Uploaded by liziyou0326

NOVA touch User Manual 05170-1.1 Rev C

advertisement
NOVAtouch™ model 1 & 2
™
and TouchWin version 1.1+
Operating Manual
Original Instructions (English)
Part Number 05170-1.1, Rev. C
© 2015-2016
Quantachrome Instruments
1900 Corporate Drive
Boynton Beach, FL 33426
USA
NOVAtouch™ Operating Manual
CONTACT INFORMATION
Quantachrome Instruments
1900 Corporate Drive
Boynton Beach, Florida, 33426
USA
Tel: 1-561-731-4999
Fax: 1-561-732-9888
E-mail: qc.service@quantachrome.com
qc.support@quantachrome.com
www.quantachrome.com
NOVAtouch™ Operating Manual
QUANTACHROME WARRANTY POLICY
Quantachrome Instruments warrants its instruments to be free from defects in material and
workmanship for a period of one year from date of shipment under normal use and conditions.
For the period commencing with the date of shipment and ending one year later, Quantachrome
will, at its option either repair or replace any part within an instrument that is found by us to be
defective in material or workmanship, without charge to the customer, at our facility or at a
customer’s facility if the instrument purchased is backed by Quantachrome’s on-site warranty as
evidenced by the sales contract.
The customer is responsible for all transportation charges to our factory.
Damages during the warranty period resulting from unstable utilities, operator error or
unauthorized repairs will not be covered by this warranty.
Parts purchases are warranted to be free of defects for 90 days from shipping date.
The following limits apply to our warranty: Glassware, including Dewar flasks, is not covered
under this warranty except if damaged during shipment. Claims for damage during shipping must
be made in writing within 10 days of receipt of the goods.
Expendable items are warranted for 90 days for other than glassware breakage. Such items
include, but are not limited to, sample tubes, lamps, fuses, valve plungers, seals, O-rings & other
seals, hoses, flexible tubing, thermocouple vacuum tubes, filters, oils and other fluids.
Products sold by Quantachrome under their own brand name are not warranted by
Quantachrome, but our best effort will be made to secure repair or replacement if found to be
defective.
Warranty is void in the event of modifications or repairs by persons other than Quantachrome’s
service personnel, unless permission is given in writing for such repairs or modifications.
Warranty is also void in the event of exposure to corrosive atmospheres of any kind.
Prior authorization must be obtained before returning any item to Quantachrome Instruments.
Items must be decontaminated before return to Quantachrome.
Any and all computer program(s), software, firmware, code, data acquisition and/or data
reduction methods, computations, graphical and/or tabular data reporting or presentation
methods (collectively referred to as software) provided with or in or loaded into any part of the
instrument or a computer supplied as part of a Quantachrome Instruments’ instrument are
provided “as is”. Quantachrome Instruments warrants the software will perform substantially in
accordance with the accompanying materials for the period of this warranty. All software is
licensed and not sold. Quantachrome Instruments retains all rights to all software. You may not
reverse engineer, decompile, or disassemble the software. You may not distribute or copy the
software, except for the purpose of storing a back-up copy. The exclusive remedy for any
software failing to perform substantially in accordance with the accompanying materials shall be,
at Quantachrome’s option, a) repair or replacement of the software; b) return of the amount paid
for the software. You are not entitled to any damages, including but not limited to incidental,
punitive, indirect, or consequential damages including but not limited to loss of profits, business
interruption, personal injury, and other pecuniary losses.
NOVAtouch™ Operating Manual
QUANTACHROME INSTRUMENTS’ TERMS AND
CONDITIONS OF SALE
Terms: Acceptance of this order is conditional on agreement by the Buyer to all terms herein.
Buyer’s agreement is assumed unless we are notified to the contrary prior to our invoicing. None
of the provisions, terms and conditions contained in this document may be added to, modified,
superseded or otherwise changed except by a written instrument signed by the Seller’s authorized
representative, regardless of any contrary or additional provisions contained in any purchase order
or other form from the buyer.
FOB Point: The FOB Point is Boynton Beach, FL unless otherwise indicated on our Quotation or
agreed to in writing by an authorized Seller representative.
Payment terms: Terms are net 30 days from date of invoice for established accounts where Seller
has received adequate financial information to warrant extending credit to Buyer. Purchase orders
from customers not meeting with Seller’s financial approval are not accepted under credit terms.
Delivery: Delivery dates are Seller’s best estimate of the time products will be shipped from our
factory and we assume no liability for loss, damage, or consequential damages due to delays. No
liability shall result from delay in performance or non-performance of this agreement, directly or
indirectly caused by fire, explosion, accidents, flood, labor trouble or shortage, war, act of God
or arising from contingencies, happenings, or causes beyond the control of the party affected.
Material Shortages: In the event of inability for any reason to supply the total demands for the
materials specified in this order, Seller may allocate its available supply among any or all
purchasers, as well as departments and divisions of the Seller, on such basis as it may deem fair
and practical, without liability for any failure or performance which may result there from.
Taxes: Buyer shall reimburse Seller for all taxes, excises or other charges which seller may be
required to pay to any government (National, State or Local) upon the sale, production or
transportation of the commodities sold hereunder and amended from time to time.
Proprietary Information: All specifications, data, drawings, designs and software are proprietary
information which are the sole and exclusive property of the Seller and purchaser agrees to
retain any and all such proprietary information as Seller may disclose to him in confidence and not
disclose it to other parties or use it except in inspection and evaluation purposes in connection
with a contract with Seller.
Software License: Title to all software provided as separate modules or embedded in the Products
("Software") shall remain the property of Quantachrome or Quantachrome's licensors.
Quantachrome grants to Buyer a non-exclusive, limited license to use the Software together with
the Products. Buyer shall not decompile, disassemble or otherwise reverse engineer the Software,
and may create derivative works only to the extent permitted by Quantachrome. Buyer shall not
sublicense, assign, copy, distribute, or disclose any portion of the Software to a third party without
the express written consent of Quantachrome. Buyer may transfer or sell its license rights to
use the Software only together with the Products to a transferee, which has accepted this provision
of Quantachrome’s Terms and Conditions of Sale in writing.
Changes: Orders accepted by the Seller cannot be changed or modified in whole or part, except
with the written consent of Seller.
NOVAtouch™ Operating Manual
Return Policy: Prior approval must be obtained from an authorized Seller representative to
return any Product. Seller will assign a return authorization number and record the reason for
return. Seller will examine returned part to determine the actual cause, if any, leading to Buyer’s
return. If Product has a manufacturing defect Seller will, at its discretion, repair or replace with
like product.
Restocking charges: Product purchased, under the sole discretion of the Seller, may be returned
for a 20% restocking fee and must be returned within 90 days of date of delivery. Product cannot
be returned if used or modified.
Claims: Claims for shortages or damage must be reported within 10 business days after receipt
of shipment. All claims for loss or damage in transit must be made against the carrier.
Warranty: Quantachrome Instruments warrants all instruments that it manufactures for a period
of twelve months from the date of delivery. This warranty includes all parts and labor.
Quantachrome does not warrant any product against damage from corrosion, contamination,
misapplication, improper specification, or wear and tear and operational conditions beyond the
Seller’s control. This warranty excludes all glassware and expendable items associated with each
instrument. Repairs made during the warranty period are guaranteed until the end of the warranty
period or 90 days, whichever is greater.
NOVAtouch™ Operating Manual
TABLE OF CONTENTS
Contents
1 INTRODUCTION ................................................................................................................................ 8
Summary of Features ............................................................................................................... 9
2 TECHNICAL SPECIFICATIONS ....................................................................................................10
Electrical ................................................................................................................................ 10
Physical ................................................................................................................................. 10
Environmental ....................................................................................................................... 10
Model Types .......................................................................................................................... 10
3 SAFETY .................................................................................................................................................11
Symbols Used in this Manual ................................................................................................ 11
Symbols Used on the Instrument ........................................................................................... 11
Safety Instructions — General .............................................................................................. 12
Safety Instructions – Cryogenic Liquids ............................................................................... 13
Safety Instructions — Heating Mantles................................................................................. 13
4 INSTALLATION .................................................................................................................................15
Unpacking and Placement ..................................................................................................... 15
Operation Voltage .................................................................................................................. 15
Analysis and Degassing Stations ........................................................................................... 15
Vacuum Connections ............................................................................................................. 16
Gas Inputs .............................................................................................................................. 16
5 INSTRUMENT COMPONENTS ....................................................................................................18
Mains Input............................................................................................................................ 18
Power Switch ......................................................................................................................... 19
Vacuum Fitting....................................................................................................................... 19
Gas Inputs .............................................................................................................................. 19
Cryogen Dewar Flask ............................................................................................................ 19
Heating Mantles and Outlets ................................................................................................. 20
Sample Cells and Glass Filler Rods ...................................................................................... 21
Cell Bulkhead Fittings ........................................................................................................... 22
Coolant Level Sensor (CLS) ................................................................................................. 24
P0 Cell .................................................................................................................................. 24
Quartz Rod Calibration Kit.................................................................................................. 24
Doors ................................................................................................................................... 25
Status Indicators .................................................................................................................. 25
6 TouchWin™ SOFTWARE ...................................................................................................................26
Installation ............................................................................................................................. 26
File ......................................................................................................................................... 30
Configure ............................................................................................................................... 34
Instrument Name (NOVAtouch) ........................................................................................... 46
Window ................................................................................................................................. 64
Help ....................................................................................................................................... 68
TouchWin™ Icon Toolbar ...................................................................................................... 69
Configuration of TouchWin™-CFR ....................................................................................... 70
Multiple Instrument Control .................................................................................................. 75
7 TOUCH SCREEN ...............................................................................................................................76
Status Bar .............................................................................................................................. 76
Page 6 of 161
NOVAtouch™ Operating Manual
TABLE OF CONTENTS
Home ..................................................................................................................................... 76
Manual Mode page ................................................................................................................ 77
Analysis Setup page .............................................................................................................. 78
Cell Calibration page ............................................................................................................. 81
Progress page ......................................................................................................................... 84
Degassing Page...................................................................................................................... 85
Results page ........................................................................................................................... 85
Maintenance page .................................................................................................................. 88
Logfile ................................................................................................................................. 95
Power Mode......................................................................................................................... 96
8 INSTRUMENT OPERATION .........................................................................................................97
Introduction ........................................................................................................................... 97
Cell Calibration ..................................................................................................................... 97
Sample Preparation................................................................................................................ 97
Unloading the Degasser ....................................................................................................... 106
Start Sample Analysis.......................................................................................................... 106
9 DATA ANALYSIS USING TouchWin™ ....................................................................................... 108
Opening Data Files .............................................................................................................. 108
Floating Menu ..................................................................................................................... 110
Tables, Graphs and Reports ................................................................................................. 110
Editing Analysis Data Information ...................................................................................... 111
Setting Data Reduction Tags................................................................................................ 111
Setting Data Reduction Parameters. .................................................................................... 113
Calculation of Heats of Adsorption ..................................................................................... 114
Review of the Data Analysis Methods ................................................................................ 114
Data Presentation ................................................................................................................. 119
10 THEORY AND DISCUSSION ..................................................................................................... 126
Surface Area ...................................................................................................................... 126
Pore Size by Gas Adsorption ............................................................................................. 130
Micropore Analysis ........................................................................................................... 137
Density Functional Theory and Monte Carlo Simulation Methods .................................. 146
Fractal Dimension Methods .............................................................................................. 149
Thermal Transpiration ....................................................................................................... 151
Heats of Adsorption .......................................................................................................... 152
11 APPENDIX ........................................................................................................................................ 154
Troubleshooting Guide ...................................................................................................... 154
Maintenance ...................................................................................................................... 156
Spare Parts List .................................................................................................................. 157
12 REFERENCES .................................................................................................................................. 160
Page 7 of 161
NOVAtouch™ Operating Manual
INTRODUCTION
1 INTRODUCTION
The NOVAtouch™ is a high-throughput surface area and pore-size analyzer. Employing
manometric measurements the NOVAtouch™ determines quantitative physisorption data (amount
of gas adsorbed/desorbed as a function of pressure) which are then used to calculate sample
characteristics such as surface area and pore size distribution. To make the measurements, a
suitable gas (referred to as the adsorbate) is quantitatively dosed, added to or removed from a
sample cell containing the solid sample (adsorbent) maintained at a constant temperature below the
critical temperature of the adsorbate – from a calibrated manifold of known volume and actively
measured temperature. As adsorption or desorption occurs the pressure in the sample cell changes
until equilibrium is established; the quantity of gas adsorbed or desorbed at the equilibrium
pressure being the difference between the amount of gas admitted or removed and the amount
required to fill the void space (that volume in the sample cell around the adsorbent and in the
connecting gas pathways to a valve isolating the cell from the manifold).
The NOVAtouch™ incorporates multiple features and enhancements for improved performance and
ease-of-use. For example, it has three adsorbate gas inputs that allow a variety of gases such as N2,
Ar and CO2 to remain connected and from which the instrument automatically selects according to
the pre-programmed protocol. Helium, used to determine the void space when operating in the socalled “classical” mode has its own dedicated input, as does the user’s gas-of-choice for flowdegassing samples and/or backfilling evacuated sample cells.
The instrument also includes everything required for sample degassing or preparation, separated
from the analysis compartment yet within the overall footprint; no separate degassing unit is
required. This capability uses either flowing gas or vacuum (LX model only, see below) together
with the provided heating mantles and a programmable temperature profile that allows for extended
and controlled sample preparation without the need for user intervention.
The NOVAtouch™ is available in LX and non-LX versions. The LX features additional dedicated
pressure transducers, one for saturation pressure (P0) and another that controls evacuation and
backfill of samples in the preparation compartment. The LX version therefore is capable of both
flow-type and vacuum degassing, whereas the non-LX version features just flow degassing. Both
types use a single, external vacuum pump (required for analysis even when flow degassing has
been used).
All NOVAtouch™ models have an interactive, full color, multi-lingual touch-screen display that
allows the user to enter sample information and initiate an analysis from a directory of user-defined
analysis protocols, follow analysis progress in real time via text messages and automatically
updated data plots, review results, check instrument status, access manual mode, start the degasser,
and much more. After the initial transfer of analysis protocols and degassing temperature profiles
from Quantachrome’s powerful multi-lingual TouchWin™ software, the NOVAtouch™ does not
require direct interaction with a PC. However, TouchWin™ not only provides real-time, remote
data acquisition and archiving, but also comprehensive off-line data reduction via an extensive
suite of calculations, graphing and reporting options. TouchWin™ runs under 64-bit versions of
Windows and communicates via Ethernet that allows for convenient placement of PC relative to
the NOVAtouch™, even across a local network remote from the unit itself.
Page 8 of 161
NOVAtouch™ Operating Manual
INTRODUCTION
Summary of Features

Fully automatic analysis of one to four samples (depending on model type, see below)
simultaneously.

Long-life 2L Dewar allows measurement of detailed isotherms for the determination of pore
size distributions.

The smaller (blue) 1L Dewar is suitable for quality control/assurance applications where a
large capacity Dewar is unnecessary, i.e., in the absence of a requirement for full isotherm
measurement (pore size distributions).

In-board sample preparation of up to four samples simultaneously.

Intuitive touchscreen used for the following operations:
• Starting/stopping degassing
• Starting analyses locally from a shortlist of protocols
• Monitoring status
• Aborting analysis
• Reviewing results
• Manual mode functions
• Calibration functions

Powerful Windows-based TouchWin™ software allows:
• Creation of multi-step degassing temperature profiles (sent to instrument)
• Creation of detailed analysis protocols (sent to instrument)
• Remote starting of analysis
• Real-time data display
• Pore size distribution calculations
• Automatic printed report generation
• Archiving of data

Comprehensive accessory and consumable kit included.
Page 9 of 161
NOVAtouch™ Operating Manual
TECHNICAL SPECIFICATIONS
2 TECHNICAL SPECIFICATIONS
Electrical
NOVAtouch™
2.1.1
Voltage:
Frequency:
Power (max):
Connection:
2.1.1.1
100, 120, 220, 240 V (see serial number plate on the side of the instrument)
50/60 Hz
350 W
Grounded, single-phase mains outlet
Heating Mantles
Voltage:
Frequency:
Power (±10%):
Connection:
110 – 120V
50/60 Hz
105 W each
Dedicated outlets in NOVAtouch™ degassing compartment
Physical
Height:
Width:
Depth:
Weight:
82.9 cm (32.7 in)
61.6 cm (24.3 in)
49.2 cm (19.4 in)
42.2 kg (93 lbs.)
Environmental
Temperature:
Max. Relative Humidity:
15 C – 40 C
80 %. It is recommended wherever possible to operate the unit in an air-
conditioned space at RH<50%.
Model Types
2.4.1
LX
Available with one up to four analysis ports.
Dedicated P0 cell
Dedicated P0 transducer
Vacuum degassing
Degassing backfill transducer
Flow degassing
2.4.2
Standard (Non-LX)
Available with one up to four analysis ports.
Dedicated P0 cell
Flow degassing
Page 10 of 161
NOVAtouch™ Operating Manual
SAFETY
3 SAFETY
Symbols Used in this Manual
HOT! This symbol denotes a possible hazard to the operator due to high
temperatures.
!
CAUTION! This symbol denotes a hazard that could result in damage to the
instrument.
!
WARNING! This symbol denotes a hazard that could result in injury to the operator.

WARNING! This symbol denotes a low-temperature hazard that could result in
injury to the operator.

NOTE! This symbol denotes an important detail.

TOOLS REQUIRED: This signifies that tools are required for the described action.

HAZARD: This denotes that an environmental hazard might be present.
Symbols Used on the Instrument
!
CAUTION! Use high purity gases. Be sure to connect gas lines to their appropriate
input fitting; incorrect gas supplies will lead to erroneous results.
HOT! Use caution at all times when handling the heating mantles.
Page 11 of 161
NOVAtouch™ Operating Manual
SAFETY
Safety Instructions — General
Read and follow all the instructions in this manual.

This instrument was designed for indoor laboratory use only. It is not designed to be
operated in a flammable atmosphere or where dust explosion hazards may be present.

Position the instrument for easy access to the on/off switch.

This instrument shall only be operated by a competent user who has received training
appropriate for the intended purpose. Management responsible for the use of this
instrument shall ensure that the user is familiar with the contents of this manual and any
necessary safety precautions before working with the instrument.

Comply with all national health and safety standards. This may include, but it is not
limited to: barrier protection, protective eyewear, gloves, and suitable laboratory attire
when operating or maintaining this or any other automated laboratory analyzer.

Inform user of hazards associated with the gas(es) used and the sample(s) being tested.

Operate this instrument only at the voltage specified on the serial number plate on the
side of the instrument.

Only use the mains power cord supplied with the unit or one that is rated for the voltage
and power stated in the specifications section of this manual.

This instrument must be disconnected from the mains for any cleaning, maintenance or
service. Do not position the instrument in such a way that it would be difficult to remove
the power cable.

Do not make any unauthorized modifications to this instrument. Use replacement fuse(s)
only as indicated in the service section of this manual or on the instrument. The fuse(s)
are not intended to be replaced by the operator but by a qualified electrical technician.

When attaching a plug to the international power cord, be sure to follow the color code:
brown = live, blue = neutral, green/yellow = earth ground.

When not in use for more than a few days, unplug the machine from the mains power
supply. If it is not going to be used for a considerable amount of time it should still be
protected from possible damage and water splash. The original packaging material is
suitable as protection for long term storage. The instrument should always be stored and
transported the “right way up”, i.e. the same orientation in which it is used.

HAZARD: Observe federal and local regulations regarding handling and
disposal of used samples.
Page 12 of 161
NOVAtouch™ Operating Manual
SAFETY
Safety Instructions – Cryogenic Liquids
!

WARNING! When filling the Dewar flask with liquid nitrogen (LN2) care must be
taken to prevent it from getting between the glass insert and the outer cover, as this
may cause the glass to implode.
WARNING! This symbol denotes a low-temperature hazard that could result in
injury to the operator.

Because the Dewar flask can shatter unexpectedly, a protective face shield, safety
glasses, and gloves should be worn when filling the flask.

When using a gas other than N2 or Ar at its respective boiling point, do not use the
Calculate P0 option while calibrating an empty cell or during analysis.
Safety Instructions — Heating Mantles
HOT! The outer surfaces of the heating mantle may become hot during use. Do not
hold hot heating mantles without wearing protective gloves. Never insert fingers
inside the pocket to determine if the mantle is heating up. Do not place a hot heating
mantle on a surface which is not heat resistant. Switch off when not in use.

!
Quantachrome heating mantles are designed only for heating sample cells for the
purpose of degassing. Quantachrome heating mantles must only be used on
Quantachrome instruments designed for the mantle type. Use heating mantles only on
instruments with properly functioning, calibrated heating mantle stations.
WARNING! The heating mantle must have a sample cell in the pocket when in
use. Ensure that no sample adheres to the outside of the sample cell. Do not use
wet, broken, cracked, or chipped sample cells. Do not insert spatulas, screwdrivers
or other objects into a heating mantle. Never place the sample to be degassed
directly into the heating mantle.

Do not allow liquids to come into contact with the heating mantle and do not handle
heating mantles with wet hands. Do not allow dust to accumulate on, nor in, a heating
mantle. Do not expose the heating mantle to a corrosive atmosphere of any kind.

Do not make any modifications to any Quantachrome heating mantle. Do not remove
the serial number tag from the heating mantle. Removing the tag will void the warranty.
Page 13 of 161
NOVAtouch™ Operating Manual
SAFETY

Insert the power plug of a heating mantle only into the socket provided for that purpose
on the NOVAtouch™. Do not insert the power plug of a heating mantle into any mains
supply socket. Do not insert the plug of any other device into the heating mantle socket
on the NOVAtouch™.

Insert the thermocouple plug of a heating mantle only into the socket provided for that
purpose on a Quantachrome instrument. Do not insert the thermocouple plug of a
heating mantle into any other socket. Do not insert the plug of any other device into the
thermocouple socket of any Quantachrome instrument.
!
!
WARNING! Always ensure that the power plug and thermocouple plug of one
heating mantle are inserted into the sockets of the same heating mantle station.
Never insert the power plug of one mantle into the power socket of a different
mantle. Use care and ensure that the thermocouples are installed on the same
station that is providing the power to the mantle.
CAUTION! Do not cross-wire the heating mantles. This will cause the mantles to
overheat.
Page 14 of 161
NOVAtouch™ Operating Manual
INSTALLATION
4 INSTALLATION
Unpacking and Placement
For best results the instrument should be installed on a level bench, suitable for supporting in
excess of 50kg, away from direct sunlight, heaters and air conditioner units. Local regulations
may require use of local ventilation and/or protective equipment. Position the instrument so that
the on/off switch can be easily accessed.
Remove the instrument from its shipping box and place it on the bench where it will be operated.
!
WARNING! Take precaution when lifting the unit since it weighs 42.2 kg (93
lbs.). Handle the instrument with two or more persons trained in proper lifting
techniques and/or a mechanical lift device.
Unpack the supplied parts and accessories and confirm them to the packing list. If any parts are
missing or damaged immediately notify Quantachrome or your local authorized representative.
4.1.1
Earthquake Zones
The NOVAtouch™ can be secured to the bench to prevent toppling in the case of a seismic event.
Contact Quantachrome for details.
Operation Voltage
The NOVAtouch™ is shipped for operation with the voltage specified in the purchase order.
However, it is designed for operation on 100V, 120V, 220V or 240V, 50/60 Hz.
!
CAUTION! The unit must not be operated at a voltage other than that indicated
on the serial number plate; the Corcom power input module must match the serial
number plate. If the unit is relocated and must be used at a different voltage than
that specified, the user must contact Quantachrome Service Department for
detailed instructions on changing the voltage settings.
Analysis and Degassing Stations
The sample cell ports in both the analysis and degassing stations must have the stainless steel
dowel inserted into the bulkhead fittings before turning on the NOVAtouch™. The stainless steel
dowels use the same O-rings and cell adapters as the 9mm cells.
Page 15 of 161
NOVAtouch™ Operating Manual
INSTALLATION
Vacuum Connections

TOOLS REQUIRED: Flat head screwdriver or 5/16’’ wrench (not supplied).
The NOVAtouch™ is supplied with a centering ring and clamp to connect the vacuum hose to the
fitting on the instrument. If the vacuum pump is supplied with the NOVAtouch™, a similar set of
hardware is supplied to connect the hose to the pump. A pump capable of reaching 10 milliTorr
should be used.
Vacuum
Connection (on
ride side of
instrument)
Ring
Clamps
90° Elbow
Centering Rings
NW-16 Stub
Hose Clamp
Vacuum Hose
Nova Touch is supplied with a 90 degree elbow (P/N 44242) to aid in
NOTE!
assembly of vacuum connections.
Page 16 of 161
NOVAtouch™ Operating Manual
INSTALLATION
Gas Inputs

TOOLS REQUIRED: 1 1/8” wrench, 7/16” wrench and 3/8” wrench (not supplied)
The NOVAtouch™ has five gas input fittings located on the right side of the instrument: three gas
input fittings reserved for adsorbates (analysis gases), one is reserved for helium and another for
degas backfill/flow gas. Helium gas must be connected to its labeled port for those analyses
programmed to use Helium Mode. It is recommended to use high purity helium and adsorbate
gases (99.99% or higher).
Attach a dual stage, stainless steel diaphragm regulator to the tank using a 1 1/8” wrench.
Connect one end of the gas input line (supplied) to the regulator with a 7/16” wrench, and the
other end to the gas-input fitting on the unit with a 3/8” wrench. It may be necessary to obtain an
adapter in order to connect the input line to the regulator. The recommended regulator assembly
is available from Quantachrome (P/N 01207, 0-25 PSI), comes complete with a shut-off valve,
CGA580 cylinder fitting and 1/8” outlet fitting. The regulator should be set to deliver 8 psig (55
kPa) for all gases.
For Helium: connect the gas input line to the regulator. Set the regulator to 8 psig and allow the
gas to flow through the line to purge out atmospheric gases. Then connect the gas input line to
the Helium gas-input fitting.
If the adsorbate is used as the backfill/flow gas, use a tee which allows the gas to be supplied to
the two fittings (adsorbate input 1 and the degas flow/backfill input) using one tank. A line with
a compression fitting on one end is provided with the instrument.
Figure 4-1, Backfill Tee
The regulator assembly with P/N: 01207 is suitable for inert gases such as
NOTE!
He, Ar, and N . The relevant P/N for CO is: 01207-CO2.
2
2
Page 17 of 161
NOVAtouch™ Operating Manual
INSTRUMENT COMPONENTS
5 INSTRUMENT COMPONENTS
Analysis
bulkhead
fittings
Touch screen
Degassing
bulkhead
fittings
CLS
Heating mantle hooks
(LX model only)
Dewar flask
Heating
mantles
Ethernet port
Power on/off
Mains voltage
Mains power
input
Figure 5-1, Instrument Components
Mains Input
The voltage indicated on the power entry module must match the voltage on the serial number
plate. If it does not, STOP, and contact Quantachrome or its local representative. The power
module accepts a standard IEC-320 detachable instrument power cord.
Page 18 of 161
NOVAtouch™ Operating Manual
INSTRUMENT COMPONENTS
Power Switch
The power switch is located on the left side panel of the NOVAtouch™. The position of the
switch indicates the status, do not rely on the light inside the switch to indicate whether the unit
is switched on.
Vacuum Fitting
The vacuum uses a standard KF-16 flange fitting; located on the right side of the NOVAtouch™.
The connecting hose, clamp, flange and centering ring are supplied with the instrument. A
vacuum pump capable of achieving at least 50 milliTorr is required, a vacuum pump capable of
10 milliTorr is recommended.
Figure 5-2, Vacuum Fitting
Gas Inputs
The NOVAtouch™ has five gas input fittings on the right side of the NOVAtouch™: three
adsorbate ports, one helium port, and one backfill port. A backfill line is provided to allow an
easy connection between the first adsorbate gas input and the backfill gas input. See Figure
4-1, Backfill Tee.
Cryogen Dewar Flask
The NOVAtouch™ is supplied with a two liter Dewar (P/N 74215), which will hold liquid
nitrogen (LN2) for more than 30 hours (capable of full mesopore size distribution measurements),
or a one liter Dewar (P/N 74218) used for short cells and is suitable for applications requiring
BET surface area/STSA measurements. The Dewar is placed in the lift-drive cup.
NOTE! The Dewar manufacturer recommends that the following procedure be
adopted:
(i)
Prior to first use, all the packing material from the inside of the Dewar must
be removed.
(ii) Wash the Dewar with hot soapy water, rinse with distilled or deionized
water and either air dry or clean with a lint free towel. This is to ensure a clean,
dry, and contamination-free Dewar.
Page 19 of 161
NOVAtouch™ Operating Manual
INSTRUMENT COMPONENTS
The Dewar flask holds the coolant used for the analysis, usually liquid nitrogen; it should be
filled to the level indicated by the Dewar clip prior to each analysis.
Figure 5-3, Dewar Flask with Coolant Level Indicator (Dewar clip) Attached
!
CAUTION! Always use short sample cells, short Po cell and short CLS with the 1L
(blue) Dewar; otherwise breakage of cells and/or Dewar is likely.
Heating Mantles and Outlets
Two heating mantle power outlets are provided in the degassing compartment. The thermocouple
plugs are inserted into the jacks located in the same compartment. The heating mantles will
not operate correctly unless both the power cord and thermocouple are plugged in.
NOTE! Always ensure that the power plug and thermocouple plug of one heating
mantle are inserted into the sockets of the same heating mantle station. Do not
insert the power plug of a mantle into the power socket of one heating mantle
station and the thermocouple plug of the same mantle into the thermocouple socket of a
different heating mantle station. For safety instructions regarding the heating mantle
see Sect. 3.5.

Page 20 of 161
NOVAtouch™ Operating Manual
INSTRUMENT COMPONENTS
Figure 5-4, Heating Mantles Positioned for Flow Degassing
Figure 5-5, Heating Mantles Ready for Vacuum Degassing (LX model only)

NOTE! The mantle hooks used to support the mantles while vacuum degassing
sample cells are provided only in the LX model.
Sample Cells and Glass Filler Rods
Sample cells are available with outside stem diameters of 6, 9 and 12 mm (internal
diameters of 4, 7, and 10 mm respectively), in “long” and “short” versions. Each cell should be
numbered in the space provided (if analyzing in NOVA mode) and used with the appropriate
glass filler rod. After performing a cell calibration, ensure that the same filler rod is always
used for that particular station. For most users, standard (bulbless) cells of a given stem
diameter can be given a single identifying number and used interchangeably.
Page 21 of 161
NOVAtouch™ Operating Manual
!
INSTRUMENT COMPONENTS
CAUTION! Never use long cell (sample and P0) with the 1L (blue) Dewar;
otherwise breakage of cells and/or Dewar is likely.
Cell Bulkhead Fittings
The fittings, designed to accept 6, 9, and 12 mm size cell stems, use an adapter sleeve and O-ring
to secure the cells in position while creating a vacuum-tight seal. The correct size O-ring and
adapter must be used with each corresponding cell. Be sure to use only one O-ring when
installing a cell into the fitting; using two O-rings (of any size) are likely to cause a leak leading
to erroneous results.
Cell stem is inserted into
the bulkhead fitting
The correct sized O-ring for
the cell stem must be used
The correct sized adapter for
the cell stem must be used
Knurled ring for
tightening/loosening the
O-ring compression fitting
Figure 5-6, Cell Fittings (not to scale)
Page 22 of 161
NOVAtouch™ Operating Manual
INSTRUMENT COMPONENTS
Bulkhead positions
.
Figure 5-7, Orientation of Sample Bulkheads
Figure 5-8, Removable Decal
The removable decal can be placed anywhere the user deems appropriate. Note the alignment of
the sample cell station number relative to the front of the instrument.
Page 23 of 161
NOVAtouch™ Operating Manual
INSTRUMENT COMPONENTS
Coolant Level Sensor (CLS)
Coolant level is maintained at a constant level around the cell, and at a position which minimizes
the so-called “cold-zone” of the cell (thereby improving sensitivity). That level is maintained by
a coolant level sensor (P/N 00080-CLS) and a sensitive electronic circuit which adjusts the
Dewar height to compensate for the evaporation of coolant during the analysis. The CLS is
installed next to the analysis cell bulkhead fittings using a BNC connector. The NOVAtouch™ is
supplied with a CLS suitable for both liquid nitrogen and liquid argon. A water float sensor (P/N
00080-H2O-NT) is available for measuring isotherms, for example, at 0oC using ice/water.
Figure 5-9, Coolant Level Sensor (CLS)
The CLS (P/N 00080-CLS) must be used; otherwise the sample will never be
NOTE!
cooled. When using short cells (sample and P ) the short CLS (P/N 00080-CLS-S)
0
must be used.
P0 Cell
The P0 cell is dedicated to measuring the saturation vapor pressure, P0, of a gas at its own boiling
point. When measuring P0 is part of an analysis protocol, the instrument liquefies a small amount
of pure adsorbate in the P0 cell when the Dewar is in the raised position. The equilibrium P0 is
measured either by a dedicated pressure transducer (LX models) or the main manifold transducer
(non-LX models). If used, the instrument safely removes the condensed gas via the vacuum
system before the Dewar is lowered.
To help stabilize saturation pressure (P0) readings, the cylindrical collar (P/N 04000-10844)
provided should be slipped over the stem of the P0 cell before it is installed. The P0 collar should
rest above the bulb.
A P cell is not required if, enter P or calculate P is chosen at the time of the
NOTE!
analysis.
0
!
0
0
CAUTION! Never use a long P0 cell with the 1L (blue) Dewar; breakage of cells
and/or Dewar is likely. Use a short P0 cell (P/N 74027-S).
Quartz Rod Calibration Kit
Manifold volume calibration is performed using the calibration kit which includes:
• Calibration cell with spring
• Calibrated quartz rod
Page 24 of 161
NOVAtouch™ Operating Manual
•
INSTRUMENT COMPONENTS
Quantachrome certificate for quartz rod volume
Doors
The NOVAtouch™ rotating doors serve to eliminate drafts (particularly in the analysis
compartment), and minimize frost build up around the Dewar. It is recommended that the doors
be kept closed whenever possible. The roller-bearings are factory-lubricated and should not
require regular maintenance.
Status Indicators
The status indicators display the present state of the heating mantles and Dewar. At the top of the
status indicators is the “Mantles Reset” button used to reset the mantles when they have reached
an over-temperature condition. The “Enabled” light will always be on when the mantles are
connected and functioning properly. The “Heat Cmd” indicates that the mantle is receiving the
command and currently heating. The indicators available for the Dewar are: Up Cmd, Send,
Down Cmd, and Received. The “Cmd” indicators will blink when the command to move the
Dewar in the respective direction is being received.
NOTE! If the heating mantles have reached an over-temperature condition they will be
placed into a shutdown state, which can only be released by resetting the mantles. The
mantles can be reset using the “Mantles Reset” button at the top of the status indicators.

Figure 5-10, Status Indicators
Page 25 of 161
NOVAtouch™ Operating Manual
SOFTWARE
6 TouchWin™ SOFTWARE
Installation
6.1.1
TouchWin™ Program CD
The installation CD contains the TouchWin™ software. It can be installed on any Windowsbased PC (Windows® Vista and higher). Ideally, before proceeding with the installation, a
backup copy of the installation CD should be made and stored in a safe location (preferably not
the same place as the original CD). TouchWin™ will need to be installed on at least one PC
before it can properly operate the NOVAtouch™ instrument.
NOTE! If upgrading the software and FULL installation is selected, it will
overwrite the configuration files regardless of date/time stamp (materials data,
reports, instrument configuration) with “factory” defaults.

6.1.2
TouchWin™ Setup
Follow the installation steps below:
1. Insert TouchWin™ installation CD in the CD-ROM drive and follow the installer
prompts.

If auto insert notification is enabled, the startup menu will launch automatically.

If not, open the CD-ROM drive in explorer, and manually run START.EXE.
Figure 6-1, TouchWin™ Setup Screen
2. Click “Next”.
Page 26 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-2, TouchWin™ “Select Additional Tasks” Screen
To choose a desktop icon or Quick Launch icon select these items. Click “Next”.
Review the items selected on the “Ready to Install” page before choosing “Install”. To change
any of the settings use the “Back” button and make the appropriate changes.
Figure 6-3, TouchWin™ “Ready to Install” Screen
After “Install” is selected, the window shown below will open showing the progress of the
software installation.
Page 27 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-4, TouchWin™ “Installing” Screen
When the software installation is completed, to launch the software automatically select the
“Launch Quantachrome TouchWin” and click “Finish”. If you do not wish to launch the
software at this time ignore the selection and click “Finish”.
Figure 6-5, Completed Installation Screen
After installation is complete TouchWin™ will be ready to run. The software may be opened
from the Quick Launch or Tool Bar (based on previous selections).
Page 28 of 161
NOVAtouch™ Operating Manual
6.1.3
SOFTWARE
Instrument – PC Connection
To begin using TouchWin™, perform the following steps:
1. Connect the NOVAtouch™ to a computer or network using an Ethernet cable
(P/N 26102-25).
2. Power on the NOVAtouch™. Allow the instrument to run through the initialization
sequence. When this sequence is finished, the Home Screen will be available on the
NOVAtouch™ display. The instrument is now available for use.
3. Navigate to the Maintenance Page on the Touchscreen and select the Network option.
(See Maintenance Page details in Sect. 7.9.5, Network Settings.)
4. Enter the desired IP address. Contact your IT department for help with assigning an IP
address.
5. Using TouchWin™, click on the Instrument name drop down menu; the default name is
“NOVAtouch” and select “Communication Parameters” to open the “Edit Profile”
window. Input the designated IP address (the same address that was used in Step 3) and
click “OK”. (See also 6.4.3, Communication Parameters.)
Figure 6-6, TouchWin™ “Setup Completion” Screen
become the new title of the “Instrument/NOVAtouch” drop down menu.
NOTE! The instrument name can be set using the “Edit Profile” window. This will
6. While still using TouchWin™, return to the Instrument drop down menu and click
“Connect” to establish communication with the NOVAtouch™ instrument.
NOTE! The TouchWin™ interface is required to setup and upload both degassing
temperature profiles and analysis protocols. These procedures can be started from
the touch screen repeatedly thereafter.

6.1.4
Working with TouchWin™
The following Sections: 6.2 – 6.7 , address the TouchWin™ software features systematically.
What follows is intended to be used as a reference.
The TouchWin™ program uses a standard Windows-style interface consisting of menus, buttons,
Page 29 of 161
NOVAtouch™ Operating Manual
SOFTWARE
and various dialog windows. The main menu bar (shown in the center of Figure 6-7) contains
five drop-down menus providing access to all program functions as described below.
Figure 6-7, TouchWin™ Drop Down Menus
File
The “File” drop down menu provides the user with access to data files. This menu also provides
“Change User” and “Exit” (close program) functions.
6.2.1
New > Physisorption Overlay
Select “New > Physisorption Overlay” to create an overlay file (.qcuPhOverlay). First, select the
primary file from the file explorer window. To add additional files in the overlay, right click and
select “Manage Isotherms”.
Page 30 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-8, Physisorption Overlay Isotherm
When the overlay manager opens, select “Add” to add files. Duplicating a selection is desirable
to modify only one instance of the graph and not the other. To remove a file, select the file’s
field then select “Remove”.
Figure 6-9, Physisorption Overlay Manager
Once multiple files have been chosen, to edit their presentation select the file’s field in the
Overlay Manager, then select “Edit”. This will prompt a contextual Floating Menu (shown
below) for the file whose field is selected.
Figure 6-10, Edit Function Floating Menu

NOTE! To use the overlay file in the future, save the file that has been created by
using either the “Save” or “Save As” functions (from the “File” drop down menu).
Page 31 of 161
NOVAtouch™ Operating Manual
6.2.2
SOFTWARE
Open
When the “Open” option is selected from the “File” drop down menu, the explorer window will
open for selection. There are two primary file types: .qcuPhysIso (for physisorption files) and
.qcuPhOverlay (for overlay files). Because the two primary file types are independent, not all of
the data will be displayed in the initial explorer window. To access the data, select the desired
file type from the “File Type” drop down menu.
Figure 6-11, Open File Explorer
6.2.3
Reopen
Hovering the cursor over the “Reopen” option opens a side menu containing a list of recently
opened documents; open any of the listed documents directly by clicking in the list. The list of
recently opened files can be deleted by clicking “Clear Recent List”.

NOTE! “Clear Recent List” is an action that cannot be undone.
Figure 6-12, Reopen a Previously Open File

NOTE! If there are no recent files because the list was cleared or the software is new,
the side menu containing recent files will not appear.
Page 32 of 161
NOVAtouch™ Operating Manual
6.2.4
SOFTWARE
Import
The import function imports files for data analysis. The software can upload .qps, .raw, .drf, and
.dat files. To change the type of Quantachrome physisorption file select the drop down list for
file of type and choose the type of file to import.
Figure 6-13
6.2.5
Save
The “Save” menu item in the “File” drop down list allows the user to save an open document to
its current file name. The menu item is only be available for selection when a change has been
made to the document.
6.2.6
Save As
The “Save As” menu item allows the user to change the location of a file and the name of a file,
but not the file type.
6.2.7
Print
The “Print” menu item launches “Print” window from which the selected graph, table, or report
file can be printed. This function does not print all open views.
6.2.8
Print Preview
The “Print Preview” function displays what will print before executing the process. The preview
window includes a zoom function and multi-page display.
Figure 6-14, Print Preview
Page 33 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Change User
6.2.9
Clicking on “Change User” opens the “Login” window. Change the User ID by simply editing
the field; the name inside the input field will be used as the User ID throughout the software
(including printed reports) until the field is updated. Select the “Autologin with this ID” box to
set the software to remember the User ID and automatically use it the next time TouchWin™ is
started.
Figure 6-15, Change User
6.2.10
Exit
This closes the program. If there are any edited files still open, and that have not been saved, the
user will be prompted to save the necessary files with the changes (click “Yes”), or not (Click
“No”… any changes will be lost), before the program is closed. To save the file under a different
name before closing the program, click “Cancel” and use the “Save As” function first.
Configure
The “Configure” drop down menu allows the user to customize most of the software’s
appearance, basic parameters, and other settings as detailed below.
6.3.1
Edit Folders
The software allows a different predefined default folder settings for each file type it recognizes.
Use this configuration window to specify the default location of the selected file type:
Figure 6-16
Page 34 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Selecting “Set Relative” generates this pop-up:
Figure 6-17, “Set Relative” Window
Selecting “Set Path” launches this window:
Figure 6-18, “Set Path” Window
Manage Reports
6.3.2

NOTE! A report is a defined collection of tables and graphs intended for printing.
The “Manage Reports” function will launch the “Manage Report Templates” window allowing
the user to create, import, or modify a report setting. The report settings created using this
function will appear as designed in the results, if selected.

Add

Edit

Delete

Export

Import
Page 35 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-19, Manage Report Templates
To create a report style using the “Manage Report Templates” window, select the desired report
style to edit. Once a field is selected, the “Edit” button will become active.
Access the “Edit” function by selecting the desired report to edit and click on the “Edit” button,
or by double-clicking the report name. After the file has been edited an asterisk will appear in the
“Mod” column to indicate that the file has been altered.
To create a new report, use the “Add” button. Enter a new, unique name for the report. The
“Report Template Editor” will open to permit changes to the report’s contents/structure.
Delete a report by selecting its name from the list, then click the “Delete” button.
Use the “Export” and “Import” buttons to save or load an individual report template as a separate
file that can be used to share report templates between TouchWin™ installations.
Click “OK” to save the changes in this session, or “Cancel” to discard them and return to the
main program. The “OK” button is only enabled when changes have been made to the list or any
items in it.
Page 36 of 161
NOVAtouch™ Operating Manual
6.3.2.1
SOFTWARE
Report Template Editor
Figure 6-20, Report Template Editor
The “Report Template Editor” window is divided into two sections: on the left, the contents of
the report template being edited are shown (item type, name, and optional parameter), while on
the right side the available items are listed in a hierarchical structure.
The user may add only one item at a time to the contents list; to add an item select it from the list
on the right hand side, then click “Add”. The contents list allows the user to delete one or more
items from the content list in one step by using the “Remove >” button. (Select multiple lines by
using either the Ctrl + click, or Shift + click functions). Multiple line selections can only be
deleted, they cannot be moved up or down. Individually selected items can be moved up or down
using the “Move Up” and “Move Down” buttons below the list.
For a graph-type item the “Param” column shows the selected print size of the item (full page,
half page, or third of the page). These sizes refer to the printout size of the whole page minus the
header. The size can be changed for graph items only. To change the graph size, right click a
graph item in the contents menu, and choose the “Set Size” command from the Floating Menu.
Figure 6-21, Change Report Graph Size
Page 37 of 161
NOVAtouch™ Operating Manual
6.3.3
SOFTWARE
Display Properties
“Display Properties” expands when hovered over with the cursor to offer “Graph Properties”
and “Report Header”.
6.3.3.1
Graph Properties
This function allows the user to alter the presentation of each graph type independently by
editing the graph’s lines, markers and axes parameters. To edit graphs, first select the document
type (Physisorption or Physisorption Overlay) from the drop down menu in the top left corner of
the window.
To modify a particular graph type click on its name in the selection box. Use the slide bar to
access the full list. The highlighted item is the one whose properties are being edited and is
represented in the preview (example) graph area. The table in the bottom half of the window lists
each of the graph-specific elements and their current settings.
Figure 6-22, Graph Properties
Page 38 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Axis Scaling
To edit the graph’s x and y-axes click the Properties button to the right of the graphs list. The
available scaling modes for both axes are: Auto (0-110%), Auto (tight), and Fixed.
Figure 6-23, Graph Parameters



Auto (0-110%): The axis values will range from zero to 110% of the maximum data
value. If multiple curves are assigned to the axis, the maximum value is calculated from
all of those curves. If any of the data values are negative, the scaling will be set to Auto
(tight).
Auto (tight): The minimum and maximum data values will set the minimum and
maximum axis values to -10% and +10% respectively of the span of data values. For
example, if the data has values from 50 to 80, the scaling of the axis will be set to 47-83.
Fixed: Use the values entered into the corresponding minimum and maximum value
fields. Note that this option can leave some (or all) of the data points outside of the
visible range.
Any changes made will update the preview example. Click “Default” to the left of the preview to
revert the axes to the original program values.
Graph Line Parameters
Use this window to specify the color and marker type for each element displayed on the graph.
Select any one ID from the list of graph elements, then click the “Properties” button. In the
“Graph Line Parameters” window, select a different color and marker for the chosen element.
Page 39 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-24
To alter the color, click “Set” to open the Select Color picker, or enter a web-standard color in
the form of #RRGGBB (red/green/blue), or simply type any web-standard color name into the
color field! (Notice the color of the text will change to the chosen value).
To alter the data point marker, simply select one from the “Marker” drop-down list.
When multiple elements are selected, clicking Properties brings up the Color Spread selector.
This allows quick assignment of different colors to multiple elements across a range from “Start”
to “End” values. The more graph lines that are being edited, the more subtle the color difference
between adjacent elements in the list. The Color Spread dialog also allows assignment of the
same marker to all graph elements from the Marker drop-down list. Selecting “Leave as is” will
ensure the individual elements retain their previously assigned markers.
Figure 6-25, Multiple Line Parameters
Page 40 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Click “Default” to the right of the element list to revert colors and markers to the original factory
values.
Click “OK” to apply all changes made and return to the main program window. Uncheck the
Save box if you do not want to make these changes permanent, i.e. they will apply only until the
program is closed.
6.3.3.2
Report Header
Click to open the “Edit Page Header” window to create a custom header which will print at the
top of every page. To add a logo to the header click the “…” button, an explorer window will
open; select the appropriate image.
Figure 6-26, Edit Page Header

NOTE! Re-start the software for the change to take effect.
Figure 6-27, Custom Logo and Text in Page Header
6.3.4
Materials
The “Materials” item in the “Configure” drop down list, allows the user to modify the model
parameters associated with adsorbates (analysis gas) referred to here as “Phys Gas” and model
parameters for solids (adsorbents) referred to here as “Phys Adsorbent” and save these changes
for later use in data reduction.
Page 41 of 161
NOVAtouch™ Operating Manual
6.3.4.1

SOFTWARE
Phys Gas parameters
NOTE! There are several predefined models (indicated with * in Fig. 6-28) that
cannot be changed. Both “Save” and “Delete” buttons are disabled for these gases.
Figure 6-28
*
*
*
*
Figure 6-28, Predefined Phys Gas Models
The user can modify or add new models by importing a file (“Import” button) or changing values
to a predefined model. To load a predefined model, click the “Select” button and choose an
adsorbate from the side-menu. The current values for each of the parameters is automatically
entered into the appropriate fields. To make changes to the model, simply edit the values in the
appropriate fields, and click “Save As” (cannot be saved as the same name as the predefined
models). Use the “Export” function to save the model as a “.qcPGAS” data file. This file can be
used to send the edited adsorbent model via e-mail (or any other means) to another location; it
can then be imported (using the “Import” button).
To delete an existing model, choose it from the “Select” button, then click “Delete” (predefined
models cannot be deleted).

NOTE! Once a model is deleted it cannot be brought back. The delete function is
irreversible.
Click on the “Close” button to leave the window and return to the main program. If there are
unsaved changes at this time, the software will prompt to save them before continuing.
Page 42 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-29, Editing Adsorbate Gas Parameters
6.3.4.2
Phys Adsorbent Parameters
There are three predefined adsorbent models in the TouchWin™ software: Carbon, Oxygen (N2SF), and Oxygen/Zeolite. To modify and save a model see above Sect. 6.3.4.1
Figure 6-30, Setting the Adsorbent Model Parameters
Page 43 of 161
NOVAtouch™ Operating Manual
6.3.5
SOFTWARE
Phys. Data Reduction Parameters
Selecting “Phys. Data Reduction Parameters” from the “Configure” drop down menu will open a
window where all of the physisorption data reduction parameters can be edited. If using a
specialized gas or application, use this function to create the correct data reduction parameters
for the analysis. Once saved, the new data reduction parameters will be available when analyzing
data.
To load a previously setup profile, select “Load” from the “Phys. Data Reduction Parameters”
window. After creating your desired adsorbate model, select “Save” to make the profile available
for analytical use.
Figure 6-31, Defining the Parameters for an Adsorbate Model
6.3.6
Units of Measurement
The “Units of Measurement” option allows the user to change units in which the data are
presented. There is not an option to return to previous settings or restore previous selections after
changes have been made. The “Use Global” button will return all of the units of measurement to
the default values.
Page 44 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-32, Units of Measurement Window
6.3.7
Manage Instrument Profiles
Selecting “Manage Instrument Profiles” will launch the “Instrument Profiles” window where
instrument profiles may be added, edited, or deleted. An instrument profile contains the name
and IP address of the instrument and can be loaded when using the “Communication Parameters”
option (available on the “NOVAtouch/Instrument Name” drop down menu. See 6.4.3,
Communication Parameters).
Figure 6-33, Manage Instrument Profiles Window
6.3.8
Control Interface
The “Control Interface” option allows the user to select their preferred instrument representation.
Either choice (Control Center or Sidebar) will provide the same information: instrument status
(Status), analysis information (Analysis), and instrument configuration details (Operation). The
software will have to be restarted before any changes to the control interface can take effect.
Page 45 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-34, Control Interface Setting
6.3.8.1
Control Center
The “Control Center” setting displays all of the instrument’s information (Operation, Status, and
Analysis) in a window that can be minimized and therefore will not use space when expanding
graphs.
Figure 6-35, TouchWin™ Instrument Control Center
6.3.8.2
Sidebar
The “Sidebar” setting displays the instrument’s information on the left hand side of the
TouchWin™ screen. The Sidebar cannot be minimized, but functions an easy reference point.
Figure 6-36, TouchWin™ Sidebar
Instrument Name (NOVAtouch)
The “Instrument Name” (“NOVAtouch” by default) drop-down menu provides important
Page 46 of 161
NOVAtouch™ Operating Manual
SOFTWARE
resources for connecting to and communicating with the instrument. The title of this drop down
menu (i.e. the instrument’s name) is an item that can be edited by changing the “Name” field, or
by loading a profile, using the “Communication Parameters” window (see Sect. 6.4.3,
Communication Parameters).
6.4.1
Connect
Selecting “Connect” from the “Instrument Name” drop down menu will automatically establish
TouchWin™ communication with the instrument, if the IP address is properly set. After the
instrument is connected, the Sidebar or Control Center will indicate that the instrument is
connected (and the option to connect to the instrument will be greyed out). For instructions on
how to physically connect to the instrument see Sect. 6.1.3, Instrument – PC Connection.
6.4.2
Disconnect
“Disconnect” is only available when the instrument is connected to the TouchWin™ software.
Selecting “Disconnect” will cease communication between the software and the instrument until
“Connect” is selected again.
6.4.3
Communication Parameters
Selecting “Communication Parameters” from the “NOVAtouch/Instrument Name” drop down
menu will open the “Edit Profile” window. From the “Edit Profile” window, the instrument name
(which will become the title of this drop down menu) and IP address may be set.

NOTE! Before connecting to the instrument, coordinate the IP address set on the
TouchWin™ software with the IP address set on the NOVA by using the Touch
Screen’s “Maintenance Page”.
Figure 6-37, Communication Parameters Window
Page 47 of 161
NOVAtouch™ Operating Manual
NOTE! Changing the name or IP address in the “Edit Profile” window will not
create a new profile or permanently create a change to the current profile. To
make permanent changes to a profile, use the “Manage Instrument Profiles” item
from the “Configure” drop down list.

6.4.4
SOFTWARE
Advanced Operations
The “Advanced Operations” Floating Menu provides the user with extended access to the
instrument firmware. Here, you can view instrument errors, upload analysis data files and
instrument logfiles, and upgrade the instrument firmware.
6.4.4.1
View Pending Errors
“View Pending Errors” is only available when errors have been recorded. If there are no errors to
view this option will be grayed-out/unavailable. Selecting this option will produce a window
showing all of the error messages that have been received.
NOTE! This symbol will be displayed on the Operation tab of the Control Center or
Sidebar when an error has occurred. Errors can only be reported by TouchWin™ if
the software is connected to the instrument.
6.4.4.2
Upload from File
The “Upload from File” option can only be used if the instrument is connected to the
TouchWin™ software. Selecting “Upload from File” will open a window showing all of the files
available on the instrument. Select a file by choosing its field, then click “Upload”. After
uploading, the file will appear on-screen as well as in the Physisorb (or designated .qcuPhysIso)
folder.
Figure 6-38, Upload Analysis Results (Field Selected)
6.4.4.3
Upload Logfile
Selecting “Upload Logfile” will display the list of instrument logfiles. To upload an instrument
logfile, click on its name then click “Upload”. The uploaded file will be saved in the designated
Page 48 of 161
NOVAtouch™ Operating Manual
SOFTWARE
directory; by default a folder named “Instrument Logs”. That is located inside the folder
designated for instrument configuration files, (see 6.3.1,”Edit Folders”).
Figure 6-39, Upload Instrument Logfile

6.4.4.1
NOTE! To delete any logfiles from the NOVAtouch™’s logfile directory, make the
modification at the instrument’s Touch Screen, using the “Maintenance” section. It
cannot be modified using TouchWin™.
Upgrade Instrument Firmware
Selecting “Upgrade Instrument Firmware” launches the “Instrument Firmware Upgrade”
window which prompts to select the desired firmware file (.qcUpgrade).
When a valid firmware upgrade file has been chosen, the software confirms the selection with
the message (in green): “Valid upgrade file. Click Next to start upgrade process.”
Follow the instructions on screen until the process is done.

NOTE! Only perform the firmware upgrade if instructed to do so by QC Service.
When speaking with QC Service be sure to report the software version to ensure
compatibility between the software and firmware.
Page 49 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-40, Select File for Instrument Firmware Upgrade
Figure 6-41, File Selected for Instrument Firmware Upgrade
Figure 6-42, End of Instrument Firmware Upgrade
Page 50 of 161
NOVAtouch™ Operating Manual
6.4.5
SOFTWARE
Backup Instrument Settings
The “Backup Instrument Settings” option allows the user to create a file from which instrument
parameters can be restored e.g., instrument configuration, calibration values (manifold and
transducer), IP address, and cell calibration. It is important to backup your instrument settings
before any maintenance occurs and before any firmware upgrades.
“Backup Instrument Settings” launches the NOVAtouch™ System Backup window. The
software will prompt to enter a unique name, and save the backup file in the .qcNTSysBkp
format.
Figure 6-43, Start of System Backup Procedure
Figure 6-44, Save Backup File
Page 51 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-45, Completed System Backup
6.4.5.1
Restore Instrument Settings
Selecting “Restore Instrument Settings” opens an explorer window, choose a “.qcNTSysBkp”
file to restore the instrument parameters.
Figure 6-46, Select File to Restore System
6.4.5.2
Heater Profile
Selecting “Heater Profile” from the Instrument drop down menu allows the user to create a
Heater Profile using the TouchWin™ software. The “Heater Profile” is used to degas samples
on the NOVAtouch™. Use the buttons on the right to add, insert, or edit a line in the heater
profile (selecting any of these corresponding buttons will produce the window seen in Figure
6-48. Each line in a heater profile consists of three components: Target (C), Rate (C/min), and
Soak Time (min). The instrument will heat the mantle to the target temperature at the rate that is
Page 52 of 161
NOVAtouch™ Operating Manual
SOFTWARE
set then hold the mantle at the target temperature for the set soak time. Selecting “Move Up” will
move the selected line up, and similarly, selecting “Move Down” will move the selected line
down. “Delete” will remove the selected line from the heating profile. Selecting the “Save”
button will save the heating profile as an “.nwOGprm” file. Selecting “Load” will allow the user
to open any “.nwOGprm” profiles. Selecting “Send” will upload the profile to the NOVAtouch™
for use.

NOTE! “Send” the degas profile to the instrument to start the degassing
procedure from the Touch Screen.

NOTE! If an existing degas profile is overwritten using the TouchWin software, it
needs to be re-loaded (
) from the Touch Screen before the changes take
effect.
Figure 6-47, Heater Profile Window
Figure 6-48, Edit Heater Profile Option
6.4.6
Analysis Parameters
The “Analysis Parameters” window allows the user to setup the common/global parameters for
an analysis, as well as, analysis file names and desired analytical points. The Analysis
Parameters can be sent to the instrument by selecting “Send” or started from TouchWin ™ by
selecting “Start” from the Analysis Parameters window.
Page 53 of 161
NOVAtouch™ Operating Manual
6.4.6.1
SOFTWARE
Common/Global Analysis Parameters
The “Analysis Parameters” contains global parameters which effect every station listed on the
“Common” tab. The global parameters include the adsorbate gas, analysis options, P0 options,
void volume mode, and evacuation crossover.
6.4.6.1.1 Adsorbate Gas
Select a pre-defined adsorbate gas using the selector button. The individual parameters can be
overridden by entering custom values here. Check the “Use as Data Reduction Adsorbate
Model” box to use your settings as the data reduction adsorbate model.
6.4.6.1.2 P0 Options
Specify the method with which the instrument will determine the P0 value for the analysis
performed.

NOTE! Not all P0 modes are available for all adsorbate gases.
Measure: Use the instrument's P0 station to measure the P0 at the beginning of the analysis.
Calculate: Automatically calculates a reasonable P0 value from the measured ambient pressure.
Daily: Use the instrument's stored Daily P0 value. This method is only recommended when using
a stable Dewar in an environment that will not be subject to an atmospheric pressure change
greater than 1 mmHg.
Continuous: The P0 is measured continuously during analysis at each relative pressure.
Depending on the instrument model, a time interval may be used for re-measure timing.
Continuous P0 method generally increases analysis time.
User: Specify the P0 value (in Torr) to be used.
6.4.6.1.3 Void Volume Mode
Specify the method with which the sample cell characteristics are determined for analysis data
correction.

Helium Measure: Calculates the warm and cold zones by helium expansion. This is the
most accurate method, however, time consuming. It is recommended for high accuracy
measurements.

NOVA Mode: A faster method than the Helium Measurement Mode. The void volume
is obtained by measuring the sample volume via the expansion of the adsorbate gas at
room temperature using the cell calibrated values of the warm and cold zones. This
method does not require He to be installed on the instrument. When using this mode the
user will be asked to specify the cell ID at each station.

NOVA He Mode: A faster method than the Helium Measurement Mode. The void
volume is obtained by measuring the sample volume via the expansion of the He gas at
room temperature using the cell calibrated values of the warm and cold zones. This
Page 54 of 161
NOVAtouch™ Operating Manual
SOFTWARE
method is used when the sample has interaction with the adsorbate at room temperature.
When using this mode the user will be asked to specify the cell ID at each station.

User Entered: Allows the user to specify the cold and warm zone volumes (instead of
having the instrument measure them). The User Entered void volume mode is only
recommended when same cell and sample are analyzed again, and the user needs to
make sure that the coolant level is identical to the analysis that the volumes were
initially obtained from.
6.4.6.1.4 Evacuation Crossover
Specify the pressure to which the system pumps down using the Fine Vacuum path before
opening up to full vacuum (to prevent sample elutriation and contamination of the instrument).
The lower this pressure value is set, the greater the time required for evacuating the cell(s) and
the slower the analysis will be completed. The following values are available:
Pellet: Crossover at 760 Torr
Powder: Crossover at 38 Torr
Fine Powder: Crossover set to 5 Torr
User Entered: Value must be between 5 and 760 Torr
6.4.6.1.5 Analysis Options

Fast Init: Instructs the instrument to bypass certain steps during initialization.

Bath Thermal Delay: Number of seconds to wait after the coolant Dewar was raised to
allow the thermal equilibrium of the immersed cells.

Backfill Mode: How to refill the analysis cells after a completed analysis for safe
removal. Available options are Helium (requires Helium source), Adsorbate or Vent
(ambient air).
Page 55 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-49, Analysis Parameters
6.4.6.1.6 Active Stations
Check those stations to be analyzed when the analysis is started. At least one station must be
selected before an analysis can be started.
6.4.6.2
Analysis Parameters
This window contains a Station tab for each station present on the instrument. Each Station tab
has four sub-tabs that affect only the analysis at that station: Admin, Points, Analysis, and
Reporting. Each sub-tab must be completed prior to starting.

NOTE! The fields of each Station sub-tab will remain filled from previous
analysis. If there is a change to analysis procedure, address the change by altering
the analysis parameter’s station sub-tab.
6.4.6.2.1 Station Admin Tab
The “Comment” field can be used to provide further information.
The Admin tab however, requires that the “Results File” field be populated. The “Verify” button,
scans existing files to ensure that there are no naming conflicts. A unique filename can be
generated automatically using a “macro string”. If using a macro, the “Admin” tab will only need
to be changed at your discretion (i.e. – when the comment becomes invalid).
Page 56 of 161
NOVAtouch™ Operating Manual
SOFTWARE
6.4.6.2.2 Results File Macros
The software allows the use of special characters as a “filename template” to automatically
generate unique filenames for each run.
Each macro is in the form of $(CC:prm), in which “CC” is the macro code, and “prm” is an
optional parameter field. For example, the format: $(ST)_$(DT:yyMMMdd)_$(TM), will
produce the filename “1_15Oct08_111529”, if the analysis was started on Station 1 on October
8th 2015 at 11:15:29”.
Page 57 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Macro for Results File Field
$(DT)
$(TM)
$(SN)
$(ST)
Date of execution, the formatting strings should be written in the form $(DT:xx),
where the x’s are a formatting string value.
Time of execution, the formatting strings should be written in the form $(TM:xx),
where the x’s are a formatting string value
Serial Number
Inserts Station ID (1, 2, 3, or 4)
Formatting Strings
yy
Year in 2 digit format
yyyy
Year in 4 digit format
M
MM
MMM
MMMM
Month (without leading zero)
Month (2 digit, with leading zero)
Month name abbreviated (Jan, Feb, etc.)
Month name
d
Day of month (no leading zero)
dd
Day of month (2 digit, with leading zero)
ddd
Day of the week abbreviated (Mon, Tue, etc.)
dddd
Day of the week, full (Monday, Tuesday, etc.)
h
hour without leading zero
hh
hour with leading zero
H
in 24 hr mode, no leading zero
HH
m
mm
in 24 hr mode, with leading zero
minute (no leading zero)
minute (with leading zero)
s
second (no leading zero)
ss
second (with leading zero)
z
millseconds (no leading zero)
zzz
millseconds (3 digit, with leading zeroes)
AP or A
AM or PM (hours will be in 12 hour format)
ap or a
am or pm (hours will be in 12 hour format)
Page 58 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-50, Analysis Parameters – Admin Tab
6.4.6.2.3 Station Points Tab
The “Points” tab defines the relative pressure, timeout, equilibration, and tolerance for each point
in that station’s analysis. The list also contains the tags assigned to each point for use in
calculations (e.g., BET, BJH, etc.). The points list is separated into adsorption and desorption
branches. Relative pressure values are sorted in increasing order for the adsorption branch points,
and decreasing order for desorption branch points.
To add individual points, ensure that a branch is selected and click on the “Add” button. In the
pop-up window select the desired parameters (multiple points can be added by listing relative
pressure values separated by a semicolon (;). Set the Equilibrium parameters as desired, check
the data reduction tag boxes, and click “OK”. When points are added to a branch they will be
inserted into the existing points list in their proper location with respect to numerical order. In
addition the point list will be cleaned up: if any two points are closer to each other than 0.0025
relative pressure one of those will be automatically removed. The point with the higher value
will be removed when effecting the adsorption branch and the lower value when effecting the
desorption branch.
Spreading points allows the user to specify a relative pressure range by means of a point count or
relative pressure interval. If using [Spread (by #)] the program will insert the requested number
of points, equally spaced within the entered pressure range. When using [(Spread (by dP)] the
points will be spaced by the specified delta. For either spread option the starting and ending
points of the relative pressure range will be included in the list.
To change pre-existing points, select the points to be altered and click on the “Change” button. In
the pop-up window select the tag operations (note that each tag value has an ON and an OFF
box, tags with none of these selected will be left alone). The tolerance band and the equilibrium
times can also be adjusted (if the corresponding box is unchecked they will be left as is).
Page 59 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-51, “Change” Pop-up Window
To change the tolerance, equilibrium, or timeout, select the check-box to the left of the field. To
delete point(s) from the list, select the points and then click the “Delete” button below the list. If
a branch node is selected ALL points will be removed from it. Click the “Reset” button to
remove all points from the list (both branches).
The DoseWizard uses data from a previous run to determine a faster dose routine. Samples must
have very similar isotherms in order for the DoseWizard to function properly. The software will
prompt for a source file (previous run) to determine the dose routine.
Figure 6-52, Analysis Parameters – Points Tab
Page 60 of 161
NOVAtouch™ Operating Manual
SOFTWARE
6.4.6.2.4 Station Analysis Tab
Use the “Analysis” tab to setup the sample cell type. There are three cell sizes: 6 mm, 9 mm, and
12 mm. For each of the cell types there is a “w/rod” and “w/o rod” option; which denotes with
and without a filler rod, respectively. It is important to specify the cell type as with or without
rod. Choose the cell type from the drop down menu.
Figure 6-53, Analysis Parameters – Analysis Tab
Figure 6-54, Cell Type Drop Down Menu
6.4.6.2.5 Station Reporting Tab
The “Reporting” page allows the user to setup the Data Reduction Parameters and report options
for the analysis. Although the “Reporting” page can be ignored, it is beneficial to utilize this tab.
Within the Data Reduction Parameters section there is an extensive list of parameters. Because
the parameters list is broad the software includes a “Jump to” function. The “Jump to” function
will automatically take the user to a location in the list.

NOTE! Ensure that the Data Reduction Parameter Set is valid. Failure to do so
could result in various software failures and loss of data.
Page 61 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-55, Analysis Parameters – Reporting Tab
“Reports at end of Analysis” determines how the analysis data is processed after completion. It
provides three choices for the final report: “Open”, “Print”, and “Save as text”.

NOTE! TouchWin™ must be connected to the instrument during the analysis for
the report functions to operate.
Figure 6-56, Reports at end of Analysis – “Open” drop down menu
To choose a report style, select the desired report format from the “Open” drop down menu. The
drop down menu will provide a list of demo files that are provided with the software, as well as
any uniquely created report styles.
Page 62 of 161
NOVAtouch™ Operating Manual

SOFTWARE
NOTE! To create a unique report style use the “Report Template Editor” function
(see Sect. 6.3.2.1).
Figure 6-57, Reports at end of Analysis – “Print” Selection Button
To have the data report print at the completion of the analysis, select the check box next to the
“Print” field. When “Print” is active a check mark will appear next to the “Print” field (this check
mark will disappear when it is inactive).
Figure 6-58, Reports at End of Analysis – “Save as text” Drop Down Menu
The “Save as text” function saves the data as a text report for later reference. The report will be
saved using the style that is chosen from the drop down list. These report options are the same
that appear in the “Open” drop down menu.
Reports at End of Analysis Options
Open
Opens the selected report template in a new window. If "— no report —"
is selected, no report will be opened upon completion of the analysis.
Print
Check this box to print the report
Save as
Text
6.4.7

Select the same or a different report template to be exported into a text file
automatically.
Cell Calibration Parameters
NOTE! Before using the helium free “NOVA” method of analysis, empty cells
must be calibrated.
Selecting “Cell Calibration Parameters” opens a window with which the basic cell calibration
Page 63 of 161
NOVAtouch™ Operating Manual
SOFTWARE
profile can be setup and started. If the information from the initial Cell Calibration window is
sent to the instrument, the Touch Screen will ask for more information at the start of the
calibration. The cell calibration can be started from the TouchWin™ program by selecting “Start”
in the “Cell Calibration Parameters” window.
Figure 6-59, Cell Calibration Parameters
Figure 6-60, TouchWin™ Cell Calibration Start Window
Window
Selecting the “Window” tab will open a menu containing items that allow quick and efficient
modification to windows that are open on the software screen.
Page 64 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-61, Window Selection Menu
6.5.1
Cascade
Selecting “Cascade” will modify the arrangement of all of the open windows and cause them to
present one in front of the other in a cascading fashion.
Figure 6-62, Cascade
Page 65 of 161
NOVAtouch™ Operating Manual
6.5.2
SOFTWARE
Tile
Selecting “Tile” will arrange all open graphs into an equally spaced format favoring one graph in
each corner.
Figure 6-63, Tile
6.5.3
Tile Horizontally
Selecting “Tile Horizontally” will arrange all open graphs parallel to each other from left to
right.
Figure 6-64, Tile Horizontally

NOTE! “Tile Horizontally” will default to the regular Tile option when more than
five graphs are open.
Page 66 of 161
NOVAtouch™ Operating Manual
6.5.4
SOFTWARE
Tile Vertically
Selecting “Tile Vertically” will arrange all of the currently open graphs parallel to each other
from top to bottom.

NOTE! “Tile Vertically” will default to the regular Tile option when more than
three graphs are open.
Figure 6-65, Tile Vertically
6.5.5
Close View
Selecting “Close View” will close the file that is currently selected (usually the foremost file).
6.5.6
Close All
Selecting “Close All” will close all of the windows that are currently open with one action.
6.5.7
Show Sidebar/Control Center
This item will be listed as either “Show Sidebar” or “Control Center” depending on which of the
two options has been selected from the Configure tabControl Interface. Selecting this item will
bring the item (either the sidebar or the control center) into view.

NOTE! In order to effect the change between Sidebar and Control Center the
program must be closed and re-started.
Page 67 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Help
The “Help” menu provides information regarding software use and properties.
6.6.1
Contents
Selecting “Contents” from the TouchWin™ “Help” menu launches the “TouchWin™ Help
System”. The Help System is a valuable tool as it can answer most questions about software
functionality.
Figure 6-66, TouchWin™ Help System
6.6.2
About
The “About” function launches a window which provides information about the software version
and build. If there are ever any matters to be discussed with the Quantachrome Service
Department they will ask for the information provided in the about window.

NOTE! The instrument F/W (firmware) version is displayed on the Operation tab
of the Control Center or Sidebar when the instrument is connected to the software.
Figure 6-67, “About” TouchWin™
Page 68 of 161
NOVAtouch™ Operating Manual
SOFTWARE
TouchWin™ Icon Toolbar
The TouchWin™ Icon Toolbar consists of four icons that allow for quick access to frequently
requested file management options.
Figure 6-68, TouchWin™ Icon Toolbar
6.7.1
Open
Selecting the “Open” toolbar icon will launch the Windows file explorer allowing selection of
the desired file and opens it in TouchWin™.
Figure 6-69, Open File Explorer
6.7.2
Save
Selecting the “Save” toolbar icon will save a file as its current name. The type of file being saved
will be determined by TouchWin™.
6.7.3
Print
Selecting the “Print” toolbar icon launches the Print window allowing selection of: printer,
printer preferences, the page range to be printed, and the number of copies to print. Selecting
“Print” in this window will print the selected document or report.
Page 69 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-70, Print Window
6.7.4
Print Preview
Selecting the “Print Preview” toolbar icon will open a window that contains the file and presents
the file as it will be printed.
Figure 6-71, Print Preview
Configuration of TouchWin™-CFR
TouchWin™-CFR users have the additional option of operating the software within full
compliance of 21 CFR Part 11 guidelines. Consequently, a full audit trail is enabled when
utilizing the full security features of TouchWin™.
6.8.1
Initial security configuration
After TouchWin™-CFR is installed for the first time, the following screen will appear:
Page 70 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-72, CFR Security Configuration
Select “Full” Security Level to enable all of the security features including: required login, full
audit trail for any changes to data files, the ability to print audits, and software timeout due to
user inactivity.

NOTE! Select Full Security Level in order for the software to operate within the
guidelines of 21 CFR Part 11.
Select “Custom” Security Level to choose which security features to incorporate. To enable the
security feature, check the box.
Selecting “None” will disable all security features.

NOTE! Once a given Security Level is selected, it cannot be changed for a given
installation. To change the Security Level, uninstall and re-install the
software.
After the security level has been configured, a Superuser account must be created. Only a
Superuser has full control of the system, including defining and managing user accounts. Create
the account in the User Account Parameter screen.
Page 71 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-73, User Account Parameters
If all of the required fields in the User Account Parameters box have been filled in correctly, the
OK button will be active. Click the OK button when finished and the Login window will appear.
Figure 6-74, Login Window

6.8.2
NOTE! You have three chances to enter a correct User ID and Password
combination. The software will automatically disable the User ID’s account
if Login is failed three times in succession. If a Superuser’s account is
disabled (and this is the only Superuser account for the given installation),
the software must be uninstalled and then re-installed.
Assigning New Users
After login as the Superuser click on Configure  SecurityUser Manager. This will open the
User Account Manager window.
Page 72 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Figure 6-75, User Account Manager
Click “Add” to open the “User Account Parameters” window (below). Enter the required
information into the fields provided. Click on Access and select a level from the list.
Figure 6-76, Adding a User Account
There are three Access levels: Superuser, Manager, and Operator.
Superuser — users in this level have full control of the system, including defining and
managing user accounts.
Manager — users in this level have the same rights as a Superuser except: managing user
accounts and deleting points from an existing data file.
Operator — users of this level are allowed to operate the software with predefined presets
established by a manger or Superuser. The operator can only enter sample information:
sample name, sample weight, cell ID, comments, etc.
Page 73 of 161
NOVAtouch™ Operating Manual
SOFTWARE
After creating a new user follow these steps to set the password and allow access for the new
user.
1. Restart the software when prompted
2. Login as a Superuser
3. Logout as a Superuser
4. Login as the new user, enter User ID only
5. Enter and confirm new password
6. Click on the OK button
6.8.3
Viewing the System Audit
Go to: Configure SecurityView System Audit. This will display the security messages in a
window like the one below.
Figure 6-77, System Audit Trail

NOTE! Only a Superuser can access the System Audit Trail.
Page 74 of 161
NOVAtouch™ Operating Manual
SOFTWARE
Multiple Instrument Control
TouchWin™ allows control of multiple instruments from a single PC. Each instrument must be
connected to a separate Ethernet port, and the host computer must have sufficient computing power
to handle simultaneous communication with all the sources. By default, the software is installed
with single instrument control. To set up for multiple-instrument control, the following steps must
be taken (outside of the software):
1. Exit TouchWin™
2. Locate and open in a plain text editor the TouchWin.cfg file (located in the Configuration
Root folder of your installation).
3. Locate the value ct=1 in the section [Instances]
4. Change it to the number of instruments to be controlled (the software supports up to 4
instruments). Save and close the file
5. In the Configuration Root folder, create sub-folders named 2, 3 and 4 (one for each
instrument). There is already one named "1".
6. Copy the file instrument.cfg from the original sub-folder (1) to each new sub-folder (2-4),
and open them in a text editor.
7. Change the item name= (in the section connection) in each sub-folder to a unique name
in order to identify the instruments. Save the files.
8. Start TouchWin™. The software will display the unique names of each instrument on the
Figure 6-7. Also, the SideBar and/or Control Center have full functionality and are color
coded to aid in identifying each instrument. See Below.
Figure 6-78, Multiple Instrument Control Displayed
Page 75 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
7 TOUCH SCREEN
Please ensure that the NOVAtouch™ is properly installed. The vacuum pump should be connected
to the instrument and stainless steel dowels should plug each station. Verify that the gas is
connected and turned on with delivery pressure set to 8 psig. Turn on the vacuum pump, lastly,
power up the NOVAtouch™.
The main screen of the NOVAtouch™ user interface consists of:
 Side Menu/Tool Box — left side.
Containing buttons for selecting operation modes/pages.
 Page Content — the remainder of the screen.
Content depends on the current operation mode.
The ten pages are: Home, Manual Mode, Analysis Setup, Cell
Calibration, Progress, Degasser, Results Page, Logfile, Tools,
and Power (see left). A grayed-out symbol indicates that the
page/mode is not available (e.g., manual mode during an
analysis).
Switch between pages by tapping on the ToolBox buttons.
Some operations require an Operator ID to be entered. To enter
data into an entry field, simply tap on the field to have the
software keyboard appear on the screen. The operator name (in
normal mode) is NOT checked against any data. The operator
name can be anything except nothing (blank).
NOTE! The Operator ID field must be populated.
Status Bar
The status bar resides along the bottom of each page and displays the current state of the
instrument’s operations, each represented by a colored circle (see Figure 7-1, Touch Screen
Home Page). In general, a blue circle means “active” and a green circle means “inactive”. A blue
circle with a tool symbol inside means that the analysis or degas is setting up and will soon be
considered “active”. An orange-filled green circle indicates that a degas station is inactive and
that an over-temperature error is the cause of the inactivity.
Home
The passive Home page displays the status of the instrument hardware: valves, Dewar bath
position, pressure reading(s) and temperatures; nothing can be actuated or entered on this page.
This page also displays the model type, firmware version and the instrument serial number in the
lower left hand corner. Closed valves are colored amber (yellow) and open valves green. Bath
position, coolant level and over-temperature appear illuminated when actuated.
Page 76 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Manifold
temperature and
pressure
Helium
Over-temperature
Analysis stations
Degassing
stations
Analysis
gases
Mantle
temperatures
Bath position and coolant
contact with level sensor
Model type, firmware version and
instrument serial number
Side menu
Instrument
status bar
Figure 7-1, Touch Screen Home Page
Manual Mode page
The Manual Mode page is visually very similar to the Home page with two distinct differences:
(i) there is a “Stop” symbol in the bath control area, and (ii) each valve has a gray switch-bar
displayed across it. Tapping the valve symbol reverses the valve’s status and the bar’s
orientation, from closed-to-open and vice-versa. Note that the position of the switch-bar displays
the last user requested state, and the color of the valve indicates the state of the valve as reported
by the firmware. When changing the valves between states there may be a ½-1 second
communication delay between changes. The Gas-tank port selection is exclusive, it does not
allow more than one gas-port to be selected; opening one gas port will close the other two. All
can be closed however by tapping on the open valve. When in manual mode the system does not
perform a purge cycle when switching between analysis gases (adsorbates).
Page 77 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Tap here to reset
the degasser’s
over-temperature
status if and
when the issue
that caused the
over-temperature
condition has
been rectified.
Tap valve symbol
to change its state
Click to remain in manual
mode (when not locked)
Select to lock the unit in
Manual mode
Figure 7-2, Manual Mode
Dewar buttons (UP, Stop, Down (DN)) control the movement of the Dewar. Up/down travel will
continue until either the stop symbol is tapped, or the mechanism reaches the travel limit
switches.
Degas port indicators (located above the degas port numbers) turn bright red if an overtemperature condition exist, in which case power to the affected mantle(s) is terminated. If such
an event occurs check the degassing setup (thermocouple connections etc.) to eliminate any
safety concerns, then tap the reset button, or press the button above the status indicators on the
panel on the right side of the instrument (see Sect. 5.13). If the over-temperature condition still
exists the indicator will NOT turn off – this is not a bypass function.
Manual mode has a three minute time-out. Without user interaction for three minutes the system
will inform the user that the idle time has expired and the user can choose to (i) remain in manual
mode (for another three minutes), or (ii) go to the Home page. If no response is given within one
minute, the instrument will leave manual mode and return to Home page status. The timer can be
disabled by toggling the Lock button (next to the Manual Mode label) allowing the user to stay
in manual mode until another mode is selected.
Analysis Setup page
The Analysis Setup page is used to start an analysis. Each section allows the loading of predesigned analysis/station presets (designed and downloaded to the instrument using TouchWin™
software). To load a preset click on the upload button
, and make a selection from the list
of presets available.
Page 78 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Enter the required parameters by tapping the appropriate entry field invoking the keyboard. The
Sample ID field has a pull-down button that allows the user to select a pre-existing Sample ID
file created in the degassing section. If there is no pre-existing Sample ID, enter a Sample ID
using the keyboard.
Each station can be deselected (or re-selected) using the check-box next to its label. Click the
Summary button to show a description of the current settings and review the analysis before
starting. When viewing the Summary, the lines in red indicate fields that are invalid (or empty)
preventing the analysis from initiating. Grayed out lines indicate unavailable or deselected
stations. The Clear button will clear all fields –all fields will have to be populated again after
clearing the Analysis Setup.
The Recall button will bring back the previous entries. The Start button will only be enabled
when all data fields are valid and the instrument is in a state where it can accept the command.
To start the analysis select Start and the analysis will begin as specified.
a clean-up automatically. This may take up 20 minutes depending
ituponwilltheperform
amount of gas remaining in the sample cell.
NOTE! If the instrument needs to establish a safe status due to aborted analysis,
suitable coolant and placed on the lift drive in the analysis compartment,
with
and P cell and level sensor installed (as appropriate) before starting the analysis.
NOTE! Samples must be in position at their desired station ports, Dewar filled
0
Page 79 of 161
NOVAtouch™ Operating Manual
Select to load the analysis gas.
(Note: If the gas port name
matches the adsorbate name
uploaded, the gas input port
will be selected automatically.)
TOUCHSCREEN
Select to load an analysis
parameters preset.
(Note: Presets must be created
using TouchWin™.)
Select to bring back the
previous entries.
Select to clear the “Analysis Setup”
Select to load a station preset.
(Note: Presets must be created
using TouchWin™.)
Select to view a
summary of the
analysis before
starting. Any missing
items will be
highlighted in red.
Select to start
the analysis.
Figure 7-3, Analysis Setup
Page 80 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Figure 7-4, Select Analysis Preset Window
Figure 7-5, Select Station Preset Window
Cell Calibration page

NOTE! A cell calibration is only required if using NOVA mode.
The Cell Calibration page is used to start a cell calibration and is necessary when setting up a
NOVA mode analysis. The Common Parameters section allows the user to select presets to be
used when running the NOVA mode cell calibration. At each of the stations the user must define
the station parameters: Cell Type and Cell ID. To overwrite a previously used Cell ID select
“Overwrite” at each station. For each station that you would like to run, select the box on the left
side of the station. Unselected stations will not run. To load a preset, click on the upload button
, and select from the list of presets available.
The options along the lower portion of the Cell Calibration page are the same as the options on
the lower portion of the Analysis Setup page: The Clear button will clear all of the fields –all
Page 81 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
fields will have to be populated after clearing the Analysis Setup. The Recall button will bring
back the previous entries; the Summary button will show the current analysis setup. The Start
button will only be enabled when all data fields are valid and the instrument is in a state where it
can accept the command.
Select to launch the
list of cell types and
choose the correct
cell type for the
analysis.
The Cell ID is
entered here.
(Note: Only
alphanumeric
combinations
may be used.)
Previously
created Cell IDs
Select to
overwrite the
Cell ID.
Select to start the
cell calibration.
Select to clear
field that is
currently
activated.
Select to fill
the fields with
the input used
in the most
recent
analysis.
Select to view the summary of the
analysis setup before starting the run.
This display items like the file names
and gas settings. It is suggested to
always review the analysis summary
before starting an analysis.
Figure 7-6, Cell Calibration


NOTE! It is not necessary to calibrate cells every time a sample is analyzed.


NOTE! Calibration is cell specific.

NOTE! If cells and filler rods have been calibrated they can be labeled using the
white spot at the top of the stem. Labelling cells and filler rods will help prevent
future errors.

Page 82 of 161
NOVAtouch™ Operating Manual


TOUCHSCREEN
NOTE! The calibration will be stored in the instrument for future use.
Select which
stations to run for
the calibration by
choosing the
appropriate boxes.
Select to load a parameters
preset. (Note: These must
be created using
TouchWin™.)
Select to launch
the list of gas
ports. Choose
the desired gas
port from the list
that opens.
Select to input the P0
value (when using
User Entered P0).
A list of all of the
present calibration
IDs.
When running a
NOVA mode
analysis this list will
be available for
selection.
Figure 7-7, Cell Calibration Continued


NOTE! Use “Calculate” P0 for adsorbates at their own boiling points.
Page 83 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Progress page
The Progress page provides up to date information about the state of the instrument during an
analysis. The upper left hand corner shows the status of the instrument hardware. The upper right
hand corner will show the status of the analysis. And, the lower component of the Progress page
shows the instrument logfile.
Current status of the instrument
Current logfile of the
instrument’s activities.
Figure 7-8, Progress Page



NOTE! The progress page is only accessible during an analysis.

NOTE! Along the Status Bar at the bottom of the Touch Screen are
colored circles indicating the instrument status. In general, blue means
inactive and green means active.
Page 84 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Degassing Page
The degassing page is used to start the degassing of the samples to prepare them for analysis.
The degassing mode (Vacuum* or Flow) is selected on the Degas Page. The mode cannot be
changed when one of the stations is loaded.
The heater profile must be setup using the TouchWin™ software, then loaded to the station
using the Touch Screen. The Sample ID is created at the Degassing Page in this step. When
loading a station select and load a Heater Profile using the upload button
. Specify Sample
ID for each of the ports in the stations that are to be analyzed, if omitted, there will be no sample
ID file created. Tap the Preview button to see a tabular representation of the heater profile.
Check the box next to Load to include this station on the “Load Selected” operation. When
removing a station, select the station to unload and tap “Unload Selected”.
NOTE!

The degasser can either Load or Unload, but cannot do both at the same
time.
*NOTE! Vacuum degassing is only available for LX models.

Figure 7-9, Degassing Page
Results page
The Results page will show up to four analysis results (one for each station) at a time. During an
analysis the running analysis for each station will be displayed (and automatically updated as
Page 85 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
new points are measured). Alternatively, a previously measured data set can be loaded and
viewed. At each station the user can click on the “Station” button to maximize/restore the station
graph.
The following commands can be selected from the “Action>” menu:

Upload – force a data retrieval during analysis

From File – select and load previous analysis Data
The user can select one of the following modes from the top toolbar:

Isotherm – Displays the isotherm plot. The “Ads” and “Des” toggle buttons on the plot
allow the user to turn the display of adsorption and desorption points off and on.

Analysis – Displays analysis information: File name, adsorbate, number of points
acquired, P0 mode, time elapsed, total volume, Sample ID, and sample weight.

BET – Displays the BET plot.

Surface Area – Displays BET surface area calculation results (tabular) including
surface area, slope, intercept, correlation coefficient, and C-constant.


NOTE! BET calculations are performed on M tagged points (if there are at least 2 of
them specified in the analysis point list) OR, if none are tagged, then on all points
between 0.05 – 0.3 P/Po.
Selecting Action will
open a side menu
which allows the user
to select a data file for
viewing.
Figure 7-10, Results Page
Page 86 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Figure 7-11, Analysis Parameters
Figure 7-12, BET Plot
Page 87 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Selecting Surface
Area will show the
BET data reduction
results for the
selected data.
Figure 7-13, Surface Area Data
Maintenance page
The Maintenance Page provides access to the instruments setups such as gas port assignment, IP
address, and system password. The components of the Maintenance Page behave as described
below:
Figure 7-14, Maintenance Page
Page 88 of 161
NOVAtouch™ Operating Manual
7.9.1
TOUCHSCREEN
Security
When “Security” is selected the user can specify a system passcode (a numeric value) that is
required to enter this Configuration window. “Secure Mode” can also be enabled, from which a
username/password combination is required for non-trivial operations and allows management of
user accounts.
Figure 7-15, Security Window
7.9.2
Time Zone
Set the time-zone (GMT ± N hours).
7.9.3
Time
Selecting “Time” will allow the user to set the instrument’s calendar and clock.
To have proper time/time zone combination, always change the time zone
first, and then the time.
NOTE!
Figure 7-16, Time Zone Select
Page 89 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Figure 7-17, Date and Time
7.9.4
Language
Selecting “Language” drop down will allow the user to select the user interface language from
the given options.
NOTE! The instrument will have to be restarted before this change can take effect.

Figure 7-18, Language Select
Page 90 of 161
NOVAtouch™ Operating Manual
7.9.5
TOUCHSCREEN
Network Settings
Provides access to set the instrument’s network IP address. The instrument IP address must be a
unique IP address because this address will be used exclusively for the communication between
the instrument and software. Contact your IT department for assistance
NOTE! The instrument will have to be restarted before this change can take effect.

Figure 7-19, Network Settings
7.9.6
Operation Mode
NOTE! The instrument will have to be restarted before this change can take effect.

Allows selection of the following options:

He tank – He required operations are enabled/disabled depending on the status
given to the Helium tank (present or not present).

Start new Logfile at analysis start – If selected, will generate a new logfile at
the start of every analysis.

Purge logfiles after – Select to choose how long the instrument will store the
logfiles. The options are as follows: 15 days, 30 days, and Never.
NOTE! It is recommended that the “Start new Logfile on analysis start” option is
selected.

Page 91 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Figure 7-20, Operation Mode
7.9.7
Gas Input Ports
Used to assign gas names to each of the three gas input ports, if these exactly match the
adsorbate names in TouchWin™ the instrument will automatically use the correct port for the
analysis (based upon the analysis parameters).
Figure 7-21, Gas Input Ports
7.9.8
Daily P0
This feature measures the P0 and stores the value for later use. Requirements to measure the P0:
CLS, P0 cell, and adsorbate gas. The cryogen must be in the Dewar at the analysis station prior to
starting a Daily P0 run. To perform a Daily P0, select the correct port for the adsorbate gas, then
select “Start”. After “Start” has been selected, the instrument will prompt to verify that a P0 cell
and the adsorbate gas are installed. The instrument will go through the process of measuring a
Daily Po after “OK” is selected.
Page 92 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Figure 7-22, Daily P0 Window
7.9.9
Calibrate
Instrument calibration (manifold volume, transducers, etc.) can be accessed by selecting
“Calibrate”. To start any of these processes, choose the desired option from the “Operation” drop
down menu then select “Start”. The instrument will provide prompts for each process.
Figure 7-23, Calibration Window
Page 93 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN

NOTE! Only perform a pressure transducer calibration if instructed to do so by
Quantachrome’s Service Department.
Figure 7-24, Transducer Calibration
NOTE! Only perform a Leak Test if instructed to do so by Quantachrome’s Service
Department. Consult Quantachrome’s Service Department when reviewing your
results.

Figure 7-25, Leak Test Instrument Part 1
Page 94 of 161
NOVAtouch™ Operating Manual
7.9.10

TOUCHSCREEN
File Manager
Allows viewing and deleting stored files in the following categories:
Analysis Station (Station parameters, points, etc.)
Analysis Preset (Analysis Common Data – adsorbate, etc.)
Analysis Result
Heating Profile
Cell Calibration Preset
Cell Calibration Data (Cell-IDs)
Sample ID – Degas profile applied to a sample
Logfiles
Logfile
The Logfile page shows the instrument record as it is being reported. The logfile is useful for
finding the exact actions of the instrument and will be requested by Quantachrome Service
Department. The instrument logfile can be uploaded to your personal computer using
TouchWin™ software (Instrument Name  Advanced Operations  Upload Logfile). The
uploaded file will typically be stored in “Instrument Logs”. Using default settings, the
“Instrument Logs” file can be found by accessing C:\QCdata\TouchWinCfg\1\Instrument Logs.
The number (1 in the example given) corresponds to the instrument number created in the
instrument’s config file.
NOTE! It is recommended that the “Start new Logfile on analysis start” option is
checked ON in the Maintenance settings.

Tapping the Logfile twice will bring up the instrument control menu allowing the user to adjust
the way that the data is presented. The menu options function as described below:
Autoscroll – Selecting Autoscroll will cause the instrument to always present the most
current line and automatically scroll as the lines build. Deselecting AutoScroll will cause
the logfile to stay in one position until the page is moved using the scroll bar on the right
side of the page.
Show Time – Selecting Show Time will list the date and time next to each line presented.
Show Source – Selecting Show Source will list which data type is being shown (Instr, St
1, St 2, St 3, St 4, Degas, Local) next to each line presented.
Data Types: Instr, St 1, St 2, St 3, St 4, Degas, Local – Use the check boxes to
show/hide specific elements.
Page 95 of 161
NOVAtouch™ Operating Manual
TOUCHSCREEN
Tapping the logfile
twice will bring up the
logfile control menu.
Current logfile of the
instrument’s activities.
Figure 7-26, Logfile Page Degassing
Power Mode
Selecting Power Mode opens up a “Select exit operation” window
Shutdown — closes the Touch Screen User Interface to prepare the instrument for being
powered off using the mains power switch on the side of the unit. When the instrument is
powered on again the User Interface will initialize automatically.
Restart — reboots the instrument and initializes the User Interface automatically.
!
WARNING! Always shutdown using the Touch Screen before switching the unit
off at the mains power switch on the left side of the unit.
Page 96 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
8 INSTRUMENT OPERATION
Introduction
The following sections describe how to prepare and analyze a sample. Please refer to other
sections throughout the manual as applicable e.g., for starting an analysis refer to Sections. 7.4
and 8.5 for sample preparation refer to Sections 7.7 and 8.3 .
If using the NOVA mode as the void volume method, the sample cells must be calibrated
prior to starting an analysis.
Cell Calibration
1. Use a cell adapter (with O-ring) and place an empty cell into the NOVAtouch™. Use
the cell appropriate to the analysis and mimic the setup exactly (i.e. – if analyzing with
a filler rod, calibrate using a filler rod.)
2. Fill a Dewar with liquid nitrogen to the lower tip of the level indicator, and ensure the
CLS is installed correctly on the BNC connector.

NOTE! It is important to start a NOVA mode analysis with the same level of liquid
nitrogen for each analysis.
3. Use TouchWin™ to provide the Cell Calibration Parameters. To access the Cell
Calibration Parameters window select the “Instrument” drop down menu, then select
“Cell Calibration Parameters”. The cell calibrations can be sent to the instrument by
selecting “Send”, allowing the calibration to be started from the Touch Screen (see 7.5
, Cell Calibration). Alternately, the cell calibration may be started from TouchWin™ by
selecting “Start” on the “Cell Calibration Parameters” window (see 6.4.7, Cell
Calibration Parameters).
4. Enter the Cell ID and Cell Type for each station to be used.
rod.
NOTE! It must be specified that the cell type has a filler rod or does not have a filler
Sample Preparation

NOTE! Every sample should be degassed by flow or vacuum before analysis.
8.3.1
Selecting a Sample Cell
There are three factors to consider when selecting a sample cell: stem diameter, sample
amount/bulb size, and overall length.
Stem diameter: Choose the narrowest diameter cell that will comfortably admit the sample. For
example, a fine powder should be analyzed in a 6 mm outer diameter (o.d.) (4 mm i.d stem cell).
Page 97 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
Use the 12 mm o.d. stem cells for large pieces that cannot be reduced in size. Larger particles
such as granules, and small pellets might require a 9 mm o.d. (7 mm i.d.) diameter stem.
Cohesive powders may be analyzed in 9 or 12 mm stem cells to reduce elutriation and facilitate
addition, removal, and cleaning.
Sample amount/Bulb size: Always use the smallest bulb that will accommodate the optimal
amount of surface area. Larger total surface areas can certainly be analyzed, but they may
lengthen the analysis. For surface area determinations only, sample amounts of at least 1 m2 to 5
m2 can be analyzed using nitrogen, but careful consideration should be given to proper degassing
and equilibrium criteria. Full adsorption and desorption isotherms should have at least 15 – 20
m2 in the cell.
NOTE! Wider stems and larger bulbs can be beneficial in reducing elutriation.
Cell length: NOVAtouch™ sample cells are available in two lengths, referred to as “long” and
“short”. Use the short CLS (P/N 00080-CLS-S) for short cells (do not mix-and-match long and
short cells in the same analysis session).
8.3.2
Methods of Sample Preparation
The sections that follow will aid in performing this step correctly. Before
 NOTE!
starting a degassing procedure, the degassing profile will have to be created using
TouchWin™ and loaded onto the instrument. (See 8.3.3, Choice of Degassing
Temperature and Time, 8.3.4, Elutriation and Its Prevention, and 8.3.5,
Programming the NOVAtouch™ .)
 NOTE! The degasser can only be started using the NOVAtouch
™
Touch Screen.
For more information regarding degassing see Sect. 7.7, Degassing, or
 NOTE!
8.3.5, Programming the NOVAtouch™

NOTE! To calculate the weight of the sample, subtract the empty cell weight
from the post degas weight (of both the cell and sample).
Page 98 of 161
NOVAtouch™ Operating Manual
8.3.2.1
INSTRUMENT OPERATION
Vacuum Degassing
Vacuum degas exposes the sample the low pressure of the vacuum pump while it is under heat to
remove the impurities that would otherwise prevent a proper analysis. To set-up a vacuum degas
on the sample(s) follow the directions below:
1. Weigh an empty cell, add sufficient sample (2-50 m2
total area).
2. Place the sample cell into the heating mantle, then
install the cell into the NOVAtouch™ degasser
(without filler rods) using the appropriate O-rings
and fittings.
3. Once the cell is installed use the hooks located at the
top of the degassing compartment to secure the
heating mantle.
4. Create a degassing profile in TouchWin™ and send it
to the instrument.
5. Ensure that the chosen degassing method is set to
vacuum.
6. Start the vacuum degassing procedure at the
NOVAtouch™.
7. After degassing is complete, allow the cell to cool.
Remove and reweigh the cell to obtain the sample
weight.
Figure 8-1, Vacuum Degassing
8.3.2.2
Flow Degas
Flow Degas directs a flow of gas (attached to the backfill/flow degas input port) over the sample
while it is heated to remove impurities. To set-up a flow degas follow the directions below:
1. Install the flow degasser assembly into the degasser station(s) using the 6 mm fittings and
O-rings. Because the degasser is split into two stations each with two ports (creating a total
of four ports) it is important that the second port associated with a station is plugged with a
dowel pin if it is not going to be used during degassing.
2. Place the cell inside of the heating mantle and place the heating mantle into the holder to
allow it to sit up right during the degassing process.
3. Insert the metal flow tube into the cell so that the tip of the metal flow tube is close to but
does not touch the sample. Use the collar stop to hold the metal flow tube at the appropriate
position.
4. Create the degassing profile in TouchWin™ and send it to the instrument.
5. Start the flow degassing procedure at the NOVAtouch™. Ensure that the chosen degassing
Page 99 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
method is set to flow.
Figure 8-2, Flow Degasser Assembly
Figure 8-3, Flow Degassing on the NOVAtouch™
8.3.3
Choice of Degassing Temperature and Time
Samples should be degassed at the highest temperature that will not cause a structural change to
the sample (up to 350 °C for standard mantles, or 450 °C if using high temperature mantles).
This will accelerate the degassing process. For instance, most carbon samples can also be
degassed at 300 °C, as can calcium carbonate. Many hydroxides must be degassed at a lower
temperature. Degassing organics must be performed with care since most have quite low
softening or glass transition points. For example, magnesium stearate, a common pharmaceutical
formulating compound, should be degassed at 40 °C according to USP requirements.
Loosely bound water (physisorbed water) will be lost at relatively low temperatures under the
influence of vacuum, but strongly bound surface water might require high temperatures. Many
zeolites, for example, will retain significant quantities of water in their micropores up to 300 °C
or higher.
Page 100 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
Use technical reference literature such as the Handbook of Chemistry and Physics (CRC, Boca
Raton, Florida) and standard methods such as those published by ASTM, DIN, ISO, etc., to
guide your selection of an appropriate degassing temperature.
If there is access to thermal analysis equipment, especially gravimetric, an analysis should be
conducted on a separate aliquot of material prior to degassing on the NOVAtouch™. A suitable
degassing temperature lies in a plateau, or weight-stable region, of the thermogram. Ideally, the
thermal analysis should be conducted under vacuum. In general, a temperature that is too low
will cause lengthy preparation, and may result in lower than expected surface areas and pore
volumes. A temperature that is too high can cause irreversible damage to the sample, which
can result in a decrease in surface area due to sintering, or an increase in surface area due to a
thermally induced decomposition.
Time for complete degassing (i.e., complete removal of unwanted vapors and gases adsorbed
on the sample surface) can only be properly determined by conducting a series of tests to
determine those conditions of temperature and time which yield reproducible data. As a general
guideline however, three hours (at temperature) should be considered a reasonable minimum.
IUPAC recommend no less than sixteen hours, which can be conveniently achieved overnight.
Samples that require low temperatures generally require the longest degas times. However, the
USP recommended degassing period for magnesium stearate is just two hours at 40 °C.
8.3.4
Elutriation and Its Prevention
Elutriation, or loss of powder out of the sample cell, is caused by too rapid a gas flow out of the
cell, and is most problematical when vacuum degassing (since it is easy to regulate the velocity
of gas when flow degassing) and analyzing very fine powders. It is particularly challenging for
low-density samples, fumed silica for example.
Wider stems and larger bulbs can be beneficial in reducing elutriation. Wider stems reduce the
velocity of the gas leaving the cell when evacuation begins, thus it is less likely to entrain
powder particles and transport them upwards and out of the cell.
The presence of a filler rod significantly increases gas velocity because of the narrowing of the
internal dimensions and can exacerbate elutriation. In problematical cases, the filler rod may be
dispensed with during analysis, but some loss of resolution and/or sensitivity may result.
Degassing should always be done without filler rod if possible.
The most dramatic elutriation problems are encountered during degassing of damp, “light”
powders. As the sample heats from ambient, the pressure over the sample decreases due to the
action of the vacuum. At some point the water “flashes” into steam. This rapid expansion of gas
volume drives powder out of the bulb and up the stem of the cell. This condition can be reduced
or eliminated by (i) pre-drying the samples in a conventional drying oven and degassing under
just vacuum or (ii) raising the temperature of the heating mantle in 20 degree steps. It is
recommended that the temperature be “paused” at 60 °C for 30 – 60 minutes under vacuum to
allow for a milder removal of moisture before increasing the temperature to 80 °C, then 100 °C
and finally maximum degas temperature.
A further flow restriction in the form of a tight fitting glass fiber filter cartridge called a Nonelutriating plug which can be inserted into the stems of sample cells so long as they don't have
hanging filler rods. Non-elutriating plugs are available for 6 mm, 9 mm and 12 mm sample cells.
Page 101 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
Contact Quantachrome for more information regarding Non-elutriating plugs.
In the most difficult cases, and the aforementioned methods have not eliminated the problem it
might be necessary to insert a small glass wool plug into the cell stem. This can be held in place
between two halves of a cut-in-two glass filler rod. This is the only time that a filler rod should
be used in the degasser.
8.3.5
Programming the NOVAtouch™ Degasser
8.3.5.1
Setup
To set-up a Degasser temperature profile using TouchWin™ follow the steps below:
1. Open TouchWin™ software, then open the Instrument’s drop down menu from the tool
bar.
Figure 8-4, Instrument Drop Down Menu
2. Select “Heater Profile” from the drop down menu. The “Heater Profiles” window will
open. From this window the user will be able to: create, modify, save, and send heater
profiles. A degassing profile can contain a total of six lines, each line is a “step” which must
include a ramp rate and target temperature. The soak time is optional. To load a degassing
procedure that has been created using TouchWin™, select the “Load” function in the
“Heater Profiles” window. Select the desired file from the File Explorer window.
Figure 8-5, Creating Degassing Profile
3.
To create the heater profile select “Add” on the right hand side of the heater profiles
Page 102 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
window. The Add function will request three fields: Target, Rate, and Soak.
 The “Target” field is the highest targeted temperature of the degassing step. The Target
value is in degrees Celsius. The minimum temperature is typically 20 degrees Celsius,
depending on the ambient temperature. The maximum temperature is 350 degrees
Celsius using standard mantles. Using a set of high temperature mantles and an
instrument configured for high temperature degassing, the maximum temperature can be
450 degrees Celsius.
 The “Rate” field is an expression of how quickly the target temperature will be reached.
The Rate field is expressed in degrees Celsius per minute. The software will not allow a
rate greater than 90 degrees Celsius per minute.
 The “Soak” field is the amount of time (in minutes) the degasser will remain at the target
temperature. The maximum amount of time that the field will accept is 2400 minutes (40
hours).
Figure 8-6, Editing/Creating an Degassing Step
4. To modify the degassing procedure, use any of the tools/functions listed on the right hand
side of the Heater Profiles window.
 Add – Create multi-stepped degassing protocols. The added line will be placed at the
bottom of the degassing protocol. It is limited to six line/step protocols.
 Insert – Using the “Insert” function will add the new line directly above the line that has
been selected.
 Edit – Selecting “Edit” will open the step editor and allows modification of the protocol
step’s details: target, rate, and soak.
 Move Up – Using the “Move Up” function will allow changes to the order of the
degassing protocol by moving the selected line up one level.
 Move Down – Using the “Move Down” function will allow changes to the order of the
degassing protocol by moving the selected line down one level.
 Delete – Using the “Delete” function will remove the selected line from the profile.
Page 103 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
Figure 8-7, Example of a Completed Degassing Profile
Loading the Degassing Profile onto the NOVAtouch™
8.3.5.2

NOTE! To calculate the weight of the sample, subtract the empty cell weight
from the post degas weight (of both the cell and sample).
1. Ensure that the NOVAtouch™ is connected. To connect to the instrument, select the
Instrument’s drop down menu from the tool bar, then select “Connect”.
2. Send the Heater Profile to the instrument by selecting “Send” on the lower portion of
the Heater Profiles window.
3. At the instrument, select the degassing Page from the Touch Screen side-bar.
4. Use the upload button
at either Station 1 or Station 2 to open the desired
degassing file. Locate the file in the explorer window when it opens.

NOTE! The degassing profile contains temperature information only. The
choice of flow (available for all models) or vacuum (LX models only) is done at
the instrument Touch Screen via the “Mode” button.
Page 104 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
Figure 8-8, Send Degassing Profile
Figure 8-9, Name Degassing Profile Window
Figure 8-10, Degasser Set for Flow
Page 105 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
Unloading the Degasser
When removing a station, select a station to unload and tap “Unload Selected” (see 7.7 ,
Degassing Page).
NOTE! The degasser can be unloaded during a sample measurement by
navigating to the Degassing page on the Touch Screen, selecting a degas
station, and selecting “Stop”.

Preferably, nitrogen should always be used as flow/backfill gas to prevent or minimize buoyancy
errors on weighing. A sample cell will weigh less when filled with helium than when filled with
air or nitrogen. The error introduced is approximately 1 mg per mL of cell volume. This can be
significant when using extremely small sample weights (< 50 mg).
the right side of the instrument at the port labelled “Degas Flow/Backfill”.
NOTE! The flow/backfill gas is determined by what is connected to the input on
Allow the mantle to cool below 80 °C before unloading the degasser. A sample cell which feels
only warm to the touch while still under vacuum can be much hotter to the touch when backfilled
with gas. This is particularly true if degassing a large mass of metal sample. Exercise caution! A
warm sample cell can also introduce weighing errors. The sample cell should be allowed to cool
to room temperature before weighing. If sample throughput permits, cool thoroughly while
attached to the NOVAtouch™, otherwise remove and cap the cell, and/or transfer to a desiccator.
Start Sample Analysis
8.5.1
Initial Steps at the Start of an Analysis
After degassing the sample, follow the initial analysis start procedure below:
1. Verify that the adsorbate gas contains at least 500 PSI in the tank.
2. Fill the Dewar to level indicator, and allow it to settle for 20 minutes, if time allows.
3. Carefully insert filler rods in the sample cells.
4. Install sample cell(s) into the analysis station, taking note of the station number each
sample is loaded into.
8.5.2
Starting an Analysis
After setting up the analysis using TouchWin™ it can be started directly from a PC or by using
the NOVAtouch™ TouchScreen.
8.5.2.1
Starting an Analysis from the TouchScreen
Follow the directions below to start an analysis at the TouchScreen (for further information see
7.4 , Analysis Setup):
1. Once an Analysis Preset has been sent to the instrument, select the Analysis Setup page
from the Touch Screen’s side menu.
2. On the Analysis Setup page load the preset that was sent to the instrument.
Page 106 of 161
NOVAtouch™ Operating Manual
INSTRUMENT OPERATION
3. Input the sample weights and sample IDs (or select “Recall” if repeating a sample), and
select the adsorbate gas port.
4. At the Touch Screen select Start from the lower right hand corner to begin the analysis.
NOTE! Selecting “Recall” will autofill the user entered fields such as sample
weight and sample ID, however it will not recall the global analysis preset used or
the station presets. Selecting “Recall” should only be done if repeating the same
samples and a macro filename has been created (thus ensuring that the previous
file will not be written over).

8.5.2.2
Starting an Analysis from TouchWin™
Follow the directions below to start an analysis using the TouchWin™ software (see E. Analysis
Parameters for further information):
1. Once the Analysis Parameters window has been completed using TouchWin™, select “Start”
to begin the analysis.
2. A second Analysis Parameters window will open; in the new Analysis Parameters window,
input: the gas type, sample ID, sample weight, description, and if running a NOVA mode
analysis select cell ID. If repeating an analysis, select “Recall” to have the software
automatically fill in all of the sample information.
NOTE! If using NOVA mode samples should be analyzed within the same cell
used when creating the cell calibration, and if applicable, using the same filler
rod.

NOTE! Samples must be in position at their desired station ports, Dewar filled
with suitable coolant and placed on the lift drive, and P0 cell and level sensor
installed before starting the analysis.

Page 107 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
9 DATA ANALYSIS USING TouchWin™
In addition to providing PC control over the measurements, TouchWin™ is also a
comprehensive program for enhanced data analysis. TouchWin™ provides features such as
graphical plots, alternative theoretical models, external data archiving, and editing capabilities.
Opening Data Files
To start using the TouchWin™ software for data analysis, first open the desired data file. To open
a data file, click the “Open Folder Icon” on the Main menu bar or click “Open” on the “File”
menu. By default, TouchWin™ stores the data files in the “Physisorb” folder.
Figure 9-1, Physisorb Folder Where Data Files are Stored
Select a file and click “Open”.
Alternatively, to reopen recent files click “Reopen” on the “File” menu or click the “Reopen
Folder Icon” on the Main menu bar.
Additionally, TouchWin™ software allows opening any known data file by simply draggingand-dropping a file from Windows® Explorer into the TouchWin™ main window. Multiple files
can be opened this way.
9.1.1
Selecting File Type
The TouchWin™ software supports three file extensions: .qcuPhysIso, .qcuPhyOverlay and
.qcuPhHoa. Click on the drop down button (left of the “File name” field) to display the
supported documents and to view the selected file type.
Page 108 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
Figure 9-2, Selecting type of file to open
Below is a list of file types that can be used with TouchWin™ software.
Table 9-1, File Types Used with TouchWin™
NOVAtouch™ File
Instrument
Extension
Access
QC PhysiSorption
NovaTouch™
.qcuPhysIso
Open
Physi Overlay
NovaTouch™
.qcuPhOverlay
New/Open
Isosteric Heats of
Adsorption
NovaTouch™
.qcuPhHoa
New/Open
Legacy Files
Instrument
Extension
Access
Quantachrome.qps
Autosorb,
Quadrasorb,
NOVAe,
Hydrosorb,
Aquadyne,
.qps
Import
Autosorb.RAW
Autosorb-1/6
.raw
Import
.drf
Import
.dat
Import
Autosorb.DRF
NOVA.DAT

9.1.2
[P/Po, volume]
Autosorb-1/6
[P, Po, vol., time]
NOVA
[P/Po, volume]
NOTE Any legacy file type that has been imported can be saved as a QC
PhysiSorption file type.
Data Processing Warnings
When opening a data file that has relative pressures (P/P0) greater than 1.0 or that has nonmonotonic points (points that do not increase in volume with increasing pressure) a warning is
displayed, see below. Data reduction is not prevented, however, some processing may not
function properly on these data files.
Page 109 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
Figure 9-3, Warning about possible problems with data
For data with relative pressures greater than one, the P0 should be checked for validity. For nonmonotonic data, the offending points may be deleted and the data saved to a new file for
processing. Usually the non-monotonic points are at very low pressures where little adsorption
has taken place and signal is in the noise range of the transducers. Deleting these points does
not affect the integrity of the data. If non-monotonic points occur at higher pressures (>0.1 P/P0)
the data should be questioned and rerun if possible.
Floating Menu
The Floating Menu is the main tool used to obtain data results. To use the contextual Floating
Menu and obtain data results, right-click on a graph or table to access the reporting options.
Tables, Graphs and Reports
Figure 9-4, Floating
Menu for Graphs,
Tables, and Reports
Page 110 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™

NOTE! The list of report options will reflect the reports that have been setup, and the
quantity of options in the Floating Menu will grow accordingly.
Editing Analysis Data Information
To confirm or change (where allowed) analysis-related information in the data file, right-click on
any graph or table and select “Edit Analysis Data” from the Floating Menu:
The resulting “Analysis Header” window contains all of the metadata and information that is
displayed in the header of each page of the report – including those which cannot be changed. If
a parameter that is used for calculating results (reduced data) is changed, such as the sample
weight, the affected results will be recalculated when “OK” is selected. Clicking “Cancel” will
disregard any changes made in this window.

NOTE! Attempting to close an edited data file without saving it, the software
will prompt to save the data before the file closes. If you do not wish to overwrite
the original information, click cancel and use the “Save As” function (with a new,
unique filename) before closing the file. Any changes to the file, including any to
the “Analysis Data” will be lost if the file is closed without using “Save” or “Save
As”.
Setting Data Reduction Tags
Some calculations rely on specific data points to be identified: those to be used (“tagged”) and
those to ignored or excluded (“untagged”). To reveal the point tags for a given data file, rightclick on an open plot or a table associated with this file to access the Floating menu, then click
“Edit Data Tags” to open the tag editor window:
Figure 9-4, Set Data Tags
Page 111 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
Refer to the “Tags” column immediately to the right of “Vol” to review which tags are currently
set for each point. The A and D tags cannot be altered since they indicate to which branch of the
isotherm they belong (Adsorption and Desorption respectively) as originally measured.
The ON/OFF/RESET area to the right serves as both legend and tag selector.
Refer to Sect. 9.6 , for guidelines regarding the appropriate relative
NOTE!
pressure ranges for each calculation.
NOTE! Not all available calculations use tags (e.g. DFT).
To make changes to points, follow the steps below:
1. To select points: click on a single point to select it (clicking on a different point
automatically deselects the first), Ctrl-click to add/remove a single point, Shift-click to
select range from last clicked item, or use the “Select All” or “Select None” buttons at the
bottom left of the Edit Data Tags window.
2. Using the selector boxes, apply tags to the selected points:
a. Check a box next to the tag to add (turn ON) or remove (turn OFF) it from the
points listed. Unchecked tags will be left in their current state.
b. To make the changes to the point(s) tag(s), click “Apply to Selected”.
c. Use the “Reset” button to deselect all tag checkboxes. The “Reset” button does
not revert the tag selections to a prior instance.
3. To delete one or more points from the data set:
a. Highlight one or more points as above
b. Click “Delete” to remove the points permanently.
4. To ignore or accept the changes made:
a. Click “Cancel” to discard all changes and return to the open file in the main
program.
b. Click “OK” to accept the changes and return to the open file in the main program.
All affected calculations will be re-evaluated, and results displayed in the open
tables and/or graphs results will be automatically updated.
NOTE! ‘P’ tags below P/P0 = 0.35 can be automatically ignored (as they should be*)
by making the appropriate selection in the Data Reduction parameters dialog. *The

option to use P tags below P/P0=0.35 is provided to allow comparison with other pore size calculations,
such as DFT.
Page 112 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
Setting Data Reduction Parameters.
Physical sample parameters on which calculations rely, and other computation options are
accessed via the “Data Reduction Parameters” (DRP) in the Floating Menu:
Figure 9-5, Data Reduction Parameters Dialog
1. Scroll down to a particular group of settings, or use the “Quick Jump” drop-down menu
for faster navigation.
2. To apply changes to the data, click “Save” to store the settings in a Physisorption Data
Reduction Parameters (*.drp) file. The files should be stored in the program’s
“TouchWinCfg”. It is recommended that you do not overwrite the pre-programmed
“default” file, but use a unique and descriptive filename.
3. Click “Load” to select a *.drp file previously created and update the data reduction
parameters.
4. To ignore or accept the changes made:
a. Click “Cancel” to discard all changes and return to the open file in the main
program.
b. Click “OK” to accept the changes and return to the open file in the main program.
All affected calculations will be re-evaluated, and results displayed in the open
tables and/or graphs results will be automatically updated.

NOTE! Results to be included in The Area/Volume Summary report are defined in the
Data Reduction Parameters window.
Page 113 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
The table below provides a general guide for the Tag selection and gives the pressure ranges in
which various calculation models are applicable.
Table 9-2, Gas Sorption Calculation Methods
P/P0 Range
Calculation Model
10-7 – 1
NLDFT, QSDFT, GCMC
0.0001 – 0.1
DR, DA
< 0.15
HK, SF
0.05 – 0.3
(Classical BET) range)
BET
Micropore BET assistant
BET for microporous samples
(Rouqerol point selection method)
> 0.15
t-plot, alpha-s, FHH, DFT
> 0.35
BJH, DH, Fractal–FHH, NK
Calculation of Heats of Adsorption
For the calculation the isosteric heat of adsorption on a sample from physisorption data, measure
at least two, preferably three, isotherms at different temperatures. To perform the calculation go
to: New on the File menu and click Isosteric Heats of Adsorption. This will open the Heats of
Adsorption window. Click Add to open and add an isotherm file to the set of isotherms to use for
the calculation of Heats of Adsorption. It is recommended to add the lowest temperature
isotherm first. Add at least two isotherms and click OK to perform the calculations and display
the results.
Review of the Data Analysis Methods
Before referring to specific calculation methods, check the Adsorbate and Adsorbent settings in
Data Reduction Parameters, to verify that the correct gas and surface chemistry are selected for
the measurement data.
Calculation “tags” are required for some of the calculations to identify the data points used.
Right click in the graph or table to bring up the Floating Menu, then click “Edit data Tags”. For
calculations that do not use tags, their settings are accessed in Data Reduction Parameters. Some
calculations (e.g. t-plot, BJH) require both tags and selections within Data Reduction Parameters.
The requirements for each calculation are outlined below.
9.2.1
BET: Multipoint
Tags : M
DRP : Adsorbate (cross sectional area)
Page 114 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
To calculate the Multi-point BET surface area for the sample, set the appropriate data points with
the M tag.
The multi-point BET surface area calculation requires a linear plot of 1/ [W (P0/P)–1] vs. P/P0,
which for most solids using nitrogen as the adsorbate is restricted to a limited region of the
adsorption isotherm. For nonporous and mesoporous materials this is usually in the P/P0 range of
0.05 to 0.35, however, this linear region is shifted to lower relative pressures for microporous
materials. The standard multipoint BET procedure requires a minimum of three points
(preferably five) in the appropriate relative pressure range. If the Y-intercept is negative and/or
the points clearly do not fall on a straight line, the point range should be adjusted as described
below. Once the desired range is selected, click “OK” to set M-tags to all adsorption points
within the highlighted range or “Cancel” to return without making any changes.
9.2.1.1
The BET Assistant (Rouquerol method)
The BET equation is certainly applicable to macro- and mesoporous solids, but, in a strict sense,
it is not applicable to microporous adsorbents. The type of porosity (i.e. macro-, meso-, or
micropores) plays an important role in the applicability of the BET equation, notwithstanding the
problems arising from the chemical and geometrical heterogeneity of a surface.
The Micropore BET Assistant [161] helps the user select points within the linear range of the
BET equation. To access the Micropore BET Assistant, select “Edit Data Tags” in the Floating
menu, then click on the “Micropore BET Assistant” button. The “Micropore BET Assistant”
window will display a plot of P/P0 vs. V(1–(P/P0)) and the recommended range for the BET
equation. The software will automatically select the range of pressures to use, but it is possible to
override the automatic selection by dragging the sliders below the graph to move the boundaries
of the BET region. Once the desired range is selected, click “OK” to set M-tags to all adsorption
points within the highlighted range or “Cancel” to return without making any changes.
9.2.1.1.1 User Actions Menu
“Unzoom”: reverts to the unzoomed display.
“Enter Values”: Pressure range boundaries can be set by entering the Start (lowest value) and
End (highest value) limits.
“Recommended”: reverts to the original recommended values.
“Currently Tagged”: reverts to the range defined by the M-tags set prior to entering the BET
Assistant.
9.2.2
BET: Single point
Tags : S
DRP : Adsorbate (cross sectional area)
Use the S tag (usually at P/P0 = 0.3) to calculate the value for the Single-point BET.
Page 115 of 161
NOVAtouch™ Operating Manual

9.2.3
DATA ANALYSIS - TouchWin™
NOTE! The single point BET calculation is not limited to just one datum point.
Langmuir Surface Area
Tags : L
DRP : Adsorbate (cross sectional area)
To calculate the Langmuir surface area for the sample, set the appropriate data points with
the L tag.
9.2.4
t-plot: Statistical Thickness Methods
Tags : T
DRP : Model type
For the t-plot method calculations, T tags must be set for the desired data points. Note that
the Carbon Black t-method calculation corresponds to ASTM D6556.
9.2.5
Total Pore Volume
Tags : V
DRP : Adsorbate (liquid density)
The Total Pore Volume calculation can be found in the Tabular Data (“Tables” portion of the
software). Assign the V tag to the desired datum point (usually the last adsorption datum point in
the isotherm, i.e. at the highest P/P0, or a point on the plateau-region of the isotherm).
NOTE! Only one V-tag.

9.2.6
Average Pore Size
Tags : V, M
DRP : Adsorbate (liquid density)
The Average Pore Size calculation can be found in the Tabular Data (“Tables” portion of the
software). The Average Pore Size calculation requires the calculation of the Multi-point BET
surface area (M tags) in addition to the V tag.
9.2.7
Tags
DRP
BJH: Barrett, Joyner & Halenda Method, DH: Dollimore Heal Method
:P
: Thickness Calculations, moving average, ignore points below P/P0 = 0.35.
Scroll or jump to the BJH/DH/Kr87 section of the “Data Reduction Parameters” window to
Page 116 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
access parameters for these methods.
The BJH and DH methods can be used to calculate pore size distribution plots and tables for the
adsorption and/or desorption branches of the isotherm. Tag the appropriate adsorption and/or
desorption data points with the P tag. This calculation uses the statistical thickness of the
adsorbed layer of the adsorbate on the surface of the sample; choose a t-method calculation
method (de Boer, Carbon Black, Halsey, or Generalized Halsey) to use in the BJH and/or DH
calculations (but do not use the T tags for these calculations).
The moving point average determines the number of adjacent, calculated pore size points
(averaged across a moving window) to smooth the calculated pore size distribution; it does not
smooth the original isotherm data points. Set this value to 1 to include all of the P-tagged data
points in the calculation without any smoothing.
Check “Ignore P tags below 0.35 P/Po” to exclude these points from the calculation.
To set interpolation ranges for the BJH/DH method calculations, click “Add” to fill in the desired
ranges in the fields provided. Click “OK” or “Cancel” to close the “Data Reduction Parameters”
window.
9.2.8
SF and HK: Saito Foley Method and Horvath Kawazoe Method
Tags : None
DRP : Tabular Data Interval
Scroll to the “SF and HK Calculations” section of the “Data Reduction Parameters” window to
access parameters for these methods. The “Tabular Data Interval” is the number of data points
that will be displayed on the data tables. For example, to tabulate every other data point a value
of 2 should be entered for the Tabulated Data Interval.

9.2.9
NOTE! The NOVAtouch™ does not have the hardware features required to
acquire the data necessary to employ the SF and HK, however, a file from an
ASiQ can be uploaded to the software to employ the SF and HK methods.
DR: Dubinin Radushkevich Method
Tags : R
DRP : β, n
Use the DR method set R Tags for the desired P/P0 points in the measurement and enter the
Affinity Coefficient (β). The value of the DR exponent (n) can be adjusted in the “Adsorbent”
section of the “Data Reduction Parameters” window. See Section 10.3.3 for more information on
the DR method.
9.2.10
DA: Dubinin Astakhov Method
Tags : R
DRP : E, n, K
For the “DA Method” enter initial values for the E and n parameters and the appropriate
Page 117 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
Interaction constant (K) for the adsorbate used. See Section 10.3.4.
9.2.11
MP Method
Tags : None
DRP : Thickness interval
Select the MP section on the Data Reduction Parameters window to access the Thickness interval
parameter. The MP method does not require any other data tags. Enter the Thickness interval
value in its field.
9.2.12
Tags
DRP
Alpha-s Method
:T
: Model type
For the alpha-s method calculations, T tags must be set for the desired data points. Select a
Standard Isotherm File from the pull-down list in Data Reduction Parameters: Alpha-S
Calculations, see below. Then select the desired P/P0 value for the calculation.
alpha–s Standard Isotherm File
Material / Experimental Conditions
_asilar
_asil1
_acarb
_aalum1
Nonporous hydroxylated silica (argon at 77K)
Nonporous hydroxylated silica (nitrogen at 77K)
Nonporous carbon (nitrogen at 77K)
Nonporous alumina (nitrogen at 77K)
9.2.13
Fractal Dimension Methods FHH: Frenkel-Halsey-Hill / NK:
Neimark Kiselev Methods
Tags : R
DRP : N/A
These two methods, Frenkel-Halsey-Hill (FHH) and Neimark Kiselev (NK) evaluate the Fractal
Dimension, D. They are intended for investigation using data acquired in the micropore region
(below a relative pressure of 1x10-3). Both of these methods can be used for the adsorption
and/or desorption branches of the isotherm. Assign R tags to the desired data points.
9.2.14
Tags
DRP
Kr(87) Thin Film Pore Size Method (Data Reduction Only)
:P
: moving average, ignore points below P/P0 = 0.35
This is a novel method for the pore size analysis of thin porous (siliceous, oxidic) films by
krypton adsorption at 87 K according to the method suggested by Thommes et al. [40].

NOTE! The NOVAtouch™ does not have the hardware features required to
acquire the data necessary to employ the Kr(87) Thin Film Pore Size method.
Page 118 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
see Powder Tech Note 39, available
 analysis,
representative or qc.support@Quantachrome.com.
NOTE! For additional information about using Krypton for thin film pore size
from your local Quantachrome
In the primary application range of this method, that is, a pore diameter range between ca. 2.5
nm and 8 nm, the obtained pore size data are traceable to NLDFT pore size data (please see
Sect.10.4 for more information). The wider application range of this Kr-87 K/thin film method
extends from ca. 0.7 nm to up to 8 nm).
In order to apply this data reduction method on Kr/87 K adsorption isotherm data (as measured
on a capable instrument, for example, Autosorb-iQ-MP).
1. Import the data file in the TouchWin™ software.
2. Select Krypton@87 as the adsorbate in Data Reduction Parameters. The file header will
not read Krypton@87K stated as the analysis gas/Adsorbate automatically. To change it,
right click to access the Floating Menu, click “Edit Analysis Data” and edit the field
labeled “Adsorbate Name”.
3. Apply "P tags" to the adsorption data points to include them in the pore size distribution
(please note that the method is not applicable to desorption data). Make sure not to include
data from potential bulk sublimation transition (indicated by a steep step in the isotherm
close to or at the saturation pressure).
4. Select Kr(87) Thin-film method (graph or table) from the Floating menu.
NOTE! A moving point average smoothing is applied to the data. The moving
point average size is set from the BJH/DH/Kr87 section of the “Data
Reduction Parameters” window.

9.2.15
Area-Volume Summary
Tags : individual methods included must have tags set
DRP : individual methods included must have data reduction parameters set/options selected
Selecting “Tables > Area-Volume Summary” from the Floating Menu generates a tabular
summary consisting of the data header information plus area and volume results calculated from
the selected models.
To select the calculations to include in the Area-Volume Summary, open the “Data Reduction
Parameters” window, using the QuickJump or scroll bar go to the “Area/Volume Summary”
section; check/uncheck those that apply. Be sure to set all the required parameters for the
selected methods.
Data Presentation
The TouchWin™ software is also used to generate graphical and tabular reports of experimental
data, and calculated results for viewing and report generation (the data file must have the
appropriate parameters i.e., specific data point tags and the necessary calculation parameters. See
9.6 Setting Data Reduction Parameters.). Tables and plots generated for a given data file can be
Page 119 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
accessed from the contextual Floating Menu. To view the list of available graphs or tables, right
click on the graph to open the Floating Menu then click “Graphs” or “Tables”. This will open the
respective submenu to select the particular item to access. In some cases this item is listed on a
lower level submenu.
If a plot or a table is selected and the data reduction parameters and/or the appropriate tags for
the calculation are not set properly, the software will display an error message. When viewing a
plot or table with the results of calculations obtained by a given method, the user can change the
number of data points or P/P0 data range (“Data Reduction Parameters “P0 Override”) via the
Floating Menu.
To modify data tags for a given data file, right-click on an open graph/table to activate the
Floating Menu. Click “Edit Data Tags” on this menu to add or remove appropriate tags in this
window.
Click “OK” when finished. This will automatically update all the open plots and tables related to
the modified data.
9.3.1
Configuring Graph Properties
To modify the features of the graphs, go to the Configure tab, Display Properties Graph
Properties. Graph properties can be set for every plot that is generated by the software. Each plot
may have different settings such as x and y axis values, data point marker style, graph colors, and
line thickness. To set values for a specific plot locate the method in the list on the left side of the
“Graph Properties” window. The selected (active) item will be highlighted.
operate the Graph Properties.
NOTE! See: Sect. 6.3.3.1, Graph Properties for more information about how to
To access this window:
 Go to: Configure tab, Display Properties Graph Properties.
 Or, click “Graph Properties” on the Floating Menu.
Page 120 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
Figure 9-6, Graph Properties
Click the “Save” checkbox before hitting “OK” to store the current settings to permanent
storage, to use these settings for all future graphs. If the user does not save the settings they will
be lost when the program is closed.
Click “OK” to return to the main window of the program, and apply all changes made. Click
“Cancel” to discard all changes made in this window.
9.3.2
Creating Overlay Plots
NOTE! An Id must be given to all plots in the overlay. See: Sect. 6.2.1
The TouchWin™ software allows the user to construct overlay plots containing different data
files.
To create a new overlay, select “File  New  Physisorption Overlay”. This will open the
explorer window prompting to choose a root file. Additional plots may be added to overlay atop
the root file by right-clicking on the root plot and accessing the overlay plot’s Floating Menu.
Right-click on the plot and select “Manage Isotherms” to access the “Overlay Manager” window.
Click “Add” and select an appropriate file from the list, click “Open”. When prompted, enter the
display label for the added data, click “OK” to view the overlay graph.
Alternately, an overlay root file can be made from any of the plots available in the software. To
make a file into a root file, right-click on the plot to overlay additional data files and select
“Create Overlay” from the Floating Menu. The user will be required to input a file name for the
overlay plot file (.qcuPhOverlay) before the overlay plot opens. After inputting the file name and
selecting “Save” an overlay plot with the name that has been designated will open automatically.
To change the graph properties, right-click on the graph, then click “Graph Properties” to open
Page 121 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
the window.
To access and modify the Data Reduction Parameters for the overlay plot, select a file from the
“Overlay Manger” window then select “Edit  Data Reduction Parameters”. To change the
label or description of a file, select the file from the list in the “Overlay Manager” window and
click “Edit  Description”. To remove a file from the graph select it on the list and click
“Remove”.
To save the overlay graph click “Save As” in the “File” menu. The graph will be saved under a
new name with the .qcuPhOverlay file type in the Physisorb Data folder.
Figure 9-7, Overlay Manager
9.3.3
Generating Custom Reports
NOTE! See also: Sect.6.3.2, Manage Reports.
To generate custom reports, go to the “Configure” tab and click “Mange Reports”. This activates
the “Manage Report Templates” window.
The files listed in this window were saved during the software installation. The list includes
sample report files (titled “demo.XXX”) which may be used or edited. To modify an existing
report file, highlight the file name and click “Edit”. The user may also create their own report file
by clicking the “Add” button. A window will appear prompting to enter a filename for the report.
Enter the filename then click “OK”. The “Report Template Editor” window will appear.
To remove unwanted report file, highlight its name and click “Delete”.

NOTE! Deleting a report file is an irreversible action. Once a report file has been
deleted, it cannot be retrieved.
Choose the desired combination of graphs, tables, and metadata (data reduction parameters and
calculation method summaries). These report items are listed in a hierarchical view grouped as
Page 122 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
follows:
Graph — available graph items further grouped by the calculation methods
Tables — tabular data lists (grouped by calculation method)
MetaData — additional text data. Including calculation result summaries, data reduction
parameters and analysis parameters
Formatting — insert a page break into the report
Figure 9-8, Report Template Editor
Highlight a selection and click “Add”, the selections will appear in the left-hand space. To
change the printing order for the report, highlight the selection and click the “Move Up” or
“Move Down” buttons. For graphs, select the size of the printed selection by right-clicking on
the “Param” field and selecting “Set Page Size” from the Floating Menu. Available choices are
“Page”, “Half Page”, and “1/3 Page”. Note that these sizes refer to the printout size of the whole
page minus the header.
Click “Remove” from the field’s Floating Menu, or “Remove >”, to remove the selection from
the report. Use the “OK” button to accept the changes, or the “Cancel” button to discard them.
NOTE! Tables do not have page size settings in the Report Manager.
The custom report can be assigned to a data file before the analysis, by requesting the report to
be automatically printed at the completion of the analysis. To assign the custom report to the
open data file, activate the contextual Floating menu by right clicking on the opened file, click
“Reports”, then click the appropriate report on the submenu. When a report is assigned to the
data file, all graphs, tables, and metadata (data reduction parameters and calculation method
summaries) will be produced by the software.
Page 123 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
NOTE! Make sure that the necessary data reduction tags have been correctly assigned
to the models that will be contained in the report. If this is not done prior to selecting
the report, an error message will appear prompting to add the required data reduction
tags.

9.3.4
Saving Tables as Text
Any table can be saved as text. Right click on the table and select “Save as text” from the
Floating Menu. This will open the “Save as text to…” window with Exports as default folder.
These text files can be imported into a spreadsheet and used in other programs for further display
and analysis. Alternatively “Export to .CSV” can be selected. This will open the “Save text as…”
window with Physisorb as the default folder. CSV files are text files with a list of data values
separated by commas.
Figure 9-9, Saving Tables as Text
9.3.5
Changing Header Information
The following is an example of a report header:
Figure 9-10, Report Header
Page 124 of 161
NOVAtouch™ Operating Manual
DATA ANALYSIS - TouchWin™
To change the information at the top of the header that is displayed on-screen and on each
printed page of the report, go to “Configure Display Properties”, then click “Report Header”
on the submenu.
Enter the header in the “Line” fields (3 lines max), add a custom logo if desired, and click “OK”
when finished.
When the new text is added Quantachrome information will remain in this portion of the Header.
Click “OK” to save changes.
Page 125 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
10 THEORY AND DISCUSSION

NOTE! While the NOVAtouch™ instrument is not equipped to perform N2/77K
adsorption analysis below a relative pressure of 1 x 10-3, its associated PC
software, TouchWin™, is capable of performing data reduction on such data
acquired on other instruments that are very-low pressure capable.
This section contains an overview of the theories on which the physisorption methods of analysis
and characterization are based. It is intended to give the reader a basic understanding of
adsorption chemistry and how it is used to characterize powders and porous materials. It is not
intended to be a comprehensive treatise on the subject. Further, while the software is capable of
performing the calculation, the NOVAtouch™ cannot provide the necessary data for pressures
below 10-3 relative pressure. For a more in-depth treatment of the subject, the reader is referred
to “Characterization of Porous Solids and Powders: Surface Area, Pore Size, and Density” [40].
Surface Area
The Brunauer-Emmett-Teller [1] (BET) method is the most widely used procedure for the
determination of the surface area of solid materials and involves the use of the following
equation:
1
𝑃
𝑊 (( 𝑃𝑜 ) − 1)
=
1
𝐶−1 𝑃
+
( )
𝑊𝑚 𝐶 (𝑊𝑚 𝐶) 𝑃𝑜
(1).
where W is the weight of gas adsorbed at a relative pressure, P/Po, and Wm is the weight of
adsorbate constituting a monolayer of surface coverage. The term C, the so-called BET C
“constant”, is related to the energy of adsorption in the first adsorbed layer and consequently its
value is an indication of the magnitude or strength of the adsorbent/adsorbate interactions.
10.1.1
Multipoint BET Method
The BET equation ( 1) requires a linear plot of 1/[W(Po/P)-1] vs. P/Po which for most solids,
using nitrogen as the adsorbate, is restricted to a limited region of the adsorption
isotherm, usually in the P/Po range of 0.05 to 0.35; this linear region being shifted to lower
relative pressure values for microporous materials. A typical BET plot is shown in Figure 10-1.
Page 126 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
Figure 10-1, Typical BET Plot
The standard multipoint BET procedure requires a minimum of three points in the
appropriate relative pressure range. The weight of a monolayer of adsorbate Wm can then be
obtained from the slope s and intercept i of the BET plot. From equation (1):
𝑠=
𝐶−1
𝑊𝑚 𝐶
(2).
𝑖=
1
𝑊𝑚 𝐶
(3).
and
Thus, the weight of a monolayer, Wm, can be obtained by combining equations (2) and (3):
𝑊𝑚 =
1
𝑠+𝑖
(4).
The second step in the application of the BET method is the calculation of the surface area.
This requires knowledge of the molecular cross-sectional area Acs of the adsorbate molecule.
The total surface area of the sample, St, can be expressed as:
𝑆𝑡 =
𝑊𝑚 𝑁𝐴𝑐𝑠
𝑀
(5).
where N is Avogadro’s number (6.0221415 × 1023molecules/mol) and M is the molar
mass (molecular weight) of the adsorbate. Nitrogen is the most widely used gas for
surface area determinations since it exhibits intermediate values for the C constant (50–250)
on most solid surfaces, precluding either localized adsorption or behavior as a two
dimensional gas. Since it has been established [2,3] that the C constant influences the value of
the cross-sectional area of an adsorbate, the acceptable range of C constants for nitrogen
makes it possible to calculate its cross-sectional area from its bulk liquid properties. For
the hexagonal close-packed nitrogen monolayer at 77 K, the accepted cross-sectional area Acs
Page 127 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
for nitrogen is 16.2 Å2.
The specific surface area S of the solid can be calculated from the total surface area St and
the sample weight ws, according to equation (6):
𝑆=
10.1.2
𝑆𝑡
𝑤𝑠
(6).
Single Point BET Method
For routine measurements of surface areas, a simplified procedure may be applied, using only
a single point on the adsorption isotherm in the linear region of the BET plot. For nitrogen, the
C-value is often sufficiently large to warrant the assumption that the intercept in the BET
equation is zero. Thus, the BET equation (1) reduces to:
𝑊𝑚 = 𝑊(1 − 𝑃⁄𝑃𝑜 )
(7).
By measuring the amount of nitrogen adsorbed at one relative pressure (preferably near P/Po
= 0.3) the monolayer capacity, Wm, can be estimated using and the total surface area, St, can then
be obtained in the similar way to (5), that is:
𝑆𝑡 =
10.1.3
𝑊𝑁𝐴𝑐𝑠
𝑃
(1 − )
𝑀
𝑃𝑜
(8).
Multipoint/Single Point Comparison
The relative error introduced by the single point versus the multipoint method for
determining surface area is a function of the BET C constant and the relative pressure used. The
magnitude of the error in the single point method can be determined from a comparison of
the monolayer weight obtained from the BET equation (1) and the single-point equation(7).
Solving equation (1) for Wm gives:
𝑃𝑜
1
1 𝑃
𝑊𝑚 = 𝑊 ( − 1) + 𝐶 − ( )
𝑃
𝐶
𝐶 𝑃𝑜
(9).
Rewriting the single point equation (7), and where W’m denotes the single point value, gives:
𝑃𝑜
𝑃
𝑊𝑚′ = 𝑊 ( − 1) ( )
𝑃
𝑃𝑜
(10).
The relative error inherent in the single point method, then, is:
𝑃
1 − (𝑃 )
𝑊𝑚 − 𝑊𝑚′
𝑜
= 𝑊
𝑃
𝑊𝑚
1 + [𝑃 (𝐶 − 1)]
(11).
𝑜
Equation (12) indicates that for a given C value, the relative error decreases with increasing
Page 128 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
relative pressure. Therefore, a relative pressure as high as possible, yet still in the linear region
of the BET plot, should be chosen for single point surface area determinations. For all
except microporous samples a P/Po of about 0.3 is preferable. For single point determinations
on microporous samples a relative pressure as high as possible on the linear BET plot should be
chosen.
𝑠
𝐶 =( )+1
𝑖
(12).
Table 10-1 gives the relative error for various C values calculated from equation (13) using P/Po
of 0.3. For example, when the C constant is 100, a two percent error is indicated.
Table 10-1
Single Point/ Multipoint Comparison
C
Relative Error
1
0.70
10
0.19
50
0.04
100
0.02
1000
0.002
∞
0
Prior to using the single point method for the determination of surface area, the C constant can be
evaluated from a multipoint BET plot. That is, where s and i are the slope and y-intercept,
respectively, of the BET plot. Subsequently, the single point method can be used on materials
having the same composition. For greater accuracy, if the C constant is known, the single point
result may be corrected using equation (12).
10.1.4
Langmuir Method
In the absence of meso and/or macropores, a sample containing micropores will exhibit a Type
I, sometimes referred to as a Langmuir isotherm (see Sect. 10.2.1). The Langmuir equation
(13) is a limiting case of the BET equation (1) i.e. the adsorption of a single molecular layer of
adsorbate:
𝑃
𝐶 (𝑃 )
𝑊
𝑜
=
𝑊𝑚 1 + 𝐶 ( 𝑃 )
𝑃
(13).
𝑜
W and Wm are the weight of adsorbate at some P/Po and the weight in a monolayer, respectively.
C is a constant associated with the energy of adsorption. Equation (13), rewritten in the form of
Page 129 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
a straight line:
𝑃
(𝑃 )
𝑃
(𝑃 )
1
=
+ 𝑜
𝑊
𝐶𝑊𝑚 𝑊𝑚
(14).
𝑜
allows the determination of the slope (1/Wm) from a plot of (P/Po)/W versus P/Po.
The weight of a monolayer Wm may then be used to calculate the total surface area of the sample
from equation (5). This method is not always applicable to composite materials containing
micropores and meso- and/or macropores. In any case, the limiting uptake of gas by a
microporous solid is not necessarily due to the formation of a monolayer, but rather a micropore
volume filling limit, and the calculated surface area can be significantly in error.
Pore Size by Gas Adsorption
It is expedient to characterize pores according to their sizes:
• Pores with openings exceeding 50nm in diameter are called “macropores”.
• The term “micropores” describes pores with diameters not exceeding 2nm.
• Pores of between 2-50nm are called “mesopores”, and to classify them according to their
geometry (shape).
10.2.1
Isotherms
The classification of isotherm shapes is useful in understanding the pore structure of the solid
under investigation. IUPAC’s updated proposals by Thommes et al. [45] are briefly described
below.
In the case of physical adsorption, sorption isotherms obtained on microporous materials are
often of (reversible) Type I. The micropore filling and therefore high uptakes are observed at
relatively low pressures, because of the narrow pore width and the high adsorption potential.
The limiting uptake (plateau) being governed by the accessible micropore volume rather than by
the internal surface area.
Type II sorption isotherms are typically obtained in case of non-porous or macroporous
adsorbent, where unrestricted monolayer-multilayer adsorption can occur. The inflection point
or knee of the isotherm (“point B”) indicates the stage at which monolayer coverage is complete
and multilayer adsorption begins.
The reversible Type III isotherm has no point B indicating that the attractive adsorbateadsorbent interactions are relatively weak but that the adsorbate-adsorbate interactions play the
most important role. Isotherms of this type are not common.
Type IV isotherms are typical for mesoporous materials, the most characteristic feature of
Type IV(a) being the hysteresis loop which is associated with the occurrence of pore
condensation. Limited uptake over a range of high P/P0 (a plateau) indicates complete pore
filling. Type IV(b) without hysteresis is associated with a mesopore network of narrow width
below a critical diameter (which depends on both adsorbate and temperature) and by conical
mesopores closed at the tapered end. The initial part of the Type IV can be attributed to
Page 130 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
Amount Adsorbed
monolayer-multilayer adsorption as in case of the Type II isotherm.
Relative Pressure (P/P0)
Figure 10-2, IUPAC Classification of Sorption Isotherms [45]
Type V isotherms exhibit pore condensation and hysteresis. However, in contrast to Type IV
the initial part of this sorption isotherm is related to adsorption isotherms of Type III,
indicating relatively weak attractive interactions between the adsorbent and the adsorbate.
The Type VI isotherm is a special case, which represents stepwise multilayer adsorption on a
uniform, non-porous surface [42], particularly by spherically symmetrical, non-polar
adsorptives. The sharpness of the steps depends on the homogeneity of the adsorbent surface,
the adsorptive, and the temperature.
Furthermore, it is widely accepted that there is a correlation between the shape of the hysteresis
loop and pore network effects and fluid metastability giving rise to delayed condensation and
evaporation.
Briefly, type H1 is found in materials which exhibit a narrow range of uniform mesopores. As
such, pore network effects are minimal and the steep, narrow loop is indicative of delayed
condensation on the adsorption branch.
The “triangular” hysteresis loops of type H2 are given by more complex pore structures in which
network effects are important. The characteristic very steep desorption branch can be attributed
either to pore-blocking/percolation in a narrow range of pore necks or to cavitation-induced
evaporation. Type H2(b) is also associated with pore blocking, but the size distribution of neck
widths is much larger.
Page 131 of 161
NOVAtouch™ Operating Manual
Amount Adsorbed
THEORY AND DISCUSSION
Relative Pressure (P/P0)
Figure 10-3, Hysteresis Types
Type H3 has two distinctive features: (i) the adsorption branch resembles a Type II isotherm, and
(ii) the lower limit of the loop is normally located at the P/P0-limiting cavitation behavior. These
isotherms are typically produced by (non-rigid) aggregates of plate-like particles, but also if the
pore network consists of macropores that are not completely filled with condensate.
H4 is somewhat similar, with the adsorption branch being a composite Type I / Type II, the more
pronounced uptake at low P/P0, compared to H3, being associated with the filling of micropores
in mixed micro-mesoporous systems.
Distinctive type H5 is associated with certain pore structures containing both open and partially
blocked mesopores. The common feature of H3, H4 and H5 is the sharp step down in desorption
where the loop closes. Generally, this is located in a narrow range of P/P0 for the particular
adsorptive and temperature (e.g., at P/P0 ∼ 0.4 – 0.5 for nitrogen at 77 K).
These hysteresis loops impose a difficulty on the pore size analysis… should the adsorption or
desorption branch be taken for calculation of the pore size distribution curve? The answer
depends very much on the reason(s) which caused the hysteresis.
10.2.2
Total Pore Volume and Average Pore Radius
The total pore volume is derived from the amount of vapor adsorbed at a relative pressure close
to unity, by assuming that the pores are then filled with liquid adsorbate. For a discussion of the
relationship between pore size and relative pressure, see Sect. 10.3.2. If the solid contains no
macropores, the isotherm will remain nearly horizontal over a range of P/P0 values approaching
unity. In this case, the limiting adsorption can be identified reliably with the total pore volume.
However, in the presence of macropores the isotherm rises rapidly near P/P0 = 1 and, in the limit
Page 132 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
of large macropores, can exhibit an essentially vertical rise and so it is not practical to evaluate
the data in terms of a true total pore volume. Nevertheless, a summed – or cumulative – pore
volume of all pores smaller than a certain size (as corresponds to the point on the isotherm
chosen for the calculation) can be determined.
The volume of nitrogen adsorbed (Vads) can be converted to the volume of liquid nitrogen (Vliq)
contained in the pores using equation (15).
That is,
𝑉𝑙𝑖𝑞 =
𝑃𝑎 𝑉𝑎𝑑𝑠 𝑉𝑚
𝑅𝑇
(15).
in which Pa and T are ambient pressure and temperature respectively, and Vm is the molar volume
of the liquid adsorbate (34.76 cm3/mol for nitrogen at 77.4K). Alternatively,
𝑉𝑙𝑖𝑞 =
𝑉𝑎𝑑𝑠 𝑀𝑊
∙
𝑉𝑀′
𝜌
(16).
where 𝑉𝑀′ is the molar volume of gas (22,414 cm3 at STP), MW is the molecular weight (mass)
of the gas (N2 = 28.0134), and ρ is the density of liquid adsorbate at the temperature of the
analysis (0.806 g/cm3 for nitrogen at 77.4K).
The average pore size can then be estimated from the pore volume. For example, assuming
cylindrical pore geometry (Type H1 hysteresis), the average pore radius rp is expressed by:
𝑟𝑝 =
2𝑉𝑙𝑖𝑞
𝑆
(17).
where Vliq is obtained from equation (15) and S is the BET surface area. For other pore
geometries knowledge of the shape of the hysteresis in the adsorption/desorption isotherm is
required.
10.2.3
Pore Size Distributions (Mesopore)
The distribution of pore volume with respect to pore size is called a pore size distribution. It is
generally accepted that the desorption branch of the isotherm is more appropriate than the
adsorption isotherm for evaluating the pore size distribution of an adsorbent. Data points on the
desorption branch of the isotherm, corresponding to the same volume of gas as adsorption points,
occur at lower relative pressures, i.e. at a lower free energy state. Thus, the desorption isotherm
is closer to true thermodynamic equilibrium. However, in certain cases, samples exhibiting Type
H2 hysteresis for example, the adsorption isotherm is recommended for pore size distribution
determinations.
TouchWin™ software has the capability to use either branch of the isotherm for the calculation.
Since nitrogen has been used extensively in gas adsorption studies, it has been well characterized
and serves as the most common adsorbate for pore size distribution measurements. Therefore,
the following discussion will apply to the use of nitrogen as the adsorbate.
Historically, mesopore size calculations are made assuming cylindrical pore geometry starting
Page 133 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
with the Kelvin equation (18) in the form
𝑟𝐾 =
−2𝛾𝑉𝑚
𝑃
𝑅𝑇𝑙𝑛 (𝑃 )
𝑜
(18).
where:
γ = the surface tension of nitrogen at its boiling point (8.85 ergs/cm2 at 77 K).
Vm = the molar volume of liquid nitrogen (34.7 cm3/mol).
R = gas constant (8.314x107 ergs/deg/mol).
T = boiling point of nitrogen (77 K).
P/P0 = relative pressure of nitrogen.
rK = the Kelvin radius of the pore.
Using the appropriate constants for nitrogen, equation (18) reduces to
𝑟𝐾 (Å) =
4.15
𝑃
𝑙𝑜𝑔 (𝑃 )
𝑜
(19).
The Kelvin radius rK is the radius of the pore in which condensation occurs at a relative pressure
of P/Po. However, prior to condensation some adsorption has already taken place on the walls
of the pore and so rK does not represent the actual pore radius. Conversely, during desorption an
adsorbed layer remains on the walls when evaporation occurs. The actual pore radius rp is given
by:
𝑟𝑝 = 𝑟𝐾 + 𝑡
(20).
where t is the thickness of that adsorbed layer. An empirical method for estimating t was
proposed by de Boer [6] in the form of equation (21) and is also used for pore size distribution
calculations in the TouchWin™ software.
0.5
13.99
𝑡(Å) = [
]
𝑃𝑜
𝑙𝑜𝑔 ( 𝑃 ) + 0.034
(21).
Other expressions for t calculations, available through the TouchWin™ software, are presented
in the Section 10-3.
TouchWin™ computes the pore size distribution using the methods proposed by Barrett,
Joyner, and Halenda [7] (BJH) and by Dollimore and Heal [8] (DH) as described in the
following sections.
Page 134 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
10.2.3.1 BJH Method
Using the starting point for the calculation, relative pressure (P/P0)1 close to unity, we can
assume all pores (of interest) are filled with liquid. The largest pore of radius rp1 has a
physically adsorbed layer of nitrogen molecules of thickness t1. Inside this thickness is an inner
capillary with radius rK from which evaporation takes place as P/P0 is lowered. The
relationship between the pore volume Vp1 and the inner capillary (Kelvin) volume VK is given
by:
2
(22).
2
V p1 = V K1 r p1 / r K1
When the relative pressure is lowered from (P/P0)1 to (P/P0)2 a volume V1 will desorb from
the surface. This liquid volume V1 represents not only emptying of the largest pore of its
condensate but also a reduction in the thickness of its physically adsorbed layer by an
amount Δt1. Across this relative pressure decrement the average change in thickness is Δt1/2.
The pore volume of the largest pore may now be expressed as:
V p1


r p1
= V 1 

 r K1 +  t 1 / 2 
(23).
2
When the relative pressure is again lowered to (P/P0)3 the volume of liquid desorbed includes
not only the condensate from the next larger size pores but also the volume from a second
thinning of the physically adsorbed layer left behind in the pores of the largest size. The
volume Vp2 desorbed from pores of the smaller size is given by:
V p2


r p2
= 

 r K2 +  t 2 / 2 
2
V 2
- V t 2

(24).
An expression for VΔt2 is:
V t 2 =  t 2 Ac1
(25).
where Ac1 is the area exposed by the previously emptied pores from which the physically
adsorbed gas is desorbed. Equation (26) can be generalized to represent any step of a stepwise
desorption by writing it in the form:
𝑛−1
(26).
𝑉∆𝑡𝑛 = ∆𝑡𝑛 ∑ 𝐴𝑐𝑗
𝑗=1
The summation in the equation above is the sum of the average area in unfilled pores down
to, but not including, the pore that was emptied in the desorption. Substituting the general value
for VΔt2 into equation (24) results in an exact expression for calculating pore volumes at various
relative pressures.
Page 135 of 161
NOVAtouch™ Operating Manual
V pn
THEORY AND DISCUSSION

r pn
= 
 r Kn +  t n /


2 
2

 V n -  t n


n -1

j=1

Ac j 

(27).
Since the exposed area (Ac) for any given emptied pore is not a constant but varies with each
decrement of P/P0, due to thinning of the adsorbed layer; this term must be evaluated. However,
the true area of each pore Ap is a constant and can be calculated from the pore volume,
assuming cylindrical pore geometry. That is:
𝐴𝑃 =
2𝑉𝑃
𝑟𝑃
(28).
Then the pore areas can be cumulatively summed so that for any step in the desorption
process Ap is known. The BJH method offers a means of computing ∑Acj from Ap for each
relative pressure decrement.
It is assumed that all pores emptied of their condensate during a relative pressure decrement have
an average radius 𝑟̅𝑐 calculated from the Kelvin equation (18) radii at the upper and lower values
of P/Po in the desorption step. The average capillary (core) radius is expressed as:
𝑟̅𝑐 = 𝑟̅𝑃 − 𝑡𝑟̅
(29).
Where 𝑡𝑟̅ is the thickness of the adsorbed layer at the average radius in the interval in the
current pressure decrement and is calculated from equation (21). The term “c” in equation (26)
then is given by:
r - tr
c = rc = p
rp
rp
(30).
Equation (27) now can be used in conjunction with equation (30) as an exact expression for
the computation of pore size distributions.
10.2.3.2 DH Method
A computationally simpler approach to evaluating mesopore size distributions was developed
by Dollimore and Heal [8]. The DH approach differs from the BJH method in that the term
Vt is calculated from:
n
𝑉∆𝑡𝑛 = ∆𝑡𝑛 Σ𝐴𝑃 − 2𝜋𝑡𝑛 ∆𝑡𝑛 Σ𝐿𝑃
(31).
where, the summations ΣAp and ΣLp represent the areas and lengths, respectively, of all the
pores emptied of condensate in previous desorption steps. Assuming cylindrical pore geometry,
the cumulative pore areas and lengths can be estimated for each desorption step by
summing the expressions:
Page 136 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
2𝑉𝑃
𝑟𝑃
(32).
𝐴𝑃
2𝜋𝑟𝑃
(33).
𝐴𝑃 =
and
𝐿𝑃 =
respectively.
Micropore Analysis
Several different approaches to micropore analysis are available using TouchWin™. While no
single treatment is applicable to all situations, enough flexibility is provided for the user to
select the analysis most suitable for a given situation or material.
10.3.1
V-t Method
The TouchWin™ software can use any of the t-methods of Halsey [9] the generally preferred one
of deBoer [10] and the STSA (carbon black) model [11] for the determination of micropore
volume in the presence of mesopores. This method involves the measurement of nitrogen
adsorbed by the sample at various pressure values. The procedure is the effectively the same as
that employed in the BET surface area measurement, but it extends that pressure range to
moderately higher pressures to permit calculation of the “external” surface area (external to
micropores that is), that is, the non-microporous part of the material (which includes the
contribution to total area by mesopores and macropores as well as the outermost convex particle
surface). A t-plot is simply a graphical plot of the volume of gas adsorbed versus t, the statistical
thickness of an adsorbed film, for each P/P0 datum point used in the calculation
In the TouchWin™ software, the t values are calculated as a function of P/P0 using either
the de Boer equation:
0.5
(34).
13.99
𝑡(Å) = [
]
𝑃
𝑙𝑜𝑔 ( 𝑃𝑜 ) + 0.034
the Carbon Black equation:
𝑃 2
𝑃
Å
𝑡( ) = 0.88 ( ) + 6.45 ( ) + 2.98
𝑃𝑜
𝑃𝑜
(35).
or the Halsey equation which, for nitrogen adsorption at 77 K can be expressed as:
𝑡(Å) = 3.54 [
5
1⁄
3
(36).
]
𝑃
2.303 𝑙𝑜𝑔 ( 𝑃𝑜 )
Page 137 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
or in a generalized Halsey form (found useful for other adsorbates and/or temperatures) as:
1
𝑡(Å) = 𝑎 [
]
𝑃𝑜
𝑙𝑛 ( 𝑃 )
1⁄
𝑏
(37).
Where the pre-exponential term, a, and the exponential term, b, are 6.0533 and 3.0 for
nitrogen adsorption at 77 K, respectively.
Typical t-plots are shown in Figure 10-4, Figure 10-5, and Figure 10-6 below, representing
various possible pore sizes. First t-plot of a sample having no micropores, as evidenced by the
ability to extrapolate the line to the origin, since the slope represents the total surface area St of
all the pores, that is:
𝑆𝑡 =
𝑆𝑇𝑃
𝑉𝑎𝑑𝑠
∙ 15.47
𝑡(Å)
(38).
VADS
(STP)
𝑆𝑇𝑃
where 𝑉𝑎𝑑𝑠
is the volume of gas adsorbed, corrected to standard conditions of temperature and
pressure, and the constant 15.47 represents the conversion of the gas volume to liquid volume.
Thickness (Å)
Figure 10-4, t-Plot of a Mesoporous Materia
Page 138 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
Thickness (Å)
VADS
(STP)
Figure 10-5, t-Plot of a Microporous Sample
Thickness (Å)
Figure 10-6, t-Plot of a microporous material
Using the slope, s, of the t-plot (as shown) above, equation (36) reduces to:
𝑆𝑡 (𝑚2 ⁄𝑔) = 𝑠 ∙ 15.47
(39).
In the absence of micropores, there is good agreement between the t-area, St, and the surface area
determined by the BET method. When micropores are present, the t-plot will exhibit a positive
intercept. t-Plots for two types of microporous materials are shown in Figure 10-5 and Figure 10Page 139 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
6. Figure 10-5 is a t-plot of a sample with pore openings smaller than 7–8 Å in diameter.
Quantitative pore volume data from t-plots cannot be obtained below t = 3.5 Å, corresponding to
pores smaller than 7 Å wide, since this t value represents the diameter of a nitrogen molecule.
The intercept, i, in the t-plot, when converted to a liquid volume, gives the micropore volume,
VMP, that is:
𝑉𝑀𝑃 = 𝑖 ∙ 1.547 × 10−3
(40).
The two linear regions of the t-plot in Figure 10-6 indicate the presence of micropores larger than
7Å and the actual pore width (2t) can be estimated at the position where the two linear plots
intersect. The slope of the upper linear portion (B) of the t-plot in Figure 10-6 gives the
mesopore surface area using equation (34), while the slope of the lower linear portion (A)
represents the total surface area of all pores. Point C indicates the presence of micropores of
about 10Å width (2t).
𝑆𝑀𝑃 = 𝑆𝐵𝐸𝑇 − 𝑆𝑡−𝑝𝑙𝑜𝑡
(41).
The linear BET region for microporous materials generally occurs at relative pressures less than
0.1. The linear t-plot range will be found at higher relative pressures and is dependent on the size
distribution of micropores. The micropore surface area, SMP, is the difference between the BET
surface area and the external surface area St-plot from the t-plot.
10.3.2
Alpha-s (αs) Method
An empirical analogue to the t-method was proposed by Sing [12], who pointed out that
comparing a given isotherm to a standard curve does not require invoking the concept of a
statistical thickness, t (which in turn depends on the BET surface area of nonporous reference
materials), see equation (138). Instead, similar trends and conclusions to those derivable from tplot figures can be reached by replacing the BET area as a normalization factor by the amount
adsorbed at some arbitrarily chosen relative pressure, usually P/P0 = 0.4. Therefore, the Alpha-s
method in principle allows a more direct comparison between actual and nonporous reference
isotherms. In practice, complications arise because the Alpha-s method is found to depend on the
exact nature of the material chosen as nonporous reference. Nonetheless, the Alpha-s method (by
analogy with the t-method) has found use for the determination of micropore volumes by
extrapolation of Alpha-s curves to αs = 0, and for the determination of nonporous surface area
contributions (from the slopes of linear portions of Alpha-s curves, since the ratio of actual to
reference BET surface areas is equal to the ratio of actual to reference slopes in these Alpha-s
plot regions), by qualitative comparison of actual isotherms with judiciously selected reference
isotherms
10.3.3
MP Method
An extension of de Boer’s t-method for micropore analysis was proposed by Mikhail,
Brunauer, and Bodor [13]. This MP method uses the fact that t can be calculated independently
of the solid by:
Page 140 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
𝑡(Å) =
104 ∙ 𝑉𝑙𝑖𝑞
𝑆𝐵𝐸𝑇
(42).
A Vliq vs. t plot is constructed from t vs. P/Po data for a sample with a similar BET C value, as
shown in Figure 10-7 below, using linear slopes constructed for t-value intervals from the origin
to 4 Å, 4 to 4.5 Å. 4.5 to 5 Å, etc.
Using the equation above, micropore surface areas can be calculated from the slopes and each
successive interval calculation represents the area of all the micropores remaining unfilled. The
calculations are continued until no further decrease in slope is found in the V vs. t plot,
indicating that all the micropores have been filled. The surface area of pores in the range of
thickness from 4 to 4.5 Å, for example, is the difference between the values calculated from the
first and second slopes. The area of pores in the thickness range from 4.5 to 5 Å is the difference
between the values calculated from the second and third slopes, and so on.
Pore volumes can similarly be calculated, using the relation
V = 10-4 S 1 - S 2   t1 +2 t 2  cm3 g -1
(43).
Where:
S1 = surface area calculated from slope 1
S2 = surface area calculated from slope 2
t1 = thickness at beginning of interval used for slope 2
t2 = thickness at end of interval used for slope 2
Each succeeding group of pores is correspondingly treated, and a distribution of pore volume is
thus obtained. While the exact pore shape is usually unknown, cylindrical pores are generally
assumed. It has been shown that this pore volume relation is equally valid for cylindrical pores or
parallel plates.
Page 141 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
Figure 10-7, Typical V vs. t-Plot with Internal Slopes
10.3.4
Dubinin-Radushkevich (DR) Method
Based on the Polanyi potential theory of adsorption [14] Dubinin and Radushkevich [15]
postulated that the fraction of the adsorption volume V occupied by liquid adsorbate at various
adsorption potentials can be expressed as a Gaussian function:
𝐴 2
𝑉 = 𝑉0 𝑒𝑥𝑝 [− (
) ]
𝛽𝐸0
(44).
where A is the free energy of adsorption, which in the early Dubinin’s works was called
adsorption potential 
𝑃𝑜
𝐴 = 𝜀 = 𝑅𝑇𝑙𝑛 ( )
𝑃
(45).
In equation (42) V0 represents micropore volume, E0 is the so-called characteristic energy of
adsorption and β is the affinity coefficient which can be approximated [16] by a ratio of the
liquid molar volumes v of a given adsorbate and benzene used as the reference liquid:
𝛽=
𝑣
𝑣𝐶6 𝐻6
(46).
Equation (42) can be written in the following linear form:
𝑅𝑇 2
𝑃𝑜 2
(𝑉
)
𝑙𝑜𝑔10 𝑉 = 𝑙𝑜𝑔10 0 − 2.303 (
) 𝑙𝑜𝑔10 ( )
𝛽𝐸0
𝑃
(47).
which shows that micropore volume V0 and E0 parameter can be calculated from the linear fit of
the isotherm data plotted as log(V) vs. [log(P0/P)]2. Intercept of the fitted straight line gives
Page 142 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
log(V0) while its slope m can be used to calculate E0.
(48).
2.303 ∙ 𝑅𝑇
𝐸0 = √
𝑚𝛽
The linear range for these plots is usually found at relative pressures of less than 10-2.
Based on empirical studies Dubinin and Stoeckli proposed [17] that E0 can be related to the
characteristic micropore width, PW, for a carbonaceous adsorbent by the following simple
formula
26(𝑘𝐽 𝑛𝑚 𝑚𝑜𝑙 −1 )
𝑃𝑊 =
𝐸0
(49).
The linear form of the DR equation was also used by Kaganer [18] to evaluate micropore
surface area from the plot intercept, log(V0), by applying equation (5).



log W = log(W0 ) - m log  K P0 
P



(50).
2
Where:
W = weight adsorbed at P
W0 = total weight adsorbed
m =slope of straight line
K =
 Pc   T 
   
 P0   T c 
2
Pc = critical pressure of adsorbate (mm Hg)
Po = saturated vapor pressure of adsorbate (mm Hg) T = adsorption temperature (K)
Tc = critical temperature of adsorbate (K)
Other modifications [19] of the DR method include:
a) Allowing the DR exponent, n, to differ from n=2 in order to provide a better data fit for
non-Gaussian pore size distributions, see next section; and
b) Introduction of a supercritical adsorption constant, K, into equation (45)
Equation (48) has been proposed as a valid correlation for adsorption data at temperatures
exceeding the critical temperature of the adsorbate.
10.3.5
Dubinin-Astakhov (DA) Method
For a large number of microporous materials, the adsorption isotherm can be well characterized
by the Dubinin-Radushkevich equation. However, for those microporous materials with
Page 143 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
heterogeneous distributions or strongly activated carbons, the Dubinin-Radushkevich equation
fails to linearize the adsorption data [20]. To describe adsorption on wider range of microporous
materials the Dubinin-Astakhov equation was proposed:
  - RT ln P / P0 n
W = W 0 exp  - 

E

 



(51).
where:
W = weight adsorbed at P/Po and T
W0 = total weight adsorbed
E = characteristic energy
n = non-integer value (typically between 1 and 3)
Whereas, equation (50) is a generalized form of the Dubinin-Radushkevich equation (n=2) and
has been found to fit adsorption data for heterogeneous micropores [21].
The Dubinin-Astakhov equation requires the parameters n and E to be calculated re-iteratively
by non-linear curve fitting to the adsorption isotherm in the low relative pressure, micropore
region. The values of n and E obtained are then used in equation (50), [22]:
d w / w0
dr
 = 3n
n
  K n - 3n
 K  -  3n +1 
exp  -   r

 r
E 
  E 



(52).
where
r = pore radius
K = interaction constant = 2.96 kJ x nm3 x mol-1 (N2), = 2.34 kJ x nm3 x mol-1 (Ar)
A plot of [d(w/w0)]/dr vs. r yields the DA Method pore size distribution.
10.3.6
Horvath-Kawazoe (HK) Method
The HK method [23] enables the calculation of pore size distribution of micropores from the
low relative pressure region of the adsorption isotherm. Many pore size distribution
methods are derived from the Kelvin equation, which describes the phenomenon of capillary
condensation. Some have questioned the reliability of the capillary condensation approach in
the small confines of micropores. The HK method is derived independent of the Kelvin
equation.
The HK method expresses the adsorption potential function within slit-like micropores as a
function of the effective pore width:
Page 144 of 161
NOVAtouch™ Operating Manual

RT ln 

THEORY AND DISCUSSION

 (53).


4
10
4
10
P
N S AS + N A AA






x
+
 = K
9
3
9
  d 3
4  -d 
P0 
d

d 
d  
9  - 
3  
9   
3   - 
2

2
2 
  2
The parameters AS, and AA can be calculated using the following equations:
AS =
AA =
6 mc2  s  A
(54).
3 mc2  A  A
2
(55).
s +  A
s
A
where,
m = mass of an electron
c = speed of light
s = polarizability of adsorbent
A = polarizability of adsorptive
s = magnetic susceptibility of adsorbent
A = magnetic susceptibility of adsorptive and
(  - ds) = effective pore width
d = ds + dA
ds = diameter of adsorbent molecule
dA = diameter of adsorbate molecule
l = distance between two layers of adsorbent
σ = 0.858d/2
K = Avogadro’s number
NS = number of atoms per unit area of adsorbent
NA = number of molecules per unit area of monolayer of adsorbate
AA = Kirkwood-Mueller constant of adsorptive
AS = Kirkwood-Mueller constant of adsorbent
By selecting effective pore widths in the micropore range, equation (51) can be used to
calculate the corresponding relative pressures. From the adsorption isotherm, the amount
of adsorption at each of these relative pressures is determined. Differentiation of weight (or
Page 145 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
volume) of gas adsorbed relative to the total uptake, W/W0, with respect to the effective
pore width yields a pore size distribution in the micropore range.
10.3.7
Saito-Foley (SF) Method
Even though the HK method is adequate for materials with a predominance of slit-like
pores (activated carbons, layered clays), certain solids (e.g. zeolites) are better represented
assuming cylindrical pore geometry. Hence, the SF method [24] was developed as an
alternative to the slit-like pore-based HK method. As with the HK method, the SF method
enables the calculation of pore size distributions of microporous materials independently
from the Kelvin Equation. The computational approach is analogous to that of the HK
method, except that cylindrical pore geometry is assumed. Accordingly, equation (51) is
replaced by
RT ln
P
P0
=
3K
x
4

 d 
G =  1- 

 D 


2k
N S AS + N A AA
 d /2  4


x

1

(56).
  k + 1  G
k=0
10
4
  21 
 d 
 d  
 ak 
 - bk 
 
 
 D 
 D  
  32 
(57).
where
  1.5  k 
ak  

k


(58).
2
and a0 = b0 = 1
(D-ds) = effective pore diameter
All of the other parameters have been defined in the previous section (HK method). The
numerical solution of the above equations is again analogous to that of the HK method,
and yields pore size distributions and cumulative pore volumes for cylindrical pores in the
micropore range.
Density Functional Theory and Monte Carlo Simulation
Methods
Classical macroscopic, theories like for instance the Dubinin-Radushkevich approach, the BJH
method, and semi empirical treatments such as those of Horvath and Kawazoe (HK), and Saito
and Foley do not give a realistic description of the filling of micropores and even narrow
mesopores. This leads to an underestimation of pore sizes.
In order to achieve a more realistic description microscopic theories, which describe the sorption
Page 146 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
and phase behavior of fluids in narrow pores on a molecular level, are necessary. Treatments
such as the Density Functional Theory (DFT) or methods of molecular simulation (Monte Carlo
simulation (MC) and Molecular Dynamics (MD)) provide a much more accurate approach for
pore size analysis. Hence, methods such as the DFT of inhomogeneous fluids [25, 26] and Monte
Carlo simulations [27, 28] bridge the gap between the molecular level and macroscopic
approaches.
The Non-Local Density Functional Theory (NLDFT) and the Grand Canonical Monte Carlo
simulation (GCMC) methods correctly describe the local fluid structure near curved solid walls;
adsorption isotherms in model pores are determined based on the intermolecular potentials of the
fluid-fluid and solid-fluid interactions. The relation between isotherms determined by these
microscopic approaches and the experimental isotherm on a porous solid can be interpreted in
terms of a Generalized Adsorption Isotherm (GAI) equation:
𝑊𝑚𝑎𝑥
𝑁(𝑃⁄𝑃𝑜 ) =
(59).
∫ 𝑁(𝑃⁄𝑃𝑜 , 𝑊 ) 𝑓(𝑊)𝑑𝑊
𝑊𝑚𝑖𝑛
where
N(P/Po) = experimental adsorption isotherm data,
N(P/Po,W) = isotherm on a single pore of width W,
f(W) = pore size distribution function
The GAI equation reflects the assumption that the total isotherm consists of a number of
individual “single pore” isotherms multiplied by their relative distribution, f(W), over a range of
pore sizes. The set of N(P/Po,W) isotherms (kernel) for a given system (adsorbate/adsorbent) can
be obtained, as indicated above, by either Density Functional Theory or by Monte Carlo
computer simulation. The pore size distribution is then derived by solving the GAI equation
numerically via a fast non-negative least square algorithm. The DFT and the Monte Carlo
simulation method have largely been applied to the characterization of micro- and mesoporous
carbons [28-30], silicas, and zeolites [28,31,32]. (Please contact qc.support@Quantachrome.com
for Powder Tech Notes 27, 31 and 36 for more details).
The NLDFT/CO2/273K/carbon kernel is based on a common, one-center, Lennard-Jones model.
This kernel is important in case a comparison with appropriate DFT-results in the literature is
needed. However, in the case of the GCMC/CO2/273K/carbon kernel, a three-center potential
function has been developed with interactions between the sites of different molecules modeled
as a sum of Lennard-Jones and electrostatic contributions. Hence, the GCMC model may serve
as a benchmark for quantitative estimates [30]. For details about the application of CO2
NLDFT/GCMC kernels please contact qc.support@Quantachrome.com for Quantachrome’s
Powder Tech Note 35.
In contrast to regular NLDFT, the QSDFT method (Quenched Solid Density Functional Theory)
[46], explicitly takes into account the effects of surface roughness and heterogeneity (i.e.
geometrical and chemical disorder) of a carbon surface. This allows one to perform pore size
analysis of carbons with different degrees of surface heterogeneity, and to assess the reliability of
the pore size analysis of unknown carbon samples. Methods based on NLDFT are however more
Page 147 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
accurate for ordered (homogenous) carbons. For more details see Quantachrome’s Powder Tech
Note 40 (contact qc.support@Quantachrome.com).
TouchWin™ software includes a number of kernels consisting of individual (P/P0,W) isotherms
derived for the systems: nitrogen-carbon, argon-carbon, nitrogen-silica, and argon-silica by
Non-Local Density Functional Theory and Grand Canonical Monte Carlo computer simulation.
Please refer to 10.4.1 for a detailed listing and recommendations for the pore width ranges
where these models are appropriate.
10.4.1
Library of DFT and GCMC Methods in Quantachrome’s Data
Reduction Software
Kernel File
Applicable Pore
Width Range
(nm)
Examples
NLDFT
N2 / 77K
Carbon,
equilibrium transition (desorption) kernel,
slit pore model
0.35-40
Activated carbons, activated carbon fibers,
novel micro/mesoporous carbons of type
CMK-1 etc.
NLDFT
N2 / 77K
Carbon,
equilibrium transition (desorption) kernel,
cylindrical pore model
0.35-40
Novel micro/mesoporous carbons (e.g.
CMK-3, carbon nanotubes, carbon aerogels)
etc.
NLDFT
N2 / 77K
Carbon,
equilibrium transition (desorption) kernel,
slit pore model for widths < 2nm,
cylindrical pore model for widths > 2nm
0.35-40
Novel micro/mesoporous carbons (some
CMK’s), certain activated carbons.
NLDFT
N2 / 77K
Silica,
adsorption branch kernel,
cylindrical pore model
0.35-100
Siliceous materials such as controlled pore
glasses, MCM-41, SBA-15, MCM-48, and
others. Accurate pore size distribution even
in case of type H2 hysteresis.
NLDFT
N2 / 77K
Silica,
adsorption branch kernel,
cylindrical pore model for diameters < 5 nm,
spherical pores model for diameters > 5 nm
0.35-40
Novel siliceous materials with hierarchically
ordered pore structure, SBA-16 silica, some
types of porous glasses and some types of
silica gels.
NLDFT
N2 / 77K
Silica,
equilibrium transition (desorption) kernel,
cylindrical pore model
0.35-100
Siliceous materials, e.g. some types of silica
gels, porous glasses, MCM-41, SBA-15,
MCM-48 and other adsorbents which show
type H1 sorption hysteresis
NLDFT
Ar / 87K
Zeolite/silica,
equilibrium transition (desorption) kernel,
cylindrical pore model
0.35-100
Zeolites with cylindrical pore channels such
as ZSM5, mordenite, and mesoporous
siliceous materials e.g., MCM-41, SBA-15,
MCM-48, some porous glasses (e.g. CPG)
and silica gels which show type H1 sorption
hysteresis.
NLDFT
Ar / 87K
Zeolite/silica,
adsorption branch kernel,
cylindrical pore model
0.35-100
Zeolites with cylindrical pore channels such
as ZSM5, Mordenite etc., and mesoporous
siliceous materials such as MCM-41, SBA15, MCM-48, porous glasses and some
silica gels etc. Accurate pore size
distribution even in case of type H2
hysteresis.
Page 148 of 161
NOVAtouch™ Operating Manual
Kernel File
THEORY AND DISCUSSION
Applicable Pore
Width Range
(nm)
Examples
NLDFT
Ar / 87K
Zeolite/silica
equilibrium transition (desorption) kernel,
spherical pore model for diameters < 2 nm
cylindrical pore model for diameters > 2 nm
0.35-100
Zeolites with cage-like structures such as
Faujasite, 13X etc.
NLDFT
Ar / 87K
Zeolite/silica
adsorption branch kernel,
spherical pore model for diameters < 2 nm
cylindrical pore model for diameters > 2 nm
0.35-40
Zeolites with cage-like structures such as
Faujasite, 13X etc.
NLDFT
Ar / 87K
Carbon
equilibrium transition (desorption) kernel,
cylindrical pore model
0.35-7
Novel micro/mesoporous carbons (e.g.
CMK-3), carbon nanotubes, carbon
aerogels, and others.
NLDFT
Ar / 87K
Carbon
equilibrium transition (desorption) kernel
slit-pore model
0.35-40
Activated carbons, activated carbon fibers,
novel micro/mesoporous carbons of type
CMK-1, and others.
NLDFT
Ar / 87K
Carbon
equilibrium transition (desorption) kernel,
slit-pore model
0.35-1.5
Activated carbons, activated carbon fibers,
novel micro/mesoporous carbons of type
CMK-1, and others.
NLDFT
CO2 /
273K
Carbon
equilibrium transition (desorption) kernel,
slit-pore model
0.35-1.5
Ultramicroporous activated carbons,
activated carbon fibers.
GCMC
CO2 /
273K
Carbon
equilibrium transition (desorption) kernel,
slit-pore model
0.35-40
Ultramicroporous activated carbons,
activated carbon fibers.
QSDFT
N2 / 77K
Carbon
equilibrium transition (desorption) kernel,
slit-pore model
0.35-40
Disordered Micro/Mesoporous carbons with
heterogeneous surface chemistry (e.g.
activated carbons, activated carbon fibers)
Fractal Dimension Methods

NOTE! Fractal Dimension is intended for investigation using data acquired in
the micropore region (below a relative pressure of 1x10-3). Thus, data from the
NOVAtouch is not appropriate for this calculation. Nevertheless, TouchWin can
be used to evaluate data from low pressure instruments such as the Autosorb-iQ.
Surface characterization methods based on fractal geometry [36-39] describe the topography of
real surfaces in terms of a "roughness exponent" known as fractal dimension, D. Ideal surfaces,
being relatively smooth, can be modeled using simple geometric concepts (e.g., 6L2 for cubes,
4πR2 for spheres, etc.). For such surfaces D=2, because the surface area is proportional to X2,
where X is some characteristic dimension of the adsorbent (e.g., X=L for squares, R for
circles, etc.). In contrast, real surfaces are generally rough because of atom packing
arrangements and defects, kinks and dislocations, and pores themselves, depending on the scale
considered.
Many real surfaces present surface irregularities that appear to be similar at different
scales. These surfaces are referred to as fractal because their magnitude is proportional to X D,
Page 149 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
where D is a fractional exponent that generally assumes values between D=2 (for smooth
surfaces) and D=3 (for surfaces so rough that they essentially occupy all available volume). The
fractal dimension D can thus be used to quantify the roughness of real surfaces in terms of
a single parameter.
Among the various approaches proposed in the literature to evaluate D, two in particular
have gained popularity because they make use of single gas sorption isotherms for their
calculations. These are the Frenkel-Halsey-Hill (FHH) method and the Neimark-Kiselev (NK)
method [37-39].
10.5.1
Frenkel-Halsey-Hill (FHH) Method
By noting that in the multilayer adsorption region the influence of surface forces tends to
be smoothed out, several independent authors derived an isotherm expression of the general
form [38].
B
log P0 = s
P V
(60).
where B is a parameter related to adsorbate-adsorbent and adsorbate-adsorbate interactions, V is
the amount of adsorbed material, and the exponent s is a constant characteristic of a
given adsorbent. Equation (60) came to be known as the Frenkel-Halsey-Hill (or FHH)
equation. Pfeifer et al. [37-39] postulated that the FHH exponent s is related to the fractal
dimension D of the adsorbent through the expression
𝐷 = 3(1 + 𝑠)
(61).
In deriving equation (61) surface tension effects were neglected. That is, the assumption was
made that the surface tension of the adsorbate at a molecular scale does not differ
appreciably from its bulk liquid value. If surface tension effects are accounted for [39], the
relationship between D and s is given by:
𝐷 =3+𝑠
(62).
where, D is the surface fractal dimension, and s is a constant characteristic of a given adsorbent.
In either case, for fractal surfaces a plot of log V vs. log [log(P0/P)] should yield a straight
line with negative slope s within the multilayer region of the isotherm.
10.5.2
Neimark-Kiselev (NK) Method
Combining thermodynamic and fractal arguments, Neimark [38] reasoned that above the
onset of capillary condensation fractal surfaces should conform to the following equation:
S lg = K

ac
 2- D
(63).
Page 150 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
where, D is the surface fractal dimension, K is a constant, ac is the mean radius of curvature
of the adsorbate-vapor interface, given by the Kelvin equation:
ac = r k =
2  Vm
R T ln PP0
 
(64).
and Slg is the adsorbate-vapor interface area, given by the Kiselev equation:
Slg 
RT

n max

n
p 
ln  o   dn
 p
(65).
with R being the universal gas constant, T the adsorption temperature, γ the adsorbate
surface tension, Vm the adsorbate molar volume, and n and nmax the amounts of gas adsorbed
at a given P/Po and at saturation, respectively. In other words, the cumulative area Slg was
taken to be equivalent to the adsorbent area measurable with a yardstick of a size
proportional to ac at any given P/Po. Accordingly, a plot of log [Slg] vs. log [ac] should also
yield a straight line within the multilayer region of the isotherm, from which the fractal
dimension D can be readily calculated.
Thermal Transpiration
At the low pressures commonly employed for nitrogen and argon micropore characterization,
the phenomenon of thermal transpiration [33-35] can result in significant pressure measurement
errors if not compensated.
Thermal transpiration results in a pressure gradient between the sample at temperature, T1
and pressure, P1 and the pressure transducer at temperature T2 and pressure P2 if the inner
diameter of the tubing between the two parts of the system is very small compared with the
mean free path of the gas. The relationship between pressure and temperature can be expressed
as follows for very low pressures:
𝑇1
𝑃1 = 𝑃2 √
𝑇2
(66).
At higher pressures, the empirical model of Liang [34, 35] can be employed to calculate
thermal transpiration corrections for measured pressures.
𝑇
𝛼𝜑 2 𝑥 2 + 𝛽𝜑𝑥 + √𝑇1
(67).
𝑃1
2
=
𝑃2
𝛼𝜑 2 𝑥 2 + 𝛽𝜑𝑥 + 1
where
P1, P2 are in Pascal
Page 151 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
x = 0.133P2d
d = diameter of connecting tube (in meters)
α = 2.52 (for helium)
β = 7.68(1 − √𝑇1 ⁄𝑇2 ) (for helium)
φ = the pressure shift factor that varies for gases relative to the value 1.00 for helium and can be
calculated by
0.27𝑙𝑜𝑔𝜑 = 𝑙𝑜𝑔𝐷 + 9.59
(68).
where D is the molecular diameter of the gas (in meters).
Heats of Adsorption
Whenever a gas molecule adsorbs on a surface, heat is (generally) released. This heat comes
mostly from the loss of molecular motion associated with a change from a 3-dimensional gas
phase to a 2-dimensional adsorbed phase. Heats of adsorption provide information about the
chemical affinity and the heterogeneity of a surface, with larger amounts denoting stronger
adsorbate-adsorbent bonds. There are at least two ways to quantify the amount of heat released
upon gas adsorption: in terms of differential heats of adsorption (q) and as an integral heat of
adsorption (Q). Both methods are described in turn below.
10.7.1
Differential Heats of Adsorption (q)
The differential heat of adsorption (q) is defined as the heat released upon adding a small
increment of adsorbate to the surface of a solid. The value of q depends on (i) the strength of the
bonds formed upon adsorption, and (ii) the degree to which a surface is already covered with
adsorbate. Hence, q is most often expressed in terms of its variation with adsorbate surface
coverage (θ). Indeed, a plot of q against θ provides a characteristic curve illustrating the
energetic heterogeneity of a solid surface. This curve not only serves as a fingerprint of the
surface energetics of a given sample, but it also allows a direct test of the validity of any Vm
evaluation method used, since each Vm evaluation method assumes a different relationship
between q and θ that is,
Langmuir method: q remains constant with increasing θ,
Temkin method: q decreases linearly with increasing θ, and
Freundlich method: q decreases exponentially with increasing θ.
Since q can vary with θ, it is convenient to express q as an isosteric heat of adsorption. (The
word isosteric implies equal surface coverage θ.) In order to evaluate this q, one must collect two
or more isotherms at different temperatures. By inspecting plots of coverage (where θ = V/Vm)
vs. pressure, one can determine those pressures corresponding to equal coverage at the different
temperatures. These pressures and temperatures can then be used to construct Arrhenius Plots of
the natural log of pressure (lnP) versus the reciprocal of the absolute temperature (1/T). Values
Page 152 of 161
NOVAtouch™ Operating Manual
THEORY AND DISCUSSION
for q at any given coverage (θ) can be calculated from the slope (m) of each Arrhenius plot using
the following equation:
q= - m R
(69)
where m = d lnP/d(1/T) and R is the universal gas constant. Differential heats of adsorption are
commonly expressed in units of kilocalories (or kcal) per mol of adsorbate taken up by the
sample.
10.7.2
Integral Heat of Adsorption (Q)
This is simply defined as the total amount of heat released when 1 gram of adsorbent takes up X
grams of adsorbate. It is equivalent to the integral of q over the adsorption range considered, i.e.
 V m   max
Q= 
  q d 
 22.4  min
(70)
where Q is expressed in calories per gram of adsorbent (or cal/g), Vm is cc(STP)/g, and q is a
function of θ, with θ ranging (ideally) from θmin = 0 to θmax = maximum coverage attained
experimentally.
Page 153 of 161
NOVAtouch™ Operating Manual
APPENDIX
11 APPENDIX
Troubleshooting Guide
Symptom
Open isotherm (desorption branch
always lies above adsorption
branch even at low relative
pressure
Reason
Action
Under-equilibrated conditions
If isotherm is unexpectedly open, extend the
desorption equilibration time. Microporous
materials often need longer desorption
equilibration times
Incomplete degassing
Extend degassing time and/or increase
degassing temperature
Leaks
Check integrity of O-ring on sample cell.
Make sure O-ring is lightly greased. Make
sure that the bulkhead does not have multiple
O-rings
Incorrect cell geometry defined
during calibration
Choose proper cell calibration file. Sample
measurement conditions must be identical
(temperature, use of filler rod) to the cell
calibration conditions
System contamination
Pump down unit overnight
Dewar goes up but analysis hangs
Upper limit switch not hit
Check for obstruction to Dewar travel
Unit fails vacuum integrity test
Vacuum hose and/or fittings are
leaking
Check hose and tighten fittings. Replace as
necessary
Sample elutriates during degassing
See “8.3.4, Elutriation and Its Prevention”
Instrument does not power up
Wall outlet has no power
Check wall outlet
Power cord is not plugged in to
the wall outlet
Plug cord in to wall outlet
Blown fuse
Replace fuse
Internal surge protector blown
Contact Quantachrome Service Department
Power switch does not illuminate
even though unit powers up
Internal light burned out
Replace switch
Touchscreen is blank
Video input cable to
Touchscreen is disconnected
Check or reseat both ends of the video cable
Crossed isotherm (desorption
branch lies below adsorption
branch at some relative pressure)
Page 154 of 161
NOVAtouch™ Operating Manual
Power input cable to
Touchscreen is disconnected.
Degassing compartment is dark
Check or reseat both ends of the power
LED light cable disconnected.
LED light burned out
Analysis compartment is dark
APPENDIX
LED light cable disconnected.
Non critical malfunction. Possible loose
connections. Contact Service Department, if
desired
LED light burned out
Heating mantle(s) not heating
when the temperature profile is
started
Heating mantles may not be
connected. Check the LED
diagnostic panel HEATING: If
not flashing, mantle might be
disconnected
Check the LED diagnostic panel (on the side
of the instrument) for the following:
A) HEAT CMD: If the heat command is not
flashing check the profile setting.
B) HEATING: If heating is not flashing.
Mantle is burned or disconnected.
C) OVERTEMP: If overtemp is on, check
the thermocouple cable
Mantle temperature already
above program start temperature
Wait for heating profile to exceed current
mantle temperature
The mantle may be burnt out.
Check the LED diagnostic panel
HEATING: If not flashing.
Mantle is burned or
disconnected
Replace mantle
Check the LED diagnostic panel
HEAT CMD: If not flashing,
check the profile setting
Edit profile in TouchWin™ and upload to
NOVAtouch™
Dewar is at the top and the
instrument seems to be locked up
The instrument may be in cleanup
Check the log for information
Powering the unit up trips the lab
circuit breaker
Electrical problem
Contact Quantachrome Service
Incomplete analysis
Insufficient LN2
Check the log for information
Analysis self-aborts
Check the log for information
Page 155 of 161
NOVAtouch™ Operating Manual
APPENDIX
Maintenance
In order to keep the instrument running at a high standard consistently, it is advised that a
maintenance plan is established using the following guidelines recommended by Quantachrome
Instruments. The service kit listed will provide the necessary parts for a maintenance routine
such as new filters and O-rings. If desired, the Quantachrome Instruments Service Department
can install the service kit. For any questions or concerns regarding the maintenance plan, please
contact the Quantachrome Instruments Service Department.
The following chart details the suggested schedule for instrument maintenance. It includes the
items that may be necessary when organizing a maintenance procedure for your instrument.
Table 11-1, Preventative Maintenance Plan
Maintenance
Item
Recommended
Frequency
Performed by
Name of Part
Needed
Part Number
Check and Inspect
Heating Mantles
Every Six Months
User
Change Oil in Oil
Pump
Install Service Kit
Every Six Months
User
Pump Oil
91029
Every Year
Service
Technician
NOVAtouch™
Service Kit
33022-NOVA
TOUCH
Replace All Valve
Disks
Every Year
Service
Technician
Valve Spider Disc
50042-4
Inspect & Grease
the Lift Drive
Every Year
Service
Technician
Vacuum Grease
91000-.25
Replace All Filters
Every Year
Every Five Years
60 micron Filtered
Screw
Valves
01568
Replace All Valves
Service
Technician
Service
Technician
50050
Page 156 of 161
NOVAtouch™ Operating Manual
APPENDIX
Spare Parts List
Description
Low Temp Heating Mantle
High Temp Heating Mantle
Input Line, 5’ SS Male
Input Line, 5’ SS Nova e
NOVA touch Backfill Line
Cable Shielded CAT 5e Ethernet
Sample Cell Cooling Rack
1/8” Nut (Female) & Ferrule Set, Brass
1/8” Nut (Male)
1/8” Ferrule Set, Brass
Long Flow Degasser Adapter Assembly
Glass Filler Rod Long 6 mm X 268.5 mm
Glass Filler Rod Short 6 mm X 131.5 mm
Glass Filler Rod Long 3 mm X 268.5 mm
Glass Filler Rod Short 3 mm X 131.5 mm
Glass Filler Rod Long 9 mm X 268.5 mm
Part Number
24020-L-JJ-CE-NT
24021-L-JJ-CE-NT
01003-V
01003-1SL
01999-10851
26102-25
01999-10856
01071
41230
40032
01308-3941L
74105-L
74105-S
74104-L
74104-S
74106-L
Glass Filler Rod Short 9 mm X 131.5 mm
9 mm Cell Adapter
6 mm Cell Adapter
12 mm Cell Adapter
Sample Cell Funnel, Delrin
2L Dewar
1L Dewar (Short Cell)
Coolant Level Sensor
Short Coolant Level Sensor
Po Cell
Short Po Cell
NOVA touch Po Collar
H2O Level Sensor
*Fuses
for 100-120V
*Contact Quantachrome’s Service Dept.
*Fuses
for 220-240V
*Contact Quantachrome’s Service Dept.
Line Cord, Domestic (USA)
74106-S
04000-3499-2
04000-3499-1
04000-3499-3
04000-3595
74215
Line Cord, International
™
TouchWin Software
Flange, Brass, Stub
Centering Ring, KF16
74218
00080-CLS
00080-CLS-S
74027
74027-S
04000-10844
00080-H2O-NT
1 A (P/N 24163), .25 A (P/N 24165),
5 A (P/N 24074)
1 A slow (P/N 24152), 1 A (P/N 24163)
.25 A (P/N 24165), 2.5 A (P/N 24073)
26007
26069
36381-1.1NTCD
44019
44028
Page 157 of 161
NOVAtouch™ Operating Manual
APPENDIX
Clamp, Alum. KF16
44030
Vardex ¾” I.D. Vacuum Tubing
46025
O-Ring Buna, 010
Clamp, Hose, 1” I.D.
Nova, 9mm, large bulb, long cell
51000-010
71018
74064
Nova, 9 mm Large Bulb short cell
74064-1
Nova, 9 mm Small Bulb long cell
74063
Nova, 9 mm Small Bulb short cell
74063-1
Nova, 6 mm Large Bulb long cell
74062
Nova, 6 mm Large Bulb short cell
74062-1
Nova, 6 mm Small Bulb long cell
74061
Nova, 6 mm Small Bulb short cell
74061-1
Nova, 12 mm Large Bulb short cell
74066-1
Nova, 12 mm Small Bulb long cell
74065
Nova, 12 mm Small Bulb short cell
74065-1
Nova, Std. 9 mm Pel. Long
74099
Nova, Std. 9 mm Pel. Short
74099-S
Nova, Std. 6 mm Pellet Cell Long
74098
Nova, Std. 6 mm Pellet Cell Short
74098-S
Nova, Std. 12 mm Pellet Cell Long
74100
Nova, Std. 12 mm Pellet Cell Short
74100-S
6mm X 12” Stems Cleaner (1 Doz.)
96036
Description
Low Temp Heating Mantle
High Temp Heating Mantle
Input Line, 5’ SS Male
Input Line, 5’ SS Nova e
NOVA touch Backfill Line
Cable Shielded CAT 5e Ethernet
Sample Cell Cooling Rack
1/8” Nut (Female) & Ferrule Set, Brass
1/8” Nut (Male)
1/8” Ferrule Set, Brass
Long Flow Degasser Adapter Assembly
Glass Filler Rod Long 6 mm X 268.5 mm
Part Number
24020-L-JJ-CE-NT
24021-L-JJ-CE-NT
01003-V
01003-1SL
01999-10851
26102-25
01999-10856
01071
41230
40032
01308-3941L
74105-L
Page 158 of 161
NOVAtouch™ Operating Manual
APPENDIX
Glass Filler Rod Long 3 mm X 268.5 mm
Glass Filler Rod Long 9 mm X 268.5 mm
74104-L
74106-L
9 mm Cell Adapter
6 mm Cell Adapter
12 mm Cell Adapter
Sample Cell Funnel, Delrin
2L Dewar
Coolant Level Sensor
Po Cell
NOVA touch Po Collar
H2O Level Sensor
*Fuses
for 100-120V
*Contact Quantachrome’s Service Dept.
*Fuses
for 220-240V
*Contact Quantachrome’s Service Dept.
Line Cord, Domestic (USA)
04000-3499-2
04000-3499-1
04000-3499-3
04000-3595
74215
00080-CLS
74027
04000-10844
00080-H2O-NT
1 A (P/N 24163), .25 A (P/N 24165),
5 A (P/N 24074)
Line Cord, International
™
TouchWin Software
Flange, Brass, Stub
1 A slow (P/N 24152), 1 A (P/N 24163)
.25 A (P/N 24165), 2.5 A (P/N 24073)
26007
26069
36381-“X.YY”NTCD
44019
Centering Ring, KF16
44028
Clamp, Alum. KF16
44030
Vardex ¾” I.D. Vacuum Tubing
46025
O-Ring Buna, 010
Clamp, Hose, 1” I.D.
Nova, 9mm, large bulb, long cell
51000-010
71018
74064
Nova, 9 mm Small Bulb long cell
74063
Nova, 6 mm Large Bulb long cell
74062
Nova, 6 mm Small Bulb long cell
74061
Nova, 12 mm Small Bulb long cell
74065
Nova, Std. 9 mm Pel. Long
74099
Nova, Std. 6 mm Pellet Cell Long
74098
Nova, Std. 12 mm Pellet Cell Long
74100
6mm X 12” Stems Cleaner (1 Doz.)
96036
Page 159 of 161
NOVAtouch™ User Manual
REFERENCES
12 REFERENCES
1. S. Brunauer, P. Emmett and E. Teller, J. Amer. Chem. Soc., 60, 309 (1938).
2. V. Kiselev and Y. A. Eltekov, World Congress on Surface Activity, Vol. II, p. 228,
Butterworths, London, 1957.
3. S. Lowell, J. Shields, G. Charalambous and J. Manzione, J. Colloid Interface Sci., 86,
191 (1982).
4. S. Brunauer, L. S. Deming, W. S. Deming and E. Teller, J. Amer. Chem. Soc., 62, 1723
(1940).
5. J. H. de Boer, “The Structure and Properties of Porous Materials,” p. 68, Butterworths,
London, (1958).
6. J. H. de Boer, B. C. Lippens, B. G. Lippens, J. C. P. Broekhoff, A. van den Heuvel and
Th. V. Osinga, J. Colloid Interface Sci., 21, 405 (1966).
7. E. P. Barrett, L. G. Joyner and P.P. Halenda, J. Amer. Chem. Soc., 73, 373 (1951).
8. D. Dollimore and G.R. Heal, J. Appl. Chem. 14, 109 (1964).
9. G.D. Halsey, J. Chem. Phys., 16, 931 (1948).
10. J. H. de Boer, B. G. Linsen, Th. van der Plas and G. J. Zondervan, J. Catalysis, 4, 649
(1965).
11. R. W. Magee, Evaluation of the External Surface Area of Carbon Black by Nitrogen
Adsorption, Presented at the meeting of the Rubber Division of the American Chem.
Soc., October, 1994.
12. K.S.W. Sing, Surface Area Determination, D.H. Everett and R.H. Ottewill, Eds.,
Butterworths, London (1970), p.25,
13. R.S. Mikhail, S. Brunauer and E.E. Bodor, J. Colloid Interface Sci., 26, 45 (1968)
14. M. Polanyi, Verh. Dtsch. Phys. Ges., 16, 1012 (1914).
15. M.M. Dubinin and L.V. Radushkevich, Dokl. Akad. Nauk. SSSR, 55, 331 (1947).
16. M.M. Dubinin, Chemistry and Physics of Carbon, Vol 2, pp. 51-118. P. L. Walker
(Ed.), Marcel Dekker, New York, 1966.
17. M.M. Dubinin and H.F. Stoeckli, J. Colloid Interface Sci., 75, 34 (1980).
18. M.G. Kaganer, Zh. Fiz. Khim. 32, 2209 (1959).
19. M.M. Dubinin, Chem. Rev. 60, 235 (1960)
20. H. Marsh and B. Rand, J. Colloid Interface Sci., 33, 101 (1970).
21. H.F. Stoeckli, Microporous Carbons and Their Characterization: the Present State of the
Art, Carbon, Vol. 28, No. 1, pp 1-6, 1990.
22. M.J.G. Janssen and C.W.M. van Oorschot, The Characterization of Zeolites by Gas
Adsorption, in Zeolites: Facts, Figures, Future, 1989, P.A. Jacobs and R.A. van Santen
(Eds.) Elsevier.
23. G. Horvath and K. Kawazoe, Method for the Calculation of Effective Pore Size
Distribution in Molecular Sieve Carbon, J. Chem. Eng. Japan, 16 (5), 470-475 (1983).
24. Saito and H.C. Foley, AIChE Journal, 37, 429 (1991).
25. R. Evans, U.M.B. Marconi and P. Tarzona, J. Chem. Soc. Faraday Trans. II, 82, 1763
(1986).
26. P.I. Ravikovitch, G.L. Haller and A.V. Neimark, Adv. Colloid and Interface Sci., 76-77,
203 (1998).
27. K.E. Gubbins, In Physical Adsorption: Experiment, Theory and Application, J.Fraissard
(ED.), Kluwer: Dordrecht, 1997.
28. L.D. Gelb, K.E. Gubbins, R. Radhakrsihnan and M. Sliwinska-Bartowiak, Rep. Prog.
Page 160 of 161
NOVAtouch™ User Manual
REFERENCES
Phys. 62, 1573 (1999).
29. N.A Seaton, J.P.R.B. Walton and N. Quirke, Carbon 27, 853 (1989).
30. Vishnyakov, P. Ravikovitch and A.V. Neimark, Langmuir 16, 2311 (2000).
31. A.V. Neimark, P.I. Ravikovitch, M. Gruen, F. Schueth and K.K. Unger, J. Colloid
Interface Sci, 207, 159 (1998).
32. P.I. Ravikovitch, D. Wie, W.T. Church, G.L. Haller and A.V. Neimark, H. Phys. Chem.
B 101, 3671 (1997).
33. S. Dushman, Scientific Foundations of Vacuum Technique, p. 65, J. Wiley & Sons,
New York (1989).
34. S.C. Liang, J. Phys. Chem., 57, 910 (1953).
35. K.F. Poulter et al., Vacuum, 33 (6) 311-316 (1983).
36. D. Avnir (Ed.), The Fractal Approach to Heterogeneous Chemistry, J. Wiley & Sons,
New York, (1989).
37. P. Pfeifer, Y.J. Wu, M.W. Cole and J. Krim, Phys. Rev. Lett. 62, 1997 (1989).
38. A.V. Neimark, Ads. Sci. Tech., 7, 210 (1991).
39. P. Pfiefer, J. Kennter and M.W. Cole, Fundamentals of Adsorption (Edited by A.B.
Mersmann and S.E. Sholl), Engineering Foundation, New York, p. 689 (1991).
40. S. Lowell, J.E. Shields, M.A. Thomas and M. Thommes, Characterization of Porous
Solids and Powders: Surface Area, Pore Size, and Density, Springer: 2004.
41. K.S.W. Sing, D.H. Everett, R.A.W. Haul, L. Moscou, R.A. Pierotti, J. Roquerol and T.
Siemieniewska, Pure Appl. Chem. 57, 603 (1985).
42. T.L. Hill, J. Phys. Chem. 59, 1065 (1955).
43. M. Thommes, et al., “Aspects of a novel method for the pore size analysis of thin silica
films based on krypton adsorption at liquid argon temperature (87.3 K),” Studies in
Surface Science and Catalysis 165, 551 (2007).
44. Rouquerol, J., Llewellyn, P. and Rouquerol, F. "Is the BET equation applicable to
microporous adsorbents?" In: Characterization of Porous Solids VII. P. Lewellyn, F.
Rodriguez-Reinoso, J. Rouquerol, N. Seaton (Eds): Studies in Surface Science and
Catalysis, 160 (2007) pp. 49-56.
45. M. Thommes, K. Kaneko, A.V. Neimark, J.P. Olivier, F. Rodriguez-Reinoso, J.
Rouquerol and K.S.W. Sing, Pure Appl. Chem. 87, 1051-1069 (2015) IUPAC Technical
Report: “Physisorption of gases, with special reference to the evaluation of surface area
and pore size distribution”.
46. P.I. Ravikovitch and A.V. Neimark, Langmuir (2006) 22, 11171-11179
Page 161 of 161
Download