Uploaded by m4310.rossi

Phywe - Wavelength and frequencies with a Quincke tube - P2150702

advertisement
Wavelength and frequencies
with a Quincke tube
1.5.0702
Related Topics
Transverse and longitudinal waves, wavelength, amplitude, frequency, phase shift, interference, velocity
of sound in air, loudness, Weber-Fechner law.
Principle
If a sound wave of a particular frequency is divided into two coherent components (like, for example, light
waves in an interferometer experiment), and if the path of one of the component waves is altered, it is
possible to calculate the wavelength of the sound wave and its frequency from the interference phenomena recorded with a microphone.
Equipment
1
1
1
1
1
1
1
1
1
1
2
5
Interference tube, Quincke type
Sound head
Measuring microphone
Digital Function Generator, USB
Digital multimeter
Vernier caliper
Connecting cable, 4 mm plug, 32 A, blue, l = 150 cm
Connecting cable, 4 mm plug, 32 A, red, l = 150 cm
Adapter, BNC-socket/4 mm plug pair
Support base -PASSSupport rod -PASS-, square, l = 630 mm
Right angle clamp -PASS-
03482-00
03524-00
03542-00
13654-99
07128-00
03010-00
07364-04
07364-01
07542-27
02005-55
02027-55
02040-55
Caution!
Don’t use an amplitude higher than 2000mV because the Sound head could collapse.
Fig. 1:
Experimental set-up: Wavelengths and frequencies with a Quincke tube.
www.phywe.com
P2150702
PHYWE Systeme GmbH & Co. KG © All rights reserved
1
Wavelength and frequencies
with a Quincke tube
1.5.0702
Tasks
1. Record the extension of a Quincke tube for given frequencies in the range 2000 Hz to 6000 Hz.
2. Calculate the frequencies from the wavelengths determined and compare them with the given
frequencies.
Set-up and Procedure
1. The experiment is set up as in Fig. 1.
2. As many intensity minima as possible should be found
for the frequencies 2000 Hz, 2200 Hz, 2400 Hz, and so on,
up to 6000 Hz, which are set on the Digital Function Generator. The relative extensions of the tube are measured
with the vernier caliper. The selection switch on the microphone is set to the position (~).
3. Use the six buttons of the Digital Function Generator to
enter your values of the signal. With the four buttons in the
middle you could choose the options of Frequency, Amplitude, Offset and Signal. Use the controller to get your values and enter with the Menu-Button.
Fig. 2:
Digital Function Generator
Theory and Evaluation
If two harmonic linear waves have the same frequency and the same or opposite direction of propagation, the superimposition of the waves is obtained from:
Ar x, t   A1 x, t   A2 x, t 
With
A1 x, t   A1e i k1xt 
and
A2 x, t   A2ei k 2 x t   
where
k1  k2  k 
2
Fig. 3:

Geometry of the Quincke tube.
are the values of the wave vectors in the x-direction or the opposite direction. The angular frequency is

2
 2f
T
.
The resultant Wave


Ar x, t   A1eik1 x  A2ei k 2 x     eit
2
PHYWE Systeme GmbH & Co. KG © All rights reserved
P2150702
Wavelength and frequencies
with a Quincke tube
1.5.0702
is likewise harmonic and has an angular frequency  . If k1 and k 2 are rectified, a progressive wave is
always obtained:
Ar x, t   Ar ei kxt 
For k1  k 2  k , with the amplitude
Ar  A1  A2 ei
The amplitude of the resultant wave is a function of the phase difference  .
If k1 are opposed (Quincke tube), the resultant wave is a superimposition of a standing wave and a progressive wave:
 
 
 




i  kx 
i  kx 
i  t 

 

2 
2  
2 
Ar  x, t   A1e ikx  A2 e i  kx    e it   A1e 
 A2 e 
e 




 
  
 




 i  kx  2 
i  kx 
i  kx 
i  t

 

2 
2  
2 








  A1  A2 e
 A2  e
e
e






 Ar1e
i  kx t 


  i  t 

 Ar 2 cos kx 
e
2 

 

2 
for k1  k 2  k .
The amplitude of the progressive wave is
Ar1  A1  A2
and that of the standing wave
Ar 2  2A2 .
It is assumed, without limitation, that A2  A1 .
In a Quincke tube, a sound wave is subdivided by the branching of the tube into two coherent components, which, after appropriate deflection, travel towards each other and are superimposed on each
other (Fig. 3). If the straight sections of the tube a and b are the same size, then, because of the pressure drop along their path, the intensities of the sound waves impinging on the measuring point (2) will
be equal A1 = A2. In this case is Ar1 = 0, so that only a standing wave exists in the region of the measuring point.
As the difference d in the lengths of the straight sections is increased, the intensity difference at the
measuring point also becomes greater A2 > A1. In this case the amplitude Ar2 of the standing wave decreases, and that of the progressive wave increases.
If the measuring point is made x  0 , the sound pressure is
p  Ar1 cos t  Ar 2 cos
 
 
 cos t 

2 
2 .
www.phywe.com
P2150702
PHYWE Systeme GmbH & Co. KG © All rights reserved
3
Wavelength and frequencies
with a Quincke tube
1.5.0702
For small values of Ar1 , the amplitude of the sound pressure is determined mainly by Ar 2 cos

. The
2
sound pressure is always at its minimum, when
 2n  1


2
2
;
n  0,  1,  2, 
The pressure minima are consequently expressed by:
dn 
2n  1

2
n  0,  1,  2, 
The gap between two minima corresponds exactly to half a wavelength:
d  d n 1  d n 

2.
When b is extended by a distance d , the path is increased by 2d . Since 2d corresponds exactly to
the wavelength, we obtain   2d (Fig. 4).
The frequency is obtained from the equation
c  f   f 
c
 (Fig. 5).
The propagation velocity c of the sound waves in air is c  331.3 ms 1 at 0°C, being dependent on temperature and air pressure. c is determined at room temperature from croomtemp.  331,3 1  0,004t at
1013 hPa, with t measured in °C.
Fig. 5:
4
Connection of wavelength

and frequency
f
Fig. 4:
Resonance wavelength dependent on the displacement d
PHYWE Systeme GmbH & Co. KG © All rights reserved
P2150702
Wavelength and frequencies
with a Quincke tube
1.5.0702
Measure
Open the PHYWE software measure and choose enter data manually. Enter Title, Number of Channels
(X and Y-Channels), Number of Values, Title, Symbol, Unit and Digits of every Channel and press continue. Then enter the measured values into the table. To plot the amplitude as a function of the distance
d , choose the channel manager from the measurement dialog. Mark the distance and press the arrow
to the right, the distance is now the x-axis. Mark your other channels and press the arrow to the right (=yaxis).From the right mouse click or the measurement menu you could choose the Display Option
Fig. 6:
Example (f= 3000 Hz)
www.phywe.com
P2150702
PHYWE Systeme GmbH & Co. KG © All rights reserved
5
Download