Bottom Section 12.5.4
English | 624165/07
Ventilation Unit Covers
12
12.5.1 Ventilation Unit Covers
Figure 12-6. Ventilation Unit Covers
A
Ventilation Unit Filter Cover (PN 160343)
D
See the External Covers and External Hardware Section
External Covers and External Hardware
B
Complete Rear Cover (MSP160481)
See the External Covers and External Hardware Section
External Covers and External Hardware
E
See the External Covers and External Hardware Section
External Covers and External Hardware
C
Battery Cover (PN 160506)
See the External Covers and External Hardware Section
External Covers and External Hardware
Hamilton Medical | HAMILTON-C2 Service Manual
Ventilator Unit Top Cover (PN 160601)
Front Cover (MSP160480)
See the External Covers and External Hardware Section
External Covers and External Hardware
F
Ventilation Unit Base
See the Metal Brackets and Frame Components Section
Metal Brackets and Frame Components
339
12
Spare Parts
12.5.2 Ventilation Unit Top Section
Figure 12-7. Ventilation Unit Top Section
A
Mainboard-C2 (MSP160382)
See the Electronic Printed Circuit Boards Section 12.8.11
B
C
Top Foam (PN 160239)
See the Metal Brackets and Frame Components Section
12.8.9
Embedded System Module (ESM) (MSP160206)
See the Electronic Printed Circuit Boards Section 12.8.11
340
English | 624165/07
Ventilation Unit Middle Section
12
12.5.3 Ventilation Unit Middle Section
Figure 12-8. Ventilation Unit Middle Section
A
Fan 12 V (PN 160346)
D
See the Electrical/Electronic Devices Section 12.8.12
B
Blower Module (MSP160250)
See the Pneumatic Parts and Assemblies Section 12.8.4
C
Hepa Filter Assembly (PN 160216)
See the Pneumatic Parts and Assemblies Section 12.8.4
E
Service Power Supply (MSP160600)
See the Electronic Printed Circuit Boards Section 12.8.11
Middle Foam (PN 160238)
See the Metal Brackets and Frame Components Section
12.8.9
Hamilton Medical | HAMILTON-C2 Service Manual
341
12
Spare Parts
12.5.4 Ventilation Unit Bottom Section
Figure 12-9. Ventilation Unit Bottom Section
A
Battery Compartment (PN 160302)
G
See the Backup Battery Pack Compartment Section 12.7.5
B
Inspiratory Valve Complete (MSP160230)
See the Electronic Printed Circuit Boards Section 12.8.11
H
See the Assembled Components Section 12.8.6
C
Qvent Flow Sensor (MSP161657, Sensirion) or (MSP399124,
TSI)
Pressure Sensor Assembly (MSP160300)
Mixer Block Assembly (MSP160608 or MSP160226)
See the Mixer Block Assembly Section 12.7.7
I
QO2 Flow Sensor
See the Pneumatic Parts and Assemblies Section 12.8.4
See the Flow Restrictors and Flow Sensors Section 12.8.5
D
Oxygen Sensor HAMILTON-C2 (coded) (PN 396200)
J
See the Electrical/Electronic Devices Section 12.8.12
E
Rinse Flow Assembly (MSP160498)
See the Assembled Components Section 12.8.6
F
Expiratory Valve Assembly (MSP160240)
See the Expiratory Valve Assembly Section 12.7.6
K
Ambient Valve Complete (MSP160290)
See the Assembled Components Section 12.8.6
Bottom Foam (PN 160237)
See the Metal Brackets and Frame Components Section
12.8.9
342
English | 624165/07
Interaction Panel Parts
12
12.6 Interaction Panel Parts
12.6.1 Interaction Panel Covers
Complete Interaction Panel MSP160325
Figure 12-10. Interaction Panel Covers
A
Interaction Panel Large Rear Cover (PN 160326)
D
See the External Covers and External Hardware Section
12.8.13
B
5x Screws (PN 420642)
See the Screws Section 12.8.14
C
Interaction Panel Small Rear Cover (PN 160402)
See the External Covers and External Hardware Section
12.8.13
E
4x Screws (PN 420642)
See the Screws Section 12.8.14
Alarm Lamp (PN 160329)
See the External Covers and External Hardware Section
12.8.13
Hamilton Medical | HAMILTON-C2 Service Manual
343
12
Spare Parts
12.6.2 Backlight Driver Board
Figure 12-11. Backlight Driver Board
A
DC/AC Inverter Board (PN 396197 (Obsolete) or PN 396229
(Obsolete)
D
2x Screws (PN 420663)
See the Screws Section 12.8.14
Backlight LED Driver Board (PN 396314, no longer available)
See the Electronic Printed Circuit Boards Section 12.8.11
B
C
2x Screws (PN 420692)
E
Cable (PN 160358) to the DC/AC Inverter Board
See the Screws Section 12.8.14
See the Electrical/Electronic Cables Section 12.8.10
2x Standoffs (PN 257076)
Bypass Cable (PN 161569) for Front Panel Board Revision 08
or higher for the LED Backlights
See the Standoffs Section 12.8.15
See the Electrical/Electronic Cables Section 12.8.10
NOTICE
Please check compatibility for HAMILTON-C2 Serial Number <1676, see Interaction Panel Section 15.2.1
344
English | 624165/07
Front Panel Board
12
12.6.3 Front Panel Board
Figure 12-12. Front Panel Board
A
Front Panel Board (MSP160196)
E
See the Electronic Printed Circuit Boards Section 12.8.11
B
5x Screws (PN 420724)
FFC Cable to Key Panel (PN 160356)
See the Electrical/Electronic Cables Section 12.8.10
F
Speaker Mount (PN 160529)
G
Loudspeaker (PN 161532)
See the Screws Section 12.8.14
C
1x Screw (PN 420641)
See the Screws Section 12.8.14
D
See the Electrical/Electronics Devices Section 12.8.12
Bypass Cable (PN 161569) to LED Backlights
See the Electrical/Electronic Cables Section 12.8.10
NOTICE
Please see Interaction Panel Section 15.2.1 for Display compatibility.
Hamilton Medical | HAMILTON-C2 Service Manual
345
12
Spare Parts
12.6.4 LCD Display and Touch Screen
Figure 12-13. LCD Display and Touch Screen
A
Display Front (MSP160362)
I
See the Electronic Printed Circuit Boards Section 12.8.11
B
Display Mounting Bracket (PN 160404)
See the Metal Brackets and Frame Components Section
12.8.9
J
See the Metal Brackets and Frame Components Section
12.8.9
C
4x Screws (PN 420724)
Cable Holder (PN 160488)
K
Graphic LCD TFT VGA 10.4” NEC (PN 380027 or 380031
(Obsolete))
USB Cover (PN 341548)
See the Clamps and Fasteners Section 12.8.2
L
See the Clamps and Fasteners Section 12.8.2
E
FDC Cable to Display (PN 160357)
See the Electrical/Electronic Cables Section 12.8.10
See the Screws Section 12.8.14
D
Display Gasket (PN 160341)
Symbol Insert
See the Electronic Printed Circuit Boards Section 12.8.11
M
2x Ferrite Sleeves (PN 393113)
N
2x Cable Clips (PN 393114)
Graphic LCD TFT VGA 10.4" NLT (PN 380046, new)
See the Electronic Printed Circuit Boards Section 12.8.11
F
3x Screws (PN 420641)
see the Screws Section 12.8.14
G
2x Washers (PN 409105)
See the Washers Section 12.8.16
H
1x Standoff (PN 257038)
ESD Isolation Sticker on Monitor (PN 160467)
See the Standoffs Section 12.8.15
NOTICE
Refer to KB ID 3315 if you have to exchange an old version of LCD Display (PN 380027 or PN380031), or an old version of
DC/AC Inverter Board (PN 396197 or PN 396229)
Please see Interaction Panel Section 15.2.1 for Display compatibility.
346
English | 624165/07
P&T Control Knob Encoder
12
12.6.5 P&T Control Knob Encoder
Figure 12-14. P&T Control Knob Encoder
A
P&T Control Knob complete (MSP160328)
See the Colored Metal Rings and Controls Section 12.8.7
B
C
Washer (PN 409908)
See the Washers Section 12.8.16
P&T Control Knob Encoder (PN 372036)
See the Electrical/Electronic Devices Section 12.8.12
Hamilton Medical | HAMILTON-C2 Service Manual
347
12
Spare Parts
12.7 Ventilation Unit Parts
12.7.1 Top Cover
Figure 12-15. Ventilation Unit Top Cover
A
Ventilator Unit Top Cover (PN 160601)
See the External Covers and External Hardware Section
12.8.13
B
C
2x Screws (PN 420682)
See the Screws Section 12.8.14
4x Screws (PN 420642)
See the Screws Section 12.8.14
348
English | 624165/07
Front Cover
12
12.7.2 Front Cover
Figure 12-16. Ventilation Unit Front Cover
A
Ventilation Unit Front Cover with Label (MSP160480)
G
See the External Covers and External Hardware Section
12.8.13
B
1x Screw (PN 420667)
See the Metal Brackets and Frame Components Section
12.8.9
H
See the Screws Section 12.8.14
C
Interaction Panel Hinge Type A Left (PN 281781)
2x Screws (PN 420727)
I
2x Screws (PN 420657)
J
2x Screw Nuts (PN 420159)
See the Screws Section 12.8.17
Hamilton Medical | HAMILTON-C2 Service Manual
2x Screw Nuts (PN 421059)
See the Nuts Section 12.8.17
K
See the Screws Section 12.8.14
F
2x Screws (PN 420657)
See the Screws Section 12.8.14
See the Screws Section 12.8.14
E
2x Screws (PN 420727)
See the Screws Section 12.8.14
See the Metal Brackets and Frame Components Section
12.8.9
D
Interaction Panel Hinge Type B Right (PN 281782)
Hamilton Medical Label (PN 160429)
See the Stickers and Labels Section 12.8.3
L
Oxygen Sensor Cover (PN 160401)
See the External Covers and External Hardware Section
12.8.13
349
12
Spare Parts
12.7.3 Rear Cover
Figure 12-17. Ventilation Unit Rear Cover
Complete Rear Cover MSP160481
A
P-Clamp / Screw (PN 361061 / PN 420724 )
G
See the Clamps and Fasteners Section 12.8.2
B
Stickers (PN 160020)
See the Clamps and Fasteners Section 12.8.2
H
See the Stickers and Labels Section 12.8.3
C
Stickers (PN 160020)
Blank Cover for Communication Board (PN 160431)
I
2x Screws (PN 420730)
J
Sticker (PN 255358)
See the Stickers and Labels Section 12.8.3
350
4x Screws (PN 420667)
See the Screws Section 12.8.14
K
See the Screws Section 12.8.14
F
Ventilator Unit Rear Cover complete (MSP160481)
See the External Covers and External Hardware Section
12.8.13
See the External Covers and External Hardware Section
12.8.13
E
RS 232 Cover (PN 341326)
See the Clamps and Fasteners Section 12.8.2
See the Stickers and Labels Section 12.8.3
D
Ethernet Cover (PN 160509)
Cable Support (PN 160408)
See the External Covers and External Hardware Section
12.8.13
L
3x Screws (PN 420684)
See the Screws Section 12.8.14
English | 624165/07
Ventilation Unit Base
12
12.7.4 Ventilation Unit Base
Figure 12-18. Ventilation Unit Base
A
Base Plate (MSP160228)
D
See the Metal Brackets and Frame Components Section
12.8.9
B
4x Screws (PN 420657)
See the Metal Brackets and Frame Components Section
12.8.9
E
See the Screws Section 12.8.14
C
Side Braces (Left PN 160426 / Right 160427)
See the Metal Brackets and Frame Components Section
12.8.9
Hamilton Medical | HAMILTON-C2 Service Manual
Base Frame (PN 160483)
Release Handle (PN 160403)
See the Metal Brackets and Frame Components Section
12.8.9
F
3x Screws (PN 420724)
See the Screws Section 12.8.14
351
12
Spare Parts
12.7.5 Backup Battery Pack Compartment
Figure 12-19. Backup Battery Pack Compartment
A
Battery Compartment (PN 160302)
See the Electrical/Electronic Devices Section 12.8.12
B
Battery LI-ION (MSP369106)
See the Electrical/Electronic Devices Section 12.8.12
NOTICE
160302/01 since SN 1500
Revision 01 has been implemented with ECO M749. SN 1500 and higher are equipped with an air slot to get the hot air out
of the casing.
352
English | 624165/07
Expiratory Valve Assembly
12
12.7.6 Expiratory Valve Assembly
Figure 12-20. Expiratory Valve Assembly
A
Expiratory Valve Complete (MSP160240)
See the Assembled Components Section 12.8.6
B
C
Expiratory Valve Silicone Membrane (PN 160500)
See the Rubber Seals and Grommets Section 12.8.8
Expiratory Valve Seal with Membrane (PN 160245)
See the Assembled Components Section 12.8.6
Hamilton Medical | HAMILTON-C2 Service Manual
353
12
Spare Parts
12.7.7 Mixer Block Assembly
Figure 12-21. Mixer Block Assembly
A
O2 Mixer Assembly Complete
G
(MSP160226 with TSI QO2 Flow Sensor)
O2 - NIST Connector (PN 160471)
See the Tubings and Fittings Section 12.8.1
(MSP160608 with Sensirion QO2 Flow Sensor)
See the Pneumatic Parts and Assemblies Section 12.8.4
B
3x Screws (PN 420657)
H
See the Screws Section 12.8.14
C
Mixer Block Cover Plate (PN 160459)
2x Screws (PN 420734)
See the Screws Section 12.8.14
I
Nebulizer Valve (PN 160400)
See the Pneumatic Parts and Assemblies Section 12.8.4
D
2x Screws (PN 420664) 12.8.14
J
2 Amp Circuit Breaker (PN 378009)
See the Electrical/Electronic Devices Section 12.8.12
E
O2 - DISS Connector (PN 160470)
See the Tubings and Fittings Section 12.8.1
F
2x Screws (PN 420734)
See the Screws Section 12.8.14
Polyurethane Tubing O2 blue 4x6 (PN 160474) (see Tubing and Fittings Section 12.8.1).
354
English | 624165/07
Detailed Parts Descriptions
12
12.8 Detailed Parts Descriptions
12.8.1 Tubing and Fittings
Table 12-1. Tubing and Fittings
Part Number
Description
160223
FORMED TUBE PATIENT CIRCUIT
160285
FORMED TUBE O2 HD
160288
CONNECTING ADAPTER
160248
CONNECTING ADAPTER
160287
TUBE RING NUT FITTING
160295
PATIENT CONNECTION
Hamilton Medical | HAMILTON-C2 Service Manual
Photo
355
12
Spare Parts
Part Number
160298
Description
Photo
OXYGEN SENSOR MOUNTING BLOCK TUBING SET (PN 160475, if the currently installed Oxygen Sensor Mounting Block is made of aluminum)
(Since > SN 2341 the OXYGEN SENSOR MOUNTING BLOCK is made out of
plastic)
160470
O2 DISS CONNECTOR
160471
O2 NIST CONNECTOR
160474
POLYURETHANE TUBING (O2 BLUE 4x6)
160475
TUBING SET (O2)
MSP160498
TUBING SET (Nebulizer Rinse Flow)
356
English | 624165/07
Clamps and Fasteners
12
12.8.2 Clamps and Fasteners
Table 12-2. Clamps and Fasteners
Part Number
Description
361000
TIE WRAP L=100 B=2.5MM
361016
CABLE HOLDER 20x20x5MM
361061
P-CLAMP D=5.0 PA 6
341458
USB COVER
341326
RS 232 COVER
160509
ETHERNET COVER
160488
CABLE HOLDER
Hamilton Medical | HAMILTON-C2 Service Manual
Photo
357
12
Spare Parts
12.8.3 Stickers and Labels
Table 12-3. Stickers and Labels
Part Number
Description
160020
STICKERS
160429
HAMILTON MEDICAL LABEL
255358
STICKER
161998
MR UNSAFE STICKER
161919
RoHS STICKER for Asia (white/green)
159696
RoHS STICKER for Asia (white/orange)
358
Photo
English | 624165/07
Pneumatic Parts and Assemblies
12
12.8.4 Pneumatic Parts and Assemblies
Table 12-4. Pneumatic Parts and Assemblies
Part Number
Description
160216
HEPA FILTER ASSEMBLY
MSP160226
O2 MIXER ASSEMBLY with TSI FLOW SENSOR (complete)
Photo
• Use Cable (PN 160387) from Oxygen Flow Sensor to Mainboard
MSP160382
• Use Cable (PN 160373) from Oxygen Flow Sensor to Mainboard
MSP160200
Includes:
• Mixer Assembly Complete
• TSI O2 Flow Sensor with Cable
• Nebulizer Valve with Cable
• Required Screws
• Cable Tie
Does not include:
• 2A Circuit Breaker
• O2 Connector
• DC Cable
NOTICE! Only compatible when the TSI Qvent Flow Sensor MSP399123
is installed.
MSP160608
O2 MIXER ASSEMBLY with SENSIRION FlOW SENSOR (complete)
NOTICE! Installed in the C2 from SN 9216
• Use Cable (PN 160387) from Oxygen Flow Sensor to Mainboard
MSP160382
• Use Cable (PN 160373) from Oxygen Flow Sensor to Mainboard
MSP160200
Includes:
• Mixer Assembly Complete
• Sensirion O2 Flow Sensor with Cable
• Nebulizer Valve with Cable
• Required Screws
• Cable Tie
Does not include:
• 2A Circuit Breaker
• O2 Connector
• DC Cable
NOTICE! Only compatible when the Sensirion Qvent Flow Sensor
MSP161657 is installed.
Hamilton Medical | HAMILTON-C2 Service Manual
359
12
Spare Parts
Part Number
Description
MSP160250
BLOWER MODULE
160497
HPO INLET SERVICE KIT
160400
NEBULIZER VALVE
MSP160262
SOUND INSULATION MAT
360
Photo
English | 624165/07
Flow Sensors
12
12.8.5 Flow Sensors
CAUTION
Always use Flow Sensors from the same brand in one ventilator (Sensirion or TSI).
Table 12-5. Flow Sensors
Part Number
MSP161657
Description
Photo
SENSIRION FLOW SENSOR AIR (Qvent)
Does not include:
• Cable (PN 160387) from Qvent Flow Sensor to Mainboard (MSP160382)
• Cable (PN 160373) from Qvent Flow Sensor to Mainboard (MSP160200)
• Tie Wrap (PN 361000)
• Cable Holder (PN 361016)
MSP399123
TSI FLOW SENSOR AIR (Qvent)
Does not include:
• Cable (PN 160387) from Qvent Flow Sensor to Mainboard (MSP160382)
• Cable (PN 160373) from Qvent Flow Sensor to Mainboard (MSP160200)
• Tie Wrap (PN 361000)
• Cable Holder (PN 361016)
Hamilton Medical | HAMILTON-C2 Service Manual
361
12
Spare Parts
12.8.6 Assembled Components
Table 12-6. Assembled Components
Part Number
Description
MSP160230
INSPIRATORY VALVE COMPLETE
MSP160240
EXPIRATORY VALVE COMPLETE
160245
EXPIRATORY VALVE SEAL WITH MEMBRANE
MSP160290
AMBIENT VALVE COMPLETE
MSP160498
RINSE FLOW ASSEMBLY
362
Photo
English | 624165/07
Colored Metal Rings and Controls
12
12.8.7 Colored Metal Rings and Controls
Table 12-7. Colored Metal Rings and Controls
Part Number
MSP160328
Description
Photo
P&T CONTROL KNOB COMPLETE
CONTAINS THE P&T KNOB PN 159449
12.8.8 Rubber Seals and Grommets
Table 12-8. Rubber Seals and Grommets
Part Number
160500
Description
Photo
EXPIRATORY VALVE SILICONE MEMBRANE 5 PIECES
Hamilton Medical | HAMILTON-C2 Service Manual
363
12
Spare Parts
12.8.9 Metal Brackets and Frame Components
Table 12-9. Metal Brackets and Frame Components
Part Number
Description
MSP160228
BASE PLATE
160237
BOTTOM FOAM
160238
MIDDLE FOAM
160239
TOP FOAM
160341
DISPLAY GASKET
160403
RELEASE HANDLE
160404
DISPLAY MOUNTING BRACKET
364
Photo
English | 624165/07
Metal Brackets and Frame Components
Part Number
Description
160426
LEFT SIDE BRACE
160427
RIGHT SIDE BRACE
160483
BASE FRAME
281781
INTERACTION PANEL HINGE TYPE A
281782
INTERACTION PANEL HINGE TYPE B
Hamilton Medical | HAMILTON-C2 Service Manual
12
Photo
365
12
Spare Parts
12.8.10 Electrical/Electronic Cables
Table 12-10. Electrical/Electronic Cables
Part Number
Description
160347
FFC CABLE TO FILTER PRESSURE BOARD
160348
CABLE TO AC INPUT AND CIRCUIT BREAKER
160349
CABLE AC INPUT TO CIRCUIT BREAKER
160370
CABLE TO BATTERY POWER
160351
FFC CABLE TO BATTERY DATA
160539
CABLE TO POWER SUPPLY
366
Photo
English | 624165/07
Electrical/Electronic Cables
Part Number
Description
160354
CABLE TO OXYGEN SENSOR
160355
FFC CABLE TO PRESSURE SENSOR BOARD
160356
FFC CABLE TO KEY PANEL
160357
FDC CABLE TO DISPLAY
160358
CABLE TO BACKLIGHT DC/AC INVERTER BOARD (PN 396197 old or PN
396229 new)
161569
CABLE TO GRAPHIC LCD TFT VGA 10.4” NLT DISPLAY BACKLIGHT LED's
12
Photo
(Bypass for the Backlight Driver Board)
160359
FFC CABLE TO BINARY VALVES
Hamilton Medical | HAMILTON-C2 Service Manual
367
12
Spare Parts
Part Number
Description
160372
CABLE DC INPUT
160373
CABLE FROM OXYGEN FLOW SENSOR TO MAINBOARD (MSP160200)
160387
CABLE FROM OXYGEN FLOW SENSOR TO MAINBOARD (MSP160382)
355198
USA POWER CABLE 2 POL 3MT C7 G (United States)
355199
GB POWER CABLE 2 POL 3MT C7 G (Great Britain)
355200
EU POWER CABLE 2 POL 3MT C7 G (European)
355308
CHINA POWER CABLE
160187
CAR ADAPTER
368
Photo
English | 624165/07
Electronic Printed Circuit Boards
12
12.8.11 Electronic Printed Circuit Boards
Table 12-11. Electronic Printed Circuit Boards
Part Number
MSP160196
Description
Photo
FRONT PANEL BOARD
(Revision 08 and higher with built-in Backlight Driver Components for
GRAPHIC LCD TFT VGA 10.4” NLT Display Backlight LED's)
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
MSP160200
MAINBOARD-C2
Includes:
• Display Cable Holder (PN 160457)
• Distance Holder 6mm (PN 362051)
MSP160382
MAINBOARD-C3
(used in the Hamilton-C2)
Includes:
• Display Cable Holder (PN 160457)
• Distance Holder 6mm (PN 362051)
• 2x Cables to Flow Sensor (QO2 and Qvent) (PN 160387)
MSP160206
EMBEDDED SYSTEM MODULE
MSP160578
ESM SHIELDING
MSP160300
PRESSURE SENSOR ASSEMBLY
Hamilton Medical | HAMILTON-C2 Service Manual
369
12
Spare Parts
Part Number
MSP160362
Description
Photo
DISPLAY FRONT
See the LCD Display and Touch Screen Section 12.6.4
Includes:
• Touch Screen
• Key Panel
• Display Gasket (PN 160341)
• Symbol Insert (PN 160376)
Does not include:
• P&T Control Knob
• LCD Display
MSP160600
SERVICE POWER SUPPLY
380027
GRAPHIC LCD TFT VGA 10.4” NEC
NOTICE! Obsolete
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
380031
GRAPHIC LCD TFT VGA 10.4” NEC
NOTICE! Obsolete
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
380046
GRAPHIC LCD TFT VGA 10.4” NLT
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
396197
DC/AC BACKLIGHT INVERTER BOARD
396229
NOTICE! Obsolete
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
396314
BACKLIGHT LED DRIVER BOARD
NOTICE! No longer available
If replaced, order Front Panel Board (MSP160196)
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
370
English | 624165/07
Electronic Printed Circuit Boards
Part Number
Description
160140
EXTENDED COMMUNICATION BOARD CO2, SPO2 (Not implemented in
HAMILTON-C2) AND NURSE CALL
160143
EXTENDED COMMUNICATION BOARD NURSE CALL
Hamilton Medical | HAMILTON-C2 Service Manual
12
Photo
371
12
Spare Parts
12.8.12 Electrical/Electronic Devices
Table 12-12. Electrical/Electronic Devices
Part Number
160302
Description
Photo
BATTERY COMPARTMENT
Includes:
• PN 160370 Cable to Battery Power
• PN 160351 FFC Cable to Battery Data
160346
FAN 12V
161532
REPLACEMENT SPEAKER
340530
AC ELECTRICAL INLET PLUG SNAP-IN CLASS II
MSP369106
BATTERY LI-ION 14.4V/6.8Ah
372036
P&T CONTROL KNOB ENCODER
378009
2 AMP CIRCUIT BREAKER
372
English | 624165/07
Electrical/Electronic Devices
Part Number
Description
380044
BACKLIGHT FOR LCD
396200
OXYGEN SENSOR (Hamilton Medical AG)
Hamilton Medical | HAMILTON-C2 Service Manual
12
Photo
373
12
Spare Parts
12.8.13 External Covers and External Hardware
Table 12-13. External Covers and External Hardware
Part Number
Description
160326
INTERACTION PANEL LARGE REAR COVER
160329
ALARM LAMP COVER
160343
VENTILATOR UNIT FILTER COVER
160506
BATTERY COVER
160401
OXYGEN SENSOR COVER
160402
INTERACTION PANEL SMALL REAR COVER
160408
CABLE SUPPORT
374
Photo
English | 624165/07
External Covers and External Hardware
Part Number
Description
160431
BLANK COVER FOR COMMUNICATIONS BOARD
MSP160480
VENTILATOR UNIT FRONT COVER with Label
MSP160481
VENTILATOR UNIT REAR COVER COMPLETE (see 12.7.3)
160601
VENTILATOR UNIT TOP COVER
160155
TUBING HOLDER DOUBLE
160156
CLIP FOR TUBING HOLDER
Hamilton Medical | HAMILTON-C2 Service Manual
12
Photo
375
12
Spare Parts
12.8.14 Screws
Table 12-14. Screws
Part Number
Description
420623
TORX ROUNDHEAD SCREW W/I-6-ROUND M4x16
420638
TORX ROUNDHEAD SCREW I-6-R M2.5x6 A4
420641
TORX ROUNDHEAD SCREW I-6-R M3x5 A4
420642
TORX ROUNDHEAD SCREW I-6-R M3x6 A4
420643
TORX ROUNDHEAD SCREW I-6-R M3x8 A4
420654
TORX CYL-SCREW I-6-R NK M3x4 A2
376
Photo
English | 624165/07
Screws
Part Number
Description
420657
TORX CYL-SCREW I-6-R NK M4x10 A2
420659
TORX C-SUNK SCREW I-6-R M3x6 A4
420663
TORX C-SUNK SCREW I-6-R M2x4 A4
420664
TORX C-SUNK SCREW I-6-R M3x8
420667
TORX CYL-SCREW I-6-R M4x8 A2
420669
TORX ROUNDHEAD SCREW I-6-R M4x30 A4
420671
TORX ROUNDHEAD SCREW I-6-R M2x3 A2
Hamilton Medical | HAMILTON-C2 Service Manual
12
Photo
377
12
Spare Parts
Part Number
Description
420682
TORX ROUNDHEAD SCREW I-6-R M4x12 A4
420684
TORX CYL-SCREW I-6-R ENK M3x6 A2
420692
TORX ROUNDHEAD SCREW I-6-R M2x4 A4
420699
TORX FILLISTER HEAD SCREW 22x5 D1=2.2, L=5, 6 IP
420710
TORX FILLISTER HEAD SCREW 25x8 D1=2.5, L=8 (Galvanized blue)
420720
TORX FILLISTER HEAD SCREW I-6R M3x10 (Galvanized blue steel)
420724
TORX FILLISTER HEAD SCREW 30x8
378
Photo
English | 624165/07
Standoffs
Part Number
Description
420727
TORX SCREW I-6-R M4x16
420730
PHILLIPS HEAD SCREW M3x20 DIN7985 PA6.6
420734
TORX CYL-SCREW I-6-R NK M3x10
12
Photo
12.8.15 Standoffs
Table 12-15. Standoffs
Part Number
Description
257038
STANDOFF SCREW M3x8
257076
STANDOFF M2x5
Hamilton Medical | HAMILTON-C2 Service Manual
Photo
379
12
Spare Parts
12.8.16 Washers
Table 12-16. Washers
Part Number
409105
WASHER M3 DIN125A
409264
WASHER M 2 DIN125A
409908
WASHER M10/1.2 BN735
411001
STAR WASHER M3 DIN6798
Description
Photo
Description
Photo
12.8.17 Screw Nuts
Table 12-17. Screw Nuts
Part Number
421059
380
SCREW SQUARE NUT M4
English | 624165/07
13
Schematics
13.1
Block Diagram......................................................................................................................................382
13.2
Wiring Diagram....................................................................................................................................383
381
13
Schematics
13.1 Block Diagram
Table 13-1. Block diagram
Diagram description
HAMILTON-C2 block diagram
Number/Revision
BD614227/00
Figure 13-1. HAMILTON-C2 block diagram
382
English | 624165/07
Wiring Diagram
13
13.2 Wiring Diagram
Table 13-2. Wiring diagram
Diagram description
HAMILTON-C2 wiring diagram
Number/Revision
WD616063/06
Figure 13-2. HAMILTON-C2 wiring diagram
Hamilton Medical | HAMILTON-C2 Service Manual
383
13
384
Schematics
English | 624165/07
14
Software Revisions, Features and
Compatibility
14.1
Introduction .........................................................................................................................................386
14.2
Software Version Information...............................................................................................................386
385
14
Software Revisions, Features and Compatibility
14.1 Introduction
NOTICE
For actual information and latest software, log in to Hamilton Medical AG Partner-net. There are the latest information
about upgrading and updating a HAMILTON-C2.
This section gives an overview of all published HAMILTON-C2 Software. It includes information about new features introduced with software upgrades and software updates information about compatibility between software of different versions.
The improvement list is not meant to be complete and covers only the most relevant changes.
14.2 Software Version Information
WARNING
The (*) indicates an obsolete software version. Mandatory update must be performed.
Table 14-1. Software Version 2.2.5
Software Version
Date Released
2.2.5
July 2018
Improvements / Corrections
The following improvements were implemented:
• Battery SoH added to the Instrument Report
– The Battery State of Health (SoH) information is added to the instrument
report in the new software
• FPGA update
– A new internal Field Programmable Gate Array (FPGA) clock distribution
improves the reliability of the integrated circuit boards
• Blower time can be updated
– The blower time can be set again as it was before
• Extend log files CPU temperature and POST will be logged in the new software
• Wrong 'positive oxygen supply failed' alarm solved
Table 14-2. Software Version 2.2.4
Software Version
Date Released
2.2.4
Not Implemented
Improvements / Corrections
Table 14-3. Software Version 2.2.3
Software Version
Date Released
2.2.3
June 2015
Improvements / Corrections
The following bugs were solved:
• Low Oxygen Alarm despite of correct FiO2
• Trend Pmean in DuoPap/APRV
• Mismatch-IBW
• Alarm criteria disconnection on Ventilator side improved
The following improvements were implemented:
• Firmware Version of the internal flow sensor (Sensirion only) now displayed in
the Instrument Report
NOTICE! For more details see Knowledge Base ID 3247.
386
English | 624165/07
Software Version Information
14
Table 14-4. Software Version 2.2.2
Software Version
Date Released
2.2.2
Not Implemented
Hamilton Medical | HAMILTON-C2 Service Manual
Improvements / Corrections
387
14
Software Revisions, Features and Compatibility
Table 14-5. Software Version 2.2.1
388
English | 624165/07
Software Version Information
Software Version
Date Released
2.2.1
January 2015
14
Improvements / Corrections
The following bugs were solved:
• False positive TF 431008 during start-up
• False positive TF 431005 during ventilation
• TF341001, TF383001, TF385002 (settings not accepted)
• TF346044, TF346043, TF385002 and TF485001 (10s after Standby)
• Increased rate of sigh in SIMV+
• Oxygen Sensor defective alarm improved
• Pinsp Minimum set to 5mbar (Japan only)
The following improvements were implemented:
• New display languages added: Croatian, Finnish, Romanian
• Once activated, CO2 monitoring remains activated until manually deactivated
• Changed the method for calibrating the infant flow sensor; a calibration adapter is now used
• Alarm loudness setting moved to Settings window and Alarm limit settings and
ranges table changed
• Default values changed for the following parameters: ETS, I:E, Rate, TI, Vt/kg
• P-ramp limitation by TI added
• Alarm tables have been updated
• HPO/LPO selection moved to Utilities window, available in Standby
• Ability to configure a minimum alarm volume (loudness) for the device
• Battery low
• Loss of external power
• Performance limited by altitude
• Battery replacement required (new)
• High PEEP (new)
• High oxygen
• Inspiratory volume limitation (now disabled in noninvasive modes)
• CO2 sensor warmup
• Function Key not operational
• Invalid Communication Board
• Replace Oxygen Sensor
• Suctioning maneuver
• Technical error
• Technical state failed
• Touch not functional
• Unknown part number
• Release valve defective
Replaced the numerical code with text and added troubleshooting information to
the Operator’s Manual for the following technical alarms:
• Battery communication error (244011)
• Battery totally discharged (444001)
• Blower fault (431001)
• Buzzer defective (243004)
• Loudspeaker defective (243003)
• Options not found (249001)
• Self test failed (485002)
• Blower lifetime
NOTICE! For more details see Knowledge Base ID 3153.
Hamilton Medical | HAMILTON-C2 Service Manual
389
14
Software Revisions, Features and Compatibility
Table 14-6. Software Version 2.2.0 *
Software Version
Date Released
2.2.0 *
December 2014
Improvements / Corrections
The following bugs were solved:
• False positive TF 431008 during start-up
• False positive TF 431005 during ventilation
• TF341001, 383001, 385002 (settings not accepted)
• TF346044, TF346043, TF385002 and TF485001 (10s after Standby)
• Oxygen Sensor defective alarm improved
The following improvements were implemented:
• New display languages added: Croatian, Finnish, Romanian
• Once activated, CO2 monitoring remains activated until manually deactivated
• Changed the method for calibrating the infant flow sensor; a calibration adapter is now used
• Alarm loudness setting moved to Settings window and Alarm limit settings and
ranges table changed
• Several alarms have been added or updated
• Replaced the numerical code with text and added troubleshooting information
to the Operator’s Manual for several technical alarms
• Default values changed for the following parameters: ETS, I:E, Rate, TI, Vt/kg
• P-ramp limitation by TI added
• Alarm tables have been updated
• HPO/LPO selection moved to Utilities window, available in Standby
• Ability to configure a minimum alarm volume (loudness) for the device
NOTICE! For more details see Knowledge Base ID 3153.
Table 14-7. Software Version 2.1.2 *
Software Version
Date Released
2.1.2 *
May 2013
Improvements / Corrections
The following bugs were solved:
The following improvements were implemented:
• AUTO function on Vthigh and Vt low only active during startup
• The Vt high and low alarm have to be set manually by the user during ventilation
– Note: The alarm message "VT high" is replaced by "Breath aborted, Vt
high".
• The parameters for the loudspeaker have been optimized
• The new software resolves the situation in which the ventilation and the alarm
system continue to operate while the display is frozen
• False positive oxygen supply missing alarms suppressed
390
English | 624165/07
Software Version Information
14
Table 14-8. Software Version 2.1.1 *
Software Version
Date Released
2.1.1 *
June 2012
Improvements / Corrections
The following bugs were solved:
• TF 431016 problem solved
• TF 331001 triggered instead of TF 431016, additionally, occurrence has been
minimized
• Oxygen Sensor calibration procedure improved (faulty Gain values should no
longer occur)
• Oxygen Sensor alarms modified (Oxygen Sensor calibration needed instead of
Oxygen Sensor defective if the Oxygen Sensor voltage is slightly out of range)
• TF 341909 problem solved (TF occurred if I:E calculator detected a problem)
• Technical state: - The Power Supply (MSP160600) can now be saved (SW 2.1.0
or lower only allowed to save the Power Supply (PN 396199))
Table 14-9. Software Version 2.1.0 *
Software Version
Date Released
2.1.0 *
November 2011
Improvements / Corrections
The following bugs were solved:
Ventilation mode:
• The ambient and the safety mode is not displayed any more as regular TF
xxx’xxx In case of a TF leading to ambient mode "Ventilation cancelled" will be
displayed.
• In case of a TF leading to safety mode "Safety ventilation" will be displayed
• Trend time has been extended to 72 hours
• Suctioning tool is now displayed in the options list
• Occurrence of watchdog failures has been reduced again
• Known technical faults during suctioning maneuver do not occur anymore
• GUI "black holes" problem solved
• RTC time will not be lost during ventilation
• Technical event: 243005 (loudspeaker sounds continual) will only be checked
during start-up
• Technical event: 232037 (Unknown alarm) during the CO2 calibration will not
appear anymore
• The sequence of the component test has been improved
• Service software:
– Technical state now displays all hardware revisions in one window
– Pressure calibration procedure is adapted to the new expiratory valve cover
with the metal inlay: With the expiratory valve cover (revision 02) the
pressure inside the breathing circuit can no longer be adjusted to 50 mbar
(51 cmH2O), because the new covers are significantly tighter than the
previous version. Therefore the pressure will be calibrated by comparing
the external reference (pressure gauge) with the internal measurement
displayed on screen (Pvent_control). Furthermore the saving of the gain
value (adjustable with turn knob on screen) will be disabled if the
difference between the Pvent_control and Pvent_monitor resp. Pressure
sensor Paw value is more than 3 mbar.
– Technical fault and event during performing the service software will be
suppressed
– Pneumatics 1 -> Exp. valve test: The test limiting values could be enlarged
– The sequence of the component test has been improved
Hamilton Medical | HAMILTON-C2 Service Manual
391
14
Software Revisions, Features and Compatibility
Table 14-10. Software Version 2.0.7 *
Software Version
Date Released
2.0.7 *
March 2011
Improvements / Corrections
The following bugs were solved:
• Watchdog Alarm Routine: With the new software version, the probability of
watchdog failures is significantly decreased
• Battery over temperature: A plausibility check has been introduced to avoid
false positive alarms about hot batteries
• DuoPAP: The alarm limit has been adapted to avoid false positive alarm "Disconnection on patient side"
• "Exhalation port occluded" in NIV: The alarm has been eliminated. Consequently the LiteCircuit has been withdrawn from the product line.
• CO2 curve transmission over RS232: The new unit for CO2 curve transmission
over RS232 is mmHg instead of cmH2O
• Language files: The following language files have been updated: Spanish, Portuguese, Chinese
• Intersurgical LiteCircuit: With the release of the software version 2.0.7 for the
HAMILTON-C2 we will withdraw the Intersurgical LiteCircuits from our product
lines for the HAMILTON-C2, Hamilton-C1 and Raphael. The use of the LiteCircuit bears the potential risk, that the whisper valve gets blocked and CO2 is not
effectively removed. The affected part numbers are:
– PN 151817 LiteCircuit Standard
– PN 151819 LiteCircuit 580
Table 14-11. Software Version 2.0.6 *
Software Version
Date Released
2.0.6 *
January 2011
Improvements / Corrections
The following bugs were solved:
• Russian label (Day/Night) corrected
The following improvements were implemented:
• With SW version 2.0.6 the HAMILTON-C2 is able to communicate with the
Capsuletec PDMS system
Table 14-12. Software Version 2.0.5 *
Software Version
Date Released
2.0.5 *
December 2010
Improvements / Corrections
The following bugs were solved:
• The software version 2.0.5 increases the oscillation stability of the HAMILTONC2. This issue becomes visible only under certain circumstances, i.e. cold temperatures, beginning of ventilation, coaxial breathing circuit without inspirations filter and our test lung. The effect can be observed especially while using
DuoPAP or APRV.
Table 14-13. Software Version 2.0.4 *
Software Version
Date Released
2.0.4 *
December 2010
Improvements / Corrections
The following bugs were solved:
• The software version 2.0.4 includes improvements of DuoPAP and APRV
regarding trigger sensitivity and pressure support on Phigh.
• The software version 2.0.3 is still considered to be safe. However, we would
advice to updated systems in the field with software version 2.0.4 for customers ventilating their patients intensively with DuoPAP or APRV.
392
English | 624165/07
Software Version Information
14
Table 14-14. Software Version 2.0.3 *
Software Version
Date Released
2.0.3 *
November 2010
Improvements / Corrections
The following bugs were solved:
• Tightness test: Issue: With SW 2.0.2, in some cases the HAMILTON-C2 faces
issues to complete the tightness test successfully Solution: Install SW 2.0.3. We
improved the algorithm significantly and made the test more reliable.
• Suctioning Tool: Issue: With SW 2.0.2, with activated sigh function in nonadaptive modes, the HAMILTON-C2 can in certain, rare cases fall into the TF
341001 during the suctioning maneuver. Solution: Install SW 2.0.3: Sighs are
suppressed during suctioning.
The following improvements were implemented:
• Labeling: We used the chance and made improvements on the software labels.
We adapted the following languages: French, Danish, Hungarian, Polish, Slovak, Serbian and Chinese.
NOTICE! The update from previous software versions to SW 2.0.3 is not
mandatory, but recommended.
Table 14-15. Software Version 2.0.2 *
Software Version
Date Released
2.0.2 *
October 2010
Improvements / Corrections
The following bugs were solved:
• In the German version SIMV+ must be replaced by DuoPAP+
• In pressure calibration window when calibrating to 50 mbar, additionally, it is
stated to calibrate to 51cmH2O when cmH2O is used in the external pressure
gauge instead of mbar
Table 14-16. Software Version 2.0.1 *
Software Version
Date Released
2.0.1 *
September 2010
Improvements / Corrections
The following bugs were solved:
• No false positive alarm "Oxygen supply failed" in the low FiO2 ranges (22% 30%) during the first breaths
• Increased stability of O2 valve leak detection (TF 231008)
• Support of new batteries in the technical file. The shown part number in the
technical state is read out from the battery itself.
• Correction of alarm rules for blower speed supervision (TF 231009)
• Reduction of false positive alarm "Exhalation port occluded" during NIV with
closed system
• Correction of consequences of internal communication error (TF 344001) (high
priority alarm instead of ambient mode)
• Correction of issues regarding connectivity with Dräger and Philips monitors
PDMS protocol sequences adapted to the latest and coming driver versions
Hamilton Medical | HAMILTON-C2 Service Manual
393
14
Software Revisions, Features and Compatibility
Table 14-17. Software Version 2.0.0 *
Software Version
Date Released
2.0.0 *
October 2011
Improvements / Corrections
The following bugs were solved:
• Quick-start-up settings: Configure 3 quick-start-up settings with the corresponding mode, control parameters, alarm limits and ventilation start-up settings. This allows you to start within a second the appropriate ventilation treatment.
• Suctioning tool: The suctioning tool recognizes a disconnection of the ventilator
during oxygen enrichment. Ventilation is immediately stopped and resumed
when a re-connection is detected. For a clean and safe suctioning procedure.
• Dimmable screen: Change the intensity of the backlight in the night to let the
patient rest.
• Automatic alarm settings: Adjusts the alarms automatically around the actual
value.
• Automatic apnea backup: The automatic apnea backup helps to adjust apnea
backup parameters correctly corresponding to the ideal body weight of the
patient.
• Improved IntelliTrig: IntelliTrig allows the HAMILTON-C2 to compensate leakages and adapt trigger sensitivity. With the new implemented algorithm, we
did a further step to increase the reactivity of the system to leaks.
Table 14-18. Software Version 1.1.4 *
Software Version
Date Released
1.1.4 *
October 2011
Improvements / Corrections
The following bugs were solved:
• The software solves a problem met during extreme stress test in house with an
external simulation system
• A decrease of the blower speed after stand-by prevents the device from false
positive "External power loss" alarms
• TF 243005: loudspeaker sounds continually: the alarm rule has been adapted
• TF 231008/TF231013: During start-up in LPO mode the false positive TF's
231008/231013 doesn't pop up anymore.
• No false positive TF 485001 during the switch off procedure any more
• Increased O2 mixer stability
• several other, minor adjustments
NOTICE! The software 1.1.4 (1.0.5 for Japan) MUST be installed to all devices
in the field according to the information you received in the partner letter April
6th, 2010.
Table 14-19. Software Version 1.1.3 *
Software Version
Date Released
1.1.3 *
December 2009
Improvements / Corrections
The following bugs were solved:
• The technical alarm TF 232007 is displayed as “Check Flow Sensor tubing”
alarm on screen during ventilation in order to give the user a better understanding of the situation. In the event and service log the technical alarm TF
232007 is displayed as Check Flow Sensor tubing alarm.
• Autozero procedure: Due to internal investigation the autozero interval takes
place more frequently to improve the flow measurement performance.
• Event Log: All technical faults generated while performing the Service Software
tests will no longer be registered in the user event log as numerical codes.
394
English | 624165/07
Software Version Information
14
Table 14-20. Software Version 1.1.2 *
Software Version
Date Released
1.1.2 *
August 2009
Improvements / Corrections
The following improvements were implemented:
• Technical Fault Handling
• Flow sensor calibration
• Auto triggering behavior
• Power Management (battery power handling)
• New Languagefiles; DuoPAP+ to PSIMV+
• Language adaption based on International Standards
Table 14-21. Software Version 1.1.1 *
Software Version
Date Released
1.1.1 *
May 2009
Improvements / Corrections
The following bugs were solved:
• Adjusting alarm limit blower service required
The following improvements were implemented:
• Internal tightness test applicable
• System test O2 mixer implemented
• System test flow implemented
Table 14-22. Software Version 1.1.0 *
Software Version
Date Released
1.1.0 *
March 2009
Improvements / Corrections
The following improvements were implemented:
• DuoPAP / APRV
• Trend
• Loops
• Asian languages
• Ferrites in Interaction Panel not necessary
Table 14-23. Software Version 1.0.2 *
Software Version
Date Released
1.0.2 *
December 2008
Improvements / Corrections
The following improvements were implemented:
• Reducing of maximum power consumption during starting the device
• Adjusting limits of the buzzer control
• New settings of O2 control low alarm
• Changing identification algorithm of the Oxygen Sensor
• Increased the divergence of flow sensor calibration
Table 14-24. Software Version 1.0.1 *
Software Version
Date Released
1.0.1 *
October 2008
Improvements / Corrections
The following improvements were implemented:
• SW 1.0.1, ready for software upgrade with USB memory stick
Hamilton Medical | HAMILTON-C2 Service Manual
395
14
396
Software Revisions, Features and Compatibility
English | 624165/07
15
Hardware Revisions, Features and
Compatibility
15.1
Introduction .........................................................................................................................................398
15.2
Component History and Improvements ................................................................................................398
397
15
Hardware Revisions, Features and Compatibility
15.1 Introduction
This section brings together information found in other parts of the service manual concerning hardware components that
have changed over time. The section also provides additional information including.
Hamilton ventilator serial numbers and the dates of changes.
15.2 Component History and Improvements
15.2.1 Interaction Panel
WARNING
If the Interaction Panel Display (PN 380027 or PN 380031) or the Backlight DC/AC Inverter (PN 396197 or PN 396229)
needs replacement and the Front Panel Board is Revision 01 to 07, please order the following parts:
1x Display (MSP380046)
1x Backlight LED Driver Board (PN 396314)
1x Cable (PN 161569)
2x Ferrite Plus (PN 393133)
2x Ferrite Clip (PN 393134)
WARNING
If the Interaction Panel Display (PN 380027 or PN 380031) or the Backlight DC/AC Inverter (PN 396197 or PN 396229)
needs replacement and the Front Panel Board is Revision 08 or higher, please order the following parts:
1x Display (MSP380046)
1x Backlight Extension Cable (PN 161569)
1x Cable (PN 161569)
2x Ferrite Plus (PN 393133)
2x Ferrite Clip (PN 393134)
Set the Front Panel Board Jumper P14 to pins 1 and 2.
398
English | 624165/07
Interaction Panel Display
15
15.2.1.1 Interaction Panel Display
Table 15-1. Interaction Panel Display (PN 380027, PN 380031 and PN 380046)
Revision
History
00
PN 380027 LCD TFT VGA 10.4” NEC Graphic Display,
1st generation with Florescent Backlights (PN
380030)
Image
Used in HAMILTON-C2 from SN1001 to SN1675
NOTICE! Obsolete
00
PN 380031 LCD TFT VGA 10.4” NEC Graphic Display
2nd generation with Florescent Backlights (PN
380044)
Used in HAMILTON-C2 from SN1676 to SN11339
NOTICE! Obsolete
00
PN 380046 LCD TFT VGA 10.4” NLT Graphic Display
With LED Backlights (no replacement defined)
Used in HAMILTON-C2 from SN11340
01
PN 380046 LCD TFT VGA 10.4” NLT Graphic Display
With LED Backlights (no replacement defined)
Hamilton Medical | HAMILTON-C2 Service Manual
399
15
Hardware Revisions, Features and Compatibility
15.2.1.2 Front Panel Board
Table 15-2. Front Panel Board (MSP160196)
Revision
History
Image
00
01
Initial release.
02
New layout, connector handling.
03
Speaker loudness.
Remove Foil Cover Sticker from the Interaction Panel
Cover (if still attached) after Rev > 03 installation.
04
Improvement of the TF 243003 problem and loudspeaker loudness and sound quality.
05
Improvement of loudspeaker lifetime.
Speaker no longer soldered, but with connector.
06
Without central red backup alarm LED.
07
Not implemented.
08
Integrated Backlight LED Components for the Graphic
LCD TFT VGA 10.4” NLT Display (PN 380046) with
LED backlights. With this Front Panel Board, the
external Backlight LED Driver Board (PN 396314) is no
longer required. An Extension Cable (PN 161569) is
required as a bypass for the Backlight Driver Board.
CAUTION! Integrated Backlight LED Components on the Front Panel Board are not compatible
with the Graphic LCD TFT VGA 10.4” NEC (Obsolete,
PN 380027 or PN 380031).
CAUTION! If the Front Panel Board is
replaced and the Graphic LCD TFT VGA 10.4” NEC
(Obsolete, PN 380027 or PN 380031) is being used,
set the Jumper P14 to pins 2 and 3 and use the DC/
AC Backlight Inverter Board (PN 396229), if available.
400
English | 624165/07
Backlight Driver Board
15
15.2.1.3 Backlight Driver Board
Table 15-3. Backlight DC/AC Inverter Board (PN 396197 and PN 396229) and Backlight LED Driver Board (PN 396314)
Revision
History
00
PN 396197 DC/AC Backlight Inverter Board for the
Graphic LCD TFT VGA 10.4” NEC Display Fluorescent
Backlight Lamps.
Image
NOTICE! Obsolete
00
PN 396229 DC/AC Backlight Inverter Board for the
Graphic LCD TFT VGA 10.4” NEC Display Fluorescent
Backlight Lamps.
NOTICE! Obsolete
00
PN 396314 Backlight LED Driver Board for the
Graphic LCD TFT VGA 10.4” NLT Display Backlight
LED's.
01
PN 396314 Backlight LED Driver Board for the
Graphic LCD TFT VGA 10.4” NLT Display Backlight
LED's.
NOTICE! No longer available
Hamilton Medical | HAMILTON-C2 Service Manual
401
15
Hardware Revisions, Features and Compatibility
15.2.2 Ventilation Unit
15.2.2.1 Mainboard (MSP160200)
Table 15-4. Mainboard (MSP160200)
Revision
History
Image
00
01
02
03
First official release.
Used in HAMILTON-C2 from SN1001 to SN1134.
04
05
06
New layout and new firmware.
Oxygen Sensor calibration without external tool.
Use in HAMILTON-C2 from SN1266 to SN2140.
07
New layout.
Blower end stage robustness.
Used in HAMILTON-C2 from SN2141 to SN2422.
08
New firmware.
Used in HAMILTON-C2 from SN 2423 to SN 2784.
09
Qvent Flow Sensor precision improved.
EMC immunity to 20 v/m increased.
Used in HAMILTON-C2 from SN2785 to SN3050.
10
Not implemented.
11
Soft boot problem of battery 2 solved.
Used in HAMILTON-C2 from SN3051 to SN3369.
12
Alarm monitor improvements.
Used in HAMILTON-C2 from SN3370 to SN4455.
13
EMC improvements and buzzer adjustability.
Neonatal implemented.
Used in HAMILTON-C2 from SN4456.
14
Exchange of discontinued parts.
Used in HAMILTON-C2 from SN5900
15
EMI improvement to conductive ESD.
Switching off problem solved.
Easy to enter service software by storing the keys.
NOTICE
The Oxygen Sensor Calibration Tool (PN 160367) is no longer required for Mainboard Revisions 06 and higher.
402
English | 624165/07
Mainboard (MSP160382)
15
15.2.2.2 Mainboard (MSP160382)
NOTICE
The Mainboard MSP160382 was previously used in the HAMILTON-C3 only. In the HAMILTON-C2, it is assembled in devices
with SN > 10’000. It is compatible from older devices as long as software version 2.2.0 or higher is installed.
Table 15-5. Mainboard (MSP160382)
Revision
History
Image
00
01
02
03
04
05
First official release in HAMILTON-C2.
Used in HAMILTON-C2 from > SN10’000.
No disconnection during O2 offset calibration
needed.
Different connector for the oxygen and air flow sensors.
Additional fan supply from power supply is no longer
needed.
Alarm lamp blinking according new standard of IEC
60601-1.
Jumper to select the device (HAMILTON-C2 or
HAMILTON-C3).
06
New PoL voltage converters for improved EMI behavior.
Oxygen Sensor calibration without disconnection.
RS232 monitoring interface bit rate up to 38.2 kBaud
for Hamilton block protocol.
07
Not implemented.
08
Increased ESD immunity against contact discharge to
fulfill the new standard for electromagnetic compatibility (EMC) IEC 60601-1-2 (2014) 4th edition.
Improved polarity protection on DC Input.
New gold caps with improved lifetime.
09
10
New blower end stage due to discontinued parts.
CAUTION! Due to an incompatibility with
older software versions, the use of a Mainboard
(MSP160382) with revision 10 or higher requires the
following software to be installed on the instrument:
HAMILTON-C2 SW 2.2.5 or higher.
NOTICE
If Mainboard (MSP160382) replaces Mainboard (MSP160200), make sure to roll up the excess Fan Supply Cable.
Hamilton Medical | HAMILTON-C2 Service Manual
403
15
Hardware Revisions, Features and Compatibility
15.2.2.3 ESM Module
Table 15-6. ESM Module (MSP160206)
Revision
History
Image
00
01
First official release.
02
New layout.
New Firmware.
SW 1.0.1.
Ready for SW Upgrade with USB memory stick.
03
Not used components dropped.
04
Not used components dropped.
05
SW 1.0.5.
No more loss of Technical State due to SW upgrade,
not used components dropped.
06
EMC improvements.
07
Pre-installed latest software 2.x.x.
08
Layout correction.
09
Pre-installed SW 2.2.1.
10
Pre-installed SW 2.2.3.
11
Clock generators replaced due to discontinued parts.
BGA components underfilled to improve mechanical
robustness.
15.2.2.4 ESM Shielding
Table 15-7. ESM Shielding (MSP160578)
Revision
History
Image
00
01
First official release.
02
Improved EMC
Used in HAMILTON-C2 from SN6976.
404
English | 624165/07
Flow Sensors
15
15.2.2.5 Flow Sensors
CAUTION
Always use Flow Sensors from the same brand in one ventilator (Sensirion or TSI).
Table 15-8. TSI Qvent Flow Sensor (MSP399123)
Revision
History
Image
00
01
02
03
Table 15-9. TSI O2 Flow Sensor (MSP399124)
Revision
History
Image
00
01
02
03
Table 15-10. Sensirion Qvent and QO2 Flow Sensors (MSP161657)
Revision
History
Image
00
01
Hamilton Medical | HAMILTON-C2 Service Manual
405
15
Hardware Revisions, Features and Compatibility
15.2.2.6 Oxygen Mixer Block Assembly
Table 15-11. Oxygen Mixer Block Assembly (MSP160226 or MSP160608)
Revision
00
History
Image
First official release
With TSI QO2 Flow Sensor
01
Features:
• Plastic Disk (A)
• Without Turbulence Filter (A)
• Manufactured with the small O2 connector seat (B)
(see Mounting Plate Section 15.2.2.7)
• Manufactured without O2 Inlet Filter (C) (see O2
Inlet Filter Section 15.2.2.8)
In HAMILTON-C2 from SN1001 to SN1271.
02
Features:
• Sinter disk (A)
• Equipped with turbulence screen (B)
• Manufactured with the small O2 connector seat (B)
from SN1590 to SN2154 (see Mounting Plate Section 15.2.2.7)
• Manufactured without O2 Inlet Filter (C) (see O2
Inlet Filter Section 15.2.2.8)
In HAMILTON-C2 from SN1272 to SN1589.
03
Features:
• The QO2 Flow Sensor (MSP161657) is part of the
mixer assembly (A, MSP160226)
• Manufactured with the small O2 connector seat
(A1) from SN1590 to SN1625
• Manufactured with the large O2 connector seat (B)
from SN1626
• Orifice (C)
• Turbulence screen (D)
In HAMILTON-C2 from SN1590 to SN2154.
04
Adaption of the manufacturing process.
In HAMILTON-C2 from SN 2155.
05
Introduction of an Injection Molded Connection Plate.
In HAMILTON-C2 from SN 5587.
406
English | 624165/07
Oxygen Mixer Block Assembly
Revision
06
History
15
Image
Additional Cable (PN 160387) for Mainboard
(MSP160387).
CAUTION! When replacing the Oxygen Mixer
Block Assembly, within the same ventilator, match
the Oxygen Mixer Block Assembly to have the same
brand QO2 Flow Sensor as the Qvent Flow Sensor (TSI
or Sensirion).
Oxygen Mixer Block Assembly (MSP160226) with TSI QO2
Flow Sensor (MSP399124).
Oxygen Mixer Block Assembly MSP160608) with Sensirion
QO2 Flow Sensor (MSP161657).
Hamilton Medical | HAMILTON-C2 Service Manual
407
15
Hardware Revisions, Features and Compatibility
15.2.2.7 Mounting Plate
Table 15-12. Mounting Plate (PN 160458)
Revision
00
History
Image
First official release.
The mounting plate with the small O2 Connector seat
(left image) is installed in all HAMILTON-C2 from
SN1001 to SN1625.
01
408
The mounting plate with the large O2 Connector seat
(right image) is installed in all HAMILTON-C2 from
SN1626.
English | 624165/07
High Pressure O2 Inlet Filter
15
15.2.2.8 High Pressure O2 Inlet Filter
Table 15-13. High Pressure O2 Inlet Filter (PN 160491)
Revision
00
History
Image
To prevent the O2 Proportional Valve (PN 160441)
from obstruction caused by particles entering the
device, the High Pressure O2 Inlet Filter (PN 160491)
has been implemented.
The High Pressure O2 Inlet Filter (PN 160491) is part
of the DISS (PN 160470) or NIST (PN 160471) Connector.
For devices which do not have a High Pressure O2
Inlet Filter installed, we strongly recommend installing
the filter to avoid device malfunction.
DISS Connector with Mounting Adapter
NIST Connector with Mounting Adapter
For this purpose, we provide the following solutions:
1.
HAMILTON-C2 with Oxygen Mixer Block
Assembly (MSP160226 Revision 00 and
Revision 01), the following items need to
be ordered for the update:
High Pressure O2 Inlet Filter
– Mounting Kit (PN 160496, includes Mounting
Adapter (PN 160494))
– O2 DISS Connector (PN 160470)
or
– O2 NIST Connector (PN 160471)
2.
HAMILTON-C2 with Oxygen Mixer Block
Assembly (MSP160226 Revision 02), the
following items need to be ordered for the
update:
– O2 DISS Connector (PN 160470)
or
– O2 NIST Connector (PN 160471)
The High Pressure O2 Inlet Filter (PN 160491) is part
of the yearly Preventive Maintenance and can be
ordered separately or order the HPO Inlet Filter Service Kit (PN 160497).
NOTICE
A number of devices with Mixer Block Assembly (MSP160226) REV 02 (SN 1590 - 1625) are not equipped with the large O2
connector seat. For these devices, it is necessary to order the Mounting Kit (PN 160496) together with the DISS or NIST connector.
Hamilton Medical | HAMILTON-C2 Service Manual
409
15
Hardware Revisions, Features and Compatibility
15.2.2.9 Pressure Sensor Module
Table 15-14. Pressure Sensor Module (MSP160300)
Revision
410
History
00
First official release.
01
Design of the Holding Plate improved.
Image
English | 624165/07
Blower Module Complete
15
15.2.2.10 Blower Module Complete
Table 15-15. Blower Module (MSP160250)
Revision
History
00
First official Release.
01
Noise reduction and Blower Turbine sealed up.
02
Assembly process improved.
03
Vibration damper included ECO M900.
04
Heat sink improvement.
05
Replacement of filter pressure board introduced.
06
Improved EMI behavior.
Image
15.2.2.11 Ambient Valve Complete
Table 15-16. Ambient Valve Complete (MSP160290)
Revision
History
Image
00
01
02
03
04
15.2.2.12 Inspiratory Valve Complete
Table 15-17. Inspiratory Valve Complete (MSP160230)
Revision
History
00
First official release in HAMILTON-C2 until SN 3004.
01
Housing and flow characteristic improved in
HAMILTON-C2 since SN 3005.
02
Stick-slip effect improved in HAMILTON-C2 since SN
4233.
Image
Neonatal implemented.
15.2.2.13 Expiratory Valve Complete
Table 15-18. Expiratory Valve Complete (MSP160240)
Revision
History
Image
00
01
Neonatal implemented.
Hamilton Medical | HAMILTON-C2 Service Manual
411
15
Hardware Revisions, Features and Compatibility
15.2.2.14 Oxygen Sensor
Table 15-19. Oxygen Sensor (PN 396200)
Revision
History
Image
00
01
First official release in HAMILTON-C2 from SN 1000.
02
Temperature compensation improved July 2013.
15.2.2.15 Oxygen Sensor Cover
Table 15-20. Oxygen Sensor Cover (PN 160401)
Revision
History
00
First official release.
01
Improved ultraviolet rays resistance.
02
Oxygen Sensor Cover improved.
Image
15.2.2.16 Battery Compartment
Table 15-21. Battery Compartment (PN 160302)
Revision
History
00
First release.
01
To prevent an increase of battery temperatures >
54°C 5, vent holes (A) are drilled on the top of the
Battery Compartment.
Image
15.2.2.17 Battery Door
Table 15-22. Battery Door PN (160344, old) or (PN 160506, new)
Revision
00
01
History
Image
PN 160344 Old Version
NOTICE! Obsolete
Old Version
00
PN 160506 New Version
If the broken Battery Cover is the Old Version (PN
160344), order the complete Rear Cover
(MSP160481), (see the Rear Cover Section 15.2.2.20).
New Version
412
English | 624165/07
Top Cover Handle
15
15.2.2.18 Top Cover Handle
Table 15-23. Top Cover Handle (PN 160342)
Revision
History
Image
00
01
To prevent damage of the device, the cylinder bolt (A)
was extended and locked against removing with a
retaining ring (B) inside.
The top cover can be replaced with PN 160601.
This improvement is implemented from HAMILTONC2 SN1271.
15.2.2.19 Front Cover
Table 15-24. Front Cover (MSP160480)
Revision
History
00
First official release.
01
Improved ultraviolet rays resistance.
02
Oxygen Sensor Cover improved.
Image
15.2.2.20 Rear Cover
Table 15-25. Rear Cover (MSP160481)
Revision
History
00
First official release.
01
Improved ultraviolet rays resistance.
02
Several further improvements like Battery Door.
Hamilton Medical | HAMILTON-C2 Service Manual
Image
413
15
414
Hardware Revisions, Features and Compatibility
English | 624165/07
16
Glossary
16.1
Definitions of Expression ......................................................................................................................416
16.2
Alarm Codes ........................................................................................................................................421
415
16
Glossary
16.1 Definitions of Expression
This glossary offers definitions of expressions not included, or extended definitions of expressions briefly included, in the glossary of the HAMILTON-C2 Operator’s Manual. It should be used together with the glossary in this guide.
Table 16-1. Definitions of Expression
Activate
An action on the Touch Screen or the P&T Control Knob to choose a function or action.
Air
Source used in the HAMILTON-C2
Alarm Buffer
An area of memory containing details of the twenty most recent alarms. The most recent six of the twenty
alarms can be accessed by activating the alarm symbol at the bottom left of the screen when the HAMILTONC2 is in normal operating mode.
Alarm Lamp
Indicates alarm conditions
• Red - High Priority Alarms and Technical Faults
• Yellow - Medium and Low Priority Alarms
Alarm Status Indicator
An area at the bottom of the Touch Screen of the HAMILTON-C2 which displays the alarm status and power
status.
Ambient State
A state that the HAMILTON-C2 uses when it cannot function normally because of an internal or external fault.
The Ambient State is often associated with Technical Faults, but can also be caused by such things as air or
oxygen supply. It is always accompanied by a high-priority patient alarm, sounded by the loudspeaker or
buzzer.
Ambient State Gas Flow
If the HAMILTON-C2 is in the Ambient State, the patient breathes unassisted.
Analog - Digital Conversion (ADC)
Converts an analog signal to a digital signal.
Audible Alarm
An alarm sounded when there is a Technical Fault or other error condition. An audible alarm is generated. A
buzzer is used as a backup if the audible alarm through the loudspeaker does not function.
Autorinse
See Rinse Flow
Autozero
A method to automatically adjust for electronic drift of a device due to temperature and environmental conditions.
Backlight
Lights used to help illuminate the screen.
Basic Input Output System (BIOS)
The basic input output system for the central processing unit (CPU). This is held on an EPROM or EEPROM
mounted on the ESM.
Basket
An accessory located on the back of the HAMILTON-C2 trolley.
Battery Charger
Circuits located in the driver poard that supply power to the primary and optional battery packs when mains
power is available.
• Primary battery pack - 14.4 VDC maximum charging voltage
• Optional battery pack - 14.4 VDC maximum charging voltage
Both battery packs can also be charged using an external charger.
Breathing Circuit
See Patient Breathing Circuit
Buzzer
A piezoelectric device mounted on the Control Board, used by the HAMILTON-C2 to sound some alarms. It
functions independently of the loudspeaker and the power supply, and typically indicates a high-priority technical faults that causes the HAMILTON-C2 to go into the ambient state. Despite its name, the buzzer makes a
high frequency sound.
NOTICE! Despite its name, the buzzer makes a high frequency sound.
Communication Board
A communication interface for the CO2, SpO2 and Nurse Call functions.
Communication Interface
The Ethernet Port on the rear of the HAMILTON-C2.
Configuration Menu
A screen used to select ventilation and other operation parameters of the HAMILTON-C2.
Control Knob
Used to select an option on the LCD screen by rotating and can be used to activate a selection on the LCD
screen by pressing.
416
English | 624165/07
Definitions of Expression
16
De-activate
An action on the Touch Screen or the P&T Control Knob to clear the selection of a function or action.
De-select
An action on the Touch Screen or the P&T Control Knob to clear the selection of a function or action.
Digital - Analog Converter (DAC)
Converts a digital signal to an analog signal.
DISS
Diameter index safety standard, a standard connector used for connection of oxygen to the HAMILTON-C2,
designed so the oxygen supply cannot be connected incorrectly.
dP Flow Sensor
Measures the differential pressure (pressure difference) across the flow sensor. The measurement is performed
by a pressure sensor inside the HAMILTON-C2, and is used to calculate airway gas flow.
EEPROM
Electrically Erasable Programmable Read Only Memory
Electrical Safety Tests
A set of electrical tests used to check the safety of a device.
Electrostatic discharge
(ESD)
Electrostatic Discharge
Emergency Buzzer Alarm
An alarm that sounds when a Technical Faults occurs that is serious enough to cause the HAMILTON-C2 to
switch to Ambient Mode.
In Ambient Mode, all valves switch to their unpowered position, and the HAMILTON-C2 is completely passive.
NOTICE! The buzzer makes a high frequency sound. It functions independently of the loudspeaker.
ESM Board
Embedded system module. A PCB containing the processor
Event Log
A record of most activity in the HAMILTON-C2. This includes user actions and internal activity such as:
• Calibration Results
• Alarms
• Technical Faults
• Controls Settings
• Configuration, Serial Numbers, Revision Numbers
• Switch-on Times
The Event Log always holds a maximum of 1,000 events. Switching OFF and ON the HAMILTON-C2 does not
delete the storage of the Event Log.
You can access a subset of the event log suited for clinical use by activating the Event Log symbol in the monitoring menu in normal operating mode.
Expiration
The act of the patient breathing out.
Expiratory Valve
A valve controlling pressure in the Patient Breathing Circuit, enabling the patient to exhale and the
HAMILTON-C2 to maintain PEEP.
Expiratory Valve Block
Connection
The plastic connector used to attach the Patient Breathing Circuit on the side of the HAMILTON-C2.
Fan Filter
A filter used to capture dust from the Cooling Fan.
Flow Restrictor
A device that restricts flow of a gas.
Front Panel Keys
Control buttons at the front of the Front Panel.
Front Panel
Part of the HAMILTON-C2 containing the user interface for interaction with the ventilator; LCD Display and
Hard Keys.
Galvanic Oxygen Sensor
See Oxygen Sensor
Gas Inlet
Connection for the inlet of oxygen to the HAMILTON-C2.
Gold Caps
One provides power for emergency buzzer alarm for 2 minutes and one for the real time clock for 3 months
when no power source is provided from mains power or battery.
Ground
Refers to the 0 electrical potential of a device.
Ground (GND)
Ground
HAMILTON-C2
Product name
HEPA
High efficiency particle air filter
Hamilton Medical | HAMILTON-C2 Service Manual
417
16
Glossary
Hot Swappable
A term used to indicate a device can be disconnected without removing power.
HPO
High Pressure Oxygen
Inspiration
The act of the patient breathing in.
Internal Temperature
The HAMILTON-C2 contains devices to measure the internal operating temperatures.
Light Emitting Diode
(LED)
Light Emitting Diode
LPO
Low Pressure Oxygen
Loudspeaker
A speaker used to indicate alarm conditions from the ventilation unit.
Main Power Switch
Powers the HAMILTON-C2 ON and OFF.
Metron EST Tester
Automated device used to perform the electrical safety test.
Microphone
In the HAMILTON-C2, a microphone is positioned near the loudspeaker as a feedback circuit to determine that
the loudspeaker is operating properly.
Nebulizer Connection
Connection on the side of the HAMILTON-C2 to attach tubing for the Nebulizer output.
Nebulizer Valve
Valve used to control the flow of oxygen to the Nebulizer Jar.
NIST
Non-interchangeable screw thread, a standard connector used for connection of oxygen to the HAMILTONC2, designed so the oxygen supply cannot be connected incorrectly.
Option Board
See Communication Board
O2 Cell
For O2 Cell, O2 Sensor or Oxygen Cell, see the Oxygen Sensor.
Check Valve
Used to restrict gas flow to only one direction.
Orifice Flow Restrictor
A device with a fixed opening which restricts the flow of a gas.
Orifice Tube
Used in the test configuration to perform adjustments in the service software.
Oxygen Sensor
A small, replaceable, plastic unit used by the HAMILTON-C2 to measure oxygen concentration.
The Oxygen Sensor reacts to the presence of oxygen, producing a voltage output in proportion to the oxygen
concentration.
The Oxygen Sensor must be replaced after a period of service, when it can no longer be calibrated. Typically,
about one year.
Oxygen Sensor Calibration
A procedure which supplies a controlled flow of air, oxygen and air/oxygen mixture to the Oxygen Sensor for
calibration.
Pambient
A sensor which measures the ambient pressure or the room pressure.
Patient Alarms
An alarm indicating that there is a problem or potential problem in ventilating the patient.There are three levels of patient alarms: high, medium and low. They are indicated by beeps on the loudspeaker, messages on
the display and the alarm lamp.
Patient Breathing Circuit
A Patient Breathing Circuit carries the air/oxygen mixture to the patient and carries the expired (exhaust) air
from the patient. It may include Tubings, Mask, Inspiratory Filter, Flow Sensor, Nebulizer Jar, Expiratory Valve
Cover and Membrane.
Paw
A sensor which measures the pressure at the proximal side of the Patient Flow Sensor.
Pfilter
A sensor which measures the pressure after the HEPA Filter; used to indicate if the HEPA filter needs to be
replaced.
Pflowsensor
A sensor which measures the differential pressure across the patient airway flow sensor.
Portable Tank
Oxygen tanks used for supply in a portable environment when the facility source is not available.
Press and Turn Control
Knob (P&T)
A button on the front of the front panel used to select and activate screen functions.
Press and Turn Encoder
(P&T Encoder)
An electrical device which provides input signals for control of screen functions.
Pressure Gauge
A mechanical device used to measure gas pressure.
Pressure Regulator
A device that regulates the pressure and restricts the flow of a gas.
Pressure Sensor Assembly
Provides components for pressure measurements used to monitor ventilation.
418
English | 624165/07
Definitions of Expression
16
Preventive Maintenance
(PM)
A term used for periodic maintenance of a device with specific planned tasks and items to check, adjust and
replace.
Principal Gas Flow
The main gas flow through the HAMILTON-C2 from the HEPA Filter for air and the oxygen inlets to the
patient, and then from the patient through the expiratory valve.
Pvent_control
A sensor which measures the pressure at the patient outlet, used for controlling.
Pvent_monitor
A sensor which measures the pressure at the patient outlet, used for monitoring.
QO2 Flow Sensor
Used to measure the oxygen flow into the Blower Module.
Qvent Flow Sensor
Used to measure the air/oxygen flow into the Patient Breathing Circuit.
Rinse Flow
A very small, continuous flow of gas through both the blue (patient side; proximal) and clear (ventilator side;
distal) tubes to the Patient Flow Sensor.
The flow minimizes the possibility of tube blockage, and hinders the potential migration of bacteria and
viruses from the patient’s expired gases through the tubes, towards the pressure sensors inside the ventilator.
Safety Mode
An emergency state that ensures a basic minute ventilation while giving the user time for corrective actions in
case of some technical fault alarms. The default inspiratory pressure is maintained, the expiratory valve opens
as needed to switch system pressure levels between PEEP and inspiratory pressure, and patient sensing is nonfunctional.
Select
An action on the touchscreen or the P&T control knob to choose a function or action.
Service Software
The software used to perform tests, adjustments and calibrations of the HAMILTON-C2. Also enables viewing
and exporting data from the Event Log. The HAMILTON-C2 cannot be used for patient ventilation when in
service software mode.
NOTICE! The HAMILTON-C2 cannot be used for patient ventilation when in Service Software Mode.
Service Software Mode
The operational status of the HAMILTON-C2 when in the service software.
Smart Battery Pack
Li-Ion battery, hot swappable which monitors and communicates the battery condition.
SoH
The State of Health (SoH) is a "measurement" that reflects the general condition of a battery and its ability to
deliver the specified performance compared with a fresh battery. It takes into account such factors as charge
acceptance, internal resistance, voltage and self-discharge. It is a measure of the long term capability of the
battery and gives an "indication" not an absolute measurement, of how much of the available "lifetime
energy throughput" of the battery has been consumed, and how much is remaining. Using the automotive
analogy, it can be compared to the "odometer"display function which indicates the number of miles traveled
since the vehicle was new.
T-Fitting
A type of tubing connector in the shape of a T.
Technical Events
A very minor fault or event recorded by the HAMILTON-C2 in the Event Log for use only by software developers.
Technical Faults
An alarm condition indicating a major malfunction of the HAMILTON-C2. (This contrasts with an alarm, that
indicates a problem with the status of a patient.) Technical Faults are intended to alert users and engineers of
the need for intervention, and are recorded in the Event Log.
Test Configuration (TC)
A particular configuration of devices and tubings use to perform the tests/calibrations in the service software.
Touch Screen
An interaction method where the LCD display screen can be touched with a finger to activate or select a function.
Trolley
A part of the HAMILTON-C2 on which the ventilator is mounted for transport within the customer location.
Update
An update is an improvement to an existing function. An update normally involves only software. A software
update is generally a revision number increment in a digit after the decimal point: for example, from 3.2 to
3.3.
Upgrade
An upgrade is the addition of new functions to a device. There are three ways to perform an upgrade:
• Add a hardware item that offers additional functions
• Upgrade to a higher software revision indicated by a higher value before the decimal point: for example
01.03 to 02.00
• Upgrade to a higher type of software
Hamilton Medical | HAMILTON-C2 Service Manual
419
16
Glossary
Variable Orifice Membrane
A variable flow restrictor as used in the Patient Flow Sensor.
Y-Fitting
A type of tubing connector in the shape of a Y.
420
English | 624165/07
Alarm Codes
16
16.2 Alarm Codes
16.2.1 Patient Alarms - 100000 Alarm Codes
Table 16-2. Patient Alarms - 100000 Alarm Codes
SID Number
CSystemConfiguration ID
enGB Text Value
132001
paBM_pawPressureLow
Loss of PEEP
132002
paBM_pawTubing
Check flow sensor tubing
132003
paBM_qawFlowSensorTubing
Check flow sensor
132004
paBM_qawFlowSensorMissing
Check flow sensor
132005
paBM_O2SensorDefect
Oxygen sensor defective
132006
paBM_O2SensorMissing
Oxygen sensor n/a
132007
paBM_O2SensorInvalid
Oxygen sensor not system compatible
132008
paBM_O2SensorCalibrationNeeded
Oxygen sensor calibration needed
132010
paBM_CO2SensorMissing
CO2 sensor disconnected
132011
paBM_CO2SensorDefect
CO2 sensor defect
132012
paBM_CO2SensorOverTemperature
CO2 sensor over temperature
132013
paBM_CO2SensorWarmUp
CO2 sensor warmup
132014
paBM_wrongFlowsensor
Wrong flow sensor
132015
paBM_FlowSensorCalibrationNeeded
Flow sensor calibration needed
Breath Monitoring
Breath Pattern Generation
132016
paBM_CO2CheckAirwayAdapter
Check CO2 airway adapter
132017
paBM_CO2CheckSamplingLine
Check CO2 sampling line
133001
paBPG_apnea
n/a
133002
paBPG_volumeLimitReached
Inspiratory volume limitation
133003
paBPG_pressureLimitReached
High pressure
133004
paBPG_CheckPatientInterface
Ventilation Mode Control
141001
paVMC_pressureLimitation
Pressure limitation
141002
paVMC_VThigh
Vt high
141003
paVMC_VTlow
Vt low
141004
paVMC_expMinVolHigh
High minute volume
141005
paVMC_expMinVolLow
Low minute volume
141006
paVMC_fTotalHigh
High frequency
141007
paVMC_fTotalLow
Low frequency
141008
paVMC_oxygenHigh
High oxygen
141009
paVMC_oxygenLow
Low oxygen
141010
paVMC_apnea
Apnea
141011
paVMC_apneaVentilationEnd
Apnea ventilation ended
141012
paVMC_apneaVentilation
Apnea ventilation
141013
paVMC_highPressureDuringSigh
High pressure during sigh
141014
paVMC_turnFlowSensor
Turn the flow sensor
Hamilton Medical | HAMILTON-C2 Service Manual
421
16
Glossary
SID Number
CSystemConfiguration ID
enGB Text Value
141016
paVMC_disconnectionVentilator
Disconnection on ventilator side
141017
paVMC_disconnectionPatient
Disconnection on patient side
141018
paVMC_exhalationObstructed
Exhalation obstructed
141019
paVMC_IRV
IRV
141020
paVMC_ASVunableToReachTarget
ASV: Cannot meet target
141021
paVMC_ASVplimitChanged
Pressure limit has changed
141023
paVMC_sensorFailMode
External flow sensor failed
141024
paVMC_pressureLow
Low pressure
141025
paVMC_PetCO2High
PetCO2 high
141026
paVMC_PetCO2Low
PetCO2 low
141027
paVMC_suctioningManoeuvre
Suctioning maneuver
141028
paVMC_performanceLimitedByHighAltitude
Performance limited by high altitude
141051
paVMC_PEEPHigh
High PEEP
141052
paVMC_PEEPLow
Low PEEP
422
English | 624165/07
Technical Alarms - 200000 Alarm Codes
16
16.2.2 Technical Alarms - 200000 Alarm Codes
Table 16-3. Technical Alarms - 200000 Alarm Codes
Error No.
Error Description
enGB Text Value
231001
taGD_pressureControllerPressureLow
Technical event
231002
taGD_pressureControllerPressureHigh
Technical event
231003
taGD_flowControllerFlowLow
Technical event
231004
taGD_flowControllerFlowHigh
Technical event
231005
taGD_inspirationValveLeak
Technical event
231006
taGD_O2ControllerFlowLow
Oxygen supply failed
231007
taGD_O2ControllerFlowHigh
Technical event
231008
taGD_O2ValveLeak
Technical event
231009
taGD_blowerControllerSpeedLow
Technical event
231010
taGD_blowerControllerSpeedHigh
Technical event
231011
taGD_ventOutputTemperatureHigh
Vent outlet temperature high
231012
taGD_qventFlowSensorDefect
Technical event
231013
taGD_qO2FlowSensorDefect
Technical event
231014
taGD_ambientValveError
Technical event
231015
taGD_autozeroPventControlOK
n/a
231016
taGD_autozeroPventControlOutOfRange
n/a
231017
taGD_blowerServiceRequired
Blower service required
231018
taGD_selftestWithError
Technical event
231019
taGD_iInspValveSensorDefect
Technical event
231020
taGD_regulatorDataLogOn
Technical event
231021
taGD_o2PresenceCheckFailed
n/a
231022
taGD_pExpValveSensorDefect
Technical event
231023
taGD_invalidFlowSensor
Technical event
231026
taGD_expValveCoverInvalid
231028
taGD_pressureControllerFlowHigh
n/a
231029
taGD_autozeroPExpValveOK
n/a
231030
taGD_autozeroPExpValveError
n/a
231031
taGD_expValveCoverUniversal
n/a
231032
taGD_expirationValveDisconnected
n/a
231033
taGD_blowerControllerSpeedLimited
n/a
231040
taGD_calibratePVentControlOK
n/a
231041
taGD_calibratePVentControlError
n/a
231044
taGD_calibrateExpValveOK
231045
taGD_calibrateExpValveError
231046
taGD_calibrateIExpValveOK
n/a
231047
taGD_calibrateIExpValveError
n/a
231099
taGD_communicationCheck
n/a
Gas Delivery
Hamilton Medical | HAMILTON-C2 Service Manual
423
16
Glossary
Error No.
Error Description
enGB Text Value
Breath Monitoring (SERVICE SOFTWARE)
232001
taBM_pressureSensorTolerance
n/a
232002
taBM_pventMonitorSensorDefect
Technical event
232003
taBM_pawSensorDefect
Technical event
232004
taBM_pressureNotReleased
Pressure not released
232005
taBM_blowerHot
Technical event
232006
taBM_blowerTemperatureSensorDefect
Technical event
232007
taBM_qawFlowSensorDefect
Check flow sensor tubing
232008
taBM_pambientSensorDefect
Technical event
232009
taBM_pventAutozeroValveError
n/a
232010
taBM_pventAutozeroValveMissing
n/a
232011
taBM_proximalAutozeroValveError
n/a
232012
taBM_proximalAutozeroValveMissing
n/a
232013
taBM_distalAutozeroValveError
n/a
232014
taBM_distalAutozeroValveMissing
n/a
232015
taBM_proximalDistalAutozeroValveMissing
n/a
232016
taBM_proximalDistalAutozeroValveError
n/a
232017
taBM_autozeroPventMonitorOK
n/a
232018
taBM_autozeroPventMonitorOutOfRange
n/a
232019
taBM_autozeroPventControlOK
n/a
232020
taBM_autozeroPventControlOutOfRange
n/a
232022
taBM_autozeroPawOutOfRange
n/a
232024
taBM_autozeroQawOutOfRange
n/a
232025
taBM_autozeroPawQawOK
n/a
232026
taBM_autozeroPawQawOutOfRange
n/a
232027
taBM_instrumentTemperatureHigh
Device temperature high
232028
taBM_pFilterPressureHigh
Technical event
232029
taBM_tinstSensorDefect
Technical event
232030
taBM_clockError
Technical event
232031
taBM_fiO2CalibrationError
n/a
232032
taBM_fiO2CalibrationOK
n/a
232034
taBM_O2SensorError
Replace Oxygen Sensor
232035
taBM_pfilterSensorDefect
Technical event
232036
taBM_co2CalibrationError
n/a
232037
taBM_co2CalibrationOK
n/a
232038
taBM_co2CalibrationNeeded
CO2 calibration needed
232040
taBM_disconnectionQuick
n/a
232041
taBM_co2CalibrationStartError
n/a
232042
taBM_co2CalibrationStartOK
n/a
232043
taBM_reconnectionQuick
n/a
232046
taBM_autozeroExpValveOK
n/a
424
English | 624165/07
Technical Alarms - 200000 Alarm Codes
Error No.
Error Description
enGB Text Value
232047
taBM_autozeroExpValveError
n/a
232048
taBM_autozeroProxValveLeak
n/a
232049
taBM_autozeroDistValveLeak
n/a
232050
taBM_pventMonitorCalibrationError
n/a
232051
taBM_pventMonitorCalibrationOK
n/a
232052
taBM_pawCalibrationError
n/a
232053
taBM_pawCalibrationOK
n/a
232054
taBM_qawCalibrationError
n/a
232055
taBM_qawCalibrationOK
n/a
232056
taBM_pAmbientPfilterMismatch
Technical event
232099
taBM_communicationCheck
n/a
233001
taBPG_autozeroPventMonitorFail
Technical event
233002
taBPG_autozeroPventControlFail
Technical event
233003
taBPG_autozeroPawFail
Technical event
233004
taBPG_autozeroQawFail
Technical event
233005
taBPG_pressureSensorTolerance
Technical event
233006
taBPG_nebulizerValveError
Technical event
233020
taBPG_autozeroPventOK
n/a
233021
taBPG_autozeroPawQawOK
n/a
233022
taBPG_autozeroVentValveOK
n/a
233023
taBPG_autozeroVentValveMissing
n/a
233024
taBPG_autozeroVentValveError
n/a
233025
taBPG_autozeroProxValveOK
n/a
233026
taBPG_autozeroProxValveMissing
n/a
233027
taBPG_autozeroProxValveError
n/a
233028
taBPG_autozeroDistValveOK
n/a
233029
taBPG_autozeroDistValveMissing
n/a
233030
taBPG_autozeroDistValveError
n/a
233031
taBPG_autozeroPventMonitorTimeout
n/a
233032
taBPG_autozeroPventControlTimeout
n/a
233033
taBPG_autozeroPawQawTimeout
n/a
233034
taBPG_autozeroPventMonitorOutOfRange
n/a
233035
taBPG_autozeroPventControlOutOfRange
n/a
233036
taBPG_autozeroPawOutOfRange
n/a
233037
taBPG_autozeroQawOutOfRange
n/a
233038
taBPG_autozeroProxDistValveMissing
n/a
233039
taBPG_autozeroProxDistValveError
n/a
233040
taBPG_autozeroExpValveOK
n/a
233041
taBPG_autozeroExpValveError
n/a
233042
taBPG_autozeroExpValveTimeout
n/a
233043
taBPG_autozeroExpValveFail
n/a
Hamilton Medical | HAMILTON-C2 Service Manual
16
425
16
Glossary
Error No.
Error Description
enGB Text Value
233044
taBPG_autozeroProxValveLeak
n/a
233045
taBPG_autozeroDistValveLeak
n/a
Ventilation Mode Control
234099
taSND_communicationCheck
n/a
241001
taVMC_replaceHEPAfilter
Replace HEPA filter
241002
taVMC_tightnessTestReleaseValveDefect
Alarming
243001
taALR_alarmSilenceError
Technical event
243002
taALR_alarmUnknown
Technical event
243003
taALR_loudspeakerDefect
Loudspeaker defective
243004
taALR_buzzerDefectAtStartup
Buzzer defective
243006
taALR_developSound
Technical event
243007
taALR_rtcReset
Realtime clock failure
244001
taPM_externalPowerLoss
Loss of external power
244002
taPM_batteryCalibrationRequiredBat1
Battery 1: Calibration required
244003
taPM_batteryCalibrationRequiredBat2
Battery 2: Calibration required
244004
taPM_batteryTemperatureHighBat1
Battery 1: Temperature high
244005
taPM_batteryTemperatureHighBat2
Battery 2: Temperature high
244006
taPM_batteryPowerLowPriorityHigh
Battery low
244007
taPM_batteryPowerLoss
Battery power loss
244008
taPM_battery1WrongBattery
Battery 1: Wrong battery
244009
taPM_battery2WrongBattery
Battery 2: Wrong battery
244010
taPM_batteryPowerLowPriorityLow
Battery low
244011
taPM_batterySystemManagerBusError
Battery communication error
244012
taPM_battery1PermanentFailure
Battery 1: Defective
244013
taPM_battery2PermanentFailure
Battery 2: Defective
244014
taPM_batteryTaskLifeSignTimeout
Technical event
244015
taPM_batteryPowerLowPriorityMedium
Battery low
244016
taPM_batteryReplacementRequiredBat1
Battery 1: Replacement required
244017
taPM_batteryReplacementRequiredBat2
Battery 2: Replacement required
244099
taPM_communicationCheck
n/a
246001
taLLS_serviceNeeded
Preventive maintenance required
246002
taLLS_cpuTemperatureHigh
Technical event
246003
taLLS_fanError
Fan failure
246004
taLLS_processorOverload
Technical event
246005
taLLS_alarmMonitorDefect
Technical event
246006
taLLS_eepromDefaults
Technical event
246007
taLLS_eepromWriteFailed
Technical event
246008
taLLS_cpuTemperatureDefect
Technical event
Power Management
Low Level Security
426
English | 624165/07
Technical Alarms - 200000 Alarm Codes
Error No.
Error Description
enGB Text Value
246009
taLLS_devWatchdogDisabled
Technical event
246010
taLLS_hardwarepmeterError
Technical event
246012
taLLS_postEthernetError
Technical event
246013
taLLS_postI2CError
Technical event
246014
taLLS_postBootloaderError
Technical event
246016
taLLS_postTouchError
Touch not functional
246017
taLLS_externalDisplayError
Technical event
249001
taCFG_OhOptionFileNotRead
Options not found
249002
taCFG_OhOptionFileNotWritten
Technical event
249003
taCFG_OhOptionFileWrongVersion
Technical event
249004
taCFG_OhOptionFileDefaultGenerated
Technical event
249010
taCFG_DscDeviceConfigFileError
Technical event
249011
taCFG_DscSetupConfigFileError
Technical event
249012
taCFG_DscLastSettingConfigFileError
Technical event
249020
taCFG_InvalidOptionBoard
Invalid Communication Board
249021
taCFG_CheckHardwareCompatibility
Invalid Communication Board
249099
taCFG_communicationCheck
n/a
255001
taIVS_PercentageMinVolHighFailed
n/a
255002
taIVS_PercentageMinVolHighLimitWrong
n/a
255003
taIVS_PercentageMinVolLowFailed
n/a
255004
taIVS_PercentageMinVolLowLimitWrong
n/a
255005
taIVS_PeepHighLimitWrong
n/a
255006
taIVS_PeepHighFailed
n/a
255007
taIVS_PeepLowLimitWrong
n/a
255008
taIVS_PeepLowFailed
n/a
255009
taIVS_OxygenLowFailed
n/a
255010
taIVS_OxygenHighFailed
n/a
255011
taIVS_PercentageMinVolChangePosFailed
n/a
255012
taIVS_PercentageMinVolChangeNegFailed
n/a
255013
taIVS_PeepChangePosFailed
n/a
255014
taIVS_PeepChangeNegFailed
n/a
255015
taIVS_OxygenChangePosFailed
n/a
255016
taIVS_OxygenChangeNegFailed
n/a
255017
taIVS_RecruitmentRunningTime
n/a
255018
taIVS_SimulationSuperviseFailed
n/a
255019
taIVS_PercentageMinVolHighFineFailed
255020
taIVS_PercentageMinVolHighLimitFineWrong
255021
taIVS_PercentageMinVolLowFineFailed
255022
taIVS_PercentageMinVolLowLimitFineWrong
255023
taIVS_CtrlOszillationFiO2
255024
taIVS_CtrlOszillationPEEP
Hamilton Medical | HAMILTON-C2 Service Manual
16
427
16
Glossary
Error No.
Error Description
enGB Text Value
255025
taIVS_CtrlOszillationPercMinVol
255026
taIVS_DataMismatch
Technical event
255027
taIVS_PetCO2FrozenFailed
Technical event
255028
taIVS_SpO2FrozenFailed
255029
taIVS_fSpontFrozenFailed
Technical event
283001
taVGUI_StartupFailed
Technical event
283003
taVGUI_languageNotLoaded
Technical event
283004
taVGUI_deviceConfigFileError
Technical event
Ventilation GUI
Technical event
283005
taVGUI_setupConfigFileError
Technical event
283007
taVGUI_lastSettingError
Technical event
283008
taVGUI_checkSetting
Check settings
283009
taVGUI_selftestFailed
Self test failed
284001
taSGUI_StartupFailed
n/a
284002
taSGUI_alarmServiceHigh
Technical event
284003
taSGUI_alarmServiceMedium
Technical event
284004
taSGUI_alarmServiceLow
Technical event
284005
taSGUI_FlowSensorCalibrationNeeded
Technical event
284006
taSGUI_RestartDeviceToEnableTestsCalib
Technical event
285001
taAGL_alarmLampsErrorDefect
Technical event
285002
taAGL_alarmLampsWarningDefect
Technical event
285003
taAGL_BacklightDefect
Technical event
285004
taAGL_HardkeyDefect
Function key not operational
Service GUI
428
English | 624165/07
Technical Failure Ending in Safety Mode - 300000 Alarm Codes
16
16.2.3 Technical Failure Ending in Safety Mode - 300000 Alarm Codes
Table 16-4. Technical Failure Ending in Safety Mode - 300000 Alarm Codes
Error No.
Error Description
Safety ventilation
Gas Delivery
331001
tfsGD_pventPressureSensorDefect
Safety ventilation
Safety ventilation
Breath Monitoring
332001
enGB Text Value
tfsBM_qawFlowSensorError
Safety ventilation
Ventilation Mode Control
Safety ventilation
341001
tfsVMC_breathSettingsNotAccepted
Safety ventilation
341002
tfsVMC_returnedBreathSettingsIncorrect
Safety ventilation
341003
tfsVMC_adaptiveSettingsInvalid
Safety ventilation
341004
tfsVMC_breathSettingsTimeout
Safety ventilation
341005
tfsVMC_controllerSettingsInvalid
Safety ventilation
341006
tfsVMC_returnedAlarmLimitsIncorrect
Safety ventilation
341007
tfsVMC_modeTransitionNotDefined
Safety ventilation
343001
tfsALR_communicationTimeout
Safety ventilation
346002
tfsLLS_watchdogFailedALR
Safety ventilation
346003
tfsLLS_watchdogFailedALR_MFmeasure
Safety ventilation
346004
tfsLLS_watchdogFailedESL
Safety ventilation
346005
tfsLLS_watchdogFailedESL_MMILog
Safety ventilation
346006
tfsLLS_watchdogFailedESL_ControlLog
Safety ventilation
346007
tfsLLS_watchdogFailedESL_BreathLog
Safety ventilation
346008
tfsLLS_watchdogFailedESL_EventSDRReader
Safety ventilation
346009
tfsLLS_watchdogFailedESL_ServiceSDRReader
Safety ventilation
346010
tfsLLS_watchdogFailedESL_EventSDRWriter
Safety ventilation
346011
tfsLLS_watchdogFailedESL_ServiceSDRWriter
Safety ventilation
346012
tfsLLS_watchdogFailedESL_MFmeasure
Safety ventilation
346013
tfsLLS_watchdogFailedGUIL_Touch
Safety ventilation
346014
tfsLLS_watchdogFailedGUIL_PTKnob
Safety ventilation
346015
tfsLLS_watchdogFailedGUIL_Hardkeys
Safety ventilation
346016
tfsLLS_watchdogFailedGUIL_ScreenShot
Safety ventilation
346017
tfsLLS_watchdogFailedLM
Safety ventilation
346019
tfsLLS_watchdogFailedLLS_HWException
Safety ventilation
346020
tfsLLS_watchdogFailedLLS_CommonServer
Safety ventilation
346022
tfsLLS_watchdogFailedLLS_AlarmStatus
Safety ventilation
346023
tfsLLS_watchdogFailedLLS_MFmeasure
Safety ventilation
346024
tfsLLS_watchdogFailedPM
Safety ventilation
346025
tfsLLS_watchdogFailedPM_MFmeasure
Safety ventilation
346026
tfsLLS_watchdogFailedQSPI
Safety ventilation
346027
tfsLLS_watchdogFailedQSPI_MMI_Server
Safety ventilation
346028
tfsLLS_watchdogFailedQSPI_BM_Server
Safety ventilation
Hamilton Medical | HAMILTON-C2 Service Manual
429
16
Glossary
Error No.
Error Description
enGB Text Value
346029
tfsLLS_watchdogFailedQSPI_BPG_Server
Safety ventilation
346030
tfsLLS_watchdogFailedQSPI_LLS_Server
Safety ventilation
346031
tfsLLS_watchdogFailedQSPI_QuadSPI
Safety ventilation
346032
tfsLLS_watchdogFailedQSPI_MFmeasure
Safety ventilation
346033
tfsLLS_watchdogFailedRTC_RealTimeClock
Safety ventilation
346034
tfsLLS_watchdogFailedRTC_AlarmClock
Safety ventilation
346035
tfsLLS_watchdogFailedSC
Safety ventilation
346036
tfsLLS_watchdogFailedSGUI
Safety ventilation
346037
tfsLLS_watchdogFailedSTU
Safety ventilation
346038
tfsLLS_watchdogFailedVMC
Safety ventilation
346039
tfsLLS_watchdogFailedVMC_VentAlarming
Safety ventilation
346040
tfsLLS_watchdogFailedVMC_VentControl
Safety ventilation
346041
tfsLLS_watchdogFailedVMC_VentMonitoring
Safety ventilation
346042
tfsLLS_watchdogFailedVMC_MFmeasure
Safety ventilation
346043
tfsLLS_watchdogFailedVGUI
Safety ventilation
346044
tfsLLS_watchdogFailedVGUI_ModeControl
Safety ventilation
346045
tfsLLS_watchdogFailedVGUI_MFmeasure
Safety ventilation
346046
tfsLLS_watchdogFailedLLS_eepromWrite
Safety ventilation
346047
tfsLLS_watchdogFailedBM_AlarmingSlow
Safety ventilation
346048
tfsLLS_watchdogFailedSND_SoundControl
Safety ventilation
346049
tfsLLS_watchdogFailedSND_Sound
Safety ventilation
346050
tfsLLS_watchdogFailedVT
Safety ventilation
346051
tfsLLS_watchdogFailedCFG_Configuration
Safety ventilation
346052
tfsLLS_watchdogFailedEXM_ComBase
Safety ventilation
346053
tfsLLS_watchdogFailedIVS
n/a
346054
tfsLLS_safetyFailureDetected
Safety ventilation
346055
tfsLLS_watchdogFailedPD_BlackBoxLogging
Safety ventilation
383001
tfsVGUI_settingsNotAccepted
Safety ventilation
383002
tfsVGUI_returnedSettingsIncorrect
Safety ventilation
383003
tfsVGUI_settingsValidation
Safety ventilation
383004
tfsVGUI_monitoringChannelObservationFailed
Safety ventilation
383005
tfsVGUI_VMCTimeout
Safety ventilation
383006
tfsVGUI_returnedDeviceSettingsIncorrect
Safety ventilation
383007
tfsVGUI_trendingChannelObservationFailed
Safety ventilation
Ventilation GUI
Application GUI Library
385001
tfsAGL_alarmingChannelObservationFailed
Safety ventilation
385002
tfsAGL_safetyFailureDetected
Safety ventilation
385003
tfsAGL_PMChannelObservationFailed
Safety ventilation
tfsGUIL_bitmapNotLoaded
Safety ventilation
GUI Library
386001
430
English | 624165/07
Technical Failure Ending in Ambient Mode - 400000 Alarm Codes
16
16.2.4 Technical Failure Ending in Ambient Mode - 400000 Alarm Codes
Table 16-5. Technical Failure Ending in Ambient Mode - 400000 Alarm Codes
Error No.
Error Description
enGB Text Value
431001
tfaGD_blowerFault
Blower fault
431002
tfaGD_blowerDisconnected
Technical fault
431004
tfaGD_inspirationValveOverCurrent
Technical fault
431006
tfaGD_expirationValveOverCurrent
Technical fault
431007
tfaGD_ventOutputOverTemperature
Technical fault
431008
tfaGD_qventFlowSensorError
Technical fault
431009
tfaGD_qO2FlowSensorError
Technical fault
431010
tfaGD_controlREGSPITimeout
Technical fault
431011
tfaGD_monitorREGSPITimeout
Technical fault
431012
tfaGD_flowSensorMeasurementImprecise
Technical fault
431013
tfaGD_calibrationReadFailed
Technical fault
431014
tfaGD_iExpValveSensorDefect
Technical fault
431015
tfaGD_emergencyOffFailed
Technical fault
431017
tfaGD_inspirationValveDisconnected
Technical fault
432001
tfaBM_blowerOverTemperature
Technical fault
432002
tfaBM_instrumentOverTemperature
Technical fault
432003
tfaBM_pressureNotReleasedAmbient
n/a
433001
tfaBPG_breathMonitoringTickTimeout
Technical fault
433002
tfaBPG_controlTickTimingError
n/a
443001
tfaALR_watchdogFailedLLS
Technical fault
444001
tfaPM_batteriesTotalDischarge
Battery totally discharged
444004
tfaPM_voltageOutOfTolerance
Technical fault
444005
tfaPM_shutdownFailed
Technical fault
446001
tfaLLS_cpuTemperatureCritical
Technical fault
446002
tfaLLS_safetyFailed
Technical fault
446003
tfaLLS_watchdogFailedGD
Technical fault
446004
tfaLLS_watchdogFailedGD_ValveRegulator
Technical fault
446005
tfaLLS_watchdogFailedGD_BlowerRegulator
Technical fault
446006
tfaLLS_watchdogFailedGD_Monitoring
Technical fault
446007
tfaLLS_watchdogFailedGD_ValveLog
Technical fault
446008
tfaLLS_watchdogFailedGD_BlowerLog
Technical fault
446009
tfaLLS_watchdogFailedGD_MFmeasure
Technical fault
446010
tfaLLS_watchdogFailedGD_GPIOFaultInput
Technical fault
446011
tfaLLS_watchdogFailedBM
Technical fault
446012
tfaLLS_watchdogFailedBM_GasDeliveryCom
Technical fault
446013
tfaLLS_watchdogFailedBM_Hardwarecontrol
Technical fault
446014
tfaLLS_watchdogFailedBM_BPGclientReceiver
Technical fault
446015
tfaLLS_watchdogFailedBM_BPGserverReceiver
Technical fault
Hamilton Medical | HAMILTON-C2 Service Manual
431
16
Glossary
Error No.
Error Description
enGB Text Value
446016
tfaLLS_watchdogFailedBM_BreathDataCalc
Technical fault
446017
tfaLLS_watchdogFailedBM_MFmeasure
Technical fault
446018
tfaLLS_watchdogFailedBPG
Technical fault
446019
tfaLLS_watchdogFailedBPG_GDclientReceiver
Technical fault
446020
tfaLLS_watchdogFailedBPG_MFmeasure
Technical fault
446021
tfaLLS_exceptionHappened
Technical fault
446022
tfaLLS_voltageError
Technical fault
446023
tfaLLS_watchdogFailedLLS_Workload
Technical fault
446024
tfaLLS_Am3v3Error
Technical fault
446025
tfaLLS_AmAdcError
Technical fault
446026
tfaLLS_AmVrefError
Technical fault
446028
tfaLLS_clockError
Technical fault
446029
tfsLLS_ambientFailureDetected
Technical fault
446030
tfaLLS_AmWatchdogOccurred
Technical fault
446031
tfaLLS_restartVentilationFailed
Technical fault
481001
tfaSTU_firstRunMigrationError
Technical fault
481002
tfaSTU_crcError
Technical fault
481003
tfaSTU_unknownPartNumber
Unknown part number
481004
tfaSTU_technicalStateError
Technical state failed
483001
tfaVGUI_StartupBasicStartupFailure
Technical fault
483002
tfaVGUI_StartupScreenFailure
Technical fault
483003
tfaVGUI_StartupLoadGuiBasicsFailure
Technical fault
483004
tfaVGUI_StartupSelftestPrologFailure
Technical fault
483005
tfaVGUI_StartupSelftestFailure
Technical fault
483006
tfaVGUI_StartupSelftestEpilogFailure
Technical fault
483007
tfaVGUI_StartupLoadingBitmapsFailure
Technical fault
483008
tfaVGUI_StartupInitializationFailure
Technical fault
484001
tfaSGUI_StartupBasicStartupFailure
Technical fault
484002
tfaSGUI_StartupScreenFailure
Technical fault
484003
tfaSGUI_StartupLoadGuiBasicsFailure
Technical fault
484004
tfaSGUI_StartupSelftestPrologFailure
Technical fault
484005
tfaSGUI_StartupSelftestFailure
Technical fault
484007
tfaSGUI_StartupLoadingBitmapsFailure
Technical fault
484008
tfaSGUI_StartupInitializationFailure
Technical fault
485001
tfaAGL_ambientFailureDetected
Technical fault
432
English | 624165/07
500000 Alarm Codes
16
16.2.5 500000 Alarm Codes
These are classed as “invisible technical faults” and are not shown on the LCD Display during operation. They refer to exception handling performed by HAMILTON-C2 during operation, and are for Hamilton Medical internal use.
Do not consult Hamilton Medical Technical Support concerning these technical faults. They have no significance for hospital
workers or Hamilton-Medical Trained Field Service Engineers.
Hamilton Medical | HAMILTON-C2 Service Manual
433
16
434
Glossary
English | 624165/07
17
Test Report
17.1
Summary of Test Report.......................................................................................................................436
435
17
Test Report
17.1 Summary of Test Report
NOTICE
For details on the tests, checks, calibrations, and maintenance procedures, refer to the relevant Chapters/Sections of the
HAMILTON-C2 Service Manual (PN 624165) and the HAMILTON-C2 Operator's Manual (PN 624131).
Table 17-1. Unit Information
HAMILTON-C2
Serial Number:
SN __ __ __ __ __ __
Inventory Number:
__ __ __ __ __ __ __
Software Version:
__ __ __ __ __ __ __
Location of Installation:
________________________________
Name of Customer:
________________________________
Date (YYYY/MM/DD):
__ __ __ __ /__ __ /__ __
Table 17-2. Component Replacement
Task
Yes
No
Not Applicable
HAMILTON-C2 Manual Reference
Oxygen Sensor replaced?
See the Operator's Manual Section
10.3.3.
Air Intake Dust Filter cleaned or replaced?
See the Operator's Manual Section
10.3.1.
Fan Filter cleaned or replaced?
See the Operator's Manual Section
10.3.1.
O2 Inlet Filter replaced?
See the Operator's Manual Section
13.3.
HEPA Filter replaced?
See the Operator's Manual Section
10.3.1.
Batteries calibrated (% error > 5%)?
—
See the Service Manual Section 5.
Batteries replaced if State of Health (SoH) < 20%
—
See the Service Manual Section 5.
Blower Module replaced (ebt ≥ 100%)?
—
See the Service Manual Section 8.9.2
and Section 10.4.10.
Table 17-3. General Maintenance
Task
All marks, symbols, and labels related to safety are complete?
Yes
No
Not Applicable
—
Trolley: are the trolley wheels and brakes functioning correctly and all the fixing screws tightened?
—
AC Battery Test successfully performed?
—
DC Battery Test successfully performed?
436
See the Operator's Manual Section 1.4.
—
Visually inspect the unit for damage and clean the unit if
necessary.
Power (battery) loss simulated: Ambient Mode test successfully performed?
HAMILTON-C2 Manual Reference
—
See the Service Manual Section 8.21.
See the Service Manual Section 8.21.
—
See the Service Manual Section 8.21.
English | 624165/07
Summary of Test Report
Task
Yes
No
Not Applicable
Buzzer Sound Test: Does the buzzer sounds for at least 2
minutes?
—
17
HAMILTON-C2 Manual Reference
See the Service Manual Section 8.21.
RS232 Test performed successfully?
See the Service Manual Section 8.21.
Nurse Call functional test performed successfully?
See the Service Manual Section 8.23.
CO2 accuracy in range?
See the Service Manual Section 8.22.
Visually check that the battery charge indicator on the IP
display is green.
—
—
Pre-operational checks performed in accordance with the
procedures in the Operator's Manual.
—
See the Operator's Manual Section 3.2.
Alarm tests performed in accordance with the procedures
in the Operator's Manual?
—
See the Operator's Manual Section 3.5.
Electrical safety tests performed successfully?
—
See the Service Manual Section 7.
—
See the Service Manual Section 8.24.
—
See the Service Manual Section 8.24.
Instrument report download performed and archived?
—
Event Log download performed and archived?
Table 17-4. Service Software
Task
Yes
No
Not Applicable
Ventilator Info: Technical State, Service, and RTC (Real Time
HAMILTON-C2 Manual Reference
—
See the Service Manual Section 8.7.
—
See the Service Manual Section 8.14.
—
See the Service Manual Section 8.15.
—
See the Service Manual Section 8.16.
Clock) checked and each status is OK?
Adjustments / Calibration: Touch Screen, Insp. Valve, Pres-
sure, Exp. Valve, Oxygen Sensor, Flow Sensor calibrated
and each status is OK?
Component Tests: Electronics, User I/F, Pneumatics 1, and
Pneumatics 2 checked and each status is OK?
System Tests: Pressure, Flow, O2 mixer, Tightness, Alarming
checked and each status is OK?
Table 17-5. Record of Test Equipment
Test Equipment
Type/Model
Serial Number
Electrical Safety Analyzer
__ __ __ __ __ __ __
SN __ __ __ __ __ __ __ __
Flow Analyzer
__ __ __ __ __ __ __
SN __ __ __ __ __ __ __ __
Pressure Gauge
__ __ __ __ __ __ __
SN __ __ __ __ __ __ __ __
Remarks
Table 17-6. Declaration of Tests and Calibration
Declaration of Tests and Calibration:
Date (YYYY/MM/DD):
Name:
Signature:
__ __ __ __/__ __/__ __
___________________________________
__________________________
Hamilton Medical | HAMILTON-C2 Service Manual
437
17
438
Test Report
English | 624165/07
18
Environmental Requirements
18.1
Environmental Requirements ................................................................................................................440
439
18
Environmental Requirements
18.1 Environmental Requirements
Table 18-1. Environmental Requirements
Temperature
Operating:
5°C to 40°C (41°F to 104°F)
Storage:
-20°C to 60°C (-4°F to 140°F), in original packaging
-15°C to 60°C (5°F to 140°F) otherwise
Altitude
-650m to 4000m (-2132ft. to 13,120 ft.) above sea level
Relative Humidity
Operating / Storage 10% to 95%, non-condensing
Water Protection
IP21
440
English | 624165/07
19
Document History
19.1
Document History ................................................................................................................................442
441
19
Document History
19.1 Document History
Table 19-1. Document History
00
October 2008
Release of the HAMILTON-C2 Service Manual
01
July, 2009
Release of the HAMILTON-C2 Service Manual.
Complete rework and new update.
02
November, 2009
Release of the HAMILTON-C2 Service Manual.
New update.
03
November, 2010
Updating of software versions.
04
Redesign and updated information of Service Manual.
05
Update for speaker and schematics.
06
February, 2015
New spare parts, software installation, updated pictures.
07
August, 2018
Complete rework and update.
Implementation into Schema ST4.
442
English | 624165/07
Hamilton Medical AG
Via Crusch 8, 7402 Bonaduz, Switzerland
( +41 58 610 10 20
info@hamilton-medical.com
www.hamilton-medical.com
Download