Uploaded by andrej.terehov0205

1 Photometric determination of Chromium III

advertisement
INSTRUMENTAL ANALYSIS METHODS, 2014.
Laboratory work No. 1
Photometric determination of chromium (III) with complexone III (EDTA)
Chromium (III) complexonate
Thermodynamic stability constant of chromium (III) complexonate CrY- is 1023.4. Because of
inertness of chromium (III) aqua complex, formation of Cr(III) complexonate in aqueous media and
under ambient temperature is very slow. In 100°C exchange with ligands becomes rapid and after
several minutes blue-purple coloured chromium (III) complexonate is made.
[Cr(H2O)n]3+ + H2Y2- + 2CH3COO- → CrY- + 2CH3COOH + nH2O
If chromium (III) complexonate is made in solutions with media pH 4-5, its absorbance maximum
is at 540nm (Figure 1). Molar absorbtion coefficient is only 140. Decreasing pH of the media it gets
even lower. To ensure optimal pH value, acetate buffer solution with ratio of components 1:1 is
used so pH=pKa=4.8.
Figure 1. Absorbtion spectrum of chromium (III) complexonate
Spectrophotometer / Photometer
Two single beam devices will be employed: Spectrophotometer Jenway 6300 with 1cm cuvette and
photometer KFK2. The principial scheme is shown in figure 2.
Figure 2. Optical scheme of the spectrophotometer
1
Main difference between both instruments is in the way how monochromatic light is obtained. In
spectrophotometer Jenway 6300 diffraction grating is used to obtain light with defined wavelenght.
In photometer KFK2 optical filters are used to give monochromatic light. (Please evaluate the
wavelength precision of both monochromators).
Diffraction grating is an optical component with a periodic structure, which splits and diffracts light
into several beams travelling in different directions. The directions of these beams depend on the
spacing of the grating and the wavelength of the light so that the grating acts as the dispersive
element.
Absorbtive filters absorb some wavelengths of light while transmitting others. Using data from table
1.1. we can find filter which will serve as monochromator.
Table 1.
Absorbtive filters
Wavelenght region absorbed,
nm
400 – 435
435 – 480
480 – 490
490 – 500
500 – 560
560 – 580
580 – 595
595 – 600
650 – 750
Colour of light absorbed
Violet
Blue
Blue-green
Green-blue
Green
Yellow-green
Yellow
Orange
Red
Complementary colour
transmitted
Yellow-green
Yellow
Orange
Red
Purple
Violet
Blue
Blue-green
Green-blue
Monochromatic light band has shape of symmetrical peak with parameters as nominal wavelenght,
maximum throughput and effective bandwidth (figure 3).
Figure 3. Monochromatic light band
2
Typical spectra of photometer filter set is shown in figure 4.
Figure 4. Transmittance of absorbtive filters (KFK-2)
The principle of light absorbance measurment
Photometers and spectrophotometers are used to measure the absorbance of coloured sample
solutions againtst the absorbance of zero standard sample. Using the zero sample absorbance is set
to 0, then the absorbance of analytical sample is measured.
Task
1. Obtain data for calibration graph. Construct calibration plot, find the equation of the trendline and
coefficient of corellation. Evaluate the quality of the trendline.
2. Find the mass (g) of chromium (III), which is in 100mL flask given by the lab assistant.
Experimental
Connect instrument to mains, turn on.
Preparation of solutions
Dilute analytical sample in 100mL volumetric flask (given by the lab assistant) with deionized
water up to the mark and mix well. With pipette transfer 10mL of obtained solution into 100mL
beaker (Marked ‘’An’’). With volumetric pipette add 10mL of deionized water, 5mL of acetate
buffer solution and 10mL of complexone III solution.
Using burette transfer 20mL of chromium (III) standard solution (γ=0.002g/mL) into 50mL
volumetric flask, dilute with deionized water up to the mark and mix well.
Label six 100mL beakers with inscriptions 0, 2, 4, 6, 8 and 10.
3
Using volumetric pipette transfer appropriate volume of previously prepared chromium (III)
standard solution into labelled beakers:
2mL → beaker #2
4mL → beaker #4
6mL → beaker #6
8mL → beaker #8
10mL → beaker #10
Then add deionized water:
20mL → beaker #0
18mL → beaker #2
16mL → beaker #4
14mL → beaker #6
12mL → beaker #8
10mL → beaker #10
To each beaker add 5mL of acetate buffer solution and 10mL of complexone III solution. Heat all 7
beakers to a boil and continue boiling for another 5 minutes and allow them to cool down.
Label seven 100mL volumetric flasks with inscriptions 0, 2, 4, 6, 8, 10 and ‘’An’’. Transfer
solutions from beakers to respective volumetric flasks. Wash each beaker with deionized water and
transfer it to respective flask (don’t pour it out!). Repeat washing for several times to be sure that
chromium (III) complexonate is completely (quantitatively) transfered to 100mL flasks.
Then dilute solutions in volumetric flasks with deionized water till the mark and mix them well.
Measurment of light absorbance
Absorbtion measurment of chromium (III) complexonate solution is performed against the 0 sample
using 1cm cuvette (KFK2) and 5cm cuvette (Jenway 6300).
Start measurment with the more diluted standard sample. Before you do each measurment, flush
cuvette with current solution. When you have finished with all the standards, flush cuvette with
deionized water, then with analytical solution ‘’An’’, then fill cuvette with analytical solution and
perform the measurment.
Spectrophotometer Jenway 6300
1. Connect the instrument to the mains, turn on and allow to condition for 15 minutes.
2. Choose the wavelength and measurment mode (T, ABS, CONC).
3. Open cuvette compartment, insert cuvette with ‘’0’’ sample solution, close compartment and
press CAL button. Absorbtion value is set to ABS=0.
4. Remove cuvette with ‘’0’’ solution, flush and fill it with next standard solution put cuvette
into holder and close compartment. Read the absorbance value.
1.
2.
3.
4.
5.
Photometer KFK-2
Connect the instrument to the mains, turn on and allow to condition for 15 minutes.
Set the appropriate light filter (controlling the filter position of 540nm mark).
Set sensitivity ‘’чувствительность’’ to position 1 (minimal), turn tuner knob ‘’установка’’
to the left till the end.
Open cuvette compartment and instert cuvette with ‘’0’’ solution, using sensitivity knob and
tuner set transmittance to 100. Into second slot put cuvette with Cr(III) containing solution.
Using handle below cuvette compartment replace both cuvettes so the sample cuvette gets
under the light beam. Read the transmittance on the upper scale or absorbance on the lower
scale.
4
University of Latvia
Faculty of Chemistry
________________________
Name Surname
Laboratory work No. 1
Photometric determination of chromium (III) with complexone III (EDTA)
1. Theoretical background
Equation of hydrated chromium (III) ion reaction with complexone III (show oxidation state
of all atoms):
2. Equipment
Draw optical scheme of KFK-2 with explanations (attach on separate sheet).
Motivate the wavelength selection:
3. Preparation of solutions
Calculate mass concentration of chromium (III) ions in solution which is obtained when 20mL
of chromium (III) standard solution (γ=2mg/mL) are diluted with deionized water in 50mL
volumetric flask. Show all members and units!
Give equation to calculate chromium (III) mass concentration in labelled 100mL volumetric
flasks ‘’2’’, ‘’4’’, ‘’6’’, ‘’8’’, ‘’10’’. Show all members and units!
Cromium (III) ion mass concentrations are:
Flask ‘’2’’: γ=
Flask ‘’4’’: γ=
Flask ‘’6’’: γ=
Flask ‘’8’’: γ=
Flask ‘’10’’: γ=
5
4. Measurments
Enter calculated data and obtained measurments in table 1.
Solution
Table 1
Mass concentration and absorbance of Cr (III) solutions
Chromium (III)
Absorbance, 1cm
Absorbance, 5cm
concentration γ, μg/mL
cuvette
cuvette
‘’2’’
‘’4’’
‘’6’’
‘’8’’
‘’10’’
An
5. Calculation of results
Use MS Excell software to construct two calibration graphs. Put chromium (III) mass concentration
on x-axis and absorbance on y-axis. If the graph shows experimental points that are very different
from those points that form a straight line, then they are gross errors and may be excluded. Use MS
Excell to find the equation of calibration curve y=A+Bx:
A1cm cuvette =
B1cm cuvette =
A5cm cuvette =
A5cm cuvette =
Use equation of calibration curve to calculate chromium (III) concentration in analytical solution
using both 1 and 5cm cuvettes:
γ1cm cuvette =
γ5cm cuvette =
Compile the equation, enter values and units, and calculate the mass of chromium (III) ions in
analytical solution:
6. Job quality rating
True mass of iron ions:
Absolute error:
Relative error:
Conclusions about the accuracy of the work. Compare obtained error to the method-specific random
deviation values. Give conclusions whether the work is vitiated by systematic error.
On horizontal line mark true and found chromium (III) ion mass. Mark lower and higher limit of
confidence interval with the brackets:
6
Assess the obtained absorbance values, explain why absorbance in 1cm cuvette is smaller than in
5cm cuvette:
Is there a need to check (calibrate) the volume of volumetric glassware? Motivate Your answer with
comparison of method specific random errors to uncertainty of class B glassware.

Mācību metodiķa paraksts par darba vietas pieņemšanu:

Work performance date: 2014.___.____________ Student’s signature:
Work counted: 2014. ___.___________________ Teacher’s signature:
7
Appendix 1
1. Chemicals
Cr2(SO4)3 · 6H2O
Cr3+, c=2mg/mL
CH3COONa, c=0,1mol/L
Complexone III, 5% solution
2. Glassware
100mL volumetric flasks, 8pcs
100mL beakers, 7pcs
10mL Mohr pipettes, 3pcs
50mL volumetric flask
10mL graduated pipette
25mL measuring cylinder
Funnel
3. Equipment
Dispenser
Jenway 6300
KFK-2
8
Download